RayTek MARATHON MM SERIES Operating Instructions Manual

Page 1
MARATHON MM
SERIES
High-Performance Infrared Thermometer
Operating Instructions
Rev. D7 Jul 2017
58201
Page 2
Page 3
Contacts
Fluke Process Instruments

Americas
Santa Cruz, CA USA Tel: +1 800 227 8074 (USA and Canada, only) +1 831 458 3900
solutions@flukeprocessinstruments.com
EMEA
Berlin, Germany Tel: +49 30 478 0080
info@flukeprocessinstruments.de
China Beijing, China Tel: +86 10 6438 4691
info@flukeprocessinstruments.cn
Worldwide Service
Fluke Process Instruments offers services, including repair and calibration. For more information, contact your local office.
www.raytek.com
© Fluke Process Instruments Specifications subject to change without notice.
Page 4
WARRANTY
The manufacturer warrants this instrument to be free from defects in material and workmanship under normal use and service for the period of two years from date of purchase. This warranty extends only to the original purchaser. This warranty shall not apply to fuses, batteries, or any product which has been subject to misuse, neglect, accident, or abnormal conditions of operation.
In the event of failure of a product covered by this warranty, the manufacturer will repair the instrument when it is returned by the purchaser, freight prepaid, to an authorized Service Facility within the applicable warranty
period, provided manufacturer’s examination discloses to its satisfaction that the product was defective. The
manufacturer may, at its option, replace the product in lieu of repair. With regard to any covered product returned within the applicable warranty period, repairs or replacement will be made without charge and with return freight paid by the manufacturer, unless the failure was caused by misuse, neglect, accident, or abnormal conditions of operation or storage, in which case repairs will be billed at a reasonable cost. In such a case, an estimate will be submitted before work is started, if requested.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY, FITNESS, OR ADEQUACY FOR ANY PARTICULAR PURPOSE OR USE. THE MANUFACTURER SHALL NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES, WHETHER IN CONTRACT, TORT, OR OTHERWISE.
SOFTWARE WARRANTY
The manufacturer does not warrant that the software described herein will function properly in every hardware and software environment. This software may not work in combination with modified or emulated versions of Windows operating environments, memory-resident software, or on computers with inadequate memory. The manufacturer warrants that the program disk is free from defects in material and workmanship, assuming normal use, for a period of one year. Except for this warranty, the manufacturer makes no warranty or representation, either expressed or implied, with respect to this software or documentation, including its quality, performance,
merchantability, or fitness for a particular purpose. As a result, this software and documentation are licensed “as is,” and the licensee (i.e., the User) assumes the entire risk as to its quality and performance. The liability of the
manufacturer under this warranty shall be limited to the amount paid by the User. In no event shall the manufacturer be liable for any costs including but not limited to those incurred as a result of lost profits or revenue, loss of use of the computer software, loss of data, the cost of substitute software, claims by third parties, or for other
similar costs. The manufacturer’s software and documentation are copyrighted with all rights reserved. It is illegal
to make copies for another person.
Specifications subject to change without notice.
Page 5

COMPLIANCE STATEMENT

The device complies with the requirements of the European Directives:
EC – Directive 2014/30/EU – EMC
EC – Directive 2011/65/EU – RoHS II
EN 61326-1: 2013 Electrical measurement, control and laboratory devices -
Electromagnetic susceptibility (EMC)
EN 50581: 2012 Technical documentation for the evaluation of electrical products with respect
to restriction of hazardous substances (RoHS)
Electromagnetic Compatibility Applies to use in Korea only. Class A Equipment (Industrial Broadcasting & Communication Equipment)
This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and is not to be used in homes.
Page 6

Content

COMPLIANCE STATEMENT .......................................................................................................................... 5
CONTENT ............................................................................................................................................................ 6
1 SAFETY INSTRUCTIONS ........................................................................................................................... 10
2 PRODUCT DESCRIPTION .......................................................................................................................... 13
3 TECHNICAL DATA ...................................................................................................................................... 14
3.1 MEASUREMENT SPECIFICATIONS ............................................................................................................... 14
3.2 OPTICAL SPECIFICATIONS .......................................................................................................................... 16
3.2.1 Variable Focus ..................................................................................................................................... 16
3.2.2 Fixed Focus ......................................................................................................................................... 17
3.3 ELECTRICAL SPECIFICATIONS .................................................................................................................... 22
3.4 ENVIRONMENTAL SPECIFICATIONS ........................................................................................................... 22
3.5 DIMENSIONS ............................................................................................................................................... 23
3.6 SCOPE OF DELIVERY ................................................................................................................................... 23
4 BASICS ............................................................................................................................................................. 24
4.1 MEASUREMENT OF INFRARED TEMPERATURE .......................................................................................... 24
4.2 EMISSIVITY OF TARGET OBJECT.................................................................................................................. 24
5 ENVIRONMENT............................................................................................................................................ 25
5.1 AMBIENT TEMPERATURE ........................................................................................................................... 25
5.2 ATMOSPHERIC QUALITY ............................................................................................................................ 25
5.3 ELECTRICAL INTERFERENCE ...................................................................................................................... 25
6 INSTALLATION ............................................................................................................................................ 26
6.1 MECHANICAL INSTALLATION ................................................................................................................... 26
6.1.1 Distance to Object .............................................................................................................................. 26
6.1.2 Variable Focus ..................................................................................................................................... 26
6.1.3 Viewing Angles................................................................................................................................... 27
6.2 ELECTRICAL INSTALLATION ...................................................................................................................... 28
6.3 COMPUTER INTERFACING .......................................................................................................................... 30
6.4 MULTIPLE SENSORS IN A NETWORK .......................................................................................................... 32
6.4.1 Wiring ................................................................................................................................................ 32
6.4.2 Addressing .......................................................................................................................................... 33
6.4.3 Configuration Procedure .................................................................................................................... 33
7 OPERATION ................................................................................................................................................... 34
7.1 CONTROL PANEL ....................................................................................................................................... 34
7.2 OPERATION MODES ................................................................................................................................... 34
7.3 SIGNAL PROCESSING .................................................................................................................................. 37
7.3.1 Averaging ........................................................................................................................................... 37
7.3.2 Peak Hold ............................................................................................................................................ 38
7.3.2.1 Reset ............................................................................................................................................. 38
7.3.2.2 Signal Slope ................................................................................................................................. 39
7.3.3 Advanced Peak Hold ........................................................................................................................... 40
7.3.4 Valley Hold ......................................................................................................................................... 41
7.3.5 Advanced Valley Hold ........................................................................................................................ 41
Page 7
7.4 INPUTS AND OUTPUTS ................................................................................................................................ 42
7.4.1 Milliamp Output ................................................................................................................................. 42
7.4.2 Relay Output....................................................................................................................................... 42
7.4.2.1 Thresholds .................................................................................................................................... 42
7.4.2.2 Deadband ..................................................................................................................................... 43
7.4.3 External Input ..................................................................................................................................... 44
7.4.3.1 Trigger .......................................................................................................................................... 44
7.4.3.2 Ambient Background Temperature Compensation ............................................................... 45
7.4.3.3 Emissivity Setting ........................................................................................................................ 47
7.5 FACTORY DEFAULTS ................................................................................................................................... 48
8 OPTIONS ......................................................................................................................................................... 49
8.1 LASER SIGHTING ......................................................................................................................................... 49
8.2 VIDEO SIGHTING ......................................................................................................................................... 50
8.3 AIR/WATER COOLED HOUSING ................................................................................................................. 53
8.3.1 Avoidance of Condensation ................................................................................................................. 54
9 ACCESSORIES ................................................................................................................................................ 55
9.1 OVERVIEW ................................................................................................................................................... 55
9.2 FIXED MOUNTING BRACKET ...................................................................................................................... 57
9.3 ADJUSTABLE MOUNTING BRACKET ........................................................................................................... 57
9.4 AIR PURGE COLLAR .................................................................................................................................... 58
9.5 SIGHT TUBE ................................................................................................................................................. 58
9.6 PIPE THREAD ADAPTER .............................................................................................................................. 59
9.7 RIGHT ANGLE MIRROR ............................................................................................................................... 60
9.8 INDUSTRIAL POWER SUPPLY ...................................................................................................................... 61
9.9 TERMINAL BOX ........................................................................................................................................... 62
9.10 LOW TEMP CABLE ..................................................................................................................................... 63
9.11 HIGH TEMP CABLE ................................................................................................................................... 64
9.12 PROTECTIVE WINDOW .............................................................................................................................. 65
9.13 THERMOJACKET ........................................................................................................................................ 66
9.14 {RESERVED}................................................................................................................................................ 66
9.15 {RESERVED}................................................................................................................................................ 66
10 PROGRAMMING GUIDE .......................................................................................................................... 67
10.1 SERIAL INTERFACE VERSUS CONTROL PANEL ......................................................................................... 67
10.2 STORING OF PARAMETERS ........................................................................................................................ 67
10.3 COMMAND STRUCTURE ............................................................................................................................ 67
10.3.1 Requesting a parameter (Poll Mode) ................................................................................................. 67
10.3.2 Setting a parameter (Poll Mode) ....................................................................................................... 67
10.3.3 Sensor response ................................................................................................................................. 68
10.3.4 Sensor notification............................................................................................................................. 68
10.3.5 Error Messages .................................................................................................................................. 68
10.4 TRANSFER MODES .................................................................................................................................... 68
10.5 CHECKSUM ................................................................................................................................................ 69
10.6 BURST MODE ............................................................................................................................................. 70
10.6.1 Speed ................................................................................................................................................. 70
10.6.2 Minimum Baud Rate......................................................................................................................... 71
10.7 SENSOR INFORMATION ............................................................................................................................. 71
10.8 SENSOR SETUP ........................................................................................................................................... 72
Page 8
10.8.1 General Settings ............................................................................................................................... 72
10.8.2 Sample Time ..................................................................................................................................... 72
10.8.3 Temperature Pre-Processing ............................................................................................................. 72
10.8.4 Temperature Range........................................................................................................................... 74
10.8.5 Emissivity Setting ............................................................................................................................ 74
10.8.6 Ambient Background Temperature Compensation ........................................................................... 74
10.8.7 Temperature Hold Functions ............................................................................................................ 74
10.9 SENSOR CONTROL .................................................................................................................................... 75
10.9.1 Current Output ................................................................................................................................ 75
10.9.2 Relay Output .................................................................................................................................... 75
10.9.3 External Input .................................................................................................................................. 76
10.9.4 Lock Mode ......................................................................................................................................... 76
10.10 RS485 COMMUNICATION ...................................................................................................................... 77
10.11 MULTIDROP MODE................................................................................................................................. 77
10.12 COMMAND LIST ..................................................................................................................................... 78
11 MAINTENANCE .......................................................................................................................................... 81
11.1 TROUBLESHOOTING MINOR PROBLEMS .................................................................................................. 81
11.2 FAIL-SAFE OPERATION ............................................................................................................................ 82
11.3 CLEANING THE LENS ............................................................................................................................... 83
12 APPENDIX..................................................................................................................................................... 84
12.1 DETERMINATION OF EMISSIVITY .............................................................................................................. 84
12.2 TYPICAL EMISSIVITY VALUES ................................................................................................................... 84
13 NOTICES ....................................................................................................................................................... 88
Page 9
Page 10
Safety Instructions
10 Rev. D7 Jul 2017 Marathon MM

1 Safety Instructions

This document contains important information, which should be kept at all times with the instrument during its operational life. Other users of this instrument should be given these instructions with the instrument. Eventual updates to this information must be added to the original document. The instrument can only be operated by trained personnel in accordance with these instructions and local safety regulations.
Acceptable Operation
This instrument is intended only for the measurement of temperature. The instrument is appropriate for continuous use. The instrument operates reliably in demanding conditions, such as in high environmental temperatures, as long as the documented technical specifications for all instrument components are adhered to. Compliance with the operating instructions is necessary to ensure the expected results.
Unacceptable Operation
The instrument should not be used for medical diagnosis.
Replacement Parts and Accessories
Use only original parts and accessories approved by the manufacturer. The use of other products can compromise the operation safety and functionality of the instrument.
Page 11
Safety Instructions
Marathon MM Rev. D7 Jul 2017 11
Safety Symbols
AC (Alternating Current)
DC (Direct Current)
Risk of danger. Important information. See manual.
Hazardous voltage. Risk of electrical shock.
Helpful information regarding the optimal use of the instrument. Earth ground
Protective ground
Fuse Normally-open (NO) relay
Normally-closed (NC) relay Switch or relay contact
DC power supply Conforms to European Union directive.
Disposal of old instruments should be handled according to professional and environmental regulations as electronic waste.
Page 12
Safety Instructions
12 Rev. D7 Jul 2017 Marathon MM
To prevent possible electrical shock, fire, or personal injury follow these guidelines:
Read all safety Information before you use the product. Use the product only as specified, or the protection supplied by the product can be
compromised.
Carefully read all instructions. Do not use and disable the product if it is damaged. Do not use the product if it operates incorrectly. Do not look directly into the laser with optical tools (for example, binoculars, telescopes,
microscopes). Optical tools can focus the laser and be dangerous to the eye.
Do not look into the laser. Do not point laser directly at persons or animals or indirectly off
reflective surfaces.
Do not use laser viewing glasses as laser protection glasses. Laser viewing glasses are used
only for better visibility of the laser in bright light.
Use the product only as specified or hazardous laser radiation exposure can occur.
Page 13
Product Description
Marathon MM Rev. D7 Jul 2017 13

2 Product Description

The Marathon MM is a multi-purpose, high performance single color infrared pyrometer with an extensive feature set, rugged industrial housing and a high level of functionality. The sensor is available in different spectral responses to accommodate the wide range of industrial applications requiring non­contact temperature measurement. It is intended to be simple to operate, with a user-friendly interface and well suited to a wide variety of industrial applications.
Each sensor has a rugged stainless steel housing, a rear membrane panel with backlit LCD display, and a standard through the lens sighting with optional laser target sighting. Alternative to the laser sighting, a video sighting is offered as an option. The Marathon MM is also available with remotely adjustable precision focus optics. Users can easily adjust the focus of measurement targets, either by push-button on the rear of the instrument, or remotely via the RS232/RS485 connection from a PC.
Each model operates as a temperature measurement subsystem consisting of optical elements, spectral filters, detector, and digital electronics. All components are water-tight NEMA-4 (IP65, IEC529) rated and are built to operate on a 100 percent duty cycle in industrial environments. Simultaneous analog and digital outputs consist of standardized signals commonly available for use with computers, controllers, recorders, alarms, or A/D interfaces.
Page 14
Technical Data
14 Rev. D7 Jul 2017 Marathon MM

