Rosemount Analytical designs, manufactures and tests its products to meet many national and
international standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them to ensure they continue to operate within their
normal specifications. The following instructions MUST be adhered to and integrated into your
safety program when installing, using, and maintaining Rosemount Analytical products. Failure to
follow the proper instructions may cause any one of the following situations to occur: Loss of life;
personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Rosemount Analytical representative
for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of
the product.
• Install your equipment as specified in the Installation Instructions of the appropriate
Instruction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and
maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by
Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe
operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result
in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified persons, to prevent electrical shock and personal
injury.
The information contained in this document is subject to change without notice.
Teflon is a registered trademark of E.I. duPont de Nemours and Co., Inc.
SNOOP is a registered trademark of NUPRO Co.
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
e-mail: gas.csc@EmersonProcess.com
613561 Schematic Diagram, Bridge Power Supply - Regulated 5 to 15V
619710 Schematic Diagram, 15V Power Supply
624003 Schematic Diagram, Temperature Controller
652813 Schematic Diagram, Isolated Current Output
652863 Schematic Diagram, Linearizer Board
654616 Schematic Diagram, Master Board
654625 Installation Drawing, Model 7D
654642 Wiring Diagram, Model 7D
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Instruction Manual
748221-M
July 2002
Model 7D
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 7D
PREFACE
The purpose of this manual is to provide information concerning the components,
functions, installation and maintenance of the Model 7D.
Some sections may describe equipment not used in your configuration. The user should
become thoroughly familiar with the operation of this module before operating it. Read
this instruction manual completely.
DEFINITIONS
The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout
this publication.
DANGER .
748221-M
July 2002
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial
property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in injury, death, or long-term health hazards of perso nnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure,
condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Instruction Manual
748221-M
July 2002
Model 7D
INTENDED USE STATEMENT
The Model 7D is intended for use as an industrial process measurement device only. It is not intended for
use in medical, diagnostic, or life support applications, and no independent agency certifications or
approvals are to be implied as covering such application.
SAFETY SUMMARY
If this equipment is used in a manner not specified in these instructions, protective systems may be
impaired.
AUTHORIZED PERSONNEL
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site
property, all personnel authorized to install, operate and service the this equipment should be
thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE
INSTRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can
cause death or serious injury. Refer servicing to qualified pe rsonnel.
For safety and proper performance this instrument must be connected to a properly grounded
three-wire source of power.
Before supplying electrical power to the analyzer, remove power to the bridge by disconnecting the
red lead from the bridge to TB1-1 or TB1-2 (depending on the bridge polarity). See drawing 654642.
To safeguard against filament damage, this lead should remain disconnected until proper gas flow
has been established.
WARNING.
ELECTRICAL SHOCK HAZARD
Do not operate the Model 7D Explosion-Proof Analyzer without the lens cover and door in place
with all bolts secured, unless location have been determined to be non-hazardous.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 7D
WARNING.
POSSIBLE EXPLOSION HAZARD
Ensure that all gas connections are made as labeled and are leak free. Improper gas connections
could result in explosion or death.
WARNING .
POSSIBLE EXPLOSION HAZARD
This analyzer is of a type capable of analysis of sample gases which may be flammable. If used for
analysis of such gases, the instruments explosion-proof encl osure must be suitable for the gas.
If explosive gases are introduced into this analyzer, the sample containment system must be carefully leak-checked upon installation and before initial startup, during routine maintenance and any
time the integrity of the sample containment system is broken, to ensure the system is in leak-proof
condition. Leak-check instructions are provided in Section 2-4 on page 2-3.
748221-M
July 2002
Internal leaks resulting from failure to observe these precautions could result in an explosion
causing death, personal injury or property damage.
CAUTION .
PRESSURIZED GAS
This analyzer requires periodic calibration with known zero and standard gases. See General Precautions for Ha ndling and Storing High Pressure Gas Cylinders, page P-4
CAUTION
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product.
Use only factory documented components for repair.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Instruction Manual
748221-M
July 2002
Model 7D
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYL INDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed
Gases" published in 1981
Compressed Gas Association
1235 Jefferson Davis Highway
Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather
and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when
located in areas where extreme temper atures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or
placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a
suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is
attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125°F (52°C). A flame should never be
permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding,
precautions must be taken to prevent striking an arc against the cylinder.
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 7D
DOCUMENTATION
The following Model 7D instruction materials are available. Contact Customer Service Center or the local
representative to order.
748221 Instruction Manual (this document)
COMPLIANCES
The explosion-proof Model 7D is approved by Factory Mutual Research Corp. (FMRC) for installation in
Class I, Groups B, C, and D, Division 1 hazardous locations as defined in the National Electrical Code
(NEC), ANSI/NFPA-70.
FM
748221-M
July 2002
APPROVED
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Instruction Manual
748221-M
July 2002
Model 7D
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
Instruction Manual
748221-M
July 2002
SECTION 1
DESCRIPTION AND SPECIFICATIONS
1-1 OVERVIEW
The Model 7D Thermal Conductivity Analyzer is
designed to continuously measure the
concentration of a single component of interest
in a flowing gas mixture. The measurement is
based on the different thermal conductivity's of
the individual components of the sample
stream. The method is especially well suited to
analysis of two-component sample streams.
However, analysis of multi-component streams
is possible if the various components of the
background gas occur in relatively constant
ratio, or have similar thermal conducti vity's.
Each Model 7D Analyzer is factory -assembled,
as ordered, for determination of a specified
component, with specified range of
concentration, contained in a background
component or background mixture of known
composition. Typical examples include: 0 to
100 % hydrogen in nitrogen; 20 to 50 % helium
in methane; and 0% to 3% carbon dioxide in air.
If so ordered, the instrument is provided with
two or three ranges; selectable via a front -panel
switch. Information specific to the individual
instrument is provided in the Data Sheet in the
rear of this manual inserted in the back of this
instruction manual.
A Model 7D Analyzer consists of an analyzer
module, Section 1-2 below, and, if ordered, an
accessory gas selector panel, Section 1-3 on
page 1-4.
1-2 ANALYZER MODULE
The analyzer module is supplied in an
explosion-proof enclosure suitable for
installation in hazardous locations classified as
Class I, Division 1, Groups B, C, and D per the
National Electrical Code (ANSI/NFPA 70) (See
DWG 654625).
a. Thermal Conductivity Cell
sample and reference gases. In all
applications, the sample passage
receives a continuous flow of sample
gas. Depending on the application, the
reference passage may receive a
continuous flow of reference gas, or
may have the reference gas sealed
within.
The sample passage contains a pair of
temperature-sensitive resistive
filaments. The reference passage
contains a similar pair. Electrically, the
filaments are connected as legs of a
Wheatstone bridge. An internal
voltage-regulated power supply is
connected via a 20-ohm dropping
resistor, to the bridge.
With the power supply output adjusted
to provide an appropriate voltage
across the bridge, an electric current
flows through the filaments, heating
them and thus increasing their electrical
resistance. The heat -dissipation rate
for each filament depends on the
thermal conductivity of the surrounding
gas.
Initially, with suitable downscale
calibration gas flowing through the
sample passage (and also through the
reference passage if of the flow-through
configuration), the bridge is balanced.
Thereafter, any change in the relative
proportions of the components passing
through the sample passage changes
the thermal conductivity of the gas
mixture, causing a temperature
differential between sample and
reference filaments. The resultant
change in filament resistance
unbalances the bridge, applying a
signal to the electronic circuitry (Section
1-2b on page 1-3).
The thermal conductivity cell is a metal
block with separate passages for the
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
Model 7D
748221-M
July 2002
ALARM 1 ALARM 2
50%
0% 100%
Rosemount Analytical
Model 7D
ZERO SPAN
RANGE
75%25%
1 3
50%
0% 100%
Model 7D
Thermal Conductivity
Analyzer
75%25%
2
ZERO SPAN
ALARM 1 ALARM 2
25%
50%
75%
RANGE
1 3
0%100%
Rosemount Analytical
Figure 1-1. Model 7D Front Panel Controls
50%
2
25%
0%100%
Thermal Conductivity
75%
Analyzer
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
Alarm Kit
±
15V Power Supply
Linearizer Board
Isolated Current Output Board
Bridge Power Supply
Temperature Control Assembly
Master Board
AC Power Input
Instruction Manual
748221-M
July 2002
b. Electronic Circuitry
The analyzer module contains solid-state
circuitry that conditions the
bridge-imbalance signal as required to
provide readout on the front -panel meter.
