See the Service Manual of EM-F3400SW (SM-860304)
except the items described in this Service Manual.
EM-V3405SW (U.S.A.)EM-V3405SW (U.S.A.)
EM-V3405SW (U.S.A.)
EM-V3405SW (U.S.A.)EM-V3405SW (U.S.A.)
oduct Code No.oduct Code No.
oduct Code No.
oduct Code No.oduct Code No.
437 550 00437 550 00
437 550 00
437 550 00437 550 00
CAUTIONCAUTION
CAUTION
CAUTIONCAUTION
WW
ARNING TO SERARNING TO SER
W
ARNING TO SER
WW
ARNING TO SERARNING TO SER
VICE TECHNICIANSVICE TECHNICIANS
VICE TECHNICIANS
VICE TECHNICIANSVICE TECHNICIANS
PRECAUTIONS TO BE OBSERVED BEFORE AND DURING
SERVICING TO AVOID POSSIBLE EXPOSURE TO
EXCESSIVE MICROWAVE ENERGY
(a) Do not operate or allow the oven to be operated with the door open.
(b) Make the following safety checks on all ovens to be serviced before activating the magnetron or other
microwave source, and make repairs as necessary :
(1) Interlock operation, (2) proper door closing, (3) seal and sealing surfaces (arcing, wear, and other damage),
(4) damage to or loosening of hinges and latches, (5) evidence of dropping or abuse.
(c) Before turning on microwave power for any service test or inspection within the microwave generating
compartments, check the magnetron, wave guide or transmission line, and cavity for proper alignment, integrity,
and connections.
(d) Any defective or misadjusted components in the interlock, monitor, door seal, and microwave generation and
transmission systems shall be repaired, replaced, or adjusted by procedures described in this manual before
the oven is released to the owner.
(e) A microwave leakage check to verify compliance with the Federal per formance standard should be performed
on each oven prior to release to the owner.
REFERENCE NO. SS-860312
CAUTIONCAUTION
CAUTION
CAUTIONCAUTION
For micrFor micr
For micr
For micrFor micr
owave enerowave ener
owave ener
owave enerowave ener
gy emissiongy emission
gy emission
gy emissiongy emission
In the area emitting the highest reading, switch the meter
to SLOW RESPONSE, and take a reading for minimum of
three (3) seconds. We recommended the pattern outline
shown below when the door surface is surveyed.
On every service calls, check for microwave energy
emission, must be made according to the following manner.
MeasurMeasur
Measur
MeasurMeasur
Measurement must be made with the microwave oven
operating at its maximum output and containing a load of
275±15 milliliters of tap water initially at 20
±9oF) placed within the cavity at the center.
NOTENOTE
NOTE : The water container must be a 600 milliliter beaker
NOTENOTE
A properly operating door and seal assembly will normally
register emission on greater than 4mW/cm2 to allow for
measurement uncertainty with the cooking shelf or tray in
place.
All repairs must be performed in such a manner All repairs must be performed in such a manner
All repairs must be performed in such a manner
All repairs must be performed in such a manner All repairs must be performed in such a manner
micrmicr
micr
micrmicr
Follow the instructions supplied with a detector being used
and performed an R.F. emission test around the door front
and edges and all edges and vent of the outer case. The
cabinet (wrapper) must be in place and the oven fully
assembled.
When performing emission survey, with the meter on FAST
RESPONSE the movement of a detector probe shall not
exceed one (1) inch per second.
ement of enerement of ener
ement of ener
ement of enerement of ener
and made of an electrically none conductive
material such as glass or plastic.
The cook tray must be in place when measuring
emission.
owave enerowave ener
owave ener
owave enerowave ener
gy emissiongy emission
gy emission
gy emissiongy emission
gy emission argy emission ar
gy emission ar
gy emission argy emission ar
e minimal.e minimal.
e minimal.
e minimal.e minimal.
o±5o
celsius (68
thatthat
that
thatthat
NOTENOTE
NOTE : Periodically check to be sure that the probe tip is
NOTENOTE
not worn or dirty.
The following U.S. standard applies to microwave ovens :
21 CFR 1030.10, Performance Standard for Microwave
Ovens.
It requires that the power density of the microwave
radiation emitted by a microwave oven shall not exceed
five (5) milliwatts per square centimeter at any point 5
centimeter (about 2 inches) or more from the external
surface of the oven.
All microwave ovens exceeding the emission level of
4mW/cm
microwave ovens and the manufacturer immediately and
the owner should be told not to use the microwave oven
until it has been repaired completely.
If a microwave oven is found to operate with the door open,
report to Dept. of Service, the manufacturer and CDRH*
immediately. Also tell the owner not to use the oven.
*CDRH : Center for Device and Radiological Health.
The interlock monitor switch acts as the final safety switch
protecting the customer from microwave radiation. If the
interlock monitor switch operates to blow the fuse with
interlocks failed, you must replace all interlock switches
primary and secondary interlock switches and the monitor
switch with new ones because the contacts of those
interlock switches may be melted and welded together.
(C) 1. Set cooking time to two (2) minutes. (“2 00”
appears in display)
2 . Touch “START” key and operate oven for exactly
two (2) minutes.
(D) 1. Take out the two bowls at once.
2 . Stir both water with thermometer and measure the
water temperature rise respectively.
(E) 1 . Get temperature rise by calculation the difference
(water temperature after cooking minus initial
temperature) in each bowl.
2 . Then calculate average value(t) of both temperature
rises in degrees centigrade.
3. Then work out :
Power Output (watt) = 70 x ∆t
2. POWER OUTPUT MEASUREMENT2. POWER OUTPUT MEASUREMENT
2. POWER OUTPUT MEASUREMENT
2. POWER OUTPUT MEASUREMENT2. POWER OUTPUT MEASUREMENT
NOTENOTE
NOTE
NOTENOTE
The Power Output specification 1050W on this model
is measured with IEC measurement. The Power Output
is measured with two (2) liters water is equivalent to
1050W in measurement with IEC, when measured with
the following Power Output.
(A) 1 . Fill two test bowls with each 1 liter water respectively.
2 . Use accurate thermometer and measure each
water temperature respectively.
(B) Place the two bowls on glass turntable.
Where ∆t is an Average Temperature Rise in degrees
centigrade.
(F)Power Output shall be in the following range :
Average Microwave
Temperature RisePower Output
Minimum
24.3oF (13.5oC) 945W
Maximum
31.1oF (17.3oC)1211W
(G) Power Output will be influenced by line voltage of
power supply. Consequently, correct Power Output
must be measured within 120V AC ± 1 volt while
unit is operating.
- 1 -
3. CIRCUIT DIAGRAM3. CIRCUIT DIAGRAM
3. CIRCUIT DIAGRAM
3. CIRCUIT DIAGRAM3. CIRCUIT DIAGRAM
WIRING COLOR
O : ORANGEB K : BLACK
R: REDBL : BLUE
W : WHITEY: YELLOW
- 2 -
Figure 1Figure 1
Figure 1
Figure 1Figure 1
4. TEST PROCEDURES4. TEST PROCEDURES
4. TEST PROCEDURES
4. TEST PROCEDURES4. TEST PROCEDURES
COMPONENT COMPONENT
COMPONENT
COMPONENT COMPONENT
TOUCH KEY
BOARD
--
-
--
CHECKOUT PROCEDURE CHECKOUT PROCEDURE
CHECKOUT PROCEDURE
CHECKOUT PROCEDURE CHECKOUT PROCEDURE
Measure the resistance between terminals of
FPC connector after removing it from S101.
