Quadtech Guardian 5000 User Manual

Guardian 5000
AC/DC/IR/GC Safety Analyzer
Instruction Manual
Form 150206/B3
©QuadTech, Inc., 1996
Maynard, Massachusetts, U.S.A. 01754-2530
November, 1999
Tel. 978-461-2100
800-253-1230 (Sales & Service)
Fax. 978-461-4295
Email: sales@quadtech.com
Web site: http://www.quadtech.com
The material in this manual is for informational purposes only and is subject to change, without notice. QuadTech assumes no responsibility for any error or for consequential damages that may result from the misinterpretation of any procedures in this publication.
IEC417
Symbol information to prevent injury or equipment damage
!
on equipment signifies that the manual contains
ii
Contents
Warranty ...........................................................................................................v
Specifications ...........................................................................................................vii
Introduction - Section 1
1.1 Unpacking and Inspection.............................................................................1-1
1.1 Product Overview..........................................................................................1-1
1.3 Front Panel Description.................................................................................1-2
1.4 Rear Panel Description..................................................................................1-6
1.5 Accessories Included.....................................................................................1-7
1.6 Accessories/Options Available......................................................................1-7
1.7 Power Requirements......................................................................................1-8
Operation - Section 2
2.1 Startup ...........................................................................................................2-1
2.2 Instrument Zeroing/Offset.............................................................................2-1
2.3 Connection to Device Under Test .................................................................2-3
2.4 Measurement Procedure................................................................................2-6
2.5 Programming Test Modes .............................................................................2-8
2.6 Programming Hipot Test...............................................................................2-10
2.7 Programming Insulation Resistance (IR) Test ..............................................2-12
2.8 Programming Ground Continuity Test..........................................................2-14
2.9 Initial Parameter Setting................................................................................2-16
2.10 Front Panel Lockout......................................................................................2-18
2.11 Software Version Display..............................................................................2-18
2.12 Clear Setup Memory......................................................................................2-18
Input/Output Interfaces - Section 3
3.1 Remote...........................................................................................................3-1
3.2 IEEE-488 Interface........................................................................................3-3
3.2.1 Description ........................................................................................3-3
3.2.2 Interface Commands..........................................................................3-4
3.2.2.1 Listener Functions..............................................................3-5
3.3.2.2 Talker Functions....................................................................3-9
3.2.3 Sample QuickBASIC Program..........................................................3-11
3.3 Scanner Interface...........................................................................................3-14
3.3.1 Scanner Accessories..........................................................................3-14
3.3.2 Scanner Card Installation ..................................................................3-14
3.3.3 Scanner Connections.........................................................................3-15
3.3.4 Scanner Programming.......................................................................3-17
3.4 G16 International Power Strip.......................................................................3-19
iii
Contents Continued
Maintenance/Calibration - Section 4
4.1 General ..........................................................................................................4-1
4.2 Instrument Return..........................................................................................4-1
4.3 Calibration.....................................................................................................4-1
4.3.1 Calibration Procedure........................................................................4-2
4.3.2 AC Voltage Calibration.....................................................................4-3
4.3.3 DC Voltage Calibration.....................................................................4-4
4.3.4 IR Voltage Calibration ......................................................................4-4
4.3.5 AC Current Calibration .....................................................................4-5
4.3.6 DC Current Calibration .....................................................................4-6
4.3.7 GC Calibration...................................................................................4-7
4.3.8 Finalize Calibration...........................................................................4-8
iv
Warranty
echuad
T
QuadTech warrants that Products are free from defects in material and workmanship and, when properly used, will perform in accordance with QuadTech's applicable published specifications. If within one (1) year after original shipment it is found not to meet this standard, it will be repaired, or at the option of QuadTech, replaced at no charge when returned to a QuadTech service facility.
Changes in the Product not approved by QuadTech shall void this warranty.
QuadTech shall not be liable for any indirect, special or consequential damages, even if notice has been given of the possibility of such damages.
This warranty is in lieu of all other warranties, expressed or implied, including, but not limited to any implied warranty or merchantability or fitness for a particular purpose.
