Puritan Bennett LP10, LP6 Plus User Manual

LP6 Plus Volume Ventilator
-And­LP10 Volume Ventilator With Pressure Limit
Clinician’s Manual
Important Note: Read this manual in its entirety before using the ventilator. Keep this manual for future reference.
equipment and related services. Or, you may contact Puritan Bennett Inc. directly.
Puritan Bennett Technical Services: 1.800.255.6774
Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Purpose of the Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Symbols and definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Electrical Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Alarm Condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Responding to Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Upper Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Lower Left Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Lower Right Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Operating Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Operating Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Assist/Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Assist/Control with Pressure Limit (LP10 Only) . . . . . . . . . . . . 18
SIMV (Synchronized Intermittent Mandatory Ventilation) . . . . 18
SIMV with Pressure Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Pressure Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Standby . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Ventilator Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Breath Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Inspiratory or I-Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Breathing Effort . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Pressure Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Pressure Limit Control (LP10 Only) . . . . . . . . . . . . . . . . . . . . . . . . . 23
Power Sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
AC Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
External Battery 12 Volt DC . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Using a Car Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Internal Battery 12 Volt DC . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Routine Safety Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Monthly Safety Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Mounting or Positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Page iii
Emergency Vehicle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Wheelchair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
AC Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
External Battery 12 Volt DC . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Battery Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Testing the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Special precautions when using an external battery . . . . . . . . . . 41
Internal Battery 12 Volt DC . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Patient Ventilator Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Exhalation Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Humidification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Short Term . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Extended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Supplemental Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Positive End Expiratory Pressure (PEEP) . . . . . . . . . . . . . . . . . . 51
Pressure Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Supplemental Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Humidification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Remote Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Cleaning and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Patient Circuit and Humidifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Inlet Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Ventilator Surface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Storage of the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Scheduled Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Service Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Keyword Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Page iv

Introduction

Purpose of the Manual

This Clinician’s Manual will help you understand the operation of the LP6 Plus and LP10 Volume Ventilators. It provides detailed information for physicians and other clinical personnel about the installation, safe use, and ver­ification of the operation of the LP6 Plus and LP10 Volume Ventilators. It provides the caregiver guidelines for safe ventilation that are specific to the ventilator. It is not a complete maintenance document. Therefore, it contains no disassembly, repair, or reassembly instructions or diagrams.
Carefully read and understand all instructions before using the venti­lator. Use the instructions contained herein in conjunction with those set by
the patient’s physician. No instruction in this manual is intended to replace accepted medical practice regarding the use of the ventilator or the care of the patient.
As you read this manual, you will notice Cautions and Warnings in boxes on many pages. Pay very special attention to these boxes. They will tell you what to do and what to avoid as you use the ventilator.
The difference between Warnings and Cautions is:
A Warning contains information about possible hazards to the patient, the care provider, or the service technician.
A Caution includes information about how to avoid equipment damage.
The Notes and Accessories sections found at the back of this manual are for your use in applying this manual’s information to a specific patient. Puritan Bennett-supplied accessories include information concerning their use with the ventilator.
Page 1

Symbols and definitions

Symbols and definitions
The following symbols appear on the LP6 Plus and LP10 ventilators.
LP6 Plus and LP10 Clinician’s Manual
O
!
V A
I
Power switch ON position, connection to mains power
Power switch OFF position, disconnection from mains power
Attention, consult accompanying manual.
Alternating current
Direct current
Volts
Amperes
Standby mode of operation
Canadian Standards Association
Battery test switch
Alarm silence switch
Alarm
Page 2
Power
Underwriters Laboratory
Patient pressure
External battery connection
Remote alarm
Battery test level
Manual reset

