Rosemount 880 NDIR Analyzer-Rev N Manuals & Guides

Page 1
Rosemount Analytical
M
ODEL
N
NFRARED ANALYZER
NSTRUCTION MANUAL
ON
-D
880
ISPERSIVE
748176-N
Page 2
OTICE
N
This Model 880 instruction manual is for use with instruments with a software version of 2.0 or greater.
The information contained in this document is subject to change without notice.
Irtran is a Registered Trademark of Eastman Kodak Co. Pyrex is a Registered Trademark of Corning Glass Works Teflon and Viton are Registered Trademarks of E.I. duPont de Nemours and Co., Inc. SNOOP is a Registered Trademark of NUPRO Co. Model SGD Series-710 Standard Gas Divider is a Registered Trademark of STEC Inc.
Manual Part Number 748176-N July 1998 Printed in U.S.A.
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
Page 3
C
ONTENTS
PREFACE
SAFETY SUMMARY ..........................................................................................P-1
SPECIFICATIONS..............................................................................................P-3
SPECIFICATIONS - OPTIONS ..........................................................................P-5
CUSTOMER SERVICE, TECHNICAL ASSISTANCE AND FIELD SERVICE....P-6
RETURNING PARTS TO THE FACTORY.........................................................P-6
TRAINING ......................................................................................................P-6
DOCUMENTATION............................................................................................P-6
QUICK STARTUP AND TROUBLESHOOTING GUIDE.....................................P-7
S
ECTION
1.1 GENERAL DESCRIPTION........................................................................1-1
1.2 AVAILABLE OPTIONS..............................................................................1-1
1. I
NTRODUCTION
SECTION 2. UNPACKING AND INSTALLATION
2.1 UNPACKING............................................................................................2-1
2.2 LOCATION...............................................................................................2-1
2.3 VOLTAGE REQUIREMENTS...................................................................2-1
2.4 ELECTRICAL CONNECTIONS................................................................2-3
2.4.1 Line Power Connections..............................................................2-4
2.4.2 Recorder Connections.................................................................2-4
2.5 SAMPLE AND REFERENCE INLET/OUTLET CONNECTIONS..............2-4
748176-N
July 1998
Rosemount Analytical
I
Page 4
ONTENTS
C
S
ECTION
2.6 CALIBRATION GAS REQUIREMENTS................................................... 2-4
2.7 SAMPLE HANDLING SYSTEM...............................................................2-4
2.8 LEAK TEST PROCEDURE...................................................................... 2-5
2.9 SAMPLE FLOW RATE ............................................................................ 2-6
2.10 DIFFERENTIAL ANALYSIS WITH FLOW-THROUGH
2.11 OPTION BOARDS................................................................................... 2-7
2.12 ORDERING OPTION KITS.................................................................... 2-11
2. (
2.11.1 Alarm Connections................................................................... 2-7
2.11.2 Current Output Option Connections......................................... 2-8
2.11.3 Calibration Gas Control Connections....................................... 2-9
2.11.4 Auto Zero/Span Connections................................................... 2-9
2.11.5 Remote Input/Output Connections........................................... 2-10
CONTINUED
REFERENCE CELL...................................................................... 2-7
)
SECTION 3. INITIAL STARTUP AND CALIBRATION
3.1 LEAK TEST.............................................................................................. 3-1
3.2 POWER VERIFICATION ......................................................................... 3-1
3.3 SOFTWARE VERSION ........................................................................... 3-1
3.4 FRONT PANEL INDICATORS AND CONTROLS ................................... 3-1
3.4.1 Display 3-1
3.4.2 Function Keys............................................................................. 3-3
3.4.3 User-Programmable Keys.......................................................... 3-4
3.4.4 Run Mode Display...................................................................... 3-5
3.4.5 General Display Information....................................................... 3-5
3.5 ACCESSING MODE DISPLAYS.............................................................. 3-6
3.6 SECURITY CODE.................................................................................... 3-6
3.7 RANGE PARAMETERS........................................................................... 3-10
3.8 ANALYZER DIAGNOSTICS .................................................................... 3-13
3.9 ZERO CALIBRATION.............................................................................. 3-13
3.9.1 Hardware and Software Zero..................................................... 3-13
3.9.2 Calibrating the Analyzer with Zero Gas...................................... 3-14
II
Rosemount Analytical
July 1998
748176-N
Page 5
SECTION 3. (CONTINUED)
3.10 ZERO CALIBRATION FOR THE ANALYZER WITH THE CAL GAS
CONTROL OPTION......................................................................3-17
3.10.1 Calibrating the Analyzer with the Calibration Gas Control
Option with Zero Gas....................................................3-17
3.11 SPAN CALIBRATION..............................................................................3-17
3.11.1 Hardware and Software Settings for Span Gas........................3-17
3.11.2 Calibrating the Analyzer with Span Gas....................................3-18
3.12 SPAN CALIBRATION FOR THE ANALYZER WITH THE CAL GAS
CONTROL OPTION......................................................................3-19
3.13 LINEARIZATION .....................................................................................3-20
3.14 ALARM OPTION .....................................................................................3-24
3.14.1 STATUS Display.......................................................................3-26
ONTENTS
C
3.15 CURRENT OUTPUT OPTION.................................................................3-26
3.16 AUTO ZERO/SPAN OPTION..................................................................3-27
3.17 REMOTE RANGE INPUT/OUTPUT OPTION .........................................3-30
S
ECTION
4.1 ROUTINE OPERATION ............................................................................4-1
4.2 RECOMMENDED CALIBRATION FREQUENCY .....................................4-1
4.3 SHUTDOWN .............................................................................................4-1
4.4 DETECTION SYSTEM THEORY..............................................................4-1
4. R
OUTINE OPERATION AND THEORY
SECTION 5. TROUBLESHOOTING
5.1 ERROR CODE SUMMARY......................................................................5-1
5.2 OSCILLATOR TUNE ADJUSTMENT.......................................................5-2
5.3 PREAMP GAIN.........................................................................................5-2
5.4 SOURCE BALANCE SHUTTER ADJUSTMENT......................................5-3
5.5 SOURCE CURRENT ADJUSTMENT.......................................................5-3
5.6 VOLTAGE CHECKS.................................................................................5-3
5.7 DIGITAL GAIN ADJUSTMENT.................................................................5-3
748176-N
July 1998
Rosemount Analytical
III
Page 6
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
SECTION 5. (CONTINUED)
5.8 CASE HEATER TEMPERATURE CONTROL.........................................5-4
5.9 DETECTOR HEATER.............................................................................. 5-4
SECTION 6. ROUTINE SERVICING
6.1 CELL REMOVAL, CLEANING AND INSTALLATION.............................. 6-1
6.2 CELL DESICCANT .................................................................................. 6-2
6.3 SOURCE REPLACEMENT...................................................................... 6-2
6.4 SOURCE BALANCE SHUTTER ADJUSTMENT..................................... 6-3
6.5 CHOPPER MOTOR ASSEMBLY (PN 658313)....................................... 6-3
6.6 MICROBOARD REPLACEMENT ............................................................ 6-3
6.7 DETECTOR REMOVAL........................................................................... 6-4
6.8 ELECTRONIC CIRCUITRY ..................................................................... 6-4
6.8.1 Oscillator Circuit Board and Associated Elements of Amplitude-
Modulation Circuit......................................................... 6-4
6.8.2 Functioning of Modulation System in TUNE Mode...................... 6-5
6.8.3 Functioning of Modulation System in Operating Mode................ 6-5
6.8.4 Radio-Frequency Demodulator................................................... 6-6
6.8.5 Signal Board (DWG 624085) ...................................................... 6-6
6.8.6 Power Supply Board (DWG 624073) .......................................... 6-6
6.8.7 Adapter Board (DWG 624127).................................................... 6-6
6.8.8 Microboard.................................................................................. 6-6
6.8.9 Optional Case Heater Temperature Control Board
(DWG 624003) ............................................................. 6-7
6.8.10 Dual Alarm/Calibration Gas Control Option Board
(DWG 624204) ............................................................. 6-7
6.8.11 Isolated Remote Range I/O Option Board (DWG 624251) ......... 6-7
6.8.12 Auto Zero/Span Option Board (DWG 624599)............................ 6-8
6.8.13 Current Output Option Board (DWG 624092)............................. 6-8
SECTION 7. REPLACEMENT PARTS
7.1 CIRCUIT BOARD REPLACEMENT POLICY............................................ 7-1
7.2 SELECTED REPLACEMENT PARTS...................................................... 7-1
IV
Rosemount Analytical
July 1998
748176-N
Page 7
ONTENTS
C
GENERAL PRECAUTIONS FOR HANDLING & STORING HIGH PRESSURE CYLINDERS WARRANTY FIELD SERVICE AND REPAIR FACILITIES
F
IGURES
2-1 Power Supply Board ..............................................................................2-2
2-2 Case Heater Temperature Control Board...............................................2-2
2-3 Rear View of Model 880.........................................................................2-3
2-4 Cable Gland ...........................................................................................2-3
2-5 Calibration Gas Control and Alarm Connections....................................2-8
2-6 Current Output Connections...................................................................2-8
2-7 Auto Zero/Span Connections................................................................2-9
2-8 Remote Input/Output Connections.........................................................2-10
3-1 Model 880 Adjustments..........................................................................3-2
3-2 Model 880 Keypad .................................................................................3-2
3-3 Run Mode Display..................................................................................3-5
3-4 Logic Flow Chart..................................................................................... 3-7
3-5 Accessing the Security Lockout Feature................................................3-9
3-6 Activating/Deactivating the Security Lockout Feature ............................3-9
3-7 Changing the Password.........................................................................3-10
3-8 Setup/Checkout of Range Parameters...................................................3-12
3-9 Analyzer Diagnostics..............................................................................3-13
3-10 Viewing the Zero Display........................................................................3-15
3-11 Changing the Zero Setting......................................................................3-15
3-12 Changing the Zero Setting for Analyzers with Calibration Gas Control..3-16
3-13 Viewing the Span Setting.......................................................................3-16
3-14 Changing the Span Setting.....................................................................3-19
3-15 Changing the Span Setting for Analyzers with Calibration Gas Control.3-20
3-16 Typical Application Linearization Curve..................................................3-22
3-17 User-Determined Linearization Curve....................................................3-22
3-18 Data Points for User-Determined Linearization Coefficients..................3-23
3-19 Setup/Checkout of Linearization Function..............................................3-24
3-20 Setup/Checkout of Alarm Option............................................................3-25
3-21 Status Display ........................................................................................3-26
3-22 Setup/Checkout of Current Output Option .............................................3-27
3-23 Setup/Checkout of Auto Zero/Span Option............................................3-29
3-24 Setup/Checkout of Remote Input/Output Option....................................3-31
4-1 Functional Diagram of Detection System...............................................4-2
5-1 Troubleshooting Chart............................................................................5-5
5-2 Signal Waveforms..................................................................................5-6
5-3 Modulation System.................................................................................5-7
748176-N
July 1998
Rosemount Analytical
V
Page 8
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
FIGURES (CONTINUED)
7-1 Case Heater Temperature Control Assembly........................................ 7-2
7-2 Motor/Source Assembly......................................................................... 7-3
7-3 Model 880 Major Components............................................................... 7-4
7-4 Optical Bench Assembly........................................................................ 7-5
7-5 Detector Assembly Removal/Installation ............................................... 7-6
T
ABLES
2-1 PURGING TIME AT ATMOSPHERIC SAMPLE PRESSURE ............... 2-6
3-1 REMOTE RANGE I/O BIT DESIGNATION............................................ 3-32
3-2 REMOTE RANGE I/O BINARY AND DECIMAL BIT CODING .............. 3-32
5-1 ERROR CODE SUMMARY................................................................... 5-1
D
RAWINGS (LOCATED IN REAR OF MANUAL
623782 Schematic Diagram, Microprocessor Board 623995 Schematic Diagram, Oscillator Board 624003 Schematic Diagram, Temperature Controller 624073 Schematic Diagram, Power Supply 624085 Schematic Diagram, Signal Board 624092 Schematic Diagram, Isolated Voltage to Current 624127 Schematic Diagram, Adapter Board 624190 Installation Drawing, Model 880 624191 Pictorial Wiring Diagram, Model 880 624204 Schematic Diagram, Dual Alarm 624251 Schematic Diagram, Isolated Remote Control 624599 Schematic Diagram, Auto Zero/Span Control
)
VI
Rosemount Analytical
July 1998
748176-N
Page 9
P
REFACE
SAFETY SUMMARY
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the Model 880 Non-Dispersive Infrared Analyzer should be thoroughly familiar with and strictly follow the instructions in this manual. Save these instructions.
DANGER is used to indicate the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored
WARNING is used to indicate the presence of a hazard which can cause severe personal injury, death, or substantial property damage if the warning is ignored.
CAUTION is used to indicate the presence of a hazard which will or can cause minor personal injury or property damage if the warning is ignored.
NOTE is used to indicate installation, operation, or maintenance information which is important but not hazard-related.
WARNING: ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
Alarm and zero/span switching relay contacts wired to separate power sources must be disconnected before servicing.
This instrument is shipped from the factory set up to operate on 115 VAC, 50/60 Hz electric power. For operation on 230 VAC, 50/60 Hz electrical power, see Section 2.3 for modifications.
748176-N
July 1998
Rosemount Analytical
P-1
Page 10
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
!
This analyzer is of a type capable of analysis of sample gases which may be flammable. If used for analysis of such gases the instrument must be protected by a continuous dilution purge system in accordance with Standard ANSI/NFPA 496-1989, Chapter 8.
If explosive gases are introduced into this analyzer, the sample containment system must be carefully leak-checked upon installation and before initial startup, during routine maintenance and any time the integrity of the sample containment system is broken, to ensure the system is in leak-proof condition. Leak-check instructions are provided in Section 2.8.
Internal leaks resulting from failure to observe these precautions could result in an explosion causing death, personal injury or property damage.
WARNING: POSSIBLE EXPLOSION HAZARD
WARNING: PARTS INTEGRITY
!
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair
!
This instrument’s internal pullout chassis is equipped with a safety stop latch located on the left side of the chassis.
When extracting the chassis, verify that the safety latch is in its proper (counter­clockwise) orientation.
If access to the rear of the chassis is required, the safety stop may be overridden by lifting the latch; however, further extraction must be done very carefully to insure the chassis does not fall out of its enclosure.
If the instrument is located on top of a table or bench near the edge, and the chassis is extracted, it must be supported to prevent toppling.
!
This analyzer requires periodic calibration with known zero and standard gases. Refer to Sections 2.6, 3.7, 3.9, 3.10 and 3.11. See also General Precautions for Handling and Storing High Pressure Cylinders, following Section 7.
CAUTION: TOPPLING HAZARD
CAUTION: HIGH PRESSURE GAS CYLINDERS
P-2
Rosemount Analytical
July 1998
748176-N
Page 11
REFACE
P
S
PECIFICATIONS
C
ATALOG NUMBERS
191809 Standard Case 191810 Extended Case
POWER REQUIREMENTS:
115/230 VAC ±10%, 50/60 ±3 Hz, 150 W; 350 W with optional case heater
O
PERATING TEMPERATURE
32°F to 113°F (0°C to 45°C) Some configuration may require optional case heater
DIMENSIONS:
8.7 in (22.0 cm) H x 19 in. (483 mm) W x 19 in. (483 mm) D, standard case
8.7 in (22.0 cm) H x 19 in. (483 mm) W x 24 IN. (610 MM) D, extended case
WEIGHT:
56 lbs (25 kg), standard case 68 lbs (31 kg), extended case
R
EPEATABILITY
1% of fullscale
N
OISE
:
1% of fullscale
1
:
:
:
Z
ERO DRIFT
:
±1% of fullscale per 24 hours
S
PAN DRIFT
:
±1% of fullscale per 24 hours
R
ESPONSE TIME
: (E
LECTRONIC
)
Variable, 90% of fullscale in 0.5 sec to 20 sec, field selectable
S
ENSITIVITY
:
100 ppm fullscale carbon monoxide at atmospheric pressure 50 ppm fullscale carbon dioxide at atmospheric pressure
S
AMPLE CELL LENGTH
:
0.04 in. (1 mm) to 15.0 in. (381 mm)
M
ATERIALS IN CONTACT WITH SAMPLE
W
INDOWS
C
ELLS
T
UBING
F
ITTINGS
O-R
:
INGS
Sapphire, quartz, Irtran
:
Gold plated Pyrex or stainless steel FEP Teflon
:
316 stainless steel
:
Viton-A
:
:
2
1
Performance specifications based on ambient temperature shifts of less than 20°F (11°C) per hour.
2
Application dependent.
748176-N
July 1998
Rosemount Analytical
P-3
Page 12
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
SPECIFICATIONS (CONTINUED)
SAMPLE PRESSURE:
Max 10 psig (69 kPa), standard Pressurized application available upon request
A
NALOG OUTPUT
:
Standard: 0 to 5 VDC Optional: 0 to 20 mA/4 to 20 mA
L
INEARIZATION
:
Keypad entered coefficients for linearizing 1, 2 or all 3 ranges
ENCLOSURE:
General purpose for installation in weather-protected area. Optional purge kit per Type Z, ANSI/NFPA 496-1989.
3
3
When installed with user-supplied components, meets requirements for Class I, Division 2 locations per National
Electrical Code (ANSI/NFPA 70) for analyzers sampling non-flammable gases. Analyzes sampling flammable gases must be protected by a continuous dilution purge system in accordance with Standard ANSI/NFPA 496-1989, Chapter 8. Consult factory for recommendations.
P-4
Rosemount Analytical
July 1998
748176-N
Page 13
S
PECIFICATIONS - OPTIONS
A
LARM
R
ELAY OUTPUTS
Two single point, field programmable high or low, deadband up
:
to 20% of fullscale. Two form C contacts rated 3A-125/250 VAC or 5A-30 VDC
(resistive)
CAL GAS CONTROL
RELAY OUTPUTS:
Two front panel actuated contact closures Two form C contacts rated 3 A, 125/250 VAC or 5 A, 30 VDC
(resistive)
A
UTO ZERO/SPAN
RELAY OUTPUTS:
Four contact closures, field programmable frequency and duration of closure. Two contact closures for indication of insufficient zero and span adjustment.
Four form C contacts rated 3A-125/250 VAC or 5A-30 VDC (resistive),
Two form A contact rated (resistive load): Max switching power: 10 Watts Max switching voltage: 30 VDC Max switching current: 0.5 A
REFACE
P
R
EMOTE
I/O
R
ELAY OUTPUTS
I
NPUTS
Three remotely changeable ranges with positive identification.
:
Binary or decimal field selectable. Eight form A contacts rated (resistive load):
:
Max switching power: 10 Watts Max switching voltage: 30 VDC Max switching current: 0.5 A
Eight optical couplers Input Range: +5 VDC to +24 VDC
748176-N
July 1998
Rosemount Analytical
P-5
Page 14
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
CUSTOMER SERVICE, TECHNICAL ASSISTANCE AND FIELD SERVICE
For order administration, replacement Parts, application assistance, on-site or factory repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned Materials Authorization (RMA) number. Please have the following information when you call: Model Number, Serial Number, and Purchase Order Number or Sales Order
Number.
Prior authorization by the factory must be obtained before returned materials will be accepted. Unauthorized returns will be returned to the sender, freight collect.
When returning any product or component that has been exposed to a toxic, corrosive or other hazardous material or used in such a hazardous environment, the user must attach an appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification that the material has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
TRAINING
A comprehensive Factory Training Program of operator and service classes is available. For a copy of the Current Operator and Service Training Schedule contact the Technical Services Department at:
Rosemount Analytical Inc.
Phone: 1-714-986-7600
FAX: 1-714-577-8006
D
OCUMENTATION
The following Model 880 Non-Dispersive Infrared Analyzer instruction materials are available. Contact Customer Service or the local representative to order.
P-6
748176 Instruction Manual (this document)
Rosemount Analytical
July 1998
748176-N
Page 15
REFACE
P
Q
UICK STARTUP AND TROUBLESHOOTING GUIDE
WARNING: ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
The purpose of this guide is to give quick and simplified instructions on getting started. Some of the non-critical details may have been left out. Refer to the appropriate section in this manual for complete descriptions of all procedures. Read this guide completely before starting up the analyzer. This guide contains step-by-step instructions for startup and includes information helpful in simplifying some procedures detailed in the manual. The analyzer has been setup and tested at the factory per the specified sales order. Generally the only settings and adjustments required are calibration gas values, zero/span calibrations, and adjustments on option boards. Refer to the appropriate section in the manual for information on option boards.
