Quadtech Sentry 30 Plus, Sentry 20 Plus, Sentry 10 Plus User Manual

Sentry Plus Series
Hipot Testers
Instruction Manual
Form 150697/A4
QuadTech, Inc., 2003
Maynard, Massachusetts, U.S.A. 01754
January 2005
Telephone: 978-461-2100 Sales: 800-253-1230 Facsimile: 978-461-4295
Website: www.quadtech.com
The material in this manual is for informational purposes only and is subject to change, without notice. QuadTech assumes no responsibility for any error or for consequential damages that may result from the misinterpretation of any procedures in this publication.
WARNING
Potentially dangerous voltages may be present on front and rear panel terminals. Follow all warnings in
this manual when operating or servicing this instrument. Dangerous levels of energy may be stored in
capacitive devices tested by this unit. Always make sure the high voltage indicator is not on when
connecting or disconnecting the device under test.
!
Product will be marked with this symbol (ISO#3864) when it is necessary for the user to refer to
the instruction manual in order to prevent injury or equipment damage.
Product marked with this symbol (IEC417) indicates presence of direct current.
Product will be marked with this symbol (ISO#3864) when voltages in excess of 1000V are
present.
Page 2 of 85
Contents
Warranty ............................................................................................................5
Specifications ............................................................................................................7
Accessories ............................................................................................................11
Safety Precautions....................................................................................................13
Condensed Operating Instructions ........................................................................15
Introduction - Section 1
1.1 Unpacking and Inspection..............................................................................21
1.2 Product Overview ..........................................................................................21
1.3 Controls and Indicators..................................................................................22
1.3.1 Front Panel Controls and Indicators .................................................22
1.3.2 Rear Panel Controls and Connectors ................................................23
1.4 Installation .....................................................................................................24
1.4.1 Dimensions ........................................................................................24
1.4.2 Instrument Positioning.......................................................................24
1.4.3 Power Requirements..........................................................................24
1.4.4 Safety Inspection................................................................................25
Operation - Section 2
2.1 Terms and Conventions.................................................................................27
2.2 Start-Up..........................................................................................................31
2.3 Programming Electrical Safety Tests.............................................................31
2.4 Programming a Ground Continuity Test .......................................................35
2.5 Programming an AC Hipot Test....................................................................37
2.6 Programming a DC Hipot Test......................................................................39
2.7 Programming an Insulation Resistance (IR) Test .........................................41
2.8 Programming a Pause (PA) in Test Sequence ..............................................43
2.9 Storing a Test Setup ......................................................................................45
2.10 Programming a Multi-Step Test.....................................................................47
2.11 PRESET Test Parameters .............................................................................48
2.11.1 AC-V FREQ ......................................................................................48
2.11.2 Software AGC ...................................................................................48
2.11.3 WV AUTO RANGE .........................................................................48
2.11.4 IR AUTO RANGE ............................................................................49
2.11.5 GFI ....................................................................................................49
2.11.6 FAIL RESTART ...............................................................................49
2.12 Instrument Offset ..........................................................................................50
2.13 Connection To Device Under Test ................................................................53
2.14 Measurement Procedure.................................................................................54
Page 3 of 85
Contents (Continued)
2.15 MENU Parameters ........................................................................................56
2.15.1 MEMORY Function .........................................................................57
2.15.2 SYSTEM Function ............................................................................59
2.15.2.1 CONTRAST ...................................................................59
2.15.2.2 BUZZER VOLUME .......................................................59
2.15.2.3 EN 50191 ........................................................................59
2.15.2.4 DC 50V AGC ..................................................................60
2.15.3 OPTION Function .............................................................................60
2.15.4 CALIBRATION Function ................................................................60
2.15.5 KEY LOCK Function .......................................................................61
2.15.6 CHANGE PASSWORD Function ....................................................63
2.15.7 ERROR LOG Function .....................................................................65
2.15.8 ABOUT Function ..............................................................................65
Interface - Section 3
3.1 Remote ..........................................................................................................67
3.2 G16 International Power Strip.......................................................................70
3.3 S07 Power Entry Adapter Cable ...................................................................71
3.4 S03 Corded Product Adapter ........................................................................72
3.5 S05 Foot Switch ............................................................................................73
3.6 S08 Gun Probe ..............................................................................................74
3.7 S50 Plus Ground Bond Tester .......................................................................75
Service & Calibration - Section 4
4.1 General...........................................................................................................77
4.2 Instrument Return ..........................................................................................77
4.3 Calibration......................................................................................................77
4.3.1 Calibration Parameters.......................................................................78
4.3.2 Enable Calibration ............................................................................79
4.3.3 AC Voltage Calibration .....................................................................80
4.3.4 DC Voltage Calibration .....................................................................80
4.3.5 IR Voltage Calibration.......................................................................81
4.3.6 AC Current Calibration......................................................................82
4.3.7 DC Current Calibration......................................................................83
4.3.8 ARC Calibration ................................................................................84
4.3.9 IR Resistor Calibration .....................................................................84
4.3.10 Ground Continuity Calibration .........................................................85
4.3.11 Contrast Calibration ..........................................................................85
4.3.12 Finalize Calibration ...........................................................................85
Page 4 of 85
Warranty
QuadTech warrants that Products are free from d efects in material and workmanship and, when properly used, will perform in accordance with QuadTech’s applicable published specifications. If within one (1) year after original shipment it is found not to meet this standard, it will be repaired, or at the option of QuadTech, replaced at no charge when returned to a QuadTech service facility.
