The ION450 Ion Analyser is dedicated for routine use. It offers two distinct user levels:
•Supervisor
Dedicated for operators who wish to edit their methods to fit their specific needs. They can
also assign a password to protect the programmed data from eventual changes.
•Routine
Dedicated for operators wishing to use the routine functions to guide them step by step
through the analyses.
The ION450 can store up to 50 methods and 30 electrodes. In addition 30 electrodes have
been pre-defined to help you save time setting up your application.
Thanks to the preprogrammed applications, the Ion Analyser is ready for use as soon as it has
been installed. Refer to "Appendix 1: Preprogrammed methods", page 153..
The ION450 also allows you to automatically sequence and repeat measurements.
The purpose of the ION450 Reference Manual is to give detailed information on the
Ion Analyser and the data displayed during operations. The information is listed in
alphabetical order for quick access and cross-references are listed in italics.
In addition to this handbook, a general User’s Guide (part no. D21M075) is available giving
descriptions and overviews of the workstation menus and operating concepts to guide you
through programming and running of the analyses.
Page 11
MeterLab
®
ION450 Reference Manual
Read me!
An important feature of this instrument interface is that it controls the presence of different
elements necessary to run the defined application for a selected method/sequence, before the
method/sequence is run.
Working in Supervisor mode
A Supervisor has access to all the libraries for creation purposes.
When programming the instrument in “SUPERVISOR” mode, it is recommended to work in
stages. These stages must be carried out in the order described below:
A.To program method
1. Define your electrode(s)
Identify electrodes (including temperature sensors) to be used for the analysis:
Electrodes can be created from the following lists:
Catalogue, see "Catalogue list", page 49.
Other, see "Others list", page 109.
Copy from, see "Copy electrode", page 57.
When creating the electrode, define if electrode calibration is required (or not), if yes specify
the "periodicity" of the calibrations and the pH, ISE or conductivity standards to be used.
Refer to "Calibrate pH electrodes", page 45.
Refer to "Calibrate ISE electrodes", page 45.
Refer to "Calibrate conductivity cells", page 45.
2. Create new method or Edit a pre-programmed one
Create the method to be used for the analyses. Enter the parameters required to calculate the
results, see "Programming methods", page 23.
When you have finished programming, select the method/sequence or pre-programmed
application, see "Select method", page 134. or see "Select sequence", page 134.
If your methods are to be performed in a sequence, program the sample stack, see "Sample
stack", page 133.
Page 12
MeterLab
®
ION450 Reference Manual
3. Check icons
The following icons indicate the exact state of your working system:
Sunny icon:
Everything is OK. Run the method or sequence.
Cloudy icon:
Electrode calibration is required within 12 or 24 hours.
Stormy icon:
Electrode calibration date elapsed or electrodes not installed.
Question mark:
Programming error.
Refer to "Electrode icons", page 85.
A Sunny icon is needed in order to run the selected method.
If a Cloudy/Stormy/Question mark icon is displayed in the Electrode window press 1
to activate the “Check” command. The ION450 will automatically guide you through
the operations required to solve the errors encountered.
B.Running methods
To run a method or sequence, see "Working in Routine mode", page 14.
Page 13
MeterLab
®
ION450 Reference Manual
Working in Routine mode
A.Access methods
A Routine operator has access to all the methods “Select method” and programmed
parameters “Display method” for checking purposes
B.Running methods
When working in “ROUTINE” mode, it is necessary to install your measurement system
according to the selected method or sequence, prior to running a method or sequence.
1. Select the method or sequence
Refer to "Select method", page 134.
Refer to "Select sequence", page 134.
2. Check icons
Refer to "Check icons", page 13.
Depending on the icon displayed, the ION450 will automatically guide you through the steps
necessary to run the analysis, see below:
a. Connect the electrode(s)
Connect/install electrodes and temperature sensors, Refer to "Electrode connection",
page 83.
b. Calibrate electrode(s)
Now, run the calibration.
Refer to "Calibrate pH electrodes", page 45.
Refer to "Calibrate ISE electrodes", page 45.
