2-1 Model 951A Controls and Adjustments.................................................... 2-4
3-1 Proper Gas Supply Pressures for Various Levels of Sample NOx........... 3-3
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
DRAWINGS (LOCATED IN REAR OF MANUAL)
619604 Schematic Diagram, Computer Interface
619710 Schematic Diagram, ±15V Power Supply
641871 Schematic Diagram, Temp Control
649819 Schematic Diagram, Amplifier Board
649822 Diagram, Tubing - Model 95A
649834 Schematic Diagram, 951A NO/NOx Analyzer
649835 Pictorial Wiring Diagram, Model 951A NO/NOx Analyzer
649958 Flow Diagram, Model 951A
652423 Schematic Diagram, Power Supply
652834 Schematic Diagram, Power Supply - Thermocooler
652838 Pictorial Wiring Diagram, Low Tempco Option
654348 Schematic Diagram, Hi Voltage Board Assembly
656313 Schematic Diagram, Valve Control Board
780350 Schematic Diagram, 4-20mA 0-5V Option
780726 Wiring Diagram, 4-20mA 0-5V Option
780727 Installation Drawing, Model 951A w/4-20mA 0-5V Option
780809 Wiring Diagram, 4-20mA 0-5V Option and Low Tempco
ONTENTS
C
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
v
ONTENTS
C
N
OTES
vi
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
P
REFACE
S
AFETY SUMMARY
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site
property, all personnel authorized to install, operate and service the Model 955A should
be thoroughly familiar with and str ictly follow the instructions in this manual. Save these
instructions.
If this equipment is used in a manner not specified in these instructions, protective
systems may be impaired.
DANGER is used to indicate the presence of a hazard which will cause severe
personal injury, death, or substantial property damage if the warning is ignored.
WARNING is used to indicate the presence of a hazard which can cause severe
personal injury, death, or substantial property damage if the warning is ignored.
CAUTION is used to indicate the presence of a hazard which will or ca n cause minor
personal injury or property damage if the warning is ignored.
NOTE is used to indicate installation, operation or maintenance information which is
important but not hazard-related.
WARNING: ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to
live parts which can cause death or serious injury. Refer servicing to qualified
personnel.
For safety and proper performance this instrument must be connected to a
properly grounded three-wire source of power.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
P1
REFACE
P
WARNING: ULTRA VIOLET LIGHT HAZARD
UV light from the ozone generator can cause permanent eye damage. DO NOT
LOOK DIRECTLY AT THE UV SOURCE IN THE OZONE GENERATOR. Use of UV
filtering glasses is recommended.
WARNING: OZONE HAZARD
This instrument generates ozone while operating. Ozone is toxic by inhalation
and is a strong irritant to the throat and lungs. Ozone is also a strong oxidizing
agent. Its presence is detected by a characteristic pungent odor.
The instrument EXHAUST outlet contains ozone and nitrogen dioxide which is
toxic by inhalation. The BYPASS outlet contains various oxides of nitrogen, and
if the sample source is from the exhaust of an internal combustion engine, it
may contain unburned hydrocarbons and carbon monoxide which is highly toxic
and, depending on duration of exposure, can cause headache, nausea, loss of
consciousness and death.
Avoid any inhalation of the internally generated ozone, sample, EXHAUST and
BYPASS discharge. Keep all tubing fittings checked for tightness to avoid
internal leaks.
Connect rear panel EXHAUST and BYPASS outlets to outside vent via se parate
lines 1/4 inch (6.3mm) or larger. Use only Teflon or stainless steel tubing.
WARNING: PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect
safety of this product. Use only factory documented components for repair.
CAUTION: AIR FLOW
Do not operate instrument without air flow to the ozonator; plugging of the filter
will result.
CAUTION: HIGH PRESSURE GAS CYLINDERS
P2
This instrument requires use of oxygen and a known standard gas in high
pressure cylinders. Refer to Handling and Storing High Pressure Gas Cylinders
located in the rear of this manual.
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
SPECIFICATIONS
REFACE
P
C
ATALOG NUMBER
R
ANGES
SENSITIVITY
L
INEARITY
R
ESPONSE TIME
S
TANDARD SAMPLE
C
APILLARY
A
UXILIARY SAMPLE
C
APILLARY
P
RECISION
S
TABILITY
Z
ERO
193702 Model 951A NO/NOx Analyzer
Selectable fullscale range of 10, 25, 100, 250, 1000, 2500 and
10,000 parts per million
0.1 ppm on 10 ppm range
±1% of fullscale
The specified linearity is obtainable throughout the operating
range, contingent upon use of an appropriate combination of
oxygen source gas, gas pressure settings and electronic
adjustments.
(E
LECTRONIC PLUS FLOW
Approximately one second on all ranges except 10 ppm.
Approximately three seconds on 10 ppm range.
Five seconds on all ranges.
For such applications as monitoring stack sources, where
comparatively slow response is desired, an internal switch
provides an optional electronic response time of approximately
10 seconds to 90% of fullscale on all ranges.
±5% of fullscale
1% of fullscale in 24 hours
)
S
PAN
D
ETECTOR OPERATING
T
EMPERATURE
R
ECORDER OUTPUT
A
MBIENT TEMPERATURE
E
LECTRICAL POWER
R
EQUIREMENTS
D
IMENSIONS
W
EIGHT
1% of fullscale in 24 hours
Atmospheric
Selectable output of 10 millivolts, 100 millivolts, 1 volt or 5 volts
40°F to 100°F (4.4°C to 37.7°C)
107 to 127 VAC, 50/60 Hz, 1000 watts
9.0 x 17.8 x 22.0 inches (228.6 x 450.9 x 558.8mm) HxWxD
76 lbs (34.5 kg)
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
P3
REFACE
P
C
USTOMER SERVICE
For order administration, replacement Parts, application assistance, on-site or factory
repair, service or maintenance contract information, contact:
, T
ECHNICAL ASSIST ANCE AND FIELD SERVICE
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned
Materials Authorization (RMA) number. Please have the following information when
you call: Model Number, Serial Number, and Purchase Order Number or Sales Order
Number.
Prior authorization by the factory must be obtained before returned materials will be
accepted. Unauthorized returns will be returned to the sender, freight collect.
When returnin g any pro duct o r compon ent t hat has be en expo sed to a toxic, corrosi ve
or other hazardous material or used in such a hazardous environment, the user must
attach an appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification
that the material has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
USA
T
RAINING
A comprehensive Factory Training Program of operator and service classes is
available. For a copy of the Current Operator and Service Training Schedule contact
the Technical Services Depart men t at:
Rosemount Analytical Inc.
Phone: 714-986-7600
FAX: 714-577-8006
D
OCUMENTATION
The following Model 951A NO/NOx Analyzer instruction materials are available.
Contact Customer Service or the local representative to order.
556383 Instruction Manual (this document)
P4
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
C
OMPLIANCES
This product may carry approvals from several certifying agencies, for use in nonhazardous, indoor locations
97-C218
REFACE
P
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
P5
REFACE
P
N
OTES
P6
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
I
NTRODUCTION
1
1.1 OVERVIEW
The Model 951A NO/NOx Analyzer continuously analyzes a flowing gas sample
performing one of two switch selectable determinations:
1. Nitric oxide (NO); or
2. Combined nitric oxide (NO) and nitrogen dioxide (NO,) designated NOx. By
definition [NOx] = [NO] + [NO2].
Typical applications include analyzing vehicular exhaust emissions from internal
combustion engines and monitoring the effluent from stationary (stack) sources
The analyzer utilizes the chemiluminescent method of detection In the nitric oxide
determination, sample NO is quantitatively converted into NO2 by gas-phase oxidation
with molecular ozone produced within the analyzer from air or oxygen supplied by an
external cylinder. A characteristic of this reaction is the elevation of approximately
10% of the NO2 molecules to an electronically-excited state, followed by immediate
reversion to the non-excited state accompanied by emission of photons.
The emitted photons impinge on a photomultiplier detector generating a low-level DC
current. The current is amplified to drive a front panel meter and an accessory
potentiometric recorder if desired.
SPAN
ZERO
PPM RANGE
250 1000
100 2500
25 10000
10 AUTO
NALYZER
CONVERTER
ADJUST
Model 951A N O /NO x A nalyzer
IGURE
F
1-1. M
ON
NO NOx
OFF
Rosemount Analytical
ODEL
951A NO/NOX A
MODE
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
1-1
NTRODUCTION
I
To minimize noise and reduce dark current, the photomultiplier tube is mounted in a
thermoelectrically-cooled housing with temperature held constant at about 59oF
(15oC). Control circuitry is contained in the 652831 Cooler Temperature
Controller/Power Supply Assembly. The power supply circuit provides a high-current
source of DC voltage. A thermistor sensor attached to the housing, and an associated
switching transistor, control a pass transistor, providing closely regulated on-off
control.
Attached to the cooler housing is thermal fuse F3, setpoint 150oF (65oC). This fuse
protects the thermoelectric cooler against the overheating that otherwise could occur
as a result of excessively high ambient temperature or failure of a fan.
CAUTION: EXCESSIVE HEAT
Do not operate this analyzer without the air duct covering the cooling fins of the
thermoelectric cooler. Excessive heat may damage the cooling devices. The
cooler indicator lamp DS1 (mounted on the upper left side of the inner flow
regulator panel and can be viewed through the window) will cycle on and off
until a control point is reached.
Analyzer functioning for the NOx determination is identical to that described above for
the NO determination except that, before entry into the reaction chamber, the sample
is routed through a converter where the NO2 component is dissociated to form NO.
Instrument response is proportional to total NO in the converted sample, that is the
sum of the NO present in the original sample plus the NO produced by dissociation of
NO2.
To minimize system response time an internal sample-bypass feature provides
high-velocity sample flow through the analyzer.
The electronic circuitry is modularized, utilizing plug-in p rinted circuit boards with solid
state components and test jacks for easy troubleshooting and replacement.
If desired, the Model 951A may be factory equipped with various optional features in
addition to the standard features of the basic instrument. Brief descriptions of the
principal options are given in the following section.
1.2 OPTIONS
ANGE
R
I.D.
1-2
This option provides contact closure signals that enable a computer or other external
device to identify the setting of the front panel PPM RANGE Switch. The Model 7D
Thermal Conductivity Analyzer is designed to continuously measure the concentration
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
NTRODUCTION
I
of a single component of interest in a flowing gas mixture. The measurement is based
on the different thermal conductivity's of the individual components of the sample
stream. The method is especially well suited to analysis of two-component sample
streams. However, analysis of multi-component streams is possible if the various
components of the background gas occur in relatively constant ratio, or have similar
thermal conductivity's.
EMOTE RANGE CHANGE
R
For applications where remote operation of the analyzer is desired, as in an emission
test console, the Remote Range Change Option may be used. This option permits
either the operator or a computer to override and disable the front panel MODE and
PPM RANGE Switches and thus to control selection of: (a) NO or NOx mode, and (b)
parts-per-million range.
The unit consists of an electrical plug connector with plug-in logic card and harness for
connection to a rear panel terminal strip on the analyzer.
ANGE
R
AND REMOTE RANGE CHANGE
I.D.
The Range I.D./Remote Range Change Option is a combination of the Range I.D. and
Remote Range Change options. This option is compatible with a user supplied remote
control system employing a 24 VDC digital output and input for analyzer range control
and analyzer range sense, respectively. It is completely integral within the analyzer
and provides a terminal strip on the rear of the analyzer for connections of the user
cable.
