Review the following safety precautions carefully before operating the instrument to
avoid any personal injuries or damages to the instrument and any products
connected to it. To avoid potential hazards use the instrument as specified by this
user’s guide only.
The instrument should be serviced by qualified personnel only.
To Avoid Fire or Personal Injury
Use Proper Power Cord. Use the power cord designed for the instrument and
authorized in your country only.
Connect and Disconnect Accessories. Do not connect or disconnect probes or
test leads while they are connected to a voltage source
Ground The Instrument. The oscilloscope is grounded through the grounding
conductor of the power cord. To avoid electric shock the instrument grounding
conductor(s) must be grounded properly. Before making connections to the input or
output terminals of the instrument.
Connect The Probe.The probes’ ground terminals are at the same voltage level of
the instrument ground. Do not connect the ground terminals to a high voltage.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and marks on the instrument. Follow the user’s guide for further ratings information
before making connections to the instrument.
Do Not Operate Without Covers. Do not operate the instrument with covers or
panels removed.
Use Proper Fuse. Use the fuse of the type, voltage and current ratings as specified
for the instrument.
Avoid Circuit or Wire Exposure. Do not touch exposed connections and
components when power is on.
Do Not Operate With Suspected Failures. If suspected damage occurs with the
instrument, have it inspected by qualified service personnel before further
operations.
Provide Proper Ventilation. Refer to the installation instructions for proper
ventilation of the instrument.
Do not operate in wet/damp conditions
Do not operate in an explosive atmosphere
Keep product surfaces clean and dry
The disturbance test of all the models meet the limit values of A in the
standard of EN 61326: 1997+A1+A2+A3, but can’t meet the limit values
of B.
Measurement Category
The DS1000E, DS1000D series Digital Oscilloscope is intended to be used for
measurements in Measurement Category I.
Measurement Category Definitions
Measurement Category I is for measurements performed on circuits not directly
connected to MAINS. Examples are measurements on circuits not derived from
MAINS, and specially protected (internal) MAINS derived circuits. In the latter case,
transient stresses are variable; for that reason, the transient withstand capability of
the equipment is made known to the user.
WARNING
IEC Measurement Category I. The input terminals may be connected to circuit
terminal in IEC Category I installations for voltages up to 300 VAC. To avoid the
danger of electric shock, do not connect the inputs to circuit’s voltages above 300
VAC. Transient overvoltage is also present on circuits that are isolated from mains.
The DS1000E, DS1000D series Digital Oscilloscopes is designed to safely withstand
occasional transient overvoltage up to 1000Vpk. Do not use this equipment to
measure circuits where transient overvoltage could exceed this level.
This book covers the following four types of DS1000E, DS1000D Series Digital
Oscilloscopes:
DS1102E, DS1052E,
DS1102D, DS1052D. (With Logic Analyzer)
RIGOL DS1000E, DS1000D Series Digital Oscilloscopes provide exceptional
waveform viewing and measurements in a compact, lightweight package. The
DS1000E, DS1000D series is ideal for production test, field service, research, design,
education and training applications involving analog/digital circuits test and
troubleshooting, as well as education and training.
Optional 16 digital channels (DS1000D series), each channel can be turned on
or off independently, or in a 8 bit group
Mono/Color TFT LCD Displays at 320× 234 resolution
USB storage and printing supports, software upgradeable via USB connectivity
Adjustable waveform intensity, more effective waveform viewing
One-touch automatic setup for ease of use (AUTO)
Saves 10 Waveforms, 10 setups, supports CSV and bitmap format
Newly designed Delayed Scan Function, easy to give attention to both details
and overview of a waveform
20 Automatic measurements
Automatic cursor tracking measurements
Waveform recorder, record and replay dynamic waveforms
User selectable fast offset calibration
Built-in FFT function, Frequency Counter
Digital filters, includes LPF, HPF, BPF, BRF
Pass/Fail Function, optically isolated Pass/Fail output
Add, Subtract and Multiply Mathematic Functions
Advanced trigger types include: Edge, Video, Pulse width, Slope, Alternative,
Adjustable trigger sensitivity
Multiple Language User Interface
Pop-up menu makes it easy to read and easy to use
Built-in Chinese and English help system
Easy-to-use file system supports Chinese & English characters input
The front panel and user interface
To inspect the instrument
To perform a functional check
To compensate probes
To use digital leads
To display a signal automatically
To understand the vertical system
To understand the horizontal system
To trigger the oscilloscope
The first thing to do with a new oscilloscope is to know its front panel. This chapter
helps to be familiar with the layout of the knobs and keys and how to use them. Read
the chapter carefully before further operations.
Figure 1- 1, Front Panel; the knobs are used most often and are similar to the knobs
on other oscilloscopes. The buttons allow you to use some of the functions directly
but also bring up soft button menus on the screen, which enable the access to many
measurement features associated with advanced functions, mathematics, and
reference or to run control features.
The front panel of DS1000E:
The front panel of DS1000D:
Figure 1- 1
Front Panel Figure of DS1000E, DS1000D Series Oscilloscope
USB Host Logic Analyzer Port Signal Input EXT Trigger Probe
Channel Input Compensation
Figure 1- 2
Front Panel Instruction
Notation definitions in this User’s Guide:
Throughout this manual, notation symbols of buttons and knobs are the same to
those on front-panel.
