Rotronic CP11 Operating Manual

OPERATION MANUAL
PORTABLE IAQ METER
CO2+RH+TEMP. IAQ METER/LOGGER
INTRODUCTION
Thank you for purchasing this portable IAQ meter. The meter measures CO level, air temp., dew point, wet bulb temp. and humidity and is an ideal instrument for indoor air quality (IAQ) diagnosis.
Poor indoor air quality is considered unhealthy because it causes tiredness, loss of ability to concentrate, and even illness(ex. Sick Building Syndrome). IAQ monitoring and survey, especially on CO level and air ventilation become widely applied in public areas such as offices, classrooms, factories, hospitals and hotels. It is also suggested in regulations of industrial hygiene in some countries. (Appendix)
The portable CO meter uses NDIR (non-
dispersive infrared) technology to ensure the reliability and long term stability. It’s useful in verifying HVAC system perform­ance and air ventilation control.
MATERIAL SUPPLIED
This package contains:
Meter 4pcs AA batteries
Operation manual Mini USB cable and software CD Soft carrying case
1
FEATURES
Big LCD display with blue backlight to use in dark area One touch to display CO2 /Dew point temperature/Wet bulb temperature/ Air temperature/Humidity in turns. Designed with NDIR (non-dispersive infrared) Waveguide technology CO2 sensor Programmable warning CO2 level Long time drift compensation on CO2 sensor Audible alarm (~80db)threshold setting Max ,Min, average functions included Mini USB output every 2 seconds to pc for analysis 99 points manually recording 30000 points automatic recording Review 99 points manual record Hold function freezes current readings Housing design surrounds with rails to help air ventilation for quick & accurate response Easy to manually calibrate at fresh air around 380-420ppm
2
POWER SUPPLY
The meter is powered by either 4 AA batteries or a DC adaptor(5V/1A output).
Install the batteries into the battery compartment on the bottom and make sure they are in correct polarity and good contact. cut off the power supply from batteries. The adaptor can’t be used as a battery charger.
When battery voltage gets low,
appear on the LCD . The CO sensor can’t work normally under low voltage, please replace with fresh b ies or connect with an adaptor.
When an adaptor is used, it will
will
(Fig.1)
atter
CO2
AM
Fig.1
3
LCD DISPLAY
AVG
DP
CAL
WBT
OO
CC
OO
CO2
HOLD
MAX
TEMP
MIN
FF
%rh%rh
AM PM
Symbols
Upper LCD Lower LCD Real time display
CO2 Carbon dioxide reading
HOLD Readings are freezed unchanged
MIN/MAX Minimum/Maximun readings
AVG Average reading
Low battery indicator
DP Dew point
TEMP Air temperature
WBT Wet bulb
%rh Unit of relative humidity
o
C/ F Celsius/Fahrenheit of temp. CAL In calibration status REC In manual/automatic logging RECALL In manual records recall mode
ON/OF F
SET
START ESC
MODE RECA LL
CO2/Temp/RH....reading
temperature
o
KEYPAD
Turns on and off the meter. Enters setup mode while meter is off. Turn on with to become non-sleep mode Exits setup/recall page. Start automatic logging.
Press to switch displayed mode. Long press to enter memory recall mode.
HOLD
4
REC Recall
temperature
HOLD
Freezes the current readings. Selects unit or increases value in setup. Cancels data hold function.
MEM
Press to manually record the reading. Selects unit or decreases value in setup.
MIN
Activates MIN,MAX,AVG function.
MAX/AV G
Saves and finishes settings.
OPERATION
POWER ON/OFF
ON/OF F
Press to turn the meter on and off.
SET
At power up, it emits a short beep and performs 30 seconds countdown(Fig.2) for meter warm up, then enters normal mode with current CO and real time displayed (Fig.3). The real time display date and time in turns.
Fig.2
Suggest to wait for 30 seconds warm up time. For quick start, you may press
MIN
MAX/AV G
for 2 seconds to end the warm up and entering normal display.
TAKING MEASUREMENT
The meter starts measurement when power on and update readings every second. In the condition of operating environment change (ex. from high to low temp.), it takes 30 sec to respond for CO sensor and 30 minutes for RH.
NOTE: Do not hold the meter close to faces in case exhalation affects CO levels.
