Note to the users of the EML105, ABL5xx, ABL SYSTEM 6xx,
ABL7xx Series, ABL800 FLEX and ABL800 BASIC analyzers
Introduction
Brief overview
of the change
This note to users outlines a change in the operator’s and reference manual for your
EML105, ABL5xx, ABL SYSTEM 6xx, ABL7xx Series, and/or ABL800 FLEX and
ABL800 BASIC analyzer.
Operator’s manual:
Limitations of use and known interfering substances:
CAUTION - Known interfering substances
Substance Interference
–
ClO
(drugs) For ClO
4
– ,
interference on cCa2+ (1.25 mmol/L
4
level) has been detected:
2+
(1.25 mmol/L level): 0.20*.
cCa
Technical
documentation
Instructions to
user
* Depending on the pH level
Reference manual:
Change/Description
Interference on
For ClO
detected. The interference results for ClO
– ,
interference on cCa2+ (1.25 mmol/L level) has been
4
–
are as follows:
4
Interference on…
+
Substance Test Conc. cK+
(4 mmol/L
level)
–
ClO
1.5 mmol/L - -
4
* Depending on the pH level
A "-" indicates that interference has not been measured on the respective parameter.
cNa
(150 mmol/L
level)
(1.25 mmol/L
2+
cCa
level)
0.20*
(110 mmol/L
cCl
level)
8-30
The manual will be updated with the above information as part of the next manual update.
Please place this note to the users in the binder of your manual.
Cleaning
solution with
cleaning
additive installation
The manuals for the ABL700 Series analyzers must be upgraded with regard to the
information on cleaning solution.
The procedure for adding cleaning additive has been changed for the abovementioned analyzers and from now on the following instructions must be followed:
1.
2.
Remove the foil from the DosiCapZip and unscrew it (Figs. 1+2).
Turn the DosiCapZip upside down and screw it onto the container again
(Figs. 3+4).
TCAUTIONT: TIf the contents of the DosiCapZip or the container have
been spilt by accident, both the container and the DosiCapZip should be
3.
4.
discarded to prevent incorrect concentrations in the solution.
Invert the container at least 20 times to dissolve the additive (Fig. 5).
Place the container horizontally so that the solution may enter the
T
DosiCapZip and leave it for 3 minutes (Fig. 6).
5.
Invert the container again at least 20 times to fully dissolve the additive
(Fig. 7).
6.
7.
Unscrew the lid from the new solution container.
Remove the used solution container by holding it on the sides and
pulling.
8.
9.
Scan the barcode of the new solution, using the barcode reader.
Place the new solution container in position on the analyzer and push it
firmly onto the connector as far as possible.
10.
For the ABL700 Series analyzers, sw. 3.836: Press Restart to restart the
analyzer.
For the ABL700 Series analyzers, sw. 6.00: Press Restart and Accept to
Cleaning
solution with
cleaning
additive –
general
information
Cleaning Solution 175 mL 944-123 S7375
T
Use: For cleaning the measuring system of the ABL700 Series analyzers.
Contains: Salts, buffer, anticoagulant, preservatives, surfactants and enzyme.
Safety Data Sheet may be obtained from your local distributor.
Item Code No. Type
Storage: At 2-10 °C.
Stability in use: The Cleaning Solution with the Cleaning Additive is stable for 2
months in use.
Analyzer: Perform cleaning every 8th hour.
T CAUTION – Risk of personal injury
TDo not breathe dust (S22). Avoid contact with skin (S24). Irritating to eyes and
skin (R36/38). Wear suitable gloves (S37). May cause sensitization by inhalation
and skin contact (R42/43). In case of accident or if you feel unwell, seek medical
advice immediately (show the label where possible) (S45).
Texts no longer
valid in the
manuals
Due to new cleaning solutions, information about Cleaning solution S7370 and
Cleaning additive S5370 is no longer valid.
Below is a list of the sections in the manual that must be ignored.
ABL700 Series reference manual:
Technical
documentation
Instructions to
user
Chapter 7: Solutions and gas mixtures:
S7370 Cleaning Solution and
S5370 Cleaning additive:
Text no longer relevant.
See instruction about the new Cleaning
Solution S7375 above instead.
Index: Cleaning Additive references no longer
relevant.
The manuals for the ABL700 Series analyzers will be updated with the above
information when reprinted.
