Rosemount 951A NO/NOx Analyzer-Rev Z Manuals & Guides

Instruction Manual
556383-Z March 2002
Model 951A
NO/NOx Analyzer
http://www.processanalytic.com

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical repre-
sentative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and mainte-
nance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate In- struction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s per­formance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Teflon® is a registered trademark of E. I. duPont de Nemours and Co., Inc. Alconox ® is a registered trademark of Alconox, Inc.
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
Model 951A
PREFACE...........................................................................................................................................P-1
Definitions ...........................................................................................................................................P-1
Intended Use Statement.....................................................................................................................P-2
Safety Summary .................................................................................................................................P-2
General Precautions For Handling And Storing High Pressure Gas Cylinders .................................P-4
Documentation....................................................................................................................................P-5
Compliances .......................................................................................................................................P-5
1-0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-1
1-1 Overview................................................................................................................................1-1
1-2 Options ..................................................................................................................................1-2
a. Range I.D. .......................................................................................................................1-2
b. Remote Range Change ..................................................................................................1-2
c. Range I.D. and Remote Range Change .........................................................................1-2
d. Sample Pump..................................................................................................................1-2
1-3 Specifications ........................................................................................................................1-3
Instruction Manual
556383-Z
March 2002

TABLE OF CONTENTS

2-0 INSTALLATION ....................................................................................................................2-1
2-1 Facility Preparation................................................................................................................2-1
a. Outline and Mounting Dimensions ..................................................................................2-1
b. Location...........................................................................................................................2-1
c. Power Requirements ......................................................................................................2-1
2-2 Unpacking..............................................................................................................................2-1
2-3 Gas Requirements.................................................................................................................2-1
a. Air Or Oxygen .................................................................................................................2-1
b. Span Gas ........................................................................................................................2-2
2-4 Sample Requirements ...........................................................................................................2-2
2-5 Gas Connections ...................................................................................................................2-2
2-6 Electrical Connections ...........................................................................................................2-6
a. Recorder Output..............................................................................................................2-6
b. Thermocouple Connections for Measuring Temperature of NO2 to NO Converter ........2-6
c. Remote Range Change ..................................................................................................2-6
d. Connections of the Range I. D. Kit..................................................................................2-6
e. Power Connections.........................................................................................................2-6
3-0 STARTUP AND OPERATION ..............................................................................................3-1
3-1 Startup Procedure .................................................................................................................3-1
3-2 Calibration..............................................................................................................................3-3
3-3 Measuring Efficiency Of No2 To No Converter And Adjusting Temperature Setpoint ..........3-4
a. Test Setup for Measurement of Conversion Efficiency...................................................3-4
b. Test Procedure................................................................................................................3-4
c. Subnormal Conversion Efficiency ...................................................................................3-8
d. Replacement of Converter ..............................................................................................3-8
e. Capillaries .......................................................................................................................3-8
f. TEA Scrubber..................................................................................................................3-9
3-4 Routine Operation .................................................................................................................3-9
3-5 Recommended Calibration Frequency..................................................................................3-9
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Instruction Manual
556383-Z March 2002
4-0 THEORY................................................................................................................................4-1
4-1 Principles Of Operation .........................................................................................................4-1
a. Nitric Oxide Determination By Chemiluminescense Method ..........................................4-1
b. NOX Determination..........................................................................................................4-1
c. Ozone Generation ...........................................................................................................4-1
4-2 Analyzer Flow System ...........................................................................................................4-1
a. Flow Of Sample Or Standard Gas To Reaction Chamber..............................................4-1
b. Flow of Air or Oxygen .....................................................................................................4-3
c. Flow System Operating Modes .......................................................................................4-3
d. Converter Bleed Flow .....................................................................................................4-4
4-3 Electronic Circuitry.................................................................................................................4-4
a. Amplifier Board and Associated Circuitry .......................................................................4-5
b. Valve Control Board, Front Panel Mode Switch and Associated Circuitry .....................4-6
c. ±15 Volt Power Supply ....................................................................................................4-6
d. High Voltage Power Supply ............................................................................................4-7
e. Converter Temperature Control Board and Associated Elements .................................4-7
f. Fan Control Circuit ..........................................................................................................4-8
g. Remote Operation Option ...............................................................................................4-8
h. Range I.D. Option ...........................................................................................................4-8
Model 951A
5-0 ROUTINE SERVICING..........................................................................................................5-1
5-1 System Checks And Adjustments .........................................................................................5-1
a. Meter Mechanical Zero ...................................................................................................5-1
b. Amplifier Zero Adjustments.............................................................................................5-1
c. Inter-Range Attenuation Correlation adjustments...........................................................5-1
d. Meter Fullscale Span Adjustment ...................................................................................5-2
e. Overall Sensitivity............................................................................................................5-3
f. Ozone Output..................................................................................................................5-3
g. Background Current........................................................................................................5-3
h. NO2 to NO Converter Temperature Adjustment .............................................................5-4
i. Flow Balance...................................................................................................................5-4
5-2 Servicing Flow System ..........................................................................................................5-5
a. Sample Capillary .............................................................................................................5-5
b. Ozone Restrictor and Capillary .......................................................................................5-6
5-3 NO2 To NO Converter............................................................................................................5-7
a. Replacing ........................................................................................................................5-7
5-4 Photomultiplier/Reaction Chamber........................................................................................5-8
a. Cleaning Reaction Chamber...........................................................................................5-8
b. Photomultiplier Tube and Housing..................................................................................5-10
5-5 Servicing Electronic Circuitry.................................................................................................5-11
6-0 REPLACEMENT PARTS ......................................................................................................6-1
6-1 Circuit Board Replacement Policy .........................................................................................6-1
6-2 Matrix .....................................................................................................................................6-1
6-3 Replacement Parts ................................................................................................................6-2
a. Pneumatics .....................................................................................................................6-7
b. Converter Assembly........................................................................................................6-8
c. Low Tempco Option........................................................................................................6-10
7-0 RETURN OF MATERIAL ....................................................................................................7-1
7-1 Return Of Material .................................................................................................................7-1
7-2 Customer Service ..................................................................................................................7-1
7-3 Training..................................................................................................................................7-1
ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model 951A
Figure 1-1. Model 951A NO/NOX Analyzer .............................................................................. 1-1
Figure 2-1. Model 951A Outline and Mounting Dimensions .................................................... 2-3
Figure 2-2. Front Panel Indicators and Controls ...................................................................... 2-4
Figure 2-3. Controls and Adjustments Located Behind Swing Out Front Panel...................... 2-4
Figure 2-4. Rear Panel............................................................................................................. 2-4
Figure 2-5. Remote Range Kit Installed................................................................................... 2-7
Figure 2-6. Typical Interconnection of Remote Range Kit ....................................................... 2-7
Figure 3-1. Amplifier Board Adjustments ................................................................................. 3-3
Figure 3-2. Measuring Efficiency of NO2 to NO Converter ...................................................... 3-7
Figure 3-3. Conversion Efficiency as a Function of Converter Temperature........................... 3-8
Figure 4-1. Schematic Flow Diagram of Model 951A .............................................................. 4-2
Figure 4-2. Functional Schematic Diagram of Electronic Signal Circuitry ............................... 4-5
Figure 5-1. Amplifier Board ......................................................................................................5-2
Figure 5-2. NO2 to NO Converter Assembly ............................................................................ 5-7
Figure 5-3. Reaction Chamber/Photomultiplier Assembly ....................................................... 5-9
Figure 5-4. Reaction Chamber Assembly and Phototube Housing ......................................... 5-9
Figure 5-5. Terminal Chassis Assembly Wiring Diagram ...................................................... 5-12
Figure 6-1. Model 951A – Exploded View................................................................................ 6-3
Figure 6-2. Model 951A – Exploded View (continued)............................................................. 6-4
Figure 6-3. Model 951A – Exploded View (continued)............................................................. 6-5
Figure 6-4. Electronics ............................................................................................................. 6-6
Figure 6-5. Front Panel Pneumatic Components..................................................................... 6-7
Figure 6-6. Converter Components.......................................................................................... 6-8
Figure 6-7. Ozone Generator Components.............................................................................. 6-9
Figure 6-8. Tempco Retrofit Component Location................................................................. 6-10
Instruction Manual
556383-Z
March 2002