(Figure 2).
NOTE NOTE
NOTE
NOTE NOTE
- When reconnecting the FPC connector, makeWhen checking “START” key, connect
sure the holes on the connector are properlyOhm-Meter as illustration below
inserted in hooks of the plastic fastener in S101.
MAMA
065
TOTAL
1
7
DEFROST
TRIX CIRCUIT FORTRIX CIRCUIT FOR
MA
TRIX CIRCUIT FOR
MAMA
TRIX CIRCUIT FORTRIX CIRCUIT FOR
TOUCH KEY BOARD TOUCH KEY BOARD
TOUCH KEY BOARD
TOUCH KEY BOARD TOUCH KEY BOARD
FPC CONNECTORFPC CONNECTOR
FPC CONNECTOR
FPC CONNECTORFPC CONNECTOR
POWER
LEVEL
QUICK
ON
SOUP /
BEVERAGE
POTATO
MORE +
LESS -
RESUL RESUL
RESUL
RESUL RESUL
Resistance
value
Ohm-MeterOhm-Meter
Ohm-Meter
Ohm-MeterOhm-Meter
TT
T
TT
WhenWhen not
touchedtouched
Less than More than
1K Ohms 1 MEG Ohms
..
.
..
2
3
4
HIGH-VOLTAGE
TRANSFORMER
8
9
STOP /
CLEAR
PIZZA
SLICE
FRESH
VEGETABLE
1)Measure the resistance :Normal readings :
With an Ohm-Meter on R x 1 scale.
a. Primary winding:Approximately 0.30 Ohms.
b. Filament winding:Less than 1 Ohm.
c. Secondary winding :Approximately 74.2 Ohms.
2)Measure the resistance :Normal readings :
With an Ohm-Meter on highest scale.
a. Primary winding to ground :Infinite Ohms.
b. Filament winding to ground :Infinite Ohms.
Ohm-MeterOhm-Meter
Ohm-Meter
Ohm-MeterOhm-Meter
QUICK
DEFROST
START
CLOCK
FROZEN
ENTREE
FROZEN
VEGETABLE
POPCORN
KITCHEN
TIMER
ADD
30 SEC
CUSTOM
PROGRAMS
TERMINALS OF FPC CONNECTORTERMINALS OF FPC CONNECTOR
TERMINALS OF FPC CONNECTOR
TERMINALS OF FPC CONNECTORTERMINALS OF FPC CONNECTOR
Figure 2Figure 2
Figure 2
Figure 2Figure 2
Note Note
Note : Remove varnish of
Note Note
measured point.
- 3 -
Figure 3Figure 3
Figure 3
Figure 3Figure 3
5.5.
EXPLODED VIEW AND PEXPLODED VIEW AND P
5.
EXPLODED VIEW AND P
5.5.
EXPLODED VIEW AND PEXPLODED VIEW AND P
CAVITY PARTS
ARAR
AR
ARAR
TS LISTTS LIST
TS LIST
TS LISTTS LIST
* * * ALL SERVICE ON MICROWAVE OVENS SHOULD BE PERFORMED BY A QUALIFIED TECHNICIAN USING
APPROVED TESTING EQUIPMENT. CUSTOMERS SHOULD NOT ATTEMPT TO REPLACE PARTS IDENTIFIED
BY A TRIPLE ASTERISK (***).
Key
No.
Part No.DescriptionQ’ty
1617 228 8321Cabinet1
2411 082 5201SCR TPG TRS 4 x 10 Z18
3411 160 6007SCR TPG TRS + SRT 4 x 10 Z12
4617 235 3234Power Cord1
5411 160 6106SCR S-T TRS + SRT 4 x 10 Z11
6617 235 2329Oven Cavity*** (Not Service Part)1
7617 124 1068Gear Motor1
8411 010 5808SCR EVR PAN 4 x 10 Z12
9617 138 5601Bottom Plate*** (Not Service Part) 1
10617 221 6089Plastic Foot with Canoe Clip4
11411 011 0802Bolt Hex + SW + W5 x 14 Z1*** 2
12617 223 3116Hinge, Lower***1
13411 089 2500Washer F 5 x 10 x 0.81
NoteNote
Note : Screws securing Cabinet cannot be removed by regular screwdriver.
NoteNote
KEY NO. 26. SPECIAL SCREW REQUIRES SPECIAL DRIVER
(PART NO. 617 209 3734) OR SPECIAL BIT (PART NO. 617 209 3741)
For Parts or Service ContactFor Parts or Service Contact
For Parts or Service Contact
For Parts or Service ContactFor Parts or Service Contact
SANYO FISHER SERSANYO FISHER SER
SANYO FISHER SER
SANYO FISHER SERSANYO FISHER SER
21605 Plummer Street,
Chatsworth, CA91311 U.S.A.
VICE COMPVICE COMP
VICE COMP
VICE COMPVICE COMP
ANYANY
ANY
ANYANY
File No :File No :
File No :
File No :File No :
SER SER
SER
SER SER
VICE MANUALVICE MANUAL
VICE MANUAL
VICE MANUALVICE MANUAL
Micr Micr
Micr
Micr Micr
owave Ovenowave Oven
owave Oven
owave Ovenowave Oven
Pr Pr
oduct Code No.oduct Code No.
Pr
oduct Code No.
Pr Pr
oduct Code No.oduct Code No.
EM-F3400SW EM-F3400SW
EM-F3400SW
EM-F3400SW EM-F3400SW
437 450 57437 450 57
437 450 57
437 450 57437 450 57
(U.S.A.)(U.S.A.)
(U.S.A.)
(U.S.A.)(U.S.A.)
CAUTIONCAUTION
CAUTION
CAUTIONCAUTION
WW
ARNING TO SERARNING TO SER
W
ARNING TO SER
WW
ARNING TO SERARNING TO SER
VICE TECHNICIANSVICE TECHNICIANS
VICE TECHNICIANS
VICE TECHNICIANSVICE TECHNICIANS
PRECAUTIONS TO BE OBSERVED BEFORE AND DURING
SERVICING TO AVOID POSSIBLE EXPOSURE TO
EXCESSIVE MICROWAVE ENERGY
(a) Do not operate or allow the oven to be operated with the door open.
(b) Make the following safety checks on all ovens to be serviced before activating the magnetron or other
microwave source, and make repairs as necessary :
(1) Interlock operation, (2) proper door closing, (3) seal and sealing surfaces (arcing, wear, and other damage),
(4) damage to or loosening of hinges and latches, (5) evidence of dropping or abuse.
(c) Before turning on microwave power for any service test or inspection within the microwave generating
compartments, check the magnetron, wave guide or transmission line, and cavity for proper alignment, integrity,
and connections.
(d) Any defective or misadjusted components in the interlock, monitor, door seal, and microwave generation and
transmission systems shall be repaired, replaced, or adjusted by procedures described in this manual before
the oven is released to the owner.
(e) A microwave leakage check to verify compliance with the Federal per formance standard should be performed
on each oven prior to release to the owner.