Changes in the Product not approved by QuadTech shall void this warranty.
QuadTech shall not be liable for any indirect, special or consequential damages, even if notice has been given of the possibility of such damages.
This warranty is in lieu of all other warranties, expressed or implied, including, but not limited to any implied warranty or merchantability of fitness for a particular purpose.
SERVICE POLICY
QuadTech’s service policy is to maintain product repair capability for a period of at least five (5) years after original shipment and to make this capability available at the then prevailing schedule
of charges.
Page 5 of 85
Page 6 of 85
Specifications
Dielectric Strength
Sentry 10, 20, & 30 Plus AC Output Voltage: Range: 0.05 to 5kV AC, in 1V steps
Regulation: ± (1% of setting +5V) Frequency: 50/60Hz selectable
Voltage Display: Accuracy: ± (1% of reading +5V) Resolution: 1Volt AC Current Display: Range: 0.001mA to 20mA AC, in 1µA steps Accuracy: ± (1.5% of reading + 5 counts) (Total)
Sentry 20 & 30 Plus DC Output Voltage: Range: 0.05 to 6kV DC, in 1V steps
Voltage Display: Accuracy: ± (1% of reading +5V)
Resolution: 1Volt DC Current Display: Range: 0.0001mA to 5mA DC Resolution: 0.1µA Accuracy: ± (1.5% of reading + 5 counts)
Insulation Resistance
Sentry 30 Plus Insulation Resistance: Voltage: 50 - 1000V DC in 1V steps
Accuracy: ± (1% of reading + 5V) Range: 0.1M - 50G (voltage dependent)
Accuracy: 0.1M - 1GΩ, ± (10% + 5counts) < 100V
0.1M - 1GΩ, ± (7% + 5counts) < 500V 1M - 1GΩ, ± (4% + 5counts) ≥ 500V 1G - 10GΩ, ± (7% + 5counts) ≥ 500V 10G - 50GΩ, ± (12% + 5counts) ≥ 500V
Page 7 of 85
Specifications (Continued)
Safety Features
Ground Continuity Test: Programmable: 0.1 to 5.0Ω ±0.2 Accuracy, or OFF
Ground Fault Interrupt (GFI): Shutdown of current imbalance when I > 0.5mA ±0.25mA, or OFF
In-Rush Current: DC Mode: Set detection limit: 0.5uA – 5mA in 0.0001mA increments
The programmable range for In-Rush current is dependent on the programmed High Limit: Range: High Limit: In-Rush: 1 0.1µA – 300µA 0.5µA – 300µA 2 0.301mA – 3.000mA 5µA – 3.000mA 3 3.01mA – 5.0mA 50µA – 5.0mA
Fast Output Cutoff: HV output voltage terminated <0.4mS after NG (Fail) result
Fast Discharge: <0.2s (Typical) Discharge of DUT upon termination of HV.
