Read this manual thoroughly and do not perform installation, operation,
maintenance, or inspection unless you fully understand all of the contents.
Keep this manual in a safe place where you can refer to it easily while
installing, operating, and carrying out maintenance or inspections.
This manual is for use by service personnel of our company or qualified to
perform maintenance services for this machine. Use by anyone except the
above personnel is not permitted.
Numerics: 6 digits, Height 25mm
Mark: Battery. Zero point, Now subtracting
tare, Stable, Weight unit (kg・lb), Over,
Under, Accept
Preset tare
40min./ 50min./ 60min.
IGB Series IGX Series
The required transformer is provided for
each 100VAC, 120VAC, 220VAC, or
240VAC.
100VAC: 57mA
120VAC: 46mA
220VAC: 28mA
240VAC: 23mA
VFD
Numerics: 5 + 1/2 digits, Height 29mm
Mark: Zero point, Now subtracting tare
LED
Over, Under, Accept
No
Manual No. 085-3435-07 4/45 IGB/IGX Series (Overseas Specifications) Service Manual
Page 6
1.3 Appearance
IGB Series (Asia/Oceania Specification)
IGX Series (Asia/Oceania Specification)
Manual No. 085-3435-07 5/45 IGB/IGX Series (Overseas Specifications) Service Manual
Page 7
1.4 Display Unit
IGB Series (Asia/Oceania Specification)
IGX Series (Asia/Oceania Specification)
Manual No. 085-3435-07 6/45 IGB/IGX Series (Overseas Specifications) Service Manual
Page 8
1.5 Outer Dimensions
(Unit: mm)
Manual No. 085-3435-07 7/45 IGB/IGX Series (Overseas Specifications) Service Manual
Page 9
memo
Manual No. 085-3435-07 8/45 IGB/IGX Series (Overseas Specifications) Service Manual
Page 10
Chapter 2 Test Mode
The Test Mode is used to check the machine when maintenance is performed or to perform settings.
2.1 Operation
The keys used in the following diagram are for Asia/Oceania specifications. For specifications for other
countries, use the corresponding keys.
2.1.1 Starting Each Mode
Scale ON
Release
ON/OFF key,
Then press one
of these keys
while all
segments are
illuminated
Start
Check
Operation Manual
Normal Mode
Voltage Check
(IGB only)
Setup Mode
Program Mode
Test Mode
2.1.2 Ending Test Mode
C1 Mode
C2 Mode
C3 to C9
Mode
In Mode
In Mode
Press for 2 seconds or more
You cannot return to Cx mode.
C3 to C9
Mode
Scale OFF
Manual No. 085-3435-07 9/45 IGB/IGX Series (Overseas Specifications) Service Manual
Page 11
2.1.3 Test Mode Flow
2.1.4 Key Functions
C1
C2
C3 to C6
C8, C9
C7
To each C1
mode setup
Key check
mode
Each mode
operation
Scale function
Setup mode
Each mode
operation
Key Function
Turns the power ON or OFF for the IGB Series.
Turns the power ON or OFF for the IGX Series.
+5V is always supplied to the circuit voltage.
Increases the numeric value (when the numeric value is set).
Advances the mode (when conditions are set).
Adjusts the zero point.
Decreases the numeric value (when the numeric value is set).
Reverses the mode (when conditions are set).
Adjusts the span point.
Determines the data in the Details Mode.
Changes between “+ Adjust” (ZERO ³ key [u display]) and “- Adjust” (TARE key
[d display]) when the span is adjusted.
Returns to each mode.
(Tact switch on the main board)
Saves the E2ROM data of each item setting of C1 mode.
Memory
switch
Manual No. 085-3435-07 10/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 12
2.1.5 Mode List
Mode Contents
*C1 Country No., Scale No., Zero point, Span adjustment
C2 Key check
C3 Display check 1 (simplified check), All LEDs light up
C4
C5 Program No. display
*C6
*C7 Settings for weighing conditions
C8 Reading E2ROM data
C9 Board check (A/D check, Interface check) for factory inspection
Display check 2 (detailed check), Each segment lights in sequence for each display
digit.