A box around the name of the key denotes MENU function buttons on
front-panel, such as Measure.
() denotes the multi-function knob.
denotes the two POSITION knobs.
denotes the two SCALE knobs.
denotes the LEVEL knob.
The name with a drop shadow denotes the menu operating key, such as
When you get a new DS1000E, DS1000D series oscilloscope, please inspect the
instrument according to the following steps:
1. Inspect the shipping container for damage.
Keep a damaged shipping container or cushioning material until the contents of
the shipment have been checked for completeness and the instrument has been
checked mechanically and electrically.
2. Check the accessories.
Accessories supplied with the instrument are listed in "
in this guide.
If the contents are incomplete or damaged, please notify your RIGOL Sales
Representative.
3. Inspect the instrument.
In case there is any mechanical damage or defect, or the instrument does not
operate properly or fails performance tests, please notify the RIGOL Sales
Representative.
If the shipping container is damaged, or the cushioning materials show signs of
stress, please notify the carrier as well as the RIGOL sales office. Keep the
shipping materials for the carrier’s inspection.
RIGOL offices will arrange for repair or replacement at RIGOL’s option without
waiting for claim settlement.
Perform this quick functional check to verify that the instrument is operating
correctly.
1. Turn on the instrument.
Use the power cord designed for the oscilloscope only.
Use a power source that delivers 100 to 240 VAC
, 45Hz to 440Hz.
RMS
Turn on the instruments, and wait until the display shows the waveform window.
Push the Storagebutton, select Storage in the top menu box and push the
Factory menu box.
Figure 1- 5
Turn on and Check the instrument
WARNNING:
To avoid electric shock, be sure the oscilloscope is properly grounded.
Perform this adjustment to match the characteristics of the probe and the channel
input. This should be performed whenever attaching a probe to any input channel for
the first time.
1. From CH1 menu, set the Probe attenuation to 10X (press CH1→Probe→10X).
Set the switch to 10X on the probe and connect it to CH1 of the oscilloscope.
When using the probe hook-tip, inserting the tip onto the probe firmly to ensure
a proper connection.
Attach the probe tip to the Probe compensator connector and the reference lead
to the ground pin, Select CH1, and then press AUTO.
2. Check the shape of the displayed waveform.
Over compensated Correctly Compensated Under Compensated
Figure1-9
Figure 1- 9
Probe Compensation
3. If necessary, use a non-metallic tool to adjust the trimmer capacitor on the probe
for the flattest square wave possible as displayed on the oscilloscope.
4. Repeat as necessary.
WARNNING: To avoid electric shock while using the probe, be sure the
perfection of the insulated cable, and do not touch the metallic portions of
the probe head while it is connected with a voltage source.
Digital leads are provided only for DS1000D series which have Logic Analyzer.
1. Switch off power supply of the device under test if necessary to avoid short
circuit. Since no voltage is applied to the leads at this step, you may keep the
oscilloscope on.
2. Connect one end of the flat cable FC1868 to the Logic Analyzer Input; connect
the other end to Logic Head LH1116. An identifier is located on each end of the
flat cable; it can only be connected in one way. It is unnecessary to switch off
power supply of your oscilloscope when connecting the cable.
Figure 1- 10
Connect the digital leads
CAUTION: Use only FC1868, LH1116, TC1100 and LC1150 made by
RIGOL for specified DS1000D series.
The oscilloscope has an automated feature to display the input signal best-fit. The
input signal should be 50Hz or higher and a duty cycle is greater than 1%.
Press the AUTO button, the oscilloscope automatically sets up VERTICAL,
HORIZONTAL and TRIGGER controls to display the input signal. Adjust the controls
manually to get the best results if necessary.
Connect a signal to the Channel 1 (CH1).
1. Connect a signal to the oscilloscope as described above.
2. Press AUTO.
The oscilloscope may change the current settings to display the signal; and adjusts
the vertical and horizontal scaling, the trigger coupling, type, position, slope, level,
and mode.
If the channel is DC coupled, measuring the DC component of the signal by
simply noting its distance from the ground symbol.
If the channel is AC coupled, the DC component of the signal is blocked, allow
you to use greater sensitivity to display the AC component of the signal.
To Understand the Vertical System
Figure 1- 14 shows the VERTICAL controls, CH1, CH2, MATH, REF , and OFF buttons
and vertical , knobs. Following the exercise of the buttons,
knobs, and the status bar to be familiar with the vertical parameters settings.
1. Center the signal on the display with the knob.
The knob moves the signal vertically, and it is calibrated. Note that
turning the knob, a voltage value is displayed for a short time
indicating its value with respect to the ground reference located at the center of the
screen. Also notice that the ground symbol on the left side of the display moves in
conjunction with the knob.
Turn the knob to change the vertical display position of
channel and press the knob to set the vertical display position
back to 0 as a shortcut key, this is especially helpful when the trace position is
far out of the screen and want it to get back to the screen center immediately.