This update kit includes a note to the user with the changes and a new date of issue
page of the manual. Please place the note to the user in the binder of the ABL700
Series reference manual and replace the date of issue page with the new
corresponding page, then discard the old page.
Reference
ABL700 Series
1. Introduction
2. Electrodes
3. The Optical System
Contents
Manual
4. User-defined Corrections
5. Performance Characteristics
6. Parameters
7. Solutions
8. Interfacing Facilities
Index
Date of Issue
SYSTEM PERFORMANCE AND WARRANTY DISCLAIM
Radiometer cannot provide or verify instrument performance characteristics and accept
warranty claims or product liability claims if the recommended procedures are not carried out,
if accessories other than those recommended by Radiometer are used, or if instrument
repairs are not carried out by authorized service representatives.
The instructions given in the Operator’s Manual for the ABL700 Series must be observed in
order to ensure proper instrument performance, and to avoid electrical hazards.
TRADEMARKS
ABL™, BMS™, FLM™, CMT™, Deep Picture™, QUALICHECK™ and RADIOMETER™ are
trademarks of Radiometer Medical ApS, Denmark.
ABL is registered in the USA.
QUALICHECK is registered in the USA and in some other countries.
COPYRIGHT
The contents of this document may not be reproduced in any form or communicated to any
third party without the prior written consent of Radiometer Medical ApS.
While every effort is made to ensure the correctness of the information provided in this
document Radiometer Medical ApS assumes no responsibility for errors or omissions which
nevertheless may occur.
Connecting a Mouse ................................................................................... 8-2
Connecting an Alphanumeric Keyboard..................................................... 8-3
Connecting the Bar Code Reader................................................................ 8-4
Connecting a Network ................................................................................ 8-6
Index
Date of Issue
ii
ABL700 Series Reference Manual 1. Introduction
1. Introduction
Overview
This section gives an introduction to the documentation that accompanies your
ABL700 Series analyzer. It describes how this particular manual is organized and
explains the different notices that appear in it.
Contents
This chapter contains the following topics.
ABL700 Series Documentation ....................................................................... 1-2
Warnings/Cautions and Notes.......................................................................... 1-3
1-1
1. Introduction ABL700 Series Reference Manual
ABL700 Series Documentation
ABL700 Series
Analyzers
Documentation
The documentation that accompanies the ABL700 Series analyzers covers the
series in general - each possible electrode combination is not considered
individually.
The table below describes the documentation that comes with each analyzer.
Documentation Description
Design of
Manual
The Operator’s
Manual
Contains all the information required for everyday
operation of the analyzer.
Describes the functions of the analyzer and how to set it up
according to customer needs and requirements.
Explains error messages and gives troubleshooting
procedures.
Contains ordering information
The Reference
Manual
Gives detailed information about the operating principles
of the analyzer.
Describes the measuring and calibrating principles.
Lists all the parameters.
Gives the equations from which the derived parameters are
calculated.
Gives information about how the performance of the
analyzer is tested.
On-line Help Summarizes the information found in the Operator’s
Manual.
Gives hands-on help at the analyzer.
Depending on the set up of your analyzer, the entire Reference Manual may not be
applicable to it. However the manual is designed in such a way that it is easy to
disregard or remove the sections that are not relevant to your instrument.
1-2
ABL700 Series Reference Manual 1. Introduction
Warnings/Cautions and Notes
Definitions
List of
WARNING/
CAUTION
Notices
The following table indicates the type of information given in warnings, cautions,
and notes:
Notice Definition
WARNING
Warnings alert users to potentially serious outcomes to
selves or to the patient (such as death, injury, or serious
them
adverse events).
PRECAUTION
Precautions alert users to exercise the special care necessary
for safe and effective use of the device. They
may include
actions to be taken to avoid effects situations on patients or
users that may not be potentially life threatening or result in
serious injury , but about which the user should be aware.
Precautions may also alert the user to adverse effects on the
device caused by use or misuse, and the care required to
avoid such effects.
NOTE
Notes give practical information.
All WARNING/CAUTION notices that appear in this manual, are listed below.
S5370 Cleaning Additive: May cause sensitization by inhalation and skin
•
contact). Do
not breathe dust. Avoid contact with skin. Wear suitable gloves.
In case of accident or if you feel unwell, seek medical advice immediately
(show the label where possible).
