Quadtech 1920 Data Sheet 12/12

Precision LCR Meters
1900 Series
The 1900 series consists of high­performance LCR Meters designed to perform fast, automated imped­ance measurements on a variety of electronic components and ma­terials. These meters have a basic accuracy specication of 0.1% for accurate test results over a wide frequency range, from 20 Hz to 1 MHz. Besides their 15 impedance parameters, the 1900 meters are also capable of measuring dc resis­tance as well as monitoring the volt­age or current in the device under test. The units incorporate a distinc­tive sequence test mode, allowing up to 6 uniquely different tests to be performed quickly on a single start command. Additionally, these me­ters include IEEE-488, RS-232, and handler interfaces, all standard.
1910 Model
The 1910 Inductance Analyzer adds bias current capability from 1mA to 1A. The addition of bias current makes the 1910 ideal for measuring inductance of power inductors under actual operating conditions.
Page 1 of 3
1900 Series Precision LCR Meter
1920 Model
The 1920 Precision LCR Meter adds bias voltage capability from 1mV to 2V. The bias voltage feature is used for testing capacitors or providing reverse bias for testing junction capacitance of diodes and similar components.
Features:
• 20 measurement parameters
• Frequency range: 20 Hz to 1 MHz
• Basic measurement accuracy: 0.1%
• Measurement speeds up to 40/sec
• Displays voltage and current measurements
• Wide measurement range with 5-digit measurement resolution
• Four-terminal kelvin connections to maintain measurement integrity
• Programmable dc bias voltage, 0-2V
• Constant voltage (voltage leveling)
• Automatic test sequencing: up to 6 individual tests in one pass
• Menu-driven interface for user-friendly operation
• Autoranging measurements or manual hold
• Averaging measurements: up to 1000
• Interfaces: IEEE-488.2, RS-232, and Handler
• Open/Short zeroing and cable­compensation
• 14 Pass/Fail bins
• Load correction
The 1900 series can measure and display any two of its 15 impedance parameters simultaneously with a basic accuracy of 0.1%. Additionally, these meters can measure the DC resistance, or display the current through or voltage across a test device ensuring the operator of the real test conditions.
Wide Frequency Range
The 1900 series offers over 27,000 user­programmable test frequencies to fully characterize devices over the range of 20 Hz to 1 MHz.
Automated Test Sequencing
For increased productivity and throughput, up to six different tests in sequence can be performed with a single push of the start button. Each test can have different measurement parameters, test conditions, and limits.
Programmable Source Impedance
Measurement results can vary substantially based solely on the source impedance of the test instrument being used. To compare measurements made on other test instruments, the 1900 series can have its source impedance set at 5, 25, 50 or 100 Ω.
1900 Rear Connectors
IET LABS, INC. in the GenRad Tradition
534 Main Street, Westbury, NY 11590
www.ietlabs.com
TEL: (516) 334-5959 • (800) 899-8438 • FAX: (516) 334-5988
1900 Datasheet/10-24-2012
Precision LCR Meters
1900 Series
SPECIFICATIONS
Measure Parameters:
Parameter Measurement Range
Ls, Lp Cs, Cp |Z|, Rs, Rp, Xs, ESR Y, Gp, Bp D Q
Θ
DCR DUT AV Voltage DUT AC Current DUT DC Voltage DUT DC Current
For more detailed accuracy information, see instruction manual
Test Frequency:
Range: 20 Hz to 1 MHz Resolution from 20 Hz to 1 kHz: 1 Hz Resolution from >1 kHz: 4 digits Accuracy: ±(0.02% + 0.02 Hz)
Ranging:
Automatic, Range Hold or user-selectable
Test Frequency:
Range: 20 Hz to 1 MHz Resolution from 20 Hz to 1 kHz: 1 Hz Resolution from >1 kHz: 4 digits Accuracy: ±(0.02% + 0.02 Hz)
Ranging:
Automatic, Range Hold or user-selectable
Trigger:
Internal (automatic) External (via RS-232, IEEE-488.2, or Handler) Manual
Source Impedance:
5 Ω, 25 Ω, 50 Ω, or 100 Ω
AC Test Signal:
20 mV to 1.0 V (open circuit) in 5 mV steps
DC Test Signal:
20 mV to 1.0 V (open circuit) in 5 mV steps, 5 Ω source impedance
0.001 nH to 99.999 kH ±0.5% ±-0.25% ±0.1%
1 pF to 9.9999 F ±0.5% ±-0.25% ±0.1%
0.0001 mΩ to 99.999 MΩ ±0.5% ±0.25% ±0.1%
10 nS to 9999.9 S ±0.5% ±-0.25% ±0.1%
0.00001 to 99.999 ±0.005 ±0.0025 ±0.001
0.00001 to 99.999 ±0.005 ±0.0025 ±0.001
-180.00 to 179.99° ±1.8° ±0.9° ±0.18°
0.1 mΩ to 100 kΩ ±0.5% ±0.25% ±0.2%
20 mV to 1.0 V ±(2% + 5 mV) @1kHz
1 µA to 150 mA ±(2% + 5 µA) @1kHz
20 mV to 1.0 V ±(2% + 5 mV)
1 µA to 150 mA ±(2% + 5 µA)
Basic Accuracy
Low Medium High
DC Bias Current: (1910 only)
Internal: 0 to 1 Amp in 1 mA steps
DC Bias Voltage: (1920 only)
Internal: 0 to 2.0 V in 1 mV steps
Power:
100 to 240 Vac 50 - 60 Hz 100 W max
Formats for Results:
Engineering or scientic format Deviation from nominal of primary parameter %Deviation from nominal of primary parameter Pass/Fail No display mode for maximum throughput
Sequencing:
Displays up to 6 sequential test results, primary and/or secondary
Display:
LCD Display with backlight Pass/Fail and Status indicators
Standard Interfaces:
IEEE-488, RS-232, Handler I/O
Measurement Delay:
Programmable from 0 to 1000 ms in 1 ms steps
Page 2 of 3
Measurement Speed:
Speed Accuracy Setting
40 meas/sec Low, No Display
25 meas/sec Low
10 meas/sec Medium
1 meas/sec High
The speed is slower at lower frequency settings and may be affected by test conditions DC measurements take 2x as long as AC measurements
Averaging:
1 to 1000, programmable
Median Value:
Averaged over the last three measurements
Setup Storage:
30 Single Tests 10 Sequential (6 tests in each)
Self-Test:
Veries critical instrument operation at power-up or when selected from a menu
Other Features:
Constant-voltage mode Cable compensation (1 m, 2 m, no cable) Open/Short zeroing Distortion check
Test Terminals:
Four bnc connection terminals are located on the front panel Optional test xture available
Mechanical:
Dimensions: 43.2 W x 13.3 H x 40.6 D cm
(17" x 5.25" x 16")
Weight: 8 kg (17.6 lbs)
Environmental Conditions:
Operating temperature: 0 to 50°C, <75% RH Storage temperature: -40 to 71°C Altitude: < 2000 m
IET LABS, INC. in the GenRad Tradition
534 Main Street, Westbury, NY 11590
www.ietlabs.com
TEL: (516) 334-5959 • (800) 899-8438 • FAX: (516) 334-5988
1900 Datasheet/05-10-2012