LIST OF ILLUSTRATIONS

LIST OF TABLES

Table 2-1. Model 951A Controls and Adjustments ................................................................. 2-5
Table 3-1. Proper Gas Supply Pressures for Various Levels of Sample NOX........................ 3-2
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Instruction Manual
556383-Z March 2002
Model 951A
DRAWINGS
619604 Schematic Diagram, Computer Interface 619710 Schematic Diagram, ±15V Power Supply 641871 Schematic Diagram, Temp Control 649819 Schematic Diagram, Amplifier Board 649822 Diagram, Tubing - Model 95A 649834 Schematic Diagram, 951A NO/NOx Analyzer 649835 Pictorial Wiring Diagram, Model 951A NO/NOx Analyzer 649958 Flow Diagram, Model 951A 652423 Schematic Diagram, Power Supply 652834 Schematic Diagram, Power Supply - Thermocooler 652838 Pictorial Wiring Diagram, Low Tempco Option 654348 Schematic Diagram, Hi Voltage Board Assembly 656313 Schematic Diagram, Valve Control Board 780350 Schematic Diagram, 4-20mA 0-5V Option 780726 Wiring Diagram, 4-20mA 0-5V Option 780727 Installation Drawing, Model 951A w/4-20mA 0-5V Option 780809 Wiring Diagram, 4-20mA 0-5V Option and Low Tempco
(Located In Rear Of Manual)
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 951A

PREFACE

The purpose of this manual is to provide information concerning the components, functions, in­stallation and maintenance of the Model 951A .
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this analyzer before operating it. Read this instruction manual completely.

DEFINITIONS

The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.
DANGER .
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.
556383-Z
March 2002
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure, condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Instruction Manual
556383-Z March 2002
Model 951A

INTENDED USE STATEMENT

The Model 951A NO/NOx Analyzer is intended for use as an industrial process measurement device only. It is not intended for use in medical, diagnostic, or life support applications, and no inde­pendent agency certifications or approvals are to be implied as covering such applications.

