-86ºC Ultra Low Temperature Freezer — MDF-U5386SC
ULTRA LOW TEMPERATURE FREEZERS
(CONVENTIONALLY INSULATED)
MDF-5386SC
__________
483L
(17.1 Cu.Ft.)
■ Ideal -86ºC freezing environment by means
of conventional insulation walls.
■ Specially designed compressorS for ultra-low
temperature applications.
320 2”
inventory
boxeS
■ Microprocessor temperature control with
digital input for precise setting and control.
■ Built-in temperature and power failure
alarms (audible/visible).
FEATURE BENEFIT
Microprocessor Comprehensive setpoint, alarm, monitoring and diagnostic functions
Controls based on SANYO-built microprocessor controller with digital display
SANYO-Brand Built by SANYO specifically for demanding ultra-low temperature applications
p r e s e r v a t i o n
Refrigeration in a laboratory environment; CFC-free refrigerants only
Integrated Cabinet High performance refrigeration system with foamed-in-place insulation
Design maximizes interior temperature uniformity
fluctuating
ambient temperatures
and protects against
TECHNOLOGY
■ Constant -86ºC ■ Microprocessor Temperature Control ■ Improved Design ■ Temperature & Power Failure Alarms ■ Digital display
SANYO MDF-Series ultra-low temperature freezers maintain internal temperatures as low as -86°C (-123°F). All models use SANYO
uniquely designed compressors for ultra-low temperature applications. Manufactured with foamed-in-place insulation, they are ideally
suited for use in laboratories and hospitals for long-term preservation and storage of blood, specimens and components, as well as materials testing.
Advanced features include a microprocessor temperature control system with digital temperature display, a platinum resistance sensor for
extra precision and reliability, a power failure warning system with built-in audible and visible indicators, double insulation polyurethane
walls and easy open/ easy-close hinged outer door latch.
Specifications
Model
Temperature range
Maximum cooling
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
Shelf
Access port
Compressor
Refrigerant
Alarm
Remote alarm contact
Accessories
MDF-U5386S
–50˚C to –86˚C (1˚C increment)
–86˚C
(Ambient temp. 30˚C)
890 x 875 x 1990mm
(35.0” x 34.4” x 78.3”)
630 x 600 x 1280mm
(24.8” x 23.6” x 50.4”)
Approx. 305kg (672 lbs.)
483 L (17.1 cu.ft.)
Stainless steel, Adjustable,
3 shelves, W608 x D533mm,
Max. load: 50kg (110 lbs.)/shelf
17mm diameter, 3 locations
(back, bottom left/right corner)
Hermetic type, Output: 1100 W
(high stage side), 1100 W (low
stage side)
HFC refrigerant
High/low temperature,
Power failure, Filter check,
Battery check
Allowable contact capacity:
DC 30V, 2A
1 set of keys, 1 scraper
*Appearance and specifications are subject to change without notice.
Pull-down characteristics
40
20
0
–20
–40
–60
–80
–100
0 2 4 6 8 105012
352520
14
1510
Temperature (˚C)
40
20
0
–20
–40
–60
–80
–100
0 2 4 6 8 10 12 14
Temperature (˚C)
Time (hour)
Time (hour)
Pull-up characteristics
during power failure
30
14
8
6
4
2
10
12
0
Time (hour)
Ambient temperature (˚C)
CO2 backup system operation
characteristics
Central air temp. in cabinet
Ambient temp. 30˚C, No-load
Ambient temp. and maintenance
of –70˚C by injection of LCO2 (30kg)
Performance Data
*Cooling performance is indicated by the temperature reached at
the center of the freezer (at ambient temperature of 30°C with no
load). In order to use the freezer at a stable temperature for a long
time, it is recommended that the temperature be set to at least
5°C higher than the indicated lowest temperature.
In addition, depending on the usage conditions, it may not be
possible to reach the indicated lowest temperature.
Dimensions [Unit : mm (inch)]
Specifications
Model
Temperature range
Maximum cooling
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
Shelf
Access port
Compressor
Refrigerant
Alarm
Remote alarm contact
Accessories
Power Requirement
MDF-U5386SC
–50°C to –86°C (1°C increment)
–86°C
(Ambient temp. 30°C)
890 x 875 x 1990mm
(35.0” x 34.4” x 78.3”)
630 x 600 x 1280mm
(24.8” x 23.6” x 50.4”)
Approx. 305kg (672 lbs.)
483 L (17.1 cu.ft.)
Stainless steel, Adjustable,
3 shelves, W608 x D533mm,
Max. load: 50kg (110 lbs.)/shelf
17mm diameter, 3 locations
(back, bottom left/right corner)
Hermetic type, Output: 1100 W
(high stage side), 1100 W (low
stage side)
HFC refrigerant
High/low temperature,
Power failure, Filter check,
Battery check
Allowable contact capacity:
DC 30V, 2A
1 set of keys, 1 scraper
220V / 60 Hz
Built-in voltage booster
*Appearance and specifications are subject to change without notice.