2
CO2
AM
Fig.3
5
CO2 (Carbon Dioxide)
MODE
Press to switch the mode to CO2.
RECA LL
User can get the CO2 reading in ppm unit on main display.(Fig.4). The lower display shows the real time clock.
CO2
TEMP
OO
CC
AM
Fig.4
AIR,DP,WBT
MODE RECA LL
Press to switch to temperatures
AM
Fig.5
display(Fig.5,6,7).The lower display shows the real time clock.
Fig.7
WBT
OO
CC
DP
OO
CC
AM
Fig.6
Humidity
MODE RECA LL
Press to switch to humidity display.
AM
(Fig.8).The lower display shows the real time clock.
Fig.8
%rh%rh
6
DATA HOLD In normal display mode, press to
HOLD
freeze the readings, "HOLD” icon is flashed on the left top of LCD(Fig.9). All current readings are kept unchanged.
HOLD
Press again to cancel the hold function.
HOLD
%rh%rh
Fig.9
NOTE: THE HOLD FUNCTION IS DISABLE IN MIN/MAX/ AVG MODE.
BACKLIGHT The backlight will be activated for 10 seconds by pressing any key.
MIN,MAX,AVG,CURRENT REVIEW This meter allows you to check the minimun, maximun, average, current value from the moment you press the
MIN
MAX/AV G
key. Under normal mode, press
MIN
MAX/AV G
to see the minimum, maximum, average and current value in turns. Each press
MIN
MAX/AV G
of ,it displays MIN, MAX, AVG, current value in sequence. (Fig.10~12). To return to normal mode, press key to escape(Fig.13)
CO2
MAX
CO2
AVG
START ESC
CO2
Fig.10 Fig.11
CO2
Fig.12
AM
Fig.13
7
In this max/min/avg/current modes, it shows the corresponding readings on main display and the accumulated time of how long you are in this mode.(Fig10 ~12). The maximum time interval for this function is 18 hours.
MODE
In this mode, press to switch the
RECA LL
displayed parameters. Fig 14. The dispalyed sequence in this mode is
min max avg current min max.....
START
Press key can immediately leave
ESC
this mode and return to normal display. Distinguishing the normal display and
current review by checking the time display format. In normal display, the date & time displayed in turns. In current review, the display is the accumulated time of how long you are in max/min mode.
MAX
%rh% rh
Fig.14
NOTE: “HOLD” & “MANUALLY RECORD” & “RECALL” FUNCTION ARE DISABLE IN MIN/MAX/AVG MODE.
MANUALLY RECORDING
The meter features 99 point memories. In normal or hold mode, Press key
to record, icon and main display
REC
MEM
flash for about 3 seconds. The main display shows the memory serial number , at most 99 points.(Fig. 15)
Each memory contains all parameters (CO2,TA,%rh,DP,WBT), not only the
8
parameter you choose to see on main display.
Fig.15
AM
If the reading changes quickly, you can
HOLD
press to freeze the reading before
REC
manually recording the data. The manually recording function is
disabled in min/max mode.
99 MEMORIES RECALL
The meter features 99 point memories review function.
In normal or hold mode, press key for more than 2 seconds until the
MODE RECA LL
Recall
icon flashes.
MEM
HOLD
Press or to scroll the memories. The memory serial number displays on main LCD fist and reading comes after. (Fig.16&17)
CO2
Recall
MODE RECA LL
Recall
Fig.16
Press key to switch the display
Fig.17
parameter.(Fig.18)
The time displayed in memory recall mode is the recording
%rh%rh
time of this memory.
Recall
Fig.18
9
To escape the memory recall mode,
START
press key to leave and return to
ESC
normal display.
ALARM The meter features audible alarm to
give warnings when CO2 concentration exceeds the limit. (See P20 in setup for setting alarm threshold). It emits beeps(Abt.80dB)when CO2 level goes  over the set value and stops only when the readings fall below the set value. It beeps again when value goes over the limit.
DATA LOGGING
The meter can automatically record readings of CO2/TEMP./RH for long time environment monitoring. The memory capacity is 10000 points for each parameter. Users can set up sampling rate from 1 second to 4hours 59 minutes and 59 seconds (See SETUP P40).
