R&S®RT-ZD02
Differential Probe
User Manual
*5800039702*
1333082102
Version 01
User Manual
Test & Measurement

RT-ZD02
Differential Probe 10 : 1
RT-ZD02
200 MHz bandwidth
high input resistance of 1 MΩ
low input capacitance of 3.5 pF
high dynamic range of ±20 V differential and ±60 V common
universal power supply using battery or USB cable
typical applications:
– differential signals like CAN Bus or LVDS
– DC-DC converter
General Information
The RT-ZD02 200 MHz Differential probe is best suited for measurements of high-speed differential
signals. These signals are commonly used in digital and power applications and especially at serial buses
like CAN. The RT-ZD02 can be powered by USB or battery and is compatible with all oscilloscopes with
50 Ω BNC inputs. Scopes without 50 Ω input need a 50 Ω feed through BNC adapter. The probe has a
high input resistance of 1 MΩ combined with 7 pF input capacitance. This secures a low loading of the
device under test. The attenuation of the RT-ZD02 is 10 : 1.
2
Subject to change without notice

SPECIFICATIONS
all data valid at 23 °C after 30 minutes warm up
Specifications
Electrical specications
Bandwidth (-3 dB): 200 MHz, typ.
Attenuation Ratio: 10 : 1
Probe Risetime (10%-90%): 1.75 ns
Gain Accuracy: ±1%, typ.
Absolute Maximum Rated Input Voltage
(each side to ground): ±60 V
Maximum Differential
Input Voltage (DC+AC Peak): ±20 V
Maximum Common
Mode Input Voltage ±60 V
Input Resistance / Input capacity
each side to ground: 500 kΩ, 7 pF
between inputs: 1 MΩ, 3.5 pF
Output Voltage Swing: ±2 V (driving 50 Ω scope input)
Offset: ±2 mV, typ.
CMRR: -80 dB at 60 Hz, typ.
-50 dB at 10 MHz, typ.
Power Requirements: one 9 V battery or USB power
adapter (5 V to 9 V, 200 mA)
1
Approx. Battery Life: 7.5 hours (alkaline battery)
Battery/Power Cord: The supplied voltage must be
less than 12 V and greater than
4.5 V or else the probe could be
damaged
Mechanical specications
Approximate Weight: 170 g
(not including battery and
accessories)
BNC Cable Length: 120 cm
Length of Input Leads: 15 cm
Dimensiones (LxWxH): 111 x 22 x 14 mm
Environmental specications
Temperature range
operating: 5 °C to +40 °C
non-operating: -20 °C to +70 °C
Altitude operating: 3,000 m
non-operating: 15,300 m
Humidity: 25 - 85% (indoor use only)
2
Set of RT-ZD02 probe consists of:
1 RT-ZD02 200 MHz differential probe 1 piece
2 Hook Clip, red 1 piece
Hook Clip, black 1 piece
3 Alligator Clip, red 1 piece
Alligator Clip, black 1 piece
3
4 USB Power Cord (2 m) 1 piece
Specications
5 9 V Battery 1 piece
Manual 1 piece
4
5
Subject to change without notice
3