Rosemount 3300HT/3300HTVP/3400HT/3400HTVP PERpH-X High Performance pH/ORP Sensors Quick Start Guide

Quick Start Guide
00825-0100-3300, Rev AA
November 2020
Rosemount™ 3300HT/3300HTVP/ 3400HT/3400HTVP
PERpH-X High Performance pH/ORP Sensors
Quick Start Guide November 2020
Essential instructions
Read this page before proceeding!
Emerson designs, manufactures, and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. You must adhere to the following instructions and integrate them into your safety program when installing, using, and maintaining Emerson's Rosemount™ products. Failure to follow the proper instructions may cause any one of the following situations to occur: loss of life, personal injury, property damage, damage to this instrument, and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Emerson representative for
clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the
product.
Install equipment as specified in the installation instructions of the appropriate Quick Start Guide
and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and
maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts
specified by Emerson. Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified people, to prevent electrical shock and personal injury.
Note
The information contained in this document is subject to change without notice.
WARNING
Pressure and temperature
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level.
Do not insert or retract retractable sensors when process pressures are in excess of 64 psig (542 kPa) for option 21 or 35 psig (343 kPa) for option 25.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
2 Emerson.com/Rosemount
November 2020 Quick Start Guide
CAUTION
Sensor/process application compatibility
The wetted sensor materials may not be compatible with process composition and operating conditions.
Application compatibility is entirely the operator's responsibility.
Contents
Install........................................................................................................................................... 5
Start up and calibrate................................................................................................................. 25
Maintenance.............................................................................................................................. 27
Rosemount pH/ORP sensor(s) product certifications..................................................................29
Declaration of Conformity..........................................................................................................36
China RoHS tables...................................................................................................................... 38
Intrinsically safe sensor installation drawing - FM........................................................................39
Quick Start Guide 3
Quick Start Guide November 2020
4 Emerson.com/Rosemount
November 2020 Quick Start Guide

1 Install

1.1 Storage

1. Emerson recommends storing electrodes in their original shipping containers until needed.
2. Do not store at temperatures below 23 °F (-5 °C).
3. Store electrodes with a protective cap containing KCl solution (PN 9210342).
4. For overnight storage, immerse the sensor in tap water or 4 pH buffer solution.
5. A pH glass electrode has a limited shelf life of one year.

