The LD9-TG features sensor technology designed to detect a 5%
hydrogen / 95% nitrogen tracer gas mixture. The LD9-TG’s full
color liquid crystal display (LCD) and sweep mode function convey
messages, graphics, and prompts to help the A/C technician locate the
source of a leak and ensure that the leak detector is always at optimal
performance. The light-emitting diode (LED) inspection lights aid the
technician in locating and inspecting suspected leak sources.
When used with the hydrogen/nitrogen tracer gas mixture, the
LD9-TG will detect leak rates equivalent to 4 g/year of R-134a in the
High setting, and 7 g/year in the Normal setting while moving at 3
inches (8 cm) per second. This complies with SAE J2970 sections
7.2 and 7.3. Testing with tracer gas also complies with European
standards EN 35422 and EN 14624.
Features
• Patented 3 LED ultraviolet (UV) lights with 395–415 nm
wavelengths optimized for A/C dye fluorescence
• Full color LCD with user-friendly message and error screens
• Sensitivity equivalent to .015 oz/yr of R-1234yf in High mode
• Sensitivity equivalent to .05 oz/yr for R-134a in High mode
1. Turn On: Press the ON/OFF button once to turn on; press again to turn off.
NOTE: The LD9-TG defaults to SAE J2970 mode on power up. Hold down
ON/OFF button to select the Sensitivity Boost mode for smaller leaks.
2. Warm Up: The detector automatically starts heating the sensor. During
the heating cycle, the LCD will display the message “WARM UP PLEASE
WAIT” with a progress bar. Warm up is usually less than 20 seconds.
3. Search: The display will show the message “READY” and then “SEARCH”
when the detector begins to search for leaks. An audible beep will begin
to sound. Move the probe tip towards a suspected refrigerant leak at the
rate of less than 2 inches (~5 cm) per second, no more than 1/4 inch
(~0.6 cm) away from the suspected source.
4. Detection: If a leak exists, the beeping will increase in rate and pitch and
the display will show the numerical indication of the leak size.
NOTE: The leak detector responds to changes in tracer gas concentration.
When detection occurs, move the probe away from the source and back
again to conrm the leak source. The detector’s audible beeping will reset
if the probe is held xed at the source (see Automatic Calibration).
Wear personal protective equipment that meets ANSI/ISEA
•
and OSHA standards.
Before leak checking with the UV light:
a. Make sure the A/C system is properly charged with sufcient dye.
(See manufacturer’s specications for proper dye charge.)
b. Run the A/C system long enough to thoroughly mix and circulate the dye
(sold separately) with the refrigerant and lubricating oil.
1. Turn on UV light by pressing the LED button once. (See control panel on page 4.)
Three UV lights will turn on (see image below).
2. Holding the leak detector approximately 10–14 inches (25–35 cm) away,
shine the UV light beam slowly over the components, hoses, and metal
ttings that make up the A/C system.
3. When the UV light shines on the fluorescent dye that has escaped from the
system, the dye will glow a bright yellow green.
3 UV Lights
1. Turn on the inspection light by pressing the LED button until the white LED
lights turn on. (See control panel on page 4.)
2. Inspect all components, hoses, and ttings for excessive wear or damage.
2 White Inspection Lights
NOTE: If not manually turned off, the UV light and inspection light will
automatically shut off after 5 minutes to preserve battery life.
Once a leak is detected, the leak size indicator
bar graph will display on the detector screen.
The number of bars will increase or decrease
depending on the amount of tracer gas detected.
The maximum value of the tracer gas will be
displayed once the leak source has been located.
Refer to the table below to determine the
approximate size of the leak.
Maximum Bars DisplayedLeak Size (oz/yr)
1–2 (green color)< 0.1
3–5 (yellow color)0.1–0.5
6–10 (red color)> 0.5
Sweep (Pinpoint) Mode
Turn on sweep mode by pressing the MODE
button. This mode allows the user to pinpoint
hard to nd small leaks. The display cursor
will sweep across the display from left to right
tracking a horizontal baseline over a 3 second
period.
When no tracer gas is detected, the display
cursor will be flat. If tracer gas is detected, the
cursor on the display will rise up and continue
to rise as the leak source is approached. If the
detector moves away from the leak source, the
cursor will drop back down.