OW TO USE THIS GUIDE
H
First, read Section 1, Introduction and install analyzer per Section 2 of this manual. Familiarize yourself with Keypad Operation. Step through Mode Functions. Go to Section 3, Initial Startup and continue (in sequence) through this guide.
CONTENTS
GAS REQUIREMENTS......................................................................................P-8
DISPLAYS ......................................................................................................P-8
FRONT PANEL KEYPAD...................................................................................P-9
MODE FUNCTION ............................................................................................P-10
INITIAL STARTUP..............................................................................................P-10
Range Mode ......................................................................................P-10
Range Number ......................................................................................P-10
Gain=1, 2, 4 or 8......................................................................................P-10
LINEARIZATION MODE.....................................................................................P-12
DIAGNOSTICS MODE.......................................................................................P-12
Osc Tune ......................................................................................P-12
Det Sig (with zero gas flow) .....................................................................P-12
DET SIG (with span gas flow) ..................................................................P-12
SCR CUR ......................................................................................P-12
ZERO CALIBRATION.........................................................................................P-13
SPAN CALIBRATION.........................................................................................P-14
SOURCE BALANCE ADJUSTMENT .................................................................P-14
SOURCE ALIGNMENT ......................................................................................P-15
TROUBLESHOOTING .......................................................................................P-15
TROUBLESHOOTING CHART..........................................................................P-17
COMPONENTS LOCATION...............................................................................P-19
748176-N
July 1998
Rosemount Analytical
P-7
Page 16
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
GAS REQUIREMENTS
Zero - Use a gas containing less than 25% of the measured component, preferably 0%. Span - Use a gas containing from 51 to 100% of the measured component, for each range.
80 to 100% is recommended. If calibration is attempted with a span gas less than 51% of the measured component, an error message will be displayed.
DISPLAYS
RUN M
ODE
ZERO M
Concentration, Engineering Units, or % Fullscale. If Linearization ON, reads in engineer ing units. If Linearization OFF, reads in % Fullscale.
XXX % ABC L1
XXX ppm ABC L2
XXX %FS ER4 R3
Code for measured component or Error Message
ODE
ZERO PS=X.X%
ZR=X.X PS=X.X%
Local, or Remote Control, Range 1, 2 or 3.
% of ZERO potentiometer used (Hardware pot)
P-8
Present RUN Mode value
SPAN M
Rosemount Analytical
ODE
Present RUN Mode value
SPAN PS=X.X%
XXX NN% MMM
% of SPAN potentiometer used (Hardware pot)
July 1998
% of ZERO potentiometer used (Hardware pot)
% of SPAN potentiometer used (Hardware pot)
Calibration Span gas value entered by user while in RANGE Mode. Calibration Gas value displayed if Linearizer is OFF.
748176-N
Page 17
REFACE
P
FRONT PANEL KEYPAD
The following flow diagram describes the basic functions of the f ront panel keypad, starting in RUN mode. Flow diagrams for other operations are located in this guide and Section 3.
MODE
MODE
Press ENTER to select the displayed mode.
MAIN MENU SELECTIONS
ZERO
Press SHIFT + ENTER to abort ZERO and return to RUN mode
ENTER
ZERO
SHIFT
Press or to change values up or down.
SHIFT
Press and to make large jumps horizontally from one display to the next.
ZERO CAL.
ZERO PS=XX%
ENTER
ENTER
ENTER
ZR=X.X PS=XX%
Press ENTER again to calculate ZERO and return to RUN mode
ENTER
Press to store in permanent memory the screen values for the displayed mode.
SHIFT
Press and to return to RUN mode.
Press or to change values up or down.
ENTER
SPAN
SPAN
Press SHIFT + ENTER to abort SPAN and return to RUN mode
SHIFT
RUN Mode
SPAN CAL.
SPAN PS=XX%
ENTER
ENTER
ENTER
RUN Mode
Press or to change values up or down.
XX.X XX% XX.X
Press ENTER again to calculate SPAN (software SPAN) and return to RUN mode
748176-N
July 1998
Rosemount Analytical
P-9
Page 18
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
MODE FUNCTION
MODE
SECURITY
RANGE
DIAGNOSTICS
LINEARIZER
ALARM
CURRENTM
AUTO-CAL
Press to enter displayed mode.
Press and to return to RUN mode.
ENTER
SHIFT ENTER
REMOTE IN/OUT
INITIAL STARTUP RANGE M
As a first step, check range mode settings and enter your calibration gas values. See Range Mode flow diagram on following page.
ANGE NUMBER
R
Setup desired operating range (1, 2 or 3) using up and down arrow keys. Remember to press ENTER after changing values to store data.
AIN
G
=1, 2, 4
Don't change from factory setting at this time. If it becomes necessary to make a change during calibration, use the following table as a guide:
ODE
OR
8
DIGITAL GAIN SETTINGS
DETECTOR SIGNAL (DET SIG) GAIN SETTING
<=7.5 V and >4 V 1 <=4 V and >2.2 V 2 <=2.2 V and >15 V 4 <=1.5 V 8
P-10
Rosemount Analytical
July 1998
748176-N
Page 19
REFACE
P
Note The detector signal voltages can be observed in the DIAGNOSTICS mode.
Some deviations from these recommended settings are permissible, if the instrument can't be spanned keeping the %PS reading between 5% and 95%. Additional information on the gain setting is located in Section 5.7.
MODE
RANGE 1
Set to 1, 2, 4 or 8 (see table
under Digital Gain Settings)
Fullscale range in % or ppm
SECURITY
RANGE
ENTER
RANGE 2 RANGE 3 CMP ABC
SHIFT
GAIN=X
SHIFT
FS=XX % or ppm
SHIFT
ENTER
ENTER
Take %FS value from non-linear
CAL GAS=XXX.X %FS
calibration curve
Zero Gas Offset as a % of fullscale
Time Constant: 0.5 to 20 sec.
ZR. OFFSET=0.00 %FS
TIME CONST=XX
Move one character at a time. If last character, moves to next screen.
SHIFT
ENTER
Move one screen at a time.
Press to store, accept
ENTER
SHIFT
ENTER
SHIFT
748176-N
July 1998
Rosemount Analytical
P-11
Page 20
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
LINEARIZATION MODE
Verify that LINEARIZER ON/OFF is in the desired position. Verify the analyzer coefficients for each range against the values supplied on the factory
calibration sheet in the rear of this manual.
Note The linearizer range positions 1, 2 and 3 allow setting coefficients for these ranges.
When the operating range is selected from the "Range" mode, the correct coefficients for that range will be selected regardless of which linearization range settings (1, 2 or 3) was used in the linearization mode. Linearization position A is an "ALL" range position, using coefficients from range 3 for all ranges. Position A can only be used if the range to range ratio is equal to or less than 3:1.
DIAGNOSTICS MODE
If the display is unstable (oscillating) in OSC, TUNE, DET SIG and/or SCR CURRENT, there probably is an ER4 error message. Refer to the Troubleshooting chart.
OSC TUNE - Tune to 75 to 80% of maximum. Turn counterclockwise from maximum.
DET SIG (
WITH ZERO GAS FLOW
) - Must be between 0.4 to 1.2 volts. If <0.4 volts: Re-
adjust source balance per Section 3.5. If >1.2 volts: See Troubleshooting Chart, Source Balance and Source Alignment.
DET SIG (
WITH SPAN GAS FLOW
) - Adjust detector signal voltage with GAIN potentiometer
R3 on the Signal Board (see Drawing 624055 in the rear of this manual). For Range 3 (least sensitive) adjust this voltage as close to 7.5 volts as possible. Do not exceed 7.5 volts. Must be between 0.4 to 1.2 volts. For low ppm applications the maximum attainable voltage may be considerably below 7.5 volts. If span gas is less than 100% fullscale, the detector voltage should be set roughly equal to (7.5 volts) X (% fullscale read off of the non­linear curve)/100.
Detector signal voltages with span gas flowing will dictate the digital GAIN settings. (see table under Range Mode, Gain.)
SRC CUR - Adjust Source Current with R9 on the Power Supply Board. Source current will be between 700 to 950 mA. The source will have a longer life on lower currents. Current setting will be lower on high concentration, short cell path applications, such as % level measurements. Will be higher on low concentration, long cell path measurements, such as 0 to 10 ppm CO2.
P-12
Rosemount Analytical
July 1998
748176-N
Page 21
REFACE
P
MODE
75 to 80% of maximum.
De-tuned in counterclockwise direction.
0.4 to 1.2 volts with zero gas flow.
1.2 to 7.5 volts with span gas flow.
7.5 volts maximum on highest range.
Must be between 700 and 950 mA. 700 to 800 mA for % to high ppm ranges. 800 to 950 mA for mid to low ppm.
SECURITY
RANGE
DIAGNOSTICS
OSC TUNE
DET SIG
SRC CUR
SHIFT
ENTER
RUN MODE
ENTER
ENTER
ENTER
ENTER
ZERO CALIBRATION
There is one hardware zero for all three ranges, and three software zeros, one for each range. Once the hardware zero has been adjusted for one of the ranges, IT SHOULD NOT BE CHANGED when doing software zero on the other two ranges. See flow chart below.
If problems are encountered, such as error messages (ER0, ER1, ER2, ER3 or ER4) go to DIAGNOSTICS mode to locate the cause of the problem. Use the Troubleshooting chart in this section, as well as Section 5, Troubleshooting.
ZERO
Using up/down arrow keys, adjust hardware zero (ZR=X.X) close to zero. Does not have to be exactly zero.
Calculates software zero, then returns to RUN mode.
ZERO PS=XX%
ZR=X.X PS=XX%
CALCULATING ZERO
XXX% or ppm ABC L3
RUN MODE
PS stands for % of digital potentiometer used.
ENTER
Limits are 5 to 95%. Press ENTER before adjusting.
ENTER
Note: Use up/down arrow keys on one range only. Take a software zero on the other ranges.
748176-N
July 1998
Rosemount Analytical
P-13
Page 22
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
SPAN CALIBRATION
There are three hardware spans and three software spans. Use the flow diagram as a guide.
If problems are encountered, such as error messages (ER0, ER1, ER2, ER3 or ER4) go to DIAGNOSTICS mode to locate the cause of the problem. Use the Troubleshooting chart in this section, as well as Section 5, Troubleshooting.
SPAN
XX.X = Actual reading in ppm, % or %FS. XX% = Percent of span potentiometer. XX.X = Span gas engineering units or non­linear value taken from curve.
Calculates software zero, then returns to RUN mode.
See Troubleshooting if error message appears. SHIFT+ENTER will abort function and return to RUN mode.
OURCE BALANCE ADJUSTMENT
S
SPAN 3 PS=XX%
XX.X XX% XX.X
CALCULATING SPAN
XXX% or ppm ABC L3
RUN MODE
PS stands for % of digital potentiometer used.
ENTER
Limits are 5 to 95%. Press ENTER before adjusting.
ENTER
Adjust hardware span with up/down arrow keys. Adjust until actual reading is close to, but less than non-linear value. Non-linear value must be over 51% of fullscale and must be <= fullscale range. Up/down arrow keys can be adjusted for all three ranges before taking software span.
This procedure should not be necessary on a new analyzer unless problems are encountered.
1. Access the DET SIG display in DIAGNOSTICS mode.
2. Flow zero gas through the sample cell for a minimum of two minutes.
3. Slightly loosen the hex standoff on the sample cell adjust screw.
4. Rotate the shutter adjust screw until a minimum reading is obtained.
5. Add 0.4 to 0.5 volts to this reading. If this reading exceeds 1.2 volts, a source alignment is required.
6. Rotate the shutter adjust screw clockwise until the display reads the value obtain in step
5.
7. Re-tighten the hex standoff. Verify that the display does not change.
8. Close analyzer. Allow one hour for analyzer readings to stabilize. If zero gas reading has changed appreciably, re-balance the analyzer.
P-14
Rosemount Analytical
July 1998
748176-N
Page 23
REFACE
P
OURCE ALIGNMENT
S
Source balance adjustment must be performed prior to source alignment. This procedure is necessary when the detector signal is greater than 1.2 volts after step 5 in
the source balance adjustment procedure.
1. Connect a DVM between TP6 and TP2 on the Signal Board.
2. Set the source balance screw halfway in. (1/2 inch of the threads are visible.)
3. Loosen the two screws holding each source in place.
4. Adjust both measurement and reference sources up and down to reach the minimum detector signal. Tighten screws.
5. Perform source balance adjustment procedure above.
ROUBLESHOOTING
T
If problems are encountered, such as error messages ER0, ER1, ER2, ER3 or ER4, go to the DIAGNOSTICS mode (page P-12) to assist in locating the cause of the problem.
Use the following Troubleshooting Chart as well as Section 5 Troubleshooting.
SPAN GAS ERRORS ER1, ER2 OR ER3
Check RANGE, CAL GAS, CL GAS
CAL GAS (Span gas) must be <=FS (fullscale range).
Span gas must be 51 to 100% of the fullscale range.
CL GAS must be read off of the non-linear curve.
The RUN mode concentration reading must be less than the CL GAS value. (See SPAN
Mode Display flow diagram, page P-8)
Note Spanning with Linearizer ON If spanning with linearizer ON, RUN MODE values will be in engineering units, and
CAL GAS values will appear in the right-side display. The microprocessor will be using the non-linear values for the initial stages of
calibration. It may calculate a CAL GAS value that is higher than the entered value. An error message will be displayed if this occurs.
748176-N
July 1998
Rosemount Analytical
P-15
Page 24
ODEL
M
880 NON-D
NY OF THE FOLLOWING MAY CORRECT THE PROBLEM
A
ISPERSIVE INFRARED ANALYZER
:
1. Re-check CAL GAS and CL GAS values entered in RANGE mode.
2. Verify that the NON-LINEAR curve is read for CL GAS entry.
3. During hardware span adjust, make the run value lower than the CAL value.
4. Re-calibrate with the linearizer OFF.
RUN M
ODE VALUE TOO LOW DURING HARDWARE
SPAN
ADJUSTMENT
Hardware potentiometer is adjusted to 95% and RUN Mode value still too low;
Check SPAN GAS bottle for correct concentration and flow
Check for leaks
Check DET SIG adjustments
Check DIGITAL GAIN adjustments
RUN M
ODE VALUE TOO HIGH DURING HARDWARE
SPAN
ADJUSTMENT
Hardware potentiometer is adjusted to 5% and RUN Mode value still higher than CL or CAL GAS display when software zero taken by pressing ENTER the error message appears.
Check SPAN GAS bottle for correct concentration and flow
Check DET SIG adjustments
Check DIGITAL GAIN adjustments
P-16
Rosemount Analytical
July 1998
748176-N
Page 25
REFACE
P
ROUBLESHOOTING CHART
T
Prior to troubleshooting, verify that OSC TUNE (oscillator tune) is properly adjusted. If reading is oscillating (and can't be correct by OSC TUNE adjustment), check DET SIG and digital GAIN adjustment.
ZERO ER0
Run ZERO Calibration (page P-13), re-adjust hardware zero
Are you unable to adjust any further (at PS=5 or 95% limits)? Have you tried adjusting the RUN mode reading (changing the hardware zero
potentiometer) but are unable to bring the reading close enough to zero? When ENTER is pressed to calculate software zero, does ER0 error message
appear?
YES
Is DET SIG >1.2 volts? Is DET SIG <0.4 volts?
YES
NO
NO
Check test points with DVM. Replace boards.
Is application for CO2 and is CO2 range <= 20 ppm @ 1 atmospheric pressure? (Reads negative, then positive on upscale gas.)
NO
YES
Is CO2 free purge connected and flowing for reference cell and
NO
source shroud?
Connect CO2 free purge with 10
YES
psig pressure. Check flow, should be 50 cc/min.
Re-balance Sources
YES
748176-N
Can't get DET SIG below 1.2 V Re-align Sources
YES
July 1998
Rosemount Analytical
P-17
Page 26
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
ER4
Digital error - Generally caused by combination of too high an analog signal for the digital gain setting used.
Check DET SIG with span gas flowing.
Are readings too unstable to read (oscillating)?
YES
Check both OSC TUNE and SRC CURRENT in DIAGNOSTICS mode (page P-12).
Are readings oscillating?
YES
Connect a DVM to TP2 (detector signal) and TP6 (ground).
Check/adjust analog detector signal gain, R3 (page P-10)
Re-Zero (page P-8)
NO
Read DET SIG in DIAGNOSTICS mode (page P-12)
P-18
Re-Span (page P-8)
Rosemount Analytical
July 1998
748176-N
Page 27
REFACE
P
EADS WRONG WAY ON UPSCALE GAS
R
Reads downscale, then upscale on low flows of upscale gas (the n egative swing might
be missed if flow is too fast) This is caused by imbalance in reference/measuring energy to detector.
Check for purge for <= 20 ppm CO2 applications
Check source balance (page P-14)
OMPONENT LOCATION
C
Display Contrast Adjust R8
Reset SW1
Micro Board
Power Supply
Gain Adjust R3
Signal Board
Source Current Adjust R9
Oscillator Tune
Detector Module
Option Boards
Motor/Source
Source Balance Adjust and Locknut
748176-N
July 1998
Rosemount Analytical
P-19
Page 28
ODEL
M
N
OTES
880 NON-D
ISPERSIVE INFRARED ANALYZER
P-20
Rosemount Analytical
July 1998
748176-N
Page 29
I
NTRODUCTION
1
1.1 DESCRIPTION
The Model 880 Non-Dispersive Infrared Analyzer is designed to continuously determine the concentration of a particular component of interest in a flowing gaseous mixture. Within the analyzer, two equal energy infrared beams are directed through two parallel optical cells, a flow-through sample cell and a reference cell. The Luft detector continuously measures the difference in the amount of infrared energy absorbed within each of the two cells. This difference is a measure of the concentration of the component of interest in the sample. Readout is on the 16 character, backlit liquid crystal display in parts per million. percent of composition or percent of fullscale. Additionally, a 0 to +5 VDC output for a potentiometric (voltage) recorder is provided as standard.
By turning the linearizer ON and entering linearizing coefficients, a calibration curve may be used to convert display or recorder readings into linearized engineering units. When this f eature is ON, the analyzer utilizes a linearizing function for linear readout of concentration values on the display and on a recorder. A diagnostics mode is provided as standard.
1.2 AVAILABLE OPTIONS
Operation of the Model 880 can be enhanced with the choice of several options:
UAL ALARMS
D
User-set dual alarms are available with configurable HI/LO designations and deadband.
SOLATED CURRENT OUTPUT
I
For normal usage, the 0 to 20 mA or 4 to 20 mA current output can be set to represent 0 to 100% of fullscale, or a suppressed range of 25% or more of fullscale may be selected.
UTO ZERO/SPAN
A
An Automatic Zero/Span is available for unattended calibration of all three ranges.
748176-N
July 1998
Rosemount Analytical
1-1
Page 30
M
ODEL
880 NON-D
ALIBRATION GAS CONTROL
C
ISPERSIVE INFRARED ANALYZER
A Calibration Gas Control allows two solenoids to be remotely actuated from the front panel, enabling one-man calibration without leaving the analyzer.
EMOTE RANGE
R
I/O
An optional remote range input/output is available.
ASE HEATER
C
A proportional temperature controller with fan assembly maintains proper operating temperature inside the case.
URGE KITS
P
An air purge kit, when installed with user-supplied components, meets Type Z requirements of standard ANSI/NFPA 496-1989 for installation in Class I, Division 2 locations as defined in the National Electrical Code (ANSI/NFPA 70) when sampling non-flammable gases. If analyzer is used to sample a flammable gases it must be protected by a continuous dilution purge system per standard ANSI/NFPA 496-1989, Chapter 8. Consult factory for further information.
1-2
Rosemount Analytical
July 1998
748176-N
Page 31
U
NPACKING AND INSTALLATION
2
2.1 UNPACKING
Examine the shipping carton carefully. If there are any signs of damage, notify the carrier immediately. Open the carton and inspect the contents for signs of damage. If there is concealed damage, save the carton and packing material and notify the carrier.
2.2 LOCATION
Locate the analyzer in a weather-protected, non-hazardous location away from vibration. For best results mount the analyzer near the sample stream to minimize sample-transport time. Refer to Installation Drawing 010-624190.