AMPLE PUMP
S
The basic Model 951A is designed to accept pressurized samples. To permit analysis
of gases at atmospheric or sub-atmospheric pressure the analyzer may be equipped
with an optional, internally mounted sample pump. This option is not available on Low
Tempco versions of the 951A. An external, accessory pump can be ordered instead.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
1-3
NTRODUCTION
I
N
OTES
1-4
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
I
NSTALLATION
2
2.1 FACILITY PREPARATION
Sections 2.1.1 through 2.1.3 provide information that may be required prior to
installation of the analyzer.
2.1.1 O
2.1.2 L
2.1.3 P
UTLINE AND MOUNTING DIMENSIONS
Significant dimensions are shown in Figure 2-1.
OCATION
Install analyzer in a clean area, not subject to excessive vibration or extreme
temperature variations .
Preferably, the analyzer should be mounted near the sample stream, to minimize
sample transport time. A circuit controlled by a thermal switch holds internal
temperature of the analyzer to the correct operating temperature for ambient
temperatures in the range 40oF to 100oF (4.4oC to 37.7oC). Temperatures outside
these limits necessitate use of special temperature controlling equipment or
environmental protection.
Preferably, the cylinders of air (or oxygen) and span gas should be located in an area
of relatively constant ambient temperature.
OWER REQUIREMENTS
Electrical power requirements are 107 to 127 VAC, 50/60 Hz, 1000 watts.
2.2 UNPACKING
Unpack instrument carefully. Preparatory to shipment, the photomultiplier tube housing
and the sample pump (if instrument is so equipped) are immobilized with hold down
screws, inserted from the bottom of the instrument and marked with red paint for
identification. The hold down screws must be removed prior to operation of the
instrument. In the event the instrument is ever returned to the factory, these screws
must be replaced to ensure safe shipment.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
2-1
NSTALLATION
[
]
[
]
I
19.0
[482]
8.7
[222]
8.7
[221]
7.0
[178]
1.25
[32]
9.0
[228]
4.5
114
2.0
[51]
6.0
[152]
20.2
[513]
18.3
[465]
1.0
[25.4]
1.25
[32]
6.6
168
4.6
[118]
2.8
[70]
DIMENSIONS
Inch
mm
3. ALLOW 16 INCHES IN FRONT OF
INSTRUMENT FOR DOOR SWING.
2. ALLOW 12 INCHES MINIMUM CLEARANCE
ABOVE INSTRUMENT FOR MAINTENANCE..
1. RECOMMENDED MOUNTING HARDWARD:
10-32 MACHINE SCREWS (SUPPLIED BY
CUSTOMER).
IGURE
F
2-1. M
ODEL
951A O
2-2
18.3
[465]
18.3
[465]
18.3
[465]
18.3
[465]
RECOMMENDED PANEL CUTO UT
UTLINE AND MOUNTING DIMENSIONS
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
8.5
[216]
∅.25
[63]
6 PLC'S
NSTALLATION
A
C
U
I
AC POWER
ON/OFF
SWITCH
IGURE
F
2-2. F
PHOTOMULTIPLIER
HIGH VOLTAGE
INTERLOCK SWITCH
RECORDER OUTPUT
SELECT SWITCH
ON
SPAN
ADJUST
250
100
25
10
1000
2500
10000
AUTO
ZERO
ADJUST
SAMPLE PRESSUR E
GAUGE
B
Y
P
A
FLOW
S
BALANCE
S
Rosemount Analytical
MODE
SWITCH
PPM RANGE
SWITCH
CONVERTER
ADJUST
CONVERTER
INDICATOR
RONT PANEL INDICATORS AND CONTROLS
RESPONSE TIME
SELECTOR SWITCH
10
5
FAST SLOW
100
1
RESPONSE
MV
V
REC
OUTPUT
R22
R34
ALIB
R16
R33
R31
R6
INP
T
R32
BYPASS NEEDLE VALVE
B
Y
P
A
FLOW
S
BALANCE
S
OZONE
OZONE
OZONE
OZONE
OZONE
PRESSURE
GAUGE
OFF
BYPASS
FLOWMETER
BYPASS
FLOWMETER
OZONE PRESSURE
REGULATOR
OZONE OFF
PUSHBUTTON
SWITCH
OFF
IGURE
F
2-3. C
THERMOCOUPLE
CONNECTOR TB9
IGURE
F
2-4. R
P
IR INLET
EXHAUST
Rosemount Analytical
FLOW BALANCE
NEEDLE VALVE
Model 951A NO/NOx Analyzer
SAMPLE BACKPRE SSURE
REGULATOR
ONTROLS AND ADJUSTMENTS LOCATED BEHIND SWING OUT FRONT
ANEL
SAMPLE INLET
RECORDER CONNECTION TB4-1
RECORDER CONNECTION TB4-2
RANGE I.D.
CONNECTOR J10
(Option)
CABLE GLAND FOR
RECORDER CABLE
AC POWER CABLE
FUSE F1FUSE F2
EAR PANEL
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
2-3
NSTALLATION
I
CONTROLFUNCTION
ON/OFF
MODE S
WITCH
METER
PPM RANGE S
CONVERTER
ADJUST
ZERO
SPAN
SAMPLE
BACKPRESSURE
REGULATOR AND
PRESSURE GAUGE
BYPASS F
AND
N
OZONE
REGULATOR AND
GAUGE
OZONE OFF
LOWMETER
EEDLE VALVE
PRESSURE
SWITCH
WITCH
Controls AC power to all components in instrument
For selection of NO or NOx
Readout of NO or NOx in ppm as selected with MODE switch.
Upper scale is graduated 0 to 100 for use with 10, 100, 1000,
and 10,000 ppm ranges. Lower scale is graduated 0 to 25 for
use with 25 and 2500 ppm ranges.
Select fullscale range for meter and recor der
LED illuminates during application of power to converter
heater. When Converter reaches temperature equilibrium,
LED will go off and on at intervals of about 2 seconds,
indicating correct temperature control.
Through-panel screwdriver adjustment of converter
temperature. Should be set for temperature that yields
optimum combination of high efficiency for the NO2 to NO
conversion and extended life for the catalytic converter.
Optimum temperature differs from one instrument to another,
see Section 3.3.
Set zero point on meter scale or recorder chart. With MODE
switch in NO position and zero air supplied to SAMPLE inlet,
ZERO Control is adjusted for zero reading.
Set upscale calibration point on meter scale or recorder chart.
MODE switch is at NO or NOx and suitably pressurized
standard gas of accurately known NO/NOx content is supplied
to SAMPLE inlet. With PPM RANGE switch set to the
appropriate range for the span gas, the SPAN control is
adjusted to correct reading on meter or recorder.
Adjustment and indication of pressure (and therefore flow) of
sample or standard gas routed through sample capillary, and
into reaction chamber. Proper setting dependent on operating
range. See Table 3-1.
Bypass flow indication and adjustment for sample or standard
gas to SAMPLE inlet. Setting of 2 L/min. is recommended to
ensure rapid response and optimum functioning of sample
backpressure regulator
Pressure adjustment and indication for air or oxygen supplied
to AIR inlet for use as oxygen source for internal ozone
generator. Oxygen is usable for all ranges. Air is usable only
for 2500 ppm range or lower. Proper pressure setting
dependent on operating rang e, see Tabl e 3- 1.
Removes AC power to ultraviolet source lamp in ozone
generator. Power is OFF when switch indicator is red.
ABLE
T
2-4
2-1. M
PAGE
ODEL
)
951A C
ONTROLS AND ADJUSTMENTS (CONTINUED ON NEXT
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
CONTROLFUNCTION
FLOW BALANCE
NEEDLE VALVE
RESPONSE TIME
SELECTOR SWITCH
RECORDER OUTPUT
SELECTOR SWITCH
M
ETER MECHANICAL
ZERO
NSTALLATION
I
Use to equalize pressure drop for NO and NOx legs of the
internal flow system. Criterion for correct adjustment is an
NO2 free nitric oxide standard gas should give equal readings
of ppm for NO and NOx modes. An alternate flow balance
check , if NO2 free standard gas is unavailable, is to verify that
the reading on the SAMPLE pressure gauge is the same for
NO and NOx modes.
Select either FAST or SLOW electronic response. When set
to FAST, electronic response time (for 0 to 90% of fullscale) is
then approximately one second for all ranges except 10 ppm,
which is approximately three seconds. When set to SLOW (as
in monitoring stack sources), electronic response time 0 to
90% of fullscale) is approximately ten seconds for all ranges.
To select output of 10 mV, 100 mV, 1V or 5 V for a
potentiometric recorder.
Located on rear of meter. With AC power OFF, meter should
read zero. If not use screw adjustment on meter.
P
ABLE
T
HOTOMULTIPLIER HIGH
V
OLTAGE INTERLOCK
S
WITCH
2-1. (
CONTINUED FROM PREVIOUS PAGE
When analyzer front panel is open, photomultiplier high
voltage power supply is automatically shut off.
)
2.3 GAS REQUIREMENTS
WARNING: HIGH PRESSURE GAS CYLINDERS
This instrument requires use of oxygen and a known standard gas in high
pressure cylinders. Refer to Handling and Storing High Pressure Gas Cylinders
located in the rear of this manual.
IR OR OXYGEN
A
This is used as both (a) oxygen source for generation of the ozone required for the
chemiluminescent reaction, and (b) standard gas for zero calibration. Gas for both
purposes may be supplied from a single cylinder and routed through a tee.
Alternatively, two separate cylinders may be used.
Oxygen is usable in all applications. Air is suitable for the oxygen source only if the
desired fullscale operating range is 2500 parts-per-million or less. Breathing grade
oxygen or air is recommended. Clean, dried ambient air containing less than 0.1
parts-per-million nitric oxide may be used, provided that its de wpoint is below -10oF (23oC). If air is insufficiently dried, or contains excessive nitric oxide, instrument
response will not be linear to 2500 ppm.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
2-5
NSTALLATION
I
PAN GAS
S
This is a standard gas of accurately known composition, used to set an upscale
calibration point.
Alternative span gases are:
1. The usual span gas is NO in a background of nitrogen.
2. For span check or adjustment in the NOx mode, the span gas may be NO2 in a
background of air or nitrogen.
3. For convenient calibration in either the NO or NOx mode, the span gas may be a
cylinder gas mixture consisting of known concentrations of both NO and NO2 in a
background of nitrogen.
Note
For maximum calibration accuracy, the concentration of NO and/or NO2 in the
span gas should be as near as possible to that in the sample gas. Also, the span
gas should be supplied to the rear panel SAMPLE inlet at the same pressure as
the sample gas. To ensure constant pressure, a pressure regulator may be
utilized immediately upstream from the SAMPLE inlet.
Preferably, each gas used should be supplied from a tank or cylinder equipped with a
clean, non-corrosive type two stage regulator. In addition, a shut-off valve is desirable.
If possible, install the gas cylinders in an area of relatively constant ambient
temperature.
2.4 SAMPLE REQUIREMENTS
The Sample must be relatively clean and dry before entering the analyzer. In general,
before admission to the analyzer, the sample should be filtered to 2 microns and
should have a dewpoint below 90oF (32oC). The factory can provide technical
assistance if desired .
Proper supply pressures for sample and span gases depend on whether or not the
analyzer is equipped with the optional 632748 Sample Pump:
For the basic Model 951A, without sample pump, sample must be supplied to the
SAMPLE inlet at a pressure of 5 to 10 psig (34.5 to 69 kPa). This ensures that the
normal bypass flow of two liters per minute will be obtainable by adjustment of the
BYPASS Needle Valve. (Proper bypass flow is essential for rapid system response
and stable flow into reaction chamber.)