Coarse/Fine Shortcut key
The Coarse/Fine vertical control can be set by simply pressing the vertical
knob.
2. Change the vertical setup and notice that each change affects the
status bar differently.
View the status bar which is on the bottom of the screen to understand the
vertical scale.
Change the vertical scale by turning the knob and notice the change
To press the knob in the horizontal control area on the front-panel
is another way to enter or exit Delayed Scan mode and it is equal to the
following menu operations, MENU→Delayed→ON.
To Understand the Horizontal System
Figure 1- 15 shows the HORIZONTAL controls: MENU button, and
knobs of horizontal system. Following the exercise to familiarize with the
buttons, knobs, and status bar.
Figure 1- 15
The horizontal system
1. Turn the knob and notice the change in the status bar.
The horizontal knob changes the sweep speed in a 1-2-5 step sequence,
and displays the value in the status bar. The time base ranges of the oscilloscope is
from 2ns/div* to 50s/div.
* NOTE: The speed of horizontal scan varies by different models.
Press the knob to set the horizontal offset to 0 as a shortcut
key, this is especially helpful when the trigger point is far out of the screen and
want it to get back to the screen center immediately.
Horizontal position control
Trig-Offset: In this setting, the trigger position will be changed horizontally
when you turn the knob.
2. The horizontal knob moves displayed signal horizontally
on waveform window
3. Press the MENU key to display the TIME menu.
To enter or exit the Delayed Scan mode, set the display to Y-T, X-Y or ROLL mode,
and turn the horizontal knob to adjust trigger offset.
Turn the knob to change trigger level value and press the
knob to set trigger level back to 0 as a shortcut key.
To Trigger the Oscilloscope
Figure 1- 16 shows the trigger control: MENU, 50%, FORCE and a trigger
level knob. Following the exercise to familiarize with the buttons, trigger level knob
and status bar.
Figure 1- 16
The trigger control window
1. Turn the trigger Level knob and notice the changes on the display.
As you turn the knob or pressing the 50% button, two things will happen
on the display for a short time.
First, the trigger level value is displayed at the bottom-left of the screen. If the
trigger is DC coupled, it is displayed as a voltage value. If the trigger is AC
coupled or LF reject, it is displayed as a percentage of the trigger range.
Second, a line is displayed showing the location of the trigger level (as long as
AC coupling or low frequency reject are not selected).
Holdoff: A time interval before the oscilloscope response to next trigger
signal. During this holdoff period, the trigger system becomes “blind” to
trigger signals. This function helps to view complex signals such as an AM
waveform. Press Holdoff button to activate () knob, then turn it to adjust
Holdoff time.
2. Change the trigger setup and notice these changes in the status bar.
Press MENU button in the Trigger control.
A soft button menu appears on the display showing the trigger setting choices as
shown in
Figure 1- 17 .
Figure 1- 17
· Press the trigger Mode button and choose Edge.
· Press the trigger Source button to select CH1.
· Press the trigger Slope button to choose .
· Press the trigger Sweep button to select Auto.
· Press the trigger Set Up button to enter secondary menu.
NOTE: The trigger type, slope and source change in conjunction with the status bar
on the top-right of the screen.
3. Press 50%
The 50% button sets the trigger level to the center of the signal.
4. Press FORCE
Starting an acquisition regardless of an adequate trigger signal, usually used in
“Normal” or “Single” trigger mode. This button has no effect if the acquisition is
already stopped.
By now, a user should understand the VERTICAL, HORIZONTAL and TRIGGER
control systems and knows how to determine the system setup by status bar of a
DS1000E, DS1000D series digital oscilloscope.
This chapter will go through all groups of front-panel buttons, knobs and menus;
and further the knowledge of the operation by hints in this guide.
It is recommended to perform all of the following exercises to get the most of the
powerful measurement capabilities of the oscilloscope.
This chapter covers the following topics:
To set up the vertical system ( CH1, CH2, MATH, REF, LA, OFF,
Vertical, Vertical)
To set up the horizontal system ( MENU, Horizontal,
Horizontal)
To set up the trigger system (, MENU, 50%, FORCE)
To set up the sampling system ( Acquire)
To set up the display system ( Display)
To save and recall waveforms, CSV format, bmp format and other setups
( Storage)
To set up utility ( Utility)
To measure automatically ( Measure)
To measure with cursors ( Cursor)
To use run control buttons ( AUTO, RUN/STOP)
Blocks the DC component of the
input Signal
Passes both AC and DC
components of the input signal
Disconnects the input signal
BW Limit
ON
OFF
Limits the channel bandwidth to
20MHz to reduce display noise.
Get full bandwidth.
Probe
1X
5X
10X
50X
100X
500X
1000X
Set this to match your probe
attenuation factor to make the
vertical scale readout correct
Digital filter
Setup digital filter (See table 2-4)
1/2
Go to the next menu page (The
followings are the same, no
more explanation)
To Set up the Vertical System
Settings of the Channels
Each channel has an operation menu and it will pop up after pressing CH1 or CH2
button. The settings of all items in the menu are shown in the table below.
