RayTek 3i User Manual

Rev H
6/98
56700-1
RAYNGER
®
3i
SERIES
OPERATOR’S MANUAL
°F
°C
T
E
S
L
C
R
G
O
L
amb
T
LOG
DC
OUT
( )
O
L
( )
I
H
RECALL
MODE
ACTIVATE
LOCK
R
E
S
A
L
SET
WARRANTY
Raytek warrants each instrument it manufactures to be free from defects in material and work­manship under normal use and service for the period of one year from date of purchase. This warranty extends only to the original purchaser. This warranty shall not apply to fuses, batteries, or any product which has been subject to misuse, neglect, accident, or abnormal conditions of operation.
In the event of failure of a product covered by this warranty, Raytek will repair the instrument when it is returned to an authorized Service Facility within one year of the original purchase, provided the warrantor’s examination discloses to its satisfaction that the product was defective. The warrantor may, at its option, replace the product in lieu of repair. With regard to any instru­ment returned within one year of the original purchase, said repairs or replacement will be made without charge. If the failure has been caused by misuse, neglect, accident, or abnormal condi­tions of operation, repairs will be billed at nominal cost. In such cases, an estimate will be sub­mitted before work is started, if requested.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MER­CHANTABILITY, FITNESS, OR ADEQUACY FOR ANY PARTICULAR PURPOSE OR USE. RAYTEK SHALL NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL, OR CONSEQUEN­TIAL DAMAGES, WHETHER IN CONTRACT, TORT, OR OTHERWISE.
WORLDWIDE HEADQUARTERS
Raytek Corporation 1201 Shaffer Road, PO Box 1820 Santa Cruz, CA 95061-1820 USA
Portable Products Division
800 866 5478 Phone 408 458 1110 Fax 408 425 4561
WORLD WIDE WEB www.raytek.com
EUROPEAN HEADQUARTERS
Raytek GmbH Arkonastrasse 45-49 D-13189 Berlin, Germany Phone 49 30 478 0080 FAX 49 30 471 0251
Offices worldwide:
BRAZIL
Raytek do Brasil Sorocaba, Brasil Phone 55 152 276556
CHINA
Raytek China Company Beijing, China Phone 86 10 437 0284
JAPAN
Raytek Japan, Inc. Tokyo, Japan Phone 03 5976 1531
MEXICO
Raytek de Mexico Puebla, Mexico Phone 52 22 30 4380
UNITED KINGDOM
Raytek UK Milton Keynes, UK Phone 44 1 908 630800
® Raytek and Raynger are registered trademarks and 3i is a trademark of Raytek Corporation © 1997 Copyright Raytek Corporation
Raynger 3i Series Operator's Manual Table of Contents
TABLE OF CONTENTS
SECTION PAGE
1.0 INTRODUCTION.............................................................................1-1
1.1 DESCRIPTION.......................................................................................................1-1
1.2 INVENTORY..........................................................................................................1-2
1.3 MODEL IDENTIFICATION.................................................................................1-3
2.0 OPERATION ......................................................................................2-1
2.1 QUICK START........................................................................................................2-2
2.2 PRINCIPLES OF OPERATION............................................................................2-3
2.3 YOUR PORTABLE INFRARED THERMOMETER...........................................2-4
2.4 OPERATION AND CONTROLS.........................................................................2-6
2.4.1 Battery/AC Adaptor Installation...................................................................2-6
2.4.2 Control Panel and Display..............................................................................2-7
2.4.3 Control System ...............................................................................................2-10
2.4.3.1 Control Loops ..............................................................................................2-10
2.4.3.2 General Information....................................................................................2-11
2.4.4 RUN Loop—To Measure Temperature .......................................................2-12
2.4.5 LOG Loop—To Measure and Store Temperature......................................2-14
2.4.6 RECALL Loop—To Recall Measured Temperature ..................................2-16
2.4.6.1 RECALL Values from RUN .......................................................................2-16
2.4.6.2 RECALL Values from LOG........................................................................2-17
2.4.7 SETUP Loop—To Setup and Activate Alarms and Features...................2-19
2.4.7.1 SETUP Values for RUN ..............................................................................2-19
2.4.7.2 SETUP Values for LOG...............................................................................2-21
2.5 USING A TRIPOD...............................................................................................2-22
2.6 DATA OUTPUTS .................................................................................................2-23
2.6.1 Digital Output.................................................................................................2-24
2.6.1.1 Data Output—RUN Loop..........................................................................2-24
2.6.1.2 Data Output—SET Loop............................................................................2-25
2.6.1.3 Data Output—LOG Run Loop..................................................................2-26
2.6.1.4 Data Output—LOG SETUP Loop.............................................................2-27
2.6.2 Analog Output................................................................................................2-28
