Rohde&Schwarz RTO-K22, RTO6-K22, RTP-K22, RTO-K23, RTO6-K23 Ethernet Compliance Test

...
R&S®RTO, R&S®RTO6, R&S®RTP Ethernet Compliance Tests User Manual
(=AÖ22)
This manual describes the Ethernet compliance test procedures with the following options:
10/100/1000BASE-T: R&S®RTO-K22 (1329.7460.02), R&S®RTO6-K22 (1801.6929.02), R&S®RTP-K22 (1337.8691.02)
2.5/5/10GBASE-T: R&S®RTO-K23 (1329.7477.02), R&S®RTO6-K23 (1801.6935.02), R&S®RTP-K23 (1337.8704.02)
100BASE-T1: R&S®RTO-K24 (1329.7483.02), R&S®RTO6-K24 (1801.6941.02), R&S®RTP-K24 (1800.6531.02)
1000BASE-T1: R&S®RTO-K87 (1337.8591.02), R&S®RTO6-K87 (1801.6970.02), R&S®RTP-K87 (1800.6554.02)
MGBASE-T1: R&S®RTO-K88 (1801.4526.02), R&S®RTO6-K88 (1801.7890.02), R&S®RTP-K88(1800.6725.02)
10BASE-T1: R&S®RTO-K89 (1801.4510.02), R&S®RTO6-K89 (1801.6987.02), R&S®RTP-K89 (1800.6719.02)
The tests require the R&S RT‑ZF2 Ethernet Test Fixture (1317.5522.02 ) and the R&S ScopeSuite soft­ware. The energy-efficient Ethernet (EEE) 100/1000BASE-T tests require the R&S RT-ZF5 Ethernet probing fix­ture (1333.0896.02). The 10BASE-Te tests require the R&S RT-ZF4 10BASE-Te fixture (1333.0880.02).
© 2021 Rohde & Schwarz GmbH & Co. KG
Mühldorfstr. 15, 81671 München, Germany
Phone: +49 89 41 29 - 0
Email: info@rohde-schwarz.com
Internet: www.rohde-schwarz.com
Subject to change – data without tolerance limits is not binding.
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
Trade names are trademarks of the owners.
1317.7202.02 | Version 16 | R&S®RTO, R&S®RTO6, R&S®RTP
Throughout this manual, products from Rohde & Schwarz are indicated without the ® symbol , e.g. R&S®ScopeSuite is indicated as
R&S ScopeSuite.
R&S®RTO, R&S®RTO6, R&S®RTP