RAM clear (Program mode data)
E2ROM clear (Program mode data and Test mode C1 data)
(Span adjustment is required)
Manual No. 085-3435-07 11/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 13
2.2 C1 Mode (Country No., Scale No., Zero point, Span adjustment)
e
2.2.1 Country No. Table
Country Name USA 1 EU AUS ASIA USA 2JPN
Country Code 0 12 3 4 56 7 8 9 1099
ADRS POS WGT Function Default Data (Change not allowed)
112 A H Start width 2 2 2 2 2 22 2 2 2 4 -
B H Stable/Re-stable count 3 33 3 3 33 3 3 3 3 -
113 A H Stable/Re-stable width 4 4 4 4 4 44 4 4 4 4 -
B H
114 A 1 Zero point mark
2 Over-scale display
4 Display less than true zero
8 Decimal point display
B 1 Over-scale range
2 Tare subtraction
4 Tare clear with ZERO key
8 Zero suppress
115 A 1 Key-in tare subtraction
Re-stable operation start width
4 44 4 4 44 4 4 4 4 -
2 80 8 2 80 8 4 0 5 -
0 66 0 0 66 0 0 6 0 -
IDV
SET
2 Zero tracking
4 Micro weight follow-up
8 Unstable width
B 1 PLU display (IGB only)
2 Stable display (IWB only)
4 Reserved
8 Reserved
1. By setting the country code, weighing conditions will function based on the default data from
Note
address 112 to 115.
2. The data is set to meet certified conditions for each country, so there is no need to change the
data.
3. Position “A” indicates the upper position of one-byte data, and “B” indicates the lower position.
4. As for Weight, a function is selected with “1”, “2”, “4”, and “8” in bit unit, and “H” indicates
Hexadecimal data.
5. Individual setting value “99” cannot be entered. (Displayed only when settings are changed in C7
mode)
6. “USA 2” code is settings to allow changing between “lb” and “kg” at anytime even in cases other
than a stable condition at zero point. (“USA 1” can change between “lb” and “kg” only when the
condition is stabilized at zero point)
When using this function, a maximum of 0.2e error may occur in weighing immediately after the
change. This error is calibrated by zero point adjustment (or zero tracking), and an accurate
weighing is guaranteed.
When selecting the weighing capacity exclusively for either “lb” or “kg”, use “USA 1”.
0 00 0 0 00 0 0 0 4
0 66 6 4 66 6 5 5 5
-
-
Refer to “Weighing Condition Data Table” for detailed function selection.
Refe re nc
Reference
Manual No. 085-3435-07 12/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 14
2.2.2 Scale No. Table
e
Scale No. Specifications
0 150kg (0.05kg/0.02kg) Multi Interval 6 C 3 8
1 60kg (0.02kg/0. 01kg) Multi Interval 2 B 3 8
2 30kg (0.01kg/0.005kg) Multi Interval B B 3 8
3 15kg (0.005kg/0.002kg) Multi Interval 7 C 3 8
4 6kg (0.002kg/0.001kg) Multi Interval 3 B 3 8
5 6000g (2g/1g) Multi Interval 0 B
6 3000g (1g/0.5g ) Multi Interval 9 B
7 120kg (0.02kg) 1/6000 Single Range 6 B 2 8
8 60kg (0.01kg) 1/6000 Single Range 2 B 2 8
9 30kg (0.005k g) 1/6000 Single Range B B 2 8
10 15kg (0.002kg) 1/7500 Single Range 7 B 2 8
11 6kg (0.001kg) 1/6000 Single Range 3 B 2 8
12 300kg (0.1kg) 1/3000 Single Range 1 B 2 8
13 150kg (0.05kg) 1/3000 Single Range A 8 2 8
14 60kg (0.02kg) 1/3000 Single Range 6 8 2 8
15 30kg (0.01kg) 1/3000 Single Range 2 8 2 8
16 15kg (0.005kg) 1/3000 Single Range B 8 2 8
17 6kg (0.002kg) 1/3000 Single Range 7 8 2 8
18 6000g (2g) 1/3000 Single Range 4 8 A 8
19 3000g (1g) 1/3000 Single Range 0 8 A 8
20 150kg/60kg (0.05kg/0.02kg) Dual Range 6 C 5 8
