Emerson designs, manufactures, and tests its products to meet many national and international
standards. Because these instruments are sophisticated technical products, you must properly install,
use, and maintain them to ensure they continue to operate within their normal specifications. You
must adhere to the following instructions and integrate them into your safety program when
installing, using, and maintaining Emerson's Rosemount™ products. Failure to follow the proper
instructions may cause any one of the following situations to occur: loss of life, personal injury,
property damage, damage to this instrument, and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Emerson representative for
clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the
product.
• Install equipment as specified in the installation instructions of the appropriate Quick Start Guide
and per applicable local and national codes. Connect all products to the proper electrical and
pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and
maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts
specified by Emerson. Unauthorized parts and procedures can affect the product's performance,
place the safe operation of your process at risk, and may result in fire, electrical hazards, or
improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified people, to prevent electrical shock and personal
injury.
Note
The information contained in this document is subject to change without notice.
WARNING
Pressure and temperature
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig
and the process temperature is lowered to a safe level.
Do not insert or retract retractable sensors when process pressures are in excess of 64 psig (542
kPa) for option 21 or 35 psig (343 kPa) for option 25.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end
users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your
system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for
all systems used within the facility.
2Emerson.com/Rosemount
November 2020Quick Start Guide
CAUTION
Sensor/process application compatibility
The wetted sensor materials may not be compatible with process composition and operating
conditions.
Application compatibility is entirely the operator's responsibility.
M. 1½-in. (38.1 mm) male pipe thread (MPT) close nipple
N. 1½-in. x 1-in. (38.1 x 25.4 mm) FPT reducing coupling
O. Ball valve (PN 23240-00) optional
Table 1-1: Rosemount 3400HT and 3400HTVP Sensor Dimensions with
Optional Ball Valve PN 23240-00
OptionDimension A
(in./mm)
2121.6/54914.0/35523.1/587
2536.1/91728.5/72437.6/955
Dimension B
(in./mm)
Dimension C
(in./mm)
Quick Start Guide9
Quick Start GuideNovember 2020
Figure 1-4: Rosemount 3400HT and 3400HTVP Sensor Dimensions with
Optional Ball Valve PN 23765-00
A. Dimension (see Table 1-2)
B. Dimension (see Table 1-2)
C. Housing titanium
D. Retraction stop collar
E. Electrode housing
F. pH electrode
G. 1¼-in. (38.1 mm) MPT close nipple
H. Ball valve
I. Packing adapter
J. Knurl retainer
K. Ball valve kit PN 23765-00
10Emerson.com/Rosemount
November 2020Quick Start Guide
Table 1-2: Rosemount 3400HT and 3400HTVP Sensor Dimensions with
Optional Ball Valve PN 23765-00
OptionA: in./mmB: in./mm
2121.6/54912.2/310
2536.1/91726.7/678
Rosemount 3300HT and 3300HTVP flow through and insertion
installation
1½-in. (38.1 mm) pipe tee (PN 2002011) with 1-in. (25.4 mm) threaded
connections.
Note
Mount the sensor at least 10 degrees above the horizon.