In addition, a field-selectable output for a
voltage-type recorder is provided as
standard. A field-selectable output of 4
to 20 mA for a current -actuated recorder
or other device is obtainable through use
of an optional plug-in circuit bo ard. A
calibration curve can be used to convert
meter or recorder readings into
concentration values. Typical calibration
curves are supplied for standard ranges.
Calibration curves for special ranges are
available as options.
To avoid use of a calibration curve in an
application where it would otherwise be
required, the analyzer may be equipped
with an optional linearizer board. If so,
the linearizer is factory set for a given
range only, and is not usable on another
range. Note that a linearizer is usable
only if nonlinearity at midscale does not
exceed approximately 20% of fullscale.
REF Hinge Side
Fuse
Cell Enclosure
Terminal Block TB6
Figure 1-2. Model 7D – Location of Major Components
Rosemount Analytical Inc. A Division of Emerson Process ManagementDescription and Specifications 1-3
Instruction Manual
CAL GAS
1 Reference Gas
Sample/Calibration
Downscale
Upscale
Sample Gas
1 Reference Gas
748221-M
July 2002
Model 7D
1-3 GAS SELECTOR PANEL
If so ordered, the analyzer module is provided
with an appropriate gas selector panel, Figure
1-3 below. The gas selector panel permits
selection, flow adjustment, and flow
measurement for the various gases: sample;
flowing reference gas, if used; and downscale
and upscale calibration gases. Proper choice
of a gas selector panel depends on:
Flow Meter
Calibration Gas
Needle Valve
DOWNSCALE
1. Configuration of the thermal conductivity
cell, i.e., flowing or sealed-in reference
gas.
2. Composition of the sample stream. For
non-corrosive streams, the gas selector
panel is assembled with brass
components. For corrosive streams,
stainless steel is used.
Gas Flow Meter
SAMPLE
Needle Valve
Calibration Gas
Needle Valve
UPSCALE
CAL GAS
REF
SAMPLE
REFERENCE
Needle Valve
1 Provided only if thermal conductivity cell uses flowing reference gas.
Figure 1-3. Typical Gas Selector Panel
DESCRIPTIONPART NUM BER
Brass and Copper construction for use with sealed reference113357
Stainless steel construction for use with sealed reference113920
Brass and Copper construction for use with flowing reference117195
Stainless steel construction for use with flowing reference118210
Table 1-1.Available Gas Selector Panels
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
1-4 SPECIFICATIONS
Reproducibility ............................. ±0.5% of fullscale
Zero Drift1................................... ±1% of fullscale per 24 hours
Span Drift1................................... ±1% of fullscale per 24 hours
Noise.......................................... Less than ±0.5% of fullscale
Cell Response Time2................... 30 seconds for 95% response, with sample flow of 250 cc/min.
Sample Flow................................ Nominal, 50 to 350 cc/min; recommended, 250 cc/min.
Calibration Gas Flow.................... Nominal, 50 to 350 cc/min; recommended, 250 cc/min.
Reference Gas Flow (If Req’d) ...... 5 to 50 cc/min.
Supply Pressure.......................... 10 to 50 psig (69 to 345 kPa)
Meter .......................................... Indicating analog meter is standard.
Operating Ranges ........................ Various zero-based and zero-suppressed ranges, from 0% to
Ambient Temperature Range ........ 32°F to 100°F (0 °C to 38°C). Case Temperature controlled at
Output Voltage
Standard Non-Linearized....... Switch selectable: 0 to 10 mV, 0 to 100 mV, 0 to 1V or 0 to 5V DC
Optional Linearized ............... Switch selectable: 0 to 10 mV, 0 to 100 mV, 0 to 1V or 0 to 5V DC
Isolated Current Output (Option) ... 4 to 20 mA, maximum load 1500 ohms
Dual Alarms (Option).................... Relay contact rating: 1.0 A, 120V AC; 5.0 A, 120V DC, resistive
Cell Materials (Standard Cell) ....... 316 stainless steel block with tungsten or Hitempco filaments.
Power Requirements.................... 115/230 VAC ±10%, 50/60 Hz, 250 Watts
Enclosure.................................... Class I, Groups B, C and D, Division 1 hazardous locations
Instruction Manual
748221-M
July 2002
100%, are available. Single range is standard; switch-selectable
dual or triple range is optional.
117°F (47°C).
loads
Corrosion-resistant filaments available on order
(ANSI/NFPA 70)
1
Zero and Span drift specifications based on ambient temperature shifts of less than 18 Fahrenheit degrees (10 Celsius de-
grees) at a maximum rate of 18 Fahrenheit degrees (11 Celsius degrees) per hour.
2
Cell response time is less than 45 seconds for 95% response, with sample flow rate of 250 cc/min, for the following gas
combinations: Argon and air, nitrogen, or oxygen; carbon dioxide and argon, nitrogen, or oxygen; helium and methane; hydrogen and methane.
Rosemount Analytical Inc. A Division of Emerson Process ManagementDescription and Specifications 1-5
Instruction Manual
748221-M
July 2002
Model 7D
1-6 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
Instruction Manual
748221-M
July 2002
SECTION 2
INSTALLATION
2-1 FACILITY PREPARATION
Sections 2-1a through 2-1d provide
information that may be required prior to
installation.
a. Installation Drawings
For outline and mounting dimensions of
the analyzer and gas selector panel
modules, refer to drawing 654625 and
Figure 2-1 on page 2-2 or Figure 2-2 on
page 2-3.
b. Customer Electrical Connections
Customer electrical connections are
shown in Figure 2-5 on page 2-6.
c. Flow Diagrams
For gas connections, refer to appropriate
flow diagram:
•Analyzer using sealed reference gas,
Figure 2-3 on page 2-4
•Analyzer using flowing reference gas,
Figure 2-4 on page 2-5.
d. Location and Mounting
Location
Proper location for the analyzer depends
on two basic considerations:
• Accessibility to the sampling point
• Protection of the instrument
Ideally, the analyzer should be located as
close to the sampling point as possible.
Short sample lines reduce time lag in
readings. In practice, however, protection
of the instrument sometimes calls for
more remote placement.
The analyzer should be mounted in a
clean, dry atmosphere. Ambient
temperature should be within the range of
32oF to 100 F (0oC to 38oC).
Mounting
The analyzer is designed for surface
mounting, utilizing the hardware provided.
Refer to drawing 654625.
2-2 UNPACKING
The Model 7D Thermal Conductivity Analyzer
is a precision instrument and should be
handled carefully. Carefully examine the
shipping carton and contents for signs of
damage. Immediately notify the carrier if the
carton or its contents are damaged. Retain
the carton and packing material until the
instrument is operational.
2-3 GAS REQUIREMENTS
a. General
The Model 7D requires cylinder gases
appropriate to the particular application
(refer to the Data Sheet inserted in the
rear of this manual). Suitable gases are
available from various suppliers.
b. Calibration Gas Requirements
For calibration, the analyzer requires a
downscale and an upscale calibration
gas, both normally specified in the Data
Sheet in the rear of this manual. Proper
choice of calibration gases for a particular
application depends on the composition of
the sample stream and the operating
range used.
c. Sample Gas Composition
In a typical application, the sample gas
consists of two components, for example:
hydrogen in nitrogen. In this example,
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Instruction Manual
10.25
8.25
Mounting Holes (4)
748221-M
July 2002
Model 7D
hydrogen is designated the “measured
component” and nitrogen constitutes the
“background gas.”
d. Suppressed-Zero Ranges
With any zero-suppressed range, the
zero-concentration point for the measured
component lies offscale, below the lower
range-limit. A typical example is 80% to
100% hydrogen in nitrogen. Here the
appropriate upscale calibration gas would
be pure hydrogen. The downscale gas
would have a composition appropriate to
establishing a calibration point slightly
above the lower range-limit e.g., 81%
hydrogen in nitrogen.
WARNING
POSSIBLE EXPLOSION HAZARD
This analyzer is of a type capable of analysis of sample gases which may be flammable. If used for analysis of such gases,
the instruments explosion-proof enclosure
must be suitable for the gas.
If explosive gases are introduced into this
analyzer, the sample containment system
must be carefully leak-checked upon installation and before initial startup, during
routine maintenance and any time the integrity of the sample containment system
is broken, to ensure the system is in leakproof condition.
Internal leaks resulting from failure to observe these precautions could result in an
explosion causing death, personal injury
or property damage.