Gas Mixtures: Not for drug use. High pressure gas. Do not puncture. Do not
•
store near heat or open flame - exposure to temperatures above 52 C
o
(125 F) may cause contents to vent or cause bursting.
Not for inhalation. Avoid
breathing gas - mixtures containing carbon dioxide
o
can increase respiration and heart rate. Gas mixtures containing less than
19.5 % oxygen can cause rapid suffocation.
Store with adequate ventilation. Avoid contact with oil and grease.
Only use with equipment rated for cylinder pressure.
Use in accordance with Safety Data Sheet.
1-3
Introduction
Contents
2. Electrodes
This chapter describes the construction, measuring principle and calibration
process for each of the electrodes in the ABL700 Series analyzers.
General sections covering the background theory used for measurements and
calibrations are also presented here.
This chapter contains the following topics.
General Construction ....................................................................................... 2-2
General Measuring Principles .......................................................................... 2-3
In this manual and other Radiometer literature, the term electrode refers to the
whole sensor unit, i.e. both the electrode and the electrode jacket. Radiometer
electrodes are cordless, thereby limiting the level of noise picked up during the
measuring process. The electrical signals from the electrodes are amplified by
preamplifiers placed in each module.
A generalized diagram of a Radiometer electrode is given below.
Electrolyte solution
Electrical contact
Color-coded ring
Electrode jacket
Membrane
The main electrode parts are described below.
Part Description
Electrical contact Provides electrical contact between the electrode and the
analyzer.
Color-coded ring Marks each electrode for easy recognition.
Electrode jacket Holds the electrolyte solution and membrane, and protects
the electrode.
Membrane A thin sheet-like material to separate the sample from the
electrode, that differentiates between the substances
allowed to pass through it towards the electrode.
2-2
Electrolyte
solution
A conducting solution to provide an electric contact
between the electrode and the sample (also known as a
salt-bridge solution).
More specific descriptions of the electrodes are found under the appropriate
electrode titles in this chapter.
ABL700 Series Reference Manual 2. Electrodes
General Measuring Principles
Introduction
There are two different measuring principles employed for electrodes in the
ABL700 Series.
• Potentiometry: The potential of an electrode chain is recorded using a
voltmeter, and related to the concentration of the sample (the Nernst equation).
• Amperometry: The magnitude of an electrical current flowing through an
electrode chain, which is in turn proportional to the concentration of the
substance being oxidized or reduced at an electrode in the chain.
These two measuring principles are described in detail on the following pages.
Potentiometric
Method
An electrode chain describes an electrical circuit consisting of a sample, electrode,
reference electrode, voltmeter, membranes, and electrolyte solutions.
Voltmeter
V
Reference
electrode
Electrolyte
solution
Membrane
Sample
Electrolyte
solution
Membrane
Electrode
Every element in the electrode chain contributes a voltage to the total potential
drop through the chain. Thus:
• When immersed in the appropriate electrolyte solution, both electrodes have
separate potentials.
• The membrane junctions between the sample and electrolyte solutions also have
separate potentials.
The complete electrode chain potential therefore, is the sum of these separate
potentials and is the quantity measured by the voltmeter.
E = E – E
totalsampleRef
where the final unknown potential (E
electrode chain potential (E
) and the reference potential (E
total
) can be calculated knowing the total
sample
that is constant
Ref
between two subsequent calibrations).
Continued on next page
2-3
2. Electrodes ABL700 Series Reference Manual
General Measuring Principles, Continued
Potentiometric
Method
(continued)
Having measured the unknown potential (E
applied to determine the activity (a
EE
) of the species under study:
x
=+
sample
0
where:
E0
R =
T
n
F =
a
= standard electrode potential
gas constant (8.3143 JK
= absolute temperature (310 K (37
= charge on the ion
Faraday constant (96487 C mol
x
= activity of
x
−1
mol−1)
The Nernst equation is rearranged to express the activity as a function of the
potential
E
. Having measured E
sample
the activity can be calculated since all
sample
other quantities are already known. Finally the analyzer converts activity to
concentration.
), the Nernst equation is then
sample
23R
.
T
log
a
n
−1
F
o
C ))
)
x
Strictly speaking, in potentiometry the potential of an electrode chain or the
magnitude of current flowing through an electrical chain is related to the activity of
a substance, and not its concentration.
Activity expresses the ‘effective concentration’ of a species, taking non-ideality of
the medium into account.