3 Technical Data

3.1 Measurement Specifications

Temperature Range
LT -40 to 800°C (-40 to 1472°F)
G7 300 to 900°C (572 to 1652°F)
G5L 250 to 1650°C (482 to 3002°F)
G5H 450 to 2250°C (842 to 4082°F)
MT 250 to 1100°C (482 to 2012°F)
3M 100 to 600°C (212 to 1112°F)
2ML 300 to 1100°C (572 to 2012°F)
2MH 450 to 2250°C (842 to 4082°F)
1ML 400 to 1740°C (752 to 3164°F)
1MH 540 to 3000°C (1004 to 5432°F)
Measurements at the low end of the temperature range for the 1M models can be effected by disturbing day light!
Spectral Response
LT 8 to 14 µm G7 7.9 µm G5 5 µm MT 3.9 µm 3M 2.1 to 2.5 µm 2M 1.6 µm 1M 1 µm
Response Time (95%)
LT, MT, G7 120 ms G5 60 ms 3M 20 ms 1M, 2M 2 ms
Exposure Time1 (95%)
1M, 2M 1 ms
1
The exposure time is the minimum time during which the measured object must be present. The output value of the sensor
can be delayed. (VDI/VDE 3511)
Page 15
Technical Data
Marathon MM Rev. D7 Jul 2017 15
System Accuracy2
LT ± 1% of reading or ± 1°C for Tmeas > 0°C (32°F)
(whichever is greater)
MT ± 1% of reading for Tmeas > 350°C (662°F)
± 2°C or ± 2% for Tmeas < 350°C (662°F)
(whichever is greater)
G5L, G5H, G7 ± 1% of reading
3M ± 1% of reading for Tmeas > 150°C (302°F)
± 5°C for Tmeas < 150°C (302°F)
2ML ± (0.3% of reading + 2°C)
2MH ± (0.3% of reading + 1°C)
1ML ± (0.3% of reading + 1°C) for Tmeas > 450°C (842°F)
± (2% of reading + 2°C) for Tmeas < 450°C (842°F)
1MH ± (0.3% of reading + 1°C) for Tmeas > 650°C (1202°F)
± (2% of reading + 2°C) for Tmeas < 650°C (1202°F)
Repeatability3
LT, MT, G5, 3M, G7 ± 0.5% of reading or ± 0.5°C, whichever is greater 2ML, 2MH ± (0.1% of reading + 1°C) 1ML ± (0.1% of reading + 1°C) for Tmeas > 450°C (842°F)
± (1% of reading + 1°C) for Tmeas < 450°C (842°F)
1MH ± (0.1% of reading + 1°C) for Tmeas > 650°C (1202°F)
± (1% of reading + 1°C) for Tmeas < 650°C (1202°F)
Temperature Resolution (mA output)
2MH, 1MH 0.2 K all other models 0.1 K
Noise Equivalent Temperature (NETD)
LT 0.1 K at Tobj = 23°C (73°F), Tamb = 23°C (74°F) 3M 0.1 K at Tobj = 200°C (392°F), Tamb = 25°C (77°F) MT, G5, G7 0.5 K at Tobj = 10% of full measurement range, Tamb = 25°C (77°F) 2M, 1M 0.5 K at Tobj = 10% of full measurement range, Tamb = 25°C (77°F)
Response time = Instrument Response time
Emissivity 0.100 to 1.150, in 0.001 increments
Signal Processing Peak hold, valley hold, averaging, advanced peak hold,
advanced valley hold, ambient background temperature compensation
2
at 23°C ±5°C (73°F ±9°F), emissivity = 1.0, and time response 1.0 s
3
at 23°C ±5°C (73°F ±9°F)
Page 16
Technical Data
16 Rev. D7 Jul 2017 Marathon MM

3.2 Optical Specifications

In all cases, make sure the target completely fills the measurement spot, see section 6.1.1 Distance to
Object, page 26.
The actual spot size for any distance, when the unit is at focus distance, can be calculated by using the
following formula. Divide the distance D by your model’s D:S number. For example, for a unit with D:S
= 300:1, if the sensor is 2200 mm (86 in.) from the target, divide 2200 by 300 (86 by 300), which gives you a target spot size of approximately 7.3 mm (0.29 in.).
Figure 1: Spot Size Chart
All target spot sizes indicated in the optical diagrams are based on 90% energy.

3.2.1 Variable Focus

Model
Focus
Focus Range
Optical
Resolution D:S *
Smallest Spot Size
LT, MT, G5, 3M
VF1
200 mm (7.9 in.) to 2200 mm (86.6 in.)
70:1
2.9 mm @ 200 mm (0.11 in. @ 7.9 in.)
G7
VF1
200 mm (7.9 in.) to 2200 mm (86.6 in.)
100:1
2 mm @ 200 mm
(0.08 in. @ 7.9 in.)
1ML, 2ML
VF1
300 mm (11.8 in.) to 2200 mm (86.6 in.)
160:1
1.9 mm @ 300 mm
(0.07 in. @ 11.8 in.)
1MH, 2MH
VF1
300 mm (11.8 in.) to 2200 mm (86.6 in.)
300:1
1 mm @ 300 mm
(0.04 in. @ 11.8 in.)
* Optical Resolution is achieved for each focal point in the focus range
Mean Time Between Failures (MTBF) 10,000 operations (at Tamb = 23°C/74°F)
D = Distance S = Spot size
Page 17
Technical Data
Marathon MM Rev. D7 Jul 2017 17

3.2.2 Fixed Focus

Optical Resolution D:S
LT, MT, G5, 3M 70:1 G7 100:1
Available Optics
LT, MT, G5, 3M, G7 SF1, SF2, SF3 LT, 3M CF1
The focus distance is measured from the front end of the sensor. For units with Air/Water Cooled Housing you have to subtract 34.5 mm (1.358 in) from the focus distance. This is very important especially for sensors with close focus optic!
Page 18
Technical Data
18 Rev. D7 Jul 2017 Marathon MM
SF … Standard Focus, CF … Close Focus * specified D:S ratio at focus point only
Table 1: Optical Diagrams for D:S = 70:1
Page 19
Technical Data
Marathon MM Rev. D7 Jul 2017 19
SF … Standard Focus, CF … Close Focus * D:S = 100:1 at focus point
Table 3: Optical Diagrams for D:S = 100:1
Page 20
Technical Data
20 Rev. D7 Jul 2017 Marathon MM
SF … Standard Focus, CF … Close Focus * D:S = 160:1 at focus point
Table 2: Optical Diagrams for D:S = 160:1
Page 21
Technical Data
Marathon MM Rev. D7 Jul 2017 21
SF … Standard Focus, CF … Close Focus * D:S = 300:1 at focus point
Table 3: Optical Diagrams for D:S = 300:1
Page 22
Technical Data
22 Rev. D7 Jul 2017 Marathon MM

3.3 Electrical Specifications

Power Supply 24 VDC ± 20%, min. 500 mA
Outputs
Analog 0 - 20 mA, 4 - 20 mA (active)
14 bit resolution max. current loop impedance: 500 Ω
RS485 Interface networkable to 32 sensors
Baud rate: 300, 1200, 2400, 9600, 19200, 38400 (default), 57600, 115200 (max. 38400 Baud in 2-wire mode) Data format: 8 bit, no parity, 1 stop bit, 4-wire mode (full-duplex) or 2-wire mode (half duplex), selectable via control panel or software
Relay Contacts max. 48 V, 300 mA, response time < 2 ms,
(software programmable)
Display 5 digit backlit LCD display
External Input
Input Voltage 0 to 5 VDC
functions: trigger, laser ON/OFF switching, ambient background temperature compensation, or emissivity setting, see section 7.4.3 External Input page 44.

3.4 Environmental Specifications

Environmental rating NEMA-4 / IEC 529, IP 65 (also with video option)
Relative Humidity 10% to 95% non-condensing
Storage Temperature -20 to 70°C (-4 to 158°F) Ambient Temperature 5 to 65°C (41 to 149°F) without cooling
with video 5 to 50°C (41 to 122°F) without cooling with air cooling 10 to 120°C (50 to 250°F) with water cooling 10 to 175°C (50 to 350°F) with ThermoJacket 10 to 315°C (50 to 600°F) water cooled
Warm up Period 20 min.
Vibration MIL-STD-810D (IEC 68-2-6) 2 G, 10 - 150 Hz, 3 axis
Mechanical Shock MIL-STD-810D (IEC 68-2-27) 5 G, 11 ms duration, 3 axis
Weight 0.7 kg (1.54 lb)
Page 23
Technical Data
Marathon MM Rev. D7 Jul 2017 23

3.5 Dimensions

Figure 2: Dimensions of Sensor
For the dimensional drawing of the Fixed Mounting Bracket, see section 9.2 Fixed Mounting Bracket, page 57.

3.6 Scope of Delivery

The scope of delivery includes the following:
Sensor with through-the-lens sighting
Operating Instructions
DataTemp Multidrop Software
Mounting nut made from stainless steel (XXXMMACMN)
Fixed mounting bracket made from stainless steel (XXXMMACFB)
Page 24
Basics
24 Rev. D7 Jul 2017 Marathon MM

4 Basics

4.1 Measurement of Infrared Temperature

Everything emits an amount of infrared radiation according to its surface temperature. The intensity of the infrared radiation changes according to the temperature of the object. Depending on the material and surface properties, the emitted radiation lies in a wavelength spectrum of approximately 1 to 20
µm. The intensity of the infrared radiation (”heat radiation”) is dependent on the material. For many
substances this material-dependent constant is known. It is referred to as ”emissivity value”, see section
12.2 Typical Emissivity Values, page 84. Infrared thermometers are optical-electronic sensors. These sensors are able to detect ”radiation of heat”. Infrared thermometers are made up of a lens, a spectral filter, a sensor, and an electronic signal processing unit. The task of the spectral filter is to select the wavelength spectrum of interest. The sensor converts the infrared radiation into an electrical signal. The signal processing electronics analyze the electrical signals and convert it into a temperature measurement. As the intensity of the emitted infrared radiation is dependent on the material, the required emissivity can be selected on the sensor. The biggest advantage of the infrared thermometer is its ability to measure temperature without touching an object. Consequently, surface temperatures of moving or hard to reach objects can easily be measured.

4.2 Emissivity of Target Object

Determine the emissivity of the target object as described in appendix 12.1 Determination of Emissivity, page 84. If emissivity is low, measured results could be falsified by interfering infrared radiation from background objects (such as heating systems, flames, fireclay bricks, etc. close beside or behind the target object). This type of problem can occur when measuring reflecting surfaces and very thin materials such as plastic films and glass. This measuring error can be reduced to a minimum if particular care is taken during installation, and the sensing head is shielded from these reflecting radiation sources.
Page 25
Environment
Marathon MM Rev. D7 Jul 2017 25

5 Environment

Sensor location depends on the application. Before deciding on a location, you need to be aware of the ambient temperature of the location, the atmospheric quality of the location, and the possible electromagnetic interference in that location, according to the sections described above. If you plan to use air purging, you need to have an air connection available. If you are installing the sensor in a ThermoJacket accessory, you should use the appropriate mounting device. Also, wiring and conduit runs must be considered, including computer wiring and connections, if used.

5.1 Ambient Temperature

The sensor is designed for measurements in ambient temperatures between 5°C and 65°C (41 to 149°F). A water or air cooled housing is available as option to extend the operating range to 120°C (250°F) with air cooling and to 175°C (350°F) with water cooling. In ambient conditions up to 315°C (600°F), the ThermoJacket housing should be used. When using the ThermoJacket, it is strongly recommended to use the supplied air purge to avoid condensation on the lens.

5.2 Atmospheric Quality

If the lens gets too dirty, it cannot detect enough infrared energy to measure accurately. It is good practice to always keep the lens clean. The air purge helps keep contaminants from building up on the lens. If you use the air purge accessory, make sure a filtered air supply with clean dry air at the correct air pressure is installed before proceeding with the sensor installation.

5.3 Electrical Interference

To minimize measurement errors due to electrical or electromagnetic interference or “noise” be aware of the following:
Mount the electronics enclosure as far away as possible from potential sources of electrical
interference such as motorized equipment producing large step load changes.
Use shielded wire for all input and output connections. Make sure the shield wire from the electronics to terminal block cable is earth grounded. For additional protection, use conduit for the external connections. Solid conduit is better than
flexible conduit in high noise environments.
Do not run AC power in the same conduit as the sensor signal wiring.
When installing the sensor, check for any high-intensity discharge lamps or heaters that may be in the field of view (either background or reflected on a shiny target). Reflected heat sources can cause a sensor to give erroneous readings.
Page 26
Installation
26 Rev. D7 Jul 2017 Marathon MM

6 Installation

6.1 Mechanical Installation

After all preparations are complete, you can install the sensor. How and where you anchor the sensor depends on the type of surface and the type of bracket you are using. You can mount the sensor through a hole, on a bracket of your own design, or on the available bracket accessory.