Figure 1-5: Straight Flow
A. 1½-in. (38.1 mm) pipe tee with 1-in. (25.4 mm) threaded connections:
PN 2002011
Quick Start Guide11
Quick Start GuideNovember 2020
Figure 1-6: Angle Flow
A. 1½-in. (38.1 mm) pipe tee with 1-in. (25.4 mm) threaded connections:
PN 2002011
B. Flow
Figure 1-7: Pipe Y Installation
A. 1½-in. x 1-in. (38.1 x 25.4 mm) reducing bushing
B. 1½-in. (38.1 mm) pipe "Y"
12Emerson.com/Rosemount
November 2020Quick Start Guide
Figure 1-8: Rosemount 3400HT and 3400HTVP Mounting Details Retraction Version
A. Weldalet (1½-in. [38.1 mm] FPT)
B. Electrode housing tip
C. Electrode
D. Flow
E. Pipe "Y"
F. Pipe tee
G. Flow
Quick Start Guide13
Quick Start GuideNovember 2020
Figure 1-9: Rosemount 3300 and 3500VP Flow through Installation
A. 1½-in. (38.1 mm) schedule 80 CPVC tee with 1-in. (25.4 mm) female
national pipe thread (FNPT) connections (code 16) straight flow shown
B. Always mount sensor at least 10 degrees above horizontal.
C. 1½-in. (38.1 mm) pipe "Y"
D. 1½-in. x 1-in. (38.1 x 25.4 mm) reducing bushing
E. Flow
F. 1½-in. (38.1 mm) schedule 80 CPVC with view or cleaning port
G. Angle flow shown
14Emerson.com/Rosemount
November 2020Quick Start Guide
Figure 1-10: Type 4 Junction Box
A. PN 23550-00 for point-to-point wiring used with extension cable. Cable
for pH, oxidation reduction potential (ORP), conductivity, or
amperometric sensors. PN 23555-00 with preamplifier for pH and ORP
sensors only.
B. PN 23820-00 carbon steel with epoxy coating. PN 23820-02 316
stainless steel
C. Flexible conduit is required.
D. 1-in. (25.4 mm) pipe by others
E. 1-in. (25.4 mm) female national pipe thread (FNPT) CPVC union: PN
9320057
F. 1-in. (25.4 mm) male national pipe thread (MNPT) connector
G. Any 1-in. (25.4 mm) threaded sensor
H. 2¾-in. (69.9 mm) O.D.
Be certain the system pressure at the sensor is below 64 psig (542 kPa
[abs]) before retracting the sensor.
Wear a face shield.
Make sure your footing is stable.
Procedure
1. Push in the sensor end or the top of the junction box and slowly
loosen the hex nut of the process end male connector.
CAUTION
Do not remove nut at this time.
2. When the hex nut is loose enough, slowly ease the sensor back
completely until it reaches the retraction stop collar.
Quick Start Guide17
Quick Start GuideNovember 2020
CAUTION
Failure to withdraw the sensor completely may result in damage to
the sensor when the valve is closed.
3. Close the ball valve slowly.
If there is resistance, the valve may be hitting the sensor. Double
check that the sensor has been retracted to the retraction stop collar.
WARNING
Leakage from the male connector threads may indicate that the
male connector is still under pressure.
Leakage through a partially open ball valve could be hazardous;
however, with the ball valve closed, some residual process fluid may
leak from the connector's pipe threads.
Before removing the sensor from the ball valve, be absolutely certain
that the ball valve is fully closed.
4. Unthread the male body connector completely from the reducing
coupling and remove the sensor for servicing.
Note
If the male connector leaks during insertion or retraction, replace the
O-ring (PN 25394-01) in the male connector.
1.4
18Emerson.com/Rosemount
Wiring
For additional wiring information on this product, including sensor
combinations not shown here, please refer to Emerson.com/
RosemountLiquidAnalysisWiring.
November 2020Quick Start Guide
Figure 1-12: Rosemount PERpH-X Sensors, Preamplifier in Analyzer
For Rosemount 3300HT, 3300HTVP, 3400HT, and 3400HTVP
Non-preamplifier sensors
Blue cable
Table 1-3: Rosemount PERpH-X Sensors, Preamplifier in Analyzer
Wire colorRosemount 56, 1056,
1057, and 1066 terminal
number
N/AInner drain (no
connection)
N/AN/A1N/A
GreenConnect to instrument
ground or as shown
WhiteResistance temperature
device (RTD) return
White/redRTD sense42
RedRTD in53
White/grayReference shield64
GrayReference in75
BlueSolution ground86
ClearmV/pH shield97
OrangemV/pH in108
N/A-5 Vdc119
N/A+5 Vdc1210
Rosemount 54e and 5081
terminal number
N/AN/A
2N/A
31
Rosemount
XMT terminal
number
Quick Start Guide19
Quick Start GuideNovember 2020
Figure 1-13: Rosemount PERpH-X Sensors, Preamplifier in Junction Box
For Rosemount 3300HT, 3300HTVP, 3400HT, and 3400HTVP
Non-preamplifier sensors
Blue cable
Extension cable PN 23646-01 or 9200273 shown.
