Quadtech GUARDIAN 6100 Data Sheet

Guardian
6100 Plus Safety Analyzer
for Electronics Production & Compliance Testing
Introduction
taneous Hipot and Ground Bond capability providing a dramatic savings in test time. Six instruments in one, the G6100 Plus provides AC Hipot, DC Hipot, Insulation Resistance, Leakage Current/ Functional Run, Ground Bond and Open/ Short meas­urements from a single test connection.Versatile and packed with features,the G6100 Plus is a cost effective solution for the full spectrum of electrical safety compliance testing.
Twin Port Technology: This patented feature allows for simulataneous Ground
Bond and Hipot Testing.Decrease test time and increase device through-put. Leakage Current:With 20A input current capability, Leakage Current can be meas-
ured from 0.1µA to 9.999mA. Perform Functional Run Tests while monitoring the Voltage, Current or Power plus perform Earth, Enclosure and Patient leakage testing in 8 possible modes including Normal, Reverse Line, Single Fault Normal and Single Fault Reverse.
Ground Bond: With test current from 1A to 30A in 0.1A steps and programmable current limits, test time, frequency and open circuit no load voltage, the G6100 Plus provides full GB testing capability. A 40Amp GB option is also available.
AC Hipot:AC dielectric testing over the voltage range from 50V to 5000VAC RMS. The maximum leakage current of 40mA RMS is ideal for testing devices with high leak-
reduction. DC Hipot: DC dielectric testing from 50V to 6000VDC with a resolution of 1V.
Leakage current can be continuously monitored from 0.1µA to 12mA DC. Insulation Resistance: The G6100 Plus is capable of measuring IR from 100k to
50G
with test voltage from 50V to 1000VDC in 1V steps.
Open/ Short Circuit Detection Mode: Avoid false pass results by checking for Open and Short connections and ensure that the DUT is properly connected not shorted.
Optional Scanners: Increase the number of test channels for multi-point or multi­device testing by adding an external scanner to the G6100 Plus. Add up to 8HV chan­nels or a combination of HV and GB channels with just one external scanner.
Standard Interfaces:
and serial data collection capability for automated applications.
The Remote I/O and RS-232 interfaces provide remote control
For more detailed specifications, visit
www.quadtech.com
1-800-253-1230
Fax 1-978-461-4295
Intl. 1-978-461-2100
HIPOT
TESTERS LCR METERS MEGOHMMETERS MILLIOHMMETERS CABLE
TESTERS
Guardian 6100 Plus
AC Output V
Voltage Display: Accuracy: ±(1% of reading + 5V)
AC Current Display: Total current, Range: 0.001 to 40mA AC
High/Low Limit Test: 1µA to 40mA AC
Arc Detection: Programmable 1-20mA,
DC Output V
Voltage Display: Accuracy: ±(1% of reading + 5V)
DC Current Display: Range: 0.1µ to 12mA DC
High/Low Limit Test: 0.0001mA to 12mA DC
Arc Detection: Programmable 1-10mA
Insulation Resist
Voltage Range: 50V to 1000V DC Voltage Accuracy: High/Low Limit Test:
Ground Bond:
Output Current:
Frequency: No Load Voltage: 6 to 15 V Programmable Resistance:
High Limit: Start Wait: 0.1 - 99.9sec, OFF
Leakage Current:
Input Voltage: Line Voltage Meter: Line Current Meter: 0 - 20A, ±(1.5% of reading + 0.1A) Power: High/Low Limits: Programmable for Voltage, Current, Power
olt
olt
age:
age:
ance:
Range: 50V to 5000V AC, 1V resolution Freq: 50-600 Hz, programmable in 1Hz steps Waveform: Sinusoidal Regulation: <(1% of setting +5V) at Rated Load
Resolution: 1Volt
Resolution: 1 or 10 A steps Accuracy: ±(1% of reading + 5cnt)
Low limit can be turned OFF
Pulse Width 40 , 20 , 10 or 4 Range: 50V to 6000V DC, 1V resolution
Regulation: <(1% of setting +5V) at Rated Load
Resolution: 1Volt
Resolution: 0.1, 1 or 10 Asteps Accuracy: ±(1% or reading + 5cnt)
Low limit can be turned OFF
Pulse Width 40 , 20 , 10 or 4 Range: 100k - 50G
Accuracy: ±5% to ±15% depending upon voltage and resistance
±(1% of setting + 5V) 100k
- 50G
Low limit can be turned OFF
Range: 1.0 to 30.0AAC, setting 0.1A/step; *40A Accuracy: ±(1% of setting + 0.3A)
50 or 60Hz Selectable
Range: 0.1m 1m Accuracy: ±(1% of reading + 3 counts)
0.1m
Range: 0V to 300V 0 - 300V AC,
0 - 4400VA
-510.0m
to 510m
±(1% of reading + 6cnts)
sec
sec
, 4 digits; Resol:
Current Display:
Current Trip Limits: 0.1µA to 9.999mA, 1µA Resolution
Measuring Circuit: 5 Human Body Models IAW UL544 NP,
Measurement Modes: Normal, Reverse, Single Fault with Ground
Max. DUT Current:
Range
0.0001-0.59mA 0.0002mA ±(2% + 5cnts)
0.6 - 9.999mA 0.003mA ±
Dep.)
