The Equipoise balance amplifies signals from a load cell, converts it to digital data and
displays it as a mass value. It is an accurate, fast and versatile series of general purpose
balances with counting and percent weighing functions.
Other features include:
• 25 mm LCD with black backlit display
• Keypads are light touch switches
• Battery provides up to 40 hours of continuous use (without backlight)
• Capacity 300 g (.661lb) to 3000 g (6.613 lb).
All units include automatic zero tracking, automatic tare and an accumulation facility
which allows the count to be stored and recalled as an accumulated total.
This manual can be viewed and downloaded from the Rice Lake Weighing
Systems website at: www.ricelake.com
Warranty information can be found on the website at www.ricelake.com/
warranties
1.1Safety
Safety Symbol Definitions
Indicates a potentially hazardous situation that, if not avoided, could
result in serious injury or death, and includes hazards that are exposed
when guards are removed.
Indicates information about procedures that, if not observed, could
result in damage to equipment or corruption to and loss of data.
Safety Precautions
Do not operate or work on this equipment unless you have read and
understand the instructions and warnings in this manual. Failure to
follow the instructions or heed the warnings could result in injury or
death. Contact any Rice Lake Weighing Systems dealer for
replacement manuals. Proper care is your responsibility.
General Safety
Failure to heed may result in serious injury or death.
Some procedures described in this manual require work inside the indicator enclosure.
These procedures are to be performed by qualified service personnel only.
Before opening the unit, ensure the power cord is disconnected from the outlet.
Disconnect all power to this unit before installing, cleaning or servicing.
Do not allow minors (children) or inexperienced persons to operate this unit.
Safety1
Page 6
Do not operate without the enclosure completely assembled.
Important
Do not use for purposes other than weight taking.
Do not place fingers into slots or possible pinch points.
Do not use this product if any of the components are cracked.
Do not exceed the rated specification of the unit, see Section 9.17 on page 74.
Do not make alterations or modifications to the unit.
Do not remove or obscure warning labels.
Do not submerge.
The balance is a precision electronic instrument. Handle it carefully.
Do not install the balance in direct sunlight.
Verify that the local voltage and receptacle type are correct for the balance.
Only use original adapter. Others could cause damage to the balance.
Balance should be located in an area near an easily accessible socket outlet.
Avoid unstable power sources. Do not place near equipment such as welders or large
motors that use large amounts of electricity
Avoid sudden temperature changes, vibration, wind and water.
Avoid heavy RF noise.
Keep the balance clean.
2Equipoise Series Technical Manual
Page 7
1.2Description
On/OffPrintModeSampleTareZero
Sample
Percent
Mode
000000
lb
0
F
NET
%
+
-
KeyDescription
Turns the balance power
On / Off
Subtracts weight of container
Sets display to Zero
Set percentage weighing
function
Set to counting mode
On/Off
Tare
Zero
Percent
Sample
Sample
Set to weighing units
Recalls Memory/Print the results
Standstill Indicator
Zero Indicator
NET
Tare Indicator
%
Percent Weighing Indicator
Battery Indicator
KeyDescription
Mode
Mode
Print
+
Keyboard
Figure 1-1. Equipoise Overlay
-
Table 1-1. Key and Display Description
Description3
Page 8
2.0Installation
3
1
2
1
2
3
4
Item No.Description
1Air Case Top Cover
2Glass Side-1 covers x 2
3Pole of glass
Item No.Description
2.1Unpacking
Use the following instructions to unpack and inspect the balance.
1. Remove the weighing balance from the carton.
2. Remove the protective covering.
3. Inspect the balance and terminal for damage.
4. Make sure all components are included.
Components
Figure 2-1. Component Parts
Item No.Description
1Balance
2Adapter
3Pan
Table 2-1. Component Parts List
Air Protection Case
The air protection case is used to protect the balance from air variations while weighing
and is available with light capacity balances only. Carefully secure glass sides 2 & 4 with
poles.
Figure 2-2. Air Protection Parts
Table 2-2. Air Protection Parts List
4Equipoise Series Technical Manual
Page 9
2.2Battery Charging
+
-
LED Light located on
display panel
Note
Adapter
Jack
RS-232
Figure 2-4 Jack Location
The balance can be operated by battery if desired. Battery life is approximately 40 hours
for continuous use without a backlight.
When the battery needs to be charged, the battery indicator ( ) will turn on.
