Rotronic HygroLog HL20 Operating Manual

Page 1
IN-E-HL20-V1_10
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
HygroLog HL20 and HL21 Data Logger
User Guide
Instruction Manual
Document Type
1 of 17
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 2
IN-E-HL20-V1_10
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
Instruction Manual
Document Type
2 of 17
Table of contents:
Overview .................................................................................................................................. 3
1.
General ................................................................................................................................ 3
1.1
General description ................................................................................................................ 4
2.
Models ................................................................................................................................. 4
2.1
Required accessories .......................................................................................................... 4
2.2
Batteries ............................................................................................................................... 4
2.3
Real time clock .................................................................................................................... 5
2.4
Memory ................................................................................................................................ 5
2.5
Function key ........................................................................................................................ 5
2.6
Status indicator (LED) .......................................................................................................... 5
2.7
Service and data connector ................................................................................................. 6
2.8
Optional display ................................................................................................................... 6
2.9
Operating limits .................................................................................................................... 6
2.10
Configurable settings ........................................................................................................... 6
2.11
Factory default settings ....................................................................................................... 8
2.12
Data logging ......................................................................................................................... 8
2.13
Operation ................................................................................................................................. 9
3.
Front panel with optional display ......................................................................................... 9
3.1
3.2 Initial start-up ..................................................................................................................... 10
Recording and displaying data (models without display) .................................................. 12
3.3
Displaying and recording data (models with display) ........................................................ 12
3.4
Practical advice for measuring temperature and humidity ................................................ 13
3.5
Maintenance .......................................................................................................................... 13
4.
Periodic calibration check .................................................................................................. 13
4.1
Firmware updates .............................................................................................................. 14
4.2
Technical data ...................................................................................................................... 14
5.
Specifications ..................................................................................................................... 14
5.1
Dew point accuracy ........................................................................................................... 16
5.2
Accessories .......................................................................................................................... 16
6.
Supporting documents ........................................................................................................ 17
7.
Document releases .............................................................................................................. 17
8.
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 3
IN-E-HL20-V1_10
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
Instruction Manual
Document Type
3 of 17
Applicability:
This manual is valid for instruments with firmware version 1.0x (check using the ROTRONIC HW4 software, Device Manager). Examples: 1.0a, 1.0b. Changes in the last character of the version number reflect minor changes in the internal software of the instrument that do not affect the manner in which the instrument should be operated.

1. Overview

1.1 General

The HygroLog HL20 data logger can record up to 20,000 pairs of relative humidity and temperature values within the range of 0 to 100 %RH and -10 to 60°C (14 to 140°F).
The HygroLog HL21 data logger can record up to 20,000 temperature values within the range of
-10 to 60°C (14 to 140°F ). Mode ls with the optional LC display are limited 0 to 50°C (32 to 122°F). Temperature is measured with a Pt100 RTD. Relative humidity is measured with a ROTRONIC
Hygromer IN1 capacitive sensor. Both sensors are internal and there is no apparent probe. Both the HL20 and HL21 require three AA batteries (1.5V) for power. Depending on both the log
interval and the display option, these data loggers can record data over a time period of up to 20 months without replacing the batteries.
Configuration of the HL20 and HL21, as well a s the downloading and viewing of recorded data, requires a PC with the ROTRONIC HW4 software installed as well as a service cable such as the AC3006
Based on the ROTRONIC AirChip 3000 digital technology the HL20 and HL21 offer the following features:
Unit system: metric or English
Real time clock (runs for up to 4 hours without the 3 AA batteries after initial charge)
Calculation of the dew or frost point (calc ula ted par ameter can be displayed but not recorde d)
Optional LC display (%RH, temperature) with battery charge indicator and message bar
Function key: starts and stops data logging, turns optional LC display on and off
Status indicator (LED): flashes green when logging, flashes red for out-of-limit values or a
logger problem
Humidity temperature calibration and adjustment
Simulator mode (used to validate the data recording function)
The ability for the user to easily update the AirChip 3000 firmware means that the HP21 can be kept up-to-date regarding any future functionality improvement.
Note: Instructions for using the HW4 software are not included in this manual. These instructions are shipped separately on the software CD ROM.
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 4
IN-E-HL20-V1_10
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
Instruction Manual
Document Type
4 of 17