Figure 2- 1 Table 2- 1 The Channel menu (Page 1/2)
To use Channel 1 as an example, input a signal containing high frequency
component.
Press CH1 →BW Limit→OFF, to set up bandwidth limit to “OFF” status. The
oscilloscope is set to full bandwidth and passing the high frequency component in
the signal.
The waveform is displayed as Figure 2- 6:
Figure 2- 6
Turn off the BW limit
Press CH1→BW Limit→ON, to set up bandwidth limit to “ON” status. It will reject
the frequency component higher than 20MHz.
The oscilloscope allows selecting the attenuation factor for the probe. The
attenuation factor changes the vertical scaling of the oscilloscope so that the
measurement results reflect the actual voltage levels at the probe tip.
To change (or check) the probe attenuation setting, press the CH1 or CH2 button
(according to which channel in using). Toggle the Probe soft button to match the
attenuation factor of the probe.
This setting remains in effect until changed again.
Figure 2-8 shows an example for using a 1000:1 probe and its attenuation factor.
Invert turns the displayed waveform 180 degrees, as respect to the ground level.
When the oscilloscope is triggered on the inverted signal, the trigger is also inverted.
Figure 2- 10 and Figure 2- 11 show the changes after inversion.
Add source A to source B
Subtract source B from source A
Multiply source B by source A
Fast Fourier Transform
Source
CH1
CH2
Define CH1 or CH2 as FFT source
Window
Rectangle
Hanning
Hamming
Blackman
Select window for FFT
Display
Split
Full screen
Display FFT waveform on half screen
Display FFT waveform on full screen
Scale
Vrms
dBVrms
Set “Vrms ” as vertical unit
Set “dBVrms ” as vertical unit
Key points for FFT
1. Signals that have a DC component or offset can cause incorrect FFT waveform
component magnitude values. To minimize the DC component, choose AC
Coupling on the source signal.
2. To reduce random noise and aliases components in repetitive or single-shot
events, set the oscilloscope acquisition mode to average.
3. To display FFT waveforms with a large dynamic range, use the dBVrms scale.
The dBVrms scale displays component magnitudes using a log scale.
Using the FFT
The FFT (Fast Fourier Transform) process converts a time-domain signal into its
frequency components mathematically. FFT waveforms are useful in the following
applications:
Measuring harmonic content and distortion in systems
Characterizing noise in DC power supplies
Analyzing vibration
Best frequency
Resolution and worst
magnitude resolution.
This is essentially the
same as no window.
Transients or bursts, the signal
levels before and after the event
are nearly equal.
Equal-amplitude sine waves with
fixed frequencies.
Broadband random noise with a
relatively slow varying spectrum.
Hanning
Hamming
Better frequency,
poorer magnitude
accuracy than
Rectangular.
Hamming has slightly
better frequency
resolution than
Hanning.
Sine, periodic, and narrow-band
random noise.
Transients or bursts where the
signal levels before and after the
events are significantly different.
Blackman
Best magnitude, worst
frequency resolution.
Single frequency waveforms, to
Find higher order harmonics.
Key points:
FFT Resolution: the quotient betweensampling rate and number of FFT
points. With a fixed FFT points, the lower sampling rate results in better
resolution.
Nyquist Frequency
The highest frequency that any real-time digitizing oscilloscope can acquire
without aliasing. It’s normally half of the sample rate. This frequency is called
the Nyquist frequency. Frequency above the Nyquist frequency will be under
sampled, causing a situation known as aliasing.
Selecting an FFT Window
The oscilloscopes provide four FFT windows. Each window is a trade-off between
frequency resolution and amplitude accuracy. What you want to measure and your
source signals characteristics help determine which window to use. Use the
following guidelines to select the best window.
Select memory location in scope
Select memory location out scope
Save
Save REF waveform
Imp./Exp.
Go to import/export menu(see table 2-10)
Reset
Reset REF waveform
Menu
Settings
Comments
Source
CH1
CH2
MATH/FFT
LA
Select channel1 as REF channel
Select channel2 as REF channel
Select Math/FFT as REF channel
Select LA as REF channel (DS1000D series)
Location
Internal
External
Select memory location in scope
Select memory location out scope
Save
Save REF waveform to outer memory
location
Import
Go to import menu(see table 2-14)
Reset
Reset REF waveform
Using REF
Reference Waveforms are saved waveforms to be selected for display. The reference
function will be available after saving the selected waveform to non-volatile memory.
Press REF button to display reference waveform menu.
Figure 2- 18 Table 2- 8 REF menu when using internal memory
Figure 2- 19 Table 2- 9 REF menu when using external memory
Select mode of whole digital channels.
The threshold voltage can set by user
when in user-defined style.
User
<Threshold
Voltage>
Set threshold voltage by turning ()
knob.
Set up LA Channel (DS1000D Series)
Single channel or group channels can be chosen ON or OFF, and also can set the size
of waveform. Change display location of digital channel on screen and select
threshold style.
Press LA function button and go to the following menu.
Figure 2- 31 Table 2- 15 The LA menu
1. Display and re-line up the digital channels
(1) Press LA→D7-D0 or D15-D8 and go to the group channel setting menu. Turn on
or turn off the display of the digital channels.
