RADIOMETER ABL 77 Series Service Manual

Page 1
1. Introduction
ABL™77 Series Blood Gas and Electrolyte Analyzer
2. Analyzer Description
3. Troubleshooting
4. Test and Calibration
6. Replacements
Service Manual
7. Re-Assembly
8. Maintenance
9. Spare Parts
Service Notes
Date of Issue
Page 2
LICENSE AGREEMENT
The software installed and provided with the instrument is comprised of Microsoft® CE Operating System and SenDx Medical, Inc. developed Software. Each program is licensed, not sold to you, to be used only under the terms of this Agreement and the Microsoft End User License Agreement. SenDx Medical, Inc. and its licensors reserve all rights not expressly granted to you. You may use the software only with the instrument in which it is installed for your internal purposes. When the instrument is transferred to others, it must be accompanied by the software and the End User License Agreements included in the Operator’s Manual. The software may not be copied. You may not modify, reverse engineer, decompile or disassemble the programs. By your use of the instrument you agree to the terms of this Agreement and the Microsoft End User License Agreement. If you do not accept or agree to the terms you should promptly contact your Radiometer representative for a return of the instrument and a refund of your money.
TRADEMARKS
ABL™, and RADIOMETER™ are trademarks of Radiometer Medical A/S, Denmark.
ABL is registered in the USA.
COPYRIGHT
The contents of this document may not be reproduced in any form or communicated to any third party without the prior written consent of Radiometer Medical A/S.
While every effort is made to ensure the correctness of the information provided in this document Radiometer Medical A/S assumes no responsibility for errors or omissions which nevertheless may occur.
This document is subject to change without notice.
© Radiometer Medical A/S, DK-2700 Brønshøj, Denmark, 2003. All Rights Reserved.
Page 3
ABL77 Service Manual Table of Contents

Table of Contents

Introduction
Contents
This manual describes how to service the ABL™77 Series analyzer.
This manual contains the following topics.
1. Introduction.................................................................................................. 1-1
Overview ............................................................................................ 1-1
Service Policy..................................................................................... 1-2
Test Equipment and Tools.................................................................. 1-5
ABL77 Identification.......................................................................... 1-6
Panel Options...................................................................................... 1-7
2. Analyzer Description ................................................................................... 2-1
Overview ............................................................................................ 2-1
Functional Description ....................................................................... 2-2
Installation ................................................................................... 2-3
SCi
Calibration .......................................................................................... 2-5
Sample Analysis ............................................................................... 2-12
Module Descriptions ........................................................................ 2-18
Upper Module......................................................................................... 2-26
Lower Module......................................................................................... 2-30
Electronics Module................................................................................. 2-34
3. Installation .................................................................................................... 3-1
Overview ............................................................................................ 3-1
General Information ................................................................................ 3-2
Cautions and Warnings....................................................................... 3-3
General Guidelines ............................................................................. 3-4
Hardware Screen ................................................................................ 3-5
Reference Table of Problems, Causes and Corrective Actions ............ 3-8
Reference Table.................................................................................. 3-9
System Messages..................................................................................... 3-16
Continued on next page
Rev. F 1
Page 4
Table of Contents ABL77 Service Manual
Table of Contents, Continued
4. Test and Calibration Procedures ................................................................ 4-1
Overview ............................................................................................ 4-1
Verify the Fluid Transport System ..................................................... 4-3
Calibration Lines Procedure ............................................................... 4-5
Testing the Valve Connections......................................................... 4-11
Valve Test Procedures...................................................................... 4-13
Check Valve Fluidics System ................................................ 4-15
Original Fluidics System........................................................ 4-19
Waste Line........................................................................................ 4-26
5 Volt Power Output Setting ............................................................ 4-29
SC/Hematocrit Circuits .................................................................... 4-31
Electronic Zero Offset ...................................................................... 4-34
Heater Circuit ................................................................................... 4-36
Inlet Flap Sensor............................................................................... 4-39
Paper Jam ......................................................................................... 4-40
Battery Charger ................................................................................ 4-43
Battery Pack...................................................................................... 4-45
Touch Screen Panel Calibration ....................................................... 4-48
Reprogramming the CPU BIOS Settings ......................................... 4-50
Re-programming the Barcode Scanner............................................. 4-52
Resetting the Analyzer Serial Number ............................................. 4-56
5. Dismantling................................................................................................... 5-1
Overview ............................................................................................ 5-1
Upper Module........................................................................................... 5-4
Upper Module..................................................................................... 5-4
Backlight Inverter Board .................................................................... 5-7
Disk Drive .......................................................................................... 5-8
LCD Display..................................................................................... 5-10
Touch Screen Panel .......................................................................... 5-12
Lower Module......................................................................................... 5-13
Printer Door...................................................................................... 5-13
Printer Module.................................................................................. 5-14
Lower Module .................................................................................. 5-16
Opt-Sensor........................................................................................ 5-19
Continued on next page
2 Rev. F
Page 5
ABL77 Service Manual Table of Contents
Table of Contents, Continued
Roller Wheel and Roller Pump......................................................... 5-20
Waste Pump Head and Waste Pump ................................................ 5-22
Original Valve Board / Manifold Assembly..................................... 5-25
Valve Board with Replaceable Valves ............................................. 5-27
Valve Board Styles ........................................................................... 5-31
Electronics Module................................................................................. 5-35
Electronics Module........................................................................... 5-35
Analog Board.................................................................................... 5-38
Interface Board ................................................................................. 5-40
Power Distribution Board................................................................. 5-42
Electronic Boards Assembly ............................................................ 5-44
LCD Adapter Board ............................................................... 5-45
Memory Module .................................................................... 5-46
ChipDisk board ...................................................................... 5-47
CPU and Touch Screen Boards.............................................. 5-48
Battery Control Board ...................................................................... 5-49
Fan Assembly ................................................................................... 5-51
6. Replacements ................................................................................................ 6-1
Overview ............................................................................................ 6-1
Main Housing............................................................................................ 6-4
Main Housing Assembly .................................................................... 6-4
Carrying Handle ................................................................................. 6-5
Upper Module........................................................................................... 6-6
Housing Bracket ................................................................................. 6-6
Lower Module........................................................................................... 6-9
Manifold Gasket ................................................................................. 6-9
Valve Board...................................................................................... 6-10
Single Valve ..................................................................................... 6-18
Valve Board Tubing ......................................................................... 6-22
Manifold ........................................................................................... 6-26
Electronics Module................................................................................. 6-29
Coin Cell Battery.............................................................................. 6-29
Rear Panel......................................................................................... 6-30
Battery .............................................................................................. 6-31
Continued on next page
Rev. F 3
Page 6
Table of Contents ABL77 Service Manual
Table of Contents, Continued
Cabling .................................................................................................... 6-33
Disk Drive Cable .............................................................................. 6-36
LCD Display Cable .......................................................................... 6-38
Backlight Inverter Cable................................................................... 6-40
Touch Screen Cable.......................................................................... 6-42
Main Analog Cable........................................................................... 6-43
Analog Power Cable......................................................................... 6-44
Power Board Cable........................................................................... 6-45
A/D Cable – Analog Board to CPU board ....................................... 6-47
Printer Data Cable ............................................................................ 6-48
Printer Power Cable.......................................................................... 6-49
A/D Cable – Analog Board to Battery Control Board...................... 6-50
COM1 Cable..................................................................................... 6-52
Power Switch Cable ......................................................................... 6-53
Barcode Reader / Keyboard Cable ................................................... 6-54
Ethernet Cable .................................................................................. 6-56
7. Re-Assembly ................................................................................................. 7-1
Overview ............................................................................................ 7-1
Upper Module........................................................................................... 7-4
Touch Screen Panel ............................................................................ 7-5
LCD Display....................................................................................... 7-8
Disk Drive .......................................................................................... 7-9
Backlight Inverter Board .................................................................. 7-12
Upper Module................................................................................... 7-13
Lower Module......................................................................................... 7-15
Valve Boards – General Information................................................ 7-15
Valve Board Assembly..................................................................... 7-20
Waste Pump And Waste Pump Head ............................................... 7-24
Roller Pump And Roller Wheel........................................................ 7-27
Opto-Sensor...................................................................................... 7-31
Lower Module .................................................................................. 7-32
Printer Module.................................................................................. 7-35
Printer Door...................................................................................... 7-36
Continued on next page
4 Rev. F
Page 7
ABL77 Service Manual Table of Contents
Table of Contents, Continued
Electronics Module................................................................................. 7-37
Rear Panel Assembly........................................................................ 7-37
Fan Assembly ................................................................................... 7-38
Electronics Chassis And Battery Cage ............................................. 7-39
Battery Control Board ...................................................................... 7-40
Electronics Shelf............................................................................... 7-42
Electronic Boards ............................................................................. 7-44
Touch Screen And CPU Boards Assembly............................. 7-46
ChipDisk Board...................................................................... 7-47
Memory Module ..................................................................... 7-48
Disk Drive Cable.................................................................... 7-49
LCD Adapter Board ............................................................... 7-50
Electronics Boards – Module Assembly .......................................... 7-51
Power Distribution Board................................................................. 7-54
Interface Board ................................................................................. 7-56
Analog Board.................................................................................... 7-58
Electronics Module........................................................................... 7-62
8. Maintenance.................................................................................................. 8-1
Overview ............................................................................................ 8-1
Valve Maintenance............................................................................. 8-2
Cleaning.............................................................................................. 8-6
Long-Term Storage ............................................................................ 8-9
Recommended Service Procedures ....................................................... 8-13
1-Year Service .................................................................................. 8-13
2-Year Service .................................................................................. 8-14
3-Year Service .................................................................................. 8-15
9. Spare Parts.................................................................................................... 9-1
Overview ............................................................................................ 9-1
Modules And Externals ...................................................................... 9-2
Upper Module Components ............................................................... 9-4
Lower Module Components ............................................................... 9-6
Main Housing ................................................................................... 9-12
Electronics Module Components...................................................... 9-15
Recommended Parts And Hardware................................................. 9-21
Continued on next page
Rev. F 5
Page 8
Table of Contents ABL77 Service Manual
Table of Contents, Continued
Service Notes
Date of Issue
6 Rev. F
Page 9
ABL77 Service Manual Chapter 1: Introduction
1. Introduction

Overview

Introduction
Contents
This chapter gives an introduction to the servicing of the ABL™77 analyzer and important information for understanding the procedures and requirements.
This chapter contains the following topics.
Overview............................................................................................... 1-1
Service Policy ....................................................................................... 1-2
Test Equipment and Tools .................................................................... 1-5
ABL77 Identification ............................................................................ 1-6
Panel Options ........................................................................................ 1-7
Rev. F 1-1
Page 10
Chapter 1: Introduction ABL77 Service Manual