ARC Detection: Detection Current: Range: 1mA – 20mA AC and 5mA DC
Pulse Width: Minimum: 10µs
General Features Time: Test*: 0.1sec – 999 sec, Continuous
Ramp: 0.1sec – 999 sec, OFF Dwell: 0.1sec – 999 sec, OFF (DC & IR Mode only) Fall: 0.1sec – 999 sec, OFF
* Te st Time is limited 60seconds when the voltage and high c urrent limit > 100VA. * Test Time for IR is 0.3sec – 999sec, Continuous
Limits: HI/LO programmable during Test Time LO can be set to OFF during Hipot Test HI can be set to OFF during IR Test
Indication: Pass/fail LEDs, audible alarm
Remote Control: Inputs: START, STOP
Characteristics: 24V active low, Pulse width ≅20ms Outputs: PASS, FAIL, UNDER TEST Characteristics: Dry contact relay, Closed if true
115V, <100mA
Connector: 9 pin male D-series & Terminal Strip
Page 8 of 85
Specifications (Continued)
General Features Setup Storage: 60 Memory Locations, 10 steps each
Standard Interfaces: Remote I/O
Connectors: Front and Rear Connection
HV OUTPUT: Custom Banana Socket RTN/LOW: Banana Socket GC (Rear Only): Binding Post/ Banana Socket
Front Panel 10 Digit Password with or without setup recall Lockout: LED Display: LOCK
Mechanical: Bench Mount
Dimensions:(w x h x d): 10.50 x 4.50 x 14.25 inches
262.5 x 112.5 x 356.25 mm
Weight: 25.0 lbs (11.5 kg) net, 28 lbs (13.0 kg) shipping Environmental: Specifications: 18°C to 28°C, 70% RH
Operating: 0°C to + 40oC, 80% RH Storage: -10°C to + 60oC, 80% RH
Warm-up Time: 15 minutes
Power:
Supplied:
Safety Agency: CE, TUV
Ordering Description Catalog No. Information: AC Hipot Tester Sentry 10 Plus
AC/DC Hipot Tester Sentry 20 Plus AC/DC/IR Hipot Tester Sentry 30 Plus
90 - 130V AC
200 - 250V AC
Instruction Manual
Calibration Certificate
Ground Continuity Lead
50 or 60Hz
300W max
Power Cable
S02 Test Leads
Page 9 of 85
Page 10 of 85
Accessories
Accessories Included
Item Quantity QuadTech P/N
AC Power Cord 1 4200-0300 Power Line Fuse 3.15A 250V SB 1 520072 Power Line Fuse 1.6A 250V SB 1 520074 High Voltage Lead Set, 1m with alligator clips 1 S02 Ground Continuity Lead 1 700100 Instruction Manual 1 150697 Calibration Certificate 1 N/A
Accessories/Options Available
Item Quantity QuadTech P/N
High Voltage Lead Set, high & low, 1m (std. with unit) 1 S02 Corded Product Adaptor, 115V 1 S03 High Voltage Lead Set, high & low, 2m 1 S04 Foot Switch 1 S05 High Voltage Probe 1 S06 Power Entry Adaptor Cable 1 S07 Gun Probe 1 S08 High Voltage Lead, 1m, unterminated 1 S09 High Voltage Lead, 2m, unterminated 1 S10 Gun Probe with Remote Start 1 S11 Load Box, resistive 1 S12 Load Box, custom resistors 1 S14 Interconnection Cable to Sentry 50 Ground Bond Tester 1 S15 Ground Continuity Lead (standard with unit) 1 700100 International Power Strip 1 G16 Corded Product Adaptor, 240V 1 G25
Page 11 of 85
Page 12 of 85
Safety Precautions
WARNING
The Sentry Plus Series Hipot Tester can provide an output voltage as high as 6000V DC (5000V
AC) to the external device under test (DUT). Although the Sentry Plus unit is designed with full
attention to operator safety, serious hazards could occur if the instrument is used improperly and
these safety instructions are not followed.
1. The Sentry Plus unit is designed for operation with its chassis connected to earth ground. The instrument is shipped with a three-prong power cord to provide this connection to ground. This power cord should only be plugged in to a receptacle that provides earth ground. Serious injury can result if the Sentry Plus is not connected to earth ground.
2. Tightly connect cable(s) to the (blue) RTN/LOW terminal. If this is not done, the DUT’s casing can be charged to the high voltage t est level and serious injur y or electri cal shock hazards could result if the DUT is touched.
3. Never touch the metal of the High Volta ge probe directl y. Touch only the insulated parts of the lead(s).
4. Never touch the test leads, test fixture or DUT in any manner (this includes insulation on all wires and clips) when the high voltage is applied and the red DANGER light is ON.
5. Before turning on the Sentry Plus unit, make sure there is no device (DUT) or fixture connected to the test leads.
6. After each test, press the [STOP] (red) button for safety if there is an y concern that HV may still be applied to the output terminals.
7. When the red DANGER LED is lit or flashing, NEVER touch the device under test, the lead wires or the output terminals.
8. Before touching the test lead wires or output terminals make sure:
a) The red [STOP] button has been pressed b) The red DANGER LED is OFF.
9. In the case of an emergency, turn OFF the POWER switch using a “hot stick” and disconnect the AC power cord from the wall. DO NOT TOUCH THE Sentry Plus INSTRUMENT.
10. If the DANGER LED does not go off when the [STOP] button is pressed, immediately stop using the tester. It is possible that the output voltage is still being delivered regardless of the TEST ON/OFF control signal.