Contents

1 R&S ScopeSuite Overview....................................................................7
2 Preparing the Measurements................................................................9
2.1 Test Equipment..............................................................................................................9
2.2 Installing Software and License.................................................................................11
2.3 Setting Up the Network...............................................................................................11
2.4 Starting the R&S ScopeSuite..................................................................................... 12
2.5 Connecting the R&S RTO/RTO6/RTP........................................................................ 12
2.6 Connecting the Arbitrary Waveform Generator....................................................... 14
2.7 Connecting the Vector Network Analyzer.................................................................15
2.8 Connecting the Spectrum Analyzer.......................................................................... 17
Contents
2.9 Report Configuration.................................................................................................. 18
3 Performing Tests..................................................................................20
3.1 Starting a Test Session...............................................................................................20
3.2 Configuring the Test................................................................................................... 21
3.3 Getting Test Results....................................................................................................23
4 1000BASE-T Tests................................................................................25
4.1 Starting 1000BASE-T Tests........................................................................................ 25
4.2 Test Configuration for Ethernet 1000BASE-T...........................................................25
4.3 1000BASE-T.................................................................................................................27
4.4 1000 BASE-T EEE Tests..............................................................................................60
5 100BASE-TX Tests............................................................................... 79
5.1 Starting 100BASE-TX Tests........................................................................................79
5.2 Test Configuration for Ethernet 100BASE-TX.......................................................... 79
5.3 100BASE-TX................................................................................................................ 81
5.4 100BASE-TX EEE...................................................................................................... 101
6 10BASE-T Tests..................................................................................110
6.1 Starting 10BASE-T Tests.......................................................................................... 110
6.2 Test Configuration for Ethernet 10BASE-T............................................................. 110
6.3 10BASE-T Tests......................................................................................................... 112
6.4 10BASE-Te Tests.......................................................................................................139
3User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
7 Ethernet 2.5/5/10 G.............................................................................169
7.1 2.5GBASE-T Tests..................................................................................................... 169
7.2 5GBASE-T Tests........................................................................................................ 192
7.3 10GBASE-T Tests...................................................................................................... 210
8 100BASE-T1 Tests..............................................................................229
8.1 Starting 100BASE-T1 Tests...................................................................................... 231
8.2 Test Configuration for 100BASE-T1........................................................................ 231
8.3 Transmitter Output Droop........................................................................................ 234
8.4 Transmitter Distortion...............................................................................................237
8.5 Transmitter Timing Jitter.......................................................................................... 241
8.6 Transmitter Power Spectral Density and Peak Differential Output...................... 244
8.7 MDI Return Loss with R&S RT-ZF2..........................................................................245
Contents
8.8 MDI Mode Conversion Loss and Return Loss........................................................247
8.9 MDI Mode Conversion Loss Adaptor Verification..................................................249
8.10 MDI Common Mode Emission..................................................................................250
9 1000BASE-T1 Tests............................................................................253
9.1 Starting 1000BASE-T1 Tests.................................................................................... 254
9.2 Test Configuration for 1000BASE-T1...................................................................... 255
9.3 Transmitter Timing Jitter (97.5.3.3)..........................................................................259
9.4 MDI Jitter and Clock Frequency (97.5.3.3/97.5.3.6)................................................ 264
9.5 Transmitter Distortion (97.5.3.2).............................................................................. 266
9.6 PSD Power Level and Output Voltage (97.5.3.4/ 97.5.3.5 ).....................................268
9.7 Maximum Output Droop (97.5.3.1)...........................................................................270
9.8 MDI Return Loss with R&S RT-ZF2 (97.7.2.1)......................................................... 272
9.9 MDI Mode Conversion Loss Test Head Verification (OA ECU)............................. 274
9.10 MDI Mode Conversion Loss and Return Loss(97.7.2.2)........................................ 275
10 10BASE-T1 Tests................................................................................278
10.1 Starting 10BASE-T1 Test.......................................................................................... 278
10.2 Test Configuration for Ethernet 10BASE-T1...........................................................278
10.3 10BASE-T1S Tests.................................................................................................... 280
10.4 10BASE-T1L Tests.....................................................................................................287
11 MGBASE-T1 Tests..............................................................................295
4User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
11.1 Starting MGBASE-T1 Test........................................................................................ 295
11.2 Test Configuration for MGBASE-T1.........................................................................295
11.3 Test Equipment..........................................................................................................297
11.4 Common Parameters................................................................................................298
11.5 Transmitter Timing Jitter Master..............................................................................299
11.6 Transmitter Timing Jitter Slave Mode..................................................................... 300
11.7 Transmitter MDI Random Jitter in Master Mode.....................................................301
11.8 Test Mode 2, JP03A...................................................................................................302
11.9 Transmit MDI Even-Odd Jitter in Master Mode.......................................................304
11.10 Transmitter Linearity.................................................................................................305
11.11 Test Mode 5................................................................................................................307
11.12 Maximum Output Droop........................................................................................... 308
11.13 MDI Return Loss........................................................................................................310
Contents
5User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
Contents
6User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP

1 R&S ScopeSuite Overview

The R&S ScopeSuite software is used with R&S RTO/RTO6/RTP oscilloscopes. It can be installed on a test computer or directly on the oscilloscope. For system require­ments, refer to the Release Notes.
R&S ScopeSuite Overview
The R&S ScopeSuite main panel has several areas:
"Settings": connection settings to oscilloscope and other instruments also default report settings
"Compliance Tests": selection of the compliance test
"Demo": accesses demo test cases that can be used for trying out the software without having a connection to an oscilloscope
"Help": opens the help file, containing information about the R&S ScopeSuite con­figuration
"About": gives information about the R&S ScopeSuite software
"Tile View": allows a personalization of the compliance test selection You can configure which tests are visible in the compliance test section and which are hidden, so that only the ones you use are displayed.
► To hide a test from the "Compliance Tests" view, do one of the following:
7User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
a) Right-click on the compliance test you want to hide.
The icon of the test changes, see Figure 1-1. Now with a left click you can hide the test.
Figure 1-1: Unpin icon
b) Click on "Title View" to show a list of the available test cases. By clicking a test
case in the show list, you can pin/unpin it from the main panel.
R&S ScopeSuite Overview
8User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP

