Congratulations on your purchase of the new portable humidity, temperature and CO2 instrument. Please read these short instruction carefully before installing the device.

General description

The CP11 is an accurate hand-held datalogger that displays & records relative humidity, temperature, CO2 and calculated parameters such as dew point and wet bulb. These short instructions are limited to a description of the main functions and installation of the device. The detailed instruction manual can be found on the internet at www.rotronic-humidity.com

Programming

Most of the settings, such as memory clear, CO2 alarm, units(C/F), sampling rate, pressure compensation and real time clock can be changed using the function keys.

5VDC adaptor port Integrated probe

PC connect port

Built-in CO2 sensor w/ calibration cap

Tripod hole

Battery

compartment

Power Supply

The CP11 uses 4pcs AA alkaline batteries or 5VDC adaptor (optional accessory). To insert (replace)the batteries, use the screw driver enclosed to loose 2 screws and put in the batteries in correct polarity with good contact.

Connection to PC or Laptop

Connect the CP11 via mini USB cable to a PC or Laptop. Recorded data download can be easily downloaded with the software enclosed.

Functions Key

ON/OFF/SETUP

-Press to turn on or turn off the instrument

-In normal mode, press longer to enter setup mode. -Press with HOLD key to disable sleep mode

 

 

HOLD/UP

 

 

 

 

-Freezes the current readings

 

 

 

 

-Cancels data hold function

 

 

 

 

-Selects unit or increases value in setup

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MODE/RECALL

START/ESC

 

 

 

-Press to switch displayed parameter

-Start automatic logging mode

 

 

 

-Press longer to enter memory recall

-Exits setup/recall/logging

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MEM/DOWN

 

 

MIN/MAX/AVG/ENTER

 

-Press to manually record the reading

 

 

-Activates MIN,MAX,AVG function

 

-Selects unit or decreases value in setup

 

 

-Saves and finishes settings

 

Display

 

 

 

CO2

TEMPDPWBT

UpperLCD Humidity/TEMP/CO2....reading

LowerLCD

Real time clock display

HOLDMAXMINAVGCAL

CO2

Carbon dioxide reading

HOLD

Readings are hold unchanged

 

O

 

 

 

C MIN/MAX Minimum/Maximum readings

 

O

AVG

Average reading

 

F

 

Low battery indicator

 

%rh DP

Dew point temperature

 

 

TEMP

Temperature

 

 

WBT

Wet bulb temperature

AM

REC

%rh

Unit of relative humidity

oC/oF

Celsius/Fahrenheit of temperature

PM

Recall

CAL

In calibration status

 

 

REC

In manual/automatic logging status

Operation

 

RECALL

In manual records recall mode

 

 

 

POWER ON/OFF

 

 

Press ON/OFFSET to turn the meter on. At power up,

 

CO2

it emits a short beep and per-forms 28 sec.

 

 

countdown (Fig.1) for meter warm up, then

 

 

enters normal mode with real time clock

 

 

displayed (Fig.2). Date and time are

 

 

displayed in turns.

 

AM

To power off, press ON/OFFSET again to turn off the

 

Fig.1

Fig.2

meter.

DATE/TIME SETUP

ON/OFF

When meter is powered on , hold down SET for more than 2 seconds until entering meter setup mode. Press HOLD or MEM key to select the program P80 (rtc) and press

MIN

MAX/AVG key to enter.

First, choose the time format as 12hour or 24hour by pressing HOLD or MEM and then

press

 

MIN

to confirm. Second, choose the date format as Month-Day-Year or Day-

MAX/AVG

 

 

 

 

 

MEM

 

MIN

 

Month-Year by pressing HOLD or

 

and then, press MAX/AVG

key to confirm. Now, start to

 

 

 

 

 

 

 

 

 

 

input the real time clock value from year, month, day, hour, minute to second. Press

HOLD or

MEM

to select and then press

 

MIN

to confirm. Press START to return to normal.

MAX/AVG

 

 

 

 

UNIT SETUP

 

 

 

 

ESC

 

 

 

 

 

 

 

 

 

 

 

ON/OFF

When meter is powered on , hold down

SET

for more than 2 seconds until entering

meter setup mode. Press HOLD or MEM

key to select the program P50 (unit) and press

MIN

 

 

 

 

 

 

 

 

 

 

 

 

 

MAX/AVG key to enter.