REFERENCE NO. SM-860304
CAUTIONCAUTION
CAUTION
CAUTIONCAUTION
For micrFor micr
For micr
For micrFor micr
On every service calls, check for microwave energy
emission, must be made according to the following manner.
MeasurMeasur
Measur
MeasurMeasur
Measurement must be made with the microwave oven
operating at its maximum output and containing a load of
275±15 milliliters of tap water initially at 20
±9oF) placed within the cavity at the center.
NOTENOTE
NOTE : The water container must be a 600 milliliter beaker
NOTENOTE
A properly operating door and seal assembly will normally
register emission on greater than 4mW/cm2 to allow for
measurement uncertainty with the cooking shelf or tray in
place.
All repairs must be performed in such a manner All repairs must be performed in such a manner
All repairs must be performed in such a manner
All repairs must be performed in such a manner All repairs must be performed in such a manner
micrmicr
micr
micrmicr
Follow the instructions supplied with a detector being used
and performed an R.F. emission test around the door front
and edges and all edges and vent of the outer case. The
cabinet (wrapper) must be in place and the oven fully
assembled.
When performing emission survey, with the meter on FAST
RESPONSE the movement of a detector probe shall not
exceed one (1) inch per second.
owave enerowave ener
owave ener
owave enerowave ener
ement of enerement of ener
ement of ener
ement of enerement of ener
and made of an electrically none conductive
material such as glass or plastic.
The cook tray must be in place when measuring
emission.
owave enerowave ener
owave ener
owave enerowave ener
gy emissiongy emission
gy emission
gy emissiongy emission
gy emissiongy emission
gy emission
gy emissiongy emission
gy emission argy emission ar
gy emission ar
gy emission argy emission ar
e minimal.e minimal.
e minimal.
e minimal.e minimal.
o±5o
celsius (68
thatthat
that
thatthat
In the area emitting the highest reading, switch the meter
to SLOW RESPONSE, and take a reading for minimum of
three (3) seconds. We recommended the pattern outline
shown below when the door surface is surveyed.
NOTENOTE
NOTE : Periodically check to be sure that the probe tip is
NOTENOTE
not worn or dirty.
The following U.S. standard applies to microwave ovens :
21 CFR 1030.10, Performance Standard for Microwave
Ovens.
It requires that the power density of the microwave
radiation emitted by a microwave oven shall not exceed
five (5) milliwatts per square centimeter at any point 5
centimeter (about 2 inches) or more from the external
surface of the oven.
All microwave ovens exceeding the emission level of
4mW/cm
microwave ovens and the manufacturer immediately and
the owner should be told not to use the microwave oven
until it has been repaired completely.
If a microwave oven is found to operate with the door open,
report to Dept. of Service, the manufacturer and CDRH*
immediately. Also tell the owner not to use the oven.
*CDRH : Center for Device and Radiological Health.
The interlock monitor switch acts as the final safety switch
protecting the customer from microwave radiation. If the
interlock monitor switch operates to blow the fuse with
interlocks failed, you must replace all interlock switches
primary and secondary interlock switches and the monitor
switch with new ones because the contacts of those
interlock switches may be melted and welded together.
NOTE : If the interlock monitor circuit operates and at the
NOTENOTE
same time the fuse blows with the door opened,
be sure to replace the control circuit board because
relay 2 on the control circuit board, the door sensing
switch and the electric circuit related on the
door sensing switch, which act as Secondary
Interlock Switch.
Door
Latch
Lever Stopper
Door
Sensing
Switch
(1) Loosen 2 screws securing the lever stopper.
(2) Adjust the lever stopper position so that it is pushed
forward and pull backward until there is about zero
gap between the latch lever and switch body on the
primary interlock switch and at the same time there
is about zero gap between the door latch and the
switch body on the door sensing switch when the
door latch is securely locked.
(3) Tighten the lever stopper screws securely.
(4) Make sure the interlock monitor switch closes after
the primary interlock switch opens when the door
is opened very slowly, according to “CHECKOUT
PROCEDURE FOR SWITCHES” on page 6.
(5) Make sure the interlock monitor switch opens before
the primary interlock switch closes when the door is
closed very slowly, according to “CHECKOUT
PROCEDURE FOR SWITCHES” on page 6.
(6) Make sure the microwave energy leakage should be
no greater than 4mW/cm2 to allow for measurement
uncertainty when measured with a detector.
(All service adjustments must be made for
minimum microwave energy leakage readings.)
Interlock
Monitor
Switch
Screws
- 1 -
Figure 1Figure 1
Figure 1
Figure 1Figure 1
Latch
Lever
Primary
Interlock
Switch
2.2.
SPECIFICASPECIFICA
2.
SPECIFICA
2.2.
SPECIFICASPECIFICA
Rated Power Consumption . . 1480W.
Microwave Output ................ 1050W.
Frequency ............................. 2,450MHz±50MHz.
Power Supply ........................120V±12V, 60Hz.
Rated Current ....................... 12.9 Amp.
Safety Devices ..................... Thermal Fuse open at
Timer .....................................Electronic Digital, up to
Overall Dimensions ................ 205/8”(W) x 161/2”(D) x 113/8”(H)
Oven Cavity Size .................. 133/4”(W) x 145/8”(D) x 81/2”(H)
Net Weight ........................... Approx. 34.5 Lbs.
3.3.
POWER OUTPUT MEASUREMENTPOWER OUTPUT MEASUREMENT
3.
POWER OUTPUT MEASUREMENT
3.3.
POWER OUTPUT MEASUREMENTPOWER OUTPUT MEASUREMENT
NOTENOTE
NOTE
NOTENOTE
The power output specification, 1050W. on this model is
measured with IEC measurement. The power output is
measured with two (2) liters water is equivalent to 1050W.
in measurement with IEC, when measured with the following
power output.
(1) Fill two beakers (glass or plastic) with each one liter of
tap water (about 20°C) and measure the water temperature.
(Use a thermometer with a 1/10 degree gauge).
(2) Place the beakers side by side in the center of the glass tray.
(3) Close the door, set the “TIME” for two minutes. (“200” in
the display window). Touch the “START” key and heat
the water exactly for two minutes.
(4) Take the beakers out, immediately stir the water and
measure the water temperature respectively.
(5) Calculate the temperature rise of water in each beaker.
Then calculate the average value of two temperature rises.
Output power can be calculated by the equation :
Power Output (W) = 70 x ∆t
Where ∆t is an average temperature rise in degrees
Centigrade.
(6) Power Output shall be in the following range :
(7) Power Output is affected by the line voltage under load.
For correct Power Output measurement, the line voltage
under load must be 120±1 volts.
TIONSTIONS
TIONS
TIONSTIONS
332°F (167°C) for Cavity.
Thermal Protector open at
252°F (122°C) for Magnetron.
Fuse (Cartridge Type 20A)
Primary Interlock Switch,
Door Sensing Switch and
Relay 2.
Interlock Monitor Switch.
99 min. 99 sec.
Average Temperature RisePower Output
4.4.
PRECAUTIONS AND REPPRECAUTIONS AND REP
4.
PRECAUTIONS AND REP
4.4.
PRECAUTIONS AND REPPRECAUTIONS AND REP
PRELIMINARPRELIMINAR
PRELIMINAR
PRELIMINARPRELIMINAR
A.A.
