Rotronic Hygrolab Specifications

Page 1
DATA LO GGE RS FO R HUM ID ITY A ND TE MPE RATU R E
ECONOMICAL & PRECISE.
BASED ON THE AIRCHIP3000 DIGITAL TECHNOLOGY.
IN N OVATI O N I N H U M ID I TY AN D T E M PE R ATU R E M E AS U R EM E NT
• Highest accuracy at 0.8 %rh and 0.2 ˚C
• Outstanding long term stability
• 20000 data point memory with date & time stamps
• Programmable alarm limits
• Programmable recording intervals
• Psychrometric calculations of dew
and frost point values available
• FDA 21 CFR Part 11 / GAMP 4 compliant
• Long battery life
Page 2
SPECIFY THE BEST:
HYGROLOG HL20/21 ADVANTAGES AT A GLANCE.
The new HygroLog HL20/21 series is the latest development of data loggers for relative humidity, temperature, dew point
and frost point. These loggers give you the accuracy and reliability to measure and record your essential data whether it
be in pharmaceutical clean rooms, storage facilities, manufacturing areas, buildings, museums or any other critical area.
The HL20 and HL21 are fully compliant with FDA 21 CFR part 11 and GAMP 4 requirements for electronic records and
electronic signatures (ERES).
AirChip3000
Temperature compensates humidity with over 30000 reference points
Dew/frost point calculation
Sensor auto-diagnostics and automatic correction
Active alarming and information
Conforms to FDA21 CFR Part 11 and GAMP4 (audit trail)
Optional display
Display of data on a clear, backlit graphic display
Battery charge indicator
Display of measured and calculated data
Function key to start and stop logging
Display resolution to 2 decimal places
Data logging function
HL20: Logs up to 20000 value pairs, relative humidity and temperature
HL21: Logs up to 20000 data values, temperature
Each record is date and time stamped (real-time clock)
Adjustable log interval
Two data recording modes: (start-stop or loop)
Alarm function
Visible alarm can be programmed for the HL20 or HL21 detects the following:
Out-of-limits value (defi ned with HW4 software)
Communication loss with the probe
Sensor failure and sensor drift
Page 3
APPLICATIONS.
Long term recording of humidity and temperature measurements is essential in many production, storage, transportation, display and test applications as well as in buildings and museums. They provide valuable information on the levels and variation of humidity and temperature which can have a signifi cant impact on people, materials and objects.
Relative humidity and temperature logger.
HygroLog HL20(D)
HygroLog HL20 HygroLog HL20D
Order codes: HL-20 (without display) HL-20D (with display)
Temperature logger.
HygroLog HL21(D)
HygroLog HL20(D)-Set
Set:
HL20 or HL20D
HW4 software
Cable AC3006
Order codes: HL-20-Set HL-20D-Set
HygroLog HL21(D)-Set
HygroLog HL21 HygroLog HL21D
Order codes: HL-21 (without display) HL-21D (with display)
Set:
HL21 or HL21D
HW4 software
Cable AC3006
Order codes: HL-21-Set HL-21D-Set
Page 4
TECHNICAL INFORMATION.
ROTRONIC Instruments (UK) Ltd, Crompton Fields, Crompton Way, Crawley, West Sussex, RH10 9EE, UK, Phone +44 (0)1293 571000, Fax +44 (0)1293 571008, www.rotronic.co.uk
Data Logging.
The values stored in the HygroLog can be displayed graphically using the HW4 software. You can determine the storage interval, the alarm limits, the recording mode and much more.
HygroLog HL20
HygroLog HL20D
Dimension
130 mm 32 mm
72 mm
Device type Humidity-temperature data logger with internal sensor(s)
Battery type 3 x AA (1.5 V) alkaline
Battery life Depends on log interval and other settings.
Examples based on a 30 minutes log interval:
- Models w/o LC display: 20 months
- Models with LC display, display sleep function set to 1 minute: 13 months
Low battery indication Yes (with HW4 software or with optional display)
Measuring range 0…100 %rh / -10...60 °C
Measurement accuracy at 23 °C ±0.8 %rh / ±0.2 °C
Repeatability 0.3 %rh / 0.05 °C
Long term stability <1 %rh / year / < 0.1 °C / year
Response time Humidity: typical 3 min.
Temperature: typical 10 min. (both depend on air fl ow)
Psychrometric calculations Dew or frost point (user confi gurable setting)
Start-up time 1.9s (typical)
Data refresh rate 1.7s (typical)
Recording capacity HL20: 20000 value pairs
HL21: 20000 values
Real time clock Yes, adjustable via PC
Logging modes
Start-stop (recording stops when memory is full) Loop: oldest data is dumped to make room for latest data
Log interval Min. 5 sec, up to1 hour in 5 sec. increments
Interface type UART (Universal Asynchronous Receiver Transmitter)
Optional display LC, 1 or 2 decimal resolution,
backlight, alarm and low battery indication
LED indicator 5 sec. green fl ash when recording data
5 sec. red fl ash for out-of-limit value or logger problem
Housing material ABS
Housing protection grade HL20: IP 40, HL21: IP65
Weight About 210 g (7.4 oz)
Solder type Lead free (RoHS directive)
FDA / GAMP directives Compatible
Operating limits -10….60 °C (models w/o display)
0….50 °C (models with display) 0…100 %rh, non condensing
Overall dimensions 130 x 72 x 32 mm (5.1 x 2.8 x 1.3”)
Subject to technical change without notice. Errors and omissions excepted.
Interested in further information? A complete and up-to-date overview of our HygroLog HL20/21 series is available on our web site www.rotronic-humidity.com. We would also be happy to advise you on the phone.
Change over now to the HygroClip2 probes with integrated AirChip3000 technology for your temperature and humidity measurements. You can find detailed information about our probes at www.rotronic-humidity.com. We would also be happy to advise you in person or on the phone.
ROT RON IC AG, Gr indel stra sse 6, CH - 8 303 Bassersd orf, Tel . +41 44 838 1 1 11 , Fa x + 41 4 4 836 44 24, w ww. rot roni c.c h
ROTRONIC Instrument Corp, 135 Engineers Road, Hauppauge, NY 11788, USA , Phone, +1 631 427-3898, Fax, +1 631 427-3902, www.rotronic-usa.com
59020E/2010- 06