HYDRIM L110w Installation Guide
Installation
Safety Information
Safe operation
The following applies to both operators and service technicians:
•Exercise caution and seek assistance when lifting or carrying the unit.
•Cleaning solutions may irritate. Avoid contact with eyes and mouth.
•Never lean on the open door. The unit may tip forward causing injury.
Water Connections
The unit must be connected to the water supply in accordance with local and national
plumbing codes. SciCan recommends a hard plumbing installation within 1.5 m / 5 ft. of
the unit. If additional distance is necessary, commercial grade plumbing hose, similar to
washing machine installation, must be used to minimize possible leaks.
Connect the inlet hose to a water tap using the enclosed parts and in accordance with the
installation instructions.
Drainage
The unit is supplied with a 1.5 m / 5 ft. flexible drain hose with an inner diameter of 2 cm /
3
/4". The hose should not be shortened or attached to any fittings that would cause a
reduction in water flow. The drain system is equipped with a non-return valve that
prevents dirty water from flowing back into the unit.
The drain hose should not be further than 1.5 m / 5 ft. from a hard plumbing drain.
If this is not possible, then commercial grade plumbing hose must be used to minimize
possible leaks.
The hose can be attached to an existing drain line with a 3.5 cm / 1
1
/2" or larger stand
pipe / P-trap combination. Alternatively, the hose can be connected directly to the
existing drain lines, provided any fittings or adapters used do not reduce the water flow.
The drain hose should not exceed 3.3 m / 13 ft. in length, or be attached to the main drain
at a point higher than 35 cm / 14" above the floor.
Water Connection
Water Pressure: • 7 - 145 psi / 0.5 - 10 bar
Water Temperature: • Cold Water < 86˚F / 30˚C
• Hot Water up to 140˚F / 60˚C