Service Policy

Introduction
Warranty disclaimer
This section outlines the policies and procedures, which must be followed when servicing Radiometer Medical A/S instruments.
The service policy for the ABL77 Series has been outlined to obtain a maximum degree of reliability with minimal repair time.
Further, a combination of comprehensive troubleshooting procedures in the service manual, a variety of self-check and service programs, and a high degree of modularization facilitates quick localization of faults in the analyzers.
Using the test equipment, procedures and spare parts listed in this service manual ensures continuous reliability.
Exceptions from the service policy are not allowed unless special permission or instructions in writing have been given beforehand by SenDx Medical, Inc.
Radiometer Medical A/S cannot guarantee the instrument’s performance specifications and safety or accept any warranty claims unless:
Non-original parts
Electrostatic discharge protection
The recommended maintenance procedures outlined in the Operator’s and Service Manuals are performed.
The accessories and spare parts specified by Radiometer Medical A/S are used.
Warranty claims for parts which suffer from physical damage, unauthorized attempted repair, or exposure to conditions other than those specified by Radiometer Medical A/S (e.g., temperature, line voltage outside specified limits) will not be accepted.
Spare parts and accessories
In order to ensure the reliability, durability and operation in accordance with technical specifications, use only original Radiometer Medical A/S parts, or parts approved by Radiometer Medical A/S.
This concerns both spare parts and accessories.
Software
Software is considered to be a spare part. Only software distributed by Radiometer may be installed in the analyzer.
In order to ensure the reliability and durability of the analyzer, ESD protection precautions must always be taken when handling or exposing ESD sensitive parts. Parts such as printed circuit boards may be damaged if handled incorrectly during storage and service. The damage is not visible, and the damaged parts may not fail immediately but several months later.
Installation test
In connection with the installation, repair and upgrade of an analyzer it is usually required that an Installation Test be carried out to verify that the analyzer calibrates correctly and measures correctly using the appropriate quality control solution.
Continued on next page
1-2 Rev. F
Page 11
ABL77 Service Manual Chapter 1: Introduction
Service Policy, Continued
Repair level
The repair level outlines the extent to which it is allowed to dismantle an analyzer in the process of troubleshooting and repair. This level is limited for several reasons such as: the need for specialized test equipment, special environmental requirements etc., and is optimized in relation to cost of parts, time for repair, etc.
The repair level is indicated in the chapters (5) Dismantling, (7) Re-Assembly and (9) Spare Parts.
Electronics Module
The major components in the electronics module are printed circuit boards. Low­level repair of these boards is not allowed in the field. Most of the boards are based on surface mounted technology, which involves a high component density and thereby a high complexity. Repair of this type of printed circuit board (PCB) requires advanced troubleshooting, repair and test facilities, which are only available at the factory. Some PCBs may be returned to the factory for replacement under the trade-in agreement for modules. These PCBs are marked with “*” next to the description in the spare parts list. Since repair is not allowed in the field, detailed circuit diagrams are not included in the service manual.
Lower Module
Most of the electromechanical and pure mechanical parts of the lower module may be repaired to the component level. Tubing, the printer, and pump motors are regarded as components (or component assemblies). The various programs facilitate checkout and adjustment of these parts and also verification of their function after repair.
Some mechanical adjustments are critical for their function. Such parts are not to be dismantled further than indicated in the spare parts list.
Upper Module
The upper module may be repaired to the component level. The various programs facilitate checkout and adjustment of these parts and also verification of their function after repair.
Some mechanical adjustments are critical for their function. Such parts are not to be dismantled further than indicated in the spare parts list.
Continued on next page
Rev. F 1-3
Page 12
Chapter 1: Introduction ABL77 Service Manual
Service Policy, Continued
WARNING/ CAUTION:
Calibration and measurement
CAUTION: Follow legal requirements and local rules for safe work practices with chemicals.
CAUTION: Working with blood gas analyzers may result in contact with blood remnants and with harsh disinfectants. During the various procedures wear suitable protection gear (gloves, face protection, and protective body clothing) and follow legal requirements and local rules for safe work practices.
CAUTION: The gloves must be free of pinholes and preferably be puncture resistant. Please consult the glove manufacturer for further information.
CAUTION: If contaminated materials come into contact with any lesion on the body, seek medical advice.
While performing calibrations and measurements, the analyzer must be fully assembled.
ECOMMENDATION: To facilitate troubleshooting procedures, consider purchasing
R one of each analyzer module to use as a test module.
1-4 Rev. F
Page 13
ABL77 Service Manual Chapter 1: Introduction

Test Equipment and Tools

Required equipment and tools
REF Description Use
902-6XX SCi sensor cassette Testing analyzer after repair
944-069 Calibration solutions pack (cal pack) Testing analyzer after repair
902-578 Electronic zeroing fixture Zeroing sensor channels
913-741 Manager setup disk Custom setup of system features
920-721 5/64” ball point Allen wrench Various applications
920-722 1/16” ball point Allen wrench Lower module
920-723 7/64” ball point Allen wrench Lower module
To carry out the procedures for servicing the analyzer, the following test equipment and tools are required: Parts marked with “N/A” in the REF column are to be purchased locally.
920-726 5/16” open-end wrench Electronics module
920-727 5/16” nut driver Electronics and Lower modules
920-728 ¼” nut driver Electronics module
920-729 Valve board calibration cassette Calibrating the Hct and SC circuit
991-260 Loctite 401 Adhesive Securing the handle to the housing
991-264 Silicone adhesive, RTV medical grade Securing cables to prevent disconnect
991-265 Loctite 222 Threadlocker Securing assorted screws and nuts
905-674 Sample path obstruction tool kit Various applications
943-906 Cleaning solution Cleaning the waste drain
N/A Deionized water (or sterile water) Various applications
N/A Soft cloth, gauze or other absorbent material Various applications
N/A Protective attire (gloves, safety glasses, and lab
Various applications
coat)
N/A Flat head screwdriver
Lower module
N/A Small flat head screwdriver Adjusting the calibration settings
N/A Phillips head screwdriver Various applications
N/A Needle nose pliers Lower module
N/A Scissors or equivalent trimming tool Various applications
N/A AT style keyboard Entering system commands
N/A Digital multimeter Testing electronics
N/A Digital voltmeter (or multimeter) 5 volt power output setting
N/A 3½” formatted diskettes Downloading data files
N/A Index card or equivalent Block the opto sensor
Rev. F 1-5
Page 14
Chapter 1: Introduction ABL77 Service Manual

ABL77 Identification

Introduction
ABL77 identification
Elements
This section explains the identification system for the ABL77 product.
Each individual analyzer consists of a number of separate elements, some using a unique identification system. These numbers are used for a large variety of purposes, e.g. tracking.
An analyzer consists of the following elements:
Element Element ID Syntax
Analyzer Serial number 200001
Sensor Cassette Lot number 52657
Serial number 041403150
Cal pack Lot number 52630
Use of identification
Electronic serial number
Serial number 52630124
Bar code scanner Part number / Revision 902-660
Battery Charger Part number / Revision 905-743
Manager Setup disk Part number / Revision 913-741
Operator’s Manual, English Part number / Revision 989-555
The analyzer serial number is the primary identification number, e.g.:
For tracking purposes (both for the analyzer and it’s modules).
For logging service actions.
As a reference on service work orders.
When ordering service components and modules.
The analyzer serial number is stored electronically in the ChipDisk. When the ChipDisk or the electronics module is replaced, the serial number must be re­programmed into the analyzer.
During service the system will recognize that there is no serial number and require serial number entry upon startup before proceeding. Ensure the serial number entered matches the serial number label located inside the printer door.
Serial number label replacement
1-6 Rev. F
The analyzer serial number label is located inside the analyzer printer door. In the event that the printer door must be replaced, a new printer door with serial number label must be ordered and produced at the factory.
Page 15
ABL77 Service Manual Chapter 1: Introduction

Panel Options

Introduction
Available options
Management
This section explains the panel options available with the ABL77 analyzer. When servicing an analyzer it should always be returned to the same settings.
The ABL77 can be configured for the following options:
Blood gas and Hematocrit Only (BG/Hct)
Electrolytes Only (Lytes)
Full Panel (BG/Hct/Lytes)
The different analyzer configurations are controlled through the analyzer software configuration at the time of purchase. The analyzer can be upgraded or downgraded after purchase.
The same configurations are also available in sensor cassette types. The different types are controlled through the sensor cassette barcode.
A full panel sensor cassette may be used on any analyzer configuration. The analyzer will report “ NI ” (Not Installed) for non-configured parameters.
A BG/Hct or Lytes sensor cassette will only report their respective configured parameters when installed on a full panel analyzer. Other parameters not supported by the sensor cassette configuration will be reported as “NI”.
When attempting to install a BG/Hct sensor on a Lytes only analyzer (or vice-versa), the analyzer will reject the installation.
Rev. F 1-7
Page 16
ABL77 Service Manual Chapter 2: Analyzer Description
2. Analyzer Description

Overview

Introduction
Contents
This chapter provides a general description of the analyzer modules, the electronic boards, the electronics wiring diagrams, and the fluidics system. This chapter includes various drawings for use during servicing. The drawings indicate the location of connection points and general identification.
This chapter contains the following topics.
Overview............................................................................................... 2-1
Functional Description.......................................................................... 2-2
SCi Installation...................................................................................... 2-3
Calibration............................................................................................. 2-5
Sample Analysis.................................................................................. 2-12
Module Descriptions ........................................................................... 2-18
Upper Module................................................................................................. 2-26
Lower Module................................................................................................. 2-30
Electronics Module......................................................................................... 2-34
Rev. F 2-1
Page 17
Chapter 2: Analyzer Description ABL77 Service Manual

Functional Description

Introduction
Overview
This section provides a general introduction to the analyzer.
The ABL77 Analyzer is an electromechanical instrument designed to measure blood gas, pH, and electrolyte concentrations of whole blood.
Intended use
Measuring principles
The heart of the ABL77 analyzer is a 486 SX single board computer. It operates on custom software in a Microsoft
®
Windows CE operating system. Peripherals include a floppy disk, VGA color monitor with touch screen, thermal printer and a custom board that performs all analog input and output functions.
An external AC to DC power supply provides system power. On board batteries enable portable operation of the analyzer. An intelligent power control board monitors system power and battery charge cycles.
After use, the analyzer may contain blood residue. Follow the Long-Term Storage Procedure in Chapter 8 to decontaminate the analyzer. Following decontamination, the ABL77 analyzer should be disposed of in a safe and proper manner.
The ABL77 pH, blood gas and electrolyte analysis system is a portable, automated analyzer that measures blood oxygen, carbon dioxide, pH, sodium, potassium, chloride, ionized calcium and hematocrit in whole blood. The ABL77 system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point of care setting.
There are three different measuring principles employed.
Potentiometry: A potential is recorded using a voltmeter, which relates to the concentration of the sample. A reference electrode is used to provide a stable, fixed potential against which other potential differences can be measured. Used for pH, CO
and electrolytes.
2
Amperometry: The magnitude of an electrical flow of current is proportional to the concentration of the substance being oxidized or reduced at an electrode. Used for O
.
2
Conductivity: The specific impedance of a sample as measured by two conducting electrodes held at a constant voltage is directly proportional to the conductive properties of the sample. Used for hematocrit and air-in-sample detection.
Operating conditions
The ABL77 system can be operated at an ambient temperature of 12-28 °C (54-82 °F).
All calibration solution packs should be stored at temperatures between 12-28 °C (54-82 °F). Do not freeze or expose to excessive heat or direct sunlight.
All sensor cassettes should be stored at temperatures between 5-32 °C (41-90 °F). Do not freeze or expose to excessive heat or direct sunlight.
Altitude conditions
2-2 Rev. F
The ABL77 system can be operated at altitudes from sea level to 1600 meters.
Page 18
ABL77 Service Manual Chapter 2: Analyzer Description

SCi Installation

Sensor cassette installation
Initialization­hydration
Initialization­calibration attempts
When a new sensor cassette is installed on the analyzer, the software will perform an initialization procedure to prepare the cassette for use. This initialization process consists of a preparatory hydration phase followed by a number of calibration attempts.
Cal1 flush is a preliminary flush to help expel air bubbles. A message “Please wait….” appears for 10 seconds.
A second Cal1 flush allows the new sensors to be conditioned with calibration solution. A message “Hydrating….” appears for 2 minutes.
Following hydration, the analyzer attempts a series of two point calibrations.
The number of calibrations performed is dependent on the results of each
attempt. The minimum number of attempts is 1 and the maximum number of attempts is 4.
Initialization completed
There are three failure conditions recognized by the software logic that determine whether the initialization process continues or terminates, and what messages are displayed. The three conditions are:
1. A single, mild sensor failure: One sensor sensitivity value falls between 50 – 100% of the minimum acceptable value. This is considered a condition that may resolve itself with additional calibration attempts.
2. A single, severe sensor failure: One sensor sensitivity value falls between 0 – 50% of the minimum acceptable value. This is considered a condition that is unlikely to resolve itself.
3. All other failure modes: Multiple sensor sensitivity values fall outside the acceptable values, no endpoint errors are present, and other conditions which may indicate air in the measuring chamber.
Following a successful initialization, the first auto calibration is scheduled for 30 minutes after initialization. Second and third calibrations are scheduled for 60 minutes after initialization and 120 minutes after initialization, respectively. From this point calibrations follow the schedule entered in the Calibration Schedule screen under Manager Setup / System Setup.
Rev. F 2-3
Page 19
Chapter 2: Analyzer Description ABL77 Service Manual
This page intentionally left blank
2-4 Rev. F
Page 20
ABL77 Service Manual Chapter 2: Analyzer Description

