SHORT INSTRUCTION MANUAL
Congratulations on your purchase of the new portable humidity, temperature and CO2 instrument. Please read these short instruction carefully before installing the device.
General description
The CP11 is an accurate hand-held datalogger that displays & records relative humidity, temperature, CO2 and calculated parameters such as dew point and wet bulb. These short instructions are limited to a description of the main functions and installation of the device. The detailed instruction manual can be found on the internet at www.rotronic-humidity.com
Programming
Most of the settings, such as memory clear, CO2 alarm, units(C/F), sampling rate, pressure compensation and real time clock can be changed using the function keys.
5VDC adaptor port Integrated probe
PC connect port
Built-in CO2 sensor w/ calibration cap
Tripod hole
Battery
compartment
Power Supply
The CP11 uses 4pcs AA alkaline batteries or 5VDC adaptor (optional accessory). To insert (replace)the batteries, use the screw driver enclosed to loose 2 screws and put in the batteries in correct polarity with good contact.
Connection to PC or Laptop
Connect the CP11 via mini USB cable to a PC or Laptop. Recorded data download can be easily downloaded with the software enclosed.
Functions Key
ON/OFF/SETUP
-Press to turn on or turn off the instrument
-In normal mode, press longer to enter setup mode. -Press with HOLD key to disable sleep mode
|
|
HOLD/UP |
|
|
|||
|
|
-Freezes the current readings |
|
|
|||
|
|
-Cancels data hold function |
|
|
|||
|
|
-Selects unit or increases value in setup |
|
|
|||
|
|
|
|
|
|
|
|
|
|
|
|
|
MODE/RECALL |
||
START/ESC |
|
|
|
-Press to switch displayed parameter |
|||
-Start automatic logging mode |
|
|
|
-Press longer to enter memory recall |
|||
-Exits setup/recall/logging |
|
|
|
|
|
|
|
|
|
|
|
||||
|
|
|
|
|
|||
MEM/DOWN |
|
|
MIN/MAX/AVG/ENTER |
|
|||
-Press to manually record the reading |
|
|
-Activates MIN,MAX,AVG function |
|
|||
-Selects unit or decreases value in setup |
|
|
-Saves and finishes settings |
|
Display |
|
|
|
CO2 |
TEMPDPWBT |
UpperLCD Humidity/TEMP/CO2....reading |
|
LowerLCD |
Real time clock display |
||
HOLDMAXMINAVGCAL |
CO2 |
Carbon dioxide reading |
|
HOLD |
Readings are hold unchanged |
||
|
O |
|
|
|
C MIN/MAX Minimum/Maximum readings |
||
|
O |
AVG |
Average reading |
|
F |
|
Low battery indicator |
|
%rh DP |
Dew point temperature |
|
|
|
TEMP |
Temperature |
|
|
WBT |
Wet bulb temperature |
AM |
REC |
%rh |
Unit of relative humidity |
oC/oF |
Celsius/Fahrenheit of temperature |
||
PM |
Recall |
CAL |
In calibration status |
|
|
REC |
In manual/automatic logging status |
Operation |
|
RECALL |
In manual records recall mode |
|
|
|
POWER ON/OFF |
|
|
|
Press ON/OFFSET to turn the meter on. At power up, |
|
CO2 |
|
it emits a short beep and per-forms 28 sec. |
|
|
|
countdown (Fig.1) for meter warm up, then |
|
|
|
enters normal mode with real time clock |
|
|
|
displayed (Fig.2). Date and time are |
|
|
|
displayed in turns. |
|
AM |
|
To power off, press ON/OFFSET again to turn off the |
|
||
Fig.1 |
Fig.2 |
||
meter. |
DATE/TIME SETUP
ON/OFF
When meter is powered on , hold down SET for more than 2 seconds until entering meter setup mode. Press HOLD or MEM key to select the program P80 (rtc) and press
MIN
MAX/AVG key to enter.