For an analyzer equipped with sample pump, the acceptable pressure range at the
SAMPLE inlet is 0 to 5 psig (0 to 34.5 kPa). The pump pressurizes the sample to
between 5 and 10 psig (34.5 to 69 kPa) for supply to the internal flow system.
2-6
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
2.5 GAS CONNECTIONS
1. Remove plugs and caps from all inlet and outlet fittings. See Figure 2-4.
2. Connect EXHAUST outlet to outside vent via tubing with O.D. of 1/4 inch (6.3mm)
or larger.
3. Connect external lines from air (or oxygen) cylinder and sample source to
corresponding rear panel inlet ports. For sample line, stainless steel tubing is
recommended.
4. Adjust regulator on air (or oxygen) cylinder for output pressure of 35 to 40 psig
(251 to 276 kPa).
Supply sample gas to rear panel SAMPLE inlet at appropriate pressure: 5 to 10 psig
(34.5 to 69 kPa) for basic analyzer without sample pump; 0 to 5 psig (0 to 34.5 kPa)
for analyzer with pump.
2.6 ELECTRICAL CONNECTIONS
NSTALLATION
I
2.6.1 R
ECORDER OUTPUT
If a recorder is used, connect leads to terminals marked REC 1 (+) and 2 (-) on TB4
(Figure 2-4). (Note that, within the analyzer, the negative recorder output terminal is
connected to ground.)
Set REC OUTPUT Selector Switch SW1 (Figure 2-3) to the recorder span: 10 mV, 100
mV, 1 VDC or 5 VDC.
2.6.2 T
HERMOCOUPLE CONNECTIONS FOR MEASURING TEMPERATURE OF
NO C
ONVERTER
Temperature of the NO2 to NO converter may be monitored by connecting a customer
supplied millivolt meter to thermocouple connector TB9 located on the rear panel
(Figure 2-4). The position of the thermocouple in the converter bundle will influence
the actual temperature readout.
2.6.3 R
EMOTE RANGE CHANGE
The Remote Range Change Kit, Figure 2-5, consists of a plug-in Adapter Board plus
attached harness and rear terminal plate.
The Adapter Board has two two-position slide switches: the MAN/AUTO MODE Switch and
the MAN/AUTO RANGE Switch.
NO
2
TO
The MAN/AUTO MODE Switch provides the choice of local or remote control of the
NO/NOx Mode Switching Solenoid Valve, assoc iated with the NO2 to NO converter. With
switch at MAN, the function is under control of the front panel MODE Switch. With switch at
AUTO, the front panel MODE Switch is disabled; the function is then remotely controlled, by
either the operator or a computer, via contact closure signals applied to terminals on TB5.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
2-7
NSTALLATION
N
I
The MAN/AUTO RANGE Switch provides the choice of local or remote selection of
operating range. With switch at MAN, range selection is under control of the front panel
PPM RANGE Switch. With switch at AUTO, the front panel PPM RANGE Switch is
disabled; range selection is remotely cont rolled, by either the operator or a computer, via
contact clos ure sign als appli ed t o ter min als o n TB6.
Control at TB5 and TB6 is accomplished by a contact closure from the terminal marked
COM (for common) to the terminal marked with the name of the desired function. Contact
closure requirements are 25 mA at 15 VDC.
FAST SLOW
5
10
RESPONSE
1
100
V
MV
OUTPUT
Computer Adapter
Board
Auto/Man Range
Switch
REC
R22
R34
CALIB
R16
R33
INPUT
R31
R32
MODE
AUTO
MA
RANGE
AUTO
MAN
R6
Rosemount Analytical
Electronics
Assembly
DC Harness (J1)
Auto/Man Mode
Switch
NOx
SPAN
ZERO
1000
2500
10,000
10
25
100
IGURE
F
IGURE
F
2-5. R
2-6. T
Computer Adapter Board installed on
Interconnect Board of the Electronics
Terminal Assembly mounted
on rear panel of Model 951A
Assembly
EMOTE RANGE KIT INSTALLED
TERMINAL
ASSEMBLY
Remote MODE
SELECT SWITCH
NOx
SPAN
ZERO
COM
100
250
10
25
Remote RANGE
SELECT SWITCH
1000
2500
10,000
COM
With all contacts of Range Select Switch
open, analyzer is ion 250 ppm range.
YPICAL INTERCONNECTION OF REMOTE RANGE KIT
2-8
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
NSTALLATION
I
2.6.4 C
ONNECTIONS OF THE RANGE
I. D. K
IT
The Range I.D. option provides contact closure signals that permit a computer or other
external device to identify the position selected manually with the front panel PPM
RANGE Switch. The cable is connected to PPM RANGE Switch SW1 and extends to
connector J10 mounted on the rear of the case (Figure 2-4).
The pin-out connections of J10 are as follows:
Pin A 10 ppm
Pin B 25 ppm
Pin C 100 ppm
Pin D 250 ppm
Pin E 1000 ppm
Pin F 2500 ppm
Pin H 10000 ppm
Pin J Wiper
2.6.5 P
OWER CONNECTIONS
Connect power cord to an AC source of 107 to 127 volts, 50/60 Hz. If power outlet
does not have third (ground) contact, use an adapter to provide proper grounding.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
2-9
NSTALLATION
I
N
OTES
2-10
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
S
TARTUP
3
Preparatory to startup and operation, a thorough familiarization with Table 2-1 and Figures
2-2, 2-3 and 2-4 is recommended. These figures give locations and brief descriptions of
operating controls. For more detailed information on controls, refer to Section Five.
3.1 STARTUP PROCEDURE
Following are detailed stepwise instructions on startup and calibration. For
convenience, condensed instructions for startup, routine calibration and normal
operation are provided at the front of this manual.
1. W ith power removed from analyzer, check front panel meter. It should read zero; if
not, adjust Mechanical Zero Screw at rear of meter for zero reading.
2. Place front panel PPM RANGE Switch at 1000.
WARNING: OZONE HAZARD
When instrument power is on, the ultraviolet source sample is energized,
converting a portion of the oxygen contained with the ozonator into ozone. With
normal flow through the ozonator, the ozonized air or oxygen is continuously
swept though the ozonator and into the reaction chamber. If flow is stopped,
however, the ozone will diffuse and will attack the sintered metallic restrictor
element in the tee fitting at the upstream end of the ozonator. Operation of the
analyzer with the restrictor element thus rusted will result in the following
consequences: Reduced flow of air or oxygen through the ozonator, ozone
deprivation within the reaction chamber and non-linear response of the analyzer
to NO/NOx.
To prevent such damage, verify that the front panel ozone ON/OFF switch is
turned off if flow of feed gas to the air inlet is terminated.
3. Set front panel MODE Switch at NO or NOx. Place POWER Switch at ON.
Electrical power is now being supplied to all circuits, including sample pump if
analyzer is so equipped. Analyzer will now require approximately one hour for
temperature equilibration before ready for calibration.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
3-1
TARTUP
S
4. Establish correct pressure for air oxygen:
a. Verify that pressure regulator on cylinder of air or oxygen is set for supply
pressure of 35 to 40 psig (242 to 276 kPa).
a. On internal gas control panel, Figure 2-3. adjust OZONE Pressure Regulator so
that OZONE Pressure Gauge indicates either 20 psig (138 kPa) or 30 psig (207
kPa), depending on the desired operating range.
5. Establish correct flow of sample gas:
a. Supply sample gas to rear panel SAMPLE inlet.
b. Adjust SAMPLE Backpressure Regulator so SAMPLE Pressure Gauge
indicates the value appropriate to the desired operating range.
c. Adjust BYPASS Needle Valve for reading of two liters per minute on BYPASS
Flowmeter.
Note
Inability to obtain a bypass flow of two liters per minute by adjustment of the
BYPASS Needle Valve usually indicates insufficient sample supply pressure at
the SAMPLE inlet.
6. Establish correct flow of zero air:
a. Supply zero air to rear panel SAMPLE inlet.
b. Note reading on SAMPLE Pressure Gauge. It should be the same as in Step
5b. If not, adjust output pressure regulator on air cylinder as required.
7. Establish correct flow of upscale standard gas:
a. Supply upscale standard gas to rear panel SAMPLE inlet.
b. Note reading on SAMPLE Pressure Gauge. It should be the same as in Step
5b. If not, adjust output pressure regulator on cylinder of upscale standard gas
as required.
Note
Supply pressures for sample and upscale standard gases must be the same;
otherwise, readout will be in error.
The Analyzer is now ready for calibration per Section 3.2.
3-2
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
TARTUP
pp
pp
pp
S
ABLE
T
Max. NOx Level
(ppm)
1000Air
2500Air
10,000Oxygen
3-1. P
ROPER GAS SUPPLY PRESSURES FOR VARIOUS LEVELS OF SAMPLE
NO
Gas Supplied
to Air Inlet
X
3.2 CALIBRATION
1. Zero Calibration.
a. Set PPM RANGE Switch for the same range that will be used during sample
analysis. Set SPAN Control at about midrange.
b. Supply zero gas to rear panel SAMPLE inlet.
c. Adjust ZERO Control for zero reading on meter or recorder, then lock ZERO
Control knob.
Ozone
Pressure Gauge Setting
20 psig (138 kPa)
Provides flow of a
cc/min to ozone generator
30 psig (207 kPa)
Provides flow of a
cc/min to ozone generator
30 psig (207 kPa)
Provides flow of a
cc/min to ozone generator
rox. 500
rox. 1000
rox. 1000
Sample
Pressure Gauge Setting
4 psig (27.6 kPa)
Provides flow of approx. 60
cc/min to reaction chamber
1.5 psig (10.3 kPa)
Provides flow of approx. 20
cc/min to reaction chamber
2. Upscale Calibration.
a. Set PPM RANGE Switch at the position appropriate to the particular span gas.
b. Supply upscale standard gas of accurately known NO/NOx content to rear
panel SAMPLE inlet.
c. Place MODE Switch at NO if nitric oxide span gas is used.
d. Adjust SPAN Control so that reading on meter or recorder is equal to the known
parts-per-million concentration of NO or NOx in the span gas. In operation at
NOx levels in the range of 2500 to 10000 parts-per-million, the correct reading
may not be obtainable, initially, by adjustment of the SPAN Control. The cause
is that the reduced sample flow required for linearity results in lowered
sensitivity and slower response. To compensate for these effects, make the
electronic adjustments of Steps (e) and (f) below.
e. If necessary, increase sensitivity by raising photomultiplier voltage per Section
6.1, Step 5.
f. To reduce observed noise, select SLOW position of Response Time Selector
Switch SW2, Figure 3-1.
g. When correct upscale reading is obtained, loc k SPAN Control knob.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
3-3
TARTUP
S
Calibration is now complete. Before Placing analyzer in operation, however, measure
efficiency of the NO2 to NO converter per Section 3.3.
FAST SLOW
5
10
100
MV
REC
R22
R6
1
V
OUTPUT
INPUT
RESPONSE
R34
CALIB
R33
CALIB
R31
CALIB
R32
CALIB
RESPONSE TIME
SWITCH
R16
IGURE
F
3-1. A
MPLIFIER BOARD ADJUSTMENTS
3.3 MEASURING EFFICIENCY OF NO2 TO NO CONVERTER AND
ADJUSTING TEMPERATURE SETPOINT
It is the responsibility of the user to measure efficiency of the NO2 to NO converter
during initial startup and thereafter at intervals appropriate to the application, normally
once a month.