2 Preparing the Measurements

2.1 Test Equipment

For Ethernet compliance tests, the following test equipment is needed:
The free-of-charge R&S ScopeSuite software, which can be installed on a com­puter or directly on the oscilloscope.
10/100/1000BASE-T Ethernet compliance tests:
R&S RTO / R&S RTO6 /R&S RTP oscilloscope with at least 600 MHz band-
width
Differential probe with at least 1 GHz bandwidth
R&S RTO / R&S RTO6 /R&S RTP-K22 10/100/1000BASE-T Ethernet compli-
ance test option (required option, installed on the oscilloscope)
R&S RT-ZF2 Ethernet test fixture set
For energy-efficient Ethernet tests, in addition:
R&S RT-ZF5 Ethernet test fixture set for 100/1000BASE-T EEE tests R&S RT-ZF4 test fixture for 10BASE-Te tests
2.5GBASE-T Ethernet compliance tests:
R&S RTO / R&S RTO6 /R&S RTP oscilloscope with at least 600 MHz band-
width For precise power spectral density and power level measurements up to 1 GHz, it is recommended to use an oscilloscope and a differential probe with 1 GHz bandwidth each.
Differential probe with at least 1GHz bandwidth
R&S RTO / R&S RTO6 /R&S RTP-K23 GBASE-T Ethernet compliance test
option (required option, installed on the R&S RTO/RTO6/RTP)
R&S RT-ZF2 Ethernet test fixture set
5GBASE-T Ethernet compliance tests:
R&S RTO / R&S RTO6 /R&S RTP oscilloscope with at least 1 GHz bandwidth
For precise power spectral density and power level measurements up to
1.5 GHz, it is recommended to use an oscilloscope and a differential probe with
1.5 GHz bandwidth each.
Differential probe with at least 1.5 GHz bandwidth
R&S RTO / R&S RTO6 /R&S RTP-K23 GBASE-T Ethernet compliance test
option (required option, installed on the R&S RTO/RTO6/RTP)
R&S RT-ZF2 Ethernet test fixture set
10GBASE-T Ethernet compliance tests:
R&S RTO / R&S RTO6 /R&S RTP oscilloscope with at least 2 GHz bandwidth
For precise power spectral density and power level measurements up to 3 GHz, it is recommended to use an oscilloscope and a differential probe with 3 GHz bandwidth each.
Preparing the Measurements
Test Equipment
9User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
Differential probe with at least 3 GHz bandwidth
R&S RTO / R&S RTO6 /R&S RTP-K23 GBASE-T Ethernet compliance test
option (required option, installed on the R&S RTO/RTO6/RTP)
R&S RT-ZF2 Ethernet test fixture set
100BASE-T1 compliance tests:
R&S RTO / R&S RTO6 /R&S RTP oscilloscope with at least 600 MHz band-
width
Differential probe with 1 GHz bandwidth
R&S RTO / R&S RTO6 /R&S RTP-K24 100BASE-T1 compliance test option
(required option, installed on the R&S RTO/RTO6/RTP)
R&S RT-ZF2 Ethernet test fixture set
R&S RT-ZF3 frequency converter for transmitter distortion tests
1000BASE-T1 compliance tests:
R&S RTO / R&S RTO6 /R&S RTP oscilloscope with at least 2 GHz bandwidth
Differential probe with 2 GHz bandwidth
R&S RTO / R&S RTO6 /R&S RTP-K87 1000BASE-T1 compliance test option
(required option, installed on the R&S RTO/RTO6/RTP)
R&S RT-ZF2 Ethernet test fixture set
R&S RT-ZF6 frequency converter for transmitter distortion tests
MGBASE-T1 compliance tests:
R&S RTO / R&S RTO6 /R&S RTP oscilloscope, with a bandwidth of:
- for 2.5 GBASE-T1 tests: at least 3.5 GHz
- for 5 GBASE-T1 tests: at least 7 GHz
- for 10 GBASE-T1 tests: at least 14 GHz
2 BNC/SMA cables
R&S RTO / R&S RTO6 /R&S RTP-K88 MGBASE-T1 compliance test option
(required option, installed on the R&S RTO/RTO6/RTP)
10BASE-T1 compliance tests:
R&S RTO / R&S RTO6 /R&S RTP oscilloscope
Optional Differential probe with 1 GHz bandwidth
R&S RTO / R&S RTO6 /R&S RTP-K89 10BASE-T1 compliance test option
(required option, installed on the R&S RTO/RTO6/RTP)
R&S RTO-B6/R&S RTP-B6/Tabor WX2182B/Tabor WX2182C arbitrary waveform generator for automatic disturber tests. For some disturber tests, also HMF2550 can be used.
R&S ZNB/ZNC/ZND/ZVL vector network analyzer for automatic return loss mea­surements. For manual measurements, also other AWGs and VNAs can be used.
Preparing the Measurements
Test Equipment
10User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP

2.2 Installing Software and License

The preparation steps are performed only once for each computer and instrument that are used for testing.
Uninstall older versions of the R&S ScopeSuite
If an older version of the R&S ScopeSuite is installed, make sure to uninstall the old version before you install the new one. You can find the version number of the current installation in "Help" menu > "About". To uninstall the R&S ScopeSuite, use the Win­dows " Control Panel" > "Programs".
For best operation results, we recommend that the installed firmware versions of the R&S ScopeSuite and the oscilloscope are the same.
To install the R&S ScopeSuite
1. Download the latest R&S ScopeSuite software from the "Software" section on the Rohde & Schwarz R&S RTO/RTO6/RTP website:
www.rohde-schwarz.com/product/rtp.html www.rohde-schwarz.com/product/rto.html
Preparing the Measurements
Setting Up the Network
2. Install the R&S ScopeSuite software:
On the computer that is used for testing, or
On the R&S RTO/RTO6/RTP.
For system requirements, refer to the Release Notes.
To install the license key on the R&S RTO/RTO6/RTP
► When you got the license key of the compliance test option, enable it on the oscil-
loscope using [Setup] > "SW Options". For a detailed description, refer to the R&S RTO/RTO6/RTP user manual, chapter "Installing Options", or to the online help on the instrument.