After sampling rate is selected, press
START
for 2 seconds under normal
ESC
mode to start logging. The icon flashes to indicate the logging status and LCD main display shows the real time CO2 value. Lower displays are the real time clock.(Fig.19)
CO2
REC
AM
10
REC
Fig.19
To terminate data logging, press for 2 seconds, icon stops flashing
While press to start the logging
START ESC
REC
START ESC
again, the previous logging records are overwrite.
MODE
Press key to switch to the mode
RECA LL
you want to see. Except the mode switch function, min/max, manual record, hold and recall functions are all disabled during logging.
AUTO POWER OFF
The meter turns off automatically after 20 minutes of inactivity. To override the function, hold down and for 2 seconds to turn on the meter until “n
ON/OF F
SET
appears.(Fig.20)
NOTE: AUTO SLEEP FUNCTION WILL BE DISABLED DURING CALIBRATION MODE.
HOLD
Fig.20
Fig.21
SETUP
ON/OF F
When the meter is off, hold down for more than 1 sec to enter setup mode. To exit setup, press in P10 ~ P60
START ESC
and it returns to normal mode. (Fig.21)
11
SET
P10: 99 MEMORIES CLEAR
When entering setup mode, P10 and “CLr” (Fig.21) are displayed on the LCD.
MIN
MAX/AVG
Press to go into P11 and decide to clear or keep all the manual records. The current set will be blinking on LCD (Fig.22).
Press or to choose NO or YES and press to confirm.
MAX/AVG
MIN
MEM
HOLD
Fig.22
START
Press to escape and return to P10.
ESC
P20: CO2 ALARM
When entering setup mode, P20 and “ALAr” (Fig.23) are displayed on LCD.
MIN
MAX/AVG
Press to go into P21 for setting CO2 alarm threshold. The current set value will be blinking on LCD(Fig.24).
HOLD
Press to increase the value or
MEM
to decrease. The selectable alarm limits are 100~9900ppm,each press
tunes 100ppm. When the preferred alarm value is set, press to save the setting
START
or without saving and return to P20.
ESC
CO2 CO2
MAX/AVG
MIN
Fig.23 Fig.24
12
P30 TEMPERATURE UNIT
When entering setup mode, P30 and “unit” (Fig.25) are displayed on the LCD.
MIN
MAX/AVG
Press to go into P31 and decide the temperature unit.
The current set will display on LCD (Fig.26).
Press or to choose C or F and press to confirm.
MAX/AVG
MIN
MEM
HOLD
TEMP
o o
TEMP
OO
CC
Fig.25
START
Press to escape and return to P30.
ESC
Fig.26
P40:LOGGING SAMPLING RATE
Go into P40 for setting sampling rate of data logging(Fig.27). The range is from 1 second to 4 hours 59 minutes and 59 seconds.
MIN
MAX/AV G
Press and it goes into setting with blinking Hour digits on the lower display. To change the digit, press to increase and to decrease. Press
MIN
MAX/AV G
to confirm and enter Minute setting . Press again to confirm and enter
MAX/AV G
MEM
MIN
Second setting (Fig.28). Press to confirm the rate setting or without saving and return to P40.
Fig.27
START ESC
MIN
MAX/AV G
Fig.28
HOLD
13
P50 PRESSURE COMPENSATION
When entering setup mode, P50 and “PrES” (Fig.29) are displayed on the LCD.
MIN
MAX/AV G
Press to set pressure compensation value for CO2 measurement.
The current set will flash on LCD (Fig.30).The barometric pressure unit
is kpa.
MEM
HOLD
Press or to adjust the pressure value and press to confirm.
CO2
MIN
MAX/AV G
CO2
Fig.29
START
Press to escape and return to P50.
ESC
Fig.30
P60:REAL TIME CLOCK
Go into P60 for setting the real time clock of this meter.
MIN
MAX/AV G
Press and it goes into P61 to set the time format as 12hour or 24 hour. The current set will flash on LCD. Press
HOLD
or to change the format and press to confirm and entering real
MIN
MAX/AV G
MEM
time clock setting.(Fig.31) The blinking Year digits on the lower
display. To change the digit, press
MEM
or to adjust. Press to confirm
MIN
MAX/AV G
HOLD
and enter Month setting. Repeat above steps to complete the month/date/hour/ minute/second setting.(Fig.32)
PM
Fig.31
Fig.32
14
CO CALIBRATION
2
The meter is calibrated at standard 400ppm CO concentration in factory.