Can also use extension cable PN 9200348; extend color scheme from sensor
or VP8 cable instead of using shown wiring scheme.
Table 1-4: Rosemount PERpH-X Sensors, Preamplifier in Junction Box
Wire
color
N/AN/AN/AInner drain (no
N/A1N/AN/A1N/A
Green2Inner drain Connect to
White3WhiteRTD return31
White/red 4White/red RTD sense42
Red5RedRTD in53
White/
gray
Gray7GrayReference in75
Blue8BlueSolution ground86
20Emerson.com/Rosemount
Junction
box
terminal
number
6White/
Extension
cable
color
gray
Rosemount 56,
1056, 1057, and
1066 terminal
number
connection)
instrument
ground as shown
Reference shield 64
Rosemount 54e
and 5081
terminal
number
N/AN/A
2N/A
Rosemount
XMT terminal
number
November 2020Quick Start Guide
Table 1-4: Rosemount PERpH-X Sensors, Preamplifier in Junction Box (continued)
Wire
color
Clear9White/
Orange10BlackmV/pH in108
N/A11Brown-5 Vdc119
N/A12Green+5 Vdc1210
Junction
box
terminal
number
Extension
cable
color
black
Rosemount 56,
1056, 1057, and
1066 terminal
number
mV/pH shield97
Rosemount 54e
and 5081
terminal
number
Rosemount
XMT terminal
number
Quick Start Guide21
Quick Start GuideNovember 2020
Figure 1-14: PERpH-X Sensors, Preamplifier in Sensor
For Rosemount 3300HTVP-70 and 3400HTVP-70
Preamplifier in sensor
Blue cable
Table 1-5: Rosemount PERpH-X Sensors, Preamplifier in Sensor
Wire colorRosemount 56,
1056, 1057, and
1066 terminal
number
N/AN/A1N/A
GreenConnect to
instrument ground or
as shown
WhiteRTD return31
White/ redRTD sense42
RedRTD in53
N/AReference shield64
GrayReference in75
BlueSolution ground86
ClearmV/pH shield97
OrangemV/pH in108
White/gray-5 Vdc119
Inner drain+5 Vdc1210
Rosemount 54e and
5081 terminal
number
2N/A
Rosemount XMT
terminal number
22Emerson.com/Rosemount
November 2020Quick Start Guide
Figure 1-15: Rosemount PERpH-X Sensors, Preamplifier in Sensor,
Junction Box Used for Extending Cable
For Rosemount 3300HTVP-70 and 3400HTVP-70
Preamplifier in sensor
Blue cable
Extension cable PN 23646-01 or 9200273 shown.
Can also use extension cable PN 9200348: white/gray is replacement wire
color for brown wire.
Table 1-6: Rosemount PERpH-X Sensors, Preamplifier in Sensor, Junction Box
Used for Extending Cable
Wire
color
N/A1N/AN/A1N/A
Green2Inner drain.
White3WhiteRTD return31
White/red4White/redRTD sense42
Red5RedRTD in53
N/A6White/gray (no
Gray7GrayReference in75
Quick Start Guide23
Junction
box
terminal
number
Extension
cable color
Connect to
instrument
ground or as
shown.