2.3 Setting Up the Network

If the R&S ScopeSuite software runs on a test computer, the computer and the testing oscilloscope require a LAN connection.
For some test cases, you need an additional instrument: arbitrary waveform generator (AWG), vector network analyzer (VNA), or spectrum analyzer. These instruments can be used in automatic or manual mode. For automatic testing, a LAN connection to the additional instrument is required.
There are two ways of connection:
LAN (local area network): It is recommended that you connect to a LAN with DHCP server. This server uses the Dynamic Host Configuration Protocol (DHCP) to assign all address information automatically.
11User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
Preparing the Measurements
Connecting the R&S
If no DHCP server is available, or if the Tabor WX2182B or WX2182C is used for automatic testing, assign fixed IP addresses to all devices.
Direct connection of the instruments and the computer or connection to a switch using LAN cables: Assign fixed IP addresses to the computer and the instruments and reboot all devices.
To set up and test the LAN connection
1. Connect the computer and the instruments to the same LAN.
2. Start all devices.
3. If no DHCP server is available, assign fixed IP addresses to all devices.
4. Ping the instruments to make sure that the connection is established.
5. If VISA is installed, check if VISA can access the instruments.
a) Start VISA on the test computer. b) Validate the VISA address string of each device.
See also:
Chapter 2.5, "Connecting the R&S RTO/RTO6/RTP", on page 12
Chapter 2.6, "Connecting the Arbitrary Waveform Generator", on page 14
Chapter 2.7, "Connecting the Vector Network Analyzer", on page 15
Chapter 2.8, "Connecting the Spectrum Analyzer", on page 17
RTO/RTO6/RTP

2.4 Starting the R&S ScopeSuite

To start the R&S ScopeSuite on the test computer or on the oscilloscope:
► Double-click the R&S ScopeSuite program icon.
To start the R&S ScopeSuite on the instrument, in the R&S RTO/RTO6/RTP firmware:
► In the "Apps" dialog, open the "Compliance" tab.

2.5 Connecting the R&S RTO/RTO6/RTP

If the R&S ScopeSuite is installed directly on the instrument, the software detects the R&S RTO/RTO6/RTP firmware automatically, and the "Oscilloscope" button is not available in the R&S ScopeSuite.
If the R&S ScopeSuite software runs on a test computer, the computer and the testing oscilloscope require a LAN connection, see Chapter 2.3, "Setting Up the Network", on page 11. The R&S ScopeSuite software needs the IP address of the oscilloscope to establish connection.
12User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
1. Start the R&S RTO/RTO6/RTP.
2. Start the R&S ScopeSuite software.
3. Click "Settings" > "Oscilloscope".
Preparing the Measurements
Connecting the R&S RTO/RTO6/RTP
4. Enter the IP address of the oscilloscope. To obtain the IP address: press the Rohde & Schwarz logo at the top-right corner of the oscilloscope's display.
5. Click "Get Instrument Information".
The computer connects with the instrument and gets the instrument data.
13User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
Preparing the Measurements
Connecting the Arbitrary Waveform Generator
If the connection fails, an error message is shown.

2.6 Connecting the Arbitrary Waveform Generator

For Ethernet tests using a disturber, an arbitrary waveform generator (AWG) is required. The tests can be performed manually or automatically, depending on the available AWG.
Automatic test execution is possible with all instruments that are listed in the R&S Sco­peSuite, in the "Instrument Settings" dialog box. In automatic mode, the R&S Scope­Suite configures the instrument and ensures that the AWG sends the required wave­forms. Automatic mode requires a LAN connection and the installation of a VISA imple­mentation (R&S VISA, see www.rohde-schwarz.com/rsvisa) on the computer that is running the R&S ScopeSuite. If the R&S ScopeSuite is installed on the R&S RTO/ RTO6/RTP, no installation is needed because VISA is already installed on the instru­ment. If the Tabor WX2182B or WX2182C is used for automatic testing, fixed IP addresses are required.
For manual test execution, it is recommended to use one of the listed AWGs, but you can also use another AWG. In manual mode, you connect the AWG to the test board and configure the instrument manually. VISA is not required. The R&S ScopeSuite uses VISA if it is installed, otherwise it uses the VXI-11 protocol.
To configure the arbitrary waveform generator for automatic testing
1. Connect the computer and the AWG.
2. Set up the LAN connection. See Chapter 2.3, "Setting Up the Network", on page 11.
14User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
3. In the R&S ScopeSuite, click "Instruments". Alternatively, you can select the "Instrument" tab in the test case configuration dia­log.
4. Click the "AWG" tab.
5. Select "Operating mode" = "Automatic".
6. Select a supported "AWG Type" and enter its IP address. For a list of the supported AWGs, see chapter "Test Equipment".
Preparing the Measurements
Connecting the Vector Network Analyzer
7. Click "Get Instrument Information".
The computer or R&S RTO/RTO6/RTP connects with the instrument and retrieves the instrument data.
8. If the connection to the arbitrary waveform generator failed, check if the IP address is assigned correctly.
To configure the AWG for manual testing
► In the "AWG" tab, enable the "Manual" operating mode.

