RIGOL DL3021, DL3031A, DL3021A, DL3031 Users guide

Page 1
RIGOL
User’s Guide
DL3000 Series Programmable DC
Electronic Load
Apr. 2019
RIGOL (SUZHOU) TECHNOLOGIES INC.
Page 2
Page 3

Guaranty and Declaration

Copyright
© 2017 RIGOL TECHNOLOGIES, INC. All Rights Reserved.
Trademark Information
RIGOL is a registered trademark of RIGOL TECHNOLOGIES, INC.
Publication Number
UGJ01107-1110
Software Version
00.01.04 Software upgrade might change or add product features. Please acquire the latest version of the manual from RIGOL website or contact RIGOL to upgrade the software.
Notices
RIGOL produ cts are cove red by P.R.C. and f oreign pa tents, issue d and pendin g. RIGOL reserves the right to modify or change parts of or all the specifications
and pricing policies at the company’s sole decision.
Information in this publication replaces all previously released materials. Information in this publication is subject to change without notice. RIGOL shall not be liable for either incidental or consequential losses in
connection with the furnishing, use, or performance of this manual, as well as any information contained.
Any part of th is d ocu ment is f orbi dden to be c opie d, ph oto copie d, o r rea r ran ged
without prior written approval of RIGOL.
Product Certification
RIGOL guarantees that this product conforms to the national and industrial standards in China as well as the ISO9001:2015 standard and the ISO14001:2015 standard. Other international standard conformance certifications are in progress.
Contact Us
If you have any problem or requirement when using our products or this manual, please contact RIGOL. E-mail: service@rigol.com Website: www.rigol.com
DL3000 User’s Guide I
Page 4
RIGOL

Safety Requirement

General Safety Summary

Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please follow the instructions specified in this manual to use the instrument properly.
Use Proper Power Cord.
Only the exclusive power cord designed for the instrument and authorized for use within the local country could be used.
Ground the Instrument.
The instrument is grounded th rou gh t he Protective Earth lead of the p ower cord. To avoid electric shock, connect the earth terminal of the power cord to the Protective Earth terminal before connecting any input or output terminals.
Connect the Probe Correctly.
If a probe is used, the probe ground lead must be connected to earth ground. Do not connect the ground lead to high voltage. Improper w ay of connection could res ult in dangerous voltages being present on the connectors, controls or other surfaces of the oscilloscope and probes, which will cause potential hazards for operators.
Observe All Terminal Ratings.
To avoid fire or shock hazard, observ e all rat ings and ma rkers on the instrume nt and check your manual for more information about ratings before connecting the instrument.
Use Proper Overvoltage Protection.
Ensure that no over voltage (su ch as that cause d by a bolt of lightni ng) can reach the product. Otherwise, the operator might be exposed to the danger of an electric shock.
Do Not Operate Without Covers.
Do not operate the instrument with covers or panels removed.
Do Not Inser t O bj e cts Into the Air Outlet.
Do not insert objects into the air outlet, as doing so may cause damage to the instrument.
Use Proper Fuse.
Please use the specified fuses.
II DL3000 User’s Guide
Page 5
Avoid Circuit or Wire Exposure.
Do not touch exposed junctions and components when the unit is powered on.
Do Not Operate With Suspected Failures.
If you suspect that any damage may occur to the instrument, have it inspected by RIGOL authorized personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.
Provide Adequate Ventilation.
Inadequate ventilation may cause an increase of temperature in the instrument, which would cause damage to the instrument. So please keep the instrument well ventilated and inspect the air outlet and the fan regularly.
Do Not Operate in Wet Conditions.
To avoid short circuit inside the instrument or electric shock, never operate the instrument in a humid environment.
Do Not Operate in an Explosive Atmosphere.
To avoid personal injuries or damage to the instrument, never operate the instrument in an explosive atmosphere.
Keep Product Surfaces Clean and Dry.
T o a void dust or moisture from af fecting the pe rformance of the inst rument, keep th e surfaces of the instrument clean and dry.
Prevent Ele c tr o static Imp act.
Operate the instrume nt i n an el ectr ostatic dischar ge protectiv e e nvi ron ment to a void damage induced by static discharges. Always ground both the internal and external conductors of cables to releas e sta t i c be fore making connections.
Use the Battery Properly.
Do not expose the battery (if available) to high temperature or fire. Keep it out of the reach of children. Improper change of a battery (lithium battery) may cause an explosion. Use the RIGOL specified battery only.
Handle with Caution.
Please handle with care during transportation to avoid damage to keys, knobs, interfaces, and other parts on the panels.
DL3000 User’s Guide III
Page 6
RIGOL
WARNING
avoided, will result in serious injury or death.
CAUTION
avoided, could result in damage to the product or loss of important data.
DANGER
It calls attention to an operation, if not correctly performed, could result in injury or hazard immediately.
WARNING
It calls attention to an operation, if not correctly performed, could result in potential injury or hazard.
CAUTION
It calls attention to an operation, if not correctly performed, could product.
Hazardous
Voltage
Safety Warning
Protective Earth
Terminal
Chassis Ground
Test Ground

Safety Notices and Symbols

Safety Notic e s in this Manua l:
Indicates a potentially hazardous situation or practice which, if not
Indicates a potentially hazardous situation or practice which, if not
Safety Terms on the Product:
result in damage to the product or other devices connected to the
Safety Symbols on the Product:
IV DL3000 User’s Guide
Page 7
CAUTION
WARNING
supply.

Care and Cleaning

Care
Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time.
Cleaning
Clean the instrument regularly according to its operating conditions.
1. Disconnect the instrument from all power sources.
2. Clean the external surfaces of the instrument with a soft cloth dampened with mild detergent or water. When cleaning the LCD , take care to av oid sca rifying it .
To avoid damage to the instr ument, do n ot expose it to caustic liquids.
To avoid short-circuit resulting from moist ure or personal injuries, ensure that the instrument is completely dry before connecting it to the power

Environmental Consideratio ns

The following symbol indicates that this product complies with the WEEE Directive 2002/96/EC.
Product End-of-Life Handling
The equipment may contain substances that could be har mful to the en vironm ent o r human health. To avoid the release of such substances into the environment and avoid harm to human health, we recommend you to recycle this product appropriately to ensure that most materials are reused or recycled properly. Please contact your local authorities for disposal or recycling information.
You can click on the following link
http://www.rigol.com/Files/RIGOL_RoHS2.0&WEEE.pdf to download the latest
version of the RoHS&WEEE certification file.
DL3000 User’s Guide V
Page 8
RIGOL

DL3000 Series Overview

DL3000 is a cost-effective programmable DC electronic load with high performance. With a user-friendly interface and superb performance specifications, DL3000 series provides variou s i nt e rf aces for remote communication to meet your divers i fi ed test requirements. It can be widely used in various industries, such as automotive electronics, aerospace, and fuel cells.
Main Features:
DL3021/DL3021A: single channel, DC 150 V/40 A, total power up to 200 W DL3031/DL3031A: sin gle c hannel, DC 150 V/60 A, total power up to 350 W Dynamic mode: up to 30 kHz Adjustable current rising speed: 0.001 A/μs to 5 A/μs Min. readback resolution: 0.1 mV, 0.1 mA 4.3-inch TFT LCD, capable of displaying multiple parameters and states
simultaneously
Overvoltage/overcurrent/overpower/overtemperature/reverse voltage
protection
4 static mod e s: CC, CV, CR, CP 3 dynamic modes: continuous, pulsed, toggled List func t ion suppor t s e diting as many a s 512 steps Battery test function, OCP test, OPP test, factory test function, etc. Short-circuit test function Power-off memory function Built-in RS232/USB/LAN communication interface USB-GPIB interface converter (optional)
VI DL3000 User’s Guide
Page 9
Tip
RIGOL (www.rigol.com).

Document Overview

Chapter 1 Quick Start
This chapter introduces some basic information that you s houl d know when you use the DL3000 series electronic load for the first time. This section contains the following contents: out -of-box inspection method, t he appeara nce and dimensions of the instrument, descriptions of the fr ont/rear panel, how to connect the power supply, how to carry out the turn-on checkout, how to replace the fuse, a brief introduction about the user interface, how to use the built-in help system, and how to set the parameters.
Chapter 2 Front Panel Ope r a tio n s
This chapter gives a detailed description about the functions of all the keys on the front panel of DL3000 series and detailed operation met ho ds.
Chapter 3 Remote Control This chapter introduces how to remotely control the instrument.
Chapter 4 Troubleshooting
This chapter introduces the p ossible failu res and solutions in using the DL3000 series electronic load.
Chapter 5 Appendix
This chapter provides the order information and warranty information about the DL3000 series electronic load.
Index
This chapter provides keyword sea rch informat ion, enabling yo u to quickly locat e the desired information.
For the latest version of this manual, download it from the official website of
Format Conventions in this Manual
1. Key
(1) The key on the front panel is denoted by the format of "Key Name (Bold) +
T e xt Box" in the manual. F or example, ON/OFF denotes the "ON/OFF" key .
(2) Use the screen shot to indicate the key. For example, denotes the
Power key.
2. Menu Softkey
The menu softkey is denoted by the format of “Menu Word (Bold) + Character
DL3000 User’s Guide VII
Page 10
RIGOL
Model
No. of Channels
Voltage
Current
Power
DL3021/DL3021A
1
DC 150 V
40 A
200 W
DL3031/DL3031A
1
DC 150 V
60 A
350 W
Shading”. For example, Interface denotes the “Interface” menu softkey under
Utility.
Content Con v entions in th is Ma n ual
The number of channels and the rated values of input parameters for the DL3000 series are listed in the following t able. Unless otherwise specified, this manual takes DL3031A as an example to illustrate the functions and operation methods of the DL3000 se ries.
VIII DL3000 User’s Guide
Page 11
Contents
Guaranty and Declaration ......................................................................... I
Safety Requirement ................................................................................ II
General Safety Summary ........................................................................... II
Safety Not ices and Symbol s ...................................................................... IV
Care and Cleaning .................................................................................... V
Environmental Considerations .................................................................... V
DL3000 Series Overview .......................................................................... VI
Document Overview ............................................................................... VII
Chapter 1 Quic k S tart ......................................................................... 1-1
General Inspection ................................................................................ 1-2
Appearance and Dime nsions ................................................................... 1-3
Front Panel ........................................................................................... 1-4
Rear Panel ........................................................................................... 1-10
Turn-on Checkout ................................................................................. 1-13
To Connect to Power Supply............................................................ 1-13
To Power On the Inst ru ment ........................................................... 1-14
Fuse Replacement ................................................................................ 1-15
User Interface ...................................................................................... 1-16
To Use the Built-in Help System ............................................................. 1-17
Parameter Setting Method ..................................................................... 1-18
Chapter 2 Front Panel Operations ...................................................... 2-1
Local/Remote Operation Mode ................................................................ 2-2
Local Operation Mode ...................................................................... 2-2
Remote Operation Mode .................................................................. 2-2
Static Operation Mode ........................................................................... 2-3
Constant Current (CC) Mode ............................................................ 2-3
Constant Voltage (CV) Mode ............................................................ 2-7
Constant Resistance (CR) Mode ....................................................... 2-11
Constant Power (CP) Mode ............................................................. 2-15
Transient Test Function ......................................................................... 2-18
CC Continuous Mode (Con) ............................................................. 2-18
CC Pulsed Mode (Pul) ..................................................................... 2-25
CC Toggled Mode (Tog) .................................................................. 2-32
List Operation Function (List) ................................................................. 2-38
Application Function ............................................................................. 2-45
OCP Test Function .......................................................................... 2-45
OPP Test Function .......................................................................... 2-53
Battery Test Function ..................................................................... 2-60
Advanced Function ............................................................................... 2-65
Factory Te st F unc tion ..................................................................... 2-65
CC+CV Function ............................................................................ 2-69
DL3000 User’s Guide IX
Page 12
RIGOL Contents
Waveform Display Function ................................................................... 2-73
Input Control ....................................................................................... 2-78
Turn On/Off the Input .................................................................... 2-78
Von Latch Function ........................................................................ 2-78
Short-circuit Operation ................................................................... 2-79
Trigger Operation .......................................................................... 2-80
Protection Function .............................................................................. 2-82
Sense Working Mode ............................................................................ 2-84
Function of Terminals on the Rear Panel ................................................. 2-86
Store and Recall ................................................................................... 2-87
Cursor .......................................................................................... 2-88
Save ............................................................................................. 2-88
Read ............................................................................................ 2-90
Delete .......................................................................................... 2-91
Copy and Paste ............................................................................. 2-91
System Utility Function ......................................................................... 2-92
System ......................................................................................... 2-93
Communication Interface Set ti ng .................................................... 2-97
System Info ................................................................................ 2-105
Reset.......................................................................................... 2-105
Option Configuration .................................................................... 2-110
Chapter 3 Remote Control ................................................................. 3-1
Remote Control via USB .......................................................................... 3-2
Remote Control via LAN .......................................................................... 3-5
Remote Control via GPIB ......................................................................... 3-8
Remote Control via RS232 .................................................................... 3-11
Chapter 4 Troubleshooting ................................................................ 4-1
Chapter 5 Appendix ........................................................................... 5-1
Appendix A: Order Information ................................................................ 5-1
Append i x B: Warranty ............................................................................. 5-4
Index ........................................................................................................ 1
X DL3000 User’s Guide
Page 13

Chapter 1 Quick Start

Contents in this chapter:
General Inspection Appearance and Dime nsions Front Panel Rear Panel Turn-on Checkout Fuse Replacement User Interface To Use the Built-in Help System Parameter Setting Method
DL3000 User’s Guide 1-1
Page 14
RIGOL Chapter 1 Quick Start

General Inspection

1 Inspect the packaging
If the packa gi ng has be en da m age d, do n ot dis po se t he da m age d pac ka gin g o r cushioning materials until the shipment has been check ed for completeness an d has passed both electrical and mechanical tests.
The consigner or carrier shall be liable for the damage to the instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the instrument.
2 Inspect the instrument
In case of any mechanical damage, missing parts, or failure in passing the electrical and mechanical tests, contact your RIGOL sales representative.
3 Check the accessories
Please check the accessories according to the packing lists. If the accessories are damaged or incomplet e , pl e a s e contact your RIGOL sales representative.
1-2 DL3000 User’s Guide
Page 15

Appearance and Dimensions

Figure 1-1 Front View Unit: mm
Figure 1-2 Side View Unit: mm
DL3000 User’s Guide 1-3
Page 16
RIGOL Chapter 1 Quick Start
Pressing this key can enter the constant current (CC) mode. You rate, and starting voltage.
Pressing this key can enter the constant voltage (CV) mode. You can set parameters for the mode, such as voltage and range.
7
12 11 10 9 8
1 2 3 4 5 6

