Rotronic HygroFlex8 Specifications

humidity and temperature calibrators
Hygrog
en
2-S-XL
REFERENCE GENERATORS FOR CALIBRATION OF HUMIDITY AND TEMPERATURE INSTRUMENTS
BASED ON AIRCHIP DIGITAL TECHNOLOGY
tranSPortaBle hUmiDitY anD temPeratUre caliBration laBoratorY
•
Generates stable humidity and temperature conditions (<±0.1 %RH, <±0.01 °C)
•
Calibration solution for the laboratory and on-site
•
Humidity equilibrium typically in only 5 minutes for HG2-S, 15 minutes for HG2-XL
•
Calibrate multiple probes simultaneously
•
«AutoCal» automated calibration (optional)
•
«Remote API» logging and control with third party software (optional)*
•
Integrated FDA 21 CFR part 11 compliant Rotronic HW4 software
humidity and temperature calibrators
hYGroGen2:
the BenchmarK for hUmiDitY caliBration
Since its launch the Rotronic HygroGen has defined the standard for portable humidity and temperature calibra­tion systems. Hundreds of users worldwide have identified that this tool for the rapid generation of stable tem­perature and humidity conditions can save significant amounts of time in performing calibrations of all types of humidity instruments from all manufacturers. The HygroGen2 satisfies the stringent demands of quality and compliance regulations with its ability to calibrate instruments over their full working range, and is embraced throughout the pharmaceutical industry as the leading instrument in its class and is favored in ISO17025 humi­dity calibration laboratories across the world. Being portable and robust, it can be set up at the installation location allowing for complete system qualification. The rock solid stability, minimal thermal gradients, and rapid time for set-point changes of HygroGen2-S are now replicated in a new family member: HygroGen2-XL, which boasts a chamber with 10 times the capacity.
Hygrogen2-S
• Chamber volume: 2 litres
• Working volume 1.5 litres
• Humidity changes (5…95 %RH, 0.1 %RH
stability): <5 minutes
• Temperature changes (23…50 °C, 0.01 °C stability): < 5 minutes
Hygrogen2-XL
• Chamber volume: 20 litres
• Working volume 17 litres
• Humidity changes (5…95 %RH, 0.1 %RH
stability): <15 minutes
• Temperature changes (23…50 °C, 0.01 °C stability): < 15 minutes
Loading...
+ 4 hidden pages