Puritan-Bennett 840 Technical Reference manual

Page 1
840
Ventilator System
Part No. 4-075609-00 Rev. G October 2006
Technical Reference Manual
Operator’s and
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s and Technical Reference Manual
Page 2
Page 3
Copyright Information
Copyright 2003, 2005, 2006 Puritan-Bennett Corporation. All rights reserved. The
TM
840
Ventilator System is manufactured in accordance with Puritan Bennett
proprietary information, covered by one or more of the following U.S. Patents and foreign equivalents: 4,954,799; 5,161,525; 5,271,389; 5,301,921; 5,319,540; 5,339,807; 5,368,019; and 5,390,666. 840, 800 Series, DualView, SandBox, SmartAlert, Flow-by, and PTS 2000 are trademarks of Puritan-Bennett Corporation.
The information contained in this manual is the sole property of Puritan-Bennett Corporation and may not be duplicated without permission. This manual may be revised or replaced by Puritan-Bennett Corporation at any time and without notice. You should ensure that you have the most current applicable version of this manual; if in doubt, contact Puritan-Bennett Corporation or visit the Puritan Bennett product manual web page at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/
While the information set forth herein is believed to be accurate, it is not a substitute for the exercise of professional judgment.
The ventilator should be operated and serviced only by trained professionals. Puritan Bennett’s sole responsibility with respect to the ventilator, and its use, is as stated in the limited warranty provided.
Nothing in this manual shall limit or restrict in any way Puritan Bennett’s right to revise or otherwise change or modify the equipment (including its software) described herein, without notice. In the absence of an express, written agreement to the contrary, Puritan Bennett has no obligation to furnish any such revisions, changes, or modifications to the owner or user of the equipment (including its software) described herein.
Rx ONLY
i
Page 4
Applicability
The information in this manual applies to
ventilator versions
840
manufactured or updated after August 2005. Some of this information may not apply to earlier versions. Contact your Puritan Bennett representative if in doubt.
Definitions
This manual uses three special indicators to convey information of a specific nature. They include:
Warning
Indicates a condition that can endanger the patient or the ventilator operator.
Caution
Indicates a condition that can damage the equipment.
NOTE:
Indicates points of particular emphasis that make operation of the ventilator more efficient or convenient.
840 Ventilator System Operator’s and Technical Reference Manual 4-075609-00 Rev. G (10/06)
ii
Page 5
Warnings, cautions, and notes
Please take the time to familiarize yourself with the following safety considerations, special handling requirements, and regulations that govern the use of the
To ensure proper servicing and avoid the possibility of physical injury, only qualified personnel should attempt to service or make authorized modifications to the ventilator.
The user of this product shall have sole responsibility for any ventilator malfunction due to operation or maintenance performed by anyone not trained by Puritan Bennett.
To avoid an electrical shock hazard while servicing the ventilator, be sure to remove all power to the ventilator by disconnecting the power source and turning off all ventilator power switches.
To avoid a fire hazard, keep matches, lighted cigarettes, and all other sources of ignition (e.g., flammable anesthetics and/or heaters) away from the
Ventilator System and oxygen
840
hoses.
Do not use oxygen hoses that are worn, frayed, or contaminated by combustible materials such as grease or oils. Textiles, oils, and other combustibles are easily ignited and burn with great intensity in air enriched with oxygen.
Ventilator System.
840
In case of fire or a burning smell, immediately disconnect the ventilator from the oxygen supply, facility power, and backup power source.
When handling any part of the
Ventilator System, always
840
follow your hospital infection control guidelines for handling infectious material.
Puritan Bennett recognizes that cleaning, sterilization, sanitation, and disinfection practices vary widely among health care institutions. It is not possible for Puritan Bennett to specify or require specific practices that will meet all needs, or to be responsible for the effectiveness of cleaning, sterilization, and other practices carried out in the patient care setting.
iii
Page 6
Patients on life-support equipment should be appropriately monitored by competent medical personnel and suitable monitoring devices.
The
Ventilator System is not intended to be a
840
comprehensive monitoring device and does not activate alarms for all types of dangerous conditions for patients on life-support equipment.
For a thorough understanding of ventilator operations, be sure to thoroughly read this manual before attempting to use the system.
Before activating any part of the ventilator, be sure to check the equipment for proper operation and, if appropriate, run SST as described in this manual.
Do not use sharp objects to make selections on the graphic user interface (GUI) display or keyboard.
US federal law restricts this device to sale by or on the order of a physician.
Check the ventilator periodically as outlined in the
840 Ventilator System Service Manual
; do not use if defective. Immediately replace parts that are broken, missing, obviously worn, distorted, or contaminated.
An alternative source of ventilation should always be available when using the
Ventilator System.
840
Warranty
The and workmanship in accordance with Puritan Bennett Medical Equipment Warranty for 2 years from the time of sale. Keep a maintenance record to ensure the validity of the warranty.
840 Ventilator System Operator’s and Technical Reference Manual 4-075609-00 Rev. G (10/06)
iv
Ventilator System is warranted against defects in material
840
Page 7
Year of manufacture
The graphic user interface (GUI), breath delivery unit (BDU), backup power source (BPS), and compressor contain a specific year of manufacture applicable only for that assembly. The year of manufacture is indicated by the fifth and sixth digits of the serial number which is located at the back panel of the GUI, BDU, and BPS, and the side panel of the compressor.
Manufacturer
Puritan-Bennett Corporation 4280 Hacienda Drive Pleasanton, CA USA
Electromagnetic susceptibility
The IEC 60601-1-2:2004 (EMC Collateral Standard), including the E­field susceptibility requirements at a level of 10 volts per meter, at frequencies from 80 MHz to 2.5 GHz, and the ESD requirements of this standard.
Ventilator System complies with the requirements of
840
v
Page 8
However, even at this level of device immunity, certain transmitting devices (cellular phones, walkie-talkies, cordless phones, paging transmitters, etc.) emit radio frequencies that could interrupt ventilator operation if operated in a range too close to the ventilator. It is difficult to determine when the field strength of these devices becomes excessive.
Practitioners should be aware that radio frequency emissions are additive, and that the ventilator must be located a sufficient distance from transmitting devices to avoid interruption. Do not operate the ventilator in a magnetic resonance imaging (MRI) environment.
Warning
Accessory equipment connected to the power receptacle, analog, and digital interfaces must be certified according to IEC 60601-1. Furthermore, all configurations shall comply with the system standard IEC 60601-1-1. Any person who connects additional equipment to the power receptacle, signal input part, or signal output part of the 840 ventilator configures a medical system, and is therefore responsible for ensuring that the system complies with the requirements of the system standard IEC 60601-1-1. If in doubt, consult Puritan Bennett Technical Services at
