KnightStar, Bi-Level, SoftFit, ADAM, and Breeze are trademarks of Nellcor Puritan Bennett Inc. For more information, contact
your Puritan Bennett representative.
The ventilator should be operated and serviced only by trained professionals. Puritan Bennett’s sole responsibility with respect
to the ventilator, and its use, is as stated in the limited warranty provided.
Nothing in this manual shall limit or restrict in any way Puritan Bennett’s right to revise or otherwise change or modify the
equipment (including its software) described herein, without notice. In the absence of an express, written agreement to the
contrary, Puritan Bennett has no obligation to furnish any such revisions, changes, or modifications to the owner or user of
the equipment (including its software) described herein.
6.2 Service information ..................................................................................................... 6-1
iiY-102942-00A Rev. G (03/04)KnightStar 330 Service Manual
SECTION
SECTION
General information1
This manual provides information needed to service the Puritan Bennett KnightStar 330® Bi-
®
Level
personnel with equivalent experience and training in servicing this type of equipment.
This section provides introductory information on the KnightStar 330 ventilator including:
1.1 Safety considerations
Please take the time to familiarize yourself with the following caveats as they cover safety
considerations, special handling requirements, and regulations that govern the use of the
KnightStar 330 ventilator.
ventilator and is intended for use by certified biomedical engineering technicians or
• General safety information
• Instructions on how to use the manual
• A description of the ventilator, its accessories, and its controls and indicators
• Detailed specifications and required tools and test equipment used for service and repair
• To ensure proper servicing and avoid the possibility of physical injury, only qualified
personnel (minimum requirement Certification for Biomedical Equipment Technician or
equivalent) should attempt to service or make authorized modifications to the ventilator.
The user of this product shall have sole responsibility for any ventilator malfunction
due to operation or maintenance performed by anyone not trained by Puritan Bennett
staff.
• To avoid an electrical shock hazard while servicing the ventilator, be sure to remove all
power to the ventilator by turning off the ventilator power switch and disconnecting the
power source.
• To avoid a fire hazard, keep matches, lighted cigarettes, and all other sources of ignition
(e.g., flammable anesthetics and/or heaters) away from the KnightStar 330 and oxygen
hoses.
Do not use oxygen hoses that are worn, frayed, or contaminated by combustible materials
such as grease or oils. (Textiles, oils, and other combustibles are easily ignited and burn
with great intensity in air enriched with oxygen.)
In case of fire or a burning smell, immediately disconnect the ventilator from the oxygen
supply, and the power source.
• When handling any part of the KnightStar 330, always follow your hospital infection
control guidelines for handling infectious material.
1
Puritan Bennett recognizes that cleaning, sterilization, sanitation, and disinfection
practices vary widely among healthcare institutions. It is not possible for Puritan Bennett
to specify or require specific practices that will meet all needs, or to be responsible for the
effectiveness of cleaning, sterilization, and other practices carried out in the patient care
setting. Specific cleaning instructions for the KnightStar 330 are given in section 5.2 on
page 1.
Puritan Bennett recommends that users of its products that require cleaning and
sterilization/disinfection consider the Center for Disease Control (CDC) publication:
KnightStar 330 Service ManualY-102942-00A Rev. G (03/04)
1-1
1General information
Guidelines for Prevention of Nosocomial Pneumonia available from the CDC Web site:
http://www.cdc.gov/publications.htm. Refer to the table on page 1-23 for a list of
approved disinfectants and cleaning agents.
• Patients on ventilation equipment should be appropriately monitored by competent
medical personnel and suitable monitoring devices.
• For a thorough understanding of ventilator operations, be sure to thoroughly read this
manual and the KnightStar 330 Clinician’s Manual before operating the device. These
manuals provide service, repair, and technical information concerning the operation and
performance of the ventilator.
• Before patient use, be sure to check the equipment for proper operation.
• Do not use sharp objects to make selections on the keypad.
• Check the ventilator periodically as outlined in this manual; do not use if defective.
Immediately replace parts that are broken, missing, obviously worn, distorted, or
contaminated.
• Federal law (U.S.) restricts this device to sale by or on the order of a physician.
