Guardian 5000
AC/DC/IR/GC Safety Analyzer
Instruction Manual
Form 150206/B3
©QuadTech, Inc., 1996
5 Clock Tower Place, 210 East
Maynard, Massachusetts, U.S.A. 01754-2530
November, 1999
Tel. 978-461-2100
800-253-1230 (Sales & Service)
Fax. 978-461-4295
Email: sales@quadtech.com Web site: http://www.quadtech.com
The material in this manual is for informational purposes only and is subject to change, without notice. QuadTech assumes no responsibility for any error or for consequential damages that may result from the misinterpretation of any procedures in this publication.
Symbol |
! |
IEC417 |
|
on equipment signifies that the manual contains |
information to prevent injury or equipment damage
ii
Contents
Warranty |
........................................................................................................... |
v |
|
Specifications |
........................................................................................................... |
vii |
|
Introduction - Section 1 |
|
||
1.1 |
Unpacking and Inspection ............................................................................. |
1-1 |
|
1.1 |
Product Overview.......................................................................................... |
1-1 |
|
1.3 |
Front Panel Description................................................................................. |
1-2 |
|
1.4 |
Rear Panel Description.................................................................................. |
1-6 |
|
1.5 |
Accessories Included..................................................................................... |
1-7 |
|
1.6 |
Accessories/Options Available...................................................................... |
1-7 |
|
1.7 |
Power Requirements...................................................................................... |
1-8 |
|
Operation - Section 2 |
|
||
2.1 |
Startup |
........................................................................................................... |
2-1 |
2.2 |
Instrument Zeroing/Offset ............................................................................. |
2-1 |
|
2.3 |
Connection to Device Under Test ................................................................. |
2-3 |
|
2.4 |
Measurement Procedure ................................................................................ |
2-6 |
|
2.5 |
Programming Test Modes ............................................................................. |
2-8 |
|
2.6 |
Programming Hipot Test ............................................................................... |
2-10 |
|
2.7 |
Programming Insulation Resistance (IR) Test .............................................. |
2-12 |
|
2.8 |
Programming Ground Continuity Test.......................................................... |
2-14 |
|
2.9 |
Initial Parameter Setting................................................................................ |
2-16 |
|
2.10 |
Front Panel Lockout ...................................................................................... |
2-18 |
|
2.11 |
Software Version Display.............................................................................. |
2-18 |
|
2.12 |
Clear Setup Memory...................................................................................... |
2-18 |
|
Input/Output Interfaces - Section 3 |
|
||
3.1 |
Remote........................................................................................................... |
3-1 |
|
3.2 |
IEEE-488 Interface........................................................................................ |
3-3 |
|
|
3.2.1 |
Description ........................................................................................ |
3-3 |
|
3.2.2 |
Interface Commands.......................................................................... |
3-4 |
|
|
3.2.2.1 Listener Functions .............................................................. |
3-5 |
|
|
3.3.2.2 Talker Functions.................................................................... |
3-9 |
|
3.2.3 |
Sample QuickBASIC Program.......................................................... |
3-11 |
3.3 |
Scanner Interface ........................................................................................... |
3-14 |
|
|
3.3.1 |
Scanner Accessories .......................................................................... |
3-14 |
|
3.3.2 |
Scanner Card Installation .................................................................. |
3-14 |
|
3.3.3 |
Scanner Connections ......................................................................... |
3-15 |
|
3.3.4 |
Scanner Programming ....................................................................... |
3-17 |
3.4 |
G16 International Power Strip....................................................................... |
3-19 |
iii
Contents Continued
Maintenance/Calibration - Section 4 |
|
||
4.1 |
General .......................................................................................................... |
4-1 |
|
4.2 |
Instrument Return.......................................................................................... |
4-1 |
|
4.3 |
Calibration ..................................................................................................... |
4-1 |
|
|
4.3.1 |
Calibration Procedure........................................................................ |
4-2 |
|
4.3.2 |
AC Voltage Calibration..................................................................... |
4-3 |
|
4.3.3 |
DC Voltage Calibration..................................................................... |
4-4 |
|
4.3.4 |
IR Voltage Calibration ...................................................................... |
4-4 |
|
4.3.5 |
AC Current Calibration ..................................................................... |
4-5 |
|
4.3.6 |
DC Current Calibration ..................................................................... |
4-6 |
|
4.3.7 |
GC Calibration................................................................................... |
4-7 |
|
4.3.8 |
Finalize Calibration ........................................................................... |
4-8 |
iv
Warranty
uadTech
QuadTech warrants that Products are free from defects in material and workmanship and, when properly used, will perform in accordance with QuadTech's applicable published specifications. If within one (1) year after original shipment it is found not to meet this standard, it will be repaired, or at the option of QuadTech, replaced at no charge when returned to a QuadTech service facility.