1. To charge the battery, insert the adapter pin into the jack, located on the right side of
the balance.
2. Plug the balance into a power source. The balance does not need to be turned on.
The battery should be charged for 12 hours for full capacity.
There is an LED to indicate the status of the battery as it is charging. When the balance is
plugged in, the internal battery will be charged.
The color of the LED indicates the following:
• Green — battery has a full charge
• Yellow — battery is nearly full of charge
• Red — battery is being charged
Figure 2-3. LED Light Display
Do not use any other type of power adapter than the one supplied with the balance.
Verify that the AC power socket outlet is properly protected.
Charge the battery before using the balance for the first time.
As the battery is used, it may fail to hold a full charge. If the battery life
becomes unacceptable, contact your distributor.
2.3Setting Up The Balance
1. Place the balance on a table or another flat
surface.
2. Plug the adapter pin into the balance
adapter jack, located on the right side of the
balance.
3. The adapter plugs into an AC power outlet.
The balance should be located near an
easily accessible socket outlet with a
protective ground/earth contact.
4. Press the
Press the key again to turn off the power.
5. The display will show the version number
and will begin a self test.
6. When the self test is complete, the normal
weighing mode will be displayed.
7. A warm-up time of 15 minutes stabilizes
the measured values after switching on.
8. Calibrate with exact calibration weights. A minimum of 1/3 of the balance capacity
should be used for calibration. See details in Section 4.1 on page 11. The balance is
now ready for use.
On/Off key to turn on the power.
Battery Charging5
Page 10
2.4Level Adjusting
Not LevelLevel
Figure 2-5 Bubble Level
Note
1. Place the balance on a table or another flat
surface.
2. Check the bubble level. If bubble is not
centered, adjust the leveling feet until it
reaches the center.
Always re-check the level when
changing the location.
6Equipoise Series Technical Manual
Page 11
3.0Operation
On/Off
On/Off
Zero
Tare
Tare
3.1Initial Start-up
Warm up time of 15 minutes stabilizes the measured values after switching on.
3.2Power On/Off
1. Press to turn the balance on. The self test will start.
2. Press again to turn balance off.
3.3Zero
Environmental conditions can lead to the balance not reading zero in spite of the pan not
taking any strain.
Press to set the display to zero.
3.4Tare
The weight of any container can be tared (subtracts the weight of the container and zeros
the balance) so that with subsequent weighing the net weight of the object being weighed
is always displayed.
1. Place a container on the pan.
2. Press . Tare is subtracted and zero is displayed. When the container is
removed from the platform, the tared weight is displayed as a negative value.
3. Press again to clear the tare.
Initial Start-up7
Page 12
3.5Weighing Units
Mode
Mode
Note
Percent
Press in the weighing mode to change the unit of measure:
Unit MarkUnit NameConversion (g)
ggram
ctnet carat0.2g
Lbpound453.59237g
OZounce28.349523125g
ddram1.7718451g
GNgrain0.06479891g
OZttroy ounce31.1034768g
dWtpenny weight1.55517384g
MMmomme3.749996g
TLTtael twn37.49995g
TLhtael chn37.799375g
TLJtroy tael37.4290018g
ttola11.6638039g
btbangladeshi tola1bt=11.6638039g
Nnewton1N=101.916g
Table 3-1. Weight Unit List
Units can be set to On/Off in the parameters settings. See Section 5.2 on
page 12. If Unit is set to off, it will not show when pressing the Mode key.
1bt=16Ana=96Roti
3.6Percent Weighing
The balance can set a base weight to be shown as 100 percent. Then when other weights
are placed on the balance, it will be displayed as a percentage of the original base weight.
For example:
1. Place base weight on the balance.
2. Press . The display will show 100.00 percent.
3. Remove the weight and wait for the display to return to zero.
4. Place the weight you want to compare to the base weight on the platform. The
display shows what percent the current weight is of the base weight.
5. The weighing may be amended on the basis of greater numbers of samples and may
improve the accuracy of percentage of large quantities.
8Equipoise Series Technical Manual
Page 13
3.7Parts Counting
Sample
Sample
Mode
Mode
Zero
Sample
Sample
Note
Tare
Zero
Sample
Sample
1. Press to enter the parts counting mode.
2. Select the sample count by pressing until the desired count is shown. Place
selected quantity of product on the balance.