The reactant material used in the converter provides the optimum combination of high
conversion efficiency and low ammonia interference. Unlike most competitive
analyzers, the Model 951A utilizes a reactant material that gradually becomes moreefficient at a given temperature , and thus after a period of use may permit operation
with a lower temperature setpoint than that required initially.
Conversion efficiency in the Model 951A is typically 95% to 98% +, considerably above
the 90% minimum accepted by the EPA. Refer to 40 CFR 60, App. A, Method 20,
Section 5.6.
3.3.1 T
EST SETUP FOR MEASUREMENT OF CONVERSION EFFICIENCY
3-4
A typical setup for measurement of conversion efficiency is shown in Figure 3-2A. The
test setup includes:
1. A cylinder of nitric oxide standard gas consisting of NO in N2. The concentration of
NO in the standard gas should be about the fullscale value of the range under test.
The test sample supplied to the analyzer should contain a concentration of NO
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
TARTUP
S
comparable to that in the samples that are to be analyzed. Alternately, a higher
concentration NO standard may be used if the test setup includes provision for
diluting it appropriately with zero air. Suitable standard gases are available from
various suppliers. Stainless steel cylinders are commonly used, but specially
treated aluminum is preferre d for low parts-per-million NO samples.
2. An ozone generator utilizing an ultraviolet lamp, not a corona discharge. A corona
discharge ozone generator is undesirable as it may produce oxygen atoms, which
can then combine with atmospheric nitrogen to form NO. The result can be an
erroneously high value for the measured conversion efficiency.
3.3.2 T
EST PROCEDURE
1. Measure converter temperature by connecting a millivolt meter to terminals 5 (+)
and 6 (-) on TB9. Within the analyzer, these terminals are connected to a Type J
thermocouple. Note present reading as a reference for comparison with
subsequent measurements. The optimum temperature differs from one analyzer to
another. but typically is in the range of 660oF to 750oF (350oC to 400oC).
2. Turn front panel CONVERTER ADJUST one turn counterclockwise, to lower
temperature setpoint slightly from the original factory setting. Wait 15 minutes for
temperature equilibration, then proceed to Step 3.
3. Refer to Figure 3-2. Connect the Model 958 NO2 Converter Efficiency Tester to the
Model 951A, and use the following procedure, adapted from 40 CFR 60, to
determine the values to be used in Equation (A):
a. Attach the NO/N2 supply at C2. the air supply at C1, and the analyzer inlet
connection at C3.
b. With the variable transformer off, place Model 951A in NO mode and close
valve MV1. Open valve MV2 until Model 951A SAMPLE Pressure Gauge
reaches operating pressure and BYPASS Flowmeter indicates some bypass
flow. Wait until stable readings are obtained at analyzer. Zero and span the
analyzer output to indicate the value of the NO concentration being used. This
value should be approximately 80% of fullscale. Record this concentration.
c. Open valve MV1 (air supply metering valve) and adjust to blend enough air to
lower the NO concentration (ii) about 10%. Record this concentration.
d. Turn on the ozonator and increase its supply voltage until the NO concentration
of (iii) is reduced to about 20 percent of (ii). NO2 is now being formed from the
NO + O3 reaction. There must always be at least 10 percent un-reacted NO at
this point. Record this concentration.
e. W hen a stable reading has been obtained from (iv), place Model 951A in NOx
mode. It will now indicate the total NOx concentration. Record this
concentration.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
3-5
TARTUP
S
f. Turn off the ozonator and allow the analyzer reading to stabilize. This reading is
the total NOx concentration of the dilute NO span ga s used at step (iii). Record
this concentration.
f. Close valve MV1. The NO concentration should be equal to or greater than the
reading of (ii). Indicating whether the NO contains any NO2.
g. Calculate the efficiency of the NO converter by substituting the concentrations
obtained during the test into the following equation.
Equation (A)
% Efficiency = 1 + X 100
v-vi
iii-iv
In the example of Figure 3-2B,
% Efficiency = 1 + X 100% = 92%
85-90
80-20
Efficiency checks should be made on each analyzer range, using an NO span
gas concentration appropriate to the instrument range.
Note
In the initial measurement, following lowering of the temperature
setpoint in Step 2, the efficiency will normally be less than 92%.
4. Turn CONVERTER ADJUST one quarter turn clockwise, thus raising temperature
setpoint slightly. Wait 15 minutes for temperature equilibration. Again measure
conversion efficiency per Step 3; it should be somewhat improved.
3-6
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
A. TYP ICA L TES T S E TUP
FLOWMETER
FM2
METERING
VALVE MV2
115
VAC
OFF
ON
VARIABLE
TRANSFORMER
T1
CONNECTOR C3
OZONATOR
UTILIZING
ULTRAVIOLET
LAMP
FLOWMETER
FM1
TO SAMPLE
INLET OF
MODEL 951A
FLOW APPROXIMATELY
3 LITERS PER MINUTE
TARTUP
S
CONNECTOR
C2
NO IN N2 BA CKGROUN D
B. TYPICAL TEST RESULTS
NO in
2
N
90
85
80
c
AIR ADDED
c-d
OZONATOR OFF
OZONATOR
ON
METERING
VALVE MV1
CONNECTOR
C2
CONNECTOR
C1
a-b
a
b
% Efficiency = 100
-
1+
-
20
0
IGURE
F
556383-Y Rosemount Analytical July 2000
3-2. M
EASURING EFFICIENCY OF
d
NO
TO NO
MODE
TO
2
X
NO C
ONVERTER
Model 951A NO/NOx Analyzer
3-7
TARTUP
S
5. Repeat Step 4 until either (a) 95% efficiency is attained, or (b) the final one quarter
turn clockwise adjustment yields an increase in efficiency of less than 1%. Then
proceed to Step 5.
6. Turn CONVERTER ADJUST one-eighth turn counterclockwise, thus lowering
temperature setpoint very slightly. Converter temperature is now set at a point on
the front edge of the plateau on the efficiency-vs.-temperature curve, Figure 3-3.
This setting should provide the optimum combination of high conversion efficiency
and low ammonia interference. Wait about fifteen minutes for temperature
equilibration, then again measure converter temperature as in Step 1. Compare
present temperature with original value. Normally, converter temperature should be
in the range of 660oF to 750oF (350oC to 400oC).
Note
Converter temperature is not a direct measure or conversion efficiency.
Temperature measurement is for reference purposes only.
IGURE
F
3-3. C
T
About
95%
Final 1/4 Turn
Clockwise adjustment of
CONVERTER ADJUST
Final 1/8-Turn
Converter
Efficiency
Converter Temperature
ONVERSION EFFICIENCY AS A FUNCTION OF CONVERTER
EMPERATURE
Counterclockwise Adjust men t
of CONVERTER ADJUST
3-8
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
TARTUP
S
3.3.3 S
UBNORMAL CONVERSION EFFICIENCY
If a measured conversion efficiency of 95% is unobtainable within the normal
temperature range, the most probable cause is depletion of the catalytic material
within the converter. However, before concluding that the converter is defective, make
sure that the conversion efficiency measurement is accurate. Even though the
measured eff iciency is less th an 95%, the actual efficiency may be somewhat higher.
An apparent subnormal efficiency can be due to a problem external to the analyzer,
located either within the test setup, or between it and the analyzer. Check for the
following:
1. Leakage.
2. Loss of NO2 between test setup and analyzer. Such loss can occur by reaction with
a rubber diaphragm in a pressure regulator or flow controller. Stainless steel
diaphragms are preferred. Loss can also occur during passage through filter
media.
3.3.4 R
EPLACEMENT OF CONVERTER
If it is determined that the subnormal conversion efficiency is real, and not due to
measurement error introduced by the test setup, the converter must be replaced per
Section 6.2.3. The usual cause of converter failure is destruction of a large part of the
catalytic material by excessive heat, due either to excessively high temperature
setpoint or failure of the converter temperature control circuit.
3.3.5 C
APILLARIES
Replacement capillaries shou ld be installed finge r-tight. Use of a wrench can constrict
capillaries, changing flow rate.
3.3.6 TEA S
CRUBBER
The TEA Scrubber accessory (PN 635741) can be used to remove residual NO2 from
the NO cylinders. Use of this accessory allows a true calibration of the analyzer.
White crystal deposits on the windows of the reaction chamber and plugging of
capillaries and vent are usually due to sample contaminates reacting with the high
ozone levels. To eliminate the contaminates, the sampling system should be reworked
or a preventive maintenance program developed (if dropout is not excessive). Another
source of crystalline formation is contaminated air.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
3-9
TARTUP
S
N
OTES
3-10
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
O
PERATION
4
4.1 ROUTINE OPERATION
After calibrating analyzer per Section 3.2, supply sample to SAMPLE inlet. Place
MODE Switch in NO or NOx position, depending on the desired determination. Set
PPM RANGE Switch in appropriate position. The instrument will now continuously
analyze the sample stream.
The Model 951A is designed for continuous operation. Normally, it is never turned off
except for servicing or for a prolonged shutdown. However, if the instrument is to
remain idle for an extended period with power on, the recommended practice is to
leave the MODE Switch in NO position to de-energize the NO/NOx Mode Switching
Solenoid Valve.
4.2 RECOMMENDED CALIBRATION FREQUENCY
After initial startup, or startup following a shutdown, the analyzer requires about one
hour for stabilization before it is ready for calibration. Maximum permissible interval
between calibrations depends on the analytical accuracy required, and therefore
cannot be specified. It is recommended that initially the instrument be calibrated at
least once every eight hours, and that this practice be continued until experience or
requirements indicate that some other interval is more appropriate.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
4-1
PERATION
O
N
OTES
4-2
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
5.1 PRINCIPLES OF OPERATION
T
HEORY
5
5.1.1 N
5.1.2 NOX D
ITRIC OXIDE DETERMINATION BY CHEMILUMINESCENCE METHOD
The chemiluminescence method for detection of nitric oxide (NO) is based on its
reaction with ozone (O3) to produce nitrogen dioxide (NO2) and oxygen (O2). Some of
the nitrogen dioxide molecules thus produced are initially in an electronically excited
state (NO2*). These revert immediately to the ground state, with emission of photons.
The reactions involved are:
NO + O3 → NO2* + O
NO2* → NO2 + hv
where
h = Planck's constant
v = frequency, Hz
As NO and O3 mix in the reaction chamber, the chemiluminescence reaction produces
light emission that is directly proportional to the concentration of NO. This emission is
measured by a photomultiplier tube and associated electronic circuitry. Refer to
Section 5.3.
ETERMINATION
2
The NOx determination is identical to the NO determination described in Section 5.1.1
except that, prior to entry into the reaction chamber, the sample is routed through a
converter where the NO2 component is converted into NO. The reaction is:
2NO2 → 2NO + O
Instrument response is proportional to total NO in the converted sample, that is, to the
sum of the NO originally present in the sample plus the NO resulting from conversion
of NO2. This sum of NO and NO2 is commonly termed NOx, i.e., ([NOx] = [NO] +
[NO2]).
556383-Y Rosemount Analytical July 2000
2
Model 951A NO/NOx Analyzer
5-1
HEORY
T
5.1.3 O
ZONE GENERATION
Ozone for the chemiluminescence reaction is produced in a flow chamber where a
stream of air or oxygen from an external cylinder is exposed to ultraviolet radiation
from a source lamp. The reaction is:
hv 3O2 → 2O
3
5.2 ANALYZER FLOW SYSTEM
The analyzer flow system is shown schematically in Figure 5-1 and pictorially in Figure
6-1. Its basic function is to deliver regulated flows of sample or calibration gas and
ozonized air or oxygen to the reaction chamber. The discharge from the reaction
chamber exits the analyzer via the EXHAUST outlet.