Refer to "Calibrate conductivity cells", page 45.
c. Run the method or the sequence
Refer to "Running a method/sequence", page 130.
Page 14
Practical examples
MeterLab
®
Programming electrodes
pH electrodes
ION450 Reference Manual
1.
Press 4.
4.
2.
Press 1.
5.
3.
Select function and ID.
6.
Select ID from Catalogue or
Others list.
Press 1 to confirm.
Press 1 to confirm the creation
of the new electrode.
For a combined or a simple or
reference electrode, enter the
potential (in mV) of the reference versus the Standard Hydrogen Electrode (SHE).
For a combined or a simple
electrode if you have selected
the Others list, enter the internal pH of the electrode.
Enter the electrode address.
If you want a message to be
displayed once a week concerning this electrode, select
Maintenance = Yes and enter
the message.
Select Fixed or Free if a calibration is required, go to step 7.
Select No, for no calibration,
press Esc to leave the menu.
Programming is completed.
Page 17
MeterLab
Calibration request = Fixed
Calibration with automatic recognition of the buffer among a list of predefined values. The buffer
values are entered during method edition.
Calibration request = Free
The buffer values are entered FREEly by the user. Use this option to calibrate pH electrode with
buffers that do not belong to the predefined list.
®
ION450 Reference Manual
7.
Enter the calibration
parameters.
10.
8.
Press 1.
11.
9.
Enter the electrode calibration
parameters.
For a Fixed calibration, press
Esc then 2. Go to step 10.
For a Free calibration, press
Esc then 3. Skip to step 11.
12.
Fixed calibration only.
Select the buffer solutions
used. Press Esc then 3.
Page 18
Enter the results parameters.
Press Esc then 4.
Enter the printouts parameters.
Press Esc twice. Electrode
programming is completed.
MeterLab
®
ISE electrodes
ION450 Reference Manual
1.
Press 4.
4.
2.
Press 1.
5.
3.
Select function and ID.
6.
Select ID from Catalogue or
Others list.
Press 1 to confirm.
Press 1 to confirm the creation
of the new electrode.
For a combined or a simple or
reference electrode, enter the
potential (in mV) of the
reference versus the Standard
Hydrogen Electrode (SHE).
IfyouhaveselectedtheOthers
list, select the valency and
enter the molar weight of the
ion under study.
Enter the electrode address.
If you want a message to be
displayed once a week
concerning this electrode,
select Maintenance = Yes and
enter the message.
Press the Left or Down arrow
key.
Select Manual if a calibration is
required then go to step 7.
Select No, for no calibration,
press Esc to leave the menu.
Programming is completed.
Page 19
MeterLab
Calibration = Manual
Calibration using 1 to 9 standards of known concentration.The standard concentrations are entered
during method edition. This method requires 1 to 9 calibration beakers.
®
ION450 Reference Manual
7.
Enter the calibration
parameters.
10.
8.
Press 1.
11.
9.
Enter the electrode calibration
parameters.
Press Esc then 2.
12.
Enter the standard solution ID
and standard concentrations.
Press Esc then 3.
Page 20
Enter the results parameters.
Press Esc then 4.
Enter the printouts parameters.
Press Esc twice. Electrode
programming is completed.
MeterLab
®
Conductivity cells
ION450 Reference Manual
1.
Press 4.
4.
2.
Press 1.
5.
3.
Select function and ID.
6.
Select ID from Catalogue or
Others list.
Press 1 to confirm.
Calibration request = Free
Use the Free calibration mode when you use a standard that does not belong to the Catalogue list
and you know the conductance of this standard at a given temperature. During a Free calibration run
and after stabilisation of the measurement, you will adjust the cell constant in order to display the
correct conductance value.
Calibration request = Fixed
With the Fixed mode when you use a standard that belongs to the Catalogue list, the cell constant is
determined as the ratio of the conductivity (known by the instrument) divided by the measured
conductance.
Press 1 to confirm the creation
of the new electrode.
IfyouhaveselectedtheOthers
list, enter the cable resistance
and capacitance.
If you want a message to be
displayed once a week concerning this electrode, select
Maintenance = Yes and enter
the message.