Warnings and Cautions

Warnings
Always follow the physician’s prescription when using the venti­lator.
Always operate and store the ventilator according to the specifi­cations and instructions set forth in this manual.
Use only Puritan Bennett-approved accessories and products with the ventilator. The use of other accessories may damage the unit and endanger the patient.
Perform daily and monthly verification of the ventilator’s opera­tion as identified in this manual.
Always stabilize and verify ventilator performance before con­necting the patient to the unit.
All alarms indicate a potential risk to patient safety. When an alarm sounds, provide immediate attention, care, and support to the patient as dictated by the situation.
The LP6 Plus and LP10 ventilators shall not be used with flam­mable anesthetic agents.
Do not use in direct sunlight.
Cautions
Refer any adjustments or procedures exceeding the scope of this manual to an Puritan Bennett Technical Service Representative. Refer to the Puritan Bennett Service Policy on page 62.
Caution: Federal Law (U.S.A.) restricts this device to sale or use by or on the order of a licensed physician.
Page 3

Electrical Interference

Electrical Interference
LP6 Plus and LP10 Clinician’s Manual
Caution
Warning
Your ventilator is an electronic instrument. Any electronic instru­ment is subject to electrical interference. Electrical interference in excess of 10 V/m may keep your ventilator from working properly.
Television sets, cordless or cellular telephones, microwave ovens, air con­ditioners, food processors, and other appliances can be sources of electri­cal interference. To avoid electrical interference between your ventilator and these appliances, you must follow these instructions:
Never place your ventilator near these appliances.
Never plug the ventilator into the same A.C. electrical outlet as
these appliances, nor into electrical outlets on the same circuit as these appliances.
Never place the cables from ventilator accessories near these appli-
ances.
Electrical interference may keep your ventilator from working properly, which may create a hazard to the patient.
Page 4
Note The ventilator is exempt under Section 15.801 (c)(5) of the no interfer-
ence regulations adopted by the FCC. If television interference does occur, contact Technical Services at Puritan Bennett, Inc. or a television repair technician for suggestions. Or, move the television to an A.C. electrical outlet that does not allow interference.

General Description

The Puritan Bennett LP6 Plus and LP10 Volume Ventilators are intended for use in a non-acute care institution or transport, on pediatric and adult patients. It is to be operated in accordance with the product labeling con­tained in this instruction manual.
The LP6 Plus and LP10 Volume Ventilators are microprocessor-controlled volume ventilators. They provide continuous respiratory support for patients with respiratory insufficiencies in a skilled nursing facility or hospital, or during transport. Because of the compact design and light weight, the units are highly portable.
The ventilators offer a wide range of delivery volumes, inspiratory times, and breathing rates. The physician or the respiratory therapist can set the appro­priate ventilation via the controls located in the recessed front panel. The magnetically latched door panel and the control knobs are designed to pre­vent tampering and accidental resetting.
Audible and visual alarms quickly identify problems. See pages 6 through 9 for a complete discussion of these alarms.
Page 5

Alarm Condition

Alarm Condition
LP6 Plus and LP10 Clinician’s Manual
Warnings
All alarms indicate a potential risk to patient safety. When an alarm sounds, provide immediate attention, care, and support to the patient as dictated by the situation.
Any device is subject to unpredictable failures. To ensure patient safety, an appropriately trained caregiver should monitor venti­lation. If the patient’s condition warrants the use of an indepen­dent secondary alarm, remote alarm, or another external monitoring device, the physician should prescribe it. The physi­cian should also determine to what level the patient may require an alternate means of ventilation.
Certain types of ventilators, including the LP Series, have a Low Inspiratory Pressure Alarm. The purpose of this alarm is to alert the clinician or caregiver when the pre-set alarm parameters are violated. As set forth in the LP Series Clinician’s and User’s Manuals, a number of environmental factors and circuit accesso­ries/components can affect the pressure in the breathing circuit. These factors may prevent circuit pressure from violating the low-pressure parameters, even in the event of a circuit being dis­connected from the patient. Therefore, it is important for the cli­nician to consider and monitor these environmental factors when establishing pressure alarm parameters. Depending on the spe­cific clinical situation (e.g., risk of disconnect perceived as high, patient is ventilator dependent) a secondary means of monitoring ventilation (e.g., pulse oximetry) should be considered.