OZONE
Pressure
Gauge
Restrictor
Bypass Valve
Air
or
Oxygen
AIR Inlet
SAMPLE
Inlet
OZONE
Pressure
Regulator
Flow
Controller
(option)
Sample
Pump
(option)
Ozone
Generator
NO
NO
Converter
Flow
Balance
Valve
End-On
Photomultiplier
Ozonized Air or Oxygen for
Chemiluminescent Reaction
Reaction
Chamber
NO/NOx Mode
Switching
2
to
Solenoid Valve
K3
Sample
Pressure
Gauge
Sample
Capillary
Pump Capillary
Window
Tube
Signal to
Electronic
Circuitry
Bleed Capillary
Bypass
Flowmeter
Sample
Backpressure
Regulator
IGURE
F
5.2.1 F
5-1. S
LOW OF SAMPLE OR STANDARD GAS TO REACTION CHAMBER
CHEMATIC FLOW DIAGRAM OF MODEL
951A
Suitably pressurized sample or standard gas is supplied to the rear panel SAMPLE
inlet.
Within the analyzer, flow rate of the selected gas into the reaction chamber is
controlled by a back-pressure regulator. It provides an adjustable, controlled pressure
on the upstream side, where gas is supplied to the calibrated, flow-limiting sample
Exhaust
5-2
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
HEORY
T
capillary. The regulator is adjusted for appropriate reading on the SAMPLE Pressure
Gauge. For operation at NOx levels below 2,500 parts-per-million, correct setting on
the SAMPLE Pressure Gauge is 4 psig (27.6 kPa), resulting in a flow of approximately
60 cc/min to the reaction chamber. For operation at NOx levels in the range of 2,500 to
10,000 parts-per-millio n, se ttin g on SAMP LE Pressure Gau ge shou ld b e 1.5 p sig (10 .3
kPa), resulting in a flow of approximately 20 cc/min to the reaction chamber.
Gas in excess of the flow required for the reaction chamber is discharged through the
BYPASS Flowmeter and out the BYPASS outlet. Bypass flow should be two liters per
minute to ensure optimum functioning of the SAMPLE Pressure Regulator and rapid
system response. Total flow of sample or standard gas into the SAMPLE inlet is
adjusted by means of the BYPASS Needle Valve (or optional flow controller).
Excessive changes, on the order of ±5 psig (±34.5 kPa), in the pressure of the sample
or standard gas will affect the bypass flow rate and can affect accuracy.
5.2.2 F
LOW OF AIR OR OXYGEN
WARNING: OZONE HAZARD
When instrument power is on, the ultraviolet source sample is energized,
converting a portion of the oxygen contained with the ozonator into ozone. With
normal flow through the ozonator, the ozonized air or oxygen is continuously
swept though the ozonator and into the reaction chamber. If flow is stopped,
however, the ozone will diffuse and will attack the sintered metallic restrictor
element in the tee fitting at the upstream end of the ozonator. Operation of the
analyzer with the restrictor element thus rusted will result in the following
consequences: Reduced flow of air or oxygen through the ozonator, ozone
deprivation within the reaction chamber and non-linear response of the analyzer
to NO/NOx.
To prevent such damage, verify that the front panel ozone ON/OFF switch is
turned off if flow of feed gas to the air inlet is terminated.
Suitably pressurized air or oxygen from an external cylinder is supplied to the rear
panel AIR inlet. Immediately downstream from the inlet, a regulator and associated
gauge provide pressure adjustment and indication for air or oxygen. Proper pressure
setting. and the corresponding flow to the ozone generator, will depend on sample
NOx level. Refer to Table 3-1.
1. For operation at NOx levels below 1,000 parts-per-million, correct air pressure
setting on OZONE Pressure Gauge is 20 psig (138 kPa), resulting in a flow of
approximately 500 cc/min of air to the ozone generator.
2. For operation at NOx levels in the range of 1,000 to 2,500 parts-per-million, airpressure setting on OZONE Pressure Gauge should be 30 psig (207 kPa),
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
5-3
HEORY
T
resulting in a flow of approximately 1 liter per minute of air to the ozone generator.
3. For operation at NOx levels in the range of 2,500 to 10,000 parts-per-million,
oxygen pressure setting on OZONE Pressure Gauge should be 30 psig (207 kPa),
resulting in a flow of approximately 1 liter per minute of oxygen to the ozone
generator.
Within the ozone generator, a portion of the oxygen in the stream is converted into
ozone by exposure to ultraviolet radiation from a lamp source. From the generator,
the flow passes into the reaction chamber for use in the chemiluminescence
reaction.
5.2.3 F
LOW SYSTEM OPERATING MODES
Downstream from the BYPASS Needle Valve, the sample or standard gas flows
through either the FLOW BALANCE Needle Valve or the NO2 to NO converter,
depending on the status of NO/NOx Mode Switching Solenoid Valve K3. Status of this
valve depends on whether the front panel MODE Switch is in NO or NOx position.
Operation for NO and NOx determination is as follows:
NO D
ETERMINATION
Sample or standard gas is supplied to the rear panel SAMPLE inlet, passes in turn
through the BYPASS Needle Valve, NO/NOx Mode Switching Solenoid Valve K3, the
sample capillary, and on to the reaction chamber. Since the NO2 to NO converter is
bypassed, any NO2 component originally present in the sample remains in the NO
form and does not contribute to the output signal. (During startup, the FLOW
BALANCE Needle Valve is adjusted to equalize flow rated for NO and NOx flow
modes.)
NOX D
ETERMINATION
Sample or standard gas supplied to the rear panel SAMPLE inlet passes in turn
through the BYPASS Needle Valve, the NO2 to NO converter, the solenoid valve, the
sample capillary and on to the reaction chamber. With the sample now routed through
the converter, the NO2 component is converted into NO prior to entry into the reaction
chamber, and thus will contribute to the output signal.
2
5-4
During startup, the FLOW BALANCE Needle Valve is adjusted to equalize flow rates
for NO and NOx flow modes. Criterion for correct adjustment is that an NO2 free nitric
oxide standard gas should give equal readings of parts-per-million for NO and NOx
modes.
An alternate method of checking flow balance, usable if NO2 free standard gas is
unavailable, is to verify that the reading on the SAMPLE Pressure Gauge is the same
for NO and NOx modes. If not, adjust FLOW BALANCE Needle Valve as required.
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
HEORY
T
5.2.4 C
ONVERTER BLEED FLOW
Even in converter bypass condition (i.e., de-energized state of the solenoid valve), a
flow of approximately 100 cc/min of sample or standard gas passes continuously
through the converter and the bleed capillary. This bleed flow pro vides f aster respo nse
and ensures that the sample within the converter is current. Bleed flow need not be
precisely 100 cc/min. However, an excessively low bleed flow will result in impaired
response. Conversely, use of excessively high bleed flow may result in failure to obtain
the required flow of 60 cc/min for the reaction chamber. Thus the nominal value of 100
cc/min represents a compromise between response time requirements and sample
supply.
5.3 ELECTRONIC CIRCUITRY
Photons emitted during the chemiluminescence reaction (Section 5.1.1) impinge on
the end-on photomultiplier tube, generating a signal current of approximately 3 x 10
amperes per parts-per-million concentration of NO in the reaction chamber. This
current is measured by an electrometer amplifier.
A functional diagram of the electronic signal circuitry is shown in Figure 5-2. For a
more detailed presentation of the signal circuitry, plus control circuitry for actuation of
NO/NOx Mode Switching Solenoid Valve, etc., refer to the overall schematic and
pictorial diagrams in the rear of this manual, respectively. Individual circuit boards plug
into the Interconnect Board, shown in Figure 7-2.
-10
+15
PHOTOMULTIPLIER
TUBE
REACTION
CHAMBER
ZERO AND SPAN POTENTIOMETERS LOCATED ON FRONT PANEL.
R6, R16, R22, R31, R32, R33 AND R34 LOCATED ON AMPLIFIER BOARD AND
ACCESSIBLE THROUGH HOLES IN AMPLIFIER BOARD COVER (SEE FIGURE 6-1).
IGURE
F
5-2. F
-
HIGH
VOLTAGE
POWER
SUPPLY
+
UNCTIONAL SCHEMATIC DIAGRAM OF ELECTRONIC SIGNAL CIRCUITRY
AR1
+
TP1
R6
R34
R33
R32
R31
PPM RANGE
TP2
R16
+
AR2
-
S
P
A
R22
5V
1V
100mV
10mV
+
REC
-
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
5-5
HEORY
T
As shown in Figure 5-2, the anode of the photomultiplier tube is connected to the
inverting input of AR1, and integrated circuit MOSFET amplifier with extremely high
input impedance to minimize loading effects, and comparatively low output impedance
to match the following stage. The signal from AR1 is directed to integrated circuit
amplifier AR2 for addition al gain.
The front panel ZERO Contro l applies an adjustable zero-biasi ng signal to the input of
AR1, to permit setting the zero point on the meter scale or recorder chart.
For adjustment of overall sensitivity, the current measuring circuitry utilizes two front
panel controls:
1. The PPM RANGE Switch, associated with AR1, provides a choice of seven
fullscale sensitivities, obtained by a combination of high/low feedback selection and
output attenuation. Potentiometers R31, R32, R33 and R34 are adjustable to
provide inter-range correlation of exactly adjusted integral multiples. Refer to
Section 6.1, Step 3.
2. The SPAN Control, associated with AR2, provides continuously variable gain
through feedback adjustment. A voltage divider network associated with AR2
provides a selectable output for a potentiometric recorder.
The following sections describe electronic circuitry of individual circuit boards.
5.3.1 A
MPLIFIER BOARD AND ASSOCIATED CIRCUITRY
The Amplifier Board provides two stages of amplification utilizing integrated circuit
amplifiers AR1 and AR2.
UNCTIONS ASSOCIATED WITH
F
AR1:
AR1 Zero Adjust Potentiometer R6: This screwdriver adjusted trimming
potentiometer, accessible through a hole in the amplifier cover, is used to eliminate
voltage offset with AR1. When input signal is zero, output signal also should be zero.
Refer to Section 6.1 for proper adjustment procedure.
Front Panel Zero Control: This control applies an adjustable zero-biasing sign al to
the input of AR1. to permit setting the zero point on the meter scale or recorder chart.
The adjustment compensates for a residual amplifier input signal resulting from a
combination of background signal and photomultiplier dark current.
Front Panel PPM RANGE: Switch and Associated Resistor Network. The PPM
RANGE Switch controls sensitivity by a combination of two methods:
5-6
Selection of high or low feedback resistance via a transistor switching circuit.
Potentiometer R34, adjustable by insertion of a screwdriver through a hole in the
amplifier cover, permits establishing exactly the correct ratio between the high and low
feedback resistances.
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
HEORY
T
Output attenuation via resistors associated with relays U1, U2 and U3 on the amplifier
board. Potentiometers R31, R32 and R33, adjustable by insertion of a screwdriver
through the corresponding holes in the amplifier cover, permit inter-range correlation of
exactly adjusted integral multiples.
UNCTIONS ASSOCIATED WITH
F
AR2:
AR2 Zero Adjust Potentiometer R16: This screwdriver adjusted trimming
potentiometer, accessible through a hole in the amplifier cover, is used to eliminate
voltage offset with AR2.
Front Panel SPAN Control: Provides continuously variable adjustment of AR2 gain.
Response Time Selection Switch SW2: Provides choice of fast or slow response:
With switch in FAST position, electronic response time (for 0 to 90% of fullscale) is
approximately one second for all ranges except 10 ppm, approximately three seconds
for 10 ppm range.