6.1.1 Distance to Object

The desired spot size on the target will determine the maximum measurement distance and the necessary focus length of the optical module. To avoid erroneous readings the target spot size must contain the entire field of view of the sensor. Consequently, the sensor must be positioned so the field of view is the same as or smaller than the desired target size. For a list indicating the available focus models and their parameters, see section 3.2 Optical Specifications, page 16.
Figure 3: Proper Sensor Placement

6.1.2 Variable Focus

The optional variable focus allows adjustment of the focus length of the sensor optics. Using sensors with this feature requires that the correct focal distance be set on the sensor. To determine the correct focal distance for the sensor, measure the distance in millimeters from the face of the sensor to the target. Set the focal distance to be equal to the measured distance. It is possible to set the focal distance either on the control panel of the sensor or through the DataTemp Multidrop Software. The factory default focal distance is 600 mm (23.6 in.).
Target greater than spot size
Target equal to spot size
Target smaller than spot size
best
critical
incorrect
Background
Sensor
Page 27
Installation
Marathon MM Rev. D7 Jul 2017 27

6.1.3 Viewing Angles

The sensor head can be placed at any angle from the target up to 30°.
Figure 4: Acceptable Sensor Viewing Angles
Best
90° to target
Good
30° to 90° to target
Bad
0° to 30° to target
Acceptable
Angles
Page 28
Installation
28 Rev. D7 Jul 2017 Marathon MM

6.2 Electrical Installation

The 12-wire connecting cable is used to wire all inputs and outputs of the sensor. The cable comes in two different temperature versions. For more information, see section see section 9.9 Terminal Box, page 62 and section 9.11 High Temp Cable, page 64. The following figure shows how to configure the drain wires of the cables before connecting to the sensor. The bare wire with the clear shrink tubing (cable shield) must be connected to the terminal labeled CLEAR.
Figure 5: Sensor Connecting Cable with Terminal Block
The factory default setting for the Marathon MM is the 4-Wire Mode!
The complete wiring must have only one common earth ground point!
Power supplies based on a IT grid structure could cause an unintended reset of the sensor. To avoid this, consider a sensor mounting isolated from the machinery!
To Sensor
Page 29
Installation
Marathon MM Rev. D7 Jul 2017 29
Figure 6: DIN Connector Pin Layout (pin side)
Pin
Cable Color
Description
A
black
RxA* B white
RxB*
C
gray
TxB** D purple
TxA** E white/drain
Shield
F
yellow
Trigger / External Input
G
orange
Relay COM
H
blue
Relay NO/NC
J
green
+ mA out
K
brown
- mA out (analog ground)
L
black
Digital ground
M
red
+24 VDC
* RxA and RxB are twisted paired ** TxA and TxB are twisted paired
Table 4: DIN Connector Wiring
Incorrect wiring can damage the sensor and void the warranty. Before applying power, make sure all connections are correct and secure!
To prevent possible electrical shock, fire, or personal injury make sure that the sensor is grounded before use.
Page 30
Installation
30 Rev. D7 Jul 2017 Marathon MM

6.3 Computer Interfacing

The distance between the sensor and a computer can be up to 1200 m (4000 ft.) via RS485 interface. This allows ample distance from the harsh environment where the sensing system is mounted to a control room or pulpit where the computer is located. The USB/RS485 Interface Converter allows you to connect your sensor to computers by using a USB interface. With auto configuration the converter is able to automatically configure RS485 signals without external switch setting. The converter is equipped with 3000 VDC of isolation and internal surge-protection to protect the host computer and the converter against high voltage spikes, as well as ground potential difference. When the converter is connected the computer gets one virtual COM port.
Technical Data
Power supply 5 VDC direct from USB port
Speed max. 256 kBit/s
RS485 4 wire (full duplex) and 2 wire (half duplex)
Terminal screwed accepts 0.05 to 3 mm² (AWG 13 to AWG 30)
USB connector type B (supplied with type A to type B cable)
Ambient Temperature 0 to 60°C (32 to 140°F), 10-90% relative humidity, non-condensing
Storage Temperature -20 to 70°C (-4 to 158°F), 10-90% relative humidity, non-condensing
Dimensions (L x W x H) 151 x 75 x 26 mm (5.9 x 2.9 x 1 in)
Figure 7: USB/RS485 Converter (XXXUSB485)
Using the 2-wire communication reduces wiring cost in comparison to the 4-wire communications. The disadvantage is, that because the data transfer can be only in one direction at the same time, 2-wire communications have a maximum baud rate of 38.4 kBaud. 2-wire communications is available for network installations, in situations where other sensors are only able to communicate via 2 wires(e.g. MI3 sensor).
The factory default setting for the Marathon MM is the 4-Wire Mode!
Page 31
Installation
Marathon MM Rev. D7 Jul 2017 31
Figure 8: Wiring the Sensor’s RS485 Interface with USB/RS485 Converter
in 4-Wire Mode (factory default)
Figure 9: Wiring the Sensor’s RS485 Interface with USB/RS485 Converter
in 2-Wire Mode
Converter
Sensor
Converter
Sensor
Page 32
Installation
32 Rev. D7 Jul 2017 Marathon MM

6.4 Multiple Sensors in a Network

6.4.1 Wiring

For an installation of two or more sensors in a network, each sensor cable is wired to its own terminal block. The RS485 terminals on each terminal block are wired in parallel. The following figures illustrate the wiring of sensors in a 4-wire and 2-wire multidrop installation.
Figure 10: 4-Wire Multidrop Wiring in a Network
Figure 11: 2-Wire Multidrop Wiring in a Network
USB/485
Interface Converter
XXXUSB485
Terminal Block
from Sensor
from additional terminal block of another sensor
USB/485
Interface Converter
XXXUSB485
Terminal Block
from Sensor
from additional terminal block(s) of other sensor(s)
Page 33
Installation
Marathon MM Rev. D7 Jul 2017 33

6.4.2 Addressing

The addressing of a sensor can be done by means of the Control Panel on the back of the sensor or the DataTemp Multidrop Software (Menu <Sensor Setup>) that came with your sensor. An alternative would be to use the specific interface commands of the sensor in conjunction with a standard terminal program (e.g. Windows HyperTerminal), see section 10.12 Command List, page 78. If you are installing two or more sensors in a multi-drop configuration, please be aware of the following:
Each sensor must have a unique address greater zero. Each sensor must be set to the same baud rate.

6.4.3 Configuration Procedure

1. Attach each unit individually to the computer.
2. Start the DataTemp Multidrop Software.
3. In the DataTemp Multidrop Startup Wizard, select the correct COM port and ASCII protocol,
then <Scan All Baud Rates> for a <Single Sensor>. DataTemp Multidrop should find the single MM unit connected to the computer serial port.
4. Once DataTemp Multidrop is running, go to the <Setup> menu and select <Sensor Setup>.
5. In the <Sensor Setup> menu select the <Advanced Setup> tab. This tab contains the
Communications Interface menu. The Interface Menu allows you to set the <Polling Address>, <Baud Rate> and <RS485 Mode>. Each unit needs a unique address, but the same <Baud Rate> and <RS485 Mode> settings.
6. Once all the units are addressed, wire up the units in the either the 2-wire or 4-wire multidrop
manner, keeping all TxA's, TxB's, RxA's and RxB's to be common.
7. Now you can run the DataTemp Multidrop Software and by selecting the baud rate that you
set, the program will quickly identify all of the units attached on the network and you're up and running.
It is also possible to address each unit without the use of the DataTemp Multidrop Software. Once the unit is powered up, use the enter and mode buttons on the back panel operator interface and toggle to the Multidrop Address field, see section 7.2 Operation Modes, page 34. Use up and down buttons to select a unique address for each unit. The units may now be installed in a multidrop network.
Page 34
Operation
34 Rev. D7 Jul 2017 Marathon MM

7 Operation

Once you have the sensor positioned and connected properly, the system is ready for continuous operation. The operation of the sensor can be done by means of the built-in control panel in the sensor’s housing or by means of the software that came with your sensor.

7.1 Control Panel

The sensor is equipped with a control panel in the sensor’s housing, which has setting/controlling buttons and an LCD display. The panel is used primarily for setting up the instrument and is sealed during operation. The buttons and the display are defined in the following sections.
Figure 12: Control Panel
The sensor has a user interface lockout feature that keeps the unit from being accidentally changed from the control panel (locked by default in multidrop mode). This lockout mode denies access to all the adjustable parameters on the control panel. Access to the display modes of the panel while in a locked condition is provided.

7.2 Operation Modes

When you first turn the unit on, the display shows the current temperature. Pushing the keys of the control panel will change the figures on the display as shown in the menu tree below.
Display
LED for
Laser Indication
Increase Value
Decrease Value
Next Mode
Previous Mode
Enter
Through-the-lens sighting
Page 35
Operation
Marathon MM Rev. D7 Jul 2017 35
Figure 13: Operation Modes
Enter
<Ambient Temp.>
E <Emissivity>
◄ ►
T <Transmission>
◄ ►
◄ ►
A <Average>
◄ ►
P <Peak Hold>
V <Valley Hold>
◄ ►
L <Low mA>
◄ ►
M <Multidrop
◄ ►
B <Baud Rate>
◄ ►
◄ ►
◄ ►
0.100 to 1.150
0.100 to 1.000
On/Off/TRG
0.0 to 300.0
0.0 to 300.0
0.0 to 300.0
Low to High Temp
--, 1 to 32
U <Temp. Unit>
◄ ►
0.3, 1.2, 2.4, 9.6,
19.2, 38.4, 57.6,
115.2
°C, °F, K
Saving of the
selected parameter
Enter
Without saving a
parameter
no action
for 10 sec
COM <Com
◄ ►
2-wire/4-wire
CP <Control
◄ ►
Lock
H <High mA>
Low to High Temp
<Object Temp.>
F <Focus>
◄ ►
0.2/0.3 - 2.2
L<Laser> / V
In the event of displaying the object's temperature and pressing the <Enter> button again, the sensor will not run through the complete menu again but directly go to the menu which was last used.
Page 36
Operation
36 Rev. D7 Jul 2017 Marathon MM
Object Temp.: The display shows the current temperature of the measured object.
Ambient Temp.: The display shows the current internal temperature of the sensor.
Emissivity: Changes the emissivity value. The emissivity is a calculated ratio of infrared
energy emitted by an object to the energy emitted by a blackbody at the same temperature (a perfect radiator has an emissivity of 1.00). For information on determining an unknown emissivity and for sample emissivities, see section 12.2 Typical Emissivity Values, page 84.
Transmission: Changes the transmission value when using protective windows. For
example, if a protective window is used with the sensor, set the transmission to the appropriate value.
Focus: Changes the focus length of the sensor optics.
Laser/Video: Switches the laser or the video (if available) on or off. With the setting <TRG>
the laser can also be switched on/off via the external input.
Average: Parameter given in seconds. Once Average is set above 0.0, it automatically
activates. Note that other hold functions (like Peak Hold or Valley Hold) cannot be used concurrently. The default value is 0.0. For further information see section 7.3.1 Averaging, page 37.
Peak Hold: Parameter given in seconds. Once Peak Hold is set above 0.0, it
automatically activates. Note that other hold functions (like Valley Hold or Averaging) cannot be used concurrently. The default value is 0.0. For further information see section 7.3.2 Peak Hold, page 38.
Valley Hold: Parameter given in seconds. Once Valley Hold is set above 0.0, it
automatically activates. Note that other hold functions (like Peak Hold or Averaging) cannot be used concurrently. The default value is 0.0. For further information see section 7.3.4 Valley Hold, page 41.
Low mA: Defines the temperature for the low current output value (0 or 4 mA).
High mA: Defines the temperature for the high current output value (20 mA).
Multidrop Addr.: Defines the address of a sensor in a network. Each sensor in a network must
have a unique address. “--“ means a standalone unit with address 0.
Baud Rate: Defines the baud rate of a sensor. Each sensor in a multidrop network must
be set to the same baud rate.
Temp. Unit: The temperature display can be set to °C , °K or °F. Note that this setting
influences the RS485 output for both object and ambient temperature. The default value is °C.
Com Mode: Selects the desired digital communication mode for the sensor, either 2-Wire
or 4-Wire.
Page 37
Operation
Marathon MM Rev. D7 Jul 2017 37
Control Panel: The control panel can be locked to avoid accidentally change of sensor
operating parameters. Once locked the control panel must be unlocked by using the control panel as follows:
1. Control Panel is locked.
2. Press the <> button to enter Control Panel menu.
3. Press the following buttons consecutively: <▲> <▼>
4. Control Panel is unlocked.
Note that the control panel is locked by default in multidrop mode and can also be unlocked through the DataTemp Multidrop Software or a programming command.

7.3 Signal Processing

Activating and adjusting the parameters for the signal processing is accomplished either through the DataTemp Multidrop Software, or the programming commands, or partially on the control panel.

7.3.1 Averaging

Averaging can be useful when an average temperature over a specific duration is desired, or when a smoothing of fluctuating temperatures is required. The signal is smoothed depending on the defined time basis. In other words, the output signal tracks the detector signal with significant time delay but noise and short peaks are damped. Use a longer average time for more accurate damping behavior. The average time is the amount of time the output signal needs to reach 90% magnitude of an object temperature jump. The following figure illustrates an averaging output signal. This function is set on the control panel, the software or by means of the programming command G.
Figure 14: Averaging
output temperature
object temperature
temperature jump
average time
90% of temperature jump
Time
Temp
Page 38
Operation
38 Rev. D7 Jul 2017 Marathon MM

7.3.2 Peak Hold

With Peak Hold, the respective last peak value is held until the next reset will occur. There are the following possibilities for a reset.
7.3.2.1 Reset
Reset by Time: The peak will be held for a certain hold time. Once the hold time is exceeded
the output signal, tracks and output the actual object temperature and the algorithm will start over again. This function is set on the control panel, the software or by means of the programming command <P>.
Figure 15: Peak Hold reset by Time
Reset by Trigger: A logical low signal for the trigger will reset the peak hold function. As long
as the input is kept at logical low level the actual object temperatures will be transferred toward the output. At the next logical high level, the hold function will be restarted. To activate the reset by trigger function, the Peak Hold must be set to 300.0 either through the control panel, or DataTemp Multidrop Software, or the programming commands <P>. For wiring the external trigger, see section 7.4.3.1 Trigger, page 44.
Figure 16: Peak Hold reset by Trigger
output temperature
object temperature
hold time
hold time
Time
Temp
output temperature
object temperature
Time
Temp
Trigger
Page 39
Operation
Marathon MM Rev. D7 Jul 2017 39
Reset by burst string (Burst Peak Hold): In burst mode, the peak will be held until a new burst
string is being sent via the digital interface, so the effective peak hold time is defined as the time difference between two sent burst strings. In the poll mode, there is a dedicated peak hold time set to <BS> + 10 ms whereby <BS> is the parameter for the burst speed. This function Burst Peak Hold is activated by means of the programming command <BP=1> and resets the commands <F>, <G>, and <P>.
Figure 17: Peak Hold reset by sent Burst String
Peak Hold time equal to burst speed <BS>
7.3.2.2 Signal Slope
Here are the following options to define the lapse for the signal slope in case of a reset.
Signal slope defined by perpendicular drop (default)
Figure 18: Perpendicular Signal Slope
Signal slope defined by a linear decay: the decay is given in the Kelvin/second. This parameter
is set by means of the programming command <XE>.
output temperature
object temperature Temp
Time
Output temperatures
object temperature
<BS>
Time
Temp
<BS>
<BS>
<BS>
<BS>
Page 40
Operation
40 Rev. D7 Jul 2017 Marathon MM
Figure 19: Signal Slope defined by Decay
Signal slope defined by an average time. The average time is the amount of time the output
signal needs to reach 90% magnitude compared to a perpendicular drop. This parameter is set by means of the programming command <AA>.
Figure 20: Averaging the Signal Slope

7.3.3 Advanced Peak Hold

This function searches the sensor signal for a local peak and writes this value to the output until a new local peak is found. Before the algorithm restarts searching for a local peak, the object temperature has to drop below a predefined threshold. If the object temperature raises above the held value which has been written to the output so far, the output signal follows the object temperature again. If the algorithm detects a local peak while the object temperature is currently below the predefined threshold the output signal jumps to the new maximum temperature of this local peak. Once the actual temperature has passed a peak above a certain magnitude, a new local peak is found. This magnitude is called hysteresis. The threshold is set by means of the programming command <C>, for hysteresis use the command <XY>.
output temperature
object temperature Temp
Time
output temperature
object temperature Temp
Time
Page 41
Operation
Marathon MM Rev. D7 Jul 2017 41
Figure 21: Advanced Peak Hold
For the advanced peak hold function, there are the same settings for reset and signal slope available like for the peak hold function, see sections 7.3.2.1 Reset, page 38 and 7.3.2.2 Signal Slope, page 39.