Activity and concentration are related by the following equation:
a
= γ c
x
x
where:
a
= the activity of the species x
x
= the activity coefficient of species x under the measurement conditions
γ
(for ideal systems
c
= the concentration of species x (mmol/L)
x
γ = 1)
NOTE:
To be exact, activity is related to the molality of species x, i.e., the number
of mmoles per kg of solvent. However molality is converted to concentration
(molarity).
ABL700 Series analyzers automatically convert activities into concentrations
[1].
The term concentration is therefore used in explanations of the measuring
principles for each of the electrodes further on in this chapter.
2-4
The potentiometric measuring principle is applied in the pH,
pCO
electrodes. It is slightly different for the
electrode, however, since the Nernst
2
equation is not directly applied.
pCO
, and electrolyte
2
Continued on next page
ABL700 Series Reference Manual 2. Electrodes
A
General Measuring Principles, Continued
Amperometric
Method
The electrode chain in amperometric measurements consists of the sample, the two
electrodes (anode and cathode), an amperemeter, a voltage source, the membranes,
and the electrolyte solutions.
Applied voltage
AnodeCathode
Electrolyte solution
Sample
mperemeter
Membrane
Part Function
Cathode Negative electrode where a reduction reaction occurs and
electrons are consumed.
Anode Positive electrode where an oxidation reaction occurs and
electrons are released.
Electrolyte
Provides electrical contact between the anode and cathode.
solution
Membrane Allows the appropriate molecules to pass through from the
sample.
Sample Contacts the membrane.
Applied voltage Applies the necessary potential for the reduction or oxidation
reaction under study.
Amperemeter Measures the current flowing through the circuit.
To simplify the description of the measuring process in an amperometric electrode,
we make the following assumptions:
• there is a species A in the sample which is reduced at the cathode to A
• there is a species X in the electrolyte which is oxidized at the anode to X
−
.
+
.
Continued on next page
2-5
2. Electrodes ABL700 Series Reference Manual
General Measuring Principles, Continued
Amperometric
Method
(continued)
The membrane is selective to the species A, allowing no other species but it to pass
through from the sample into the electrolyte solution.
As an appropriate potential is applied across the electrodes, the species
A is
reduced at the cathode according to the following reaction:
A + e
−
→ A
−
The reduction of A produces a flow of electrons, i.e. an electrical current.
To complete the electrical circuit an oxidation reaction where electrons are
released is necessary. Therefore species
X is oxidized according to the following
reaction:
+
X
→ X
+ e
−
The magnitude of the current flowing through the circuit is proportional to the
concentration of the species being reduced, in this case species
thereby automatically calculates the concentration of
A in the sample.
A. The analyzer
The amperometric measuring principle is applied in the
electrodes.
pO
, glucose, and lactate
2
2-6
ABL700 Series Reference Manual 2. Electrodes
−
−
−
Calibration
Actual Electrode
Condition
Calibration Line
The electrodes are active elements and must be calibrated regularly as the signals
from the electrodes change with, e.g. age or deposits on the membrane.
Calibration relates the electrode signals during the calibration sequence to the
values of the calibrating solutions and must be performed at regular intervals so
that the accuracy can be constantly refined after inevitable minor changes in the
electrodes’ behavior.
Actual electrode condition is described by status/zero point and sensitivity and
compared with theoretical conditions for an "ideal" electrode. In addition to status
and sensitivity, an electrode condition is described by drift.
The calibration line expresses the relationship between the potential (or current)
measured at an electrode, and the concentration of the species specific to the
electrode. The calibration line forms the basis of the scale used by the analyzer to
convert electrode chain potentials to concentrations. Each electrode has a different
calibration line.
The pH electrode is used as an example to illustrate how this line is derived from
two calibration solutions of known pH.
• Cal 1 solution has a pH of 7.398 that gives potential reading of −100 mV.
• Cal 2 solution has a pH of 6.802 that gives a potential reading of −64 mV.
These two values are plotted on a graph.
The relationship between potential and pH is linear so a line can be drawn between
the two points, as shown below:
Measured
potential
(mV)
64
Calibration line
7.398
pH of Cal 1 sol.
pH
100
97
6.802
pH of Cal 2 sol.
7.346
pH of sample
The calibration line now forms the scale used to convert the potential measured at
the pH electrode during sample analysis to an actual pH value.