Select Fixed or Free if a
calibration is required, go tostep 7.
Select No, for no calibration,
press Esc to leave the menu.
Programming is completed.
Page 21
MeterLab
®
ION450 Reference Manual
7.
Enter the calibration
parameters.
10.
8.
Press 1.
11.
9.
Enter the electrode calibration
parameters.
For a Fixed calibration, press
Esc then 2. Go to step 10.
For a Free calibration, press
Esc then 3. Skip to step 11.
12.
Fixed calibration only.
Select the standard solution
used. Press Esc then 3.
Enter the results parameters.
Press Esc then 4.
Enter the printouts parameters.
Press Esc twice. Electrode
programming is completed.
Page 22
MeterLab
®
Programming methods
Creating and editing a method
ION450 Reference Manual
1.
Press 4.
4.
2.
Press 1.
5.
3.
Enter ID.
Press 1 to confirm.
6.
Enter method parameters.
Specify the Mode, see "Mode",
page 105.
Press 1.
Press ✓ and select the
electrode(s) and temperature
sensor from the lists.
Enter the other method
parameters.
Press Esc then 3.
Page 23
MeterLab
®
ION450 Reference Manual
7.
Enter the results parameters.
Press Esc then 4.
8.
Enter the printouts parameters. If a QC sample has been
defined in step 4, press Esc
then 5.
9.
Enter the QC data. Press Esc
twice. Method programming is
completed.
Page 24
MeterLab
®
For a Coupled method
ION450 Reference Manual
1.
Press 4.
4.
2
Press 1.
3.
Enter the Method ID and press
1 to confirm.
Select Mode = Coupled.
Enter the method to be linked.
Press Esc twice.
Method programming is
completed.
Page 25
MeterLab
®
ION450 Reference Manual
Page 26
MeterLab
®
Programming sequences
ION450 Reference Manual
1.
Select:
Sequence:tocreatea
sequences of methods.
SAC Sequence:tocreatea
sequences of methods to be
performed using a Sample
Changer.
Define the Sample
Changer in the
Configuration menu
before selecting SAC
Method or SAC
Sequence.
2
Press 2.
3.
Enter a name for the sequence.
4.
Press 3.
5.
Press 1 to add a method.
6.
Select the type of method.
Page 27
MeterLab
®
ION450 Reference Manual
7.
Selectamethodinthelistof
available methods.
10.
8.
Press 1 to add the method to
the sequence.
9.
If Sample has been selected in
step 6, enter the number of
samples (number of times you
wish to repeat the method in
the sequence).
Press 1 to add a second
method to the sequence.
Repeat steps 6 to 9.
Up to 10 methods can be
chained in a sequence.
After having added the last
method, press Esc twice.
Sequence programming is
completed.
Page 28
MeterLab
®
ION450 Reference Manual
Programming tips
•Do not forget to declare electrode(s) when programming your method parameters
•If a Sample Changer is used, do not forget to declare one in the Configuration menu.
•If a printer is used, do not forget to declare one in the Configuration menu.
If no sun icon appears after the method has been selected, check the following points:
1.Install electrode(s) for selected method, see "Check electrodes", page 53.
2.If required, calibrate electrode.
Refer to "Calibrate pH electrodes", page 45..
Refer to "Calibrate ISE electrodes", page 45..
Refer to "Calibrate conductivity cells", page 45.
If a Sunny icon appears:
Everything is OK. A sunny icon is required to run the selected method.
If a Cloudy icon appears:
An electrode calibration should be performed within 24 hours.
This is a simple warning, it will not stop you from running the analysis.
If a Stormy icon appears:
Electrode required in the selected method is not installed. Electrode required in
the selected method has not been calibrated.
If a Question mark icon appears:
It is a programming error, electrode is not defined in the selected method. Revise
the method programming.
When a Stormy or a Question mark icon appears, press 1 “Check” . The ION450
will automatically guide you through the operations necessary to solve the errors
encountered.
Page 29
MeterLab
®
ION450 Reference Manual
Page 30
Loading...
+ 140 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.