 

 

 

 

 

 

 

 

 

Press

HOLD

or

MEM

to select

o

C or

o

F and then press

MIN

to confirm. Press

START

to return

MAX/AVG

 

 

 

 

 

ESC

to normal.

TAKING MEASUREMENT

The meter starts measurement when power on and update readings every second. In the condition of operating environment change, it takes 30 seconds to respond for CO2 sensor.

NOTE: Do not hold the meter close to faces in case exhalation affects CO2

Humidity

MODE

Press RECALL to switch to humidity display.(Fig.3).The lower display shows the real time clock.

TEMP,DP,WBT

Press RECALLMODE to switch to temperatures display(Fig.4,5,6). The lower display shows the real time clock.

CO2 (Carbon Dioxide)

MODE

Press RECALL to switch the mode to CO2.User can get the CO2 reading in ppm unit on main display.(Fig.7). The lower display shows the real time clock.

DATA HOLD

In normal display mode, press HOLD to freeze the readings, "HOLD” icon is flashed on the left top of LCD (Fig.8). All current readings are kept unchanged. Press HOLD again to cancel the hold function.

NOTE:

The hold function is disabled in Min/Max/Avg mode.

BACKLIGHT

The backlight will be activated for 10 seconds by pressing any key.

MIN,MAX,AVG,CURRENT REVIEW

This meter allows you to check the minimun, maximum, average, current value from the moment you press the

 

MIN

key. Under normal mode, each press of

MIN

displays

MAX/AVG

MAX/AVG

 

 

 

 

 

 

 

MIN, MAX, AVG and current value in sequence.

 

(Fig.9~11). To return to normal mode, press STARTESC

to

escape(Fig.12)

 

In this max/min/avg/current modes, it shows the

 

corresponding readings on main display and the accumulated time of how long you are in this mode. The maximum time interval for this function is 18 hours.

MODE

Press RECALL can switch the displayed parameters. (Fig.13)

Distinguishing the normal display and current review by checking the time display format. In normal display, date & time are displayed in turns. In current review, the lower display is the accumulated time of how long you are in max/min/avg mode.

%rh Fig.3

TEMP

O

C

Fig.4

AM

DP

O

C

Fig.5

AM

WBT

O

C

Fig.6

AM

CO2

Fig.7

AM

HOLD

%rh Fig.8

NOTE:

“HOLD” & “MANUALLY RECORD” & “RECALL” functions are disabled in Min/ Max/Avg mode.

CO2

CO2

CO2

CO2

 

MAX

AVG

 

 

MAX

 

 

 

 

%rh

 

 

 

AM

 

Fig.9

Fig.10

Fig.11

Fig.12

Fig.13

ALARM

The meter features audible alarm to give warnings when CO2 concentration exceeds the limit. (See “Meter Setup” section for setting alarm threshold). It emits beeps (Abt. 80dB) when CO2 level goes over the set value and stops only when the readings fall below the set value. It beeps again when value goes over the limit.

AUTO POWER OFF

The meter turns off automatically after 20 minutes of inactivity. To override the function, first, the meter is in power off status. Hold down ON/OFFSET and HOLD for 2 seconds to turn on the meter until “n” appears.(Fig.14)

NOTE:

Auto sleep function will be disabled during calibration & auto recording mode. Auto sleep function WILL NOT be

disabled automatically via using adaptor. Fig.14

MANUALLY RECORDING

MEM

The meter features a 99 point memory. In the normal or hold mode, press key to start recording, the REC icon and the main display will be flashed 3 times. The

memory sequential number will show on the main display with 99 points in maximum. The data in the memory contains all parameters information and not only the displayed parameter. When all 99 points memory are occupied, a “FULL” message will be displayed on the screen. If you think the reading is changing too fast, you can press

HOLD key to freeze current reading first then manually record the data. This recording function will be disabled in the Min/Max mode.

99 MEMORIES RECALL

MODE

In normal or hold mode, press RECALL key for more than 2 seconds until the Recall icon flashes. Press HOLD or MEM to scroll the memories. The memory sequential number displays on main LCD first and reading comes after. (Fig16&17). Press key to switch the display parameter.(Fig.18).The time displayed in memory recall mode is the recording time of this memory.

 

 

CO2

 

 

 

 

%rh

AM

REC

Rec all

 

Fig.15

Rec all

Rec all

Fig.16

Fig.17

Fig.18

DATA LOGGING

The meter can automatically record Humidity/TEMP./CO2 for long term environment monitor. The memory capacity is 6000 points for each parameter. Users can set up sampling rate from 1 second to 4hours 59 minutes and 59 seconds. The factory preset rate is 30 second.