(2) Press LA→current and choose digital channel by turning () knob. The chosen
channel will display in red color.
(3) Turn vertical knob to re-position the channel in screen.
Display 8 channels in a single screen
Display 16 channels in a single screen
Reset
Reset waveform of channel D7-D0
Menu
Settings
Comments
channel
D15-D8
Turn on or off single channel of D15-D8
D15-D8
Turn on
Turn off
Turn on or off 8 channels together
Size
Display 8 channels in a single screen
Display 16 channels in a single screen
Reset
Reset waveform of channel D15-D8
Set up Channel Group
Press LA→D7-D0 or D15-D8; turn on/off the channel single, or in a group. Also
you can change the size of waveforms in 8 bits as a group. See table 2-16 and 2-17
Figure 2- 34 Table 2- 16 The Digital Channel menu (Page 1)
Figure 2- 35 Table 2- 17 The Digital Channel menu (Page 2)
Press LA →D7-D0→Channel, and choose the wanted channel by turning () knob.
Press No. 1 soft button or push down () knob to turn on /off the channel. When
the channel is on, we can see the mark ( ). When the channel is turned off; the
mark will display as ( ).
As figure 2-36 shows.
Figure 2- 36
Turn on or off the digital channel
2. Force turn on or off all logic channels
Press LA →D7-D0→Turn On / Turn Off (or D15-D8 →Turn On / Turn Off) will force
to turn all the channels on/off. If you want to turn on/off any single channel instead,
select the Channel by turning () knob, then press No. 1 soft button or () knob.
3.Set up the viewing size of logic channels:
Press LA→D7-D0→Size, or D15-D8→Size, to select wave size of logic channels.
Selectto view 8 channels on the screen;Selectto view all of the 16
channels on the screen.
4.Reset the logic channels display:
Press LA→D7-D0→Reset, or D15-D8→Reset to reset the display of logic channels.
The CH1, CH2, Ext. Trigger and LA (DS1000D series) channels are input channels.
All functionalities applied will be based on operating the instrument with channels.
So MATH and REF can be regarded as relatively isolated channels.
To turn on/off any one of the channels, press the corresponding button on the front
panel. The key backlight indicates the channel is currently active. Press the button
again to turn the channel off. Or when channel is currently selected, press OFF will
turn the channel off as well, and the key backlight also goes off.
Table 2- 18 Status of the channels
NOTE:
The channel status symbol is displayed at the lower-left of the screen. Pressing LA
will turn all the digital channels on/off.
You can use the vertical controls to display waveforms, adjust vertical
and , and set input parameters.
1. Using verticalknob.
The vertical control changes the position of signal waveforms in
all channels (including MATH and REF). The resolution changes according to the
vertical level set. Pressing this knob will clear the channel offset to zero. (The
function is available for DS1000D series, but not includes digital channel.)
2. Using verticalknob.
The vertical can change the vertical sensitivity of waveforms in all
channels (including MATH and REF, excluding LA). If the Volts/Div is set to
“Coarse”, the waveform scales in a 1-2-5 step sequence from 2 mV to 5 V. If the
Volts/Div is set to “Fine”, it scales to small steps between the coarse settings.
3. Channels can be adjusted by the verticalandonly when
they are selected.
4. During the vertical position, a position message is displayed on the left bottom
of the screen, in the same color as the corresponding channel. The unit is V
(Volts).
The oscilloscope shows the time per division in the scale readout. Since all active
waveforms use the same time base, the oscilloscope only displays one value for all
the active channels, except when using Delayed Scan, or Alternative Trigger.
The horizontal controls can change the horizontal scale and position of waveforms.
The horizontal center of the screen is the time reference for waveforms. Changing
the horizontal scale causes the waveform to expand or contract about the screen
center.
Horizontal position changes the displayed waveform position, relative to the trigger
point.
The Horizontal Knobs
: The horizontal knob adjusts the horizontal position of
all channel (include Math) waveforms. The resolution of this control
varies with the time base. Pressing this button clears trigger offset
and moves the trigger point to the horizontal center of the screen.
: Use to select the horizontal time/div (scale factor) for the
main or the Delayed Scan time base. When Delayed Scan is enabled, it
changes the width of the window zone by changing the Delayed Scan
time base.
Horizontal Menu.
Press the horizontal MENU button to display the horizontal menu. The settings of
this menu are listed in the following table.
Turn on Delayed Scan mode
Turn off the Delayed Scan mode
Time Base
Y-T
X-Y
Roll
Show the relative relation between
vertical voltage and horizontal
time.
Show CH1 value at X axis; CH2
value at Y axis.
In Roll Mode, the waveform display
updates from right to left.
Y-T: The conventional oscilloscope display format. It shows the voltage of a
waveform record (on the vertical axis) as it varies over time (on the
horizontal axis).
X-Y: XY format displays channel 1 in the horizontal axis and channel 2 in the
vertical axis.
Roll Mode: In this mode, the waveform display rolls from right to left. No trigger
or horizontal offset control of waveforms is available during Roll Mode, and
it’s only available when set to 500 ms/div or slower.