Calibration

Introduction
Hydration
The user can define the frequency of automatic two-point calibrations. The possible intervals are 1, 2, 3 or 4 hours. The Calibration option on the Main Menu is used to initiate a manual two- point calibration as desired.
If the calibration is more than 30 minutes past the scheduled time, a preliminary flush of Cal1 solution is performed prior to the two-point calibration. A message “hydrating….” will be displayed.
Continued on next page
Rev. F 2-5
Page 21
Chapter 2: Analyzer Description ABL77 Service Manual
Calibration, Continued
Calibration­phase 1
During the first phase of the calibration process, the roller wheel is activated along with valves 1 and 3. This flushes Cal2 solution through the sensor cassette measuring chamber (see Figure 2-1).
A measurement is then made on this solution.
The waste pump is also activated to transport residual flush solution to the waste
pouch.
Figure 2-1
Continued on next page
2-6 Rev. F
Page 22
ABL77 Service Manual Chapter 2: Analyzer Description
Calibration, Continued
Calibration­phase 1 (continued)
Part Function
1. Cal1 solution (Calibration solution – Level 1)
2. Cal2 solution (Calibration solution – Level 2)
3. Waste bag
4. Cal Pack
5. Main waste line
6. Manifold
7. Manifold luers
8. Side waste line
9. Valves L1, L2 and L3
10. Cassette measuring chamber
11. Roller wheel
12. Inlet probe
13. Waste drain
14. Waste pump
15. Cassette luer
Continued on next page
Rev. F 2-7
Page 23
Chapter 2: Analyzer Description ABL77 Service Manual
Calibration, Continued
Calibration – phase 2
During the second phase of the calibration, the roller wheel is activated along with valves 2 and 3. This flushes Cal1 solution through the sensor cassette measuring chamber (see Figure 2-2).
A measurement is made on this solution.
The waste pump is also activated to transport residual flush solution to the waste
pouch.
Figure 2-2
Continued on next page
2-8 Rev. F
Page 24
ABL77 Service Manual Chapter 2: Analyzer Description
Calibration, Continued
Calibration – phase 2 (continued)
Part Function
1. Cal1 solution (Calibration solution – Level 1)
2. Cal2 solution (Calibration solution – Level 2)
3. Waste bag
4. Cal Pack
5. Main waste line
6. Manifold
7. Manifold luers
8. Side waste line
9. Valves L1, L2 and L3
10. Cassette measuring chamber
11. Roller wheel
12. Inlet probe
13. Waste drain
14. Waste pump
15. Cassette luer
Continued on next page
Rev. F 2-9
Page 25
Chapter 2: Analyzer Description ABL77 Service Manual
Calibration, Continued
Sensitivity values
Calibration sensitivity values are calculated based on these two measurements.
These sensitivity values are compared to the acceptable range.
If the values are acceptable, the calibration is complete.
If one or more sensitivity values are outside the acceptable range, the
calibration is automatically repeated. If repeat calibration also results in one or more parameter sensitivity values outside the acceptable range, the software will suggest the appropriate next step. Suggestions include investigating the fluidic pathways of the analyzer, inactivating the sensor, or replacing the sensor cassette.
Parameter Acceptable Sensitivity Range
pH 40.0 61.5
CO2 30.0 61.5
Failed calibration status
O2 0.002 0.052
Na+ 40.0 61.5
K+ 40.0 61.5
Ca2+ 20.0 31.0
Cl– 30.0 61.5
Hct 10.0 25.0
If the calibration fails, the system will revert to a mandatory four-hour cycle of two-point calibrations (without repeating) to maintain the fluidic system. A successful calibration will return the analyzer to the calibration interval entered into the Calibration Schedule.
2-10 Rev. F
Page 26
ABL77 Service Manual Chapter 2: Analyzer Description
This page intentionally left blank
Rev. F 2-11
Page 27
Chapter 2: Analyzer Description ABL77 Service Manual

Sample Analysis

Introduction
Analysis – initial flush
Analysis – phase 1 aspiration
The Analysis option on the Main Menu allows the user to initiate a sample analysis.
Pressing the Analysis icon initiates a brief flush of Cal1 solution to ensure the tip of the inlet probe is filled with fluid.
To begin the sample analysis, the inlet probe flap is lifted to the syringe or capillary position.
The Aspirate button is then pressed.
The primary sample pump then draws in 70 microLiters of sample (see
Figure 2-3).
Figure 2-3
Continued on next page
2-12 Rev. F
Page 28
ABL77 Service Manual Chapter 2: Analyzer Description
Sample Analysis, Continued
Analysis – phase 1 aspiration (continued)
Part Function
1. Cal1 solution (Calibration solution – Level 1)
2. Cal2 solution (Calibration solution – Level 2)
3. Waste bag
4. Cal Pack
5. Main waste line
6. Manifold
7. Manifold luers
8. Side waste line
9. Valves L1, L2 and L3
10. Cassette measuring chamber
11. Roller wheel
12. Inlet probe
13. Waste drain
14. Waste pump
15. Cassette luer
Continued on next page
Rev. F 2-13
Page 29
Chapter 2: Analyzer Description ABL77 Service Manual
Sample Analysis, Continued
Analysis – phase 2 aspiration
The analyzer prompts the user to remove the sample and lower the inlet probe
The analyzer then advances the sample a second time to complete the positioning
of the sample (see Figure 2-4).
Figure 2-4
Continued on next page
2-14 Rev. F
Page 30
ABL77 Service Manual Chapter 2: Analyzer Description
Sample Analysis, Continued
Analysis – phase 2 aspiration (continued)
Part Function
1. Cal1 solution (Calibration solution – Level 1)
2. Cal2 solution (Calibration solution – Level 2)
3. Waste bag
4. Cal Pack
5. Main waste line
6. Manifold
7. Manifold luers
8. Side waste line
9. Valves L1, L2 and L3
10. Cassette measuring chamber
11. Roller wheel
12. Inlet probe
13. Waste drain
14. Waste pump
15. Cassette luer
Continued on next page
Rev. F 2-15
Page 31
Chapter 2: Analyzer Description ABL77 Service Manual
Sample Analysis, Continued
Analysis – sample flush
Following the sample measurement, the roller wheel is activated along with valves 2 and 3 to flush cal 1 solution through the sensor cassette (see Figure 2-5).
A measurement is performed using this cal solution prior to final results being displayed ensuring that no drift has occurred.
The waste pump is activated to transport the flush solution to the waste pouch.
Figure 2-5
Continued on next page
2-16 Rev. F
Page 32
ABL77 Service Manual Chapter 2: Analyzer Description
Sample Analysis, Continued
Analysis – sample flush (continued)
Part Function
1. Cal1 solution (Calibration solution – Level 1)
2. Cal2 solution (Calibration solution – Level 2)
3. Waste bag
4. Cal Pack
5. Main waste line
6. Manifold
7. Manifold luers
8. Side waste line
9. Valves L1, L2 and L3
10. Cassette measuring chamber
Quality control
11. Roller wheel
12. Inlet probe
13. Waste drain
14. Waste pump
15. Cassette luer
Routine quality control measurements are suggested to evaluate the performance of the analyzer. This ensures the reliability, accuracy and precision of patient sample results.
QC sample measurement follows similar steps as those used for blood sample analysis.
Rev. F 2-17
Page 33
Chapter 2: Analyzer Description ABL77 Service Manual

Module Descriptions

Overview
ABL77 analyzer
Front view of analyzer
This section describes the analyzer, individual modules and external components. Detailed diagrams of the modules and connections are included with each module.
This sub-section describes the analyzer, external components and disposables.
Figure 2-6 shows the front view and sensor cassette of the ABL77 analyzer.
Figure 2-6
Part Function
1. Upper module
2. Sensor cassette
3. Lower module
Continued on next page
2-18 Rev. F
Page 34
ABL77 Service Manual Chapter 2: Analyzer Description
Module Descriptions, Continued
Rear view of analyzer
Figure 2-7 shows the rear view and cal pack of the ABL77 analyzer.
Figure 2-7
Part Function
1. Main housing assembly
2. Cal pack
3. Electronics module
Continued on next page
Rev. F 2-19
Page 35
Chapter 2: Analyzer Description ABL77 Service Manual
Module Descriptions, Continued
Battery charger
The external battery charger is for use with the ABL77 analyzer only. It provides the source of AC power to the system when plugged into an active electrical socket (see Figure 2-8).
Battery pack
Figure 2-8
The battery pack is for use with the ABL77. It provides battery power to the analyzer when AC power is not available or not desired (see Figure 2-9).
Figure 2-9
Continued on next page
2-20 Rev. F
Page 36
ABL77 Service Manual Chapter 2: Analyzer Description
Module Descriptions, Continued
Barcode scanner
The barcode scanner, shown below, is for use with the ABL77 analyzer only (see Figure 2-10).
Barcode scanner holder
Figure 2-10
The barcode scanner holder may be installed on the analyzer in any location that is convenient (see Figure 2-11).
Figure 2-11
Continued on next page
Rev. F 2-21
Page 37
Chapter 2: Analyzer Description ABL77 Service Manual
Module Descriptions, Continued
Disposables
Sensor cassette
The ABL77 has two key disposable components, a sensor cassette and a cal pack. These disposable items are designed for a limited number of uses throughout a specified in-use life. After use, these components are considered a biohazard and should be disposed of in a safe and proper manner.
The sensor cassette contains the sensor array within a sample flowcell. It transmits sensor data to the analyzer for interpretation (see Figure 2-12).
Figure 2-12
Part Function
1. Sensor window
2. Peristaltic pump tube
3. Inlet flap
4. Inlet probe
5. Release latch
Continued on next page
2-22 Rev. F
Page 38
ABL77 Service Manual Chapter 2: Analyzer Description
Module Descriptions, Continued
Cal pack
The cal pack contains the solutions used to calibrate and flush the sensor cassette (see Figure 2-13). The cal pack also contains a waste bag to collect all samples, solutions and other fluids that are processed through the analyzer.
1
3
2
Figure 2-13
Part Function
1. Cal2 pouch
2. Cal1 pouch
3. Waste bag
Continued on next page
Rev. F 2-23
Page 39
Chapter 2: Analyzer Description ABL77 Service Manual
Module Descriptions, Continued
ABL77 block diagram
Figure 2-14 displays the inter-connection of primary and secondary components of the ABL77 analyzer.
Figure 2-14
2-24 Rev. F
Page 40
ABL77 Service Manual Chapter 2: Analyzer Description
This page intentionally left blank
Rev. F 2-25
Page 41
Chapter 2: Analyzer Description ABL77 Service Manual