Front Panel

Figure 1-3 Front Panel
1. LCD
4.3-inch TFT LCD, used to display the system status, input parameters, menu setting, prompt message, etc.
2. Waveform Display Key
Pressing this key can ent er the w a vef orm displa y int erface . You can observe the change s of parameters from the dynamic wavefo r m.
3. Function Keys
Provide four static operating modes and three dynamic test functions. In addition, there is also a key used for list operation. (1) CC key
can set parameters for the mode, such as current, range, slew
(2) CV key
1-4 DL3000 User’s Guide
Page 17
Pressing this key can enter the constant resistance (CR) mode. range.
Pressing this key can enter the constant power (CP) mode. You of the voltage.
Pressing this key can enter the continuous (Con) mode. You can rising slew rate, and falling slew rate.
Pressing this key can enter the pulsed (Pul) mode. You can set slew rate, and falling slew rate.
Pressing this key can enter the toggled (Tog) mode. You can set slew rate, and falling slew rate.
Pressing this key can enter the list operation interface. You can
is cycled, a nd the number of steps.
In remote operation mode, when you press this key, you can switch to the local mode.
System
Configures para met ers f or rem ot e commu n icat io n in te rface s
(3) CR key
You can set parameters for the mode, such as resistance and
(4) CP key
can set parameters for the mode, such as power and upper limit
(5) Co n key
set parameters for the mode, such as range, Level A, Level B,
(6) Pul key
parameters for the mode, such as range, Level A, Level B, rising
(7) Tog key
parameters for the mode, such as range, Level A, Level B, rising
(8) List ke y
set parameters such as mode, range, the number of times the list
4. System Function Keys
(1) Local/remote switch key
(2) System utility function key
Sets the system language, power-on value, brightness, short-circuit function, digital input/output, beeper (on/off), Sense function, log, Von Latch function, voltage monitoring output terminal, and current monitoring output terminal.
Interface
(GPIB/USB/RS232/LAN).
DL3000 User’s Guide 1-5
Page 18
RIGOL Chapter 1 Quick Start
System Info
hardware versions, FPGA version, boo t version, sys t e m boot
settings, For default settings, refer to Table 2-3.
Provides advanced fun ctions. Currently, the factory test function
Saves, reads, deletes, copies, and pastes the files. The
files can be stored in the instrument.
Enters the built-in help system. You can view releva nt help Enables or disables the sink function of the load.
Turns on or off the transient trigger function.
Enables or disables the short circuit function of the load.
Displays the manufacturer, device model, software and times, calibration date, and product serial number.
Reset
Restores the instrument settings to its factory default
(3) Advanced function key
and CC+CV function are supported.
(4) Storing and recalling operation key
instrument provides an internal non-volatil e memo ry ("Disk C") and supports an external memory ("Disk D"). A maximum of 100
(5) Help key
information for any key or menu key on the front p an e l .
5. Knob
(1) Switch the parameters or parameter values. (2) Increases (rotate clockwise) or decreases (rotate counterclockwise) the
value at the cursor.
Note: Rotating the knob in different i nte rfaces can produce different operation results. Here introduces its common usage. For other specific usages and operation methods, refer to details in relevant chapters.
6. Input Control Keys
(1) Input on/off key
(2) Transient trigger key
(3) Short-circuit key
1-6 DL3000 User’s Guide
Page 19
Switch the parameter focus.
arrow key to decre ase the val ue at the cu rsor.
Numer ic keys
Backspace key
7. Control Keys
Control keys include the arrow keys, numeric keys, and confirmation key. Pressing the control keys in different interfaces can produce different oper ati o n results.
Note: Here introdu ces t heir com mon usa ge. F o r the ir other specific usages an d operation methods, refer to details in relevant chapters.
(1) Arrow keys (Up/Down/Left/Right arrow key)
M ove t h e cu rsor. Switch th e parameters or parameter values.
When setting parameters, use the Up arrow key to
increase the value at the cursor or use the Down
(2) Confirmation key
Confirms the parameter setting.
(3) Numeric k eys and backspace k ey
8. Application Key
Enables one of the following three application functions: OCP test, OPP test, or Battery test. You can press this key to switch among the three functions.
9. Channel Terminals
Channel terminals include the channel input terminals and Sense terminals, as shown in Figure 1-4.
The numeric keys contain numbers (0-9) and a decimal point. You can use the numeric keys to enter a value for the parameter.
Backsp ace key
Deletes the unwanted characters or numerical values that have been entered.
DL3000 User’s Guide 1-7
Page 20
RIGOL Chapter 1 Quick Start
CAUTION
than 70 V.
Channel Input Terminals
Sense Terminals
A
Outer Nut
Shield
Figure 1-4 Channel Terminals
(1) Channel input terminal: Connects with the device under test (DUT, such as
battery and power supply), which is used for inputting voltage and current.
Note: Connect the positive polarity of the load to the (+) terminal of the
channel output, and the negative polarity of the load to the (-) terminal of
the channel output. (2) Sense terminal: When the DUT outputs large current, the sense terminal
can be used to accurately measure the voltage across the output terminals
of the DUT to compensate for the voltage drop on the load.
Connection methods for the input terminals (see Figure 1-5):
a) Rotate the oute r nut of the in put t erminal counte r clockwise t o remo ve it.
b) Connect the test lead to Position A of the input terminal.
c) Fasten the outer nut of the input terminal clockwise.
d) Secure the shield to the terminal and then tighten the screw.
1-8 DL3000 User’s Guide
Figure 1-5 Connect Input Terminals
To avoid electric shock, ensure that you have installed the shield when the DC voltage output from the DUT is equal to or greater
Page 21
10. Menu Keys
You can press the specified menu key to execute the corresponding operation displayed at the bottom of the interface on the screen.
11. USB Host
(1) The instrument is regarded as the " master device". Y ou can use the data line
to con nect it to t he USB storage device (e.g. U-d isk) to realize external storage or recalling.
(2) Us e the USB-GPIB interface converter to extend the GPIB interface for the
load, and then use the GPIB cable to connect the load to the PC to realize remote contr ol .
12. Power Key
Turn on or off the instrument.
DL3000 User’s Guide 1-9
Page 22
RIGOL Chapter 1 Quick Start
9
11
1 2 3
4

Rear Panel

10
Figure 1-6 Rear Panel
1. Air Outlet
Decrease the temperature inside the instrument to ensure its performance. When you place the instru ment on the wo rkbench or install it i nto the ra ck, kee p the air outlet from a distance of 10 cm to ensure adequate ventilation.
2. Digital I/O
The digital I/O interface supp orts the di gital input (Terminal 1, input voltage 3.3 V or 5 V) and output (Terminal 2, output voltage 3. 3 V) .
Note: This function is a s tandard configuration for DL3031A/DL3021A, and it is an option for DL3031/DL3021. Therefore, for the DL3031/DL3021 model, you need to purchase the digital I/O opti on (with the orde r No. DIGITALIO-DL3).
3. LAN Interface
LAN interface is of the Non-Auto-MDIX type. The instrument can be connecte d to the Local A rea Netw ork (L AN) via the interf ace, so that y ou can realize re mote control for the instrument. The instrument is in compliance with the standards specified in LXI Device Sp ecification 2011. It can be used to set up a test syste m with other standard devices to easily realize the system integration.
4. AC Selector
Selects the voltage that matches with the actual AC input power. For details,
5 6
7 8
1-10 DL3000 User’s Guide
Page 23
CAUTION
scale by using the AC selector.
CAUTION
To avoid electric sho ck or fire disaster , please use th e s pecified fuse and ensure that no short circuit occurs to the fuse holder.
refer to "Fuse Rating" on the rear panel or Table 1-2. The electronic load supports two AC voltage ratings: 115 V and 230 V.
Please
select the proper voltage scale based on the AC rating in your country. When the switch lever is located at a different position, it indicates that a
different voltage rating is selected. When the switch lever is at the left, it indicates that 230 V is se lected. When the switch lev er is at t he right, it in dicates that 115 V is selected.
5. AC Power Cord Connector
AC power input interface. Plug the specified power cord available in the accessories into the AC power cord connector of the instrument, and then connect the instrument to the AC power.
Before connecting the AC power, please select the proper voltage
6. Fuse
When leaving the factory, the instrument has installed a fuse that conforms to the local standard. For fuse replacement, refer to "Fuse Replacement".
7. Fuse Rating
The required fuse rating is related to the actual input voltage, please refer to th e "fuse rating" on the rear panel of the load or refer to Table 1-2.
8. RS232 Interface
Serial communication interface. Use the 9-pin RS232 cable to build communication with the PC to realize remote control.
9. USB Device
The instrument can be regarded as the "standby device" to connect to the external USB dev ice (e.g. PC) to rea l ize remote cont rol.
10. Voltage Monitorin g Output Terminal
The voltage monitoring output terminal outputs the signal in analog quantity. You can connect a device such as a digital oscilloscope through the terminal to display the voltage output from the output terminal, so as to monitor the changes of input voltage of the load.
DL3000 User’s Guide 1-11
Page 24
RIGOL Chapter 1 Quick Start
11. Current Monitoring Output Terminal
The current monitoring output terminal outputs the signal in analog quantity. You can connect a device such as a digital oscilloscope through the terminal to display the current output from the output terminal, so as to monitor the changes of input current of the load.
1-12 DL3000 User’s Guide
Page 25
AC Input Power
AC Selector
±(10% of AC input + 115 Vac), 50 Hz to 60 Hz
115
±(10% of AC input + 230 V ac) (max: 250 V ac), 50 Hz to 60 Hz
230
WARNING
grounded.

Turn-on Checkout

To Connect to Power Supply

Two kinds of AC power inputs are supported by the DL3000 series. When it is connected to different power sources, the AC selector setting on its rear panel is different, as shown in Table 1-1.
Table 1-1 AC Input Power Specification and AC Selector Setting
Please strictly follow the steps below to connect the DL3000 electronic load to the power supply.
1 Check the input power
Ensure that the AC power to be connected t o t he load conforms to the AC input power requirement specified in Table 1-1.
2 Check the AC selector
Ensure that the AC selector setting (115 V or 230 V) on the rear panel of the load matches the actual AC input power (For the match relationship, refer to Table 1-1).
3 Check the fuse
When leaving the factory, the load has installed a fuse that conforms to the destination country standard. Please ref er to the "fuse rating" on the rear panel of the load or refer to Table 1-2 to ensure that the fuse matches the actual AC input power.
4 Connect the AC power
Plug the specified power cord available in the accessories into the AC power supply connector of the instrument, and then connect the instrument to the properly grounded AC power.
To avoid electric shock, ensure that the instrument is correctly
DL3000 User’s Guide 1-13
Page 26
RIGOL Chapter 1 Quick Start
CAUTION
Ensure that the AC selecto r setting on the rear panel of the instrument be burned out.
CAUTION
then enable the input of the electronic load. Otherw is e, the instrument or the device un der test (DUT) wi l l be da maged.
CAUTION
burned out.
Tip
After powerin g off t he elect ronic l oad, plea se wai t for at leas t 1 s bef ore you power it on again.

To Power On the Instrument

After the instrument is con nected t o the power sour ce, press the Power key the left bottom of the front panel to power on the instrument. When the instrument is turned on, it will undergo a self-test. If the instrument passes the self-test, the welcome interface is displayed; otherwise, self-test failure informat ion will be displayed. At this time, please conta ct RIGOL distributors.
matches the actual AC input voltage , otherwise, the electronic load will
After starting the instrument, please connect the cable properly, and
Please pay attention to the positive and negative polarities of the electronic load to avoid wrong connection. Otherwise, the load will be
at
1-14 DL3000 User’s Guide
Page 27
Input Voltage
Fuse Rating
115 Vac
T0.315 A/250 V
230 Vac
T0.20 A/250 V
WARNING
supply before replacing the fuse; to avoid electric shock or fire disaster,
conforms to the voltage specification.
Fuse Holder
Fuse

Fuse Replacement

The fuse specification is related to the actual input voltage, as shown in Table 1-2. Table 1-2 Fuse Rating
When leaving the factory, the instrument has installed a fuse that conforms to the local standard. If the fuse is required to be replaced, select a fuse that matches the actual input voltage and perform the following steps.
1. Turn off the instrument, and then disconnect the instrument from the power source. Remove the power cord.
2. Insert a slotted screwdriver into the slot of the fuse holder to pry it out, as shown in Figure 1-7.
Figure 1-7 Fuse Replacement
3. Take out the fuse and replace it according to the fuse rating (refer to the "Fuse Rating" on the rear panel of the instrument or Table 1-2).
4. Mount the fuse holder (pay attention to its direction when mounting it).
To avoid personal injuries, t urn off the instrume nt and cut off the power before connecting to the AC power, select the voltage specification that
matches the actual input voltage and replace a proper fuse that
DL3000 User’s Guide 1-15
Page 28
RIGOL Chapter 1 Quick Start
No.
Name
Description
1
RIGOL
Company Logo
: indicates that the instrument is remotely controlled.
Displays the function state of the load in a real -time manner.
Actual Input Value
Displays the function menu. Press the specified key below to select the specified menu item.
1
2 3 4
5

User Interface

The user interface of the DL3000 series consists of main interface, guide interface, and function interface. In the main interface, you can set and view the information about the channel input; in the guide interface, you can directly set the parameters according to the guide diagram and view the parameter information; in the function interface (including the waveform interface), you can set and view the information about the functions. When the instru ment powers on , it e nters t he main interf ace by default. This section mainly introduces the main interface of the electronic load. The guide interface and function interface will be introduced in "Front Panel Operations".
The main interface is sh own in Figure 1-8. For the des criptions of the main inte rface, refer to Table 1-3.
Figure 1-8 Main Interface
Table 1-3 Main Interface Description
: indicates that the overtemperature protection occurs.
: indicates that the USB device is recognized.
: indicates that the network is connected. : indicates that GPIB is connected.
2
System Status Icon
: indicates that the beeper is off. : indicates that the beeper is on.
3 Function State 4
Displays the actual input voltage, sink current, etc.
5 Menu Item
1-16 DL3000 User’s Guide
Page 29

To Use the Built-in Help System

The built-in help system provides help information for any key on the front panel (except parameter input area) and menu keys, which are convenient for you to get the descriptions of the function keys or menus, as shown in Figure 1-9.
Figure 1-9 Help Information Interf ace
1. Get the built-in help information Press Help, and then "Help" will be displayed at the top of the interface, being highlighted. At this time, press the corresponding function key or menu key to enter the corresponding help information interface.
2. Page up/down opera tion If the help info rmation is displa yed in seve ral pa ges, press Previous to go to the previous page or press Next to go to the next page. Y ou can also use the arrow keys or the knob to page up/down the help information.
3. Exit the current help information interface When the help information is displayed, press OK to exit the current help information interface and return to the previous interface.
DL3000 User’s Guide 1-17
Page 30
RIGOL Chapter 1 Quick Start

Parameter Setting Method

Most parameters can be set by operating the keys on the front panel. The common setting methods are listed below. The setting method for certain parameters is different from the methods below, refer to the relevant chapters of this manual for the further explanation.
Method 1: Use the numeric keys
1. In the main interface, press the specified menu key to switch the parameter focus; in the guide int erf ace, use the arrow keys to switch the parameter focus.
2. Enter a value by using the numeric key. While entering a name, press delete the unwanted character if necessary.
3. When setting resistance, press the specified menu key and select "Ω" or "" to be the unit. When setting period/width, press the specified menu key and select "ms" or "s" to be the unit. When setting frequency, press the specified menu key and select "Hz" or "kHz" to be the unit. When setting other parameters, press OK to confirm the input.
Method 2: Use the knob or the arrow keys
1. In the main interface, press the specified menu key to switch the parameter focus.
2. Press the Left/Right arrow key to move the cursor to a desired position.
3. Press the Up/Down arrow key or rotate the knob to modify the value.
to
1-18 DL3000 User’s Guide
Page 31

Chapter 2 Front Panel Operations

Contents in this chapter:
Local/Remote Operation Mode Static Operation Mode Transient Test Function List Operation Function (List) Application Function Advanced Function Waveform Display Function Input Control Protection Function Sense Working Mode Function of Terminals on the Rear Panel Store and Recall System Utility Function
DL3000 User’s Guide 2-1
Page 32
RIGOL Chapter 2 Front Panel Operations
Tip
Factory test mode and APP functions (OCP/OPP) do not support re mote operat ion.