Pull-down characteristics
40
20
0
–20
–40
–60
–80
–100
0 2 4 6 8 105012
352520
14
1510
Temperature (°C)
40
20
0
–20
–40
–60
–80
–100
0 2 4 6 8 10 12 14
Temperature (°C)
Time (hour)
Time (hour)
Pull-up characteristics
during power failure
30
14
8
6
4
2
10
12
0
Time (hour)
Ambient temperature (°C)
CO2 backup system operation
characteristics
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –70°C by injection of LCO2 (30kg)
Performance Data
956 (37.6)
890 (35)
18 (0.7)
48
(1.9)
630 (24.8)
1540 (60.6)450 (17.7)
1990 (78.3)
605.5 (23.8)
105
(4.1)
1280(50.4)
130 (5.1)130 (5.1)
867 (34.1)
11
(0.4)
1625 (64)
795 (31.3)
140
(5.5)
ISO 9001 & 14001 certified
Options
Circular recorder
MTR-C954
MTR-85H
Recorder mounting
bracket
MDF-S3085
(for MTR-85H)
LCO2 backup system
CVK-UB2
LN2 backup system
CVK-UBN2
Inventory rack
SUR-624-I
Rack Capacity: 24
*Cooling performance is indicated by the temperature reached at
the center of the freezer (at ambient temperature of 30C with no
load). In order to use the freezer at a stable temperature for a long
time, it is recommended that the temperature be set to at least 5C
higher than the indicated lowest temperature.
In addition, depending on the usage conditions, it may not be
possible to reach the indicated lowest temperature.
Specifications
Model
Temperature range
Maximum cooling
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
Shelf
Access port
Compressor
Refrigerant
Alarm
Remote alarm contact
Accessories
Power Requirement
MDF-U5386SC
–50°C to –86°C (1°C increment)
–86°C
(Ambient temp. 30°C)
890 x 875 x 1990mm
(35.0” x 34.4” x 78.3”)
630 x 600 x 1280mm
(24.8” x 23.6” x 50.4”)
Approx. 305kg (672 lbs.)
483 L (17.1 cu.ft.)
Stainless steel, Adjustable,
3 shelves, W608 x D533mm,
Max. load: 50kg (110 lbs.)/shelf
17mm diameter, 3 locations
(back, bottom left/right corner)
Hermetic type, Output: 1100 W
(high stage side), 1100 W (low
stage side)
HFC refrigerant
High/low temperature,
Power failure, Filter check,
Battery check
Allowable contact capacity:
DC 30V, 2A
1 set of keys, 1 scraper
220V / 60 Hz
Built-in voltage booster
EVOLUTIONARY DESIGN
Specifications
Model
Temperature range
Maximum cooling
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
Shelf
Access port
MDF-U5386SC
–50°C to –86°C (1°C increment)
–86°C
(Ambient temp. 30°C)
890 x 875 x 1990mm
(35.0” x 34.4” x 78.3”)
630 x 600 x 1280mm
(24.8” x 23.6” x 50.4”)
Approx. 305kg (672 lbs.)
483 L (17.1 cu.ft.)
Stainless steel, Adjustable,
3 shelves, W608 x D533mm,
Max. load: 50kg (110 lbs.)/shelf
17mm diameter, 3 locations
(back, bottom left/right corner)
Pull-down characteristics
40
20
40
20
0
–20
–40
–60
–80
–100
0 2 4 6 8 10 12 14
Temperature (°C)
Time (hour)
Pull-up characteristics
during power failure
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Performance Data
Back
cover
Compressors
Front
Rear
Air
■
The newly developed refrigeration system and freezer
structure minimizes intrusive operating noise.
■
Achieves 10% higher energy conservation than competing
models.
■
The newly developed back cover combined with new
aerodynamically designed and placed components in the
refrigeration compartment provide superior air flow, making
it possible to drastically reduce the stress to the freezer and
p r e s e r v a t i o n
contributing to excellent durability.
■
Two independent and insulated inner doors ensure maximum interior chamber uniformity.
ACCESSIBILITY
■
A new rugged inner door latch tightly closes the inner door
and makes accessing chamber contents easier.
■
The condenser filter is located at the bottom right side of
the front panel to make filter removal and cleaning easier.
■
Alarm lamp and buzzer offer secure warning of power
failure or abnormal temperature increase.
■
New rugged, one-handed outer door latch has a hasp to
allow a padlock for additional security
SAFETY
■
High and low temperature warning provides an audible and
visual alarm when the temperature deviates more than ±5°C
to ±20°C (adjustable) from the set point.
■
Alarm ring back function ensures that buzzer will resume
operation, should alarm conditions continue after it is
silenced.
■
Microprocessor-controlled filter-clog check function
protects the refrigeration circuit.
■
Control panel with digital display for easy operation.
*Appearance and specifications are subject to change without notice.
*Cooling performance is indicated by the temperature reached at the center of the freezer (at ambient temperature of 30°C with no load). In order to use the freezer at a
stable temperature for a long time, it is recommended that the temperature be set to at least 5°C higher than the indicated lowest temperature. In addition, depending on the
usage conditions, it may not be possible to reach the indicated lowest temperature.
w w w . s a n y o b i o m e d i c a l . c o m
1062 Thorndale Avenue, Bensenville, IL 60106 USA
Toll Free USA 800-858-8442 • Fax 630-238-0074
1- 300 Applewood Crescent, Concord, Ontario L4K 5C7
905-760-4025 • Fax 905-760-9945
SANYO E&E America Company
www.sanyobiomedical.com
SANYO Canada, Inc.
DMREV606.1