1.2 Electromagnetic susceptibility
The KnightStar 330 equipment has been tested and found to comply with the limits for
medical devices to IEC 60601-1-2:2001 (or Medical Device Directive 93/42/EEC). This testing
shows that the device provides reasonable protection against harmful interference in a typical
medical installation. There is, however, no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to other devices or
is negatively affected by other devices, the user is encouraged to try to correct the interference
by one or more of the following measures:
• Re-orient or relocate the devices
• Increase the separation between the devices
• Connect the equipment to an outlet on a different circuit
• Contact the manufacturer or your local representative for help
1.3 Customer assistance
For further assistance, or for questions regarding the applicability of the information in this
manual, contact Puritan Bennett Technical Support at 1.800.255.6774 (within the USA) or
your local Puritan Bennett representative (outside the USA).
1.4How to use this manual
While this manual covers the ventilator configurations currently supported by Puritan
Bennett, some product upgrades may be available prior to a corresponding revision of this
manual. The current revision of this manual is available on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/PBProductManuals.html.
Puritan Bennett recommends that you become familiar with this manual and the KnightStar 330Clinician’s Manual before attempting to operate or service the ventilator. These manuals
provide service, repair, and technical information concerning the operation and performance
of the Puritan Bennett KnightStar 330 bi-level ventilator.
1-2 Y-102942-00A Rev. G (03/04)KnightStar 330 Service Manual
General information1
1.4.1 Definitions
This manual uses three special indicators to convey information of a specific nature.
They include:
Warning
Indicates a condition that can endanger the patient or the ventilator operator.
Caution
Indicates a condition that can damage the equipment or other property.
NOTE:
Indicates points of particular interest that make operation of the ventilator more efficient or
convenient.
1.4.2 Using the manual to troubleshoot the KnightStar 330
NOTES:
• At a minimum, the repair technician should be a Certified Biomedical Equipment
Technician (CBET) or possess equivalent experience and training before performing
any of the service instructions described in this manual
• Due to specific design interactions between the Main PCBA, blower, and pitot
tube, individual field replacement of any of these parts is not possible. If required,
please return the KnightStar 330 to a Puritan Bennett factory service center for
repair.
1Refer to Section 4of this manual to diagnose the problem. Using the troubleshooting
guides in this section, determine if the problem can be corrected without returning the
device to a factory service center.
2Use Table 1-14 to identify and order the required service parts.
3Follow instructions in Section 5 to disassemble, repair, and reassemble the KnightStar
330.
4Perform and document performance verification tests described in Section 3.
1.5 General product description
The Puritan Bennett KnightStar 330 is a continuous, bi-level ventilator that provides
noninvasive ventilation for the treatment of respiratory insufficiency and obstructive sleep
apnea that may occur in the home. It is also indicated for the treatment of respiratory failure
in institutional environments. It is intended to assist the ventilation of spontaneously
breathing patients who are over 30 kg (66 lb) in weight.
The KnightStar 330 is a microprocessor-controlled pressure generator capable of monitoring
the air flow and controlling the pressure delivered to the patient. The following are some of
its operating features:
• Provides three breathing modes, including Continuous Positive Airway Pressure (CPAP),
Inspiratory/Expiratory Positive Airway Pressure (I/E PAP), and Assist Control (A/C).
• Monitors pressure, tidal volume, respiratory rate, air leaks, peak flow, and I:E ratio.
KnightStar 330 Service ManualY-102942-00A Rev. G (03/04)1-3
1General information
• Provides precise respiratory support and patient comfort via adjustable inspiratory and
expiratory trigger sensitivity.
• Uses audible and visual indicators to alert users to power failure, system leaks, and other
conditions that could affect device performance.
• Allows a maximum pressure setting of 30 cmH
for a single-fault condition.
• Compensates for delivered pressure within specification for altitudes from 0 to 8,000 feet
(2438 meters) at 3 to 30 cmH
1.6 Configuration information
The KnightStar 330 is available in six configurations—North American and five international
versions. The major differences between ventilators are listed below:
North American: The North American version has Mode and Settings control panel keys and
some displayed information identified in English, and includes English labels, software, and
manuals. The power cord provided fits a standard 115 V AC outlet. Alarm volume may be set
from 0 (off) to 3 (maximum).
All other versions: All versions of the KnightStar 330 other than North American have
control panel keys and displayed information identified using symbols. Labels, software and
manuals are provided in various language configurations. A power cord with the appropriate
plug end is provided according to the configuration ordered. Alarm volume may be set from
0 (off) to 3 (maximum), except for Japanese. The Japanese KnightStar 330 alarm volume may
be set from 1 (minimum) to 3 (maximum) and cannot be set to “off.”
Figure 1-2 on page 1-12 shows the KnightStar 330’s control panel.
1.7 Accessories
The following accessories are either required or can be used with the KnightStar 330. See
Tab le 1-1 for ordering information.