SERVICE POLICY
QuadTech policy is to maintain product repair capability for a period of at least five (5) years
after original shipment and to make this capability available at the then prevailing schedule of
charges.
v
Specifications
Dielectric Strength Output Voltage: Range: 0.1 to 5 KV AC, in 10V/steps, 50 - 600Hz
Accuracy: +/- (1% of reading + 5 counts) Regulation: <1% + 5V, rated load
Breakdown: Detection:
Leakage Current: Accuracy: +/- (1.5% + 5 counts), unspecified 0.01 - 0.3mA
Test Time: Ramp: 0.1 to 99.9sec (+/-20ms) Hold: 0.1 to 999sec (+/-20ms)
Insulation Resistance
Measurements: Range: 10M - 10G Test Voltage: 100 - 1000V DC Accuracy: Measurement: +/- 5% 500V, 10M - 1G +/- 15% 500V, 1G - 10G +/- 10%<500V, 1 - 200M +/- 15%<200V, 200M - 1G Voltage: +/- (1% of setting + 5 counts)
Measure Time: 0.1 - 999sec (+/-20ms) Limit Delay: 0.3 - 99.9sec (+/-20ms)
Voltage
0.1 to 6 KV DC, in 10V/steps
Imax: 0.3 - 40mA AC, 0.3 - 20mA DC
I: 0.5 to 40mA, 20µsec
1000V
500V
100V
vi
10M>
20M 50M>>
Specification Range
Resistance
>>>500M 1G 2G
Specifications Continued
Ground Continuity
Output Current:Range: 1.0 to 30.0A AC, setting 0.1A/step
Accuracy: +/- (1% of setting + 0.3A) +/- (1% of reading + 3 counts) Frequency: 50 or 60Hz, 6 to 15V Resistance: Range: 0 - 500.0m, 4 digits, Hi Limit Accuracy: +/- (1% of reading + 3 counts) Offset Function: 0 to 100mohm offset, user selectable
Test Time: 0.5 - 999sec (+/-20ms)
Current
30A
Specification Range
10A
1A
Resistance
>> >>
100m 150m
500m10m
General
Safety Features: Fast Cutoff (<0.4msec), Fast Discharge, Panel Lock Setup Storage: 50 Memory Groups Remote Control: Remote start & reset, Go/NoGo output, IEEE-488 Intf. Pass/Fail: Display indication & alarm Mechanical: Bench mount
Dimensions: (w x h x d):17x8.5x22in (430x215x550mm) Weight: 50 lbs (23kg) net, 60 lbs (27kg) shipping
Environmental: Meets MIL-T-28800E, Type 3, Class 5 Operating: 0 to + 45o C Humidity: <75% Storage: - 10 to + 60o C Power:
115 V/230V +/- 10% 47 - 63Hz 650W max
Supplied: Instruction Manual Power Cable
Calibration Certificate Test Leads
Ordering Information:
Description Catalog No. AC/DC/IR/GC Safety Analyzer Guardian 5000
vii
Section 1 Introduction
1.1 Unpacking and Inspection
Inspect the shipping carton before opening, if damaged contact the carriers agent immediately. Inspect the instrument for any damage. If the instrument appears damaged or fails to meet specifications notify QuadTech (refer to instruction manual front cover) or its local representative. Retain the shipping carton and packing material for future use such as returning for recalibration or service.
1.2 Product Overview
The Guardian 5000 Safety Analyzer combines four critical safety tests into a single instrument, these being AC hipot, DC hipot, insulation resistance measurements and ground continuity test. The hipot test can be programmed over a voltage range of 0.1 to 5KV AC and 0.5 to 6KV DC with a min/max leakage current detection range of 300µA to 40mA AC and 20mA DC. Insulation resistance measurements are possible to 10G at programmable DC test voltages between 100 and 1000V. A ground continuity test to 30A AC is also possible. The instrument comes standard with internal storage of up to 50 memory groups and an IEEE-488.2 interface for remote control operation and communication with other instrumentation.