SECTION PAGE
Table of Contents Raynger 3i Series Operator's Manual
2.7 SIGHTING SYSTEMS..........................................................................................2-29
2.7.1 Single Laser Sighting .....................................................................................2-31
2.7.2 Dual Laser Sighting........................................................................................2-31
2.7.3 Crossed Laser Sighting..................................................................................2-32
2.7.4 Scope Sighting ................................................................................................2-32
3.0 SPECIFICATIONS ............................................................................3-1
3.1 THERMAL..............................................................................................................3-1
3.2 OPERATIONAL.....................................................................................................3-2
3.3 ELECTRICAL .........................................................................................................3-3
3.4 ENVIRONMENTAL..............................................................................................3-4
3.5 PHYSICAL..............................................................................................................3-4
3.6 DEFAULT VALUES ...............................................................................................3-5
3.7 REGULATORY.......................................................................................................3-6
4.0 MAINTENANCE...............................................................................4-1
4.1 BATTERY REPLACEMENT.................................................................................4-1
4.2 CLEANING ............................................................................................................4-1
4.2.1 Front Window Cleaning..................................................................................4-1
4.2.2 Cleaning the Housing......................................................................................4-2
4.3 LASER MAINTENANCE.....................................................................................4-2
SECTION PAGE
Raynger 3i Series Operator's Manual Table of Contents
APPENDIX A: OPTICAL...........................................................................A-1
HOW TO READ THE OPTICAL CHARTS...................................................................A-1
OPTICAL CHARTS ..........................................................................................................A-2
APPENDIX B: OBJECT EMISSIVITY.......................................................B-1
HOW TO DETERMINE OBJECT EMISSIVITY.............................................................B-1
TYPICAL EMISSIVITY VALUES.....................................................................................B-2
APPENDIX C: TROUBLESHOOTING ....................................................C-1
APPENDIX D: OPTIONS AND ACCESSORIES...................................
D-1
APPENDIX E: TRACEABILITY OF
INSTRUMENT CALIBRATION...............................................................E-1
APPENDIX F: CE CONFORMITY FOR THE
EUROPEAN COMMUNITY......................................................................F-1
GLOSSARY OF TERMS
INDEX
Table of Contents Raynger 3i Series Operator's Manual
Raynger 3i Series Operator’s Manual 1-1
1.0 INTRODUCTION
1.1 DESCRIPTION
The Raynger
®3i™
series of instruments are portable infrared temperature measure­ment devices. Each model is rugged and easy to use for making fast, noncontact, non­destructive temperature measurements. They can measure operating temperatures of mechanical, electrical, or production equipment without removing the equipment from service. They can also measure product temperatures during manufacturing or storage without contaminating or marring the product. Table 1-1 lists the standard Raynger 3i models.
Table 1-1: Raynger 3i Models
TEMPERATURE OPTICAL SPECTRAL MODEL RANGE RESOLUTION RANGE SIGHTING
LTDL2 & LTDL3
LTSC
LTCL2 & LTCL3
-30 to 1200°C
(-20 to 2200°F)
-30 to 1200°C
(-20 to 2200°F)
-30 to 1200°C
(-20 to 2200°F)
75:1
75:1
75:1
8-14 µ
8-14 µ
8-14 µ
dual laser
scope
crossed laser
LRSC
LRSCL2
LRL2 & LRL3
P7DL2 & P7DL3
G5SC
1MSC
1ML2 & 1ML3
2MSC
2ML2 & 2ML3
-30 to 1200°C
(-20 to 2200°F)
-30 to 1200°C
(-20 to 2200°F)
-30 to 1200°C
(-20 to 2200°F)
10 to 800°C
(50 to 1450°F) 150 to 1800°C
(300 to 3275°F)
600 to 3000°C
(1100 to 5400°F)
600 to 3000°C
(1100 to 5400°F)
200 to 1800°C
(400 to 3275°F)
200 to 1800°C
(400 to 3275°F)
120:1
105:1
120:1
25:1
50:1
180:1
180:1
90:1
90:1
8-14 µ
8-14 µ
8-14 µ
7.9 µ
5.0 µ
1.0 µ
1.0 µ
1.6 µ
1.6 µ
scope
single laser with scope
single laser
dual laser
scope
scope
single laser
scope
single laser
1-2 Raynger 3i Series Operator’s Manual
Each model is molded from rugged, high-strength, solvent resistant plastic and is actuated by a two-stage trigger (second stage is used for data logging only). Each model has the following:
High quality optical system
Infrared detector
Circuit board assembly with microprocessor
LCD display with backlighting feature
Touch-sensitive membrane switches for changing loops and modes
Battery compartment for four AA batteries
Power input jack for AC adaptor
Signal output jack (analog 1mV per degree/digital RS232)
A rugged, padded pouch for easy carrying
There are four types of laser sighting models:
Single laser—shows the center of the measurement area.