Span Gas In
To AnalyzerZero Gas In
Sample In
[210]
NO. 8 Flat Head Screw
[260]
ZEROSPAN
SAMPLE
Figure 2-1. Gas Selector Panel for Thermal Conductivity Cell with Sealed-In Reference Gas
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
Upscale
To Analyzer
Downscale
1.25
To Analyzer
3.25
.12
Mounting Holes (4)
10.25
8.25
Instruction Manual
748221-M
July 2002
Sample Gas
Sample Inlet
Ref Gas
Ref Gas Inlet
Calibration Gas
Calibration Gas
[83]
[3]
[32]
[210]
NO. 8 Flat Head Screw
DOWNSCALE
CALIBRATION GAS
UPSCALE
CALIBRATION GAS
[260]
REF SAMPLE
Figure 2-2. Gas Selector Panel for Thermal Conductivity Cell Using Flowing Re ference Gas
SAMPLE
REF
2-4 LEAK CHECK
Pressurize the system with air or inert gas
such as nitrogen, making sure not to exceed
specified pressure limitation.
Liberally cover all fittings, seals and other
possible sources of leakage with leak test
liquid such as SNOOP (PN 837801).
Bubbling or foaming indicates leakage, which
MUST be corrected before introduction of
flammable-sample and/or application of
electrical power.
2-5 GAS CONNECTIONS
The analyzer and gas selector panel modules
must be interconnected according to the flow
diagram specified in the Data Sheet at the
rear of this manual. Gas fittings on both
modules are tagged as to use. Fittings are
1/4-inch NPT for 1/4-inch (6.3 mm) tubing.
For interconnection, use 1/4-inch (6.3 mm)
copper or stainless steel tubing, depending on
whether the sample stream is corrosive.
Refer to Figure 2-3 on page 2-4, or Figure 2-4
on page 2-5 and drawing 654625. Connect
sample/calibration gas lines to fittings tagged
“INLET” on bottom of analyzer. Connect
appropriate vent line to fitting labeled
“OUTLET.”
Follow similar procedure for reference gas, if
any.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
Sample In
Needle
Flowmeter
To Vent
Thermal Conductivity Cell
Reference Gas
748221-M
July 2002
Model 7D
Model 7D
Sealed-In
Zero
Standard
Gas
Span
Standard
Gas
Valves
113357 or 113920
Gas Selector Panel
Thermal Conductivity Analyzer
Figure 2-3. Connection of Analyzer Using Sealed-In Reference Gas to Associated Gas
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
Upscale
Sample Gas
Model 7D
To
Thermal Conductivity Cell
117195 or 118210
Reference
Needle
Sample
To
Downscale
Gas
Upscale
Sample Gas
Reference
Model 7D
To
Thermal Conductivity Cell
117195 or 118210
Downscale
Used as
Needle
Sample
To
A. DIFFERENT REFERENCE GAS AND CALIBRATION GAS
Instruction Manual
748221-M
July 2002
Gas
Downscale
Calibration
Gas
Calibration
Gas
Valves
Gas Selector Panel
Reference
Gas
Flowmeter
Gas
Flowmeter
B. REFERENCE GAS ALSO USED AS CALIBRATION GAS
Thermal Conductivity
Analyzer
Sample
Vent
Reference
Vent
Calibration Gas
Calibration
Calibration
Gas
Valves
Gas Selector Panel
Gas
Flowmeter
Gas
Flowmeter
Thermal Conductivity
Analyzer
Sample
Vent
Reference
Vent
Figure 2-4. Connection of Analyzer Using Flowing Reference Gas to Associated Gas Selector Panel
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
10
862209
11
748221-M
July 2002
OPTIONAL DUAL ALARMS
13
14
15
17
19
TB10
12
16
18
DETECTOR ASSEMBLY
Model 7D
Heater
7
5
4
3
1
TEMPERATURE CONTROL
ASSEMBLY
Fan
RECORDER VOLTAGE
OUTPUT AND CURRENT
OUTPUT
TB4
4
(-) CURRENT
3
(+) CURRENT
2
(-) VOLTAGE
1
(+) VOLTAGE
AC POWER INPUT
Fuse
TB6
1 2 3
L1 L2 GND
DC POWER SUPPLIES
Figure 2-5. Electrical Connections
MASTER
BOARD
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
Instruction Manual
748221-M
July 2002
CAUTION
Do not plug or restrict vents.
2-6 RECORDER OUTPUT SELECTION AND CA-
BLE CONNECTIONS
If a recorder, controller, or other output device
is used, connect it to the analyzer (refer to
drawing 654642) via a number 22 or number
24 AWG two-conductor shielded cable. Route
the cable through conduit to the analyzer, and
into the case through the appropriate opening
shown in drawing 654625.
NOTE
Route recorder cable through a separate
conduit, not with power cable.
Output selection and cable connections for
voltage-actuated and current -actuated
devices are explained in Sections 2-6a below
through 2-6e on page 2-10.
a. Standard (Non-linearized) Voltage
Output
1. On the Master Board (Figure 2-6 on
page 2-8):
a. Verify that TB2-1 is jumpered to
TB2-2.
b. Set S1 for desired voltage: 5V,
1V, .1V, or .01V.
b. Linearized Voltage Output (Optional)
1. Verify that Voltage-Output Linearizer
Board (PN 633756) is properly
inserted in J102.
2. On the Master Board (Figure 2-6 on
page 2-8):
a. Verify that TB2-1 is jumpered to
TB2-4, and TB2-2 is jumped to
TB2-5.
b. Set S1 for desired voltage: 5V,
1V, .1 V, or .01 V.
c. Connect the recorder cable to
the TB4 terminals labeled VOLT
OUT: TB4-1 (+) and TB4-2 (-).
NOTE
Take the usual precautions to avoid
AC pickup. DO NOT GROUND EITHER
LEAD.
3. Connect the cable to the recorder
input terminals; ensure that polarity
is correct.
c. Isolated 4 to 20 mA Current Output
(Optional)
1. Verify that the Isolated 4 to 20 mA
Current Output Board (PN 652816) is
properly inserted in J103.
c. Connect the recorder cable to
the TB4 terminals labeled VOLT
OUT: TB4-1(+) and TB4-2(-).
NOTE
Take the usual precautions to avoid AC
pickup. DO NOT GROUND EITHER LEAD.
2. Connect the cable to input terminals
of the recorder; ensure that polarity
is correct.
3. Ground shield on one end only.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
2. On the Master Board (Figure 2-6 on
page 2-8):
a. Verify that TB2-1 is jumpered to
TB2-2 and TB2-2 is jumpered to
TB2-6.
b. Connect the recorder cable to
the TB4 terminals labeled CUR
OUT: TB4-3 (+) and TB4-4 (-).
3. Connect the cable to the recorder
input terminals; ensure that polarity
is correct. Total resistance of the
output device and associated cable
must not exceed 1500 ohms.
Instruction Manual
FAN
115V 230V
S2
U1
S1
TB2
6 5 4
3 2 1
BOARD
J101
J100
J103
J102
TB3
TB5
TP1
TP2
TB1
6 5 4
3 2 1
OUT
OUT
AC
HEATER
C2
R12
C4
R11
R15
AR1
R27R1R5C3C1
R3R2C5C6R23
AR3C7 R19
R21 R18
R17
R20J7J211
748221-M
July 2002
Model 7D
POWER
POWER
CUR
J4
- + - +
VOLT
R4 R9 R7 R22 R26 R6 R8
TB4
F13
1
R10
R14
J6
1
AR2
R24
R16
OPTIONS
RESISTOR
SELECT
RANGE
SPAN
METER
1
2
3
4
1
2
3
4
5
6
7
8
11
TP3 TP4 TP5
1
1
DETECTOR
R100
S1Used to select voltage output range: 5V, 1V, 0.1V, or .01V
AR1 gain adjust. Permits adjustment of AR1 gain from X1 to
R4
X100, to establish the sensitivity desired for Range 1. This is
the highest sensitivity range.
AR1 zero adjust. Used to eliminate voltage offset within AR1
and Bridge, and provide zero suppression.
Setting determines attenuation factor applicable to AR2 output, Range 3.
Setting determines attenuation factor applicable to AR2 output Range 2.
Permits adjusting meter fullscale to agree with recorder fullscale
.01V .1V 1V 5V
DETAIL OF S1
R6, R8, R26
R9
R7
R22
R15Used to eliminate voltage offset within AR2.
R16
Sets the stable ±10 V source.
Figure 2-6. Master Board
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
Deadband Potentiometer
Deadband Potentiometer
PIN
OUTPUT
PIN
OUTPUT
Instruction Manual
748221-M
July 2002
d. Dual Alarms (Optional)
The alarm module is installed in the
instrument as shown in Figure 1-2 on
page 1-3 and electrical connections are
shown in drawing 654642. The analyzer
is factory configured and requires no user
adjustments.
