1715 LCR Meter
HIPOT • LCR METERS • MEGOHMMETERS • MILLIOHMMETERS
Passive Component Tester with RS-232
USES:
• Production Testing of LCR Components
• Component Design and Evaluation
• Component Inspection and Screening
• Quality Assurance Testing
• Material Testing
FEATURES:
• 11 Measurement Parameters
• 0.2% Basic Measurement Accuracy
• Selectable Test Frequency: 100Hz,
120Hz, 1kHz, 10kHz
• Selectable Test Voltage: 0.25V, 1.0V
• Measurement Speed to 25/second
• LCD Display
• RS-232 Interface Standard
•IEEE and Handler Interfaces, Optional
• Automatic Open/Short Zeroing
• Automatic Hi/Lo Comparator Limits
• Pass/Fail Sorting (8 Bins)
Introduction
The QuadTech 1715 LCR Digibridge is an economically priced, easily programmable, micro-processor based passive component tester for fast, precise
impedance measurements. It measures eleven different parameters with a
basic accuracy of 0.2%. The 1715 provides four different test frequencies and
two test voltages for testing flexibility. For remote operation and automated
system applications, the unit comes standard with an RS-232 interface, plus
IEEE-488 and Handler interfaces are available as options.
Description
11 Measurement Parameters:Measure and display two of 11 impedance
parameters simultaneously with a basic accuracy of 0.2%. Test results can also
be shown as Deviation or %Deviation from nominal value.
Test Signal:Four user selectable test frequencies: 100Hz, 120Hz, 1kHz, and
10kHz (9.6kHz), plus two test voltages: 0.25V and 1.0V.
Display:40 x 4 character LCD display for user friendly menu programming
and measured results. Simultaneously displays primary/secondary values and
binning or comparator results.
Setup Storage: Storage of 1 setup with parameter, binning and compare information saved. Auto recall of this setup upon power-up of instrument.
Auto Ranging:For maximum measurement integrity the 1715 automatically
selects the appropriate measurement range and instrument source impedance
(25
Ω, 100Ω, 1kΩ, 10kΩ or 100kΩ), dependent upon the impedance of the
device under test.
Pass/Fail Testing:The 1715 provides a programmable Hi/Lo comparator
function as well as eight sorting bins for categorization of components.
• Storage of 1 Setup
• Automatic Source Impedance Selection
• Keypad Lockout
• Easy Operation
• Programmable Delay Time
Open/Short Zeroing: Reduce the effects of error sources (stray
admittance/residual impedance) between the device under test and the calibrated connection to the measuring instrument.
Interfaces:For remote control, or adaption to a production environment, the
1715 instrument includes an RS-232 interface as standard equipment. Both an
IEEE-488 and a Handler interface are available as optional interfaces.
For more detailed
specifications, visit
www.quadtech.com
For more information about special
purchase, rent & lease options, call
1-800-253-1230
Fax 1-978-461-4295
Intl. 1-978-461-2100

1715 Digibridge LCR Features
Parameter Measurement Range
Ls, Lp 0.01uH - 9.999kH
Cs, Cp 0.01pF - 99.999mF
Rs, Rp 0.1mΩ - 99.99MΩ
D 0.0001 - 9999
Q 0.0001 - 9999
lZl 0.1mΩ - 99.99MΩ
Xs 0.1mΩ - 99.99MΩ
Phase -180.00° - +180.00°
Accuracy: Basic LCR ±0.2%
Basic DQ ±0.002
Test Frequency: 100Hz, 120Hz, 1kHz and 10kHz (9.6kHz)
Accuracy ±0.2%
Test Signal: 0.25V or 1.0V RMS, ±(10% + 3mV)
Measurement Speed: Fast: 25 measurements/second
Medium: 8 measurements/second
Slow: 2 measurements/second
Measurement Mode: Continuous or Trigger
Trigger: Internal
External (IEEE or Handler)
Manual
Trigger Delay: 0 - 9999ms
Ranging: Automatic or Hold Range
Equivalent Circuit: Series or Parallel
Zeroing: Open and short circuit compensation
Source Impedance: 25Ω, 100 Ω,1kΩ, 10kΩ or 100k Ω
(based on DUT impedance)
Comparator: Primary Nominal
Primary Hi/Lo Limits (Value or %)
Secondary Hi/Lo Limuts (Value or %)
Indication: Audible Alarm programmable: HI LO or
OFF for Pass or Fail Result
Binning: Primary Nominal
Primary Hi/Lo Limits (8 Bins in %)
Secondary Hi/Lo Limits (Value)
Display: 40 by 4 Character LCD display
Results Format: Value
Deviation
% Deviation
Bin Number
Setup Storage: Store 1 setup, Auto recall upon power-up
Lock: Keypad Lockout
Interfaces (Standard): RS-232
Interfaces (Optional): IEEE-488 and Handler
Test Terminals: Front: 4 BNC & 1 GND
Dimensions: (w x h x d): 12.5 x 4.0 x 8.0in
(312.5 x 100.0 x 200.0mm)
Weight: 10.85 lbs. (5kg) net, 15.2 lbs. (7kg) ship
Environmental: Specifications: +15°C to +35°C, 75% RH
Operating: 10°C to +40°C
Storage: 0°C to +50°C
Humidity: 10 -90% RH
Pollution Degree 2
Installation Category II
Power: • 90 - 125V AC
• 190 - 250V AC
• 48 - 62 Hz
• 50VA max
Ordering Information
1715 LCR Digibridge
Includes:
QuadTech P/N Item
150710 1715 Instruction Manual
1700-03 Lead Set: 4 BNC Connectors to 2 Kelvin Clips
4200-0300 AC Power Cord
520023/520025 Power Line Fuses (0.630A 250V/0.315A 250V)
Calibration Certificate Traceable to NIST
For more detailed specifications, visit www.quadtech.com
For more information about special purchase, rent & lease options, call
1-800-253-1230 • Fax 1-978-461-4295 • Intl. 1-978-461-2100
5 Clock Tower Place, 210 East, Maynard MA 01754
Optional Accessories:
B/A Calibration Data
1700-01 Axial/Radial Component Test Fixture
1700-02 Axial/Radial Remote Test Fixture
1700-03 Lead Set: 4 BNC Connectors to 2 Kelvin Clips
1700-04 Lead Set: 4 BNC Connectors to 4 Banana Plugs
1700-05 Lead Set: 4 BNC Connectors to Chip Component Tweezers
7000-07 Low Voltage Chip Component Test Fixture
700171 IEEE-488 & Handler Interface
P/N 030126/A1