RayTek 3i User Manual

Rev H
6/98
56700-1
RAYNGER
®
3i
SERIES
OPERATOR’S MANUAL
°F
°C
T
E
S
L
C
R
G
O
L
amb
T
LOG
DC
OUT
( )
O
L
( )
I
H
RECALL
MODE
ACTIVATE
LOCK
R
E
S
A
L
SET
WARRANTY
Raytek warrants each instrument it manufactures to be free from defects in material and work­manship under normal use and service for the period of one year from date of purchase. This warranty extends only to the original purchaser. This warranty shall not apply to fuses, batteries, or any product which has been subject to misuse, neglect, accident, or abnormal conditions of operation.
In the event of failure of a product covered by this warranty, Raytek will repair the instrument when it is returned to an authorized Service Facility within one year of the original purchase, provided the warrantor’s examination discloses to its satisfaction that the product was defective. The warrantor may, at its option, replace the product in lieu of repair. With regard to any instru­ment returned within one year of the original purchase, said repairs or replacement will be made without charge. If the failure has been caused by misuse, neglect, accident, or abnormal condi­tions of operation, repairs will be billed at nominal cost. In such cases, an estimate will be sub­mitted before work is started, if requested.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MER­CHANTABILITY, FITNESS, OR ADEQUACY FOR ANY PARTICULAR PURPOSE OR USE. RAYTEK SHALL NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL, OR CONSEQUEN­TIAL DAMAGES, WHETHER IN CONTRACT, TORT, OR OTHERWISE.
WORLDWIDE HEADQUARTERS
Raytek Corporation 1201 Shaffer Road, PO Box 1820 Santa Cruz, CA 95061-1820 USA
Portable Products Division
800 866 5478 Phone 408 458 1110 Fax 408 425 4561
WORLD WIDE WEB www.raytek.com
EUROPEAN HEADQUARTERS
Raytek GmbH Arkonastrasse 45-49 D-13189 Berlin, Germany Phone 49 30 478 0080 FAX 49 30 471 0251
Offices worldwide:
BRAZIL
Raytek do Brasil Sorocaba, Brasil Phone 55 152 276556
CHINA
Raytek China Company Beijing, China Phone 86 10 437 0284
JAPAN
Raytek Japan, Inc. Tokyo, Japan Phone 03 5976 1531
MEXICO
Raytek de Mexico Puebla, Mexico Phone 52 22 30 4380
UNITED KINGDOM
Raytek UK Milton Keynes, UK Phone 44 1 908 630800
® Raytek and Raynger are registered trademarks and 3i is a trademark of Raytek Corporation © 1997 Copyright Raytek Corporation
Raynger 3i Series Operator's Manual Table of Contents
TABLE OF CONTENTS
SECTION PAGE
1.0 INTRODUCTION.............................................................................1-1
1.1 DESCRIPTION.......................................................................................................1-1
1.2 INVENTORY..........................................................................................................1-2
1.3 MODEL IDENTIFICATION.................................................................................1-3
2.0 OPERATION ......................................................................................2-1
2.1 QUICK START........................................................................................................2-2
2.2 PRINCIPLES OF OPERATION............................................................................2-3
2.3 YOUR PORTABLE INFRARED THERMOMETER...........................................2-4
2.4 OPERATION AND CONTROLS.........................................................................2-6
2.4.1 Battery/AC Adaptor Installation...................................................................2-6
2.4.2 Control Panel and Display..............................................................................2-7
2.4.3 Control System ...............................................................................................2-10
2.4.3.1 Control Loops ..............................................................................................2-10
2.4.3.2 General Information....................................................................................2-11
2.4.4 RUN Loop—To Measure Temperature .......................................................2-12
2.4.5 LOG Loop—To Measure and Store Temperature......................................2-14
2.4.6 RECALL Loop—To Recall Measured Temperature ..................................2-16
2.4.6.1 RECALL Values from RUN .......................................................................2-16
2.4.6.2 RECALL Values from LOG........................................................................2-17
2.4.7 SETUP Loop—To Setup and Activate Alarms and Features...................2-19
2.4.7.1 SETUP Values for RUN ..............................................................................2-19
2.4.7.2 SETUP Values for LOG...............................................................................2-21
2.5 USING A TRIPOD...............................................................................................2-22
2.6 DATA OUTPUTS .................................................................................................2-23
2.6.1 Digital Output.................................................................................................2-24
2.6.1.1 Data Output—RUN Loop..........................................................................2-24
2.6.1.2 Data Output—SET Loop............................................................................2-25
2.6.1.3 Data Output—LOG Run Loop..................................................................2-26
2.6.1.4 Data Output—LOG SETUP Loop.............................................................2-27
2.6.2 Analog Output................................................................................................2-28
SECTION PAGE
Table of Contents Raynger 3i Series Operator's Manual
2.7 SIGHTING SYSTEMS..........................................................................................2-29
2.7.1 Single Laser Sighting .....................................................................................2-31
2.7.2 Dual Laser Sighting........................................................................................2-31
2.7.3 Crossed Laser Sighting..................................................................................2-32
2.7.4 Scope Sighting ................................................................................................2-32
3.0 SPECIFICATIONS ............................................................................3-1
3.1 THERMAL..............................................................................................................3-1
3.2 OPERATIONAL.....................................................................................................3-2
3.3 ELECTRICAL .........................................................................................................3-3
3.4 ENVIRONMENTAL..............................................................................................3-4
3.5 PHYSICAL..............................................................................................................3-4
3.6 DEFAULT VALUES ...............................................................................................3-5
3.7 REGULATORY.......................................................................................................3-6
4.0 MAINTENANCE...............................................................................4-1
4.1 BATTERY REPLACEMENT.................................................................................4-1
4.2 CLEANING ............................................................................................................4-1
4.2.1 Front Window Cleaning..................................................................................4-1
4.2.2 Cleaning the Housing......................................................................................4-2
4.3 LASER MAINTENANCE.....................................................................................4-2
SECTION PAGE
Raynger 3i Series Operator's Manual Table of Contents
APPENDIX A: OPTICAL...........................................................................A-1
HOW TO READ THE OPTICAL CHARTS...................................................................A-1
OPTICAL CHARTS ..........................................................................................................A-2
APPENDIX B: OBJECT EMISSIVITY.......................................................B-1
HOW TO DETERMINE OBJECT EMISSIVITY.............................................................B-1
TYPICAL EMISSIVITY VALUES.....................................................................................B-2
APPENDIX C: TROUBLESHOOTING ....................................................C-1
APPENDIX D: OPTIONS AND ACCESSORIES...................................
D-1
APPENDIX E: TRACEABILITY OF
INSTRUMENT CALIBRATION...............................................................E-1
APPENDIX F: CE CONFORMITY FOR THE
EUROPEAN COMMUNITY......................................................................F-1
GLOSSARY OF TERMS
INDEX
Table of Contents Raynger 3i Series Operator's Manual
Raynger 3i Series Operator’s Manual 1-1
1.0 INTRODUCTION
1.1 DESCRIPTION
The Raynger
®3i™
series of instruments are portable infrared temperature measure­ment devices. Each model is rugged and easy to use for making fast, noncontact, non­destructive temperature measurements. They can measure operating temperatures of mechanical, electrical, or production equipment without removing the equipment from service. They can also measure product temperatures during manufacturing or storage without contaminating or marring the product. Table 1-1 lists the standard Raynger 3i models.
Table 1-1: Raynger 3i Models
TEMPERATURE OPTICAL SPECTRAL MODEL RANGE RESOLUTION RANGE SIGHTING
LTDL2 & LTDL3
LTSC
LTCL2 & LTCL3
-30 to 1200°C
(-20 to 2200°F)
-30 to 1200°C
(-20 to 2200°F)
-30 to 1200°C
(-20 to 2200°F)
75:1
75:1
75:1
8-14 µ
8-14 µ
8-14 µ
dual laser
scope
crossed laser
LRSC
LRSCL2
LRL2 & LRL3
P7DL2 & P7DL3
G5SC
1MSC
1ML2 & 1ML3
2MSC
2ML2 & 2ML3
-30 to 1200°C
(-20 to 2200°F)
-30 to 1200°C
(-20 to 2200°F)
-30 to 1200°C
(-20 to 2200°F)
10 to 800°C
(50 to 1450°F) 150 to 1800°C
(300 to 3275°F)
600 to 3000°C
(1100 to 5400°F)
600 to 3000°C
(1100 to 5400°F)
200 to 1800°C
(400 to 3275°F)
200 to 1800°C
(400 to 3275°F)
120:1
105:1
120:1
25:1
50:1
180:1
180:1
90:1
90:1
8-14 µ
8-14 µ
8-14 µ
7.9 µ
5.0 µ
1.0 µ
1.0 µ
1.6 µ
1.6 µ
scope
single laser with scope
single laser
dual laser
scope
scope
single laser
scope
single laser
1-2 Raynger 3i Series Operator’s Manual
Each model is molded from rugged, high-strength, solvent resistant plastic and is actuated by a two-stage trigger (second stage is used for data logging only). Each model has the following:
High quality optical system
Infrared detector
Circuit board assembly with microprocessor
LCD display with backlighting feature
Touch-sensitive membrane switches for changing loops and modes
Battery compartment for four AA batteries
Power input jack for AC adaptor
Signal output jack (analog 1mV per degree/digital RS232)
A rugged, padded pouch for easy carrying
There are four types of laser sighting models:
Single laser—shows the center of the measurement area.
Dual laser—shows the spot diameter of the measurement area.
Crossed laser—the point where the two laser beams cross is the location of the minimum diameter measurement spot.
Single laser with scope
1.2 INVENTORY
Your Raynger 3i package contains the following:
Raynger 3i
Carrying Case
Operator’s Manual
Four (4) AA batteries
Warranty card
Raynger 3i Series Operator’s Manual 1-3
1.3 MODEL IDENTIFICATION
Refer to Table 1-1 for a list of standard models along with their temperature ranges, optical resolutions, spectral ranges, and sighting systems.
You can determine the exact model number of your unit by looking at the manufac­turing label on the underside of the unit. On the label is an area for model designa­tion. The model type is printed in the following format:
XXXYYYZZZZZ
where XXX is an abbreviation of the company name, YYY is the product (or abbrevia­tion of the product name), and ZZZZZ is the model type. (Model type may be four or five characters long. Refer to Table 1-1 to compare the label to model type.)
1-4 Raynger 3i Series Operator’s Manual
Raynger 3i Series Operator’s Manual 2-1
2.0 OPERATION
This portion of the manual contains the following sections:
Quick Start—To use your unit right away, follow the brief instructions on basic
operating procedures in this section.
Principles of Operation—A short introduction to infrared thermometry.
Your Portable Infrared Thermometer—Describes and illustrates the thermome-
ter’s control panel, display, and features.
How to Operate—A detailed user guide that describes each of the operating
modes. It includes descriptions of the RUN, RECALL, SETUP, and Data Logger loops.
Data Outputs—How to use the analog and digital (RS232) outputs to connect the
thermometer to a printer, chart recorder, or computer.
WARNING—LASER SAFETY
Models with laser sighting produce visible laser radiation that may be harmful to the human eye. Be aware of the following:
Avoid direct exposure of human eyes to laser light. Eye damage can result.
Use extreme caution when operating.
Never point the unit at another person.
Keep out of the reach of children.
Refer to the FDA laser label on the unit for specific information.
IMPORTANT
1. If the unit is exposed to significant changes in ambient temperature (hot to cold or cold to hot), allow 45 minutes for temperature stabi­lization before taking measurements.
2. Do not operate the unit near large electrical or magnetic fields such as arc welders and induction heaters. These fields can cause mea­surement errors.
3. For the short wavelength units (e.g., 2 µm and below)—Avoid taking temperature measurements in bright sunlight. High levels of ambi­ent light may produce apparently valid high-temperature readings when no target is in the thermometer’s field-of-view.
2-2 Raynger 3i Series Operator’s Manual
RUN
LOG
SET
LOG
HI
RCL
( )
LO
SET
( )
°C °F
T
amb
Raynger 3i Series Operator’s Manual 2-3
2.2 PRINCIPLES OF OPERATION
An infrared thermometer and the human eye are very similar. An infrared ther­mometer has a lens that focuses infrared radiation from an object onto a detector. The eye focuses light onto the retina. The detector is stimulated by the incoming infrared energy and produces a signal that is transmitted to the circuitry. The retina is stimu­lated by incoming light and sends a signal to the brain. The circuitry processes this signal and computes the temperature of the object.
The intensity of an object’s emitted infrared energy increases or decreases in propor­tion to its temperature. The higher the temperature of the target, the greater the inten­sity of infrared radiation.