Responding to Alarms

The ventilator has visual and audible alarms. The audible alarm is usu­ally a pulsating tone. Both the ventilator and the remote alarm emit these tones. Flashing or steady light(s) on the ventilator indicate the source of the problem.
A Pulsating Audible Alarm and Flashing Light(s):
The Low Power, High Pressure, Setting Error, and Power Switchover alarms all use this type of alert signal.
Page 6
LP6 Plus and LP10 Clinician’s Manual
Note In Assist/Control and SIMV modes, the High Pressure Alarm sounds
Responding to Alarms
A Steady Audible Alarm and Steady Lights:
This combination indicates a detected microprocessor error in the venti­lator.
Single Reminder Tone:
A single tone sounds every five minutes when the internal battery pow­ers the ventilator.
whenever the air pressure exceeds the selected high pressure limit. In the Pressure Cycle mode, however, the High Pressure Alarm sounds only when the air pressure exceeds the selected high pressure limit by 10 cmH2O/hPa.
When an alarm sounds:
First, attend to the patient immediately. Then, check the flashing or
steady light(s) on the ventilator to identify the source of the problem.
You may press the Alarm Silence/Reset button to silence the alarm. This turns off the signal for one minute. If the alarm condition is corrected during that minute, the alarm light will turn off.
A microprocessor error cannot be silenced. You cannot silence an alarm before it occurs.
Note If a High Pressure, or Setting Error alarm condition is corrected before
you press Alarm Silence/Reset, the audible alarm will stop but the light will continue to flash. Press Alarm Silence/Reset to turn off the light.
If a Low Pressure/Apnea, Low Power or Power Switchover alarm condi­tion is corrected before you press Alarm Silence/Reset, both the audible and visual alarms will continue. You must press Alarm Silence/Reset to turn off the audible alarm and the light.
Warning
If alarms continue to sound, provide another means of ventila­tion and contact your homecare dealer.
Page 7

Troubleshooting Guide

Troubleshooting Guide
Conditions Probable Cause Solution
LP6 Plus and LP10 Clinician’s Manual
All lights turn on and audi­ble alarm sounds
Low Pressure/Apnea Alarm: Pulsating audible tone with flashing light
Normal condition. Alarms test when unit is turned on.
Normal; manual alarm test. Alarms will stop in one second.
Microprocessor error. Turn vent off and set mode to Standby.
The patient is not breathing. Check the patient for breathing effort.
Leaks or loose connections in the patient circuit.
Water in small-bore tubing. Inspect and remove water from small-
Crimped small-bore tubing. Uncrimp the small-bore tubing.
PEEP pressure set higher than the Low Alarm control setting.
The patient’s breathing effort is less than the Breathing Effort control setting.
Alarms will stop in two seconds.
Wait a few seconds. Return switches to prescribed settings. If alarm persists, provide another means of ventilation.
Check connection of the patient circuit to the ventilator; check all connections for leaks and tightness, especially at the humidifier, trach tube, and exhalation valve.
bore tubing.
Set Low Alarm control setting higher than the PEEP pressure.
Set Breathing Effort so the patient’s breathing effort turns on the Breathing Effort light.
Page 8
Patient speech or other activities lower patient airway pressure.
Low alarm setting is higher than Pres­sure limit setting. (LP10 only)
Volume set below patient’s tidal vol­ume.
Pressure Limit level is set too low. (LP10
only)
Incorrect control settings. Reset to prescribed values.
Leaks or obstructions in the patient cir­cuit.
Other causes. Notify your physician and your home-
Low pressure alarm sounds whenever low pressure limit is not reached for two consecutive breaths. Review the section on alarms.
Correct to the prescribed value.
Correct to the prescribed value.
Correct to the prescribed value.
Check for leaks or crimped tubing.
care dealer
LP6 Plus and LP10 Clinician’s Manual
Conditions Probable Cause Solution
Troubleshooting Guide
Low Power Alarm: Pulsating audible tone with flashing light
High Pressure Alarm: Pulsating audible tone with flashing light
Setting Error Alarm Inappropriate setting or settings beyond
Failure to recharge the Internal battery. Operate the ventilator on AC power for
Water in the tubing. Remove water from tubing.
Crimped tubing. Straighten crimped tubing.
Coughing or other high-flow expira­tory efforts.
Patient inspiratory resistance or compli­ance changes.
A sticky Pressure Limit control. Occlude the end of the patient circuit to
Airway obstruction Check for trach obstruction or for a
Malfunction in the exhalation manifold. See the manifold manufacturer’s
Pressure Limit setting is higher than the High Alarm setting. (LP10 only)
the capabilities of the machine.
at least three hours, or place ventilator in Standby Mode while on AC power; use backup ventilator.
Treat patient’s cough. The alarm is appropriate for these conditions.
Have physician determine new ventila­tor settings.
free the valve.
condition in which the patient requires suctioning.
instructions.
Reset both to the prescribed values.
Readjust settings to the physician’s pre­scription.
Dirty inlet filter. Replace filter.
Internal Battery light flashes Unit has not switched to external bat-
tery.
DC circuit breaker is open. Reset by pushing in protruding rod.
Single tone Unit is operating on internal battery. Check for unconnected or miscon-
Green AC Power light does not glow
Warning
AC circuit breaker is open. Turn it back ON.
AC power cord is not connected. Plug in the cord.
No power at the wall outlet. Use an active outlet.
If the problems continue, provide another means of ventilation
Check for unconnected or miscon­nected battery cable. Check for blown fuse in the battery cable. Use another external battery.
nected battery cable. Check for blown fuse in the battery cable. Use another external battery.
and contact your homecare dealer.
Page 9