DisplayCount number
sp 10 pcsFor 10 pieces
sP 20 pcsFor 20 pieces
sP 50 pcsFor 50 pieces
sP 100 pcsFor 100 pieces
sP 200 pcsFor 200 pieces
Table 3-2. Display/Count
3. Press to confirm.
4. Add more items. The weight display will show the total number of parts.
5. Press to exit counting mode.
The balance retains the sample rate until exiting the counting mode.
3.7.1Parts Counting With A Container
1. Place a container on the platform.
2. Press to zero the balance.
3. Select the parts quantity as shown in Table 3-2 .
4. Place the specified number of parts in the container.
5. Press to confirm. The display will show ---- then the quantity.
6. Add product on the platform. The display will update the parts quantity
automatically to total number of parts in the container.
7. Press to exit counting mode.
Parts Counting9
Page 14
3.8 Accumulation
Print
Note
Note
Print
Print
Sample
Sample
Before setting the balance to accumulate, it should be stable () and the display should
be at zero ( ). Accumulation is available only when weight is more than 20 divisions (or
20 times the resolution of the balance).
Accumulation Operation
1. Place the load on the platform.
2. When the stable indicator () is displayed, press .
3. The display will show ACC x (x = total number of accumulations) for three
seconds.
The total saved value will then display for three seconds.
If the optional RS-232 interface is installed, the weight data will be sent to the
printer.
4. Remove the weight from the pan.
5. When the display returns to zero and the stable () indicator is shown, place
another weight on balance.
6. Repeat steps 2-5 for all items to be weighed.
Display will show ADDERR when memory is full.
3.8.1 Memory Recall
To recall the memory, press .
Display will show ACC x (x = total number of accumulations) for three seconds.
The total saved value will then display for three seconds.
3.8.2 Memory Clear
To clear the memory:
1. Press .
2. While accumulation is displayed, press .
The display will return to normal mode; all accumulated memory will be cleared.
10 Equipoise Series Technical Manual
Page 15
4.0Calibration
Zero
Zero
Zero
Mode
Mode
Mode
Mode
Zero
Sample
Sample
Print
Tare
Zero
Print
Sample
Sample
Percent
4.1Normal Calibration
1. Press and hold to enter calibration. unload is displayed.
2. Remove any weight from the platform.
3. After the stable indicator () is displayed, press to confirm. Load will
display.
4. Place the calibration weight (1/3 of the capacity) on the platform.
5. When the scale is stable, press key to confirm.
4.2Linear Calibration
1. Turn on the balance. Press during the self test. F1 UNT will display.
2. Press several times to scroll to tech, then press to enter calibration.
N will display.
3. Press
4. Press to enter calibration.
5. Press in sequence.
6. Ensure the platform is empty. After the stable () and zero ( ) indicators come
LOAD 1 will display.
on,
7. Place the first calibration weight (1/3 of the capacity) on the platform. After the
stable () and zero ( ) indicators come on,
8. Place the next calibration weight (2/3 of the capacity) on the platform. After the
stable () and zero ( ) indicators come on,
9. Place the last calibration weight (full capacity) on the platform. After the stable
() and zero ( ) indicators come on, the balance will start self test and return to
normal weighing mode.
in sequence. P 1 LIN will display.
PI N will display.
load 0 will display.
load 2 will display.
load 3 will display.
PI
Normal Calibration 11
Page 16
5.0Parameter Settings
On/Off
Mode
Mode
Mode
Mode
Zero
Mode
Mode
Print
Print
Zero
Mode
Mode
Zero
Print
Zero
Mode
Mode
Zero
Print
Note
5.1Enter Parameter Settings
1. Press to turn the balance on.
2. Press during the self test to enter parameter settings mode.
3. Press to scroll to desired parameter.
4. Press to enter parameter and scroll through choices.
5. Press to change setting.
6. Press the screen will display EX F1.
7. Press to return to weighing mode.
5.2Set Weighing Unit
Enter parameter settings (see Section 5.1 ).
1. At F1 unt, press to enter unit settings.
2. Press to set each unit to On/Off.
3. Press to advance to the next unit.
4. After all units have been set, press to return to weighing mode.
5.3Set Backlight
Enter parameter settings (see Section 5.1 ).
1. At F2 El, press to enter backlight setting.
2. Press to select backlight mode. See Table 5-1 on page 14.
3. Press to accept.
4. Press to return to weighing mode.
When the battery is low, the backlight function is not available.