1.800.255.6774 or your local representative.
This manual describes possible ventilator alarms and what to do if they occur. Consult with your institution’s biomedical engineering department in case of interrupted ventilator operation, and before relocating any life support equipment.
Customer assistance
For further assistance call Puritan Bennett Technical Services at
1.800.255.6774 or contact your local Puritan Bennett
representative.
840 Ventilator System Operator’s and Technical Reference Manual 4-075609-00 Rev. G (10/06)
vi
Page 9
Preface
This manual is divided into two parts: the operator’s manual and the technical reference. The operator’s manual describes how to operate the Puritan Bennett provides product specifications and accessory order numbers. The technical reference includes background information about how the ventilator functions, including details on its operating modes, self-tests, and other features. In the table of contents and index, the prefix OP- identifies page numbers in the operator’s manual, and the prefix TR- identifies page numbers in the technical reference.
Any references to the software options
Plus
®
(VV+)
which includes VC+ and VS breath types
Proportional Assist Ventilation
that are made in this manual assume that the option has been installed on the ventilator. If these options aren’t installed, then references to their functions do not apply.
While this manual covers the ventilator configurations currently supported by Puritan Bennett, it may not be all-inclusive and may not be applicable to your ventilator. Within the USA, contact Puritan Bennett at 1.800.255.6774 for questions about the applicability of the information.
Ventilator System. It also
840
®
, Volume Ven t ilation
Tube Compensation (TC)
®
(PAV+),
BiLevel
and
, NeoMode®,
vii
Page 10
This page is intentionally blank.
840 Ventilator System Operator’s and Technical Reference Manual 4-075609-00 Rev. G (10/06)
viii
Page 11
Contents
Operator’s Manual
1 Introduction OP 1-1
1.1 Technical description . . . . . . . . . . . . . . . . . . . . . . . . . OP 1-3
1.1.1 General background . . . . . . . . . . . . . . . . . . . . . . OP 1-3
1.1.2 Pressure and flow triggering . . . . . . . . . . . . . . . . OP 1-5
1.1.3 Breathing gas mixture. . . . . . . . . . . . . . . . . . . . . OP 1-5
1.1.4 Inspiratory pneumatics . . . . . . . . . . . . . . . . . . . . OP 1-6
1.1.5 Patient circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 1-6
1.1.6 AC mains and backup power system. . . . . . . . . . OP 1-7
1.1.7 Ventilator emergency states . . . . . . . . . . . . . . . . OP 1-8
1.2 Graphic user interface. . . . . . . . . . . . . . . . . . . . . . . . . OP 1-9
1.3 User interface controls and indicators . . . . . . . . . . . . . OP 1-10
1.3.1 Onscreen symbols and abbreviations . . . . . . . . . OP 1-15
1.4 Ventilator system labeling symbols . . . . . . . . . . . . . . . OP 1-22
2 How to set up the 840 ventilator OP 2-1
2.1 How to connect the electrical supply . . . . . . . . . . . . . OP 2-4
2.2 How to connect the air and oxygen supplies. . . . . . . . OP 2-8
2.3 How to connect the patient circuit components . . . . . OP 2-10
2.3.1 How to select and connect a patient circuit . . . . OP 2-11
2.3.2 How to install the expiratory filter and
collector vial . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 2-14
2.3.3 How to install the flex arm . . . . . . . . . . . . . . . . . OP 2-17
2.3.4 How to install the humidifier. . . . . . . . . . . . . . . . OP 2-18
2.3.5 How to use the ventilator cart . . . . . . . . . . . . . . OP 2-20
3 How to run Short Self Test (SST) OP 3-1
3.1 Introduction to SST. . . . . . . . . . . . . . . . . . . . . . . . . . . OP 3-1
3.2 When to run SST. . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 3-2
3.3 SST components and requirements. . . . . . . . . . . . . . . OP 3-4
3.4 SST Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 3-5
3.5 SST Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 3-14
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s and Technical Reference Manual
ix
Page 12
Contents
3.5.1 How to interpret individual SST test results. . . . . .OP 3-15
3.5.2 SST outcomes . . . . . . . . . . . . . . . . . . . . . . . . . . .OP 3-16
4 How to use the 840 ventilator OP 4-1
4.1 Structure of user interface . . . . . . . . . . . . . . . . . . . . . .OP 4-2
4.2 Patient setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .OP 4-3
4.2.1 How to ventilate with most recent control
parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . .OP 4-4
4.2.2 How to ventilate with new control parameters . . .OP 4-4
4.2.3 Patient data and current settings . . . . . . . . . . . . .OP 4-7
4.2.4 Ideal Body Weight (IBW) . . . . . . . . . . . . . . . . . . .OP 4-10
4.3 How to change the main ventilator
control parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . .OP 4-13
4.4 Mode, breath type, and other changes . . . . . . . . . . . .OP 4-14
4.5 How to select a constant timing variable during
respiratory rate changes . . . . . . . . . . . . . . . . . . . . . . . .OP 4-15
4.6 How to change apnea ventilation settings . . . . . . . . . .OP 4-17
4.7 How to set alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . .OP 4-18
4.8 How to change other settings . . . . . . . . . . . . . . . . . . .OP 4-19
4.9 Expiratory pause maneuvers. . . . . . . . . . . . . . . . . . . . .OP 4-21
4.10 Inspiratory pause maneuvers . . . . . . . . . . . . . . . . . . .OP 4-22
4.11 How to interpret inspiratory pause maneuver results
for static compliance and resistance. . . . . . . . . . . . . .OP 4-24
4.12 How to use NIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . .OP 4-25
4.12.1 NIV intended use . . . . . . . . . . . . . . . . . . . . . . . .OP 4-25
4.12.2 NIV breathing interfaces. . . . . . . . . . . . . . . . . . .OP 4-25
4.12.3 NIV setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .OP 4-26
4.12.4 High spontaneous inspiratory time limit setting .OP 4-30
4.12.5 Apnea setup. . . . . . . . . . . . . . . . . . . . . . . . . . . .OP 4-30
4.12.6 Alarm setup . . . . . . . . . . . . . . . . . . . . . . . . . . . .OP 4-31
4.12.7 Changing patient from INVASIVE to NIV
Vent Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . .OP 4-33
4.12.8 Changing patient from NIV to INVASIVE
Vent Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . .OP 4-34
4.12.9 NIV patient data. . . . . . . . . . . . . . . . . . . . . . . . .OP 4-34
840 Ventilator System Operator’s and Technical Reference Manual 4-075609-00 Rev. G (10/06)
x
Page 13
Contents
5 How to handle alarms OP 5-1
5.1 Ventilator alarm classifications. . . . . . . . . . . . . . . . . . . OP 5-1
5.2 Alarm silence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 5-2
5.3 Alarm reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 5-4
5.4 Alarm log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 5-5
5.5 Alarm volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 5-6
5.6 Alarm messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 5-7
6 How to view graphics OP 6-1
6.1 Graphics display function . . . . . . . . . . . . . . . . . . . . . . OP 6-1
6.2 How to set up a graphics display . . . . . . . . . . . . . . . . OP 6-2
6.3 Graphics display details and calculations. . . . . . . . . . . OP 6-3
6.4 How to adjust displayed graphics . . . . . . . . . . . . . . . . OP 6-4
6.5 The graphics display FREEZE function . . . . . . . . . . . . . OP 6-5
6.6 How to print patient data graphics . . . . . . . . . . . . . . . OP 6-6
