Quadtech 6000 Data Sheet

Guardian
USES:
• Production and Compliance Testing of Appliances, Instruments and Informa tion Technology Equipment in Accordance with UL, CSA, IEC, TUV and Other Standards such as EN60335, EN60950, EN61010, CSA C22.2 No. 1010.1, UL3111 and UL60950
• Electric Motor Safety Testing
• Power Supply Safety Testing
• Verification of the Ground Connection on Products with a Three Prong Power Cord
FEATURES:
• Twin Port - Simultaneous Hipot & GB
• Programmable Output Voltage to 5KV AC and 6KV DC
• Ground Bond Testing to 30A AC (to 40A with Optional Transformer)
• Insulation Resistance Measurements from 100kΩto 50G
• Open/Short Circuit Detection Mode
• Programmable Ramp, Dwell & Test Times
• Storage of 100 Test Setups, 50 Steps Each
• Continuous Leakage Current Monitoring
• Programmable Trip Current to 40mA AC and 12mA DC
• Front Panel Lockout via Password
• Standard RS-232 & Remote I/O Interfaces
• Optional IEEE-488, Printer Interfaces
• Optional Built-in 8 Channel Scanner
• Optional CaptivATE Automation Software for G6000 Plus
Introduction
The Guardian 6000 Plus is the industry’s first safety analyzer with simultaneous Hipot and Ground Bond capability providing a dramatic savings in test time. Five instru­ments in one, the G6000 Plus provides AC Hipot, DC Hipot, Insulation Resistance, Ground Bond and Open/Short measurements from a single test connection. Versatile and packed with features, the G6000 Plus is a cost effective solution to the full spec­trum of electrical safety compliance testing.
Description
AC Hipot: AC dielectric testing over the voltage range from 50V to 5000VAC RMS. The maximum leakage current of 40mA RMS is ideal for testing devices with high leak­age currents such as power supplies which have large filter or “Y” capacitors for noise reduction.
DC Hipot: DC dielectric testing from 50V to 6000VDC with a resolution of 1V. Leakage current can be continuously monitored from 0.1µA to 12mA DC.
Insulation Resistance: The G6000 Plus is capable of measuring IR from 100k to 50Gwith test voltage from 50V to 1000VDC in 1V steps.
Quick Discharge: In both DC and IR modes, the device under test is discharged through the HV transformer resulting in a rapid and safe discharge.
Ground Bond: With test current from 1A to 30A in 0.1A steps and programmable current limits, test time, frequency and open circuit no load voltage, the G6000 Plus provides full GB testing capability. The offset feature automatically compensates for any lead resistance. A 40A GB option is also available.
Twin Port Technology: This patented feature allows for simulataneous Ground Bond and Hipot Testing. Decrease test time and increase device through-put.
Open/Short Circuit Detection Mode: Check for Opens and Shorts against a standard capacitance value, Cs, and have confidence the DUT is properly connected.
Optional Scanners: One internal and/or multiple external scanners may be added to the G6000 Plus for testing multiple devices simultaneously. Add up to 8 HV channels or a combination of HV and GB channels.
Standard Interfaces: The Remote I/O and RS-232 interfaces provide remote control and serial data collection capability for automated applications.
1-800-253-1230
Fax 1-978-461-4295
Intl. 1-978-461-2100
For more detailed specifications,
visit
www.quadtech.com
HIPOT TESTERS  LCR METERS  MEGOHMMETERS  MILLIOHMMETERS  CABLE TESTERS
6000 Plus
with Revolutionary Twin Port Technology
Electrical Safety Analyzer
AC Output Voltage:
Range: 50V to 5000V AC, 1V resolution Freq: 50-600 Hz, programmable in 1Hz steps Waveform: Sinusoidal Regulation: <(1% of setting +5V) at Rated Load
Voltage Display: Accuracy: ±(1% of reading + 5V)
Resolution: 1Volt
AC Current Display: Total current
Range: 0.001 to 40mAAC Resolution: 1 or 10µA steps Accuracy: ±(1% of reading + 5cnt)
High/Low Limit Test: 1µA to 40mAAC
Low limit can be turned OFF
Arc Detection: Programmable 1-20mA,