First, choose the time format as 12hour or 24hour by pressing HOLD or MEM and then
press |
|
MIN |
to confirm. Second, choose the date format as Month-Day-Year or Day- |
||||||
MAX/AVG |
|||||||||
|
|
|
|
|
MEM |
|
MIN |
|
|
Month-Year by pressing HOLD or |
|
and then, press MAX/AVG |
key to confirm. Now, start to |
||||||
|
|
|
|
|
|
|
|
|
|
input the real time clock value from year, month, day, hour, minute to second. Press
HOLD or |
MEM |
to select and then press |
|
MIN |
to confirm. Press START to return to normal. |
||
MAX/AVG |
|||||||
|
|
|
|
||||
UNIT SETUP |
|
|
|
|
ESC |
||
|
|
|
|
|
|||
|
|
|
|
|
|
ON/OFF |
|
When meter is powered on , hold down |
SET |
for more than 2 seconds until entering |
meter setup mode. Press HOLD or MEM |
key to select the program P50 (unit) and press |
||||||||||||
MIN |
|
|
|
|
|
|
|
|
|
|
|
|
|
MAX/AVG key to enter. |
|
|
|
|
|
|
|
|
|
||||
Press |
HOLD |
or |
MEM |
to select |
o |
C or |
o |
F and then press |
MIN |
to confirm. Press |
START |
to return |
|
MAX/AVG |
|||||||||||||
|
|
|
|
|
ESC |
to normal.
TAKING MEASUREMENT
The meter starts measurement when power on and update readings every second. In the condition of operating environment change, it takes 30 seconds to respond for CO2 sensor.
NOTE: Do not hold the meter close to faces in case exhalation affects CO2
Humidity
MODE
Press RECALL to switch to humidity display.(Fig.3).The lower display shows the real time clock.
TEMP,DP,WBT
Press RECALLMODE to switch to temperatures display(Fig.4,5,6). The lower display shows the real time clock.
CO2 (Carbon Dioxide)
MODE
Press RECALL to switch the mode to CO2.User can get the CO2 reading in ppm unit on main display.(Fig.7). The lower display shows the real time clock.
DATA HOLD
In normal display mode, press HOLD to freeze the readings, "HOLD” icon is flashed on the left top of LCD (Fig.8). All current readings are kept unchanged. Press HOLD again to cancel the hold function.
NOTE:
The hold function is disabled in Min/Max/Avg mode.
BACKLIGHT
The backlight will be activated for 10 seconds by pressing any key.
MIN,MAX,AVG,CURRENT REVIEW
This meter allows you to check the minimun, maximum, average, current value from the moment you press the
|
MIN |
key. Under normal mode, each press of |
MIN |
displays |
||
MAX/AVG |
MAX/AVG |
|
||||
|
|
|
|
|
|
|
MIN, MAX, AVG and current value in sequence. |
|
|||||
(Fig.9~11). To return to normal mode, press STARTESC |
to |
|||||
escape(Fig.12) |
|
|||||
In this max/min/avg/current modes, it shows the |
|
corresponding readings on main display and the accumulated time of how long you are in this mode. The maximum time interval for this function is 18 hours.
MODE
Press RECALL can switch the displayed parameters. (Fig.13)
Distinguishing the normal display and current review by checking the time display format. In normal display, date & time are displayed in turns. In current review, the lower display is the accumulated time of how long you are in max/min/avg mode.
%rh Fig.3
TEMP
O
C
Fig.4
AM
DP
O
C
Fig.5
AM
WBT
O
C
Fig.6
AM
CO2
Fig.7
AM
HOLD
%rh Fig.8
NOTE:
“HOLD” & “MANUALLY RECORD” & “RECALL” functions are disabled in Min/ Max/Avg mode.
CO2 |
CO2 |
CO2 |
CO2 |
|
MAX |
AVG |
|
|
MAX |
|
|
|
|
%rh |
|
|
|
AM |
|
Fig.9 |
Fig.10 |
Fig.11 |
Fig.12 |
Fig.13 |
ALARM
The meter features audible alarm to give warnings when CO2 concentration exceeds the limit. (See “Meter Setup” section for setting alarm threshold). It emits beeps (Abt. 80dB) when CO2 level goes over the set value and stops only when the readings fall below the set value. It beeps again when value goes over the limit.