6.7 Automatic display of graphics . . . . . . . . . . . . . . . . . . . OP 6-6
6.8 When graphics are not accessible . . . . . . . . . . . . . . . . OP 6-7
7 Preventive maintenance OP 7-1
7.1 How to dispose of used parts . . . . . . . . . . . . . . . . . . . OP 7-1
7.2 How to clean, disinfect and sterilize parts . . . . . . . . . . OP 7-2
7.2.1 How to clean components . . . . . . . . . . . . . . . . . OP 7-6
7.3 Disinfection and sterilization. . . . . . . . . . . . . . . . . . . . OP 7-6
7.4 Preventive maintenance procedures for the operator . OP 7-8
7.4.1 Total operational hours. . . . . . . . . . . . . . . . . . . . OP 7-9
7.4.2 Inspiratory and expiratory bacteria filters . . . . . . OP 7-12
7.4.3 Daily or as required: collector vial and drain bag. OP 7-14
7.4.3.1 How to remove the collector vial . . . . . . . . OP 7-14
7.4.3.2 How to remove the drain bag . . . . . . . . . . . OP 7-15
7.4.4 Daily or as required: in-line water traps. . . . . . . . OP 7-16
7.4.5 Every 250 hours: compressor inlet filter. . . . . . . . OP 7-16
7.4.6 Every year: ventilator inspection . . . . . . . . . . . . . OP 7-17
7.4.7 Every 2 years or as necessary: oxygen sensor. . . . OP 7-17
7.4.7.1 Oxygen sensor replacement procedure. . . . OP 7-18
xi
Page 14
Contents
7.5 Additional preventive maintenance procedures . . . . . .OP 7-24
7.6 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .OP 7-26
7.7 Repacking and shipping. . . . . . . . . . . . . . . . . . . . . . . .OP 7-26
A Specifications OP A-1
A.1 Physical characteristics . . . . . . . . . . . . . . . . . . . . . . . . .OP A-2
A.2 Environmental requirements . . . . . . . . . . . . . . . . . . . .OP A-4
A.3 Pneumatic specifications . . . . . . . . . . . . . . . . . . . . . . .OP A-4
A.4 Electrical specifications. . . . . . . . . . . . . . . . . . . . . . . . .OP A-6
A.5 Compliance and approvals. . . . . . . . . . . . . . . . . . . . . .OP A-9
A.5.1 Manufacturer’s Declaration . . . . . . . . . . . . . . . . .OP A-10
A.6 Technical specifications . . . . . . . . . . . . . . . . . . . . . . . .OP A-18
A.7 Ranges, resolutions, and accuracies . . . . . . . . . . . . . . .OP A-24
A.7.1 Recommended limits . . . . . . . . . . . . . . . . . . . . . .OP A-24
A.7.2 Software options . . . . . . . . . . . . . . . . . . . . . . . . .OP A-25
B Part numbers OP B-1
C Pneumatic schematic OP C-1
D Alarm and oxygen sensor calibration testing OP D-1
D.1 Alarm test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .OP D-1
D.2 Oxygen sensor calibration test. . . . . . . . . . . . . . . . . . .OP D-7
E Remote alarm and RS-232 ports OP E-1
E.1 Remote alarm port . . . . . . . . . . . . . . . . . . . . . . . . . . . .OP E-2
E.2 RS-232 port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .OP E-3
E.3 How to configure the RS-232 ports. . . . . . . . . . . . . . . .OP E-4
E.4 Printers and cables . . . . . . . . . . . . . . . . . . . . . . . . . . . .OP E-5
E.5 RS-232 port commands . . . . . . . . . . . . . . . . . . . . . . . .OP E-6
840 Ventilator System Operator’s and Technical Reference Manual 4-075609-00 Rev. G (10/06)
xii
Page 15
Contents
Technical Reference
1 Introduction to breath delivery TR 1-1
2 Detecting and initiating inspiration TR 2-1
2.1 Internally triggered inspiration . . . . . . . . . . . . . . . . . . TR 2-2
2.1.1 Pressure sensitivity . . . . . . . . . . . . . . . . . . . . . . . TR 2-2
2.1.2 Flow sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . TR 2-3
2.1.3 Time-cycled inspiration. . . . . . . . . . . . . . . . . . . . TR 2-5
2.2 Operator-triggered inspiration . . . . . . . . . . . . . . . . . . TR 2-6
3 Detecting and initiating exhalation TR 3-1
3.1 Internally initiated exhalation . . . . . . . . . . . . . . . . . . . TR 3-1
3.1.1 Time-cycled exhalation . . . . . . . . . . . . . . . . . . . . TR 3-1
3.1.2 End-inspiratory flow method. . . . . . . . . . . . . . . . TR 3-2
3.1.3 Airway pressure method . . . . . . . . . . . . . . . . . . . TR 3-2
3.2 Backup limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 3-3
3.2.1 Time limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 3-4
3.2.2 High circuit pressure limit . . . . . . . . . . . . . . . . . . TR 3-4
3.2.3 High ventilator pressure limit . . . . . . . . . . . . . . . TR 3-4
4 Mandatory breath delivery TR 4-1
4.1 Comparison of pressure- and volume-based
mandatory breaths . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 4-1
4.2 Compliance compensation for volume-based
mandatory breaths . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 4-3
4.3 BTPS compensation for volume-based
mandatory breaths . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 4-5
4.4 Manual inspiration . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 4-5
5 Spontaneous breath delivery TR 5-1
xiii
Page 16
Contents
6 Assist/control (A/C) mode TR 6-1
6.1 Breath delivery in A/C . . . . . . . . . . . . . . . . . . . . . . . . .TR 6-1
6.2 Rate change during A/C. . . . . . . . . . . . . . . . . . . . . . . .TR 6-3
6.3 Changing to A/C mode . . . . . . . . . . . . . . . . . . . . . . . .TR 6-3
7 Synchronous intermittent mandatory
ventilation (SIMV) TR 7-1
7.1 Breath delivery in SIMV . . . . . . . . . . . . . . . . . . . . . . . .TR 7-2
7.2 Apnea ventilation in SIMV . . . . . . . . . . . . . . . . . . . . . .TR 7-4
7.3 Changing to SIMV mode . . . . . . . . . . . . . . . . . . . . . . .TR 7-5
7.4 Rate change during SIMV. . . . . . . . . . . . . . . . . . . . . . .TR 7-7
8 Spontaneous (SPONT) mode TR 8-1
8.1 Breath delivery in SPONT . . . . . . . . . . . . . . . . . . . . . . .TR 8-1
8.2 Changing to SPONT mode. . . . . . . . . . . . . . . . . . . . . .TR 8-1
9 Apnea ventilation TR 9-1
9.1 Apnea detection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TR 9-1
9.2 Transition to apnea ventilation . . . . . . . . . . . . . . . . . . .TR 9-3
9.3 Key entries during apnea ventilation. . . . . . . . . . . . . . .TR 9-3
9.4 Resetting apnea ventilation . . . . . . . . . . . . . . . . . . . . .TR 9-3
9.4.1 Resetting to A/C. . . . . . . . . . . . . . . . . . . . . . . . . .TR 9-4
9.4.2 Resetting to SIMV. . . . . . . . . . . . . . . . . . . . . . . . .TR 9-4
9.4.3 Resetting to SPONT . . . . . . . . . . . . . . . . . . . . . . .TR 9-4
9.5 Phasing in new apnea intervals. . . . . . . . . . . . . . . . . . .TR 9-5
10 Detecting occlusion and disconnect TR 10-1
10.1 Occlusion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TR 10-1
10.2 Disconnect. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TR 10-3
10.3 Occlusions and disconnect annunciation . . . . . . . . . .TR 10-5
840 Ventilator System Operator’s and Technical Reference Manual 4-075609-00 Rev. G (10/06)
xiv
Page 17
Contents
11 Phasing in setting changes TR 11-1
12 Ventilator settings TR 12-1
12.1 Apnea ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 12-1
12.2 Circuit type and Ideal Body Weight (IBW) . . . . . . . . TR 12-2
12.3 Disconnect sensitivity (D
12.4 Expiratory sensitivity (E
12.5 Expiratory time (T
) . . . . . . . . . . . . . . . . . . . . . . . . . TR 12-4
E
12.6 Flow pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 12-4
12.7 Flow sensitivity (V
SENS
12.8 High spontaneous inspiratory time limit (2T
12.9 Humidification type . . . . . . . . . . . . . . . . . . . . . . . . . TR 12-6
12.10 I:E ratio. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 12-7
12.11 Ideal body weight (IBW) . . . . . . . . . . . . . . . . . . . . . TR 12-7
12.12 Inspiratory pressure (P
12.13 Inspiratory time (T
) . . . . . . . . . . . . . . . . . . . . . . . . TR 12-8
I
12.14 Mode and mandatory breath type . . . . . . . . . . . . . TR 12-8
12.15 O
%. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 12-11
2
12.16 Peak inspiratory flow (V