O, with a pressure limitation of 40 cmH2O
2
O, and compensates for leaks up to 60 liters per minute.
2
Calibration Shell: The calibration shell is required for Performance Verification testing on
the KnightStar 330.
Patient circuit: Puritan Bennett recommends using the KnightStar 330 with 1.8 m (6 ft) or
2.4 m (8 ft) tubing and approved interfaces.
Oxygen Adapter: An optional O
or outlet air filter to enable the use of supplemental oxygen. Refer to the KnightStar 330Clinician’s Manual for more information.
Humidification device: The KnightStar 330 supports the use of an optional Fisher & Paykel
HC100 Humidifier, or equivalent.
External Battery: The KnightStar 330 may be powered by an external battery. A 32 ampere-
hour or a 7 ampere-hour battery are available. Separate cables are required to connect the
KnightStar 330 to an external battery or to a car or truck cigarette lighter outlet.
Other accessories: A carrying case and rolling stand are also available for the KnightStar 330.
adapter may be connected to the KnightStar 330 outlet port
2
1-4 Y-102942-00A Rev. G (03/04)KnightStar 330 Service Manual
General information1
Table 1-1 lists the ordering information for KnightStar 330 accessories.
Table 1-1: KnightStar 330 accessories
DescriptionPart number
Calibration ShellS-231702-00B
KnightStar 330 Patient tube, 6 ftY-261000-27
KnightStar 330 Patient tube, 8 ftY-261000-30
Oxygen AdapterY-616155-03B
Passover humidifier kit (includes 13 in. hose, base,
and reservoir)
Table 1-2 lists the technical specifications of the KnightStar 330 ventilator.
Physical characteristics
Y-102938-00
Y-C GVPD
Y-C GVPE
Y-CGVP7120
Table 1-2: Ventilator specifications
Weight1.21 kg (2.7 lb)
Dimensions9.52 cm x 20.95 cm x 14.27 cm (3.75 in x 8.25 in x 5.62 in)
Air Outlet Port
Connector
Device Airway
Volume
Patient Circuit
Volume
Environmental specifications
Temperature and
Humidity
Altitude Operating: 0 to 2438 m (0 to 8,000 ft)
KnightStar 330 Service ManualY-102942-00A Rev. G (03/04)1-5
Standard 22-mm conical male
65 mL
695 mL (1.8 m/6 ft)
927 mL (2.4 m/8 ft)
Operating: 5 to 40 ° C (41 to 104 ° F) at 15 to 95% relative humidity, noncondensing
Storage: -40 to 70 ° C (-40 to 158 ° F) at 10 to 95% relative humidity, noncondensing
1General information
Table 1-2: Ventilator specifications (continued)
Performance specifications
Working Pressure3 to 30 cmH2O (1 cmH2O = 0.098 kPa)
Pressure Limit40 cmH
Static Flow Pressure
Regulation
For flow rates up to 60 L/min:
CPAP ± 0.5 cmH2O for settings from 4 to 20 cmH2O
Bi-level ± 1.0 cmH
Noise≤ 30 dBA for IPAP/EPAP = 10 cmH
O
2
O for settings from 4 to 30 cmH2O
2
O (measured 1 m in front of device)
2
Electrical specifications
Rated Mains/AC
100 – 240 V AC nominal (85 – 264 V AC operating range)
Input Voltage
Rated Input
50 – 60 Hz
Frequency
Rated Input Power140 W
Displayed Patient Parameter Accuracy
Tidal Volume (Vt)± 20 mL +20% of reading (between 50 mL and 2000 mL)
Peak Flow (V)± 5 LPM +20% of reading (between 1 and 100 LPM)
Leak (L)± 5 LPM +20% of reading (between 1 and 100 LPM)
Respiratory Rate± 1 BPM (between 1 and 50 BPM)
I:E Ratio± 15% of reading (between 1:1 and 1:9.9)
Pressure± 1 cmH
O + 10% of reading (3 to 35 cmH2O)
2
Circuit Resistance
Inspiratory0.2 cmH2O at 30 L/min
0.9 cmH
Expiratory4.1 cmH
5.0 cmH
O at 60 L/min
2
O at 30 L/min
2
O at 60 L/min
2
External Battery Specifications
Rated Input Voltage12 V DC
Rated Input Current6.0 A
Rated Input Power140 W
Operating Time32 Ampere-hour: approximately 8 hours
7 Ampere-hour: approximately 3 hours
NOTE:Actual usage times depend upon patient’s prescription settings.