11. When the Sentry Plus instrument is used in remote control mode, be extremel y careful. The High Voltage Output is being turned on and off with an external signal.
Page 13 of 85
Page 14 of 85
Condensed Operating Instructions
WARNING
High Voltage is applied to the white HV Output Terminal anytime the red DANGER LED is ON
or flashing. Always make sure the DANGER LED i s OFF when connecting or disconnecting
the device under test (DUT).
General Information
The Sentry Plus Series Hipot Tester is a measuring instrument for direct readout of hipot output voltage and leakage current and insulation resistance. The voltage applied to the device under test is adjustable from 50V to 5kV AC and 50V to 6kV DC. The trip current limit is programmable from 1uA to 20mA AC in 1uA steps and from 0.1uA to 5mA DC in 0.1uA steps. The output voltage for Insulation Resistance tests is 50V to 1000V DC over a measurement range of 100k to 50GΩ.
Start-Up
The Sentry Plus Series unit can be operated from a power source betw een 90 and 250VAC at a power line frequency of 50 or 60Hz. The standard Sentry Plus Series unit is shipped from QuadTech with a 3.15A fuse in place for AC 90 -130V operation. (A 1.6A fuse is included for 200-250V operation). The Sentry Plus unit is shipped with the line voltage selector set for 120V. Refer to paragraph 1.4.3 for instructions on changing the fuse or line voltage selector.
Connect the Sentry Plus Series unit AC power cord to the source of p roper voltage. Operate the Sentry Plus Series instrument with its chassis connected to earth ground. The Sentry Plus instrument is shipped with a three-prong power cord to provide this connection to ground. This power cord should only be plugged into a receptacle that provides earth ground. Serious injur y may result if the Sentry Plus Series instrument is not connected to earth ground.
Press the [POWER] button on the front panel to apply power. To switch the powe r off, press the [POWER] button again or if measurements are to be made proceed with the Test Parameter Setup in Table COI-1. The Sentry Plus Series instrument should warm up for 15 minutes prior to use.
NOTE
Please read this instruction manual in its entirety before operating this instrument.
These condensed operating instructions are not a substitute for all the information provided in the
remainder of this manual.
NOTE
Refer to paragraphs 2.3 through 2.9 for a full description of programming test parameters and
instruction on how to store the test setup. Test parameters must be set before the Sentry Plus
Series instrument can be zeroed.
Page 15 of 85
Condensed Operating Instructions (Continued )
There are numerous menus within the Sentry Plus Series instruments. Familiarize yourself with these menus prior to programming a test. Figure COI-1 illustrates the STAND BY display and lists the functions that can be accessed by pressing the [F1] through [F4] keys.
"STAND BY" or "Power-Up" Display
Function of F1 - F4 Keys
Instrument
Status
M1
STEP 1/1
1.250kV
15.00mA
3.0s
RMT ERROFSTLOCKSTAND BY
PROGRAM MOREMENUPRESET
MORE..
Insert Delete
NEXT
STEP: 1-10 MODE: AC/DC/IR/GC/PA Voltage High Limit Test Time Low Limit ARC Limit Ramp Time Dwell Time Fall Time In Rush Current
(DC, IR)
(DC)
AC-V Freq. Memory Soft AGC. WV Auto Range IR Auto Range GFI FAIL Restart
Figure COI-1: Sentry Plus Series Menus
LOWAC ARC RAMP FALL
: OFF : OFF : OFF : OFF
F3 F4F2F1
Recall Delete
System
Contrast Buzzer Volume EN 50191 DC 50V AGC
Option
Multi-Link
Calibration
(Need password to enter)
Key Lock
(Need password to enter)
Change Password
(Need password to enter)
Error Log About
Manufacturer
Software version
PROGRAM
PRESET
MENU
MORE..
View Test Steps (Tabular Format) OffsetStore
To enter programming mode.
F1
To view/change preset (initial) settings.
F2
To view/change: memory, system, option,
F3
cal., key lock, pw, error & about. To view the programmed test setups and
F4
access offset function.
Page 16 of 85
Condensed Operating Instructions (Continued )
With the Sentry Plus Series instrument in “STAND BY” (or power-up display) status, follow the steps in Table COI-1 to program an AC, DC, IR or GC test or insert a Pause in the test sequence.
"STAND BY"
DISPLAY
0.000kV
0.500mA
3.0s
LOWACSTEP 1/1 ARC RAMP FALL
RMT ERROFSTLOCKSTAND BY
OFF
:
OFF
:
OFF
:
OFF
:
PROGRAM PRESET MENU MORE..
To enter programming mode.
F1
To view/change preset (initial)
F2
test parameters. To view/change system parameters:
F3
memory, system, option, calibration, key lock, password, error, & about. To view the programmed test setups and
F4
access offset function.