2.7 Connecting the Vector Network Analyzer

The vector network analyzer (VNA) is required to perform Ethernet return loss mea­surements.
Similar to the AWG, the VNA can be used in automatic or manual mode. You can use the automatic mode only with supported instruments. A LAN connection and a VISA
15User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
installation on the computer that is running the R&S ScopeSuite is required. If the R&S ScopeSuite is installed on the R&S RTO/RTO6/RTP, no installation is needed because VISA is already installed on the instrument.
For manual test execution, it is recommended to use one of the listed VNAs. Moreover, any VNA can be used that meets the following requirements:
S11 parameter measurements are possible
Can export trace data in Touchstone (*.s1p) or *.csv format
Supports frequency range 1 MHz to 500 MHz
In manual mode, you connect the vector network analyzer to the test board and config­ure the instrument manually.
To connect the vector network analyzer for automatic testing
1. Connect the computer and the VNA. Set up the LAN connection, see Chapter 2.3,
"Setting Up the Network", on page 11.
2. In the R&S ScopeSuite, click "Instruments".
Preparing the Measurements
Connecting the Vector Network Analyzer
3. Click the "VNA" tab.
4. Select the "Automatic" operating mode.
5. Select the "VNA Type" and enter its IP address.
6. Click "Get Instrument Information".
The computer or R&S RTO/RTO6/RTP connects with the instrument and retrieves the instrument data.
16User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP

2.8 Connecting the Spectrum Analyzer

A spectrum analyzer is required to perform Ethernet 10GBASE-T transmitter linearity tests.
Similar to the AWG, the spectrum analyzer can be used in automatic or manual mode. Automatic mode is supported with R&S FSV/FSQ/FSW/ESRP. It requires a LAN con­nection and a VISA installation on the computer that is running the R&S ScopeSuite. If the R&S ScopeSuite is installed on the R&S RTO/RTO6/RTP, no installation is needed because VISA is already installed on the instrument.
For manual test execution, it is recommended to use one of the listed spectrum ana­lyzers. Moreover, any spectrum analyzer can be used that meets the following require­ments:
Frequency range from 1 MHz to 400 MHz
RF input attenuation greater than 35 dB
In manual mode, you connect the spectrum analyzer to the test board and configure the instrument manually.
Preparing the Measurements
Connecting the Spectrum Analyzer
To connect the spectrum analyzer for automatic testing
1. Connect the computer and the spectrum analyzer.
2. Set up the LAN connection. See Chapter 2.3, "Setting Up the Network", on page 11.
3. In the R&S ScopeSuite, click "Instruments".
4. Click the "SA" tab.
5. Select the "Automatic" operating mode.
6. Select the "SA Type" and enter its IP address.
17User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
Preparing the Measurements
Report Configuration
7. Click "Get Instrument Information".
The computer or R&S RTO/RTO6/RTP connects with the instrument and retrieves the instrument data.

2.9 Report Configuration

In the "Report Configuration" menu, you can select the format of the report and the details to be included in the report. You can also select an icon that is displayed in the upper left corner of the report.
Also, you can enter common information on the test that is written in the "General Infor­mation" section of the test report.
18User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
Preparing the Measurements
Report Configuration
19User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP

3 Performing Tests

3.1 Starting a Test Session

Performing Tests
Starting a Test Session
After you open a compliance test, the "Session Selection" dialog appears. In this dia­log, you can create new sessions, open or view existing report.
The following functions are available for handling test sessions:
Function Description
"Add" Adds a new session
"Open" Opens the selected session
"Remove" Removes the selected session
"Rename" Changes the "Session Name"
"Comment" Adds a comment
"Show report" Generates a report for the selected session
To add a test session
1. In the R&S ScopeSuite window, select the compliance test.
2. In the "Session Selection" dialog press "Add".
3. If necessary change the "Session Name"
20User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
To open a test session
1. In the R&S ScopeSuite window, select the compliance test.
2. In the "Session Selection" dialog, select the session you want to open and double click on it. Alternatively, select the session and press "Open".
To show a report for a test session
1. In the R&S ScopeSuite window, select the compliance test.
2. In the "Session Selection" dialog, select the session you want the report for and press "Show report".