Local/Remote Operation Mode

The load provides two operation modes: local and remote.

Local Operation M ode

After you power on the instrument, it ent ers the local operation mode by def ault. In the local operation mode, all the keys on the front panel are available for yo u t o use.

Remote Operation Mode

In the remote operation mode, you can send the programming commands from the PC via any one of the interfaces (GPIB, USB, RS232, or LAN).
In remote operation mode, all the keys (except the Power key key) were disabled. You can only control the load by progr a m ming commands. To return to the lo cal operation mo de , press Local on the front panel.
and the Local
2-2 DL3000 User’s Guide
Page 33
I V Load Current
Set Current
Input Voltage
Front Panel
- +
DUT
- +

Static Operation Mode

The static operation modes include the following 4 modes:
Constant Current (CC) Mode  Constant Voltage (CV) Mode  Constant Resistance (CR) Mode  Constant Power (CP) Mode

Constant Current (CC) Mode

In CC mode, the electronic load will sink a current in accordance with the programmed value regardless of the input voltage, as shown in Figure 2-1.
Figure 2-1 Constant Current (CC) Mode
Operation Procedures:
1. Power on the instrument
Press the Power key
2. Connect the DUT and the channel input terminals of the load
Connect the DUT and t he channel input terminals on the front panel of the load as shown in Figure 2-2.
on the front panel to power on the instrument.
=
Figure 2-2 Connect to the Terminals
DL3000 User’s Guide 2-3
Page 34
RIGOL Chapter 2 Front Panel Operations
CAUTION
the instrument or the DUT.
Tip
from the DUT, short circuit occurs to the DUT.
While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. A misconnection with the terminals may cause damage to
3. Set channel parameters
Press CC to enter the main interfa ce of CC mo de, as sh own in Figure 2-3
. At the top of the main interface, "CC" and "STATIC" are active. Of which, "STATIC" is highlighted.
Figure 2-3 Main Interface of CC Mode
In CC mode, you need to set the following parameters: Current, Range, Slew, Von, V_Limit, C_Limit.
Set Current
The constant current value in CC mode. The default unit for the current is A. (1) In the main interface, press Current to switch the parameter focus to
"Current".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to t he descriptions of the help information or ref er to the descriptions in "Par a m e te r Setting Method".
In CC mode, when the set load current is greater than the current output
Set Range
The working ranges of current in CC mo de are as f oll ows. Diffe rent mo dels hav e different ranges. DL3021/DL3021A: low r a nge (0 to 4 A); high range (0 to 40 A) DL3031/DL3031A: low range (0 to 6 A); high range (0 to 60 A) (1) In the main interface, press Range to switch the parameter focus to
"Range".
2-4 DL3000 User’s Guide
Page 35
CAUTION
re switching the curre nt r ange, please disa ble th e channel in put
to avoid causing damage to the instrument or the DUT.
Tip
90%; from 90% to 10%) changes.
(2) Press Range to switch the current working range. You can also use the
Left/Right arrow key or the knob to switch it.
Note:
The low range provides better resolution and accuracy at low current
settings.
If the set current value is greater than the maximum value of the low range,
you must select a high range.
Befo
Set Slew
The current slew rate for the positive transitions in CC mode. The slew rate refers to the rate at which the input current to a module changes to a new programmed value. The default unit for the slew rate is A/μs. (1) In the main interface, press Slew to switch the parameter focus to "Slew". (2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help info rmation or refer to the descriptions in "Par a m e te r Setting Me th o d ".
The range of the slew rate is determined by the current ratio (from 10% to
Set Von
When the input voltage is greater than the set starting voltage (Von), the load starts to sink the cu rrent. T he def ault unit for Von is V, and it ranges from 0 V to 150 V. The default Von is 0 V.
To enable or disable this function, refer to
descriptions in "Enable/Disable the Von Latch Function". (1) In the main interface, press Von to switch the parameter focus to "Von". (2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help i nformation or refe r to the descriptions in "Par a m e te r Setting Me th o d ".
Set V_Limit
The upper limit of the voltage working in CC mode. The default unit for V_Limit is V, and it ranges from 0 V to 155 V. (1) In the main interface, press V_Limit to switch the parameter focus to
"V_Limit".
(2) Enter a value by using the numeric keys, ar row keys, or the knob. For setti ng
methods, refer to the descriptions of the help information or refer to the descriptions in "Parameter Setting Method".
DL3000 User’s Guide 2-5
Page 36
RIGOL Chapter 2 Front Panel Operations
WARNING
input.
CAUTION
message is displayed, "Fan stops running!"
Set C_Limit
The upper limit of the current working in CC mode. The default unit for C_Limit is A, and it ranges from 0 A to 70 A. (1) In the main interface, press C_Limit to switch the parameter focus to
"C_Limit".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to t he descriptions of the help information or ref er to the descriptions in "Par a m e te r Setting Me th o d ".
4. Turn on the channel input
Press On/Off to turn on the channel input. At this time, the actual input voltage, current, resistance, and power will be displayed in the main interface.
Note: Once the channel input is turned on, the load will not start to sink the current until the input voltage is greater than the starting voltage.
To avoid electric shock, ensure that the DUT is connected to the input terminals of the load properly before you turn on the channel
When the fan stops working, the channel is disabled. Then, a
5. View waveform display
Press the waveform display key
to enter the wa vef orm displa y interfa ce , as shown in Figure 2-4. By default, the current w avef orm is displaye d. When the input voltage changes, the load will sink a constant current. For detailed operations, refer to the de scriptions in "Wav eform Dis play Func tion" section. Press the waveform display key
again to exit the waveform display
interface and return to the main interface of CC mode.
Figure 2-4 Waveform Display Interface of CC Mode
6. Exit CC mode
Press any key on the front panel of the load to exit CC mode.
2-6 DL3000 User’s Guide
Page 37
CAUTION
the instrument or the DUT.
V
I
Input Voltage
Load Current
Set Voltage

Constant Voltage (CV) Mode

In CV mode, the load will s ink enough current to control the in put voltage to the set value. The load acts as a shunt voltage regulator when operating in CV mode , as shown in Figure 2-5.
Figure 2-5 Constant Voltage (CV) Mode
Operation Procedures:
1. Power on the instrument
Press the Power key
2. Connect the DUT and the channel input terminals of the load
Connect the DUT and t he channel input terminals on the front panel of the load as shown in Figure 2-2.
on the front panel to power on the instrument.
While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. A misconnection with the terminals may cause damage to
3. Set channel parameters
Press CV to enter the main i nterfa ce of CV mode , as sh own in Figure 2-6
. At the top of the main interface, "CV" and "STATIC" are active. Of which, "STATIC" is highlighted.
DL3000 User’s Guide 2-7
Page 38
RIGOL Chapter 2 Front Panel Operations
CAUTION
to avoid causing damage to the instrument or the DUT.
Tip
from the DUT, broken circuit occurs to the DUT.
Figure 2-6 Main Interface of CV Mode
In CV mode, you need to s et the followin g parameters: Voltage, Ran ge, V_Limit, C_Limit.
Set Voltage
The constant voltage value in CV mode. The default unit for the voltage is V. (1) In the main interface, press Voltage to switch the parameter focus to
"Voltage".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For the
setting methods, refer t o t he des criptions of the b uilt -in hel p inf ormatio n or refer to the descriptions in "Parame te r Se tting Metho d ".
In CV mode, when the set load voltage is greater than the voltage output
Set Range
Working r ange for the voltage in CV mode . Two ranges are available: low range (0-15 V) and high ra nge (0-150 V). (1) In the main interface, press Range to switch the parameter focus to
"Range".
(2) Press Range to switch the voltage working range. You can also use the
Left/Right arrow key or the knob to switch it.
Note:
The low range provides better resolution and accuracy at low voltage
settings.
If the set v oltage value is greater tha n the maximum value of the low ra nge,
you must select a high range.
Before switching the v olt age range, please disable the channel input
2-8 DL3000 User’s Guide
Page 39
WARNING
input.
CAUTION
message is displayed, "Fan stops running!"
Set V_Limit
The upper limit of the voltage working in CV mode. The default unit for V_Limit is V, and it ranges from 0 V to 155 V. (1) In the main interface, press V_Limit to switch the parameter focus to
"V_Limit".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For the
setting methods, refer t o t he des criptions of the b uilt -in hel p inf ormatio n or refer to the descriptions in "Parameter Setting Method".
Set C_Limit
The upper limit of the current working in CV mode. The default unit for C_Limit is A, and it ranges from 0 A to 70 A. (1) In the main interface, press C_Limit to switch the parameter focus to
"C_Limit".
(2) Enter a value by using the numeric keys, arrow ke ys, or the knob. Fo r setting
methods, refer to the descriptions of the help information or refer to the descriptions in "Parameter Setting Method".
4. Turn on the channel input
Press On/Off to turn on the channel input. At this time, the actual input voltage, current, resistance, and power will be displayed in the main interface.
To avoid electric shock, ensure that the DUT is connected to the input terminals of the load properly before you turn on the channel
When the fan stops working, the channel is disabled. Then, a
5. View waveform display
Press the waveform d isp lay k e y
to enter the wavef orm display interf ace, as shown in Figure 2-7. When you select "U" under "Data", you can view the voltage waveform. When changes occur to the current, the input voltage remains to be the set voltage, unchanged. For detailed operations, refer to the descriptions in "Waveform Disp la y Fun ction" section. Press the waveform display key
again to exit the wav eform displa y int erface a nd retu rn to t he
main interface of CV mode.
DL3000 User’s Guide 2-9
Page 40
RIGOL Chapter 2 Front Panel Operations
Figure 2-7 Waveform Display Interface of CV Mode
6. Exit CV mode
Press any key on the front panel of the load to exit CV mode.
2-10 DL3000 User’s Guide
Page 41
CAUTION
the instrument or the DUT.
I
Load Current
V
Set Resistance (R=V/I)
Input Voltage

Constant Resistance (CR) Mode

In CR mode, the load is regarded as a constant resistance, and changes the current linearly proportional to the input voltage, as shown in Figure 2-8.
Figure 2-8 Constant Resistance (CR) Mode
Note: When in CR mode, please ensure that the input device works in the constant voltage mode.
Operation Procedures:
1. Power on the instrument
Press the Power key
2. Connect the DUT and the channel input terminals of the load
Connect the DUT and the channel input terminals on t he front panel of t he load as shown in Figure 2-2.
on the front panel to power on the instrument.
3. Set channel parameters
Press CR to enter the main interfa ce of CR mode, as shown i n Figure 2-9 top of the main interface, "CR" and "STATIC" are active. Of which, "STATIC" is highlighted.
DL3000 User’s Guide 2-11
While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. A misconnection with the terminals may cause damage to
. At the
Page 42
RIGOL Chapter 2 Front Panel Operations
CAUTION
input to avoid causing damage to the instrument or the DUT.
Figure 2-9 Main Interf ace of CR Mode In CR mode, you need to set the following parameters: RES, Range, V_Limit, C_Limit.
Set RES
The constant resistance value in CR mode. The resistance unit is: Ω or kΩ. (1) In the main interface, press RES to switch the parameter focus to "RES". (2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the des criptions of the help information or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Note: When the loop current is greater than 15 mA, the resistance set in CR mode stays stable and the accuracy is good. When the loop current is smaller than 15 mA, the accuracy becomes poor. Moreover, the smaller the loop current, the poorer the accuracy.
Set Range
Working range for the resistance in CR mode. Two resistance ranges are available: low range (0.08 Ω to 15 Ω); high range (2 Ω to 15 kΩ) (1) In the main interface, press Range to switch the parameter focus to
"Range".
(2) Press Range to switch the resistance working range. You can also use the
Left/Rig h t arrow key or the knob to switch it.
Note:
The low range provides better resolution and accuracy at low resistance
settings.
If the set resistance value is greater than the maximum value of the low
range, you must select a high ran ge.
When the input voltage in CR mode is greater than 10 V, the low range
cannot work normally.
Before switching the resistance range, please disable the channel
2-12 DL3000 User’s Guide
Page 43
WARNING
input.
CAUTION
message is displayed, "Fan stops running!"
Set V_Limit
The upper limit of the voltage working in CR mode. The default unit for V_Limit is V, and it ranges from 0 V to 155 V. (1) In the main interface, press V_Limit to switch the parameter focus to
"V_Limit".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help infor mation or refe r to the descriptions in "Par a m e te r Setting Me th o d ".
Set C_Limit
The upper limit of the current working in CR mode. The default unit for C_Limit is A, and it ranges from 0 A to 70 A. (1) In the main interface, press C_Limit to switch the parameter focus to
"C_Limit".
(2) Enter a value by using the numeric keys, ar row keys, or the knob. For setti ng
methods, refer to the descriptions of the help information or refer to the descriptions in "Parameter Setting Method".
Set Von
When the input voltage is greater than the set starting voltage (Von), the load starts to sink the cu rrent. T he def ault unit for Von is V, and it ranges from 0 V to 155 V. The default Von is 0 V. To enable or disable this function, refer to descriptions in "Enable/Disable the Von Latch Function". (1) In the main interface, press Von to switch the parameter focus to "Von". (2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to t he descriptions of the hel p information or ref er to the descriptions in "Par a m e te r Setting Me th o d ".
4. Turn on the channel input
Press On/Off to turn on the channel input. At this time, the actual input voltage, current, resistance, and power will be displayed in the main interface.
Note: Once the channel input is turned on, the load will not start to sink the current until the input voltage is greater than the starting voltage.
To avoid electric shock, ensure that the DUT is connected to the input terminals of the load properly before you turn on the channel
When the fan stops working, the channel is disabled. Then, a
5. View waveform display
Press the waveform display key
DL3000 User’s Guide 2-13
to enter the wa vef orm displa y interfa ce ,
Page 44
RIGOL Chapter 2 Front Panel Operations
as shown in Figure 2-10. When you select "R" under "Data", you can view the resistance waveform. When changes occur to the input voltage, the current changes linearly to keep the resistance value unchanged. For detailed operations, refer to the de scriptions in "Wav eform Dis play Func tion" section. Press the waveform display key
again to exit the waveform display
interface and return to the main interface of CR mode.
Figure 2-10 Wave form Display Inte r face of CR Mode
6. Exit CR mode
Press any key on the front panel of the load to exit CR mode.
2-14 DL3000 User’s Guide
Page 45
CAUTION
the instrument or the DUT.
V
I
V1
V2
I1
I2
Input Voltage
Load Current
Set Power