To calibrate a noncontact temperature measurement instrument, the manufacturer uses a blackbody. A blackbody is a perfect emitter because it absorbs and emits all radiant energy but reflects or transmits none. The emissivity value of a blackbody is
1.00. Figure 2-2 shows the radiant emittance values of a blackbody at various tem­peratures and wavelengths.
Most objects have emissivities that are less than 1.00 but are reasonably uniform at all wavelengths of the infrared spectrum. These are called graybodies. The non-ideal (less then 1.00) emissivity values of different materials can be compensated for, by the emissivity controls, so that accurate temperature readings can be obtained. Emissivity values for many common materials (both metals and non-metals) are listed in Appendix B.
Figure 2-2: Blackbody Radiation Curves
2
10
2
1
10
1
-1
10
-2
10
-3
10
Blackbody Radiant Emittance (Watt/cm )
-4
10
12345678
0
Wavelength (Microns)
9 10111213
1500°C (2730°F) 1000°C (1830°F) 542°C (1000°F)
260°C (490°F)
20°C (70°F)
14
2-4 Raynger 3i Series Operator’s Manual
2.3 YOUR PORTABLE INFRARED THERMOMETER
Portable infrared thermometers measure surface temperatures without touching the surface. They collect the infrared energy radiated by a target and compute its surface temperature. They also compute the running average, maximum, minimum, and dif­ferential temperatures and present them on a digital display in either degrees Celsius or Fahrenheit. A digital/analog output allows data recording, instrumentation or process control, and/or remote display of temperature and emissivity. The instru-6(. The instr)7.6contr is batmety powerared/ocan be powerared byionptatiitaAC adaprato7354.6(. he)]TJ0 -2.4158 Tg0 TwInmetiitamemoty cirart aum, anacs danalgnninfea-he
Raynger 3i Series Operator’s Manual 2-5
Your portable thermometer has the following:
Trigger—Two-stage trigger. The first stage activates the unit to take temperature
readings. The second stage is functional only in the datalog mode. To store a tem­perature reading, pull the trigger all the way in until you hear the tone (the tone signals that the reading has been stored). When you release the trigger, the unit goes to sleep.
Control Panel and Display—All controls (except the trigger) are located on the
control panel. The display shows temperature and setup values, mode and loop status, and operating information.
Sighting System—Laser or scope sighting is provided with each model.
Note: Read the laser warning label before operating the laser.
Analog Output—Connects the instrument to analog recording/printing devices
such as chart recorders and printers.
Digital Output—An RS232 interface to connect the instrument to a computer or
directly to a printer’s RS232 port.
DC In—AC adaptor connection.
2-6 Raynger 3i Series Operator’s Manual
2.4 OPERATION AND CONTROLS
This section instructs you in the operation of the instrument. It describes battery and/or AC adaptor installation and the controls and functions of the different control loops and operating modes.
2.4.1 Battery/AC Adaptor Installation
The instrument may be powered by batteries or an AC adaptor. Battery power is sup­plied by 4 “AA” batteries. The batteries are located in the base of the handle. AC power is supplied by an optional AC power adaptor (DIN VDE 0551 approved). NiCad batteries can also be used, but battery life will be substantially reduced. Figure 2-3 shows the location of the battery compartment, battery orientation, and the loca­tion of the AC adaptor power connection.
Table 2-1 shows approximate battery life (for alkaline batteries) for various operating conditions.
Notes: Battery types and brands vary in length of usable life. The values in Table 2-1 are
approximate for new alkaline batteries. The instrument will continue to read accurate­ly up to 4 hours after the low battery icon displays if the laser and backlight are off.
Remove the batteries if the unit is not used for long periods of time.
220 VAC adaptors must have DIN VDE 0551 approval to be used with IEC Class 2 laser units.
Table 2-1: Battery Life (Alkaline)
CONDITION
Laser OFF Backlight OFF Laser ON Backlight OFF Laser OFF Backlight ON Laser ON Backlight ON "Sleep" mode (trigger off)
HOURS OF
CONTINUOUS USE
25
12.5
12.5 10
1 year
Raynger 3i Series Operator’s Manual 2-7
Figure 2-4: The Control Panel and Display
2.4.2 Control Panel and Display
Figure 2-4 shows the display and controls. Descriptions of these, in alphabetical order, follow the illustration.
Backlight on/off button
Run/Log selector switch
Setup button
Mode and Recall button
Laser on/off button
Trigger lock icon
Backlight icon
RUN
RUN
LOG
LOG
SET
Data logger
icon
RUN
RUN
LOG
LOG
SET
LOG
HI
RECALL
MODE
MODE
RECALL
RCL
LOG
HI
RCL
SET
( )
LO
LASER
Recall
loop icon
SET
( )
( )
LO
°C °F
( )
T
amb
ACTIVATE
ACTIVATE
LOCK
LOCK
°C °F
T
amb
Increase/up
button
Decrease/down
button
Activate and trigger
lock button
Setup
loop icon
Celsius/Fahrenheit
icons
Temperature
display
High and low
alarm icons
Ambient temperature
compensation icon
Low battery icon
Mode display
Laser on/off LED
RECALL
MODE
MODE
RECALL
LASER
ACTIVATE
ACTIVATE
LOCK
LOCK
Emissivity display
Mode value display
2-8 Raynger 3i Series Operator’s Manual
Activate button—Press the activate button to activate HAL, LAL, or TAM, or to tog­gle between DIG/ANA, or °C/°F, in the SET loop.
Backlight button and icon—The display has a backlight for working in low lighting conditions. Press the backlight button to activate or deactivate the backlight. The backlight icon is activated when the backlight is on. To save battery power, use the backlight only when necessary. Note that if the battery voltage falls below 4.0 V, the backlight will automatically turn off.
Low battery icon—The instrument is powered by four “AA” batteries. When the bat­tery voltage falls below 4.6 V, the low battery icon is activated.
Note: Turning off the backlight and laser will extend the battery life (refer to Table 2-1 for
battery life under various conditions).
Celsius/Fahrenheit icons—The °C and °F icons indicate which temperature scale has been selected.
LOG button and icon—Press the LOG/RUN button while you have the trigger pulled to toggle between the LOG and RUN loops. The LOG icon is activated when the instrument is in the LOG loop.
Emissivity display—The emissivity display shows the emissivity value selected in either the RUN, RECALL, or LOG loops.
and ▼buttonsincreases and decreases the emissivity settings, DOI rate,
HAL or LAL set-points, TAM, or the LOG location number.
High and Low alarm icons—These are activated when the corresponding alarms are activated. The HAL icon flashes and the buzzer sounds when measured temperature is greater than or equal to HAL and HAL is active. The LAL icon flashes and the buzzer sounds when the measured temperature is less than or equal to LAL and LAL is active. Note that when a high or low alarm condition is met and the unit is sending out RS232 data while in the Digital Output mode, the buzzer will make two tones: a normal tone followed by a higher pitched tone.
Laser on/off button and LED—Press the laser button to activate or deactivate the laser (RUN and LOG loops only). The laser LED is activated when the laser is activat­ed. (Not applicable for the scope sighting model.) Note that if the battery voltage falls below 4.3 V, the laser will automatically turn off. Also, if the unit is in the LOCK mode, the laser will go off when the trigger is released.
Temperature display—Shows the current temperature (while the trigger is pulled) or the last temperature measured (when the RECALL button is pressed).
Raynger 3i Series Operator’s Manual 2-9
Mode button—Press the mode button to change modes in any of the four loops.
Mode display and Mode value display—The Mode display shows the current mode
selected. The mode value display shows the temperature, set-point, or LOG location value for the mode selected.
Recall button and icon—Press the RCL button to activate the RECALL loop. The RECALL loop may be used to recall values from either the RUN or LOG loops. The RCL icon is activated when the instrument is in the RECALL loop.
RUN button—Press the RUN/LOG button, when the trigger is pulled, to toggle between the RUN and LOG loops.
Setup button and icon—Press the SET button to activate the SETUP loop. The SETUP loop may be used to set values in either the RUN or LOG loops. The SET icon is activated when the instrument is in the SET loop.
T
amb
icon—This is activated when the ambient temperature compensation function is
activated. This feature only affects readings in the RUN loop; readings in the LOG loop are not affected.
Trigger lock icon—The small padlock is the trigger lock icon and is activated when LOCK is pressed while the trigger is pulled. To unlock the trigger, simply press the LOCK button again.
2-10 Raynger 3i Series Operator’s Manual
2.4.3 Control System
The control system consists of four loops: SET, RUN, RECALL and LOG. The instru­ment may be cycled to any of the four loops by using the trigger or control panel but­tons, as shown below.
RCL
LOG
SET
HI
( )
LASER
( )
LO
°C °F
T
amb
ACTIVATE
LOCK
RUN
LOG
SET
LOG
HI
MODE
RECALL
RCL
( )
LASER
SET
( )
LO
°C °F
T
amb
ACTIVATE
LOCK
RUN
LOG
SET
MODE
RECALL
RUN
LOG
SET
LOG
HI
MODE
RECALL
RCL
( )
LASER
SET
( )
LO
°C °F
T
amb
ACTIVATE
LOCK
Raynger 3i Series Operator’s Manual 2-11
2.4.3.1 Control Loops
Each control loop has several modes, which are described in the following sections.
2.4.3.2 General Information
Before you begin using your portable instrument, you should be aware of the follow­ing general operating rules:
The instrument goes to “sleep” (low power consumption, no display, no laser, no backlight) after different times for each loop, as follows:
SET Loop (after last button use): 10 seconds – RUN Loop (after releasing the trigger): immediately – RECALL Loop (after last button use): 10 seconds _ LOG Loop (after releasing trigger): immediately
The backlight can be turned on or off in any mode.
The laser can be used only in the RUN and LOG loops.
The trigger has two stages. The first stage is for measuring temperature. The sec­ond stage, when the trigger is pulled all the way in, is for storing temperature val­ues in the LOG loop.
The °C or °F icon flashes when the instrument is measuring temperature in the RUN or LOG loops. It does not flash when values are recalled.
You can lock the trigger in the RUN loop and the LOG loop.
You must release the trigger to go into the SET or RECALL loops.
When you simultaneously press the MODE and ACTIVATE buttons while in the RUN loop, the instrument is reset to the RUN loop factory default settings (stored data is unaffected). In the LOG loop, pressing the MODE and ACTIVATE buttons not only resets the instrument to its LOG loop factory default settings, but also clears all previously stored data. In both cases, the instrument will “beep” after reinstalling the default settings. Section 3.6 lists default setting for each model as well as non-model-specific default settings and range values.
IMPORTANT
For the short wavelength units (e.g., 2 µm and below)—Avoid taking temperature measurements in bright sunlight. High levels of ambient light may produce apparently valid high-temperature readings when no target is in the thermometer’s field-of-view.
2.4.4 RUN Loop–To Measure Temperature
The RUN loop is for taking temperature measurements. Figure 2-6 illustrates the RUN loop.
To take a measurement, do the following:
1. Point the instrument at the target.
2. Pull the trigger (press the RUN/LOG button, if necessary, so that the LOG icon is not activated).
Figure 2-6: RUN Loop
2-12 Raynger 3i Series Operator’s Manual
RUN
RUN
LOG
LOG
SET
RUN
RUN
LOG
LOG
SET
MAX Mode
RCL
LOG
( )
HI
RECALL
MODE
MODE
RECALL
LASER
LASER
MODE
RCL
LOG
( )
HI
SET
( )
LO
ACTIVATE
SET
( )
LO
°C °F
T
amb
ACTIVATE
LOCK
LOCK
°C °F
T
amb
Current temperature
MODE
Turn the trigger lock ( ) on or off with the Lock button
The °C (or °F) icon flashes when the trigger is pulled
MODE
RUN
RUN
LOG
LOG
SET
RUN
RUN
LOG
LOG
SET
MIN Mode
RCL
LOG
( )
HI
RECALL
MODE
MODE
RECALL
LASER
MODE
RCL
LOG
( )
HI
LO
LASER
LO
SET
°C °F
( )
T
amb
ACTIVATE
ACTIVATE
LOCK
LOCK
SET
°C °F
( )
T
amb
Change emissivity
RECALL
MODE
MODE
RECALL
LASER
LASER
ACTIVATE
ACTIVATE
LOCK
LOCK
with the and buttons in any mode
RECALL
MODE
MODE
RECALL
LASER
LASER
ACTIVATE
ACTIVATE
LOCK
LOCK
DIF ModeAVG Mode
TRIGGER
Pulled or Locked
BUTTON
RUN
LOOP ICON
None
Raynger 3i Series Operator’s Manual 2-13
3. Press the laser button to activate the laser (if equipped with laser sighting).
4. Carefully aim using the laser or scope.
Note: The laser or scope indicate the target (see Sections 2.7.1 through 2.7.4). Make sure the
spot you are measuring fills the target.
5. Read the temperature from the display.
RUN contains 4 modes: MAX, MIN, DIF, and AVG (as shown in Figure 2-7). In RUN, the current temperature and emissivity setting along with either the MAX, MIN, DIF, or AVG temperature can be displayed. Note that AVG is a weighted average of all readings taken since the trigger was pulled. The unit uses the following formula to calculate the running average (while the trigger is pulled):
R
1
+ R2+ R3+ ...R
n
= AVG
n
where R = a reading and n = the total number of readings.