21 60kg/30kg (0.02kg/0.01kg) Dual Range 2 B 5 8
22 30kg/15kg (0.01kg/0.005kg) Dual R ange B B 5 8
23 15kg/6kg (0.005kg/0.002kg) Dual Range 7 C 5 8
24 6kg/3kg (0.002kg/0.001kg) Dual Range 3 B 5 8
25 6000g/3000 (2g/1g) Dual Range 0 B D 8
26 3000g/1500 (1g/0.5g) Dual Range 9 B D 8
27 150kg/150k (0.1kg/0.05k g Dual Range
Fishery specification
28 150kg (0.05kg) 1/3000 Single Range
Body weight specification
29 30kg (0.01kg) 1/3000 Single Range
Baby scale specification
30 300lb/150kg (1/3000) Multi Interval 6 C 7 8
31 150lb/60kg (1/ 3000) Multi Interval 2 B 7 8
32 60lb/30kg (1/3000) Multi Interval B B 7 8
33 30lb/15kg (1/3000) Multi Interval 7 C 7 8
34 15lb/6kg (1/3000) Multi Interval 3 B 7 8
99 Individual scale settings - - - -
1. By setting the Scale No., the scale will function acc ording to default data of addresses 110 and
111.
2 The above data determines each specificati on, so changing the data is prohibited.
3 Data A indicates the upper position of the on e-byte data, and B indicates the lower position.
Refer to “Weighing Condition Data Table” for data details.
Referenc
Note
Reference
Address
110A 110B 111A 111B
Default Data
B
B
8
8
A 8 5 8
A 8 4 2
2 8 4 2
Manual No. 085-3435-07 13/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 15
2.2.3 Operation Procedure
Operation Display
1. Starting Test Mode (C1)
Press ON/OFF.
Press and hold ZERO(+) while all segments are lit.
Press SET.
2. Setting Country Number
To increase the number, press ZERO(+).
To decrease the number, press TARE(-).
Press SET.
[Example] Asia = 3
3. Setting Scale Number
When setting “60kg”, select “08”.
Press SET.
4. Selecting Weight at Span Adjustment
The following weight can be selected:
kg: 1/1.0→1/2.0→1/2.5→1/5.0 of weighing capacity
lb: 1/1.0→1/2.0→1/2.5→1/5.0 of weighing capacity
・ Pressing ZERO(+) selects the next item in the above
sequence, and pressing TARE(-) selects the previous item.
kg unit: Same weight is used as weighing
capacity.
[Example] Selecting “kg” unit and “1/2.0” for weighing
capacity “150kg” will require the total 75kg
weight.
To change between “lb” and “kg”, press ZERO(+) and
TARE(-) at the same time.
5. Zero Point Adjustment
After selecting the weight amount, ensure that nothing is
placed on the weighing platter, then press SET.
The previously set zero point A/D data will be displayed.
Zero point indicator W will light up.
Press ZERO(+) to set “20000” counts forcibly.
Zero point indicator W will go off.
kg unit: Half weight is used of weighing
capacity.
lb unit: Same weight is used as weighing
capacity.
Manual No. 085-3435-07 14/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 16
Operation Display
6. Span Adjustment
Place the selected weight on the weighing platter, then press
TARE(-).
[Example] 60kg Scale: Weight selection: “kg” unit and “1/1”
Place a 60kg weight, 30000 + Zero point 20000 =
50000 counts
In case of “kg” unit “1/2” for a weight selection, place
a 30kg weight, then 15000 + Zero point 20000 =
35000 counts
・ One graduation equals 5 counts.
Processing is automatically executed in the following order:
[Calib] (Calibration) Computing process executing
[C- OK] (Calibration OK) Computing result succeeded
[No. of adjusted counts] Computing result
[C-Err] will be displayed, indicating a computing result error
when span adjustment is performed without placing a weight.