Constant Power (CP) Mode

The load sinks a constant power in CP mode. The load current will change linearly proportional to the input voltage to keep sinking a constant power (P = V x I), as shown in Figure 2-11.
Figure 2-11 Constant Power (CP) Mode
Operation Procedures:
1. Power on the instrument
Press the Power key
2. Connect the DUT and the channel input terminals of the load
Connect the DUT and t he channel input terminals on the front panel of the load as shown in Figure 2-2.
on the front panel to power on the instrument.
While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. A misconnection with the terminals may cause damage to
3. Set channel parameters
Press CP to enter the main interface of CP mode, as shown in Figure 2-12
. At the top of the main interface, "CP" and "STATIC" are active. Of w hich, "STATIC" is highlighted.
DL3000 User’s Guide 2-15
Page 46
RIGOL Chapter 2 Front Panel Operations
Figure 2-12 Main Interface of CP Mode
In CP mode, you need to set the following par ameters: P ower, V_Limit, C_Limit.
Set Power
The constant power value in CP mode. The default unit for power is W. (1) In the main interface, press Power to switch the parameter focus to
"Power".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For the
setting methods, refer t o t he des criptions of the b uilt -in hel p inf ormatio n or refer to the descriptions in "Parame te r Se tting Metho d ".
Set V_Limit
The upper limit of the voltage working in CP mode. The default unit for V_Limit is V, and it ranges from 0 V to 155 V. (1) In the main interface, press V_Limit to switch the parameter focus to
"V_Limit".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For the
setting methods, refer to the descriptions of the built-in help information or refer to the descriptions in "Parameter Setting Method".
Set C_Limit
The upper limit of the current working in CP mode. The default unit for C_Limit is A, and it ranges from 0 A to 70 A. (1) In the main interface, press C_Limit to switch th e parameter focus to
"C_Limit".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For the
setting methods, refer to the descriptions of the built-in help information or refer to the descriptions in "Parameter Setting Method".
4. Turn on the channel input
Press On/Off to turn on the c hannel input. At this time, the actual input voltage, current, resistance, and power will be displayed in the main interface.
2-16 DL3000 User’s Guide
Page 47
WARNING
input.
CAUTION
message is displayed, "Fan stops running!"
To avoid electric shock, ensure that the DUT is connected to the input terminals of the load properly before you turn on the channel
When the fan stops working, the channel is disabled. Then, a
5. View waveform display
Press the waveform display key
to enter the wa vef orm displa y interfa ce , as shown in Figure 2-13. When you select "P" under "Data", you can v i ew the power waveform. W hen changes occur to the input voltage, the current changes linearly to keep the power value unchan ged. For detailed operations, ref er to the descriptions in "Wave fo rm Display Function" section. Press the waveform display key
again to exit the waveform display interface and return to the
main interface of CP mode.
Figure 2-13 Wave form Display Inte r face of CP Mode
6. Exit CP mode
Press any key on the front panel of the load to exit CP mode.
DL3000 User’s Guide 2-17
Page 48
RIGOL Chapter 2 Front Panel Operations
CAUTION
the instrument or the DUT.
T A B

Transient Test Function

Transient test f uncti on ena bles the l oad to periodi cally swit ch between tw o set l ev els (Level A and Level B). It can be used to test the tra nsient characteristics of the DUT. Transient operation can only be used in CC mode and supports the following 3 operation modes:
CC continuous mode (Con) CC pulsed mode (Pul) CC toggled mode (Tog)
Before performing the transient operation, first set parameters for the transient operation, such as Level A, Level B, frequency, duty, e tc.

CC Continuous Mode (Con)

In continuous operation, when you enable the transient test, the load current will continuously switch over between Level A and Level B, as shown in Figure 2-14.
Figure 2-14 CC Continuous Mode (Con)
Operation Procedures:
1. Power on the instrument
Press the Power key
2. Connect the DUT and the channel input terminals of the load
Connect the DUT and t he channel input terminals on the front panel of the load as shown in Figure 2-2.
on the front panel to power on the instrument.
While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. A misconnection with the terminals may cause damage to
2-18 DL3000 User’s Guide
Page 49
CAUTION
Before switching the current range, please disa ble th e channel input to avoid causing damage to the instrument or the DUT.
Parameter
Guide
3. Set channel parameters in the guide interface
Press Con to enter the Con guide interface, as shown in Figure 2-15
. In the guide interface, you can configure parameters in the parameter configuration list and view the configuration diagram at the left of the co nfiguration list in the real-time manner.
Diagram
Configuration List
Figure 2-15 Guide Interface of Con Operation
The parameters for t he continuous operation mode include range, Level A, Level B, rising slew rate, falling slew rat e, f req uenc y/ period, width/duty, and trigger source.
Set Range
Working range for the current in Con mode. Two ranges are available. DL3021/DL3021A: low range (0 to 4 A); high range (0 to 40 A) DL3031/DL3031A: low range (0 to 6 A); high range (0 to 60 A) (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Range".
(2) Use the Left/Right arrow key or the knob to switch the current working
range.
Note:
The low range provides better resolution and accuracy at low current
settings.
If the set current value is greater than t he maximum v alue of the low range,
you must select a high range.
Set A_Level
The sink current toggles between a high value and a low value in Con mode. A_Level indicates a high value. The default unit for A_Level is Ampere (A). (1) In the guide interface, press t he Up/Do wn arrow key to switch the
parameter focus to " A_Level".
DL3000 User’s Guide 2-19
Page 50
RIGOL Chapter 2 Front Panel Operations
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Para meter Settin g Method".
Note: The input value of A_Level should be within the set range.
Set B_Level
The sink current toggles between a high value and a low value in Con mode. B_Level indicates a low value. The default unit for B_Level is Ampere (A). (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus t o "B_Level".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Note: The set A_Level and B_Level for a specified model should be within its specified high and low ranges for the model.
Set Slew_RIS
Sets the sink current slew rate for the positive t ransitions f rom Level B to Level A in Con mode. The default unit for the rising slew rate is A/μs. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Slew_RIS".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set Slew_FAL
Sets the sink current sle w rate for the negative trans itions from Level A to Level B in Con mode. The default unit for the falling slew rate is A/μs. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Slew_FAL".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set FREQ/PER
Period: The sum of time during which the si nk current st ays at Level A and Level B. The unit is s or ms. Frequency: The reciprocal of period. The unit is Hz or kHz. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " FREQ/PER".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set Width/Duty
Width: The time du rin g which the sink current stays at Level A when it switches to Level A in Con mode. Its unit is s or ms. Duty: The ratio of duration of Level A to the period when the sink current switches to Level A in Con mod e.
2-20 DL3000 User’s Guide
Page 51
(1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Width/Duty". Then, press the Left/Rig ht arrow key or the knob to select Width or Duty.
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set TRIG_SRC
Three triggers are avai lable in Con mode: BUS trigger, TRAN trigger, and DIGIO trigger. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " TRIG_SRC".
(2) Press the Left/Righ t arro w key or rotate the knob to select the type of the
trigger source.
4. Enter the ma in interface
In the guide interface, after y ou configure the parameters, press Apply to enter the main interface, as shown in
Figure 2-16. At this time, the parameters are successfully configured. "CC" and "Continuous" are activ e at the top of the main interface. "Continuous" is highlighted. In the guide interface, if you give up configuring parameters , you can press Cancel to exit the guide interface and enter the main interface.
Figure 2-16 Main Interface of Con Operation
The menus of the main interfac e ar e displayed in three pages.
Menu ite ms on Page 1: Range, A_Level, B_Level, PER/FREQ, Duty/Width, and Next. Menu items on Pa ge 2: Sle w_RIS, Slew _F AL , TRIG_SRC, Guide, Next, and Back. Menu items on Page 3: Read, Save, and Back. To go to t he next page, p r es s Next; to exit the current page and go back to the previous page, press Back; to go back to the guide interface from the main interface, press Guide.
In the main interface, you can also set parameters based on your own needs. The setting methods for the parameters are as follows:
DL3000 User’s Guide 2-21
Page 52
RIGOL Chapter 2 Front Panel Operations
Set Range
(1) In the main interface, press Range to switch the parameter focus to
"Range".
(2) Press Range to sw itch the curr ent working range. You can also use the
Left/Right arrow key or the knob to switch it.
Set A_Level
(1) In the main interface, press A_Level to s witch th e parameter focus to
"A_Level".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the desc riptions of the help information or ref er to the descriptions in "Par a m e te r Setting Me th o d ".
Set B_Level
(1) In the main interface, press B_Level to switch the parameter focus to
"B_Level".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set PER/FREQ
(1) In the main interface, press PER/FREQ to switch the parameter focus to
"PER/FREQ". (2) Press PER/FREQ to switch between Period and Frequency. (3) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help infor mation or refe r to
the descriptions in "Par a m e te r Setting Me th o d ".
Set Duty/Width
(1) In the main interface, press Duty/Width to switch the parameter focus to
"Duty/Width". (2) Press Duty/Width to switch between Duty and Width. (3) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to
the descriptions in "Par a m e te r Setting Me th o d ".
Set Slew_RIS
(1) In the main interface, press Slew_RIS to switch the parameter focus to
"Slew_RIS". (2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to
the descriptions in "Parameter Set tin g Method".
Set Slew_FAL
(1) In the main interface, press Slew_FAL to switch the parameter focus to
"Slew_FAL".
2-22 DL3000 User’s Guide
Page 53
WARNING
input terminals of the load properly before you enable the trigger.
CAUTION
message is displayed, "Fan stops running!"
Tip
In the guide interface, yo u can only use the numeri c keys to input the v alue for the
he numeri c keys, t he arrow keys, or
the knob to input the value for the parameter.
Tip
The load sinks the current of Level B, and then the input will be updated
sinks the current of Level B, and then waits trigger to o ccur. After the trigger is
interface.
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to t he descriptions of the hel p information or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set TRIG_SRC
(1) In the guide interface, press TRIG_SRC to switch the parameter focus to
"TRIG_SRC".
(2) Press TRIG_SRC to select the type of the trigger source. You can also use
the Left/Right arrow key or the knob to select it.
parameter; in the main interface, y ou can use t
5. Enable trigger
After you enable the trigger, the actual input voltage, current, resistance, and power of the load will be displayed in the main interface.
Note: In Con mode, if the trigger is always enabled, the load will run based on the set parameters.
You can enable the trigger first and then the load in put is automatically turned
on ( according to the preset value). You can also press the ON/OFF key on the front panel first to turn on the input before you enable the trigger (The load
enabled, the input will be updated according to th e prese t value) .
After the trigger is enabled, the load current displayed in the main interface
may stay at a constant value, because it is very fast for the current to transition from Level A to Level B. At this time, you can observe that the load current toggles between Level A and Level B in the dynamic wa veform displa y
To avoid electric shock, ensure that the DUT is connected to the
When the fan stops working, the channel is disabled. Then, a
DL3000 User’s Guide 2-23
Page 54
RIGOL Chapter 2 Front Panel Operations
6. View waveform display
After you enter the main interface, press the waveform display key
to enter the waveform display interface, as shown in Figure 2-17. By default, the waveform is displayed in Fast mode. You can observe that the load current toggles continuously between Level A and Level B. For detailed operations, refer to the descriptions in "Wa veform Display Function" section. Press the waveform display key
again to exit the waveform display interface and
return to the main interface of Con mode.
Figure 2-17 Wav eform Display Interface of Con Mode
7. Save and read files
In the guide interface and main interface of Con mode, you can press Save to save the parameters for Con mode to the internal or external memory, and you can read and recall it if necessary by pressing Read.
(1) Save
After you complete the parameter setting, press Save to enter the storage and recall interface. The file is saved in ".CON" format. For detailed saving operation, refer to the descriptions in "
Save".
(2) Read
Press Read to enter the storage and recall interface. Recall and read the files in ".CON" format or the ".RSF" files beginning with "CON_". For detailed reading operation, refer to descriptions in "
Read". The files to be
recalled and read are allowed to be edited.
How to press the Store key on the front to perform the saving and reading operation, refer to the descriptions in "Store and Recall".
8. Exit Con mode
Press any key on the front panel of the load to exit Con mode.
2-24 DL3000 User’s Guide
Page 55
CAUTION
the instrument or the DUT.
A
B
TRIG
TRIG
TWD
TWD

CC Pulsed Mode (Pul)