2. General description

2.1 Models

Basic models without display
o HygroLog HL-20: relative humidity and temperature logger, function key, status indicator
(LED), internal sensors protected by a slotted cover.
o HygroLog HL-21: temperature data logger, function key, status indicator (LED), internal
Pt100 RTD
Models with display
o HygroLog HL-20D: same as HL20, LC display with backlight, message bar and battery
charge indicator
o HygroLog HL-21D: same as HL21, LC display with backlight, message bar and battery
charge indicator

2.2 Required accessories

The following accessories are required in order to be able to configure the logger and to download and view the recorded data:
- Cable AC3006 or equivalent (plugs into a USB port of the PC)
- PC with the ROTRONIC HW4 software installed (version 2.4.0 or higher)

2.3 Batteries

All models ship wit h thr ee AA batt eries which must be inserted first (see battery replacement).
Note
: To conserve battery power, models w ith the LC displ a y are shipp ed with the Disp lay Sleep function enabled and set to 1 minute (see Software Functions). When the function key has not been pressed for some time (in this case 1 minute), the display sleep function automatically turns the display off. The display can be temporarily turned on at any time by pressing on the function key.

2.3.1 Battery lifetime

The battery lifetime depends on the manner in which the HL20 or HL21 is being used. The following examples are based on the following assumptions:
Fresh set of AA batteries (2900 mAh)
Display backlight turned off (if applicable)
Log interval set to 30 minutes
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 5
IN-E-HL20-V1_10
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
Models w/o LC display 20
Models with LC displ ay, display sleep function set to 1 minute 13
Models with LC display, display sleep function disabled 5
Battery lifetime in months
Instruction Manual
Document Type
5 of 17

2.4 Real time clock

After inserting a fresh set of batteries for the first time, the real time clock must be adjusted prior to using the HL20 or HL21 (s ee “Operation”). The real time clock is powered from a large capacitor and can run for up to 4 hours without the batteries. R emoving the battery for an extended period of time will cause the logger to completely lose track of the date and time.
Because of the requirement to always keep track of the date and time, the HL20 and HL21 do not have a power on/off switch.

2.5 Memory

All models use a non-volatile EEPROM internal memory to hold the measurement data. The memory cannot be removed from the data logger.
Memory capacity is as follows: HL20: 20,000 pairs of relative humidity and temperature values
HL21: 20,000 temperature values

2.6 Function key

The function key is used as follows:
Long key press: press the key for a few seconds to start or stop data logging.
Short key press: depending on the configuration settings of the display (display sleep
function) a short key press turns on and off the optional LC display.
For more details, see “Operation”

2.7 Status indicator (LED)

The status indicator consists of a green and red LED. The green side of the LED flashes every 5 seconds when logging is active. The red side of the LED flashes every 5 seconds when a measured value is out of limits (see “Configurable settings”) or when the logger requires attention (low battery, memory full, major sensor failure – such as open or shorted sensor)
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 6
IN-E-HL20-V1_10
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
Instruction Manual
Document Type
6 of 17

2.8 Service and data connector

The service / data connector (mini-USB type) can be accessed after removing the back cover (see Operation). The connector gives access to the data logger UART digital interface (Universal Asynchronous Receiver Transmitter).
Configuring the data logger or retrieving the recorded data requires cable AC3006 or equivalent. One end of the cable plugs into the service / data connector. The other end plugs into the USB of a PC with the ROTRONIC HW4 software installed. See “Initial start-up” for the location of the service / data connector.

2.9 Optional display

The optional LC display has a backlight which can be enabled or disabled when configuring the data logger.
o HL20D: depending on the configuration of the logger, the LC display shows relative humidity
and temperature or the dew / frost point and temperature. The top line displays humidity
(%RH or the dew / frost point) o HL21D: the LC display shows temperature. The LC display can be configured to show the measured or calculated values with either one or
two decimals. Note
: the LC display is automatically turned on when the data logger is connected to the USB
port of a PC by means of cable AC3006 or equivalent.