7.3.4 Valley Hold

This function works similar to the peak hold function, except it will search the signal for a minimum.

7.3.5 Advanced Valley Hold

This function works similar to the advanced peak hold function, except it will search the signal for a local minimum.
output temperature
object temperature
hysteresis
threshold
Temp
Time
Page 42
Operation
42 Rev. D7 Jul 2017 Marathon MM

7.4 Inputs and Outputs

7.4.1 Milliamp Output

The milliamp output is an analog output you can connect directly to a recording device (e.g., chart recorder), PLC, or controller. The mA output can be forced to a specific value through the DataTemp Multidrop software or a programming command according to section 10.9.1 Current Output, page 75. This feature is useful for testing or calibrating connected equipment.
Figure 22: Wiring the Terminal

7.4.2 Relay Output

The relay output is used as an alarm for failsafe conditions, see section 11.2 Fail-Safe Operation, page 82, or as a threshold relay. Relay output relates to the currently displayed temperature on the control panel display. The relay output can be used to indicate an alarm state or to control external actions. The relay contacts can be set either to N.C. (normally closed: relay contacts are closed while in the home position) or N.O. (normally open: relay contacts are open while in the home position). The relay can also be forced on or off for testing connected equipment through the DataTemp Multidrop software or a programming command, see section 10.9.2 Relay Output, page 75.
7.4.2.1 Thresholds
The relay output has two user selectable thresholds. The two thresholds are set by default to the bottom temperature range. Activating and adjusting the threshold value is accomplished through the DataTemp Multidrop Software. Once one or both thresholds are activated the relay changes state as the current measured temperature passes the threshold temperature.
Shield
to Sensor
Page 43
Operation
Marathon MM Rev. D7 Jul 2017 43
Figure 23: Relay Output Example
7.4.2.2 Deadband
Deadband is a zone of flexibility around the threshold. The alarm does not go abnormal until the temperature exceeds the threshold value by the number of set deadband degrees. Thereafter, it does not go normal until the temperature is below the threshold by the number of set deadband degrees. The deadband is factory preset to ±2 K of threshold value. Adjusting to other values is accomplished through DataTemp Multidrop Software. For information on the sensor’s communication protocols, see section 10 Programming Guide, page 67.
High Threshold
Deadband
Low Threshold
Deadband
Relay Output
Object Temperature
Time
Object Temperature
Page 44
Operation
44 Rev. D7 Jul 2017 Marathon MM

7.4.3 External Input

The external input can be used to provide the following functions:
digital input for triggering digital input for On/Off switching of the laser analog input for compensating the ambient background temperature analog input for setting the emissivity
Only one input function can be active at a given time. Before changing to another input function (e.g. remote emissivity setting) the currently set function (e.g. ambient background compensation) needs to be switched off explicitly!
See the DataTemp Multidrop Software help for set up instructions, or refer to the required parameter commands in section 10.9.3 External Input, page 76.
Figure 24: Digital (left) and Analog (right) Using of External Input
7.4.3.1 Trigger
The trigger is activated by shorting the external input to digital ground (pin GROUND on the terminal block) for a minimum duration pulse of 10 ms. That can be done with an external switch, relay, transistor, or TTL gate.
Sensor
Sensor
TRIGGER
Analog ground
(pin –mA OUT)
TRIGGER
Digital ground
(pin GROUND)
Page 45
Operation
Marathon MM Rev. D7 Jul 2017 45
7.4.3.2 Ambient Background Temperature Compensation
The sensor is capable of improving the accuracy of target temperature measurements by taking into account the ambient or background temperature. This feature is useful when the target emissivity is below 1.0 and the background temperature is not significantly lower than the target temperature. For instance, the higher temperature of a furnace wall could lead to too-high temperatures being measured especially for lower emissivity targets. A built in ambient background temperature compensation utility compensates for the impact of the reflected radiation in accordance to the reflective behavior of the target. Due to the surface structure of the target, some amount of ambient radiation will be reflected and therefore added to the thermal radiation that is collected by the sensor. The ambient background temperature compensation compensates the final result by subtracting the amount of ambient radiation measured from the sum of thermal radiation the sensor is exposed to.
The ambient background temperature compensation should always be activated in case of lower emissivity targets in conjunction with targets cooler than the surrounding environment or heat sources near to the target!
Three possibilities for ambient background temperature compensation are available:
The internal sensor temperature is utilized for compensation assuming that the ambient
background temperature is more or less represented by the internal sensor temperature. This is the default setting.
If the background ambient temperature is known and constant, the user may input the known
ambient temperature as a constant temperature value.
Ambient background temperature compensation from a second temperature sensor (infrared
or contact sensor) ensures extremely accurate results. An analog voltage signal at the external input (0 to 5 VDC) is utilized for real time compensation. The voltage input signal is wired to the trigger input terminal of the Marathon terminal block. If an infrared temperature sensor is used for background compensation, both sensors must be set on the same temperature range.
Page 46
Operation
46 Rev. D7 Jul 2017 Marathon MM
All ambient background temperature compensation functions are enabled through the DataTemp Multidrop software, see the software help for set up instructions, or refer to the required command protocol in section 10.8.6 Ambient Background Temperature Compensation, page 74.
Figure 25: Ambient Background Temperature Compensation with Second Infrared Sensor
Sensor 2
targeted
to ambient
Sensor 1
targeted
to object
Thermal radiation of ambient
Thermal radiation of target
0 – 5 VDC analog signal at external
input for ambient compensation
Furnace wall
Target object
Page 47
Operation
Marathon MM Rev. D7 Jul 2017 47
7.4.3.3 Emissivity Setting
The external input (trigger input) can be configured to accept an analog voltage signal (0 to 5 VDC) to provide real time emissivity setting. This function is enabled through the DataTemp Multidrop Software, see the software help for set up instructions, or refer to the required command protocol in section 10.8.5 Emissivity Setting, page 74. The following table shows the relationship between input voltage and emissivity.
U in V
0.00
0.24
0.48
0.71
0.95
1.19
1.43
1.67
1.90
2.14
2.38
2.62
2.86
3.10
3.33
3.57
3.81
4.05
4.29
4.52
4.76
5.00
Emissivity
0.10
0.15
0.20
0.25
0.30
0.35
0.40
0.45
0.50
0.55
0.60
0.65
0.70
0.75
0.80
0.85
0.90
0.95
1.00
1.05
1.10
1.15
Table 5: Ratio between Analog Input Voltage and Emissivity
Figure 26: Adjustment of Emissivity at External Input (Example)
To the external input (trigger input) of the sensor
position “product 1”
position “product 2”
3.1 V (ε=0.75)
1.9 V (ε=0.5)
Page 48
Operation
48 Rev. D7 Jul 2017 Marathon MM

7.5 Factory Defaults

The following table provides the values in case of a factory default. The factory default can only be initiated by using the DataTemp Multidrop Software. The Multidrop address and the baud rate will not change from the last value when factory default is done.
Parameter
Factory defaults
Display mode
°C, TEMP- Display
Emissivity
0.95
Transmission
1.00
Focus
600 mm (23.6 in.)
Laser
Off
Average
000.0 (off)
Peak Hold
000.0 (off)
Valley Hold
000.0 (off)
Low mA (4 mA)
Minimum temperature of range
High mA (20 mA)
Maximum temperature of range
Multidrop Address
not changed (0 with delivery)
Baud Rate
not changed (38400 with delivery)
Temperature Unit
°C
Relay alarm output
controlled by unit
Current Output
4 – 20 mA
Control Panel
unlocked
Serial Communication
4-wire
RS485 Transfer Mode
Poll mode
Output String (RS485)
UTEIEC = temperature unit, target temperature, emissivity, internal temperature, error code
Figure 27: Factory Defaults
Page 49
Options
Marathon MM Rev. D7 Jul 2017 49

8 Options

Options are items that are factory installed and must be specified at time of order. The following are available:
Laser Sighting (…L) or Video Sighting (…V)
Air/Water Cooled Housing including air purge
Manufacturer's Calibration Certificate based on certified probes traced on national standards, e.g.
DAkkS (XXXMMCERT)
Variable Focus (…VF1), see section 3.2.1 Variable Focus, page 16

8.1 Laser Sighting

The laser sighting allows fast and precise aiming at small, rapidly moving targets, or targets passing at irregular intervals. The laser is specially aligned with the sensor’s lens to provide accurate, non-parallax pinpointing of targets. The laser comes as a small, bright red spot indicating the center of the area being measured.
Figure 28: Spot Size Indication
For activating the laser sighting see Figure 13: Operation Modes on page 35.
The laser is a Class II, AlGaInP type laser with an output power less than 1 mW, and an output wavelength of 650 nm. The laser complies with FDA Radiation Performance Standards, 21CFR, subchapter J, and meets IEC 825, Class 2 specifications.
To preserve laser longevity, the laser automatically turns off after approximately 10 minutes of constant use!
WARNING! Avoid exposure to laser light! Eye damage can result. Use extreme caution when operating! Never point at another person!
Laser dot
Spot marker
Measured spot at the inner of the circle
Page 50
Options
50 Rev. D7 Jul 2017 Marathon MM

8.2 Video Sighting

The Marathon MM unit has optional video sighting. Video sighting is intended as a convenient way to verify correct sighting of the Marathon MM unit. Video sighting also allows for either video or frame capture of the process, enhancing process documentation.
Video Specifications: Pixels: 510 x 492 Field of View 8° Focus equal the focal distance of the IR channel
(for units with fixed and variable focus) Composite Video Output Format: NTSC (analog) Signal – Noise Ratio: 40 dB Fixed Noise: 0.03% Vpp Ambient Temperature Range: 5 to 50°C (41 to 122°F) Minimum Required Illumination: 5 Lux Impedance: 75  Cable Connection Type: BNC Maximum Analog Video Cable Run: 100 m (328 ft.)
The video sighting option is integral to the sensor electronics, so no additional installation is required. The unit utilizes an industry-standard BNC connector for the video output.
Figure 29: Wiring the Video Output
Analog Display Monitors
The analog NTSC video output can be feed directly to any monitor that accepts this video format. If NTSC video monitors are not locally available, there are commercially available devices that convert NTSC format into PAL or SECAM formats.
Manufacturer
supplied cable
(2 m / 6.6 ft.)
Female BNC
Connector
Male BNC
Connector
Male BNC Connector
Connection to Video devices, e.g.:
Analog Display Monitors Analog to Digital Video converters
Analog Format converters
Page 51
Options
Marathon MM Rev. D7 Jul 2017 51
Digital Video
In order to utilize the frame capture functionality of the MultiDrop Software, the analog video signal must be converted into a digital signal imported via a USB port to the MultiDrop Software. The USB port on the PC must fulfill at least the USB 2.0 specification! An analog to digital video converter must be purchased locally. As a suitable converter is the TERRATEC G3.
Consider the operating instructions for the selected converter!
In case of installation problems it is recommended to deactivate possible other video devices via the control panel of the PC temporally!
Once converted and imported to a PC, the Video Icon in the MultiDrop Software’s toolbar automatically detects and displays the video stream.
Figure 30: Video Icon in the MultiDrop Software
The video image window can be formatted in the <Setup Video View> window. This window can be opened by right clicking on the default video image window.
Figure 31: Formatting the Video Image
The <Auto Save> tab in the <Setup Video View> window is used to define the parameters for video frame capture and file path.
Select the video device
You can only see the video device for a PC
connected converter!
The list specifies multiple video devices
alphabetically.
The DataTemp MultiDrop software does not save the selection of a video device. A restart of the software requires, if necessary, the re-selection of
the desired video device!
Page 52
Options
52 Rev. D7 Jul 2017 Marathon MM
Figure 32: Setting the <Auto Save> function
Page 53
Options
Marathon MM Rev. D7 Jul 2017 53