1.2 Installation

For sensor dimensions, see Figure 1-1 through Figure 1-4.
For sensor orientation and installation, see Figure 1-5 through Figure 1-11.
Quick Start Guide 5
Quick Start Guide November 2020
Figure 1-1: Rosemount 3300HT Sensor Dimensional Drawing
A. Sensor body B. 2-in. (50.8 mm) Swagelok® fitting process connector assembly PN
23166-00/01
C. 1-in. (25.4 mm) national pipe thread (NPT)
6 Emerson.com/Rosemount
November 2020 Quick Start Guide
Figure 1-2: Rosemount 3300HTVP Sensor Dimensional Drawing
A. Cable with Variopol receptacle (female) connector B. Variopol plug (male) connector C. Sensor body
D. Retraction stop collar
Quick Start Guide 7
Quick Start Guide November 2020
Figure 1-3: Rosemount 3400HT and 3400HTVP Sensor Dimensions with Optional Ball Valve PN 23240-00
A. Dimension (see Table 1-1) B. Dimension (see Table 1-1) C. Dimension (see Table 1-1)
D. Cable
E. Cable bushing polypropylene F. 1-in. by 1-in. (25.4 x 25.4 mm) Swagelok fitting kit (PN 23166-00 or
23166-01) required to connect sensor directly to process or to ball valve
G.
WARNING
Residual pressure and process may remain trapped between ball valve and male connector.
Maximum pressure at retraction:
Option 21: 64 psig (441.2 kPa) Option 25: 35 psig (241.3 kPa)
8 Emerson.com/Rosemount
November 2020 Quick Start Guide
H. O.D. titanium housing
I. O.D. retraction stop collar J. Electode housing Ryton
K. pH electrode
L. 1½-in. (38.1 mm) female pipe thread (FPT) ball valve PN 9340065
M. 1½-in. (38.1 mm) male pipe thread (MPT) close nipple
N. 1½-in. x 1-in. (38.1 x 25.4 mm) FPT reducing coupling O. Ball valve (PN 23240-00) optional
Table 1-1: Rosemount 3400HT and 3400HTVP Sensor Dimensions with Optional Ball Valve PN 23240-00
Option Dimension A
(in./mm)
21 21.6/549 14.0/355 23.1/587
25 36.1/917 28.5/724 37.6/955
Dimension B (in./mm)
Dimension C (in./mm)
Quick Start Guide 9
Quick Start Guide November 2020
Figure 1-4: Rosemount 3400HT and 3400HTVP Sensor Dimensions with Optional Ball Valve PN 23765-00
A. Dimension (see Table 1-2) B. Dimension (see Table 1-2) C. Housing titanium
D. Retraction stop collar
E. Electrode housing
F. pH electrode G. 1¼-in. (38.1 mm) MPT close nipple H. Ball valve
I. Packing adapter J. Knurl retainer
K. Ball valve kit PN 23765-00
10 Emerson.com/Rosemount
November 2020 Quick Start Guide
Table 1-2: Rosemount 3400HT and 3400HTVP Sensor Dimensions with Optional Ball Valve PN 23765-00
Option A: in./mm B: in./mm
21 21.6/549 12.2/310
25 36.1/917 26.7/678
Rosemount 3300HT and 3300HTVP flow through and insertion installation
1½-in. (38.1 mm) pipe tee (PN 2002011) with 1-in. (25.4 mm) threaded connections.
Note
Mount the sensor at least 10 degrees above the horizon.
Figure 1-5: Straight Flow
A. 1½-in. (38.1 mm) pipe tee with 1-in. (25.4 mm) threaded connections:
PN 2002011
Quick Start Guide 11
Quick Start Guide November 2020
Figure 1-6: Angle Flow
A. 1½-in. (38.1 mm) pipe tee with 1-in. (25.4 mm) threaded connections:
PN 2002011
B. Flow
Figure 1-7: Pipe Y Installation
A. 1½-in. x 1-in. (38.1 x 25.4 mm) reducing bushing B. 1½-in. (38.1 mm) pipe "Y"
12 Emerson.com/Rosemount
November 2020 Quick Start Guide
Figure 1-8: Rosemount 3400HT and 3400HTVP Mounting Details ­Retraction Version
A. Weldalet (1½-in. [38.1 mm] FPT) B. Electrode housing tip C. Electrode
D. Flow
E. Pipe "Y" F. Pipe tee
G. Flow
Quick Start Guide 13
Quick Start Guide November 2020
Figure 1-9: Rosemount 3300 and 3500VP Flow through Installation
A. 1½-in. (38.1 mm) schedule 80 CPVC tee with 1-in. (25.4 mm) female
national pipe thread (FNPT) connections (code 16) straight flow shown B. Always mount sensor at least 10 degrees above horizontal. C. 1½-in. (38.1 mm) pipe "Y"
D. 1½-in. x 1-in. (38.1 x 25.4 mm) reducing bushing
E. Flow F. 1½-in. (38.1 mm) schedule 80 CPVC with view or cleaning port
G. Angle flow shown
14 Emerson.com/Rosemount
November 2020 Quick Start Guide
Figure 1-10: Type 4 Junction Box
A. PN 23550-00 for point-to-point wiring used with extension cable. Cable
for pH, oxidation reduction potential (ORP), conductivity, or amperometric sensors. PN 23555-00 with preamplifier for pH and ORP sensors only.
B. PN 23820-00 carbon steel with epoxy coating. PN 23820-02 316
stainless steel
C. Flexible conduit is required.
D. 1-in. (25.4 mm) pipe by others
E. 1-in. (25.4 mm) female national pipe thread (FNPT) CPVC union: PN
9320057
F. 1-in. (25.4 mm) male national pipe thread (MNPT) connector G. Any 1-in. (25.4 mm) threaded sensor H. 2¾-in. (69.9 mm) O.D.
Quick Start Guide 15
Quick Start Guide November 2020
Figure 1-11: Submersion Installation: Handrail Mounting Accessory (PN 11275-01)
A. Sensor cable B. 1½-in. (38.1 mm) PVC pipe schedule 80 C. 1½-in. (38.1 mm) pipe clamp, three places
D. 45-degree sweep ell extension pipe
E. Regularly check to make sure connections are water tight
F. Unistrut 1-in. x 1-in. (41.3 x 41. 3 mm) aluminum G. 1½-in. (38.1 mm) pipe clamp, two places H. Can be any convenient dimension
I. Customer handrail, two places J. Mounting channel aluminum, two places
K. Locking pin with bead chain
Note
Unless otherwise specified.
16 Emerson.com/Rosemount
November 2020 Quick Start Guide