Place a weight and press TARE(-).
7. Saving Data
To save Country No., Scale No., Zero point, and Span data in
E2ROM.
・ Remove the weight from the weighing platter.
・ Remove the seal covering the opening on the rear case of
display unit.
・ Insert a thin rod such as the inner shaft of a ball-point pen
into the hole, and push the Memory Switch located on the
main board.
When writing has finished normally, [S-OK] will be
displayed.
・ To release the above status, press COLUMN.
Then, the display will return to A/D data.
¾
¾
(Calibration error)
8. Finishing Procedure
OFF status
・ Press ON/OFF to finish this procedure.
Manual No. 085-3435-07 15/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 17
When Performing Fine Span Adjustment
1. Trimming
After span adjustment, press COLUMN while the A/D data
is displayed.
[Tri-] (Trimming) display will appear.
(This is possible even with the weight placed)
2. Trimming Up
Press ZERO(+), then [Tri -U] will be displayed.
Press ZERO(+) for the desired number of times to increase
the span.
3. Trimming Down
Press TARE(-), then [Tri-d] will be displayed.
Press TARE(-) for the desired number of times to decrease
the span.
Note
The change amount for one time may differ
depending on each machine; however,
approximately one count for one time is average.
4. Repeating Steps
Press COLUMN, then A/D data will be displayed.
Place a weight on the weighing platter, and perform
adjustment by repeating steps 1, 2, and 3 above.
5. Saving Data
After completing the adjustment, press COLUMN to save
the data.
Manual No. 085-3435-07 16/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 18
Memory Switch
The memory switch is located on the main board.
IGB Series
Memory switch
Memory switch
IGX Series
Pushing memory switch
Manual No. 085-3435-07 17/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 19
2.3 C2 Mode: Key Check
Operation Display
1. Starting Test Mode (C1)
Press ON/OFF.
Press and hold ZERO(+) while all segments are lit.
2. Entering C2 Mode
Press ZERO(+).
3. Checking Keys
Press SET.
[KEY-0] indicates that there is no key entry.
4. Checking Keys
ZERO(+)p 1 TARE(-)p 2 ∗ p 3 SETp 4 EXT. INPUTp 5 MEMORY SW p 6
5. Finishing Procedure
Press ON/OFF to finish this procedure.
2.4 C3 Mode: Display Check 1 (Simplified Check)
Operation Display
1. Starting Test Mode (C1)
Press ON/OFF.
Press and hold ZERO(+) while all segments are lit.
2. Entering C3 Mode
Press ZERO(+) twice.
3. Checking Displays
Press SET.
The display shows self-diagnostic check.
OFF status
¼
¼
4. Returning to Test Mode
Press COLUMN to return to Test Mode.
Manual No. 085-3435-07 18/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 20
2.5 C4 Mode: Display Check 2 (Detailed Check)
Operation Display
1. Starting Test Mode (C1)
Press ON/OFF.
Press and hold ZERO(+) while all segments are lit.
2. Entering C4 Mode
Press ZERO(+) three times.
3. Checking Displays
Press SET.
The display shows self-diagnostic check.
4. Returning to Test Mode
Press COLUMN to return to Test Mode.
¼
2.6 C5 Mode: Program No Display
Operation Display
1. Starting Test Mode
Press ON/OFF.
Press and hold ZERO(+) while all segments are lit.
2. Entering C5 Mode
Press ZERO(+) four times.
3. Displaying Program Number
Press SET.
The installed program number (D005) will be displayed.
4. Returning to Test Mode
Press COLUMN to return to Test Mode.
Manual No. 085-3435-07 19/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 21
2.7 C6 Mode: RAM Clear and E2ROM Clear
Operation Display
1. Starting Test Mode
Press ON/OFF.
Press and hold ZERO(+) while all segments are lit.
2. Entering C6 Mode
Press ZERO(+) five times.
3. Press SET.
4. RAM Clear
(Preset data of Program Mode is cleared)
Press ZERO(+).
Press ZERO(+).