Press the MODE button to change modes.
Press the or buttons to change emissivity.
Press the backlight button if a brighter display is necessary.
Note: When you simultaneously press the MODE and ACTIVATE buttons while in the
RUN loop, the instrument is reset to its RUN loop factory default settings (stored data is unaffected). The instrument will “beep” after reinstalling the default settings. Section 3.6 lists default setting for each model as well as non-model-specific default set­tings and range values.
2-14 Raynger 3i Series Operator’s Manual
2.4.5 LOG Loop–To Measure and Store Temperature
The LOG loop is for making temperature measurements and storing them in the data logger locations. Figure 2-7 illustrates the LOG loop.
To use the modes and functions of the LOG loop, do the following:
1. Point the instrument at the target.
2. Pull the trigger and press the RUN/LOG button, if necessary, so that the LOG icon is activated.
3. Press the or buttons to select the LOG location number where you want to store the measurement. To change the emissivity setting for that location, press
the MODE button and use the or buttons .
4. Press the laser button to activate the laser (if equipped with laser sighting).
5. Carefully aim using the laser or scope.
6. Read the temperature from the display.
Figure 2-7: LOG Loop
LOC Mode MAX Mode
RCL
RUN
RUN
LOG
LOG
SET
LOG
SET
°C °F
( )
( )
HI
LO
T
amb
MODE
RUN
RUN
LOG
LOG
SET
LOG
RCL
SET
°C °F
( )
( )
HI
LO
T
amb
Lock and unlock the trigger with the lock button
RECALL
MODE
RECALL
MODE
ACTIVATE
ACTIVATE
LOCK
LOCK
LASER
LASER
TRIGGER
Pulled or locked
Use the and buttons to change the log location (LOC) number
BUTTON
LOG
RECALL
MODE
RECALL
MODE
LASER
LASER
LOOP ICON
LOG
ACTIVATE
ACTIVATE
LOCK
LOCK
Use the and buttons to change the emissivity
Raynger 3i Series Operator’s Manual 2-15
7. Pull the trigger further until you hear the “beep” indicating that the measurement has been stored. Both the current and MAX temperatures are stored in memory.
Notes: The instrument has a 2-stage trigger. The first stage is activated by pulling the trigger
a small amount. The second stage (operational in LOG only) is activated when the trigger is pulled all the way.
To protect the data from being accidentally overwritten (trigger accidently pulled), press the RUN/LOG button to exit the LOG loop (the LOG icon is off).
8. Release the trigger.
LOG contains two modes: LOC (location) and MAX (as shown in Figure 2-7). In LOG, the current temperature and emissivity setting along with either the location number or MAX temperature can be displayed. Note the following:
Press the MODE button to change between LOC and MAX modes.
Press the or buttons to change the location number or the emissivity set-
ting.
Press the backlight button if a brighter display is needed.
Notes: If you change the emissivity after storing data, both the target temperature and MAX
temperature will reset to zero. The instrument will “beep” to indicate this.
IMPORTANT
When you simultaneously press the MODE and ACTIVATE buttons while in the LOG loop, all previously stored data is cleared. The instrument also resets to its LOG loop factory default settings. The instrument will “beep” after reinstalling the default settings. Section
3.6 lists default setting for each model as well as non-model-specific default settings and range values.
2-16 Raynger 3i Series Operator’s Manual
2.4.6 RECALL Loop–To Recall Measured Temperatures
The RECALL loop is for recalling values from either the RUN or LOG loops.
2.4.6.1 RECALL Values from RUN
Figure 2-8 illustrates RECALL for RUN values.
RUN
LOG
SET
RUN
LOG
SET
LOG
RCL
( )
HI
LASER
MODE
SET
LO
LASER
ACTIVATE
LOCK
°C °F
( )
T
amb
ACTIVATE
LOCK
MODE
MODE
If the display goes out, press the Recall button again to continue the Recall function
RUN
LOG
SET
RUN
LOG
SET
LOG
LOG
HI
HI
RCL
( )
LO
LASER
MODE
RCL
( )
LO
LASER
SET
°C °F
( )
T
amb
ACTIVATE
LOCK
SET
°C °F
( )
T
amb
ACTIVATE
LOCK
Raynger 3i Series Operator’s Manual 2-17
To recall values from the RUN loop, do the following:
1. Release or unlock the trigger, if necessary.
2. Press the RECALL button. The RCL icon will be activated.
3. Press the RUN/LOG button, if necessary, so that the LOG icon is not activated.
4. Read the recalled temperature from the display.
RECALL contains four modes: MAX, MIN, DIF, and AVG (as shown in Figure 2-8). In RECALL, the last temperature and emissivity setting along with either the MAX, MIN, DIF, or AVG temperature can be displayed. Note the following:
Press the MODE button to change modes.
Press the or button to change emissivity. (This shows the effect that a dif-
ferent setting would have. Note that if the target’s temperature is known, you can determine its emissivity by this method. MAX, MIN, DIF, or AVG are not updated.)
Press the backlight button if a brighter display is needed.
2.4.6.2 RECALL Values from LOG
Figure 2-9 illustrates RECALL for LOG values.
Figure 2-9: RECALL Loop–Recalling LOG Values
LOC Mode MAX Mode
RCL
RUN
RUN
LOG
LOG
SET
RECALL
Press the Recall button again, if display goes out, to continue Recall function
LOG
MODE
RECALL
MODE
SET
°C °F
( )
( )
HI
LO
T
amb
ACTIVATE
LASER
LASER
ACTIVATE
LOCK
LOCK
Use the and buttons to change the log location (LOC) number
TRIGGER
Released & Unlocked
MODE
Stored temperature. If no temperature
is stored, the display shows all zeros.
BUTTON
RECALL & LOG
RUN
RUN
LOG
LOG
SET
Press the Mode button to toggle between the LOC and MAX modes.
LOOP ICONS
RCL
LOG
( )
HI
RECALL
MODE
RECALL
MODE
LASER
LASER
LOG RCL
SET
( )
LO
°C °F
T
amb
ACTIVATE
ACTIVATE
LOCK
LOCK
Stored temperatures. If no temperatures
are stored, the displays show all zeros.
To recall values from the LOG loop, do the following:
1. Release or unlock the trigger, if necessary.
2. Press the RECALL button. The RCL icon will be activated.
3. Press the RUN/LOG button, if necessary, so that the LOG icon is activated.
4. Read the recalled temperature from the display.
5. Press the or buttons (LOC mode) to recall values from other LOG locations.
RECALL contains two modes: LOC (location) and MAX (as shown in Figure 2-9). In RECALL, the stored temperature and emissivity setting along with either the location number or MAX temperature can be displayed for each of the 100 LOG locations.
2-18 Raynger 3i Series Operator’s Manual
Raynger 3i Series Operator’s Manual 2-19
2.4.7 SETUP Loop–To Setup and Activate Alarms and Features
The SETUP loop is for setting up and activating alarms and features in either the RUN or LOG loops.
2.4.7.1 SETUP Values for RUN
Figure 2-10 illustrates SETUP for the RUN loop.
Figure 2-10: SETUP Loop–RUN Values
The display changes to show temperature in °C or °F
ANA
Selected
RUN
RUN
LOG
LOG
SET
F/C Mode
LOG
HI
RECALL
MODE
MODE
RECALL
RUN
RUN
LOG
LOG
SET
RCL
( )
LASER
LASER
MODE
SET
( )
LO
RECALL
°C °F
T
amb
ACTIVATE
ACTIVATE
LOCK
LOCK
MODE
LOG
HI
MODE
MODE
RECALL
DOI Mode
RCL
SET
( )
( )
LO
ACTIVATE
LASER
LASER
DIG
Selected
MODE
Press the ACTIVATE button to toggle between °C or °F
°C °F
T
amb
ACTIVATE
LOCK
LOCK
Turn the High Alarm (HAL) on or off with the ACTIVATE Button
Time (in seconds) between data transmissions
Press or to change the interval (from 1 to 9999 seconds)
RUN
RUN
LOG
LOG
SET
RUN
RUN
LOG
LOG
SET
HAL Mode
RCL
LOG
( )
HI
RECALL
MODE
MODE
RECALL
LASER
LASER
MODE
RCL
LOG
( )
HI
RECALL
MODE
MODE
RECALL
LASER
LASER
MODE
SET
( )
T
LO
ACTIVATE
ACTIVATE
LOCK
LAL Mode
SET
( )
LO
T
ACTIVATE
ACTIVATE
LOCK
°C °F
amb
LOCK
°C °F
amb
LOCK
Change the High Alarm setpoint with the and buttons
Turn the Low Alarm (LAL) on or off with the ACTIVATE Button
Change the Low Alarm setpoint with the and buttons
Last temperature value changes if
is activated
T
amb
RCL
LOG
SET
( )
HI
LO
RECALL
MODE
MODE
RECALL
LASER
LASER
TAM Mode
LOOP ICON
SET
°C °F
( )
T
amb
ACTIVATE
ACTIVATE
LOCK
LOCK
Change the TAM setpoint with the and buttons
Turn the Ambient Temperature (T on or off with the ACTIVATE Button
amb
)
Press the ACTIVATE button to toggle between DIG or ANA
RCL
LOG
RUN
RUN
LOG
LOG
( )
HI
LO
SET
RECALL
MODE
MODE
RECALL
LASER
LASER
DIG/ANA Mode
Released & Unlocked
SET
°C °F
( )
T
amb
ACTIVATE
ACTIVATE
LOCK
LOCK
TRIGGER
MODE
BUTTON
SET & RUN
RUN
RUN
LOG
LOG
SET
To setup values for the RUN modes and functions, do the following:
1. Release or unlock the trigger, if necessary.
2. Press the SET button. The SET icon will be activated.
3. Press the RUN/LOG button, if necessary, so that the LOG icon is not activated.
4. Press ACTIVATE to toggle between °C or °F for the display and data output.
5. Press the MODE button to switch between HAL, LAL, TAM, and DOI. Press the
or buttons to change the HAL, LAL, TAM, and DOI settings.
6. Press ACTIVATE to activate the HAL, LAL, or TAM.
7. Press ACTIVATE to toggle between DIG (digital) or ANA (analog) outputs.
8. Press the or buttons to set DOI (Digital Output Interval) if DIG was selected.
SET contains six modes: °C/°F, HAL, LAL, TAM, DIG/ANA, and DOI (as shown in Figure 2-10). In SET, the temperature scale can be set for °C or °F, the HAL, LAL, and
TAM setpoints can be displayed and set, the digital (DIG) or analog (ANA) output can be selected, and the Digital Output Interval (DOI) can be set. Note the following:
Press the MODE button to change modes.
Press the backlight button if a brighter display is needed.
Notes on setting T
amb
(Ambient temperature compensation)—The T
amb
icon is activated when the ambient temperature compensation function is active. Targets that have low emissivities will reflect energy from nearby objects. This additional reflected energy is added to the target’s own emitted energy and may result in inaccu­rate readings. In some situations objects near the target (machines, furnaces, or other heat sources) have a temperature much higher than that of the target. In these situa­tions it is necessary to compensate for the reflected energy from those objects. Note that the T
amb
feature is disabled if the emissivity is set to 1.00.
To set or change the ambient temperature compensation, complete the following:
1. Go to the RUN loop (pull trigger, press RUN/LOG button, if necessary).
2. Set the emissivity value to 1.0.
3. Press the MODE button until the mode function indicator displays AVG (average
temperature).
4. Pull the trigger and scan across objects and surfaces that face the target. Read the
average temperature value in the MODE value display. This is the value to be entered as the reflected ambient temperature in Step 7 below. Release the trigger.
5. Press the SET button (do not pull the trigger) to enter the Setup mode
6. Press the MODE button until TAM shows in the mode function indicator.
2-20 Raynger 3i Series Operator’s Manual
Raynger 3i Series Operator’s Manual 2-21
7. Set the TAM value by pressing the and buttons.
8. Press the ACTIVATE button to turn on the ambient temperature compensation function. The Tamb icon will be displayed indicating that it is active. (Pressing ACTIVATE again turns off the function.)
9. Pull trigger and reset the emissivity to the proper value for the target. You can now take the target’s temperature using normal procedures. The current tempera­ture and all computed values, as well as the analog and digital outputs, will now be based on the compensated measurement.
Note: TAM only affects readings in the RUN loop; LOG loop readings are not affected.
2.4.7.2 SETUP Values for LOG
Figure 2-11 illustrates SETUP for the LOG loop.
To setup values for the LOG modes and functions, do the following:
1. Release or unlock the trigger, if necessary.
2. Press the SET button. The SET icon will be activated.
3. Press the RUN/LOG button, if necessary, so that the LOG icon is activated.
4. Press the or buttons to select the LOG location number (LOC mode).
Figure 2-11: SETUP Loop–LOG Values
LOC Mode HAL Mode
RUN
RUN
LOG
LOG
SET
Press the Set button again, if the display goes out, to continue the SETUP functions
LOG
RECALL
MODE
RECALL
MODE
SET
°C °F
( )
( )
HI
LO
T
amb
ACTIVATE
LASER
LASER
ACTIVATE
LOCK
LOCK
TRIGGER
Released & Unlocked
MODE
Press the and buttons to change the Log Location (LOC) number.