1.3 Retraction with kit PN 23240-00

WARNING
System pressure may cause the sensor to blow out with great force.
Take care when removing the sensor.

1.3.1 Retract Rosemount 3400HT/3400HTVP-21 (21-in. [533.4 mm] tube)

WARNING
Be certain the system pressure at the sensor is below 64 psig (542 kPa [abs]) before retracting the sensor.
Wear a face shield. Make sure your footing is stable.
Procedure
Push in on the sensor at the top of the junction box and slowly loosen the hex nut of the process end male connector.

1.3.2 Retract Rosemount 3400HT/3400HTVP-25 (36-in. [914,4 mm]) tube

WARNING
Be certain the system pressure at the sensor is below 64 psig (542 kPa [abs]) before retracting the sensor.
Wear a face shield. Make sure your footing is stable.
Procedure
1. Push in the sensor end or the top of the junction box and slowly loosen the hex nut of the process end male connector.
CAUTION
Do not remove nut at this time.
2. When the hex nut is loose enough, slowly ease the sensor back completely until it reaches the retraction stop collar.
Quick Start Guide 17
Quick Start Guide November 2020
CAUTION
Failure to withdraw the sensor completely may result in damage to the sensor when the valve is closed.
3. Close the ball valve slowly. If there is resistance, the valve may be hitting the sensor. Double
check that the sensor has been retracted to the retraction stop collar.
WARNING
Leakage from the male connector threads may indicate that the male connector is still under pressure.
Leakage through a partially open ball valve could be hazardous; however, with the ball valve closed, some residual process fluid may leak from the connector's pipe threads.
Before removing the sensor from the ball valve, be absolutely certain that the ball valve is fully closed.
4. Unthread the male body connector completely from the reducing coupling and remove the sensor for servicing.
Note
If the male connector leaks during insertion or retraction, replace the O-ring (PN 25394-01) in the male connector.
1.4
18 Emerson.com/Rosemount