12 Equipoise Series Technical Manual
Page 17
5.4Communications
Zero
Mode
Mode
Zero
Mode
Mode
Zero
Mode
Mode
Zero
Mode
Mode
1. At F3 Con, press to enter the communications setting.
2. Press to modify the print setup. See Table 5-1 on page 14.
3. Press to accept the baud rate settings.
4. Press to toggle between choices.
5. Press to select printer.
6. Press to toggle between TP (tape printer) and LP-50 (not used).
7. Press to accept.
8. Press to return to the weighing mode.
Communications 13
Page 18
Menu Sub MenuDescription
F1 untg / ct / lb / oz / d / gn / ozt /
dwt / mm / tl.T / tl.c / tl.t / t,.
F2 elEl auOnWhen weight is on the pan, the backlight
Off
El onOnBacklight is always on when the balance is
Off
El offOnBacklight is always off when the balance is
Off
F3 comP prtBy pressing the Print key, the weighing
P contSends data continuously.
P autoAutomatic accumulation — Individual
Set BAUD rate:
After setting the RS-232 mode, the display will be shown with the current
baud rate b XXX.
Baud rate Options: b600, b1200, b2400, b4800 and b9600
Set Printer type:
After setting the BAUD rate, the display will show the printer type.
Lp 50Not available
TpTicket printer
F4 ACCACC ONTurn on accumulation.
ACC OFTurn off accumulation.
techpinFactory access only
P 1 linLinear calibration
P 2 CalNormal calibration
P 3 cntxxxxThis display will display XXXXX for
P 4 a 2n0 .5 dSet auto zero
1 d
2 d
4 d
P 5 graXxxxSet local gravity
Each unit can be set to on or off.
will turn on.
turned on.
turned on.
value will be added to the memory and
sent to the printer if connected.
weighing values are automatically added.
indicating the internal counts.
Table 5-1. Parameter Settings
14 Equipoise Series Technical Manual
Page 19
6.0RS-232 Output
6.1Specifications
• RS-232 output of weighing data
• ASCII code
• 600~9600 baud
• 8 data bits
• No parity
6.2RS-232 (9-pin D Type Connector)
PinSignalDescription
Pin 2RXDInputReceiving data
Pin 3TXDOutputTransmission data
Pin 5GND--Signal ground
Table 6-1. 9-Pin Connector Types
9-pin D Connector:
Balance Computer / Printer
Pin 2:Pin 3
Pin 3:Pin 2
Pin 5:Pin 5
Table 6-2. Balance to Printer Connector
Continuous Output Protocol (Weighing Mode)
ST , GS-/kgCRLF
Header 1Header 2Weight DataWeight
Unit
Table 6-3. Continuous Output Protocol
Te rm i n at o r
6.3RS-232 Serial Commands
1. R= Send
2. T= Tare
3. Z= Zero
Specifications 15
Page 20
7.0Maintenance
7.1General
Use the following instructions to complete general maintenance on the Equipoise balance.
7.1.1Balance Does Not Operate Properly
Determine whether the problem is constant or alternate. Be aware that problems can be
caused by mechanical or electrical influences.
Check for the following:
• Water
• Corrosive materials
• Vibrations, temperature or wind
• Physical damage
Check the balance cables for damage. Check all connections and connectors for any loose
contact or incorrect connection.
7.1.2Cleaning
Use the following instructions to clean the Equipoise:
• Disconnect the power before cleaning.
• Use a clean and soft cloth with mild cleaning agents.
• Make sure that the fluid is not able get into the device.
16 Equipoise Series Technical Manual
Page 21
7.2Error Codes
Zero
Error
Code
Err 1Date set error--Check the input data
Err 2Time set error--Check the input data
Err 3Calibrationweight is too light--Check the calibration
Err 4Zero range exceeded, due to
turning on or by pressing
Err 5Keyboard errorHolding a key over
Err 6A/D Count out of the rangePlatform is not
Err 7Pecentage setting error--0.01% weight must be
Err 8External calibration weight errorMore than 4% error
Err 9Unstable, can’t perform zero or
tare
Err 10No data RS-232, continous
Err 11Won’t accept communication
protocal
Err 12Accumulation errorMax accmulate
Err 13Lack of unit weightCounting: unit
Err 14Lack of sampleCounting: sample
Err 15Error gravity value settingGravity value range
DescriptionPossible CauseSolution
weight
Goods on the
platform
Remove good from the
platform.
overloaded, when
zeroing the balance
Improper calibration
Load cell problem
PCB problem
Press the key for
30 seconds
shorter length of time.