8.3 Air/Water Cooled Housing

The Air/Water Cooled Housing allows the sensor to be used in ambient temperatures up to 120°C (250°F) with air cooling, and 175°C (350°F) with water cooling. The cooling media should be connected using 1/8” NPT stainless steel fittings requiring 6 mm (0.24 in) inner diameter and 8 mm (0.31 in) outer diameter for the tube. Air flow should be 1.4 to 2.5 l/sec at 25°C (77°F). Water flow should be approximately 1.0 to 2.0 l/min (water temperature between 10 and 27°C / 50 to 80.6°F). The maximal pressure limit is 5 bar (73 PSI). Chilled water below 10°C (50°F) is not recommended, see section 8.3.1 Avoidance of Condensation, page 54.
The Air/Water Cooled Housing is delivered with appropriate fittings.
Figure 33: Air/Water Cooled Housing
For ambient temperatures exceeding 175°C (350°F), the ThermoJacket can be used. This accessory allows operation at ambient temperatures up to 315°C (600°F)!
Page 54
Options
54 Rev. D7 Jul 2017 Marathon MM

8.3.1 Avoidance of Condensation

If environmental conditions makes water cooling necessary, it is strictly recommended to check whether condensation will be a real problem or not. Water cooling also causes a cooling of the air in the inner part of the sensor, thereby decreasing the capability of the air to hold water. The relative humidity increases and can reach 100% very quickly. In case of a further cooling, the surplus water vapor will condense out as water. The water will condense on the lenses and the electronics resulting in possible damage to the sensor. Condensation can even happen on an IP65 sealed housing.
There is no warranty repair possible in case of condensation within the housing!
To avoid condensation, the temperature of the cooling media and the flow rate must be selected to ensure a minimum device temperature. The minimum sensor temperature depends on the ambient temperature and the relative humidity. Please consider the following table.
Relative Humidity [%]
Ambient Temperature [°C/°F]
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
5/
41
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
5/
41
10/
50
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
5/
41
5/
41
5/
41
5/
41
5/
41
10/
50
15/
59
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
5/
41
5/
41
5/
41
5/
41
10/
50
10/
50
10/
50
10/
50
10/
50
15/
59
20/
68
0/
32
0/
32
0/
32
0/
32
0/
32
0/
32
5/
41
5/
41
5/
41
10/
50
10/
50
10/
50
10/
50
15/
59
15/
59
15/
59
15/
59
15/
59
20/
68
25/
77
0/
32
0/
32
0/
32
0/
32
5/
41
5/
41
10/
50
10/
50
10/
50
10/
50
15/
59
15/
59
15/
59
20/
68
20/
68
20/
68
20/
68
20/
68
25/
77
30/
86
0/
32
0/
32
0/
32
5/
41
5/
41
10/
50
10/
50
15/
59
15/
59
15/
59
20/
68
20/
68
20/
68
20/
68
25/
77
25/
77
25/
77
25/
77
30/
86
35/
95
0/
32
0/
32
5/
41
10/
50
10/
50
15/
59
15/
59
20/
68
20/
68
20/
68
25/
77
25/
77
25/
77
25/
77
30/
86
30/
86
30/
86
30/
86
35/
95
40/
104
0/
32
5/
41
10/
50
10/
50
15/
59
20/
68
20/
68
20/
68
25/
77
25/
77
25/
77
30/
86
30/
86
30/
86
35/
95
35/
95
35/
95
35/
95
40/
104
45/
113
0/
32
10/
50
15/
59
15/
59
20/
68
25/
77
25/
77
25/
77
30/
86
30/
86
35/
95
35/
95
35/
95
35/
95
40/
104
40/
104
40/
104
40/
104
45/
113
50/
122
5/
41
10/
50
15/
59
20/
68
25/
77
25/
77
30/
86
30/
86
35/
95
35/
95
35/
95
40/
104
40/
104
40/
104
45/
113
45/
113
45/
113
45/
113
50/
122
60/
140
15/
59
20/
68
25/
77
30/
86
30/
86
35/
95
40/
104
40/
104
40/
104
45/
113
45/
113
50/
122
50/
122
50/
122
50/
122
50/
122
50/
122
50/
122
60/
140
70/
158
20/
68
25/
77
35/
95
35/
95
40/
104
45/
113
45/
113
50/
122
50/
122
50/
122
50/
122
50/
122
60/
140
60/
140
60/
140
60/
140
60/
140
60/
140
80/
176
25/
77
35/
95
40/
104
45/
113
50/
122
50/
122
50/
122
60/
140
60/
140
60/
140
60/
140
60/
140
90/
194
35/
95
40/
104
50/
122
50/
122
50/
122
60/
140
60/
140
60/
140
100/ 212
40/
104
50/
122
50/
122
60/
140
60/
140
Tab. 6: Minimum device temperatures [°C/°F]
Temperatures higher than 60°C (140°F) are not recommended due to the temperature limitation of the sensor.
Example:
Ambient temperature = 50 °C Relative humidity = 40 % Minimum device temperature = 30 °C
The use of lower temperatures is at your own risk!
Page 55
Accessories
Marathon MM Rev. D7 Jul 2017 55

9 Accessories

9.1 Overview

Accessories include items that may be ordered at any time and added on-site.
Fixed Mounting Bracket (XXXMMACFB) Adjustable Mounting Bracket (XXXMMACAB) Air Purge Collar (XXXMMACAP...) Sight Tube (XXXTST…) Pipe Thread Adapter (XXXMMACPA) Right Angle Mirror (XXXMMACRA...) USB/RS485 Converter (XXXUSB485) see section 6.3 Computer Interfacing, page 30. Industrial Power Supply (XXXSYSPS) Terminal Box (RAYMAPB) Low Temp Cable with Terminal Block(XXX2CLTCB…) High Temp Cable with Terminal Block(XXX2CCB…) Terminal Block (XXXMATB) Terminal Box with terminal block and power supply (RAYMAPB) Protective Window (XXXMMACTW…) ThermoJacket (RAYTXXTJ4) {reserved} {reserved}
Page 56
Accessories
56 Rev. D7 Jul 2017 Marathon MM
Figure 34: Accessories
Sensor
Adjustable Mounting Bracket
XXXMMACAB
Air Purge Collar XXXMMACAP...
Sight Tube XXXTST…
Fixed Mounting Bracket
XXXMMACFB
Pipe Adapter
XXXMMACPA
Pipe Adapter
XXXMMACPA
Protection Window XXXMMACTW…
Mounting Nut XXXMMACMN
Adjustable Pipe Adapter
XXXTXXAPA
ThermoJacket
RAYTXXTJ4
Right Angle Mirror XXXMMACRA...
Page 57
Accessories
Marathon MM Rev. D7 Jul 2017 57

9.2 Fixed Mounting Bracket

Figure 3: Fixed Mounting Bracket in Stainless Steel (XXXMMACFB)

9.3 Adjustable Mounting Bracket

Figure 35: Adjustable Mounting Bracket in Stainless Steel (XXXMMACAB)
Page 58
Accessories
58 Rev. D7 Jul 2017 Marathon MM

9.4 Air Purge Collar

The Air Purge Collar accessory is used to keep dust, moisture, airborne particles, and vapors away from
the optical head’s lens. It can be installed before or after the bracket. It must be screwed in fully. Air flows into the 1/8” NPT fitting and out the front aperture. Air flow should be a maximum of 0.5 - 1.5
liters/sec (1 - 3 cfm). Clean (filtered) or “instrument” air is recommended to avoid contaminants from settling on the lens. Do not use chilled air below 10°C (50°F). The air purge collar can only be used either with the sensor alone or with the Air/Water Cooled Housing. The Air Purge Collar is rotatable in steps of 120°. The Air Purge Collar is available in Aluminum (XXXMMACAP) or stainless steel (XXXMMACAPS).
Figure 36: Air Purge Collar

9.5 Sight Tube

Use a protection tube in temperature measurement environments where reflected energy is a problem.
Stainless Steel Protection Tube up to 800°C (1472°F) (XXXTST12) Ceramic (Alumina) Protection Tube up to 1500°C (2732°F) (XXXTSTC12)
When using a customer supplied protection tube, use caution in specifying the inside diameter and length. Your sensing head determines what diameter/length combinations are possible without impeding the optical field of view!
Figure 37: Sight Tube
Page 59
Accessories
Marathon MM Rev. D7 Jul 2017 59

9.6 Pipe Thread Adapter

The pipe thread adapter must be used to connect the sight tube with the sensor housing. It is made from stainless steel.
Figure 38: Pipe Adapter (XXXMMACPA)
Figure 39: Sensor with Sight Tube (XXXTST…) Pipe Adapter (XXXMMACPA),
and Fixed Mounting Bracket (XXXMMACFB)
Sensor
Page 60
Accessories
60 Rev. D7 Jul 2017 Marathon MM

9.7 Right Angle Mirror

The Right Angle Mirror is used to turn the field of view by 90° against the sensor axis. It is recommended when space limitations or excessive radiation do not allow to directly align the sensor to the target. In dusty or contaminated environments, the air purging should be used to keep the mirror surface clean. The right angle mirror is available in Aluminum (XXXMMACRA) or stainless steel (XXXMMACRAS).
Figure 40: Right Angle Mirror
Page 61
Accessories
Marathon MM Rev. D7 Jul 2017 61

9.8 Industrial Power Supply

The DIN-rail mount industrial power supply delivers isolated dc power and provides short circuit and overload protection.
To prevent electrical shocks, the power supply must be used in protected environments (cabinets)!
Technical data:
Protection class prepared for class II equipment Environmental protection IP20 Operating temperature range -25°C to 55°C (-13°F to 131°F) AC Input 100 – 240 VAC 44/66 Hz DC Output 24 VDC / 1.3 A Cross sections input/output
0.08 to 2.5 mm² (AWG 28 to 12)
4
Figure 41: Industrial Power Supply (XXXSYSPS)
4
Copyright Wago®
Page 62
Accessories
62 Rev. D7 Jul 2017 Marathon MM

9.9 Terminal Box

The terminal box is designed to provide IP66 (NEMA-4) protection to the terminal block, see section
6.2 Electrical Installation, page 28, and a power supply for the sensor. The box should be surface mounted using the flanges and holes provided. It should be mounted in such a manner to allow the free flow of air around the unit. Ambient temperatures should be kept within the range of 0 to 50°C (32 to 120°F).
Technical data for the power supply:
AC input 100 – 240 VAC 50/60 Hz DC output 24 VDC / 1.1 A Operating temperature -20 to 60°C (-4 to 140°F) Humidity 20 to 90%, non-condensing
Figure 42: Terminal Box (RAYMAPB)
Page 63
Accessories
Marathon MM Rev. D7 Jul 2017 63

9.10 Low Temp Cable

The 12-wire low temp cable (XXX2CLTCB…) is used for wiring the sensor with the 24 VDC power supply, all outputs, and the RS485 interface. The cable is PUR (Polyurethane) coated and withstands ambient temperatures form -40 to 105°C (-40°F to 221°F). PUR coated cables are flexible and have good to excellent resistance to against oil, bases, and acids.
Temperature: -40 to 105°C (-40°F to 221°F) Cable material PUR- 11Y (Polyurethane), Halogen free, Silicone free Cable diameter: 7.2 mm (0.283 in) nominal Conductors:
Power supply 2 wires (black/red)
Conductor: 0.2 mm² (AWG 24), 7x32 tinned copper Isolation: PE- 2YI1 Shield: none
RS485 interface 2 twisted pairs (black/white and purple/gray)
Conductor: 0,2 mm² (AWG 24), 7x32 tinned copper Isolation: PE- 2YI1 Shield: CDV-15, 85% covered
Outputs and Ground 6 wires (green/brown/blue/orange/yellow/clear)
Conductor: 0,2 mm² (AWG 24), 7x32 tinned copper Isolation: PE- 2YI1 Shield: none
Further information for wiring the cable can be found in section 6.2 Electrical Installation, page 28.
The low temp cable can be purchased from the manufacturer in the following lengths: 4 m, 8 m, 15 m, 30 m, 60 m (13 ft., 26 ft., 49 ft., 98 ft., 197 ft.)
If you cut the cable to shorten it, notice that both sets of twisted-pair wires have drain wires inside their insulation. These drain wires (and the white wire that is not part of the twisted pair) must be connected to the terminal labeled CLEAR.
If you purchase your own RS485 cable, use wire with the same specifications as those listed above. Maximum RS485 cable length is 1.200 meters (4000 ft).
Page 64
Accessories
64 Rev. D7 Jul 2017 Marathon MM

9.11 High Temp Cable

The 12-wire cable (XXX2CCB…) is used for wiring the sensor with the 24 VDC power supply, all outputs, and the RS485 interface. The cable is Teflon coated and withstands ambient temperatures form
-80 to 200°C (-112°F to 392°F). Teflon coated temperature cables have good to excellent resistance to oxidation, heat, weather, sun, ozone, flame, water, acid, alkalis, and alcohol, but poor resistance to gasoline, kerosene, and degreaser solvents.
Power supply 2 wires (black/red)
Conductor: 0.3 mm² (AWG 22), 7x30 tinned copper Isolation: FEP 0.15 mm wall (0.006 in) Shield: none
RS485 interface 2 twisted pairs (black/white and purple/gray)
Conductor: 0,22 mm² (AWG 24), 7x32 tinned copper Isolation: FEP 0.15 mm wall (0.006 in) Shield: Aluminized Mylar with drain wire
Outputs and Ground 6 wires (green/brown/blue/orange/yellow/clear)
Conductor: 0,22 mm² (AWG 24), 7x32 tinned copper Isolation: FEP 0.15 mm wall (0.006 in) Shield: none
Cable diameter: 7 mm (0.275 in) nominal Temperature: UL-rated at -80 to 200°C (-112°F to 392°F)
Teflon develops poisonous gasses when it comes into contact with flames!
Further information for wiring the cable can be found in section 6.2 Electrical Installation, page 28.
High temp 12-wire cable can be purchased from the manufacturer in the following lengths: 4 m, 8 m, 15 m, 30 m, 60 m (13 ft., 26 ft., 49 ft., 98 ft., 197 ft.)
If you cut the cable to shorten it, notice that both sets of twisted-pair wires have drain wires inside their insulation. These drain wires (and the white wire that is not part of the twisted pair) must be connected to the terminal labeled CLEAR.
If you purchase your own RS485 cable, use wire with the same specifications as those listed above. Maximum RS485 cable length is 1.200 m (4000 ft).
Page 65
Accessories
Marathon MM Rev. D7 Jul 2017 65

9.12 Protective Window

Protective windows can be used to protect the sensor’s optics against dust and other contamination. The following table provides an overview of the available protective windows recommended for the spectral models. All protective windows have a transmission below 100%.
To avoid erroneous readings, ensure that the transmission for the appropriate protective window must be set in the sensor, see section 7.2 Operation Modes, page 34!
Order number
Identification
Model
Material
Transmission
XXXMMACTWL
no dot
Focus SF/VF:
LT, 3M, MT, G5
Zinc Sulphide
0.75 ±0.05
XXXMMACTWLS
no dot
Focus SF/VF:
LT, 3M, MT, G5
Zinc Sulphide
stainless steel
0.75 ±0.05
XXXMMACTWLCF
1 dot
Close Focus CF:
LT, 3M
Zinc Sulphide
0.75 ±0.05
XXXMMACTWLCFS
1 dot
Close Focus CF:
LT, 3M
Zinc Sulphide
stainless steel
0.75 ±0.05
XXXMMACTWLF1
no dot
LT
Plastic Foil
stainless steel
0.75 ±0.05 XXXMMACTWGP
2 dots
1M, 2M
Fused Silica
0.93 0.05
XXXMMACTWGPS
2 dots
1M, 2M
Fused Silica
stainless steel
0.93 0.05
Tab. 7: Protective Windows
For special requirements, please contact your local vendor or representative about our range of special protective windows.
Figure 43: Dimensions for Protective Window(left: SF version, right: CF version)
Page 66
Accessories
66 Rev. D7 Jul 2017 Marathon MM

9.13 ThermoJacket

The ThermoJacket gives you the ability to use the sensor in ambient temperatures up to 315°C (600°F). The ThermoJacket’s rugged cast aluminum housing completely encloses the sensor and provides water and/or air cooling and air purging in one unit. The sensor can be installed or removed from the ThermoJacket housing in its mounted position.
Figure 44: ThermoJacket (RAYTXXTJ4) with Mounting Base (XXXTXXMB)
For more information see the ThermoJacket’s manual.