Continued on next page
2-7
2. Electrodes ABL700 Series Reference Manual
Calibration, Continued
Calibration Line
(continued)
A blood sample gives a potential reading of -97 mV at the pH electrode. Reading
off from the calibration line shown below, this potential corresponds to a pH of
7.346.
Measured
potential
(mV)
64
−
Theoretical
Calibration Line
Calibration line
7.398
pH of Cal 1 sol.
pH
−
97
−
100
6.802
pH of Cal 2 sol.
7.346
pH of sample
The calibration line is updated at every calibration. Drift describes the variation in
the calibration line between consecutive calibrations.
The theoretical calibration line is the relationship between potential and
concentration in a potentiometric measurement, or the relationship between current
and concentration in an amperemetric measurement.
In the ABL700 Series the theoretical calibration line for pH is defined by the
following two points:
pHElectrode potential
(vs. Ref. potential)
6.800
7.400
Measured
potential (mV)
−75.5 mV
−112.4 mV
2-8
−75.5
−112.4
6.8
7.4
pH
The position and slope of the calibration line compared to the theoretical
calibration line are described by the status and sensitivity respectively.
Continued on next page
ABL700 Series Reference Manual 2. Electrodes
Calibration, Continued
Sensitivity
Electrode sensitivity expresses how a real electrode measures compared with the
specified values of the calibration material; it illustrates the slope of the calibration line
derived from a 2-point calibration as a percentage (or fraction) of the slope of the
theoretical calibration line, as determined by the Nernst equation of the ion in question.
Calculating the sensitivity is a way of monitoring the deviation of the electrode
sensitivity from the theoretical value.
Calculation of the sensitivity is shown, using the pH electrode as an example.
A theoretical calibration line for the pH electrode with a slope of
−61.5 mV/pH is
drawn:
Measured
potential
(mV)
75.5
−
Theoretical calibration line
112.4
−
6.8
7.4
pH
The calibration line from a 2-point calibration is superimposed on the same graph:
The sensitivity of the electrode is calculated as the ratio between the slope of the 2point calibration line and that of the theoretical line, expressed as a percentage or
fraction.
If the theoretical calibration line is assumed to have a sensitivity of 100 %, the 2point calibration line shown in the example will have a sensitivity of
approximately 95 %.
Continued on next page
2-9
2. Electrodes ABL700 Series Reference Manual
Calibration, Continued
Sensitivity
(continued)
Status
The sensitivity limits for the calibration are set for each electrode. If the sensitivity
of any electrode falls outside the allowed limits, the message
out of range
appears in the Calibration Messages, with the particular electrode
Calibration Sensitivity
specified.
The electrode status is a measure of zero point of a complete electrode chain.
Status of a real electrode reflects deviations from the conditions of a theoretical
electrode, and define the position of the calibration line.
Calculating the status value of an electrode is a way of monitoring the position of
the calibration line despite the fact that only a 1-point calibration has been carried
out. The calculation of the status is shown, using the pH electrode as an example.
A calibration line with the same slope as the theoretical calibration line (
−61.5
mV/pH) is drawn through this point. This theoretical calibration line is used since
no 2-point calibration is performed which would otherwise give an actual
calibration line.
Measured
potential
(mV)
Theoretical calibration
line with a slope = −61.5
mV/pH drawn through
E
Cal 1
pH
A pH of 7.400, the nominal pH of Calibration Solution 1 (pH
corresponding potential (
E
) is read off the theoretical calibration line that
Cal 1 nom
the point from a 1-point
calibration.
pH
Cal 1
) is chosen. Its
Cal 1 nom
was drawn through the 1-point calibration point.
Measured
potential
(mV)
Theoretical calibration line for the
theoretical pH electrode with known
potential of −112.4 mV at a pH =
7.400.
2-10
E
Cal 1 nom(theo)
E=−112.4 mV
E
Cal 1 nom
E
Cal 1
∆pH
E
∆
Theoretical calibration
line drawn through the 1point calibrat ion point.
pH
Status
pH
Cal 1 nom
(7.400)
pH
Cal 1
pH
Continued on next page
ABL700 Series Reference Manual 2. Electrodes
−
Calibration, Continued
Status
(continued)
A line from pH
corresponding potential (
is extrapolated up to the theoretical calibration line, and the
Cal 1 nom
E
Cal 1 nom(theo)
) read off the theoretical calibration line.