If equipped with P/N 624446 optional air purge kit and installed with user-provided components per Instructions 015-748157, the analyzer may be located in a Class I, Division 2 area as defined by the National Electrical Code (ANSI/NFPA 70). This kit is designed to provide Type Z protection in accordance with Standard ANSI/NFPA 496-1989, Chapter 2, when sampling nonflammable gases. For flammable samples the instrument must be equipped with a continuous dilution purge system in accordance with ANSI/NFPA 496-1989 Chapter 8 or IEC Publication 79-2 (1983) Section Three. Consult Factory for recommendations on sample flow limitations and minimum purge flow requirements for your particular application.
2.3 VOLTAGE REQUIREMENTS
WARNING: ELECTRICAL SHOCK HAZARD
For safety and proper performance, this instrument must be connected to a properly grounded three-wire source of electrical power.
This instrument was shipped from the factory set up to operate on 115 VAC, 50/60 Hz electric power. For operation on 230 VAC, 50/60 Hz the installer must position voltage select switches S1 and S2 located on power supply board and switch S3 located on case temperature heater assembly (optional) to the 230 VAC position (refer to Figures 2-1 and 2-2).
748176-N
July 1998
Rosemount Analytical
2-1
Page 32
M
G
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
Power consumption is less than 150 watts without optional case heater or 350 watts with optional case heater.
Set switch windows for v o ltage required
S2 S1
J16
1
115 115
TEMP CONTROL CASE
J5
1
R11
S2 S1
230V
115V
DETECTOR
HEATER
J4
1
K1
4 3 2 1
TP1
C12
C11
C9
R12 R13
+
+
C10
+
+
TP2
C6
C8
+
C5
C7
+
R1 R2
S3
SS LUFT
8
1
2
7 6
3
45
+
C4
+
CR4
CR3
CR2
+
C2
+
+
C3
+
+
C1
FAN
J2
1
VR2
VR4
G
O
VR3
I
O
I
I
G
O
VR1
I
G
G
O
R3 R10
R8 R4 R5 R6 R7
AR1
- CR1 +
R9
R37 R34 R35
R36
C14
C13
U1
+
+
C
E
Q3
AR4
C19
J
U2
AR5
CR8
R14
E
C
AK
R21
10
C21
R 38
R17
1
R25 R29 R26
B Q2 C15
+
R24 R23
Q1
R22
Q4
R19
R15
C20
C16
R31 R30
R27 R28 R32
CR6 CR7
R33 B
CR5
J14
1
J11
J8
1
J15
1
1
J13
F
IGURE
2-1. P
655135 POWER SUPPLY BD
OWER SUPPLY BOARD
SENSOR J18
400A 880 951E
R10 R11 R7 R8
C2
CR1
C
B
Q2
R17R16 R12 CR2
R4
R3
C1
E
+
Q1
K
G
A
TEMP CONTROL BD
1
J7
Heater LED (CR5)
Set switch w in do w fo r volta g e required
S3
POWER SUPPLY
1
R2R1
R6
C3
AR1
R13
R9 R5
CR3
1
E
B
Q3
TEST
C
J19
J11
R15
POWER LINE J5
C4
3 2 1
R14
1
1 2 1 2 3
T.I.F. HEATER
U2
2
3
1
U1
S3
230
115
J17
115
F
IGURE
2-2
2-2. C
ASE HEATER TEMPERATURE CONTROL BOARD
Rosemount Analytical
July 1998
748176-N
Page 33
NPACKING AND INSTALLATION
U
PURGE GAS
OUTLET
Auto Zero/Span Control Board Isolated V/I Board Dual Alarm Board Isolated Remote Control Board Calibration Gas Control Board
OPTION BOARDS INSTALLED WITH COMPONENT SIDE TO THE LEFT.
REFERENCE
OUT IN
SAMPLE
OUT IN
OPTIONS CONNECTIONS
RECORDER
OUTPUT POWER
F1 4A RESETTABLE CIRCUIT BREAKER
+ - GND
TB2
PURGE GAS
TB1
GND
NEUT/L2
HOT/L1
Recorder Output Hook-Up
IN
Power Hook-UP
F
IGURE
F
IGURE
2-3. R
2-4. C
EAR VIEW OF MODEL
INTERIOR EXTERIOR
N u t G la n d N u t
Cable
ABLE GLAND
880
Case Wall
2.4 ELECTRICAL CONNECTIONS
The power, recorder and option board cable glands are shipped loose in the shipping kit to allow cable installation to connectors or terminal strips. Route the cable through the cable gland parts and connect to a connector or terminal strip as shown in Figure 2-4, then tighten the gland.
CABLE GLAND PART NUMBER
P
OWER
899330
R
ECORDER
O
PTION BOARD
Remove the rear cover to access the terminals.
748176-N
899329 899329
July 1998
Rosemount Analytical
2-3
Page 34
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
2.4.1 L
INE POWER CONNECTIONS
If this instrument is located on a bench or table top or is installed in a protected rack, panel or cabinet, power may be connected to it via a 3-wire flexible power cord, minimum 18 AWG (max. O.D. 0.480", min. O.D. 0.270") through hole "H" (refer to Drawing 010-624190) utilizing the connector gland (P/N 899330) provided in the ship kit. Accessory kits are available which include a 10 foot North American power cord set plus four enclosure support feet for bench top use (P/N 654008) or the power cord only (P/N 634061) or the four feet only. If the instrument is permanently mounted in an open panel or rack, use electrical metal tubing or conduit.
Refer to Figure 2-4. Route the power cable through the cable clamp and connect the leads to TB1. Tighten the cable clamp after connecting the leads. Since the rear terminals do not slide out with the chassis. no excess power cable slack is necessary.
2.4.2 R
ECORDER CONNECTIONS
Recorder connections are made to the rear panel. Refer to Drawing 624190. Route the recorder cable through the cable clamp and connect the end to TB2.
Recorder Interconnection Cable:
Distance recorder to analyzer - maximum 1000 feet (305 meters). Input impedance - greater than 5000 ohms. Customer supplied two-conductor shielded cable, 20 AW G min. Voltage output: 0 to +5 VDC.
2.5 SAMPLE AND REFERENCE INLET/OUTLET CONNECTIONS
The standard Model 880 is intended for atmospheric pressure operation only and must be vented to atmosphere. Pressurized cells are available for special applications. Sample inlet and outlet and, when used, flowing reference inlet and outlet connections are located on the rear panel. All connections are 1/4-inch bulkhead fittings.
2.6 CALIBRATION GAS REQUIREMENTS
Analyzer calibration consists of setting a zero point and one or more upscale points. All applications require a zero standard gas to set the zero point on the display or
recorder chart. If the factory Calibration and Data Sheet (included with the drawings at the end of the manual) specifies a background gas, use this as the zero gas. If a background gas is not specified, use dry nitrogen for the zero gas.
2.7 SAMPLE HANDLING SYSTEM
Many different sample handling systems are available, either assembled completely or as loose components. The type used depends on the requirements of the particular application and the preferences of the individual user. Typically, the sample handling
2-4
Rosemount Analytical
July 1998
748176-N
Page 35
NPACKING AND INSTALLATION
U
system incorporates such components as pumps and valves to permit selection of sample, zero standard, or upscale standard gas; needle valve in sample-inlet line for flow adjustment; flowmeter for flow measurement and/or indication of flow stoppage; and filter(s) to remove particulate matter.
2.8 LEAK TEST PROCEDURE
!
This analyzer is of a type capable of analysis of sample gases which may be flammable. If used for analysis of such gases, the instrument must be either in an explosion-proof enclosure suitable for the gas, or protected by a continuous dilution purge system in accordance with Standard ANSI/NFPA-496-1986 (Chapter 8) or IEC Publication 79-2-1983 (Section Three).
If explosive gases are introduced into this analyzer, the sample containment system must be carefully leak checked upon installation and before initial startup, during routine maintenance and any time the integrity of the sample containment system is broken, to ensure the system is in leak proof condition.
Internal leaks resulting from failure to observe these precautions could result in an explosion causing death, personal injury or property damage.
The following test is designed for sample pressure up to 10 psig (69 kPa).
1. Supply air or inert gas such as nitrogen at 10 psig (69 kPa) to analyzer via a flow indicator with a range of 0 to 250 cc/min and set flow rate at 125 cc/min to the sample inlet.
WARNING: POSSIBLE EXPLOSION HAZARD
2. Seal off with a cap.
3. Use a suitable test liquid such as SNOOP® (P/N 837801) to detect leaks. Cover all fittings. seals. or possible leak sources.
4. Check for bubbling or foaming which indicates leakage and repair as required. Any leakage must be corrected before introduction of sample and/or application of electrical power.
5. If the instrument incorporates a flow-through reference, the above test must be repeated for that gas system.
Note Do not allow test liquid to contaminate cells or detector and source windows.
Should this occur, the cells should be cleaned (Section 6.1).
748176-N
July 1998
Rosemount Analytical
2-5
Page 36
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
2.9 SAMPLE FLOW RATE
For best results, the sample flow rate must be in the range of 1 to 2 SCFH (500 to 1000 cc/min). A subnormal flow rate will result in an undesirable time lag. However, an excessive flow rate will result in cell pressurization.
Assume that two cell volumes are required to flush any cell. Table 2-1 indicates approximate flushing time at atmospheric sampling pressure, i.e., the outlet of the cell venting to atmosphere for various cell lengths.
Flushing time is inversely proportional to flow rate. The primary effect of flow rate, other than flushing time, is cell pressure. Due to the
restriction of the exit tubing, an increasing flow rate increases sample pressure in the cell. For a 9-inch (232 mm) cell venting to atmosphere, the cell pressure rises from 0 psig (0 kPa) at no flow, essentially linearly, by 1 mm Hg per CFH flow up to at least 20 CFH (10 L/min).
TIME FOR 2
CELL LENGTH MM INCH
CELL VOLUME IN CC
WITHOUT INLET
TUBE
TOTAL VOLUME IN CC
CELL WITH INLET
TUBE
VOLUMES
AT 2 SCFH (1 L/MIN)
AT 750 MM HG
3 0.118 0.85 12 2 sec. 4 0.157 1.14 12 2 sec.
8 0.315 2.28 13 2 sec. 16 0.630 3.56 16 2 sec. 32 1.25 9.12 20 2 sec. 64 2.52 18.24 25 3 sec.
128 4.03 35.48 44 3 sec. 232 9.13 65.12 73 6 sec. 343 13.50 97.76 105 13 sec. 381 15.00 108.60 116 14 sec.
T
ABLE
2-1. P
URGING TIME AT ATMOSPHERIC SAMPLE PRESSURE
At 7.5 to 8.0 CFH (3.8 to 4 L/min), therefore, the pressure is increased by about 1%, and the output signal is thereby increased by about the same 1% over static conditions. In all cases, the effect of pressure on readout is eliminated if the same flow rate is used for the measured sample as well as for the zero gas and span gas.
Note that at higher flow rates the non-linearity of the calibration curve increases, because of increase in sample cell pressure. Therefore, if higher flow rates are required, the calibration curve should be redrawn at the higher rate.
2-6
Rosemount Analytical
July 1998
748176-N
Page 37
NPACKING AND INSTALLATION
U
At 2 CFH (1 L/min) gaseous sample temperatures are equilibrated to instrument temperature regardless of stream temperature. At extremely high flow rates, this may not be true, but no such effect has been noted up to 18 CFH (9 L/min).
2.10 DIFFERENTIAL ANALYSIS WITH FLOW-THROUGH REFERENCE CELL
In some applications the analyzer is used to measure the difference between the concentration of the component of interest in two sample streams. If so, the reference side of the analyzer, as well as the sample side, utilizes a flow-through cell. The sample cell receives the sample stream, which contains the higher concentration of the component of interest. The reference cell receives the stream containing the lower concentration of this component. The flow rate through both cells should be monitored with a flowmeter and kept the same.
2.11 OPTION BOARDS
The following option boards may be ordered factory installed, or may be ordered as kits from the factory at a later date: Alarm, Current Output, Calibration Gas Control, Auto Zero/Span and Remote Range I/O. The boards are equipped with mating plugs for field wiring attached to the connector at the edge of each board. Attach the cable (customer supplied) to the plug and socket connector according to the schematic for each option board.
If an option board has been ordered installed at the factory, this board will be in one of the five slots inside the rear of the analyzer. Each option will require a cable (user-provided) which connects to a female plug. The female plug comes attached to the appropriate terminal block on the option board. If the instrument came equipped with one option, the interconnect cable will be in place for all options.
The Alarm, Auto Zero/Span, Calibration Gas Control and Remote Range Change Boards have jumper-selectable addresses (Figures 2-5, 2-7, 2-8 and 2-9).
2.11.1 A
LARM CONNECTIONS
Refer to Drawing 624204 and Figure 2-5. Connect cable (customer supplied) to the 6-pin connector J2. The Dual Alarm Option consists of two form C contacts rated 3A­125/250 VAC or 5A-30 VDC (resistive) .
Run the cable through the cable gland and tighten once attached to connector (Figure 2-4).
This board may be configured to provide any one of three functions:
Fail Safe Alarm - Jumpers E1, E2, E6, E7, E8, E10 Dual Alarm - Jumpers E1, E2, E5, E7, E9, E10 Calibration Gas Control - Jumpers E1, E4, E5, E7, E9, E10.
748176-N
The hook wiring remains the same on each.
July 1998
Rosemount Analytical
2-7
Page 38
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
Outlet Cable
J2
R5 R4 R3
A
C
K1
624419 CTR L
B B
CAL
E
Q1
FT2FT1
A
E
Q2
C
624207 ALA RM
R7
DUAL
C
B
C
B
E8
E6
K2
E4 E2 E 1
R1 R2 R8 R6
CR1
1
1
CR2
1
E10
E9
E7
1
E5
1
FAIL
654398 SAF E A LAR M
C1
PR1
C1
U1
U2
C3
U3
U4
J1
1
+
Jumper Selectable Address
Interconnect
R9
Cable
Note: The Dual Alarm, Fail Safe Alarm and Calibration Gas Control use the same board. However, the jumpers locations are different.
Cal Gas Control: E1, E4, E5 - E7 and E9 - E10 Dual Alarm: E1, E2, E5 - E7 and E9 - E10 Fail Safe Alarm: E1, E2, E6 - E7 and E8 - E10
F
IGURE
2.11.2 C
2-5. C
ALIBRATION GAS CONTROL AND ALARM CONNECTIONS
URRENT OUTPUT OPTION CONNECTIONS
Refer to Drawing 624291 and Figure 2-6. Connect cable (customer supplied) to the 2-pin connector J2. The voltage-to-current board has a fixed address at the top of the board. Run the cable through the cable gland and tighten once the connector has been made up (Figure 2-4).
Outlet Cable
C1 C2
SPAN
R1
OFFSET
R2
+ CR1
J2
1
1
VR1
VR2
AR1
624095 ISOLATED V TO I BD
U3
1
O
G
I
O
G
I
I
C3
1
R6 R5 R4 R3
1
C4 C 5 C 6
U4
U1
U2
C8 +
++
C7
J1
1
Interconnect Cable
F
IGURE
2-8
2-6. C
URRENT OUTPUT CONNECTIONS
Rosemount Analytical
July 1998
748176-N
Page 39
NPACKING AND INSTALLATION
U
2.11.3 C
ALIBRATION GAS CONTROL CONNECTIONS
Refer to Drawing 624204 and Figure 2-5. Connect cable (customer supplied) to the 6-pin connector J2. The Calibration Gas Control Option consists of two form C contacts rated 3A-125/250 VAC or 5A-30 VDC (resistive).
Run the cable through the cable gland and tighten once the connector has been made up (Figure 2-4).
2.11.4 A
UTO ZERO/SPAN CONNECTIONS
Refer to Drawing 624202 and Figure 2-7. Connect cable (customer supplied) to the 9-pin connectors J2 and 13. The Auto Zero/Span Option consists of four form C contacts rated 3A-125/250 VAC or 5A-30 VDC (resistive) and two form A contacts rated at 10 watts maximum switching power, 200 VDC maximum switching voltage and 0.5 A maximum switching current.
Run the cable through the cable gland and tighten once the connector has been made up (Figure 2-4).
If installed, this board can also be activated from the keyboard (Zero/Span) for the selected range.
F
IGURE
Outlet Cable
2-7. A
FT1 K1
J2
FT2 K2
FT3
J3
C
Q1
CR1
C
Q2
CR2
C
Q1
CR3K3R6
B
E
R4
B
E
R5
B
E
UTO ZERO/SPAN CONNECTIONS
Jumper Selectable Address
K4FT4
B
B
R8
RP1
E
R10
E
1
U3
U1
C3
1
1
R1R1 R2 R3
1
U2
K6
K5
C
Q5
CR5
C
Q1
C
Q4
CR4
B
E
C1
R7
E4 E2 E1
C2
+
J1
1
Interconnect Cable
1 U4
748176-N
July 1998
Rosemount Analytical
2-9
Page 40
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
2.11.5 R
EMOTE INPUT/OUTPUT CONNECTIONS
Refer to Drawing 624249 and Figure 2-9. Connect cable (customer supplied) to the 9-pin connectors J2 and J3.
The signal output is at J2 which consists of eight form A contacts rated (resistive load) 10 watts, maximum switching power, 200 VDC maximum switching voltage and 0.5 A maximum switching current.
The signal input is at J3 which consists of eight opto-couplers, operated from a user-supplied 24 VDC power source.
Run the cable through the cable gland and tighten once the connector has been made up (Figure 2-4).
Jumper Selectable Address
CR1 R13
E4 E2
C5
E1
U1
1
K5
RP2
C4
U7
C3
R12
C1
J1
+
1
Outlet Cable
J2
E5 E6 E7
R11
R2 R1
E9E8
K1
Interconnect Cable
F
IGURE
2-9. R
J3
K2
K3
K4
R3 R4 R5 R6
624254 654416 ISOLA TE D R EM O T E C O NT R OL B D
K6
K7
K8
R7 R8 R9
R10
C2
1
11
EMOTE INPUT/OUTPUT CONNECTIONS
1
1
1
U2
U3
U4
RP1
U5U6
2-10
Rosemount Analytical
July 1998
748176-N
Page 41
NPACKING AND INSTALLATION
U
2.12 ORDERING OPTION KITS
Options not ordered from the factory at the time of purchase may be ordered as the following kits:
624422 Isolated Remote Control Kit 624423 Dual Alarm Kit 624424 Auto Zero/Span Control Kit 624425 Isolated Current Output Kit 624426 Calibration Gas Control Kit
The option kit consists of the circuit board, a cable gland and two circuit card guides which press into predrilled holes in the card cage. Mount the option in the board as shown and follow the wiring directions in section 2.4. There are five connectors on the interconnect cable. It is important for the slot to be connected to the correct connector on the interconnect cable.
To install any of the above kits, the Common Parts Kit, P/N 624414, must be ordered. if not originally ordered with the analyzer. This kit consists of a card cage which mounts in the rear as shown in Figure 2-3 and three interconnect cables that plug in as shown on DWG 026-624191. Once this kit is installed, it need not be ordered again for other kits.
748176-N
July 1998
Rosemount Analytical
2-11
Page 42
ODEL
M
880 NON-D
NOTES
ISPERSIVE INFRARED ANALYZER
2-12
Rosemount Analytical
July 1998
748176-N
Page 43
I
NITIAL STARTUP AND CALIBRATION
3
Prior to shipment this instrument was subjected to extensive factory performance testing, during which all necessary optical and electrical adjustments were made. The following instructions are recommended for initial startup and subsequent standardization of the analyzer.
3.1 LEAK TEST
Perform the Leak Test Procedure in Section 2.8.
3.2 POWER VERIFICATION
1. Verify power switch settings are for available power (115 VAC/230 VAC). Refer to Section 2.
2. Apply power. On the Power Supply Board, verify that heater LED (CR5) is ON. Refer to Figure 2-1 and Drawing 624073.
3.3 SOFTWARE VERSION
When power is first applied to the Model 880 analyzer, the display will read [INITIALIZING]. Next, the display will show the current software revision number, [VERSION 2.XX]. This manual is intended only for use with instruments with a software version of 2.0 or greater.