2.10 Operating limits

Models without display:
-10…60°C with the factory supplied alkaline batteries
0…100 %RH, non condensing
Models with display:
0…50°C with the factory supplied alkaline batteries
0…100 %RH, non condensing

2.11 Configurable se ttings

Configuration of the data logger by the user and access to its functions requires a PC with the ROTRONIC HW4 software (version 2.4.0 or higher) installed. Service cable AC3006 is used to connect the service / dat a connector to a USB port of the PC.
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 7
IN-E-HL20-V1_10
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
Instructions regarding the configuration of the data logger and access to its functions are provided in the following manuals:
E-M-HW4v2-F2-018 E-M-HW4v2-Main (§ 6.5) E-M-HW4v2-A2-001
Instruction Manual
Document Type
7 of 17

2.11.1 Relative humidity symbol and temperature unit

Both the symbol representing relative humidity and the temperature can be configured to suit local requirements.

2.11.2 Calculated parameter

HL20: the data logger can be set to compute one of the following psychrometric parameters.
o Dew point (Dp)
o Frost point (Fp)
Notes:
When the optional display is configured to show the c a lc ulated par ameter, either the symbol Dp or the symbol Fp appear to the left of the numerical value, depending on whether the dew point or frost point was selected. For values below freezing, the symbol Fp indicates that the value is a frost point (above ice) as opposed to being a dew point (above liquid).
The data logger cannot be used to record either the dew point or the frost point

2.11.3 Alarm function (out-of-limit value)

The alarm function can be enabled or disabled for each of the measured or calculated parameters. High and low value can be associated with each parameter so as to define out-of­limits values. A hysteresis value can also be entered for each parameter. The hysteresis value is used for both the low and the high alarm conditions.
The data logger has a red LED whic h flashes when the value of any parameter is out-of-limits. The LED stops flashing when the value returns to normal..
Log files do not record the existence of an alarm condition. However, when the contents of a log file are viewed as a data table with the HW4 software, values that correspond to an alarm condition are printed over a red background, based on the alarm settings that HW4 downloaded from the logger.

2.11.4 Display-Sleep function

Using the Display Sleep function can substantially extend the lifetime of the battery. When the Display Sleep function is active, the LC display of the data logger s tays blank.
The optional display of the data logger can be set to be always on. In this case, the display can be turned on and off manually with the function key. The display can also be set to automatically go blank after 1, 5, 10 or 20 minutes.
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 8
IN-E-HL20-V1_10
Configurable Settings
Factory default
Unit system
Metric, except USA: English
Psychrometric calculation
No calculation
Display resolution
1 decimal
Display Sleep function
Enabled, display automatically turns off after 1 minute
Display backlight
Off
Displayed parameters
%RH and temperature or temperature only
Communication protocol
RO-ASCII
Device name
Instrument model
Functions
Factory default
Humidity / temperature adjustment
Enabled
Device write protection
Disabled
Limit humidity to 100 %RH
Enabled
Out-of-limit value alarm
Disabled
Data recording
Stopped (loop mode – 10 min. interval)
Major sensor failure alarm
Enabled (cannot be disabled)
Simulator mode
Disabled
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
The data logger suspends the Display Sleep function whenever the function key is pressed. The function is also suspended when the data logger is connected to the USB port of a PC by means of cable AC3006.
Instruction Manual
Document Type
8 of 17

2.12 Factory default settings

2.13 Data logging

Function overview:
The main characteristics of the data logging function are as follows:
- Relative humidity values are recorded with a resolution of 0.1%RH
- Temperature values are recorded with a resolution of 0.05˚C / 0.09˚F
- Memory capacity: 20,000 single values or value pairs
- Adjustable log interval: from 5 seconds to 1 hour, in increments of 5 seconds
- Two data recording modes: Loop and Start-Stop
Note
: Any recorded data present in the data logger memory is automatically erased whenever a
new recording session is started.
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 9
IN-E-HL20-V1_10
Status indicator
Function key
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
Data recording modes: Loop : when the memory is full the logger dumps the oldest data sample and keeps
recording. The memory retains the most recent data. Start-Stop : the logger stops recording data as soon as the memory is full Data recording can be sto p ped at an y tim e by pressing on the function key (long key press) or
with the ROTRONIC HW4 software.
Recorded parameters:
Depending on the model, the data logger records either temperature only or relative humidity and temperature. Calculated values such as dew point cannot be recorded.
Date and time:
The data logger has an internal real time clock that is used to associate a date and time stamp with each data record.
When data recording is either started or stopped using the ROTRONIC HW4 software, the real time clock is automatically synchronized with the PC clock.
Log file format (data download to a PC with HW4) :
The recorded data can be saved as a file to a PC using either HW4 or another communications software. When using HW4, the file can be either formatted as a text file (extension TXT) or as an encrypted binary file (extension LOG). The LOG file format can only be read with HW4 and complies with the FDA / GAMP rules regarding electronic records and electronic signatures.
Instruction Manual
Document Type
9 of 17