Pins for AC power (1, 2, and 3), alarm setpoint
control (8, 10, 18, and 19), and signal from
analyzer circuitry (7 and 9) are wired at the
factory. Connections of the remaining pins
depend on individual application
requirements.
If the instrument has this option, the analyzer
will have the following setpoint adjustment
potentiometers: Alarm 1 (the Low Alarm) and
Alarm 2 (the High Alarm) on the front panel
FRONT VIEW OF ALARM MODULE
( Figure 1-1 on page 1-2) and Deadband on
the alarm module (see Figure 2-7 below).
The power and input signals have been
wired to the terminal strip into which the
alarm module plugs. The wiring to the
contacts (18 or 20 AWG) is routed
through a conduit hole (see Drawing
654625).
The alarm relay is in energized mode
when power is applied. Wire output to the
appropriate contact (see Figure 2-5 on
page 2-6 and Table 2-1 below). The
Form C relay contacts are rated at 5 A,
120 VDC and 1 A, 120 VAC, resistive
loads.
A lit LED next to the Deadband pots
indicates the alarm is activated.
Contact Rating: 1.0 A, 120 VAC; 5.0 A, 120 VAC. Form C, resistive loads.
Table 2-1.Alarm Output Connections
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-9
Instruction Manual
748221-M
July 2002
Model 7D
NOTE:
The Zero and Span for setting the input
voltage from the analyzer has been set at
the factory. To check it, see Section 5-2e
on page 5-3.
The following is recommended:
1. A fuse should be inserted into the line
between the customer-supplied power
supply and the alarm module
terminals on the Alarm Relay
Assembly.
2. If the alarm contacts are connected to
any device that produces radio
frequency interference (RFI), the
device should be arc -suppressed
(P/N 858728 Arc Suppressor is
recommended).
3. The analyzer and any RFI-producing
device should operate on different AC
power sources to avoid RFI.
Removal of AC power from the analyzer,
as in a power failure, de-energizes both
alarm module relays, setting an alarm
condition. Switching characteristics of the
ALARM 1 and ALARM 2 relays are as
follows:
corresponds to setpoint minus deadband
(see Figure 2-8 on page 2-11 ).
Fail-safe Applications
By making the appropriate connections to
the double-throw relay contacts, the
operator can obtain either 1) a contact
closure or contact opening for an
energized relay, or 2) a contact closure or
contact opening for a de-energized relay.
For fail-safe applications, the operator
must understand which circuit conditions
are required to achieve relay
de-energization in the event of power
failure.
e. Linearized Voltage, Two Ranges
(Optional)
1. Verify that the Voltage Output
Linearizer Board (PN 633756) is
properly inserted in both J102 and
J103.
2. Verify that the range switch on the
door is properly connected: Position
(Range) 1 to TB2-5, Position 2 to
TB2-3, Wiper to TB2-2.
3. On the Master Board (Figure 2-6 on
page 2-8):
Alarm 1 Relay
The ALARM 1 relay coil is de-energized
when the meter needle moves downscale
through the value that corresponds to
setpoint minus deadband. This relay coil
is energized when the needle moves
upscale through the value that
corresponds to setpoint plus deadband
(see Figure 2-8 on page 2-11 ).
Alarm 2 Relay
The ALARM 2 relay coil is de-energized
when the meter needle moves upscale
through the value that corresponds to the
setpoint plus deadband. This relay coil is
energized when the needle moves
downscale through the value that
2-10 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Take the usual precautions to avoid AC
pickup. DO NOT GROUND EITHER LEAD.
a. Verify that TB2-1 is jumpered to
TB2-4, and TB2-4 is jumped to
TB2-6.
b. Select desired range on S1.
c. Connect recorder cable to the
TB4 terminals labeled VOLT
OUT: TB4-1 (+) and TB4-2 (-).
NOTE
4. Connect the cable to the recorder
input terminals; ensure that polarity
is correct.
Model 7D
4030205550
45
A. Typical ALARM 1 Setting (LOW)
DEADBAND SET FOR
20% OF FULLSCALE
B. Typical ALARM 2 Setting (HIGH)
DEADBAND SET FOR
10% OF FULLSCALE
INPUT SIGNAL
Percent of Fullscale
INPUT SIGNAL
Percent of Fullscale
Instruction Manual
748221-M
July 2002
When input signal moves upscale through this point, the
coil of ALARM 1 relay (K1) is energized, providing
continuity between the common and normally-closed
contacts of the relay.
ALARM 1 Setpoint
When input signal moves downscale through this point, the
coil of ALARM 1 relay (K1) is de-energized, providing
continuity between the common and normally-open
contacts of the relay.
When input signal moves upscale through this point, the
coil of ALARM 2 relay (K2) is de-energized, providing
continuity between the common and normally-open
contacts of the relay.
ALARM 2 Setpoint
When input signal moves upscale through this point, the
coil of ALARM 2 relay (K2) is energized, providing
continuity between the common and normally-closed
contacts of the relay.
Figure 2-8. Typical Alarm Settings
f. Linearized Voltage and Isolated 4 to 20
mA Current Output (Optional)
1. Verify that the Voltage Output
Linearizer Board (PN 633756) is
properly inserted in J102.
2. Verify that the Isolated 4 to 20 mA
Current Output Board (PN 652816) is
properly inserted in J103.
3. On the Master Board (Figure 2-6 on
page 2-8):
a. Verify that TB2-1 is jumped to
TB2-4, TB2-2 is jumped to
TB2-5, and TB2-5 is jumped to
TB2-6.
b. Set S1 for the desired voltage: 5
V, 1 V, .1 V, or .01 V.
c. Connect the recorder cable to
the TB4 terminals labeled VOLT
OUT: TB4-1 (+) and TB4-2 (-).
NOTE
Take the usual precautions to avoid AC
pickup. DO NOT GROUND EITHER LEAD.
4. Connect the cable to recorder
VOLTAGE input terminals; ensure
that polarity is correct.
5. On the Master Board (Figure 2-6 on
page 2-8):
a. Connect the recorder cable to
the TB4 terminals labeled CUR
OUT: TB4-3 (+) and TB4-4 (-).
b. Connect the cable to the
recorder CURRENT input
terminals; ensure that polarity is
correct.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-11
Instruction Manual
SENSOR
TEST
HEATER
J17
115
TEMP CONTROL BD
115
748221-M
July 2002
Model 7D
2-7 ELECTRICAL POWER CONNECTIONS
WARNING
ELECTRICAL SHOCK HAZARD
For safety and proper performance, this
instrument must be connected to a properly grounded three-wire source of power.
Before supplying electrical power to analyzer, disconnect the red lead from the
bridge to TB1-1 or TB1-2 (depending on
bridge polarity). This action disconnects
power to the bridge (see drawing 654642).
To safeguard against filament damage,
this lead should remain disconnected until
proper gas flow has been established.
The analyzer is supplied, as ordered, for
operation on 107 to 127 or 214 to 254 VAC,
50/60 Hz, 250 watts. Verify that the power
source conforms to the requirements of the
individual instrument as noted on the name
rating plate. Ensure that switches S2 on the
Master Board (Figure 2-6 on page 2-8) and
S3 on the case heater temperature control
board (see Figure 2-9 below) are set to
required voltage.
Electrical power is supplied to the analyzer via
a customer-supplied three-conductor cable,
type SJT, minimum wire size 18 AWG. Route
the power cable through conduit and into the
appropriate opening in the instrument case
(see drawing 654625). On TB6, Figure 2-5 on
page 2-6, connect cable leads to terminals 1
(HOT/L1), 2 (NEUT/L2), and 3 (GND).
Set switch window for voltage required
S3
400A 880 951E
C2R10 R11 R7 R8
CR1
C1
C
E
B
Q2
R17R16 R12 CR
R4
R3
J18
1
AR1
R13
R2R1
Q1
K
G
A
POWER
SUPPLY
R6
C3
R9 R5
CR
1
E
B
Q3
C
J19
J11
R15 R14
POWER
LINE
J5
C4 3 2 1
1
1 21 2 3
T.I.F.
U2
2
3
1
U1
Figure 2-9. Case Heater Temperature Control Board
S3
2-12 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
Instruction Manual
748221-M
July 2002
SECTION 3
STARTUP AND OPERATION
3-1 ANALYZER CONTROLS AND ADJUS T-
MENTS
Normal operation of the analyzer involves
adjustments of only the front-panel controls:
ZERO and SPAN, and RANGE Switch, if
provided. See Figure 1-1 on page 1-2.
The various internal adjustments are factory
set and normally do not require readjustment
except after replacement of a circuit board or
major component. Refer to Section 5,
Service and Maintenance
3-2 GAS SELECTOR PANEL CONTROLS
The controls provided on the optional gas
selector panel will depend on the application.