AUTO POWER OFF
The meter turns off automatically after 20 minutes of inactivity. To override the function, first, the meter is in power off status. Hold down ON/OFFSET and HOLD for 2 seconds to turn on the meter until “n” appears.(Fig.14)
NOTE:
Auto sleep function will be disabled during calibration & auto recording mode. Auto sleep function WILL NOT be
disabled automatically via using adaptor. Fig.14
MANUALLY RECORDING
MEM
The meter features a 99 point memory. In the normal or hold mode, press key to start recording, the REC icon and the main display will be flashed 3 times. The
memory sequential number will show on the main display with 99 points in maximum. The data in the memory contains all parameters information and not only the displayed parameter. When all 99 points memory are occupied, a “FULL” message will be displayed on the screen. If you think the reading is changing too fast, you can press
HOLD key to freeze current reading first then manually record the data. This recording function will be disabled in the Min/Max mode.
99 MEMORIES RECALL
MODE
In normal or hold mode, press RECALL key for more than 2 seconds until the Recall icon flashes. Press HOLD or MEM to scroll the memories. The memory sequential number displays on main LCD first and reading comes after. (Fig16&17). Press key to switch the display parameter.(Fig.18).The time displayed in memory recall mode is the recording time of this memory.
|
|
CO2 |
|
|
|
|
%rh |
AM |
REC |
Rec all |
|
Fig.15 |
Rec all |
Rec all |
|
Fig.16 |
Fig.17 |
Fig.18 |
DATA LOGGING
The meter can automatically record Humidity/TEMP./CO2 for long term environment monitor. The memory capacity is 6000 points for each parameter. Users can set up sampling rate from 1 second to 4hours 59 minutes and 59 seconds. The factory preset rate is 30 second.
ON/OFF
To setup the sampling rate, when meter is powered on , hold down SET for more than
|
|
|
HOLD |
MEM |
key to select the program |
2 seconds until entering meter setup mode. Press |
or |
||||
|
MIN |
key to enter. |
|
|
|
P60 (rAtE) and press MAX/AVG |
|
|
|||
|
|
|
|
|
|
Press HOLD or MEM to select the sample rate from hour, minute to second and then press
MIN |
to confirm. Press START |
to return to normal. |
MAX/AVG |
||
|
ESC |
|
After sampling rate is selected, press STARTESC for 2 seconds under normal mode to start logging. The REC icon flashes to indicate the logging status and LCD main display
shows the real time value. Lower displays are real time clock.(Fig.19)
Battery low indicator
A battery low icon “ “ appears when the supplied power is too low to maintain a normal operation. Please replace the batteries for accurate measuring. Suggest to use adaptor (optional part: AC1214) for long term measuring & datalogging.
Humidity Calibration
This meter can be calibrated either via 35% & 80% salt bottle or humidity generator. The ambient condition is recommended as stable 25oC. While the meter is powered off, plug the sensor probe into 35% salt. Press MEM simultaneously for 3 seconds to turn on the meter and enter calibration mode. Press HOLD or MEM to select
MIN
35.0% calibration and press MAX/AVG to start. “CAL” and calibrating value are blinking on the LCD.
CAUTION:
Do not calibrate the humidity without the default calibration salt. Otherwise, it will cause permanent damage. Contact ROTRONIC for calibration salt or services.
Single point calibration will cause error code E11 or non accurate reading. Always do dual point calibration to complete a process. Suggest to calibrate low humidity first.