3. Operation

3.1 Front panel with optional dis play

© 2010; Rotronic AG IN-E-HL20-V1_10
(LED)
Page 10
IN-E-HL20-V1_10
3-segment battery Message bar
Cover screw
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
The LC display has a backlight which can enabled or disabled as part of the data logger configuration. When the backlight is enabled, it turns on for a few seconds after pressing on the function key.
The upper line corresponds to relative humidity or dew / frost point and the bottom line corresponds to temperature. In the event of an alarm (out-of-limit value) the display shows the symbol [ ! ] to the right of the value and the LED indicator flashes red.
In addition to the measured values, the display also provides information such as battery charge and data logger status.
For configuration instructions see the following HW4 manual: E-M-HW4v2-F2-018.
charge indicator
Instruction Manual
Document Type
10 of 17

3.2 Initial start-up

3.2.1 Battery compartment

The battery compartment is located on the back panel and can be opened after loosening the screw that holds the cover in place
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 11
IN-E-HL20-V1_10
Service and data connector (mini USB type)
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
Insert three fresh AA alkaline batteries in the battery compartment. Be sure to observe the polarity.
Instruction Manual
Document Type
11 of 17

3.2.2 Initial configuration (all models)

Required Hardware and software
PC with the ROTRONIC HW4 software installed (version 2.4.0 or higher)
Cable AC3006 or equivalent (connects the data logger service / data connector to a PC
USB port)
IMPORTANT After installing the HW4 software
of the PC. The PC will detect the cable and ask for the appropriate driver. The ROTRONIC USB driver (ftdibus.inf) is located in a folder named USB_Driver which is in the HW4 installation folder. For detailed instructions see documents E-M-HW4v2-Main (§ 6.3: USB port) and E-M-HW4v2- F2-018
Setting the Real Time Clock (required configurati on step)
The real time clock of the data logger should be set to agree with the date and time of the HW4 PC.
Use cable AC3006 to connect the data logger to a USB port of the HW4 PC. Start HW4 and discover the data logger by selecting with the mouse the following command in the menu bar: Devices and Groups > Search for Master Devices > USB Masters.
When the Data logger icon appears, left click with the mouse on the (–) sign located to the left of the icon. This displays the functional modules available for the data logger. Click on the Device Manager module. For additional instructions, see document E-M-HW4v2-F2-018.
:
and prior to running HW4
, connect cable AC3006 to a USB port
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 12
IN-E-HL20-V1_10
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
Instruction Manual
Document Type
12 of 17
Other initial configuration steps
Other initial configuration steps include the following:
- Temperature measurement unit
- Log interval
- Recording mode (start-stop or loop)
For additional instructions, see document E-M-HW4v2-F2-018.

3.2.3 Optional display configuration

To conserve battery power, models with a display are shipped with the Display Sleep function enabled (set to 1 minute) and with the display backlight disabled. When configured in this manner the data logger can record data for extended periods of time
For additional instructions regarding the configuration of the display, see document E-M-HW4v2-F2-018.

3.3 Recording and displayi ng da t a (m odel s wi t hout display)

To start or stop recording data, press on the function key for a few seconds. When the data logger is recording data, the status indicator flashes green every 5 seconds
The ROTRONIC HW4 software is required to view the measured or calculated data on a PC monitor. Connect the data logger to the HW4 PC as explained under “Initial Start-up”. HW4 is also required to download and save the recorded data to disk. For additional instructions, see document E-M-HW4v2-F2-018.