Wiring

For additional wiring information on this product, including sensor combinations not shown here, please refer to Emerson.com/
RosemountLiquidAnalysisWiring.
November 2020 Quick Start Guide
Figure 1-12: Rosemount PERpH-X Sensors, Preamplifier in Analyzer
For Rosemount 3300HT, 3300HTVP, 3400HT, and 3400HTVP
Non-preamplifier sensors
Blue cable
Table 1-3: Rosemount PERpH-X Sensors, Preamplifier in Analyzer
Wire color Rosemount 56, 1056,
1057, and 1066 terminal number
N/A Inner drain (no
connection)
N/A N/A 1 N/A
Green Connect to instrument
ground or as shown
White Resistance temperature
device (RTD) return
White/red RTD sense 4 2
Red RTD in 5 3
White/gray Reference shield 6 4
Gray Reference in 7 5
Blue Solution ground 8 6
Clear mV/pH shield 9 7
Orange mV/pH in 10 8
N/A -5 Vdc 11 9
N/A +5 Vdc 12 10
Rosemount 54e and 5081 terminal number
N/A N/A
2 N/A
3 1
Rosemount XMT terminal number
Quick Start Guide 19
Quick Start Guide November 2020
Figure 1-13: Rosemount PERpH-X Sensors, Preamplifier in Junction Box
For Rosemount 3300HT, 3300HTVP, 3400HT, and 3400HTVP
Non-preamplifier sensors
Blue cable
Extension cable PN 23646-01 or 9200273 shown.
Can also use extension cable PN 9200348; extend color scheme from sensor or VP8 cable instead of using shown wiring scheme.
Table 1-4: Rosemount PERpH-X Sensors, Preamplifier in Junction Box
Wire color
N/A N/A N/A Inner drain (no
N/A 1 N/A N/A 1 N/A
Green 2 Inner drain Connect to
White 3 White RTD return 3 1
White/red 4 White/red RTD sense 4 2
Red 5 Red RTD in 5 3
White/ gray
Gray 7 Gray Reference in 7 5
Blue 8 Blue Solution ground 8 6
20 Emerson.com/Rosemount
Junction box terminal number
6 White/
Extension cable color
gray
Rosemount 56, 1056, 1057, and 1066 terminal number
connection)
instrument ground as shown
Reference shield 6 4
Rosemount 54e and 5081 terminal number
N/A N/A
2 N/A
Rosemount XMT terminal number
November 2020 Quick Start Guide
Table 1-4: Rosemount PERpH-X Sensors, Preamplifier in Junction Box (continued)
Wire color
Clear 9 White/
Orange 10 Black mV/pH in 10 8
N/A 11 Brown -5 Vdc 11 9
N/A 12 Green +5 Vdc 12 10
Junction box terminal number
Extension cable color
black
Rosemount 56, 1056, 1057, and 1066 terminal number
mV/pH shield 9 7
Rosemount 54e and 5081 terminal number
Rosemount XMT terminal number
Quick Start Guide 21
Quick Start Guide November 2020
Figure 1-14: PERpH-X Sensors, Preamplifier in Sensor
For Rosemount 3300HTVP-70 and 3400HTVP-70
Preamplifier in sensor
Blue cable
Table 1-5: Rosemount PERpH-X Sensors, Preamplifier in Sensor
Wire color Rosemount 56,
1056, 1057, and 1066 terminal number
N/A N/A 1 N/A
Green Connect to
instrument ground or as shown
White RTD return 3 1
White/ red RTD sense 4 2
Red RTD in 5 3
N/A Reference shield 6 4
Gray Reference in 7 5
Blue Solution ground 8 6
Clear mV/pH shield 9 7
Orange mV/pH in 10 8
White/gray -5 Vdc 11 9
Inner drain +5 Vdc 12 10
Rosemount 54e and 5081 terminal number
2 N/A
Rosemount XMT terminal number
22 Emerson.com/Rosemount
November 2020 Quick Start Guide
Figure 1-15: Rosemount PERpH-X Sensors, Preamplifier in Sensor, Junction Box Used for Extending Cable
For Rosemount 3300HTVP-70 and 3400HTVP-70
Preamplifier in sensor
Blue cable
Extension cable PN 23646-01 or 9200273 shown.
Can also use extension cable PN 9200348: white/gray is replacement wire color for brown wire.
Table 1-6: Rosemount PERpH-X Sensors, Preamplifier in Sensor, Junction Box Used for Extending Cable
Wire color
N/A 1 N/A N/A 1 N/A
Green 2 Inner drain.
White 3 White RTD return 3 1
White/red4 White/red RTD sense 4 2
Red 5 Red RTD in 5 3
N/A 6 White/gray (no
Gray 7 Gray Reference in 7 5
Quick Start Guide 23
Junction box terminal number
Extension cable color
Connect to instrument ground or as shown.
connection)
Rosemount 56, 1056, 1057, and 1066 terminal number
N/A 2 N/A
Reference shield
Rosemount 54e and 5081 terminal number
6 4
Rosemount XMT terminal number
Quick Start Guide November 2020
Table 1-6: Rosemount PERpH-X Sensors, Preamplifier in Sensor, Junction Box Used for Extending Cable (continued)
Wire color
Blue 8 Blue Solution ground 8 6
Clear (orange shield)
Orange 10 Black pH in 10 8
White/ gray
ID (inner drain)
Junction box terminal number
9 White/black pH shield 9 7
11 Brown -5 Vdc 11 9
12 Green +5 Vdc 12 10
Extension cable color
Rosemount 56, 1056, 1057, and 1066 terminal number
Rosemount 54e and 5081 terminal number
Rosemount XMT terminal number
24 Emerson.com/Rosemount
November 2020 Quick Start Guide

2 Start up and calibrate

2.1 Prepare electrode

Procedure
1. Remove electrode from shipping container.
2. Remove the protective boot covering the electrode bulb.
3. Rinse away salt film with clean water; then shake the electrode so that the internal solution fills the bulb, thus removing any air trapped there.
WARNING
The buffer in the protective boot may cause skin or eye irritation.