ON/OFF

To setup the sampling rate, when meter is powered on , hold down SET for more than

 

 

 

HOLD

MEM

key to select the program

2 seconds until entering meter setup mode. Press

or

 

MIN

key to enter.

 

 

P60 (rAtE) and press MAX/AVG

 

 

 

 

 

 

 

 

Press HOLD or MEM to select the sample rate from hour, minute to second and then press

MIN

to confirm. Press START

to return to normal.

MAX/AVG

 

ESC

 

After sampling rate is selected, press STARTESC for 2 seconds under normal mode to start logging. The REC icon flashes to indicate the logging status and LCD main display

shows the real time value. Lower displays are real time clock.(Fig.19)

Battery low indicator

A battery low icon “ “ appears when the supplied power is too low to maintain a normal operation. Please replace the batteries for accurate measuring. Suggest to use adaptor (optional part: AC1214) for long term measuring & datalogging.

Humidity Calibration

This meter can be calibrated either via 35% & 80% salt bottle or humidity generator. The ambient condition is recommended as stable 25oC. While the meter is powered off, plug the sensor probe into 35% salt. Press MEM simultaneously for 3 seconds to turn on the meter and enter calibration mode. Press HOLD or MEM to select

MIN

35.0% calibration and press MAX/AVG to start. “CAL” and calibrating value are blinking on the LCD.

CAUTION:

Do not calibrate the humidity without the default calibration salt. Otherwise, it will cause permanent damage. Contact ROTRONIC for calibration salt or services.

Single point calibration will cause error code E11 or non accurate reading. Always do dual point calibration to complete a process. Suggest to calibrate low humidity first.

CO2 Calibration

The sensor has an automatic calibration function to secure long-term accuracy. If a

calibration is desired, press

ON/OFF

START

MEM

simultaneously for 3 seconds to turn on the

SET

ESC

 

meter and enter calibration mode. Press HOLD or

 

MEM

key to select 400 ppm CO2

 

 

 

 

 

 

 

MIN

to start and see“CAL” and CO2 value

calibration (Outside Air calibration). Press MAX/AVG

 

 

 

 

 

 

 

 

 

 

blinking on LCD. Wait about 10 minutes until the blinking stops and the calibration is completed and back to normal mode. While choose 0 ppm CO2 calibration, remove the cap from meter rear side and plug the CO2 CALIBRATOR pipe into the sensor hole beneath. Wait about 10 min until the blinking stops and the calibration is completed and back to normal mode.

CAUTION:

Do not calibrate the meter in the air with unknown CO2 concentration. Otherwise, it will be calibrated as 0 or 400ppm by default and leads to inaccurate measurements.

Ensure the batteries are with full voltage during the calibration to prevent from interruption or failed calibration. When the accuracy becomes a concern after a long time usage, return to dealers for standard calibration.

Repeat above to start another run of logging. While the logging

CO2

 

Trouble Shooting

 

 

 

 

 

 

 

 

space is full, it shows “FULL”. To terminate data logging, press

 

 

 

 

 

 

 

 

 

 

STARTESC for 2 seconds, REC icon stops flashing.

 

 

 

 

Error

Messages

 

Solution

 

 

 

Press

MODE

key to switch to the mode you want to see. Except the

 

 

E01

CO2 sensor is out of order

Turn off meter and re-start again

RECALL

 

 

E33

CO2 sensor is out of order

Retry CO2 calibration

mode switching function, auto power off, min/max, manual

 

 

 

 

E02

measured value is under range

 

Put meter in normal condition

record, hold and recall functions are all disabled during logging. AM

REC

 

E03

measured value is over range

 

Put meter in normal condition

 

 

 

 

 

 

 

 

 

Meter Setup

 

 

 

Fig.19

 

E04

Original data error results in

 

Solve the original data error

When meter is powered on , hold down

ON/OFF

 

 

 

 

E11

this error (DP, WB)

 

 

Retry humidity calibration

SET for more than 2 seconds until entering

RH calibration error

 

 

meter setup mode. To exit setup, press

START

 

 

 

 

E31

Temp. sensor or AD damaged

 

Return for repair

 

 

ESC to return to normal mode.