SAFETY SUMMARY

If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.

AUTHORIZED PERSONNEL

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, do not operate or service this instrument before reading and understanding this instruction manual and receiving appropriate training. Save these instructions.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure.. Installation requires access to live parts which can cause death or serious injury.
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
DANGER.
ULTRA VIOLET LIGHT HAZARD
UV light from the ozone generator can cause permanent eye damage. DO NOT LOOK DIRECTLY AT THE UV SOURCE IN THE OZONE GENERATOR. Use of UV filtering glasses is recommended.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 951A
DANGER.
OZONE HAZARD
This instrument generates ozone while operating. Ozone is toxic by inhalation and is a strong irri­tant to the throat and lungs. Ozone is also a strong oxidizing agent. Its presence is detected by a characteristic pungent odor.
The instrument EXHAUST outlet contains ozone and nitrogen dioxide which is toxic by inhalation. The BYPASS outlet contains various oxides of nitrogen, and if the sample source is from the ex­haust of an internal combustion engine, it may contain unburned hydrocarbons and carbon mon­oxide which is highly toxic and, depending on duration of exposure, can cause headache, nausea, loss of consciousness and death.
Avoid any inhalation of the internally generated ozone, sample, EXHAUST and BYPASS discharge. Keep all tubing fittings checked for tightness to avoid internal leaks.
Connect rear panel EXHAUST and BYPASS outlets to outside vent via separate lines 1/4 inch (6.3mm) or larger. Use only Teflon or stainless steel tubing.
556383-Z
March 2002
WARNING.
PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely affect the safety of this instrument. Use only factory approved components for repair.
CAUTION
AIR FLOW
Do not operate instrument without air flow to the ozonator; plugging of the filter will result.
CAUTION.
PRESSURIZED GAS
This unit requires periodic calibration with a known standard gas. It also may utilizes a pressur­ized carrier gas, such as helium, hydrogen, or nitrogen. See General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-4.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Instruction Manual
556383-Z March 2002
Model 951A
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981
Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202 Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125°F (52°C). A flame should never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 951A

DOCUMENTATION

The following Model 951A NO/NOx Analyzer instruction materials are available. Contact Customer Serv­ice Center or the local representative to order.
556383 Instruction Manual (this document)

COMPLIANCES

This product may carry approvals from several certifying agencies, for use in non-hazardous, indoor loca­tions.
97-C218
556383-Z
March 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Instruction Manual
556383-Z March 2002
Model 951A
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 951A
Instruction Manual
556383-Z
March 2002
SECTION 1

DESCRIPTION AND SPECIFICATIONS

1-1 OVERVIEW

The Model 951A NO/NOX Analyzer continu­ously analyzes a flowing gas sample per­forming one of two switch selectable determinations:
1. Nitric oxide (NO); or
2. Combined nitric oxide (NO) and nitrogen dioxide (NO tion [NO
) designated NOX. By defini-
2
] = [NO] + [NO2].
X
Typical applications include analyzing ve­hicular exhaust emissions from internal com­bustion engines and monitoring the effluent from stationary (stack) sources
The analyzer utilizes the chemiluminescent method of detection In the nitric oxide deter­mination, sample NO is quantitatively con­verted into NO
by gas-phase oxidation with
2
molecular ozone produced within the analyzer from air or oxygen supplied by an external cylinder. A characteristic of this reaction is the elevation of approximately 10% of the NO molecules to an electronically-excited state, followed by immediate reversion to the
non-excited state accompanied by emission of photons.
The emitted photons impinge on a photomul­tiplier detector generating a low-level DC cur­rent. The current is amplified to drive a front panel meter and an accessory potentiometric recorder if desired.
To minimize noise and reduce dark current, the photomultiplier tube is mounted in a ther­moelectrically-cooled housing with tempera­ture held constant at about 59
o
F (15oC). Control circuitry is contained in the 652831 Cooler Temperature Controller/Power Supply Assembly. The power supply circuit provides a high-current source of DC voltage. A ther­mistor sensor attached to the housing, and an associated switching transistor, control a pass transistor, providing closely regulated on-off control.
Attached to the cooler housing is thermal fuse F3, setpoint 150
o
F (65oC). This fuse protects
the thermoelectric cooler against the over-
2
heating that otherwise could occur as a result of excessively high ambient temperature or failure of a fan.
SPAN
ZERO
ON
OFF
Rosemount Analytical
MODE
NO NOx
PPM RANGE
250 1000
100 2500
25 10000
10 AUTO
CONVERTER
ADJUST

Model 951A NO/NOx Analyzer

Figure 1-1. Model 951A NO/NOX Analyzer
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
556383-Z March 2002
Model 951A
CAUTION
EXCESSIVE HEAT
Do not operate this analyzer without the air duct covering the cooling fins of the ther­moelectric cooler. Excessive heat may damage the cooling devices. The cooler indicator lamp DS1 (mounted on the upper left side of the inner flow regulator panel and can be viewed through the window) will cycle on and off until a control point is reached.
Analyzer functioning for the NOX determina­tion is identical to that described above for the NO determination except that, before entry into the reaction chamber, the sample is routed through a converter where the NO
2
component is dissociated to form NO. In­strument response is proportional to total NO in the converted sample, that is the sum of the NO present in the original sample plus the NO produced by dissociation of NO
.
2
To minimize system response time an internal sample-bypass feature provides high-velocity sample flow through the analyzer.
the different thermal conductivity's of the individual components of the sample stream. The method is especially well suited to analysis of two-component sam­ple streams. However, analysis of multi-component streams is possible if the various components of the background gas occur in relatively constant ratio, or have similar thermal conductivity's.