It’s suggested to do manual calibration regularly to maintain good accuracy.
Note:
When the accuracy becomes a concern after a long time usage, return to dealers for standard calibration.
CAUTION:
Do not calibrate the meter in the air with unknown CO concentration.
Otherwise, it will be calibrated as 400ppm by default and leads to inaccurate measurements.
The manual calibration is suggested to to be done in fresh outdoor air that is well ventilated and in sunny weather.
Place the meter in the calibration site. Press , , simultaneously for 3 seconds to turn on the meter and
START ESC
MEM
ON/OF F
SET
enter calibration mode (Fig.33). Two calibration menu is available: Humidity /CO2.
HOLD
Press key to select the CO2 calibration. 400ppm and “ “ icon are blinking on the LCD. Press
CAL
MIN
MAX/AV G
to start calibration. (Fig.34) Wait about 10 minutes until the blinking
stops and the calibration is completed automatically. To abort the calibration, turn off the meter at any time.
NOTE:
Ensure the batteries are with full voltage during the calibration to prevent from interruption or failed calibration.
15
CAL
CO2
%rh%rh
CAL
Fig.33
Fig.34
RH CALIBRATION
The meter defaults to be calibrated the humidity with 33% and 75% salt solution. The ambient condition is recommended to be at 25 C and stable humidity(better to be close to the calibrating value). To abort calibration, just turn off the meter.
CAUTION:
Do not calibrate the humidity without the default calibration salt. Otherwise, it will cause permanent damage. Contact the dealer for calibration salt or services.
33% calibration
Plug the sensor probe into 33% salt bottle. neously for 3 seconds to turn on the meter and enter calibration mode (Fig.
33.). bration. ¡§CAL¡¨ and calibrating value (32.8% if at are blinking on the LCD.
Meter is now calibrating, and will finish in about 60 minutes when ¡§CAL¡¨ and humidity stop blinking. bration, turn off the meter at any time.
75% calibration
Plug sensor probe into 75% salt bottle and entering calibration mode. In calibration mode, press to select 75% calibration(Fig.35), press to start.
o
MEM
START
Press , , simulta-
ESC
MIN
o
25 C)
MAX/AV G
Hold down to enter 33% cali-
ON/OF F
SET
To abort the cali-
HOLD
MIN
MAX/AV G
16
“CAL” and calibrating value (75.3% if at
o
25 C) are blinking on the LCD. Meter is now calibrating, and will finish
in about 60 minutes when “CAL” and humidity stop blinking. To abort the cali­bration, turn off the meter at any time.
CAL
%rh%rh
Fig.35
TROUBLESHOOTING
? Can’t power on
Press for more than 0.3 seconds and try again. Check whether batteries are in good contact and correct polarity, or the adaptor is well plugged. You may also remove the batteries for > 10 mins and then install the batteries again.
? Fixed readings
Check whether data hold function was activated. (HOLD icon at the left top)
? Slow response
Check whether the air flow channels on the rear were blocked.
? Error messages
E01/E33: CO sensor is out of order. Retry CO2 calibration. E02: The value is under range. E03: The value is over range. E04: The original data error results in this error (DP, WB) E11: Retry humidity calibration. E31: Temperature sensor or AD damaged. E32: Memory IC damaged. E33: Humidity sensor or circuit damaged.
ON/OF F
SET
17
PC CONNECTION
Connect the logger to PC, users can do logging setting and data transmission with the included software.
Plug the USB cable into the socket at the right side of the meter and the other port to PC. Then install the software in your PC with the following procedure.
1.Insert the CD Rom and run installation. Select a preferred directory and click “Next” step by step and finish it.
2. After the software installation is completed, it will run USB driver installation next.
18
Software Operation
AutoConnect
Start the software and it will detect
logger connection automatically and
indicate the COM Port information at
the bottom of the main screen as well
as the COMPort setting column.
Logger Settings
To set up logging plan, click “Setting” icon and select “Logger”. The setting page is opened.
Users can set up the logger identifier
number. Click OK for setting and Exit
the screen to confirm the settings.
19
Data Transmission
To transmit auto recorded data from the meter, click “Download” icon on the left top side of the main screen. All auto­recorded data in the logger will be transmitted. The raw data with time stamp will display at the lower part of the screen and the graph in the middle.