Changes in the Product not approved by QuadTech shall void this warranty.
QuadTech shall not be liable for any indirect, special or consequential damages, even if notice has been given of the possibility of such damages.
This warranty is in lieu of all other warranties, expressed or implied, including, but not limited to any implied warranty or merchantability or fitness for a particular purpose.
SERVICE POLICY
QuadTech policy is to maintain product repair capability for a period of at least five (5) years after original shipment and to make this capability available at the then prevailing schedule of charges.
v
|
|
|
Specifications |
Dielectric Strength |
|
|
|
Output Voltage: |
Range: • 0.1 to 5 KV AC, in 10V/steps, 50 - 600Hz |
||
|
|
|
• 0.1 to 6 KV DC, in 10V/steps |
|
Accuracy: |
+/- (1% of reading + 5 counts) |
|
Breakdown: |
Regulation: <1% + 5V, rated load |
||
Detection: • Imax: 0.3 - 40mA AC, 0.3 - 20mA DC |
|||
Leakage Current: |
|
|
• ∆I: 0.5 to 40mA, ≤ 20µsec |
Accuracy: +/- (1.5% + 5 counts), unspecified 0.01 - 0.3mA |
|||
Test Time: |
Ramp: |
0.1 to 99.9sec (+/-20ms) |
|
|
Hold: |
0.1 to 999sec (+/-20ms) |
|
Insulation Resistance |
|
|
|
Measurements: |
Range: 10MΩ - 10GΩ |
||
Accuracy: |
Test Voltage: 100 - 1000V DC |
||
Measurement: +/- 5% ≥ 500V, 10MΩ - 1GΩ |
|||
|
|
|
+/- 15% ≥ 500V, 1GΩ - 10GΩ |
|
|
|
+/- 10%<500V, 1 - 200MΩ |
|
|
|
+/- 15%<200V, 200MΩ - 1GΩ |
Measure Time: |
Voltage: |
+/- (1% of setting + 5 counts) |
|
0.1 - 999sec (+/-20ms) |
|||
Limit Delay: |
0.3 - 99.9sec (+/-20ms) |
Voltage
1000V
500V |
Specification Range |
|
100V
Resistance
10M > 20M> 50M > |
500M> 1G > 2G > |
vi
|
|
Specifications Continued |
Ground Continuity |
|
|
Output Current:Range: |
1.0 to 30.0A AC, setting 0.1A/step |
|
|
Accuracy: +/- (1% of setting + 0.3A) |
|
|
|
+/- (1% of reading + 3 counts) |
Resistance: |
Frequency: 50 or 60Hz, 6 to 15V |
|
Range: |
0 - 500.0mΩ, 4 digits, Hi Limit |
|
Offset Function: |
Accuracy: +/- (1% of reading + 3 counts) |
|
0 to 100mohm offset, user selectable |
||
Test Time: |
0.5 - 999sec (+/-20ms) |
Current
30A
Specification Range
10A
1A
Resistance
10m > |
100m > |
150m > |
|
|
500m > |
|
General |
|
|
|
|
|
|
Safety Features: |
Fast Cutoff (<0.4msec), Fast Discharge, Panel Lock |
|||||
Setup Storage: |
50 Memory Groups |
|
|
|
|
|
Remote Control: |
Remote start & reset, Go/NoGo output, IEEE-488 Intf. |
|||||
Pass/Fail: |
Display indication & alarm |
|
|
|
||
Mechanical: |
Bench mount |
|
|
|
|
|
|
Dimensions: (w x h x d):17x8.5x22in (430x215x550mm) |
|||||
|
Weight: 50 lbs (23kg) net, 60 lbs (27kg) shipping |
|
||||
Environmental: |
Meets MIL-T-28800E, Type 3, Class 5 |
|
||||
|
Operating: 0 to + 45o C |
|
|
|
||
|
Humidity: <75% |
|
|
|
|
|
Power: |
Storage: - 10 to + 60o C |
|
|
|
||
• 115 V/230V +/- 10% |
• 47 - 63Hz |
• 650W max |
||||
Supplied: |
• |
Instruction Manual |
• |
Power Cable |
|
|
|
• |
Calibration Certificate |
• |
Test Leads |
|
|
Ordering |
Description |
|
|
|
Catalog No. |
|
Information: |
AC/DC/IR/GC Safety Analyzer |
|
|
Guardian 5000 |
vii
Section 1 |
Introduction |
1.1Unpacking and Inspection
Inspect the shipping carton before opening, if damaged contact the carriers agent immediately. Inspect the instrument for any damage. If the instrument appears damaged or fails to meet specifications notify QuadTech (refer to instruction manual front cover) or its local representative. Retain the shipping carton and packing material for future use such as returning for recalibration or service.