2.2 Calibrate with two point pH buffer solution

Select two stable buffer solutions, preferably pH 4.0 and pH 7.0. You can use buffers other than pH 4.0 and pH 7.0 as long as the pH values are least two pH units apart.
Note
A pH 7 buffer solution reads an mV value of approximately zero, and pH buffers read approximately ±59.1 mV for each pH unit above or below pH 7. Check the pH buffer manufacturer specifications for milivolt values at various temperatures, since temperature may affect the actual value of the pH buffer solution mV/pH value.
Procedure
1. Immerse the sensor in buffer solution. Make sure the buffer solution contacts the metal housing of the
sensor, which acts as the solution ground contact.
2. Allow sensor to equilibrate to the buffer temperature and wait for the reading to stabilize. The transmitter can now acknowledge the value of the buffer.
3. Rinse the buffer solution off of the sensor with distilled or deionized water.
4. Repeat steps Step 1 through Step 3 using the second buffer solution.
The theoretical slope value, according to the Nernst equation for calculating pH, is approximately 59.17 mV/pH. Over time the sensor will age, both in the process and in storage, and this will result in reduced slope values. To
Quick Start Guide 25
Quick Start Guide November 2020
ensure accurate readings, Emerson recommends replacing the electrode when the slope value falls below 47 to 49 mV/pH.

2.3 Standardize pH sensor

For maximum accuracy, you can standardize the sensor in-line or with a process grab sample after performing a buffer calibration and conditioning the sensor to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization does not change the sensor's slope, but simply adjusts the transmitter's reading to match that of a known process pH.
26 Emerson.com/Rosemount
November 2020 Quick Start Guide

3 Maintenance

3.1 Clean electrodes

Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that you may need to clean or replace the electrodes.
Procedure
1. To remove oil deposits, clean the electrodes with a mild, non­abrasive detergent.
2. To remove scale deposits, soak electrodes for 30 to 60 minutes in a five percent hydrochloric acid solution.

3.2 Replace reference junction and recharge sensor electrolyte

The reference junction and reference fill gelled solution are replaceable to facilitate longer sensor life. Over time, the electrolyte depletes and the junction may become plugged or contaminated.
Prerequisites
Use the junction replacement kit and replacement fill gel.
Procedure
1. Remove the junction cap by turning it counter clockwise.
2. Remove the liquid junction by pulling the junction straight out.
WARNING
The reference electrolyte may cause skin or eye irritation
Avoid contact with skin and eyes.
3. Rinse with water to remove the old reference fill gel.
4. Fill the reference fill chamber with the reference fill gel using the syringe, and remove any air bubbles. Top off the reference fill chamber until it is completely filled.
5. Replace the junction O-ring and liquid junction by sliding them over the glass electrode. Excess reference gel should flow out.
6. Replace junction cap by turning clockwise. Hand tighten the junction cap only; do not use pliers to tighten the
cap.
Quick Start Guide 27
Quick Start Guide November 2020
7. Buffer check and calibrate the sensor as described in Calibrate with
two point pH buffer solution.
28 Emerson.com/Rosemount
November 2020 Quick Start Guide
4 Rosemount pH/ORP sensor(s) product
certifications
Rev 0.5

4.1 European directive information

A copy of the EU Declaration of Conformity can be found at the end of the Quick Start Guide. The most recent revision of the EU Declaration of Conformity can be found at Emerson.com/Rosemount.

4.2 Ordinary location certification

As standard, the transmitter has been examined and tested to determine that the design meets the basic electrical, mechanical, and fire protection requirements by a nationally recognized test laboratory (NRTL) as accredited by the Federal Occupational Safety and Health Administration (OSHA).

