Quadtech 1715 Data Sheet

1715 LCR Meter
HIPOT • LCR METERS • MEGOHMMETERS • MILLIOHMMETERS
Digibridge
Digibridge
Passive Component Tester with RS-232
USES:
• Production Testing of LCR Components
• Component Design and Evaluation
• Component Inspection and Screening
• Quality Assurance Testing
• Material Testing
FEATURES:
• 0.2% Basic Measurement Accuracy
• Selectable Test Frequency: 100Hz, 120Hz, 1kHz, 10kHz
• Selectable Test Voltage: 0.25V, 1.0V
• Measurement Speed to 25/second
• LCD Display
• RS-232 Interface Standard
•IEEE and Handler Interfaces, Optional
• Automatic Open/Short Zeroing
• Automatic Hi/Lo Comparator Limits
• Pass/Fail Sorting (8 Bins)
Introduction
The QuadTech 1715 LCR Digibridge is an economically priced, easily pro­grammable, micro-processor based passive component tester for fast, precise impedance measurements. It measures eleven different parameters with a basic accuracy of 0.2%. The 1715 provides four different test frequencies and two test voltages for testing flexibility. For remote operation and automated system applications, the unit comes standard with an RS-232 interface, plus IEEE-488 and Handler interfaces are available as options.
Description
11 Measurement Parameters:Measure and display two of 11 impedance
parameters simultaneously with a basic accuracy of 0.2%. Test results can also be shown as Deviation or %Deviation from nominal value.
Test Signal:Four user selectable test frequencies: 100Hz, 120Hz, 1kHz, and 10kHz (9.6kHz), plus two test voltages: 0.25V and 1.0V.
Display:40 x 4 character LCD display for user friendly menu programming and measured results. Simultaneously displays primary/secondary values and binning or comparator results.
Setup Storage: Storage of 1 setup with parameter, binning and compare infor­mation saved. Auto recall of this setup upon power-up of instrument.
Auto Ranging:For maximum measurement integrity the 1715 automatically selects the appropriate measurement range and instrument source impedance (25
, 100Ω, 1kΩ, 10kΩ or 100kΩ), dependent upon the impedance of the
device under test. Pass/Fail Testing:The 1715 provides a programmable Hi/Lo comparator
function as well as eight sorting bins for categorization of components.
• Storage of 1 Setup
• Automatic Source Impedance Selection
• Keypad Lockout
• Easy Operation
• Programmable Delay Time
Open/Short Zeroing: Reduce the effects of error sources (stray
admittance/residual impedance) between the device under test and the cali­brated connection to the measuring instrument.
Interfaces:For remote control, or adaption to a production environment, the 1715 instrument includes an RS-232 interface as standard equipment. Both an IEEE-488 and a Handler interface are available as optional interfaces.
For more detailed
specifications, visit
www.quadtech.com
For more information about special
purchase, rent & lease options, call
1-800-253-1230
Fax 1-978-461-4295
Intl. 1-978-461-2100
1715 Digibridge LCR Features
Parameter Measurement Range
Ls, Lp 0.01uH - 9.999kH Cs, Cp 0.01pF - 99.999mF Rs, Rp 0.1m - 99.99M D 0.0001 - 9999 Q 0.0001 - 9999 lZl 0.1m - 99.99M Xs 0.1m - 99.99M Phase -180.00° - +180.00°
Accuracy: Basic LCR ±0.2%
Basic DQ ±0.002
Test Frequency: 100Hz, 120Hz, 1kHz and 10kHz (9.6kHz)
Accuracy ±0.2%
Test Signal: 0.25V or 1.0V RMS, ±(10% + 3mV) Measurement Speed: Fast: 25 measurements/second
Medium: 8 measurements/second Slow: 2 measurements/second
Measurement Mode: Continuous or Trigger Trigger: Internal
External (IEEE or Handler) Manual
Trigger Delay: 0 - 9999ms Ranging: Automatic or Hold Range Equivalent Circuit: Series or Parallel Zeroing: Open and short circuit compensation Source Impedance: 25, 100 ,1kΩ, 10kor 100k
(based on DUT impedance)
Comparator: Primary Nominal
Primary Hi/Lo Limits (Value or %) Secondary Hi/Lo Limuts (Value or %)
Indication: Audible Alarm programmable: HI LO or
OFF for Pass or Fail Result
Binning: Primary Nominal
Primary Hi/Lo Limits (8 Bins in %) Secondary Hi/Lo Limits (Value)
Display: 40 by 4 Character LCD display Results Format: Value
Deviation % Deviation Bin Number
Setup Storage: Store 1 setup, Auto recall upon power-up Lock: Keypad Lockout
Interfaces (Standard): RS-232 Interfaces (Optional): IEEE-488 and Handler Test Terminals: Front: 4 BNC & 1 GND Dimensions: (w x h x d): 12.5 x 4.0 x 8.0in
(312.5 x 100.0 x 200.0mm)
Weight: 10.85 lbs. (5kg) net, 15.2 lbs. (7kg) ship Environmental: Specifications: +15°C to +35°C, 75% RH
Operating: 10°C to +40°C Storage: 0°C to +50°C Humidity: 10 -90% RH Pollution Degree 2 Installation Category II
Power: • 90 - 125V AC
• 190 - 250V AC
• 48 - 62 Hz
• 50VA max
Ordering Information
1715 LCR Digibridge
Includes:
QuadTech P/N Item 150710 1715 Instruction Manual 1700-03 Lead Set: 4 BNC Connectors to 2 Kelvin Clips 4200-0300 AC Power Cord 520023/520025 Power Line Fuses (0.630A 250V/0.315A 250V)
Calibration Certificate Traceable to NIST
For more detailed specifications, visit www.quadtech.com
For more information about special purchase, rent & lease options, call
1-800-253-1230 Fax 1-978-461-4295 Intl. 1-978-461-2100
5 Clock Tower Place, 210 East, Maynard MA 01754
Optional Accessories:
B/A Calibration Data 1700-01 Axial/Radial Component Test Fixture 1700-02 Axial/Radial Remote Test Fixture 1700-03 Lead Set: 4 BNC Connectors to 2 Kelvin Clips 1700-04 Lead Set: 4 BNC Connectors to 4 Banana Plugs 1700-05 Lead Set: 4 BNC Connectors to Chip Component Tweezers 7000-07 Low Voltage Chip Component Test Fixture 700171 IEEE-488 & Handler Interface
P/N 030126/A1