SciCan Hydrim L11w Quick Reference Manual

Hydrim®L110 w Quick Reference Guide
For best results:
Use Stat-Dri Plus rinse aid (part number 8OZPLUS) to prevent spotting.
P0 should be used to keep instruments moist when they will not be washed within one hour.
P1 and P2 should be used to wash instruments in baskets.
P3 and P4 should be used to wash instruments in cassettes and heavily soiled instruments.
Shines when cleaning solution needs changing
Caution:
Flashes when rinse-aid needs filling
1. Hydrim Cleaning Solution is compatible with solid
2. Carbon steel, aluminum, nickel and rhodium-plated
instruments may corrode in Hydrim Cleaning Solution. Do not process burrs or endofiles in the Hydrim.
3. Do not process stainless steel and carbon steel
instruments at the same time.
Shines when salt needs refilling
*Actual cycle time will vary depending on
incoming water temperature and pressure
8 min.* 20 min.* 30 min.* 30 min.* 40 min.*
Replacing the Hydrim Cleaning Solution:
When the chemical indicator light shines continuously, and Ed appears on the display, the solution level is low and the container is in need of replacement.
1. Open the cleaning solution door.
2. Put a new cleaning solution box into the drawer upside down.
3. Te a r o p en the tab.
4a.Pull out the spout and remove the protective cap.
4b.Attach the hose tightly to the connector on the cleaning solution box.
5. Place the box in the cleaning solution drawer and then close.
1
order number 01-108367S
2 5
4a
3
4b
Filter Maintenance:
1. Check the filters daily for debris and
remove if necessary.
2. After undoing the coarse filter, the fine
filter can be removed. Remove any debris from the filter cavity and rinse the filters under running water.
3. Insert the filter units and tighten the
coarse filter.
coarse filter
fine filter
lower wash arm upper wash arm
Refilling the Rinse Aid:
When the chemical indicator light flashes, you will need to refill the rinse-aid container with Stat-Dri Plus.
1. Fold back the lid of the rinse
aid container.
2. Pour the rinse-aid into the rinse-aid
container until the level indicator darkens.
3. Close the lid ensuring you that you
hear it snap.
salt container lid
Wash Arm Maintenance:
1. Inspect the nozzles and wash arms
weekly for blocked holes due to debris.
2. Pull the lower arm upwards and lift off.
3. Unscrew the upper wash arm collar
and remove the upper wash arm.
4. Clean both wash arms under
running water.
5. Refit the wash arms. Ensure that the
lower wash arm has locked into place
upper wash arm collar
chemical indicator light
and the upper one is screwed tight.
level
lid
indicator
rinse-aid containersalt indicator light
Refilling the Water Softener:
If you have set the water softener to any setting above 0, the SALT indicator flashes when you need to refill the salt container.
1. Open the salt container lid.
2. Upon first use of the Hydrim L110w, pour approximately
1 litre (1 quart) of water into the salt container. It is not necessary to add water during subsequent refills of the salt container.
3. Using the accompanying salt jug, fill the salt container
to the top (maximum of 1kg / 2.2 lbs.)
4. Close the salt container lid.
United States: 1-800-572-1211, Canada: 1-800-870-7777, International: (416) 446-4500
Hydrim is a registered trademark of SciCan, Division of Lux & Zwingenberger Ltd.
For all service and repair inquiries:
Email: techservice.ca@scican.com
P.N.96-108366 Rev. 1.0
Loading...