Upper Module

Introduction
Components
This section provides a description of the upper module of the ABL77 analyzer. The upper module contains the display components and disk drive (see the following figures).
Resistive Touch Screen: Located in front of the LCD display. It reacts to gentle pressure from a finger or wand to supply the ABL77 with user input.
LCD Display: This is a 640 x 480 resolution, backlit, VGA, color display used as the primary method to provide display information to the user.
Backlight Inverter Board: This board converts then supplies the power to the LCD display and the backlight bulb.
Diskette Drive: This drive is a standard 3.5 inch floppy drive for 1.44 M-byte diskettes. Its primary use is to download records and settings, and is used to upgrade software.
Continued on next page
2-26 Rev. F
Page 42
ABL77 Service Manual Chapter 2: Analyzer Description
Module Descriptions, Continued
Diagrams
Figure 2-15
Part Function
1. Upper housing
2. Disk drive
3. Disk drive cable
4. Touch screen cable
5. LCD display cable
6. Backlight inverter cable
7. Backlight inverter
8. Insulating tape
9. Housing bracket
10. Touch screen panel
11. LCD display
Continued on next page
Rev. F 2-27
Page 43
Chapter 2: Analyzer Description ABL77 Service Manual
Module Descriptions, Continued
Diagrams (continued)
Figure 2-16
Continued on next page
2-28 Rev. F
Page 44
ABL77 Service Manual Chapter 2: Analyzer Description
Module Descriptions, Continued
Diagrams (continued)
Part Function
1. Backlight inverter board screws
2. Insulating tape
3. Backlight inverter board
4. Housing screws
5. Housing bracket
6. Copper tape
7. Disk drive screws
8. Disk drive washers
9. Copper cloth tape
10. Disk drive
11. LCD display
12. Touch screen panel
13. Upper housing
Rev. F 2-29
Page 45
Chapter 2: Analyzer Description ABL77 Service Manual

Lower Module

Introduction
Components
This section provides a description of the lower module of the ABL77 analyzer. The lower module contains the following components and related mechanisms.
The Valve Board routes all fluids through a system of three 3-way valves. The valve board contains the sensor cassette connector and connections to the analog board, door sensor, roller pump and waste pump motor.
The Printer allows printing of sample test and QC reports, using a roll of thermal paper.
The Roller Pump draws calibration solutions and other fluids into and out of the sensor cassette in carefully controlled step increments for accurate placement of fluids.
The Waste Pump transports waste fluids from the waste drain to the waste pouch in the cal pack.
The Sensor Cassette Nest provides a cavity to accommodate the sensor cassette. This nest contains the electrical and fluid connections for the sensor cassette, allowing the analyzer to receive data from the sensor cassette and to move fluids through the sensor cassette.
The Manifold is the connection point between the valve board and the cal pack and provides a fluid pathway. The manifold is considered a separate part from the lower module but is identified here to show the proper tubing connections.
Continued on next page
2-30 Rev. F
Page 46
ABL77 Service Manual Chapter 2: Analyzer Description
Module Descriptions, Continued
Valve board
Layout of the Valve Board, REF: 902-812 (Figure 2-17).
J5
J5
12 1
13 24
J2
K1
J6
J6
J1
J3
SENSOR
LUER
41449 / B
SENSOR BOSS
L2 L1 L3
R2
R18
J4
J4
J9
J9
2587
PN 902-812
J8
41477
SERVICE BOARD ABL70-77
ASSY REV :
REVISION
J8
J10
J10
J7
J7
Figure 2-17
The Valve Board Assembly (REF: 902-811) includes the valves and tubing (Figure 2-18).
J5
J5
12 1
13 24
J2
K1
J6
J6
J3
SENSOR
LUER
41449 / B
J1
COM
N.C.
SENSOR BOSS
N.O.
L2 L2 L1 L3
J4
J4
J9
J9
2587
R2
R18
PN 902-811
J8
Figure 2-18
COM
N.O.
41477
ASSY REV :
REVISION
N.C.
L3
COM
N.C.
L1
SERVICE BOARD AB L70-77
J8
N.O.
J10
J10
J7
J7
Continued on next page
Rev. F 2-31
Page 47
Chapter 2: Analyzer Description ABL77 Service Manual
Module Descriptions, Continued
Lower module ­Front
Figure 2-19
Part Function
1. Waste drain
2. Waste pump
3. Sensor cassette nest
4. Roller pump wheel
5. Printer
Continued on next page
2-32 Rev. F
Page 48
ABL77 Service Manual Chapter 2: Analyzer Description
Module Descriptions, Continued
Lower module ­Rear
Figure 2-20
Part Function
1. Printer
2. Sensor luer connector
3. Roller pump motor
4. I/O cable connector
5. Valve board
6. Sensor cable connector
7. Opto sensor
8. Manifold
9. Waste pump
Rev. F 2-33
Page 49
Chapter 2: Analyzer Description ABL77 Service Manual

Electronics Module

Introduction
Components
This section provides a description of the electronics module of the ABL77 analyzer.
Electronics Boards
Analog board
Power distribution board
Interface board
ChipDisk
LCD adapter board
Memory module
CPU board
Touch screen controller board
Battery control board
Rear Panel with External Connections
Barcode scanner / keyboard connector
COM1 port
Ethernet port
Power switch
Charger port
Battery Pack
Continued on next page
2-34 Rev. F
Page 50
ABL77 Service Manual Chapter 2: Analyzer Description
Module Descriptions, Continued
Electronics Boards
Analog board
The analog board (REF: 902-682) contains a microprocessor and custom algorithms that control valve, motor, and sensor operations (see Figure 2-21). The onboard processor performs various operations such as thermistor control, roller pump motor control, and sensor signal A to D conversions. This remote intelligence enables the main computer to focus primarily on the operator interface. Commands are received from the main computer to turn on valves, motors, begin sensor heating cycle, etc. The analog board returns a character string once a second to the main computer that contains the output states and sensor characteristics.
Figure 2-21
Continued on next page
Rev. F 2-35
Page 51
Chapter 2: Analyzer Description ABL77 Service Manual
Module Descriptions, Continued
Power distribution board
The power distribution board (REF: 902-838) receives power from the battery control board and routes the power to the electronic boards assembly and the analog board (see Figure 2-22).
Figure 2-22
Interface board
ChipDisk
The interface board (REF: 902-681) contains the internal speaker and the coin cell battery that supplies the CPU with power to maintain the computer settings See Figure 2-23). This board also receives signals from the barcode scanner and routes data to the printer to produce reports.
SPK
BAT 1
SN
J3
Revision
ASSY 41407
Figure 2-23
The ChipDisk (REF: 365-436) is the mass storage media for all records, settings and software (see Figure 2-24). It is an electronic based “computer hard drive” The ChipDisk is a component of the CPU board.
Figure 2-24
Continued on next page
2-36 Rev. F
Page 52
ABL77 Service Manual Chapter 2: Analyzer Description
Module Descriptions, Continued
LCD adapter board
Memory module
The LCD adapter board (REF: 902-678) converts signals from the CPU for use by the LCD display (see Figure 2-25). The LCD adapter board is a component of the CPU board.
Figure 2-25
The memory module (REF: 365-438) is the active memory (RAM) that the CPU uses for calculations and command processing (see Figure 2-26). The memory module is a component of the CPU board.
Figure 2-26
Continued on next page
Rev. F 2-37
Page 53
Chapter 2: Analyzer Description ABL77 Service Manual
Module Descriptions, Continued
CPU board
The CPU board (REF: 902-679) consists of a 486 SX microcomputer with at least 8 MB memory, VGA, serial and parallel ports, and floppy disk control (see Figure 2-27). The interface board connects the microcomputer to its battery support and speaker as well as interfacing with the printer and the COM 1 port. Valves, motors, and sensor states are controlled by commands sent via the serial connection to the custom analog board.
Touch screen controller
Figure 2-27
The touch screen controller (REF: 365-437) processes all signals from the touch screen and sends these signals to the CPU (see Figure 2-28).
Figure 2-28
Continued on next page
2-38 Rev. F
Page 54
ABL77 Service Manual Chapter 2: Analyzer Description
Module Descriptions, Continued
Battery control board
The battery control board (REF: 902-680) monitors system power requirements and diverts the incoming AC power to and from the battery pack accordingly (see Figure 2-29). The batteries are charged in a controlled manner to prevent over­charging. The battery charge status is sent to the analog board and then retransmitted to the main computer so that it can be displayed on the screen for the operator.
J7
J1
R28
J5
U4/1
D7 D6
D1
U4/2
C2
U4/3
U4/4
J2
D2
R19
C1
R12
R23 R24
D9
J3
D10
D4 D3
U2
C13
R2
R21 R22
D5
C14 C15
R20
R9
Q6 Q5
P4
U1
R3
R4
R25
R6
R8
Q3Q4
R27
P5 P7 P6
P1
R5 R10
Q8 Q7
C8
R17
R11
R7
D11
R13
D14
C3
C7
R15
C9
D13
J 4
C5
Q1
R14
U5
C10 C11
D8
P8 P3 P2
U4/9
U4/8
U6
C4
U4/7
C12
U4/6
U4/5
J8J6
Q2
VB5
D12
R18
C6
U3
C19
U7
R1
ASSY 41174
Figure 2-29
Continued on next page
Rev. F 2-39
Page 55
Chapter 2: Analyzer Description ABL77 Service Manual
Module Descriptions, Continued
Electronics module assembly
Figure 2-30
Continued on next page
2-40 Rev. F
Page 56
ABL77 Service Manual Chapter 2: Analyzer Description
Module Descriptions, Continued
Rear Panel with External Connections
Diagrams
Figure 2-31
Part Function
1. Barcode / keyboard port cable
2. Power switch cable
3. Ethernet port cable
4. COM1 port cable
5. Fan assembly cable
Continued on next page
Rev. F 2-41
Page 57
Chapter 2: Analyzer Description ABL77 Service Manual
Module Descriptions, Continued
Diagrams (continued)
Figure 2-32
Part Function
1. COM1 port
2. Barcode / keyboard port
3. Fan
4. Charger port
5. Ethernet port
6. Power switch
7. Battery door
Continued on next page
2-42 Rev. F
Page 58
ABL77 Service Manual Chapter 2: Analyzer Description
Module Descriptions, Continued
Barcode / keyboard connector
COM1 port
The barcode scanner port is a 5 pin DIN connector that sends the signals to the interface board. This port is also used for connecting a keyboard to the analyzer for testing purposes.
The COM1 port is a serial connection to the CPU board. It allows interaction with external computers. The RS232 port contains 9 pins. The following table lists the pin designations and transmission format.
Pin I/O Description
Ethernet port
1.
2.
3.
4.
5.
6.
7.
8.
9.
DCD (Data Carrier Detect)
RX (Receive Data)
TX (Transmit Data)
DTR (Data Terminal Ready)
GND (Signal Ground)
DSR (Data Set Ready)
RTS (Request To Send)
CTS (Clear To Send)
RI (Ring Indicator)
— Connector
The Ethernet port is a connection to the CPU. It allows bi-directional communication with external computer systems.
Power switch
The power switch is a 2-position switch that places the analyzer in either an On or Standby mode. There is no Off mode as the analyzer is always charging when it is connected to an AC power source.
Charger port
The charger port is a 6 pin DIN connector. It allows the ABL77 charger to supply AC power to the battery control board.
Continued on next page
Rev. F 2-43
Page 59
Chapter 2: Analyzer Description ABL77 Service Manual
Module Descriptions, Continued
Battery Pack
Battery pack
The battery pack contains 12 nickel metal hydride (NiMH) cells (see Figure 2-33). The pack has a nominal DC voltage of 14.4 and a rated capacity of 3.8 Ah. The battery pack stores power for use by the analyzer when not connected to an AC power source. After use, the battery pack should be disposed of in a safe and proper manner consistent with disposal of any battery.
Charge cycle of the battery pack
Battery pack temperature
Figure 2-33
The analyzer controls the charge to the battery pack using a counter and charge controller. The counter value is reflected in the value displayed in the Bat field on the Hardware screen.
The input charge increments the charge counter, which reflects the charge capacity of the pack. The charge rate is a function of the current control circuit. As the battery charges, the current applied will decrease in proportion to the battery voltage. As the battery voltage increases, the temperature of the pack increases. This increased temperature increases the internal pressure of the cells, causing a higher voltage across the pack. This results in varying charge rates based on the charge state of the battery pack. To ensure maximum charging of the pack, the charge controller (on the Battery Control Board) provides a six-hour charge cycle. For safety purposes, the maximum allowed charge current is 650 mA.
The counter decrements based on the normal operation of the analyzer on battery power. The counter also decrements based on the self-discharge rate of the battery with storage and age. The self-discharge rate is accelerated with increased temperatures.
The battery pack contains an internal safety switch that is triggered based on the current output over time and monitored by an internal thermistor. This switch prevents unsafe levels of operation. During charging, if the temperature of the battery pack rises above 45°C, charging is terminated. In addition, there is an integral temperature cutoff switch which opens if the pack reaches 70 ± 5 °C. If this switch is triggered, the battery pack will no longer be operational.
Continued on next page
2-44 Rev. F
Page 60
ABL77 Service Manual Chapter 2: Analyzer Description
Module Descriptions, Continued
Optimal battery pack life
The lifetime of a battery pack is affected by various conditions. Elevated temperatures, long periods of inactivity and incomplete charging will diminish the expected lifetime and operation of the battery pack.
NiMH batteries operate at their optimal capacity when regularly exercised and allowed to discharge to a level other than full capacity. It is recommended that the analyzer be operated on battery power routinely or at regular intervals to maintain optimal performance of the battery pack
Rev. F 2-45
Page 61
ABL77 Service Manual Chapter 3: Troubleshooting
3. Troubleshooting