0.1
A to 6.000mA for UL544NP
UL544P, UL1563, UL2601-1,IEC60601-1, IEC 950, UL1950, UL3101 Standards
ON/OFF, Earth Line Leakage, Patient Line Leakage and Patient Auxiliary Leakage.
20A
Res.
Accuracy
(2% + 5cnts)
(Range
Common Features:
Open/Short Circuit Mode: Check for Open & Short against a standard
Offset Function: 0 to 100m offset, user selectable Test Time: Ramp: 0.3 to 999s (±20ms), AC/DC/IR
Remote Control: Inputs: Start,Stop
Test Setups: 100 Test Setups with 50 Steps each,
Connectors: Display: Front Panel Lockout: Key Lock w/ Password; Fail Lock w/
Safety Features: Fast Cutoff (<0.4ms) & Fast Discharge
Miscellaneous:
Indication:
Standard Interfaces: RS-232, Remote I/O Optional Interfaces: IEEE-488, Printer Dimensions:
Weight: 53 lbs (24kg) - Net, 60 lbs (27kg) Shipping Environmental:
Power:
90 - 130V AC 50 or 60Hz
capacitance value (Cs); <100V, 600Hz
Dwell: 0 to 999s (±20ms), DC only Start Wait: 0.1 to 99.9s (±20ms), GB only Test: 0.3 to 999s (±20ms) & Continuous
Characteristics: Optically Isolated, Low, Pulse Width >1ms. Outputs: Pass/Fail/Under Test Characteristics: Dry Contact relay Electrical Characteristics: 120V 100mAmax. Logic: Closed if True Connector: Terminal Strip and 9 pin D Series
13 character Alpha-Numeric Label Front and Rear 4- terminal connection 320 x 240 enhanced LCD, status indicators
Password
(0.2s) Adjustable Discharge: .05-5.1kV DC
Fail Retest; Continue on Fail PAUSE Mode: Program pause between steps
Pass/Fail lights, audible sound; Remote, Lock, Offset & Error status indicators
(w x h x d):17x5.25x18.5in (430x133x470mm)
Operating: 0 to + 40° C; Humidity: <75% Storage: - 20 to + 70° C
200 - 250V AC 800W max
Ordering Information
Guardian 6100 Plus Safety Analyzer
Includes:
150799 Instruction Manual S02 HV Lead Set, 1m G15 Ground Continuity Lead Set G30 Corded Product Adapter G33 Power Entry Adapter 4200-0300 AC Power Cable 520157 8A 250V Power Line Fuse 520053 4A 250V Power Line Fuse 520124 20A 250V Power Line Fuse:
N/A Calibration Cert. Traceable to NIST
For more detailed specifications, visit www.quadtech.com For more information about special purchase,
Scanner
1-800-253-1230
5 Clock Tower Place, 210 East, Maynard, MA 01754
rent & lease options, call:
Fax 1-978-461-4295 Intl. 1-978-461-2100
*GB to 40Awith optional transformer
Optional Accessories
N/A Calibration Data 5000-01 External Scanner, 8 Channel: 8HV 5000-02 External Scanner, 8 Channel: 8HV,
5000-03 5000-04 S04 HV Lead Set 2m S05 Foot Switch S08 Gun Probe S09 HV Lead, 1m, Unterminated S10 HV Lead, 1m, Unterminated G13 Corded Product Adapter (115V)
4GB External Scanner, 8 Channel: 8HV External Scanner, 8 Channel: 8HV, 3GB
G14 Power Entry Adapter G16 International Power Strip G24 Scanner Cable (5000 scanners) G25 Corded Product Adapter (240V) G31 Isolation Transformer, 500VA G32 Isolation Transformer, 1000VA G38 Printer Interface (replaces IEEE) G39 IEEE-488 Interface G43 Rack Mount Kit G44 Barcode Scanner G45 40A Ground Bond Transformer 630157 RS232, 9-pin (M-F), 10 length
P/N: 030147/