12.17 PEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 12-12
12.17.1 PEEP restoration . . . . . . . . . . . . . . . . . . . . . . . TR 12-13
12.18 Plateau time (T
) . . . . . . . . . . . . . . . . . . . . . . . . . . TR 12-13
PL
12.19 Pressure sensitivity (P
12.20 Pressure support (P
SUPP
12.21 Respiratory rate (f) . . . . . . . . . . . . . . . . . . . . . . . . . TR 12-15
12.22 Rise time %. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 12-15
12.23 Safety ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . TR 12-16
12.24 Spontaneous breath type . . . . . . . . . . . . . . . . . . . . TR 12-17
12.25 Tidal volume (V
) . . . . . . . . . . . . . . . . . . . . . . . . . . TR 12-18
T
12.26 Vent type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 12-18
) . . . . . . . . . . . . . . . . . . TR 12-3
SENS
) . . . . . . . . . . . . . . . . . . . TR 12-4
SENS
) . . . . . . . . . . . . . . . . . . . . . . . TR 12-5
) . TR 12-6
I SPONT
) . . . . . . . . . . . . . . . . . . . . . TR 12-7
I
) . . . . . . . . . . . . . . . . . TR 12-12
MAX
) . . . . . . . . . . . . . . . . . . . TR 12-14
SENS
) . . . . . . . . . . . . . . . . . . . . . TR 12-14
xv
Page 18
Contents
13 Alarms TR 13-1
13.1 Alarm handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TR 13-1
13.1.1 Alarm messages . . . . . . . . . . . . . . . . . . . . . . . . .TR 13-3
13.1.2 Alarm summary . . . . . . . . . . . . . . . . . . . . . . . . .TR 13-5
13.2 AC POWER LOSS alarm . . . . . . . . . . . . . . . . . . . . . . .TR 13-23
13.3 APNEA alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TR 13-23
13.4 CIRCUIT DISCONNECT alarm. . . . . . . . . . . . . . . . . . .TR 13-24
13.5 DEVICE ALERT alarm. . . . . . . . . . . . . . . . . . . . . . . . . .TR 13-24
13.6 High circuit pressure (
13.7 High delivered O
13.8 High exhaled minute volume (
13.9 High exhaled tidal volume (
13.10 High inspired tidal volume alarm (
V
TI SPONT
). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TR 13-27
13.11 High respiratory rate (
13.12 INSPIRATION TOO LONG alarm . . . . . . . . . . . . . . .TR 13-28
13.13 Low circuit pressure alarm (
13.14 Low delivered O
13.15 Low exhaled mandatory tidal volume (
V
TE MAND
) alarm . . . . . . . . . . . . . . . . . . . . . . . . . .TR 13-30
13.16 Low exhaled spontaneous tidal volume (
V
TE SPONT
) alarm . . . . . . . . . . . . . . . . . . . . . . . . . .TR 13-30
13.17 Low exhaled total minute volume (
13.18 PROCEDURE ERROR alarm . . . . . . . . . . . . . . . . . . . .TR 13-31
P
) alarm . . . . . . . . . . . . . .TR 13-25
PEAK
% (↑O2%) alarm . . . . . . . . . . . . . .TR 13-26
2
V
VTE) alarm . . . . . . . . . . .TR 13-27
f
) alarm. . . . . . . . . . . . . .TR 13-28
TOT
P
PEAK
% (↓O2%) alarm . . . . . . . . . . . . . .TR 13-29
2
) alarm . . . . . .TR 13-26
ETOT
V
TI
, V
TI MAND
,
). . . . . . . . . . . . . .TR 13-29
) alarm. .TR 13-31
V
E TOT
14 Patient data TR 14-1
14.1 Delivered O2%. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TR 14-1
14.2 End expiratory pressure (PEEP) . . . . . . . . . . . . . . . . . .TR 14-2
14.3 End inspiratory pressure (P
14.4 Exhaled minute volume (V
14.5 Exhaled tidal volume (V
TE
14.6 I:E ratio (I:E) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TR 14-4
14.7 Intrinsic (auto) PEEP (PEEP
840 Ventilator System Operator’s and Technical Reference Manual 4-075609-00 Rev. G (10/06)
xvi
). . . . . . . . . . . . . . . . .TR 14-2
I END
) . . . . . . . . . . . . . . . .TR 14-3
E TOT
) . . . . . . . . . . . . . . . . . . . . .TR 14-4
) and total PEEP (PEEP
I
).TR 14-5
TOT
Page 19
Contents
14.8 Mean circuit pressure (P
14.9 Peak circuit pressure (P
14.10 Plateau pressure (P
) . . . . . . . . . . . . . . . . . . . . . . . TR 14-6
PL
14.11 Spontaneous minute volume (V
14.12 Static compliance and resistance (C
14.13 Total respiratory rate (f
) . . . . . . . . . . . . . . . . . TR 14-5
MEAN
) . . . . . . . . . . . . . . . . . . . TR 14-5
PEAK
E SPONT
). . . . . . . . . . . . . . . . . . . TR 14-13
TOT
) . . . . . . . . . TR 14-6
STAT
and R
) . TR 14-7
STAT
15 Safety net TR 15-1
15.1 Patient problems. . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 15-1
15.2 System faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 15-2
15.3 Ongoing background checks . . . . . . . . . . . . . . . . . . TR 15-3
15.4 Hardware monitoring circuitry . . . . . . . . . . . . . . . . . TR 15-4
15.5 Power on self test (POST) . . . . . . . . . . . . . . . . . . . . . TR 15-5
15.6 Short self test (SST). . . . . . . . . . . . . . . . . . . . . . . . . . TR 15-5
15.7 Extended self test (EST). . . . . . . . . . . . . . . . . . . . . . . TR 15-5
15.8 Oxygen sensor calibration. . . . . . . . . . . . . . . . . . . . . TR 15-6
15.9 Exhalation valve calibration. . . . . . . . . . . . . . . . . . . . TR 15-6
15.10 Ventilator inoperative test. . . . . . . . . . . . . . . . . . . . TR 15-6
15.11 Flow sensor offset calibration . . . . . . . . . . . . . . . . . TR 15-6
15.12 Atmospheric pressure transducer calibration. . . . . . TR 15-6
16 Power on self test (POST) TR 16-1
16.1 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 16-1
16.2 POST characteristics . . . . . . . . . . . . . . . . . . . . . . . . . TR 16-2
16.3 POST following power interruptions . . . . . . . . . . . . . TR 16-3
16.4 POST fault handling . . . . . . . . . . . . . . . . . . . . . . . . . TR 16-4
16.5 POST system interface . . . . . . . . . . . . . . . . . . . . . . . TR 16-4
16.6 POST user interface. . . . . . . . . . . . . . . . . . . . . . . . . . TR 16-5
17 Short self test (SST) TR 17-1
xvii
Page 20
Contents
18 Extended self test (EST) TR 18-1
18.1 EST results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TR 18-2
18.2 EST failure handling . . . . . . . . . . . . . . . . . . . . . . . . . .TR 18-3
18.3 EST safety considerations . . . . . . . . . . . . . . . . . . . . . .TR 18-3
19 RS-232 commands TR 19-1
19.1 RSET command . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TR 19-1
19.2 SNDA command . . . . . . . . . . . . . . . . . . . . . . . . . . . .TR 19-1
Glossary
Index
840 Ventilator System Operator’s and Technical Reference Manual 4-075609-00 Rev. G (10/06)
xviii
Page 21
Figures
Operator’s Manual
Figure 1-1. 840 Ventilator System block diagram . . . . . . . . . . . . . . . OP 1-4
Figure 1-2. 840 Ventilator System graphic user interface (GUI) . . . . . OP 1-9
Figure 2-1. How to lift the ventilator components . . . . . . . . . . . . . . . OP 2-2
Figure 2-2. How to connect the ventilator power cord . . . . . . . . . . . OP 2-5
Figure 2-3. Ventilator power switch, AC indicator, and AC panel. . . . OP 2-6
Figure 2-4. Power cord storage on the cart . . . . . . . . . . . . . . . . . . . . OP 2-7
Figure 2-5. How to connect the air and oxygen supplies . . . . . . . . . . OP 2-9
Figure 2-6. How to connect the patient circuit . . . . . . . . . . . . . . . . . OP 2-13
Figure 2-7. How to install the expiratory filter and collector vial . . . . OP 2-15
Figure 2-8. How to use the collector vial with or
without the drain bag . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 2-16
Figure 2-9. How to install the flex arm. . . . . . . . . . . . . . . . . . . . . . . . OP 2-17
Figure 2-10. How to install the humidifier
(Fisher & Paykel version shown) . . . . . . . . . . . . . . . . . . . OP 2-19
Figure 2-11. How to lock and unlock the cart’s front wheels . . . . . . . . OP 2-20
Figure 3-1. Test button location . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 3-6
Figure 4-1. Touch screen user interface . . . . . . . . . . . . . . . . . . . . . . . OP 4-2
Figure 4-2. Ventilator Startup screen . . . . . . . . . . . . . . . . . . . . . . . . . OP 4-3
Figure 4-3. Touch screen appearance during normal ventilation
(shown with alarm silence and
100% O Figure 4-4.