You can select and log data in any location number from 1 to 100.
BUTTON
SET & LOG
RUN
RUN
LOG
LOG
SET
Turn High Alarm on or off with the Activate button
RCL
LOG
( )
HI
RECALL
MODE
RECALL
MODE
LASER
LASER
LOOP ICON
LOG SET
SET
( )
LO
ACTIVATE
ACTIVATE
RCL
°C °F
T
amb
LOCK
LOCK
Change the High Alarm temperature value with the and buttons
5. Press the MODE button.
6. Press the or buttons to change the HAL setting (HAL mode).
7. Press ACTIVATE to activate the HAL setting for the selected location.
SET contains two modes: LOC and HAL (as shown in Figure 2-11). In SET, the LOG location can be selected and the high alarm value (HAL) can be set for each location (independent of the setting for the RUN loop). Note the following:
Press the MODE button to change modes.
Press the backlight button if a brighter display is needed.
2.5 USING A TRIPOD
The instrument is equipped with a standard camera-type tripod mount fitting at the base of the handle. You can set it up for continuous use by mounting it on a sturdy tripod. This allows the instrument to be connected to a digital or analog recording or controlling device (computer, printer, chart recorder, line or temperature controller) to monitor temperatures over long periods of time.
Mounting the instrument is easy. Just attach it to the tripod by screwing the tripod’s mounting screw into the fitting at the bottom of the handle.
You can also connect a digital or analog recording or controlling device to the digi­tal/analog connector (labeled OUT) on the side of the instrument. For continuous monitoring, pull the trigger and press the LOCK button.
Note that battery life is shortened by continuous use. If you plan to monitor tempera­tures over long intervals, you should connect an AC adaptor to the DC IN connector.
2-22 Raynger 3i Series Operator’s Manual
Raynger 3i Series Operator’s Manual 2-23
2.6 DATA OUTPUTS
Data outputs from the instrument provide a direct interface to chart recorders, print­ers, and computers. All models are equipped with an output jack capable of provid­ing analog and digital signals, which are user selectable in the SET loop. The format of these signals are as follows:
Digital: RS-232
Format: ASCII data Baud Rate: 9600 Data Format: 8 bits, 1 stop bit, no parity
Analog: 1 mV/° (°C or °F) for all models except the 1M
1 mV/°C or 0.5 mV/°F for the 1M
Figure 2-12: Data Outputs
F0072 E095 MAX 0080
F0072 E095 MAX 0080
F0072 E095 MAX 0080
F0072 E095 MAX 0080
F0072 E095 MAX 0080
F0072 E095 MAX 0080
F
0
0
7
2
E
0
F
9
0
5
0
7
M
2
A
X
E
0
0
F
9
0
0
5
8
0
0
7
M
2
A
X
E
0
0
9
0
5
8
F
0
0
M
0
A
7
X
2
0
E
0
0
8
9
F
0
5
0
0
M
7
A
2
X
E
0
0
0
9
8
F
5
0
0
0
M
7
A
2
X
E
0
0
0
F
9
8
0
5
0
0
7
M
2
A
X
E
0
0
9
0
F
5
8
0
0
0
M
7
A
2
X
E
0
0
0
F
9
8
0
5
0
0
7
M
2
A
X
E
0
0
9
0
5
8
0
M
A
X
0
0
8
0
RS232
RS232
Analog
2.6.1 Digital Output
The following sections describe the digital data outputs for each mode.
2.6.1.1 Data Output—RUN Loop
In the RUN loop, the temperature scale, current temperature, emissivity, mode tem­perature (MAX, MIN, DIF, or AVG), HAL or LAL, a carriage return, and a line feed are sent out at intervals determined by the DOI (set in the SET loop) when the trigger is pulled. HAL or LAL is sent out if the temperature is above HAL or below LAL (when they are activated). Three blank spaces replace HAL or LAL if they do not meet this condition. There are 5 characters for the mode temperature. The fifth char­acter is always a space for MAX, MIN, or DIF. For AVG, the fifth character is a space if the average temperature is 1000° or above. Average temperatures up to 999.9° use all 5 characters. A total of 29 characters are sent out.
Table 2-2 shows a sample output. The headings at the top of the table explain the con­tents of each column (these are not part of the output).
2-24 Raynger 3i Series Operator’s Manual
Table 2-2: RUN Mode Output Example
S p a
°C
c e
or °F Temp Emissivity Value Mode Temperature 3 Blank Spaces Return Feed
S p a
Main E for Emissivity Mode HAL, LAL, or
c e
S p a c e
S p a c e
S p a c e
S p a c e
Carriage Line
F 0072 E 0 . 95 MAX 0072 <CR> <LF> F 0073 E 0 . 95 MAX 0073 <CR> <LF> F 0090 E 0 . 95 MAX 0090 <CR> <LF> F 0070 E 0 . 95 MAX 0090 <CR> <LF> F 0071 E 0 . 95 MAX 0090 HAL <CR> <LF> F 0081 E 0 . 95 MAX 0090 <CR> <LF> F 0080 E 0 . 95 MAX 0090 <CR> <LF> F 0070 E 0 . 95 MIN F 0070 E 0 . 95 DIF F 0081 E 0 . 95 AVG F –018 E 0 . 95 AVG F 0421 E 0 . 95 AVG F 1480 E 0 . 95 AVG
0070 <CR> <LF>
0020 <CR> <LF>
088.4 <CR> <LF>
–11.5 <CR> <LF>
100.0 <CR> <LF>
999.9 <CR> <LF>
F 1900 E 0 . 95 AVG 1000 <CR> <LF> F 0071 E 0 . 95 MAX 0090 <CR> <LF> F 0081 E 0 . 95 MIN F 0081 E 0 . 95 MIN C
0027 E 0 . 95 MIN –028 <CR> <LF>
C
0027 E 1 . 00 MIN –028 <CR> <LF>
–018 LAL <CR> <LF>
–018 <CR> <LF>
Notes
(see below)
1
2 3 4 5 6 7 8
9
10 11
Raynger 3i Series Operator’s Manual 2-25
Notes:
1. Indicates HAL is active and temperature HAL value.
2. Mode changed to MIN.
3. Mode changed to DIF.
4. Mode changed to AVG.
5. Main and Average temperatures went negative.
6. Average temperature goes positive and is 100°F.
7. Average temperature is up to 999.9°F.
8. Average temperature is now greater than 999.9°F so decimal point is no longer
displayed.
9. Indicates LAL is active and temperature LAL value.
10. Scale is changed from °F to °C.
11. Emissivity is changed to 1.00.
2.6.1.2 Data Output—SET Loop
After entering the SET loop (SET icon activated), setup parameters can be sent out by pressing the SET button again. This information can be sent out at any time while still in the SET loop. Unactivated setpoints are printed in lowercase (e.g., hal) and activat­ed setpoints are printed in uppercase (e.g., HAL). Hi Alarm, Low Alarm, T. Ambient, Temperature Scale, and DOI are sent out and terminated by a carriage return and line feed. A total of 39 characters are sent out. The SET values will be printed any time you press the SET button. Two examples follow:
hal 0085 lal 0065 tam 0100 F;DOI 3600<CR><LF>
(hal, lal, tam not activated)
HAL 0085 lal 0065 tam 0100 F;DOI 3600<CR><LF> (HAL activated)
The above lines indicate Hi Alarm = 85, Low Alarm = 65, T. Ambient = 100, the tem­perature scale is °F, and DOI = 3600 seconds. (The <CR> and <LF> do not print; these denote the carriage return and line feed.)
Note: DOI does not apply to the SETUP mode.
2.6.1.3 Data Output—LOG Run Loop
When you press the trigger (first stage only) in the LOG Loop, data will be sent out only for those locations that have stored data. The temperature scale, stored main temperature, stored emissivity, stored MAX temperature, location number, HAL, a carriage return, and a line feed are sent out when the trigger is pressed. HAL is sent out if the stored temperature is above or equal to the HAL value for that location, and HAL is activated. Three spaces replace HAL if this condition is not satisfied. A total of 32 characters are sent out.
Table 2-3 shows a sample output. The headings at the top of the table explain the con­tents of each column (these are not part of the output).
Notes:
1. Current and MAX temperature stored at Location 1.
2. Current and MAX temperature stored, and the emissivity changed to 1.00 at Location 6.
3. Current and MAX temperature stored at Location 7, HAL is active, and the stored temperature is HAL value. When the 2-stage trigger is pulled all the way to store the current and maximum temperatures at Location 7, the current temperature also exceeded the HAL value, which was activated. Therefore, HAL was sent out at the end of the string.
Notes: Note that in the above example nothing was stored at locations 2, 3, 4, 5, and 8 through
99, so these locations were not sent out. There will be a pause in printing between locations 1 and 6 and between locations 7 and 100 because the unit is searching through the locations sequentially for stored data.
DOI does not apply to the LOG mode.
2-26 Raynger 3i Series Operator’s Manual
Table 2-3: LOG Run Mode Output Example
S p a
°C
c e
or °F Temp Emissivity Value Mode Temperature Number HAL Return Feed
S p a
Main E for Emissivity Stored MAXMAX Location
c e
S p a c e
S p a c e
S p a c e
S p a c e
F 0072 E 0 . 95 MAX 0090 001 <CR> <LF> F 0095 E 1 . 00 MAX 0100 006 <CR> <LF> F 0070 E 0 . 95 MAX F 0090 E 0 . 97 MAX
0090 007 HAL <CR> <LF>
0100 100 <CR> <LF>
S p a c e
Carriage Line
Notes
(see below)
1 2 3
2.6.1.4 Data Output—LOG SETUP Loop
After entering the LOG SETUP loop (both the LOG and SET icons are activated), setup parameters are sent out when the SET button is pressed again. This informa­tion can be sent out at any time in the LOG SETUP mode. HAL, temperature scale, and the location number are sent out and terminated by a carriage return and line feed. A total of 16 characters are sent out. Go to the appropriate location number in the LOG loop and press the SET button to send out the Hi Alarm value for that loca­tion. Activated Hi Alarm locations are uppercase (HAL), and unactivated Hi Alarm locations are lowercase (hal). Only the HAL value at the present location is sent out. For additional HAL values, go to the appropriate location and press SET again while still in any of the two screens in the LOG SETUP mode. To get the LOG SETUP para-
meters for all 100 log locations, press the SET button for about 3 seconds.
Table 2-4 shows a sample output. The headings at the top of the table explain the con­tents of each column (these are not part of the output).
Notes:
1. For Location 1, HAL is not activated and is 1000°F.
2. For Location 2, HAL is activated and is 1000°F.
3. For Locations 3, 4, 5, 6, and 7, HAL is not activated and is 85°F.
4. For Location 8, HAL is activated and is 65°F.
Note: DOI does not apply to the LOG SETUP mode.
Raynger 3i Series Operator’s Manual 2-27
Table 2-4: LOG SETUP Mode Output Example
S
S p a
hal
c e
or HAL Value or °F Number Return Feed
S
p
p
a
a
HAL °C Location
c
c
e
e
Carriage Line
Notes
(see below)
hal 1000 F 001 <CR> <LF> HAL 1000 F 002 <CR> <LF> hal 0085 F 003 <CR> <LF> hal 0085 F 004 <CR> <LF> hal 0085 F 005 <CR> <LF> hal 0085 F 006 <CR> <LF> hal 0085 F 007 <CR> <LF> HAL 0065 F 008 <CR> <LF>
1 2 3 3 3 3 3 4
2.6.2 Analog Output
The analog output is made up of the following:
Output: 1 mV/° (°C or °F) for all models except 1M
1 mV/°C or 0.5 mV/°F for the 1M
Output impedance: 1.3 K
W
Use the signal output jack accessory.
The analog output represents the current temperature of the object being measured, regardless of the mode used. If T
amb
compensation is activated, the analog output
will be representative of the compensated temperature values.
2-28 Raynger 3i Series Operator’s Manual
Raynger 3i Series Operator’s Manual 2-29
2.7 SIGHTING SYSTEMS
The aiming options for this instrument are laser(s), scope, or laser and scope combina­tion. Laser sighting is available in single or dual spot models or in a crossed laser model.
Two different laser power levels are available for this product. Please refer to the label on your unit and to the following label diagrams (Figure 2-13) to determine the one that you have. The following table shows the specifications for both laser types.
CAUTION
Use of controls or adjustments, or performance of procedures other than those specified herein, may result in hazardous laser radiation exposure.
WARNING—LASER SAFETY
Models with laser sighting produce visible laser radiation that may be harmful to the human eye. Be aware of the following:
Avoid direct exposure of human eyes to laser light. Eye damage can result.