SINCE NEARLSINCE NEARL
A.
SINCE NEARL
A.A.
SINCE NEARLSINCE NEARL
CUITS OF THIS MICROWCUITS OF THIS MICROW
CUITS OF THIS MICROW
CUITS OF THIS MICROWCUITS OF THIS MICROW
BE CARRIED OUT WITH GREABE CARRIED OUT WITH GREA
BE CARRIED OUT WITH GREA
BE CARRIED OUT WITH GREABE CARRIED OUT WITH GREA
B.B.
TO ATO A
B.
TO A
B.B.
TO ATO A
ENERGY LEAKAGE, THE FOLLOWING PRECAUTIONSENERGY LEAKAGE, THE FOLLOWING PRECAUTIONS
ENERGY LEAKAGE, THE FOLLOWING PRECAUTIONS
ENERGY LEAKAGE, THE FOLLOWING PRECAUTIONSENERGY LEAKAGE, THE FOLLOWING PRECAUTIONS
MUST BE TMUST BE T
MUST BE T
MUST BE TMUST BE T
(1) Before the power is applied :
(a) Open and close door several times to make sure the
(b) Make sure the perforated screen and the choke die-
(2) After the power is applied :
(a) Open and close the door to see if the interlock mech-
(b) Check microwave energy leakage with a leakage
(3) Do not operate the unit until it is completely repaired,
if any of the following conditions exists :
(a) Door does not close firmly against the cavity front.
(b) The hinge is broken.
(c) The choke dielectric or the door seal is damaged.
(d) The door is bent or warped, or there is any other
NOTE :
(e) Make sure that there are no defective parts in the
(f) Make sure there are no defective parts in the
(4) The following items should be checked after the unit
is repaired :
(a) The interlock monitor switch is connected correctly
(b) The magnetron gasket on the magnetron is properly
(c) Waveguide and oven cavity are intact (no leakage of
(d) The door can be properly closed and the safety
(e) The oven must be stopped when the door is opened
The oven must not be operated with any of the aboveThe oven must not be operated with any of the above
The oven must not be operated with any of the above
The oven must not be operated with any of the aboveThe oven must not be operated with any of the above
components rcomponents r
components r
components rcomponents r
YY
Y
YY
Y 2,000 VOLY 2,000 VOL
Y 2,000 VOL
Y 2,000 VOLY 2,000 VOL
VOID POSSIBLE EXPOSURE TO MICROWVOID POSSIBLE EXPOSURE TO MICROW
VOID POSSIBLE EXPOSURE TO MICROW
VOID POSSIBLE EXPOSURE TO MICROWVOID POSSIBLE EXPOSURE TO MICROW
AKEN BEFORE SERAKEN BEFORE SER
AKEN BEFORE SER
AKEN BEFORE SERAKEN BEFORE SER
primary interlock switch, the interlock monitor switch
and the door sensing switch operate properly.
(Listen for the clicking sound from the switches).
Make sure the interlock monitor switch closes after
the primary interlock switch is opens when the door
is opened. (See pages 1 and 6).
lectric of the door are correctly mounted.
anism operates properly.
detector and confirm the energy leakage should be
no greater than 4mW/cm
uncertainty.
visible damage to the oven that may cause microwave
energy leakage.
Always keep the seal clean.Always keep the seal clean.
Always keep the seal clean.
Always keep the seal clean.Always keep the seal clean.
interlock mechanism.
microwave generating and transmission assembly.
(especially waveguide).
and firmly.
positioned.
microwave energy).
switches work properly.
or the time is up.
emoved or bypassed.emoved or bypassed.
emoved or bypassed.
emoved or bypassed.emoved or bypassed.
AIR SERAIR SER
AIR SER
AIR SERAIR SER
TS EXISTS IN SOME CIR-TS EXISTS IN SOME CIR-
TS EXISTS IN SOME CIR-
TS EXISTS IN SOME CIR-TS EXISTS IN SOME CIR-
AA
VE OVEN, REPVE OVEN, REP
A
VE OVEN, REP
AA
VE OVEN, REPVE OVEN, REP
T CARE.T CARE.
T CARE.
T CARE.T CARE.
VICING.VICING.
VICING.
VICING.VICING.
2
to allow for measurement
VICE TIPSVICE TIPS
VICE TIPS
VICE TIPSVICE TIPS
AIRS SHOULDAIRS SHOULD
AIRS SHOULD
AIRS SHOULDAIRS SHOULD
AA
VEVE
A
VE
AA
VEVE
- 2 -
5. CIRCUIT DIAGRAM
DNOCTINOI
Figure 2Figure 2
Figure 2
Figure 2Figure 2
KCOLRETNIYRADNOCES
YRAMIRP
EDAMHCTIWS
•
NEPOROOD
ESOLCROOD
•••
KCOLRETNI
HCTIWS
MOCMOCMOCMOC
ONCNONON
•
- 3 -
KCOLRETNI
ROTINOM
HCTIWS
ROOD
GNISNES
HCTIWS
2YALER
6. TEST PROCEDURES AND TROUBLESHOOTING6. TEST PROCEDURES AND TROUBLESHOOTING
6. TEST PROCEDURES AND TROUBLESHOOTING
6. TEST PROCEDURES AND TROUBLESHOOTING6. TEST PROCEDURES AND TROUBLESHOOTING
CAUTIONCAUTION
CAUTION
CAUTIONCAUTION
- DISCONNECT THE POWER SUPPL- DISCONNECT THE POWER SUPPL
- DISCONNECT THE POWER SUPPL
- DISCONNECT THE POWER SUPPL- DISCONNECT THE POWER SUPPL
THE WTHE W
ALL OUTLET WHENEVER REMOVING THEALL OUTLET WHENEVER REMOVING THE
THE W
ALL OUTLET WHENEVER REMOVING THE
THE WTHE W
ALL OUTLET WHENEVER REMOVING THEALL OUTLET WHENEVER REMOVING THE
CABINET FROM THE UNITCABINET FROM THE UNIT
CABINET FROM THE UNIT
CABINET FROM THE UNITCABINET FROM THE UNIT
TESTS ONLTESTS ONL
TESTS ONL
TESTS ONLTESTS ONL
VOLVOL
TT
VOL
T
VOLVOL
TT
WIRES FROM THE PRIMARWIRES FROM THE PRIMAR
WIRES FROM THE PRIMAR
WIRES FROM THE PRIMARWIRES FROM THE PRIMAR
HIGH VOLHIGH VOL
HIGH VOL
HIGH VOLHIGH VOL
A.A.
TEST PROCEDURESTEST PROCEDURES
A.
TEST PROCEDURES
A.A.