Overview

Introduction
Contents
This chapter provides troubleshooting information for errors or messages that may occur during the operation of the ABL77 analyzer.
This chapter contains the following topics.
Overview............................................................................................... 3-1
General Information ........................................................................................ 3-2
Cautions and Warnings ......................................................................... 3-3
General Guidelines................................................................................ 3-4
Hardware Screen ................................................................................... 3-5
Reference Table of Problems, Causes and Corrective Actions .................... 3-8
Reference Table .................................................................................... 3-9
System Messages............................................................................................. 3-16
3-1
Page 62
Chapter 3: Troubleshooting ABL77 Service Manual

General Information

Introduction
This section provides information regarding cautions and warnings as well as information on normal operation that can be fundamental to understanding error conditions.
3-2
Page 63
ABL77 Service Manual Chapter 3: Troubleshooting

Cautions and Warnings

Introduction
WARNING/ CAUTION:
This topic emphasizes important safety issues. Please review before proceeding with any troubleshooting steps.
WARNING: Always turn the power off and unplug the system when cleaning the analyzer.
CAUTION: Always remember to exercise Universal Precautions when handling contaminants and biohazardous materials (OSHA standard 1910.1030).
CAUTION: Be prepared to collect fluid flowing from open ports and luers when flushing analyzer parts. Always exercise Universal Precautions when handling contaminants and biohazardous materials.
CAUTION: Pushing the syringe plunger too hard can damage internal tubing connections and cause leakage. When using syringes to clear path obstructions, whether filled with air or liquid, very gently push the syringe plunger to avoid detaching internal tube connections.
CAUTION: Do not inject solution or air into the cal pack ports. Injecting solution or air into the cal pack will alter the calibrant values and may cause erroneous measurements and/or failure messages.
CAUTION: The cal pack contains two calibration fluid pouches and a waste pouch. With use, the waste pouch may contain blood and other biohazardous fluids. Disposal of a used cal pack should comply with all required biohazard regulations.
CAUTION: Dispose of used cal packs immediately. If immediate disposal is not possible, lay the cal pack on its side or back and replace the foil cover over the ports to prevent potentially contaminated ports from contacting other surfaces. Laying a used cal pack on its side or back can also prevent leakage in the unlikely event that a cal pack port develops a slight seepage.
3-3
Page 64
Chapter 3: Troubleshooting ABL77 Service Manual

General Guidelines

Introduction
Suggested tools
Warranty sensor cassette returns
This topic describes additional information pertinent to the use of the ABL77 analyzer.
The following tools are recommended for various service procedures:
A Sample Path Obstruction Tools (REF: 905-674) kit can assist in removing
fluid path obstructions.
A Zeroing Cassette (REF: 902-578) is used to zero the sensor channels
A Valve Board Calibration Cassette (REF: 920-729) is used to calibrate the
Hct and SC channels between the valve board and analog board.
Any sensor cassette that fails during use should be returned via the distributor to SenDx Medical, Inc. Please return all the following items for full warranty credit and to assist in a complete failure investigation:
Analyzer component returns
Cassette pump tubing
Sensor cassette
Foil Pouch
Printout of the failed Event Record. Only the Event Record contains a complete
set of information important for failure investigation. An Event Record is printed from the Recall / Event screen.
Highlight the failed calibration, initialization or QC analysis
Press the
icon to print an Event Record of this failed event.
Any analyzer component returned for credit or for failure investigation should be accompanied by a copy of the service work order. Please be as complete as possible in describing the events that led up to the failure, any troubleshooting steps taken, and resolution of the problem. Please include the serial number of the analyzer repaired.
During calibrant pump cycles, while the roller wheel is spinning, the cassette pump tubing can be "snapped" repeatedly to help dislodge any air bubbles residing in the cassette tubing (see Figure 3-1). This simple step can alleviate a number of the more common calibration and initialization failure experiences.
Figure 3-1
3-4
Page 65
ABL77 Service Manual Chapter 3: Troubleshooting

Hardware Screen

Introduction
Description of Hardware screen
This topic describes the various features of the Hardware screen that can assist in troubleshooting.
The Hardware screen allows manual control of basic analyzer functions.
The three soft keys across the bottom of the screen allow the user to initiate a
single flush of Cal1 solution, Cal2 solution, or a single aspiration through the inlet probe (see Figure 3-2).
Figure 3-2
Continued on next page
3-5
Page 66
Chapter 3: Troubleshooting ABL77 Service Manual
Hardware Screen, Continued
Description of Hardware screen (continued)
Activate the heater circuit by selecting the check box labelled Heater ON (see Figure 3-3). The temperature reading as measured by the thermistor (located in the sensor cassette) is displayed and labelled Th.
Figure 3-3
The oxygen sensor, which requires an applied current to function, can be energized by selecting the check box labeled O2(nA).
The Hematocrit sensors also require an applied current which is activated by selecting the Hct(AD) choice on the drop-down menu which initially displays OFF(AD).
The other selection in this drop-down window refers to the air-in-sample circuit. SC(AD) is the solution conductivity (SC) value. The units of measure for both Hct and SC are analog to digital counts.
The voltage output of the other sensors are displayed and labeled accordingly.
Continued on next page
3-6
Page 67
ABL77 Service Manual Chapter 3: Troubleshooting
Hardware Screen, Continued
Description of Hardware screen (continued)
A relative charge reading for the battery pack is displayed (Bat) as well as the function of the battery charger (Pak) (see Figure 3-4). An inlet flap sensor indicator is also displayed (Door).
Figure 3-4
3-7
Page 68
Chapter 3: Troubleshooting ABL77 Service Manual
Reference Table of Problems, Causes and
Corrective Actions
Introduction
This section summarizes, in table form, the various types of problems and error conditions which may be encountered during the operation of the ABL77 analyzer.
3-8
Page 69
ABL77 Service Manual Chapter 3: Troubleshooting