Calibration
Materials
The difference between
E
Cal 1 nom
and E
Cal 1 nom(theo)
which corresponds to ∆E on the
graph, represents the potential that should theoretically be obtained for a solution
with pH = 7.400. This potential difference (
∆E) thus describes the deviation of the
actual pH reference electrode system from a theoretical electrode system.
Similarly
∆pH describes the deviation in pH values that would be produced
between measurements with an actual electrode system and measurements with a
theoretical electrode system.
The status of the pH electrode, pH(Status), is then calculated as:
pH(Status) = 7.4 +
Cal 1 nomCal 1 nom(theo)
EE
61.5
The status limits of the calibration are set for each electrode. If the status for any
electrode falls outside the allowed limits, the message
appears in the Calibration Messages, with the particular electrode specified.
limits
Calibration status out of
The following calibration materials are used:
Calibration MaterialUsed for...
Calibration Solutions 1 and 2: the exact composition of
the calibration solutions is given in the bar code on the
bottle label, which can be read into the analyzer using the
Calibration of the pH,
and electrolyte
electrodes
bar code reader, or entered manually via the keyboard.
Calibration Solution 1: Calibration of the
metabolite electrodes
and optical system
Gas 1 and Gas 2: each gas has a precise composition
essential for determining the accuracy of the analyzer in
each pCO
and pO
2
measurement.
2
Calibration of the
pCO
and pO
2
2
electrodes
tHb Calibration Solution: Calibration of the
optical system
The Chemical Reference Laboratory at Radiometer is responsible for the accuracy
of the calibrating solutions and gases.
Traceability certificates for individual solutions are presented in Chapter 7 of this
manual.
Continued on next page
2-11
2. Electrodes ABL700 Series Reference Manual
t
d
t
d
t
t
d
t
d
Calibration, Continued
Drift
Drift is defined as the difference measured by the electrodes during last and
previous calibrations, and is a measure of the electrode stability.
Drift 1 is obtained on Calibration Solution 1 and/or Gas 1 and is calculated as
follows, using the pH electrode as an example:
Measured
potential
Calibration line from last
2-point calibration
(mV)
n
2
1-point calibration
n
2
Drift 1 value
s
1
Drift 1 value
s
1
1-point calibration
2-point calibration
pH
Calibration Solution 2
pH
Calibration Solution 1
pH
Drift 2 is obtained after 2-point calibration. The pH electrode is used as an
example and the calibration schedule is set so that each 2-point calibration is
separated by two 1-point calibrations.
Measured
potential
(mV)
Drift 2
Calibration line from 2
Slope of 1
Calibration line from 1
n
2-point calibration
s
2-point calibration line
s
2-point calibration
n
2
1-point calibration
s
1
1-point calibration
pH
Calibration Solution 2
pH
Calibration Solution 1
pH
Drift tolerances express the extent to which drift values for an electrode can
fluctuate before the electrode is deemed unstable and thus incapable of providing
reliable calibrations.
The drift tolerances for each electrode are set in the analyzer’s Setup programs.
Radiometer recommends the use of the default drift tolerances, as too narrow drift
tolerances will cause electrode drift errors even for normal electrode fluctuations.
If the drift tolerances are too wide, significant measurement errors will result
without warning.
Continued on next page
2-12
ABL700 Series Reference Manual 2. Electrodes
Calibration, Continued
Drift (continued)
If the drift values for any electrode fall outside the drift tolerances, the message
Calibration drift out of range appears in the Calibration Messages, with the
particular electrode specified.
No drift values are reported for startup calibrations as there are no previous
calibrations available for comparison.
2-13
2. Electrodes ABL700 Series Reference Manual
Electrode Measuring Time and Updatings
Measuring Time
Updatings
In the ABL700 Series analyzers the measuring time of the electrode is independent
of the electrode type. Electrode signals are registered at 0.982 second intervals
during both calibrations and measurements. The registration of each electrode
signal begins after the samples, calibration solutions, and calibration gases are in
position in the measuring modules.
The duration of each calibration is predetermined, as is the number of updatings of
the electrodes’ signals.
In general, the updatings from an electrode response are numbered from 1 to upd last, where updating number 1 is the first updating and upd last is the last. The
diagram below schematically illustrates the electrode response that is calculated on
uncorrected electrode updating values in the ABL700 Series.
Signal
Upd 1 Upd last
Updatings
2-14
Loading...
+ 204 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.