3.4 FRONT PANEL INDICATORS AND CONTROLS
3.4.1 D
ISPLAY
The display consists of a 16-character, backlighted liquid crystal display. The contrast on the display may be adjusted so that the display can be read from any vertical angle. This adjustment is made by loosening the two screws on the front of the case and sliding the front panel forward (up to 8.2 inches), then turning the potentiometer (R8) adjust the contrast (Figure 3-1) until the best view of the display is obtained.
In the normal RUN mode of operation, the display will show current process value, component name, control mode and range. In other modes, relevant information will be displayed as is necessary.
748176-N
July 1998
Rosemount Analytical
3-1
Page 44
M
ODEL
880 NON-D
Display Contrast Adjust R8
Reset SW1
ISPERSIVE INFRARED ANALYZER
Isolated Current
Jumper/Test Point
Power Supply
Micro Board
Gain Adjust R3
Signal Board
Source Current Adjust R9
Oscillator Tune
Detector Module
Option Boards
Motor/Source
Source Balance Adjust and Locknut
F
IGURE
3.1 M
ODEL
880 A
DJUSTMENTS
ZERO F1
SPAN F2
STATUS MODE SHIFT ENTER
Rosemount Analytical
F
IGURE
3-2
3-2. M
Rosemount Analytical
ODEL
880 K
July 1998
EYPAD
748176-N
Page 45
NITIAL STARTUP AND CALIBRATION
I
3.4.2 F
UNCTION KEYS
The Model 880 has twelve function keys (Figure 3-2). Each key must be pressed firmly to insure that the microprocessor recognizes the keystroke. The definitions for these keys are as follows:
ZERO
SPAN
STAT US
SHIFT
To activate the manual zero calibration of the analyzer.
To activate the manual span calibration of the analyzer.
To display the configuration and the status of alarms and error messages.
Used in conjunction with left and right or up and down arrows, F1, F2 and ENTER keys. Pressing the SHIFT key in any display except Run Mode, Zero
Setting, Span Setting and Status causes a to be displayed at the far right hand position. Pressing will then move the cursor 16 characters to the right, pressing will move the cursor 16 characters to the left, and, if a displayed parameter is being modified, pressing will access the highest value allowed for that parameter and pressing will access the lowest value allowed for that parameter.
Software programmable keys for quick access to mode functions.
F1
F2
When used in conjunction with the SHIFT key, two additional functions
are available: SHIFT/F1 and SHIFT/F2. The computer acknowledges the keystrokes by flashing [** KEY SAVED **] on the display. These four functions can immediately access a particular display for the following modes: Range, Diagnostics, Linearization, Alarm, Current Output, Auto Zero/Span or Remote Range I/O.
MODE
To display instrument functions. The standard functions are security, range, diagnostics, and linearization. Additional functions (in conjunction with option boards) are Auto Zero/Span, Remote Range I/O, Current Output, and Alarm.
The up and down arrow keys are used to modify the data in the
display. Press either the up or down arrow to change the values
displayed. When used in one of the editing modes, pressing SHIFT will display the highest value allowed in a fu nction. Pressin g SHIFT will display the lowest value. In the zero and span calibrations, these keys will display the highest and lowest potentiometer settings.
Press the arrow key once to change one digit; press and hold either key to scroll (continuous value change), thereby reducing the time required to make large value changes.
748176-N
July 1998
Rosemount Analytical
3-3
Page 46
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
To move cursor one position at a time or, when used in conjunction with the SHIFT key moves the cursor 16 characters, one full display, at a time.
To access a function, to store a value in nonvolatile memory or to return to run
ENTER
mode from span, zero and security screens. The computer acknowledges
ENTER by momentarily flashing [** DATA STORED **] on the display when display when used to store a setting in non-volatile memory. Use ENTER to engage the span and zero functions, which are initiated by the SPAN and ZERO keys. [CALCULATING SPAN] or [CALCULATING ZERO] will then be momentarily displayed. Instruments with the Calibration Gas Control option or Auto Zero/Span option, ENTER also turns ON and OFF a solenoid valve for zero and span gas.
SHIFT ENTER
Mode. In the event of a power outage, items placed in volatile memory
will be lost. First press the SHIFT key, followed by the ENTER key. SHIFT+ENTER does not shut off the solenoid valve for instruments with the Calibration Gas Control or Auto Zero/Span.
The SHIFT+ENTER combination is the Escape feature.
The ENTER key in conjunction with the SHIFT key will return to Run
3.4.3 U
SER-PROGRAMMABLE KEYS
F1, F2, SHIFT/F1 and SHIFT/F2 are software-programmable keys which can be user-programmed to access any frequently used display or sub-menu for the following modes: Range, Diagnostics, Linearization, Auto Zero/Span, Remote Range I/O or Alarm, provided the option board selected is still present.
To use this feature, the function keys must be preprogrammed by the user through the following steps:
1. Access a display or sub-menu that will be frequently used.
2. Access a display or submenu by following the steps in the particular set of
instructions given in Figures 3-7 through 3-24 until the desired display is obtained.
3. Press F1, F2, SHIFT/F1 or SHIFT/F2 to program the analyzer to return to this
display from the RUN mode. This will assign F1, F2, SHIFT/F1 or SHIFT/F2 to this particular display, and will retain those assignments until the key or combination of keys is reprogrammed using the same procedure described in this section. The analyzer acknowledges this command by flashing [**KEY SAVED**] on the display.
4. Exit to the RUN mode display by completing the remaining steps in the figure
chosen in Step 2.
5. When the analyzer returns to the RUN mode display, press the key or keys
selected in Step 3 (F1, F2, SHIFT/F1 or SHIFT/F2) to check the setup. The analyzer will return to the display or sub-menu selected in Step 1.
3-4
Rosemount Analytical
July 1998
748176-N
Page 47
NITIAL STARTUP AND CALIBRATION
I
6. Press SHIFT+ENTER to return to the RUN mode. To reprogram the key or keys selected in Step 2, repeat Steps 1 through 5 for another
display or sub-menu. For example, if the GAIN is frequently changed, access the RANGE sub-menu to
access the GAIN display and press the F1 key. Press SHIFT+ENTER to return to the RUN mode. To get to the GAIN display from the RUN mode display, press the F1 key. To reprogram the F1 key, go to another display other than the RUN mode display and press the F1 key. This will reprogram the F1 key to the new display.
3.4.4 R
UN MODE DISPLAY
The RUN mode is the normal mode of operation. In this mode the display will show current process value component designation, control mode and range. Should an error condition or an alarm condition occur, [ER?] (where ? is an alphanumeric character) or [AL#] (where # is either the num ber 1 or 2) will flash on the display in the component name location. A list of error messages is located in Section 5.1. Refer to Figure 3-3 for the different run mode displays.
XXX % ABC L1
XXX ppm ABC L2
XXX %FS ER4 R3
Concentration Engineering Unit s , or % Fullscale. If Linear ization ON, reads in engineer i ng uni t s. If Linearization OFF, reads in % Fullscale.
Code for measured component or Error Message
Local, or Rem ote Control, Range 1, 2 or 3.
F
IGURE
3.4.5 G
3-3. R
UN MODE DISPLAY
ENERAL DISPLAY INFORMATION
The following features are present to the right of all display sequences:
→→→→
The beginning of a sub-menu is indicated by in the extreme right position of the display. This arrow indicates that there will be more information in subsequent displays which can be obtained by either pressing the key until the next display is obtained, or pressing SHIFT to move 16 characters, one full display, at a time.
748176-N
July 1998
Rosemount Analytical
3-5
Page 48
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
* Indicates that there are subsequent displays which can be accessed by pressing the key to view a new display or the key to return to a previous display. To move 16 characters, one full display at a time, press SHIFT or SHIFT ←.
←←←←
The last display of a routine is indicated by the To return to other d isplays in the routine, press the → key or SHIFT ← to move 16 characters, one full display at a time.
Note: At any point in the sequence, a sub-menu may be exited by pressing
SHIFT+ENTER.
3.5 ACCESSING MODE DISPLAYS
Ensure that all MODE displays are functional and that all options ordered from the factory are present by following the flow chart in Figure 3-4. To follow the logic flow chart, use the following steps:
Note: For more detailed instructions, refer to Figures 3-7 through 3-24.
1. Press MODE.
2. Use the key to move to the desired sub-menu (SECURITY, RANGE, DIAG-
NOSTICS, LINEARIZER, ALARM, CUR-RENT OUT PUT, AUTO-CAL or REMOTE I/O) and press ENTER.
3. Press SHIFT then to move through each sub-menu.
4. At the end of each routine, press SHIFT+ENTER to return to the RUN mode.
5. Repeat steps 1 through 4 to check the next function.
3.6 SECURITY CODE
The Model 880 is equipped with a security code feature, which is deactivated when the instrument is shipped from the factory. When the security feature is activated, only the STATUS and MODE function keys are active to access the STATUS and SECURITY displays. A valid password must be entered to activate the rest of the keyboard.
INITIAL PASSWORD IS “880”
This password may be changed to any three character group as shown in Figure 3-7. Entering the correct password activates the keyboard.
To gain access, follow the steps in the appropriate figure in this section. Once access has been gained, the procedure described in Sections 3.6 through 3.16 may be performed.
3-6
Rosemount Analytical
July 1998
748176-N
Page 49
NITIAL STARTUP AND CALIBRATION
I
In the event the password is misplaced, the operator may return to the initial password (880) through the following steps:
1. Press and release the RESET push-button switch on the Micro Board (see Figure 3-1).
2. Press and hold the ENTER key until the RUN mode display appears.
IR-880 VER. 2.0
ZERO
ZERO PS=30% ZR=0.0 PS-30% NOTE: Shift-enter from screen #2 leaves
valve on Time Constant = 1 sec.
SPAN
STATUS
SPAN PS=57%
98.5 57% 100.0 NOTE: Shift-enter from screen #2 leaves valve
on Time Constant = 1 sec.
SECURITY ENABLED/DISABLED EO-ZERO POT LMTS EI-SPAN #1 LMTS E2-SPAN #2 LMTS E3-SPAN #3 LMTS E4-ADC SATURATED E5-ZERO DRIFT E6-SPAN DRIFT ALARM 1 MN ON/OFF ALARM 2 AT RMT L/R BIN/DEC CURRENT 4/0 SP ON/OFF AUTOCAL ON/OFF
F
IGURE
748176-N
3-4. L
MODE
(continue to Figure 3-4B)
OGIC FLOW CHART (CONTINUED ON NEXT PAGE
July 1998
)
Rosemount Analytical
3-7
Page 50
M
ODEL
880 NON-D
MODE
ISPERSIVE INFRARED ANALYZER
SECURITY
PASSWORD: ?? SECURITY: ON/OFF NEW PSWRD: ??? Note: <reset> button plus <enter> disables security and sets password to "880"
DIAGNOSTICS
DIAG: OSC.T.=96 DET.SIG.=0.57 SCR.CUR.=825mA +5V=5.1 +15V=15.0 +12V=11.9
-15V+14.9
ALARM
ALARM: 1 MAN OFF/ON RANGE: 1 /2,3 LOW SET: 500 /HIGH DEADBAND: 10% ALARM: 2 AUTO /MAN RANGE: 1 /2,3 HIGH SET: 4500 /LOW DEADBAND: 5% Note: Each alarm can be set to different range; only one range per alarm. The value for the setpoint can be in ppm, % or %FS dependent on the selected range. The deadband value is in % of the selected range.
CUR. OUTPUT
CURRENT: 4 20mA /0 ZR/SPN SUPR ON /OFF 25 TO 75 %FS Note: SPN SUPR - ZR SUPR must be >= to 25 %FS
RANGE
RANGE: 1 CMP CO2 /2.3 GAIN = 1 /2, 4, 8 FS = 5000 ppm /% CAL GAS = 5000 ppm /% CL GAS = 100.0 % FS ZR-OFST 0.00 %FS (only) TIME CONST = 2 /ln SEC. Note: Zero offset works for both linearizer "ON" and linearizer "OFF".
LINEARIZER
LIN:ON/OFF RANGE 1,2,3 AO=0.0000000 A1=1.0000000 A2=0.0000000 A3=0.0000000 A4=0.0000000 Note: Range A, linearizer coefficients from A are equal to Range 3 and used for all three ranges. Maximum dynamic range 3:1. Coefficient accuracy=5 digits. Changing the linearizer ranges resets the drift capture, flag.
AUTO-CAL
AUTO-CAL: ON /OFF SH-V:Y SH-I:N RANGE: 1Y 2Y 3N DELAY: 1 HR PURGE: 2 MIN RPT ZERO: 24 HR RPT SPAN: 48 HR DRIFT LIMIT: OFF /ON ZR-DFT: ±±±±10 %FS SP-DFT: R1 ±±±±12 %FS Note: Switching DRIFT-LIMIT from OFF to ON enables the drift check and causes an alarm if the limit is reached. The starting point is the first Auto-Cal sequence after Drift-Limit is switched from OFF to ON. Remote request for ACAL is only valid when ACAL is set to OFF. Changing state of any of the ranges enables resets the Drift capture flag if drift is enabled.
F
IGURE
3-8
REMOTE I/O
LOCAL/REMOTE: R /I IN/OUT: DEC/BIN Note: IN/OUT (Bin/Bin) (Dec/Dec) (Bin/Dec) (Dec/Bin) (SP) special BIT­6 (I/O) for ACAL request for status. BIT-7 (0) for RHT/LOCAL status. In SP mode BIT-7 (0) is to switch the power voltage. When set to "special" BITS 0-5 are suppressed; BITS 6 & 7 are output as usual.
3-4. L
Rosemount Analytical
OGIC FLOW CHART (CONTINUED FROM PREVIOUS PAGE
July 1998
)
748176-N
Page 51
NITIAL STARTUP AND CALIBRATION
I
STEP
1MODE
2ENTER
3ENTER
4
KEYBOARD
ENTRY
DIGITAL DISPLAY EXPLANATION
Pressing the MODE key accesses the security feature. If the security feature is activated as indicated by at the right of the display,
SECURITY
PASSWORD: ???
PASSWORD VALID
(or)
INVALID PASSWORD
X.X %FS CH4 L1 XXXX ppm CH4 L1 X.X % CH4 L1
continue with Step 2. If the security feature is not activated as indicated by the at the right of the display, use the to access the other functions in Figures 3-7 through 3-24.
Pressing ENTER accesses the PASSWORD display. at the right of the display indicates that the security feature is activated. Use the or k ey to enter the first pass word character. Use the key to move the cursor to the next position and enter the next character. Repeat for the third character.
Pressing ENTER completes the authorization sequence, displays whether the password is valid or invalid and returns the analyzer to the run mode. If the password is valid, the user may now access the other analyzer functions in Figures 3-7 through 3-24.
This is the RUN mode. One of these three displays will be present on the digital display.
F
IGURE
STEP
F
IGURE
3-5. A
1MODE
2ENTER
3
4 ENTER **DATA STORED**
5 SHIFT+ENTER
3-6. A
CCESSING THE SECURITY LOCKOUT FEATURE
KEYBOARD
ENTRY
DIGITAL DISPLAY EXPLANATION
SECURITY
PASSWORD: ???
SECURITY: ON *
X.X %FS CH4 L1 XXXX ppm CH4 L1 X.X % CH4 L1
CTIVATING/DEACTIVATING THE SECURITY LOCKOUT FEATURE
Pressing the MODE key accesses the security feature.
Pressing ENTER accesses the password display. If the security feature is engaged, as indicated by at the right of the display, enter the password as in Figure 3-5, then repeat step
1. Use the or key to toggle SECURITY
ON/OFF. To activate the security feature, toggle SECURITY ON.
Pressing ENTER stores the data in non-volatile memory.
Pressing SHIFT+ENTER returns the user to the RUN mode.
748176-N
July 1998
Rosemount Analytical
3-9
Page 52
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
STEP
1MODE
2ENTER
3
4
5 ENTER PSWRD IS NOW ABC
KEYBOARD
ENTRY
NEW PSWRD: ???
DIGITAL DISPLAY EXPLANATION
SECURITY
PASSWORD: ???
SECURITY: OFF *
X.X %FS CH4 L1 XXXX ppm CH4 L1 X.X % CH4 L1
Pressing the MODE key accesses the security feature.
Pressing ENTER accesses the password display. If the security feature is OFF (PASSWORD: ??? ), use the to move to the next display.
If the security feature is ON, as indicated by at the right of the display, enter the password as in Figure 3-5, then proceed to step 3.
Use the to move to the next display. A new password may be entered in the NEW
PSWRD: ??? display. Use the or key to change the first character. Use the key to move the cursor to the next position and change the second and third characters. Valid characters are "space", "0 ... 9" and "A ... Z".
Pressing ENTER stores the new password in non-volatile memory and returns the analyzer to RUN mode.
This is the RUN mode. One of these three displays will be present on the digital display.
F
IGURE
3-7. C
HANGING THE PASSWORD
3.7 RANGE PARAMETERS
There are several range parameters that may be changed. The first display [RANGE:# CMP NNN ] allows RANGE1, RANGE2 or RANGE3 to be selected with the or key. Of these three independent ranges, RANGE3 should always be the least sensitive range.
The Model 880 Analyzer allows a different set of linearizing coefficients (Section 3.13) to be entered for each range. Or, if desired, one set of linearizing coefficients may be used for all three ranges when the dynamic range ratio is 3:1 or less. When using one set of coefficients, this set should always be entered in Range 3. Coefficients placed in Range 3 will automatically be used for Range A (All).
The component of interest is designated by a three digit group of letters or numbers. This gas name or designation may be selected for each range by placing the cursor under the desired digit [NNN] and selecting a letter or number with the or key. Valid characters are "space", "0"..."9" and "A"..."Z". This name will appear on the display when the analyzer is in the run mode.
3-10
Rosemount Analytical
July 1998
748176-N
Page 53
NITIAL STARTUP AND CALIBRATION
I
On the [GAIN=X] display, an amplifier gain of 1, 2, 4 or 8 can be selected for each range with the or key depending on the sensitivity desired. Refer to Sections 5.7 and 5.10. Range 3 is normally the least sensitive range. Other ranges are generally set with gains that are proportional to their relative fullscale spans. Thus, if range 1 is 0 to 10 % CO and range 3 is 0 to 100 % CO, then the respective gains will usually be 8 and 1.
On the [F S=XXXX pp m * ] or [FS= XX.X % *] dis play, u p to a fo ur digit f ulls cale va lue is entered in ppm (parts per million) or % (percent) for each linearized range.
On the [CAL GAS=XXXX ppm *] or [CAL GAS=XX.X % *] display, up to a f our digit calibration gas value is entered in ppm or % (percent) for each linearized range. Each
calibration gas concentration must be between 75 to 100 % of fullscale. On the [CL GAS=XXX.X%FS] display, the calibration gas in percent fullscale is
entered for each range. This value is obtained from the non-linear Response
Curve for Each Range located at the back of the manual. This curve will be different for each different application. Locate the calibration gas value on the bottom scale and find the corresponding non-linearized Recorder Deflection value on the side scale. This is the value that should be entered in [CL GAS=XXX.X%FS].
On the [ZR-OFFSET:X.XX *] display, the amount of zero offset in percent of fullscale is entered for each range. The zero offset feature compensates for impurities in zero calibration gas. If there are no impurities in the zero gas, set ZR-OFFSET to 0.00. Otherwise, the value should be obtained from the Response Curve for Each Range located at the back of the manual. Locate the amount of zero offset desired on the bottom scale and find the corresponding Recorder Deflection value on the side scale. This is the value that should be entered in [ZR-OFFSET:X.XX *].
On the [TIME CONST=XX ] display, the value of the TIME CONSTANT can be changed for each range. This TIME CONSTANT is responsible for the amount of time (in seconds) in which the analyzer responds to change. A different TIME CONSTANT can be selected for each range.
To change or check the settings of the different range parameters, press the keys in the following sequence:
Note After changing a setting, press ENTER to retain the new setting in
nonvolatile memory. Should a power outage occur, settings stored in nonvolatile memory will be saved.
At any point in the sequence, the routine may be exited by pressing SHIFT+ENTER.
The analyzer must be in local mode (L1, L2 or L3 will show in the run mode display) to change ranges in STEP 3 of Figure 3-8.