1.2Product Overview
The Guardian 5000 Safety Analyzer combines four critical safety tests into a single instrument, these being AC hipot, DC hipot, insulation resistance measurements and ground continuity test. The hipot test can be programmed over a voltage range of 0.1 to 5KV AC and 0.5 to 6KV DC with a min/max leakage current detection range of 300µA to 40mA AC and 20mA DC. Insulation resistance measurements are possible to 10GΩ at programmable DC test voltages between 100 and 1000V. A ground continuity test to 30A AC is also possible. The instrument comes standard with internal storage of up to 50 memory groups and an IEEE-488.2 interface for remote control operation and communication with other instrumentation.
Figure 1-1
Guardian 5000
Introduction 1-1
1.3Front Panel Description
Figure 1-2 shows the controls and indicators on the front panel of the Guardian 5000. Table 1-1 identifies them with descriptions and functions.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
26 |
25 24 |
23 |
22 |
21 |
20 |
19 |
18 |
17 |
Figure 1-2
Front Panel Controls & Indicators
Introduction 1-2
|
|
Table 1-1 |
|
|
|
Front Panel Controls and Indicators |
|
Fig 1-2 |
|
|
|
Ref. # Item |
Function |
|
|
1 |
Step |
Display indicates step number selected in the auto test sequence |
|
|
|
(1 to 15). Indicates memory group number (1 to 50) when RCL |
|
|
|
(recall) or STO (store) is selected. |
2Mode RΩ/W-ACV W-DCV/IR
Light indicates that tester is in Ground Continuity, AC Hipot, DC Hipot or Insulation Resistance mode status.
3 |
RMT |
Light indicates when instrument is in remote mode. |
|
|
Indicator |
|
|
4 |
Voltage or |
Display shows voltage/voltage setting value when test is not in |
|
|
Current |
process and shows output voltage/current when test is in process. |
|
|
Display |
0.1 - 5KV for AC Hipot, 0.5 - 6KV for DC Hipot, |
|
|
|
100 - 1000V for IR, and 1.0 - 30.0A for GC, |
|
5 |
LOCK |
Light indicates when instrument front panel keys are locked |
|
|
Indicator |
|
|
6 |
Display Units |
Indicates the units for the current/voltage display (AC or DC, A, |
|
|
|
KV for hipot, V for IR or A for GC) |
|
7 |
OFFSET |
Light indicates when instrument offset function is enabled. |
|
|
Indicator |
|
|
8 |
Limit |
Indicates the function of the limit shown on the Limit Display |
|
|
Indicator |
(HI, LO, or ARC) |
|
9 |
Limit/Measure Display shows limit setting when test is not in process and |
||
|
|
shows measured value when test is in process. |
|
10 |
Result |
Lights indicate test results based on set limits. |
|
|
Indicators |
HI NG- |
Measured value exceeded high limit |
|
|
LO NG - |
Measured value less than low limit |
|
|
ARC NG - Measured value exceeded arc limit |
|
|
|
PASS - |
Measured value is good based on all set limits |
Introduction 1-3
|
Table 1-1 |
|
Front Panel Controls and Indicators (continued) |
Fig 1-2 |
|
Ref. # Item |
Function |
11Limit/Measure Indicates the units for the entered limits or the measured display Display Units (mA for hipot, MΩ for IR or mΩ for GC)
12 |
Timer |
Indicates when test time shown is RAMP time |
|
Indicator |
|
13 |
CAL |
For use by qualified service personnel during instrument |
|
ENABLE |
calibration. |
14 |
Timer |
Display shows set time or time countdown when test is in |
|
Display |
process. A test time of 0 (zero) places the instrument in a |