1-6Y-102942-00A Rev. G (03/04)KnightStar 330 Service Manual
General information1
Table 1-2: Ventilator specifications (continued)
Communication Port Specifications
Communication
capabilities
RS-232 (serial) port (Figure 1-1). A 9-pin female connector provides for RS-232 serial
communications, I/E PAP digital triggering, and calibration EEPROM programming. The RS-232
function operates at signal levels of at least 3 V into a standard load at a data rate of 9.6 kbps and
19.2 kbps. Pin 9 is used for calibration EEPROM programming at the manufacturing and service sites,
and during normal bi-level operation, supplies a 0 or 5-volt signal indicating a respective exhalation
or inhalation trigger. Applying 24 ± 1 V to pin 9 enables the write function of the calibration EEPROM
(U3). Any voltage less than 15 V will not enable the EEPROM write function. The calibration EEPROM
contains calibration constants including those for flow and pressure control.
1 2 3 4 5
8
6
9
7
Figure 1-1. KnightStar 330 RS-232 serial port pinout
NOTES:
• The RS-232 connector on devices manufactured prior to the year 2004 is rotated 180o from
the configuration shown above.
• If you would like to use the KnightStar 330 in a special application requiring serial
communications, contact Puritan Bennett Technical Support at 1.800.255.6774 or your
local Puritan Bennett representative.
PinSignal
1Not connected
2Transmit data (TxD) to PC
3Receive data (RxD) from PC
4Not connected
5Ground (GND)
6Not connected
7Not connected
8Not connected
9Dual use I/E PAP trigger-out and program-
enable input
KnightStar 330 Service ManualY-102942-00A Rev. G (03/04)1-7
1General information
1.9 Compliance and approvals
The KnightStar 330 was developed in accordance with pertinent FDA guidances and North
American and ISO international standards (Table 1-3). The manufacturing facility for this
product is ISO 13485 certified.
The ventilator’s IEC 60601-1/EN 60601-1 classification is Protection class II, Type BF,
externally powered, IPX1 drip-proof equipment.
Table 1-3: Compliance with standards
Standard TypeStandard NumberDescription
Quality systemEN ISO 13485:2000Quality Systems – Medical Devices –
Particular Requirements for the application
of EN ISO 9001:1994
FDA Ventilator GuidanceFDA Reviewer Guidance for Ventilators,
Draft (July 1995)
FDA Guidance
FDA Medical Electrical
Safety Guidance
FDA Reviewer Guidance for Pre-market
Notification Submissions, November 1993
draft, Anesthesiology and Respiratory
Devices Branch
Safety, USAUL 2601-1Medical Electrical Equipment, General
Requirements for Safety, (2
Safety, CanadaCAN/CSA C22.2 No.
601.1- M90
EN 60601-1 (compliant
with all applicable
collateral standards and
particular requirements)
EN 60601-1-1Collateral Standard: Safety Requirements
Safety, Europe &
International
EN 60601-1-2 (compliant
with all applicable tests)
JIST-1001-1Safety Requirements for Medical Electrical
CISPR 11EMC Disturbance Characteristics – Limits
EN 475Electrically generated alarm signals
Safety of Medical Electrical Equipment,
General Requirements (Supplement
1:1994), (A2:1998)
Medical Electrical Equipment, Part 1:
General Requirements for Safety, 1
Edition, 1988. (A1:1993), (A2:1995)
for Medical Electrical Systems, 2
2001
Medical Electrical Equipment, Collateral
Standard: EMC – Requirements and Test,
nd
Edition, 2001
2
Equipment in Japan
and Methods of Measurement, Industrial
Scientific and Medical (ISM) RF Equipment,
Edition 3.1, 1999-08
nd
Edition)
st
nd
Edition,
This device complies with the requirements of Medical Device Directive 93/42/EEC concerning medical
devices.
1-8 Y-102942-00A Rev. G (03/04)KnightStar 330 Service Manual
General information1
1.10 Breathing modes and settings
The KnightStar 330 offers three breathing modes: CPAP, I/E PAP, and A/C. Each breathing
mode enables a different set of system settings. Table 1-4 lists the modes and the adjustable
settings within each mode. Table 1-5 describes each setting, its adjustable range, and how the
setting can be accessed.