connection)
Rosemount 56,
1056, 1057,
and 1066
terminal
number
N/A2N/A
Reference
shield
Rosemount
54e and 5081
terminal
number
64
Rosemount
XMT terminal
number
Quick Start GuideNovember 2020
Table 1-6: Rosemount PERpH-X Sensors, Preamplifier in Sensor, Junction Box
Used for Extending Cable (continued)
Wire
color
Blue8BlueSolution ground 86
Clear
(orange
shield)
Orange10BlackpH in108
White/
gray
ID (inner
drain)
Junction
box
terminal
number
9White/blackpH shield97
11Brown-5 Vdc119
12Green+5 Vdc1210
Extension
cable color
Rosemount 56,
1056, 1057,
and 1066
terminal
number
Rosemount
54e and 5081
terminal
number
Rosemount
XMT terminal
number
24Emerson.com/Rosemount
November 2020Quick Start Guide
2Start up and calibrate
2.1Prepare electrode
Procedure
1. Remove electrode from shipping container.
2. Remove the protective boot covering the electrode bulb.
3. Rinse away salt film with clean water; then shake the electrode so
that the internal solution fills the bulb, thus removing any air trapped
there.
WARNING
The buffer in the protective boot may cause skin or eye irritation.
2.2Calibrate with two point pH buffer solution
Select two stable buffer solutions, preferably pH 4.0 and pH 7.0. You can use
buffers other than pH 4.0 and pH 7.0 as long as the pH values are least two
pH units apart.
Note
A pH 7 buffer solution reads an mV value of approximately zero, and pH
buffers read approximately ±59.1 mV for each pH unit above or below pH 7.
Check the pH buffer manufacturer specifications for milivolt values at
various temperatures, since temperature may affect the actual value of the
pH buffer solution mV/pH value.
Procedure
1. Immerse the sensor in buffer solution.
Make sure the buffer solution contacts the metal housing of the
sensor, which acts as the solution ground contact.
2. Allow sensor to equilibrate to the buffer temperature and wait for the
reading to stabilize.
The transmitter can now acknowledge the value of the buffer.
3. Rinse the buffer solution off of the sensor with distilled or deionized
water.
4. Repeat steps Step 1 through Step 3 using the second buffer solution.
The theoretical slope value, according to the Nernst equation for calculating
pH, is approximately 59.17 mV/pH. Over time the sensor will age, both in
the process and in storage, and this will result in reduced slope values. To
Quick Start Guide25
Quick Start GuideNovember 2020
ensure accurate readings, Emerson recommends replacing the electrode
when the slope value falls below 47 to 49 mV/pH.
2.3Standardize pH sensor
For maximum accuracy, you can standardize the sensor in-line or with a
process grab sample after performing a buffer calibration and conditioning
the sensor to the process. Standardization accounts for the sensor junction
potential and other interferences. Standardization does not change the
sensor's slope, but simply adjusts the transmitter's reading to match that of
a known process pH.
26Emerson.com/Rosemount
November 2020Quick Start Guide
3Maintenance
3.1Clean electrodes
Electrodes should respond rapidly. Sluggishness, offsets, and erratic
readings are indicators that you may need to clean or replace the electrodes.
Procedure
1. To remove oil deposits, clean the electrodes with a mild, nonabrasive detergent.
2. To remove scale deposits, soak electrodes for 30 to 60 minutes in a
five percent hydrochloric acid solution.
3.2Replace reference junction and recharge sensor
electrolyte
The reference junction and reference fill gelled solution are replaceable to
facilitate longer sensor life. Over time, the electrolyte depletes and the
junction may become plugged or contaminated.
Prerequisites
Use the junction replacement kit and replacement fill gel.
Procedure
1. Remove the junction cap by turning it counter clockwise.
2. Remove the liquid junction by pulling the junction straight out.
WARNING
The reference electrolyte may cause skin or eye irritation
Avoid contact with skin and eyes.
3. Rinse with water to remove the old reference fill gel.
4. Fill the reference fill chamber with the reference fill gel using the
syringe, and remove any air bubbles. Top off the reference fill
chamber until it is completely filled.
5. Replace the junction O-ring and liquid junction by sliding them over
the glass electrode.
Excess reference gel should flow out.
6. Replace junction cap by turning clockwise.
Hand tighten the junction cap only; do not use pliers to tighten the
cap.