|
|
continuous measure mode. |
15 |
TEST ON |
When indicator light is blinking high voltage is present on the |
|
Indicator |
output terminals. |
16 |
HIGH |
High voltage output terminal (Hipot or Insulation Resistance |
|
VOLTAGE |
tests) |
17 |
LOW SENSE Low potential terminal (SENSE) for Ground Continuity test. |
|
18 |
LOW |
Ground reference terminal for all tests |
|
DRIVER |
|
19 |
HIGH SENSE High potential terminal (SENSE) for Ground Continuity test. |
|
20 |
HIGH |
High current terminals (DRIVER) for Ground Continuity test. |
|
DRIVER |
|
21 |
BUZZER |
Audible output for pass/fail results. |
Introduction 1-4
|
|
|
|
Table 1-1 |
||
Fig 1-2 |
|
Front Panel Controls and Indicators (continued) |
||||
|
|
|
|
|
|
|
Ref. # Item |
Function |
|
|
|
||
22 |
CURSOR |
Under STOP status: |
|
|
||
|
keys |
UP |
or DOWN |
to examine 1 - 15 steps |
||
|
|
Under Hipot test On: |
|
|
||
|
|
UP |
or DOWN |
to increase or decrease test voltage in |
||
|
|
continuous mode |
|
|
||
|
|
Under PROG status: |
|
|
||
|
|
LEFT |
or RIGHT |
to select parameter for programming |
||
|
|
UP |
or DOWN |
to change status status |
||
23 |
Data Entry |
[0] [.] .....[9] Numerical keys to input test parameters |
||||
|
Keys |
[OFF] |
|
For turning off test parameters (limits, arc, etc.) |
||
|
|
[PROG] |
To enter and exit parameter setting status |
|||
|
|
[DELE] |
To delete a test step, those below move up |
|||
|
|
[CLEAR] |
To cancel parameter number and input again |
|||
|
|
[ENTER] |
Press to confirm enter parameter |
|||
|
|
[RCL] |
|
Recalls stored parameters (1 - 50) from memory |
||
|
|
[STO] |
|
Stores parameter (1 - 50) into memory |
||
|
|
[OFFSET] Initiates zeroing offset function |
||||
24 |
START |
Pressing this button starts a test and applies high voltage/current |
||||
|
button |
to the test terminals. |
|
|
||
25 |
STOP |
Pressing this button stops a test in process and must be pressed |
||||
|
button |
as a reset function before a test can be started (after a failed test |
||||
|
|
or after test conditions have been reprogrammed). |
||||
26 |
OFF/ON |
Applies AC power to the tester, 0 is off, 1 is on. |
Introduction 1-5
1.4Rear Panel Description
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
1 |
2 |
3 |
4 |
5 |
6 |
7 |
|
|
||||
|
|
|
|
|
|
|
|
Figure 1-3 |
|
|
|
||
|
|
|
|
|
|
|
Rear Panel View |
|
|
|
|||
|
|
|
|
|
|
|
|
Table 1-2 |
|
|
|
||
Fig 1-3 |
|
|
|
|
Rear Panel Connectors and Controls |
||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Ref. # Item |
|
Function |
|
|
|
|
|
|
|
||||
1 |
IEEE |
|
Input/output connections according to IEEE |
||||||||||
|
|
|
STD-488.2. 24 pin socket for standard |
|
|
|
|||||||
|
|
|
IEEE-488 cable. Refer to paragraph 3.2 |
||||||||||
2 |
SCAN |
|
Input/output scanner interface for QuadTech 5000-01 or 5000-02 |
||||||||||
|
|
|
Multiport Scanner Matrix. Refer to Paragraph 3.3. |
||||||||||
3 |
Voltage |
|
Switches for selecting range of AC power source |
||||||||||
|
Selector |
|
- Set to 100V position for |
90 - 100VAC operation |
-Set to 120V position for 110 - 130VAC operation
-Set to 220V position for 200 - 240VAC operation
-Set to 240V position for 220 - 250VAC operation
4 |
! AC Inlet |
Fuse drawer and 3-wire connection for AC power source. Module |
|
Contains 7A, 250V, 5x20mm fuse for 90-130V operation or |
|
|
