Quadtech 7600 Data Sheet

USES:
• Production Testing of LCR Components
• Frequency Response, Component & Sensor Characterization
• Component Screening
• Material Testing
• Quality Assurance Testing
FEATURES:
• Frequency Range 10 Hz to 2 MHz
• 0.05% Basic Measurement Accuracy
• 7 Digit Measurement Resolution
• Programmable Test Voltage and Current
• Auto Ranging
• Test Setup and Measurement Data Storage
• Four-Terminal Kelvin Connections
• Standard USB Host Port, RS-232, Handler, Parallel Printer Interfaces
• Optional IEEE-488
• Graphical and Tabular Display of Swept Frequency, Voltage and Current Measurements
• Sequence Testing of Up To 6 Individual Tests
• Load Correction
• Binning (15)
• Built-in Auto Calibration Routine
The 7600 Plus LCR Meter performs precision impedance measurements over a fre­quency range of 10 Hz to 2 MHz. The instrument can measure 14 different imped­ance parameters with 0.05% accuracy, meeting today’s requirements for component and material testing. The ease of use and user-friendly menu programming makes the 7600 Plus ideal for applications in product development, incoming inspections, or production line testing.
14 Different Impedance Parameters: Measure and display any two parameters simultaneously to achieve coverage and flexibility not previously available.
Automatic T est Sequencing: Run up to six different tests in sequence with a single push of the start button. Each test can have different conditions and limits.
Swept Measurements: Fast and accurate swept parameter measurements, graphical or tabular, for verification of component and material response to changes in AC test frequency, AC test voltage, or AC test current without the need for complex pro­gramming or an external controller.
Program and Data Storage: Test setups can be stored and recalled from either internal memory or from standard USB memory stick. The front panel can be locked out with password protection to ensure procedures are run the same way every time. Measured data can be stored on a USB memory stick and then trans­ferred to PC for data reduction and analysis.
Load Correction: Substantially improves instrument accuracy by performing measurements on a known standard and applying correction to subsequent meas­urements. Ideal for repetitive testing of identical devices at like test conditions.
Automatic Calibration Procedure: The 7600 Plus can be calibrated without returning the unit to the factory using the NIST traceable QuadTech Calibration Kit, reducing downtime and calibration costs.
Easy to Use: Large LCD graphics display and user friendly menu driven interface allows the 7600 Plus to be put on line fast, providing useful measurements by oper­ators with little or no training.
1-800-253-1230
Fax 1-978-461-4295
Intl. 1-978-461-2100
For more information
about special purchase,
rent & lease options,
call
For more specifications, see
www.quadtech.com
HIPOT TESTERS • LCR METERS • MEGOHMMETERS • MILLIOHMMETERS • CABLE TESTERS
7600 Plus Precision LCR Meter
PRELIMINARY
7600 Plus
Measured Parameters: Any two of 14 parameters measured and
displayed simultaneously, user selectable
Test Frequency: Range: 10 Hz to 2 MHz, continuous
Resolution: 0.1 Hz from 10 Hz to 10 kHz, 5 digits>10kHz Accuracy: +/-(0.01% + 0.10Hz)
Measurement Speed: Fast: 25 meas/sec
Medium: 8 meas/sec Slow: 1 meas/sec
Ranging: Automatic, Range Hold or user selectable Trigger: Internal (automatic), External (RS-232,
IEEE-488.2 or Handler interfaces) and Manual
AC Test Signal: Voltage: 20 mV to 5.0 V (open circuit)
up to 500kHz in 5 mV steps 20 mV to 1.0 V (open circuit) 500kHz-1MHz in 5 mV steps 20 mV to 0.5 V (open circuit) >1MHz in 5 mV steps
Current: 250 µA to 100 mA(short circuit) in
50 µ A steps Max. Compliance 3V < 500kHz.