For such applications as monitoring of stack sources, where slow response is desired,
switch is placed in SLOW position. Electronic response time (0 to 90% of fullscale) is
then approximately 10 seconds for all ranges. Closure of the switch connects
capacitors C7 and C8 across resistor R17, thus increasing the RC time constant.
Recorder Output Selection Switch SW1: The desired recorder output is obtained by
closing the corresponding contact on Switch SW1, thus selecting the appropriate tap
on a voltage divider connected across the output of AR2.
Meter Span Adjustment Potentiometer R22: Potentiometer R22, adjustable by
insertion of a screwdriver through the hole in the amplifier cover, permits adjusting the
fullscale span of the meter so that meter readout agrees with recorder readout.
5.3.2 V
ALVE CONTROL BOARD
IRCUITRY
C
RONT PANEL MODE SWITCH AND ASSOCIATED
, F
As shown in Drawings 649834 and 649835, actuation of NO/NOx Switching Solenoid
Valve K3, associated with the NO2 to NO converter, is controlled by front panel MODE
Switch SW2 via the valve driver circuit on the Valve Control Board. With MODE Switch
in NO position, the solenoid valve is de-energized, placing the converter in bypass
condition. With switch in NOx position, the solenoid valve is energized, placing the
converter in in-line condition.
Drawing 656313 shows internal circuitry of the Valve Control Board. The valve driver
circuit consists of a single-pole, single-throw relay and associated TRIAC element. The
relay is used only to switch the gate current for the TRIAC; therefore, power carried by
the relay contacts is very small.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
5-7
HEORY
T
Each relay is connected to AC neutral while the associated TRIAC is connected to 115
VAC. When the relay closes, in response to closure of the associated contacts on the
MODE Switch, gate current flows through the relay contacts and causes the TRIAC to
fire. Resistance of the TRIAC becomes very small, causing application of 115 VAC
across the solenoid. When the relay contacts open, the gate current is interrupted and
the TRIAC stops conducting. Thus the relay controls the power applied to the solenoid.
5.3.3 ±±±±15 V
OLT POWER SUPPLY
The ±15 Volt Power Supply provides power for the various circuits. As shown in
Drawing 619710, power transformer T1 has three secondarys that are used as follows:
38 VAC center-tapped secondary - Powers both 15 volt supplies through diode
bridge CR1 and filter capacitors C1 and C4.
The adjustable positive regulator, VR1, is set by voltage divider R1, R2 and R3; and its
output is applied to pin A of the circuit board and to test point TP1. Potentiometer R2
should be adjusted to +15.5 VDC ±50 mVDC.
The negative DC, regulated by VR2, is applied to pin D of the circuit board.
The center tap is the common reference for both the +15 and the -15 volt supplies and
is applied to pin R of the circuit board and test point TP2.
Both outputs are used for individual amplifiers on the various circuit boards and for the
zero-biasing circuit associated with the front panel ZERO Control (Section 5.3.1).
90 V secondary - Not used in this instrument.
9.5 VAC secondary - Drives a +5 V supply not used in this instrument.
5.3.4 H
As shown in Drawings 649834 and 649835, DC high voltage for the photomultiplier
tube is provided by the 630922 High Voltage Power Supply. It operates from the ±15
Volt Power Supply (Section 5.3.3).
Drawings 652423 and 654348 shows internal circuitry of the high voltage power
supply. It consists of a voltage regulator circuit utilizing non-inverting operational
amplifier AR1 and a flux-switching DC to AC converter. Potentiometer R4 provides an
adjustment range of approximately -650 V to -2100 V for the regulated DC voltage
supplied to the photomultiplier. Nominal setting is for -900 V; however, during factory
checkout of the individual instrument, R4 is adjusted as required for overall system
sensitivity. If necessary, R4 is readjusted per Section 6.1, Step 5.
5-8
IGH VOLTAGE POWER SUPPLY
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
HEORY
T
5.3.5 C
ONVERTER TEMPERATURE CONTROL BOARD AND ASSOCIATED ELEMENTS
The temperature control circuit for the NO2 to NO converter consists of:
Temperature sensor R3, a platinum resistance element m ounted in the converter.
Converter Temperature Control Board.
CONVERTER ADJUST potentiometer R6 mounted on the front panel, Figure 2-2.
Resistive heating element within a lace-on jacket.
During startup, temperature of the platinum sensor is initially below the setpoint
selected with the CONVERTER ADJUST potentiometer. Controller now switches to
“on” condition, and applies power to the header via TRIAC element Q5. The controller
remains in “on” condition until temperature of the platinum sensor has risen sufficiently
to come within the range of the proportional band of the controller. The controller then
switches alternately off and on during a two-second time base. to maintain the
temperature of the heater at the setpoint.
Clockwise rotation of the CONVERTER ADJUST raises the setpoint; counterclockwise
rotation lowers it. Setpoint adjustment is explained in Section 3.3.
Note
When installing or reinstalling the lace-on healer jacket on the converter, make
sure that the yellow fiberglass sleeve on the temperature sensor leads is
exposed and is not enclosed within the jacket. Otherwise, the insulation may
burn off the sensor leads, causing short-circuiting of the sensor and
temperature runaway or the converter heater.
ONVERTER TEMPERATURE READOUT THERMOCOUPLE
C
The Converter has an associated Type J thermocouple to permit temperature
monitoring. The thermocouple connections, i.e., terminals 5(+) and 6(-) on TB9, are
used to measure converter temperature in the range of 660oF to 750oF (350oC to
400oC). The optimum temperature differs from one analyzer to another. Refer to
Section 3.3.
Note
Converter temperature is not a direct measure of conversion efficiency.
Temperature measurement is for reference purposes only. An exact
determination of conversion efficiency is performed by the procedure given in
section 3.3.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
5-9
HEORY
T
5.3.6 F
AN CONTROL CIRCUIT
The fans in current production instruments run continuously. In older instruments, the
fan control circuit consists of thermistor R4, TRIAC switching circuit and blower fan B.
When internal temperature of the analyzer reaches approximately 100oF (38oC), the
fan is turned on.
5.3.7 R
EMOTE OPERATION OPTION
The Remote Operation Option, Figure 2-5, consists of a Plug-In Adapter Board with
attached harness that terminates on a rear panel plate.
The adapter board has two two-position slide switches: the MAN/AUTO MODE Switch
and the MAN/AUTO RANGE Switch.
The MAN/AUTO MODE Switch provides the choice of local or remote control of the
NO/NOx Mode Switching Solenoid Valve, associated with the NO2 to NO converter.
With switch at MAN, the function is under control of the front panel MODE Switch
(Section 5.3.2). With switch at AUTO, the front panel MODE Switch is disabled; the
function is then remotely controlled, by either the operator or a computer, via
contact-closure signals applied to terminals on TB5.
The MAN/AUTO RANGE Switch provides the choice of local or remote selection of
operating range. With switch at MAN, range selection is under control of the front
panel PPM RANGE Switch. With switch at AUTO, the front panel PPM RANGE Switch
is disabled; range selection is remotely controlled by either the operator or a computer,
via contact-closure signals applied to terminals on TB6.
Control at TB5 and TB6 is accomplished by a contact closure from the terminal
marked COM (for common) to the terminal marked with the name of the desired
function. Contact closure requirements are 25 mA at 15 VDC.
5.3.8 R
ANGE
I.D. O
PTION
The Range I.D. Option provides contact closure signals that permit a computer or
other external device to identify the position selected on the front panel PPM Switch.
This cable is connected to PPM Switch SW1 and extends to connector J10 mounted
on the rear of the case.
5-10
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
S
ERVICE AND MAINTENANCE
6
WARNING: ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to
live parts which can cause death or serious injury. Refer servicing to qualified
personnel.
For safety and proper performance this instrument must be connected to a
properly grounded three-wire source of power.
WARNING: ULTRAVIOLET LIGHT HAZARD
UV light from the ozone generator can cause permanent eye damage. DO NOT
LOOK DIRECTLY AT THE UV SOURCE IN THE OZONE GENERATOR. Use of UV
filtering glasses is recommended.
WARNING: LIGHT EXPOSURE
The photomultiplier tube must not be exposed to ambient light. Unless
appropriate precautions are observed, light can strike the tube upon removal of
fittings from the reaction chamber.
6.1 SYSTEM CHECKS AND ADJUSTMENTS
The following procedures may be used to determine the cause of unsatisfactory
instrument performance, or to make adjustments following replacement of
components. If a recorder is available, use it for maximum convenience and accuracy
in the various tests.
ETER MECHANICAL ZERO
M
With electrical power removed from the analyzer, connect a shorting jumper across the
terminals of the front panel readout meter. Meter should read zero; if not, adjust screw
at rear of meter as required. Remove jumper.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
6-1
ERVICE AND MAINTENANCE
S
MPLIFIER ZERO ADJUSTMENTS
A
Refer to Figure 6-1 for location of Amplifier Board components noted in the following
procedure.
1. Open front panel to turn off high voltage; set SPAN Control at 1000; set ZERO
Control at 1000. Then perform the following checks and adjustments:
2. With PPM RANGE Switch at 250, obtain zero reading on meter or recorder by
screwdriver adjustment of R16, accessible through hole in cover of amplifier board.
3. With PPM RANGE Switch at 1000, obtain reading of 0 (±1 division) on meter or
recorder by screwdriver adjustment of R6, accessible through hole in cover of
amplifier board.
4. Repeat Step a. If the specified zero readings are unobtainable by adjustment of
the corresponding potentiometers, replace amplifier board.
NTER-RANGE ATTENUATION CORRELATION ADJUSTMENTS
I
Refer to Figure 6-1 for location of Amplifier Board components noted in the following
procedure.
1. During the following adjustments, observe the analyzer output on a digital voltmeter
or strip-chart recorder. If neither is available, use the front panel meter. Perform
the following adjustments:
2. Supply 10 parts-per-million standard to analyzer. Set PPM RANGE Switch at 10.
Obtain fullscale reading by adjustment of front panel SPAN Control.
3. Move PPM RANGE Switch to 250. Obtain reading of 4% of fullscale by
screwdriver adjustment of R33, accessible through hole in cover of amplifier board.
4. Move PPM RANGE Switch to 25. Obtain reading of 40% of fullscale by
screwdriver adjustment of R31, accessible through hole in cover of amplifier board.
5. Move PPM RANGE Switch to 100. Obtain reading of 10% of fullscale by
screwdriver adjustment of R32, accessible through hole in cover of amplifier board.
6. Supply 100 parts-per-million standard to analyzer. Adjust front panel SPAN
Control for fullscale reading.
7. Move PPM RANGE Switch to 1000. Obtain reading of 10% of fullscale by
screwdriver adjustment of R34, accessible through hole in cover of amplifier board.
ETER FULLSCALE SPAN ADJUSTMENT
M
If a recorder is used, and has been properly zeroed, it should agree with the meter
reading. If not, obtain agreement by screwdriver adjustment of R22, accessible
through hole in cover of amplifier board (Figure 6-1). If agreement cannot be obtained,
check recorder. If recorder is functioning properly, replace amplifier board.
6-2
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
10
100
MV
REC
R6
R22
5
1
V
OUTPUT
FAST SLOW
RESPONSE
R34
CALIB
INPUT
R33
CALIB
R31
CALIB
R32
CALIB
R16
ERVICE AND MAINTENANCE
S
IGURE
F
6-1. A
VERALL SENSITIVITY
O
MPLIFIER BOARD
Principal factors that determine overall sensitivity of the analyzer are: (a) sample flow
rate to the reaction chamber, and (b) sensitivity of the photomultiplier tube. If specified
fullscale readings are unobtainable by adjustment of the SPAN Control, sensitivity is
subnormal. The cause of the reduced sensitivity may be in either the flow system
(Section 6.2) or the electronic circuitry (Section 6.3).