748176-N
July 1998
Rosemount Analytical
3-11
Page 54
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
STEP
1MODE 2
3ENTER
4
5
6
7
8
9
10 SHIFT+ENTER
KEYBOARD
ENTRY
TIME CONST=XX
DIGITAL DISPLAY EXPLANATION
SECURITY RANGE
RANGE=1 CMP CO
GAIN=X
FS=XXX.X % FS=XXXX ppm
CAL GAS=XXXX ppm
CL GAS=XXX.X%FS
ZR-OFFSET: X.XX
X.X %FS CH4 L1 XXXX ppm CH4 L1 X.X % CH4 L1
Pressing the MODE key accesses the security feature.
Use the to move to the next display. Pressing ENTER allows the range parameters
to be changed or viewed. Range 1, 2, or 3 can be selected with the or keys. Use the key to move the cursor to the c omponent of interest section of the display. Use the or keys to enter up to three characters. Store any changes made in non-volatile memory by pressing ENTER.
Use the or keys to select a GAIN of 1, 2, 4, or 8. Store the new GAIN setting in non-volatile memory for the selected range by pressing ENTER.
XXXX is fullscale value in ppm or %. This value may be changed by moving the cursor to a digit with the and using the or keys to obtain the desired value. In the linear mode, ppm can be toggled with % with the or keys, depending on the engineering units to be used in the RUN mode. Store the new fullscale value in non-volatile memory by pressing ENTER.
XXXX is the calibration gas value in ppm or %. This value may be changed by moving the cursor to a digit with the key and using the or keys to obtain the desired value. Store the new calibration gas value (in % fullscale) in non­volatile memory by pressing ENTER.
XXX.XX% is the amount of non-linearized recorder deflection f or the calibration gas value on the Response Curve at the back of the manual. Use the or keys to obtain the desired value. Store the new value (in % fullscale) in non-volatile memory by pressing ENTER.
Use the or keys to select the amount of zero offset.
Us the or keys to change the TIME CONSTANT to a value between 0.5 and 20 seconds. Store the new TIME CONSTANT setting in non-volatile memory by pressing ENTER.
Pressing SHIFT+ENTER returns the analyzer to RUN mode. One of these three dis plays will be present on the digital display.
F
IGURE
3-12
3-8. S
Rosemount Analytical
ETUP/CHECKOUT OF RANGE PARAMETERS
July 1998
748176-N
Page 55
NITIAL STARTUP AND CALIBRATION
I
3.8 ANALYZER DIAGNOSTICS
Diagnostics is selectable through the mode function. This function allows the oscillator tune, detector signal. source current and four power supply values to be viewed. It is recommended that the values for oscillator tune, detector signal and source current be recorded when the diagnostic display is first accessed.
STEP
1MODE 2
3
4ENTER
5
6
7 8 9
10 11 SHIFT+ENTER X.X %FS CO2 L1
KEYBOARD
ENTRY
→ →
→ → → → -15V=XX.X
DIGITAL DISPLAY EXPLANATION
SECURITY RANGE
DIAGNOSTICS
DIAG:OSC.T.=XXX
DET. SIG.=X.XX
SRC.CUR.=XXXXmA
+5V=X.X A power supply voltage. +15V=XX.X A power supply voltage. +12V=XX.X A power supply voltage
Pressing the MODE key accesses the security feature.
Use the to move to the next display. Press to move to the next display. Pressing ENTER allows the analyzer diagnostic
displays to be viewed. This display gives the current oscillator tune reading. The value may be changed by making a hardware adjustment (Section 5.2).
DET. SIG. is the detector s ignal value in volts. The value may be changed by making a hardware adjustment (Section 5.4).
SRC CUR is the value of the source current in milliamps. The value may be changed by making a hardware adjustment (Section 5.5).
A power supply voltage Pressing SHIFT+ENTER returns the analyzer to
RUN mode.
F
IGURE
3-9. A
NALYZER DIAGNOSTICS
3.9 ZERO CALIBRATION
3.9.1 H
748176-N
ARDWARE AND SOFTWARE ZERO
The Model 880 Analyzer has both a hardware and a software zero. The hardware zero is the adjustment of the zero potentiometer with the up and down arrows when the analyzer is in the zero setpoint mode. Using the up and down arrows adjusts the percentage of the zero potentiometer currently being used. The zero potentiometer should be adjusted so that the zero value [ZR=X.X] in the zero display [ZR=X.X PS=XX% ] is as close as possible to absolute zero while keeping the potentiometer status [PS=XX%] between 5% and 95%. Values outside this range will not be accepted.
July 1998
Rosemount Analytical
3-13
Page 56
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
Note There is only one hardware zero for all three ranges. However, there are
three software zeros, one for each range. When the hardware zero is engaged, the amplifier GAIN is automatically set to one and the TIME CONSTANT is set to one second. Engaging the software zero resets the GAIN and TIME CONSTANT to the values selected in Range Parameters (Figures 3.8, Step 9).
Pressing the up or down arrow sets the present value of the software zero to absolute zero. Pressing ENTER after the hardware zero has been set, engages the software zero. The analyzer can use up to ±500 mV to compensate for the hardware zero. In the run mode, the analyzer will now read zero. This new zero value set by software will be used as zero for the range in which it was calibrated until the analyzer is recalibrated.
While there is only one hardware zero for all three ranges, there are three software zeros. When calibrating more than one range, after the hardware and software zero has been set for the first range, the operator should select a second range that will be used and re-enter the zero mode. Pressing ENTER twice while flowing zero gas without making a potentiometer adjustment will complete a software zero for the second range. This step should be repeated for the third range, should this range be used.
Note When entering this function, make sure that there is zero calibration gas
flowing through the analyzer. When entering this function for viewing purposes only, press SHIFT+ENTER to exit. Press ENTER to exit the function only if a calibration has been made.
Note If the zero potentiometer has been changed, pressing SHIFT+ENTER
instead of ENTER will leave the software zero set to absolute zero. This will cause the analyzer to give faulty readings unless the hardware zero gives an absolute zero reading.
3.9.2 C
ALIBRATING THE ANALYZER WITH ZERO GAS
1. Allow system to warm up a minimum of one and one half hours.
2. Connect zero gas to the sample cell inlet at the back of the analyzer. Flow the gas at a flow rate of 500 cc/min, as read on a flowmeter, through the analyzer for at least two minutes.
3. Press the keys in the following sequence to calibrate the zero setting for the analyzer for each range desired.
3-14
Rosemount Analytical
July 1998
748176-N
Page 57
NITIAL STARTUP AND CALIBRATION
I
STEP
1 ZERO ZERO PS=XX%
2 SHIFT+ENTER
3
F
IGURE
STEP
1NONE
2 ZERO ZERO PS=XX%
3 ENTER ZR=X.X PS=XX%
4ENTER
KEYBOARD
ENTRY
3-10. V
KEYBOARD
ENTRY
DIGITAL DISPLAY EXPLANATION
X.X % CO2 L2 XXXX ppm CO2 L2 X.X %FS CO2 L2
X.X % CO2 L2 XXXX ppm CO2 L2 X.X %FS CO2 L2
IEWING THE ZERO DISPLAY
DIGITAL DISPLAY EXPLANATION
X.X % CO2 L2 XXXX ppm CO2 L2 X.X %FS CO2 L2
CALCULATING ZERO X.X % CO2 L2
XXXX ppm CO2 L2 X.X %FS CO2 L2
then
ZERO indicates that the instrument is in the zero setting mode. PS=XX% is the zero potentiometer status. XX is the percent of the zero potentiometer being used. indicates beginning of routine.
SHIFT+ENTER returns the analyzer to the RUN mode without engaging the software or hardware zero. To avoid un -calibrating th e analyzer, do not
use the ENTER key to exit this function u nless a change has been made.
To make a c hange in the zero setting, go to Figure 3­11 for the standard analyzer. Figure 3-12 for the analyzer with Calibration Gas Control or Figure 3-23 for the analyzer with Auto Zero/Span option.
This is the RUN mode display. Verify that zero calibration gas is connected to the rear of the analyzer and turn the gas ON.
Zero indicates that the analyzer is in the zero setting mode. PS=XX% is the zero potentiometer status. This is the percent of the zero potentiometer being used.
Pressing ENTER sets the gain amplification to one, bypasses the linearizer if this feature is ON and allows the zero setting to be changed. ZR=X.X is the present numeric value of the zero signal in %FS for the non-linear mode. The next two digits ar e the zero potentiometer status. Use the or key to change the zero potentiometer (PS=XX%) and set the zero signal (ZR=X.X) as close to absolute zero as possible. The first time the or k ey is pressed, the software zero is set to absolute zero. Subsequent moves change the setting of the zero potentiometer . Valid settings for the zero potentiometer are between 5% and 95%.
Pressing ENTER when the desired value has been obtained to exit the functions changes the GAIN to the value set in the range parameters (Figure 3-8, Step 4), engages the linearizer if this f eature is ON, engages the software zero, stores the new zero value in non-volatile memory and returns the analyzer to the RUN mode.
F
IGURE
748176-N
3-11. C
HANGE THE ZERO SETTING
July 1998
Rosemount Analytical
3-15
Page 58
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
STEP
KEYBOARD
ENTRY
1NONE
2 ZERO ZERO PS=XX%
3 ENTER ZR=X.X PS=XX%
4ENTER
DIGITAL DISPLAY EXPLANATION
X.X % H2O L1 XXXX ppm H2O L1 X.X %FS H2O L1
CALCULATING ZERO
then X.X % H2O L1 XXXX ppm H2O L1 X.X %FS H2O L1
This is the RUN mode display. Verify that zero calibration gas is connected to the rear of the analyzer and turn the gas ON.
Zero indicates that the analyzer is in the zero setting mode. PS=XX% is the zero potentiometer status. This is the percent of the zero potentiometer being used.
Pressing ENTER turns ON the solenoid valve (customer supplied) for zero calibration gas, sets the GAIN amplification to one, bypasses the linearizer if this f eature is ON, and allows the zero setting to be changed. Allow the zero gas to flow for two minutes before proceeding with the rest of the calibration. ZR=X.X is the pres ent num eric value of the zero signal in %FS. The next two digits are the zero potentiometer status. Use the or key to change the zero potentiometer (PS=XX%) and set the zero signal (ZR=X .X) as close to absolute zero as possible. The f irst time the or key is pressed, the software zero is set to absolute zero. Subsequent moves change the setting of the zero potentiometer. Valid settings for the zero potentiom eter are between 5% and 95%.
Pressing ENTER when the desired value has been obtained to exit the functions changes the GAIN and the TIME CONSTANT to the values set in the range parameters (Figure 3-8, Steps 4 and 9), engages the linearizer if this feature is ON, engages the software zero, stores the new zero value in non-volatile memory and turns OFF the solenoid valve for zero gas and returns the analyzer to the RUN mode. Pressing SHIFT+ENTER to exit this
function will NOT turn the zero gas OFF or set the software zero.
F
IGURE
3-12. C
G
STEP
1 SPAN
2 SHIFT+ENTER
3
F
IGURE
3-16
KEYBOARD
ENTRY
3-13. V
Rosemount Analytical
HANGING THE ZERO SETTING FOR ANALYZERS WITH CALIBRATION
AS CONTROL
DIGITAL DISPLAY EXPLANATION
SPAN indicates that the analyzer is in the span mode for the range selected in the range parameters. PS=XX% is the percent of span potentiometer being used.
X.X %FS SO2 L2 XXXX ppm SO2 L2
X.X % SO2 L2 X.X %FS SO2 L2
XXXX ppm SO2 L2 X.X % SO2
L2
SHIFT+ENTER returns the analyzer to the RUN mode without engaging the software span. Do not use the
ENTER key to exit this function unless a calibration is desired.
To make a change in the span setting, go to Figure 3­14 for the standard analyzer, Figure 3-15 for the analyzer with the Cal Gas Control option or Figure 3­23 for the analyzer with Auto Zero/Span option.
IEWING THE SPAN SETTING
July 1998
748176-N
Page 59
I
NITIAL STARTUP AND CALIBRATION
3.10 ZERO CALIBRATION FOR THE ANALYZER WITH THE CAL GAS CONTROL OPTION
The Calibration Gas Control Option allows one-man calibration. This option consists of two form C contacts rated 3A-125/250 VAC or 5A-30 VDC, (resistive). These contacts are connected to a solenoid valve (customer supplied) which will turn zero and span calibration gases on and off when activated. Pressing ENTER at the [ZERO PS=XX%] display activates the solenoid valve, turning on the zero calibration gas. Pressing ENTER to exit the function at the [ZR=X.X PS=XX%] display deactivates the solenoid valve. turning off the zero calibration gas.
Note For instruments with Calibration Gas Control Option or Auto Zero/Span Option,
pressing ENTER at the [ZERO PS=XX%] display turns on the solenoid valve (customer supplied) for zero calibration gas. To turn this valve off, press ENTER to exit this function. Pressing SHIFT+ENTER to exit this function will NOT turn off the relay for this valve.
Note When entering this function for viewing purposes only, press SHIFT+ENTER to
exit. Press ENTER to exit the function only if a calibration has been made.
3.10.1 C
ALIBRATING THE ANALYZER WITH CALIBRATION GAS CONTROL OPTION
WITH
ERO GAS
Z
1. Allow system to warm up a minimum of one and one half hours.
2. Connect the solenoid valve for the zero gas to the two form C contacts. Connect the zero gas to the sample cell inlet located on the back of the analyzer. The gas should flow at a flow rate of 500 cc/min, as read on a flowmeter.
3. Press the keys in the following sequence to calibrate the zero setting for the analyzer for each range desired:
3.11 SPAN CALIBRATION
3.11.1 H
The Model 880 Analyzer has both a hardware and a software span for all three ranges. The hardware span is the adjustment of the span potentiometer with the up and down arrows when the analyzer is in the span setting mode. The display for the span setting mode is [X.XX NN% MMM ] where X.XX is the run mode value, NN% is the percentage of span potentiometer in use and MMM is the span gas value.
ARDWARE AND SOFTWARE SETTINGS FOR SPAN GAS
748176-N
July 1998
Rosemount Analytical
3-17
Page 60
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
There are three span potentiometers, one for each range, that can be calibrated. These span potentiometers should be adjusted with the up or down arrows at the [X.XX NN% MMM ] display so that the run mode value [X.XX] is between 51 % and 99 % of the span gas value [MMM] while keeping the span potentiometer [NN %] between 5 % and 95 %. Span potentiometer values outside this range will not be accepted. Also, if the run mode value is not between 51 % and 99 % of the span gas value, the software span cannot be made and an error message of ER1, ER2 or ER3 will flash on the run mode screen. If the setting cannot be made while keeping within these ranges, the digital GAIN must be changed. Refer to Section 5.7 and Figure 3-8, STEP 4.
Pressing the up or down arrow for the first time resets the present software span factor to one. Pressing ENTER after the hardware span has been set engages the software to multiply the run mode value by the span factor. This new span factor will be stored in nonvolatile memory until the analyzer is recalibrated. Pressing SHIFT+ENTER does not engage the software span.
Note If the span potentiometer has been changed, pressing SHIFT+ENTER instead of
ENTER will cause the span factor to be one. This will cause the analyzer to give faulty readings unless the span gas value is the same as the run mode value.
For instruments with the Calibration Gas Control Option, pressing ENTER at the [SPAN PS=XX%] display turns on the solenoid valve (customer supplied) for span calibration gas. To turn this valve off, press the ENTER key when exiting this function. Pressing SHIFT+ENTER to exit this function will NOT turn off the relay for this valve.
When entering this function, make sure that there is span calibration gas connected to the analyzer. When entering this function for viewing purposes only, press SHIFT+ENTER to exit. Press ENTER to exit the function only if a calibration is desired.
When the hardware span is engaged, the TIME CONSTANT is set to one second. Engaging the software span, resets the TIME CONSTANT to the value selected in Range Parameters (Figure 3.8, STEP 9).
3.11.2 C
ALIBRATING THE ANALYZER WITH SPAN GAS
1. Allow system to warm up a minimum of one and one half hours.
2. Connect span gas to the sample cell inlet at the back of the analyzer. Flow the gas at a flow rate of 500 cc/min, as read on a flowmeter, through the analyzer for at least two minutes.
3. Press the keys in the following sequence for those ranges being calibrated.
3-18
Rosemount Analytical
July 1998
748176-N
Page 61
I
NITIAL STARTUP AND CALIBRATION
STEP
1NONE
2 SPAN SPAN PS=XX%
3 ENTER X.XX NN% MMM.M
4ENTER
KEYBOARD
ENTRY
DIGITAL DISPLAY EXPLANATION
X.X %FS H2O L3 XXXX ppm H2O L3 X.X %FS H2O L3
CALCULATING SPAN
then X.X %FS H2O L3 XXXX ppm H2O L3 X.X % H2O L3
This is the RUN mode display. Verify that zero calibration gas is connected to the rear of the analyzer and turn the gas ON.
Span indicates that the analyzer is in the span mode for the current range. PS=XX% is the percent of the span potentiometer being used.
The first set of digits (X.XX ) is the present RUN mode value. The next digits (NN%) are the percent of the span potentiometer being used. The thir d set of digits (MMM) is the calibration span gas value. Use the or key to change the percent of the span potentiometer used so that the RUN mode value (X.XX) is between 51% and 99% of the span gas value (MMM) while keeping the span potentiometer (NN%) between 5% and 95%. The first time the or key is pressed, it engages the hardware span. Subsequent moves changes the setting of the span potentiometer. Span potentiometer settings outside the range of 5% to 95% are not allowed.
Pressing ENTER when the desired value has been obtained to exit the function engages the software span, stores the new factor in non- volatile m em or y and returns the analyzer to RUN mode.
Repeat Steps 1 through 4 for Ranges 2 and 3, if used.
F
IGURE
3-14. C
HANGING THE SPAN SETTING
3.12 SPAN CALIBRATION FOR THE ANALYZER WITH THE CAL GAS CONTROL OPTION
The Calibration Gas Control Option allows one-man calibration. This option consists of two form C contacts rated 3A-125/250 VAC or SA-30 VDC, (resistive). These contacts are connected to a solenoid valve (customer supplied) which will turn zero and span calibration gases on and off when activated. Pressing ENTER at the [SPAN PS=XX%] display activates the solenoid valve, turning on the span calibration gas. Pressing ENTER to exit the function at the [X.XX NN % MMM] display deactivates the solenoid valve. turning off the span calibration gas.
Note For instruments with the Calibration Gas Control Option, pressing ENTER at the
[SPAN1 PS=XX%] display turns on the solenoid valve (customer supplied) for span calibration gas. To turn this valve off, press the ENTER key when exiting this function. Pressing SHIFT+ENTER to exit this function will NOT turn off the relay for this valve.
748176-N
July 1998
Rosemount Analytical
3-19
Page 62
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
STEP
1NONE
2 SPAN SPAN PS=XX%
3 ENTER X.XX NN% MMM
4ENTER
KEYBOARD
ENTRY
DIGITAL DISPLAY EXPLANATION
X.X %FS HEX L1 XXXX ppm HEX L1 X.X % HEX L1
CALCULATING SPAN
then X.X %FS HEX L1 XXXX ppm HEX L1 X.X % HEX L1
This is the RUN mode display. Verify that zero calibration gas is connected to the rear of the analyzer.
Span indicates that the analyzer is in the span mode for the current range. To change ranges, refer to Figure 3-8. PS=XX% is the percent of the span potentiometer being used for current range.
Pressing ENTER turns ON the solenoid valve (customer supplied) f or s pan calibr ation gas and allows the span setting to be changed. Allow the span gas to flow for two minutes before proceeding with the calibration. The first set of digits (X.XX) is the present RUN mode value. The second set of digits (NN%) is the percent of span potentiometer being used. The third set of digits (MMM) is the calibration span gas value. Use the or key to change the percent of the span potentiometer used so that the RUN mode value (X.XX) is between 51% and 99% of the span gas value (MMM) while keeping the span potentiometer (NN%) between 5% and 95%. The first time or key is pressed engages the hardware span and sets the software span to one. Subsequent moves change the setting of the span potentiometer. Span potentiom eter settings outside the range of 5% and 95% are not allowed.
Pressing ENTER when the desired value has been obtained to exit the function engages the software span, stores the new factor in non-volatile memory, turns the solenoid valve for span gas OFF, and returns the analyzer to RUN mode. Pressing SHIFT+ENTER
to exit this function will NOT turn the span gas OFF or set the software span.