Front Panel

Front Panel
LP6 Plus and LP10 Clinician’s Manual
Page 10
The Front Panel of the ventilator has three sections:
The upper section has small lights, two touch button pads, and a
meter.
The lower left section has the operating controls. The physician pre-
scribes their setting. To prevent accidental resetting, they are behind a closed panel.
The lower right section has the Pressure Limit control and the con-
nections for the Patient Circuit.
LP6 Plus and LP10 Clinician’s Manual
Upper Section 1. Alarm Lights: When flashing or continuously lit, they identify a
Front Panel
condition that demands immediate attention. There is also an audi­ble tone when these lights begin flashing.
2. Alarm Silence/Reset Button: This has five uses.
Push to test the alarms.
Push to silence alarms for 60 seconds.
Push to reset the alarm after correcting the problem.
Push simultaneously with the Battery Test Button for operating
hours. (See the Scheduled Maintenance portion of the manual, page
60.)
Use this button with other controls to start the self-test. (See page
30.)
3. Power Source Lights:
The top light is green when the ventilator is AC powered.
The middle light is amber when an external battery powers the ven-
tilator.
The bottom light flashes amber when the ventilator’s internal bat-
tery is in use. A single tone also beeps every five minutes.
4. Battery Test Button: It has four uses.
When pressed, the Pressure Meter displays the charge status of the
battery in use (internal or optional external battery).
Push simultaneously with the Alarm Silence/Reset button for an
indication of operating hours. (See the Scheduled Maintenance por­tion of this manual, page 56.)
Press the button to print a report from an attached printer.
Use this button with other controls to start the self test. (See page
30.)
5. Breathing Effort Light: This light turns green whenever the venti-
lator senses the patient’s effort to breathe. The Breathing Effort con­trol sets the sensitivity.
6. Patient Pressure Meter: The meter displays three pieces of infor-
mation:
Pressure at the Exhalation Manifold.
The number of hours of ventilator operation.
The charge status of the internal or attached external battery.
Page 11
Front Panel
Lower Left Section
LP6 Plus and LP10 Clinician’s Manual
1. Control Panel Door: This door is latched magnetically to pro-
tect the controls from accidental resetting.
2. Alarm Reference Guide: Consult this guide for a summary of
alarms and the action you should take. You will find the Guide on the inside of the Control Panel Door.
3. Control Knobs: They are behind the closed Control Panel Door.
The patient’s physician prescribes the settings for these controls. See pages 16 through 22 for details.
Page 12
LP6 Plus and LP10 Clinician’s Manual
Lower Right Section
Front Panel
1. Patient Air Tube: The Patient air hose connects to this tube. The
ventilator delivers air through this tube.
2. Exhalation Valve Port: The Exhalation Pressure Tube of the
Patient Circuit connects to this port.
3. Patient Pressure Port: The Patient Pressure Tube of the Patient
Circuit connects to this port.
4. Pressure Limit Control (LP10 Only): This control sets the air pres-
sure limit during a forced or assisted breath. For use in Assist/Con­trol or SIMV modes only; use in other modes may not allow effective ventilation. See pages 23 through 25.
Page 13