3.2 Configuring the Test

Performing Tests
Configuring the Test
1. In the R&S ScopeSuite window, select the compliance test to be performed:
"100BASE-T1"
"Ethernet"
"MGBASE-T1"
"Ethernet 2.5/5G/10G"
"1000BASE-T1"
"10BASE-T1"
2. Open a test session, see Chapter 3.1, "Starting a Test Session", on page 20.
3. Adjust the "Properties" settings for the test cases you want to perform.
4. Click "Limit Manager" and edit the limit criteria, see Chapter 3.2.1.1, "Limit Man-
ager", on page 23.
5. If you want to use special report settings the "Report Config" tab to define the for­mat and contents of the report. Otherwise the settings defined in "RSScopeSuite" > "Settings" > "Report" are used. See Chapter 2.9, "Report Configuration", on page 18.
6. Click "Test Checked"/"Test Single" and proceed as described in the relevant test case chapter.
21User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP

3.2.1 General Test Settings

Performing Tests
Configuring the Test
Each session dialog is divided into several sections:
"Properties": shows the settings that can be made for the test case selected on the left side of the dialog. You can differentiate between the "All" and the sub test prop­erties In the "All" > "Properties" tab you can configure the settings for all test cases in the current session. Once you change and save a setting in this tab, the changes will be done for all test in the sessions. At the same time, there will be a special mark­ing for the functions that have different settings for different sub tests.
"Limit Manager": sets the measurement limits that are used for compliance testing, see Chapter 3.2.1.1, "Limit Manager", on page 23.
"Results": shows an overview of the available test results for this session.
"Instruments": defines instruments settings for connecting to external devices, that are specific for this test session. When a session is first created the global settings ("RSScopeSuite" > "Settings" > "Instruments") are copied to the session. This "Instruments" tab can be used to change those copied defaults.
"Report Config": defines the format and contents of the report for this session. When a session is first created the global settings ("RSScopeSuite" > "Settings" > "Report") are copied to the session. This "Report Config" tab can be used to change those copied defaults.
"Test Checked"/ "Test Single": starts the selected test group.
22User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
3.2.1.1 Limit Manager
The "Limit Manager" shows the measurement limits that are used for compliance test­ing.
Each limit comprises the comparison criterion, the unit, the limit value A, and a second limit value B if the criterion requires two limits.
You can set the values to defaults, change the values in the table, export the table in xml format, or import xml files with limit settings.
► Check and adjust the measurement limits.
Performing Tests
Getting Test Results

3.3 Getting Test Results

For each Ethernet test, the test data - report, diagrams and waveform files - is saved in the following folder:
%ProgramData%\Rohde-Schwarz\RSScopeSuite\3.0\Sessions\Ethernet\ <Ethernet category>\<Session_Name>
23User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
For each 100BASE-T1 test, the test data - report, diagrams and waveform files - is saved in the following folder:
%ProgramData%\Rohde-Schwarz\RSScopeSuite\3.0\Sessions\ 100BASE-T1\<Session_Name>
If you resume an existing session, new measurements are appended to the report, new diagrams and waveform files are added to the session folder. Existing files are not deleted or replaced. Sessions data remain until you delete them in the "Results" tab of the session.
The report format can be defined in "RSScopeSuite" > "Settings" > "Report" for all compliance tests (see also Chapter 2.9, "Report Configuration", on page 18). If you want to use special report settings for a session, you can define the format and con­tents of the report in the "Report Config" tab of the session.
All test results are listed in the "Results" tab. Reports can be provided in PDF, MSWord, or HTML format. To view and print PDF reports, you need a PDF viewer, for example, the Acrobat Reader.
The test report file can be created at the end of the test, or later in the "Session Selec­tion" dialog.
Performing Tests
Getting Test Results
To show a test report
1. In the R&S ScopeSuite window, select the compliance test to be performed.
2. Select the session name in the "Session Selection" dialog and click "Show report".
The report opens in a separate application window, depending on the file format. You can check the test results and print the report.
To delete the results, diagrams and waveform files of a session
1. In the "Session Selection" dialog select the session and open it.
2. In the "Results" tab, select the result to be deleted.
3. Click "Remove".
24User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP

4 1000BASE-T Tests

1000BASE-T Ethernet compliance tests require option R&S RTO/RTO6/RTP-K22. 1000BASE-T EEE tests require also option R&S RTO/RTO6/RTP-K86.