F
IGURE
3-15. C
G
AS CONTROL
HANGING THE SPAN SETTING FOR ANALYZERS WITH CALIBRATION
3.13 LINEARIZATION
The Model 880 Analyzer can be operated in the linear and non-linear mode. Linearization can be toggled ON/OFF with the 1' or J, key. In the OFF position, linearization is disabled for all ranges. In the linear mode the component of interest is measured in engineering units, either ppm (parts per million) or 5 (percent of composition), in the non-linear mode measurement is in %FS (percent of fullscale).
The analyzer is linearized with a fourth-order polynomial.
4
3-20
Y = A0 + A1X + A2X2 + A3X3 + A4X
Rosemount Analytical
July 1998
748176-N
Page 63
I
NITIAL STARTUP AND CALIBRATION
where X is the non-linear input, A0, A1, A2, A3 and A4 are the linearization coefficients and Y is the linear output.
A typical response curve is located on the application sheet at the back of the manual. This curve is plotted with X as the concentration and Y as the output deflection.
Each range can be linearized using a separate curve, or the same curve can be used for all three ranges (Range A=All) when the dynamic range ratio is 3:1 or less. If Range A is selected, the microp rocessor will use the coefficients in Ran ge 3. Also, if the linearizer is ON and in Range A, the microprocessor will use the GAIN and TIME CONSTANT from Range 3, regardless of the GAIN and TIME CONSTANT selected for Ranges 1 and 2.
When ordered, the linearization coefficients are entered in the appropriate range(s) at the factory. If one set of linearization coefficients have been ordered and a range has not been specified, these coefficients will be for Range 3.
To operate the analyzer with a combination of linearized and non-linearized ranges, turn the linearizer ON and enter the appropriate coefficients for the range(s) to be linearized. For the non-linearized range(s). enter the following coefficients:
A0 = 0 A1 = 1 A2 = 0 A3 = 0 A4 = 0
Using these coefficients will make the output equal to the input (Y=X). For reasons of accuracy, this procedure is only recommended for narrow ranges that span a linear section of the standard curve such as 0 to 100 ppm CO. In the range parameters select ppm (Figure 3-8, STEP 5) and enter the span gas in ppm (Figure 3-8, STEP 6). In the run mode, the display will read ppm.
To calculate linearization coefficients other than those installed at the factory, either 11 or 21 data points must be taken. These data points are obtained by using a precision gas divider with an accuracy of ±0.5 % such as the Model SGD Series-710 Standard Gas Divider™ and following the STEPS in Figure 3-18. These data points can be entered into any program capable of computing a fourth order polynomial curve fit. The curve obtained with these data points (Figure 3-17) will have X as the output (recorder) voltage and Y as the signal concentration. This curve will be the mirror image of the curve on the application sheet at the back of the manual (Figure 3-16), however the linearization coefficients will be different. Use the coefficients calculated with the curve in Figure 3-17 for linearization coefficients.
The coefficients in Figure 3-16 may be used to calculate any Recorder Deflection for a given Concentration. Use these coefficients to solve the equation and calculate the calibration gas value in Range Parameters [CL GAS XXX %F] Fi gure 3-8. STEP 7.
748176-N
July 1998
Rosemount Analytical
3-21
Page 64
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
Y = A0 + A1X + A2X2 + A3X3 + A4X
4
For coefficients determined for user-specific gas, after taking the data points either use any program capable of calculating a fourth order curve fit or call the factory to have the specific coefficients calculated using the data determined in Figure 3-18.
When entering user-determined coefficients, note that the microprocessor only recognizes five significant figures to the right of the decimal point, for example,
0.12345 or 0.00012345, disregard any additional digits.
C12 5% CO (Curve 1)
1.00
0.90
0.80
0.70
F
IGURE
RECORDER NORMALIZED
3-16. T
0.60
0.50
0.40
0.30
0.20
0.10
0.00
0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00
CONCENTRATION NORMALIZED
YPICAL APPLICATION LINEARIZATION CURVE
A0 = 0.007196
A1 = 2.453977
A2 = 2.998715
A3 = 2.270009
A4 = 0.733518
F
IGURE
3-22
1.00
0.90
0.80
0.70
0.60
0.50
0.40
0.30
0.20
CONCENTRATION NORMALIZED
0.10
0.00
3-17. U
SER-DETERMINED LINEARIZATION CURVE
Rosemount Analytical
C12 5% CO (Curve 2)
A0 = 0.000155
A1 = 0.313933
A2 = 0.712239
A3 = 0.890872
A4 = 0.864127
0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00
RECORDER NORMALIZED
July 1998
748176-N
Page 65
I
NITIAL STARTUP AND CALIBRATION
STEP
10
11
12
13 SHIFT+ENTER XX.X %FS CH4 L1
14 ZERO ZERO PS=XX%
15 SPAN XXXX %FS CH4 L3
16
17
18
19
KEYBOARD
ENTRY
1MODE 2
3
4
5ENTER
6 SHIFT+ENTER
7 MODE SECURITY 8
9ENTER
DIGITAL DISPLAY EXPLANATION
SECURITY RANGE *
DIAGNOSTICS * LINEARIZER *
LIN:OFF RANGE:N
XX.X %FS CH4 L1
RANGE *
RANGE:N CMP CO
GAIN=X
FS=XX.X % * FS=XXXX ppm *
CAL GAS=XXXX ppm *
Pressing MODE allows the mode settings to be viewed or changed
moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display
Pressing ENTER places the analyzer in the linearization mode. Use the or key to toggle the linearizer OFF. Use the key to move to the range number. Select Range 1, 2, or 3 with or keys.
Pressing SHIFT+ENTER returns the analyzer to the RUN mode.
Pressing MODE allows the mode settings to be viewed or changed
moves the cursor to the next display Pressing ENTER allows the range parameters to be
changed. Range 1, 2 or 3 can be selected with the or
key. moves the cursor to the next display.
XXXX is fullscale value in ppm or %. This value may be changed by moving the cursor to a digit with the and using the or key to obtain the desired value. In the linear mode, ppm can be toggled to % with the or key, depending on the engineering units to be used in the RUN mode. Enter the span gas value as fullscale value.
XXXX is the calibration gas value in ppm or %. This value may be changed by moving the cursor to a digit with the key and using the or key to obtain the desired value. Enter the span gas value.
Pressing SHIFT+ENTER returns the analyzer to RUN mode.
Using zero and span gas, zero and span the analyzer as described in Figures 3-11 and 3-14 for the standard equipment or 3-12 and 3-15 for the analyzer with Calibration Gas Control option or Auto Zero/Span option.
Using a gas divider, deliver 10% span gas to the analyzer and note the RUN mode value.
Repeat Step 15 for 20% through 100% span gas in increments of 10%. This will produce 11 data points (0% through 100%).
For greater accuracy of lower range readings, take 10 more data points between 0% and 10% for a total of 21 data points
Use a computer program to calculate the new coefficients, or call factory for assistance.
When the new coefficients have been calculated, enter their values into memory (Figure 3-19).
F
IGURE
748176-N
3-18. D
ATA POINTS FOR USER-DETERMINED LINEARIZATION COEFFICIENTS
July 1998
Rosemount Analytical
3-23
Page 66
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
STEP
10
11 ENTER
12 SHIFT+ENTER
KEYBOARD
ENTRY
1MODE 2
3
4
5ENTER
6ENTER
7
8ENTER
9ENTER
SHIFT LIN:OFF RANGE N
DIGITAL DISPLAY EXPLANATION
SECURITY RANGE *
DIAGNOSTICS * LINEARIZER *
LINE:OFF RANGE:N
AO=(-)X.XXXXXXXX
A1=(-)X.XXXXXXXX
A2=(-)X.XXXXXXXX A2=(-)X.XXXXXXXX A2=(-)X.XXXXXXXX
LIN:ON RANGE N XXXX ppm CH4 L3
XX.X % CH4 L3
Pressing MODE allows the mode settings to be viewed or changed
moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display
Pressing ENTER places the analyzer in the linearization mode. Use the or key to toggle the linearizer OFF. Use the key to move to the range number. Select Range 1, 2, or 3 with or keys.
Pressing ENTER stores the range num ber (1, 2 or 3) in non-volatile memory.
Accessing this display allows the linearizing coeff icients to be changed. Use the or key to select A0. Use the key to move to the next position and toggle the ­(minus) sign ON/OF F with the or keys. Using the key to move to the correct position, enter a nine digit coefficient with the or key.
Pressing ENTER stores the f irst linearization coefficient in non-volatile memory and causes the display to prom pt the user to enter the next coefficient. Enter the linearizing coefficient for A1 as in Step 7.
Repeat Steps 7 and 8 and enter the linearizing coefficients for A2, A3 and A4.
If linearizing coefficients are to be entered for another range, repeat Steps 5 through 9 for the next range. If not, turn the linearizer back on with the or keys.
Pressing ENTER stores the linear izer ON setting in non­volatile memory.
Pressing SHIFT+ENTER returns the analyzer to RUN mode.
F
IGURE
3-19. S
ETUP/CHECKOUT OF LINEARIZATION FUNCTION
3.14 ALARM OPTION
The Alarm Option consists of two single point, field-programmable high or low outputs with a deadband up to 20% of fullscale. The two alarm setpoints are programmable for one range selected in Figure 3-20, STEP 6, and are dimensionless. The alarms can be set with one alarm HIGH and one alarm LOW, both alarms HIGH or both alarms LOW. This option is completely user configurable.
The Status Display will reflect an alarm condition, should one occur. When the instrument is in alarm condition (exceeding the alarm setpoint), the latch associated with the alarm is set. W hen the alarm condition clears, (run mode value is less than the alarm setpoint plus the deadband) the latch is reset.
3-24
Rosemount Analytical
July 1998
748176-N
Page 67
I
g
g
y
(
y
g
y
g
g
y
g
g
g
g
NITIAL STARTUP AND CALIBRATION
The high alarm is determined when run mode value exceeds the alarm setpoint. The alarm is cleared when run mode value is less than alarm setpoint minus the deadband.
The low alarm is determined when the run mode value is less than the alarm setpoint. This alarm is cleared when the run mode value is greater than the alarm setpoint plus the deadband.
ALARM1 and ALARM2 can be toggled with the up and down arrows to either AT. (automatic) or MAN (manual). In the AUTO (automatic) setting, an alarm relay will be activated should an alarm condition occur. Alarms are calculated in the AUTO mode on the basis of parameter settings. The MANUAL mode is the test mode and alarms are not scanned by CPU. In the MANUAL (test) mode, the ALARM ON/OFF can be toggled with the up and down arrows to set and reset the alarm latch.
STEP
1MODE 2
3 4 5
6ENTER
7ENTER
8
9ENTER
10
11 ENTER 12 13 ENTER 14 SHIFT+ENTER X.X %FS CO L1
KEYBOARD
ENTRY
→ → → →
DEADBAND:XX %
DIGITAL DISPLAY EXPLANATION
SECURITY RANGE *
DIAGNOSTICS * LINEARIZER * ALARM *
ALARM:1 MAN OFF
RANGE: 1 *
HIGH SET:XXXX *
Pressin or changed
moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display
Pressin MODE. Use the and ALARM2. Use the ke next characters, then use the between MAN ke key to toggle between OFF/ON. Pressin memory. Use the alarm. The alarm is onl Selectin ranges. Pressin memory. Use the LOW SET . Move the cursor with the ke di keys. Pressin memory. Use the up to 20% of fullscale. Pressin memory. Pressin mode.
MODE allows the mode settings to be viewed
ENTER accesses the ALARM SETPOINT
or key to toggle between ALARM1
to move the cursor to the
or key to toggle
manual) or AUTO (automatic). Use the
to move to the next character, then use the ↑ or
ENTER stores the setting in non-volatile
or key to select Range 1, 2 or 3 for the
valid for the selected range.
a range disables the alarm for the other two
ENTER stores the setting in non-volatile
or key to toggle between HIGH SET and
to the four
its and change to the desired value with the ↑ or
ENTER stores the setting in non-volatile
or key to change the value of the deadband ENTER stores the setting in non-volatile SHIFT+ENTER returns the analyzer to RUN
F
IGURE
748176-N
3-20. S
ETUP/CHECKOUT OF ALARM OPTION
July 1998
Rosemount Analytical
3-25
Page 68
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
3.14.1 STATUS D
ISPLAY
The STATUS display shows the alarms, error messages and security lockout status. An explanation of error messages is given in Section 5.1. The STATUS display can be used to check the following alarm setpoints without entering one of the MODE functions: HIGH/LOW, AUTO/MANUAL and ON/OFF. Refer to Figure 3-21.
The order of priority for error messages, security status and alarms is as follows:
[SECURITY ENABLED/DISABLED] [E0-ZERO POT LMTS] [E1-SPAN #1 LMTS] [E2-SPAN #2 LMTS] [E3-SPAN #3 LMTS] [E4-ADC SATURATED] [E5-ZERO DRIFT] [E6-SPAN DRIFT] [RMT: R/L] [ALARM 1 AUTO/{MAN ON/OFF}] [ALARM 2 AUTO/{MAN ON/OFF}] [CAL-CTL PRESENT] [AUTOCAL: ON/OFF] [CURRENT 0/4 SP ON/OFF]
Alarm MN (Manual) or AT (Automatic)
ALARM 1 MN OFF
Alarm ON or OFF
F
IGURE
Alarm 1 (15 seconds)
followed by
Alarm 2 (15 seconds)
3-21. S
TATUS DISPLAY
3.15 CURRENT OUTPUT OPTION
The Model 880 Analyzer has an optional 0 to 20 or 4 to 20 mA current output with zero span suppression. With this option, the current output can represent any suppressed range with at least a 25% span. For example, a valid range could be 0 to 25%, 28 to 61% or 33 to 100%. When the Zero Span Suppression is off (NO), the analyzer defaults to the 0 to 100% range. Refer to Figure 3-22.
3-26
Rosemount Analytical
July 1998
748176-N
Page 69
I
g
g
g
g
y
y
g
y
g
y
y
y
g
g
NITIAL STARTUP AND CALIBRATION
STEP
10 ENTER
11
12 ENTER 13 SHIFT+ENTER X.X %FS CO2 L1
KEYBOARD
ENTRY
1MODE 2
3
4
5
6
7ENTER
8ENTER
9
XX TO XXX %FS
DIGITAL DISPLAY EXPLANATION
SECURITY RANGE *
DIAGNOSTICS * LINEARIZER * ALARM * CURRENT OUTPUT *
CURRENT:4 20mA
ZR/SPN SUPR:NO *
Pressin or changed
moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display
Pressin chan mA.
Pressin memory.
moves the cursor to the nex t displa ke off).
Pressin memory.
moves the cursor to the next displa zero suppression ran with the ke ke be in percent fullscale, even in the linear mode. An range may be selected as long as the span is at least 25% of fullscale.
Pressin memory.
Pressin mode.
MODE allows the mode settings to be viewed
ENTER allows the Current Output option to be
ed. Use the ↑ or key to toggle between 0 and 4
ENTER stores the setting in non-volatile
. Use the or
s to toggle between YES/NO (zero suppression on or
ENTER stores the setting in non-volatile
. Change the
e by moving the cursor to a digit
and changing the setting with the ↑ or
. The value of the upper and lower lim its will always
ENTER stores the setting in non-volatile SHIFT+ENTER returns the analyzer to RUN
F
IGURE
3-22. S
ETUP/CHECKOUT OF CURRENT OUTPUT OPTION
3.16 AUTO ZERO/SPAN OPTION
The Auto Zero/Span Option allows automatic unattended calibration at set intervals. The option has six contact closures, four of which are field programmable for frequency and duration of the calibration cycle (span 1. span 2, span 3 and zero), while the other two contact closures indicate insufficient zero and span adjustments, and also drift limits for zero and span, if activated.
The auto zero/span [ACAL: ON] display allows the user to select ON or OFF to turn the Auto Zero/Span Option "on" or "off". Toggling from OFF to ON resets the timers for the Auto Zero/Span Option. To reset the timers when the Auto Zero/Span Option is "on", toggle from ON to OFF to ON.
The sample and hold [SH: YES] display allows the user to select YES or NO to turn the automatic sample and hold "on" or "off". When the sample and hold feature is "on", the recorder and Current Output Option do not get updated until the calibration sequence is completed.
748176-N
July 1998
Rosemount Analytical
3-27
Page 70
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
The range selection [RANGE: 1Y 2Y 3Y] display allows the user to select the ranges which will be automatically calibrate d with span gas by using the arrow to move the cursor to the desired range and using the or , key to select Y (yes) or N (no) for each range. The zero for all three ran ges will be calibrated at each interval regard less of range(s) selected.
The initial delay [DELAY nnn HR] display allows the user to select the amount of time until the first automatic calibration occurs. This is the initial delay until the automatic cycle starts. At this time a zero and span calibration is made, regardless of selection. If a zero delay is selected, there will be an autom a t ic two min ute delay.
The purge [PURGE: nnn MIN] display allows the user to select the amount of time for the calibration gas to flow through the analyzer before the calibration starts for zero and span, or the amount of time for sample gas to flow through the analyzer before run mode values are recorded when the sample and hold feature is selected. The analyzer is calibrated during the final minute of purge time. During the remaining purge time, the signal is modified according to previous calibration data.
The repeat zero [RPT ZERO nnn HR] display allows the user to select the amount of time between zero calibrations. This is the amount of time after the initial calibration before the zero calibration is repeated without repeating the span calibration.
Note Each time a span calibration is made, a zero calibration is also made regardless
of selection. The keyboard is disabled during the auto zero/span sequence. During the auto zero/span sequence, the time constant is set to one second. Upon completion of the calibration sequence, the time constant is reset to the value chosen in Range Parameters, Figure 3-8, STEP 9. In order to engage the Auto-Cal function with the Remote Range I/O Option, the Auto-Cal function must be disabled by toggling AUTO-CAL to OFF in the [AUTO-CAL:OFF] display, STEP 8, Figure 3-23.
The repeat span [RPT SPAN nnn HR] display allows the user to select the amount of time between span calibrations. This is the amount of time after the initial span calibration before this calibration is repeated.
The [DRIFT LIMIT: ON] display allows the user to determine the maximum amount of span and zero drift allowable. The [ZR-DFT: +/- XX%] or [SP-DFT: =/- XX%] displays allow the user to select the percentage of fullscale by which the analyzer is allowed to drift from the reference span or zero calibration values. The maximum zero drift limit is 10% fullscale and the maximum span drift is 15% fullscale.
In the linearized mode, these values should be obtained from the Response Curve for Range located at the back of the manual. For the linear mode, locate the amount of span or zero drift limit desired on the bottom scale and find the corresponding Recorder Deflection value on the side scale. These are the values that should be entered in [ZR-DFT: +/-XX % *] or [S-DFT:R# +/- XX% ←].
3-28
Rosemount Analytical
July 1998
748176-N
Page 71
I
g
g
g
play
y
g
y
g
play
g
y
g
y
g
play
y
g
NITIAL STARTUP AND CALIBRATION
The reference span or zero calibration is the first calibration after the drift feature is toggled to "ON" in the DRIFT LIMIT display (Figure 3-23, STEP 8) or the first calibration after a range is reset from "N" (off) to "Y" (on) in the [RANGE: 1Y 2Y 3Y] display (Figure 3-23, STEPS 12, 14 and 16) if the DRIFT LIMIT has been toggled to ON in the [DRIFT LIMIT: ON] display.
STEP
1MODE 2
3 4 5 6 7
8ENTER
9ENTER
10
11 ENTER
12
13 ENTER 14 15 ENTER 16 17 ENTER 18 19 ENTER
20
21 ENTER
KEYBOARD
ENTRY
→ → → → → →
DIGITAL DISPLAY EXPLANATION
SECURITY RANGE *
DIAGNOSTICS * LINEARIZER * ALARM * CURRENT OUTPUT * AUTO-CAL *
AUTO-CAL: OFF
SH-V:N SH-I:N *
SH-V:N SH-I:N *
RANGE 1Y 2Y 3Y *
RANGE 1Y 2Y 3Y *
RANGE 1Y 2Y 3Y *
DELAY nn HR *
Pressin or changed
moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display
Pressin set or checked.. Use the ↑ or ↓ key to toggle between ON/OFF (Auto Zero/Span feature ON/OFF). This activates the Auto Zero/Span Option. Toggling from OFF to ON resets the CPU and the timers.
Pressin memory.
moves the cursor to the nex t dis ke hold feature for current for voltage).
Pressin memory.