3.5A, 250V, 5x20mm fuse for 200-250V operation |
|
5 |
SIGNAL |
Contact closures for test in process, test results good, and test |
|
OUTPUT |
results no good. Maximum contact rating is 115V at 0.3A. |
6 |
Ground |
Chassis ground connection |
7 |
Fan |
Temperature control fan |
|
|
- on for > 50oC |
|
|
- off for < 45oC |
Introduction 1-6
1.5 |
Accessories Included |
|
|
|
|
Table 1-3 |
|
|
|
Item |
Quantity |
|
|
Instruction Manual |
1 |
|
|
Calibration Certificate |
1 |
|
|
Power Cord |
1 |
|
|
Power Line Fuse, 7A |
2 |
|
|
S02, Test Lead Set, Hipot (1 red, 1 black, with alligator clips) |
1 |
|
|
G15, Test Lead Set, GC (1 red, 1 black, with spring clips) |
1 |
|
|
G14, Test Lead Set, Hipot (115V Receptacle) |
1 |
1.6 Accessories/Options Available
Table 1-4 |
|
Item |
Part Number |
Multiport Scanner Matrix (8 channel high V) |
5000-01 |
Multiport Scanner Matrix (8 channel high V, 4 GC) |
5000-02 |
Multiport Scanner Matrix, rack (8 channel high V) |
5000-03 |
Multiport Scanner Matrix, rack (8 channel high V, 4 GC) |
5000-04 |
High Voltage Lead Set, high & low, 1m (std w/unit) |
S02 |
High Voltage Lead Set, high & low, 2m |
S04 |
Foot Switch |
S05 |
High Voltage Probe |
S06 |
Gun Probe |
S08 |
High Voltage Lead, 1 meter, unterminated |
S09 |
High Voltage Lead, 2 meters, unterminated |
S10 |
Gun Probe with remote start |
S11 |
Load Box, resistive |
S12 |
Load Box custom resistors |
S14 |
Corded Product Adaptor |
G13 |
Power Entry Adaptor Cable, GC (std w/unit) |
G14 |
Ground Continuity Lead Set (std w/unit) |
G15 |
International Power Strip |
G16 |
Rear Panel Connection (factory installed) |
G22 |
Rack Mount Flanges (factory installed) |
G23 |
Corded Product Adaptor (240V) |
G25 |
RS232 Interface |
G26 |
Introduction 1-7
1.7Power Requirements
! The Guardian 5000 can be operated from a power source of 90 to 130 VAC or 200 to 250 VAC. Power connection is via the rear panel through a standard receptacle. Before connecting the 3-wire power cord between the unit and AC power source make sure the voltage selection switches on the rear panel (as indicated) and fuses are in accordance with the power source being used. 7A, 250V, 5x20mm, for 90-130V source and 3.5A, 250V, 5x20mm, for 200-250V source. Always use an outlet which has a properly connected protection ground.
To change the fuse proceed as follows:
WARNING
MAKE SURE THE UNIT HAS BEEN DISCONNECTED FROM ITS AC POWER SOURCE FOR AT LEAST FIVE MINUTES BEFORE PROCEEDING.
Remove the fuse drawer, by inserting a flat head screwdriver behind the small tab located just below the 3 prong receptacle, and force outward.
Once the fuse drawer has been removed from the instrument snap the fuse from the holder and replace. Make sure the new fuse is of the proper rating. Note that the fuse drawer can also be used to store a spare fuse.
Install the fuse drawer back in the inlet module (fuse down) by pushing in until it locks securely in place.
Introduction 1-8
Section 2 |
Operation |
2.1Startup
Check to make sure the Voltage Selector Switch on the rear panel agrees with the power source available (Depending on the power source the switch positions should be in the up or down positions as shown).
WARNING
NEVER TOUCH THE TEST LEADS IN ANY MANNER (this includes insulation on all wires and clips) WHEN THE HIGH VOLTAGE IS APPLIED AND RED DANGER LIGHT ON.