Figure 1-1
Guardian 5000
Introduction 1-1
1.3 Front Panel Description
Figure 1-2 shows the controls and indicators on the front panel of the Guardian 5000. Table 1-1 identifies them with descriptions and functions.
12345 678 9 10 11 12 13 14 15 16
Introduction 1-2
Figure 1-2
Front Panel Controls & Indicators
17181920212223242526
Table 1-1
Front Panel Controls and Indicators
Fig 1-2 Ref. # Item Function
1 Step Display indicates step number selected in the auto test sequence (1 to 15). Indicates memory group number (1 to 50) when RCL (recall) or STO (store) is selected.
2 Mode Light indicates that tester is in Ground Continuity, AC Hipot, RΩ/W-ACV DC Hipot or Insulation Resistance mode status. W-DCV/IR
3 RMT Light indicates when instrument is in remote mode. Indicator
4 Voltage or Display shows voltage/voltage setting value when test is not in Current process and shows output voltage/current when test is in process. Display 0.1 - 5KV for AC Hipot, 0.5 - 6KV for DC Hipot, 100 - 1000V for IR, and 1.0 - 30.0A for GC,
5 LOCK Light indicates when instrument front panel keys are locked Indicator
6 Display Units Indicates the units for the current/voltage display (AC or DC, A, KV for hipot, V for IR or A for GC)
7 OFFSET Light indicates when instrument offset function is enabled. Indicator
8 Limit Indicates the function of the limit shown on the Limit Display Indicator (HI, LO, or ARC)
9 Limit/Measure Display shows limit setting when test is not in process and shows measured value when test is in process.
10 Result Lights indicate test results based on set limits. Indicators HI NG - Measured value exceeded high limit LO NG - Measured value less than low limit ARC NG - Measured value exceeded arc limit PASS - Measured value is good based on all set limits
Introduction 1-3
Table 1-1
Front Panel Controls and Indicators (continued)
Fig 1-2 Ref. # Item Function
11 Limit/Measure Indicates the units for the entered limits or the measured display Display Units (mA for hipot, M for IR or m for GC)
12 Timer Indicates when test time shown is RAMP time Indicator
13 CAL For use by qualified service personnel during instrument ENABLE calibration.
14 Timer Display shows set time or time countdown when test is in Display process. A test time of 0 (zero) places the instrument in a continuous measure mode.
15 TEST ON When indicator light is blinking high voltage is present on the Indicator output terminals.
16 HIGH High voltage output terminal (Hipot or Insulation Resistance VOLTAGE tests)
17 LOW SENSE Low potential terminal (SENSE) for Ground Continuity test.
18 LOW Ground reference terminal for all tests DRIVER
19 HIGH SENSE High potential terminal (SENSE) for Ground Continuity test.
20 HIGH High current terminals (DRIVER) for Ground Continuity test. DRIVER
21 BUZZER Audible output for pass/fail results.
Introduction 1-4
Table 1-1
Front Panel Controls and Indicators (continued) Fig 1-2 Ref. # Item Function
22 CURSOR Under STOP status: keys UP or DOWN to examine 1 - 15 steps Under Hipot test On: UP or DOWN to increase or decrease test voltage in continuous mode Under PROG status:
LEFT or RIGHT to select parameter for programming UP
23 Data Entry [0] [.] .....[9] Numerical keys to input test parameters
Keys [OFF] For turning off test parameters (limits, arc, etc.) [PROG] To enter and exit parameter setting status [DELE] To delete a test step, those below move up [CLEAR] To cancel parameter number and input again [ENTER] Press to confirm enter parameter [RCL] Recalls stored parameters (1 - 50) from memory [STO] Stores parameter (1 - 50) into memory [OFFSET] Initiates zeroing offset function
24 START Pressing this button starts a test and applies high voltage/current button to the test terminals.
25 STOP Pressing this button stops a test in process and must be pressed button as a reset function before a test can be started (after a failed test or after test conditions have been reprogrammed).