(or TH) selected as the constant during rate change. . . OP 4-16
T
I
Figure 4-5. Alarm setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 4-19
Figure 4-6. New patient setup screen — NIV. . . . . . . . . . . . . . . . . . . OP 4-27
Figure 4-7. NIV ventilator settings screen . . . . . . . . . . . . . . . . . . . . . OP 4-29
Figure 4-8. New patient default alarm settings . . . . . . . . . . . . . . . . . OP 4-31
Figure 4-9. More patient data screen — NIV . . . . . . . . . . . . . . . . . . . OP 4-35
Figure 5-1. Alarm indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 5-2
Figure 5-2. Alarm Silence in Progress indicator (lower screen) . . . . . . OP 5-3
Figure 5-3. Alarm log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 5-5
Figure 5-4. Alarm message format. . . . . . . . . . . . . . . . . . . . . . . . . . . OP 5-8
Figure 6-1. Pressure-volume loop . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 6-2
Figure 7-1. How to empty the collector vial and seal the drain bag . . OP 7-15
Figure 7-2. 806 compressor with inlet filter. . . . . . . . . . . . . . . . . . . . OP 7-17
/CAL in progress) . . . . . . . . . . . . . . . . . . . . . . . OP 4-9
2
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s and Technical Reference Manual
xix
Page 22
Figures
Figure 7-3. Dislodge the O2 sensor access cover . . . . . . . . . . . . . . . . OP 7-20
Figure 7-4. Open O Figure 7-5. Locate O
Figure A-1. Recommended patient circuit configurations . . . . . . . . . OP A-23
Figure B-1. Ventilator accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . OP B-2
Figure C-1. Pneumatic schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . OP C-1
Figure E-1. Remote alarm and RS-232 ports . . . . . . . . . . . . . . . . . . . OP E-1
Figure E-2. Remote alarm port pinout (view from back of GUI). . . . . OP E-2
Figure E-3. RS-232 serial port pinout . . . . . . . . . . . . . . . . . . . . . . . . OP E-3
Technical Reference
Figure 2-1. Declaring inspiration using pressure sensitivity . . . . . . . . TR 2-3
Figure 2-2. Declaring inspiration using flow sensitivity . . . . . . . . . . . TR 2-4
Figure 2-3. Time-cycled inspiration. . . . . . . . . . . . . . . . . . . . . . . . . . TR 2-6
Figure 3-1. Initiating exhalation using the
end-inspiratory flow method. . . . . . . . . . . . . . . . . . . . . . TR 3-2
Figure 3-2. Initiating exhalation using the airway pressure method. . TR 3-3
Figure 6-1. A/C mode, no patient effort detected. . . . . . . . . . . . . . . TR 6-2
Figure 6-2. A/C mode, patient effort detected . . . . . . . . . . . . . . . . . TR 6-2
Figure 6-3. A/C mode, VIM and PIM breaths . . . . . . . . . . . . . . . . . . TR 6-2
Figure 7-1. SIMV breath cycle
(mandatory and spontaneous intervals) . . . . . . . . . . . . . TR 7-1
Figure 7-2. SIMV breath cycle, PIM delivered within
mandatory interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 7-2
Figure 7-3. SIMV breath cycle, PIM not delivered within
mandatory interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 7-2
Figure 7-4. Apnea ventilation in SIMV. . . . . . . . . . . . . . . . . . . . . . . . TR 7-4
Figure 9-1. Apnea interval equals breath period . . . . . . . . . . . . . . . . TR 9-2
Figure 9-2. Apnea interval greater than breath period. . . . . . . . . . . . TR 9-2
Figure 9-3. Apnea interval less than breath period . . . . . . . . . . . . . . TR 9-2
Figure 12-1. 840 ventilator modes and breath types. . . . . . . . . . . . . . TR 12-10
Figure 13-1. Alarm message format (upper GUI screen) . . . . . . . . . . . TR 13-3
sensor access port . . . . . . . . . . . . . . . . . . . . . . OP 7-21
2
sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 7-22
2
840 Ventilator System Operator’s and Technical Reference Manual 4-075609-00 Rev. G (10/06)
xx
Page 23
Tables
Operator’s Manual
Table 1-1. 840 Ventilator System GUI controls and indicators . . . . . . OP 1-10
Table 1-2. BDU indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 1-15
Table 1-3. 840 Ventilator System symbols and abbreviations. . . . . . . OP 1-16
Table 2-1. Patient circuit and IBW values. . . . . . . . . . . . . . . . . . . . . . OP 2-12
Table 3-1. SST test sequence. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 3-9
Table 3-2. Individual SST test results . . . . . . . . . . . . . . . . . . . . . . . . . OP 3-15
Table 3-3. Overall SST outcomes. . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 3-16
Table 4-1. IBW based on patient height . . . . . . . . . . . . . . . . . . . . . . OP 4-11
Table 4-2. Patient circuit and IBW values. . . . . . . . . . . . . . . . . . . . . . OP 4-13
Table 4-3. Monitored ventilator control parameters . . . . . . . . . . . . . OP 4-14
Table 4-5. Automatic settings changes—INVASIVE to NIV . . . . . . . . . OP 4-33
Table 4-6. Automatic settings changes—NIV to INVASIVE . . . . . . . . . OP 4-34
Table 5-1. Alarm messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 5-9
Table 7-1. Procedures to clean, disinfect, and sterilize parts. . . . . . . . OP 7-3
Table 7-2. Disinfection and sterilization procedures. . . . . . . . . . . . . . OP 7-7
Table 7-3. Operator preventive maintenance procedures
and frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP 7-10
Table 7-4. Service preventive maintenance procedures and intervals. OP 7-25
Table A-1. Physical characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . OP A-2
Table A-2. Environmental requirements . . . . . . . . . . . . . . . . . . . . . . . OP A-4
Table A-3. Pneumatic specifications . . . . . . . . . . . . . . . . . . . . . . . . . . OP A-4
Table A-4. Electrical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . OP A-6
Table A-5. Compliance and approvals . . . . . . . . . . . . . . . . . . . . . . . . OP A-9
Table A-6. Electromagnetic Emissions . . . . . . . . . . . . . . . . . . . . . . . . OP A-11
Table A-7. Electromagnetic Immunity . . . . . . . . . . . . . . . . . . . . . . . . OP A-12
Table A-8. Electromagnetic Immunity – conducted and radiated RF . OP A-14 Table A-9. Recommended separation distances between portable
and mobile RF communications equipment
and the 840 Ventilator System . . . . . . . . . . . . . . . . . . . . . OP A-16
Table A-10. Compliant cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP A-17
Table A-11. Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . OP A-18
Table A-12. Ventilator settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP A-25
Table A-13. Alarm settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP A-42
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s and Technical Reference Manual
xxi
Page 24
Tables
Table A-14. Patient data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OP A-47
Table A-15. Other Screens — displayed data. . . . . . . . . . . . . . . . . . . . OP A-54
Table B-1. Ventilator parts and accessories . . . . . . . . . . . . . . . . . . . . OP B-3
Technical Reference
Table 4-1. Comparison of pressure- and volume-based
mandatory breaths TR 4-2
Table 4-2. Compliance volume factors . . . . . . . . . . . . . . . . . . . . . . . TR 4-5
Table 5-1. Spontaneous breath delivery characteristics . . . . . . . . . . TR 5-1
Table 12-1. 840 ventilator modes and breath types . . . . . . . . . . . . . . TR 12-9
Table 13-1. Alarm urgency levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 13-2
Table 13-2. Alarm summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 13-5
Table 13-3. Applicability of high inspired tidal volume alarm symbols. TR 13-27
Table 14-1. Inspiratory pause maneuver displays . . . . . . . . . . . . . . . . TR 14-9
Table 19-1. MISCA response. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TR 19-3
840 Ventilator System Operator’s and Technical Reference Manual 4-075609-00 Rev. G (10/06)
xxii
Page 25
C
HAPTER
1