In the Pul mode, when you enable the transient test operation, the load current will switch from Level B to Level A once the load receives a trigger signal. After maintaining Level A for a specified pulse width, it will switch to Level B, as shown in Figure 2-18.
Figure 2-18 CC Pulsed Mode (Pul)
Operation Procedures:
1. Power on the instrument
Press the Power key
2. Connect the DUT and the channel input terminals of the load
Connect the DUT and t he channel input terminals on the front panel of the load as shown in Figure 2-2.
on the front panel to power on the instrument.
While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. A misconnection with the terminals may cause damage to
3. Set channel parameters in the guide interface
Press Pul to enter the Pul guide interface, as shown in Figure 2-19
. In the guide interface, you can config ure parameters in the parameter configuration list and view the configuration diagram at the left of the configuration list in the real-time manner.
DL3000 User’s Guide 2-25
Page 56
RIGOL Chapter 2 Front Panel Operations
CAUTION
Before switching the current range, please disa ble th e channel input to avoid causing damage to the instrument or the DUT.
Figure 2-19 Guide Interface of Pul Operation
The parameters for the pulsed operation mode include range, Level A, Level B, rising slew rate, falling slew rate, width, and trigger source.
Set Range
The current working range in Pul mode. Two current ranges are available: DL3021/DL3021A: low range (0 to 4 A); high range (0 to 40 A) DL3031/DL3031A: low range (0 to 6 A); high range (0 to 60 A) (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Range".
(2) Use the Left/Right arrow key or the knob to switch the current working
range.
Note:
The low range provides better resolution and accuracy at low current
settings.
If the set current value is greater than t he maximum v alue of the low range,
you must select a high range.
Set A_Level
The sink current toggles between a high value and a low value in Pul mode. A_Level indicates a high value. The default unit for A_Level is Ampere (A). (1) In the guide interface, press the Up/Down arrow key to swi tch the
parameter focus to " A_Level".
(2) Enter a value by using the numeric keys. For the setting methods, refer to
the descriptions in "Parameter Set tin g Method".
Note: The input value of A_Level should be within the set range.
2-26 DL3000 User’s Guide
Page 57
Set B_Level
The sink current toggles between a high value and a low value in Pul mode. B_Level indicates a low value. The default unit for B_Level is Ampere (A). (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " B_Level".
(2) Enter a value by using the numeric keys. For the setting methods, refer to
the descriptions in "Par a m e te r Setting Me th o d ".
Note: The set A_Level and B_Level for a specified model should be within its specified high and low ranges for the model.
Set Slew_RIS
Sets the sink current slew rate for the positive t ransitions f rom Level B to Level A in Pul mode. The default unit for the rising slew rate is A/μs. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Slew_RIS".
(2) Enter a value by using the numeric keys. For the setting methods, refer to
the descriptions in "Par a m e te r Setting Me th o d ".
Set Slew_FAL
Sets the sink current slew r ate for the negati v e transitions from Lev el A to Level B in Pul mode. The default unit for the falling slew rate is A/μs. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Slew_FAL".
(2) Enter a value by using the numeric keys. For the setting methods, refer to
the descriptions in "Parameter Setting Method".
Set Width
The time during whi ch the sink current st ays at Level A when it switches to Level A in Pul mode. Its unit is s or ms. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Width".
(2) Enter a value by using the numeric keys. For the setting methods, refer to
the descriptions of the built-in help information or refer to the descriptions in "Parameter Setting Method".
Set TRIG_SRC
Three triggers are available in Pul mode : B US trigger, TRAN trigger, and DIGIO trigger. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " TRIG_SRC".
(2) Press the Left/Right arrow key to select the type of the trigger source.
4. Enter the ma in interface
In the guide interface, after you configure the parameters, press Apply to enter the main interface, as shown in
Figure 2-20. At this time, the parameters are
DL3000 User’s Guide 2-27
Page 58
RIGOL Chapter 2 Front Panel Operations
successfully configured. "CC" and "Pulse" are active at the top of the main interface. Of which, "Pulse" is highlighted. In the guide interface, if you give up configuring parameters , you can press Cancel to exit the guide interface and enter the main interface.
Figure 2-20 Main Interface of Pul Operation
The menus of the main interfac e ar e displayed in two pages .
Menu ite ms on Page 1: Range, A_Level, B_Level, Slew_RIS, Slew_FAL, and Next. Menu ite ms on Page 2: Width, TRIG_SRC, Guide, Read, Save, and Back. To go to the next page, press Next; to exit the current page and go back to the previous page, press Back; to go back to the guide interface from the main interface, press Guide.
In the main interface, you can also set parameters based on your own needs. The setting methods for the parameters are as follows:
Set Range
(1) In the main interface, press Range to switch the parameter focus to
"Range".
(2) Press Range to switch the current working range. You can also use the
Left/Right arrow key or the knob to switch it.
Set A_Level
(1) In the main interface, press A_Level to switch the parameter focus to
"A_Level".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For the
setting methods, refer t o t he des criptions of the b uilt -in hel p inf ormatio n or refer to the descriptions in "Parameter S etting Meth o d".
Set B_Level
(1) In the main interface, press B_Level to switch the parameter focus to
"B_Level".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to t he descriptions of the hel p information or refer to the descriptions in "Par a m e te r Setting Me th o d ".
2-28 DL3000 User’s Guide
Page 59
Tip
keys, or the knob to input the value for the parameter.
Tip
the input will be updated according to the preset value).
Set Slew_RIS
(1) In the main interface, press Slew_RIS to switch the parameter focus to
"Slew_RIS".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set Slew_FAL
(1) In the main interface, press Slew_FAL to switch the parameter focus to
"Slew_FAL".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set Width
(1) In the main interface, press Width to switch the parameter focus to
"Width".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help information or ref er to the descriptions in "Parameter Set tin g Method".
Set TRIG_SRC
(1) In the main interface, press TRIG_SRC to switch the parameter focus to
"TRIG_SRC".
(2) Press TRIG_SRC to select the type of the t ri gge r source. You can also use
the Left/Right arrow key or the knob to select it.
In the guide interface, you can only use the numeric keys to input the value for the parameter; in the main interface, you can use the numeric keys, the arrow
5. Enable trigger
After you enable the trigger, the actual input voltage, current, resistance, and power of the load will be displayed in the main interface.
You can ena ble the trigger first and then the load i nput is aut omatically turne d on (The load sinks the current of Level B, and then the input will be updated according to the preset value). You can also p ress the ON/OFF key on the front panel first to turn on the input before you enable the trigger (The load sinks the current of Level B, and then waits trigger to occur. After the trigger is enabled,
DL3000 User’s Guide 2-29
Page 60
RIGOL Chapter 2 Front Panel Operations
CAUTION
message is displayed, "Fan stops running!"
WARNING
input terminals of the load properly before you enable the trigger.
To avoid electric shock, ensure that the DUT is connected to the
When the fan stops working, the channel is disabled. Then, a
6. View waveform display
After you enter the main interface, press the waveform display key
to enter the waveform display interface, as shown in Figure 2-21. After t he load current stays at Level A for a specified time (reach the set width), the trigger is disabled automatically. Then, the current automatically switches to Level B. For detailed operations, refer to the descriptions in "Waveform Display Function" section. Press the waveform display key
again to exit the
waveform display interface and return to the main interface of Pul mode.
7. Save and read files
2-30 DL3000 User’s Guide
Figure 2-21 Waveform Display Interface of Pul Mode
In the guide interface and main interface of Pul mode, you can press Save to save the parameters for Pul mode to the internal or external memory, and you can read and recall it if necessary by pressing Read.
(1) Save
After you complete the parameter setting, press Save to enter the storage and recall interface. The file is saved in ".PUL" format. For detailed saving operation, refer to the descriptions in "
Save".
(2) Read
Press Read to enter the storage and recall interface. Recall and read the files in ".PUL" format or the ".RS F" files beginning with " PUL_". F or detailed reading operation, refer to descriptions in "
Read". The files to be recalled
and read are allowed to be edited.
How to press the Store key on the front to perform the saving and reading
Page 61
operation, refer to the descriptions in "Store and Recall".
8. Exit Pul mode
Press any key on the front panel of the load to exit Pul mode.
DL3000 User’s Guide 2-31
Page 62
RIGOL Chapter 2 Front Panel Operations
CAUTION
the instrument or the DUT.
TRIG
TRIG
A
B

CC Toggled Mode (Tog)

In Tog mo de, when y ou e nable the t ransient te st, th e load cu rrent will t oggle f or one time between Level A and Level B once the load re ceives a trigger si gnal, as shown in Figure 2-22.
Figure 2-22 CC Toggled Mode (Tog)
Operation Procedures:
1. Power on the instrument
Press the Power key
2. Connect the DUT and the channel input terminals of the load
Connect the DUT and t he channel input terminals on the front panel of the load as shown in Figure 2-2.
on the front panel to power on the instrument.
3. Set channel parameters in the guide interface
2-32 DL3000 User’s Guide
While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. A misconnection with the terminals may cause damage to
Press Tog to enter the Tog guide interface, as shown in Figure 2-23
. In the guide interface, you can configure parameters in the parameter configuration list and view the configuration diagram at the left of t he configuration list in the real-time manner.
Page 63
CAUTION
to avoid causing damage to the instrument or the DUT.
Figure 2-23 Guide Interface of T o g Operation
The parameters for the to ggle d ope r at io n m ode i ncl ude r a n ge, Lev el A, Level B,
rising slew rate, falling slew rate, and trigger source.
Set Range
The current working range in Tog mode. Two current ranges are available: DL3021/DL3021A: low range (0 to 4 A); high range (0 to 40 A) DL3031/DL3031A: low range (0 to 6 A); high range (0 to 60 A) (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Range".
(2) Use the Left/Right arrow key or the knob to switch the current working
range.
Note:
The low range provides better resolution and accuracy at low current
settings.
If the set current value is greater than t he maximum v alue of the low range,
you must select a high range.
Before switching the current ran ge, please disable the channel input
Set A_Level
The sink current toggles between a high value and a low value in Tog mode. A_Level indicates a high value. The default unit for A_Level is Ampere (A). (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " A_Level".
(2) Enter a value by using the numeric keys. For the setting methods, refer to
the descriptions in "Par a m e te r Setting Me th o d ".
Note: The input value of A_Level should be within the set range.
DL3000 User’s Guide 2-33
Page 64
RIGOL Chapter 2 Front Panel Operations
Set B_Level
The sink current toggles between a high value and a low value in Tog mode. B_Level indicates a low value. The default unit for B_Level is Ampere (A). (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to "B_Level".
(2) Enter a value by using the numeric keys. For the setting methods, refer to
the descriptions in "Par a m e te r Setting Me th o d ".
Note: The set A_Level and B_Level for a specified model should be within its specified high and low ranges for the model.
Set Slew_RIS
Sets the sink current sle w rate for t he positive t ransitions from Level B to Level A in Tog mode. The default unit for the rising slew rate is A/μs. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Slew_RIS".
(2) Enter a value by using the numeric keys. For the setting methods, refer to
the descriptions in "Para me te r Setting Me th o d ".
Set Slew_FAL
Sets the sink current s lew rate f or t he negative tr ansitions from LevelA to Leve lB in Tog mode. The default unit for the falling slew rate is A/μs. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Slew_FAL".
(2) Enter a value by using the numeric keys. For the setting methods, refer to
the descriptions in "Par a m e te r Setting Me th o d ".
Set TRIG_SRC
Three triggers are available in Pul mode: BUS trigger, TRAN trigger, and DIGIO trigger. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " TRIG_SRC".
(2) Press the Left/Right arrow key to select the type of the trigger source.
4. Enter the ma in interface
In the guide interface, after y ou configure the parameters, press Apply to enter the main interface, as shown in
Figure 2-24. At this time, the parameters are successfully configured. "CC" and "Toggle" are active at the top of the main interface. Of which, "Toggle" is highlighted. In the guide interface , if you giv e up configuring parameters , you can press Cancel to exit the guide interface and enter the main interface.
2-34 DL3000 User’s Guide
Page 65
Figure 2-24 Main Interface o f Tog Operation
The menus of the main interfac e ar e displayed in two pages .
Menu ite ms on Page 1: Range, A_Level, B_Level, Slew_RIS, Slew_FAL, and Next. Menu ite ms on Page 2: TRIG_SRC, Gu i de, Read, Save, and Back. To go to t he next page, p r es s Next; to exit the current page and go back to the previous page, press Back; to go back to the guide interface from the main interface, press Guide.
In the main interface, you can also set parameters based on your own needs. The setting methods for the parameters are as follows:
Set Range
(1) In the main interface, press Range to switch the parameter focus to
"Range".
(2) Press Range to switch the current working range. You can also use the
Left/Rig h t arrow key or the knob to switch it.
Set A_Level
(1) In the main interface, press A_Level to switch the parameter focus to
"A_Level".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For the
setting methods, refer t o t he des criptions of the b uilt -in hel p inf ormatio n or refer to the descriptions in "Parameter Setting Method".
Set B_Level
(1) In the main interface, press B_Level to switch the parameter focus to
"B_Level".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For th e
setting methods, refer t o t he des criptions of the b uilt -in hel p inf ormatio n or refer to the descriptions in "Parameter Setting Method".
Set Slew_RIS
(1) In the main interface, press Slew_RIS to switch the parameter focus to
"Slew_RIS".
DL3000 User’s Guide 2-35
Page 66
RIGOL Chapter 2 Front Panel Operations
WARNING
input terminals of the load properly before you enable the trigger.
CAUTION
message is displayed, "Fan stops running!"
Tip
arrow keys, or the knob to input the value for the parameter.
Tip
enabled, the input will be updated according to the preset value).
(2) Enter a value by using the numeric keys , arrow keys , or the kn ob. F or setting
methods, refer to the descriptions of the help information or refer to the descriptions in "Parameter Setting Method".
Set Slew_FAL
(1) In the main interface, press Slew_FAL to switch the parameter focus to
"Slew_FAL".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set TRIG_SRC
(1) In the main interface, press TRIG_SRC to switch the parameter focus to
"TRIG_SRC".
(2) Press TRIG_SRC to select the t y pe of the trigger source. You can also use
the Left/Right arrow key or the knob to s e l e ct it.
In the guide interface, you can only use the numeric keys to input the value for the parameter; in the main interface, you can use the numeric keys, the
5. Enable trigger
After you enable the trigger, the actual input voltage, current, resistance, and power of the load will be displayed in the main interface.
You can enable the trigger first and then the load inp ut is automatic ally turned on (The load sinks the current of Level B, and then the input will be updated according to the preset value). You can also press the ON/OFF key on the front panel first to turn on the input before you enable the trigger (The load sinks the current of Level B, and then waits trigger to occur. After the trigger is
To avoid electric shock, ensure that the DUT is connected to the
When the fan stops working, the channel is disabled. Then, a
2-36 DL3000 User’s Guide
Page 67
6. View waveform display
After you enter the main interface, press the waveform display key
to enter the waveform display interf ace , as sh own in Figure 2-25. The load current will always sta y at Level A and it will not swit ch to Le vel B unless you disable the trigger manually. For detailed operations, refer to the descriptions in "Waveform Display Function" section. Press th e wavef orm display key
again to exit the wav eform display inte rface and return to the main interfac e
of Tog mode.
Figure 2-25 Wave fo rm Display Interface of Tog Mode
7. Save and read files
In the guide interface and main interface of Tog mode, you can press Save to save the parameters for Tog mode to the internal or external memory, and you can read and recall it if necessary by pressing Read.
(1) Save
After you complete the parameter setting, press Save to enter the storage and recall interface. The file is saved in ".TOG" format. For detailed saving operation, refer to the descriptions in "
Save".
(2) Read
Press Read to enter the storage and recall interface. Recall and read the files in ".TOG" format or the ".RSF" f iles beginning with "TOG_". For detailed reading operation, refer to descriptions in "
Read". The files to be
recalled and read are allowed to be edited.
How to press the Store key on the front to perform the saving and reading operation, refer to the descriptions in "Store and Recall".
8. Exit Tog mode
Press any key on the front panel of the load to exit Tog mode.
DL3000 User’s Guide 2-37
Page 68
RIGOL Chapter 2 Front Panel Operations
CAUTION
the instrument and the devices connected to it.
0
TRIG
Circle=1
Circle=2
0
TRIG
Circle=1
Circle=2
1
2 3 4 5 0
TRIG
Circle=1
Circle=2