LEAK SIZE
MED SENS
LEAK LOCATION
HI SENS
AUTO
AUTO
NOTE: The default sensitivity level in the sweep
mode is High.
After a leak is identied, the detector will recalibrate itself either
automatically (default) or manually to the ambient air and resume
audibly beeping as soon as the leak is detected again.
In Automatic mode, the detector will automatically recalibrate if the
probe is held xed at the source of the leak, and the detector will not
beep again until the probe is moved away from the source and back
again. In Manual mode, the detector will continue to beep if tracer gas
is detected until the user presses the SENS button to recalibrate.
To use the detector in Manual mode, press and hold the SENS button
and release when the AUTO icon is replaced with MANUAL on the
display. To return to Automatic Calibration, press and hold the SENS
button and release when the AUTO icon is displayed.
NOTE: The sensitivity levels can only be changed in Automatic
mode. To change sensitivity levels while in Manual mode, switch to
Automatic mode and select the desired sensitivity level, then return
back to Manual mode.
Adjusting Sensitivity Levels
While in automatic mode, the detector can be set to 3 different
sensitivity levels (LO, MED, HI). If the detector continues to beep after
moving it away from the source of the leak, the sensitivity level can be
adjusted so the detector will only sound when the probe is close to the
leak source.
The detector will default to the MED sensitivity level automatically
once the unit comes out of the warm up cycle. To change sensitivity
levels, press the SENS once for HI sensitivity and again for LO
sensitivity.
To mute the audible beep, press the MUTE button. To restore the audible
beep, press the MUTE button again. (NOTE: There is a few seconds lapse
to restore sound if the MUTE button is pressed in rapid succession.)
Leak Test Vial
The leak detector comes with a leak test vial that allows the user to
verify that the detector is performing properly. Check the expiration date
on the vial before testing the leak detector.
1. Remove the colored label dot on the center of the screw cap to
expose the vent hole.
2. Turn on the detector and allow the unit to complete the warm up
cycle. Set sensitivity level to HIGH.
3. Place the probe tip close to the hole in the leak test vial. The beep
rate should increase and the leak size indicator should display 3–6
bars, indicating that the sensor and electronics are working properly.
Replacing Batteries
Replace the batteries when the display shows the message “REPLACE
BATTERIES.”
1. Unscrew battery cover located at the base of the unit as shown.
2. Insert all four batteries into the battery compartment, noting the
polarity mark on the inside of the battery compartment for proper
battery orientation.
becomes visibly dirty or when the display shows
“REPLACE SENSOR FILTER.” Unscrew probe
tip as shown to replace lter. The LD9-TG will
keep track of the number of hours of usage and
advise the user when it is time to replace the
lter.
To Replace Sensor: Remove the sensor by
pulling it out of the socket. Install the new
sensor by aligning the grooves in the sensor
cover with the raised grooves on the sensor
socket holder (see gure below).
NOTE: Do not force the sensor into the socket.
Misalignment can damage the sensor pins.
Push straight on (do not twist)TO INSTALL sensor.
PAG E 9
REPLACE
SENSOR
FILTER
CHECK
SENSOR
CONNECTION
OR REPLACE
SENSOR
Filter
Pull straight out (do not twist)
TO REMOVE sensor.
CAUTION: The detector’s software is designed to alert the user if the
sensor is dislodged or defective. If the sensor is not fully inserted into the
six-pin socket, or if it is defective, the unit will not come out of the warm
up mode for proper operation when the power button is pressed. In this
case, the message screen “CHECK SENSOR CONNECTION OR REPLACE
SENSOR” will be displayed. Additionally, if the unit becomes unstable
during operation, it is an indication that the sensor may be defective.
NOTE: If the leak detector has been out of use for an extended period of
time, the following action is recommended. Power on the unit and allow
it to come out of warm up, then run it at sensitivity level HI for several
minutes before testing it with the leak test vial. This action will help
ensure that the sensor is fully conditioned for maximum response to the
tracer gas.