Introduction

The intended use of the Puritan Bennett for acute and subacute care of infant, pediatric, and adult patients. Software options, available from Puritan Bennett, provide additional ventilation functions.
The management, offers selectable modes of breath delivery, and assists the practitioner in the selection of the most appropriate ventilator control parameters for the patient. The user interface is intuitive and easy to operate for those with prior knowledge of ventilator operation.
The user interface includes monitored patient data, for easy assessment of the patient’s condition. The touch screens also display the current ventilator control parameters.
The preview the selected ventilator control parameters prior to active ventilation of the patient.
The specific information about the cause, and prompts the user with actions to resolve the reported condition(s).
Ventilator System facilitates work of breathing
840
SandBox
SmartAlert
1
Ventilator System is
840
DualView
™ area on the touch screen allows the practitioner to
™ system intercepts alarms, or events, provides
™ touch screens that display
The breath delivery unit (BDU) comprises the pneumatics and the patient circuit.
The ventilator uses two independent Central Processing Units (CPUs):
Breath delivery unit (BDU) CPU
Graphic user interface (GUI) CPU
The BDU CPU uses the ventilator control parameters, selected by the practitioner, to deliver breaths to the patient. The BDU CPU also runs continuous and extensive operational background checks to ensure proper operation of the ventilator.
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 1-1
Page 26
OP 1
Introduction
The GUI CPU monitors the ventilator and the ventilator/patient interaction. The GUI CPU also monitors the operation of the BDU CPU and prevents simultaneous failure of control and monitor functions when a single fault is reported.
The breaths with a preset level of positive end-expiratory pressure (PEEP), trigger sensitivity, and oxygen concentration. A mandatory breath can either be pressure- or volume-controlled, but it is always pressure-controlled in the optional spontaneous breath allows patient inspiratory flows of up to 200 L/min, with or without pressure support.
The optional 806 Compressor unit provides compressed air to the BDU, and can be used in place of wall or bottled air. The compressor unit is powered through and communicates with the BDU.
The 802 Backup Power Source (BPS) provides DC power to the BDU and GUI in the event that AC power is lost. A new, fully charged BPS runs the ventilator (without a compressor or a humidifier) for at least 30 minutes, which allows transport of the patient and the ventilator within the healthcare facility.
This manual tells you how to operate and perform simple maintenance for the this manual and accompanying labels before attempting to operate or maintain the ventilator.
To ensure optimum performance of the Puritan Bennett strongly recommends that certified biomedical engineering technicians, or other personnel with equivalent experience and training in the service of this type of equipment, perform periodic maintenance on the ventilator. For more information, contact Puritan Bennett Technical Services in the U.S. at 1.800.255.6774. Outside of the U.S., contact your local Puritan Bennett representative.
Ventilator System supplies mandatory or spontaneous
840
BiLevel
Ventilator System. Become familiar with
840
Ventilator System,
840
mode. A
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 1-2
Page 27
Introduction

1.1 Technical description

1.1.1 General background

The practitioner uses the GUI touch screens, the off-screen keys, and GUI knob to select the ventilator control parameters and input data (see Figure 1-1). The GUI CPU processes this information and stores it in ventilator memory. The BDU CPU uses this stored information to control and monitor the flow of gas to and from the patient. The two CPUs communicate to transfer and verify any new ventilator control parameters or alarm limits. Each CPU then performs continuous background verification of operational and data integrity.
OP 1
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 1-3
Page 28
OP 1
Graphic user
interface (GUI)
Introduction
Collector
vial
Exhalation module:
Active exhalation valve Pressure transducer Flow sensor
Expiratory filter
(Expiratory
limb)
(Inspiratory limb)
regulator
Air
supply
Oxygen
supply
Inspiratory module:
PSOLs Safety valve Oxygen sensor Pressure transducers Flow sensors
Air
Oxygen regulator
Inspiratory
filter
Patient
circuit
Humidification device
8-00001
Figure 1-1. 840 Ventilator System block diagram
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 1-4
Page 29
Introduction
OP 1

1.1.2 Pressure and flow triggering

The ventilator uses flow or pressure triggering to recognize patient effort. When
pressure triggering
pressure in the patient circuit. As the patient draws gas from the circuit and airway pressure drops by at least the value selected for pressure sensitivity, the ventilator delivers a breath.
When
flow triggering (Flow-by
monitors the difference between the inspiratory and expiratory flow sensor measurements. As the patient inhales, the ventilator measures less exhaled flow while the delivered flow remains constant. The result is an increase in the difference between the inspiratory and expiratory flows. When the difference is at least the operator-selected value for flow sensitivity, the ventilator delivers a breath.
If the patient is not inhaling, any difference between the delivered and exhaled flow is due to sensor inaccuracy or leaks in the patient system. To compensate for leaks in the patient system which can cause autotriggering, the operator can increase the flow sensitivity setting.
As a backup method of triggering inspiration, a pressure sensitivity of 2 cmH
O is also in effect. This setting is the most
2
sensitive setting that is still large enough to avoid autotriggering, yet will trigger with acceptable patient effort.
is in effect, the ventilator monitors
) is in effect, the ventilator

1.1.3 Breathing gas mixture

Air and oxygen from cylinders, wall supplies, or compressor (air only) enter the ventilator through hoses and fittings (the fittings are available in several versions). Once inside the ventilator, air and oxygen are regulated to pressures appropriate for the ventilator, then mixed according to the selected O
The ventilator delivers the mixed air and oxygen through the
inspiratory module
and out to the patient. The oxygen concentration of the delivered gas is monitored here, using a galvanic oxygen sensor. The galvanic sensor generates a voltage proportional to the oxygen concentration. The ventilator reports an alarm if the O
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
sensor is enabled and monitored oxygen
2
%.
2
OP 1-5
Page 30
OP 1
Introduction
concentration is more than seven percent above or below the O
% setting, or below 18% after the concentration stabilizes.
2
The inspiratory manifold also includes a safety valve to relieve patient pressure if necessary (for example, if the patient circuit is kinked or occluded). The inspiratory module also corrects for gas temperature and humidity, based on the practitioner-set humidification type.

1.1.4 Inspiratory pneumatics

Ventilator inspiratory pneumatics consist of two parallel circuits: one for oxygen and one for air. The primary elements of the inspiratory pneumatics are two proportional solenoid valves (PSOLs), which control the flow of gas delivered to the patient. Air and oxygen flow sensors, along with pressure signals from the patient circuit, provide feedback that the BDU CPU uses to control the PSOLs.
As a result, the ventilator supplies mixed breathing gas to the patient, based on the practitioner-set ventilator control parameters. The mixed air and oxygen passes through the patient circuit external to the ventilator. The system delivers the breathing gas mixture to the patient at the patient wye, located in the external patient circuit.