For use of sealed-in reference gas, refer to
Figure 2-1 on page 2-2 and Figure 2-3 on
page 2-4. For flowing reference gas, refer to
Figure 2-2 on page 2-3 and Figure 2-4 on
page 2-5.
3-3 STARTUP PROCEDURE
WARNING
POSSIBLE EXPLOSION HAZARD
If explosive gases are introduced into this
analyzer, the sample containment system
must be carefully leak-checked upon installation and before initial startup, during
routine maintenance and any time the integrity of the sample containment system
is broken, to ensure the system is in leakproof condition. Leak-check instructions
are provided in Section 2-4 on page 2-3.
Internal leaks resulting from failure to observe these precautions could result in an
explosion causing death, personal injury
or property damage.
WARNING
FILAMENT DAMAGE
Never apply power to analyzer without gas
flowing. The filaments in the cell tend to
deteriorate faster than normal.
NOTE:
Before supplying electrical power to the
analyzer, disconnect the red lead from the
bridge to TB1-1 or TB1-2 on the Master
Board (depending on detector bridge polarity). This action disconnects power to
the bridge. To safeguard against filament
damage, this lead should remain disco nnected until proper gas flow has been es tablished. Refer to drawing 654642.
After performing a leak check, start up the
analyzer as follows:
1. Remove the connector from J2 on the
Master Board (Figure 2-6 on page 2-8).
Verify that this line goes to TB6, located
at the lower left of the instrument (see
Figure 1-2 on page 1-3 ). If it does not,
make corrections using drawing 654642
as a reference.
2. Set regulators on the gas cylinders for a
supply pressure of 10 to 50 psig f(69 to
345 kPa).
3. Provide a sample flow of 50 to 350
cc/minute through the analyzer. A rate
of 250 cc/minute is recommended unless
faster flow is desired to reduce sample
transport time.
4. If the thermal conductivity cell uses
flowing reference gas, provide a
reference flow of 5 to 50 cc/minute.
Rosemount Analytical Inc. A Division of Emerson Process ManagementStartup and Operation 3-1
Instruction Manual
748221-M
July 2002
Model 7D
5. Refer to drawing 654642 and Figure 2-6
on page 2-8. Connect the cable from
TB6 to J2 on the Master Board to provide
AC power to the Master Board.
Reconnect the red lead of TB1 (or TB2)
to provide power to the detector. Apply
power to the analyzer. The filaments will
now begin to heat. Verify proper flow of
sample gas and flowing reference gas, if
used.
6. Allow the analyzer to warm-up for a
minimum of six hours to ensure
temperature equilibrium.
NOTE:
If ambient temperature is below 60°F
(15.6°C), or if the temperature will go below
60°F (15.6°C) within the next six hours, al-
low a minimum of 12 hours for the instrument to stabilize.
Startup is now complete; the analyzer is ready
for calibration per Section 3-4 below.
3-4 CALIBRATION
1. Set downscale calibration point as
follows:
a. On the analyzer, set front-panel
SPAN control to midscale (five turns).
If the analyzer has front-panel
RANGE Switch, set it at Range 1, the
highest sensitivity range.
b. Admit downscale calibration gas to
the analyzer at the same flow rate as
is used for sample gas. Wait for the
reading on the meter or recorder to
stabilize.
c. On the analyzer, adjust the front
panel ZERO control so that the
reading on the front panel meter or
recorder is appropriate to the
downscale calibration gas.
2. Set upscale calibration point as follows:
a. If the analyzer has a front panel
RANGE switch, set it for the desired
range.
b. Admit upscale calibration gas to the
analyzer at the same flow rate as is
used for sample gas. Wait for the
reading on the meter or recorder to
stabilize.
c. On the analyzer, adjust the front
panel SPAN control so that the
reading on the front panel meter or
recorder is appropriate to the upscale
calibration gas.
If a proper reading is unobtainable by
adjustment of the SPAN control, refer to
Section 5, Service and Maintenance.
After the downscale and upscale calibration
points have been established, the analyzer is
ready for routine operation per Section 3-5
below.
If the analyzer has Suppressed-Zero Ranges
and does not calibrate properly, refer to
Section 5-3 on page 5-3 for Suppressed Zero
Adjustment.
3-5 ROUTINE OPERATION
First, complete startup per Section 3. If the
analyzer has more than one range, turn front
panel RANGE Switch to desired position.
Admit sample gas at the previously selected
flow rate. The analyzer will continuously
indicate the concentration of the measured
component in the sample stream.
A calibration curve can be used to convert
meter or recorder readings to concentration
values. Typical calibration curves are
supplied for standard ranges. Calibration
curves for special ranges are available as
options.
d. If a proper reading is unobtainable by
adjustment of the ZERO control, refer
to Section 5-2b on page 5-2.
3-2 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
To avoid use of a calibration curve in an
application where it would otherwise be
required, the analyzer may be equipped with
an optional linearizer board. If so, the
Model 7D
linearizer board is factory set for a given
range only, and is not usable on another
range. Note that a linearizer is usable only if
nonlinearity at midscale does not exceed 20%
of fullscale.
3-6 RECOMMENDED CALIBRATION FRE-
QUENCY
Provided that the instrument remains in
continuous operation with power on, it is
necessary only to calibrate once a week, by
the procedure of Section 3-4 on page 3-2.
To restart the analyzer after power turn-off,
repeat the startup procedure of Section 3-3 on
page 3-1 and calibrate per Section 3-4 on
page 3-2.
3-7 SHUTDOWN
Instruction Manual
748221-M
July 2002
Before turning off sample gas, and flowing
reference gas, if used, disconnect power from
the analyzer. This precaution minimizes the
risk of filament damage.
Rosemount Analytical Inc. A Division of Emerson Process ManagementStartup and Operation 3-3
Instruction Manual
748221-M
July 2002
Model 7D
3-4 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
Instruction Manual
748221-M
July 2002
SECTION 4
THEORY
4-1 THERMAL CONDUCTIVITY CELL AND AS-
SOCIATED BRIDGE ADJUSTMENTS
Within the thermal conductivity cell are four
resistive filaments suspended in individual
cavities of a metal block (Figure 4-1A on page
4-3) and connected electrically as legs of a
Wheatstone bridge (Figure 4-1B on page 4-3).
Although physically the cell block is one piece,
functionally it may be considered to have two
sides as shown:
Sample Side
Two filaments that constitute opposite legs of
the bridge are positioned in a passage that
receives a continuous flow of the sample gas.
Reference Side
The remaining two filaments are positioned in
a passage filled with the reference gas.
Depending on the application, the reference
gas may flow continuously through the
passage, or it may be sealed within the cell.
resultant change in filament resistance
unbalances the bridge.
The bridge-imbalance signal is routed to the
Master Board (PN 654620), where it is
processed to drive the front -panel meter and
recording device, if used (see Section 4-2a
below).
Periodically, downscale calibration gas is
passed through the cell, and the front panel
ZERO Pot is adjusted for an appropriate
reading on the m eter or recorder.
4-2 ELECTRONIC CIRCUITRY
Electronic circuitry of the Model 7D is shown
in the schematic diagram of drawing 654616.
Internal circuitry of plug-in boards and other
electronic assemblies is shown in separate
schematic diagrams, found at the end of this
manual, and is described in the following
sections.
a. Master Board
The Bridge Voltage Power Supply (PN
613560) is connected, via a 20-ohm dropping
resistor, to the bridge (See Section 4-2d on
page 4-4). The power supply output is
adjusted to provide an appropriate voltage
across bridge terminals 1(+) and 2(-). An
electric current flows through the filaments,
heating them and thus increasing their
electrical resistance. The heat -dissipation
rate for each filament depends on the thermal
conductivity of the surrounding gas. Initially,
with downscale calibration gas flowing
through the sample and reference sides of the
flow-through configuration, R26 is set for zero
bridge-output signal. During subsequent
analysis of the sample stream, any change in
the relative proportions of the components
passing through the sample side changes the
thermal conductivity of the gas mixture,
causing a temperature differential between
sample and reference filaments. The
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-1
The Master Board (PN 654620), Figure
2-6 on page 2-8, provides two stages of
amplification utilizing integrated-circuit
amplifiers AR1 and AR2.
Functions Associated with AR1
AR1
Gain Adjust Potentiometer R4. This
screwdriver-adjustable, factory -set
trimming potentiometer determines
feedback resistance for AR1 and thus
permits adjustment of AR1 gain from X1
to X100. This adjustment sets the
sensitivity for Range 1, i.e., the
highest-sensitivity range.
