Puritan Bennett KnightStar 330 User Manual

KnightStar ® 330
®
Bi-Level
Ventilator
Y-500008-00 Rev. H
Clinician’s Manual
KnightStar®, Bi-Level®, Companion®, and SoftFit® are registered trademarks of Puritan-Ben­nett Corporation. Sullivan® is a registered trademark of ResMed, Inc. ADAM™ and Breeze™ are trademarks of Puritan-Bennett Corporation. For more information, contact your Puritan Bennett representative.
iiii
ii

Contents

Contents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .i
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Patient and Clinician Access Levels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Warnings, Cautions, and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Control Panel Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Control Panel Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Control Panel Indicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Air Outlet Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Inlet Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
System Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Power On Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Operating Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Modes/Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Display Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Display Preferences. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Changing Device Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Alarm Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Setting Prescription Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Sensitivity Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Inspiratory Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Expiratory Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Rise Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Clinical Application and Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Connecting the Device to the Patient. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Titrating Therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Using the Optional Humidifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Using Supplemental Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Connecting Oxygen to the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
i
Rebreathing of Carbon Dioxide . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
Cleaning Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Cleaning the Exterior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Cleaning the Inlet Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Replacing the Optional Air Outlet Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Appendix A: KnightStar 330 Setup Checklist. . . . . . . . . . . . . . . . . . .49
Appendix B: KnightStar 330 Specifications . . . . . . . . . . . . . . . . . . . .52
Appendix C: What the Patient and Caregiver Must Know . . . . . . . . .53
Appendix D: Service Information . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
Appendix E: Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
ii

Introduction The Puritan Bennett KnightStar 330 is a continuous bi-level ventilator