If either the High Voltage Board or the Phototube/Reaction Chamber Assembly has
been replaced, a readjustment of R4 on the High Voltage Board will probably be
required to obtain the correct overall sensitivity. Turn the adjustment in the
appropriate direction: clockwise to increase photomultiplier high voltage and therefore
sensitivity; or, counterclockwise to decrease voltage and sensitivity. Adjustment range
is approximately -650 volts to -2100 volts for the regulated DC voltage applied to the
photomultiplier. Nominal setting is for -1100 volts; however, the adjustment should be
set as required for overall system sensitivity.
ZONE OUTPUT
O
Maximum concentration of NO or NOx which the analyzer will measure with linearity
depends on both the ozone generator output and the sample flow rate. If the analyzer
seems to be ozone limited at high concentrations of NO or NOx, possible causes are:
incorrect sample flow rate, incorrect flow rate through ozone generator, malfunction in
ozone generator or associated transformer.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
6-3
ERVICE AND MAINTENANCE
S
WARNING: TOXIC GAS
Use extreme caution in troubleshooting ozone generator. Ozone is toxic.
CAUTION: OZONE DAMAGE
When instrument power is on, the ultraviolet source sample is energized,
converting a portion of the oxygen contained with the ozonator into ozone. With
normal flow through the ozonator, the ozonized air or oxygen is continuously
swept though the ozonator and into the reaction chamber. If flow is stopped,
however, the ozone will diffuse and will attack the sintered metallic restrictor
element in the tee fitting at the upstream end of the ozonator. Operation of the
analyzer with the restrictor element thus rusted will result in the following
consequences: Reduced flow of air or oxygen through the ozonator, ozone
deprivation within the reaction chamber and non-linear response of the analyzer
to NO/NOx.
To prevent such damage, verify that the front panel ozone ON/OFF switch is
turned off if flow of feed gas to the air inlet is terminated.
ACKGROUND CURRENT
B
With zero air supplied to rear panel SAMPLE inlet, and MODE Switch at either NO or
NOx, excessive background current is evidenced by inability to obtain zero meter
reading by adjustment of the ZERO Control. If background current is excessive, the
cause must be found and corrected. The fault may be in either the electronic circuitry
or the sample flow system.
First, establish proper performance of the electronic circuitry. Turn off high voltage by
opening door. Turn on analyzer power. Verify that ZERO Control and amplifier are
functioning properly. Close door and check for the following:
1. Excessive photomultiplier dark current. To check, shut off all flow to the ozone
generator. Supply cylinder air to rear panel SAMPLE inlet. With MODE switch at
NO, note response on meter or recorder. If background is still excessive, possible
causes are: (1) leakage of ambient light to photomultiplier tube, or (2) defective
photomultiplier tube.
2. Contamination of reaction chamber or sample flow system. Refer to Sections
6.2.1, 6.2.2 and 6.2.4
6-4
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
ERVICE AND MAINTENANCE
S
2
NO
TO
NO C
ONVERTER TEMPERATURE ADJUSTMENT
Use procedure of Section 3.3
The FLOW BALANCE Control must now be adjusted to equalize pressure drops for
the NO and NOx legs of the internal flow system. In the usual method, NO2 free nitric
oxide standard gas is supplied to the analyzer and the FLOW BALANCE Control is
adjusted for equal reading of parts-per-million for the NO and NOx modes, as
explained in Step 9.
An alternate method of flow balance, useable if NO2 free nitric oxide standard gas is
unavailable consists of checking SAMPLE pressure reading for NO and NOx modes,
as described in Step 10.
LOW BALANCE USING FRONT PANEL METER INDICATION WITH
F
ITRIC OXIDE STANDARD GAS
N
NO
2
FREE
1. Supply to the SAMPLE inlet a suitable NO2 free nitric oxide span gas, for example
2500 parts-per-million NO. And NO span gas may be used, since balance is
independent of concentration. However, it is important that the span gas be
entirely free of NO2; otherwise, the supposed flow balance achieved will be
erroneous. A NO span gas known to be free of NO2 can be purchased as a
certified standard gas. Set SAMPLE Pressure Regulator so that SAMPLE
Pressure Gauge indicated the value appropriate to the desired operating range.
Refer to Figure 3-3. Set BYPASS Needle Valve for reading of 2 liters per minute
on BYPASS Flowmeter.
2. Supply to the AIR inlet suitable air or oxygen, depending on the operating range.
Refer to Section 2.3. Adjust OZONE Pressure Regulator so that OZONE Pressure
Gauge indicates the value appropriate to the desired operating range. Refer to
Figure 3-3.
3. Set MODE Switch at NO. Leave ZERO Control at previous setting. Turn PPM
RANGE Switch to a setting appropriate to the nitric oxide content of the particular
span gas. Adjust SPAN Control for reading of approximately 90% on meter or
recorder.
4. Move MODE Switch from NO to NOx; compare meter or recorder readings for the
two modes. The NOx reading should be equal to the NO reading. If the readings
are within acceptable agreement, the flows are properly balanced.
If the readings for the two modes are not in acceptable agreement, adjust the front
panel FLOW BALANCE Control as required.
LTERNATE FLOW BALANCE METHOD
A
NO
AND
NOX M
ODES
SING SAMPLE PRESSURE READING FOR
, U
If NO2 free nitric oxide standard gas is unavailable, verify that reading on SAMPLE
Pressure Gauge is the same for NO and NOx modes. If not, adjust FLOW BALANCE
Needle Valve as required.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
6-5
ERVICE AND MAINTENANCE
S
6.2 SERVICING FLOW SYSTEM
To facilitate servicing and testing, the Model 951A has front, rear and top access.
Refer to Drawings 649958 and 649842 in the rear of this manual for details of flow
system.
6.2.1 S
AMPLE CAPILLARY
Depending on when the instrument was made, one of two types of capillaries may be
present: a) a black tubing with white Teflon inner sleeve, or b) a one-piece black
Teflon tubing.
If clogging of sample capillary is suspected, measure flow rate as described below:
1. Turn off instrument power and shut off all gases.
2. Cover and shade the fittings on the reaction chamber with a dark cloth or other
light-shielding material. While maintaining the area properly darkened, remove the
fitting associated with the sample capilla ry (red-bande d tube) and place a cap o ver
the open fitting to prevent entry of stray light.
Note
If the opened fitting is inadvertently exposed to ambient light, the instrument will
temporarily give a highly noisy background reading. If so, this condition may be
corrected by leaving the instrument on, with high voltage on, for several hours.
3. With instrument power off, supply suitable test gas (dry nitrogen or air) to rear
panel SAMPLE inlet.
6-6
4. Connect a flowmeter to open end of sample capillary. Per Table 3-1, adjust
SAMPLE Pressure Regulator to normal operating setting, and verify that flowmeter
indicates appropriate flow.
If flow is correct, restore analyzer to normal operation. If flow is low, the capillary
requires cleaning; proceed with the following steps:
1. On NO/NOx Mode Switching Solenoid Valve K3, remove fitting from port
associated with sample capillary.
2. Clean capillary with alcohol, then purge with dry nitrogen for one minute.
Reconnect capillary to K3.
3. W ith the photomultiplier still covere d, slowly insert the free end of the capillary into
the corresponding fitting on the reaction chambe r. (If capillary is of the two-piece
variety, the bare Teflon tube at the end of the capillary must be inserted blindly and
rotated until it enters a channeled internal passage, which can be detected by feel.
The end of the Teflon tube should not project into the interior of the chamber; it
should be slightly below the surface of the inner face. If tube is too long, trim back
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
ERVICE AND MAINTENANCE
S
slightly as required.) P ush the cap illary in unt il it bottoms again st the interna l fittin g,
then tighten fitting.
Note
After capillary cleaning, instrument readout may be noisy for a short time, but
should then become quiet. However, if bare Teflon tube is improperly inserted
in Step 7, noisy readout will persist.
6.2.2 O
ZONE RESTRICTOR AND CAPILLARY
CAUTION: OZONE DAMAGE
When instrument power is on, the ultraviolet source sample is energized,
converting a portion of the oxygen contained with the ozonator into ozone. With
normal flow through the ozonator, the ozonized air or oxygen is continuously
swept though the ozonator and into the reaction chamber. If flow is stopped,
however, the ozone will diffuse and will attack the sintered metallic restrictor
element in the tee fitting at the upstream end of the ozonator. Operation of the
analyzer with the restrictor element thus rusted will result in the following
consequences: Reduced flow of air or oxygen through the ozonator, ozone
deprivation within the reaction chamber and non-linear response of the analyzer
to NO/NOx.
To prevent such damage, verify that the front panel ozone ON/OFF switch is
turned off if flow of feed gas to the air inlet is terminated.
With instrument power off, supply suitable test gas (dry nitrogen or air) to rear panel
AIR inlet. Cover photomultiplier housing with a dark cloth. At the fitting on the reaction
chamber, disconnect the ozone tube (yellow banded) and place a cap over the open
fitting to prevent entry of ambient light. Connect a flowmeter to open end of ozone
tube. Adjust the OZONE Pressure Regulator so that OZONE Pressure Gauge
indicates normal operating pressure, as listed in Table 3-1; verify that test flowmeter
indicates appropriate flow, as listed in Table 3-1.
Subnormal flow indicates clogging in the flow path that supplies air or oxygen to the
ozone generator. This path contains two pressure reducing elements:
1. A Restrictor, consisting of a metal fitting with internal fritted (metal membrane)
restrictor, the fitting is upstream from the inlet port of the ozone generator. If the
internal restrictor becomes plugged, the Restrictor must be replaced, as it cannot
normally be cleaned satisfactorily.
2. A capillary connected from the outlet of the ozone generator to the ozone inlet of
the reaction chamber (coded with a yellow dot). This capillary may be cleaned if
desired. Remove the ozone capillary by a procedure similar to that given for the
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
6-7
ERVICE AND MAINTENANCE
S
sample capillary in Section 6.2.1, observing all cautions cited. Clean ozone
capillary by same method used for sample capillary.
Test for Ozone Efficiency
Perform the following steps:
1. Span the instrument with sample at 5 lb.
2. Decrease the pressure in 1 lb. steps. The response should be equal steps.
If the steps are not equal, there is an ozone deficiency. The ozone transformer lamp
should be checked.
6.2.3 R
EPLACING
NO
2
TO
NO C
ONVERTER
Converter replacement is recommended upon failure to achieve a conversion
efficiency of 95% durin g the test of Section 3.3
To remove and replace converter, refer to Figure 6-2A. Within the outer lace on jacket
heater, the converter tube is wrapped in aluminum foil, with the temperature sensor in
contact with the glass converter tube. In removing foil, note sensor and thermocouple
position to ensure correct repositioning during reassembly. Assemble sensor to
replacement converter, as shown in Figure 6-2B, using new aluminum foil.
Note
When installing or reinstalling the lace on heater jacket on the converter, make
sure that the yellow fiberglass sleeve on the temperature sensor leads is
exposed and is not enclosed within the jacket. See Figure 6-2C.
6-8
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
Lace-On Heater
Thermocouple
Temperature Sensor
Converter Tube
Connector (2)
Tube (2)
B. ATTACHING TEMPERATURE SENSO R TO CO NVERTER TUBEA. EXPLODED VIEW
Temperature
Sensor
NOx -to-NO
Converter Tube
Top of
Converter Assembly
Thermocouple
2
NOTE
Orient converter tube as shown
and install sensor flush with
outside edge of tube as shown.