Quick Start Guide27
Quick Start GuideNovember 2020
7. Buffer check and calibrate the sensor as described in Calibrate with
two point pH buffer solution.
28Emerson.com/Rosemount
November 2020Quick Start Guide
4Rosemount pH/ORP sensor(s) product
certifications
Rev 0.5
4.1European directive information
A copy of the EU Declaration of Conformity can be found at the end of the
Quick Start Guide. The most recent revision of the EU Declaration of
Conformity can be found at Emerson.com/Rosemount.
4.2Ordinary location certification
As standard, the transmitter has been examined and tested to determine
that the design meets the basic electrical, mechanical, and fire protection
requirements by a nationally recognized test laboratory (NRTL) as accredited
by the Federal Occupational Safety and Health Administration (OSHA).
4.3Installing equipment in North America
The US National Electrical Code® (NEC) and the Canadian Electrical Code
(CEC) permit the use of Division marked equipment in Zones and Zone
marked equipment in Divisions. The markings must be suitable for the area
classification, gas, and temperature class. This information is clearly defined
in the respective codes.
4.4
USA
4.4.1FM Intrinsic Safety
Certificate
Standards
Markings
Specific Conditions for Safe Use (X):
1. Sensors with Model 1700702 preamplifier:
Quick Start Guide29
FM17US0198X
FM Class 3600:1998, FM Class 3610:2010, FM Class 3611:
2004, FM Class 3810: 2005
IS/I,II,III/1/ABCDEFG/T6 Ta = –20 °C to 60 °C
I/0/AEx ia IIC/T6 Ta = –20 °C to 60 °C
NI/I/2/ABCD/T6 Ta = –20 °C to 60 °C
S/II,III/2/EFG/T6 Ta = –20 °C to 60 °C
a. Model 385+-a-b-c. Triple junction pH/ORP sensor
b. Model 389-a-b-c-d-e. pH/ORP sensor
c. Model 389VP-a-b-c-d. pH/ORP sensor
d. Model 396VP-a-b-c-d. Submersion/insertion pH/ORP sensor
Quick Start GuideNovember 2020
e. Model 396P-a-b-c-d-e. Submersion/insertion pH/ORP sensor
f. Model 396PVP-a-b-c-d-e. Submersion/insertion pH/ORP
sensor
g. Model 396RVP-a-b-c-d-e. Retraction/submersion/insertion
pH/ORP sensor
h. Model 398RVP-a-b-c-d-e-f. pH/ORP sensor
i. Model 3200HP-00. High purity water pH sensor
j. Model 3300HTVP-a-b-c-d. High performance pH and ORP
sensor
k. Model 3400HTVP-a-b-c-d-e. High performance pH and ORP
sensor
l. 3500P-a-b-c-d-e-f. High performance pH and ORP sensor
m. 3500VP-a-b-c-d-e-f. High performance pH and ORP sensor
n. Model 3900-a-b-c. General purpose pH/ORP sensor
o. Model 3900VP-a-b. General purpose pH/ORP sensor
The polymeric surface of all the apparatus listed above may store
electrostatic charge and become a source of ignition. Clean surface
should only be done with a damp cloth.