Use extreme caution when operating.
Never point the unit at another person.
Keep out of the reach of children.
Refer to the FDA laser label on the unit for specific information.
To operate the laser, do the following:
1. Point the instrument toward the spot being measured. (Do not point it at anyone.)
2. Pull the trigger.
3. Press the Laser button on the control panel.
4. Aim accordingly. (Refer to the following sections for laser aiming information.)
Note: The laser turns off automatically when the trigger is released. It cannot be locked on when
the trigger is locked. It cannot be turned on in the Recall and Setup loops.
Table 2-5: Laser Specifications
FDA Class IIIa FDA Class II and IEC Class 2
Type Wavelength Power Operating Range
(depending on ambient light level)
Gallium Arsenide 630-670 nm <5 milliwatt Up to 30 m (100 ft)
Gallium Arsenide 630-670 nm <1 milliwatt Up to 15 m (50 ft)
2-30 Raynger 3i Series Operator’s Manual
FDA Class II
Figure 2-13: Laser Labels
AVOID EXPOSURE ­LASER RADIATION IS EMITTED FROM THIS APERTURE.
CAUTION
LASER RADIATION - DO NOT STARE INTO BEAM OUTPUT < 1mW WAVELENGTH 630-670 nm CLASS II LASER PRODUCT
COMPLIES WITH
FDA 21 CFR
SUBCHAPTER J
FDA Class IIIa
IEC Class 2
POR ESTA ABERTURA SE EMITE RADIACION LASER. EVITE EXPONER LA VISTA.
AVOID EXPOSURE ­LASER RADIATION IS EMITTED FROM THIS APERATURE.
<1mW/630–670nm
IEC 825/93
PELIGRO
RADIACION LASER – EVITE EXPONER LA VISTA SALIDA < 5mW
`
LONGITUD DE ONDA 630-670 nm PRODUCTO LASER CLASE IIIa
DANGER
LASER RADIATION - AVOID DIRECT EYE EXPOSURE OUTPUT < 5mW WAVELENGTH 630-670 nm CLASS IIIa LASER PRODUCT
LASER LIGHT
DO NOT STARE INTO BEAM
CLASS 2 LASER
RAYONNEMENT LASER
NE PAS EXPOSER L'OEIL
AU RAYON LASER
LASER DE CLASSE 2
CUMPLE CON LAS
NORMAS FDA 21 CFR
SUB CAPITULO J
COMPLIES WITH
FDA 21 CFR
SUBCHAPTER J
LASERLIGHT
NICHT IN DEN STRAHL
BLICKEN
LASER KLASSE 2
RAYO LASER
NO FIJAR LA VISTA
EN EL RAYO
LASER CLASE 2
Raynger 3i Series Operator’s Manual 2-31
2.7.1 Single Laser Sighting
A single laser unit (see Figure 2-14) indicates the center of the area being measured, not the diameter of the spot. (To find out the diameter of the spot being measured, refer to the optical charts in Appendix A.)
2.7.2 Dual Laser Sighting
A dual laser unit indicates the diameter of the spot being measured. To take a mea­surement, aim the unit at the target and move closer or farther until the target is with­in the laser dots.
Note: The IR and laser spot diameters are not the same at close distances. The distance
between the laser beams is slightly greater than the spot being measured. Refer to Figure 2-15 for a comparison of the IR spot and laser beam diameters.
Figure 2-15: Dual Laser and IR Spot Diameters
Figure 2-14: Single Laser Sighting
Dual Laser Focus Point D:S = 75:1 @ 15 m (50 ft) Far Field D:S = 68:1
DISTANCE: SENSOR TO OBJECT (ft)
0
SPOT DIA. (in)
SPOT DIA. (mm)
02
IR Spot Diameter at Lens = 23 mm (0.9 in)
Laser Diameter at Lens = 40 mm (1.6 in)
10 20 30 40 50
16
12
8
4 0
75
150
225 300
4 6 8 10 12 14 16
DISTANCE: SENSOR TO OBJECT (m)
IR
LASER
2-32 Raynger 3i Series Operator’s Manual
2.7.3 Crossed Laser Sighting
The distance at which the two laser dots overlap is the point where the smallest area is measured (Focus Point). To find this distance, aim the unit at the target and move closer or farther until the laser beams overlap. Refer to Figure 2-16 for a comparison of the IR spot and laser beam diameters.
2.7.4 Scope Sighting
Scope sighting models (Figure 2-17) have parallax-free sighting (at the focus distance). Look through the scope. There are one or two circular reticles, depending on model. The inner reticle indicates the area that is measured at the focus point (see Appendix A for the focus point distance for your model). The outer reticle indicates an area greater than or equal to the area that is measured at all other distances.
Figure 2-16: Crossed Laser and IR Spot Diameters
Figure 2-17: Scope Sighting
Crossed Laser Focus Point D:S = 75:1 Far Field D:S = 20:1
DISTANCE: SENSOR TO OBJECT (in)
0
24 36 60 120 180
12
Focus Point
0.3 in @ 24 in
SPOT DIA. (in)
Focus Point
8 mm @ 610 mm
SPOT DIA. (mm)
0 610
1000 1500 3000 5000
DISTANCE: SENSOR TO OBJECT (mm)
9
6
3 0
75
150
225
300
Raynger 3i Series Operator’s Manual 2-33
2.7.5 Scope with Laser Sighting
The scope and single laser sighting model (Figure 2-18) combines the parallax-free sight­ing of the scope with the convenience of the laser. The combination unit has a single class 2 laser. It indicates the center of the target being measured, not the diameter of the spot being measured. See section 2.7.1 and Figure 2-14 for further details on the laser. The scope has one reticle and it indicates the target area that is measured at the focus point (see Appendix A for the focus point distance of your model). See section 2.7.4 and Figure 2-17 for further details on the scope.
Figure 2-18: Scope with Laser Sighting
2-34 Raynger 3i Series Operator’s Manual
Raynger 3i Series Operator’s Manual 3-1
3.0 SPECIFICATIONS
This section covers the following specifications:
Thermal
Operational
Electrical
Environmental
Physical
Also covered are factory default settings for each model as well as factory default set­tings and range values for all models.
3.1 THERMAL
Table 3-1 lists thermal specifications for each model.
Table 3-1: Thermal Specifications
* The instrument may measure temperature a few degrees below the minimum and
above the maximum specified temperature range.
Note: Hi Alarm, Lo Alarm, and T. Ambient ranges are the same as measurement ranges in
Table 3-1.
1M 2M G5 P7
600 to 3000°C
1.0 micron
Silicon diode
±0.5% of reading or ±1°C (±1°F), whichever is greater
200 to 1800°C
(400 to 3275°F)
1.6 micron
InGaAs diode
±3mV of display reading
150 to 1800°C
(300 to 3275°F)
5 micron
Thermopile
10 to 800°C
(50 to 1450°F)
7.9 micron
Thermopile
Measurement Range
Spectral Response
Detecting Element
Display Accuracy
Analog Output Accuracy
Repeatability Temperature
Coefficient
LT, LR, LRSCL2
-30 to 1200°C
(-20 to 2200°F)
8 to 14 micron
Thermopile
±1% of reading (±0.5% for 1M) or ±1°C (±1.5°F), whichever is greater, or ±2°C (±4°F)
for targets below 0°C (32°F), at 23°C ±5°C (73°F ±9°F) ambient operating temp.
(1100 to 5400°F)
±0.1°C (±0.1°F) change per 1°C (1°F) change in ambient temperature
3-2 Raynger 3i Series Operator’s Manual
3.2 OPERATIONAL
Table 3-2 lists operational specifications for each model.
Table 3-2: Operational Specifications
Temperature Display
Display and Digital Output Resolution
Display Controls
Emissivity DOI (Digital
Output Interval) Analog Output
Resolution LOG Locations Response Time
(95% response)
LT, LR, LRSCL2
°C or °F (selectable), multifunctional 4-digit backlit LCD
1°C or 1°F (0.1°C or 0.1°F in AVG Mode for temperatures up to 999.9°)
Adjustable from 0.10 to 1.00 in 0.01 increments (default = 0.95)
700 msec 550 msec 700 msec 700 msec550 msec
1M 2M G5 P7
Membrane Switch Panel
1 to 9999 seconds
1°C or 1°F
1 to 100
Raynger 3i Series Operator’s Manual 3-3
Table 3-3: Electrical Specifications
3.3 ELECTRICAL
Table 3-3 lists electrical specifications for each model.
LT, LR, LRSCL2 1M 2M G5 P7
Analog Output Analog Output
Range Analog Output
Impedance Digital Output
Digital Output Range
Power Requirements
1mV/°C (1mV/°F) 1mV/°C (1mV/°F)
Four (4) AA alkaline or rechargeable batteries or 6 to 9 volt, 200 mA, DC power supply
1mV/°C (0.5mV/°F)
Same as Measurement Range in Table 3-1
1.3 K
RS232, 9600 baud
Same as Measurement Range in Table 3-1
3-4 Raynger 3i Series Operator’s Manual
3.4 ENVIRONMENTAL
Table 3-4 lists environmental specifications for each model.
Table 3-4: Environmental Specifications
3.5 PHYSICAL
Table 3-5 lists environmental specifications for each model.
Table 3-5: Physical Specifications
Note: The laser will turn off automatically if the instrument’s internal temperature exceeds
45° C (113° F).
LT, LR, LRSCL2
Relative Humidity
Storage Temperature
Ambient Operating Temperature
1M
10 to 95% at up to 30°C (86°F)
under non-condensing conditions
-20 to 50°C (-4 to 120°F)
0 to 50°C (32 to 120°F)
2M G5 P7
without battery
SCOPE & LASER
244 mm (9.6 in) H
257 mm (10.1 in) L
71 mm (2.8 in) W
1000 g (2.21 lbs)
Dimensions
Weight
SCOPE MODELSLASER MODELS
208 mm (8.2 in) H
257 mm (10.1 in) L
71 mm (2.8 in) W
794 g (1.75 lbs) 1000 g (2.21 lbs)
244 mm (9.6 in) H
257 mm (10.1 in) L
71 mm (2.8 in) W
Raynger 3i Series Operator’s Manual 3-5
3.6 DEFAULT VALUES
Table 3-6 lists the factory default values for each model.
Table 3-6: Model Specific Factory Default Values
Table 3-7 lists factory default settings that are valid for all models.
Table 3-7: Non-Model Specific Factory Default Values
Hi Alarm Lo Alarm T. Ambient
LTDL, LTCL, LTSC, LRSC, LRL, LRSCL2
1200°C
(2200°F)
0°C
(0°F)
23°C
(73°F)
1MSC, 1ML
P7DL
G5SC
2MSC, 2ML
3000°C
(5400°F)
800°C
(1450°F)
1800°C
(3275°F)
1800°C
(3275°F)
600°C
(1100°F)
10°C
(50°F) 150°C
(300°F)
200°C
(400°F)
600°C
(1100°F)
23°C
(73°F) 150°C
(300°F)
200°C
(400°F)
Function Default
Emissivity DOI (Digital Output Interval) Digital/Analog Output Trigger LOCK Laser Backlight HAL LAL TAM Scale: U.S. Scale: Outside U.S. Loop Setup Screen (RUN Loop) Setup Screen (LOG Loop) LOG Location
0.95 60 seconds Analog Output off off off off off off degrees F degrees C RUN Loop, MAX Mode F/C LOC (Location) 1
LOG Data–All Locations
Current Temperature MAX Temperature Hi Alarm Value Hi Alarm State Emissivity
0 degrees C, or 0 degrees F 0 degrees C, or 0 degrees F See Table 3-6 off
0.95
3-6 Raynger 3i Series Operator’s Manual
Note: When you simultaneously press the MODE and ACTIVATE buttons while in the
RUN loop, the instrument is reset to its RUN loop factory default settings (stored data is unaffected). In the LOG loop, pressing the MODE and ACTIVATE buttons not only resets the instrument to its LOG loop factory default settings, but also clears all previ­ously stored data. In both cases, the instrument will “beep” after reinstalling the default settings. Section 3.6 lists default setting for each model as well as non-model­specific default settings and range values.
3.7 REGULATORY
The appropriate regulatory approvals and certificates have been issued as follows:
FDA Class II Laser Certification
FDA Class IIIa Laser Certification
IEC Class 2 Laser Certification
IEC801-3 (EN50082) for EMI susceptibility
Raynger 3i Series Operator’s Manual 4-1
4.0 MAINTENANCE
4.1 BATTERY REPLACEMENT
When the battery icon comes on, you need to replace the batteries. To open the battery compartment, press gently on the middle of the sliding panel (located on the bottom of the handle) and slide it to the rear of the unit. Remove the batteries and replace with four AA batteries. Be sure to insert the new batteries so they point in the proper direction. (A symbol on the side of the handle shows the proper battery orien­tation.) Replace the battery compartment cover.
4.2 CLEANING
You should periodically clean the instrument’s front window and housing.