4.3 Installing equipment in North America

The US National Electrical Code® (NEC) and the Canadian Electrical Code (CEC) permit the use of Division marked equipment in Zones and Zone marked equipment in Divisions. The markings must be suitable for the area classification, gas, and temperature class. This information is clearly defined in the respective codes.
4.4

USA

4.4.1 FM Intrinsic Safety

Certificate
Standards
Markings
Specific Conditions for Safe Use (X):
1. Sensors with Model 1700702 preamplifier:
Quick Start Guide 29
FM17US0198X
FM Class 3600:1998, FM Class 3610:2010, FM Class 3611: 2004, FM Class 3810: 2005
IS/I,II,III/1/ABCDEFG/T6 Ta = –20 °C to 60 °C I/0/AEx ia IIC/T6 Ta = –20 °C to 60 °C NI/I/2/ABCD/T6 Ta = –20 °C to 60 °C S/II,III/2/EFG/T6 Ta = –20 °C to 60 °C
a. Model 385+-a-b-c. Triple junction pH/ORP sensor
b. Model 389-a-b-c-d-e. pH/ORP sensor
c. Model 389VP-a-b-c-d. pH/ORP sensor
d. Model 396VP-a-b-c-d. Submersion/insertion pH/ORP sensor
Quick Start Guide November 2020
e. Model 396P-a-b-c-d-e. Submersion/insertion pH/ORP sensor
f. Model 396PVP-a-b-c-d-e. Submersion/insertion pH/ORP
sensor
g. Model 396RVP-a-b-c-d-e. Retraction/submersion/insertion
pH/ORP sensor
h. Model 398RVP-a-b-c-d-e-f. pH/ORP sensor
i. Model 3200HP-00. High purity water pH sensor
j. Model 3300HTVP-a-b-c-d. High performance pH and ORP
sensor
k. Model 3400HTVP-a-b-c-d-e. High performance pH and ORP
sensor
l. 3500P-a-b-c-d-e-f. High performance pH and ORP sensor
m. 3500VP-a-b-c-d-e-f. High performance pH and ORP sensor
n. Model 3900-a-b-c. General purpose pH/ORP sensor
o. Model 3900VP-a-b. General purpose pH/ORP sensor
The polymeric surface of all the apparatus listed above may store electrostatic charge and become a source of ignition. Clean surface should only be done with a damp cloth.
2. Sensors without Model 1700702 preamplifier (simple apparatus):
a. Model 385-a-b-c-d-e. Retractable pH/ORP sensor
b. Model 385+-a-b-c Triple junction pH/ORP sensor
c. Model 389-a-b-c-d-e. pH/ORP sensor
d. Model 389VP-a-b-c. pH/ORP sensor
e. Model 396-a-b-c. Submersion/insertion pH sensor
f. Model 396VP-a-b. Submersion/insertion pH sensor
g. Model 396P-a-b-c-d-e. Submersion/insertion pH/ORP sensor
h. Model 396PVP-a-b-c-d. Submersion/insertion pH/ORP sensor
i. Model 396R-a-b-c-d-e. Retraction/submersion/insertion
pH/ORP sensor
j. Model 396RVP-a-b-c-d. Retraction/submersion/insertion
pH/ORP sensor
k. Model 397-a-b-c-d-e. pH sensor
l. Model 398-a-b-c-d-e. pH/ORP sensor
30 Emerson.com/Rosemount
November 2020 Quick Start Guide
m. Model 398VP-a-b-c. pH/ORP sensor
n. Model 398R-a-b-c-d-e-f. pH/ORP sensor
o. Model 398RVP-a-b-c-d-e-f. pH/ORP sensor
p. Model 3200HP-00. High purity water pH sensor
q. Model 3300HT-a-b-c-d. High performance pH and ORP sensor
r. Model 3300HTVP-a-b-c-d. High performance pH and ORP
sensor
s. Model 3400HT-a-b-c-d-e-f. High performance pH and ORP
sensor
t. Model 3400HTVP-a-b-c-d-e-f. High performance pH and ORP
sensor
u. Model 3500P-a-b-c-d-e-f. High performance pH and ORP
sensor
v. Model 3500VP-a-b-c-d-e-f. High performance pH and ORP
sensor
w. Model 3800-a. Autoclaveable and steam sterilizable pH
sensors
x. Model 3800VP-a. Autoclaveable and steam sterilizable pH
sensors
y. Model 3900-a-b-c. General purpose pH/ORP sensor
z. Model 3900VP-a-b. General purpose pH/ORP sensor
The polymeric surface of all the apparatus listed above may store electrostatic charge and become a source of ignition. Clean surface should only be done with a damp cloth.