List Operation Function (List)

In list operation mode, you can perform the complex current/voltage/resistance/power modes accurately and rapidly, which may be synchronized with internal or external signals. In this way, you can complete the sophisticated test. In this mode, when receiving a trigger signal, the load will operate according to the set execution mode and sequence. As shown in Figure 2-26, the load simulates the complex sequences of in put chan ges base d on the li st pa ra meters that y ou e dit . The list function supports CC, CV, CR, and CP modes.
Figure 2-26 List Operation Function (List)
When selecting different trigger sources, you can generate complex sequences by editing the setting value for each step, the dwell time, and slew rate (the slew rate can only be edited in CC mode) to meet the complex test demands. All listed data is stored in a n on-v olatile memory. It means that the data will not be lost when th e load is turned off.
Operation Procedures:
1. Power on the instrument
Press the Power key
2. Connect the DUT and the channel input terminals of the load
Connect the DUT and t he channel input terminals on the front panel of the load as shown in Figure 2-2.
on the front panel to power on the instrument.
While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. A misconnection with the terminals may cause damage to
2-38 DL3000 User’s Guide
Page 69
Actual Input Parameter
Parameter Diagram
3. Set channel parameters
Press List to enter the list function interface, as shown in Figure 2-27
.
Preview
Values
List
Figure 2-27 List Function Interface
The menus of the list function interface are displayed in two pages. Menu items on Page 1: Mode, Range, Cycles, Steps, EndState, and Next. Menu ite ms on Page 2: TRIG_SRC, Save, Read, Recor d, and Back . To go to t he next page, p r es s Next; to exit the current page and go back to the previous page, press Back.
The setting methods for the parameters are as follows:
Set Mode
Sets the working modes that the List function supports. The modes include CC, CV, CR, and CP. (1) In the List function interface, p ress Mode to switch the parameter focus to
"Mode".
(2) Press Mode to switch the working mode. Yo u can also use the Left/Right
arrow key or the knob to swi t ch it.
Set Range
Selects a different ran ge based o n the current w orking mode of List . There is no range available in CP mod e . If "CC" is sele cted to be the working mode, the ran ge selection is as follows:
DL3021/DL3021A: low range (0 A to 4 A); high range (0 A to 40 A) DL3031/DL3031A: low range (0 A to 6 A); high range (0 A to 60 A)
If "CV" is selected to be the working mo de, the ran ge selection is as follows:
low range (0 V to 15 V); high range (0 V to 150 V)
If "CR" is s elected to be th e working mode , the rang e selection is as follows:
low range (0.08 Ω to 15 Ω); high range (2 Ω to 15 kΩ)
(1) In the List function interfa ce, press Range to switch the parameter focus to
"Range".
DL3000 User’s Guide 2-39
Page 70
RIGOL Chapter 2 Front Panel Operations
CAUTION
avoid causing damage to the instrument or the DUT.
Tip
You can edit up to 512 groups of data.
(2) Press Range to switch the workin g ran ge for dif ferent modes. You can al so
use the Left/Right arrow key or the knob to switch it.
Note:
The low range provides better resolution and accuracy at low settings. If the set value is greater than the maximum value of the low range, you
must select a high range.
Before switching the range, please disable the channel input to
Set Circle
Sets the number of times t he list is cycled whe n t iming input is completed based on the preset current/voltage/resistance/power. Its range is from 0 to 99999. When the value is 0, the value under "Circle" is displayed to be "Infinite", and then the load will be switched to infinite mode. (1) In the List function interface, press Circle to switch the parameter focus to
"Circle".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set Steps
Sets the number of steps for the input preset current/voltage/resistance/power in each cycle. Its range is from 2 to 512. (1) In the List function interface, press Steps to switch the parameter focus to
"Steps".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For the
setting methods, refer to the descriptions of the built-in help information or refer to the descriptions in "Parameter Setting Method".
Total Steps = Steps x Circle
Set EndState
Last state: when the number of execution times (Circle) is a finite value, last state refers to the state the load stays after the load completes the total steps of current/voltage/resistance/power inputs. Off: indicates that the input will be turned off automatically after the input is completed. Last: k eeps the input state of the last group after the input is comp l e te d. (1) In the List function interface, press EndState to switch the parameter
focus to " EndState".
2-40 DL3000 User’s Guide
Page 71
Tip
disabled and grayed out.
(2) Press EndState to switch the last working state. You can also use the
Left/Right arrow key or the knob to switch it.
When the number of execut ion times (Circle) is set to "Infinite", "EndState" is
Set TRIG_SRC
Three triggers are available in List mode: BUS trigger, TRAN trigger, and DIGIO trigger. (1) In the main interface, press TRIG_SRC to switch the parameter focus to
"TRIG_SRC".
(2) Press TRIG_SRC to select the type of the trigger source. You can also use
the Left/Right arrow key or the knob to select it.
Set Parameter List
The parameter list consists of the following parameters: Step, Value, Duration, Slew (only available in CC mode), V_Sampling (V)/C_Sampling (A). Step indicates the number of steps; Value indicates the value set for each step; Duration indicates the dwell time for each step; Slew indicates the slew rate for each step; V_Sampling (V) indicates the readback sampling voltage; and C_Sampling (A) indicates the readback sampling current. If the scrollbar is shown at the right side of the para meter list, it indicates tha t not all parameters are displayed in the cur rent interface. At this time, press the Up/Down arrow ke y to go to the next page.
(1) Select Step
The Step value in t he para meter list is relat ed to the setting of "Steps" in the menu. For example, press Steps to set it to 4, then the first 4 steps in the parameter list and the relevant parameters are highlighted.
If the number of steps that you set is a small value, you only need to use the Left/Right arrow key or the knob to directly select the desired number of steps. If the number of steps that you set is a large value, you can input the desired steps by using the numeric keys first. After that, you will di rectly go to the column under the desired steps. For example, if you set Steps to 300, and you want to set the parameters for Step 250, then use the numeric keys to input 250 and you will go to Step 250 directly. At this time, you can set the relevant parameters.
If the parameter focus is currently not on the "Step" line, first press the Up/Down arrow key to switch the parameter focus to the "Step" line, and then select the desired steps.
DL3000 User’s Guide 2-41
Page 72
RIGOL Chapter 2 Front Panel Operations
(2) Enter Value
In CC mode, the set value is for current, and its default unit is A. In CV mode, the set value is for voltage, and its default unit is V. In CR mode, the set value is for resistance, and its default unit is Ω. In CP mode, the set value is for power, and its default unit is W.
a) In the main interface, press the Up/Down arrow key to switch the
parameter focus to the "Value" line.
b) Enter a value by using the numeric keys. For setting methods, refer to
the descriptions in "Paramete r Setting Me t h od ".
Note: If the parameter focus is on the "Value" line, press the Left/Right arrow key or rotate the knob to set the value for the adjacent step.
(3) Set Duration
The default unit for duration i s s . Its range is 0.00005 s to 3600 s. Press the Up/Down arrow key to switch the parameter focus to the
"Duration" line.
Enter a value by using the numeric keys. For setting methods, refer to
the descriptions in "Par a m e te r Setting Me th o d ".
Note: When the current parameter focus is on the "Duration" line, press the Left/Right arrow key or rotate the knob to switch the parameter focus to the duration for the adjacent step.
(4) Set Slew (only available in CC mode)
The default unit for the slew rate is A/μs. In the main interface, press the Up/Down arrow key to switch the
parameter focus to the "Slew" line.
Enter a value by using the numeric keys. For setting methods, refer to
the descriptions in "Par a m e te r Setting Me th o d ".
Note: If the parameter focus is on the "Slew" line, press the Left/Right arrow key or rotate the knob to set the value for each step.
(5) Set V_Sampling (V)/C_Sampling (A)
After the list operation is finished, the readback value for each step in the last cycle (that is, the last sampling value stably displayed for each step in the last cycle) will be automatically displayed in the parameter list. In CC mode, readback voltage (V_Sampling (V)) is displa yed; in CV mode, readback current (C_Sampling (A)) is displayed; and in CR and CP modes, both readba ck volta ge (V_Sam pling (V )) and rea dback cur rent (C_S amplin g (A)) are displayed.
4. Enable trigger
After you complete the parameter setting, turn on the trigger, and then the load
2-42 DL3000 User’s Guide
Page 73
Tip
Before enabling the trigger, ensure that the changes of the input will not
e trigger. During
the list operation, you are not allowed to modify the parameters.
WARNING
input terminals of the load properly before you enable the trigger.
CAUTION
message is displayed, "Fan stops running!"
will input base d on the c urrent set tings. If the trigger is tu rned off during the lis t operation, the next t ime you turn on the t ri gger, the load will continue inputting the last parameter where you stop the list operation.
The actual input voltage, current, resistance, and power of the load will be
displayed in the main interface in a real-time manner. In the parameter preview area, you can view the dynami c changes of the parameters in the parameter list.
In the parameter list, the editing box is highlighted in green. If the number of execution times ("Cycles") is set to be a specif ied value,
after completing the number of inputs (total steps), the load will terminate inputting. At this time, the end state of the load is determined by "EndState". When th e number of execution times ("Cycles") is set to "Infinite", the load makes consecutive inputs based on the current parameter settings. At this time, the end state is invalid.
After the trigger is enabled, the load input is automatically enabled (The
load first inputs the data for the first step and then inputs according to the preset values). Y o u can also press the ON/OFF key on the front panel first to turn on the input before you enable the trigger (The load fi r s t inpu ts th e data for the first step and then wait for trigge r to occ u r. After the trigger is enabled, the load inputs according to the preset values).
affect the DUT.
Please complete the pa rameter set ting bef ore enabling t h
To avoid electric shock, ensure that the DUT is connected to the
When the fan stops working, the channel is disabled. Then, a
5. Save and read files
In the List function inter face, you ca n press Save to save the parameters for List mode to the internal or external memory, and you can read and recall it if necessary by pressing Read.
(1) Save
After you complete the parameter setting, press Save to enter the storage
DL3000 User’s Guide 2-43
Page 74
RIGOL Chapter 2 Front Panel Operations
Tip
Connect the external memory U Disk to t he PC , and r ead the data files in CSV format from the U Disk.
and recall interface. The file is saved in ".LST" format. For detailed saving operation, refer to the descriptions in "Save".
(2) Read
Press Read to enter the storage and recall interface. Recall and read the files in ".LST" format or the ".RSF" files beginning with "LST_". For detailed reading operation, refer to descriptions in "
Read". The files to be recalled
and read are allowed to be edited.
How to press the Store key on t he front pa nel of the load t o pe rfo rm the sa vi ng and reading operation, refer to the descriptions in "Store and Recall".
6. Record data file
In the List function interface, you can press Record to record the data file and save it to the external memory in CSV format. During recording, "Record" is displayed at the t op of the interface. Press Record again to stop recording, and then "Record" is not displayed at the top of the interface.
7. Exit List mode
Press any key on the front panel of the load to exit List mode.
2-44 DL3000 User’s Guide
Page 75
Tip
test status is associated with the digital output of I/O interface on the rear panel.
t V Output Voltage of DUT
Load Current
I t V
I
Starting Current
Step Current
Step Delay
End Current
t V I

Application Function

The load supports the following three application functions. Press APP to switch among the three functions.
OCP test function  OPP test function  Battery test function