4.2.1 Front Window Cleaning
A dirty front window can cause temperature measurement errors. The window is fragile, and care should be taken when cleaning it to prevent scratching. Use camera lens or eye glass tissues to clean the window.
Periodic cleaning can be done by completing one or more of the following:
Blow loose particles off the front window with clean air.
Note: Unfiltered compressed air, as well as your own breath, can cause condensation on
the front window, which can trap dust particles instead of removing them.
Gently brush off particles with a soft camel-hair brush.
Thicker contaminants can be cleaned with water and a camera lens tissue.
For finger prints or grease, use either isopropyl alcohol, Ethanol, Xylene, Acetone, or Kodak
®
Lens Cleaner. Apply to the front window, and wipe gen­tly with a camera lens tissue until you see colors on the surface, then allow to air dry. Do not wipe the surface dry, as this may scratch the surface.
CAUTION
Do not use any ammonia, or cleaners with ammonia, bleach, acids, or strong bases. This can cause severe damage to the front window.
4-2 Raynger 3i Series Operator’s Manual
4.2.2 Cleaning the Housing
To clean the instrument’s housing, simply use soap and water or a mild commercial cleaner. Wipe with a damp sponge or soft rag. Use a soft rag to gently wipe the dis­play.
4.3 LASER MAINTENANCE
If the laser (laser models only) does not operate properly, call your infrared ther­mometer supplier. DO NOT open the instrument’s main housing.
Raynger 3i Series Operator’s Manual A-1
APPENDIX A: OPTICAL
HOW TO READ THE OPTICAL CHARTS
The optical charts indicate the nominal target spot diameter at various distances from the sensing head. Information on the top line of each chart shows the different spot diameters at the distances from the sensor given on the bottom line (see Figure A-1). All measurements are in millimeters or meters and inches.
Figure A-1: How to Read the Optical Charts
Target spot size
at focus point
Model
Type
LTCL2 LTCL3
Distance from sensor to object in inches or feet
DISTANCE: SENSOR TO OBJECT IN INCHES
0
36
Focus Point
0.3 in @ 24 in
Spot Diameter (in)Spot Diameter (mm)
IR Spot Diameter at Lens = 23 mm (0.9 in) Laser Diameter at Lens = 40 mm (1.6 in)
Focus Point D:S = Distance to spot divided by spot diameter at the focus point Far Field D:S = Ratio at distances greater than 10x the focus distance
0.9
23
8 mm @ 610 mm
0
DISTANCE: SENSOR TO OBJECT IN MILLIMETERS
FOCUS POINT D:S = 75:1 FAR FIELD D:S = 20:1
60
1.9
120
4.6
48
118
300015001000
IR Beam Profile
Laser Profile
180
7.3
203
5000
Diameter of target spot size in inches
Diameter of target spot size in millimeters
Distance from sensor to object in millimeters or meters
How to calculate spot sizes
To calculate a spot size between two known points on an optical chart, use the following formula:
Sx = Sn + x (Sf - Sn)
Sx = Unknown spot size Sn = Smaller known spot size Sf = Larger known spot size Dx = Distance to unknown spot Dn = Distance to smaller known spot Df = Distance to larger known spot
(Dx - Dn) (Df - Dn)
A-2 Raynger 3i Series Operator’s Manual
OPTICAL CHARTS
Note that the optical resolution values are stated at minimum 90% energy (95% for 1M).
Figure A-2: Optical Charts
DISTANCE: SENSOR TO OBJECT IN INCHES
LTDL2 LTDL3
DISTANCE: SENSOR TO OBJECT IN FEET
10
0
Spot Diameter (in)Spot Diameter (mm)
2.0
1.3
20
3.5
5.0
0.9
60
36
1.9
23
48
120
4.6
118
300015001000
IR Beam Profile
Laser Profile
0
Focus Point
LTCL2 LTCL3
305
6.5
40
50
8.0
0.3 IN @ 24 IN
Spot Diameter (in)Spot Diameter (mm)
8 mm @ 610 mm
0
DISTANCE: SENSOR TO OBJECT IN MILLIMETERS
FOCUS POINT D:S = 75:1 FAR FIELD D:S = 20:1
IR Spot Diameter at Lens = 23 mm (0.9 in) Laser Diameter at Lens = 40 mm (1.6 in)
180
7.3
203
5000
33
51
3
0
DISTANCE: SENSOR TO OBJECT IN METERS
FOCUS POINT D:S = 75:1 @15 m (50 ft) FAR FIELD D:S = 68:1
IR Spot Diameter at Lens = 23 mm (0.9 in) Laser Diameter at Lens = 40 mm (1.6 in)
DISTANCE: SENSOR TO OBJECT IN FEET
10
0
0
Focus Point
1.0 IN @ 15 FT
0.80.5
1813
25 mm @ 4.6 m
3
1ML2 1ML3 1MSC
Spot Diameter (in)Spot Diameter (mm)
DISTANCE: SENSOR TO OBJECT IN METERS
FOCUS POINT D:S = 180:1 FAR FIELD D:S = 140:1
IR Spot Diameter at Lens = 7 mm (0.3 in)
1.5
86
124
6
20
305
2.3
91.5
IR Beam Profile
Laser Profile
165
3.2
203
12
15
DISTANCE: SENSOR TO OBJECT IN INCHES
36
1.3
1.0
27
38 mm @ 1900 mm
60
Focus Point
1.5 IN @ 75 IN
32
120
2.9
74
300015001000
0
G5SC
40
50
4.0
Spot Diameter (in)Spot Diameter (mm)
0
DISTANCE: SENSOR TO OBJECT IN MILLIMETERS
180
4.9
137
5000
FOCUS POINT D:S = 50:1 FAR FIELD D:S = 30:1
IR Spot Diameter at Lens = 23 mm (0.9 in)
36
57
6
91.5
80
12
100
15
Raynger 3i Series Operator’s Manual A-3
DISTANCE: SENSOR TO OBJECT IN FEET
10
0
20
30
40
50
LTSC
DISTANCE: SENSOR TO OBJECT IN INCHES
36
0.8
20
60
Focus Point
1.0 IN @ 75 IN
0.9
23
120
2.1
53
0
Spot Diameter (in)Spot Diameter (mm)
25 mm @ 1900 mm
0
300015001000
DISTANCE: SENSOR TO OBJECT IN MILLIMETERS
FOCUS POINT D:S = 75:1 FAR FIELD D:S = 40:1
IR Spot Diameter at Lens = 23 mm (0.9 in)
180
5000
3.6
LRSCL2
Spot Diameter (in)Spot Diameter (mm)
1.9
47
3
0
2.8
71
3.8
95
6
9
4.8
119
12
5.7
143
15
DISTANCE: SENSOR TO OBJECT IN METERS
FOCUS POINT D:S = 105:1 @ 15 m (50 ft) FAR FIELD D:S = 90:1
IR Spot Diameter at Lens = 23 mm (0.9 in)
97
DISTANCE: SENSOR TO OBJECT IN FEET
10
0
20
30
40
50
LRL2 LRL3 LRSC
Spot Diameter (in)Spot Diameter (mm)
1.7
43
2.4
61
3.3
81
4.1
102
5.0
124
2ML2 2ML3 2MSC
DISTANCE: SENSOR TO OBJECT IN FEET
10
0
Focus Point
2.0 IN @ 15 FT
Spot Diameter (in)Spot Diameter (mm)
1.5
38
51 mm @ 4.6 m
3
0
20
3.0
74
30
4.9
122
6
9
DISTANCE: SENSOR TO OBJECT IN METERS
FOCUS POINT D:S = 90:1 FAR FIELD D:S = 60:1
IR Spot Diameter at Lens = 23 mm (0.9 in)
171
6.8
3
0
6
9
12
15
DISTANCE: SENSOR TO OBJECT IN METERS
FOCUS POINT D:S = 120:1 @ 15 m (50 ft) FAR FIELD D:S = 100:1
IR Spot Diameter at Lens = 23 mm (0.9 in)
40
50
8.8
DISTANCE: SENSOR TO OBJECT IN FEET
10 15
218
12
15
P7DL2 P7DL3
0
Spot Diameter (in)Spot Diameter (mm)
5.1
2.8
71
127
0
20
10.0
7.5
208
248
3
5
6
14.5
363
305
91.5
19.5
483
40
12
50
24.0
603
15
DISTANCE: SENSOR TO OBJECT IN METERS
FOCUS POINT D:S = 25:1 @15 m (50 ft) FAR FIELD D:S = 23:1
IR Spot Diameter at Lens = 23 mm (0.9 in) Laser Diameter at Lens = 40 mm (1.6 in)
IR Beam Profile
Laser Profile
A-4 Raynger 3i Series Operator’s Manual
Raynger 3i Series Operator’s Manual B-1
APPENDIX B: OBJECT EMISSIVITY
HOW TO DETERMINE OBJECT EMISSIVITY
Emissivity is a measure of an object’s ability to absorb, transmit, and emit infrared energy. It can have a value from 0.0 (perfect mirror) to 1.0 (blackbody). When setting the emissivity value on your sensor, if you set a higher than actual emissivity value, the output will read low (provided the target temperature is above the local’s ambient temperature). For example, if you set 0.95 and the actual target emissivity is 0.9, the sensor’s reading will be lower than the target’s true temperature (when the target temperature is above ambient).
To determine an object’s emissivity, you can use any one of the following methods:
1. Determine the actual temperature of the object using a sensor such as a Resistance Temperature Device (RTD), thermocouple, or another suitable method. Next, use your portable sensor to measure the object temperature and adjust the emissivity setting until it shows the actual temperature. This is the correct emissivity for the measured material.
2. For objects with relatively low temperatures, up to 260°C ( 500°F), place a piece of tape, such as masking, on the object. Make sure the tape is large enough for the­measurement spot diameter. You do not want the spot to go beyond the tape’s edges. Next, measure the tape temperature using an emissivity setting of 0.95. Finally, measure an adjacent area on the object and adjust the emissivity setting until you reach the same temperature. This is the correct emissivity for the mea­sured material.
3. If a portion of the surface of the object can be coated, use a flat black paint. Next, measure the painted area using an emissivity setting of 0.93. Finally, measure an adjacent area on the object and adjust the emissivity setting until you reach the same temperature. This is the correct emissivity for the measured material.
B-2 Raynger 3i Series Operator’s Manual
TYPICAL EMISSIVITY VALUES
The following tables provide references for estimating emissivity and can be used when none of the previous three determining steps are practical. Emissivity values shown in the tables are only approximate. Any or all of the following parameters can affect the emissivity of an object:
1. Temperature
2. Angle of measurement
3. Geometry (plane, concave, convex, etc.)
4. Thickness
5. Surface quality (polished, rough, oxidized, sandblasted)
6. Spectral region of measurement
7. Transmissivity (i.e., thin film plastics)
Table B-1: Emissivity Values for Metals
Material Emissivity
1.0µm 1.6µm 8-14µm
Aluminum
Unoxidized 0.1-0.2 0.02-0.2 n.r. Oxidized 0.4 0.4 0.2-0.4 Alloy A3003,
Oxidized n.r. 0.4 0.3 Roughened 0.2-0.8 0.2-0.6 0.1-0.3 Polished 0.1-0.2 0.02-0.1 n.r.
Brass
Polished 0.8-0.95 0.01-0.05 n.r. Burnished n.r. n.r. 0.3 Oxidized 0.6 0.6 0.5
Chromium 0.4 0.4 n.r. Copper
Polished n.r. 0.03 n.r. Roughened n.r. 0.05-0.2 n.r. Oxidized 0.2-0.8 0.2-0.9 0.4-0.8 Electrical Terminal Blocks n.r. n.r. 0.6
Gold 0.3 0.01-0.1 n.r. Haynes
Alloy 0.5-0.9 0.6-0.9 0.3-0.8
Inconel
Oxidized 0.4-0.9 0.6-0.9 0.7-0.95 Sandblasted 0.3-0.4 0.3-0.6 0.3-0.6 Electropolished 0.2-0.5 0.25 0.15
Continued on next page
n.r. = not recommended
Raynger 3i Series Operator’s Manual B-3
Material Emissivity
1.0µm 1.6µm 8-14µm
Iron
Oxidized 0.4-0.8 0.5-0.9 0.5-0.9 Unoxidized 0.35 0.1-0.3 n.r. Rusted n.r. 0.6-0.9 0.5-0.7 Molten 0.35 0.4-0.6 n.r.