9.14 {reserved}

9.15 {reserved}

Page 67
Programming Guide
Marathon MM Rev. D7 Jul 2017 67

10 Programming Guide

This section explains the sensor’s communication protocol. A protocol is the set of commands that define all possible communications with the sensor. The commands are described along with their associated ASCII command characters and related message format information. Use them when writing custom programs for your applications or when communicating with your sensor with a terminal program.

10.1 Serial Interface versus Control Panel

Since the sensor includes a control panel, the possibility exists for the user to make manual changes to parameter settings. To resolve conflicts between manual parameter settings and settings by means of the serial interface, the sensor observes the following rules:
Command precedence: the most recent parameter change is valid, whether originating from
control panel or the serial interface.
If a manual parameter change on the control panel is made, the sensor will transmit a
“notification” string to the host (e.g. a PC). Notification strings are suppressed in multidrop
mode.

10.2 Storing of Parameters

All sensor parameters, which are changed via the serial interface, are changed in the sensor internal EEPROM memory. The EEPROM memory will retain all information after powering off the sensor.

10.3 Command Structure

After transmitting one command, it is obligatory to wait for the response from the sensor before sending another. Make sure that a command sent was completely transmitted from the sender before the next command can be sent.
All commands must be entered in upper case (capital) letters!

10.3.1 Requesting a parameter (Poll Mode)

?E<CR> “?” is the command for “request” “E” is the parameter requested <CR> carriage return (0Dh) is closing the request

10.3.2 Setting a parameter (Poll Mode)

E=0.975<CR> “E” is the parameter to be set “=” is the command for “set a parameter” “0.975” is the value for the parameter <CR> carriage return (0Dh) is closing the setting
Page 68
Programming Guide
68 Rev. D7 Jul 2017 Marathon MM

10.3.3 Sensor response

!E0.975<CR><LF> “!” is the parameter for “answer” “E” is the parameter “0.975” is the value for the parameter <CR> <LF> (0Dh 0Ah) is closing the answer
For processing the received commands, the device needs up to 500 ms in the normal case. For certain commands, this time can be even longer, see <Time out> parameter given in section 10.12 Command
List, page 78.

10.3.4 Sensor notification

With a notification the sensor informs the host, that the sensor or the firmware was reset. #XI1<CR><LF> “#” is the parameter for “Notification” “XI1” is the value for the notification (e.g. “XI1firmware reset) <CR> <LF> (0Dh 0Ah) is closing the notification To track the sensor’s reset, XI remains unequal to zero as long as XI=0 is set manually via the digital interface.
With a notification the sensor also informs the host, that a parameter was set on the control panel manually. #XL1<CR><LF> “#” is the parameter for “Notification” “XL1” is the value for the notification (e.g. “XL1laser switched on) <CR> <LF> (0Dh 0Ah) is closing the notification

10.3.5 Error Messages

An asterisk * will be transmitted back to the host in the event of an “illegal” instruction. An illegal instruction is considered to be one of the following:
“*Unknown Command” – any non-used or non-allowed character (e.g. lower case characters) “*Range Error” – an “out-of-range” parameter value “*Syntax Error” – a value entered in an incorrect format “*Function impossible” – unit not in correct modus to execute the requested function

10.4 Transfer Modes

There are two possible transfer modes for the serial interface. Poll Mode: The current value of any individual parameter can be requested by the host. The
sensor responds once with the value at the selected baud rate.
Burst Mode: A pre-defined data string, a so called “burst string”, will be transferred continuously
as long as the burst mode is activated.
V=P “P” starts the poll mode V=B “B” starts the burst mode $=UTIEECCS “$” sets the content of the burst string:
“U” for temperature unit “T” for target temperature “I” for internal temperature of the sensor “E” for emissivity value “EC” for error code “CS” adds a checksum
Page 69
Programming Guide
Marathon MM Rev. D7 Jul 2017 69
?$ gives the burst string parameters while in poll mode, e.g. “UTIE” ?X$ gives the burst string content while in poll mode, e.g. “UC T0150.3 I0027.1 E0.950”
Return from burst mode to poll mode
4-Wire Communication: send “V=P” 2-Wire Communication: send “V=P”. It could be necessary to send the command more than one
times.

10.5 Checksum

The destination for the burst string is often a Windows PC. The Windows operating system is non deterministic and known to let the buffer of the serial communication overflow. To test the burst string a checksum can be added to it by writing <CS> to the burst string definition string ($=...).
The <CS> command can also be used in the polling mode. <CS=1> activates the checksum transfer as a 3 digit integer number.
Example: ?CS //asks for the checksum status !CS0 //answer: no checksum activated CS=1 //enable the checksum calculation !CS1 CS048 //answer: checksum is activated, checksum added, checksum in decimal format
Character Transmission Check
You can check the correctness of transmitted strings by evaluating the block check character (BCC), the checksum. This can be done by taking the exclusive OR (XOR) of the ASCII values of all the characters transmitted from the start character (here: “!”) up to the “S” but excluding the checksum value itself (here: 127), carriage return <CR> and line feed <LF>. For example, the shaded characters are included as follows:
! E 0 . 5
<space>
C S 127
<CR>
<LF>
Please note, the start character can be one of the following: “!”, “?”, “*”, “#”, <multidrop address>, <first character of the burst string>. You can evaluate the received string easily by converting the ASCII values to Binary and progressively adding the Binary values together, minding the following rules for taking the XOR:
0 1 1 0
0 1 0 1
0 0 1 1
The sent checksum from the exemplary line above needs to be identical to the calculated checksum below.
Page 70
Programming Guide
70 Rev. D7 Jul 2017 Marathon MM
ASCII
HEX
Binary
Dec
!
21
0010 0001
E
45
0100 0101
0
30
0011 0000
.
2E
0010 1110
5
35
0011 0101
<space>
20
0010 0000
C
43
0100 0011
S
53
0101 0011
7F
0111 1111
127
calculated checksum via XOR operation
Algorithm for a character wise executed XOR operation given in exemplary Basic programming code:
Bcc = 0 For I = 1 To Len(String)
Bcc = Bcc Xor Asc(Mid$(String, I, 1))
Next I

10.6 Burst Mode

10.6.1 Speed

The communications of the sensor are capable of operating in either burst mode or poll mode. Poll mode requires that a query be issued by a host PC and the unit will respond to the query. In burst mode the unit transmits a pre-configured burst string at a certain rate.
LT, G5, MT, 3M, and G7 units produce a new temperature read (sample time) every 20 ms. 2M and 1M units produce a new temperature read (sample time) every 1 ms.
However, the communications of the sensor in burst mode are limited by the computer baud rate as well as the length and content of the burst string.
The Standard burst mode cycle time is 50 ms. If the burst string contains characters other than
“T”, “I”, or “XT” then the unit will transmit the string every 50 ms. Example burst programmed so that $=TIXTE produces the string <T0150.3 I0027.1 XT00 E0.950> every 50 ms.
The burst mode cycle time of the sensor can be reduced by shortening the length of the burst
string. A Faster burst mode cycle time can be achieved when the burst string is configured to contain no other characters than “T”, “I”, and “XT”. Example 1: $=TIXT results in <T0150.3 I0027.1 XT00> Example 2: $=TI results in <T0150.3 I0027.1>. When the burst string is configured in this way the unit will transmit the string with 20 ms for LT, G5, MT, 3M, and G7 models and 5 ms for 2M and 1M models). Due to this faster rate of data transmission, communications will now be limited by the baud rate of the host. If the host computer does not have a fast enough baud rate then data will be lost and the effective data transmission rate will be slower than the cycle time of the unit.
Because the required baud rate is a function of the number of characters in the burst string, the
Fastest burst mode cycle time can effectively be achieved by reducing the number of
Page 71
Programming Guide
Marathon MM Rev. D7 Jul 2017 71
characters in the burst string. Using this mode automatically sets the burst string to contain
only the values for the “T”, “I” and “XT” parameters. However, the “T”, “I” and “XT”
characters are removed from the burst string.. , Issuing the command $=$ sets the unit into the fastest burst and changes the format of the burst string from <T0150.3 I0027.1 XT00> to <0150.3
0027.1 00>. Because of the burst string is now shorter, the required baud rate will be slower and the effective transmission rate will be faster.

10.6.2 Minimum Baud Rate

The minimum required interface baud rate depends on the number of transferred characters and the read cycle time of the burst mode (standard, fast, or fastest). It may be computed as the following:
cycle
charbit
t
nnb
where:
b = minimum required interface baud rate in [bit/s] nbit = the number of necessary bits for one transferred character including stop bit, is always 9 (8 data bits and 1 stop bit), given in [bit/char] nchar = the number of characters in the burst string including blanks, <CR>, and <LF>, given in [char] tcycle = read cycle time given in [s]
Example:
The burst string in the format “$=TIXT” + “$=$” gives the content with 12 characters, e.g. “1234.5 46 0<CR>”. What is the minimum required interface baud rate in the very fast burst mode with 1 ms read
cycle time?
s
bit
s
bit
schar
charbit
b 115200108000
101
129
3

10.7 Sensor Information

The sensor information is factory installed as read only values.
Command
Description
Answer (example)
?XU
Name of the sensor
“!XUMMLT”
?DS
Additional remark, e.g. for special numbers
“!DSRAY
?XV
Serial number of the sensor
“!XV2C027“
?XR
Firmware revision number
“!XR2.08“
?XH
Maximum temperature of the sensor
“!XH0800.0“
?XB
Minimum temperature of the sensor
“!XB-040.0“
Table 9: Sensor Information
Page 72
Programming Guide
72 Rev. D7 Jul 2017 Marathon MM

10.8 Sensor Setup

10.8.1 General Settings

U=C sets the physical unit for the temperature value (C or F or K). In case of a changed
physical unit all temperature related parameters (e.g., thresholds) are converted
automatically. E=0.950 sets the emissivity according to the setting of “ES” command A=250 sets the ambient background temperature compensation according to the setting of
“AC” command XG=1.000 sets the transmission
?T asks for the target temperature ?I asks for the internal temperature of the sensor ?Q asks for the energy value of the target temperature

10.8.2 Sample Time

The sample time roughly defines the time between the update of the analog output. On all units this is factory defined to be equal to the stated response time of the unit. However, LT, G5, MT, 3M, and G7 sensors have the capability to update the analog output at rates faster than the stated response time. Reducing the sample time will reduce the time that the AD converter is averaging the detector signal. This reduced averaging can result in considerably more noise in the analog output signal, thus this setting should not be changed under most circumstances. The sample time can be set to the following values:
ST=20000 sets sample time to 20 ms, for complete suppressing of 50 Hz noise ST=16666 sets sample time to 16.6 ms, for complete suppressing of 60 Hz noise ST=2000 sets sample time to 2 ms, for fastest possible analog update rate that provides a
smooth step response, but with higher noise.

10.8.3 Temperature Pre-Processing

The samples from the AD converter can be processed before or after temperature calculation. The following filters are available for temperature pre-processing: FF=0 0 0 switches the filter off. This command is for 2M and 1M devices and allows for
transient events as short as 900 µs in duration to be captured. This mode greatly increases the amount of noise on the unit’s output.
FF=1 <threshold> 0 averages the last 16 samples. Over the <threshold> no averaging is applied. The
<threshold> is given in AD counts. Use the command Q for getting AD counts. (default value: <threshold> = 750).
FF=2 0 0 sets the filter to consider the detector response curve for LT, G5, MT, 3M, and G7
devices to allow for transient events as short as 20 ms in duration to be captured. This mode greatly increases the amount of noise on the unit’s output.
Page 73
Programming Guide
Marathon MM Rev. D7 Jul 2017 73
Figure 49: Typical step responses at the output depending on the setting
of FF and ST (example for LT sensor)
FF = 1 ST = 20000
FF = 1 ST = 2000
FF = 2 ST = 20000
FF = 2 ST = 2000
Page 74
Programming Guide
74 Rev. D7 Jul 2017 Marathon MM

10.8.4 Temperature Range

The device has a standard range in which it is calibrated. This range may be extended by the internal reserve. RT=S switches to the standard temperature range RT=E switches to the extended temperature range
All parameters of the technical specification (e.g. accuracy) are valid only inside the standard temperature range – there is no specification for the extended temperature range!