Rear Panel

Rear Panel
Vent
AC Plug
Carrying
Handle
External
12 Volt DC
Battery
LP6 Plus and LP10 Clinician’s Manual
Remot e Ala rm
Connector
Mounting
Rails
14
Inlet
Filter
1
11
Pressure
Relief
10
9
8
6
7
DC Circuit
Breaker
12
Communi-
cation Port
4
5
Serial Num-
ber Plate
Page 14
2
Cord Wrap
Rear Feet
13
Voltage
Select
3
AC Power
Switch
LP6 Plus and LP10 Clinician’s Manual
Rear Panel
1. Inlet Filter: The ventilator draws in air through this filter.
Warning
Do not block the inlet filter. (Keep away from curtains.)
2. Cord Wrap and Rear Feet: There is a foot at each corner.
3. AC Power Switch/Circuit Breaker: This is the ON/OFF switch
for AC power. It also has a built-in circuit breaker. 1 is power con­nected to mains and 0 is power disconnected from mains.
4. Communications Port: A special cable fits here and leads to an
optional printer.
Note The ventilator and printer should be turned off before connection or dis-
connection of the printer.
5. Serial Number Plate: This has Puritan Bennett’s identification
number for the ventilator. It also lists the unit’s power requirements.
6. External 12 Volt DC Battery Connector: This is where you plug
in an external battery cable.
7. Remote Alarm Connection: A remote alarm (optional) can sum-
mon the caregiver when an alarm sounds.
8. Rear Carrying Handle Recess
9. AC Plug Holder
10. Vent: Warm air from the unit’s circuitry leaves the ventilator and
cool air enters through this vent. This cools the ventilator.
Warning
Warning
Caution
Do not block rear panel vent
11. Pressure Relief: This prevents the air pressure from exceeding
approximately 100 cmH
Do not block the pressure relief valve.
12. DC Circuit Breaker: This circuit breaker protects the ventilator
O/hPa.
2
when it is powered by an external battery.
13. Voltage Select Switch: This switch selects 110 or 220 volts when it
is powered by an external battery
An incorrect switch setting may damage your ventilator.
14. Mounting Rails: These are used to mount and connect accessories to
the ventilator.
Page 15