Use the key to move the cursor to SH-I:N. Use the or ke
feature for current "ON") or N (NO) (sample and hold feature for current "OFF").
Pressin memory.
moves the cursor to the nex t dis keys to toggle between Y (YES) or N (NO) for Range 1.
Pressin memory.
Use the ke Y or N.
Pressin memory.
Use the ke Y or N.
Pressin memory.
moves the cursor to the nex t dis ke calibration
Pressin memory.
MODE allows the mode settings to be viewed
ENTER allows the Auto Zero/Span option to be
ENTER stores the setting in non-volatile
s to toggle between Y (YES) or N (NO) (sample and
ENTER stores the setting in non-volatile
s to toggle between Y (YES) (sample and hold
ENTER stores the setting in non-volatile
ENTER stores the setting in non-volatile
to move the cursor to Range 2 and select
ENTER stores the setting in non-volatile
to move the cursor to Range 3 and select
ENTER stores the setting in non-volatile
s to select the number of hours (0...99) until the initial
ENTER stores the setting in non-volatile
. Use the or
. Use the or
. Use the or
F
IGURE
748176-N
3-23. S
NEXT PAGE
ETUP/CHECKOUT OF AUTO ZERO/SPAN OPTION (CONTINUED ON
)
July 1998
Rosemount Analytical
3-29
Page 72
M
play
y
g
play
y
p
y
y
g
play
y
y
g
y
g
y
g
play
y
g
g
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
STEP
22
23 ENTER
24
25 ENTER
26
27 ENTER
28
29 ENTER
30
31 ENTER
32
33 ENTER
34
35 SHIFT+ENTER XXXX ppm H20 L2
KEYBOARD
ENTRY
S-DFT:R1 +/-15%
S-DFT:R2 +/-15%
DIGITAL DISPLAY EXPLANATION
PURGE: nn MIN *
RPT ZERO: nnn HR *
RPT SPAN: nnn HR *
DRIFT LIMIT: OFF *
ZR-DFT: +/-10%
S-DFT:R1 +/-15%
moves the cursor to the nex t dis ke
s to select the number of minutes (2...99) for the
analyzer to be purged with sample or calibration gas. Pressin
memory. moves the cursor to the nex t dis
ke s made each time a span calibration is made, enter onl the number of hours between each zero calibration made without an accompanying span calibration.
Pressin memory.
moves the cursor to the nex t dis ke zero calibration. A zero calibration is automaticall each time a span calibration is made.
Pressin memory.
Use the ke Use the ON/OFF.
Pressin memory.
Use the ke display. Use the ↑ or ↓ keys to select the percent fullscale of zero drift (1...50) allowable for all three ranges.
Pressin memory.
moves the cursor to the nex t dis ke the cursor to the right and select the percent fullscale of span drift (1...50) allowable for the selected range.
With the cur sor under the value, pressing ENTER stores the setting in non-volatile memory.
Use the key to return to range selection and select another ran the cursor under the value to store the setting in non­volatile memory. Repeat for third range.
Pressin mode.
ENTER stores the setting in non-volatile
s to select the number of hours (0...99) between each
an calibration. Since a zero calibration is automatic all
ENTER stores the setting in non-volatile
s to select the number of hours (0...99) between each
ENTER stores the setting in non-volatile
to move the cursor to the next display.
or ↓ keys to toggle the DRIFT LIMIT feature
ENTER stores the setting in non-volatile
to move the cursor move to the next
ENTER stores the setting in non-volatile
s to select range 1, 2 or 3. Use the key to move
e. Repeat step 30, pressing ENTER with
SHIFT+ENTER return the analyzer to RUN
. Use the or
. Use the or
. Use the or
made
. Use the or
F
IGURE
3-23. S
ETUP/CHECKOUT OF AUTO ZERO/SPAN OPTION (CONTINUED FROM
PREVIOUS PAGE
)
3.17 REMOTE RANGE INPUT/OUTPUT OPTION
The Model 880 Analyzer has optional remote input/output capability. When the Remote Range Input/Output Option is switched to REMOTE, in the run mode the range indicator at the right corner of the display will be R# instead of L#. Refer to Table 3-2 for explanations of BIN (binary) and DEC (decimal). When SPECIAL (Figure 3-24, STEP 11) is selected, only autocal status and remote/local output on bits 6 and 7, respectively.
3-30
Rosemount Analytical
July 1998
748176-N
Page 73
I
g
g
p
g
play
y
(
g
g
NITIAL STARTUP AND CALIBRATION
STEP
10 ENTER
11
12 ENTER 13 SHIFT+ENTER X.X %FS CO2 R1
KEYBOARD
ENTRY
1MODE 2
3
4
5
6
7
8
REMOTE I/O
9ENTER
DIGITAL DISPLAY EXPLANATION
SECURITY RANGE *
DIAGNOSTICS * LINEARIZER * ALARM * CURRENT OUTPUT * AUTO-CAL *
CNTRL:REMOTE
IN/OUT:BIN/BIN
Pressin or changed
moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display moves the cursor to the next display
Pressin remote or local in between LOCAL or REMOTE.
Pressin memory.
moves the cursor to the nex t dis ke
binary/decimal), SPECIAL or DEC/DEC
(decimal/decimal). Pressin
memory. Pressin
mode.
MODE allows the mode settings to be viewed
ENTER allows the analyzer to be placed in
ut/output. Use the ↑ or ↓ key to toggle
ENTER stores the setting in non-volatile
. Use the or
s to select BIN/BIN (binary/binary), BIN/DEC
ENTER stores the setting in non-volatile SHIFT+ENTER returns the analyzer to RUN
F
IGURE
3-24. S
ETUP/CHECKOUT OF REMOTE INPUT/OUTPUT OPTION
748176-N
July 1998
Rosemount Analytical
3-31
Page 74
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
OUTPUT
BIT DESIGNATION
0 RANGE I.D. 1 RANGE I.D.
2 RANGE I.D. 3 Not used
4 Not used 5 Not used 6 AUTO-CAL STATUS 7 REMOTE/LOCAL STATUS
INPUT
BIT DESIGNATION
0 RANGE SELECTION IN REMOTE 1 RANGE SELECTION IN REMOTE 2 RANGE SELECTION IN REMOTE 3 Not used 4 Not used 5 Not used 6 AUTO-CAL REQUEST 7 NOT USED
Note: The Auto-Cal request BIT is level triggered and therefore, it is the responsibility
of the user to verify that the BIT is brought low before the analyzer completes the Auto-Cal process.
T
ABLE
T
ABLE
3-1. R
EMOTE RANGE
I/O B
IT DESIGNATION
MODE RANGE BIT 2 BIT 1 BIT 0
BINR3011 BINR2010 BINR1001
DECR3100 DECR2010 DECR1001
3-2. R
EMOTE RANGE
I/O B
INARY AND DECIMAL BIT CODING
3-32
Rosemount Analytical
July 1998
748176-N
Page 75
R
OUTINE OPERATION AND THEORY
4
4.1 ROUTINE OPERATION
First set the range for desired operating range: 1, 2, or 3. Then follow the steps for zero and span (Sections 3.8 and 3.9). Next supply sample gas through the instrument. The Model 880 will now automatically and continuously analyze the sample stream.
As a check of instrument performance, it is recommended that the operator keep a log of the zero/span status.
4.2 RECOMMENDED CALIBRATION FREQUENCY
Maximum permissible interval between calibrations depends on the analytical accuracy required and cannot, therefore, be specified. It is recommended initially that the instrument be calibrated once every 8 hours and that this practice be continued unless experience indicates that some other interval is more appropriate.
A change in cell pressure of 1 inch of mercury (3.38 kPa) will result in a readout error of approximately 3% of reading. Therefore, if barometric pressure changes significantly, it is advisable to recheck the calibration against an upscale standard gas.
4.3 SHUTDOWN
The Model 880 will retain settings during prolonged shutdown. Recalibrate the instrument upon restart.
4.4 DETECTION SYSTEM THEORY
As shown in Figure 4-1, infrared radiation is produced from two separate energy sources. This radiation is interrupted by a chopper at 5 Hz. Depending on the application, the radiation may then be optically filtered to reduce background interference from other infrared-absorbing components.
Each infrared beam passes through a cell, one containing a continuously flowing sample and the other cell sealed or with a continuously flowing reference gas.
During analysis, a portion of the infrared radiation is absorbed by the component of interest in the sample, with the quantity of infrared radiation absorbed being proportional to the component concentration. The detector is a “gas microphone” based on the Luft principle. It converts the difference in energy between sample and reference cells to a capacitance change. This capacitance change, proportional to component concentration, is processed and indicated on the display.
748176-N
July 1998
Rosemount Analytical
4-1
Page 76
M
ODEL
880 NON-D
INFRARED
SOURCE
ISPERSIVE INFRARED ANALYZER
CHOPPER
SAMPLE IN
REFERENCE
CELL
DETECTOR
STATIONARY
BUTTON
SIGNAL
SIGNAL
CONDITIONING
CIRCUITRY
SAMPLE CELL
SAMPLE OUT
DIAPHRAGM DISTENDED
COMPONENT OF INTEREST
OTHER MOL ECULES
F
IGURE
4-2
4-1. F
Rosemount Analytical
UNCTIONAL DIAGRAM OF DETECTION SYSTEM
July 1998
748176-N
Page 77
T
ROUBLESHOOTING
5
5.1 ERROR CODE SUMMARY
In the Run Mode, the error codes described in Table 5-1 may appear on the display. These messages also are shown on the STATUS display in a slightly different f ormat. Error messages in Table 5-1 are listed in order of priority.
Note: When using the DIAGNOSTICS display to help make hardware
adjustments, set the analyzer GAIN to one. Otherwise, the ADC may be in saturation resulting in a false value in the DIAGNOSTICS display.
RUN
MODE
DISPLAY
STATUS DISPLAY EXPLANATION
ERO
ER1 [E1-SPAN #1 LMTS]
ER2 [E2-SPAN #2 LMTS]
ER3 [E3-SPAN #3 LMTS]
ER4
ER5 [E5-ZERO DRIFT]
ER6 [E6-SPAN DRIFT]
Note: If any of the above error messages occur, software will restore previous values.
T
ABLE
[EO-ZERO POT LMTS]
[E4-ADC SATURATED]
5-1. E
RROR CODE SUMMARY
Zero Potentiometer setting is such that mV is required to make a software zero.
Zero must be reset.
Span errors for Range 1, Range 2, or Range 3. Software span is outside limits so that the rum mode value is not between 55% and 99% of the span gas value while in the Span Mode.
Signal into ADC is greater than fullscale rating. Refer to Figure 3-6 and reduce the digital GAIN setting by one value, i.e. 8 to 4, 4 to 2 or 2 to 1. If the GAIN is initially on 1, switch from High to Low gain.
Zero drift limit exceeded. To clear, recalibrate or toggle the drift limit OFF and then ON. See Figure 3-14.
Span drift limit exceeded. T o clear, recalibrate or t oggle the drift limit OFF and then ON. See Figure 3-14.
more
than 500
748176-N
July 1998
Rosemount Analytical
5-1
Page 78
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
5.2 OSCILLATOR TUNE ADJUSTMENT
This procedure should not be performed on a routine basis:
1. Refer to Section 3.8, and access the Oscillator Tune display.
2. Adjust coil knob (Oscillator Tune, located on top of the detector housing) clockwise or counterclockwise until a maximum reading is obtained on the display.
3. Adjust the coil knob counterclockwise until the unit reads between 75 and 80% of the maximum value.
5.3 PREAMP GAIN
The preamp gain is used to adjust the fullscale value at TP2 and the diagnostic display to 7.5V. To prevent saturation, this value must never be higher than 7.5V, fullscale. If this value is too low or is above 7.5V, adjust the preamp gain.
1. Refer to Section 3.8, and access the Detector Signal display.
2. Flow SPAN calibration gas for the least sensitive range through the SAMPLE CELL for a minimum of two minutes.
Example:
Range 1 - 500 ppm CO Range 2 - 2000 ppm CO Range 3 - 5000 ppm CO
In this case, the least sensitive range would be 5000 ppm.
3. If the calibration gas is not equal to fullscale, find the percent fullscale of the calibration gas by looking at the application curve at the back of the manual.
4. Multiply this value by 7.5 and record the resultant value for Step 5. For example, if the SPAN gas is 67 % of fullscale, then:
0.67 x 7.5 = 5
In this case, the value to be used in Step 5 is 5.
5. Adjust the displayed value with preamplifier gain potentiometer R3, located on the Signal Board, for the value obtained in Step 4. THIS VALUE SHOULD NEVER BE
HIGHER THAN 7.5.
Note: For applications with very low concentrations (for example Range 3 = 400
ppm CO), the fullscale value at TP2 and the Diagnostics Display may be considerably less than 7.5V.
5-2
Rosemount Analytical
July 1998
748176-N
Page 79
ROUBLESHOOTING
T
5.4 SOURCE BALANCE SHUTTER ADJUSTMENT
Note: These adjustments are part of the factory checkout and are not normally
required for routine operation, but must be performed whenever the optical system is disturbed (i.e., removal of cells for cleaning).
1. Access Diagnostics Mode (See Section 3.8), and access the Detector Signal display.
2. Flow zero gas (nitrogen) through the SAMPLE CELL for a minimum of two minutes.
3. Slightly loosen the threaded hex standoff on the sample cell shutter adjust screw. The shutter adjust screw is located on top of the motor source assembly (Figure 7-
1)
4. Using a screwdriver, rotate the shutter adjust screw until a minimum reading on the display is obtained. A typical reading is 0.2 to 0.5. Add 0.4 to this value. Use this value for Step 5.
5. Rotate the shutter adjust screw clockwise (viewed from the screw head) until the display reads the value obtained in Step 4.
6. Re-tighten the threaded hex standoff. Ensure that the display does not change.
5.5 SOURCE CURRENT ADJUSTMENT
1. Follow steps 1 through 6 in Figure 3-9 to access the Source Current display.
Refer to DWG 624073. Adjust the trim potentiometer (R9) located on the Power Supply Board to view the desired current on the digital display until the value on the display is within ±10 of the value on the application data sheet. Clockwise adjustment of R9 will increase the value. Counterclockwise will decrease the value.
5.6 VOLTAGE CHECKS
Refer to Section 3.8 and ensure that the voltages for the detector signal and the three power supplies are correct.
5.7 DIGITAL GAIN ADJUSTMENT
The digitally controlled GAIN amplifier does not normally need adjustment, however, in the event that the analyzer cannot be spanned, the GAIN must be adjusted as follows:
1. Follow the STEPS for Spanning the Analyzer in Sections 3.14 (standard analyzer) or 3.15 (analyzer with the Calibration Gas Control Option) and span the analyzer. If the Run Mode value is not between 51% and 100% of the span gas value, while keeping the span potentiometer between 5% and 95%, then the digital GAIN should be adjusted. (The ideal span pot setting is 50%.) Note the final value of the PS (potentiometer status) for Step 4.
748176-N
July 1998
Rosemount Analytical
5-3
Page 80
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
2. Exit to Run Mode.
3. Follow steps 1 through 4 in Figure 3-8 to obtain the GAIN display in the RANGE parameters menu.
4. Change the GAIN setting to a value higher or lower than the original value. The GAIN may be changed to 1, 2, 4, or 8. If the span potentiometer status (PS) was at the top of its range in Step 2 (95%), then the GAIN should be raised. If the span potentiometer status (PS) was at the bottom of its range in Step 2 (5%), the GAIN should be lowered.
5. Press SHIFT/ENTER to return to Run Mode.
6. Repeat Step 1. If the analyzer still cannot be spanned, repeat steps 3 through 5 for a new GAIN value.
5.8 CASE HEATER TEMPERATURE CONTROL
Malfunction in this option can occur in three sections:
EATER
H
Check with an ohmmeter for continuity. The heater resistance is approximately 113 ohms.
EMPERATURE SENSOR
T
This is an RTD and should have approximately 550 ohms at 25°C. Check with ohmmeter for continuity.
VER TEMPERATURE FUSE
O
This is a thermal fuse that opens above 72°C. Check for continuity with an ohmmeter. If the above are functional, refer to Drawing 624073 for circuit diagram and
troubleshoot board.
5.9 DETECTOR HEATER
There are three sections that can cause a malfunction
EATER
H
Check with an ohmmeter for continuity. The heater resistance is approximately 113 ohms.
EMPERATURE SENSOR
T
This is an RTD and should have approximately 550 ohms at 25°C. Check with ohmmeter for continuity.
VER TEMPERATURE FUSE
O
This is a thermal fuse that opens above 72°C. Check for continuity with an ohmmeter.
5-4
Rosemount Analytical
July 1998
748176-N
Page 81
ROUBLESHOOTING
T
PROBLEM PROBABLE CAUSE ACTION TO BE TAKEN
Check fuse; reset if necessary.
DIGITAL DISPLAY DOES NOT
LIGHT UP
UNABLE TO SPAN
ANALYZER
U
NSTABLE, NOISY SIGNAL
E
XCESSIVE DRIFT WITH CHANGING AMBIENT TEMPERATURE
L
OW SENSITIVITY
No power to analyzer
Check input power and connections.
Check Power Supply Board. CPU Board Digital GAIN needs
adjustment Preamplifier gain needs
adjustment Source current needs
adjustment
Check power connector to CPU
Board.
See Section 5.7
See Section 5.3
See Section 5.5 Oscillator tune See Section 5.2
Time constant too low See Section 3.7 Temperature Control
Assembly Sources need to be
replaced
See Section 5.8
The resistance of both sources
should be 24 ±3 ohms.
B
ASELINE DRIFT
E
RROR MESSAGES WHEN
ATTEMPTING TO CALIBRATE
U
NABLE TO SPAN OR ZERO
ANALYZER
F
IGURE
5-1. T
ROUBLESHOOTING CHART
Loss of source current See Section 5.5 Source balance See Section 5.4
Detector signal should be
between 0.4 and 1 volt with zero Check Diagnostics Display
gas, 7.5 volts, maximum, on
high range with 100% fullscale
span gas. Adjust GAIN
potentiometer, see Section 5.3 Calibration parameters
incorrect
The zero value must be close to
zero before pressing ENTER.
The concentration display (left
side of display) must be less
than the span gas displayed in
the right side of the display, but
greater than 50% of the
fullscale range.
748176-N
July 1998
Rosemount Analytical
5-5
Page 82
M
OU
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
OSCILLATOR BOARD
Information Signal (5 Hz capacitance change)
SOURCE
Detector (Modulator)
SIGNAL BOARD
BUFFER
AMPLIFIER
LOWPASS
FILTER
DIGITALLY CONTROLLED
OSCILLATOR
AMPLIFIED
UNFILTERED
5 Hz SIGNAL
ZERO
ZERO
SUPPRESSION
REF
AMPLITUDE-MODULATED
10 MHz CARRIER
WITH NOISE
5 Hz
BANDPASS
FILTER
DIGITAL
GAIN
X 1 X 2 X 4 X 8
FILTERED OUT
SPAN 1, 2, 3 DIGITALLY CONTROLLED
VOLTAGE
D
BLER
5 Hz
FULLWAVE RECTIFIER
MUX ADC MICRO
HALFWAVE RECTIFIED CARRIER
FULLWAVE
RECTIFIED 5 Hz
SIGNAL
TIME
CONSTANT SOFTWARE
SPAN
DAC
DACLIN
SOFTWARE
ZERO
F
IGURE
5-6
5-2. S
Rosemount Analytical
IGNAL WAVEFORMS
July 1998
748176-N
Page 83
A. Functional Diagram - Circuitry in Tune Mode
Tank Circuit Voltage Doubler Rectifier
Diaphragm Undistended
DETECTOR
Stationary
L1
10 MHz Oscillator
Metal Button
Reference Chamber
Sample Chamber
ROUBLESHOOTING
T
Detector Oscillator Tune Output
OSC. TUNE
B. Tank Circuit Resonance Curves
Carrier
Amplitude
Max. Obtainable Amplitude
75% to 80% of max.
Diaphragm at maximum Distention
Curve 1. OSCILLATOR TUNE Control set for maximum obtainable meter reading.
Curve 2. OSCILLATOR TUNE Control set for 75% to 80% of maximum obtainable meter reading.
Curve 3. Sample beam blocked, causing maximum distention of diaphragm.