Re-install platform.
installed
Load cell problem
Check the load cell.
PCB problem
>0.5d
--
as linear calibration
----
--
communication
Can get signal, but
--
can’t support
-time 999 or total
weight 999999 has
been reached.
-weight < 0.5d
-weight < 20d
--
0.96-1.04
Table 7-1. Error Codes
Error Codes 17
Page 22
Error
Code
Err 16No paper----
Err 17Tare OperationMinus weight or
Err 18Incorrect pre-tare operationMinus weight or
Err PPrinter no response two
seconds later
OKSet password OK----
PASSCalibration OK----
FailCalibrate or password setting
fail
NGRS-232 Command error----
--OL-OverloadMax +9d (NTEP:
--LO-Minus weight-5% capacity--
Bat loBattery voltage too low<5.9v--
DescriptionPossible CauseSolution
overload
overload
----
----
105% capacity)
Table 7-1. Error Codes
Measuring PointsResistance
Red (+ Exc) to White ( –Exc) 420 ±20Ω
Green (+Sig) to Black ( –Sig) 350Ω ±5Ω
Table 7-2. Measuring Points
--
--
--
18 Equipoise Series Technical Manual
Page 23
7.3Replace Battery
Use the following steps to replace the battery.
1. Remove the pan.
2. Remove the top cover.
3. Remove the two screws from the battery clamp.
4. Remove the connectors from the battery
terminals.
5. Remove the used battery and replace with a
new a battery.
6. Connect the connectors to the battery terminals.
7. Install the two screws to secure battery clamp.
8. Install top cover and pan.
9. Turn on the balance.
10. Charge the battery before using the balance for
the first time.
Replace Battery 19
Page 24
8.0Troubleshooting
ProblemsPossible causeCommon Solutions
Display is
blank.
No self test.
Blank display
after self test
OL or -------Maximum capacity exceeded
------- or
NULL
Display is
unstable
Weight value
incorrect
Cannot use
full capacity
Platform
Corner
Weight
different
Battery is not
charging
Main power is turned off.
Power supply faulty or not
plugged in.
Internal battery is not charged.
On/Off switch is faulty.
Pan not installed
Unstable weight
Load cell damaged
Load cell or mechanics
damaged
Power supply faulty
Weight is on the platform is
below permissible limit.
Pan not installed correctly.
Power supply faulty.
Load cell or mechanism faulty
Goods touching somewhere.
Air variation or any vibrations.
Temperature changed .
Load cell or connections faulty.
Power supply faulty
Calibration error.
Platform of load cell touching
somewhere.
Wrong weighing unit
Over load protection stoppers or
transport locks are not removed.
Parameters are set incorrectly.
AD problem.
Load cell or mechanism
damaged
Over load protection stoppers or
transport locks are not removed.
Load cell or mechanism
damaged
Main voltage problem
Charging circuit problem
Battery Problem
Ensure there is power to the balance
and that the on/off switch is working.
Verify the voltages, which are located on
the power labels.
Ensure the pans are installed correctly.
Try to turn on the balance again.
Ensure the platform is installed correctly.
Try to turn on the balance again.
Do the calibration again.
Ensure the platform is installed correctly.
Try to turn on the balance again.
Do the calibration again.
Ensure the balance is in acceptable
location.
Check the connecters and load cell.
Check the power supply and battery
Use accurate weight for doing the
calibration.
Ensure the pan and load cell are
properly installed and are touching.
Check the parameter settings.
Check the load cell and connecters.
Check the stoppers and locks under the
platform.
Check the weighing unit and parameter
settings.
Check the load cell.
Check the stoppers and locks under the
platform.
Use accurate weight for doing the
calibration.
Check the load cell.
Check the main voltage and adapter.
Check the battery.
Check the charging circuit.
20 Equipoise Series Technical Manual
Page 25
Page 26
230 W. Coleman St. • Rice Lake, WI 54868 • USA
U.S. 800-472-6703 • Canada/Mexico 800-321-6703 • International 715-234-9171 • Europe +31 (0)26 472 1319