10.8.5 Emissivity Setting

The emissivity setting is selected by means of the “ES” command. ES=I sets emissivity by a constant number ES=E sets the emissivity by an analog voltage on the external input. The ambient
background temperature compensation will be set off (“AC=0”) when it was in
external mode before (“AC=2”). For more information see section 7.4.3 External
Input, page 44.
?E asks for the current emissivity value

10.8.6 Ambient Background Temperature Compensation

In case the ambient background temperature is not represented by the internal sensor temperature, you must set the ambient background temperature values as follows: A=250.0 current ambient background temperature according to the setting of “AC”
command AC=0 no compensation (internal sensor temperature equal to ambient temperature) AC=1 compensation with a constant temperature value set with command “A” AC=2 compensation with an analog voltage signal at the external input,
0 – 5 VDC corresponds to a temperature range set by using the commands AL and
AH.
Resulting temperature is read out by command “A”. The emissivity setting will be
set to internally (“ES=I”). For more information see section 7.4.3 External Input,
page 44.

10.8.7 Temperature Hold Functions

The following table lists the various temperature hold functions along with their resets and timing values. Use this table as a guide for programming your sensor and adjusting the hold times. Please note, the setting of some commands is not possible by using of the control panel, these commands are only available by means of the software. For further information see section 7.3 Signal Processing, page 37.
Hold Function
RESET by
Peak Time
Valley Time
Thresold
Hystersis
Decay Rate
Protocol code
P F C
XY
XE
none
none
000.0
000.0
-*
-*
-*
Peak Hold
timer
000.1-299.9
000.0
000.0
-*
000.0
Peak Hold
trigger
300.0**
000.0
000.0
-*
000.0
Peak Hold
sent burst string
Peak Hold with decay
timer
000.1-299.9
000.0
000.0
-*
0001-3000
Advanced Peak Hold
trigger or threshold
300.0**
000.0
Temp. range
-*
0000
Advanced Peak Hold
timer or threshold
000.1-299.9
000.0
Temp. range
-*
0000
Page 75
Programming Guide
Marathon MM Rev. D7 Jul 2017 75
Advanced Peak Hold with decay
timer or threshold
000.1-299.9
000.0
Temp. range
-*
0001-3000
Valley Hold
timer
000.0
000.1-299.9
000.0
-*
000.0
Valley Hold
trigger
000.0
300.0**
000.0
-*
000.0
Valley Hold with decay
timer
000.0
000.1-299.9
000.0
-*
0001-3000
Advanced Valley Hold
trigger or threshold
000.0
300.0**
Temp. range
-*
0000
Advanced Valley Hold
timer or threshold
000.0
000.1-299.9
Temp. range
-*
0000
Advanced Valley Hold with decay
timer or threshold
000.0
000.1-299.9
Temp. range
-*
0001-3000
* Value does not affect the function type ** Holds infinitely or until triggered
As alternative to the linear decay an averaged decay was implemented. For selection see the following table.
Hold Function
Decay Rate [K/s]
Adv. Average Hold Time [s]
Adv...Hold with Linear Decay
XE = 0001 - 3000
0
Adv...Hold with Averaged Decay
*
AA = 0.1 - 999
Table 10: Decay Functions
Please note the different meanings of the parameters from “XE” and “AA”:
The linear decay rate (XE) is given in K/s The averaged decay (AA) gives the time when the signal will have reached 90% of the final
temperature.

10.9 Sensor Control

10.9.1 Current Output

The current output corresponds to the target temperature value. The output can be set to 0 – 20 mA or 4 – 20 mA. XO=4 sets the current output range to 4 – 20 mA H=500 sets the temperature for the top current output value to 500 (in current scale)
e.g., the top current output value of 20 mA shall represent 500°C
L=0 sets the temperature for the bottom current output value to 0 (in current scale)
e.g., the bottom current output value of 4 mA shall represent 0°C
The minimum temperature span between “H” and “L” command values is 20 K.
For testing purposes the output can be forced to provide a constant current value. O=13.57 output of a constant current at 13.57 mA O=60 switches back to the target temperature controlled output

10.9.2 Relay Output

The relay output (alarm output) can be triggered:
by target temperature by internal sensor temperature manually (command controlled)
The alarm output can be set either to N.C. (normally closed: relay contacts are closed while in the home position) or N.O. (normally open: relay contacts are open while in the home position).
K=2 alarm output triggered by target temperature and internal sensor temperature, N.O.
normally open
Page 76
Programming Guide
76 Rev. D7 Jul 2017 Marathon MM
K=3 alarm output triggered by target temperature and internal sensor temperature, N.C.
normally closed K=4 alarm output triggered by internal sensor temperature, N.O. normally open K=5 alarm output triggered by internal sensor temperature, N.C. normally closed K=6 alarm output triggered by target temperature, N.O. normally open K=7 alarm output triggered by target temperature, N.C. normally closed XS=125.3 sets the alarm threshold to 125.3 in current scale. The alarm threshold is used for the
target temperature only (see command XS). If the internal sensor temperature is
selected as the trigger source (K=4, K=5), use the command <DA> for defining the
alarm threshold.

10.9.3 External Input

The external input is shared by several functions, see section 7.4.3 External Input, page 44. The following commands will change the input type automatically:
P=300, F=300 external input as trigger, resets the (advanced) Peak/Valley Hold function ES=E external input as voltage input (0 V – 5 V) for setting the emissivity AC=2 external input as voltage input (0 V – 5 V) for setting the ambient temperature XL=T external input for On/Off switching of the laser
One of the basic rules for the used protocol here does forbid cross dependencies between commands. So the user has to take care not to use two functions the same time. Using more than one function will not destroy the device electrically but may result in malfunction.
?XT gives the actual trigger state ?TV gives the measured input voltage

10.9.4 Lock Mode

The access to the sensor is possible via serial interface or via direct user input on the control panel. With a command it is possible to lock the mode buttons. This allows access to the sensor only via serial interface.
J=L direct user input via mode buttons on control panel denied
Page 77
Programming Guide
Marathon MM Rev. D7 Jul 2017 77

10.10 RS485 Communication

The serial RS485 communication can be either in 2-wire or 4-wire mode.
HM=2 sets the sensor into 2-wire communication mode. HM=4 sets the sensor into 4-wire communication mode
For setting the baud rate, one of the following two commands must be selected.
D=576 sets the baud rate to 57600, baud rate must be given with 3 numbers (003, 012, 024,
096, 192, 384, 576, 115).
BR=57600 sets the baud rate to 57600, baud rate must be given with up to 6 numbers (300, 1200,
2400, 9600, 19200, 38400, 57600, 115200)
In case of using the sensor in 2-wire mode the baud rate is limited to 38400.

10.11 Multidrop Mode

Up to 32 devices can be connected within an RS485 multidrop network, see section 6.4 Multiple Sensors
in a Network, page 32. To direct a command to one sensor among the 32 possible, it is necessary to
“address” a command. Therefore, a 3-digit number is set prior the command. The 3-digit number is determined between 001 and 032. A unit with the address 000 is a single unit and not in multidrop mode.
XA=024 sets the device to address 24
Changing an address:
(e.g., the address is change from 17 to 24) command answer
“017?E” “017E0.950” // asking one sensor on address 17 “017XA=024” “017XA024” // setting of a new address “024?E” “024E0.950” // asking same sensor now on address 24
If a command is transferred, starting with the 3-digit number 000, all units (with addresses from 001 to
032) connected will get this command - without to send an answer.
command answer
“024?E” “024E0.950” “000E=0.5” will be executed from all sensors, no answer “024?E” “024E0.500” “012?E” “012E0.500”
When a sensor has a multidrop address between 001 and 032, its control panel is automatically locked. It can be unlocked with the command “xxxJ=U”, where xxx is the multidrop address!
Page 78
Programming Guide
78 Rev. D7 Jul 2017 Marathon MM

10.12 Command List

P ... Poll, B ... Burst, S ... Set, N ... Notification
(1) n = number, X = uppercase letter
Description
Char
Format
P
B
S
N
legal values
factory default
LCD Char
Poll parameter
?
?X/?XX
?T
Set parameter
=
X/XX=...
E=0.85
Multidrop addressing
001?E
answer: 001!E0.950
000
Error message
* *Syntax error
Acknowledge message
! !P010
Burst string format
$
UTEI
Ambient radiation correction
A
nnnn.n
0 – max. range
Advanced hold averaging time
AA
nnn.n
0 = no averaging
0.1 … 999.0 s
0
Ambient compensation control
AC
n
0 = no compensation, 1 = by command <A> 2 = external input (0V5V)
0 Low range for ambient control with external input at 0V
AL
nnnn.n
min range – max range
min range
High range for ambient control with external input at 5 V
AH
nnnn.n
min range – max range
max range
Burst Peak Hold for 1M/2M models only
BP
n
0 = off (no Peak Hold) 1 = on Polling: Peak Hold time over <BS> + 10 ms Burst: Peak Hold time over <BS>
0 Baud rate
BR
nnnnnn
9600, 19200, 38400, 57600, 115200 Time out: 2000 ms
38400 with delivery, no change with factory default
Burst speed
BS
integer
number in milliseconds 50ms ... 20s for 1M/2M models down to 5 ms but sample time guaranteed steps in 5 ms
50
Advanced hold threshold
C
nnnn.n
in current scale (°C/°F/°K)
<= min.range
--advanced off
min. range
Internal temp. alarm threshold
DA
nn.n
-10 to 65°C (14 to 149°F)
65°C (149°F)
Device special
DS
XXX
()
e.g. !DSRAY
Set at production
Emissivity internal
E
n.nnn
0.1 - 1.15
0.95
E
Page 79
Programming Guide
Marathon MM Rev. D7 Jul 2017 79
Description
Char
Format
P
B
S
N
legal values
factory default
LCD Char
Error code
EC
nnnn
hex value of ErrCode:
BIT0 Object temp. over range BIT1 Object temp. under range BIT2 Internal temp. over range BIT3 Internal temp. under range BIT4 ADC initialisation error BIT5 EEPROM error user space BIT6 EEPROM error calibration BIT7 Device in init phase BIT8 Focus motor error BIT9 Focus zero position lost BITA Focus motor moving BITB mA output over range BITC mA output under range
Emissivity source
ES
X
I = Emissivity from Internal (by command) E = Emissivity from External analog input (0V5V)
I
Valley hold time
F
Nnn.n
  
0.0 - 299.9s (300 = )
0.0s
V
Focal distance
FC
nnn.n
 
within the focus range, in meter Time out: 1000 ms (Polling) Time out: 2000 ms (Setting)
0.6 F Activate pre-calculation Filter
FF
n 0 0
0 = off 1 = Filter Average 2 = Filter Detector
1
Average time
G
nnn.n
  
0.0 – 999.0 s
0.0 s
A
Top of mA range
H
nnnn.n
  
H
RS485 mode
HM
n
 
2 = 2-wire, 4 = 4-wire
4
COM
Sensor internal temp.
I
nnn.n
in current scale (C/F)
Switch panel lock
J
X
 
L = locked U = unlocked
unlocked
CP
Relay alarm output control
K
n
0 = off 1 = on 2 = Target and Internal norm. open 3 = Target and Internal norm. closed 4 = Internal, normal open 5 = Internal, norm. closed 6 = Target, norm. open 7 = Target, norm. closed
2
Bottom of mA range
L
nnnn.n
  
L
Output current
O
nn.nn
0.00 – 20.00 current in mA 21 = over range; 60 = controlled by unit
60 After reset = 60
Peak hold time
P
nnn.n
  
0.0 - 299.9 s (300 = )
0.0 s
P
AD counts
Q
nnnnnnn
Reset
RS
Reset of the firmware, acknowledged with: !RS[CR][LF] #XI1[CR][LF] Time out: 12000 ms
Temperature range
RT
E = Extended range S = Standard range (= <bottomTemp> <topTemp>)
S
-
Page 80
Programming Guide
80 Rev. D7 Jul 2017 Marathon MM
Description
Char
Format
P
B
S
N
legal values
factory default
LCD Char
Sample time
ST
X
()
For LT, G5, MT, 3M it can be set to the following times (in µs): 2000, 10000, 16666, 20000 or 33333
20000
-
Target temperature
T
nnnn.n
in current scale (C/F)
Thermal shock control
TS
n
Y / N reserved for special models only
N
Voltage at trigger pin
TV
float
in Volt
Temperature unit
U
X
  
C / K / F
C
U
Poll/burst mode
V
X
P=poll B=burst
poll mode
Handle video control
VI
n
 
0=OFF, 1=ON N=not installed
0
UART counter for 1M/2M models only
W
n
0 = off, 1 = on (for Poll Mode) responds with 4 digit hex code (max x7FFF) Reset to 1 with activating the burst mode V=B Reset to 0 with resetting of firmware or device
0
Burst string content
X$
Multidrop address
XA
0nn
 
0 – 32 (0 single unit mode)
0
Device bottom range limit
XB
nnnn.n
()
Deadband
XD
nn
1 – 55°C/K; 1-99°F
2 Decay rate
XE
nnnn
1-3000K/s
0
Restore factory defaults
XF
Time out: 12000 ms
Transmission
XG
n.nnn
0.1 - 1.0
1.0 T Device high range limit
XH
nnnn.n
()
Sensor initialisation
XI
n
  
1 after RESET, 2 if the internal watchdog caused the reset, 0 if XI=0
Laser control
XL
X 
0=OFF, 1=ON, N=not installed, Y=installed, T=external input is used to switch the laser
0
Analog output mode
XO
n
0 = 0...20mA, 4 = 4...20mA
4
Second relay threshold
XP
nnnn.n
XP = bottom range switches the alarm mode off
bottom range Firmware revision
XR
e.g. 1.01
Set in FW
Alarm threshold for target temp.
XS
nnnn.n
XS = bottom range switches the alarm mode off
bottom range Trigger
XT
n
 
0 = inactive, 1 = active
0 Unit identification
XU
()
e.g. !XUMM
Serial number
XV
()
e.g. 98123
Advanced hold hysteresis
XY
nnnn
0 – 3000 K
2
Timer internal for 1M/2M models only
Z  0 – 9999 ms
Page 81
Maintenance
Marathon MM Rev. D7 Jul 2017 81

11 Maintenance

Our sales representatives and customer service are always at your disposal for questions regarding application assistance, calibration, repair, and solutions to specific problems. Please contact your local sales representative if you need assistance. In many cases, problems can be solved over the telephone. If you need to return equipment for servicing, calibration, or repair, please contact our Service Department before shipping. Phone numbers are listed at the beginning of this document.
To prevent possible electrical shock, fire or personal injury follow these guidelines:
Have an approved technician repair the product.