 

 

 

 

Press HOLD

 

 

MIN

 

 

 

E32

Memory IC damaged

 

 

Return for repair

 

 

or MEM key to select the program and press MAX/AVG key to select.

 

E33

RH sensor or circuit damaged

 

Return for repair

 

 

The programmable setting flashes on display, press

HOLD

or MEM

key to select and

 

 

 

Technical Data

 

 

 

 

 

 

 

 

press

MIN

key to confirm. To leave without saving, press

START

 

 

 

 

 

 

 

 

 

 

MAX/AVG

 

 

 

 

 

 

 

 

 

 

 

ESC key to return.

 

 

 

 

 

 

 

 

 

 

 

 

 

Note:

 

 

Humidity/accuracy:

0.1% to 99.9%/ ±3.0%(10~95%@25 °C), ±5% (other)

P10, 99 points MEMORY clear

 

 

 

Temperature/accuracy:

-20...60

o

C/+0.3

o

C@5~40

o

C

 

 

 

 

 

 

 

 

 

 

“is displayed while

CO2 /accuracy:

0...9999ppm/+(30ppm+5% of reading)@0~5000ppm

 

 

memory is cleared.

P11, choose yes or no

Storage and transit:

-20...60oC/10...90%rh, non condensing.

 

 

P20, auto LOG data clear

 

Note:

Operating limit at electronics:0...50oC for CO2, -20...60oC for rest/ non condensing.

 

Memory:

99 records manually, 18000 records auto logging.

 

 

“is displayed while

 

 

Dimensions:

234(L)x77(W)x42(H),mm

P21, choose yes or no

memory is cleared.

Weight:

200gr.

 

 

 

 

P30, CO2 ALARM setup

 

Note:

Battery type:

4pcs AA alkaline

 

-Factory preset 1400ppm.

DC power type:

5VDC, 250mA. Optional available.

 

 

P31, choose from 100

-Adjustable scale is every

PC connection:

mini USB port

to 9900

 

100ppm

 

 

 

P40, CO2 ABC setup

 

Note:

Optional Accessory

 

P41, Choose on or off

-Factory preset ABC

 

Part Number

 

function as ON

5VDC universal adaptor:

AC1214

 

 

 

 

P50, Temp. UNIT setup

 

Note:

CO2 calibrator:

CO2 CALIBRATOR

 

 

 

 

-Factory preset oC

 

 

 

oC

 

 

P51, choose oC or oF

oF

 

 

 

P60, logging sampling RATE

Note:

 

 

-Factory preset at 30 sec.

 

 

 

 

 

 

P61, choose from

 

-The format is

 

 

04:59:59 to

 

 

hour:minute:second

 

 

00:00:01

 

Note:

 

 

P70, PRESSURE compensation

 

 

-Factory preset at 1013hpa

ROTRONIC AG, CH-8303 Bassersdorf

 

 

P71, choose from

 

-The adjustable scale is

Tel. +41 44 838 11 44, www.rotronic.com

700 to 1990hpa

 

 

every 1hpa.

ROTRONIC Messgeräte GmbH, D-76275 Ettlingen

 

 

Note:

Tel. +49 7243 383 250, www.rotronic.de

P80, REAL TIME CLOCK setup

ROTRONIC SARL, 56, F - 77183 Croissy Beaubourg

 

 

-Factory preset at 12H &

Tél. +33 1 60 95 07 10, www.rotronic.fr

 

 

P81, choose 12 or 24

dd-mm-year format.

ROTRONIC Italia srl, I- 20157 Milano

hour format

 

 

 

Tel. +39 2 39 00 71 90, www.rotronic.it

choose DATE

 

 

 

ROTRONIC Instruments (UK) Ltd, West Sussex RH10 9EE

 

 

 

Phone +44 1293 571000, www.rotronic.co.uk

format

 

 

 

year

 

 

ROTRONIC Instrument Corp, NY 11788, USA

 

 

 

Input Year/

 

 

Phone +1 631 427-3898, www.rotronic-usa.com

month

 

 

Month/ Day

date

or

 

ROTRONIC Instruments Pte Ltd, Singapore 159836

Input Hour/

Hour

 

Phone +65 6376 2107, www.rotronic.com.sg

Minute

 

 

ROTRONIC Shanghai Rep. Offi ce, Shanghai 200233, China

Min./Sec.

Second

 

 

Phone +86 40 08162018, www.rotronic.cn