TEST PROCEDURESTEST PROCEDURES
Y AFTER DISCHARGING THE HIGHY AFTER DISCHARGING THE HIGH
Y AFTER DISCHARGING THE HIGH
Y AFTER DISCHARGING THE HIGHY AFTER DISCHARGING THE HIGH
AGE CAPAGE CAP
AGE CAP
AGE CAPAGE CAP
ACITOR AND REMOVING THE LEADACITOR AND REMOVING THE LEAD
ACITOR AND REMOVING THE LEAD
ACITOR AND REMOVING THE LEADACITOR AND REMOVING THE LEAD
TT
AGE TRANSFORMER. (SEE FIGURE 3)AGE TRANSFORMER. (SEE FIGURE 3)
T
AGE TRANSFORMER. (SEE FIGURE 3)
TT
AGE TRANSFORMER. (SEE FIGURE 3)AGE TRANSFORMER. (SEE FIGURE 3)
, PROCEED WITH THE
Y CORD FROMY CORD FROM
Y CORD FROM
Y CORD FROMY CORD FROM
, PROCEED WITH THE, PROCEED WITH THE
, PROCEED WITH THE, PROCEED WITH THE
Y WINDING OF THEY WINDING OF THE
Y WINDING OF THE
Y WINDING OF THEY WINDING OF THE
Secondary
Windings
Filament Windings
Primary
Windings
Figure 3Figure 3
Figure 3
Figure 3Figure 3
COMPONENT COMPONENT
COMPONENT
COMPONENT COMPONENT
MAGNETRON
CHECKOUT PROCEDURE CHECKOUT PROCEDURE
CHECKOUT PROCEDURE
CHECKOUT PROCEDURE CHECKOUT PROCEDURE
1) Check for resistance : Across the filamentNormal reading :
terminals of the Magnetron with an Ohm-MeterLess than 1 Ohm.
on R x 1 scale.
Ohm-MeterOhm-Meter
Ohm-Meter
Ohm-MeterOhm-Meter
Figure 4Figure 4
Figure 4
Figure 4Figure 4
2) Check for resistance : Between each filamentNormal reading :
terminal of the Magnetron and the chassisInfinite Ohms.
ground with an Ohm-Meter on highest scale.
Ohm-MeterOhm-Meter
Ohm-Meter
Ohm-MeterOhm-Meter
Figure 5Figure 5
Figure 5
Figure 5Figure 5
1) Measure the resistance :Normal readings :
With an Ohm-Meter on R x 1 scale.
a. Primary winding :Approximately 0.38 Ohms.
b. Filament winding :Less than 1 Ohm.
c.Secondary winding :Approximately 82.0 Ohms.
RESUL RESUL
RESUL
RESUL RESUL
TT
T
TT
HIGH-VOLTAGE
TRANSFORMER
2) Measure the resistance :Normal readings :
With an Ohm-Meter on highest scale.
a. Primary winding to ground.Infinite Ohms.
bFilament winding to ground.Infinite Ohms.
Ohm-MeterOhm-Meter
Ohm-Meter
Ohm-MeterOhm-Meter
Figure 6Figure 6
Figure 6
Figure 6Figure 6
Note Note
Note : Remove varnish of
Note Note
measured point.
- 4 -
COMPONENT COMPONENT
COMPONENT
COMPONENT COMPONENT
HIGH-VOLTAGE
CAPACITOR
including
BLEEDER
RESISTOR
CHECKOUT PROCEDURE CHECKOUT PROCEDURE
CHECKOUT PROCEDURE
CHECKOUT PROCEDURE CHECKOUT PROCEDURE
Measure the resistance : Across two terminalsNormal reading :
with an Ohm-Meter on highest scale.Momentarily indicates several Ohms,
Abnormal reading :
Ohm-MeterOhm-Meter
Ohm-Meter
Ohm-MeterOhm-Meter
Figure 7Figure 7
Figure 7
Figure 7Figure 7
Measure the resistance : Across two terminalsNormal reading :
with an Ohm-Meter on R x 10,000 scale.Indicates about the middle position in
Ohm-MeterOhm-Meter
Ohm-Meter
Ohm-MeterOhm-Meter
RESUL RESUL
RESUL
RESUL RESUL
and gradually returns to 10 Meg-Ohms.
Indicates continuity or 10 Meg-Ohms
from the beginning.
one direction (forward direction) and
infinite ohms in the reverse direction,
using meter which is provided with a
9 volt battery.
TT
T
TT
HIGH-VOLTAGE
DIODE
CONTROL
CIRCUIT
BOARD
Figure 8Figure 8
Figure 8
Figure 8Figure 8
Measure the voltage : Between test point TP-1,
TP-2 and Ground. (See Control Circuit Board on
page 19).
- Proceed with the check of the Step-Down
Transformer, to see if any one of the measured
values is different from the specified values.
NOTENOTE
NOTE
NOTENOTE
NOTENOTE
NOTE
NOTENOTE
- Some digital meter may show over
even in a forward direction because
low measuring voltage of meter does
not allow the meter current to pass
through the High Voltage Diode.
Abnormal reading :
Indicates continuity or infinite Ohms
in both directions.
Test
Point
Voltage
TP-1TP-2
-5V-15V
DC DC
- 5 -
COMPONENT COMPONENT
COMPONENT
COMPONENT COMPONENT
CHECKOUT PROCEDURE CHECKOUT PROCEDURE
CHECKOUT PROCEDURE
CHECKOUT PROCEDURE CHECKOUT PROCEDURE
RESUL RESUL
RESUL
RESUL RESUL
TT
T
TT
WhenWhen not
touchedtouched
Less than More than
1K Ohms 1 MEG Ohms
TOUCH KEY
Measure the resistance between terminals of
FPC connector after removing it from S101.
(Figure 9).
NOTE NOTE
NOTE
NOTE NOTE
Resistancevalue
BOARD- When reconnecting the FPC connector, makeWhen checking “START” key, connect
sure the holes on the connector are properlyOhm-Meter as illustration below.
inserted in hooks of the plastic fastener in S101.
Ohm-MeterOhm-Meter
Ohm-Meter
MAMA
065
TOTAL
1
2
7
8
DEFROST
TRIX CIRCUIT FORTRIX CIRCUIT FOR
MA
TRIX CIRCUIT FOR
MAMA
TRIX CIRCUIT FORTRIX CIRCUIT FOR
TOUCH KEY BOARD TOUCH KEY BOARD
TOUCH KEY BOARD
TOUCH KEY BOARD TOUCH KEY BOARD
FPC CONNECTORFPC CONNECTOR
FPC CONNECTOR
FPC CONNECTORFPC CONNECTOR
POWER
LEVEL
QUICK
QUICK
DEFROST
ON
SOUP /
BEVERAGE
POTATO
FROZEN
ENTREE
MORE +
LESS -
KITCHEN
TIMER
Ohm-MeterOhm-Meter
3
4
Disconnect the lead wires from the switches and check for the continuity of the switches, connecting an Ohm-Meter
to its terminals.