OCP Test Function

OCP test principle (as shown in Figure 2-28) When the OCP test s ta rts, after the input v oltage reaches Von, the load will del a y f or a period, then starts to sink the current, increasing with incrementing steps at a certain interval. Meanwhile, the load will compare the detected voltage value with the OCP protection voltage value. If higher than the protection voltage value, the load proceeds to run and increases with incrementing steps until reaching the protection voltage. If OCP occurs to the DUT, judge whether the current under test is within the maximum and minimum current range. If yes, the load passes the OCP test. Digital output of I/O interface on the rear panel is 0. If no, the load fails the test. Digital output of I/O interface on the rear panel is 1. If no OCP occurs to the DUT and the set protection time times out, the load automatically stops sinking. The OCP test fails, and the test ends.
Figure 2-28 OCP Test Function
In the system setting menu, set the digital output to be "Disable", then the OCP
DL3000 User’s Guide 2-45
Page 76
RIGOL Chapter 2 Front Panel Operations
CAUTION
the instrument or the DUT.
Operation Procedures:
1. Power on the instrument
Press the Power key
on the front panel to power on the instrument.
2. Connect the DUT and the channel input terminals of the load
Connect the DUT and t he channel input terminals on the front panel of the load as shown in Figure 2-2.
While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. A misconnection with the terminals may cause damage to
3. Set channel parameters in the guide interface
Press APP to switch to the g uide inter face of the O CP test function , as s hown in Figure 2-29. In the guide interface, you can configure parameters in the parameter configuration list and view the configuration diagr am at the left of th e conf iguration list in the real-time manner. If the scrollbar appears at the right side of the parameter list, it indicates that n ot all paramet ers are displaye d in the current interface. At this time, press the Up/Down arrow key to go to the next page.
Figure 2-29 Guide Interface of OCP Test Function
The parameters for the OCP test function include range, starting voltage, delay time, starting current, step current, step delay, protection voltage, maximum current, minimum current, and protection time.
Set Range
The current working range for the OCP test function. Two current ranges are available: DL3021/DL3021A: low range (0 to 4 A); high range (0 to 40 A) DL3031/DL3031A: low range (0 to 6 A); high range (0 to 60 A) (1) In the guide interface, press the Up/Down arrow key to switch the
2-46 DL3000 User’s Guide
Page 77
parameter focus to " Range".
(2) Use the Left/Right arrow key or the knob to switch the current working
range.
Note:
The low range provides better resolution and accuracy at low current
settings.
If the set current v alue is g reater than the ma ximum value of t he low range ,
you must select a high range.
Set Von
When the input voltage is greater than the set starting voltage (Von), the load starts to sink the cu rrent. T he def ault unit for Von is V, and it ranges from 0 V to 150 V. The default Von is 0.01 V. (1) In the guide interface, press the Up/Down arr ow key to switch the
parameter focus to " Von".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set Delay_Von
Sets the time period for how long the load starts to sink the current after the input voltage reaches Von. The default unit for the delay time is ms. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Delay_Von".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set C_Start
The current value when the load starts in OCP test mode. The default unit for the value is A. (1) In the guide interface, press the Up/Dow n arrow key to switch the
parameter focus to " C_Start".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set C_Step
The step current in OCP test mode. The default unit for the value is A. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " C_Step".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set Delay_Step
The time interval during which the current moves from step to step in OCP test mode. The default unit for the value is ms.
DL3000 User’s Guide 2-47
Page 78
RIGOL Chapter 2 Front Panel Operations
(1) In the guide interface, press t he Up/Do wn arrow key to switch the
parameter focus to " Delay_Step".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set OCP_V
The protection voltage in the OCP test mo de. T he default unit for the value is V. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " OCP_V".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set C_Max
The maximum value of th e protection current in the OC P test mode. The defa ult unit for the value is A. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " C_Max".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set C_Min
The minimum value of the protection current in the OCP test mode. The def ault unit for the value is A. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " C_Min".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set T_Limit
The maximum overcurrent protection time in OCP mode. If no action is performed during the protection time, the load will automatically stop sinking the current. The default unit for the value is
μs.
(1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " T_Limit".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
4. Enter the ma in interface
In the guide interface, after you configure the parameters, press Apply to enter the main interface, as shown in
Figure 2-30. At this time, "OCP" is active at the top of the main interface, which indicates that the parameters have been successfully configured. In the guide interface, if you give up configuring parameters, you can press Cancel to exit the guide i nterface and ent er the main interface.
2-48 DL3000 User’s Guide
Page 79
Figure 2-30 Main Interface o f O CP Test Function
The menus of the main interfac e ar e displayed in three pages.
Menu items on Page 1: Range, Von, Delay_Von, C_Start, C_Step, and Next . Menu items on Page 2: Delay_Step, OCP_V, C_Max, C_Min, Back, and Next. Menu ite ms on Page 3: T_Limit, Guide, Read, Save, and Back. To go to t he next page, p r es s Next; to exit the current page and go back to the previous page, press Back; to go back to the guide interface from the main interface, press Guide.
In the main interface, you can also set parameters base d on your own nee ds . The setting methods for the parameters are as follows:
Set Range
(1) In the main interface, press Range to switch the parameter focus to
"Range".
(2) Press Range to switch the current working range. You can also use the
Left/Right arrow key or the knob to switch it.
Set Von
(1) In the main interface, press Von to switch the parameter focus to "Von". (2) Enter a value by using the numeric keys, arrow keys, or the knob. For the
setting methods, ref er to t he descriptions of the built -in he lp information or refer to the descriptions in "Parameter Setting Method".
Set Delay_Von
(1) In the main interface, press Delay_Von to switch the parameter focus to
"Delay_Von".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set C_Start
(1) In the main interface, press C_Start to switch the parameter focus to
"C_Start".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to
DL3000 User’s Guide 2-49
Page 80
RIGOL Chapter 2 Front Panel Operations
the descriptions in "Parameter Setting Method".
Set C_Step
(1) In the main interface, press C_Step to switch the parameter focus to
"C_Step".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For the
setting methods, ref er to t he descriptions of the built -in he lp informati on o r refer to the descriptions in "Parameter Settin g Method".
Set Delay_Step
(1) In the main interface, press Delay_Step to switch the parameter focus to
"Delay_Step".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For the
setting methods, refer t o t he des criptions of the b uilt -in hel p inf ormatio n or refer to the descriptions in "Parameter Setting Method".
Set OCP_V
(1) In the main interface, press OCP_V to switch the parameter focus t o
"OCP_V".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For the
setting methods, refer t o t he des criptions of the b uilt -in hel p inf ormatio n or refer to the descriptions in "Parameter Setting Method".
Set C_Max
(1) In the main interface, press C_Max to switch the parameter focus to
"C_Max".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For the
setting methods, refer t o t he des criptions of the b uilt -in hel p inf ormatio n or refer to the descriptions in "Parameter Setting Method".
Set C_Min
(1) In the main interface, press C_Min to switch th e parameter focus to
"C_Min".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set T_Limit
(1) In the main interface, press T_Limit to switch the parameter focus to
"T_Limit".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Par a m e te r Setting Me th o d ".
2-50 DL3000 User’s Guide
Page 81
WARNING
input terminals of the load properly before you turn on the channel input.
CAUTION
message is displayed, "F an stops running!"
Tip
In the guide interface, yo u can only use the numeri c keys to input the v alue for the parameter; in the main interface, y ou can use the numeric keys, the arrow keys , or the knob to input the value for the parameter.
5. Turn on the channel input
Press On/Off to turn on the channel input. At this time, the actual input voltage, current, resistance, and power will be displayed in the main interface. If the load passes the test, a prompt message is displayed, "OCP test passed!" The DUT will automatically turn off output. If the load fails the test, one of the following prompt messages is displayed: "Upper limit exceeded. OCP test failed!" "Lower limit exceeded. OCP test failed!" Or "Timeout! OCP test failed!" Then, the channel input of the load will be automatically turned off.
To avoid electric shock, ensure that the DUT is connected to the
When the fan stops working, the channel is disabled. Then, a
6. View waveform display
After you enter the main interface, press the waveform display key enter the waveform display interface, as shown in Figure 2-31. The current will have a stepped increase. If the OCP test is successful, the current will increase until it reaches the protection current of the DUT; if the OCP test fails, the current will increase until it reaches the set maximum current. For detailed operations, refer to the de scriptions in "Waveform Display Function" section. Press the waveform display key
again to exit the waveform display
interface and return to the main interface of OCP test function.
Figure 2-31 OCP Waveform Display Interface
to
DL3000 User’s Guide 2-51
Page 82
RIGOL Chapter 2 Front Panel Operations
7. Save and read files
In the guide interface an d main interface of OCP test mo de, you can press Save to save the parameters for the OCP test mode to the internal or external memory, and you can r ead and recall it if necessary by pressing Read.
(1) Save
After you complete the parameter setting, press Save to enter the storage and recall interface. The file is saved in ".OCP" format. For detailed saving operation, refer to the descriptions in "
Save".
(2) Read
Press Read to enter the storage and recall interface. Select the files that are in ".OCP" format or the ".RSF" files beginning with "OCP_" to read and recall. For detailed reading operat ion, refer to descriptions in "
Read". The
files to be recal led and read are allowed to b e e dited.
How to press the Store key on t he front pa nel of the load t o pe rfo rm the sa vi ng and reading operation, refer to the descriptions in "Store and Recall".
8. Exit the OCP test function
Press any key on the front panel of the load to exit the OCP test mode.
2-52 DL3000 User’s Guide
Page 83
Tip
test status is associated with the digital output of I/O interface on the rear panel.
t V Output Voltage of DUT
Load Power
P t V
Starting Power
Step Power
Step Delay
End Power
t
V

OPP Test Function

OPP test principle (as shown in Figure 2-32): When the OPP test starts, after the input voltage reaches Von, the load will dela y for a period, then the DUT starts to work according to the starting power. The power increases with increasing with incrementing steps at a certain interval. Meanwhile, the load will com pare the detected v oltage v alue wi th the protection voltage v al ue of the OPP test. If higher than the protection voltage value, the load proceeds to run and increases with incrementing steps until reaching the protection voltage.
If OPP oc curs to the DUT, judge whether the power under test is within the set maximum and minimum power range. If yes, the load passes the OPP test. Digital output of I/O interface on the rear panel is 0. If no, the load fails the test. Digital output of I/O interface on the rear panel is 1.
If no OPP occurs to the DUT and the set protection time times out, the load automatically stops sinking. The OPP test fails, and the test ends.
Figure 2-32 OPP Test Function
In the system setting menu, set the digital output to be "Disable", then the OPP
Operation Procedures:
1. Power on the instrument
Press the Power key
on the front panel to power on the instrument.
2. Connect the DUT and the channel input terminals of the load
Connect the DUT and t he channel input terminals on the front panel of the load
DL3000 User’s Guide 2-53
Page 84
RIGOL Chapter 2 Front Panel Operations
CAUTION
the instrument or the DUT.
as shown in Figure 2-2.
While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. A misconnection with the terminals may cause damage to
3. Set channel parameters in the guide interface
Press APP to switch to the guide inte rface of the OP P test fu nction, as sh own i n Figure 2-33. In the guide interface, you can configure parameters in the parameter configuration list and view the configuration diagr am at the left of th e conf iguration list in the real-time manner. If the scrollbar appears at the right side of the parameter list, it indicates that not all pa rameters are displayed in th e current interface. At this time, press the Up/Down arrow key to go to the next page.
2-54 DL3000 User’s Guide
Figure 2-33 Guide Interface of OPP Test Function
The parameters for the OPP test function include starting voltage, delay time, starting power, step power, step delay, protection voltage, maximum power, minimum power, and protection time.
Set Von
When the input voltage is greater than the set starting voltage (Von), the load starts to sink the cu rrent. T he def ault unit for Von is V, and it ranges from 0 V to 150 V. The default Von is 0.01 V. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Von".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set Delay_Von
Sets the time period for how long the load starts to sink the current after the input voltage reaches V o n. The default unit for the delay time is ms. (1) In the guide interface, press the Up/Down arrow key to switch the
Page 85
parameter focus to " Delay_Von".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set P_Start
The power value when the load starts in O PP test mo de. The def ault unit for t he value is W. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " P_Start".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set P_Step
The step power in OPP test mode. The default unit for the value is W. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " P_Step".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set Delay_Step
The time interval during which the power moves from step to step in OPP test mode. The default unit for the value is ms. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " Delay_Step".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set OPP_V
The protection voltage in the OPP test mode . The default unit f o r the value is V. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " OPP_V".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set P_Max
The maximum value of the protection power in the OPP test mode. The default unit for the value is W. (1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " P_Max".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set P_Min
The minimum value of the protection power in the OPP test mode. The default unit for the value is W. (1) In the guide interface, press the Up/Down arrow key to switch the
DL3000 User’s Guide 2-55
Page 86
RIGOL Chapter 2 Front Panel Operations
parameter focus to "P_Min".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set T_Limit
The maximum overpower protection time in OPP mode. If no action is performed during the protection time, the load will automatically stop sinking the current. The default unit for the value is
μs.
(1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to " T_Limit".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Para meter Settin g Method".
4. Enter the ma in interface
In the guide interface, afte r you configure the param eters, press Apply to enter the main interface, as shown in
Figure 2-34. At this time, "OPP" is active at the top of the main interface, which indicates that the parameters have been successfully configured. In the guide interface, if you give up configuring parameters, you can press Cancel to exit the guide interface and enter the main interface.
Figure 2-34 Main Interface of OPP Test Function
The menus of the main interfac e ar e displayed in three pages.
Menu items on Pa ge 1: Von, Delay_ V on, P_Start, P_ Step, Dela y_Ste p, and Next. Menu ite ms on Page 2: OPP_ V, P_Max, P_Min, T_Limit, Back, and Next. Menu items on Page 3: Guide, Read, Save, and Back. To go to t he next page, p r es s Next; to exit the current page and go back to the previous page, press Back; to go back to the guide interface from the main interface, press Guide.
In the main interface, you can also set parameters based on your own needs. The setting methods for the parameters are as follows:
Set Von
(1) In the main interface, press Von to switch the parameter focus to "Von". (2) Enter a value by using the numeric keys, arrow keys, or the knob. For
2-56 DL3000 User’s Guide
Page 87
setting methods, refer to the des criptions of the help information or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set Delay_Von
(1) In the main interface, press Delay_Von to switch the parameter focus to
"Delay_Von".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set P_Start
(1) In the main interface, press P_Start to switch the parameter focus to
"P_Start".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Paramete r Setting Me th o d ".
Set P_Step
(1) In the main interface, press P_Step to swi tc h the parameter focus to
"P_Step".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help inf ormation or refe r to the descriptions in "Par a m e te r Setting Me th o d ".
Set Delay_Step
(1) In the main interface, press Delay_Step to switch the parameter focus to
"Delay_Step".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set OPP_V
(1) In the main interface, press OPP_V to switch the parameter focus to
"OPP_V".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set P_Max
(1) In the main interface, press P_Max to switch th e parameter focus to
"P_Max".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help information or refe r to the descriptions in "Par a m e te r Setting Me th o d ".
DL3000 User’s Guide 2-57
Page 88
RIGOL Chapter 2 Front Panel Operations
WARNING
input.
CAUTION
message is displayed, "Fan stops running!"
Tip
In the guide interface, yo u can only use the numeri c keys to input the v alue for the parameter; in the main interface, y ou can use the numeric keys, the arrow keys , or the knob to input the value for the parameter.
Set P_Min
(1) In the main interface, press P_Min to switch th e parameter focus to
"P_Min".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Parameter Set tin g Method".
Set T_Limit
(1) In the main interface, press T_Limit to switch the parameter focu s to
"T_Limit".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Parameter Setting Method".
5. Turn on the channel input
Press On/Off to turn on the channel input. At this time, the actual input voltage, current, resistance, and power will be displayed in the main interface. If the load passes the test, a prompt message is displayed, "OPP test passed!" The DUT will automatically turn off outp ut. If the load fails the test, one of the following prompt messages is displayed: "Upper limit exceeded. OPP test failed!" "Lower limit exceeded. OPP test failed!" Or "Timeout! OPP test failed!"
To avoid electric shock , ensure that the DUT is connected to the input terminals of the load properly before you turn on the channel
When the fan stops working, the channel is disabled. Then, a
6. View waveform display
After you enter the main interface, press the waveform display key enter the waveform display interface, as shown in Figure 2-35. The power will have a stepped increase. If the OPP test is successful, the power will increase until it reaches the p r otection power of the DUT; if the OP P te st f ai ls, t he power will increase until it reaches the set maximum power. For detailed operations, refer to the descriptions in "Waveform Display Function" section. Press the waveform display key
2-58 DL3000 User’s Guide
again to exit the waveform display interface and
to
Page 89
return to the main interface of OPP test function.
Figure 2-35 OPP Wavef or m Displa y Inte rface
7. Save and read files
In the guide inte rface a nd main interfa ce of OPP test mode, you can p ress Save to save the parameters for the OPP test mode to the internal or external memory, and you can read and recall it if necessary by pressing Read.
(1) Save
After you complete the parameter setting, press Save to enter the storage and recall interface. The file is saved in ".OPP" format. For detailed saving operation, refer to the descriptions in "
Save".
(2) Read
Press Read to enter the storage and recall interface. Select the files that are in ".OPP" format or the ".RSF" files beginning with "OPP_" to read and recall. For detailed reading operat ion, refer to descriptions in "
Read". The
files to be recal led and read are allowed to b e edited.
How to press the Store key on t he front pa nel of the load t o pe rfo rm the sa vi ng and reading operation, refer to the descriptions in "Store and Recall".
8. Exit the OPP test function
Press any key on the front panel of the load to exit the OPP test function.
DL3000 User’s Guide 2-59
Page 90
RIGOL Chapter 2 Front Panel Operations
CAUTION
the instrument or the DUT.
V
Battery Voltage
Load Current
Discharge Time
Cut-off V ol t a ge
t