PROGRAM MODE DISPLAYS
AC
Display
DC
Display
IR
Display
GC
Display
VOLT HIGH TIME
VOLT HIGH TIME
VOLT LOW TIME
CURR HIGH DWELL
0.000kV
:
0.500mA
:
3.0s
:
0.000kV
:
0.500mA
:
3.0s
:
:
0.000kV
1.0M
::3.0s
: 0.0A :
1.0
:
3.0s
LOWACSTEP 1/1 ARC RAMP FALL
RMT ERROFSTLOCKSELECT MODE
LOWDCSTEP 1/1 ARC RAMP
FALL I-RUS
RMT ERROFSTLOCKSELECT MODE
HIGHIRSTEP 1/1 RAMP DWELL : OFF
FALL
RMT ERROFSTLOCKSELECT MODE
GCSTEP 1/1
LOW
OFF
:
OFF
:
OFF
:
OFF
:
UP DOWN NEXT EXIT
OFF
:
OFF
:
OFF
: :
OFFDWELL OFF
:
OFF
:
UP DOWN NEXT EXIT
OFF
:
OFF
:
OFF
:
UP DOWN NEXT EXIT
OFF
:
UP DOWN NEXT
To change value in highlighted box.
F1
To toggle choices of highlighted box.
F2
To move highlighted box around
F3
display to select parameter to change To exit programming mode.
F4
F1
To select test mode: AC/DC/IR/GC/PA
F2
F3
F4
To select test mode: AC/DC/IR/GC/PA
F1
F2
F3
F4
F1
To select test mode: AC/DC/IR/GC/PA
F2
F3
F4
F1
To select test mode: AC/DC/IR/GC/PA
F2
F3
F4
PA
Display
PAUSE UNDER TEST SIGNAL
PAUSE MODE
::OFF
RMT ERROFSTLOCKSELECT MODE
PASTEP 1/1
EXIT
UP DOWN NEXT
RMT ERROFSTLOCKSELECT MODE
EXIT
Figure COI-2: “STAND BY” and PROGRAM Displays
Page 17 of 85
Condensed Operating Instructions (Continued )
Table COI-1: Test Parameter Setup
Step Test Para meter AC Hipot DC Hipot IR GC Range
1 To enter
programming mode
2 Select Test Step [F1] = UP
3 Select Test Mode [F1] = UP
4 Set Test Voltage
Set Current (GC)
5 Set High Limit *
Set Low Limit (IR)
6 Set Test Time
Set Dwell (GC) ***
7 Set Low Limit **
Set High Limit (IR)
8 Set ARC Limit [F1] = INC.
9 Set Ramp Time [F1] = INC.
10 Set Dwell Time*** [F1] = INC.
11 Set Fall Time [F1] = INC.
12 Set In-Rush Current [F1] = INC.
13a To program next test
step OR
13b To exit
programming mode
* High limit decision is made throughout test time. ** Unit will make low limit decision at the end of test. * ** Limits will be checked at end of dwell time.
[F1] = PROGRAM
[F3] = NEXT
[F3] = NEXT [F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT
[F3] = NEXT
[F3] = NEXT
[F3] = NEXT
[F1] = NEW
OR
[F4] = EXIT [F4] = EXIT [F4] = EXIT Exit programm i ng
[F1] = PROGRAM [F1] = UP [F3] = NEXT [F1] = UP [F3] = NEXT [F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT [F1] = INC. [F3] = NEXT
[F3] = NEXT [F1] = INC. [F3] = NEXT
[F3] = NEXT [F1] = NEW
OR
[F1] = PROGRAM [F1] = UP [F3] = NEXT [F1] = UP [F3] = NEXT [F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT
1-20mA AC
[F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT [F1] = INC. [F3] = NEXT
0, 0.5uA-5mA DC
[F1] = NEW
OR
[F1] = PROGRAM
[F1] = UP [F3] = NEXT [F1] = UP [F3] = NEXT [F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT
[F1] = INC. [F3] = NEXT
0-999s AC
0-999s DC
0-999s AC
[F1] = NEW
OR
1-10
AC, DC, IR, GC, PA
0.05-5kV AC
0.05-6kV DC
0.05-1kV IR
0.1A GC
0.001-20mA AC
0.0001-5mA DC
0.1-50000M IR 0, 0.1-5 GC 0, 0.1-999s AC 0, 0.1-999s DC 0, 0.1-999s I R
0.1-1s GC 0-20mA AC 0-5mA DC 0-50G IR 0-5 GC
1-5mA DC
0-999s DC 0-999s IR
0-999s IR
0-999s DC 0-999s IR
Program next step
OR
Page 18 of 85
Condensed Operating Instructions (Continued )
Offset
After setting your test parameters, zero the Sentry Plus Series instrument by using the automatic offset. With no device connected, connect the appropriate cable (or other fixture) into the OUTPUT connectors. Refer to paragraph 2.13 cable connections based on test to be performed. Return and HV test leads should not be connected together (open circuit) for AC and DC hipot tests. Return and GC test leads should be connected (short circuit) for a GC test. There is no offset in an IR test.