b. Remote Range Change

For applications where remote operation of the analyzer is desired, as in an emis­sion test console, the Remote Range Change Option may be used. This option permits either the operator or a computer to override and disable the front panel MODE and PPM RANGE Switches and thus to control selection of: (a) NO or NO mode, and (b) ppm range.
The unit consists of an electrical plug connector with plug-in logic card and har­ness for connection to a rear panel termi­nal strip on the analyzer.

c. Range I.D. and Remote Range Change

X
The electronic circuitry is modularized, utiliz­ing plug-in printed circuit boards with solid state components and test jacks for easy troubleshooting and replacement.
If desired, the Model 951A may be factory equipped with various optional features in ad­dition to the standard features of the basic in­strument. Brief descriptions of the principal options are given in the following section.

1-2 OPTIONS

a. Range I.D.

This option provides contact closure sig­nals that enable a computer or other ex­ternal device to identify the setting of the front panel PPM RANGE Switch. The Model 951A is designed to continuously measure the concentration of a single component of interest in a flowing gas mixture. The measurement is based on
The Range I.D./Remote Range Change Option is a combination of the Range I.D. and Remote Range Change options. This option is compatible with a user supplied remote control system employing a 24 VDC digital output and input for analyzer range control and analyzer range sense, respectively. It is completely integral within the analyzer and provides a termi­nal strip on the rear of the analyzer for connections of the user cable.

d. Sample Pump

The basic Model 951A is designed to ac­cept pressurized samples. To permit analysis of gases at atmospheric or sub­atmospheric pressure the analyzer may be equipped with an optional, internally mounted sample pump. This option is not available on Low Tempco versions of the 951A. An external, accessory pump can be ordered instead.
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 951A

1-3 SPECIFICATIONS

Catalog Number ............................ 193702 Model 951A NO/NOX Analyzer
Ranges .......................................... Selectable fullscale range of 10, 25, 100, 250, 1000, 2500 and
Sensitivity ...................................... 0.1 ppm on 10 ppm range
Linearity ......................................... ±1% of fullscale
Response Time (Electronic Plus Flow)
Standard Sample Capillary.... Approximately one second on all ranges except 10 ppm.
Auxiliary Sample Capillary..... Five seconds on all ranges.
Precision........................................ ±5% of fullscale
Stability
Zero ....................................... 1% of fullscale in 24 hours
Span ...................................... 1% of fullscale in 24 hours
Detector Operating Temperature .. Atmospheric
Recorder Output............................ Selectable output of 10 millivolts, 100 millivolts, 1 volt or 5 volts
Ambient Temperature.................... 40°F to 100°F (4.4°C to 37.7°C)
Electrical Power Requirements ..... 107 to 127 VAC, 50/60 Hz, 1000 watts
Dimensions.................................... 9.0 x 17.8 x 22.0 inches (228.6 x 450.9 x 558.8mm) HxWxD
Weight ........................................... 76 lbs. (34.5 kg)
Instruction Manual
556383-Z
March 2002
10,000 parts per million
The specified linearity is obtainable throughout the operating range, contingent upon use of an appropriate combination of oxygen source gas, gas pressure settings and electronic adjustments.
Approximately three seconds on 10 ppm range.
For such applications as monitoring stack sources, where compara­tively slow response is desired, an internal switch provides an op­tional electronic response time of approximately 10 seconds to 90% of fullscale on all ranges.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Instruction Manual
556383-Z March 2002
Model 951A
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 951A
Instruction Manual
556383-Z
March 2002
SECTION 2

INSTALLATION

2-1 FACILITY PREPARATION

Sections 2-1a through 2-1c provide informa­tion that may be required prior to installation of the analyzer.

a. Outline and Mounting Dimensions

Significant dimensions are shown in Fig­ure 2-1.

b. Location

Install analyzer in a clean area, not sub­ject to excessive vibration or extreme temperature variations .
Preferably, the analyzer should be mounted near the sample stream, to minimize sample transport time. A circuit controlled by a thermal switch holds inter­nal temperature of the analyzer to the cor­rect operating temperature for ambient temperatures in the range 40
o
(4.4
C to 37.7oC). Temperatures outside
these limits necessitate use of special temperature controlling equipment or en­vironmental protection.
Preferably, the cylinders of air (or oxygen) and span gas should be located in an area of relatively constant ambient tem­perature.

c. Power Requirements

Electrical power requirements are 107 to 127 VAC, 50/60 Hz, 1000 watts.

2-2 UNPACKING

Unpack instrument carefully. Preparatory to shipment, the photomultiplier tube housing and the sample pump (if instrument is so equipped) are immobilized with hold down screws, inserted from the bottom of the in­strument and marked with red paint for identi­fication. The hold down screws must be
o
F to 100oF
removed prior to operation of the instrument. In the event the instrument is ever returned to the factory, these screws must be replaced to ensure safe shipment.