3.4 Displaying and recording data (models with display)

Displaying data
When the display sleep function is enabled, the display goes blank after some time in order to conserve battery power.
To view the measured or calculated data, press briefly on the function key. When the display sleep function is not enabled, the display can be manually turned off by pressing briefly on the function ke y.
Start or stop recording data
To start or stop recording data, press on this key for a few seconds. When the data logger is recording data, this is indicated by the display message bar and by the LED indicator flashes green every 5 seconds
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 13
IN-E-HL20-V1_10
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
The ROTRONIC HW4 software is required to view the measured or calculated data on a PC monitor. Connect the data logger to the HW4 PC as explained under “Initial Start-up”. HW4 is also required to download and save the recorded data to disk. For additional instructions, see document E-M-HW4v2-F2-018.
Instruction Manual
Document Type
13 of 17

3.5 Practical advice for measuring temperature and humidity

The most common source of error when measuring relative humidity is a difference between the temperature of the probe and the temperature of the environment. At a humidity condition of 50 %RH, a temperature difference of 1°C (1.8 °F) typically results in an error of 3 %RH on relative humidity.
When using the data logger, it is good practice to allow for sufficient for temperature equilibration with the environment. The larger the initial temperature difference between the probe and the environment to be measured, the more time temperature equilibration requires. This can be monitored with the optional display.
In extreme situations, condensation may occur on the sensors when the data logger is colder than the environment. As long as the humidity / temperature limits of the humidity sensor are not exceeded, condensation does not alter the calibration of the sensor. However, the sensor has to dry out before it can provide a valid measurement.
Non-moving air is an excellent insulator. When there is no air movement, surprising differences in temperature and humidity can noted over short distances. Air movement at the probe generally results in measurements that are both faster and more accurate.

4. Maintenance

4.1 Periodic calibration check

Both the Pt 100 RTD temperature sensor and associated electronics are very stable and should not require any calibration after the initial factory adjustment.
Long term stability of the ROTRONIC Hygromer humidity sensor is typically better than 1 %RH per year. For maximum accuracy, calibration of the HP21 should be verified every 6 to 12 months. Applications where the data logger is exposed to significant pollution may require more frequent verifications.
Note
: calibrating and adjusting the data logger typically require a humidity-temperature
generator or other specialized equipment that may not be available to all users.
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 14
IN-E-HL20-V1_10
General
HL20 / HL21
Device type
Humidity-temperature data logger with internal sensor(s)
Battery type
3 x AA (1.5 V) alkaline
- Models with LC display, display sleep function set to 1 minute: 13 months
Low battery indication
Yes (with HW4 software or with optional display)
Humidity measurement
HL20
Measuring range
0…100 %RH
Measurement accuracy at 23 °C
±0.8 %RH
Repeatability
0.3 %RH
Long term stability
< 1 %RH / year
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
Instruction Manual
Document Type
14 of 17
Procedure for adjusting the data l o g g er with the ROTRONIC HW4 software:
Use cable AC3006 to connect the service connector of the data logger to the USB port of a PC with the ROTRONIC HW4 software installed. Note that the ROTRONIC USB driver must be installed on the PC as explained in the HW4 manual E-M-HW4v2-Main
Start HW4 on the PC and search for the data lo gger (HW4 Main Menu Bar > Devices and Groups > Search for USB Masters).
After finding the data log ge r with HW4, expand the device tree to see the data logger functions. Select Probe Adjustment.
For further instructions see HW4 manual E-M-HW4v2-A2-001

4.2 Firmware updates

Firmware updates will be available on the ROTRONIC website for downloading. Firmware files are given a name that shows both to which device the file applies and the version number of the firmware. All firmware files have the extension HEX. Procedure for updating the firmware:
Use cable AC3006 to connect the service connector of the data logger to the USB port of a PC with the ROTRONIC HW4 software installed. Note that the ROTRONIC USB driver must be installed on the PC as explained in the HW4 manual E-M-HW4v2-Main
Copy the firmware update file from the ROTRONIC website to the PC.
Start HW4 software on the PC and search for the data logger (HW4 Main Menu Bar >
Devices and Groups > Search for USB Masters).
After finding the data log ge r , expand the device tree to see the data logger functions. Select Device Manager. In the Device Manager menu bar select Tools > Firmware Update. For instructions see document E-M-HW4v2-F2-018