In the Graph display, the Y-axis indicates Temp, RH,DP, WBT, Co2 level in different line colors. And the X-axis can be switched to show Time or Recorded No. Stamp.
Data Review Function
The following functions help to view the GRAPH data in more detailed way.
Under View function, there are 4 tools to enlarge the Graph data for detailed data review.
Zoom in: Hold down “Ctrl” key on the PC keyboard and click any target point on the graph. It enlarges the point in each click.
20
Zoom window: Hold down “Ctrl” key and drag click the left button of the mouse to select an area on the Graph and the selected area will be enlarged .
Zoom Xaxis: Hold down “Ctrl” key and drag click the left button of the mouse to select an area on the Graph and the X-axis of the selected area will be enlarged.
Zoom Yaxis: Hold down “Ctrl” key and drag click the left button of the mouse to select an area on the Graph and the Y-axis of the selected area will be enlarged.
Three icons on the right top side of the main screen:
Redraw: After any review and zooming on the Graph, click “Redraw” to reset it to original format.
Clear: Click “Clear” to delete all data and Graph on the window. A warning box shows up for confirmation. Press “Yes” and all data will be cleared.
Cursor: Switch Cursor ON and it shows the raw data with the cursor. In any click on the graph, the digital data displayed. Switch OFF to disable the function.
21
Data Processing The downloaded data can be Saved or
Printed via “File” processing functions. Load: To retrieve saved files, just click
“Load” and select a desired file and it will be loaded in the main screen with the file details at the left top side.
SPECIFICATION
Measuring range CO2
Temperature Relative Humidity DP(Dew point temp.)
WB(Wet bulb temp.) Resolution
Accuracy CO2
ROTRONIC CP11
0~9999 ppm,
(5001~9999 ppm out of scale range)
o o
-20~60 C (-5~140 F)
0.1%~99.9%RH
o
-20.0~59.9 C
o
-5.0~59.9 C
o o
1ppm , 0.1 C/ F, 0.1%RH
+30ppm+5% of reading( 0~5000ppm)
Other ranges are not specified
Temperature Relative Humidity CO2 Warm-up time Response time CO2
Tair RH LCD / Meter size (mm) Operating condition
(avoid condensation) Storage condition Power supply Battery life Weight
Standard package
o o
+0.3 C/+0.6 F at 5~40 C
o o
+3%RH (at 25 C, 1 0~90%RH) ; +5%RH ( at 25 C , o ther ran ge )
-20 to 50 °C (CO2 sensor)
-20 to 60 °C (for the rest parameters)
-20~ 60 C,10~90%RH(avoid condensation)
Meter, manual ,AA bat.
30 seconds
<30 seconds(90% step change)
<2 mins (90% step change)
<10 minutes ( 90% step change)
40(H)x45(L), 173(L)x77(W)x42(H)
o
AA x 4pcs or 5V adaptor
> 24 hours (Alkaline battery)
200g
22
CO2 LEVELS AND GUIDELINES
Non-Enforced Reference levels
NIOSH recommendations 250-350 ppm: normal outdoor ambient
concentrations
600 ppm: minimal air quality complaints 600-1000 ppm: less clearly interpreted 1000 ppm: indicates inadequate ventilation;
complaints such as headaches, fatigue, and eye/throat irritation will be more widespread. 1000 ppm should be used as an upper limit for indoor levels.
EPA Taiwan: 600ppm and 1000ppm Type 1 indoor areas such as department
stores, theaters, restaurants, libraries, the acceptable CO2 concentration of 8 hours avarge is 1000ppm. Type 2 indoor areas with special requirements of good air quality such as schools, hospitals, day care centers, the suggested CO2 level is 600ppm.
Regulatory exposure limit
ASHRAE Standard 62-1989: 1000ppm
CO2 concentration in occupied building should not exceed 1000ppm.
Building bulletin 101 (BB101): UK standards for schools say that CO2 at averaged over the whole day(i.e. 9am to 3.30 pm) should not exceed 1500ppm.
OSHA: 5000ppm Time weighted average over five 8-hour work days should not exceed 5000ppm.
Germany, Japan, Australia, UK...: 5000ppm 8 hours weighted average in occupational exposure limit is 5000ppm.
1500ppm
23
Loading...