Rice Lake IQ plus 390-DC User Manual

IQ plus 390-DC Digital Weight Indicator
Live Weighing Option
PNs 53852 and 65738
The live weighing option extends the function of the IQ plus 390-DC indicator to provide the specialized weight sampling, display, and print functions needed for animal weighing applications. This addendum describes the conÞguration and operation of the live weighing option only. See the IQ plus 390-DC complete installation, conÞguration, and calibration information.
NOTE: Live Weighing Option 53852 uses CPU board PN 48816, software version AW1.01; Option 65738 uses
CPU board PN 52380 with software version AW1.02. The external function of the two software versions is identical, but each must be paired with the correct CPU board.
Configuration
The extended conÞguration for the live weighing option is contained in the PROGRM menu. All other conÞguration menus are identical to the standard IQ plus 390-DC menus.
The ANIMAL submenu on the PROGRM menu (see Figure 1 on page 2) provides parameters to do the following:
¥ Enable the live weighing option ¥ Set the sampling time for acquiring animal
weight ¥ Set the display time for the acquired weight ¥ Select whether the animal weight is
automatically printed
¥ Select whether the indicator display shows the
current scale weight (live animal weight) or is blank (no weight displayed) when the speciÞed sample and display times have expired
Table 1 on page 2 describes each of the ANIMAL submenu parameters.
NOTE:
All features of the standard indicator are included in the option software. When the live weighing option is disabled, the indicator functions as a standard IQ plus 390-DC. Use of piece count mode is not recommended when the live weighing option is enabled.
Installation Manual for
Operation
With the indicator conÞgured for live weighing and calibrated, do the following to operate in live weighing mode:
1. Zero the scale. Acquire tare, if any.
PRINT
2. Lead animal onto scale, then press acquire the animal weight. The indicator display shows the message acquires the sample weight.
3. When the sample time (speciÞed on the SMPTIM parameter) expires, the display shows the animal weight acquired over that time and ßashes the units annunciator to indicate acquired weight display. The acquired weight is displayed for the length of time speciÞed on the DSPTIM parameter.
¥ If AUTOPRT is ON, a print ticket is sent to
the serial port as soon as the weight is acquired.
July 2000 53673
–SMP–
to
as it
¥ If AUTOPRT is OFF, press
the acquired weight is displayed to send
data to the serial port.
4. When the display time (DSPTIM parameter value) expires, the indicator display returns to the display mode speciÞed on the WEIMOD parameter:
¥ If WEIMOD is NORMAL, the display
shows the current scale weight until the next weight sample is acquired by pressing
PRINT
the
¥ If WEIMOD is BLANK, the numeric digits
on the display are blanked out (no weight display). The only weight values shown in this mode are the acquired animal weights. The display remains blank until the next weight sample is acquired by pressing the
PRINT
key.
key.
PRINT
while
2
Serial Data Considerations
When operating in live weighing mode with AUTOPRT set OFF, ensure that the acquired weight display time (DSPTIM parameter) is set to a large enough value to allow the operator time to press the
PRINT
required. In live weighing mode, the
key to generate a demand print ticket, if
PRINT
key can send data to the serial port only while the acquired weight is displayed; pressing the
PRINT
key at any
other time acquires a new weight sample.
Live Weighing Program Menu
ANIMAL
ACCESS
ENABLE
DSABLE
SMPTIM
000040
000001–
002400
sample and display times
specified in 0.1-second intervals
DSPTIM
000080
000001–
999999
AUTPRT
OFF
ON
Demand print data (SERIAL menu STREAM parameter set OFF) uses the NFMT ticket format if a tare is in the system, GFMT if no tare is acquired or entered. The <G> format command on the GFMT ticket and the <N> format command on the NFMT ticket both represent the animal weight.
When streaming continuous data (SERIAL menu STREAM parameter set to LFT or INDUST), the weight data displayed or printed depends on the value set for the WEIMOD parameter: If WEIMOD is set to BLANK and no weight data is shown on the current display, streamed weight data is also blank.
XXXXXXXXXXXXXX
WEIMOD
NORMAL
BLANK
VERSPROGRM PFORMTSERIALCALIBRCONFIG FORMAT
PWRUPM
GO
DELAY
•••
(same as standard PROGRM menu)
Figure 1. ANIMAL Submenu on IQ plus 390-DC PROGRM Menu
PROGRM Menu / ANIMAL Submenu
Parameter
Level 3 Submenus
ACCESS
SMPTIM
DSPTIM
AUTPRT
WEIMOD
Choices Description
ENABLE
DSABLE
000040
number
000080
number
OFF
ON
NORMAL
BLANK
Table 1. ANIMAL Submenu Parameters on PROGRM Menu
Animal mode access. Specify ENABLE to use the live weighing option; specify DSABLE to operate as standard indicator.
Sample time. Specifies, in 0.1-second intervals, the sample time during which animal weight is acquired. The default value (000040) is 4 seconds.
Display time. Specifies, in 0.1-second intervals, how long the acquired animal weight is displayed. The default value (000080) is 8 seconds.
Automatic print. Specifies whether animal weight data is automatically sent to the serial port once the weight is acquired.
Weigh mode. Specifies whether the indicator display shows the current scale weight (NORMAL) or is blank (no weight data) when the acquired weight display time expires.