If the chamber temperature is higher
than the temperature at which the
high temperature alarm is activated.
If the chamber temperature is lower
than the temperature at which the
low temperature alarm is activated.
In the case of power failure.
When power switch is turned OFF.
When the power to the unit is
disconnected.
When there is no key pressing in
each setting mode for 90 seconds.
If the thermal sensor is disconnected.
If the thermal sensor is short-circuited.
If the protective sensor for compressor
is disconnected.
If the protective sensor for compressor
is short-circuited.
When the battery switch is OFF at the
time of alarm test.
In the case of failure of compressor
cooling fan motor.
In the case of abnormal high
temperature due to the dust on the
condenser.
In the case of abnormal high ambient
temperature.
Note:
1. The above power failure alarm is available when the battery switch is ON and the battery is charged.
If the battery switch is OFF or the battery is discharged, only the remote alarm is activated.
2. The power failure alarm can be kept about 12 hours with the battery charged completely. 2-day
operation of the freezer is needed to charge the battery full.
3. The chamber temperature is displayed for 5 seconds if the BUZZER key is depressed during the power
failure alarm. After that, the alarm stops and the alarm lamp blinks.
4. The remote alarm is silenced by pressing BUZZER key as the remote alarm is operated in conjunction
with the buzzer, except for the power failure alarm.
5. After power failure, the operation is resumed with the condition before power failure since the
temperature setting and alarm temperature setting are memorized in a nonvolatile memory.
ALARM lamp is flashed.
Temperature indicator is
flashed.
ALARM lamp is flashed.
Temperature indicator is
flashed.
ALARM lamp is flashed. Intermittent tone Remote alarm.
Chamber temperature is
displayed.
ALARM lamp is flashed.
E01 and chamber temp. are
displayed alternately.
ALARM lamp is flashed.
E02 and chamber temp. are
displayed alternately.
ALARM lamp is flashed.
E03 and chamber temp. are
displayed alternately.
ALARM lamp is flashed.
E04 and chamber temp. are
displayed alternately.
E09 is flashed. Intermittent tone Remote alarm.
E10 is flashed. Intermittent tone
Intermittent tone with
15 minutes delay.
Intermittent tone with
15 minutes delay.
Intermittent tone
Intermittent tone
Intermittent tone
Intermittent tone
-----
Remote alarm with 15
minutes delay.
Remote alarm with 15
minutes delay.
Finishing of each
Setting mode.
Change of setting is
disable.
Remote alarm.
Continuous running.
Remote alarm.
Continuous running.
Remote alarm.
Normal operation.
Remote alarm.
Normal operation.
Remote alarm.
Compressor stops
running.
18
SETTING OF ALARM RESUME TIME
The alarm buzzer and remote alarm are silenced by pressing BUZZER key on the control panel during
alarm condition. The buzzer and remote alarm will be activated again after certain suspension if the
alarm condition is continued. The suspension time can be set by following the procedure shown in the
Table 6 below.
The example in the table is based on the assumption that the desired duration is 20 minutes.
Note: The duration is set in 30 minutes at the factory.
Table 6. Setting procedure for alarm resuming time (change from 30 minutes to 20 minutes)
Description of operation Key operatedIndication after operation
1 ----
2 Press key for 5 seconds.
Set the figure to F25 with the
3
key and key.
4 Press SET key. SET
Set the figure to 020 with the
5
key.
6 Press SET key. SET
• The settable alarm resume time are 10, 20, 30, 40, 50, or 60 minutes. The buzzer and remote alarm
would not reset if the reset time is set in 000.
• The setting of alarm reset time cannot be changed during the defrosting.
• The buzzer and remote alarm during power failure or alarm testing cannot be silenced.
• The set mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing SET key is not memorized.
The first digit is flashed.
When pressed, the figure of settable
The current chamber temperature is
displayed.
Pressing the key shifts the digit which can be set.
digit increases.
The current reset time is displayed.
The middle digit is flashed.
When pressed, the figure of settable
digit increases.
Alarm temperature is memorized and
the current chamber temperature is
displayed.
19
ROUTINE MAINTENANCE
WARNING
Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in
order to prevent electric shock or injury.
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of
maintenance. These may be harmful to your health.
CAUTION
Always put on the dry gloves to protect the hands at the time of maintenance. No gloves may cause
cut of the finger by the edge or corner.