2-3 GAS REQUIREMENTS

CAUTION
HIGH PRESSURE GAS CYLINDERS
This instrument requires use of oxygen and a known standard gas in high pres­sure cylinders. Refer to Handling and Storing High Pressure Gas Cylinders lo­cated on page P-4.

a. Air or Oxygen

This is used as both (a) oxygen source for generation of the ozone required for the chemiluminescent reaction, and (b) stan­dard gas for zero calibration. Gas for both purposes may be supplied from a single cylinder and routed through a tee. Alter­natively, two separate cylinders may be used.
Oxygen is usable in all applications. Air is suitable for the oxygen source only if the desired fullscale operating range is 2500 ppm or less. Breathing grade oxygen or air is recommended. Clean, dried ambient air containing less than 0.1 ppm nitric ox­ide may be used, provided that its dew­point is below -10 insufficiently dried, or contains excessive nitric oxide, instrument response will not be linear to 2500 ppm.
o
F (-23oC). If air is
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Instruction Manual
556383-Z March 2002
Model 951A

b. Span Gas

This is a standard gas of accurately known composition, used to set an up­scale calibration point.
Alternative span gases are:
1. The usual span gas is NO in a back­ground of nitrogen.
2. For span check or adjustment in the NO
mode, the span gas may be NO
X
in a background of air or nitrogen.
3. For convenient calibration in either the NO or NO
mode, the span gas
X
may be a cylinder gas mixture con­sisting of known concentrations of both NO and NO
in a background of
2
nitrogen.
NOTE
For maximum calibration accuracy, the concentration of NO and/or NO span gas should be as near as possi­ble to that in the sample gas. Also, the span gas should be supplied to the rear panel SAMPLE inlet at the same pressure as the sample gas. To ensure constant pressure, a pressure regula­tor may be utilized immediately up­stream from the SAMPLE inlet.
in the
2
tory can provide technical assistance if de­sired .
Proper supply pressures for sample and span gases depend on whether or not the analyzer is equipped with the optional 632748 Sample Pump:
For the basic Model 951A, without sample pump, sample must be supplied to the SAM­PLE inlet at a pressure of 5 to 10 psig (34.5 to 69 kPa). This ensures that the normal bypass
2
flow of two liters per minute will be obtainable by adjustment of the BYPASS Needle Valve. (Proper bypass flow is essential for rapid system response and stable flow into reaction chamber.)
For an analyzer equipped with sample pump, the acceptable pressure range at the SAM­PLE inlet is 0 to 5 psig (0 to 34.5 kPa). The pump pressurizes the sample to between 5 and 10 psig (34.5 to 69 kPa) for supply to the internal flow system.

2-5 GAS CONNECTIONS

1. Remove plugs and caps from all inlet and outlet fittings. See Figure 2-4.
2. Connect EXHAUST outlet to outside vent via tubing with O.D. of 1/4 inch (6.3mm) or larger.
Preferably, each gas used should be sup­plied from a tank or cylinder equipped with a clean, non-corrosive type two stage regulator. In addition, a shut-off valve is desirable. If possible, install the gas cylin­ders in an area of relatively constant am­bient temperature.