Dual laser—shows the spot diameter of the measurement area.
Crossed laser—the point where the two laser beams cross is the location of the minimum diameter measurement spot.
Single laser with scope
1.2 INVENTORY
Your Raynger 3i package contains the following:
Raynger 3i
Carrying Case
Operator’s Manual
Four (4) AA batteries
Warranty card
Raynger 3i Series Operator’s Manual 1-3
1.3 MODEL IDENTIFICATION
Refer to Table 1-1 for a list of standard models along with their temperature ranges, optical resolutions, spectral ranges, and sighting systems.
You can determine the exact model number of your unit by looking at the manufac­turing label on the underside of the unit. On the label is an area for model designa­tion. The model type is printed in the following format:
XXXYYYZZZZZ
where XXX is an abbreviation of the company name, YYY is the product (or abbrevia­tion of the product name), and ZZZZZ is the model type. (Model type may be four or five characters long. Refer to Table 1-1 to compare the label to model type.)
1-4 Raynger 3i Series Operator’s Manual
Raynger 3i Series Operator’s Manual 2-1
2.0 OPERATION
This portion of the manual contains the following sections:
Quick Start—To use your unit right away, follow the brief instructions on basic
operating procedures in this section.
Principles of Operation—A short introduction to infrared thermometry.
Your Portable Infrared Thermometer—Describes and illustrates the thermome-
ter’s control panel, display, and features.
How to Operate—A detailed user guide that describes each of the operating
modes. It includes descriptions of the RUN, RECALL, SETUP, and Data Logger loops.
Data Outputs—How to use the analog and digital (RS232) outputs to connect the
thermometer to a printer, chart recorder, or computer.
WARNING—LASER SAFETY
Models with laser sighting produce visible laser radiation that may be harmful to the human eye. Be aware of the following:
Avoid direct exposure of human eyes to laser light. Eye damage can result.
Use extreme caution when operating.
Never point the unit at another person.
Keep out of the reach of children.
Refer to the FDA laser label on the unit for specific information.
IMPORTANT
1. If the unit is exposed to significant changes in ambient temperature (hot to cold or cold to hot), allow 45 minutes for temperature stabi­lization before taking measurements.
2. Do not operate the unit near large electrical or magnetic fields such as arc welders and induction heaters. These fields can cause mea­surement errors.
3. For the short wavelength units (e.g., 2 µm and below)—Avoid taking temperature measurements in bright sunlight. High levels of ambi­ent light may produce apparently valid high-temperature readings when no target is in the thermometer’s field-of-view.
2-2 Raynger 3i Series Operator’s Manual
RUN
LOG
SET
LOG
HI
RCL
( )
LO
SET
( )
°C °F
T
amb
Raynger 3i Series Operator’s Manual 2-3
2.2 PRINCIPLES OF OPERATION
An infrared thermometer and the human eye are very similar. An infrared ther­mometer has a lens that focuses infrared radiation from an object onto a detector. The eye focuses light onto the retina. The detector is stimulated by the incoming infrared energy and produces a signal that is transmitted to the circuitry. The retina is stimu­lated by incoming light and sends a signal to the brain. The circuitry processes this signal and computes the temperature of the object.
The intensity of an object’s emitted infrared energy increases or decreases in propor­tion to its temperature. The higher the temperature of the target, the greater the inten­sity of infrared radiation.