Operating Controls

Operating Controls
LP6 Plus and LP10 Clinician’s Manual
Warning
Warning
Periodically check the control settings to be sure they are at the prescribed setting. Always verify that the controls are set correctly before connecting and using the ventilator. Do not change them without a physician’s orders.
1. Mode: This control selects the operating mode for the ventila-
tor.
2. Volume: This sets the amount of air the patient receives for each
breath. To change the setting, push in the control and then turn it to the selected setting. Any change made during operation results in a maximum change of 100 milliliters from breath to breath until the new volume is reached.
For more precise accuracy in setting the volume control (indica­tor), the use of an external volume measuring device is recom­mended.
3. Breath Rate: This setting controls the minimum number of breaths
per minute (BPM) delivered by the ventilator.
4. Inspiratory Time: This sets the time it takes for the ventilator to
deliver a breath.
Page 16
LP6 Plus and LP10 Clinician’s Manual
Operating Controls
5. Breathing Effort: This adjusts the ventilator’s sensitivity to the
patient’s breathing effort. When the patient’s effort reaches the set­ting, the Breathing Effort light turns on and the ventilator delivers a breath. Push in the control knob to change the setting.
6. High Alarm/Limit: This sets the point at which the High Pressure
alarm will sound. If the pressure exceeds this limit, the High Pres­sure alarm sounds (except in the Pressure Cycle mode). Delivery of this breath stops after the pressure reaches this limit. The audible alarm is automatically silenced if the following breath does not exceed the setting.
Often, when the high alarm sounds, the patient circuit has an obstruction. Clearing the circuit will stop the alarm.
7. Low Alarm: This establishes the pressure which must be exceeded
with “controlled” or "assisted" breath. The alarm sounds only when two consecutive breaths do not reach the selected limit, or if the limit is reached but the pressure fails to return to a level below the limit. Normally, this setting is just below the pressure you need for proper ventilation.
If this alarm sounds, look for a leak in the patient circuit. A leak or disconnection of the circuit is often the cause for the alarm.
Note Some circuit components will prevent a Low Pressure alarm by keeping
the pressure in the circuit above the alarm limit. Examples of these com­ponents include hydrated heat and moisture exchangers (HMEs) and tra­cheostomy tubes. If the patient circuit is disconnected from the patient, but still connected to these components, a Low Pressure alarm may not sound.
Where such disconnections from a ventilator-dependent patient are pos­sible, you must set the Low Pressure alarm to a level that permits an alarm to sound. To do this, simulate the disconnection; if a Low Pres­sure/Apnea alarm does not sound after two breath cycles, increase the alarm limit until an alarm sounds.
8. Pressure Limit (LP10 Only): This limits the maximum pressure
developed for each breath. For use with Assist/Control or SIMV modes only; use in other modes may not allow effective ventilation. See Operating Modes section, below, for details.
Page 17

Operating Modes

Operating Modes
Assist/Control In this mode, if the breathing effort is strong enough to trigger the
Breathing Effort light, the ventilator assists breathing. The ventilator then delivers the selected prescribed volume of air. If the patient makes no effort to breathe and, thus, fails to activate the Breathing Effort light, the ventilator takes control. It delivers breaths at the selected rate.
Breath Rate set at 1 - 5 BPM
If the patient does not start a breath on his/her own for 10 seconds, the Apnea alarm sounds and alerts the caregiver. Meanwhile, the ventilator delivers breaths at 10 BPM at the selected volume.
Breath Rate set at 6 BPM or more
No alarms sound if the patient fails to start a breath. The ventilator con­tinues to deliver breaths at the selected rate and volume.
LP6 Plus and LP10 Clinician’s Manual
Assist/Control with
Pressure Limit (LP10 Only)
Note Since the pressure limit function bleeds off air to limit pressure, the vol-
SIMV (Synchronized
Intermittent Mandatory
Ventilation)
The ventilator functions as described under the Assist/Control mode. The only difference is that the ventilator will limit the pressure during a delivered breath.
ume of air delivered will be less than the set value.
In this mode, the patient can breathe unassisted and on his/her own between ventilator delivered breaths. The ventilator monitors these spontaneous breaths.
If the patient’s effort is not strong enough to turn on the Breathing Effort light or the patient makes no effort, the ventilator delivers a con­trolled breath. All this depends on the breath rate setting:
Breath Rate set at 1 - 5 BPM
If the patient does not start a breath on their own for 20 seconds, the Apnea alarm sounds and alerts the caregiver. Meanwhile, the ventilator delivers breaths at 10 BPM at the selected volume.
Breath Rate set at 6 BPM or more
No alarms sound if the patient fails to start a breath. The ventilator con­tinues to deliver breaths at the selected rate and volume.
Page 18
Loading...
+ 50 hidden pages