26 OFF/ON Applies AC power to the tester, 0 is off, 1 is on.
or DOWN to change status status
Introduction 1-5
1.4 Rear Panel Description
12 3 456 7
Figure 1-3
Rear Panel View
Table 1-2
Rear Panel Connectors and Controls Fig 1-3 Ref. # Item Function
1 IEEE Input/output connections according to IEEE STD-488.2. 24 pin socket for standard IEEE-488 cable. Refer to paragraph 3.2
2 SCAN Input/output scanner interface for QuadTech 5000-01 or 5000-02 Multiport Scanner Matrix. Refer to Paragraph 3.3.
3 Voltage Switches for selecting range of AC power source Selector - Set to 100V position for 90 - 100VAC operation
- Set to 120V position for 110 - 130VAC operation
- Set to 220V position for 200 - 240VAC operation
- Set to 240V position for 220 - 250VAC operation
!
4
AC Inlet Fuse drawer and 3-wire connection for AC power source. Module
Contains 7A, 250V, 5x20mm fuse for 90-130V operation or
3.5A, 250V, 5x20mm fuse for 200-250V operation 5 SIGNAL Contact closures for test in process, test results good, and test
OUTPUT results no good. Maximum contact rating is 115V at 0.3A. 6 Ground Chassis ground connection 7 Fan Temperature control fan
- on for > 50oC
- off for < 45oC
Introduction 1-6
1.5 Accessories Included
Table 1-3 Item Quantity
Instruction Manual 1 Calibration Certificate 1 Power Cord 1 Power Line Fuse, 7A 2 S02, Test Lead Set, Hipot (1 red, 1 black, with alligator clips) 1 G15, Test Lead Set, GC (1 red, 1 black, with spring clips) 1 G14, Test Lead Set, Hipot (115V Receptacle) 1
1.6 Accessories/Options Available
Table 1-4 Item Part Number
Multiport Scanner Matrix (8 channel high V) 5000-01 Multiport Scanner Matrix (8 channel high V, 4 GC) 5000-02 Multiport Scanner Matrix, rack (8 channel high V) 5000-03 Multiport Scanner Matrix, rack (8 channel high V, 4 GC) 5000-04
High Voltage Lead Set, high & low, 1m (std w/unit) S02 High Voltage Lead Set, high & low, 2m S04 Foot Switch S05 High Voltage Probe S06 Gun Probe S08 High Voltage Lead, 1 meter, unterminated S09 High Voltage Lead, 2 meters, unterminated S10 Gun Probe with remote start S11 Load Box, resistive S12 Load Box custom resistors S14 Corded Product Adaptor G13 Power Entry Adaptor Cable, GC (std w/unit) G14 Ground Continuity Lead Set (std w/unit) G15 International Power Strip G16 Rear Panel Connection (factory installed) G22 Rack Mount Flanges (factory installed) G23 Corded Product Adaptor (240V) G25 RS232 Interface G26
Introduction 1-7
1.7 Power Requirements
!
The Guardian 5000 can be operated from a power source of 90 to 130 VAC or 200 to
250 VAC. Power connection is via the rear panel through a standard receptacle. Before connecting the 3-wire power cord between the unit and AC power source make sure the voltage selection switches on the rear panel (as indicated) and fuses are in accordance with the power source being used. 7A, 250V, 5x20mm, for 90-130V source and 3.5A, 250V, 5x20mm, for 200-250V source. Always use an outlet which has a properly connected protection ground.
To change the fuse proceed as follows:
WARNING
MAKE SURE THE UNIT HAS BEEN DISCONNECTED FROM ITS AC POWER
SOURCE FOR AT LEAST FIVE MINUTES BEFORE PROCEEDING.
Remove the fuse drawer, by inserting a flat head screwdriver behind the small tab located just below the 3 prong receptacle, and force outward.
Once the fuse drawer has been removed from the instrument snap the fuse from the holder and replace. Make sure the new fuse is of the proper rating. Note that the fuse drawer can also be used to store a spare fuse.
Install the fuse drawer back in the inlet module (fuse down) by pushing in until it locks securely in place.