2-4 SAMPLE REQUIREMENTS

The Sample must be relatively clean and dry before entering the analyzer. In general, be­fore admission to the analyzer, the sample should be filtered to 2 microns and should have a dewpoint below 90
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
o
F (32oC). The fac-
3. Connect external lines from air (or oxy­gen) cylinder and sample source to corre­sponding rear panel inlet ports. For sample line, stainless steel tubing is rec­ommended.
4. Adjust regulator on air (or oxygen) cylin­der for output pressure of 35 to 40 psig (251 to 276 kPa).
5. Supply sample gas to rear panel SAMPLE inlet at appropriate pressure: 5 to 10 psig (34.5 to 69 kPa) for basic analyzer without sample pump; 0 to 5 psig (0 to 34.5 kPa) for analyzer with pump.
Model 951A
[
]
[
]
Instruction Manual
556383-Z
March 2002
19.0
[482]
8.7
[222]
8.7
[221]
DIMENSIONS
Inch
mm
3. ALLOW 16 INCHES IN FRONT OF INSTRU­MENT FOR DOOR SWING.
2. ALLOW 12 INCHES MINIMUM CLEARANCE ABOVE INSTRUMENT FOR MAINTENANCE..
1. RECOMMENDED MOUNTING HARDWARD: 10-32 MACHINE SCREWS (SUPPLIED BY CUSTOMER).
Figure 2-1. Model 951A Outline and Mounting Dimensions
[228]
7.0
[178]
1.25 [32]
9.0
4.5
114
18.3
[465]
2.0
[51]
18.3
[465]
18.3
[465]
18.3
[465]
20.2
[513]
6.0
[152]
18.3
[465]
RECOMMENDED PANEL CUTOUT
1.25 [32]
1.0
[25.4]
2.8
[70]
8.5
[216]
4.6
[118]
[63]
6.6
168
.25
6 PLC'S
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
A
556383-Z March 2002
Model 951A
AC Power On/off Switch
Mode Switch
Photomultiplier High Voltage Interlock Switch
Recorder Output Select Switch
Zero Adjust
ON
Rosemount Analytical
PPM Range Switch
Span Adjust
SPAN
ZERO
FLOW
PPM RANGEMODE
250 100
25 10
1000 2500 10000 AUTO
CONVERTER
ADJUST
BALANCE
Converter Adjust
Converter Indicator
Model 951A NO/NOx Analyzer
Figure 2-2. Front Panel Indicators and Controls
Response Time Selector Switch
10
5
FAST SLOW
100
1
RESPONSE
MV
V
REC
OUTPUT
R22
R34
CALIB
R16
R33
R31
R6
INPUT
R32
Bypass Needle Valve
FLOW
BALANCE
Sample Pressure Gauge
B Y P A S
S
B Y P A S
S
OZONE
OZONE
OFF
OZONE
OZONE
OFF
Ozone Pressure Gauge
Bypass Flowmeter
Bypass Flowmeter
Ozone Pressure Regulator
Ozone Off Push­button Switch
Figure 2-3. Controls and Adjustments Located Behind Swing Out Front Panel
Thermocouple Connector TB9
ir Inlet
Exhaust
Rosemount Analytical
Flow Balance Needle Valve
Sample Inlet
Recorder Connection TB4-2
Fuse F1 Fuse F2
Figure 2-4. Rear Panel
Model 951A NO/NOx Analyzer
Sample Backpressure Regulator
Recorder Connection TB4-1
Range I.D. Connec­tor J10 (option)
Cable Gland For Re­corder Cable
AC Power Cable
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 951A
CONTROL FUNCTION
ON/OFF Controls AC power to all components in instrument MODE Switch For selection of NO or NOX
Readout of NO or NOX in ppm as selected with MODE switch. Upper scale is
Meter
PPM RANGE Switch Select fullscale range for meter and recorder
CONVERTER
ADJUST
ZERO
SPAN
SAMPLE backpressure regu­lator and pressure gauge
BYPASS Flowmeter and Nee­dle Valve
OZONE pressure regulator and gauge
OZONE OFF switch
FLOW BALANCE needle valve
RESPONSE TIME selector switch
RECORDER OUTPUT selector switch
Meter mechanical zero
Photomultiplier High Voltage Interlock Switch
graduated 0 to 100 for use with 10, 100, 1000, and 10,000 ppm ranges. Lower scale is graduated 0 to 25 for use with 25 and 2500 ppm ranges.
LED illuminates during application of power to converter heater. When Con­verter reaches temperature equilibrium, LED will go off and on at intervals of about 2 seconds, indicating correct temperature control. Through-panel screwdriver adjustment of converter temperature. Should be set for temperature that yields optimum combination of high efficiency for the NO NO conversion and extended life for the catalytic converter. Optimum tempera­ture differs from one instrument to another, see Section 3-3, page 3-4. Set zero point on meter scale or recorder chart. With MODE switch in NO posi­tion and zero air supplied to SAMPLE inlet, ZERO Control is adjusted for zero reading. Set upscale calibration point on meter scale or recorder chart. MODE switch is at NO or NO NO/NO
and suitably pressurized standard gas of accurately known
X
content is supplied to SAMPLE inlet. With PPM RANGE switch set to
X
the appropriate range for the span gas, the SPAN control is adjusted to correct reading on meter or recorder. Adjustment and indication of pressure (and therefore flow) of sample or standard gas routed through sample capillary, and into reaction chamber. Proper setting dependent on operating range. See Table 3-1, page 3-2. Bypass flow indication and adjustment for sample or standard gas to SAMPLE inlet. Setting of 2 L/min. is recommended to ensure rapid response and opti­mum functioning of sample backpressure regulator Pressure adjustment and indication for air or oxygen supplied to AIR inlet for use as oxygen source for internal ozone generator. Oxygen is usable for all ranges. Air is usable only for 2500 ppm range or lower. Proper pressure setting de­pendent on operating range, see Table 3-1, page 3-2. Removes AC power to ultraviolet source lamp in ozone generator. Power is OFF when switch indicator is red.
Use to equalize pressure drop for NO and NOX legs of the internal flow system. Criterion for correct adjustment is an NO2 free nitric oxide standard gas should give equal readings of ppm for NO and NO check , if NO
free standard gas is unavailable, is to verify that the reading on
2
modes. An alternate flow balance
X
the SAMPLE pressure gauge is the same for NO and NO Select either FAST or SLOW electronic response. When set to FAST, electronic response time (for 0 to 90% of fullscale) is then approximately one second for all ranges except 10 ppm, which is approximately three seconds. When set to SLOW (as in monitoring stack sources), electronic response time 0 to 90% of fullscale) is approximately ten seconds for all ranges.
To select output of 10 mV, 100 mV, 1V or 5 V for a potentiometric recorder.
Located on rear of meter. With AC power OFF, meter should read zero. If not use screw adjustment on meter. When analyzer front panel is open, photomultiplier high voltage power supply is automatically shut off.
Instruction Manual
556383-Z
March 2002
to
2
modes.
X
Table 2-1. Model 951A Controls and Adjustments
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
556383-Z March 2002
Model 951A