To calibrate a noncontact temperature measurement instrument, the manufacturer uses a blackbody. A blackbody is a perfect emitter because it absorbs and emits all radiant energy but reflects or transmits none. The emissivity value of a blackbody is
1.00. Figure 2-2 shows the radiant emittance values of a blackbody at various tem­peratures and wavelengths.
Most objects have emissivities that are less than 1.00 but are reasonably uniform at all wavelengths of the infrared spectrum. These are called graybodies. The non-ideal (less then 1.00) emissivity values of different materials can be compensated for, by the emissivity controls, so that accurate temperature readings can be obtained. Emissivity values for many common materials (both metals and non-metals) are listed in Appendix B.
Figure 2-2: Blackbody Radiation Curves
2
10
2
1
10
1
-1
10
-2
10
-3
10
Blackbody Radiant Emittance (Watt/cm )
-4
10
12345678
0
Wavelength (Microns)
9 10111213
1500°C (2730°F) 1000°C (1830°F) 542°C (1000°F)
260°C (490°F)
20°C (70°F)
14
2-4 Raynger 3i Series Operator’s Manual
2.3 YOUR PORTABLE INFRARED THERMOMETER
Portable infrared thermometers measure surface temperatures without touching the surface. They collect the infrared energy radiated by a target and compute its surface temperature. They also compute the running average, maximum, minimum, and dif­ferential temperatures and present them on a digital display in either degrees Celsius or Fahrenheit. A digital/analog output allows data recording, instrumentation or process control, and/or remote display of temperature and emissivity. The instru-6(. The instr)7.6contr is batmety powerared/ocan be powerared byionptatiitaAC adaprato7354.6(. he)]TJ0 -2.4158 Tg0 TwInmetiitamemoty cirart aum, anacs danalgnninfea-he
Raynger 3i Series Operator’s Manual 2-5
Your portable thermometer has the following:
Trigger—Two-stage trigger. The first stage activates the unit to take temperature
readings. The second stage is functional only in the datalog mode. To store a tem­perature reading, pull the trigger all the way in until you hear the tone (the tone signals that the reading has been stored). When you release the trigger, the unit goes to sleep.
Control Panel and Display—All controls (except the trigger) are located on the
control panel. The display shows temperature and setup values, mode and loop status, and operating information.
Sighting System—Laser or scope sighting is provided with each model.
Note: Read the laser warning label before operating the laser.
Analog Output—Connects the instrument to analog recording/printing devices
such as chart recorders and printers.
Digital Output—An RS232 interface to connect the instrument to a computer or
directly to a printer’s RS232 port.
DC In—AC adaptor connection.
2-6 Raynger 3i Series Operator’s Manual
2.4 OPERATION AND CONTROLS
This section instructs you in the operation of the instrument. It describes battery and/or AC adaptor installation and the controls and functions of the different control loops and operating modes.
2.4.1 Battery/AC Adaptor Installation
The instrument may be powered by batteries or an AC adaptor. Battery power is sup­plied by 4 “AA” batteries. The batteries are located in the base of the handle. AC power is supplied by an optional AC power adaptor (DIN VDE 0551 approved). NiCad batteries can also be used, but battery life will be substantially reduced. Figure 2-3 shows the location of the battery compartment, battery orientation, and the loca­tion of the AC adaptor power connection.
Table 2-1 shows approximate battery life (for alkaline batteries) for various operating conditions.
Notes: Battery types and brands vary in length of usable life. The values in Table 2-1 are
approximate for new alkaline batteries. The instrument will continue to read accurate­ly up to 4 hours after the low battery icon displays if the laser and backlight are off.
Remove the batteries if the unit is not used for long periods of time.
220 VAC adaptors must have DIN VDE 0551 approval to be used with IEC Class 2 laser units.
Table 2-1: Battery Life (Alkaline)
CONDITION
Laser OFF Backlight OFF Laser ON Backlight OFF Laser OFF Backlight ON Laser ON Backlight ON "Sleep" mode (trigger off)
HOURS OF
CONTINUOUS USE
25
12.5
12.5 10
1 year
Raynger 3i Series Operator’s Manual 2-7
Figure 2-4: The Control Panel and Display
2.4.2 Control Panel and Display
Figure 2-4 shows the display and controls. Descriptions of these, in alphabetical order, follow the illustration.