Source Impedance: 25, 400, 6.4k, or 100k, range dependent DC Bias Voltage: Internal: 2.0 V
External: 0 to +/-200V
Display: LCD Graphics with back light and adjustable contrast
Result Formats: Engineering or scientific notation
% Deviation from nominal of primary parameter Deviation from nominal of primary parameter Pass/Fail Binning summary No Display for maximum throughput
Sweep Result: Primary parameter vs. frequency, voltage or current
Graphical or Tabular Format Up to 200 measurement points per sweep
Sequencing Result: Displays up to 6 sequential test results, primary and/or
secondary
AutoAcc: Automatic calculation and display of overall instrument
accuracy for selected settings, test conditions, and device under test
Interfaces: Standard: USB Host Port, RS-232, Handler, Printer Port
Optional: IEEE-488.2
Charged Capacitor Protection:
8/C for Vmax <
250 V; √2/C for Vmax < 1000V
C = Capacitance in farads of the device under test
Measurement Delay: Programmable from 0 - 1000 ms in 1 ms steps Averaging: Programmable from 1 - 1000
Median value mode
Data Storage: ASCII format Program Storage: Memory internal
USB Host Port ASCII format
Calibration: Recommended Calibration Interval 1 year
Complete NIST Traceable Calibration using QuadTech 7000-09 Cal Kit Built-in automatic calibration procedure
Usage & Cal Data: Displays last calibration date, standard values used in
calibration and # of hours operation
Self-Test Routine: Verifies critical instrument operation at power-up
or when selected from menu
Contact Check: Time to detect, 2ms Test Terminals: Front panel, four terminal (BNC) guarded
Mechanical: Bench mount with tilt bail
Dimensions: (w x h x d): 16 x 6 x 14in
(410 x 150 x 360mm)
Weight: 17 lbs (8kg) net, 23 lbs (10.5kg) shipping
Environmental: Meets MIL-T-28800E, Type 3, Class 5, Style E & F
Operating: 0oC to + 50oC Humidity: <75% for 11oC to 30oC Operating Storage: - 10o C to + 60oC
Power: • 90 - 250Vac • 47/63Hz
• 100W max
Safety: IEC61010-1: 2001
CAT1, Pollution Degree 2
EMC: 89/336/EEC, 92/31/EEC, 93/68/EEC
For detailed specifications, visit www.quadtech.com •
For more information about special purchase, rent & lease options, call
1-800-253-1230 Fax 1-978-461-4295 Intl. 1-978-461-2100
5 Clock Tower Place, 210 East, Maynard, MA 01754
Product Specifications
PN030250/A1
7600 Plus Precision LCR Meter
Includes:
150905 Instruction Manual 700700 Power Cord P/N N/A Calibration Certificate Traceable to NIST
Optional Accessories:
7000-00 Rack Mount Kit 7000-01 BNC Cable Set, 1 meter 7000-02 BNC Cable Set, 2 meters 7000-03 Kelvin Clip Leads 7000-04 Alligator Clip Leads 7000-05 Chip Component Tweezers
7000-06 Axial/Radial Lead Component Test Fixture 7000-07 Chip Component Test Fixture 7000-09 Calibration Kit
Ordering Information
*At optimum test signal levels, frequencies, DUT values and without calibration uncertainty.
Capacitance (Cs/Cp), Inductance (Ls/Lp), Resistance (Rs/Rp), Dissipation (D) and Quality (Q) Factors, Impedance Z, Admittance Y, Phase Angle (0-), Equivalent Series Resistance (ESR), Conductance (Gp), Reactance (Xs), Susceptance (Bp)
Note: s = series, p = parallel, ESR equivalent to Rs
Basic Measurement A c c ur ac y*
Speed
Parameter Measurement Rang e
Fast
Medium
Slow Cs, Cp
00000.01 fF to 9. 999999 F
±0.5%
±0.25%
±0.05%
Ls, Lp
0000.001 nH to 99.99999 H
±0.5%
±0.25%
±0.05%
D
.0000001 to 99.99999
±0.005
±0.0025
±0.0005
Q
.0000001 to 999999.9
±0.005
±0.0025
±0.0005 Z, Rs, Rp,
ESR,Xs
000.0001 m to 99.99999 M
±0.5%
±0.25%
±0.05% Y, Gp, Bp
00000.01 µS to 9.999999 MS
±0.5%
±0.25%
±0.05%
Phase Angle
-180.0000° to +179. 9999°
±1.8°
±0.9°
±0.18°