Table 1-4: Adjustable settings in each breathing mode
CPAPI/E PAPA/C
CPAPIPAPIPAP
Alarm volumeEPAPEPAP
Leak alarmInspiratory sensitivityRespiratory rate and backup
Internal ClockClock used by device24-hour clockRS-232
Pressure at which the device
starts ramp sequence
Patient interface purge hole
leak rate
0–30 minutes (in increments
of 5 minutes)
0–30 minutes (increments of
5 minutes)
3–20 cmH
1 cmH
1–6 (1 is the lowest leak
value, and 6 is the highest)
O (increments of
2
O)
2
Top panel, RS-232
Top panel, RS-232
Top panel, RS-232
Top panel, RS-232
NOTE: For information on using the RS-232 port, contact Puritan Bennett Technical
Support at 1.800.255.6774 or your local Puritan Bennett representative.
1-10 Y-102942-00A Rev. G (03/04)KnightStar 330 Service Manual
General information1
1.11 Initial factory settings
The KnightStar 330’s initial factory settings are listed in Table 1-6.
Table 1-6: Initial factory settings
SettingValue
MODEA/C
IPAP5 cm H
EPAP3 cm H
O (same setting for CPAP)
2
O
2
Respiratory rate10
I:E ratio1:2.0
I Sensitivity3
E Sensitivity3
Rise time3
Alarm volume3
Leak alarm100 L/min
Low pressure alarm4 cmH
High pressure alarm6 cmH
O
2
O
2
Delay time0
Ramp duration0
Ramp start pressure3 cm H
O
2
Mask leak2
Over-pressure alarm40 cmH
2
O (not adjustable)
KnightStar 330 Service ManualY-102942-00A Rev. G (03/04)1-11
1General information
1.12 Controls, indicators, and symbols
Refer to Figure 1-2 through Figure 1-4 and Table 1-7 through Table 1-10 for ventilator
controls, indicators, and symbols.
P
URITAN BENNETT
TM
Bi-Level âVentilator
1
2
3
4
5
6
1
2
3
8
7
9
10
12
14
13
Figure 1-2. KnightStar 330 Control Panel (International version shown above, North
American version shown below)
P
URITAN BENNETT
TM
Bi-Level âVentilator
Mode
Set
Delay
Ramp
Alarm
Silence
1-12 Y-102942-00A Rev. G (03/04)KnightStar 330 Service Manual
General information1
Table 1-7: KnightStar 330 control panel keys and indicators
Index
(Figure 1-2
on page 1-
12)
1Lockout Mode Indicator.
LabelingFunction
Indicates the KnightStar 330 control panel is locked. In Lockout
mode, the patient can change only the delay time, ramp duration,
and start pressure settings.
Indicates the KnightStar 330 control panel is unlocked. When
Lockout mode is inactive, the clinician may change any ventilator
settings.
NOTE:
To change the Lockout state, press and hold the Mode and Up Arrow
keys simultaneously for approximately 2 seconds.
2
1
2
3
Mode key.
Pressing the Mode key repeatedly allows you to scroll through CPAP, I/E, and
A/C modes. Press the Mode key to leave Settings mode when finished
adjusting settings.
The Mode key does not function when the KnightStar 330 control panel is
International
locked.
or
Mode
North American
3
Settings key.
Press the Settings key repeatedly to scroll through available parameters for
each breathing mode. When the KnightStar 330 control panel is locked, the
Settings key can only be used to change the delay time, ramp duration, and
ramp start pressure.
International version
or
Set
North American version
KnightStar 330 Service ManualY-102942-00A Rev. G (03/04)1-13
1General information
Table 1-7: KnightStar 330 control panel keys and indicators (continued)
Index
(Figure 1-2
on page 1-
12)
LabelingFunction
4
International version
or
Delay
Ramp
North American version
5
Down Arrow & Delay/Ramp key.
Use the Down Arrow key to decrease a selected setting value in Settings
mode. If not in Settings mode, use this key to start or stop the Delay/Ramp
function.
Up Arrow & Alarm Silence key.
Use the Up Arrow key to increase a selected setting value in Settings mode. If
not in Settings mode, use this key to mute an active alarm for one minute.
NOTE:
In A/C or I/E mode, when the main display screen is shown, pressing
this key displays V and I:E ratio if there are no active alarms.
International version
or
Alarm
Silence
North American version
6On/Off key. Turn the KnightStar 330 system on with a quick press and release
action. To turn the device off, press and hold the On/Off key for 3 seconds.
The KnightStar 330 retains in memory the prescription settings last entered.
7Green LEDWhen illuminated, indicates the presence of power, whether from Mains/AC
or external battery (Stand-by mode).