Iron, Cast
Oxidized 0.7-0.9 0.7-0.9 0.6-0.95 Unoxidized 0.35 0.3 0.2 Molten 0.35 0.3-0.4 0.2-0.3
Iron, Wrought
Dull 0.9 0.9 0.9
Lead
Polished 0.35 0.05-0.2 n.r. Rough 0.65 0.6 0.4
Oxidized n.r. 0.3-0.7 0.2-0.6 Magnesium 0.3-0.8 0.05-0.3 n.r. Mercury n.r. 0.05-0.15 n.r. Molybdenum
Oxidized 0.5-0.9 0.4-0.9 0.2-0.6
Unoxidized 0.25-0.35 0.1-0.35 0.1 Monel (Ni-Cu) 0.3 0.2-0.6 0.1-0.14 Nickel
Oxidized 0.8-0.9 0.4-0.7 0.2-0.5
Electrolytic 0.2-0.4 0.1-0.3 n.r. Platinum
Black n.r. 0.95 0.9 Silver n.r. 0.02 n.r. Steel
Cold-Rolled 0.8-0.9 0.8-0.9 0.7-0.9
Ground Sheet n.r. n.r. 0.4-0.6
Polished Sheet 0.35 0.25 0.1
Molten 0.35 0.25-0.4 n.r.
Oxidized 0.8-0.9 0.8-0.9 0.7-0.9
Stainless 0.35 0.2-0.9 0.1-0.8 Tin (Unoxidized) 0.25 0.1-0.3 n.r. Titanium
Polished 0.5-0.75 0.3-0.5 n.r.
Oxidized n.r. 0.6-0.8 0.5-0.6 Tungsten n.r. 0.1-0.6 n.r.
Polished 0.35-0.4 0.1-0.3 n.r. Zinc
Oxidized 0.6 0.15 0.1
Polished 0.5 0.05 n.r.
Material Emissivity
Table B-1 (continued): Emissivity Values for Metals
n.r. = not recommended
B-4 Raynger 3i Series Operator’s Manual
1.0µm 5.0µm 7.9µm 8-14µm
Asbestos 0.9 0.9 0.95 0.95 Asphalt n.r. 0.95 0.95 0.95 Basalt n.r. 0.7 0.7 0.7 Carbon
Unoxidized 0.8-0.95 0.8-0.9 0.8-0.9 0.8-0.9
Graphite 0.8-0.9 0.7-0.9 0.7-0.8 0.7-0.8 Carborundum n.r. 0.9 0.9 0.9 Ceramic 0.4 0.85-0.95 0.95 0.95 Clay n.r. 0.85-0.95 0.95 0.95 Concrete 0.65 0.9 0.95 0.95 Cloth n.r. 0.95 0.95 0.95 Glass
Plate n.r. 0.98 0.85 0.85
“Gob” n.r. 0.9 n.r. n.r. Gravel n.r. 0.95 0.95 0.95 Gypsum n.r. 0.4-0.97 0.8-0.95 0.8-0.95 Ice n.r. 0.98 0.98 Limestone n.r. 0.4-0.98 0.98 0.98 Paint (non-Al.) 0.9-0.95 0.9-0.95 Paper (any color) n.r. 0.95 0.95 0.95 Plastic (opaque, n.r. 0.95 0.95 0.95
over 20 mils) n.r. Rubber n.r. 0.9 0.95 0.95 Sand n.r. 0.9 0.9 0.9 Snow n.r. 0.9 0.9 Soil n.r. 0.9-0.98 0.9-0.98 Water n.r. 0.93 0.93 Wood, Natural n.r. 0.9-0.95 0.9-0.95 0.9-0.95
To optimize surface temperature measurement accuracy consider the following:
1. Determine the object emissivity for the spectral range of the instrument to be used
for the measurement.
2. Avoid reflections by shielding object from surrounding high temperature sources.
3. For higher temperature objects use shorter wavelength instruments, whenever
possible.
4. For semi-transparent materials such as plastic film and glass, assure that the back-
ground is uniform and lower in temperature than the object.
5. Hold instrument perpendicular to surface whenever emissivity is less than 0.9. In
all cases, do not exceed angles more than 30 degrees from incidence.
6. For 1M and 2M models, avoid measurements in high ambient light conditions (see
Page 2-11).
Table B-2: Emissivity Values for Non-Metals
n.r. = not recommended
Raynger 3i Series Operator’s Manual C-1
APPENDIX C: TROUBLESHOOTING
The portable instrument is equipped with a failsafe program that lets you know if it is over or under range while in the MAX, MIN, DIF, or AVG modes. The failsafe pro­gram also lets you know the instrument has a failed component. All failsafe informa­tion displays on the control panel and is sent to the RS-232 and analog outputs. Figure C-1 shows how the control panel displays the over range and under range fail­safe codes.
MAX Mode: If the object’s temperature is over range, the panel’s main and mode dis-
plays and RS-232 output are sent the -O- failsafe code. The analog out­put is sent the maximum temperture for your model.
If the object’s temperature is under range, the panel’s main display and RS-232 output are sent the -U- failsafe code. The mode display and RS­232 ouput are sent either the correct MAX temperature or the -U- failsafe code. The analog output is sent the minimum temperature for your model.
RCL
RUN
RUN
LOG
RCL
SET
°C °F
RUN
LOG
SET
°C °F
LOG
LOG
SET
HI
RECALL
MODE
MODE
RECALL
( )
LO
LASER
LASER
( )
T
amb
ACTIVATE
ACTIVATE
LOCK
LOCK
LOG
SET
HI
MODE
RECALL
( )
LO
LASER
( )
T
amb
ACTIVATE
LOCK
C-2 Raynger 3i Series Operator’s Manual
MIN Mode: If the object’s temperature is over range, the panel’s main display and
RS-232 output are sent the -O- failsafe code. The mode display and RS­232 ouput are sent either the correct MIN temperature or the -O- failsafe code. The analog output is sent the maximum temperature for your model.
If the object’s temperature is under range, the panel’s main and mode displays and RS-232 output are sent the -U- failsafe code. The analog output is sent the minimum temperture for your model.
DIF Mode: If the object’s temperature is over range, the panel’s main and mode dis-
plays and RS-232 output are sent the -O- failsafe code. The analog out­put is sent the maximum temperture for your model.
If the object’s temperature is under range, the panel’s main and mode displays and RS-232 output are sent the -U- failsafe code. The analog output is sent the minumum temperture for your model.
AVG Mode: If the object’s temperature is over range, the panel’s main and mode dis-
plays and RS-232 output are sent the -O- failsafe code. The analog out­put is sent the maximum temperture for your model.
If the object’s temperature is under range, the panel’s main and mode displays and RS-232 output are sent the -U- failsafe code. The analog output is sent the minumum temperture for your model.
All Modes: If the internal temperature is less than 0°C or greater than 50°C, the pan-
el’s main and mode displays and the RS232 output are sent the -U- or ­O- failsafe code respectively. The analog output is sent the minimum or maximum temperature for your model.
Note: On laser models, if the internal temperature is less than 0°C (32°F) or greater than
45°C (113°F), the laser will automatically turn off. The laser will also automatically turn off if the battery is too low (when the low battery icon displays).
Raynger 3i Series Operator’s Manual C-3
FAILED COMPONENT
If the microprocessor fails, the panel’s main and mode displays and RS-232 output are sent a 7777 failsafe code (see Figure C-2). The analog output is sent the maximum temperature for your model.
NOISY OR RANDOM READINGS OR SYSTEM LOCKUP
Noisy or random readings or a system lockup may be caused by electromagnetic interference from nearby unshielded equipment. To clear the instrument’s system, release then press the trigger. If this does not clear the problem, slide the battery door off then back on. (This disconnects the battery power.)
Figure C-2: Failed Component Failsafe Code
RCL
RUN
RUN
LOG
LOG
SET
LOG
HI
SET
°C °F
( )
( )
T
LO
amb
RECALL
MODE
MODE
RECALL
LASER
LASER
ACTIVATE
ACTIVATE
LOCK
LOCK
C-4 Raynger 3i Series Operator’s Manual
APPENDIX D: OPTIONS AND ACCESSORIES
ACCESSORIES
A full range of accessories for various applications and industrial environments is available. Accessories may be ordered at any time and added on-site. The following accessories are available:
Hard shell carrying case with die-cut foam interior
Variable brightness filter (for scope models)
DataTemp 2
software for real-time graphic temperature display and datalog-
ger downloading
Portable/battery operated printer
110 V/60 Hz or 220 V/50 Hz voltage adaptors (DIN VDE 0551 approved)
Computer cable 1.5 m (5 ft) w/9-pin connector and mini plug
Printer cable 1.5 m (5 ft) w/25-pin connector and mini plug
Analog cable 1.5 m (5 ft) w/banana plugs and mini plug
Replacement operator’s manual
Laser glasses
OPTIONS
A full range of options for various applications and environments is available. Options are factory installed and must be ordered with base model units. The follow­ing options are available:
NIST certification
Raynger 3i Series Operator’s Manual D-1
D-2 Raynger 3i Series Operator’s Manual
Raynger 3i Series Operator’s Manual E-1
APPENDIX E: TRACEABILITY OF INSTRUMENT CALIBRATION
The temperature sources (blackbodies) used to calibrate this instrument are traceable to the U.S. National Institute of Standards and Technology (NIST).
The calibration sources for this instrument were certified by a NIST certified calibra­tion laboratory and are traceable to NIST primary standards. The certificate describes the equipment used for calibration and any corresponding NIST report numbers. In addition, the certificate lists test accuracy data and the next calibration date.
NIST certificates are available as an option (must be ordered with the instrument). Contact the manufacturer (not NIST) to order this option.
Figure E-1: Traceability Flowchart
National Institute of Standards and Technology
(NIST)
Manufacturer
High Temperature Models
1200 - 3000°C (2200 - 5400°F)
Certified Radiation Transfer
Standard
Calibration Source
(Blackbody)
Emissivity 1.0
NIST Certified Calibration Lab: calibration in
accordance with MIL-STD-45662 and against
standards traceable to NIST
Medium Temperature Models
200 - 1200°C (400 - 2200°F)
Certified RTD
Instrumentation & NIST
Calibrated Radiation Transfer
Standard Instrument
Calibration Source
(Blackbody)
Emissivity < 1
Calibrated Product
Low Temperature Models
-18 - 870°C (0 - 1600°F) Certified RTD
Instrumentation
Measured
Source
Emissivity
E-2 Raynger 3i Series Operator’s Manual
Raynger 3i Series Operator’s Manual F-1
APPENDIX F: CE CONFORMITY FOR THE EUROPEAN
COMMUNITY
This instrument conforms to the following standards:
EN50081-1:1992, Electromagnetic Emissions
EN50082-1:1992, Electromagnetic Susceptibility
Tests were conducted over a frequency range of 27–500 MHz with the instrument in three orientations. The instrument’s average error in this frequency range for the three orientations is 3.1° C at an electric field strength of 3 V/m. At some frequencies the instrument may not meet its stated accuracy.
F-2 Raynger 3i Series Operator’s Manual
Raynger 3i Series Operator's Manual Glossary-1
Absolute Zero
The temperature (0 Kelvin) of an object defined by the theoretical condition where the object has zero energy.
Accuracy
Maximum deviation, expressed in tempera­ture units, or as a percentage of the tempera­ture reading, or as a percentage of the full scale temperature value, or as a percentage of the target temperature, indicating the differ­ence between a temperature reading given by an instrument under ideal operating condi­tions, and the temperature of a calibration source (per the ASTM standard test method E 1256-88).
Activate
Pressing the activate button (available on vari­ous units) while in different modes can acti­vate alarms and TAM, or toggle between digi­tal and analog outputs or between °C and °F.
Ambient Derating
Refer to Temperature Coefficient.
Ambient Operating Range
Range of the ambient temperature conditions over which the thermometer is designed to operate.
Ambient Temperature
Room temperature or temperature surround­ing the instrument.
Ambient Temperature Compensation (TAMB)
Refer to Reflected Energy Compensation.
ASTM
American Society for Testing and Materials.
Atmospheric Windows
Infrared spectral bands in which the atmos­phere best transmits radiant energy. Two pre­dominant windows are located at 2-5 µm and at 8-14 µm.
AVG
Average. Units with this feature have the abil­ity to display the average of measured temper­atures. AVG is a weighted average of all read­ings taken since the trigger was pulled.
Background Temperature
Temperature behind and surrounding the target, as viewed from the instrument.
Blackbody
A perfect emitter; an object that absorbs all the radiant energy incident on it at all wave­lengths and reflects and transmits none. A surface with emissivity of unity (1.00).
°C (Celsius)
Temperature scale based on 0° (zero degrees) as the freezing point of water, and 100° as the vaporization point of water, at standard pressure. °C = (°F - 32) ÷ 1.8.
Calibration
A methodical measurement procedure to determine all the parameters significantly affecting an instrument’s performance.
Calibration Source
A source (blackbody, hot plate, etc.) of known and traceable temperature and emis­sivity. Usually NIST traceable in the USA, with other recognized standards available for international customers.
Colored Body
See Non-Gray Body.
D:S
Distance to size ratio. See Optical Resolution.
Detector
Transducer which produces a voltage or cur­rent proportional to the IR energy incident upon it. See also thermopile, pyroelectric, and Si detectors.
Dielectric Withstand Voltage (Breakdown Voltage)
The voltage up to which a dielectric (insula­tor) can endure before conduction through the material occurs.