2-6 ELECTRICAL CONNECTIONS

a. Recorder Output

If a recorder is used, connect leads to terminals marked REC 1 (+) and 2 (-) on TB4 (Figure 2-4, page 2-4). (Note that, within the analyzer, the negative recorder output terminal is connected to ground.)
Set REC OUTPUT Selector Switch SW1 (Figure 2-3, page 2-4) to the recorder span: 10 mV, 100 mV, 1 VDC or 5 VDC.
b. Thermocouple Connections for Meas-
uring Temperature of NO Converter
Temperature of the NO2 to NO converter may be monitored by connecting a cus­tomer supplied millivolt meter to thermo­couple connector TB9 located on the rear panel (Figure 2-4, page 2-4). The position of the thermocouple in the converter bun­dle will influence the actual temperature readout.

c. Remote Range Change

to NO
2
The MAN/AUTO RANGE Switch provides the choice of local or remote selection of operating range. With switch at MAN, range selection is under control of the front panel PPM RANGE Switch. With switch at AUTO, the front panel PPM RANGE Switch is disabled; range selec­tion is remotely controlled, by either the operator or a computer, via contact clo­sure signals applied to terminals on TB6.
Control at TB5 and TB6 is accomplished by a contact closure from the terminal marked COM (for common) to the termi­nal marked with the name of the desired function. Contact closure requirements are 25 mA at 15 VDC.

d. Connections of the Range I. D. Kit

The Range I.D. option provides contact closure signals that permit a computer or other external device to identify the posi­tion selected manually with the front panel PPM RANGE Switch. The cable is con­nected to PPM RANGE Switch SW1 and extends to connector J10 mounted on the rear of the case (Figure 2-4, page 2-4).
The Remote Range Change Kit, Figure 2-5 page 2-7, consists of a plug-in Adapter Board plus attached harness and rear terminal plate.
The Adapter Board has two two-position slide switches: the MAN/AUTO MODE Switch and the MAN/AUTO RANGE Switch.
The MAN/AUTO MODE Switch provides the choice of local or remote control of the NO/NO associated with the NO
Mode Switching Solenoid Valve,
X
to NO converter.
2
With switch at MAN, the function is under control of the front panel MODE Switch. With switch at AUTO, the front panel MODE Switch is disabled; the function is then remotely controlled, by either the op­erator or a computer, via contact closure signals applied to terminals on TB5.
The pin-out connections of J10 are as follows:
Pin A 10 ppm Pin B 25 ppm Pin C 100 ppm Pin D 250 ppm Pin E 1000 ppm Pin F 2500 ppm Pin H 10000 ppm Pin J Wiper

e. Power Connections

Connect power cord to an AC source of 107 to 127 volts, 50/60 Hz. If power outlet does not have third (ground) contact, use an adapter to provide proper grounding.
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 951A
y
Computer Adapter Board
Auto/Man Range Switch
10
5
FAST SLOW
100
1 MV REC
RANGE AUTO MAN
RESPONSE V OUTPUT
R22
R34
CALIB
R16
R33
INPUT
R31
R32
MODE
AUTO
MAN
R6
Rosemount Analytical
Electronics Assembly
DC Harness (J1)
Auto/Man Mode Switch
NOX
SPAN
ZERO
Instruction Manual
10
25
100
1000
500
2
10,000
556383-Z
March 2002
Computer Adapter Board installed on Interconnect Board of the Electronics Assembly
Figure 2-5. Remote Range Kit Installed
Terminal
Assembl
Remote Mode Select Switch
NOx
SPAN
ZERO
COM
10
25
100
250
1000
2500
10,000
COM
Terminal Assembly mounted on rear panel of Model 951A
Remote Range
Select Switch
With all contacts of Range Select Switch open, analyzer is on 250 ppm range.
Figure 2-6. Typical Interconnection of Remote Range Kit
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
556383-Z March 2002
Model 951A
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 951A
Instruction Manual
556383-Z
March 2002
SECTION 3

STARTUP AND OPERATION

Preparatory to startup and operation, a thorough familiarization with Table 2-1 (page 2-5), Figure 2-2 (page 2-4), Figure 2-3 (page 2-4) and Figure 2-4 (page 2-4) is recommended. These figures give locations and brief descriptions of operating controls. For more detailed information on con­trols, refer to Section 4.