1-14 Y-102942-00A Rev. G (03/04)KnightStar 330 Service Manual
General information1
Table 1-7: KnightStar 330 control panel keys and indicators (continued)
Index
(Figure 1-2
on page 1-
12)
LabelingFunction
8Yellow LEDWhen steadily illuminated and accompanied by the symbol displayed in
the lower left corner of the LCD panel, indicates a LOW PRIORITY alarm
condition.
When steadily illuminated, with no symbol displayed on the LCD panel,
indicates a full compliance log (see Table 4-4 on page 4-11).
When flashing, indicates a MEDIUM PRIORITY alarm condition accompanied
by an audible alarm signal (3 beeps at intervals of approximately 25
seconds).
Refer to Section 4 for causes and corrections for alarm conditions.
9Red LEDWhen flashing, indicates a HIGH PRIORITY alarm condition accompanied by
an audible alarm signal (a series of 3 beeps, then 5 beeps, then 2 beeps at
intervals of approximately 6 seconds). Refer to Table 4 for causes and
corrections for alarm conditions.
10Liquid Crystal Display
(LCD)
The LCD provides an easy-to-read format for mode, settings, and patient
data. A backlight illuminates the display when the Mode, Settings, or Up
arrow key is pressed. The display will remain illuminated for approximately
60 seconds after the last key is pressed.
f
KnightStar 330 Service ManualY-102942-00A Rev. G (03/04)1-15
1General information
1
2
3
1
2
back.eps
3
Figure 1-3. KnightStar 330 rear view
Table 1-8: KnightStar 330 rear view
Index
(Figure 1-3
on page 1-
16)
1Mains/AC power electrical input connector. The KnightStar 330 operates on
LabelingFunction
100V to 240 V AC at 50 or 60 Hz.
1-16 Y-102942-00A Rev. G (03/04)KnightStar 330 Service Manual
General information1
Table 1-8: KnightStar 330 rear view (continued)
Index
(Figure 1-3
on page 1-
16)
2RS-232 (serial) port. The KnightStar 330 is capable of serial communications
3External battery connector. Used for connecting an optional external 12 V
LabelingFunction
with other devices such as a personal computer (PC). Contact your Puritan
Bennett representative for more information.
NOTE:
PCs used with the KnightStar 330 must meet regulatory standards for Safety
of Information Technology Equipment (i.e. UL 60950 or EN 60950). Available
from many sources, these standards specify requirements intended to reduce
risks of fire, electrical shock, and injury to the operator or service person who
comes into contact with the equipment.
DC battery, or for use with a 12 V automobile cigarette lighter adapter when
Mains/AC power is not available.
KnightStar 330 Service ManualY-102942-00A Rev. G (03/04)1-17
1General information
2
1
1
2
7
3
6
3
cover.eps
5
8-00197
Figure 1-4. KnightStar 330 front view
Table 1-9: KnightStar 330 front view
Index
(Figure 1-4 on
page 1-18)
1N/APatient pressure connector.
2Air outlet connector.
3
LabelingFunction
N/AAlarm speaker.
1-18 Y-102942-00A Rev. G (03/04)KnightStar 330 Service Manual
General information1
7
6
5
8
1
2
3
4
Figure 1-5. KnightStar 330 bottom panel (International version shown)
Table 1-10: Symbols found on the KnightStar 330 bottom panel
ItemLabelingDescription
1Class II equipment
(Per IEC 60601-1: Equipment in which protection against electric shock
does not rely on BASIC INSULATION only, but in which additional safety
precautions such as DOUBLE INSULATION or REINFORCED INSULATION
are provided, there being no provision for protective earthing or reliance
upon installation conditions.)
2Type BF equipment.
bottomlabel.eps
3Drip proof.
4Attention! Consult accompanying documents.
!
5Direct current (battery power).
KnightStar 330 Service ManualY-102942-00A Rev. G (03/04)1-19
1General information
Table 1-10: Symbols found on the KnightStar 330 bottom panel (continued)
ItemLabelingDescription
6This device complies with the requirements of Medical Device Directive
93/42/EEC concerning medical devices.
7Alternating current (Mains/AC power from wall outlet).
8Authorized to bear the CSA certification mark, signifying the product has
been evaluated to the applicable CSA standards for use in the US and
Canada.
UL2601-1
CAN/CSA C22.2 No. 601.1- M90
1-20 Y-102942-00A Rev. G (03/04)KnightStar 330 Service Manual
Loading...
+ 54 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.