Slow Scan Mode: This mode is available when the horizontal time base is set to
50ms/div or slower. In this mode, the oscilloscope acquires sufficient data
for the left part to the trigger point, then wait for trigger, when trigger
occurs, it continues to draw the rest part from the trigger point to the end of
the right side. When choosing this mode to view low frequency signals, it is
recommended that the channel coupling be set as DC.
Time/Div: Horizontal scale. If the waveform acquisition is stopped (using the
RUN/STOP button), the Time/Div control expands or compresses the
waveform.
Marks Indicator
① The current waveform window’s position in the memory.
② The trigger position in the memory.
③ The trigger position in the current waveform windows.
④ The horizontal time base (main time base).
⑤ The trigger’s horizontal offset according to the center of the window.
The Delayed Scan is a magnified portion of the main waveform window. Use
Delayed Scan to locate and horizontally expand part of the main waveform window
for a more detailed (higher horizontal resolution) analysis of signal. The Delayed
Scan time base setting cannot be set slower than the Main time base setting.
Figure 2- 39
Delayed Scan window
The following steps show you how to use Delayed Scan.
1. Connect a signal to the oscilloscope and obtain a stable display.
2. Press horizontal MENU→Delayed→ON or press horizontal knob to
enter Delayed Scan mode.
The screen splits into two parts. The upper half displays the main waveform window
and the lower half displays an expanded portion of the main waveform window. This
expanded portion of the main window is called the Delayed Scan window. Two
blocks shaded at the upper half; the un-shaded portion is expanded in the lower half.
The horizontal and knobs control the size and position of
the Delayed Scan. The value at bottom of the screen is the main time base and the
This format is useful for studying phase relationships between two signals.
Channel 1 in the horizontal axis(X) and channel 2 in the vertical axis(Y), the
oscilloscope uses a none-trigger acquisition mode, data is displayed as dots.
Figure 2- 40
X-Y display format
NOTE:In Y-T format, all sample rates are available. But in X-Y format, 100 MSa/s
is not available. In common, deceasing the sample rate can display the wavefrom
better.
The following modes or functions will not work in X-Y format.
LA Function (DS1000D series)
Automatic Measurements
Cursor Measurements
REF and MATH Operations
Delayed Scan Mode
Vector Display Mode
Horizontal knob
Trigger Controls
The trigger determines when the oscilloscope starts to acquire data and display a
waveform. When a trigger is set up properly, it can convert unstable displays or
blank screens into meaningful waveforms.
When the oscilloscope starts to acquire a waveform, it collects enough data so that it
can draw the waveform to the left of the trigger point. The oscilloscope continues to
acquire data while waiting for the trigger condition to occur. After it detects a trigger,
the oscilloscope continues to acquire enough data so that it can draw the waveform
to the right of the trigger point.
The trigger control area on the front panel includes a knob and three buttons:
: The knob that set the trigger level; press the knob and the level will
reset to zero.
50%: The instant execute button setting the trigger level to the vertical
midpoint between the peaks of the trigger signal
FORCE: Force to create a trigger signal and the function is mainly used in
Normal and Single mode
MENU: The button that activates the trigger controls menu.
Select CH1 as trigger signal
Select CH2 as trigger signal
Select EXT TRIG as trigger signal
Select power line as trigger signal
Select a digital channel in D15-D0 as
trigger source (for DS1000D series)
Slope
Rising
Falling
Rising &
Falling
Trigger on rising edge
Trigger on falling edge
Trigger on both ring & falling edge
Sweep
Auto
Normal
Single
Acquire waveform even no trigger
occurred
Acquire waveform when trigger occurred.
When trigger occurs, acquire one
waveform then stop
Set up
To go to Set Up menu, see table 2-38
Settings for Edge Trigger
An edge trigger determines whether the oscilloscope finds the trigger point on the
rising or the falling edge of a signal. Select Edge trigger Mode to trigger on Rising
edge, falling edge or rising & falling edge.
Selects CH1 as trigger signal
Select CH2 as trigger signal
Select EXT TRIG as trigger signal
Select a digital channel in
D15-D0 as trigger source( Only
for DS1000D series)
When
(+Pulse width
less than)
(+Pulse width
more than)
(+Pulse width
equal to)
(-Pulse width
less than)
(-Pulse width
more than)
(-Pulse width
equal to)
To select pulse condition
Settings
<Width>
Set required pulse width
Settings for Pulse Width Trigger
Pulse trigger occurs according to the width of pulse. The abnormal signals can be
detected through setting up the pulse width condition.
Figure 2- 43 Table 2- 21 The Pulse Trigger menu (page 1/2)
Set channel 1 as trigger source
Set channel 2 as trigger source
Set EXT. channel as trigger source
When
To select the pulse condition
Time
<Time Set >
To set slope time
Menu
Settings
Comments
Vertical
Select the level that can be adjusted
by
Sweep
Auto
Normal
Single
Acquire waveform even when no
trigger condition is met.
Acquire waveform when trigger
condition is met.
When trigger condition is met, acquire
one waveform and then stop
Set Up
To go to set up menu. See table 2-38
Slope Trigger
Slope trigger sets the oscilloscope as the positive/negative slope trigger within the
specified time.