1.1.5 Patient circuit

The
patient circuit
ventilator that route gas between the ventilator and the patient. These components include:
•an
a humidification device (optional) in line with the patient
the inspiratory and expiratory limbs of the patient circuit that
•a
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 1-6
inspiratory filter
between the patient and ventilator
circuit
conduct the breathing gas to and from the patient
collector vial
bulk moisture in the exhaled gas
comprises the components external to the
that protects against contamination
that protects the expiratory pneumatics from
Page 31
Introduction
OP 1
•an
expiratory filter
particulates in the patient’s exhaled gas into the room air or inside the ventilator exhalation pneumatics
The ventilator actively controls the exhalation valve that the software accurately positions throughout the patient’s inspiration and exhalation. The exhalation valve allows the ventilator to deliver aggressive breaths while pressure overshoots are minimized, PEEP is controlled, and excess patient pressures are relieved. The exhalation system monitors the exhaled gas leaving the patient circuit for spirometry.
Throughout the respiratory cycle, pressure transducers monitor inspiratory, expiratory, and atmospheric pressures. The temperature of the exhaled gas is heated to a temperature above its dew point to prevent condensation in the exhalation compartment. Refer to Appendix C for a detailed diagram of the ventilator’s pneumatic system and patient circuit.
that limits the escape of microorganisms and

1.1.6 AC mains and backup power system

The ventilator derives its power to operate from the AC mains (wall) power or the backup power system (BPS). The design of the BDU integral power supply protects against excessive voltages, temperatures, or current draws. A power cord retainer prevents accidental disconnection of the BDU from the AC mains. A power switch cover on the front face of the BDU protects against spills and accidental AC power-off.
The ventilator connects to the 802 BPS, which supplies DC power to the ventilator if AC power is lost. A fully charged BPS operating under nominal ambient conditions, can power the ventilator for at least 30 minutes (the BPS does not power the compressor unit or the humidifier, if present). The GUI indicates when the ventilator is operating on the BPS, rather than AC mains.
When AC power is connected, it recharges the BPS. The BPS continues to recharge from the AC power during normal ventilator operation.
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 1-7
Page 32
OP 1
Introduction

1.1.7 Ventilator emergency states

Emergency states include
safety valve open (SVO)
always includes the independent of a
The following describe the two ventilator emergency states:
Safety valve open (SVO)
air and oxygen supplies are lost, or an occlusion is detected, or the ventilator enters the
The safety valve open ( room air unassisted by the ventilator. The ventilator remains in the
SVO
state is corrected.
When the ventilator enters the on the front face of the BDU illuminates, and a high-urgency alarm sounds.
In case of a malfunction that prevents software from opening the safety valve, there is also an analog circuit that opens the safety valve if system pressure exceeds 100 to 120 cmH
Ventilator inoperative (Vent Inop)
ventilator inoperative condition if a hardware failure or critical software error occurs that could compromise safe ventilation of the patient.
Vent I n o p
state until the condition that caused the emergency
ventilator inoperative (Vent Inop)
. When a
state. A
SVO
: The ventilator enters a
Vent Ino p
SVO
condition.
Vent Inop
) state allows the patient to breathe
SVO
condition occurs, it
state can also occur
condition.
state, the
SVO
: The ventilator declares a
SVO
SVO
and
state if both
indicator
O.
2
When a the front face of the BDU illuminates and the ventilator enters the
If a ventilator from use until qualified service personnel evaluate and correct the
If the ventilator declares a test (POST) must first verify that power levels to the ventilator are acceptable and that the functions of the major electronics systems are satisfactory before normal ventilation can resume. Qualified service personnel must repair the ventilator to correct the problem and execute EST successfully before normal ventilation is allowed.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 1-8
Vent Inop
state, which in turns sounds a high-urgency alarm.
SVO
Vent Inop
condition occurs, the
condition occurs, immediately remove the
Vent Ino p
condition.
Vent Inop
Vent I n o p
state, the power on self
indicator on
Page 33
Introduction

1.2 Graphic user interface

This section describes the graphic user interface (GUI), the GUI keys, the GUI indicators, and the symbols you see on the GUI.
OP 1
Upper screen:
monitored
information
(alarms,
patient data)
Lower screen:
ventilator
control
parameters
The graphic user interface (GUI) of the comprises the
DualView
touch screens, the off-screen keys located
Ventilator System
840
below the touch screens, and a knob. Use the knob to set a given ventilator control parameter to its desired value. Press the
ACCEPT
key—the off-screen key above and right of the knob—to enter the selected value or parameter into memory.
Figure 1-2 identifies the components of the GUI, and the location of information on the
Vital patient data
Alarms and ventilator status
Assorted patient data,
including graphical displays
Active alarm log, if applicable
Primary patient parameters
Setup of ventilator control parameters, alarm limits, breath timing parameters, and other parameters
Symbol definitions
DualView
touch screens.
Prompt area
Status indicators
Off-screen keys
CLEAR key
ACCEPT key
Knob
8-00025
Figure 1-2. 840 Ventilator System graphic user interface (GUI)
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 1-9
Page 34
OP 1