CO2 Calibration
The sensor has an automatic calibration function to secure long-term accuracy. If a
calibration is desired, press |
ON/OFF |
START |
MEM |
simultaneously for 3 seconds to turn on the |
|||||
SET |
ESC |
|
|||||||
meter and enter calibration mode. Press HOLD or |
|
MEM |
key to select 400 ppm CO2 |
||||||
|
|
||||||||
|
|
|
|
|
MIN |
to start and see“CAL” and CO2 value |
|||
calibration (Outside Air calibration). Press MAX/AVG |
|||||||||
|
|
|
|
|
|
|
|
|
|
blinking on LCD. Wait about 10 minutes until the blinking stops and the calibration is completed and back to normal mode. While choose 0 ppm CO2 calibration, remove the cap from meter rear side and plug the CO2 CALIBRATOR pipe into the sensor hole beneath. Wait about 10 min until the blinking stops and the calibration is completed and back to normal mode.
CAUTION:
Do not calibrate the meter in the air with unknown CO2 concentration. Otherwise, it will be calibrated as 0 or 400ppm by default and leads to inaccurate measurements.
Ensure the batteries are with full voltage during the calibration to prevent from interruption or failed calibration. When the accuracy becomes a concern after a long time usage, return to dealers for standard calibration.
Repeat above to start another run of logging. While the logging |
CO2 |
|
Trouble Shooting |
|
|
|
|
|
|
|
|
||||||
space is full, it shows “FULL”. To terminate data logging, press |
|
|
|
|
|
|
|
|
|
|
|||||||
STARTESC for 2 seconds, REC icon stops flashing. |
|
|
|
|
Error |
Messages |
|
Solution |
|
|
|
||||||
Press |
MODE |
key to switch to the mode you want to see. Except the |
|
|
E01 |
CO2 sensor is out of order |
Turn off meter and re-start again |
||||||||||
RECALL |
|
|
E33 |
CO2 sensor is out of order |
Retry CO2 calibration |
||||||||||||
mode switching function, auto power off, min/max, manual |
|
|
|||||||||||||||
|
|
E02 |
measured value is under range |
|
Put meter in normal condition |
||||||||||||
record, hold and recall functions are all disabled during logging. AM |
REC |
|
|||||||||||||||
E03 |
measured value is over range |
|
Put meter in normal condition |
||||||||||||||
|
|
|
|
|
|
|
|
|
|||||||||
Meter Setup |
|
|
|
Fig.19 |
|
E04 |
Original data error results in |
|
Solve the original data error |
||||||||
When meter is powered on , hold down |
ON/OFF |
|
|
|
|
E11 |
this error (DP, WB) |
|
|
Retry humidity calibration |
|||||||
SET for more than 2 seconds until entering |
RH calibration error |
|
|
||||||||||||||
meter setup mode. To exit setup, press |
START |
|
|
|
|
E31 |
Temp. sensor or AD damaged |
|
Return for repair |
|
|
||||||
ESC to return to normal mode. |
|
|
|
|
|||||||||||||
Press HOLD |
|
|
MIN |
|
|
|
E32 |
Memory IC damaged |
|
|
Return for repair |
|
|
||||
or MEM key to select the program and press MAX/AVG key to select. |
|
E33 |
RH sensor or circuit damaged |
|
Return for repair |
|
|
||||||||||
The programmable setting flashes on display, press |
HOLD |
or MEM |
key to select and |
|
|
|
|||||||||||
Technical Data |
|
|
|
|
|
|
|
|
|||||||||
press |
MIN |
key to confirm. To leave without saving, press |