Reference Table

Introduction
This table offers a convenient reference to problems, their causes and the recommended corrective action.
The most frequent cause for each problem is listed first followed by less frequent causes.
Problem Probable Cause Corrective Action
Calibration Failure – this also
includes failed initializations. The important first step is to identify whether the failure is due to a sensor failure or failure of the analyzer's fluidic system
Fluidic system blockage in analyzer. This prevents fresh calibration solution from reaching the sensor array or combines fresh cal solution with
air. to properly transport the calibration fluid to the sensor cassette measuring chamber.
Follow the instructions on the
Sensor failure Inactivate the failed parameter analyzer screen to help identify the underlying cause.
Calibration, Initialization, or QC failures with multiple sensor
Valve board – worn or corroded
pin receptacle cassettes over time
Valve board – component
failure
Analog board – component
failure
Proceed to Chapter 4, Test and Calibration Procedures to investigate the fluidics system using the Hardware screen and the Calibration Lines procedure
or replace the SCi.
Replace the valve board. Proceed to Chapter 5 Dismantling.
Replace the valve board. Proceed to Chapter 5 Dismantling.
Replace the analog board. Proceed to Chapter 5 Dismantling.
Zero calibration failure Check analyzer connector for
corrosion or wear. If OK, proceed to Chapter 4, Test and Calibration Procedures to set the electronic zero of the sensor channels.
Valve board - single valve
failure
Proceed to Chapter 4, Test and Calibration Procedures to test the valves on the valve board. If a valve is malfunctioning, replace it. See Chapter 6
Replacements.
Continued on next page
3-9
Page 70
Chapter 3: Troubleshooting ABL77 Service Manual
Reference Table, Continued
Problem Probable Cause Corrective Action
Calibration, Initialization, or
QC failures of the Hct channel with multiple sensor cassettes over time. Also multiple false
Insufficient 5V supply of power Proceed to Chapter 4, Test and
Calibration Procedures to check the 5 volt power supply.
Adjust as necessary. air-in-sample messages with multiple sensor cassettes over time
Calibration of SC/Hct circuit is not possible. Unable to attain target calibration values.
Insufficient 5V supply of power Proceed to Chapter 4, Test and
Calibration Procedures to
check the 5 volt power supply.
Adjust as necessary.
Computer – Incomplete boot­up
CPU Board Malfunction Replace CPU Board. See
Chapter 5, Dismantling
Chip Disk Malfunction Replace Chip Disk. See Chapter
5, Dismantling
Computer – clock malfunction CPU Board Malfunction
(software versions 1.30 or
Replace CPU board. See
Chapter 5, Dismantling
lower)
Analog board real-time clock
malfunction or clock battery on
Replace analog board. See
Chapter 5, Dismantling
analog board depleted (software versions 1.32 or higher)
Disk Drive
Disk drive failure Replace the disk drive. See
Chapter 5, Dismantling
Disk drive cable Check the connection. Replace
as necessary. See Chapter 6,
Replacements
Main CPU Replace the main CPU. See
Chapter 5, Dismantling
Floppy diskette damaged, full,
or has write-protect activated.
Leak – at manifold Worn manifold gasket. Can
cause fluid leakage into printer area.
Use another diskette or
inactivate the write-protect.
Replace gasket. Refer to
specific instructions in Chapter
5, Dismantling. The first step is
to identify the correct version of
gasket that is being replaced.
Also, check the printer to ensure
there is no corrosion.
Continued on next page
3-10
Page 71
ABL77 Service Manual Chapter 3: Troubleshooting
Reference Table, Continued
Problem Probable Cause Corrective Action
Power – no AC power A/C not reaching the analyzer
due to poor power cord connection or inactive power
Verify active power source and
ensure all cord connections are
secure
source
Battery charger malfunction See Battery Charger in Chapter
4. Replace as necessary.
Battery Control Board
malfunction
Replace the battery control
board. See Chapter 5,
Dismantling
Power – no battery power Battery Pack has aged Replace battery pack. See
Battery Pack in Chapter 4.
Battery Pack is not fully
See Battery Pack in Chapter 4
charged
Battery Charger malfunction See Battery Charger in Chapter
4. Replace as necessary
Power – intermittent power loss Battery Control Board
malfunction
Replace battery control board.
See Chapter 5, Dismantling
Printer – non functional Corrosion on printer module Replace printer module. See
Chapter 5, Dismantling.
Investigate source of fluid
contamination
Printer Board malfunction Replace board. See Chapter 5,
Dismantling
Printer Power Cable Check the connection. Replace
as necessary. See Chapter 6,
Replacements
Printer Data Cable Check the connection. Replace
as necessary. See Chapter 6,
Replacements
Thermal Printer malfunction Replace the thermal printer. See
Chapter 5, Dismantling
Continued on next page
3-11
Page 72
Chapter 3: Troubleshooting ABL77 Service Manual
Reference Table, Continued
Problem Probable Cause Corrective Action
Printer – Other
Printout light or missing pixels / not sensing paper: the thermal
printer can have a component malfunction or this problem can be the result of corrosion
Printout light / poor quality: thermal paper has been exposed to high temperature
Printout is un-readable or shows streaks: the printer data
cable pins are making contact with the screw that holds the printer module together.
Paper continually advances: BIOS setting incorrect
No printout: The printer must be
enabled in the software.
Replace printer. See Chapter 5,
Dismantling. If corrosion is
present, investigate source of
fluid contamination
Use a fresh roll of thermal
paper.
Replace the metal screw with a
nylon screw (REF: 023-013).
See Chapter 5 - Printer Module
Reset the BIOS setting in the
main CPU. See Chapter 4, Test
and Calibration Procedures.
This option is under Manager
Setup/ Report Setup / Print
Setup. The auto print feature
must be selected for patient,
calibration, and QC
Paper Jam
Paper installed backwards
Paper shifts to the left: This is a
normal printer response when a new roll of paper is installed
See Chapter 4, Test and Calibration Procedures
Instruct user on proper steps when replacing the paper roll which includes:
Advancing the leading edge of the paper out the printer door
Ensuring the operator removes paper by tearing up or down, not sideways
Orient the paper correctly. Refer to the diagram on the inner surface of the printer door
Advancing the paper roll 18 to 24 inches (~0.5 meter) when installing a new roll of paper will allow the paper to seat itself properly within the printer
3-12
Continued on next page
Page 73
ABL77 Service Manual Chapter 3: Troubleshooting
Reference Table, Continued
Problem Probable Cause Corrective Action
QC - results out of range
If failures are occurring with multiple sensor cassettes over time see "Calibration, Initialization, or QC failures
Incorrect handling Review the user's QC technique.
Stress the importance of thorough mixing and aspirating the sample immediately after opening.
with multiple sensor cassettes over time"
Incorrect temperature of
ampoule
Equilibrate ampoule to a known temperature. Ensure the correct QC Temperature value is entered into the analyzer
Incorrect ranges Use correct ranges. Ensure
proper ranges are entered into analyzer
Incorrect lot Verify the proper lot of QC
material is being tested
Faulty sensor function
Perform a manual calibration and repeat the QC measurement
If QC fails repeatedly, activate QC lockout, inactivate parameter or replace sensor cassette
Screen - blank/dark display
Description: the screen remains dark and blank despite signs of power (beeps, fan running etc)
LCD Display Cable Check the cable connection. If
securely connected, replace the cable. See Chapter 6, Replacements
Backlight Inverter Check the connection. If
securely connected, replace the backlight inverter. See Chapter 5, Dismantling
LCD Adapter Board
malfunction
Replace the LCD Adapter Board. See Chapter 5, Dismantling
Screen - blank / illuminated
Description: the screen lights up (glows but there is no text or graphics)
The application software has not successfully loaded due to a CPU failure
LCD Adapter Board malfunction
Replace the main CPU. See Chapter 5, Dismantling
Replace the LCD Adapter Board. See Chapter 5,
Dismantling
Continued on next page
3-13
Page 74
Chapter 3: Troubleshooting ABL77 Service Manual
Reference Table, Continued
Problem Probable Cause Corrective Action
Screen – Touch Screen non
functional
Description: there is no response when the touch screen is pressed
Main CPU Replace the main CPU. See
Chapter 5, Dismantling
LCD Display Cable Check the connection. Replace
as necessary. See Chapter 6, Replacements
Touch Screen Controller Board Replace the board. See Chapter
5, Dismantling
Touch Screen Panel malfunction Replace the panel. See Chapter
5, Dismantling
Touch Screen cable Check the connection. Replace
as necessary. See Chapter 6, Replacements
Barcode scanner connection or
cable
Remove barcode scanner. Shut down unit. Power up unit and verify the touch screen is operating correctly. If so, reconnect barcode scanner and verify the touch screen is still operating correctly. If not, replace the barcode scanner.
Screen – Touch Screen display malfunction
Description: variety of incorrect displays (e.g. missing pixels or no text)
Screen – Touch Screen Description: pressing in one area of the screen produces an effect on another area of the screen
LCD Display Panel Replace the panel. See Chapter
5, Dismantling
LCD Display Cable Check the connection. Replace
as necessary. See Chapter 6, Replacements
Screen is not properly calibrated Proceed to Chapter 4, Test and
Calibration Procedures to recalibrate the touch screen.
Continued on next page
3-14
Page 75
ABL77 Service Manual Chapter 3: Troubleshooting
Reference Table, Continued
Problem Probable Cause Corrective Action
Temperature Errors
SCi temperature circuit failure Replace the sensor cassette. See
Chapter 4, Test and Calibration Procedures for further
instructions on checking the heater circuit
Valve Board - worn or corroded
pin receptacle
Analog Board - component
failure
Waste Drain – waste fluids overflowing out the waste drain
Blockage in the waste drain tubing
during flush cycles
Replace the valve board. See Chapter 5, Dismantling
Replace the analog board. See Chapter 5, Dismantling
Flush the waste drain. See Waste Drain in Chapter 4 further in this chapter
Waste Pump Motor failure Replace the waste pump (a
"waste pump kit" includes the motor and pump head). See Chapter 5, Dismantling
Waste Pump head is worn Replace the waste pump head
Cal Pack waste port is blocked
Replace the cal pack
or pouch has been over-filled
3-15
Page 76
Chapter 3: Troubleshooting ABL77 Service Manual