5. PressCOLUMN to return to Test Mode.
E2ROM Clear (Initializing E2ROM data)
6. Press SET.
7. Push the Memory Switch (SW1) located on the main
board.
(This does not work when there has been no data
change)
C1 data and setting data (F01toF17) have been
Note
8. Press ON/OFF.
After initialization, nothing except “C1” will be displayed.
C1 data settings (Country No., Scale No., Zero point,
Span adjustment) are required.
initialized.
Settings are required.
Manual No. 085-3435-07 20/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 22
2.8 C7 Mode: Weighing Condition Setting
e
Setting Country No. and Scale No. determine the weighing conditions for securing certified
Note
1. Starting Test Mode
Press ON/OFF.
Press and hold ZERO(+) while all segments are lit.
2. Entering C7 Mode
Press ZERO(+) six times.
Or, press TARE(-) three times.
3. Press SET.
To change data:
Select the desired data by pressing ZERO(+) to increase,
or TARE(-) to decrease the address.
To fix the data or advance the address:
Press SET.
specifications of the country.
Individual contents can be changed in this mode.
Specifications must be selected when used overseas.
Operation Display
3-digit address + A (Upper data) - Data
4. After changing data
Push the Memory Switch (SW1) located on the main
board.
(This does not work when there was no data change)
5. PressCOLUMN twice to return to Test Mode.
Addresses for weighing condition settings are from 110 to 115.
Refer to the next page.
Refe re nc
Reference
3-digit address + B (Lower data) - Data
Manual No. 085-3435-07 21/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 23
Weighing Condition Setting Table
Address Weight Item Data
110A 1 Decimal point position 0: 0 1: 0.0
2 2: 0.00 3: 0.000
4 Minimum graduation 0: 1 (1-2 change in Dual range/ Multi Interval)
4: 2 (2-5-10 change in Dual range/ Multi Interval)
8 8: 5 (5-10-20 change in Dual range/ Multi Interval)
Changing method 0: Single range LED stable display only (IGX)
1: 1/3000 Multi interval Stabe/Upper/Lower display (IGB)
2: Single range Upper/Lower display (IGX)
2
4
Weight unit
8
(Valid only for
communications)
1
2
Filter setting
Do not change the data.
4
8
3: 1/3000 Multi interval Stabe/Upper/Lower display (IGB)
4: Single range (Body weight specification)
5: 1/3000 Dual range (A/B range change)
6: Single range (lb/kg change)
7: 1/3000 Multi interval (lb/kg change)
<111A> <111B>
0 + 0: kg (kg - lb)
8 + 0: g
0 + 1: lb
8 + 1: oz
Note: Part numbers may change without notice due to product improvements.
Manual No. 085-3435-07 29/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 31
3.1.1 Load Cell
Load cell
Top of base
Bottom of base
Manual No. 085-3435-07 30/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 32
3.2 Electric Parts (IGB Series)
3.2.1 IGB Block Diagram
Display Unit
Main Board (PS-018)
XH 2p
Ferrite Core
Front Case
Rear Case
Power Board
(PS-019)
AC Adaptor
JACK
AC Adaptor
Scale Unit
XH 2p
MK 2p
Battery
Battery
Load Cell
Manual No. 085-3435-07 31/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 33
3.2.2 Main Board (PS-018)
Parts Surface of Main Board (The LCD and tact keys are installed on the soldering surface)
(1) SW1: Memory Switch
Used to write the data when performing initialization in Test Mode.
Used to save (write) the data in E2ROM after changing Country No. and Scale No., and performing
Span Adjustment.
(2) Connector XJ1
This connector is not used.
Pin No. Signal Name Pin No.Signal Name
1 IN1 6 NC
2 OUT4 7 RESET
3 OUT3 8 VPP
4 OUT2 9 GND
5 OUT1 10 VCC
(3) Connector XJ2
This connector is used for DC input.
Pin No. Signal Name
1 +DC
2 -DC
Manual No. 085-3435-07 32/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 34
(4) Connector XJ3
This connector is not used.