Shows raw signal from sensor
Unit stays in HI and Manual mode
Mute is turned on
HI SENSMAN
Model: LD9-TG
DETECTION
SEARCH
HI
No Sound
Warm up
MAN
Press once for
UV LED
WARM UP
PLEASE WAIT
Rapid beep rate when
leak is detected
LEAK SIZE
MED SENS
AUTO
Press once to
MUTE
Press again to
UNMUTE
SEARCH
HI
AUTO
Normal beep rate;
Unit is ready
after 20 seconds
SENS
Press to change
sensitivity level
LO, MED, HI
SEARCH
HI
AUTO
Press twice for
Inspection LED
READY
SEARCH
HI
AUTO
Press for
Leak Location
mode
SEARCH
HI
AUTO
Sensor Clearing Message
NOTE: The “SENSOR CLEARING MESSAGE” is displayed when
the sensor becomes saturated with a very large concentration
of gas. Recovery is normally less than 10 seconds, during which
time the sensor will not function optimally.
Replace Sensor Filter Message
NOTE: The “REPLACE SENSOR FILTER” message is displayed
when the detector’s timer registers approximately 30 hours of
accumulated use. Press the appropriate button when prompted
“DONE” or “LATER” on the display. If “DONE” is selected, the
detector will reset to zero hours. If “LATER” is selected, the
detector will continue to prompt the user to replace the lter after
Some automotive solvents and chemicals have similar hydrocarbon
properties as R134a and may elicit a positive response from the detector.
Before leak checking, clean up any chemicals in the list below that elicit a
positive response.
Chemical Name/BrandResponse
Rain-X Windshield Wash FluidYes
Ford Spot Remover (wet)Yes
Ford Rust InhibitorYes
Ford Gasket Adhesive (wet)Yes
Loctite Natural Blue Degreaser (diluted)Yes
Ford Brake Parts CleanerYes
Ford Silicone Rubber (uncured)No
Motorcraft Antifreeze heated to 160 degrees FNo
Gunk Liquid WrenchYes
Ford Silicone LubricantNo
Ford Pumice Lotion (with solvent)Yes
Ford Motorcraft Brake FluidYes
Ford Carburetor CleanerYes
Dextron Transmission Fluid heated to 160 degrees F No
1. Determine the refrigerant charge of the A/C system using
refrigerant gauges (not included with this tool).
2. If there is a suspected refrigerant leak, use the SAE certied
refrigerant leak detector models LD7, LD5, or LD3 to locate the
leak(s).
3. Using an SAE certied A/C recovery machine (not included
with this tool), vacuum any remaining refrigerant out of the A/C
system.
4. Connect the psi regulator to the tracer gas tank (not included with
this tool) to charge the A/C system with tracer gas.
5. Connect the appropriate R134a or R1234yf service port coupler
tting to the hose and verify that the coupler is in the closed
position.
NOTE: Refrigerant may be used in some electric or hybrid vehicles
to cool the batteries. Refer to the appropriate shop manual
procedures to recover refrigerant from the battery coolant system,
or to charge refrigerant into the system.
6. Open the valve on the tracer gas tank to charge the A/C system
and adjust the psi regulator to 60 psi.
NOTE: Do not exceed the maximum pressure of the A/C
system being serviced. Refer to the appropriate shop manual
specifications.
7. Purge the air from the hose by loosening the hose fitting to the
R134a or R1234yf service port coupler and then retightening.
8. Connect the service port coupler to the high side service port on
the vehicle and open the service port coupler.
9. Allow the tracer gas to fill the A/C system. Refer to the appropriate
shop manual for pressure specifications.
10. Use the model LD9-TG Leak Detector that is included with the
LD9-TGKIT to check the system for leaks. Refer to the operating
instructions for using the LD9-TG Leak Detector listed on page 4.
11. Sweep the probe tip slowly over the components, hoses, and metal
ttings that make up the A/C system.
NOTE: The hydrogen/nitrogen tracer gas mixture rises; it does not
sink like R134a or R1234yf refrigerant.
12. Discharge the tracer gas into the atmosphere after the system
leak(s) are located. Do Not recover the tracer gas into an A/C
recovery machine.
13. Charge the system with the appropriate R134a or R1234yf
refrigerant only after the leak(s) are repaired and the A/C system
retested with the tracer gas and LD9-TG Leak Detector.