4.4.2 CSA Intrinsic Safety

Certificate
Standards
Markings
Quick Start Guide 31
70164066
C22.2 No 0-10, C22.2 No 0.4-M2004, C22.2 No 94-M1991, C22.2 No 142 – M1987, C22.2 No 157-M1992, CAN/CSA E60079-0:07, CAN/CSA E60079-11:02, UL 50-11th Ed, UL 508-17th Ed, UL 913-7th Ed, UL 60079-0: 2005, UL 60079-11: 2002
Preamplifier assembly: Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups ABCD; ambient temperature rating –20 °C to +60 °C; Ex ia IIC; T6: Class I, Zone 0, AEx ia IIC ; T6
Quick Start Guide November 2020
Sensor apparatus with preamplifier: Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups ABCD; ambient temperature rating –20 °C to +60 °C; Ex ia IIC; T6: Class I, Zone 0, AEx ia IIC ; T6
Sensor apparatus: Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups ABCD; Ex ia IIC; T6; ambient temperature rating –20 °C to +60 °C: (simple apparatus)

4.5 Canada

4.5.1 CSA Intrinsic Safety

Certificate
Standards
Markings
70164066
C22.2 No 0-10, C22.2 No 0.4-M2004, C22.2 No 94-M1991, C22.2 No 142 – M1987, C22.2 No 157-M1992, CAN/CSA E60079-0:07, CAN/CSA E60079-11:02, UL 50-11th Ed, UL 508-17th Ed, UL 913-7th Ed, UL 60079-0: 2005, UL 60079-11: 2002
Preamplifier assembly: Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups ABCD; ambient temperature rating –20 °C to +60 °C; Ex ia IIC; T6: Class I, Zone 0, AEx ia IIC ; T6
Sensor apparatus with preamplifier: Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups ABCD; ambient temperature rating –20 °C to +60 °C; Ex ia IIC; T6: Class I, Zone 0, AEx ia IIC ; T6
Sensor apparatus: Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups ABCD; Ex ia IIC; T6; ambient temperature rating –20 °C to +60 °C: (simple apparatus)

4.6 Europe

4.6.1 ATEX Intrinsic Safety

Certificate
Standards
Markings
32 Emerson.com/Rosemount
Baseefa10ATEX0156
EN 60079-0: 2012+A11: 2013, EN 60079-11: 2012
pH/ORP sensors with no preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +60 °C)
November 2020 Quick Start Guide
pH sensors with integral smart preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +60 °C)
ORP sensors with integral standard preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +80 °C) Ex ia IIC T5 Ga (–20 °C to +40 °C) pH sensors with integral standard preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +80 °C) Ex ia IIC T5 Ga (–20 °C to +40 °C)
Specific Conditions for Safe Use (X):
1. All pH/ORP sensor models with a plastic enclosure or exposed plastic parts may provide an electrostatic ignition hazard and must only be cleaned with a damp cloth to avoid the danger of ignition due to build-up of electrostatic charge.
2. All pH/ORP sensor models with a metallic enclosure may provide a risk of ignition by impact or friction. Care should be taken during installation to protect the sensor from the risk.
3. External connections to the sensor must be suitably terminated and provide a degree of protection of at least IP20.
4. All pH/ORP sensor models are intended to be in contact with the process fluid and may not meet the 500V r.m.s. test to earth. This must be taken into consideration at installation.
4.7