SWITCHES SWITCHES
SWITCHES
SWITCHES SWITCHES
(See Figure 1 on page 1) (See Figure 1 on page 1)
(See Figure 1 on page 1)
(See Figure 1 on page 1) (See Figure 1 on page 1)
Primary Interlock
Door Sensing
Interlock MonitorTerminals “COM” and “NC”
9
STOP /
CLEAR
PIZZA
SLICE
FRESH
VEGETABLE
START
CLOCKPOPCORN
CHECKOUT PROCEDURE FOR SWITCHESCHECKOUT PROCEDURE FOR SWITCHES
CHECKOUT PROCEDURE FOR SWITCHES
CHECKOUT PROCEDURE FOR SWITCHESCHECKOUT PROCEDURE FOR SWITCHES
CHECKOUT PROCEDURES CHECKOUT PROCEDURES
CHECKOUT PROCEDURES
CHECKOUT PROCEDURES CHECKOUT PROCEDURES
Terminals “COM” and “NO”
FROZEN
VEGETABLE
ADD
30 SEC
CUSTOM
PROGRAMS
TERMINALS OF FPC CONNECTORTERMINALS OF FPC CONNECTOR
TERMINALS OF FPC CONNECTOR
TERMINALS OF FPC CONNECTORTERMINALS OF FPC CONNECTOR
DOOR OPEN DOOR OPEN
DOOR OPEN
DOOR OPEN DOOR OPEN
DOOR CLOSED DOOR CLOSED
DOOR CLOSED
DOOR CLOSED DOOR CLOSED
Figure 9Figure 9
Figure 9
Figure 9Figure 9
CAUTION CAUTION
CAUTION :After checking the switches, make sure that the interlock monitor switch is properly connected according
CAUTION CAUTION
to the CIRCUIT DIAGRAM on page 3.
- 6 -
TT
TT
T
RESULRESUL
RESULRESUL
RESUL
Runs
Replace it
if necessary.
PROBLEM
1-b
Runs
Normal Control
Circuit Board.
Correct seating of
connector S101 or
replace Control
Circuit Board.
Each resistance
OK.
Runs
Normal Touch
Key Board.
Replace Touch
Key Board.
Resistance
incorrect
(ON or OFF).
Runs
Normal Control
Circuit Board.
Replace Control
Circuit Board.
SEQUENCESEQUENCE
SEQUENCE
STEPSTEP
STEPSTEP
STEPSEQUENCESEQUENCE
(1) PROBLEM
No display or mis-display in
B. TROUBLESHOOTINGB. TROUBLESHOOTING
B. TROUBLESHOOTINGB. TROUBLESHOOTING
B. TROUBLESHOOTING
the display window by
a
touching the keys.
Thermal Protector for
Magnetron, Thermal Protector
for Cavity and AC Power Cord.
If the Fuse is blown, see “E,
REMOVING FUSE” on page 13.
0 volt.Check continuity of Fuse,
Check 120VAC power supply
of connector S1 between Pin
1 and Pin 3 after removing
connector S1 from Control
Circuit Board.
120 volts.
- 7 -
Touch Key Board.
b
Check resistance of
Touch Key Board.
(See page 6)
Each voltage
OK.
Control Circuit
Board.
Voltage
Check power supply
voltage on Control Circuit
Board. (See page 5)
incorrect.
TT
TT
T
RESULRESUL
RESULRESUL
RESUL
PROBLEM
3-a
Runs
Normal
contact.
Runs
PROBLEM
Normal Door
Sensing Switch.
Adjust or replace
Door Sensing Switch.
(See page 1)
3-b
Runs
Normal Control
Replace Control
Circuit Board.
Circuit Board.
SEQUENCESEQUENCE
SEQUENCE
Contact
OK.
Check contact
of Connector
Continuity
OK.
Thermal Protector for Cavity.
2. Primary Interlock Switch.
3 . Thermal Protector for Magnetron and
Correct
Poor
S102.
seating.
contact.
No
continuity.
AC 120 Volts.
0 Volts.
STEPSTEP
STEPSTEP
STEPSEQUENCESEQUENCE
(2) PROBLEM
Cooking operation will not
CHECK : 1 . Power Supply to Oven.
a
start.
Door Sensing Switch.
Check continuity of connector
S102 between Pins (O and BL)
with the door closed after
removing connector S102 from
- 8 -
Control Circuit Board.
(3) PROBLEM
Oven does not heat up.
a
Touch “START” key and
measure voltage between
white wire lead and black wire
lead for primary winding of
H.V. transformer after removing
the wire leads from terminals of
H.V. Transformer.
TT
TT
T
RESULRESUL
RESULRESUL
RESUL
Runs
Normal
Magnetron.
Replace
Magnetron.
Runs
Normal H.V.
Capacitor or
H.V. Diode.
Capacitor or
Replace H.V.
H.V. Diode.
Runs
Normal H.V.
Transformer.
PROBLEM
3-b
Runs
Normal thermal protector for
Magnetron or Thermal
Protector for Cavity.
Runs
Normal Blower
Motor.
Replace
Blower
Motor.
Resistance
incorrect.
Runs
Normal Control
Circuit Board.
Replace Control
Circuit Board.
Normal
resistance
SEQUENCESEQUENCE
SEQUENCE
Normal H.V.
Capacitor or
H.V. Diode.
Defective
or H.V. Diode.
H.V. Capacitor
Replace H.V.
Check H.V. Capacitor or H.V.
Diode with an Ohm-Meter after
removing lead wires.
(See page 5)
OK.
Resistance
Resistance
Transformer.
incorrect.
Normal
operation.
Replace Thermal Protector
for Magnetron or Thermal
Protector for Cavity.
Normal
operation.
of Blower Motor
Measure resistance
No
operation.
Temporarily short
across terminals of
Thermal Protector for
Magnetron or Thermal
Protector for Cavity
by using jumper lead.
No
operation.
Approx. 63 ohms.
STEPSTEP
STEPSTEP
STEPSEQUENCESEQUENCE
b
H.V. Transformer.
Check resistance of each
winding after removing
lead wires. (See page 4)
Check operation of Blower
Motor when “START” key
(4) PROBLEM
Low microwave power
output.
a
is touched.
- 9 -
TT
TT
T
RESULRESUL
RESULRESUL
RESUL
No problem.
Measure water
temperature
rise exactly.
Runs
Normal Control
Circuit Board
Replace Control
Circuit Board
Runs
Normal Control
Replace Control
Circuit Board
Circuit Board
SEQUENCESEQUENCE
SEQUENCE
STEPSTEP
STEPSTEP
STEPSEQUENCESEQUENCE
Voltage cycles
on and off.
H.V. Transformer.
120V
Touch “START” key and
measure voltage between
white wire lead and black wire
lead for primary winding of
H.V. Transformer after
removing the wire leads from
Set power level
at “50”.
terminals of H.V. Transformer.
Check normal operation
of Control Circuit Board.
(5) PROBLEM
The Magnetron operates on
high level when a lower
a
cook power is selected.
- 10 -
(6) PROBLEM
No buzzing by touching keys
a
and at the end of cooking.
7.7.
DISASSEMBLDISASSEMBL
7.
DISASSEMBL
7.7.