DIF
Differential. When this mode is activated, the difference between the temperatures of the last series of measured temperatures dis­plays.
Glossary Of Terms
Glossary-2 Raynger 3i Series Operator's Manual
DIG/ANA
Digital/Analog switch. On units with this feature, a user can switch between digital and analog outputs.
Digital Data Bus
A means for transmitting coded digital data on a common buss in accordance with a stan­dard format such as RS-232 or IEEE-488.
Digital Output Interval (DOI)
The time interval between variable length digital message transmissions containing temperature and system status information.
DIN
Deutsches Institut für Normung (DIN). The German standard for many instrumentation products.
Display Resolution
The level of precision to which a temperature value can be displayed, usually expressed in degrees or tenths of degrees.
Drift
The change in instrument indication over a long period of time, not caused by external influences on the device (per the ASTM stan­dard test method E 1256-88).
EMC
Electro-Magnetic Compatibility. Resistance to electrical signal disturbances within IR thermometers.
Emissivity
The ratio of infrared energy radiated by an object at a given temperature and spectral band to the energy emitted by a perfect radi­ator (blackbody) at the same temperature and spectral band. The emissivity of a per­fect blackbody is unity (1.00).
EMI/RFI Noise
Electro-Magnetic Interference/Radio Frequency Interference. EMI and RFI may cause disturbances to electrical signals with­in IR thermometers. EMI and RFI noise is most commonly caused by devices with switching motors (air conditioners, power tools, refrigeration systems, etc.).
°F (Fahrenheit)
Temperature scale where °F = (°C x 1.8) + 32 = °R - 459.67.
Far Field
A measured distance substantially greater than the focus distance of the instrument; typically greater than 10 times the focus dis­tance.
Field of View (FOV)
The region, at the target, measured by the IR thermometer. Typically presented by giving the spot diameter as a function of distance from the instrument. Also presented as the angular size of the spot at the focus point. See Optical Resolution.
Filter (Optical)
Refer to Spectral Filter or Neutral Density Filter.
Focus Point (or Distance)
The distance from the instrument where the optical resolution is greatest.
Full Scale
The maximum of the temperature range or output signal.
Full Scale Accuracy
A convention for expressing the accuracy as a percentage of an instrument’s (highest) full scale temperature.
Gray Body
A radiating object whose emissivity is in con­stant ratio (not unity) at all wavelengths to that of a blackbody at the same temperature, and does not transmit infrared energy.
HAL
High Alarm. Units with this feature can sound an alarm when they sense that a user­defined high temperature has been reached.
Hertz (Hz)
Units in which frequency is expressed. Synonymous with cycles per second.
IEC
International Electrotechnical Commission; a European organization that coordinates and sets standards among the European Community.
IEEE-488
A communications format standard. See dig­ital data bus.
Raynger 3i Series Operator's Manual Glossary-3
Infrared (IR)
The portion of the electromagnetic spectrum extending from the far red visible at approximately 0.75 µm, out to 1000 µm. However, because of instrument design considerations and the atmospheric windows, most infrared measurements are made between 0.75 µm and 20 µm.
Infrared Thermometer
An instrument that converts incoming IR radiation from a spot on a target surface to a measurement value that can be related to the temperature of that spot.
Insulation Resistance
The electrical resistance of the insulating material as measured by the ratio of the applied voltage (applied for example between a conducting wire and the case or chassis ground) to the leakage current, and normally expressed in megohms.
Intrinsically Safe
A standard for preventing explosions in haz­ardous areas by limiting the electrical energy available to levels that are insufficient to cause ignition of explosive atmospheres dur­ing normal operation.
J,K t/c
Thermocouple types available, depending upon the temperature range to be measured.
JIS
Japanese Industrial Standard; sets standards for determining or establishing the accuracy of an IR thermometer.
K (Kelvin)
The unit of absolute or thermodynamic tem­perature scale where 0 K is absolute zero and
273.15 K is equal to 0° C. There is no (°) sym­bol used with the Kelvin scale, and K = °C + 273.15.
LAL
Low Alarm. Units with this feature can sound an alarm when they sense that a user­defined low temperature has been reached.
Laser
Single or dual lasers are used in some units for aiming and/or locating the optimum tem­perature measurement point.
LOC
Location. Units with the datalogging feature store data in numbered locations, which can be recalled and reviewed on the display when necessary.
LOCK
Locks or unlocks the trigger on portable instruments.
LOG
Datalogging. Units with the datalogging fea­ture can store data when in the LOG mode.
Loop
A cycle of operations within a selected mode. For example, the RUN Loop cycles through standard running operations; the LOG Loop cycles through datalogging operations; and the RECALL Loop cycles through stored operations and displays the data.
mA
Milli-ampere, or 0.001 amp.
MAX
Maximum. When this mode is activated, the maximum temperature of the last series of measured temperatures displays.
Micron (or µm)
10
-6
meters (m), or 0.000001 m.
MIN
Minimum. When this mode is activated, the minimum temperature of the last series of measured temperatures displays.
Minimum spot size
The smallest spot an instrument can accu­rately measure.
Mode
Modes are various user-selectable operations within Loops.
mV
Milli-volts, or 0.001 volt.
NET
Noise Equivalent Temperature. Peak to peak system electrical noise normally measured at the output (display or analog) expressed in °F or °C.
Glossary-4 Raynger 3i Series Operator's Manual
Neutral Density Filter
An optical element used to restrict the amount of energy reaching an instrument’s detector by ideally attenuating the energy at all wavelengths by the same amount.
NIST Traceability
Calibration in accordance with and against standards traceable to NIST (National Institute of Standards and Technology, USA). Traceability to NIST is a means of ensuring that reference standards remain valid and their calibration remains current.
Non-Gray Body
A radiating object that is partly transparent to infrared (transmits infrared energy at cer­tain wavelengths); also called Colored Bodies. Glass and plastic films are examples of non-gray bodies.
Optical Pyrometer
A system that, by comparing a source whose temperature is to be measured to a standard­ized source of illumination (usually com­pared to the human eye), determines the temperature of the former source.
Optical Resolution
The distance to size ratio (D:S) of the IR mea­surement spot, where the distance is usually defined at the focus distance, and the size is defined by the diameter of the IR energy spot at the focus (typically at the 90% IR energy spot diameter). Optical resolution may also be specified for the far field by using values of far field distance and spot size.
Pyroelectric Detector
Infrared detector which behaves as a current source with an output proportional to the rate of change of the incident IR energy.
°R (Rankine)
Temperature scale where °R = 1.8 x K, or also °R=°F+460.
Radiation Thermometer
A device that calculates an object’s tempera­ture (given a known emissivity) from mea­surement of either visible or infrared radia­tion from that object.
Recall (RCL)
When the Recall loop is activated, stored val­ues can be recalled from either the RUN or LOG loops.
Reference Junction
Refers to the “cold” or ambient thermocouple junction, which is held at a known tempera­ture. Also see thermocouple.
Reflectance
The ratio of the radiant energy reflected off a surface to that incident on the surface; for a gray body this is equal to unity minus emit­tance; for a perfect mirror this approaches unity; and for a blackbody the reflectance is zero.
Reflected Energy Compensation
Correction feature used to achieve greater accuracy when, due to a high uniform background temperature, IR energy is reflected off the target into the instrument. If the background temperature is known the instrument reading can be cor­rected by using this feature.
Relative Humidity
The ratio, expressed as a percent, of the amount of water vapor actually present in a sample of air to the greatest amount of water vapor possible at the same temperature.
Repeatability
The degree to which a single instrument gives the same reading on the same object over successive measures under the same ambient and target conditions (per the ASTM standard test method E 1256-88).
Resolution
See Temperature Resolution or Optical Resolution.
Response Time
A measure of an instrument’s change of out­put corresponding to an instantaneous change in target temperature, generally expressed in milli-seconds, for 95 percent of full scale temperature indication (per the ASTM standard test method E 1256-88). The specification for Raytek instruments also includes the average time required for soft­ware computations.
Raynger 3i Series Operator's Manual Glossary-5
RS-232 (1-way)
A 1-way transfer of digital information from a digital output. RS-232 is a standardized format for asynchronous serial data transfer.
RTD
Resistance Temperature Device. A contact measurement device whose resistance varies with temperature.
RUN
The RUN Loop is used when standard spot measurements need to be taken.
Scatter
See Size of Source Effect.
Setpoint
Temperature setting which when crossed by the actual temperature value will trigger an event and/or cause a relay to change state.
Setup (SET)
When in the Setup Loop, you can set values (emissivity, alarms, etc.) in either the RUN or LOG Loops.
Shock Test
An impact test per MIL-STD-810D where a force is applied to any axis of an object over a specified duration. The force is usually mea­sured in g’s (1g = 9.81m/s2= 32.2 ft./s2), and the duration is usually measured in msec.
Silicon (Si) Detector
A photodiode detector typically used in high temperature IR thermometers.
Size of Source Effect
An undesirable increase in temperature read­ing caused by IR energy outside the spot reaching the detector. The effect is most pro­nounced when the target is much larger than the field of view.
Sleep
Units with this feature go to “sleep” either immediately or after a period of inactivity (depending on the Loop). Sleep is a period of low power consumption, no display, no laser, and no backlight.
Spectral Filter
An optical element used to restrict the spec­tral band of energy reaching an instrument’s detector.
Spectral Response
The wavelength region in which the IR Thermometer is sensitive.
Spot
The diameter of the area on the target where the temperature determination is made. The spot is defined by the circular aperture at the target which allows typically 90% of the IR energy to be collected by the instrument, as compared with the 100% spot diameter which is defined by the IR energy collected from a very large target. The actual size and distance to the target for the 100% spot diam­eter is specified in the calibration procedure for each instrument.
Stare
A saturation effect caused by aiming a sensor at a “hot” target for an extended period of time and then quickly aiming at a target at a “lower” temperature. The increase in time (beyond the normal system response) for the sensor to return to within 5 % of the lower temperature is defined as the “stare” time.
Storage Temperature Range
Ambient temperature range that the ther­mometer can safely withstand in a non-oper­ating mode, and subsequently, operate with­in published performance specifications.
TAM
T-ambient (Ambient temperature compensa­tion). Targets that have low emissivities will reflect energy from nearby objects, which may result is inaccurate readings. Sometimes objects near the target (machines, furnaces, or other heat sources) have a tem­perature much higher than that of the target. In these situations it is necessary to compen­sate for the reflected energy from those objects. (TAM has no effect if the emissivity is 1.0.)
Target
The object upon which the temperature determination is intended to be made.
Temperature
A degree of hotness or coldness of an object measurable by a specific scale; where heat is defined as thermal energy in transit, and flows from objects of higher temperature to objects of lower temperature.
Glossary-6 Raynger 3i Series Operator's Manual
Temperature Coefficient (or Ambient Derating)
An indication of the instruments ability to maintain accuracy when the ambient condi­tions are subject to a slow change or drift. The temperature coefficient is usually expressed as the percent change in accuracy per degree change in ambient temperature. For a rapid change in ambient conditions refer to Thermal Shock.
Temperature Resolution
The minimum simulated or actual change in target temperature that gives a usable change in output and/or indication (per the ASTM standard test method E 1256-88).
Thermal Shock
A short term error in accuracy caused by a transient ambient temperature change. The instrument recovers from its accuracy error when it comes back into equilibrium with the new ambient conditions.
Thermistor
A semiconductor material whose resistivity changes with temperature.
Thermocouple
A junction, comprising two dissimilar met­als, that develops a small voltage dependent upon the temperature of the junction. Typical thermocouples types include:
J iron/constantan K chromel/alumel T copper/constantan E chromel/constantan R platinum/platinum-30% rhodium S platinum/platinum-10% rhodium B platinum-6% rhodium/platinum-30%
rhodium G tungsten/tungsten-26% rhenium C tungsten-5% rhenium/tungsten-26%
rhenium D tungsten-3% rhenium/tungsten-25%
rhenium
Thermopile
An arrangement of thermocouples in series such that the alternate junctions are at the measuring temperature and at the reference temperature. This arrangement increases the thermoelectric voltage.
Time Constant
The time it takes for a sensing element to respond to 63.2% of a step change at the tar­get.
Transfer Standard
A precision radiometric measurement instru­ment with NIST traceable calibration in the USA (with other recognized standards avail­able for international customers), used to cal­ibrate radiation reference sources.
Transmittance
The ratio of IR radiant energy transmitted through an object to the total IR energy received by the object for any given spectral range; the sum of emittance, reflectance, and transmittance is unity.
Warm-Up Time
Time, after turn on, until the instrument will function within specified repeatability (per the ASTM standard test method E 1256-88).
Verification
Confirmation of accuracy.
Vibration Test
An oscillatory or repetitive motion test of the instrument per MIL-STD-810D or IEC 68-2-6, often specified as an acceleration in g’s (1g =
9.81m/s2= 32.2 ft./s2), over a frequency range typically measured in Hertz (sec-1).
Loading...