Cleaning of cabinet
1. Clean the unit once a month. Regular cleaning keeps the unit looking new.
2. Use a dry cloth to wipe off small amounts of dirt on the outside and inside of the unit and all
accessories. If the unit is very dirty, use a neutral detergent.
3. After cleaning, wipe away the cleaner completely with a cloth washed in clean water.
4. Never pour water onto or into the unit. Doing so can damage the electric insulation and cause failure.
・The compressor and other mechanical parts are completely sealed. This unit requires absolutely no
lubrication.
・Remove dust from the power supply plug periodically.
Defrosting
This freezer is direct-cooling type and the frost is built on the chamber wall during long term operation.
The excessive frost possibly make some gap between the cabinet and door gasket, which may cause
poor cooling. Remove the frost inside the chamber once a month. Following shows the procedure for
removing the chamber frost.
1. Temporarily move all the contents of freezer chamber to another freezer.
2. Press DEF key for 5 seconds to stop the refrigerating operation. While the refrigerating operation is
stopped, the current chamber temperature and dF is displayed on the control panel alternatively.
3. After a several hours, check visually that all defrost was removed completely.
4. Remove the cap of drain port on the bottom of the chamber and drain the accumulated water.
5. Wipe up the water remaining in the freezer chamber and then replace the cap of drain port.
6. Press DEF key so that the refrigerating operation can be started.
7. Once the chamber temperature has dropped to the desired temperature, place the original contents
back in the freezer chamber.
Note:
After the defrosting, the refrigerating operation is never resumed automatically. Make sure to press DEF
key to start the freezer operation after defrosting.
While the freezer stops refrigerating operation for defrosting, neither high temperature alarm nor low
temperature alarm is effective.
20
TROUBLE SHOOTING
If the unit malfunctions, check out the following before calling for service.
The chamber is not cooled at all
1. The circuit breaker of power source is active.
2. The voltage is too low. (In this case, call an electrician.)
3. The power switch is not ON.
4. The large amount of articles (load) is stored in the chamber at one time.
5. The freezer is in defrost condition.
The cooling is poor
1. The ambient temperature is too high.
2. The door is not closed firmly.
3. The large amount of frost is built on the chamber wall.
4. The air intake vent is blocked.
5. The set temperature is not inputted properly.
6. The freezer is in the direct sunlight.
7. There is any heating source near the freezer.
8. A rubber cap and insulation for the access port are not set correctly.
9. You put too many unfrozen articles into the freezer compartment.
When the unit does not accept changes of set-point temperature
1. The key lock is set in “ON” mode.
Noise
1. The freezer is not installed on the sturdy floor.
2. The freezer is not leveled with the leveling feet.
3. There is anything touching the frame.
4. The freezer is in the status immediately after start up.
The unit sometimes causes a noise when the chamber temperature is high due to the large load. The
noise gets less and less accompanying with the cooling of the chamber.
DISPOSAL OF UNIT
WARNING
If the unit is to be stored unused in an unsupervised area for an extended period ensure that children
do not have access and doors cannot be closed completely.
The disposal of the unit should be accomplished by appropriate personnel. Always remove
doors to prevent accidents such as suffocation.
21
SPECIFICATIONS
NameBiomedical Freezer
ModelMDF-136 MDF-236 MDF-436
External dimensionsW640 x D687 x H881 (mm)W905 x D687 x H881 (mm) W1265 x D807 x H905 (mm)
Internal dimensionsW525 x D440 x H715 (mm)W790 x D440 x H715 (mm) W1140 x D550 x H735 (mm)
Effective capacity138 L 221 L 426 L
ExteriorPolyester finish baked on zinc galvanized steel
InteriorColored aluminum plate
DoorPolyester finish baked on zinc galvanized steel
InsulationRigid polyurethane foamed-in place (CFC-FREE)
Baskets Polyethylene coated steel wire
Access port Diameter 17 mm, Right side and bottom left
Compressor Hermetic type, 175 W Hermetic type, 350 W
Condenser Wire and tube type
Evaporator Aluminum tube on sheet type
RefrigerantR-404A
Temperature controller Electronics controller (between -18 and -40oC)
Temperature display Digital display (between -50 and +50oC)
Temperature sensor Thermister sensor
Temperature alarm Flash of digital indicator and alarm lamp, Buzzer, (Remote alarm)