Precision LCR Meters
1900 Series
The 1900 series consists of highperformance LCR Meters designed
to perform fast, automated impedance measurements on a variety
of electronic components and materials. These meters have a basic
accuracy specication of 0.1% for
accurate test results over a wide
frequency range, from 20 Hz to
1 MHz. Besides their 15 impedance
parameters, the 1900 meters are
also capable of measuring dc resistance as well as monitoring the voltage or current in the device under
test. The units incorporate a distinctive sequence test mode, allowing
up to 6 uniquely different tests to be
performed quickly on a single start
command. Additionally, these meters include IEEE-488, RS-232, and
handler interfaces, all standard.
1910 Model
The 1910 Inductance Analyzer adds
bias current capability from 1mA to 1A.
The addition of bias current makes the
1910 ideal for measuring inductance of
power inductors under actual operating
conditions.
Page 1 of 3
1900 Series Precision LCR Meter
1920 Model
The 1920 Precision LCR Meter adds bias
voltage capability from 1mV to 2V. The
bias voltage feature is used for testing
capacitors or providing reverse bias for
testing junction capacitance of diodes and
similar components.
Features:
• 20 measurement parameters
• Frequency range: 20 Hz to 1 MHz
• Basic measurement accuracy: 0.1%
• Measurement speeds up to 40/sec
• Displays voltage and current
measurements
• Wide measurement range with 5-digit
measurement resolution
• Four-terminal kelvin connections to
maintain measurement integrity
• Programmable dc bias voltage, 0-2V
• Constant voltage (voltage leveling)
• Automatic test sequencing: up to 6
individual tests in one pass
• Menu-driven interface for user-friendly
operation
• Autoranging measurements or
manual hold
• Averaging measurements: up to 1000
• Interfaces: IEEE-488.2, RS-232, and
Handler
• Open/Short zeroing and cablecompensation
• 14 Pass/Fail bins
• Load correction
20 Measurement Parameters
The 1900 series can measure and display
any two of its 15 impedance parameters
simultaneously with a basic accuracy
of 0.1%. Additionally, these meters can
measure the DC resistance, or display the
current through or voltage across a test
device ensuring the operator of the real test
conditions.
Wide Frequency Range
The 1900 series offers over 27,000 userprogrammable test frequencies to fully
characterize devices over the range of 20 Hz
to 1 MHz.
Automated Test Sequencing
For increased productivity and throughput,
up to six different tests in sequence can
be performed with a single push of the
start button. Each test can have different
measurement parameters, test conditions,
and limits.
Programmable Source Impedance
Measurement results can vary substantially
based solely on the source impedance
of the test instrument being used. To
compare measurements made on other test
instruments, the 1900 series can have its
source impedance set at 5, 25, 50 or 100 Ω.
1900 Rear Connectors
IET LABS, INC. in the GenRad Tradition
534 Main Street, Westbury, NY 11590
www.ietlabs.com
TEL: (516) 334-5959 • (800) 899-8438 • FAX: (516) 334-5988
1900 Datasheet/10-24-2012

Precision LCR Meters
1900 Series
SPECIFICATIONS
Measure Parameters:
Parameter Measurement Range
Ls, Lp
Cs, Cp
|Z|, Rs, Rp, Xs, ESR
Y, Gp, Bp
D
Q
Θ
DCR
DUT AV Voltage
DUT AC Current
DUT DC Voltage
DUT DC Current
For more detailed accuracy information, see instruction manual
Test Frequency:
Range: 20 Hz to 1 MHz
Resolution from 20 Hz to 1 kHz: 1 Hz
Resolution from >1 kHz: 4 digits
Accuracy: ±(0.02% + 0.02 Hz)
Ranging:
Automatic, Range Hold or user-selectable
Test Frequency:
Range: 20 Hz to 1 MHz
Resolution from 20 Hz to 1 kHz: 1 Hz
Resolution from >1 kHz: 4 digits
Accuracy: ±(0.02% + 0.02 Hz)
Ranging:
Automatic, Range Hold or user-selectable
Trigger:
Internal (automatic)
External (via RS-232, IEEE-488.2, or
Handler)
Manual
Source Impedance:
5 Ω, 25 Ω, 50 Ω, or 100 Ω
AC Test Signal:
20 mV to 1.0 V (open circuit) in 5 mV steps
DC Test Signal:
20 mV to 1.0 V (open circuit) in 5 mV steps,
5 Ω source impedance
0.001 nH to 99.999 kH ±0.5% ±-0.25% ±0.1%
1 pF to 9.9999 F ±0.5% ±-0.25% ±0.1%
0.0001 mΩ to 99.999 MΩ ±0.5% ±0.25% ±0.1%
10 nS to 9999.9 S ±0.5% ±-0.25% ±0.1%
0.00001 to 99.999 ±0.005 ±0.0025 ±0.001
0.00001 to 99.999 ±0.005 ±0.0025 ±0.001
-180.00 to 179.99° ±1.8° ±0.9° ±0.18°
0.1 mΩ to 100 kΩ ±0.5% ±0.25% ±0.2%
20 mV to 1.0 V ±(2% + 5 mV) @1kHz
1 µA to 150 mA ±(2% + 5 µA) @1kHz
20 mV to 1.0 V ±(2% + 5 mV)
1 µA to 150 mA ±(2% + 5 µA)
Basic Accuracy
Low Medium High
DC Bias Current: (1910 only)
Internal: 0 to 1 Amp in 1 mA steps
DC Bias Voltage: (1920 only)
Internal: 0 to 2.0 V in 1 mV steps
Power:
100 to 240 Vac
50 - 60 Hz
100 W max
Formats for Results:
Engineering or scientic format
Deviation from nominal of primary parameter
%Deviation from nominal of primary
parameter
Pass/Fail
No display mode for maximum throughput
Sequencing:
Displays up to 6 sequential test results,
primary and/or secondary
Display:
LCD Display with backlight
Pass/Fail and Status indicators
Standard Interfaces:
IEEE-488, RS-232, Handler I/O
Measurement Delay:
Programmable from 0 to 1000 ms in 1 ms
steps
Page 2 of 3
Measurement Speed:
Speed Accuracy Setting
40 meas/sec Low, No Display
25 meas/sec Low
10 meas/sec Medium
1 meas/sec High
The speed is slower at lower frequency settings and may be
affected by test conditions
DC measurements take 2x as long as AC measurements
Averaging:
1 to 1000, programmable
Median Value:
Averaged over the last three measurements
Setup Storage:
30 Single Tests
10 Sequential (6 tests in each)
Self-Test:
Veries critical instrument operation at
power-up or when selected from a menu
Other Features:
Constant-voltage mode
Cable compensation (1 m, 2 m, no cable)
Open/Short zeroing
Distortion check
Test Terminals:
Four bnc connection terminals are located
on the front panel
Optional test xture available
Mechanical:
Dimensions: 43.2 W x 13.3 H x 40.6 D cm
(17" x 5.25" x 16")
Weight: 8 kg (17.6 lbs)
Environmental Conditions:
Operating temperature: 0 to 50°C, <75% RH
Storage temperature: -40 to 71°C
Altitude: < 2000 m