3-1 STARTUP PROCEDURE

Following are detailed stepwise instructions on startup and calibration. For conven­ience, condensed instructions for startup, routine calibration and normal operation are provided at the front of this manual.
WARNING
OZONE HAZARD
When instrument power is on, the ultra­violet source sample is energized, con­verting a portion of the oxygen contained with the ozonator into ozone. With normal flow through the ozonator, the ozonated air or oxygen is continu­ously swept though the ozonator and into the reaction chamber. If flow is stopped, however, the ozone will diffuse and will attack the sintered metallic re­strictor element in the tee fitting at the upstream end of the ozonator. Opera­tion of the analyzer with the restrictor element thus rusted will result in the following consequences: Reduced flow of air or oxygen through the ozonator, ozone deprivation within the reaction chamber and non-linear response of the analyzer to NO/NOX.
To prevent such damage, verify that the front panel ozone ON/OFF switch is turned off if flow of feed gas to the air inlet is terminated.
1. With power removed from analyzer, check front panel meter. It should read zero; if not, adjust Mechanical Zero Screw at rear of meter for zero reading.
2. Place front panel PPM RANGE Switch at 1000.
3. Set front panel MODE Switch at NO or
NO
. Place POWER Switch ON. Electri-
X
cal power is now being supplied to all circuits, including sample pump if ana­lyzer is so equipped. Analyzer will now require approximately one hour for tem­perature equilibration before ready for calibration.
4. Establish correct pressure for air oxy­gen:
a. Verify that pressure regulator on cyl-
inder of air or oxygen is set for sup­ply pressure of 35 to 40 psig (242 to 276 kPa).
b. On internal gas control panel, Figure
2-3. adjust OZONE Pressure Regu­lator so that OZONE Pressure Gauge indicates either 20 psig (138 kPa) or 30 psig (207 kPa), depend­ing on the desired operating range.
5. Establish correct flow of sample gas:
a. Supply sample gas to rear panel
SAMPLE inlet.
b. Adjust SAMPLE Backpressure
Regulator so SAMPLE Pressure Gauge indicates the value appropri­ate to the desired operating range.
c. Adjust BYPASS Needle Valve for
reading of two liters per minute on BYPASS Flowmeter.
Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Operation 3-1
Instruction Manual
556383-Z March 2002
Model 951A
NOTE
Inability to obtain a bypass flow of two liters per minute by adjustment of the BYPASS Needle Valve usually indicates insufficient sample supply pressure at the SAMPLE inlet.
6. Establish correct flow of zero air:
a. Supply zero air to rear panel SAM-
PLE inlet.
b. Note reading on SAMPLE Pressure
Gauge. It should be the same as in Step 5b. If not, adjust output pres­sure regulator on air cylinder as re­quired.
Max. NOX Level
(ppm)
1000 Air
2500 Air
10,000 Oxygen
Gas Supplied
to Air Inlet
Provides flow of approx. 500 cc/min to ozone generator
Provides flow of approx. 1000 cc/min to ozone generator
Provides flow of approx. 1000 cc/min to ozone generator
7. Establish correct flow of upscale standard gas:
a. Supply upscale standard gas to rear
panel SAMPLE inlet.
b. Note reading on SAMPLE Pressure
Gauge. It should be the same as in Step 5b. If not, adjust output pres­sure regulator on cylinder of upscale standard gas as required.
Supply pressures for sample and up­scale standard gases must be the same; otherwise, readout will be in error.
The Analyzer is now ready for calibration per Section 3-2, page 3-3.
Ozone
Pressure Gauge Setting
20 psig (138 kPa)
30 psig (207 kPa)
30 psig (207 kPa)
NOTE
Sample
Pressure Gauge Setting
4 psig (27.6 kPa)
Provides flow of approx. 60 cc/min to reaction chamber
1.5 psig (10.3 kPa)
Provides flow of approx. 20 cc/min to reaction chamber
Table 3-1. Proper Gas Supply Pressures for Various Levels of Sample NO
3-2 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
X
Model 951A
Instruction Manual
556383-Z
March 2002

3-2 CALIBRATION

1. Zero Calibration.
a. Set PPM RANGE Switch for the
same range that will be used during sample analysis. Set SPAN Control at about midrange.
b. Supply zero gas to rear panel
SAMPLE inlet.
c. Adjust ZERO Control for zero
reading on meter or recorder, then lock ZERO Control knob.
2. Upscale Calibration.
a. Set PPM RANGE Switch at the po-
sition appropriate to the particular span gas.
b. Supply upscale standard gas of ac-
curately known NO/NO rear panel SAMPLE inlet.
c. Place MODE Switch at NO if nitric
oxide span gas is used.
d. Adjust SPAN Control so that read-
ing on meter or recorder is equal to
content to
X
the known ppm concentration of NO or NO eration at NO
in the span gas. In op-
X
levels in the range
X
of 2500 to 10000 ppm, the correct reading may not be obtainable, ini­tially, by adjustment of the SPAN Control. The cause is that the re­duced sample flow required for linearity results in lowered sensitiv­ity and slower response. To com­pensate for these effects, make the electronic adjustments of Steps e and f below.
e. If necessary, increase sensitivity by
raising photomultiplier voltage per Section 5-1e, page 5-3.
f. To reduce observed noise, select
SLOW position of Response Time Selector Switch SW2, Figure 3-1.
3. When correct upscale reading is ob­tained, lock SPAN Control knob.
Calibration is now complete. Before Placing analyzer in operation, however, measure efficiency of the NO
to NO converter per
2
Section 3-3, page 3-4.
10
5
R22
1 V OUTPUT
FAST SLOW
RESPONSE
R34
CALIB
INPU T
R33 CALIB
R31 CALIB
R32 CALIB
RESPONSE TIME SWITCH
R16
100 MV REC
R6
Figure 3-1. Amplifier Board Adjustments
Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Operation 3-3
Loading...
+ 62 hidden pages