START |
|
|
|
|
|
|
|
|
|
|
||||
MAX/AVG |
|
|
|
|
|
|
|
|
|
|
|||||||
|
ESC key to return. |
|
|
|
|
|
|
|
|
|
|||||||
|
|
|
|
Note: |
|
|
Humidity/accuracy: |
0.1% to 99.9%/ ±3.0%(10~95%@25 °C), ±5% (other) |
|||||||||
P10, 99 points MEMORY clear |
|
|
|
Temperature/accuracy: |
-20...60 |
o |
C/+0.3 |
o |
C@5~40 |
o |
C |
||||||
|
|
|
|
|
|
|
|
|
|
“ |
“is displayed while |
CO2 /accuracy: |
0...9999ppm/+(30ppm+5% of reading)@0~5000ppm |
|
|
|
memory is cleared. |
||||
P11, choose yes or no |
Storage and transit: |
-20...60oC/10...90%rh, non condensing. |
||||
|
|
|||||
P20, auto LOG data clear |
|
Note: |
Operating limit at electronics:0...50oC for CO2, -20...60oC for rest/ non condensing. |
|||
|
Memory: |
99 records manually, 18000 records auto logging. |
||||
|
|
“ |
“is displayed while |
|||
|
|
Dimensions: |
234(L)x77(W)x42(H),mm |
|||
P21, choose yes or no |
memory is cleared. |
|||||
Weight: |
200gr. |
|||||
|
|
|
|
|||
P30, CO2 ALARM setup |
|
Note: |
Battery type: |
4pcs AA alkaline |
||
|
-Factory preset 1400ppm. |
DC power type: |
5VDC, 250mA. Optional available. |
|||
|
|
|||||
P31, choose from 100 |
-Adjustable scale is every |
PC connection: |
mini USB port |
|||
to 9900 |
|
100ppm |
||||
|
|
|
||||
P40, CO2 ABC setup |
|
Note: |
Optional Accessory |
|
||
P41, Choose on or off |
-Factory preset ABC |
|
Part Number |
|||
|
function as ON |
5VDC universal adaptor: |
AC1214 |
|||
|
|
|
|
|||
P50, Temp. UNIT setup |
|
Note: |
CO2 calibrator: |
CO2 CALIBRATOR |
||
|
|
|
||||
|
-Factory preset oC |
|
|
|||
|
oC |
|
|
|||
P51, choose oC or oF |
oF |
|
|
|
||
P60, logging sampling RATE |
Note: |
|
|
|||
-Factory preset at 30 sec. |
|
|
||||
|
|
|
|
|||
P61, choose from |
|
-The format is |
|
|
||
04:59:59 to |
|
|
hour:minute:second |
|
|
|
00:00:01 |
|
Note: |
|
|
||
P70, PRESSURE compensation |
|
|
||||
-Factory preset at 1013hpa |
ROTRONIC AG, CH-8303 Bassersdorf |
|||||
|
|
|||||
P71, choose from |
|
-The adjustable scale is |
Tel. +41 44 838 11 44, www.rotronic.com |
|||
700 to 1990hpa |
|
|
every 1hpa. |
ROTRONIC Messgeräte GmbH, D-76275 Ettlingen |
||
|
|
Note: |
Tel. +49 7243 383 250, www.rotronic.de |
|||
P80, REAL TIME CLOCK setup |
ROTRONIC SARL, 56, F - 77183 Croissy Beaubourg |
|||||
|
|
|||||
-Factory preset at 12H & |
Tél. +33 1 60 95 07 10, www.rotronic.fr |
|||||
|
|
|||||
P81, choose 12 or 24 |
dd-mm-year format. |
ROTRONIC Italia srl, I- 20157 Milano |
||||
hour format |
|
|
|
Tel. +39 2 39 00 71 90, www.rotronic.it |
||
choose DATE |
|
|
|
ROTRONIC Instruments (UK) Ltd, West Sussex RH10 9EE |
||
|
|
|
Phone +44 1293 571000, www.rotronic.co.uk |
|||
format |
|
|
|
|||
year |
|
|
ROTRONIC Instrument Corp, NY 11788, USA |
|||
|
|
|
||||
Input Year/ |
|
|
Phone +1 631 427-3898, www.rotronic-usa.com |
|||
month |
|
|
||||
Month/ Day |
date |
or |
|
ROTRONIC Instruments Pte Ltd, Singapore 159836 |
||
Input Hour/ |
Hour |
|
Phone +65 6376 2107, www.rotronic.com.sg |
|||
Minute |
|
|
ROTRONIC Shanghai Rep. Offi ce, Shanghai 200233, China |
|||
Min./Sec. |
Second |
|
|
Phone +86 40 08162018, www.rotronic.cn |