DISASSEMBLDISASSEMBL
••
OVEN MUST BE DISCONNECTED FROM ELECTRICALOVEN MUST BE DISCONNECTED FROM ELECTRICAL
•
OVEN MUST BE DISCONNECTED FROM ELECTRICAL
••
OVEN MUST BE DISCONNECTED FROM ELECTRICALOVEN MUST BE DISCONNECTED FROM ELECTRICAL
OUTLET WHEN MAKING REPLACEMENTS, REPOUTLET WHEN MAKING REPLACEMENTS, REP
OUTLET WHEN MAKING REPLACEMENTS, REP
OUTLET WHEN MAKING REPLACEMENTS, REPOUTLET WHEN MAKING REPLACEMENTS, REP
ADJUSTMENTS AND ADJUSTMENTS AND
ADJUSTMENTS AND
ADJUSTMENTS AND ADJUSTMENTS AND
PROCEEDING WITH ANY REPPROCEEDING WITH ANY REP
PROCEEDING WITH ANY REP
PROCEEDING WITH ANY REPPROCEEDING WITH ANY REP
DISCONNECTING, WDISCONNECTING, W
DISCONNECTING, W
DISCONNECTING, WDISCONNECTING, W
THE CAPTHE CAP
THE CAP
THE CAPTHE CAP
HAS FULLHAS FULL
HAS FULL
HAS FULLHAS FULL
••
WHEN REPLACING ANY DOOR MICROSWITCH,WHEN REPLACING ANY DOOR MICROSWITCH,
•
WHEN REPLACING ANY DOOR MICROSWITCH,
••
WHEN REPLACING ANY DOOR MICROSWITCH,WHEN REPLACING ANY DOOR MICROSWITCH,
REPLACE WITH THE SAME TYPE SWITCH SPECIFIEDREPLACE WITH THE SAME TYPE SWITCH SPECIFIED
REPLACE WITH THE SAME TYPE SWITCH SPECIFIED
REPLACE WITH THE SAME TYPE SWITCH SPECIFIEDREPLACE WITH THE SAME TYPE SWITCH SPECIFIED
ON THE PON THE P
ON THE P
ON THE PON THE P
A.A.
REMOVING DOOR SENSING SWITCHREMOVING DOOR SENSING SWITCH
A.
REMOVING DOOR SENSING SWITCH
A.A.
REMOVING DOOR SENSING SWITCHREMOVING DOOR SENSING SWITCH
ACITOR IN THE HIGH-VOLACITOR IN THE HIGH-VOL
ACITOR IN THE HIGH-VOL
ACITOR IN THE HIGH-VOLACITOR IN THE HIGH-VOL
Y DISCHARGED.Y DISCHARGED.
Y DISCHARGED.
Y DISCHARGED.Y DISCHARGED.
ARAR
AR
ARAR
Y INSTRUCTIONSY INSTRUCTIONS
Y INSTRUCTIONS
Y INSTRUCTIONSY INSTRUCTIONS
CONTINUITY CHECKS BEFORECONTINUITY CHECKS BEFORE
CONTINUITY CHECKS BEFORE
CONTINUITY CHECKS BEFORECONTINUITY CHECKS BEFORE
AIR WORK. AFTERAIR WORK. AFTER
AIR WORK. AFTER
AIR WORK. AFTERAIR WORK. AFTER
AIT AAIT A
T LEAST 1 MINUTE, UNTILT LEAST 1 MINUTE, UNTIL
AIT A
T LEAST 1 MINUTE, UNTIL
AIT AAIT A
T LEAST 1 MINUTE, UNTILT LEAST 1 MINUTE, UNTIL
TS LISTTS LIST
..
TS LIST
.
TS LISTTS LIST
..
TT
AGE AREAAGE AREA
T
AGE AREA
TT
AGE AREAAGE AREA
AIRS,AIRS,
AIRS,
AIRS,AIRS,
(Figures 1 on page 1, 10-a)
(1) Disconnect all wire leads from the door sensing
switch. (Figure 1 on page 1)
(2) Remove 2 screws securing the lever stopper.
(3) Push the door sensing switch upward while pressing
adjacent upper switch stopper of lever stopper and
lower switch stopper down (Figure 10-a).
(4) Then remove it by lightly pressing the switch toward
(1) Disconnect all wire leads from primary interlock and
Interlock monitor switch.
(2) Pull the primary interlock switch or Interlock monitor
switch upward at the same time while pushing the
switch stoppers to allow a space between the
switch body and catches.
(3) Make necessary adjustments or replacement of
switch by the reversing step (2) and check microwave energy leakage according to “1. ADJUSTMENT
PROCEDURES” on page 1, after it is replaced with
new one, and check proper operation of it according
to “CHECKOUT PROCEDURE FOR SWITCHES” on
page 6.
Switch Stoppers
(5) Make necessary adjustments or replacements of the
door sensing switch by reversing step (3) and check
microwave energy leakage according to “1.
ADJUSTMENT PROCEDURES” on page 1, after it is
replaced with new one, and check proper operation
of it according to “CHECKOUT PROCEDURE FOR
SWITCHES” on page 6.
Upper
Switch
Stopper
Door
Sensing
Switch
Lower
Switch
Stopper
Figure 10-aFigure 10-a
Figure 10-a
Figure 10-aFigure 10-a
Interlock
Monitor
Switch
Primary
Interlock
Switch
Switch Stoppers
Figure 10-bFigure 10-b
Figure 10-b
Figure 10-bFigure 10-b
Interlock Switch Replacement - when replacing
faulty switches, be sure switch mounting tabs are
not bent, broken or otherwise deficient in their
ability to secure the switches in place.
- 11 -
C.C.
REMOVING BLOWER MOTORREMOVING BLOWER MOTOR
C.
REMOVING BLOWER MOTOR
C.C.
REMOVING BLOWER MOTORREMOVING BLOWER MOTOR
(Figures 11 (Top View) and 12).
(1) Remove screw securing the stay.
(2) Disconnect all lead wires from the blower motor
and H.V. capacitor.
(3) Remove 3 screws securing the blower base and
disengage 3 hooks from the rear plate of cavity
(Figure 12).
(4) Remove 1 screw securing the blower motor with
the blower base.
D.D.
REMOVING MAGNETRONREMOVING MAGNETRON
D.
REMOVING MAGNETRON
D.D.
REMOVING MAGNETRONREMOVING MAGNETRON
(Figure 11 (Top View))
After removing the blower motor :
(1) Remove duct (mag. exhaust). Remove the screw
securing the stay.
(2) Remove 1 screw securing the thermal protector.
(3) Disconnect 2 lead wires from the magnetron.
(4) Remove 4 screws securing the magnetron to the
waveguide.
Duct (mag. exhaust)
Thermal Protector
Stay
FigurFigur
Figur
FigurFigur
Screw
Duct (mag. intake)
e 11 (Te 11 (T
e 11 (T
e 11 (Te 11 (T
op Vop V
op V
op Vop V
iew)iew)
iew)
iew)iew)
Blower Base
(5) Take out Magnetron VERY CAREFULLY.
1. When removing the magnetron from the cavity or
wave guide, use a proper care so that the dome of
the magnetron does not hit any adjacent parts of
microwave oven.
2. Make sure that the contact face of the magnetron
gasket is free from any damage or deformation.
3. Adjust the position of the magnetron properly, so
that it correctly sits in place and the magnetron
gasket is in contact with the mounting rim evenly.
4. While holding the magnetron under this condition,
tighten mounting screws or nuts with your fingers
temporarily.
5. Further tighten the screws or nuts with a box
wrench, giving one or two turns to each of the
screws (or nuts) alternatively so that the
magnetron is mounted on to the bracket uniformly.
6. After replacing the magnetron, be sure to check the
microwave energy leakage with a leakage detector
and confirm the leakage is below 4mW/cm
NOTENOTE
NOTE
NOTENOTE
2
.
Blower Motor
Screw
Figure 12Figure 12
Figure 12
Figure 12Figure 12
Hook
- 12 -
E.E.
REMOVING FUSEREMOVING FUSE
E.
REMOVING FUSE
E.E.