NOTE:
If Ground Continuity (GC) is turned ON when performing an offset, the continuity lead must be connected to the
return (RTN/LOW) terminal, otherwise turn the continuity test OFF.
Prior to performing the OFFSET function:
Allow the instrument to warm up for 15 minutes.
Connect the Test cables (or fixture) to the front panel OUTPUT and RTN/LOW connectors.
Program the test steps.
With the instrument in STAND BY status:
Press [F4] = MORE
Press [F3] = OFFSET
Follow instructions on display: i.e.: connect OPEN across OUTPUT terminal.
Press green [START] button.
Wait while instrument gets OFFSET value.
The OFST block at the bottom of the display is now highlighted (back lit).
Press [F4] = MORE to return to STAND BY status.
STOP
START
01
When ready, press [START] to get offset values
Q
uadTech
1.
2.
Figure COI-3: Zero/Offset OPEN Configuration
Sentry 30 Plus AC/DC/IR Hipot Tester
Please open the HV output terminal
No offset in IR mode.
PRESS START KEY TO G ET OFFSET..
CAL UPDATE
DANGERPASS FAIL
!
CAUTION
Max 5kVAC
6kVDC
F1
F2
F3
F4
RTN/LOW
OPEN
F1
F2
F3
F4
OUTPUT
+-
Page 19 of 85
Condensed Operating Instructions (Continued )
Connection to Device under Test (DUT)
Figure COI-4 illustrates the connection of the Sentry Plus Series unit to a single DUT using the S02 1-meter HV cable set that comes standard with the instrument. The custom white banana plug/red alligator clip is connected between the OUTPUT terminal on the Sentry Plus Series unit and the high side of the device under test. The black banana plug/alligator clip is connected between the RTN/LOW terminal on the Sentry Plus Series unit to the low side of the DUT.
STOP
START
Q
uadTech
M1
1.250kV
0.500mA
3.0s
STAND BY OFST
01
Sentry 30 Plus AC/DC/IR Hipot Tester
STEP 1/5
CAL UPDATE
AC
RMT LOCK ERR
LOW: OFF ARC: OFF RAMP: OFF
FALL: OFF
CAUTION
6kVDC
F1
F2
F3
F4
RTN/LOW
S02
Cable
Set
DUT
PROGRAM PRESET MENU
MORE..
DANGERPASS FAIL
!
Max 5kVAC
COI-4: Connection to Device under Test
Measurement Mode
1. Turn [POWER] ON. Allow Sentry Plus Series instrument a 15-minute warm up time.
2.
3. Connect S02 Black ground cable to Sentry Plus Series unit RTN/LOW terminal
4. Connect S02 White/red HV cable to Sentry Plus Series unit OUTPUT terminal.
5.
Press [F1] = PROGRAM and enter test parameters. When finished programming, press [F4] = EXIT to return to STAND BY status. Press [F4] = MORE to access Offset function. Press [F3] = OFFSET. Follow Offset
6. instructions. When Offset is complete, press [F4] = MORE to return to STAND BY.
7. Connect device under test (DUT) to test leads.
8.
Press [START].
9. Record measurement. Press [STOP].
10.
F1
F2
F3
F4
OUTPUT
+-
Page 20 of 85
Section 1: Introduction
1.1 Unpacking and Inspection
Inspect the shipping carton before opening. If damaged, contact the carrier agent immediately. Inspect the Sentry Plus Series instrument for an y damage. If the inst rument appears dam aged or fails to meet specifications notify QuadTech (refer to instruction manual front cover) or its local representative. Retain t he original shipping carton and packing mat erial for future use such as returning the instrument for recalibration or service.
1.2 Product Overview
The Sentry Plus Series is available in three models, the 10, 20 and 30, all of which provide AC Hipot testing capability. Additionally, the Sentry 20 & 30 Plus instruments provide DC Hipot testing. The Sentry 30 Plus unit provides Insulation Resistance testing. The hipot test can be programmed over a voltage range of 0.05 to 5kV AC and 0.05 to 6kV DC with a min/max leakage current detection range of 0.001 to 20mA AC and 0.0001 to 5mA DC. Insulation resistance measurements are possible to 50G at programmable DC test voltages between 50 and 1000V. Each instrument comes standard with programmable ground continuity, internal storage containing 60 memory locations (10 steps each) and a remote interface with start/stop inputs & pass/fail outputs.