Pin No. Signal Name
1 Buzzer
2 DTR
3 RxD
4 TxD
5 GND
6 FG
(5) Program Memory Media
Mask ROM 256KB (Writing or replacement is not possible)
There is no compatibility between the IGB Series main board (PS-018) and the IGX Series
main board (PS-016); However, basic operations are same for these models.
3.2.3 Power Board (PS-019)
+
−
Batteries
Jack center +
Jack center −
CN2
2
1
GND
PS-019
CN3
1
To main board
2
When the DC plug of the AC adapter is inserted into the DC jack, the battery GND will be interrupted.
Manual No. 085-3435-07 33/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 35
3.3 Electric Parts (IGX Series)
3.3.1 IGX Block Diagram
Display Unit
Main Board (PS-016)
XH 7p
Front Case
Rear Case
Earth Board
(steel)
Fuse
holder
Transformer
EL-R 2P
EL-P 2P
AC Input
Scale Unit
Load Cell
Manual No. 085-3435-07 34/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 36
3.3.2 Main Board (PS-016)
Parts Surface of Main Board (The LCD and tact keys are installed on the soldering surface).
(1) SW1: Memory Switch
Used to write the data when performing initialization in Test Mode.
Used to save (write) the data in E2ROM after changing Country No. and Scale No., and performing
Manual No. 085-3435-07 35/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 37
(4) Connector XJ3
This connector is not used.
Pin No. Signal Name
1 FG
2 GND
3 TxD
4 RxD
5 DTR
6 Buzzer
(5) Program Memory Media
Mask ROM 256KB (Writing or replacement is not possible)
There is no compatibility between the IGB Series main board (PS-018) and the IGX Series main
board (PS-016). However, basic operations are same for these models.
Manual No. 085-3435-07 36/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 38
Chapter 4 Maintenance
4.1 Disassembly Procedure for Display Unit
Disassembly procedure
If batteries are being used, make sure to remove them, or when using an AC adapter,
1
extract the cable from the power outlet. (IGB Series only) When using AC power, make
sure to extract the cable from the power outlet. (IGX Series only)
Remove the two angle-adjusting knobs.
2
Remove the Void Seal covering the sealing screw hole.
3
Remove the 8 screws (including the sealing screw mentioned above) that hold the front and
4
rear cases together.
Manual No. 085-3435-07 37/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 39
Open the front and rear cases.
5
IGB Series
Note: Reverse the procedure to re-assemble the display unit, then affix a new Void Seal
(Part No.: 040-9585-01 Size:
φ
21)
IGX Series
4.2 Replacement of Main Board
4.2.1 IGB Series
(1) The procedure for opening the Front and Rear Cases is described in the previous section.
(2) Main Board (PS-018)
Extract the cable from the XJ2
connector.
Desolder the four lead wires of the
Load Cell cable.
Manual No. 085-3435-07 38/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 40
Remove the four screws holding the main
board.
Reverse this procedure to install a new main board.
Note
After installing a new main board, carry out RAM Clear in C3 mode, adjust the scale
and set the user operation setup and program modes.
4.2.2 IGX Series
(1) The procedure for opening the Front and Rear Cases is described in the previous section.
(2) Main Board (PS-016)
Extract the cable from the XJ2 connector
Desolder the four lead wires of the Load Cell
cable
Manual No. 085-3435-07 39/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 41
Remove the screw holding the grounding
cable to the rear case
Remove the four screws holding the main
board
Reverse this procedure to install a new main board.
Note
After installing a new main board, carry out RAM Clear in C3 mode, adjust the scale
and set the user operation setup and program modes.
Manual No. 085-3435-07 40/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Using an allen wrench, remove the four hexagon-headed bolts and spacers fixing the load
1
cell unit from the top of the base.
Remove the four hexagon-headed bolts and spacers from the bottom of the base.
2
Rated
Capacity
Rated
Output
Resistance
Input
Output
Resistance
insulation
Resistance
5GΩ
5GΩ
Manual No. 085-3435-07 41/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 43
Remove the load cell unit. Then, install a new load cell unit and reverse the above
3
procedure.