USE ALL PRECAUTIONS NECESSARY TO AVOID TOUCHING THE DEVICE UNDER TEST WHEN THE RED DANGER LIGHT IS ON OR BLINKING.
Connect the instrument power cord to the source of proper voltage. The instrument is to be used only with three wire grounded outlets.
Power is applied to the Guardian 5000 by pressing the front panel POWER switch to ON (1 position).
WARNING
DO NOT TURN INSTRUMENT POWER ON OR OFF WITH TEST DEVICES CONNECTED.
2.2Instrument Zeroing/Offset
The Guardian 5000 provides automatic zeroing/offset for lead or fixture effects. During the zeroing/offset process a correction is made (subtracted out) as the result of lead leakage current and stored in instrument memory to be applied to ongoing measurements. For maximum measurement accuracy it is recommended that the unit be zeroed after power-up and any time the test leads or fixture is changed (the offset is not saved under setup storage, 1 - 50 setups, but is saved on a power down and power back up. The instrument should warm-up for at least 15 minutes before zeroing.
Operation 2-1
Proceed as follows for automatic zeroing/offset:
•Plug the appropriate cable (or other fixture) into the front panel OUTPUT connectors, with no device connected. Refer to paragraph 2.3 for cable configurations based on test to be made. Test leads used for an AC Hipot, DC Hipot or Insulation Resistance measurement should be open and those leads used for the ground continuity measurement be shorted together before performing the zeroing offset.
•With the instrument in the power-up state (or in any other desired setup, refer to paragraph 2.4 for recalling setups) press the OFFSET key.
STEP |
R > |
OFSt |
|
A |
HI |
GEt |
m > |
|
TIMER |
SEC |
|
DC |
G> |
|
|||||||||
1 |
W-ACV |
V |
ARC |
M > |
RAMP 5.0 |
||||||
|
W-DCV |
|
AC |
KV |
LO |
|
|
uA |
mA |
TEST |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
I-R |
|
|
|
|
|
|
|
|
|
|
|
|
RMT LOCK |
|
OFFSET |
HI-NG |
LO-NG |
ARC-NG |
PASS |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Offset Key |
OFF ON |
|
7 |
8 |
9 |
PROG |
RCL |
|
|
|
|
|
STOP START |
|
|
|
|
|
|
|
|
|
||
|
|
4 |
5 |
6 |
DELE |
STO |
|
|
|
|
|
|
|
1 |
2 |
3 |
CLEAR OFFSET |
|
CURSOR |
|
|||
|
|
|
|
|
|
|
|
|
|
0. OFF ENTER
•Display should read OFSt and GEt. Push the START button to initiate the automatic storage of the offset value for all tests within a sequence.
•The offset will remain on until turned off by pressing the OFFSET key twice. If the instrument is powered down with the offset on it remains in effect when the unit is again powered up.
The following formulas apply to the offset function:
For AC offset current < 80µA:
Display current = (current read)2 −(offset current)2
For DC offset or AC offset current ≥ 80µA:
Display current = (current read) - (offset current)
Operation 2-2
2.3Connection to Device Under Test
Before connecting the device for test Press the STOP key and make sure the red DANGER light is not on.
Depending on the test to be conducted (Hipot, IR or Ground Continuity) connect the test cables to the front panel OUTPUT connectors. Refer to the following figures to determine the correct configuration. When using the black cable, with the metal retaining bracket, make sure it is locked behind the connector to prevent this cable from accidentally coming loose.
WARNING
NEVER TOUCH THE TEST LEADS OR THE DEVICE UNDER TEST WHEN THEY ARE CONNECTED TO THE INSTRUMENT AND THE RED DANGER LIGHT IS ON OR BLINKING.
HIGH-VOLTAGE |
|
|
|
|
|
|
Shorted |
UNKNOWN |
Alligator Clip Lead |
|
Together |
|
|
||
HIGH LOW |
|
|
|
(Red) |
|
|
|
SENSE WV |
|
|
|
DRIVER |
With Retaining Bracket |
Device Under |
|
|
Test |
(DUT) |
|
|
Alligator Clip Lead |
||
|
(Black) |
|
|
Figure 2-1 Connection for Hipot/IR Test (Using S02 Two Lead Test Set)
Operation 2-3