that provides noninvasive ventilation for the treatment of respiratory insufficiency and obstructive sleep apnea that may occur in the home.
The KnightStar 330 is also indicated for the treatment of respiratory failure in institutional environments, and is intended to assist the ventilation of spontaneously breathing patients who are over 30 kg (66 lbs) in weight.
CAUTION:
Read this manual and the KnightStar 330 User’s Manual thoroughly before operating the device. The manuals pro­vide clinical as well as technical infor­mation concerning the operation and performance of the Puritan Bennett KnightStar 330 bi-level ventilator.
The KnightStar 330 is a microprocessor-controlled pressure generator capable of monitoring the air flow and controlling the pressure delivered to the patient. The KnightStar 330 possesses the following features:
Provides three operation modes: CPAP, I/E PAP, and Assist Control (A/C).
Monitors pressure, tidal volume, respiratory rate, air leaks, peak flow, and the I:E ratio.
Provides adjustable inspiratory and expiratory trigger sensitivity.
Provides precise respiratory support and patient comfort.
Provides audible and visual indicators to alert users to power failure, system leaks, device performance.
Allows a maximum pressure setting of 30 cmH2O; with a pres­sure limitation of 40 cmH
O for a single-fault condition.
2
Compensates for delivered pressure within specification for alti­tudes from 0 to 8,000 feet (2438 meters), at 4 to 30 cmH
O; and
2
compensates for leaks up to 60 liters per minute.
1
There are certain limitations and instructions that must be understood by the clinician and patient before using the KnightStar 330. Refer to Appendix C: What the Patient and Caregiver Must Know.
2
Patient and
The KnightStar 330 features two access levels:
Clinician Access Levels
Patient access (Lockout mode “Active”)
Clinician access
The patient access level enables the patient to turn the device on and off, and to adjust comfort settings.
The clinician access level enables the caregiver to access all of the prescription settings and device controls, as well as the patient access features.
In the sleep lab, the KnightStar 330 can be operated with either the optional remote control, or the control panel on the device. The device’s controls enable the user to input the patient’s prescription settings and review the estimated tidal volume, estimated peak flow, estimated leak, respiratory rate, I:E ratio, IPAP, and EPAP settings.
For home care applications, the home care provider can set the patient’s prescribed parameters. All prescription parameters programmed by the home care provider are stored in the KnightStar 330’s memory.
If the prescription settings are corrupted, the KnightStar 330 will not operate. Instead, an alarm symbol and error number appear on the display, and the audible alarm will sound.
3
Figure 1 shows the components that make up the KnightStar 330
system.
KnightStar 330 bi-level ventilator
Power cord
Tubing and proximal pressure line
Puritan
Bennett
KnightStar 330
Bi-level Ventilator
Inlet air filter (with spare)
Optional outlet air filter
Clinician’s Manual
Figure 1. KnightStar 330 Components
It is recommended to use the KnightStar 330 with 6-ft (1.8 m) or 8-ft (2.4 m) tubing and Puritan Bennett nasal interfaces, and Breeze™ or ADAM™ circuits.
Three modes of operation are available:
CPAP (continuous positive airway pressure)
I/E PAP (inspiratory and expiratory positive airway pressure)
A/C (Assist Control)
In the CPAP mode, the system delivers a continuous positive regulated airway pressure throughout the breath cycle (normal operating range is from 3 cmH2O to 20 cmH2O).
4
In the I/E PAP mode, the system tracks patient breathing effort and provides two levels of pressure—a higher level of pressure for inspiration (normal operating range from 3 to 30 cmH pressure for expiration (normal operating range from 3 to 20 cmH
O) and a lower
2
O).
2
In the A/C mode, the system delivers the same two levels of pressure as described for the I/E PAP mode with the addition of a backup breath rate (normal operating range from 3 to 30 breaths per minute) and an I:E ratio (normal operating range from 1:1.0 to 1:4.0).
When Lockout mode is active, the settings available to the patient are:
•Delay Time
•Ramp Time
Ramp Starting Pressure
CAUTION:
Before using the Knightstar 330, read all warnings and cautions in the next section.
5
Warnings, Cautions, and
Information about specific hazards or special significance are presented in the following formats:
Notes
WARNING: Indicates a condition that can endanger the patient or the device operator.
CAUTION: Indicates a condition that can damage the device and/or other property.
NOTE:
Indicates information of particular interest for more efficient and convenient device operation.
WARNINGS:
Clinical research indicates that CPAP therapy may be CONTRAINDICATED for patients with the following pre­existing conditions:
Bullous lung disease
Pneumothorax
Severe cardiac rhythm disturbances
Extremely low blood pressure
Pneumocephalus or pre-existing CSF leaks or head trauma (Chest 1989; 96: 1425 - 1426)
Acute sinus or middle ear infection (may be an indication to suspend CPAP therapy temporarily)
Unstable airway
Acute facial trauma
The physician’s prescription should be based upon the appropriate diagnostic testing. The prescribed nasal pres­sure should only be adjusted by trained, authorized person­nel in accordance with the physician’s prescription.
Use only interfaces and accessories that are approved by Puritan Bennett.
6
WARNINGS (continued): The physician’s prescription should be followed in accor­dance with established medical protocols.
An alternate means of ventilation must be available when patients are being treated for respiratory failure.
Alarms should never be disabled for patients who could be injured due to ineffective or interrupted ventilation. The phy­sician should determine secondary or independent alarms.