11.1 Troubleshooting Minor Problems

Symptom
Probable Cause
Solution
No output
No power to instrument
Check the power supply
Erroneous temperature
Faulty sensor cable
Verify cable continuity
Erroneous temperature
Field of view obstruction
Remove the obstruction
Erroneous temperature
Window lens
Clean the lens
Erroneous temperature
Wrong emissivity
Correct the setting
Temperature fluctuates
Wrong signal processing
Correct Peak/Valley Hold or Average settings
Table 11: Troubleshooting
Page 82
Maintenance
82 Rev. D7 Jul 2017 Marathon MM

11.2 Fail-Safe Operation

The Fail-Safe system is designed to alert the operator and provide a safe output in case of any system failure. Basically, it is designed to shutdown the process in the event of a set-up error, system error, or a failure in the sensor electronics.
The Fail-Safe circuit should never be relied on exclusively to protect critical heating processes. Other safety devices should also be used to supplement this function!
When an error or failure does occur, the display indicates the possible failure area, and the output circuits automatically adjust to their lowest or highest preset level. The following table shows the values displayed on the control panel and transmitted over the serial interface.
Symptom
Error Code
Priority
0 – 20 mA Output
4 – 20 mA Output
Internal temperature over range
EIHH
1 (high)
21 to 24 mA
21 to 24 mA
Internal temperature under range
EIUU
2
0 mA
3.5 mA
Target temperature under range
EUUU
3
0 mA
3.5 mA
Target temperature over range
EHHH
4 (low)
21 to 24 mA
21 to 24 mA
Table 12: Error Codes and Current Output Values
The relay is controlled by the temperature selected on the display. If any failsafe code appears on the
display, the relay changes to the “abnormal” state. This causes the relay to change state, leaving a
normal numerical temperature output.
If two errors occur simultaneously, the higher priority error is the one that is presented on the digital and analog outputs. For example, if the internal temperature is too high and the target temperature is over range, the unit outputs EIHH on the display and digital output and 21 mA on the analog output.
Page 83
Maintenance
Marathon MM Rev. D7 Jul 2017 83

11.3 Cleaning the Lens

Keep the lens clean at all times. Care should be taken when cleaning the lens. To clean the window, do the following:
1. Lightly blow off loose particles with “canned” air (used for cleaning computer equipment) or a
small squeeze bellows (used for cleaning camera lenses).
2. Gently brush off any remaining particles with a soft camel hair brush or a soft lens tissue
(available from camera supply stores).
3. Clean remaining “dirt” using a cotton swab or soft lens tissue dampened in distilled water. Do
not scratch the surface.
For finger prints or other grease, use any of the following:
Denatured alcohol Ethanol Kodak lens cleaner
Apply one of the above to the lens. Wipe gently with a soft, clean cloth until you see colors on the surface, then allow to air dry. Do not wipe the surface dry, this may scratch the surface. If silicones (used in hand creams) get on the window, gently wipe the surface with Hexane. Allow to air dry.
Do not use any ammonia or any cleaners containing ammonia to clean the lens. This
may result in permanent damage to the lens’ surface!
Page 84
Appendix
84 Rev. D7 Jul 2017 Marathon MM

12 Appendix

12.1 Determination of Emissivity

Emissivity is a measure of an object’s ability to absorb and emit infrared energy. It can have a value between 0 and 1.0. For example a mirror has an emissivity of 0.1, while the so-called “Blackbody“ reaches an emissivity value of 1.0. If a higher than actual emissivity value is set, the output will read low, provided the target temperature is above its ambient temperature. For example, if you have set
0.95 and the actual emissivity is 0.9, the temperature reading will be lower than the true temperature. An object’s emissivity can be determined by one of the following methods:
1. Determine the actual temperature of the material using an RTD (PT100), a thermocouple, or any
other suitable method. Next, measure the object’s temperature and adjust emissivity setting until
the correct temperature value is reached. This is the correct emissivity for the measured material.
2. For relatively low temperatures (up to 260°C / 500°F) place a plastic sticker ( 38 mm / 1.5 in.,
XXXRPMACED) on the object to be measured. This sticker should be large enough to cover the
target spot. Next, measure the sticker’s temperature using an emissivity setting of 0.95. Finally,
measure the temperature of an adjacent area on the object and adjust the emissivity setting until the same temperature is reached. This is the correct emissivity for the measured material.
3. If possible, apply flat black paint to a portion of the surface of the object. The emissivity of the paint
is 0.95. Next, measure the temperature of the painted area using an emissivity setting of 0.95. Finally, measure the temperature of an adjacent area on the object and adjust the emissivity until the same temperature is reached. This is the correct emissivity for the measured material.

12.2 Typical Emissivity Values

The following table provides a brief reference guide for determining emissivity and can be used when one of the above methods is not practical. Emissivity values shown in the table are only approximate, since several parameters may affect the emissivity of a material. These include the following:
1. Temperature
2. Angle of measurement
3. Geometry (plane, concave, convex)
4. Thickness
5. Surface quality (polished, rough, oxidized, sandblasted)
6. Spectral range of measurement
7. Transmission (e.g. thin films plastics)
Page 85
Appendix
Marathon MM Rev. D7 Jul 2017 85
METALS
Emissivity
Material
1 µm
1.6 µm
2.3 µm
3.9 µm
5 µm
7.9 µm
8 – 14 µm
Aluminum
Unoxidized
0.1-0.2
0.02-0.2
0.02-0.2
0.02-0.2
0.02-0.2
0.03-0.15
0.02-0.1
Oxidized
0.4
0.4
0.2-0.4
0.2-0.4
0.2-0.4
0.20-0.55
0.2-0.4
Alloy A3003, Oxidized
0.4
0.4
0.4
0.4 0.3 Roughened
0.2-0.8
0.2-0.6
0.2-0.6
0.1-0.4
0.1-0.4
0.1-0.3
Polished
0.1-0.2
0.02-0.1
0.02-0.1
0.02-0.1
0.02-0.1
0.02-0.1
Brass
Polished
0.1-0.3
0.01-0.05
0.01-0.05
0.01-0.05
0.01-0.05
0.03-0.15
0.01-0.05
Burnished
0.4
0.3
0.3 0.3
Oxidized
0.6
0.6
0.6
0.5
0.5 0.5
Chromium
0.4
0.4
0.05-0.3
0.03-0.3
0.03-0.3
0.10-0.20
0.02-0.2
Oxidized
0.60-0.85
Copper
Polished
0.03
0.03
0.03
0.03
0.03-0.15
0.03
Roughened
0.05-0.2
0.05-0.2
0.05-0.15
0.05-0.15
0.05-0.1
Oxidized
0.2-0.8
0.2-0.9
0.7-0.9
0.5-0.8
0.5-0.8
0.40-0.80
0.4-0.8
Gold
0.3
0.01-0.1
0.01-0.1
0.01-0.1
0.01-0.1
0.02-0.15
0.01-0.1
Haynes
Alloy
0.5-0.9
0.6-0.9
0.6-0.9
0.3-0.8
0.3-0.8
0.3-0.8
Inconel
Oxidized
0.4-0.9
0.6-0.9
0.6-0.9
0.6-0.9
0.6-0.9
0.80-0.90
0.7-0.95
Sandblasted
0.3-0.4
0.3-0.6
0.3-0.6
0.3-0.6
0.3-0.6
0.3-0.6
polished
0.2-0.5
0.25
0.25
0.15
0.15
0.10-0.25
0.15
Iron Oxidized
0.4-0.8
0.5-0.8
0.7-0.9
0.6-0.9
0.6-0.9
0.80-0.95
0.5-0.9
Unoxidized
0.35
0.1-0.3
0.1-0.3
0.05-0.25
0.05-0.25
0.05-0.2
Rusted
0.6-0.9
0.6-0.9
0.5-0.8
0.5-0.8
0.5-0.7
Molten
0.35
0.4-0.6
0.4-0.6
Iron, Cast
Oxidized
0.7-0.9
0.7-0.9
0.7-0.9
0.65-0.95
0.65-0.95
0.10-0.95
0.6-0.95
Unoxidized
0.35
0.3
0.1-0.3
0.25
0.25
0.10-0.15
0.2
Molten
0.35
0.3-0.4
0.3-0.4
0.2-0.3
0.2-0.3
0.2-0.3
Iron, Wrought
Dull
0.9
0.9
0.95
0.9
0.9 0.9
Lead
Polished
0.35
0.05-0.2
0.05-0.2
0.05-0.2
0.05-0.2
0.05-0.1
Rough
0.65
0.6
0.5
0.4
0.4 0.4
Oxidized
0.3-0.7
0.3-0.7
0.2-0.7
0.2-0.7
0.2-0.6
Magnesium
0.3-0.8
0.05-0.3
0.05-0.2
0.03-0.15
0.03-0.15
0.02-0.1
Mercury
0.05-0.15
0.05-0.15
0.05-0.15
0.05-0.15
0.05-0.15
Molybdenum
Oxidized
0.5-0.9
0.4-0.9
0.4-0.9
0.3-0.7
0.3-0.7
0.2-0.6
Unoxidized
0.25-0.35
0.1-0.35
0.1-0.3
0.1-0.15
0.1-0.15
0.10-0.25
0.1
Page 86
Appendix
86 Rev. D7 Jul 2017 Marathon MM
METALS
Emissivity
Material
1 µm
1.6 µm
2.3 µm
3.9 µm
5 µm
7.9 µm
8 – 14 µm
Monel (Ni-Cu)
0.3
0.2-0.6
0.2-0.6
0.1-0.5
0.1-0.5
0.10-0.25
0.1-0.14
Oxidized
0.60-0.85
0.7-0.9
Nickel
Oxidized
0.8-0.9
0.4-0.7
0.4-0.7
0.3-0.6
0.3-0.6
0.80-0.95
0.2-0.5
Electrolytic
0.2-0.4
0.1-0.3
0.1-0.2
0.1-0.15
0.1-0.15
0.05-0.15
Platinum
Black
0.95
0.95
0.9
0.9 0.9
Silver
0.02
0.02
0.02
0.02
0.03-0.15
0.02
Steel
Cold-Rolled
0.8-0.9
0.8-0.9
0.8-0.9
0.8-0.9
0.7-0.9
Ground Sheet
0.6-0.7
0.5-0.7
0.5-0.7
0.4-0.6
Polished Sheet
0.35
0.25
0.2
0.1
0.1
0.10-0.25
0.1
Molten
0.35
0.25-0.4
0.25-0.4
0.1-0.2
0.1-0.2
Oxidized
0.8-0.9
0.8-0.9
0.8-0.9
0.7-0.9
0.7-0.9
0.80-0.95
0.7-0.9
Stainless
0.35
0.2-0.9
0.2-0.9
0.15-0.8
0.15-0.8
0.10-0.25
0.1-0.8
Tin (Unoxidized)
0.25
0.1-0.3
0.1-0.3
0.05
0.05 0.05
Titanium
Polished
0.5-0.75
0.3-0.5
0.2-0.5
0.1-0.3
0.1-0.3
0.05-0.2
Oxidized
0.6-0.8
0.6-0.8
0.5-0.7
0.5-0.7
0.5-0.6
Tungsten
0.1-0.6
0.05-0.5
0.05-0.5
0.03
Polished
0.35-0.4
0.1-0.3
0.1-0.3
0.05-0.25
0.05-0.25
0.05-0.20
0.03-0.1
Zinc
Oxidized
0.6
0.15
0.15
0.1
0.1 0.1
Polished
0.5
0.05
0.05
0.03
0.03
0.15-0.25
0.02
Tab. 13: Typical Emissivity Values
Page 87
Appendix
Marathon MM Rev. D7 Jul 2017 87
NON-METALS
Emissivity
Material
1 µm
1.6 µm
2.3 µm
5 µm
7.9 µm
8 – 14 µm
Asbestos
0.9 0.8
0.9 0.95
Asphalt
0.95
0.95-1.00
0.95
Basalt
0.7 0.7
Carbon
Unoxidized
0.8-0.95
0,8-0,9
0.8-0.9
0.8-0.9
Graphite
0.8-0.9
0.8-0.9
0.7-0.9
0.45-0.70
0.7-0.8
Carborundum
0.95
0.9 0.9
Ceramic
0.4 0.8-0.95
0.8-0.95
0.95
Clay
0.8-0.95
0.85-0.95
0.95
Coke
0.95-1.00
0.95-1.00
0.95-1.00
0.95-1.00
0.95-1.00
0.95-1.00
Concrete
0.65 0.9
0.9 0.95
Cloth
0.95 0.95
Glass
Plate
0.2
0.98
0.98
0.85
“Gob”
0.4-0.9
0.9
Gravel
0.95 0.95
Gypsum
0.4-0.97
0.8-0.95
Ice 0.98
Limestone
0.4-0.98
0.98
Paint (non-al.)
0.90-1.00
0.9-0.95
Paper (any color)
0.95
0.90-1.00
0.95
Plastic, opaque at 500 µm thickness (20 mils)
0.95 0.95
Rubber
0.9
0.95-1.00
0.95
Sand
0.9 0.9
Snow
0.9
Soil 0.9-0.98
Water
0.93
Wood, Natural
0.9-0.95
0.90-1.00
0.9-0.95
Tab. 14: Typical Emissivity Values
To optimize surface temperature measurements, consider the following guidelines:
Determine the object emissivity using the instrument which is also to be used for the
measurements.
Avoid reflections by shielding the object from surrounding temperature sources. For higher temperature objects use instruments with the shortest wavelength possible. For translucent materials such as plastic foils or glass, assure that the background is uniform
and lower in temperature than the object.
Page 88
Notices
88 Rev. D7 Jul 2017 Marathon MM

13 Notices

Loading...