2. Sensors without Model 1700702 preamplifier (simple apparatus):
a. Model 385-a-b-c-d-e. Retractable pH/ORP sensor
b. Model 385+-a-b-c Triple junction pH/ORP sensor
c. Model 389-a-b-c-d-e. pH/ORP sensor
d. Model 389VP-a-b-c. pH/ORP sensor
e. Model 396-a-b-c. Submersion/insertion pH sensor
f. Model 396VP-a-b. Submersion/insertion pH sensor
g. Model 396P-a-b-c-d-e. Submersion/insertion pH/ORP sensor
h. Model 396PVP-a-b-c-d. Submersion/insertion pH/ORP sensor
i. Model 396R-a-b-c-d-e. Retraction/submersion/insertion
pH/ORP sensor
j. Model 396RVP-a-b-c-d. Retraction/submersion/insertion
pH/ORP sensor
k. Model 397-a-b-c-d-e. pH sensor
l. Model 398-a-b-c-d-e. pH/ORP sensor
30Emerson.com/Rosemount
November 2020Quick Start Guide
m. Model 398VP-a-b-c. pH/ORP sensor
n. Model 398R-a-b-c-d-e-f. pH/ORP sensor
o. Model 398RVP-a-b-c-d-e-f. pH/ORP sensor
p. Model 3200HP-00. High purity water pH sensor
q. Model 3300HT-a-b-c-d. High performance pH and ORP sensor
r. Model 3300HTVP-a-b-c-d. High performance pH and ORP
sensor
s. Model 3400HT-a-b-c-d-e-f. High performance pH and ORP
sensor
t. Model 3400HTVP-a-b-c-d-e-f. High performance pH and ORP
sensor
u. Model 3500P-a-b-c-d-e-f. High performance pH and ORP
sensor
v. Model 3500VP-a-b-c-d-e-f. High performance pH and ORP
sensor
w. Model 3800-a. Autoclaveable and steam sterilizable pH
sensors
x. Model 3800VP-a. Autoclaveable and steam sterilizable pH
sensors
y. Model 3900-a-b-c. General purpose pH/ORP sensor
z. Model 3900VP-a-b. General purpose pH/ORP sensor
The polymeric surface of all the apparatus listed above may store
electrostatic charge and become a source of ignition. Clean surface
should only be done with a damp cloth.
4.4.2CSA Intrinsic Safety
Certificate
Standards
Markings
Quick Start Guide31
70164066
C22.2 No 0-10, C22.2 No 0.4-M2004, C22.2 No 94-M1991,
C22.2 No 142 – M1987, C22.2 No 157-M1992, CAN/CSA
E60079-0:07, CAN/CSA E60079-11:02, UL 50-11th Ed, UL
508-17th Ed, UL 913-7th Ed, UL 60079-0: 2005, UL 60079-11:
2002
Preamplifier assembly:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups
EFG; Class III; Class I, Division 2, Groups ABCD; ambient
temperature rating –20 °C to +60 °C; Ex ia IIC; T6: Class I, Zone
0, AEx ia IIC ; T6
Quick Start GuideNovember 2020
Sensor apparatus with preamplifier:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups
EFG; Class III; Class I, Division 2, Groups ABCD; ambient
temperature rating –20 °C to +60 °C; Ex ia IIC; T6: Class I, Zone
0, AEx ia IIC ; T6
Sensor apparatus:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups
EFG; Class III; Class I, Division 2, Groups ABCD; Ex ia IIC; T6;
ambient temperature rating –20 °C to +60 °C: (simple
apparatus)
4.5Canada
4.5.1CSA Intrinsic Safety
Certificate
Standards
Markings
70164066
C22.2 No 0-10, C22.2 No 0.4-M2004, C22.2 No 94-M1991,
C22.2 No 142 – M1987, C22.2 No 157-M1992, CAN/CSA
E60079-0:07, CAN/CSA E60079-11:02, UL 50-11th Ed, UL
508-17th Ed, UL 913-7th Ed, UL 60079-0: 2005, UL 60079-11:
2002
Preamplifier assembly:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups
EFG; Class III; Class I, Division 2, Groups ABCD; ambient
temperature rating –20 °C to +60 °C; Ex ia IIC; T6: Class I, Zone
0, AEx ia IIC ; T6
Sensor apparatus with preamplifier:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups
EFG; Class III; Class I, Division 2, Groups ABCD; ambient
temperature rating –20 °C to +60 °C; Ex ia IIC; T6: Class I, Zone
0, AEx ia IIC ; T6
Sensor apparatus:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups
EFG; Class III; Class I, Division 2, Groups ABCD; Ex ia IIC; T6;
ambient temperature rating –20 °C to +60 °C: (simple
apparatus)
4.6Europe
4.6.1ATEX Intrinsic Safety
Certificate
Standards
Markings
32Emerson.com/Rosemount
Baseefa10ATEX0156
EN 60079-0: 2012+A11: 2013, EN 60079-11: 2012
pH/ORP sensors with no preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +60 °C)
November 2020Quick Start Guide
pH sensors with integral smart preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +60 °C)
ORP sensors with integral standard preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
pH sensors with integral standard preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
Specific Conditions for Safe Use (X):
1. All pH/ORP sensor models with a plastic enclosure or exposed plastic
parts may provide an electrostatic ignition hazard and must only be
cleaned with a damp cloth to avoid the danger of ignition due to
build-up of electrostatic charge.