Backlight on/off button
Run/Log selector switch
Setup button
Mode and Recall button
Laser on/off button
Trigger lock icon
Backlight icon
RUN
RUN
LOG
LOG
SET
Data logger
icon
RUN
RUN
LOG
LOG
SET
LOG
HI
RECALL
MODE
MODE
RECALL
RCL
LOG
HI
RCL
SET
( )
LO
LASER
Recall
loop icon
SET
( )
( )
LO
°C °F
( )
T
amb
ACTIVATE
ACTIVATE
LOCK
LOCK
°C °F
T
amb
Increase/up
button
Decrease/down
button
Activate and trigger
lock button
Setup
loop icon
Celsius/Fahrenheit
icons
Temperature
display
High and low
alarm icons
Ambient temperature
compensation icon
Low battery icon
Mode display
Laser on/off LED
RECALL
MODE
MODE
RECALL
LASER
ACTIVATE
ACTIVATE
LOCK
LOCK
Emissivity display
Mode value display
2-8 Raynger 3i Series Operator’s Manual
Activate button—Press the activate button to activate HAL, LAL, or TAM, or to tog­gle between DIG/ANA, or °C/°F, in the SET loop.
Backlight button and icon—The display has a backlight for working in low lighting conditions. Press the backlight button to activate or deactivate the backlight. The backlight icon is activated when the backlight is on. To save battery power, use the backlight only when necessary. Note that if the battery voltage falls below 4.0 V, the backlight will automatically turn off.
Low battery icon—The instrument is powered by four “AA” batteries. When the bat­tery voltage falls below 4.6 V, the low battery icon is activated.
Note: Turning off the backlight and laser will extend the battery life (refer to Table 2-1 for
battery life under various conditions).
Celsius/Fahrenheit icons—The °C and °F icons indicate which temperature scale has been selected.
LOG button and icon—Press the LOG/RUN button while you have the trigger pulled to toggle between the LOG and RUN loops. The LOG icon is activated when the instrument is in the LOG loop.
Emissivity display—The emissivity display shows the emissivity value selected in either the RUN, RECALL, or LOG loops.
and ▼buttonsincreases and decreases the emissivity settings, DOI rate,
HAL or LAL set-points, TAM, or the LOG location number.
High and Low alarm icons—These are activated when the corresponding alarms are activated. The HAL icon flashes and the buzzer sounds when measured temperature is greater than or equal to HAL and HAL is active. The LAL icon flashes and the buzzer sounds when the measured temperature is less than or equal to LAL and LAL is active. Note that when a high or low alarm condition is met and the unit is sending out RS232 data while in the Digital Output mode, the buzzer will make two tones: a normal tone followed by a higher pitched tone.
Laser on/off button and LED—Press the laser button to activate or deactivate the laser (RUN and LOG loops only). The laser LED is activated when the laser is activat­ed. (Not applicable for the scope sighting model.) Note that if the battery voltage falls below 4.3 V, the laser will automatically turn off. Also, if the unit is in the LOCK mode, the laser will go off when the trigger is released.
Temperature display—Shows the current temperature (while the trigger is pulled) or the last temperature measured (when the RECALL button is pressed).
Raynger 3i Series Operator’s Manual 2-9
Mode button—Press the mode button to change modes in any of the four loops.
Mode display and Mode value display—The Mode display shows the current mode
selected. The mode value display shows the temperature, set-point, or LOG location value for the mode selected.
Recall button and icon—Press the RCL button to activate the RECALL loop. The RECALL loop may be used to recall values from either the RUN or LOG loops. The RCL icon is activated when the instrument is in the RECALL loop.
RUN button—Press the RUN/LOG button, when the trigger is pulled, to toggle between the RUN and LOG loops.
Setup button and icon—Press the SET button to activate the SETUP loop. The SETUP loop may be used to set values in either the RUN or LOG loops. The SET icon is activated when the instrument is in the SET loop.
T
amb
icon—This is activated when the ambient temperature compensation function is
activated. This feature only affects readings in the RUN loop; readings in the LOG loop are not affected.
Trigger lock icon—The small padlock is the trigger lock icon and is activated when LOCK is pressed while the trigger is pulled. To unlock the trigger, simply press the LOCK button again.