Amplitude with Sample Beam Blocked
F
IGURE
748176-N
Tank Circuit Resonant Frequency
5-3. M
Crystal Frequency (10 MHz)
ODULATION SYSTEM
Decrease in inductance and/or capacitance in tank circuit shifts resonance curve to right, decreasing carrier amplitude.
July 1998
Note:
Rosemount Analytical
5-7
Page 84
ODEL
M
N
OTES
880 NON-D
ISPERSIVE INFRARED ANALYZER
5-8
Rosemount Analytical
July 1998
748176-N
Page 85
R
OUTINE SERVICING
WARNING: ELECTRICAL SHOCK HAZARD
Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
Note: Before servicing analyzer, disconnect power and shut off sample flow to
unit.
6.1 CELL REMOVAL, CLEANING AND INSTALLATION
1. Slide chassis out.
2. Remove sample lines:
6
a. From the end cap assemblies (long cell). b. From the sample cell (short cell).
3. Remove the two hold-down screws on the motor/source assembly (Figure 7-1).
4. Gently move the motor/source assembly away from the detector.
5. For long cell models, support the cell while performing Step 4. The cell and its O-rings will now be free.
6. For short cell models. the three retaining screws holding the two end caps and sample cell must be removed. The cell, end caps and O-rings will now be free.
7. Rinse the cell with acetone. If this does not remove all foreign matter, use a soft brush. Do not use any metallic object inside the cell because it will scratch the gold plating.
8. After all matter has been removed, rinse the cell with distilled water and allow to air dry. Do not use towels.
9. Inspect the cell inside by holding it up to a bright light. If particles are seen, repeat Steps 7 and 8 as often as necessary.
10. After cleaning cell, examine O-rings at the source, detector, and end caps. If damaged, replace with new O-rings.
748176-N
July 1998
Rosemount Analytical
6-1
Page 86
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
11. Remove any contaminants from windows with a lint free cloth soaked in acetone. Do not use alcohol or other solvents. Allow to air dry.
12. To replace the cell perform the following steps:
a. Long cell - fit it into position. Make sure that the O-rings seat properly. b. Short cell - replace the three retaining screws holding two end caps. Make sure
that the O-rings seat properly.
c. Move the motor/source assembly back into position. Make sure that the
O-rings seat properly.
13. Replace the two hold-down screws on the motor/source assembly. Do not overtighten.
14. Check for leaks as instructed in Condensed Startup and Standardization Procedure (Section A). Take corrective action if necessary.
GAS DESICCANT PART NUMBER
2
CO
Cardoxide 096218
4)2
2
2
2
2
2
096217 096217 096217 096217 096217
096217/096218
T
ABLE
6-1. T
CO Mg (CI04) H2OMg (CI0 SOS Mg (CI04)
4
CH Hexane Mg (CI04) CO + CO
2
Mg (CI04)
Cardoxide + Mg (CI04)
YPES OF DESICCANT
6.2 CELL DESICCANT
The reference cell may use a flowing reference. If so, desiccant is required. A desiccant holder is used on the inlet and outlets to keep moisture from entering the
reference cell. The desiccant should be replaced each time the cell is opened. To determine the type of desiccant used, refer to Table 6-1.
6.3 SOURCE REPLACEMENT
Refer to Figure 6-2. Sources are marked with the resistance value, for example, 11.5
- 11.6 in matched pairs. Install the higher value as the reference source.
Note: Refer to Figure 7-1. Observe how the parts are disassembled so that the
reverse procedure can be used for reassembly.
1. Loosen the two screws on the front of the case and slide the front panel forward.
6-2
Rosemount Analytical July 1998
748176-N
Page 87
OUTINE SERVICING
R
2. Remove the two screws holding the source housing to the chopper housing.
3. Remove the two screws holding the source to the source housing. Note how the
source is mounted. There is a front and back side.
4. If replacing the source, insure that its orientation is exactly the same as the old
one. Each source is marked on the back. Install the source with the higher designation at the reference site.
5. Reverse the procedure outlined above to reinstall the new source assembly,
6.4 SOURCE BALANCE SHUTTER ADJUSTMENT
When the sources are replaced, follow the Source Balance Procedure in Section 5.4 to adjust the source balance shutter.
6.5 CHOPPER MOTOR ASSEMBLY (P/N 658313)
1. Remove Source Assembly front cover.
2. Remove chopper blade.
3. Remove two screws from rear of motor and remove motor.
6.6 MICROBOARD REPLACEMENT
All calibration constants and settings stored in nonvolatile memory must be changed when the microboard is replaced. These procedures are given in Section Three, Initial Startup and Calibration.
Additionally, the zero potentiometer and the three span potentiometers must be resynchronized with software. To resynchronize the zero potentiometer follow the steps listed below:
1. Follow the instructions in Figure 3-11 for the standard instrument or 3-12 for the
instrument with Calibration Gas Control Option and access the [ZR=X.X PS=XX% ] display.
2. Press SHIFT then ↓.
3. Proceed with the zero calibration. The span potentiometer must be resynchronized for all three ranges as follows:
1. Choose the first range by selecting a range using the range parameters in Figure
3-8.
2. Follow the instructions in Figure 3-14 for the standard instrument or 3-15 for the
instrument with Calibration Gas Control Option and access the [X.XX XX% 100 ↓] display.
748176-N
July 1998
Rosemount Analytical
6-3
Page 88
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
3. Press SHIFT then ↓.
4. Proceed with the span calibration.
6.7 DETECTOR REMOVAL
Refer to Figure 7-1, Part 5. Slide chassis out.
1. Remove sample lines:
a. From the end cap assemblies (long cell). b. From the sample cell (short cell).
2. Remove top cover from detector housing.
3. Remove entire detector assembly by removing the four screws from the base
plate.
4. For long cell configurations, support the cell while performing Step 4. The cell and
its O-rings will now be free.
5. Remove oscillator board.
6. Remove dual end cap assembly for long cells or sample cell and cell assembly for
short cell.
7. Remove two screws holding detector to detector base.
8. Remove detector.
9. Wipe off all the white heat sink compounds from the base plate.
10. Apply new heat sink compound to the detector (supplied with the new detector).
11. Replace detector and reverse the removal process.
Note When replacing detector, insure that the thermal fuse and temperature
sensor mounted in the base plate is in good thermal contact with the base plate.
6.8 ELECTRONIC CIRCUITRY
6.8.1 O
SCILLATOR CIRCUIT BOARD AND ASSOCIATED ELEMENTS OF
MPLITUDE-MODULATION CIRCUIT
A
In the Oscillator Circuit (DWG 623995) the 10 MHz carrier wave is generated by a crystal-controlled radio-frequency oscillator using crystal Y1 and transistors Q1 and Q2.
6-4
Rosemount Analytical July 1998
748176-N
Page 89
OUTINE SERVICING
R
The modulation circuit is driven by the detector. the sensing element of the analyzer. Mechanical functioning of the detector is explained in Paragraph 4.4. Considered electronically. the detector is a two-plate variable capacitor. The modulator is coupled inductively, through one winding of inductance T1, to the oscillator. Amplitude of the 10 MHz carrier thus varies with the 5 Hz modulation signal.
6.8.2 F
UNCTIONING OF MODULATION SYSTEM IN
TUNE M
ODE
Preparatory to oscillator tuning, access Oscillator Tune (OT=XX) in the Diagnostic Display (Figure 39). In this mode the display indicates the rms value of the halfwave-rectified carrier. The tank circuit is now adjusted in the following two-step sequence:
1. Tuning: Initially, the OSC TUNE adjustment is set somewhat counterclockwise
from its correct setting. Then, it is rotated clockwise to move the slug into the core, thus increasing inductance and decreasing resonant frequency. The adjustment is set for maximum obtainable reading. At this setting, tank-circuit resonant frequency is the same as oscillator frequency (i.e., nominal 10 MHz). See Resonance Curve Number 1, Figure 5-2.
2. Detuning: By counterclockwise rotation of the OSC TUNE adjustment, the slug is
partially withdrawn from the core, thus decreasing inductance and increasing resonant frequency. The adjustment is set so reading decreases to between 75% and 80% of the maximum obtainable value noted in Step 1, above. See Resonance Curve Number 2, Figure 5-2. This curve has the same shape as that obtained in Step 1, immediately preceding, but is displaced to the right.
6.8.3 F
UNCTIONING OF MODULATION SYSTEM IN OPERATING MODE
Overall sensitivity of the analyzer system may now be checked by placing span gas in the sample beam to simulate absorption of sample-beam energy and thus provide the maximum obtainable 5 Hz detector-output signal. During that portion of the chopping cycle, while the chopper is not blocking the sample and reference beams, the diaphragm distends away from the metal button, thus decreasing detector capacitance and shifting the tank-circuit resonance curve to the right. At the moment the diaphragm reaches maximum distention, the curve reaches the position of Curve 3, Figure 5-3.
The diaphragm now pulses cyclically, causing the resonance curve to move continuously back and forth within the limits defined by Curves 2 and 3 of Figure 5-2. Carrier amplitude decreases as the curve moves to the right and increases as it moves to the left. Thus, the response characteristics of the system depend on the location of Curve 2. Position of this curve depends on the degree of tank-circuit detuning used. By detuning to 70% to 75% of the maximum obtainable carrier amplitude and operating on the portion of the curve thus obtained. maximum slope yields highest sensitivity and minimum curvature provides best linearity.
748176-N
July 1998
Rosemount Analytical
6-5
Page 90
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
6.8.4 R
ADIO-FREQUENCY DEMODULATOR
The amplitude-modulated 10 MHz carrier from the detector/oscillator circuit is applied to the radio-frequency demodulator. This circuit is a voltage-doubler type rectifier utilizing diodes CR1, CR2, CR3, CR4, and capacitor C7. The circuit gives approximately double the output voltage of a conventional halfwave rectifier. This result is obtained by charging a capacitor during the normally wasted half-cycle, and then discharging it in series with the output voltage during the next half-cycle.
6.8.5 S
IGNAL BOARD
(DWG 624085)
The 5 Hz sine wave detector signal goes through an AC amplifier U1A and associated resistor. The output signal goes through bandpass filter network U2 and U4 to remove harmonics and distortion. The signal next goes through a precision signal rectifier U3 and Q1 and then through low pass filter U5. This output goes through an RC low pass filter R29, C13 and U7 and then to inverting buffer amplifier U8 with zero control U11. The signal goes through a range amplifier consisting of eight bit DAC U9 and the amplifier U10.
The GAIN is digitally controlled via U15, U16 and U18 resulting in a selectable digital GAIN of X1, X2, X4 and X8.
The span is controlled for the three ranges with SPAN 1 (U12), SPAN 2 (U13) and SPAN 3 (U14).
The eight-channel multiplexer (U17) selects the input signals by commands applied to the switch driver and feeds the selected signal to pin two of J2.
6.8.6 P
OWER SUPPLY BOARD
(DWG 624073)
The power supply board supplies the different voltages to the various boards. Additionally. the power supply board includes an adjustable source driver circuit. a chopper motor driver circuit, a proportional temperature controller circuit and a DC to AC converter for backlight.
6.8.7 A
DAPT OR BOARD
(DWG 624127)
The adaptor board which includes a resettable fuse is used for line power distribution. The adaptor board also serves as an interface board for all the option boards and provides the recorder output on TB2.
6.8.8 M
ICROBOARD
The microboard is a self-contained circuit assembly which includes an advanced microprocessor and multiple I/O functions with a complete analog domain consisting of analog-to-digital converters and digital-to-analog converters. Multiple output registers allow the transmission of digital data to and from the board under program control. The circuit board can be used alone or in conjunction with I/O boards that satisfy special interfacing requirements such as the following:
6-6
Rosemount Analytical July 1998
748176-N
Page 91
OUTINE SERVICING
R
1. Current Output
2. Bi-directional Remote Range I/O board
3. Dual Alarm assembly
4. Auto Zero/Span
5. Calibration Gas Control The board is configured with an analog domain that allows the processing of analog
signals directly with a 12-bit plus sign ADC. In addition, two independent DAC's, each 12 bytes wide, allow the presentation of analog voltages for peripheral functions immediately.
6.8.9 O
PTIONAL CASE HEATER TEMPERATURE CONTROL BOARD
(DWG 624003)
This is a proportional temperature controller. which works on a variable time method. Resistors R7, R8, R9, R10, R11 and the sensor form a bridge which feeds a
comparator, AR1. AR1 operates in an ON/OFF mode to drive transistor Q3. The sensor is a resistor with a positive temperature coefficient (1.925 ohms/oC). The resistance is 500 ohms at 25 oC.
Resistors R1 through R6, Q1, Q2 and C1 provide the circuit for the time proportioning action: C1 charges until the voltage on C1 reaches 9.0 V. Q1 then discharges C1, and the charging process repeats itself. The emitter of Q2 follows the voltage on C1, which is essentially a sawtooth. This is injected into the bridge, which causes the setpoint to bump on a variable time basis. Q3 (through LED CR1 ) triggers optical coupler U1 which gates TRIAC (U2). U2 allows fullwave VAC to flow through the case heater element.
6.8.10 D
UAL ALARM/CALIBRATION GAS CONTROL OPTION BOARD
(DWG 624204)
This board is used for both dual alarm and calibration gas control, depending on the position of the jumper in the jumper-selectable address. This is a peripheral circuit function which communicates with the computer via an 8-bit buss arrangement. This option consists of two form C contacts rated 3A 125/250 VAC or 5A 30 VDC. (resistive). This circuit board satisfies a dual alarm requirement, as it provides two medium power relays that can be independently controlled from the central processor. Also, the board can be used to connect user-supplied solenoid valves to zero and span calibration gases for one-man calibration. Provision is made to assign a specific address in the range 0 through 7 using jumpers.
6.8.11 I
SOLATED REMOTE RANGE
I/O O
PTION BOARD
(DWG 624251)
The Remote Range I/O board is a peripheral circuit function which communicates with the computer via an 8-bit buss arrangement. This assembly provides isolated two way communication between the host instrument and external user devices. Provision is made to assign a specific address in the range 0 through 7 using jumpers.
748176-N
July 1998
Rosemount Analytical
6-7
Page 92
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
6.8.12 A
UTO ZERO/SPAN OPTION BOARD
(DWG 624599)
The auto zero/span board is a peripheral circuit function which communicates with the computer via an 8-bit bus. With the appropriate software it satisfies the auto zero/span requirement. The assembly provides 6 form C relay contact outputs, 4 of which are suitable for medium power requirements, the remaining two are relegated to alarm or indicator functions. Snubbers are provided for the medium power relays. Provision is made to assign a specific address in the range 0 through 7 using jumpers. The auto­cal request bit is level triggered and therefore, the request line must be brought low before the analyzer completes the Auto-Cal process.
6.8.13 C
URRENT OUTPUT OPTION BOARD
(DWG 624092)
This board changes the instrument voltage output to an isolated current output for use with external recorders or data gathering systems.
6-8
Rosemount Analytical July 1998
748176-N
Page 93
R
EPLACEMENT PARTS
7
The following parts are recommended for routine maintenance and troubleshooting of the Model 880 Non-Dispersive Infrared Analyzer. If the troubleshooting procedures do not resolve the problem, contact your local Rosemount Analytical service office. A list of Rosemount Analytical Service Centers is located in the back of this manual. Figures 7-1 through 7-5 show locations of components and assemblies.
!
WARNING: PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory-documented components for repair.
7.1 CIRCUIT BOARD REPLACEMENT POLICY
In most situations involving a malfunction of a circuit board, it is more practical to replace the board than to attempt isolation and replacement of the individual component. The cost of test and replacement will exceed the cost of a rebuilt assembly. As standard policy, rebuilt boards are available on an exchange basis.
Because of the exchange policy covering circuit boards the following list does not include individual electronic components. If circumstances necessitate replacement of an individual component, which can be identified by inspection or from the schematic diagrams, obtain the replacement component from a local source of supply.
7.2 SELECTED REPLACEMENT PARTS
While the following sections list parts which are common to all Model 880 applications, the configuration number is required when ordering parts which are specific to an individual application. The configuration number (8801-XX or 8802-XX) is on the Data Sheet in the rear of this manual.
748176-N
623998 Oscillator Board 624006 Temperature Control Board 624076 Power Supply Board 624088 Signal Board
July 1998
Rosemount Analytical
7-1
Page 94
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
623785 Micro Board 622733 Fan, Heater (if used) 622732 Heater (if used) 624433 Thermal Fuse, Fan Heater (if used) 658313 Chopper Motor 624442 Source (Matched Pair) 898733 Detector Thermal Fuse 620298 Detector Heater 622917 Detector Temperature Sensor
Sensor 622917
Temperature Control Board 624006
F
IGURE
7-2
Fan Assembly 622733
7-1. C
Rosemount Analytical July 1998
ASE HEATER TEMPERATURE CONTROL ASSEMBLY
Heater Assembly 622732
748176-N
Page 95
Chopper Motor
EPLACEMENT PARTS
R
Source (Matched Pair)
Source Housing 622706
Shutter Adjustment Screw
Screw
O-Ring 900406
Jam Nut
Teflon Spacers
Washer, Nylon
Chopper Housing 624166 (Sapphire Windows) 624411 (Irtran Windows)
Chopper Blade 620300
Chopper Cover 624167
F
IGURE
748176-N
7-2. M
OTOR/SOURCE ASSEMBLY
July 1998
Rosemount Analytical
7-3
Page 96
M
ODEL
880 NON-D
ISPERSIVE INFRARED ANALYZER
Power Supply Board 624076
Case Heater Temperature Control Option (See Figure 7-1, Part 1)
Transformer 622751
Power Amplifier 898724
Optical Bench (See Figure 7-1, Part 5) Note: Long Cell shown.
Resistor Assembly 622700
F
IGURE
7-4
Micro Board 623785
7-3. M
Rosemount Analytical July 1998
ODEL
880 M
AJOR COMPONENTS
Signal Board 624088
748176-N
Page 97
)
A. LONG CELLS
g
g
End Cap/Optical Filter Assembly
Detector Assembly (See Figure 7-1, Part 5)
Desiccant Holders
Sample In
Sample Out
End Cap Assembly
Motor/Source Assembly (See Figure 7-1, Part 2)
O-Rings
EPLACEMENT PARTS
R
Motor/Source Assembly to Optical Bench Plate mountin
screws
Detector Cover removed for clarity
B. SHORT CELLS
Desiccant
Detector Assembly (See Figure 7-1, Part 5
Holder
Detector Cover removed for clarity
Oscillator Tune Adjust
Detector Cover
Oscillator Tune Adjust
Detector Cover
Sample In/Out (interchangeable)
O-Rings
Motor/Source Assembly to Optical Bench Plate mountin
Motor/Source Assembly (See Figure 7-1, Part 2)
End Cap/Filter Assembly, Sample Cell to Detector
Cell Assembly
Optical Bench
Cells
Optical Bench
screws
Plate Assembly
Compression Gaskets
End Cap/Optical Filters Assembly
O-Rings
F
IGURE
748176-N
7-4. O
PTICAL BENCH ASSEMBLY
July 1998
Rosemount Analytical
7-5
Page 98
ODEL
M
880 NON-D
A. LONG CELLS
ISPERSIVE INFRARED ANALYZER
Detector Cover
Desiccant Holder
Sample Li ne
B. SHORT CELLS
Detector Assembly
Detector Cover
Detector Heater 620298
Detector Plate
Desiccant Holder
O-Rings
End Cap/Optical Filter Assembly
O-Rings
Sample Lines
Cell Assembly
End Cap/Optical Filters Assembly
O-Rings
O-Rings
Compression Gaskets
Sample Li ne
End Cap Assembly
Optical Bench Plate
C. DETECTOR ASSEMBLY
F
IGURE
7-5. D
ETECTOR ASSEMBLY REMOVAL/INSTALLATION
Oscillator Board 623998
Thermal Fuse 898733
Pad
Detector Base
Detector
Detector Pad
Pad
Temperature Sensor 622917
Optical Bench Plate
7-6
Rosemount Analytical July 1998
748176-N
Page 99
ENERAL PRECAUTIONS FOR HANDLING AND
G
TORING HIGH PRESSURE GAS CYLINDERS
S
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981 Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202 Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125°F (52°C). A flame should never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
4125 E
AST LA PALMA AVENUE
• A
Rosemount Analytical Inc.
J
ULY
NAHEIM
ALIFORNIA
1997 • 748525-C • P
, C
92807-1802 • 714-986-7600 • FAX 714-577-8006
RINTED IN
USA
Page 100
(blank)
Loading...