4.1 Starting 1000BASE-T Tests

Before you run the test, complete the following actions:
Initial setup of the equipment, see Chapter 2.2, "Installing Software and License", on page 11
LAN connection of the oscilloscope and the computer running the R&S Scope­Suite, see Chapter 2.5, "Connecting the R&S RTO/RTO6/RTP", on page 12
AWG connection for tests with disturber, see Chapter 2.6, "Connecting the Arbitrary
Waveform Generator", on page 14.
VNA connection for MDI Return Loss and Transmitter Power Spectral Density tests, see Chapter 2.7, "Connecting the Vector Network Analyzer", on page 15.
1000BASE-T Tests
Test Configuration for Ethernet 1000BASE-T
1. Select "Ethernet" in the R&S ScopeSuite start window.
2. In the "Session Selection" dialog, set "Select Type" > "1000BASE-T".
3. Add a new test session and open it, see Chapter 3.1, "Starting a Test Session", on page 20.
4. Check the test configuration settings and adjust, if necessary. See:
Chapter 4.2, "Test Configuration for Ethernet 1000BASE-T", on page 25
Chapter 3.2.1.1, "Limit Manager", on page 23
5. Select/check the test cases you want to run and click "Test Single"/"Test checked".
6. A step-by step guide explains the following individual setup steps. When you have finished all steps of the step-by-step guide, the compliance test runs automatically.

4.2 Test Configuration for Ethernet 1000BASE-T

Open a "1000BASE-T" session.
The test configuration consists of some test-specific configuration settings.
25User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
1000BASE-T Tests
Test Configuration for Ethernet 1000BASE-T
Figure 4-1: Configuration for 1000BASE-T compliance tests
Pair
Select which cable pair is used.
Signals
Selects the oscilloscope channels for the pair testing.
Average Count
Defines the number of waveforms which the oscilloscope acquires to calculate the average waveform (average count).
Increasing the number of waveforms results in more accurate measurement results but also in longer test execution time. If you are unsure, use the default values. The valid range is 5 to 200. The default value is 10.
Measurement Time
Sets the measurement time. The valid range is 50E-9 to 100E-3, the default value is 100E-3.
Voltage Attenuation
The voltage attenuation is used for all 1000BASE-T EEE test cases.
Voltage attenuation sets how many times the voltage attenuates after going through directional coupler on the test fixture. The default value is a theoretical value and works well for R&S RT-ZF5. It is not suggested to be changed for all time unless a value is confirmed from careful calibration.
26User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
Expert Mode
If enabled, the "Expert Mode" allows you to bypass the guided steps of the test case.
Test Fixture
Selects if the fixture that is used for the measurement has a SMA input or not.
This setting is used for the "Common-mode Output Voltage" test cases.