System Messages

Introduction
System messages
Error # Message Interpretation Corrective Action
This section lists system messages, provides an interpretation and any recommended corrective action. This section groups system messages in categories. Reference error messages using its unique error number.
This table provides an interpretation of the message and suggests corrective action.
1000 System Errors
1001 Battery is low Battery voltage is below
minimum charge
Connect analyzer to A/C power. If battery does not re-charge refer to Chapter 4,
Test and Calibration Procedures to test the
battery. Replace the battery pack as necessary.
1002 Unable to allocate memory Insufficient memory Cycle the analyzer power
off then back on to re-boot the computer.
1003 Calculation error Unable to calculate
sensitivity values due to impossible mathematical values
Repeat the measurement. If this error persists, investigate possible sources of air contamination.
1004 The file can not be opened Memory is unavailable. Shutdown the analyzer
power through the Main Menu Shutdown button then turn the analyzer back on to re-boot the computer
1005 A file has been corrupted The file being read has been
corrupted.
Investigate a possible main CPU failure
1006 No data from analog board. The analyzer does not
recognize any signal from the analog board
3-16
Check cable connections. Replace the A/D ribbon cable between the CPU and analog board as necessary. Secondarily it may be necessary to replace the Analog board.
Continued on next page
Page 77
ABL77 Service Manual Chapter 3: Troubleshooting
Reference Table, Continued
Error # Message Interpretation Corrective Action
1100 Printer Errors
1101 Printer access denied General failure of the
operating system.
Shutdown the analyzer power through the Main Menu Shutdown button then turn the analyzer back on to re-boot the computer
1102 The printer is out of paper. The printer sensor does not
recognize the presence of paper
Load paper in the printer. If the printer is not sensing the presence of paper, replace the printer
1103 The printer head is up. The print head lever is in the
Lower the print head lever. up position, which lifts the printer head off the paper.
1104 Printer Error Hardware Error Replace the printer module
1105 Printer Error Hardware Error Replace the printer module
1106 Printer Error Hardware Error Replace the printer module
1200 Detection Errors
1201 Warning! Possible air in
sample.
Air-in-sample detection system has identified air in the measuring chamber during sample analysis. The air may affect sample
Carefully evaluate sample
analysis results for integrity
and acceptable clinical
correlation. Repeat analysis
suggested. results.
1202 Air in calibration fluid. Air-in-sample detection
system has identified air in the measuring chamber during the Cal1 phase of sample analysis. The air may affect sample result calculations
1203 Air detection is unavailable
for this analysis.
Air-in-sample detection circuit does not meet stability requirements for this analysis.
1204 Air detection is unavailable. Air-in-sample detection
circuit has malfunctioned or is unable to determine reference point due to air in calibration fluid prior to analysis.
Carefully evaluate sample
analysis results for integrity
and acceptable clinical
correlation. Perform a
manual calibration and
repeat the sample analysis.
Perform a manual
calibration and repeat the
sample analysis.
Perform a manual calibration and repeat the sample analysis
If the condition persists, re-install the current sensor cassette
Continued on next page
3-17
Page 78
Chapter 3: Troubleshooting ABL77 Service Manual
Reference Table, Continued
Error # Message Interpretation Corrective Action
1300 Sensor Errors
1301 Sensor cassette is not
connected.
Analyzer does not sense the presence of a sensor cassette
1302 Sample inlet flap is open The analyzer is preparing to
flush and the inlet probe is not down
1303 Sensor cassette not
initialized.
The analyzer has not successfully installed a sensor cassette either because the initialization process was not performed, was not allowed to be completed, or because the initialization was not successful.
Re-seat sensor cassette. If message persists after re­seating try another sensor cassette. If message continues, replace the valve board
Lower the sample inlet flap. If the inlet flap message persists with the flap down, refer to Chapter 4, Test and Calibration Procedures to investigate the opto sensor.
Complete the initialization. If a failure occurs, investigate the origin of the failure. The important first step is to identify whether the failure is due to a sensor failure or failure of the analyzer's fluidic system to properly transport the calibration fluid to the sensor cassette measuring chamber See Hardware and Calibration Lines in Chapter
4 Test and Calibration Procedures.
1304 Sensor cassette has expired. The currently installed
sensor cassette is beyond its in-use life or the cassette being installed is expired.
1305 No reportable parameters All parameters have been
inactivated or all active parameters are locked out due to the QC lockout feature.
3-18
Replace the sensor cassette
Re-activate desired parameters under Active Parameters, perform successful QC analysis, or turn off the QC lockout feature, as applicable.
Continued on next page
Page 79
ABL77 Service Manual Chapter 3: Troubleshooting
Reference Table, Continued
Error # Message Interpretation Corrective Action
1306 QC Level 1 is required
before analysis can be performed
QC Level 2 is required
before analysis can be performed
QC Level 3 is required
before analysis can be performed
QC Level 4 is required
before analysis can be performed
QC Hct Level 1 is required
before analysis can be performed
Mandatory QC interval has been reached. Analysis will not be allowed until BG/Lytes QC Level 1 analysis is performed
Mandatory QC interval has been reached. Analysis will not be allowed until BG/Lytes QC Level 2 analysis is performed.
Mandatory QC interval has been reached. Analysis will not be allowed until BG/Lytes QC Level 3 analysis is performed.
Mandatory QC interval has been reached. Analysis will not be allowed until BG/Lytes QC Level 4 analysis is performed.
Mandatory QC interval has been reached. Analysis will not be allowed until Hct QC Level 1 analysis is performed.
Perform the required BG/Lytes QC measurement or inactivate the required QC interval
Perform the required BG/Lytes QC measurement or inactivate the required QC interval.
Perform the required BG/Lytes QC measurement or inactivate the required QC interval.
Perform the required BG/Lytes QC measurement or inactivate the required QC interval.
Perform the required Hct QC measurement or inactivate the required QC interval.
QC Hct Level 2 is required
before analysis can be performed
Mandatory QC interval has been reached. Analysis will not be allowed until Hct QC Level 2 analysis is performed.
1308 No tests remaining. All allowable sample
analyses for the current sensor cassette have been performed.
1309 Hct: No End Point Stabilization criteria not met
for Hct
1310 This cassette has already
been installed. Please install a new sensor cassette.
The sensor cassette has been identified as previously installed.
Perform the required Hct QC measurement or inactivate the required QC interval.
Replace the sensor cassette.
Repeat measurement. If message persists, inactivate Hct or replace sensor cassette.
Install a new sensor cassette.
Continued on next page
3-19
Page 80
Chapter 3: Troubleshooting ABL77 Service Manual
Reference Table, Continued
Error # Message Interpretation Corrective Action
1311 Calibration required before
analysis.
Maximum allowable time between calibrations has been reached.
1312 QC Level 1 out of range At least one parameter fell
outside the acceptable range for the most recent BG/Lytes Level 1 QC measurement
QC Level 2 out of range At least one parameter fell
outside the acceptable range for the most recent BG/Lytes Level 2 QC measurement
QC Level 3 out of range At least one parameter fell
outside the acceptable range for the most recent BG/Lytes Level 3 QC measurement
Perform manual calibration or allow auto calibration to complete.
Perform a manual calibration and repeat the BG/Lytes Level 1 measurement
If results remain out of range inactivate this parameter or replace the sensor cassette.
Perform a manual calibration and repeat the BG/Lytes Level 2 measurement
If results remain out of range inactivate this parameter or replace the sensor cassette.
Perform a manual calibration and repeat the BG/Lytes Level 3 measurement
If results remain out of range inactivate this parameter or replace the sensor cassette.
QC Level 4 out of range At least one parameter fell
outside the acceptable range for the most recent BG/Lytes Level 4 QC measurement
3-20
Perform a manual calibration and repeat the BG/Lytes Level 4 measurement
If results remain out of range inactivate this parameter or replace the sensor cassette.
Continued on next page
Page 81
ABL77 Service Manual Chapter 3: Troubleshooting
Reference Table, Continued
Error # Message Interpretation Corrective Action
1312
continued
QC Hct Level 1 out of range At least one parameter fell
outside the acceptable range for the most recent Hct Level 1 QC measurement
QC Hct Level 2 out of range At least one parameter fell
outside the acceptable range for the most recent Hct Level 2 QC measurement
1400 Calibration Errors
1407 Initialization failed At least one sensitivity value
is outside the acceptable range during a cassette initialization
Perform a manual calibration and repeat the Hct Level 1 measurement
If results remain out of range inactivate this parameter or replace the sensor cassette.
Perform a manual calibration and repeat the Hct Level 2 measurement
If results remain out of range inactivate this parameter or replace the sensor cassette.
Follow the instructions on the screen. Investigate the origin of the failure. The important first step is to identify whether the failure is due to a sensor failure or failure of the analyzer's fluidic system to properly transport the calibration fluid to the sensor cassette measuring chamber
1408 Calibration failed At least one sensitivity value
is outside the acceptable range during a calibration
Follow the instructions on the screen. Investigate the origin of the failure. The important first step is to identify whether the failure is due to a sensor failure or failure of the analyzer's fluidic system to properly transport the calibration fluid to the sensor cassette measuring chamber
Continued on next page
3-21
Page 82
Chapter 3: Troubleshooting ABL77 Service Manual
Reference Table, Continued
Error # Message Interpretation Corrective Action
1500 Cal Pack Errors
1501 Cal pack not installed. There is no currently
installed cal pack on the analyzer
1502 Cal pack has expired. The currently installed cal
pack is beyond its in-use life.
1503 Cal 1 is empty. The current cal pack has
consumed all available Cal 1 solution.
1504 Cal 2 is empty. The current cal pack has
consumed all available Cal 2 solution.
1505 This cal pack has already
been installed. Please install a new cal pack.
1506 Cal pack creation date is a
future date
The cal pack being installed has previously been installed.
The date the cal pack was manufactured is a date in the future
1507 Insufficient Cal1 cycles There are not enough Cal1
cycles remaining to ensure completion of the desired action
Install a new cal pack.
Install a new cal pack.
Replace the cal pack.
Replace the cal pack.
Install a new cal pack.
Check the accuracy of the date in the analyzer.
Replace the cal pack
1508 Insufficient Cal2 cycles There are not enough Cal2
cycles remaining to ensure completion of the desired action
Replace the cal pack
Continued on next page
3-22
Page 83
ABL77 Service Manual Chapter 3: Troubleshooting
Reference Table, Continued
Error # Message Interpretation Corrective Action
1600 Temperature Errors
1601 Sample: Temperature Not
Stable
Temperature stabilization criteria not met during sample analysis
1602 Flush: Temperature Not
Stable
Temperature stabilization criteria not met during sample flush
1603 QC Sample: Temperature
Not Stable
Temperature stabilization criteria not met during QC sample analysis
1604 QC Flush: Temperature Not
Stable
Temperature stabilization criteria not met during QC sample flush.
1605 Cal 1: Temperature Not
Stable
Temperature stabilization criteria not met during Cal 1 measurement
1606 Cal 2: Temperature Not
Stable
Temperature stabilization criteria not met during Cal 2 measurement
1700 Upload Errors
1701 Connection timed out Maximum time exceeded for
uploading data to remote host computer
See Chapter 4, Test and Calibration Procedures to investigate the heater circuit
See Chapter 4, Test and Calibration Procedures to investigate the heater circuit
See Chapter 4, Test and Calibration Procedures to investigate the heater circuit
See Chapter 4, Test and Calibration Procedures to investigate the heater circuit
See Chapter 4, Test and Calibration Procedures to investigate the heater circuit
See Chapter 4, Test and Calibration Procedures to investigate the heater circuit
Check connections. Verify the remote host system is working properly.
1702 Receiver not ready Remote host system has not
acknowledged it is ready to receive data.
1703 Receiver cancelled Remote host system has
cancelled the communication.
Wait a few moments then retry. If problem persists, verify the remote host system is working properly
Wait a few moments then retry. If problem persists, check connections. Verify the remote host system is working properly.
Continued on next page
3-23
Page 84
Chapter 3: Troubleshooting ABL77 Service Manual
Reference Table, Continued
Error # Message Interpretation Corrective Action
1704 Session terminated The analyzer terminated the
upload session due to unsuccessful transmission after 6 attempts
1705 Receiver was not found Communication signal was
sent from the analyzer but the remote host system was not found
1706 The network settings are
invalid
The network address entered is not valid
The address already exists in the system
Using DHCP and the system was not connected when turned on so an IP address was not automatically assigned
Check connections. Verify the remote host system is working properly.
Verify the proper host address or name and port number in Communications Setup
Verify proper operation of the remote host system
Review the addresses entered into the analyzer and remote host system
If using DHCP, check the connections and re-boot
If using a Computer Name, check the integrity of the connection (it is not necessary to re-boot)
Using a Computer Name and the system was not connected so this name could not be interpreted as an IP address
1707 Network error A general hardware or
software related problem
1708 Query failed The patient query found no
matches within the queried external database
1709 This query command is not
enabled on Radiance
This query feature in Radiance must be enabled
Evaluate the status of the remote host system
Shutdown the analyzer power through the Main Menu Shutdown button then turn the analyzer back on to re-boot the computer
If it persists, contact your IT resource
Check the data input for accuracy
Execute an internal database search
Enable the external query command in Radiance.
3-24
Continued on next page
Page 85
ABL77 Service Manual Chapter 3: Troubleshooting
Reference Table, Continued
Error # Message Interpretation Corrective Action
1710 Internal query failed The patient query found no
matches within the analyzer's database
1711 Connection terminated by
user
1712 No connection license
available
The external query was cancelled by the user
The connection license has not been acquired
1713 Wrong RAML version There is an incompatible
RAML version being used in the Radiance communication
1800 System Errors
1801 Analyzer has been locked The analyzer is in a locked
condition
Check the data input for accuracy
Check the network load for excessive delays
Contact the Radiometer Sales Representative
Review the RAML version and contact the appropriate Radiance representative to install a compatible version.
The system can be unlocked either on the analyzer (Options / Data
Communications / Analyzer Lock) or if
connected to Radiance the system can be unlocked within Radiance (see Analyzer Control).
3000 Input Errors
3001 Input error The value(s) input are
beyond the acceptable input range.
3002 Invalid password Password entered was not
found in the User List
3003 Sensor cassette model is
incompatible with the analyzer parameter configuration.
A sensor cassette is being installed with a parameter configuration incompatible with the analyzer's allowable parameters.
3004 Data Bits can not be set to 5
while the Stop bits is set to
2.
Under Communication Setup, only certain combinations of data bits and stop bits can be selected.
Re-enter values within the acceptable input range.
Re-enter password. If problem persists, review password list and add password as appropriate.
Install the correct model of sensor cassette, matching it with the analyzer configuration.
Select the proper combination of data bits to stop bits
Continued on next page
3-25
Page 86
Chapter 3: Troubleshooting ABL77 Service Manual
Reference Table, Continued
Error # Message Interpretation Corrective Action
3005 Data bits can not be set to 6,
7, or 8 while the Stop bits is set to 1.5.
Under Communication Setup, only certain combinations of data bits and stop bits can be selected.
3006 Please select the settings
you wish to transfer.
Under Settings Transfer, the user must select at least one setting to transfer
3007 Order date cannot be a
future date.
Order Date field under Patient Information had a date entered that is beyond the current date.
3008 Invalid bar code The bar code has been
entered in an unacceptable format.
3011 Incorrect barcode scanned The barcode scanned is not
for the ABL77,
3012 Bar code scan unsuccessful The scanning process did
not result in a complete bar code scan.
Select the proper combination of data bits to stop bits.
Select at least one setting before attempting to transfer settings.
Re-enter the correct order date. If necessary, verify the correct current date on the analyzer.
Re-enter the bar code correctly.
Ensure you are scanning the ABL77 (not ABL70) barcode on the cal pack.
Re-scan the barcode. Ensure you are scanning the proper bar code and the red scan light covers all barcode lines.
3013 Invalid solution ID The Radiometer solution ID
within the QC barcode is not compatible for use with the ABL77.
3014 Password Expired The user password has
expired.
Ensure you are using the correct quality control material and scanning the correct, associated bar code.
Reset the user password in the User Setup screen.
Continued on next page
3-26
Page 87
ABL77 Service Manual Chapter 3: Troubleshooting
Reference Table, Continued
Additional messages
This table provides an interpretation of additional messages that may appear.
Message Interpretation Corrective Action
I/A Parameter is inactive. Turned off
by user.
L/O Parameter locked out due to QC
results outside acceptable limits.
N/A Not applicable. Appears in the
Hct field when Other Fluids is selected for sample type.
N/D Parameter not derived because
measurement parameters necessary for this calculation are not available.
N/E No endpoint. Stabilization criteria
not met for one or more parameters during sample measurement or calibration.
N/I Not installed. A BG/Hct – only or
Lytes – only sensor cassette has been installed on a full panel analyzer.
No action required.
Perform a successful QC measurement or deactivate the QC lockout feature.
If Hct results are desired, repeat the analysis selecting any sample type other than Other Fluids.
Activate necessary measured parameters or select necessary measured parameters for reporting.
No endpoints suggest air in the measuring chamber. Investigate possible sources of air. Then perform a manual calibration and repeat the measurement.
If all parameters are desired, install a full panel sensor cassette.
N/R Not reported. This parameter was
not selected for reporting by the user.
N/C Not calculated. Hct cannot be
calculated due to a failed Na+ calibration
*** Measured parameter value is out
of the analyzer display range.
# # # Sensor response is out of the
established response range.
Select all desired parameters from the parameter profile and repeat the analysis.
Repeat the calibration. If the
Na
+
failure persists, either continue testing without Hct or replace the sensor cassette.
Evaluate the integrity of the sample. Perform a manual calibration and repeat the sample analysis.
Initiate a manual calibration. Repeat the sample analysis.
Continued on next page
3-27
Page 88
Chapter 3: Troubleshooting ABL77 Service Manual
Reference Table, Continued
Additional messages
(continued)
Message Interpretation Corrective Action
The sample analysis result is above or below the user-defined
No action required for sample analysis.
reference value.
The quality control result is above
or below the acceptable QC range.
The sample analysis result has fallen into the user-defined critical range.
The sample analysis result has fallen outside the reportable range.
? Air was detected in the sample or
the associated sample flush solution.
The last QC measurement for this parameter was not acceptable. The QC Lockout feature was not activated
Perform a manual calibration then repeat the QC measurement.
No action required.
Repeat the analysis if appropriate.
If air was detected in the sample, repeat the analysis. If air is detected in the flush solution, perform a manual calibration then repeat the analysis.
Perform a successful QC measurement for the failed parameter and level.
3-28
Page 89
ABL77 Service Manual Chapter 4: Test and Calibration Procedures
4. Test and Calibration Procedures