Alarm volume should be set in accordance with ambient noise level. Respond immediately to all alarm conditions.
Patients receiving supplemental oxygen should be advised of the hazards of combustible materials and flames or sparks in the presence of oxygen. Do not smoke in the presence of oxygen.
To prevent oxygen from accumulating in the device and tub­ing, turn on the device before turning on the oxygen supply; shut off the oxygen before turning off the device.
Patients receiving supplemental oxygen and nasal pressure therapy should be monitored for arterial blood oxygen satu­ration.
Configure the KnightStar 330 system as shown in this man­ual for safe and effective operation.
Always place the KnightStar 330 upright on a firm, flat sur­face. Setting KnightStar 330 on uneven surfaces or in a posi­tion that is not upright may result in damage to the unit.
Do not use the KnightStar 330 with antistatic or electrically conductive tubing.
The KnightStar 330 should never be operated in the presence of anaesthetic gases. The equipment is not suitable for use in the presence of a flammable anaesthetic mixture with air, or with oxygen or nitrous oxide. Placing the unit in such an area may result in an explosion.
7
WARNINGS (continued): The KnightStar 330 device should never be operated where the air intake might draw in hazardous gases from external sources such as gas stoves, engine exhaust, or anesthesia machines. Placing the unit in such an area may result in asphyxiation of the patient.
Do not set the device on or within 3 feet (1 m) of electric or electronic appliances, such as space heaters, electric blan­kets, or televisions. Do not operate cordless phones near the device. Doing so may result in device malfunction.
The KnightStar 330 should be used with care to avoid over­heating the patient when the room temperature exceeds 90 °F (32.2 °C), since under certain conditions the patient outlet gas flow can be as much as 6.7 °F (3.7 °C) degrees warmer than room temperature.
The KnightStar 330 should be used only with interfaces rec­ommended by the device’s manufacturer. An interface should not be used unless the device is turned on and oper­ating properly. The purge hole(s) associated with the inter­face should never be blocked. The purge hole(s) allow a continuous flow of air out of the interface. When the device is turned on and operating properly, fresh air from the device flushes most of the expired air out through the interface purge hole(s). However, when the device is not operating, a substantial proportion of expired air and carbon dioxide may be rebreathed. Rebreathing of carbon dioxide can increase levels of CO
to become somnolent and may even result in death.
, and in some circumstances cause the patient
2
This device must never be operated with an obstructed air­way circuit. Prevent foreign matter from entering the airway circuit. Failure to do so could result in asphyxiation of the patient.
Should the patient experience excessive nasal or airway dry­ness, skin sensitivity, runny nose, ear pain, sinus discomfort, daytime sleepiness, mood change, disorientation, or memory lapse when using this device, discontinue use and call the physician.
8
WARNINGS (continued): Do not block or restrict airflow around the device. Unimpeded airflow is necessary to maintain proper pressure and flow to the patient.
For patient health and comfort, it is important to clean the patient interface regularly. Follow the cleaning instructions that came with your patient interface.
Contact the home care company if the equipment malfunc­tions in any way. Do not attempt to open the device case. Only qualified personnel may service this equipment.
To reduce the risk of strangulation, ensure that the patient tubing is routed away from the patient’s head.
The KnightStar 330 equipment has been tested and found to comply with the limits for medical devices to IEC 601-1­2:1993 (or EN 60601-1-2:1994 or Medical Device Directive 93/ 42/EEC). This testing shows the device provides reasonable protection against harmful interference in a typical medical installation. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to other devices or is nega­tively impacted by other devices, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the devices
Increase the separation between the devices
Connect the equipment to an outlet on a different circuit
Consult the manufacturer or your local representative for help
The factory default setting for Lockout mode is inactive. The clinician is responsible for activating Lockout mode.
Under certain conditions, some alarms may not occur. For example: (1) The leak alarm may not occur if patient breath efforts are not detected, as in the case of excessively large leaks; and (2) The low pressure alarm may not occur under conditions such as an incorrect alarm threshold setting, or air pathway resistance. Check all alarms and settings for correct alarm operation prior to use.
9
WARNINGS (continued): Be careful when handling the KnightStar 330 during or imme­diately after use. Under specified operating conditions, some surfaces of the unit may become hot to the touch. This is a normal occurrence and is typical of this type of device.
CAUTIONS:
Federal (USA) Law restricts this device to sale by or on the order of a physician.
The KnightStar 330 will discontinue operation upon loss of A/C power. The optional, external 12 V battery may be used as an alternate power source, but it is not intended for emer­gency backup power. Either A/C or external battery power may be connected to the KnightStar 330, but not both simul­taneously. Refer to the battery instruction sheet.
Inspect the inlet air filter often. Remove the foam filter from the rear panel and clean it at least once per week.
NOTES:
At the end of the KnightStar 330’s useful life, return the device to the manufacturer for proper disposal.
10