Battery Test Function

Battery test principle (as shown in Figure 2-36): A test for dischar ging the batte ry in C C mode. You can se t the termination condition s: cut-off voltage value, battery capacity value, and discha rge time . When any one of the three conditions is met, the discharge stops. Then the load automatically stops sinking. When only one or two condition(s) is/are selected as the termination condition(s) for the battery test, set the rest termination condi tions not in use to be "OFF".
Figure 2-36 Battery Test Function
The battery test can reflect the reliability of the batte ry and the remaining battery lif e. Therefore,
Operation Procedures:
1. Power on the instrument
Press the Power key
2. Connect the b a t te ry to the channel input ter minals of the lo ad
Connect the battery to the channel input terminals on the front panel of the loa d as shown in Figure 2-2.
before replacing the battery, do the battery test first .
on the front panel to power on the instrument.
While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. A misconnection with the terminals may cause damage to
3. Set channel parameters
Press APP to switch to the main interface of Battery test function, as shown in Figure 2-37. "Battery" is active at the top of the main interface.
2-60 DL3000 User’s Guide
Page 91
CAUTION
avoid causing damage to the instrument or the battery.
Figure 2-37 Main Interface of Battery Test Function
The parameters for the battery test function include current, range, cut-off voltage, battery capacity, discha r ge time, and starting voltage.
Set Current
The discharge current. Its default unit is A. (1) In the main interface, press Current to switch the parameter focus to
"Current".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set Range
Working range for the current in battery test mode. Two ranges are available. DL3021/DL3021A: low range (0 to 4 A); high range (0 to 40 A) DL3031/DL3031A: low range (0 to 6 A); high range (0 to 60 A) (1) In the main interface, press Range to switch the parameter focus to
"Range".
(2) Press Range to switch the current working range. You can also use the
Left/Right arrow key or the knob to switch it.
Note:
The low range provides better resolution and accuracy at low current
settings.
If the set current value is greater than the maximum v alue of the low range,
you must select a high range.
Before switching the range, please disable the channel input to
Set V_Stop
When the cut-off voltage is set to the termination condition, after the battery voltage reaches the cut-off voltage value, the load stops sinking automatically. The battery stops discharging. The default unit for the cut-off voltage is V. If yo u set the cut-off voltage to be the termination condition:
DL3000 User’s Guide 2-61
Page 92
RIGOL Chapter 2 Front Panel Operations
Tip
conf irms that the battery reaches the set value.
(1) I n the main interface, press V_Stop to switch the param eter focu s to
"V_Stop".
(2) Press V_Stop to switch to "ON", and select the cut-off volt age to b e
the termination condition.
(3) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help information or refer to the descriptions in "Parameter Setting Method".
If you do not nee d to set the cut-off voltage to be the termination condition:
(1) In the main interface, press V_Stop to switch the parameter focus to
"V_Stop".
(2) Press V_Stop to switch to "OFF", and the input box is disabled, and
you are unable to input any value.
If the cut-off voltage i s set to be the termination condition, when the battery voltage is too low, the system will automatically discontinue test once it
Set C_Stop
When the battery capacity is set to the termination condition, after the battery capacity reaches the se t value, the load stops sinki ng automa tically. The battery stops discharging. The default unit is mAh. If you set the battery capacity to be the termination condition:
(1) In the main interface, press C_Stop to switch the parameter focus to
"C_Stop".
(2) Press C_Stop to switch to "ON", and select the battery capacity to be
the termination condition.
(3) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help information o r refer to the descriptions in "Parameter Setting Method".
If you do not need to set the battery capacity to be the termination
condition: (1) In the main interface, press C_Stop to switch the parameter focus to
"C_Stop".
(2) Press C_Stop to switch to "OFF", and the input box is disabled, and
you are unable to input any value.
Set T_Stop
When the discharge time is set to the te rmination condition, after the discharge time reaches the set value, the load stops sinking automatically. The battery stops discharging. The default unit for the discharge time is s. If yo u set the discharge time to be the termination condition:
(1) In the main interface, press T_Stop to switch the parameter focus to
2-62 DL3000 User’s Guide
Page 93
WARNING
input.
"T_Stop".
(2) Press T_Stop to switch to "ON", and select the discharge time to be
the termination condition.
(3) Enter a value by using the numeric keys, arrow keys, or the knob. For
the setting methods, refer to the descriptions of the built-in help information or refer to the descriptions in "Parameter Setting Method".
If you do not need to set the discharge time to be the termination
condition:
(1) In the main interface, press T_Stop to switch the parameter focus to
"T_Stop".
(2) Press T_Stop to switch to "OFF", and the input box is disabled, and
you are unable to input any value.
Set Von
When the input voltage is greater than the set starting voltage (Von), the load starts to sink the cu rrent. T he def ault unit for Von is V, and it ranges from 0 V to 150 V. The default Von is 0.5 V. To enable or disable this function, refer to descriptions in "Enable/Disable the Von Latch Function". (1) In the main interface, press Von to switch the parameter focus to "Von". (2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Parameter Set tin g Method".
4. Turn on the channel input
Press On/Off to turn on the channel input. At this time, the actual current, discharge voltage, current discharge time, discharged capacitance, and battery energy will be displayed in the main interface. If the termination condition for the battery discharge is met, after complet i n g the test, a prompt message is displayed, "Battery test completed!" Press OK to exit.
Note: Once the channel input is turned on, the load will not start to sink the current until the input voltage is greater than the starting voltage.
To avoid electric shock, ensure that the battery is connected to the input terminals of the load properly before you turn on the channel
5. View waveform display
Press the waveform display key
to enter the wa vef orm displa y interface, as shown in Figure 2-38. By default, it displa ys the current wav eform . When you select "I" under "Data", the battery is discharged in CC mode. For detailed operation, refer to descriptions in "Waveform Display Function". Press the
DL3000 User’s Guide 2-63
Page 94
RIGOL Chapter 2 Front Panel Operations
waveform display key again to exit the waveform display interface and return to the main interface of Battery test function.
Figure 2-38 Battery W aveform Dis pla y Interface
6. Exit the batt e r y test function
Press any key on the front panel of the load to exit the battery test function.
2-64 DL3000 User’s Guide
Page 95
CAUTION
the instrument or the DUT.

Advanced Function

The advanced functions include factory test function and CC+CV function. Press Option to enter the advanced function interface, as shown in Figure 2-39
Figure 2-39 Option Interface

Factory Test Function

In CC/CV mode, the load supports the factory test function, which is use d to te st the stability of the current and voltage of the DUT.
Operation Procedures:
1. Power on the instrument
Press the Power key
2. Connect the DUT and the channel input terminals of the load
Connect the DUT and t he channel input terminals on the front panel of the load as shown in Figure 2-2.
on the front panel to power on the instrument.
.
3. Set channel parameters
Press Option to enter the advanced function interface. Press OK or OK to ente r the main interface of factory test function, as shown in Figure 2-40. "Factory" is displayed at the top of the main interface.
DL3000 User’s Guide 2-65
While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. A misconnection with the terminals may cause damage to
Page 96
RIGOL Chapter 2 Front Panel Operations
Figure 2-40 Main Interface of Factory Test Function
The menus of the main interface are displayed in two pages. Menu items on Page 1: Mode, Current, Range, Max, Min, and Next. Menu items on Page 2: Pass_Tip, Fail_Tip, Pass_Sound, Fail_Sound, and Back. To go to the next page, press Next; to exit the current page and go back to the previous page, press Back.
Set Mode
Factory test function supports CC/CV mode. (1) In the main interf ace, press Mode to switch the parameter focus to "Mode". (2) Press Mode to switch the factory test mode. You can also use the
Left/Right arrow key or the knob to switch it.
Set Current/Voltage
Sets the working voltage or current in the current mode based on the currently selected working mode for the factory test function. The default unit for the current is A, and that for the voltage is V. If you select "CC" mode:
(1) In the main interfa ce, press Current to switch the parameter focus to
"Current".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help information or refer to the descriptions in "Parame te r Se tting Metho d ".
If you select "CV" mode:
(1) In the main interface, press Voltage to switch the parameter focus to
"Voltage".
(2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help information or refer to the descriptions in "Parame te r Se tting Metho d ".
Set Range
Sets the range in the current mode based on the currently selected working mode of the factory test function. If you select "CC" mode:
2-66 DL3000 User’s Guide
Page 97
CAUTION
input to avoid causing damage to the instrument or the DUT.
(1) In the main interface, press Range to switch the parameter focus to
"Range".
(2) Press Range to switch the current working range . Y ou can also use t he
Left/Right arrow key or the knob to switch it.
If you select "CV" mode:
(1) In the main interface, press Range to switch the parameter focus to
"Range".
(2) Press Range to switch the voltage working range. Y ou can also use the
Left/Rig h t arrow key or the knob to switch it.
Note:
The low range provides better resolution and accuracy at low current or
voltage settings.
If the set current or voltag e value is great er than the maximum v alue of t he
low range, yo u must select a high range.
Before switching the current range, please disable the channel
Set Max
Sets the upper limit of the working v oltage o r current in t he current mo de based on the currently selected working mode for the factory test function. (1) In the main interface, press Max to switch the parameter focus to "Max". (2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Par a m e te r Setting Me th o d ".
Set Min
Sets the lower limit of the working v oltage or current in the cur rent mode based on the currently selected working mode for the factory test function. (1) In the main interface, press Min to switch the parameter focus to "Min". (2) Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help informat ion or refer to the descriptions in "Parameter Set tin g Method".
Set Pass_Tip
If the test is passed, you can determine whether to display the prompt message that reports the test pass. (1) In the main interface, press Pass_Tip to switch the parameter focus to
"Pass_Tip".
(2) Press Pass_Tip to enable or disable the display of the prompt message.
You can also use the Left/Right arrow key or the knob to switch the on/off state.
DL3000 User’s Guide 2-67
Page 98
RIGOL Chapter 2 Front Panel Operations
WARNING
input.
CAUTION
message is displayed, "Fan stops running!"
Set Fail_Tip
If the test is failed, you can determine whether to display the prompt message that indicates the test failure. (1) In the main interface, press Fail_Tip to switch the parameter focus to
"Fail_Tip".
(2) Press Fail_Tip to ena ble or disabl e the displa y of the pro mpt message . Y ou
can also use the Left/Right arrow key or the knob to switch the on/off stat e.
Set Pass_Sound
If the test is passed, you can determine whether to sound the alarm for test pass. (1) In the main interface, press Pass_Sound to switch the parameter focus to
"Pass_Sound".
(2) Press Pass_Sound to enable or disable the alarm sound once the test is
passed. Y ou can also use the Left/Right a rrow ke y or t he kno b to sw itch th e on/off state.
Set Fail_Sound
If the test is failed, you can determine whether to sound the alarm for test failure. (1) In the main interface, press Fail_Sound to switch the parameter focus to
"Fail_Sound".
(2) Press Fail_Sound to enable or disable the alarm sound once the test is
failed. You can also use the Left/Right arrow key or the knob to switch the on/off state.
4. Turn on the channel input
Press On/Off to turn on the channel input for the factory test function. At this time, the actual input v oltage, current, resistance, and power will be displa yed in the main interface.
To avoid electric shock, ensure that the DUT is connected to the input terminals of the load properly before you turn on the channel
When the fan stops working, the channel is disabled. Then, a
5. Exit the factory test function
Press any key on the front panel of the load to exit the factory test function.
2-68 DL3000 User’s Guide
Page 99
CAUTION
the instrument or the DUT.

CC+CV Function

The load supports CC+CV function that simulates two sta ges of chargin g: The load i s set to be in CC mode and simulates constant current (CC) charging. When the readback voltage is detected and the detected value reaches the preset voltage, the load switches to the CV mode to simulate constant voltage (CV) charging. This function is used to test the charger and similar devices.
Operation Procedures:
1. Power on the instrument
Press the Power key
2. Connect the DUT and the channel input terminals of the load
Connect the DUT and t he channel input terminals on the front panel of the load as shown in Figure 2-2.
While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. A misconnection with the terminals may cause damage to
on the front panel to power on the instrument.
3. Set channel parameters
Press Option to enter the advanced function interface, and then use the Up/Down arrow keys or the knob to switch to "CC+CV". Press OK or OK to enter the main interface of CC+CV function, as shown in Figure 2-41. "CC+CV"
is displayed at the top of the main interface.
Figure 2-41 Main Interface of CC+CV Mode
In CC+CV mode, you need to set the following parameters: Current, C_Range, Voltage, V_Range, V_Limit, and C_Limit.
DL3000 User’s Guide 2-69
Page 100
RIGOL Chapter 2 Front Panel Operations
CAUTION
input to avoid causing damage to the instrument or the DUT.
Tip
In CC+CV mode, when the set load current is greate r than the cur rent output from the DUT, short circuit occurs to the DUT.
Tip
output from the DUT, broken circuit occurs to the DUT.
Set Current
The constant current in CC+CV mode. The default unit for the current is A. (1) In the main interface, press Current t o switch the parameter focus to "Current". (2) Enter a value by using the numeric keys, arrow keys, or the knob. For setting
methods, refer to the descriptions of the help information or refer to the descriptions in "Parameter Setting Method".
Set C_Range
The working ranges of cur rent in C C+CV mo de a re as foll ows. Diff erent mo dels ha v e different ranges. DL3021/DL3021A: low range (0 to 4 A); high range (0 to 40 A) DL3031/DL3031A: low range (0 to 6 A); high range (0 to 60 A) (1) In the main interface, press C_Range to switch the parameter focus to
"C_Range".
(2) Press C_Range to switch the current working range. You can also use the
Left/Right arrow key or the knob to switch it.
Note:
The low range provides better resolution and accuracy at low current settings. If the set current val ue is greate r than the ma ximum value o f the low range, you
must select a high range.
Before switching the current range, please disable the channel
Set Voltage
The constant voltage in CC+CV mode. The default unit for the voltage is V. (1) In the main interface, pres s Voltage to switch the par ameter focus to "Voltage". (2) Enter a value by using the numeric keys, arrow keys, or the knob. For setting
methods, refer to the descriptions of the help information or refer to the descriptions in "Parameter Setting Method".
In CC+CV mode, wh e n th e set load voltage is greater than the voltage
Set V_Range
Working range for the voltage in CC+CV mode. Two ranges are available: low range
2-70 DL3000 User’s Guide