2-10 Raynger 3i Series Operator’s Manual
2.4.3 Control System
The control system consists of four loops: SET, RUN, RECALL and LOG. The instru­ment may be cycled to any of the four loops by using the trigger or control panel but­tons, as shown below.
RCL
LOG
SET
HI
( )
LASER
( )
LO
°C °F
T
amb
ACTIVATE
LOCK
RUN
LOG
SET
LOG
HI
MODE
RECALL
RCL
( )
LASER
SET
( )
LO
°C °F
T
amb
ACTIVATE
LOCK
RUN
LOG
SET
MODE
RECALL
RUN
LOG
SET
LOG
HI
MODE
RECALL
RCL
( )
LASER
SET
( )
LO
°C °F
T
amb
ACTIVATE
LOCK
Raynger 3i Series Operator’s Manual 2-11
2.4.3.1 Control Loops
Each control loop has several modes, which are described in the following sections.
2.4.3.2 General Information
Before you begin using your portable instrument, you should be aware of the follow­ing general operating rules:
The instrument goes to “sleep” (low power consumption, no display, no laser, no backlight) after different times for each loop, as follows:
SET Loop (after last button use): 10 seconds – RUN Loop (after releasing the trigger): immediately – RECALL Loop (after last button use): 10 seconds _ LOG Loop (after releasing trigger): immediately
The backlight can be turned on or off in any mode.
The laser can be used only in the RUN and LOG loops.
The trigger has two stages. The first stage is for measuring temperature. The sec­ond stage, when the trigger is pulled all the way in, is for storing temperature val­ues in the LOG loop.
The °C or °F icon flashes when the instrument is measuring temperature in the RUN or LOG loops. It does not flash when values are recalled.
You can lock the trigger in the RUN loop and the LOG loop.
You must release the trigger to go into the SET or RECALL loops.
When you simultaneously press the MODE and ACTIVATE buttons while in the RUN loop, the instrument is reset to the RUN loop factory default settings (stored data is unaffected). In the LOG loop, pressing the MODE and ACTIVATE buttons not only resets the instrument to its LOG loop factory default settings, but also clears all previously stored data. In both cases, the instrument will “beep” after reinstalling the default settings. Section 3.6 lists default setting for each model as well as non-model-specific default settings and range values.
IMPORTANT
For the short wavelength units (e.g., 2 µm and below)—Avoid taking temperature measurements in bright sunlight. High levels of ambient light may produce apparently valid high-temperature readings when no target is in the thermometer’s field-of-view.
2.4.4 RUN Loop–To Measure Temperature
The RUN loop is for taking temperature measurements. Figure 2-6 illustrates the RUN loop.
To take a measurement, do the following:
1. Point the instrument at the target.
2. Pull the trigger (press the RUN/LOG button, if necessary, so that the LOG icon is not activated).
Figure 2-6: RUN Loop
2-12 Raynger 3i Series Operator’s Manual
RUN
RUN
LOG
LOG
SET
RUN
RUN
LOG
LOG
SET
MAX Mode
RCL
LOG
( )
HI
RECALL
MODE
MODE
RECALL
LASER
LASER
MODE
RCL
LOG
( )
HI
SET
( )
LO
ACTIVATE
SET
( )
LO
°C °F
T
amb
ACTIVATE
LOCK
LOCK
°C °F
T
amb
Current temperature
MODE
Turn the trigger lock ( ) on or off with the Lock button
The °C (or °F) icon flashes when the trigger is pulled
MODE
RUN
RUN
LOG
LOG
SET
RUN
RUN
LOG
LOG
SET
MIN Mode
RCL
LOG
( )
HI
RECALL
MODE
MODE
RECALL
LASER
MODE
RCL
LOG
( )
HI
LO
LASER
LO
SET
°C °F
( )
T
amb
ACTIVATE
ACTIVATE
LOCK
LOCK
SET
°C °F
( )
T
amb
Change emissivity
RECALL
MODE
MODE
RECALL
LASER
LASER
ACTIVATE
ACTIVATE
LOCK
LOCK
with the and buttons in any mode
RECALL
MODE
MODE
RECALL
LASER
LASER
ACTIVATE
ACTIVATE
LOCK
LOCK
DIF ModeAVG Mode
TRIGGER
Pulled or Locked
BUTTON
RUN
LOOP ICON
None
Loading...
+ 54 hidden pages