Figure 1-1: Sentry 30 Plus AC/DC/IR Hipot Tester
UL Requirements
The Sentry Plus Series instruments meet the requirements outlined by U L for Hipot Testers. The Sentry Plus indicates the test potential (test voltage), has both visual and audible indication of failure and the STOP switch must be manually pressed prior to another measurement being made. The Sentry Plus Series instrument has a 100VA output [(5000V * 20mA) = 100VA]. The Sentry Plus Series instrument also measures and displays the output voltage directly at the output terminals during the test.
Introduction Page 21 of 85
1.3 Controls and Indicators
1.3.1 Front Panel Controls and Indicators
Figure 1-2 illustrates the controls and indicators on the front panel of the Sentry Plus Series AC/DC/IR Hipot Tester. Table 1-1 identifies them with description and function.
54
STOP
Q
uadTech
3
START
2
01
1
Sentry 30 Plus AC/DC/IR Hipot Tester
CAL UPDATE
12
11 10 9
F1
F2
F3
F4
DANGERPASS FAIL
!
CAUTION
Max 5kVAC
8
RTN/LOW
6kVDC
7
F1
F2
F3
F4
OUTPUT
Figure 1-2: Sentry 30 Plus Front Panel Controls & Indicators
Table 1-1: Sentry 30 Plus Front Panel Controls & Indicators
Reference Number Figure 1-2
1 Power Green Push Button Apply AC Power: 1=ON, 0=OFF 2 START Green Push Button Initiate Test: HV applied to OUTPUT terminal 3 STOP Red Push Button Stop Test: HV terminated at OUTPUT terminal 4 Display LCD Pro gram Menu, Test Setup, Measurement Results,
5 F1, F2, F3
6 OUTPUT White Custom
7 RTN/LOW Blue Banana Socket RTN: Low voltage reference terminal
8 DANGER Red LED When lit, high voltage is present at OUTPUT terminals 9 FAIL Red LED When lit, DUT judged as FAIL. Output voltage is
10 PASS Green LED When lit, DUT judged as PASS 11 UPDATE Recessed P-B Qualified Service Personnel Only 12 CAL Recessed P-B Enable/Disable Instrument Calibration
Name Type Function
Memory Contents, Calibration
Gray Push Buttons Select Instru men t Functions
and F4
Keys perform different functions under different menus. Right side of display shows corresponding key function. High Voltage (Potential) Terminal
Banana Socket
LOW: Common ground reference terminal
immediately cut off. Press [STOP] to disable FAIL LED
6
Page 22 of 85 Introduction
1.3.2 Rear Panel Controls and Connectors
Figure 1-3 illustrates the controls and connectors on the rear panel of the Sentry Plus Series AC/DC/IR Hipot Tester. Table 1-2 identifies them with description and function.
3254
WARNING: FOR CONTINUED PROTECTION AGAINST FIRE HAZARD, REPLACE ONLY WITH THE SAME TYPE AND RATING OF FUSE AS
!
SPECIFIED FOR THE LINE VOLTAGE BEING UTILIZED.
CAUTION: NO O PERATOR SERVICEABLE PARTS INSIDE. REFER SERVICING TO QUALIFIED
!
PERSONNEL. POLLUTION DEGREE 2 INSTALL ATION CATEGORY II
VOLTAGE SELECTOR
1
MODEL NO. SERIAL NO.