Pulse Width 40µ, 20µ, 10µ or 4µsec
DC Output Voltage: Range: 50V to 6000V DC, 1V resolution
Regulation: <(1% +5V) at Rated Load
Voltage Display: Accuracy: ±(1% of reading + 5V)
Resolution: 1Volt
DC Current Display: Range: 0.1µ to 12mA DC
Resolution: 0.1, 1 or 10µA steps Accuracy: ±(1% or reading + 5cnt)
High/Low Limit Test: 0.0001mA to 12mA DC
Low limit can be turned OFF
Arc Detection: Programmable 1-10mA
Pulse Width 40µ, 20µ, 10µ or 4µsec
Insulation Resistance:Range: 100k- 50G
Accuracy: ±5% to ±15% depending upon voltage and resistance
Voltage Range: 50V to 1000V DC Voltage Display: ±(1% of reading + 5V) High/Low Limit Test: 100k- 50G
Low limit can be turned OFF
Ground Bond:
Output Current: Range: 1.0 to 30.0AAC, setting .01A/step;
*40A Regulation: ±(1% of setting + 0.3A) Accuracy: ±(1% of reading + 3 counts)
Frequency: 50 or 60Hz Selectable No Load Voltage: 6 to 15 V Programmable Resistance: Range: 0.1m- 510.0m, 4 digits
Accuracy: ±(1% of reading + 3 counts) Resolution: 0.1m
High Limit: 0.1mto 510m Start Wait: 0.1 - 99.9sec
Common Features:
Open/Short Circuit Mode: Check for Open & Short against a standard
capacitance value (Cs); <100V, 600Hz
Offset Function: 0 to 100moffset, user selectable
AC Programmable Offset, 0-25mA
Test Time: Ramp: 0.3 to 999s (±20ms), (AC/DC/IR)
Dwell: 0 to 999s (±20ms), (DC only) Start Wait: 0.1 to 99.9s (±20ms), (GB only) Test: 0.3 to 999s (±20ms) or Continuous (ALL) Ramp, Dwell & Start Wait can be set to OFF
Remote Control: Inputs: Start,Stop
Characteristics: Optically Isolated, Low, Pulse Width >1ms. Outputs: Pass/Fail/Under Test Characteristics: Dry Contact relay Electrical Characteristics: 120V 100mAmax. Logic: Closed if True Connector: Terminal Strip and 9 pin D Series
Test Setups: 100 Test Setups with 50 Steps each,
13 character Alpha-Numeric Label
Connectors: Front and Rear 4- terminal connection Display: 320 x 240 enhanced LCD with status indica-
tors
Front Panel Lockout: Key Lock with Password
Fail Lock with Password
Safety Features: Fast Cutoff (<0.4ms) & Fast Discharge
(0.2s) Adjustable Discharge: .05-5.1kV DC
Miscellaneous: Fail Retest; Continue on Fail
PAUSE Mode: Program pause between steps
Indication: Pass/Fail lights, audible sound; Remote,
Lock, Offset & Error status indicators
Buzzer Level: 1,2,3 and Off Standard Interfaces: RS-232, Remote I/O
Data bits: 8, Stop bits: 1; Parity: None Selectable Baud Rate: 300 - 19200
Optional Interfaces: IEEE-488, Printer Dimensions: (w x h x d):17x5.25x18.5in
(430x133x470mm)
Weight: 53 lbs (24kg) - Net, 60 lbs (27kg) Shipping Environmental: Operating: 0 to + 40° C,
Humidity: <75% Storage: - 20 to + 70° C Warm-up Time: 1minute
Power: • 90 - 130V AC • 50 or 60Hz
• 200 - 250V AC • 800W max
Guardian 6000 Plus Electrical Safety Analyzer
Includes:
150798 Instruction Manual S02 HV Lead Set, 1m G15 Ground Continuity Lead Set G14 Power Entry Adapter 4200-0300 AC Power Cable 520157 8A 250V Power Line Fuse 520053 4A 250V Power Line Fuse N/A Calibration Cert. Traceable to NIST
Optional Accessories
N/A Calibration Data 6000-01 Scanner, 8 Channel 5HV,3GC 6000-02 Scanner, 8 Channel 3HV,5GC 6000-03 Scanner, 8 Channel HV S04 HV Lead Set 2m S05 Foot Switch S08 Gun Probe S09 HV Lead, 1M, Unterminated S10 HV Lead, 1M, Unterminated
G13 Corded Product Adapter (115V) G14 Power Entry Adapter G16 International Power Strip G24 Scanner Cable (5000 scanners) G25 Corded Product Adapter (240V) G38 Printer Interface (replaces IEEE) G39 IEEE-488 Interface G43 Rack Mount Kit G44 Barcode Scanner G45 40A Ground Bond Transformer
Ordering Information
For more detailed specifications, visit www.quadtech.com • For more information about special
purchase, rent & lease options, call:
1-800-253-1230 Fax 1-978-461-4295 Intl. 1-978-461-2100
5 Clock Tower Place, 210 East, Maynard, MA 01754
P/N: 030145/A2
Guardian 6000 Plus
Product Specifications
*GB to 40A with optional transformer
Loading...