Symbols Table 1 lists descriptions for the various symbols that appear on the

KnightStar 330.
Table 1. Symbols
Symbol Description
!
Attention, consult accompanying manual
Alternating current (AC power from wall outlet)
Direct current (battery power)
Type BF equipment, degree of protection against electrical shock
Class 2 equipment, double insulation design
Alarm condition
CE Mark: This device complies with the requirements of Medical Device Directive 93/42/EEC concerning medical devices
A/C power cord connection
Air outlet connector (blower connector)
External Battery/DC power connector
RS-232 communications connector
UL mark, classified by Underwriters Laboratories Inc. with respect to electric shock, fire, and mechanical hazards only in accordance with standards UL2601-1 and CAN/CSA C22.2 No. 601.1-M90.
11
Table 1. Symbols (continued)
Symbol Description
IPX1 Drip proof
SN Serial Number
REF Product model number
Max
Min
Storage temperature range
12

System Description

Control Panel Display

Control Panel Buttons

The Liquid Crystal Display (LCD) shown in Figure 2 provides an easy-to-read format for mode, settings, and patient parameters. A backlight illuminates the display when the Mode or Set button is pressed. The display remains illuminated for approximately 60 seconds after the last button is pressed.
The control panel buttons are shown in Figure 2 and listed in Table 2.
PURITAN BENNETT
TM
Control panel buttons
Mode
Figure 2. KnightStar 330 Control Panel
Set
13
Delay Ramp
Alarm
Silence
Table 2. Control Panel Buttons
Symbol Name Function
On/Off Turns the KnightStar 330 on or off.
Turn the KnightStar 330 on with a quick press and release of the On/Off button. The device retains the prescription settings last entered.
To turn the device off, press and hold the On/Off button for 3 seconds.
Mode
Scrolls through various device modes.
Mode
On + Mode +
Mode +
Set
Leave Settings Mode
Autoclear three-button combination
Lockout Mode and Toggle
Set Scrolls through the available parameters. Press Set once to sh ow the
Press the Mode button to scroll through various modes, as follows: CPAP, I/EPAP, A/C.
Note: If the Lockout function is active, the Mode button will not operate.
Performs a device Autoclear. When the KnightStar 330 is in the Stand-by mode (plugged in to AC
power but not operating), perform an Autoclear by pressing and holding the following three buttons simultaneously for approximately 20 seconds:
On, Mode, and (Up Arrow) Pressing this three-button combination clears the updatable “flash”
memory and restores the device’s defaul t values. Within approximately 20 seconds after simultaneously releasing the
buttons, you will recognize that this process is occurring by the “Xs” that appear on the display (in place of the patient ID) during Power On Self Test (POST).
Changes the Lock or Unlock position. If the Lockout mode is active, the patient may only change the delay, start pressure, and ramp duration functions.
To change the Lock or Unlock position, hold the Mode button and the (Up Arrow) button simultaneously for approximately two seconds.
patient-settable parameters (delay, start pressure, ramp duration). Press Set again to scroll through the remaining parameters.
If the
Lockout mode is inactive, you may scroll through all of the available parameter settings. If the Lockout mode is active, you may only scroll through the patient-settable parameters (delay, start pressure, ramp duration).
14
Table 2. Control Panel Buttons (continued)
Symbol Name Function
Delay/Ramp Starts or stops the Delay/Ramp function. Press the /Delay/Ramp
(Down Arrow/Delay/Ramp) button to start the Delay/Ramp function, if inactive; press the /Delay/Ramp button to stop this function, if active.
Delay Ramp
Down Arrow Decreases a selected setting value when in Settings mode. Press the
/Delay/Ramp (Down Arrow/Delay/Ramp) button once to decrease a setting value by one decrement.
Alarm
Silence
Alarm Silence
Up Arrow Increases a selected setting value when in Settings mode. Press the
Display Secondary Screen
Mutes an active alarm. Press the Silence) button once to silence an active alarm for one minute.
/Alarm Silence (Up Arrow/Alarm Silence) button once to increase a
setting value by one increment. Displays and I:E ratio.
In AC or I/E mode, when the main display screen is shown, pressing this button displays and I:E ratio if there are no active alarms.
V
V
/Alarm Silence (Up Arrow/Alarm
/Delay/Ramp Button. This button is used to activate the Delay
feature. When Delay is activated, both inspiratory and expiratory pressures will decrease to the Ramp Start pressure. The time of delay can be set from 0 (no delay) to 30 minutes. After the delay time has elapsed, pressure will slowly ramp up to the prescription pressures.
The Delay mode can be cancelled by again pressing the ▼/Delay/ Ramp button. Once activated, Delay can be restarted by again pressing the ▼/Delay/Ramp button.

Control Panel Indicators

The KnightStar 330 control panel features visual indicators (shown in
Figure 3) that illuminate in the presence of power and in response to
specific device or tubing circuit problems.
The presence of power, whether from A/C or external battery, is indicated by an illuminated green LED.
15
A low priority condition is indicated by a steadily illuminated
yellow LED (without an audible alarm).
A medium priority condition is indicated by a flashing yellow LED, along with an audible alarm that beeps three times at intervals of 25 seconds.
A high priority alarm is indicated by a flashing red LED, along with an audible alarm that beeps five times at intervals of 10 seconds.
Mode
Green
Figure 3. Control Panel Indicators
Set
Delay Ramp
Yellow
Control Indicators (LEDs)
Red
Alarm
Silence
NOTE:
An audible alarm will sound under both medium and high priority alarm conditions.
Refer to the Troubleshooting section on page 45 for possible causes and corrective actions for visual and audible indicators.
WARNING:
The KnightStar 330 does not have an audible alarm to indi­cate that the patient has stopped breathing.
16

Air Outlet Assembly

The KnightStar 330’s air outlet assembly consists of the air outlet and the optional outlet air filter.
Air Outlet. 22-mm conical port where the optional outlet air filter and tubing circuit are connected.
Optional Outlet Air Filter. This optional, single-patient filter removes contaminants and microbes as small as 0.2 microns from the outlet air. It is disposable, and must be replaced between patients. Be sure to inspect the filter regularly and replace it when noticeably dirty or discolored. Refer to “Replacing the Optional Air Outlet Filter” on
page 44. Frequency of replacement can vary, depending on usage and
environmental conditions. Contact your home care provider for replacement filters.
Optional outlet air filter
Figure 4. Optional Outlet Air Filter

Inlet Air Filter The Inlet Air Filter prevents large contaminants (dust and lint) in the

incoming air from entering the device. This filter has an efficiency of 90% or greater at 20 microns. It is reusable, and a spare filter is also provided. Refer to “Cleaning Instructions” on page 43.
17
Loading...
+ 47 hidden pages