Figure 2- 50 Table 2- 26 The Slope Trigger menu (Page 1/2)
Figure 2- 51 Table 2- 27 The Slope Trigger menu (Page2/2)
NOTE: Slope time can be set from 20ns to 10s. When a signal meets the trigger
condition, scope will execute the acquisition. You can adjust LEVEL A/ LEVEL B or
both simultaneous by turning the knob.
Set trigger mode for Channel 1
Set trigger mode for Channel 2
Type
Edge
Set Edge Trigger as the trigger
type
Slope
(Rising)
(Falling)
Trigger on rising edge
Trigger on falling edge
Set Up
To go to set up menu. See table
2-38
Alternative Trigger
When alternative trigger is on, the trigger sources come from two vertical channels.
This mode can be used to observe two non-related signals. You can choose two
different trigger modes for the two vertical channels. The options are as follows:
Edge, Pulse, Slope and video. The info of the trigger level of the two channels will be
displayed on the upper-right of the screen.
Figure 2- 52 Table 2- 28 The Alternative menu (Trigger Type: Edge)
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform
and then stop
Set Up
To go to set up menu, see table 2-40
Pattern Trigger (DS1000D Series)
Pattern trigger identifies trigger terms by checking appointed code. The code is
logical relationship of all channels, with high (H), low (L) and ignore(X).
Figure 2- 59 Table 2- 35 The Pattern Trigger menu
Key Points:
H (High): Logic high: voltage is higher than threshold setting.
L (Low): Logic low: voltage is lower than threshold setting.
X (Ignore): Don’t care. If all the channels are ignored, the oscilloscope won’t be
triggered.
Rising Edge () or Falling Edge (): Set the code as an edge of the channel,
rising edge or falling edge. When the edge is appointed, if code settings of other
channels are all true, the oscilloscope will be triggered on the appointed edge. If no
edge is appointed, the oscilloscope will be triggered on the last edge whose code is
You can only appoint one code as edge. If you have appointed an edge, then
appointed another edge in a different channel, and the first appointed edge will
be set to X (Ignore).
Allow all signals pass
Block DC signals
Reject high frequency signals
Reject DC and low frequency
signals
Sensitivity
<Sensitivity Setting>
Set trigger sensitivity
Holdoff
<Holdoff Setting>
Set time slot before another
trigger event
Holdoff
Reset
Reset Holdoff time to 100ns
Menu
Settings
Comments
Sensitivity
<Sensitivity Setting>
Set trigger sensitivity
Holdoff
<Holdoff Setting>
Set time slot before
another trigger event
Holdoff
Reset
Reset Holdoff time to
100ns
Trigger Setup
Set up different trigger settings according to different trigger modes. When choosing
source as D15-D0 in the mode of Edge and Pulse (DS1000D series), only Holdoff is
adjustable. When source is non-digital channel and in slope trigger, only trigger
coupling, trigger sensitivity and Holdoff can be set. For video trigger, Sensitivity and
Holdoff can be set. For the pattern trigger and the duration trigger (DS1000D series),
only Holdoff is adjustable.
Figure 2- 62 Table 2- 38 The Trigger Set Up menu (Settings for trigger coupling,
trigger sensitivity and holdoff)
Figure 2- 63 Table 2- 39 The Trigger Set Up menu (Settings for sensitivity and
Trigger Holdoff can stabilize complex waveform, such as the pulse range. Holdoff
time is the oscilloscope’s waiting period before starting a new trigger. During Holdoff,
oscilloscope will not trigger until Holdoff ends. For instance: To trigger on the first
pulse on a group of them, users can set the holdoff time to Pulse cluster width.
Figure 2- 65
Trigger Holdoff
To use trigger Holdoff:
1. Press the trigger MENU button to display Trigger Menu.
2. Press Set Up key to display trigger set up menu.
3. Turn the multi function knob () to change Holdoff time until waveform is
stable.
4. Pushing Trigger Hold off reset can reset the Holdoff time to its default value.
Trigger can occur from several sources: Input channels (CH1 and CH2), AC Line,
Ext.
CH1 or CH2:
It is the most commonly used trigger source.The channel works when selected
as a trigger source whatever displayed or not.
Ext Trig:
The instrument can be triggered from a third source while acquiring data from
CH1 and CH2. For example, to trigger from an external clock or with a signal
from another part of the test circuit. The Ext trigger sources use an external
trigger signal connected to the EXT TRIG connector. Ext uses the signal directly;
it has a trigger level range of -1.2V to +1.2V.
AC Line:
AC power can be used to display signals related to the power line frequency,
such as lighting equipment and power supply devices. The oscilloscope gets
triggered on its AC power input; an AC trigger signal is not required. When AC
Line is selected as trigger source, the oscilloscope automatically set coupling to
DC, set trigger level to 0V.
2. Sweep Mode:
The sweep mode determines how the oscilloscope behaves in the absence of a
trigger event. The oscilloscope provides three trigger modes: Auto, Normal, and
Single.