4.3.1 Checking and Adjusting Gap of Four-corner Limit
In the two spots of the four-corner limit front and of the four-corner limit rear (display pole side), adjust
the gaps to the following values.
Manual No. 085-3435-07 42/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 44
4.3.2 Performing Zero Point and Span Adjustments
e
Refe re nc
Reference
Refer to 2.2 “C1 Mode” of Chapter 2 “Test Mode”.
Manual No. 085-3435-07 43/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 45
4.4 Troubleshooting
Symptoms Causes Remedies
The display does not appear when
ON/OFF is pressed.
Each ON/OFF key is shown below
for the IGB or IGX Series.
or
“Err” is displayed when the
ON/OFF key is pressed.
1. AC power is not supplied.
(IGX Series only)
2. Battery voltage is low.
(IGB Series only)
3. The ON/OFF key is
defective.
4. The main board is
defective.
1. E2ROM data is garbled.
• Check the voltage of the main
power outlet.
• Check the power cable, and replace
if necessary.
• Check the FUSE (1A 250V)
• Check the TRANS ASSY
• Check the main board PS-016
connector XJ2.
• Check batteries, and replace if
necessary.
• When using the AC adapter, confirm
that the adapter voltage is within the
range of 3.2 to 6VDC.
• Using the tester, check that the key
is conducting, and replace if
necessary.
• Replace the main board PS-016
(IGX) or PS-018 (IGB) with a normal
one, then check.
• Initialize E2ROM (all) in Test Mode
C6.
• (Span adjustment is required after
initialization)
• If recovery is not possible, the main
board may be defective. Replace if
necessary.
Test mode “C1” is displayed when
the ON/OFF key is pressed.
“-----“ is displayed after the
ON/OFF key is pressed and the
display is checked.
Scale data is in initialized
state.
1. A/D value at Zero point is
outside of start range, or
unstable.
Out of AD initial value
Load cell is defective.
2. Main board is defective.
• Perform span adjustment.
• Remove items from the weighing
platter.
• Check A/D value and adjust.
Perform Test Mode C1, and if the
value is unstable, first replace the
main board, then replace the load
cell.
If outside of the start range, perform
Zero and Span adjustments.
When performing Zero and Span
adjustments, if the weight that is
less than the weighing capacity
reaches the lower limit, replace the
load cell.
• Replace the main board PS-016
(IGX) or PS-018 (IGB) with a
normal one, then check.
Manual No. 085-3435-07 44/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 46
Symptoms Causes Remedies
Zero point or weight is unstable.
No response or input difficulty in
key operation.
1. Vibration due to wind or
conditions at place of
installation.
2. Interference to Weighing
platter, Platter support, or
Load cell.
3. Main board or Load cell is
defective.
4. Extraordinary
electromagnetic wave
1. Defective key
2. Bad clearance between
Keysheet and Keys (Main
board)
• If Weighing platter is subject to
wind, move the scale to the place
where there is no wind, or provide
something to block the wind.
If there is any vibration at the place
of installation, move the scale to the
place where there is no vibration.
Perform visual inspection to check
whether something is touching the
Weighing platter, Platter support, or
Load cell, and remove if any.
• Perform Test mode C1 and check
A/D value, and replace A/D board
and Load cell in this order, if
necessary.
• Determine the source and remove
it, or move the scale to a place
where the scale will not be subject
to the electromagnetic waves.
• Replace the tact key soldered to the
Main board.
• Install Main board along Keysheet
face.
Non-displaying part, double
display digit, or double segment.
3. Defective Main board
1. Defective Main board
• Replace Main board.
• Replace Main board.
Manual No. 085-3435-07 45/45 IGB/IGX Series (Overseas Specifications) Service Man ual
Page 47
ISHIDA CO.,LTD
44 SANNO-CHO,SHOGOIN,SAKYO-KU
KYOTO,606-8392 JAPAN
PHONE:81-75-771-4141
FACSIMILE:81-75-751-1634
ULR:http://www.ishida.co.jp
Design and specifications are subject change without notice.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.