Range Provisions
In the basic single-range Model 7D, a
jumper is connected from TB3-1 to TB3-4,
thus routing the unattenuated output from
AR1 directly to the non-inverting input of
Instruction Manual
POSITION
748221-M
July 2002
Model 7D
AR2. During factory assembly of a
dual-range or triple-range instrument, the
jumper is omitted and a front panel
RANGE Switch is connected as shown
below. The output from AR1 is then
routed to AR2 through a network that
provides adjustable attenuation for ranges
2 and 3.
RANGE
Table 4-1.Range Switch Connections
AR1 OUTPUT
ATTENUATION
1X1, fixed11
2adjustable by R722
3adjustable by R933
Coarse zero and zero-suppression
Pots R26, R6 or R8, depending upon
range number, provide the zero correction
at the non-inverting input of AR1.
Functions Associated with AR2
AR2 Zero Adjustment Potentiometer
R1
This screwdriver-adjustable, factory -set
trimming potentiometer is used to
eliminate voltage offset within AR2.
When the input signal is zero, R15 is
adjusted so that the output signal also is
zero.
RANGE
SWITCH
wiper4
TB3
POSITION
Front-Panel M eter
The meter is connected from TB3-7 to
TB3-8. Potentiometer R22 permits
adjusting meter sensitivity so that meter
fullscale agrees with recorder fullscale.
Output Selection Switch S1
The desired output is obtained by
appropriate selection of switch contacts:
5V, 1V, .1V, or .01V.
b. Voltage Output Linearizer Board
(Optional)
The output signal from the Master Board
is a function of the degree of imbalance in
the bridge circuit, but is not linear with
respect to the concentration of the
measured component. Providing that the
nonlinearity of the calibration curve does
not exceed 20% at midscale (for 50 % H
max.), the Voltage Output Linearizer
Board (PN 633756) may be used to equip
a given operating range for linear readout
of concentration on the meter and on a
potentiometric recorder.
Straightening of the concentration vs
output curve is accomplished by
sequential adjustment of eight odd
numbered trimming potentiometers
designated R19 through R33. Each pot
controls the gain of an associated
operational amplifier.
2
Front-Panel SPAN Control
This potentiometer, connected across
TB3-5 and TB3-6, provides continuously
variable adjustment of closed-loop gain
for AR2, to permit establishing an upscale
calibration point on the meter scale or
recorder chart. With upscale calibration
gas flowing through the analyzer, the
SPAN Control is adjusted for the
appropriate reading.
4-2 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
During factory checkout of a linearizer
circuit board, potentiometers R19 through
R33 are initially set at midrange. With
zero input signal applied to the linearizer
circuit, ZERO potentiometer R35 is
adjusted for zero output. Then an
appropriate low-level signal is applied to
the input, and R19 is adjusted for fullscale
output. The procedure is repeated as
many times as required to obtain properly
linearized output.
NOTE: Cell Block Sectioned
Through Sample Side.
Section Through
Reference side is
Similar.
Reference Flow
Resistor values are in ohms.
Sample Flow
Bridge Power
Supply (+)
Bridge Power
Supply (-)
Figure 4-1. Thermal Conductivity Cell
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-3
Instruction Manual
748221-M
July 2002
Model 7D
c. Isolated 4 to 20 mA Current Output
Board (Optional)
This option provides isolated current
output for applications which require 4 to
20 mA into a maximum load of 1500
ohms. This output is NOT linearized. A
description of the board and the functions
of its main components follows. Refer to
schematic 652813 at the rear of this
manual.
The purpose of the Isolated Current
Output Board (PN 625816) is to convert
an input signal of 0 to -5 V to an isolated
output signal of 4 to 20 mA. With a zero
voltage at the input, R6 is adjusted so that
the output at AR1 is 1 V. The gain of AR1
is .8 so that an input of 0 to -5 V is
converted to 1 to 5 V. This voltage is fed
to a variable output level-inverter
consisting of AR2, Q1, Q2, and T1. T1
has two identical output windings, each
with a rectified DC output. One is used to
provide feedback to the inverter input of
AR2. The other is used as an isolated
output to drive AR3 and Q3 with a 1 to 5
V signal across R16. Trim-pot R17 is
adjusted to provide a span of 4 to 20 mA.
This current is presented to the collector
of Q3 and is capable of driving loads of up
to 1500 ohms.
e. ±15 Volt Power Supply
The ±15 Volt Power Supply (PN 619714)
plugged into J101 of the Master Board,
provides power for the various circuits.
As shown in drawing 619710, power
transformer T1 has three secondary that
are used as follows:
38 VAC center-tapped secondary
Powers both 15 volt supplies through
diode bridge CR1 and filter capacitors C1
and C4.
The adjustable positive regulator, VR1, is
set by voltage divider R1, R2 and R3 and
its output is applied to pin A of the circuit
board and to test point TP1.
Potentiometer R2 should be adjusted to
+15.5 VDC ±50 mVDC.
The negative DC, regulated by VR2 is
applied to pin D of the circuit board.
The center tap is the common reference
for both the +15 and -15 volt supplies and
is applied to pin R of the circuit board and
to test point TP2.
Both outputs are used for individual
amplifiers on the various circuit boards.
d. Bridge Power Supply
The regulated, adjustable voltage
required for the thermal conductivity
bridge (Section 4-1 on page 4-1) is
provided by the Bridge Power Supply (PN
613560). It consists of a power
transformer, fullwave rectifiers CR1 and
CR2, voltage regulator Q1, and an RC
filter network. Bridge voltage, as
measured between bridge terminals 1(+)
and 2(-), is adjustable from 5 to 13 VDC
via R2. Proper setting depends primarily
on filament material: 3 to 4 VDC for
tungsten; 5 to 12 VDC for Hitempco.
Bridge voltage is factory -set as required
for the application (see Data Sheet in the
rear of this manual) and normally does
not require readjustment unless the power
supply is replaced.
4-4 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
90-volt center-tapped secondary
Drives a rectifier circuit on the optional
Current Output Board (PN 652816). The
transformer winding and the associated
circuit constitute a floating power supply
for the emitter-follower stage. Refer to
Section 4-2c above.
9.5 VAC secondary
Drives a +5 VDC supply not used in this
instrument.
f. Detector Blocks
There are two types of detector
configurations; flowing reference and
sealed nitrogen reference. The sealed
reference with nitrogen can be recharged
by flowing nitrogen through the reference
Model 7D
Instruction Manual
748221-M
July 2002
and closing the plugs. Three filament
types are available: tungsten, Hitempco,
and gold-sheathed tungsten.
g. Case Temperature Controller
Assembly
The case Temperature Controller
Assembly (PN 652270) maintains an
approximate 117 °F (47°C) temperature
within the instrument.
Changes in case temperature affect the
resistance of the sensor connected to
J18. This in turn changes the bias of
amplifier AR1, which controls the input to
the case heater via switch U2. The
heater receives power through pins 3, 4
and/or 5, depending on the voltage
selected at switch S3. A thermal fuse at
J5 prevents the case from overheating.
h. Dual Alarms (Option)
Instruments with optional Dual Alarms
have two pots on the front door which are
used to select setpoints. The scales on
the nameplate designate setpoints from 0
to 100%.
The alarm module is shown in Figure 6-2
on page 6-5, and the factory connections
are shown in drawing 654642. The
module plugs into a socket, which is part
of a terminal block mounted onto the
chassis of the instrument. A marker on
the alarm module indicates the function of
each socket pin.
This unit is factory configured and
requires no user adjustment. Pins for AC
power (1, 2, and 3), alarm setpoint control
(8, 10, 18, and 19), and signal from
analyzer circuitry (7 and 9) are wired at
the factory.
Connections of the remaining pins
depend upon the application. The relays
are rated at 1.0 A, 120 VAC and 5.0 A,
120 VDC, form C, resistive loads.
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-5
Instruction Manual
748221-M
July 2002
Model 7D
4-6 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
Instruction Manual
748221-M
July 2002
SECTION 5
SERVICE AND MAINTENANCE
WARNING
POSSIBLE EXPLOSION HAZARD
If explosive gases are introduced into this
analyzer, the sample containment system
must be carefully leak-checked upon installation and before initial startup, during
routine maintenance and any time the integrity of the sample containment system
is broken, to ensure the system is in leakproof condition. Leak-check instructions
are provided in Section 2-4 on page 2-3.
Internal leaks resulting from failure to observe these precautions could result in an
explosion causing death, personal injury
or property damage.
In troubleshooting, the basic approach is to
isolate the analyzer from the sample and the
sample-handling system.
First admit downscale and upscale standard
gases to analyzer and note response:
1. If performance is normal with standard
gases, although not with sample gas, the
sample and the sample-handling system
are suspect. Check these areas.