Introduction 1-8
Section 2 Operation
2.1 Startup
Check to make sure the Voltage Selector Switch on the rear panel agrees with the power source available (Depending on the power source the switch positions should be in the up or down positions as shown).
WARNING
NEVER TOUCH THE TEST LEADS IN ANY MANNER (this includes insulation on all
wires and clips) WHEN THE HIGH VOLTAGE IS APPLIED AND RED DANGER
LIGHT ON.
USE ALL PRECAUTIONS NECESSARY TO AVOID TOUCHING THE DEVICE
UNDER TEST WHEN THE RED DANGER LIGHT IS ON OR BLINKING.
Connect the instrument power cord to the source of proper voltage. The instrument is to be used only with three wire grounded outlets.
Power is applied to the Guardian 5000 by pressing the front panel POWER switch to ON (1 position).
WARNING
DO NOT TURN INSTRUMENT POWER ON OR OFF WITH TEST DEVICES CONNECTED.
2.2 Instrument Zeroing/Offset
The Guardian 5000 provides automatic zeroing/offset for lead or fixture effects. During the zeroing/offset process a correction is made (subtracted out) as the result of lead leakage current and stored in instrument memory to be applied to ongoing measurements. For maximum measurement accuracy it is recommended that the unit be zeroed after power-up and any time the test leads or fixture is changed (the offset is not saved under setup storage, 1
- 50 setups, but is saved on a power down and power back up. The instrument should
warm-up for at least 15 minutes before zeroing.
Operation 2-1
Proceed as follows for automatic zeroing/offset:
Plug the appropriate cable (or other fixture) into the front panel OUTPUT connectors,
with no device connected. Refer to paragraph 2.3 for cable configurations based on test to be made. Test leads used for an AC Hipot, DC Hipot or Insulation Resistance measurement should be open and those leads used for the ground continuity measurement be shorted together before performing the zeroing offset.
With the instrument in the power-up state (or in any other desired setup, refer to
paragraph 2.4 for recalling setups) press the OFFSET key.
STEP
1
>R W-ACV W-DCV
I-R
A
HI
KV
OFSt
RMT LOCK OFFSET HI-NG LO-NG ARC-NG PASS
AC DC
LO
GEt
ARC
V
>m uA
M >
TIMER
mA
TEST
>G
RAMP
SEC
5.0
Offset Key
OFF ON
STOP START
7 8 9 PROG
45
1
0
6
23
.
CLEAROFFSET
OFF ENTER
RCL
STODELE
CURSOR
Display should read OFSt and GEt. Push the START button to initiate the automatic
storage of the offset value for all tests within a sequence.
The offset will remain on until turned off by pressing the OFFSET key twice. If the
instrument is powered down with the offset on it remains in effect when the unit is again powered up.
The following formulas apply to the offset function: For AC offset current < 80µA:
22
Display current = current read offset current
()( )
For DC offset or AC offset current 80µA: Display current = (current read) - (offset current)
Operation 2-2
2.3 Connection to Device Under Test
Before connecting the device for test Press the STOP key and make sure the red DANGER light is not on.
Depending on the test to be conducted (Hipot, IR or Ground Continuity) connect the test cables to the front panel OUTPUT connectors. Refer to the following figures to determine the correct configuration. When using the black cable, with the metal retaining bracket, make sure it is locked behind the connector to prevent this cable from accidentally coming loose.
WARNING
NEVER TOUCH THE TEST LEADS OR THE DEVICE UNDER TEST WHEN THEY
ARE CONNECTED TO THE INSTRUMENT AND THE RED DANGER LIGHT IS
ON OR BLINKING.
HIGH-VOLTAGE
UNKNOWN
HIGH LOW
SENSE
DRIVER
Connection for Hipot/IR Test
(Using S02 Two Lead Test Set)
Alligator Clip Lead
(Red)
WV
With Retaining Bracket Alligator Clip Lead
(Black)
Figure 2-1
Shorted Together
Device Under Test
(DUT)
Operation 2-3
Loading...
+ 43 hidden pages