Auto:
This sweep mode allows the oscilloscope to acquire waveforms even when it
does not detect a trigger condition. If no trigger condition occurs while the
oscilloscope is waiting for a specific period (as determined by the time-base
setting), it will force itself to trigger.
When forcing invalid triggers, the oscilloscope cannot synchronize the
waveform, and the waveform seems to roll across the display. If valid triggers
occur, the display becomes stable on the screen.
Any factor results in the un-stability of waveforms can be detected by Auto
NOTE: When horizontal control is set under 50 ms/div, Auto mode allows the
oscilloscope not to capture trigger signal.
Normal:
The Normal mode allows the oscilloscope to acquire a waveform only when it is
triggered. If no trigger occurs, the oscilloscope keeps waiting, and the previous
waveform, if any, will remain on the display.
Single:
In Single mode, after pressing the RUN/STOP key, the oscilloscope waits for
trigger. While the trigger occurs, the oscilloscope acquires one waveform then
stop.
3. Coupling:
Trigger coupling determines which signal component passing to the trigger circuit.
Coupling types include AC, DC, LF Reject and HF Reject.
AC: AC coupling blocks DC components and attenuates the signal below
10Hz.
DC: DC coupling passes both AC and DC components.
LF Reject: LF Reject coupling blocks DC component, and attenuates all
signal with a frequency lower than 8 kHz.
HF Reject: HF Reject coupling attenuates all signals with a frequency
higher than 150 kHz.
4. Pre-trigger/delayed trigger:
The data collected before and after trigger.
The trigger position is typically set at the horizontal center of the screen. In the
full-screen display the 6div data of pre-trigger and delayed trigger can be
surveyed. More data (14div) of pre-trigger and 1s delayed trigger can be
surveyed by adjusting the horizontal knob.
This feature is very useful to study the events that led up to the trigger point.
Everything to the right of the trigger point is called post-trigger information. The
delay range (pre-trigger and post-trigger information) depends on the sweep
speed selected.
5. Adjustable trigger sensitivity
To avoid the influence of noise from the physical world, and get the stable trigger,
the trigger circuit has adopted Stickiness. In DS1000E, DS1000D series, the
stickiness is adjustable from 0.1div-1.0div, which means when it sets to 1.0div,
the trigger circuit will not affect any signal with peak-peak amplitude less than
1.0div, so as to avoid the influence of the noise.
To Avoid signal aliasing, select Peak Detect Acquisition.
Figure 2- 70
RIGOL
Signal with Peak Detect Acquisition
The Peak Detect effect is shown as the figure above.
Stop Acquisition: When the scope is acquiring waveforms, the waveforms is in a
live status; when acquisition is stopped, frozen waveform will be displayed, the
position and scale can still be adjusted by vertical control and horizontal control.
The oscilloscope has Real-time sampling rate up to 1GSa/s. At the time base 50ns
or faster, the oscilloscopes use the sine(x)/x interpolation to expand the horizontal
time base.
Equivalent sampling:
Known as Repetitive sampling to get up to 40ps of horizontal resolution (equivalent
25Gsa/s). This mode is good for observing repetitive signals, and it is not
recommended for single-shot or pulse.
Normal:
Oscilloscope acquires signal byequal time interval.
Average Acquisition:
Apply averaging to your signal to remove uncorrelated noise and improve
measurement accuracy. Reduces random or uncorrelated noise in the signal
display. The averaged waveform is a running average over a specified number of
acquisitions from 2 to 256.
Peak Detect:
Peak Detect mode captures the maximum and minimum values of a signal. Finds
highest and lowest record points over many acquisitions.
Display grids and coordinates on
the screen
Turn off the grids
Turn off the grids and coordinates
Brightness
< percentage >
Set up grid brightness
Menu
Display
1s
2s
5s
10s
20s
Infinite
Set the time before menu fades
away. The menu will be hidden
after the set time of last button
pressing.
Adjusting waveform intensity
Default setup of multi-function knob () is adjusting waveform intensity.
Figure 2- 73 Table 2- 43 The Display menu (Page 2/2)
Key points:
Display type: Display type includes Vector and Dot. In vectors type, oscilloscope
connects dots through digital interpolation including both linearity and sin(x)/x.
Sin(x)/x interpolation is suitable for Real-time sampling and will be more effective at
50ns or faster time base.
Refresh rate: It is an important performance of digital oscilloscopes. It means the
number of display refreshing per second and it will affect the ability to observe
signal.
Store or recall waveform
Store or recall instrument setups
Create or delete bit map files
Create or delete CSV files
Recall factory setups
Internal
Go to menu for internal memory
operation (see table 2-48)
External
Go to menu for external memory
operation (see table 2-49)
Disk Mana.
Go to disk manage menu (see
table 2-50)
Storage setup button
RIGOL
To Store and Recall
Figure 2- 74 shows the menu button for the storage system on the front panel.
Figure 2- 74
The Front Panel MENU
Press the Storage button to show the menu for the settings of the storage system.
Waveforms and setups can be stored in and recalled from, both internal memory
and external memory. The waveform file, setup file, bitmap and CSV file in external
memory can be created and deleted. System supports English/Chinese key in.