REMOVING FUSEREMOVING FUSE
Remove the 20A fuse with a screwdriver.
--When replacing the 20A fuse, be sure to use an exact
repair part.
--If the 20A fuse blows immediately, check the primary
interlock switch, the relay 2 (on the control circuit board)
and the interlock monitor switch according to,
“CHECKOUT PROCEDURE“CHECKOUT PROCEDURE
“CHECKOUT PROCEDURE
“CHECKOUT PROCEDURE“CHECKOUT PROCEDURE
And make sure to check the microwave energy leakage
according to,
1, when the primary interlock switch, the relay 2 or the
interlock monitor switch is adjusted or replaced.
--If the primary interlock switch, the relay 2 and the interlock
monitor switch operate properly, determine which of the
following is defective : control circuit board, blower motor,
high voltage transformer, high voltage capacitor, high voltage
diode or magnetron.
FF
..
REMOVING CONTROL CIRCUIT BOARDREMOVING CONTROL CIRCUIT BOARD
F
.
REMOVING CONTROL CIRCUIT BOARD
FF
..
REMOVING CONTROL CIRCUIT BOARDREMOVING CONTROL CIRCUIT BOARD
NOTENOTE
NOTE
NOTENOTE
FOR SWITCHES”FOR SWITCHES”
FOR SWITCHES” on page 6.
FOR SWITCHES”FOR SWITCHES”
“1. ADJUSTMENT“1. ADJUSTMENT
“1. ADJUSTMENT
“1. ADJUSTMENT“1. ADJUSTMENT
PROCEDURES”PROCEDURES”
PROCEDURES” on page
PROCEDURES”PROCEDURES”
(See exploded view on page 18 and Figure 13)
(1) Remove the connector S1, S102 and the connector for
the relay 2 from the control circuit board.
(2) Remove 1 screw securing the control panel complete to
the oven cavity.
(3) Remove the FPC connector from the connector S101
while grasping up both lever ends of the plastic fastener.
Control Sheet
Control Frame
Control Base
Figure 13Figure 13
Figure 13
Figure 13Figure 13
Control Frame
(4) Remove 2 screws securing the control circuit board to
the control base.
(5) Lift up the control circuit board from its lower side and
take it out from the control base.
G.G.
REMOVING TOUCH KEY BOARDREMOVING TOUCH KEY BOARD
G.
REMOVING TOUCH KEY BOARD
G.G.
REMOVING TOUCH KEY BOARDREMOVING TOUCH KEY BOARD
(Figure 14)
After removing the control circuit board :
(1) Remove control base while lifting it up at the right side
and take it out from the control frame.
(2) Remove control sheet which is glued to touch key board.
(3) Remove the touch key board which is held on the control
base bracket with the adhesive tape.
Touch Key
Board
Figure 14Figure 14
Figure 14
Figure 14Figure 14
- 13 -
H.H.
REMOVING DOORREMOVING DOOR
H.
REMOVING DOOR
H.H.
REMOVING DOORREMOVING DOOR
(1) Remove 2 hex nuts securing the upper hinge.
(2) Tilt the top of the door toward you.
(3) Lift up the door to remove it.
J.J.
REMOVING CAREMOVING CA
J.
REMOVING CA
J.J.
REMOVING CAREMOVING CA
VITY COVERVITY COVER
VITY COVER
VITY COVERVITY COVER
(Figure 16)
(1) Remove a screw from the cavity compartment.
- After replacing the door, be sure to check that the
primary interlock switch, the door sensing switch
and the interlock monitor switch operate normally.
(See page 1).
- After replacing the door, check for microwave energy
leakage with a leakage detector. Microwave energy
leakage must be below the limit of 4mW/cm
allow for measurement uncertainty.
I.I.
DISASSEMBLING DOORDISASSEMBLING DOOR
I.
DISASSEMBLING DOOR
I.I.
DISASSEMBLING DOORDISASSEMBLING DOOR
NOTENOTE
NOTE
NOTENOTE
(Figure 15)
(1) Insert a thin flat-blade screwdriver between the
choke dielectric and the door main frame and lift
up the choke dielectric to release hooks one by
one. (Figure 15).
(2) Remove the choke dielectric.
(3) To detach the door cover, insert a thin flat-blade
screwdriver between door cover and door panel
and release the projections inside door cover.
(4) To detach the door panel, insert a thin flat-blade
screwdriver between the door cover and door
main frame.
Hooks
Cavity Cover
2
to
Screw
Figure 16Figure 16
Figure 16
Figure 16Figure 16
K.K.
REMOVING TURNTREMOVING TURNT
K.
REMOVING TURNT
K.K.
REMOVING TURNTREMOVING TURNT
ABLE MOTOR COVERABLE MOTOR COVER
ABLE MOTOR COVER
ABLE MOTOR COVERABLE MOTOR COVER
(Figure 17)
(1) Turn over the oven on its back.
(2) Cut the 6-joints of the bottom plate using diagonal
pliers (nipper).
(3) Separate the motor cover from bottom plate.
(4) Remove 2 screws securing turntable motor to the
cavity and take it out.
Hooks
Door
Main
Frame
Choke
Dielectric
- The choke dielectric may be damaged when
it is removed. When reinstalling it replace it
with new ones if it is damaged.
- After installing the door in place, check for
microwave energy leakage with a detector.
Microwave energy leakage must be below the
limit of 4mW/cm2 to allow for measurement
uncertainty.
NOTENOTE
NOTE
NOTENOTE
Hooks
Figure 15Figure 15
Figure 15
Figure 15Figure 15
Hooks
- 14 -
NoteNote
Note : Bent the cut joints inside slightly for safety
NoteNote
and be careful of sharp edge.
6 Joints
Figure 17Figure 17
Figure 17
Figure 17Figure 17
Motor Cover
Bottom Plate
8.8.
EXPLODED VIEW AND PEXPLODED VIEW AND P
8.
EXPLODED VIEW AND P
8.8.
EXPLODED VIEW AND PEXPLODED VIEW AND P
CAVITY PARTS
ARAR
AR
ARAR
TS LISTTS LIST
TS LIST
TS LISTTS LIST
* * * ALL SERVICE ON MICROWAVE OVENS SHOULD BE PERFORMED BY A QUALIFIED TECHNICIAN USING
APPROVED TESTING EQUIPMENT. CUSTOMERS SHOULD NOT ATTEMPT TO REPLACE PARTS IDENTIFIED
BY A TRIPLE ASTERISK (***).
Key
No.
Part No.DescriptionQ’ty
1617 228 8321Cabinet1
2411 082 5201SCR TPG TRS 4 x 10 Z18
3411 160 6007SCR TPG TRS + SRT 4 x 10 Z12
4617 171 6863Power Cord1
5411 160 6106SCR S-T TRS + SRT 4 x 10 Z11
6617 228 8345Oven Cavity*** (Not Service Part)1
7617 221 3590Gear Motor1
8411 010 5808SCR EVR PAN 4 x 10 Z12
9617 138 5601Bottom Plate*** (Not Service Part) 1
10617 221 6089Plastic Foot with Canoe Clip4
11411 011 0802Bolt Hex + SW + W5 x 14 Z1*** 2
12617 180 6137Hinge, Lower***1
13411 089 2500Washer F 5 x 10 x 0.81