2. If analyzer gives offscale or erratic
readings with standard gases, as well as
with sample gas, the problem might be
the filaments or the electronic circuitry.
To isolate the malfunction, substitute
fixed precision resistors of appropriate
value for the filaments. Hitempco
filaments (P/N 25499) have a cold
resistance of 72 ohms. Tungsten
filaments (P/N 811993) have a cold
resistance of 18 ohms. Filament
connections are shown in drawing
654616. With appropriate resistors
substituted for the filaments, attempt to
balance the bridge as follows:
a. If bridge balance is obtained with the
fixed resistors, the filaments are
probably defective and should be
replaced.
b. If bridge balance is unobtainable with
the fixed resistors, the problem might be
the electronic circuitry. Substitute each
circuit board, in turn, until proper
operation is obtained.
5-1 THERMAL CONDUCTIVITY CELL
The thermal conductivity cell and associated
elements of the bridge are mounted inside the
analyzer case, within a thermally-insulated
compartment.
Depending on the bridge current level, the
filaments gradually become mismatched, and
eventually burn out. Normal progression of
symptoms is baseline drift. Sealed reference
cells must be serviced by the factory.
If the cell uses flowing reference gas, the
filaments may be replaced by the user. Refer
to Figure 4-1 on page 4-3 and the parts list in
Section 6.
Filaments are sold as matched sets. If one
must be replaced, its mate also must be
replaced. Electrically, the filaments are mated
across the diagonals of the cell. Refer to
Figure 4-1B. The filament across points 3 and
1 corresponds to the filament across 8 and 1,
and the filament across points 6 and 2
corresponds to the filament across 5 and 2.
5-2 ELECTRONIC CIRCUITRY
a. Amplifier Zero Adjustments
The zero adjustments on the Master
Board are factory set and normally do not
require readjustment except after
replacement of a major component. If
Rosemount Analytical Inc. A Division of Emerson Process Management Service and Maintenance 5-1
Instruction Manual
748221-M
July 2002
Model 7D
readjustment becomes necessary, use
following procedure:
AR2 Zero Adjustment:
1. Open the input to AR2. In a
single-range instrument, this is done
by disconnecting the jumper from
TB3-1 to TB3-4.
2. Ground the input of AR2 by
connecting a jumper from TB3-4 to
TP1 (GND).
3. Adjust R15, (Figure 2-6 on page 2-8 ),
for 0 VDC output signal at TB4-2.
4. Restore connections to normal.
b. Bridge Balance and Range Sensitivity
Adjustments
Perform start -up procedure in Section 3-3
on page 3-1, then proceed as follows:
1. On the analyzer module, set the front
panel ZERO and SPAN
potentiometers to their midpoints (five
turns). If the analyzer has a front
panel RANGE Switch, set it to Range
1, the highest-sensitivity range.
2. Admit downscale calibration gas to
the analyzer at the same flow rate as
is used for the sample gas. Wait for
the reading on the meter or recorder
to stabilize.
3. Refer to Figure 2-6 on page 2-8 and
set the Balance Adjustment on the
Master Board, (range 1 or single
range R8, range 2 R6, range 3 R26)
so that the reading on the meter or
recorder is zero or an offset value.
Refer to Data Sheet in the rear of this
manual.
Sensitivity Adjustment(s) are:
1. Admit upscale calibration gas to the
analyzer at the same flow rate as is
used for the sample gas. Wait for
reading on meter or recorder to
stabilize.
2. On the Master Board, set AR2 Gain
Adjustment R4 so that reading on the
recorder, if used, or on the front -panel
meter is appropriate to the upscale
calibration gas. Refer to Data Sheet
in the rear of this manual or
calibration curve. Range 1 sensitivity
is now properly adjusted.
3. If analyzer has more than one range,
set the front panel RANGE Switch for
Range 2. This is the reduced
sensitivity range. Then, on the
Master Board, set Range 2
Attenuation Adjustment R7 so that the
reading on the recorder, if used, or on
the front panel meter is appropriate to
the upscale calibration gas.
4. If the analyzer has three ranges, set
the front panel RANGE Switch for
Range 3. This is the least sensitive
range. Then, on the Master Board,
set Range 3 Attenuation Adjustment
R9 so that the reading on the
recorder, if used, or on the front panel
meter is appropriate to the upscale
calibration gas.
5. If previous readings were obtained on
a recorder, set the Meter 100%
Adjustment R22 so that the meter
fullscale matches the recorder
fullscale. If the analyzer has more
than one range, this adjustment
should be made on Range 1.
c. Bridge Voltage Adjustment
Bridge voltage is factory set as required
for the application (see Data Sheet in the
rear of this manual) and normally does
not require readjustment unless the
bridge power supply is replaced. Bridge
voltage is measured between terminals
1(-) and 2(+), and is adjustable via R2,
DWG 613561, on the bridge power
supply.
5-2 Service and MaintenanceRosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
Instruction Manual
748221-M
July 2002
d. Case Temperature Controller
Refer to Figure 6-3 on page 6-6.
Malfunction in this option can occur in
three areas:
Heater
Check continuity with ohmmeter. Verify
that resistance is approximately 113 ohms
at 25°C.
Temperature Sensor
This is an RTD and should have
approximately 550 ohms at 25°C. Check
for continuity with ohmmeter.
Thermal Fuse
The fuse opens at temperatures above
72°C. Check continuity with ohmmeter.
e. Dual Alarm Module (Optional)
This module is not user-serviceable. If
problems occur, contact Rosemount
Analytical Customer Service Center (see
page 7-1).
The module has Zero and Span
potentiometers that set the 0 to 5 V input.
To check these, adjust the analyzer for
zero gas, set the Low Alarm (ALARM 1 on
the front panel) set point to 0 %, and turn
the Deadband counterclockwise. Adjust
the Zero pot to trigger the alarm. Do the
same for High Alarm (ALARM 2 on the
front panel). The alarm span should
agree with the analyzer meter span.
Readjust the deadbands to the desired
level. (See Figure 2-7 on page 2-9 and
Figure 2-8 on page 2-11.)
5-3 SUPPRESSED ZERO ADJUSTMENT
If the instrument has suppressed-zero ranges
and does not calibrate properly, adjust the
potentiometers located on the Master Board
(see Figure 2-6 on page 2-8). Follow the
startup procedure in Section 3-3 on page 3-1.
After the analyzer has stabilized, use
appropriate calibration gases to make the
following adjustments. Refer to Section 2-3b
on page 2-1 for information on selecting
appropriate gases.
NOTE
For instruments with more than one range,
RANGE 1 is the most sensitive range; it
must be set for the highest gain. This
range should be used for the smallest
span. Always start procedure with RANGE
1 (i.e., RANGE 1 ... 95 to 100%, RANGE 2 ...
90 to 100%, RANGE 3 ... 60 to 80%).
1. Set the front panel ZERO and SPAN pots
to their midpoints (five turns).
2. If the instrument has a Range Switch, set
it to RANGE 1.
3. Connect a digital voltmeter between
TB3-1 and TP1 (polarity does not matter).
4. Flow downscale calibration gas through
the sample inlet of the analyzer. Adjust
R8 until the reading on the voltmeter is
zero. The front panel meter should also
read zero.
5. Flow upscale calibration gas through the
sample inlet of the instrument. Adjust R4
until the front panel meter reads 100 %
(fullscale).
6. Repeat steps 4 and 5 above until no
adjustment is required.
If the analyzer has more than one range,
repeat steps 2 through 6 with the following
alterations:
Range 2
Set the Range Switch to RANGE 2. In step 4,
adjust R6, and in step 5, adjust R7.
Range 3
Set the Range Switch to RANGE 3. In step 4,
adjust R26, and in step 5, adjust R9.
Rosemount Analytical Inc. A Division of Emerson Process Management Service and Maintenance 5-3
Instruction Manual
748221-M
July 2002
Model 7D
5-4 Service and MaintenanceRosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
Instruction Manual
748221-M
July 2002
SECTION 6
REPLACEMENT PARTS
WARNING
PARTS INTEGRITY
Tampering with or unauthorized substitution of
components may adversely affect safety of this
product. Use only factory-approved components
for repair.
6-1 CIRCUIT BOARD REPLACEMENT POLICY
In most situations involving a malfunction of a
circuit board, it is more practical to replace the
board than to attempt isolation and
replacement of the individual component, as
the cost of test and replacement will exceed
the cost of a rebuilt assembly. As standard
policy, rebuilt boards are available on an
exchange basis.