Overview

Introduction
WARNING/ CAUTION:
This chapter details the procedures for checking and adjusting the analyzer circuitry.
The tools required for carrying out the procedures are listed in Chapter 1; Test
Equipment and Tools.
CAUTION: Follow legal requirements and local rules for safe work practices with
chemicals.
CAUTION: Working with blood gas analyzers may result in contact with blood remnants and with harsh disinfectants. During the various procedures wear suitable protection gear (gloves, face protection, and protective body clothing), following legal requirements and local rules for safe work practices.
CAUTION: The gloves must be free of pinholes and preferably be puncture resistant. Please consult the glove manufacturer for further information.
CAUTION: If contaminated material comes into contact with any lesion on the body, seek medical advice.
Continued on next page
Rev. F 4-1
Page 90
Chapter 4: Test and Calibration Procedures ABL77 Service Manual
Overview, Continued

Contents

This chapter contains the following topics.
Electrostatic discharge protection
Overview............................................................................................... 4-1
Verify the Fluid Transport System........................................................ 4-3
Calibration Lines Procedure.................................................................. 4-5
Testing the Valve Connections ........................................................... 4-11
Valve Test Procedures......................................................................... 4-13
Check Valve Fluidics System ................................................ 4-15
Original Fluidics System........................................................ 4-19
Waste Line .......................................................................................... 4-26
5 Volt Power Output Setting............................................................... 4-29
SC/Hematocrit Circuits ....................................................................... 4-31
Electronic Zero Offset......................................................................... 4-34
Heater Circuit...................................................................................... 4-36
Inlet Flap Sensor ................................................................................. 4-39
Paper Jam ............................................................................................ 4-40
Battery Charger ................................................................................... 4-43
Battery Pack ........................................................................................ 4-45
Touch Screen Panel Calibration.......................................................... 4-48
Reprogramming the CPU BIOS Settings............................................ 4-50
Re-programming the Barcode Scanner ............................................... 4-52
Resetting the Analyzer Serial Number................................................ 4-56
To ensure the reliability and durability of the analyzer, ESD protection precautions must always be taken when handling or exposing ESD sensitive parts. Parts such as printed circuit boards may be damaged if handled incorrectly during storage and service. The damage is not visible, and the damaged parts may not fail immediately and can take up to several months before damage is apparent.
Calibration and measurement
4-2 Rev. F
While performing certain calibrations and measurements, the analyzer must be fully assembled. A cal pack and sensor cassette may be required.
Page 91
ABL77 Service Manual Chapter 4: Test and Calibration Procedures

Verify the Fluid Transport System

Introduction
Verify the fluidics system
The Hardware screen may be used to manually initiate fluidic events. Observing these events can help identify whether the analyzer’s fluidic system is effectively transporting calibration fluids from the cal pack to the sensor cassette measuring chamber.
Follow the steps below to verify proper flow of calibration fluids.
Step Action
Ensure a cal pack and a sensor cassette are installed on the analyzer.
1.
2.
Select Options from the Main Menu
Enter a Password that will allow access to the Hardware screen.
Access to this screen requires a password. If a user password is not available, the generic MK592 may be used.
NOTE: This password is for Radiometer use only. Do not share this password with customers.
Select Hardware from the Options menu.
RESULT: The following screen will appear.
Press the Aspirate button to draw air into the sensor cassette
3.
measuring chamber.
Carefully observe the measuring chamber. Ensure you can recognize the different visual characteristics of the measuring chamber when filled with fluid versus air.
Continued on next page
Rev. F 4-3
Page 92
Chapter 4: Test and Calibration Procedures ABL77 Service Manual
Verify the Fluid Transport System, Continued
Verify the fluidics system
(continued)
Step Action
Press the Pump Cal1 button to flush Cal1 solution through the
4.
measuring chamber.
Carefully observe the measuring chamber. The chamber should be completely filled with fluid. There should be no air bubbles of any size.
If air is observed, flush Cal1 solution a second time. While the pump wheel is turning, snap repeatedly on the pump tubing to dislodge any air bubbles in the system.
If fluid is being transported, it is also possible to see the fluid dripping from the tip of the inlet probe down into the waste receptacle.
Press the Aspirate button a second time to remove all Cal1 solution
5.
Press the Pump Cal2 to flush Cal2 solution through the measuring
6.
chamber.
Carefully observe the measuring chamber. The chamber should be completely filled with fluid. There should be no air bubbles of any size.
If air is observed, flush Cal2 solution a second time. While the pump wheel is turning, snap repeatedly on the pump tubing to dislodge any air bubbles in the system.
7.
If both Cal1 and Cal2 solution fill the measuring chamber without air, the fluid transport is functioning properly
If one or both Cal solutions are mixed with air or do not flow at all, continue to investigate the fluidic transport system. This includes investigating:
The cal pack. Ensure it is properly seated on the manifold and has not been installed on more than one analyzer. Installing a cal pack on multiple analyzers will cause inaccurate tracking of cal cycles and cause cal solutions to be completely depleted without warning.
The sensor cassette. Ensure it is properly seated onto the analyzer and free of blockage
The internal fluidic pathways. To check, perform the following Calibration Lines procedure.
4-4 Rev. F
Page 93
ABL77 Service Manual Chapter 4: Test and Calibration Procedures

Calibration Lines Procedure

Introduction
Calibration lines procedure
The Calibration Lines procedure is used to evaluate the function of the internal tubing and valves that are used to transport calibration solution from the cal pack to the sensor cassette measuring chamber. If the calibration lines cannot be flushed, one or more valves may be functioning improperly.
Follow the steps below to perform the Calibration Lines procedure.
Step Action
Select Options from the Main Menu.
1.
2.
Enter a Password as necessary.
Press OK.
Press Fluidics.
3.
RESULT: The following screen will appear.
4.
From the pop-up window press Calibration Lines.
Press Yes to continue.
Continued on next page
Rev. F 4-5
Page 94
Chapter 4: Test and Calibration Procedures ABL77 Service Manual
Calibration Lines Procedure, Continued
Calibration lines procedure
(continued)
Step Action
If a sensor cassette is…
5.
Attached
The analyzer will remove any fluid, then prompt the operator to remove the sensor cassette. The analyzer will sense when the cassette has been removed and will automatically proceed to the next screen.
not attached
The analyzer will automatically proceed to the next step.
RESULT: The following screen will appear.
Ensure there is a cal pack on the analyzer to collect the fluids exiting
6.
the waste line.
Introduce 1 cc of Cleaning Solution (REF: 943-906) into the waste
7.
drain (black opening located directly below the inlet probe when a cassette is in place).
Continued on next page
4-6 Rev. F
Page 95
ABL77 Service Manual Chapter 4: Test and Calibration Procedures
Calibration Lines Procedure, Continued
Calibration lines procedure
(continued)
Step Action
8.
Activate the waste pump by pressing Pump. This will clear the fluid, carrying it through the waste line and into the cal pack.
Press Continue.
RESULT: The following screen will appear.
9.
Remove the cal pack and press Continue.
10.
CAUTION: The cal pack must
be removed before proceeding. Performing the following steps with a cal pack in place will result in contamination of the calibration solution.
RESULT: The following screen will appear.
11.
Continued on next page
Rev. F 4-7
Page 96
Chapter 4: Test and Calibration Procedures ABL77 Service Manual
Calibration Lines Procedure, Continued
Calibration lines procedure
(continued)
Step Action
12.
Fill a syringe (with luer adapter) from a Sample Path Obstruction
Tools kit (REF: 905-674) with deionized water or sterile water.
Attach the syringe to the cassette luer of the analyzer (see
Figure 4-1).
Figure 4-1
Be prepared to collect fluids exiting the cal pack manifold at the rear
13.
of the analyzer (see Figure 4-2).
Figure 4-2
Continued on next page
4-8 Rev. F
Page 97
ABL77 Service Manual Chapter 4: Test and Calibration Procedures
Calibration Lines Procedure, Continued
Calibration lines procedure
(continued)
Step Action
Press Cal1.
14.
NOTE: This will activate internal valves and open a pathway between the cassette luer and the Cal1 manifold luer.
RESULT: The following screen will appear.
While the screen displays “Cal1 Open”, GENTLY introduce water
15.
into the luer. The Cal1 lines will remain open for 10 seconds. The water should exit the Cal1 manifold luer.
NOTE: If any resistance is felt while trying to introduce water into this line DO NOT force the syringe plunger. This will cause internal tubing connections to weaken and possibly disconnect.
The operator may need to repeat this flush attempt to soak free any
16.
dried calibrant blockage.
NOTE: If unable to open up a blockage by flushing, the cause of the blockage is likely a valve. See Valve Test Procedure further in this chapter and replace the valves as necessary.
Continued on next page
Rev. F 4-9
Page 98
Chapter 4: Test and Calibration Procedures ABL77 Service Manual
Calibration Lines Procedure, Continued
Calibration lines procedure
(continued)
Step Action
Press Continue.
17.
RESULT: The following screen will appear.
18.
Refill the syringe with deionized or sterile water and flush the Cal2
line in a similar manner, pressing Cal2 to open a pathway between the cassette luer and the Cal2 manifold luer.
Repeat as necessary.
Press OK when flushing is complete.
19.
Re-attach the cal pack and sensor cassette (if one is currently
installed).
Press OK.
RESULT: The system will return to the Options menu.
Perform a manual calibration to ensure proper function prior to sample
20.
analysis.
4-10 Rev. F
Page 99
ABL77 Service Manual Chapter 4: Test and Calibration Procedures

Testing the Valve Connections

Introduction
NOTE:
Follow these steps to test the valve connections after replacing the valve board tubing.
NOTE: Verify that the manifold is properly installed to prevent other fluidic problems from arising.
NOTE: Flush water or air from a syringe through the sensor cassette luer ONLY while the waste pump is running. Flushing without activating the waste pump can disconnect internal tubing.
Step Action
Apply power to the analyzer.
1.
Select the Hardware screen from the Options menu. Use password
2.
MK592 when entering the Options screen to gain access to the Hardware screen.
NOTE: This password is for Radiometer use only. Do not share this password with customers.
Insert an index card (or equivalent) into the door slot. The door
3.
indicator in the Hardware screen will change from 1 to 0. This will simulate the presence of a sensor cassette and allow the Pump Cal1 and Pump Cal2 buttons to function.
Place a soft cloth or gauze over the manifold ports to collect the water
4.
from the following steps.
NOTE: Be prepared to collect all fluids that exit the manifold.
NOTE: Observe the valve board and connections for leaks during
this process.
Fill a syringe, with luer adapter, with DI water.
5.
a) Place the syringe at the sensor cassette fitting luer.
6.
b) Press Aspirate and flush the waste lines with de-ionized water.
Fluid should exit the waste port of the manifold.
a) Refill the syringe with DI water and replace it onto the cassette
7.
luer
b) Press Pump Cal1
c) Flush the Cal1 line. Fluid should exit the Cal1 port of the
manifold.
a) Refill the syringe with DI water and replace it onto the cassette
8.
luer
b) Press Pump Cal2
c) Flush the Cal2 line. Fluid should exit the Cal2 port of the
manifold.
Continued on next page
Rev. F 4-11
Page 100
Chapter 4: Test and Calibration Procedures ABL77 Service Manual
Testing the Valve Connections, Continued
NOTE:
(continued)
Step Action
Remove the index card, or equivalent, from the door slot and the soft
9.
cloth or gauze from the manifold ports.
The calibration resistors may have been inadvertently affected during
10.
this procedure. To verify proper calibration settings, Refer to the procedure “Calibrating the SC and Hct Circuits”.
Shut down and re-assemble the analyzer.
11.
4-12 Rev. F
Loading...