NOTE
Sensor leads to be
outside heater jacket.
ERVICE AND MAINTENANCE
S
C. HEATER JACKET PROPERLY INSTALLED ON CONVERTER TUBE
Temperature Sensor
Leads
Thermocouple
Lead
Converter Tube
Inlet/Outlet Connections
Yellow Fiberglass Sleeving
Exposed Outside of Jacket
Lace-On
Heater Jacket
IGURE
F
556383-Y Rosemount Analytical July 2000
6-2. NO2
TO
NO C
ONVERTER ASSEMBLY
Model 951A NO/NOx Analyzer
6-9
ERVICE AND MAINTENANCE
A
y
S
6.2.4 C
LEANING REACTION CHAMBER
CAUTION: LIGHT DAMAGE
Photomultiplier will develop temporary electronic noise if exposed to ambient
light with high voltage on. Damage may be permanent if exposure occurs with
high voltage off. A temporary noisy condition may be corrected by leaving
instrument on, with high voltage on, for several hours. The required recovery
time depends on intensity and duration of the exposure. Noise level on the most
sensitive range usually drops to normal within 24 hours.
If sample gas is properly filtered, the reaction chamber should not require cleaning. In
event of carryover or contamination, however, the chamber should be disassembled to
permit cleaning the quartz window and the optical filter. The following procedure is
recommended.
1. Turn off instrument power and shut off all gases.
2. Cover and shade the Reaction Chamber Photomultiplier Assembly, Figure 6-3, with
a dark cloth or other light shielding material.
CAUTION: HAND PROTECTION
Always wear surgical rubber gloves when handling reaction chamber.
Reaction Chamber Assembly
Phototube Housing
Magnetic
Phototube
1
O-rings (1 or 2, depending on Phototube length.
(Lubricate with silicone grease)
IGURE
F
Phototube Socket
ssembl
Proper orientation ofphotomultiplier in socketmay be determined by measuring the
1
resistance from the centerpin toground (should be approximately 11 megohm).
6-3. R
EACTION CHAMBER/PHOTOMULTIPLIER ASSEMBLY
6-10
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
ERVICE AND MAINTENANCE
S
3. At the front of the Reaction Chamber Photomultiplier Assembly, remove the three
screws that secure the reaction chamber mounting bracket. See Figure 6-3. The
middle screw is a 1/4-turn quick-release type; the two adjacent screws are the
conventional continuous thread type.
4. With reaction chamber now free, detach its three tube fittings. See Figure 6-3.
5. Slide off and remove the reaction chamber from the housing. Take care not to let
light enter.
6. Unscrew plastic end cap, Figure 6-4, thus freeing the quartz window and the red
plastic optical filter.
7. Clean the reaction chamber by the appropriate one of the following two methods,
“a” or “b”. Method “a” is applicable in most cases. Alternate method “b” is
applicable when the instrument has shown high residual fluorescence, as indicated
by high residual currents on a zero gas and high differentials between zero gas
readings obtained with the ozone lamp on and off.
CAUTION: SOLVENT USAGE
For a solvent, use only distilled water, do not use alcohol or acetone.
Reaction Chamber Assembly
876351
O-ring
836486
O-ring
Quartz Window
IGURE
F
6-4. R
Phototube Housing
Optical Filter
(Red Plastic)
Cap
EACTION CHAMBER ASSEMBLY AND PHOTOTUBE HOUSING
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
6-11
ERVICE AND MAINTENANCE
S
a. Standard Cleaning Procedure
Using a stiff plastic bristle brush, such as a toothbrush, scrub the Teflon surface
and gas ports of the reaction chamber with clean distilled water and Alconox
detergent (634929). Alconox detergent is included in the shipping kit provided
with the Model 951A NO/NOx Analyzer and is also available from
Sargent-Welch Scientific Company under its Catalog Number S-19650-A.
Using Alconox and clean, soft facial tissue (NOT industrial wipes), carefully clean
the quartz window, vigorously flush reaction chamber and quartz window with
clean distilled water. Blow out all possible water from internal passages of
reaction chamber. Dry reaction chamber and quartz window in a warm oven
at 125oF to 150oF (52oC to 66oC) for 30 to 45 minutes or purge dry the parts
with dry cylinder air or nitrogen to eliminate all moisture.
b. Alternate Cleaning Procedure - For High Residual Fluorescence
Holding the reaction chamber by the tube fittings, and using appropriate caution.
immerse the white Teflon part of the chamber in 50% concentrated Reagent
Grade HCl. After five minutes, rinse thoroughly with deionized water, then air
dry as in method “a” above.
8. Place parts in position and press on end cap so that mating threads engage
properly, without cross-threading. Turn mating parts in one continuous motion until
the parts bottom. Do not over torque.
9. With reaction chamber now assembled, replace and reconnect it in the reverse
order of that used for removal.
6.2.5 P
HOTOMULTIPLIER TUBE AND HOUSING
The photomultiplier tube operates at high DC voltages (nominal range 600 to 1200
volts) and generates small currents that are highly amplified by the signal conditioning
circuitry. It is therefore important that ambient humidity and condensed water vapor be
excluded from the interior of the photomultiplier housing. Ambient humidity can result
in electrical leakage, evidenced as abnormally high dark current. Water vapor or
condensed moisture in contact with the thermoelectrically-cooled photomultiplier may
be evidenced by abnormally high noise level during instrument readout on zero air or
upscale standard gas.
The Reaction Chamber Photomultiplier Assembly, Figure 6-3, incorporates several
features for exclusion of humidity and moisture. The photomultiplier socket assembly
is potted with a high impedance compound and is sealed from external influences with
epoxy and rubber gasket material. The socket assembly and the reaction chamber
are sealed, with o-rings, into opposite ends of the tubular photomultiplier housing. The
socket end of the housing may be sealed with either one or two o-rings, depending on
the length of the phototube.
6-12
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
ERVICE AND MAINTENANCE
A
y
S
In addition, a rolled-up piece of blotter-like desiccant paper is inserted between the
shield external to the housing. The desiccant paper ensures hermetic isolation of the
housing despite possible minor leakage around the o-rings, and eliminates the effects
of any humidity possibly trapped during assembly of the socket and the reaction
chamber to the housing. If the photomultiplier housing is exposed to ambient air for a
period of tie, the desiccant paper will absorb water. Before reassembly, the paper
must then be heated in an oven at about 300oF (149oC) for at least two hours.
Within the housing is a humidity indicator which turns pink in the presence of humidity.
6.2.6 T
EFLON LINER IN LAMP HOUSING OF OZONE GENERATOR
WARNING: MERCURY HAZARD
Handle lamp assembly with extreme caution. Lamp contains mercury vapor.
CAUTION: LAMP DAMAGE
Do not touch lamp. Fingerprints could cause a decrease in lamp output.
A Teflon liner was used inside the ozone lamp housing during early production runs of
the Model 951A. A bare lamp housing is now used in lieu of the liner.
Inlet
Outlet
O-Ring
Ozone Lamp
ssembl
Connector Plug
For replacement of lamp only, order Ozone Lamp Kit 658156 (includes O-Ring).
For replacement of complete Ozone Generator, order Kit 654793.
Refer to instructions provided with kit.
IGURE
F
556383-Y Rosemount Analytical July 2000
6-5. O
ZONE GENERATOR
Model 951A NO/NOx Analyzer
6-13
ERVICE AND MAINTENANCE
S
6.3 SERVICING ELECTRONIC CIRCUITRY
Refer to the appropriate schematic and pictorial diagrams in the rear of this manual
when servicing. The electronic system, Figure 7-2, utilizes plug in printed circuit
boards with solid state components. After a malfunction is traced to a particular board,
the recommended procedure is to return it to the factory for repair. Your Service
Representative will arrange a Returned Goods Authorization.
6-14
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
R
y
1
EPLACEMENT PARTS
7
The following parts are recommended for routine maintenance and troubleshooting of
your instrument. If the troubleshooting procedures do not resolve the problem, contact
your local service office.
7.1 CIRCUIT BOARD REPLACEMENT POLICY
In most situations involving a malfunction of a circuit board, it is more practical to
replace the board than to attempt isolation and replacement of the individual
component, as the cost of test and replacement will exceed the cost of a rebuilt
assembly. As standard policy, rebuilt boards are available on an exchange basis.
Because of the exchange policy covering circuit boards, the following list does not
include individual electronic components. If circumstances necessitate replacement of
an individual component, which can be identified by inspection or from the schematic
diagrams, obtain the replacement component from a local source of supply.
7.2 REPLACEMENT PARTS
621095±15V Power Supply
649815Amplifier Board
652842Ballast Transformer, Ozone Gener ator
623719Capillary, Sam ple
623743Capillary, Pump
634398Capillary, Bleed
630960Coaxial Cable, Amplifier
632783Converter Assembl
861273Fan
809374Fuse 3/4 Amp
777010Fuse 5 Amp (Package of 15)
658992High Voltage Power Supply Board
630920Interconnect Board
630931Meter
654793Ozone GeneratorSee Section 7.2.3
658156Ozone Generator Lamp Kit (Lamp Only)
657842Power Supply Board
632748Pump (Option)
635664Pump Repair Kit
636343Reaction Chamber
902315Signal Converter 4-20mA (Option)
641620Temperature Control Board
624593Valve Control Board
See Section 7.2.2
1
See section noted for components of assembly.
556383-Y Rosemount Analytical July 2000
Model 951A NO/NOx Analyzer
7-1
EPLACEMENT PARTS
R
Pneumatics (see Figure 7-3)
Pump (Option)
Ozone Generator
Interlock Switch
Fan
Power
Switch
Meter
Power Switch Shield
Ballast
Transformer
Converter Assembly
(see Figure 7-4)
IGURE
F
7-2
7-1A. M
ODEL
951A
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
Coaxial
Amplifier
Cable
Electronics
(see Figure 7-2)
Terminal Chassis Assembly
EPLACEMENT PARTS
R
Reaction
Chamber
Thermocouple
Connector
IGURE
F
556383-Y Rosemount Analytical July 2000
7-1B. M
ODEL
951A
Model 951A NO/NOx Analyzer
7-3
EPLACEMENT PARTS
R
IGURE
F
7-4
7-1C. M
ODEL
951A
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
For replacement of lamp only, order Ozone Lamp Kit 658156 (includes O-Ring).
For replacement of complete Ozone Generator, order Kit 654793.
Refer to instructions
IGURE
F
7-5. O
rovided with kit.
ZONE GENERATOR COMPONENTS
7-8
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer
EPLACEMENT PARTS
R
7.2.3 L
OW TEMPCO OPTION
652831Power Supply Assembly
654831Reaction Chamber
861040Sensor, Cooler Temperature
884791Thermistor
654943Thermoelectric Cooler Assembly
652836Transformer, Thermo el ectri c Cooler
BALLAS
TOROID
TRANSFORME
INDUCTOR
OZONE
GENERATOR
THERMO COOLER
POWER
THERMO COOLER
INDICATOR
FAN
CHASSIS
TERMINA
BAFFL
IGURE
F
556383-Y Rosemount Analytical July 2000
7-6. T
EMPCO RETROFIT COMPONENT LOCATION
THERMO COOLER ASSY
ELECTRONICS
CONVERTER ASSY
Model 951A NO/NOx Analyzer
7-9
EPLACEMENT PARTS
R
N
OTES
7-10
July 2000 Rosemount Analytical 556383-YModel 951A NO/NOx Analyzer