4.3 1000BASE-T

The following table shows an overview of the needed equipment for performing the 1000BASE-T test cases. A prerequisite is also a R&S RTO / R&S RTO6 /R&S RTP with at least 600 MHz bandwidth, the R&S ScopeSuite software and a 1000BASE-T DUT.
Table 4-1: Equipment for 1000BASE-T Ethernet compliance tests
1000BASE-T Tests
1000BASE-T
Transmitter Distortion
(40.6.1.2.4)
No disturber and No TX_TCLK
Transmitter Distortion
(40.6.1.2.4)
No disturber and with TX_TCLK
Transmitter Distortion
(40.6.1.2.4)
With disturber and No TX_TCLK
Transmitter Distortion
(40.6.1.2.4)
With disturber and With TX_TCLK
Differen-
tial probe
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
Differential type Single-
Test fixture
R&S RT-ZF2
Section:
Resistive Load
R&S RT-ZF2
Section:
Resistive Load
R&S RT-ZF2
Section:
Resistive Load
R&S RT-ZF2
Section:
Resistive Load
Ended type
SMA
- Optional:
1
at least 1
GHz band-
width
- Optional:
1
at least 1
GHz band-
width
Additional instruments
R&S RTO / R&S RTO6 /R&S RTP-K17
Optional:
R&S RTO / R&S RTO6 /R&S RTP-K17
R&S RTO / R&S RTO6 /R&S RTP-K17
ARB:
R&S RTO-B6 or
Tabor WX2182B/ WX2182C or
HMF2550
Optional:
R&S RTO / R&S RTO6 /R&S RTP-K17
ARB:
R&S RTO-B6 or
Tabor WX2182B/ WX2182C or
HMF2550
27User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
1000BASE-T Tests
1000BASE-T
Peak Output Voltage
(40.6.1.2.1)
No disturber
Peak Output Voltage
(40.6.1.2.1)
With disturber
Maximum Output Droop
(40.6.1.2.2)
No disturber
Maximum Output Droop
(40.6.1.2.2)
With disturber
Differen-
tial probe
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
Differential type Single-
Test fixture
R&S RT-ZF2
Section:
Resistive Load
R&S RT-ZF2
Section:
Resistive Load
R&S RT-ZF2
Section:
Resistive Load
R&S RT-ZF2
Section:
Resistive Load
Ended type
SMA
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
Additional instruments
Optional:
R&S RTO / R&S RTO6 /R&S RTP-K17
Optional:
R&S RTO / R&S RTO6 /R&S RTP-K17
ARB:
R&S RTO-B6 or
Tabor WX2182B/ WX2182C or
HMF2550
Optional:
R&S RTO / R&S RTO6 /R&S RTP-K17
Optional:
R&S RTO / R&S RTO6 /R&S RTP-K17
ARB:
R&S RTO-B6 or
Tabor WX2182B/ WX2182C or
HMF2550
Differential Output Tem­plates
(40.6.1.2.3)
No disturber
Differential Output Tem­plates
(40.6.1.2.3)
With disturber
Jitter Master Mode and Clock Frequency
(40.6.1.2.5 and
40.6.1.2.6 )
No TX_TCLK
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
R&S RT-ZF2
Section:
Resistive Load
R&S RT-ZF2
Section:
Resistive Load
R&S RT-ZF2
Section:
Jitter Slave Test 1000BaseT
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
- Total of 2 DUTs required
Optional:
R&S RTO / R&S RTO6 /R&S RTP-K17
Optional:
R&S RTO / R&S RTO6 /R&S RTP-K17
ARB:
R&S RTO-B6 or
Tabor WX2182B/ WX2182C or
HMF2550
28User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
1000BASE-T Tests
1000BASE-T
Jitter Master Mode and Clock Frequency
(40.6.1.2.5 and
40.6.1.2.6 )
With TX_TCLK
Jitter Slave Mode
(40.6.1.2.5)
No TX_TCLK
Jitter Slave Mode
(40.6.1.2.5)
With TX_TCLK
MDI Return Loss
(40.8.3.1)
Differential type Single-
Differen-
tial probe
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
1
at least 1
GHz band-
width
Test fixture
R&S RT-ZF2
Section:
Jitter Slave Test 1000BaseT
R&S RT-ZF2
Section:
Jitter Slave Test 1000BaseT
R&S RT-ZF2
Section:
Jitter Slave Test 1000BaseT
- R&S RT-ZF2
Section:
Twisted Pair Model
Additional instruments
Ended type
SMA
1
at least 1
GHz band-
width
- Total of 2 DUTs required
1
at least 1
GHz band-
width
- VNA:
Total of 2 DUTs required
Gigabit Ethernet jitter cable:
R&S RT-ZF2C
Total of 2 DUTs required
R&S ZNB or
R&S ZNC or
R&S ZND or
R&S ZVL
Common-mode Output Voltage
(40.8.3.3)
at least 1
GHz band-
width
Transmitter Distortion..............................................................................................29
Peak Output Voltage...............................................................................................35
Maximum Output Droop..........................................................................................39
Differential Output Templates..................................................................................43
Jitter Master Mode.................................................................................................. 49
Jitter Slave Mode.................................................................................................... 53
MDI Return Loss..................................................................................................... 56
Common-mode Output Voltage...............................................................................58

4.3.1 Transmitter Distortion

The " Transmitter Distortion" test verifies that the peak distortion of the transmitter is less than 10 mV - with/without a disturbing signal, and with/without a DUT transmit clock TX_CLK.
1
R&S RT-ZF2
Section:
Common Mode Volt­age
- Optional:
R&S RTO / R&S RTO6 /R&S RTP-K17
29User Manual 1317.7202.02 ─ 16
R&S®RTO, R&S®RTO6, R&S®RTP
4.3.1.1 No Disturber
Test Equipment
Table 4-2: Equipment for Transmitter Distortion test, no disturber
Item Description, model Quantity
1000BASE-T Tests
1000BASE-T
Rohde & Schwarz oscilloscope R&S RTO / R&S RTO6 /R&S RTP with at
least 600 MHz bandwidth
Ethernet test fixture, "Resistive Load" section R&S RTZF2 1
Single-ended probe (for test "With TX_CLK") at least 1 GHz bandwidth 1
Differential probe at least 1 GHz bandwidth 1
DUT the device you want to test 1
Performing the Tests
1. Start the test as described in Chapter 4.1, "Starting 1000BASE-T Tests", on page 25.
2. Select the test case:
"No TX_CLK" test:
"No Disturber" > "No TX_CLK" > "Transmitter Distortion (40.6.1.2.4)"
1
"With TX_CLK" test:
"No Disturber" > "With TX_CLK" > "Transmitter Distortion (40.6.1.2.4)"
30User Manual 1317.7202.02 ─ 16
Loading...
+ 281 hidden pages