Guardian
6100 Plus Safety Analyzer
for Electronics Production & Compliance Testing
Introduction
taneous Hipot and Ground Bond capability providing a dramatic savings in test time.
Six instruments in one, the G6100 Plus provides AC Hipot, DC Hipot, Insulation
Resistance, Leakage Current/ Functional Run, Ground Bond and Open/ Short measurements from a single test connection.Versatile and packed with features,the G6100
Plus is a cost effective solution for the full spectrum of electrical safety compliance
testing.
Description
Twin Port Technology: This patented feature allows for simulataneous Ground
Bond and Hipot Testing.Decrease test time and increase device through-put.
Leakage Current:With 20A input current capability, Leakage Current can be meas-
ured from 0.1µA to 9.999mA. Perform Functional Run Tests while monitoring the
Voltage, Current or Power plus perform Earth, Enclosure and Patient leakage testing
in 8 possible modes including Normal, Reverse Line, Single Fault Normal and Single
Fault Reverse.
Ground Bond: With test current from 1A to 30A in 0.1A steps and programmable
current limits, test time, frequency and open circuit no load voltage, the G6100 Plus
provides full GB testing capability. A 40Amp GB option is also available.
AC Hipot:AC dielectric testing over the voltage range from 50V to 5000VAC RMS.
The maximum leakage current of 40mA RMS is ideal for testing devices with high leak-
reduction.
DC Hipot: DC dielectric testing from 50V to 6000VDC with a resolution of 1V.
Leakage current can be continuously monitored from 0.1µA to 12mA DC.
Insulation Resistance: The G6100 Plus is capable of measuring IR from 100k to
50G
with test voltage from 50V to 1000VDC in 1V steps.
Open/ Short Circuit Detection Mode: Avoid false pass results by checking for
Open and Short connections and ensure that the DUT is properly connected not
shorted.
Optional Scanners: Increase the number of test channels for multi-point or multidevice testing by adding an external scanner to the G6100 Plus. Add up to 8HV channels or a combination of HV and GB channels with just one external scanner.
Standard Interfaces:
and serial data collection capability for automated applications.
The Remote I/O and RS-232 interfaces provide remote control
For more detailed specifications, visit
www.quadtech.com
1-800-253-1230
Fax 1-978-461-4295
Intl. 1-978-461-2100
HIPOT
TESTERS LCR METERS MEGOHMMETERS MILLIOHMMETERS CABLE
TESTERS

Guardian 6100 Plus
AC Output V
Voltage Display: Accuracy: ±(1% of reading + 5V)
AC Current Display: Total current, Range: 0.001 to 40mA AC
High/Low Limit Test: 1µA to 40mA AC
Arc Detection: Programmable 1-20mA,
DC Output V
Voltage Display: Accuracy: ±(1% of reading + 5V)
DC Current Display: Range: 0.1µ to 12mA DC
High/Low Limit Test: 0.0001mA to 12mA DC
Arc Detection: Programmable 1-10mA
Insulation Resist
Voltage Range: 50V to 1000V DC
Voltage Accuracy:
High/Low Limit Test:
Ground Bond:
Output Current:
Frequency:
No Load Voltage: 6 to 15 V Programmable
Resistance:
High Limit:
Start Wait: 0.1 - 99.9sec, OFF
Leakage Current:
Input Voltage:
Line Voltage Meter:
Line Current Meter: 0 - 20A, ±(1.5% of reading + 0.1A)
Power:
High/Low Limits: Programmable for Voltage, Current, Power
olt
olt
age:
age:
ance:
Range: 50V to 5000V AC, 1V resolution
Freq: 50-600 Hz, programmable in 1Hz
steps
Waveform: Sinusoidal
Regulation: <(1% of setting +5V) at Rated
Load
Resolution: 1Volt
Resolution: 1 or 10 A steps
Accuracy: ±(1% of reading + 5cnt)
Low limit can be turned OFF
Pulse Width 40 , 20 , 10 or 4
Range: 50V to 6000V DC, 1V resolution
Regulation: <(1% of setting +5V) at Rated
Load
Resolution: 1Volt
Resolution: 0.1, 1 or 10 Asteps
Accuracy: ±(1% or reading + 5cnt)
Low limit can be turned OFF
Pulse Width 40 , 20 , 10 or 4
Range: 100k - 50G
Accuracy: ±5% to ±15% depending upon
voltage and resistance
±(1% of setting + 5V)
100k
- 50G
Low limit can be turned OFF
Range: 1.0 to 30.0AAC, setting 0.1A/step;
*40A
Accuracy: ±(1% of setting + 0.3A)
50 or 60Hz Selectable
Range: 0.1m
1m
Accuracy: ±(1% of reading + 3 counts)
0.1m
Range: 0V to 300V
0 - 300V AC,
0 - 4400VA
-510.0m
to 510m
±(1% of reading + 6cnts)
sec
sec
, 4 digits; Resol:
Current Display:
Current Trip Limits: 0.1µA to 9.999mA, 1µA Resolution
Measuring Circuit: 5 Human Body Models IAW UL544 NP,
Measurement Modes: Normal, Reverse, Single Fault with Ground
Max. DUT Current:
Range
0.0001-0.59mA 0.0002mA ±(2% + 5cnts)
0.6 - 9.999mA 0.003mA ±
Dep.)
