Rotronic CP11 Operating Manual

OPERATION MANUAL
PORTABLE IAQ METER
CO2+RH+TEMP. IAQ METER/LOGGER
INTRODUCTION
Thank you for purchasing this portable IAQ meter. The meter measures CO level, air temp., dew point, wet bulb temp. and humidity and is an ideal instrument for indoor air quality (IAQ) diagnosis.
Poor indoor air quality is considered unhealthy because it causes tiredness, loss of ability to concentrate, and even illness(ex. Sick Building Syndrome). IAQ monitoring and survey, especially on CO level and air ventilation become widely applied in public areas such as offices, classrooms, factories, hospitals and hotels. It is also suggested in regulations of industrial hygiene in some countries. (Appendix)
The portable CO meter uses NDIR (non-
dispersive infrared) technology to ensure the reliability and long term stability. It’s useful in verifying HVAC system perform­ance and air ventilation control.
MATERIAL SUPPLIED
This package contains:
Meter 4pcs AA batteries
Operation manual Mini USB cable and software CD Soft carrying case
1
FEATURES
Big LCD display with blue backlight to use in dark area One touch to display CO2 /Dew point temperature/Wet bulb temperature/ Air temperature/Humidity in turns. Designed with NDIR (non-dispersive infrared) Waveguide technology CO2 sensor Programmable warning CO2 level Long time drift compensation on CO2 sensor Audible alarm (~80db)threshold setting Max ,Min, average functions included Mini USB output every 2 seconds to pc for analysis 99 points manually recording 30000 points automatic recording Review 99 points manual record Hold function freezes current readings Housing design surrounds with rails to help air ventilation for quick & accurate response Easy to manually calibrate at fresh air around 380-420ppm
2
POWER SUPPLY
The meter is powered by either 4 AA batteries or a DC adaptor(5V/1A output).
Install the batteries into the battery compartment on the bottom and make sure they are in correct polarity and good contact. cut off the power supply from batteries. The adaptor can’t be used as a battery charger.
When battery voltage gets low,
appear on the LCD . The CO sensor can’t work normally under low voltage, please replace with fresh b ies or connect with an adaptor.
When an adaptor is used, it will
will
(Fig.1)
atter
CO2
AM
Fig.1
3
LCD DISPLAY
AVG
DP
CAL
WBT
OO
CC
OO
CO2
HOLD
MAX
TEMP
MIN
FF
%rh%rh
AM PM
Symbols
Upper LCD Lower LCD Real time display
CO2 Carbon dioxide reading
HOLD Readings are freezed unchanged
MIN/MAX Minimum/Maximun readings
AVG Average reading
Low battery indicator
DP Dew point
TEMP Air temperature
WBT Wet bulb
%rh Unit of relative humidity
o
C/ F Celsius/Fahrenheit of temp. CAL In calibration status REC In manual/automatic logging RECALL In manual records recall mode
ON/OF F
SET
START ESC
MODE RECA LL
CO2/Temp/RH....reading
temperature
o
KEYPAD
Turns on and off the meter. Enters setup mode while meter is off. Turn on with to become non-sleep mode Exits setup/recall page. Start automatic logging.
Press to switch displayed mode. Long press to enter memory recall mode.
HOLD
4
REC Recall
temperature
HOLD
Freezes the current readings. Selects unit or increases value in setup. Cancels data hold function.
MEM
Press to manually record the reading. Selects unit or decreases value in setup.
MIN
Activates MIN,MAX,AVG function.
MAX/AV G
Saves and finishes settings.
OPERATION
POWER ON/OFF
ON/OF F
Press to turn the meter on and off.
SET
At power up, it emits a short beep and performs 30 seconds countdown(Fig.2) for meter warm up, then enters normal mode with current CO and real time displayed (Fig.3). The real time display date and time in turns.
Fig.2
Suggest to wait for 30 seconds warm up time. For quick start, you may press
MIN
MAX/AV G
for 2 seconds to end the warm up and entering normal display.
TAKING MEASUREMENT
The meter starts measurement when power on and update readings every second. In the condition of operating environment change (ex. from high to low temp.), it takes 30 sec to respond for CO sensor and 30 minutes for RH.
NOTE: Do not hold the meter close to faces in case exhalation affects CO levels.
2
CO2
AM
Fig.3
5
CO2 (Carbon Dioxide)
MODE
Press to switch the mode to CO2.
RECA LL
User can get the CO2 reading in ppm unit on main display.(Fig.4). The lower display shows the real time clock.
CO2
TEMP
OO
CC
AM
Fig.4
AIR,DP,WBT
MODE RECA LL
Press to switch to temperatures
AM
Fig.5
display(Fig.5,6,7).The lower display shows the real time clock.
Fig.7
WBT
OO
CC
DP
OO
CC
AM
Fig.6
Humidity
MODE RECA LL
Press to switch to humidity display.
AM
(Fig.8).The lower display shows the real time clock.
Fig.8
%rh%rh
6
DATA HOLD In normal display mode, press to
HOLD
freeze the readings, "HOLD” icon is flashed on the left top of LCD(Fig.9). All current readings are kept unchanged.
HOLD
Press again to cancel the hold function.
HOLD
%rh%rh
Fig.9
NOTE: THE HOLD FUNCTION IS DISABLE IN MIN/MAX/ AVG MODE.
BACKLIGHT The backlight will be activated for 10 seconds by pressing any key.
MIN,MAX,AVG,CURRENT REVIEW This meter allows you to check the minimun, maximun, average, current value from the moment you press the
MIN
MAX/AV G
key. Under normal mode, press
MIN
MAX/AV G
to see the minimum, maximum, average and current value in turns. Each press
MIN
MAX/AV G
of ,it displays MIN, MAX, AVG, current value in sequence. (Fig.10~12). To return to normal mode, press key to escape(Fig.13)
CO2
MAX
CO2
AVG
START ESC
CO2
Fig.10 Fig.11
CO2
Fig.12
AM
Fig.13
7
Loading...
+ 16 hidden pages