HIGH VOLTAGE Max 5kVAC/6kVDC
HV OUTPUT
90V - 110V ~ 108V - 132V ~ 198V - 242V ~ 216V - 250V ~
RTN/LOW
3.15AT 250V
1.6AT 250V
15
69
CONT. CHECK OPTION
100V/120V/220V/240V~
50/60Hz 300W MAX
START
RESET
COM
INTER LOCK
6
FAN 50o C AUTO ON/OFF
10 9
78
Figure 1-3: Rear Panel Sentry Plus Series Instrument
Table 1-2: Sentry Plus Series Rear Panel Controls & Connectors
Reference # Figure 1-3
1 VOLTAGE
2 Remote Silver 9-pin
3 CONT CHK OPT. Black banana plug Connection for Ground Continuity Check 4 TUV CE Sticker Instrument Safety Agency Listing 5 Remote Black 5-screw
6 Fan SF11580AT
7 Ground Silver Banana Plug Instrument Chassis Ground Connection 8 AC Line Input Black 3-wire inlet
9 RTN/LOW Blue Banana Socket RTN: Low voltage reference terminal
10 HV OUTPUT White Custom
Name Type Function
SELECTOR
2 Red 2-position Slide Switches
Select Voltage Level corresponding to AC Source 90V – 110V: 3.15A 250V Slow Blow 110V – 130V: 3.15A 250V Slow Blow 200V – 240V: 1.6A 250V Slow Blow 220V – 250V: 1.6A 250V Slow Blow Remote Connection: Inputs: Start, Reset
D-Type Connector
Outputs: Pass, Fail, Und e r Test
Remote Connection: Start, Reset, Com, Interlock
Terminal Strip
Cool Unit: T≥50°C = ON, T<45°C = OFF
115V 50/60Hz 0.10A
Connection to AC power source
module & fuse holder
Fuse Drawer: 3.15A 250V or 1.6A 250V (see #1)
LOW: Common ground reference terminal Rear panel High Voltage (Potential) Terminal
Banana Socket
Introduction Page 23 of 85
1.4 Installation
1.4.1 Dimensions
The Sentry Plus series unit is supplied in a bench configuration, i.e., in a cabinet with resilient feet for placement on a table. Flip feet are provided under the front f eet so that the Sentry Plus instrument can be tilted up for convenient operator viewing.
4.50"
STOP
START
01
Sentry 30 Plus AC/DC/IR Hipot Tester
QuadTech
CAL UPDATE
10.50"
262.5mm
F1
F2
F3
F4
DANGERPASS FAIL
CAUTION
Max 5kVAC
RTN/LOW
6kVDC
F1
F2
F3
F4
OUTPUT
14.25"
356.25mm
112.50mm
Figure 1-4: Sentry Plus Series Instrument Dimensions
1.4.2 Instrument Positioning
The Sentry Plus unit contains a graphic display for direct readout of measured parameters. The optimum angle for viewing is slightly down and about 10 degrees either side of center. For bench operation the front flip feet should always be used to angle the instrument up. In bench or rack mount applications the instrument should be positioned with consideration for ample air flow around the rear panel fan ventilation hole. An open space of at least 3 inches (75mm) is recommended behind the rear panel. Testing should be performed on a non-conductive surface. An ESD mat is not a recommended test platform.
1.4.3 Power Requirements
The Sentry Plus instrument can be operated from a power source of 90 to 132V AC or 198 to 250V AC. Power connection is via the rear panel through a standard receptacle. Before connecting the 3-wire power cord between the unit and AC power sour ce, make sure the voltage selection switches on the rear panel (Fi gure 1-5) are in accordance with the power sour ce being used. For a 90-132V source, use a 3.15A 250V fuse. For a 198-250V so urce, use a 1.6A 250V fuse. Always use an outlet that has a properly connected protection ground.
Page 24 of 85 Introduction
WARNING
MAKE SURE THE UNIT HAS BEEN DISCONNECTED FROM ITS AC POWER SOURCE
FOR AT LEAST FIVE MINUTES BEFORE PROCEEDING.
Procedure For Changing A Sentry Plus Series Fuse
Remove the fuse drawer, by pressing the black tab located at the center of the extended fuse drawer, just below the 3-prong receptacle, and pull outward.
Once the fuse drawer has been removed from the instrument slide out the fuse from the holder and replace. Make sure the new fuse is of the proper rating. Note that the fuse drawer can also be used to store a spare fuse.
Install the fuse drawer back in the inlet module by pushing in until it locks securely in place.
VOLTAGE SELECTOR
90V - 110V ~
108V - 132V ~
198V - 242V ~
216V - 250V ~
3.15AT 250V
1.6AT 250V
100V/120V/220V/240V~
50/60Hz 300W MAX
FUSE Drawer
Figure 1-5: Close-Up of Sentry Plus Series Rear Panel
1.4.4 Safety Inspection
Before operating the instrument inspect the power inlet module on the rear of the Sentry Plus to ensure that the properly rated fuse i s in place, otherwise damage to the unit is possible. M ake sure that the voltage selector swit ches are set in accordance with the power source in use. R efer to paragraph 1.4.3 and Figure 1-5.
The Sentry Plus instrument is shipped with a standard U.S. power cord, QuadTech P/N 4200­0300 (with Belden SPH-386 socket or equivalent, and a 3-wire plug conforming to IEC 320). Make sure the instrument is only used with these cables (or other approved international cord set) to ensure that the instrument is provided with connection to protective earth ground.
The surrounding environment should be free from excessive dust to prevent contamination of electronic circuits. The surrounding environment should also be free from excessive vibration. Do not expose the Sentry Plus instrument to direct sunlight, extreme temperature or humidity variations, or corrosive chemicals.
Introduction Page 25 of 85