Because of the exchange policy covering
circuit boards, the following list does not
include individual electronic components. If
circumstances necessitate replacement of an
individual component, which can be identified
by inspection or from the schematic diagrams,
obtain the replacement component from a
local source of supply.
6-2 REPLACEMENT PARTS
The following parts are recommended for
routine maintenance and troubleshooting of
the Model 7D. If the troubleshooting
procedures do not resolve the problem,
contact Rosemount Analytical Customer
Service Center (see page 7-1).
Figure 6-1 through Figure 6-3 show locations
of components and assemblies.
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-1
Parts are selected or are optional depending on instrument application. Refer to Application Data Sheet and ordering info r-
mation to determine which parts are applicable to your instrument.
2
Refer to section listed for replacement parts of assembly.
3
Sealed reference cells must be serviced at factory.
6-2 Replacement PartsRosemount Analytical Inc. A Division of Emerson Process Management
Model 7D
Hydrogen in Carbon Dioxide
Instruction Manual
748221-M
July 2002
6-3 MATRIX
Each analyzer is configured per the customer
sales order. Below is the Model 7D sales
matrix which lists the various configurations
available.
7D MODEL 7D TC ANALYZER (EXPLOSION PROOF ENCLOSURE)
Parameter and Background Gas
CodeCode
10Hydrogen in Air21Helium in Nitrogen
11Hydrogen in Nitrogen22Helium in Oxygen
12Hydrogen in Methane23Helium in Methane
13Hydrogen in Argon24Helium in Argon
14
20Helium in Air26Helium in Argon & Oxygen
50Nitrogen in Hydrogen (0-5000 ppm Range Only) Range Code #31
60Refinery (15% Hydrocarbon - 85% N2; 100% N2; 1% Hydrogen in 99% N2) Range Code #60
61Utilities (BD Special) (80-100% H2 in Air, 0-100% H2 in CO2 and 0-100% Air in CO2) Range Code #61
90Argon in Air
91Argon in Nitrogen
92Argon in Oxygen
93Carbon Dioxide in Air
94Carbon Dioxide in Nitrogen
95Carbon Dioxide in Oxygen
96Nitrogen in Argon
97Corrosive Gas Mixture Application (Zero Based Ranges Only) Range Code #97
98Corrosive Gas Mixture Application (Suppressed Range Only) Range Code #98
99Special
Code Range
1485-100%XXX
1595-100%XXX
1699-100%XXX
1799.5%-100%XXX
19Special Suppressed Range
300-500 ppmXXX
310-5000 ppmXXX
010-1%XXX
020-5%XXX
030-10%XXX
040-15%XXX
050-25%XXX
060-50%XXX
070-75%XXX
080-100%XXX
09Special Zero Based Range
1020-50%XXX
1145-55%XXX
1260-80%XXX
1350-100%XXX
60RefineryXXX
61UtilitiesXXX
97Corrosive Gas Mixture - Application Zero Based RangeXGold Sheathed
98Corrosive Gas Mixture - Application Suppressed RangeXGold Sheathed
Code Single & Dual Range Analyzer
01 One Range, Voltage Output
02 One Range, Linearized Voltage Output
03 One Range with 4-20 mA, Isolated Current Output
04 One Range, Linearized and 4-20 mA, Isolated Current Output
05 Two Ranges, Voltage Output
06 Two Ranges, Range One Linearized Voltage Output
Code Second Range: Dual Range Analyzer
XX Range Selections same as position 5, 6; 00 if second range is not selected.
7D1203020003Example
1,23,45,67,89,1011,12 Matrix positions
25Helium in Carbon Dioxide
Code Power, Alarm, Tropicalization
01 115V, 50/60 Hz
02 115V, 50/60 Hz with Dual Alarm
03 115V, 50/60 Hz with Tropicalization
04 115V, 50/60 Hz with Dual Alarm and Tropicalization
05 230V, 50/60 Hz
06 230V, 50/60 Hz with Dual Alarm
07 230V, 50/60 Hz with Tropicalization
08 230V, 50/60 Hz with Dual Alarm and Tropicalization
Not
Applicable
To identify the configuration of an analyzer,
locate the analyzer name-rating plate. The
sales matrix identifier number appears on the
analyzer name-rating plate.
LinearizationReferenceFilament
Optional
Not
Available
Sealed Flowing Tungsten Hytempco
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-3
Instruction Manual
898672
(Kit 654648)
638426
748221-M
July 2002
Model 7D
193311 Meter
REF Hinge Side
Flame Arrestor
ALARM 1 Potentiometer
(Kit 654648)
RANGE Switch
(Kit 654892)
ALARM 2 Potentiometer
ZERO Potentiometer
901917
SPAN Potentiometer
Figure 6-1. Analyzer Assembly - Door and Pneumatic Components
6-4 Replacement PartsRosemount Analytical Inc. A Division of Emerson Process Management
Model NGA 2000 TO2
Alarm Kit
±
15V Power Supply
Linearizer Board (option)
Isolated Current Output Board
Bridge Power Supply
Fuse
Temperature Control Assembly
Master Board
652816
Instruction Manual
748221-M
July 2002
616443
652270
654620
619714
613560
REF Hinge Side
000516 (115 VAC)
801566 (230 VAC)
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-5
6-6 Replacement PartsRosemount Analytical Inc. A Division of Emerson Process Management
Model NGA 2000 TO2
RETURN OF MATERIAL
Instruction Manual
748221-M
July 2002
SECTION 7
7-1 RETURN OF MATERIAL
If factory repair of defective equipment is
required, proceed as follows:
1. Secure a return authorization from a
Rosemount Analytical Inc. Sales Office or
Representative before returning the
equipment. Equipment must be returned
with complete identification in accordance
with Rosemount instructions or it will not
be accepted.
Rosemount CSC will provide the shipping
address for your instrument.
In no event will Rosemount be
responsible for equipment returned
without proper authorization and
identification.
2. Carefully pack the defective unit in a
sturdy box with sufficient shock absorbing
material to ensure no additional damage
occurs during shipping.
3. In a cover letter, describe completely:
•The symptoms that determined the
equipment is faulty.
•The environment in which the
equipment was operating (housing,
weather, vibr ation, dust, etc.).
•Site from where the equipment was
removed.
•Whether warranty or non-warranty
service is expected.
•Complete shipping instructions for the
return of the equipment.
4. Enclose a cover letter and purchase order
and ship the defective equipment
according to instructions provided in the
Rosemount Return Authorization, prepaid,
to the address provided by Rosemount
CSC.
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
If warranty service is expected, the defective
unit will be carefully inspected and tested at
the factory. If the failure was due to the
conditions listed in the standard Rosemount
warranty, the defective unit will be repaired or
replaced at Rosemount’s option, and an
operating unit will be returned to the customer
in accordance with the shipping instructions
furnished in the cover letter.
For equipment no longer under warranty, the
equipment will be repaired at the factory and
returned as directed by the purchase order
and shipping instructions.
7-2 CUSTOMER SERVICE
For order administration, replacement Parts,
application assistance, on-site or factory
repair, service or maintenance contract
information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
7-3 TRAINING
A comprehensive Factory Training Program of
operator and service classes is available. For
a copy of the Current Operator and ServiceTraining Schedule contact the Technical
Services Department at:
Rosemount Analytical Inc.
Customer Service Center
1-800-433-6076
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 7-1
Instruction Manual
748221-M
July 2002
Model 7D
7-2 Return of MaterialRosemount Analytical Inc. A Division of Emerson Process Management
WARRANTY
Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from
defects in workmanship and material under normal use and service for a period of twelve (12)
months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid
junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and
material under normal use and service for a period of ninety (90) days from date of shipment by
Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or
material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the
goods, part(s) or consumables are returned to Seller's designated factory, transportation charges
prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in
the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in
effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day
warranty in the case of consumables. A defect in goods, part(s) and consumables of the
commercial unit shall not operate to condemn such commercial unit when such goods, part(s)
and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly
or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from
any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS,
SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER
WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED
WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY
DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF
WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD
WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF
ACTION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC
GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER
AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE
INCIDENTAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL
INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE,
LOSS OF REVENUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY
OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN
ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD
PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE
PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS
PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond
Seller's direct control.
Instruction Manual
Emerson Process Management
ASIA - PACIFIC
Fisher-Rosemount GmbH & Co.
EUROPE, MIDDLE EAST, AFRICA
West Sussex PO22 9SH
LATIN AMERICA
748221-M
July 2002
Model 7D
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
E gas.csc@emersonprocess.com
Fisher-Rosemount
Singapore Private Ltd.
1 Pandan Crescent
Singapore 128461
Republic of Singapore
T 65-777-8211
F 65-777-0947