2. All pH/ORP sensor models with a metallic enclosure may provide a
risk of ignition by impact or friction. Care should be taken during
installation to protect the sensor from the risk.
3. External connections to the sensor must be suitably terminated and
provide a degree of protection of at least IP20.
4. All pH/ORP sensor models are intended to be in contact with the
process fluid and may not meet the 500V r.m.s. test to earth. This
must be taken into consideration at installation.
4.7
International
4.7.1IECEx Intrinsic Safety
Certificate
Standards
Markings
Quick Start Guide33
IECEx BAS 10.0083X
IEC 60079-0: 2011, IEC 60079-11: 2011
pH/ORP sensors with no preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +60 °C)
pH sensors with integral smart preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +60 °C)
ORP sensors with integral standard preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
pH sensors with integral standard preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
Quick Start GuideNovember 2020
Specific Conditions for Safe Use (X):
1. All pH/ORP sensor models with a plastic enclosure or exposed plastic
parts may provide an electrostatic ignition hazard and must only be
cleaned with a damp cloth to avoid the danger of ignition due to
build-up of electrostatic charge.
2. All pH/ORP sensor models with a metallic enclosure may provide a
risk of ignition by impact or friction. Care should be taken during
installation to protect the sensor from the risk.
3. External connections to the sensor must be suitably terminated and
provide a degree of protection of at least IP20.
4. All pH/ORP sensor models are intended to be in contact with the
process fluid and may not meet the 500V r.m.s. test to earth. This
must be taken into consideration at installation.
4.8China
4.8.1Nepsi Intrinsic Safety
4.9
Certificate
Standards
Markings
Specific Conditions for Safe Use (X):
1. It is strictly forbidden to rub the plastic shell parts of the product to
2. When the product shell contains light metals, it should be prevented
Technical Regulations Customs Union (EAC)
GYB19.1035X
GB 3836.1-2010, GB 3836.4-2010, GB 3836.20-2010
Ex ia II C T4 Ga (–20 °C to +60 °C)
prevent the risk of static ignition.
in a zone 0 environment.
4.9.1EAC Intrinsic Safety
Certificate
Markings
TC RU C-US .MIO62. B.06011
pH/ORP sensors with no preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +60 °C)
pH sensors with integral smart preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +60 °C)
ORP sensors with integral standard preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
pH sensors with integral standard preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +80 °C)
34Emerson.com/Rosemount
November 2020Quick Start Guide
Ex ia IIC T5 Ga (–20 °C to +40 °C)
Specific Condition for Safe Use (X):
See certificate for special conditions.
Quick Start Guide35
Quick Start GuideNovember 2020
5Declaration of Conformity
36Emerson.com/Rosemount
November 2020Quick Start Guide
Quick Start Guide37
Quick Start GuideNovember 2020
6China RoHS tables
Figure 6-1: Rosemount 3300HT/3300HTVP China RoHS Table
Figure 6-2: Rosemount 3400HT/3400HTVP China RoHS Table
The Emerson logo is a trademark and service
mark of Emerson Electric Co. Rosemount is a
mark of one of the Emerson family of companies.
All other marks are the property of their
respective owners.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.