International

4.7.1 IECEx Intrinsic Safety

Certificate
Standards
Markings
Quick Start Guide 33
IECEx BAS 10.0083X
IEC 60079-0: 2011, IEC 60079-11: 2011
pH/ORP sensors with no preamplifier fitted Ex ia IIC T4 Ga (–20 °C to +60 °C)
pH sensors with integral smart preamplifier fitted Ex ia IIC T4 Ga (–20 °C to +60 °C)
ORP sensors with integral standard preamplifier fitted Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C) pH sensors with integral standard preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +80 °C) Ex ia IIC T5 Ga (–20 °C to +40 °C)
Quick Start Guide November 2020
Specific Conditions for Safe Use (X):
1. All pH/ORP sensor models with a plastic enclosure or exposed plastic parts may provide an electrostatic ignition hazard and must only be cleaned with a damp cloth to avoid the danger of ignition due to build-up of electrostatic charge.
2. All pH/ORP sensor models with a metallic enclosure may provide a risk of ignition by impact or friction. Care should be taken during installation to protect the sensor from the risk.
3. External connections to the sensor must be suitably terminated and provide a degree of protection of at least IP20.
4. All pH/ORP sensor models are intended to be in contact with the process fluid and may not meet the 500V r.m.s. test to earth. This must be taken into consideration at installation.

4.8 China

4.8.1 Nepsi Intrinsic Safety

4.9
Certificate
Standards
Markings
Specific Conditions for Safe Use (X):
1. It is strictly forbidden to rub the plastic shell parts of the product to
2. When the product shell contains light metals, it should be prevented

Technical Regulations Customs Union (EAC)

GYB19.1035X
GB 3836.1-2010, GB 3836.4-2010, GB 3836.20-2010
Ex ia II C T4 Ga (–20 °C to +60 °C)
prevent the risk of static ignition.
in a zone 0 environment.

4.9.1 EAC Intrinsic Safety

Certificate
Markings
TC RU C-US .MIO62. B.06011
pH/ORP sensors with no preamplifier fitted Ex ia IIC T4 Ga (–20 °C to +60 °C)
pH sensors with integral smart preamplifier fitted Ex ia IIC T4 Ga (–20 °C to +60 °C)
ORP sensors with integral standard preamplifier fitted Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C) pH sensors with integral standard preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +80 °C)
34 Emerson.com/Rosemount
November 2020 Quick Start Guide
Ex ia IIC T5 Ga (–20 °C to +40 °C)
Specific Condition for Safe Use (X):
See certificate for special conditions.
Quick Start Guide 35
Quick Start Guide November 2020

5 Declaration of Conformity

36 Emerson.com/Rosemount
November 2020 Quick Start Guide
Quick Start Guide 37
Quick Start Guide November 2020

6 China RoHS tables

Figure 6-1: Rosemount 3300HT/3300HTVP China RoHS Table
Figure 6-2: Rosemount 3400HT/3400HTVP China RoHS Table
38 Emerson.com/Rosemount
November 2020 Quick Start Guide
7 Intrinsically safe sensor installation drawing -
FM
Quick Start Guide 39
*00825-0100-3300*
EMERSON AUTOMATION SOLUTIONS
6021 Innovation Blvd. Shakopee, MN 55379
+1 866 347 3427 +1 952 949 7001 RMTNA.RCCPO@Emerson.com
00825-0100-3300, Rev. AA
Quick Start Guide
November 2020
NORTH AMERICA
Emerson Automation Solutions 8200 Market Blvd Chanhassen, MN 55317
Toll Free +1 800 999 9307 F +1 952 949 7001 RMTNA.RCCPO@Emerson.com
MIDDLE EAST AND AFRICA
Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai, United Arab Emirates, P.O. Box 17033
+971 4 811 8100 +971 4 886 5465 RMTNA.RCCPO@Emerson.com
LinkedIn.com/company/Emerson-
Automation-Solutions
Twitter.com/rosemount_news
Facebook.com/Rosemount
Youtube.com/RosemountMeasurement
EUROPE
Emerson Automation Solutions Neuhofstrasse 19a PO Box 1046 CH-6340 Baar Switzerland
+41 (0) 41 768 6111 +41 (0) 41 768 6300 RMTNA.RCCPO@Emerson.com
ASIA-PACIFIC
Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Republic of Singapore
+65 6 777 8211 +65 6 777 0947 RMTNA.RCCPO@Emerson.com
©
2020 Emerson. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.
Loading...