Rosemount 371 pH Sensor with Variopol Connector Manuals & Guides

Model 371 pH Sensor with VP 6.0
13.5 EuroSenz
Instruction Sheet
PN 51A-371pHVP/rev.E January 2011
SPECIFICATIONS
pH Range: 0-14 pH
Resistance (GPLR)*
Percent Linearity: 96% @ 0-4 pH, 99% @ 4-12
pH, 97% @ 12-13 pH, 92% @ 13-14 pH
Wetted Materials: Polypropylene, EP, and Glass
Process Connection: Standard PG 13.5 Thread
Electrode Connection: Variopole (VP 6.0) for TC
Temperature Compensation: Optional Pt100 or
Pt1000
Temperature Range: 0° - 95°C (32° - 203°F)
Pressure Range: 100 - 790 kPa abs (0 - 100
psig)
STORAGE
1. It is recommended that electrodes be stored in their original shipping containers until needed.
2. Do not store at temperatures below -5°C (23°F).
3. Electrodes should be stored with a protective cap containing KCl solution (PN 9210342).
4. For overnight storage, immerse the sensor in tap water or 4 pH buffer solution.
5. A pH glass electrode has a limited shelf life of one year.
FIGURE 1. Sensor Dimensions
MILLIMETER
INCH
For additional information, please visit our website at
www.emersonprocess.com/raihome/liquid/.
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!
WARNING
The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the
responsibility
of the user.
CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY
ATEX DIRECTIVE
Special Conditions for safe use
1. All pH/ORP sensors have a plastic enclosure which must only be cleaned with a damp cloth to avoid the danger due to a build up of an electro­static charge.
2. All pH/ORP sensor Models are intended to be in contact with the process fluid and may not meet the 500V r.m.s. a.c. test to earth. This must be taken into consideration at installation.
FIGURE 2. Sensor Orientation
MOUNTING
For sensor orientation, see Figure 2.
WIRING
For wiring, see Figures 3 to 8.
FIGURE 3. VP 6.0 Cable Description
CABLE
FIGURE 4. Wiring to Models 54, 81, 3081, and
remote junction box PN 23555-00
FIGURE 5. Wiring to Model 2081
WIRING
ELECTRODE PREPARATION
1. Remove electrode from shipping container.
2. Remove the protective boot covering the electrode bulb.
3. Rinse away salt film with clean water, then shake the electrode so that the internal solution fills the bulb, thus removing any air trapped there.
NOTE
Do not allow lubricant to coat electrode bulb or reference junction. If it does, wipe it clean before installation.
TWO POINT BUFFER CALIBRATION
Select two stable buffer solutions, preferably pH 4.0 and 10.0 (pH buffers other than pH 4.0 and pH 10.0 can be used as long as the pH values are at least two pH units apart).
NOTE
A pH 7 buffer solution reads a mV value of approx. zero, and pH buffers read approximate­ly ± 59.1 mV for each pH unit above or below pH
7. Check the pH buffer manufacturer specifica­tions for millivolt values at various temperatures since it may affect the actual value of the buffer solution mV/pH value.
1. Immerse sensor in the first buffer solution. Allow sensor to equilibrate to the buffer temperature (to
avoid errors due to temperature differences between the buffer solution and sensor tempera­ture) and wait for reading to stabilize. Value of buffer can now be acknowledged by analyzer/transmitter.
2. Once the first buffer has been acknowledged by the analyzer/transmitter, rinse the buffer solution off of the sensor with distilled or deionized water.
3. Repeat steps 1 and 2 using the second buffer solu­tion.
4. The theoretical slope value, according to the Nernst equation for calculating pH, is approximately 59.17 mV/pH. Over time the sensor will age, both in the process and in storage, and will result in reduced slope values. To ensure accurate readings, it is rec­ommended that the electrode be replaced when the slope value falls below 47 to 49 mV/pH.
FIGURE 6. Wiring to Models 1054 series and 2054
FIGURE 7. Wiring to Model 54e
FIGURE 8. Wiring to Model 1055-22-32
RECOMMENDED pH SENSOR STANDARDIZATION
For maximum accuracy, the sensor can be standardized on-line or with a process grab sample after a buffer cali­bration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sensor’s slope but will sim­ply adjust the analyzer’s reading to match that of the known process pH.
MAINTENANCE
Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement.
1. To remove oil deposit, clean the electrode with a mild non-abrasive detergent.
2. To remove scale deposits, soak electrodes for 30 to 60 minutes in a 5% hydrochloric acid solution.
3. Temperature effect on life expectancy: If glass electrode life expectancy is 100% @ 25°C (77°F), then it will be approximately 25% @ 80°C (176°F), and approximately 5% @ 120°C (248°F).
Credit Cards for U.S. Purchases Only.
The right people, the right answers, right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Specifications subject to change without notice.
Emerson Process Management
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2011
8
Loading...