0.1
A to 6.000mA for UL544NP
UL544P, UL1563, UL2601-1,IEC60601-1,
IEC 950, UL1950, UL3101 Standards
ON/OFF, Earth Line Leakage, Patient Line
Leakage and Patient Auxiliary Leakage.
20A
Res.
Accuracy
(2% + 5cnts)
(Range
Common Features:
Open/Short Circuit Mode: Check for Open & Short against a standard
Offset Function: 0 to 100m offset, user selectable
Test Time: Ramp: 0.3 to 999s (±20ms), AC/DC/IR
Remote Control: Inputs: Start,Stop
Test Setups: 100 Test Setups with 50 Steps each,
Connectors:
Display:
Front Panel Lockout: Key Lock w/ Password; Fail Lock w/
Safety Features: Fast Cutoff (<0.4ms) & Fast Discharge
Miscellaneous:
Indication:
Standard Interfaces: RS-232, Remote I/O
Optional Interfaces: IEEE-488, Printer
Dimensions:
Weight: 53 lbs (24kg) - Net, 60 lbs (27kg) Shipping
Environmental:
Power:
90 - 130V AC 50 or 60Hz
capacitance value (Cs); <100V, 600Hz
Dwell: 0 to 999s (±20ms), DC only
Start Wait: 0.1 to 99.9s (±20ms), GB only
Test: 0.3 to 999s (±20ms) & Continuous
Characteristics: Optically Isolated, Low,
Pulse Width >1ms.
Outputs: Pass/Fail/Under Test
Characteristics: Dry Contact relay
Electrical Characteristics: 120V 100mAmax.
Logic: Closed if True
Connector: Terminal Strip and 9 pin D
Series
13 character Alpha-Numeric Label
Front and Rear 4- terminal connection
320 x 240 enhanced LCD, status indicators
Password
(0.2s)
Adjustable Discharge: .05-5.1kV DC
Fail Retest; Continue on Fail
PAUSE Mode: Program pause between
steps
Pass/Fail lights, audible sound; Remote,
Lock, Offset & Error status indicators
(w x h x d):17x5.25x18.5in
(430x133x470mm)
Operating: 0 to + 40° C; Humidity: <75%
Storage: - 20 to + 70° C
200 - 250V AC 800W max
Ordering Information
Guardian 6100 Plus Safety Analyzer
Includes:
150799 Instruction Manual
S02 HV Lead Set, 1m
G15 Ground Continuity Lead Set
G30 Corded Product Adapter
G33 Power Entry Adapter
4200-0300 AC Power Cable
520157 8A 250V Power Line Fuse
520053 4A 250V Power Line Fuse
520124 20A 250V Power Line Fuse:
N/A Calibration Cert. Traceable to NIST
For more detailed specifications, visit www.quadtech.com For more information about special purchase,
Scanner
1-800-253-1230
5 Clock Tower Place, 210 East, Maynard, MA 01754
rent & lease options, call:
Fax 1-978-461-4295 Intl. 1-978-461-2100
*GB to 40Awith optional transformer
Optional Accessories
N/A Calibration Data
5000-01 External Scanner, 8 Channel: 8HV
5000-02 External Scanner, 8 Channel: 8HV,
5000-03
5000-04
S04 HV Lead Set 2m
S05 Foot Switch
S08 Gun Probe
S09 HV Lead, 1m, Unterminated
S10 HV Lead, 1m, Unterminated
G13 Corded Product Adapter (115V)
4GB
External Scanner, 8 Channel: 8HV
External Scanner, 8 Channel: 8HV, 3GB
G14 Power Entry Adapter
G16 International Power Strip
G24 Scanner Cable (5000 scanners)
G25 Corded Product Adapter (240V)
G31 Isolation Transformer, 500VA
G32 Isolation Transformer, 1000VA
G38 Printer Interface (replaces IEEE)
G39 IEEE-488 Interface
G43 Rack Mount Kit
G44 Barcode Scanner
G45 40A Ground Bond Transformer
630157 RS232, 9-pin (M-F), 10 length
P/N: 030147/
A1