SciCan Hydrim L110w Installation Manual

Page 1
HYDRIM L110w Installation Guide
Installation
Dimensions:
Height (free-standing) 850 mm / 33.5" Height (built-in) 810 mm / 32" Width 600 mm / 23.75" Depth 600 mm / 23.75" Depth with Door Open 1143 mm / 45" Weight 85 kg / 187 lbs Running Noise 58 dB wash
78 dB dry Hot Water Connection 60˚C max / 158˚F Rinse aid dispenser 80 ml / 2.7 U.S. fl. oz. capacity Water softener 1.0 kg / 2.2 lbs salt capacity Filling system 4.5 L / 135 U.S. fl. oz. capacity Wash Temperature 50˚C / 122˚F Rinse and hold 30˚C / 86˚F
Electrical Consumption:
North America 208 - 240 V 60 Hz 15 A Rated Load 2.5 kW
Other:
Equipment Pollution Degree 2 Equipment Installation Category II Maximum relative humidity 80 % for temp. up to 31˚C / 88˚F
50 % for temp. up to 40˚C / 104˚F Mains supply + / - 10 % of nominal Operating mode Free-standing or Built in
Specifications
Hydrim L110w instrument washer:
Page 2
HYDRIM L110w Installation Guide
Installation
Safety Information
Safe operation
The following applies to both operators and service technicians:
•Exercise caution and seek assistance when lifting or carrying the unit.
•Cleaning solutions may irritate. Avoid contact with eyes and mouth.
•Never lean on the open door. The unit may tip forward causing injury.
Water Connections
The unit must be connected to the water supply in accordance with local and national plumbing codes. SciCan recommends a hard plumbing installation within 1.5 m / 5 ft. of the unit. If additional distance is necessary, commercial grade plumbing hose, similar to washing machine installation, must be used to minimize possible leaks.
Connect the inlet hose to a water tap using the enclosed parts and in accordance with the installation instructions.
Drainage
The unit is supplied with a 1.5 m / 5 ft. flexible drain hose with an inner diameter of 2 cm /
3
/4". The hose should not be shortened or attached to any fittings that would cause a reduction in water flow. The drain system is equipped with a non-return valve that prevents dirty water from flowing back into the unit.
The drain hose should not be further than 1.5 m / 5 ft. from a hard plumbing drain. If this is not possible, then commercial grade plumbing hose must be used to minimize possible leaks.
The hose can be attached to an existing drain line with a 3.5 cm / 1
1
/2" or larger stand pipe / P-trap combination. Alternatively, the hose can be connected directly to the existing drain lines, provided any fittings or adapters used do not reduce the water flow. The drain hose should not exceed 3.3 m / 13 ft. in length, or be attached to the main drain at a point higher than 35 cm / 14" above the floor.
Water Connection
Water Pressure: • 7 - 145 psi / 0.5 - 10 bar Water Temperature: • Cold Water < 86˚F / 30˚C
• Hot Water up to 140˚F / 60˚C