5. Technical da ta

5.1 Specifications

Depends on log interval and other settings. Examples based on a 30 minutes log
Battery life time
Sensor ROTRONIC Hygromer ® IN1
Response time Typical 3 min. (depends on air flow)
© 2010; Rotronic AG IN-E-HL20-V1_10
interval:
- Models w/o LC display: 20 months
Page 15
IN-E-HL20-V1_10
Temperature measurement
HL20 / HL21
Measuring range
-10...60°C
Measurement accuracy at 23 °C
±0.2 °C
Repeatability
0.05°C
Long term stability
< 0.1°C / year
Calculated parameters
HL20
Psychro metric calculations
Dew or frost point (user configurable setting)
Start-up time and data refresh rate
HL20 / HL21
Start-up time
1.9s (typical)
Data refresh rate
1.7s (typical)
Data recording and download
HL20 / HL21
Recorded parameter(s)
HL20: relative humidity and temperature, HL21: temperature
Recording capacity
HL20: 20,000 value pairs, HL21: 20,000 values
Real time clock
Yes, adjustable via PC
Start-stop (recording stops when memory is full Loop: oldest data is dumped to make room for latest data
Log interval
Min. 5 sec, up to1 hour in 5 sec. increments
Service and data connector
HL20 / HL21
Interface type
UART (Universal Asynchronous Receiver Transmitter)
Maximum service cable length
5 m (16.4 ft)
General specifications
HL20 / HL21
5 sec, green flash when recording data 5 sec. red flash for out-of-limit value or logger problem
Housing material
ABS
Overall dimensions
130 x 72 x 32 mm (5.1 x 2.8 x 1.3”)
Weight
About 210 g (7. 4 oz)
Conformity with standards
HL20 / HL21
Solder type
Lead free (RoHS directive)
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
Sensor Pt100 RTD, IEC 751 1/3 class B
Response time Typical 10 min. (depends on air flow)
Recording media Internal EEPROM non volatile, not removable
Instruction Manual
Document Type
15 of 17
Logging modes
Interface for data download Internal connector, located under protective cover
Optional display LC, 1 or 2 decimals resolution, backlight, alarm and low battery indication LED indicator
Housing protection grade HL20: IP 40, HL21: IP65
CE / EMC imm unity
FDA / GAMP direc tives compatible
EMC Directive 2004/108/EG: EN 61000-6-1: 2007, EN 61000-6-2: 2005 EN 61000-6-3: 2007, EN 61000-6-4: 2007
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 16
IN-E-HL20-V1_10
Environment a l limits
HL20 / HL21
-10….60 °C (models w/o display), 0….50 °C (m odels with displ ay) 0…100 %RH, non condensing
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title
Instruction Manual
Document Type
16 of 17
Operating limits Critical environments Humidity sensor: as per DV04-14.0803.02 - Critical chemicals

5.2 Dew point accuracy

The HL20 can be configured to calculate either the dew point or frost point based on the measurement of relative humidity and temperature. The accuracy of this conversion varies, depending on the humidity and temperature conditions as sho wn in the grap h belo w:
Example: at a temperature of 20 ˚C, a dew point value of -37 ˚C is measured with an accuracy of ± 1.0 ˚C or better.

6. Accessories

For accessories and parts such as the HW4 configuration software, service cables, calibration accessories and spare dust filters, please see document E-M-HC2-accessories
© 2010; Rotronic AG IN-E-HL20-V1_10
Page 17
IN-E-HL20-V1_10
Document File Name
Contents
HW4 software version 2: General instructions and functions common to all devices
HW4 software version 2: HL20 / HL21 data logger Device configuration and data recording
HW4 software version 2: Probe Adjustment function AirChip 3000 devices
Temperature and humidity calibration basics Instructions for using the ROTRONIC humidity standards
Document code Unit
Rotronic AG Bassersdorf, Switzerland
HygroLog HL20 and HL21 data logger: User Guide
Document title

7. Supporting documents

E-M-HC2-accessories
E-M-HW4v2-DIR
E-M-HW4v2-Main
E-M-HW4v2-F2-018
E-M-HW4v2-A2-001
E-M-AC3000-CP
Accessories and parts for probes, indicators and transmitters
List of the HW4 manuals
AirChip 3000 Communication Protocol
Instruction Manual
Document Type
17 of 17
E-M-CalBasics
E-T-HumiDefs
Note: All document file names have an extension corresponding to the document release number. This extension is not shown in the above table.
Humidity Definitions

8. Document releases

Doc. Release Date Notes
_10 Mar. 5, 2010 Original release
© 2010; Rotronic AG IN-E-HL20-V1_10