1.3 User interface controls and indicators

Descriptions of the controls and indicators on the graphic user interface are given in Table 1-1 below.
Table 1-1: 840 Ventilator System GUI controls and indicators
Control or indicator Function
Screen lock key: When the yellow light on the screen lock key
is lit, the screen or off-screen controls (including the knob and ACCEPT key) have no effect when touched until you press the screen lock key again. New alarms automatically unlock the screen and controls.
The screen lock allows you to clean the touch screen and prevents inadvertent changes to settings and displays.
Alarm volume key: Allows you to adjust the alarm volume when you hold down this key while turning the knob. You cannot turn off the alarm volume.
Alarm silence key: Turns off the audible alarm for two minutes. The yellow light on the alarm silence key illuminates during the silence period. An ALARM SILENCE IN PROGRESS indicator displays on the lower touch screen, along with a CANCEL button, if there is not a higher-priority alarm display active. To exit out of the alarm silence, touch the CANCEL button.
The system automatically exits the alarm silence when the two-minute interval times out. High-urgency alarms such as Device Alerts, Safety Valve Open, Occlusion, and loss of either gas supply cancel the alarm silence.
Each time you press the alarm silence key, the silence period resets to two minutes. Each time you press the alarm silence key (whether or not there is an active alarm), the keypress is recorded in the alarm log.
Introduction
Alarm reset key: Clears active alarms or resets high-urgency alarms and cancels an active alarm silence. Each time you press the reset key, it is recorded in the alarm log, if there is an active alarm. You cannot reset a DEVICE ALERT alarm.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 1-10
Page 35
Introduction
OP 1
Table 1-1: 840 Ventilator System GUI controls and indicators (cont)
Control or indicator Function
Information key: Displays basic operating information about
the ventilator. Press the key to display a menu of information topics, then touch the button corresponding to the desired topic. Browse topical information using the , ,
, and buttons located in the information header.
GO BACK
Oxygen sensor calibration key: Delivers 100% oxygen (if available) for two minutes and calibrates the oxygen sensor. The green light on this key illuminates and a message (100%
Cal in Progress) on the lower touch screen indicates that
O
2
100% O
delivery is active. If you press the O2 key again, the
2
system restarts the two-minute delivery interval. Press CANCEL to stop the calibration.
Use the procedure in Appendix D.2 to test the oxygen sensor calibration.
Manual inspiration key: In A/C, SIMV, and SPONT modes, delivers one manual breath to the patient in accordance with the current mandatory breath parameters. In BILEVEL mode, transitions from Low PEEP (PEEP
) to High PEEP (PEEPH) (or
L
vice versa). To avoid breath stacking, a manual inspiration is not delivered during inspiration or during the restricted phase of exhalation.
You can use the MANUAL INSP key to supplement minute volume or to assist measurement of a patient data parameter, such as peak inspiratory pressure, or to run an INSP PAUSE maneuver in SPONT mode.
DIAGRAM
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 1-11
Page 36
OP 1
Introduction
Table 1-1: 840 Ventilator System GUI controls and indicators (cont)
Control or indicator Function
Expiratory pause key: Causes the ventilator to seal the
patient’s breathing circuit when the expiratory phase of a designated breath, mandatory or spontaneous, is followed by a time-cycled mandatory inspiration. An expiratory pause is used to estimate PEEP
The ventilator performs two types of pause maneuver: automatic, which you initiate by a momentary press of the
EXP PAUSE
continuous press on the key. An automatic pause performs the maneuver until the pressure stabilizes, then takes its measurements. The pause lasts at least 0.5 second and does not exceed 3.0 seconds.
During a manual pause, the ventilator takes its measurements as soon as the pressure stabilizes or the pause ends. The ventilator continues the maneuver until you release the PAU SE key. The pause cannot exceed 20 seconds. Section 4.9 describes, in detail, how to use the EXP PAUSE key.
key, and manual, which you control by a
and PEEPI (autoPEEP).
TOT
EXP
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 1-12
Page 37
Introduction
OP 1
Table 1-1: 840 Ventilator System GUI controls and indicators (cont)
Control or indicator Function
Inspiratory pause key: Causes the ventilator to seal the
patient’s breathing circuit at the conclusion of the gas delivery phase of a designated, volume- or pressure-based mandatory inspiration. This inspiratory pause maneuver provides a means to measure the patient’s static lung-thoracic compliance
), static resistance (R
(C
STAT
The inspiratory pause maneuver maintains the inflated state of the lungs.
The ventilator performs two types of pause maneuver: automatic, which is initiated by the momentary press of the
INSP PAUSE
continuous press on the key. An automatic pause performs the maneuver until the pressure
stabilizes, then the system takes its measurements. The pause event lasts at least 0.5 second but no longer than
2.0 seconds. In a manual pause, the maneuver continues until you release
the INSP PAUSE key, but cannot exceed 7 seconds. The ventilator computes C and displays the values at the end of the maneuver. P computed and updated continuously during the plateau, and its value is frozen at the end of the plateau. Section 4.10 describes, in detail, how to use the INSP PAUSE key.
key, and manual, which you control by a
STAT
), and plateau pressure (PPL).
STAT
and R
at the end of the plateau
STAT
is
PL
Knob: Adjusts the value of a setting. A highlighted button on a touch screen means that the knob is linked to that setting. Where applicable, a clockwise turn of the knob increases the highlighted value, and a counterclockwise turn of the knob decreases the highlighted value.
Clear: Cancels a proposed ventilator parameter value change.
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 1-13
Page 38
OP 1
Introduction
Table 1-1: 840 Ventilator System GUI controls and indicators (cont)
Control or indicator Function
Accept: Applies and saves new ventilator parameter value(s).
Red high-urgency alarm indicator ( ! ! ! ): This alarm
indicator blinks rapidly if active; it is steadily lit if autoreset. Yellow medium-urgency alarm indicator ( ! ! ): This alarm
indicator blinks slowly if active; it turns off if autoreset. Yellow low-urgency alarm indicator ( ! ): This indicator is
steadily lit if active; it turns off if autoreset. Green normal ventilator operation indicator: When
ventilation is active and no alarm states exist, this indicator is steadily lit. This indicator is off if the ventilator is not in a ventilation mode, for example, during service mode or short self test (SST).
VENT
INOP
SAFETY
VALVE
OPEN
BATTERY
READY
BATTERY
ON
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 1-14
Red ventilator inoperative (VENT INOP) indicator: The
ventilator cannot support ventilation and requires service. The ventilator enters the safe state (safety ventilation) and discontinues detection of new patient data or alarm conditions. Qualified service personnel must repair the ventilator to correct the problem and execute EST successfully before normal ventilation is allowed. This indicator is accompanied by an audio signal and cannot be reset.
Red safety valve open (SVO) indicator: The ventilator has entered its safe state and opened its safety valve to allow the patient to breathe unassisted from room air.
Green BPS ready indicator: The ventilator senses that the BPS is installed, operational, and has at least two (2) minutes of estimated run time.
Battery on power indicator: When the yellow bar to the right of a lit BPS ready indicator is lit, the ventilator is operating on BPS, and AC power is insufficient to support ventilator operation. During BPS operation, power to the compressor unit and the humidifier outlet is off.
Page 39
Introduction
OP 1
Table 1-1: 840 Ventilator System GUI controls and indicators (cont)
Control or indicator Function
COMPRESSOR
READY
COMPRESSOR
ON
The indicators on the breath delivery unit are shown in Table 1-2.
VENT
INOP
SAFETY
VALV E
OPEN
Green compressor ready indicator: The compressor cable and air supply hose are connected to the ventilator. The compressor is up to operating pressure but not supplying gas to the ventilator. The compressor motor turns on intermittently to keep the compressor chamber pressurized.
Green compressor operating indicator: When lit, compressor is supplying air to the ventilator. This indicator does not light unless the compressor is actually supplying air to the ventilator.
Table 1- 2: BDU in dicators
Red ventilator inoperative (VENT INOP) indicator: The
ventilator cannot support ventilation and requires service. The ventilator enters the safe state (safety ventilation) and discontinues detection of new patient data or alarm conditions. Qualified service personnel must repair the ventilator to correct the problem and execute EST successfully before normal ventilation is allowed. This indicator is accompanied by an audio signal and cannot be reset.
Red safety valve open (SVO) indicator: The ventilator has entered its safe state and opened its safety valve to allow the patient to breathe unassisted from room air.
DISPLAY
(GUI) INOP
Red loss of GUI indicator: The ventilator has detected a malfunction that prevents the GUI from reliably displaying or receiving information.

1.3.1 Onscreen symbols and abbreviations

Touch an onscreen symbol to display its definition in the lower left corner of the lower screen. Table 1-3 summarizes the symbols and abbreviations the ventilator uses.
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 1-15
Page 40
OP 1
For example, if you touch:
V
MAX
21.8
The symbol definition area shows this message:
V
MAX
Table 1-3: 840 Ventilator System symbols and abbreviations
L
min
= Peak flow
Introduction
Symbol or
abbreviation
(blinking)
P
%
Definition
Additional alarms that relate to the monitored information are active. The symbol blinks when there is not enough screen area to display all active alarms.
The upper alarm limit
The lower alarm limit
Press to access the alarm log
Alarm log contains events that you have not yet viewed
Rise time percent
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 1-16
Page 41
Introduction
OP 1
Table 1-3: 840 Ventilator System symbols and abbreviations (cont)
Symbol or
abbreviation
RAMP SQUARE
Definition
Flow pattern
The value you selected for a ventilator control parameter exceeds its recommended limit (soft bound) and requires acknowledgement to continue
or
The value selected exceeds its allowable minimum or maximum limit (hard bound)
Press to view more patient data
Press to view patient data graphics
Press to view additional screens
X-axis (time or pressure) adjust of patient data graphics
Y-axis (pressure, volume, or flow) adjust of patient data graphics
Baseline pressure (PEEP) adjust
A/C Assist control ventilation mode
AV Apnea ventilation
C
STAT
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
Static compliance
OP 1-17
Page 42
OP 1
Introduction
Table 1-3: 840 Ventilator System symbols and abbreviations (cont)
Symbol or
abbreviation
E
SENS
Spontaneous expiratory sensitivity percentage
Definition
EST Extended Self Test
f Respiratory rate (ventilator control parameter)
f
f
TOT
TOT
Total respiratory rate (monitored)
High respiratory rate alarm
GUI Graphic user interface
HME Heat-Moisture Exchanger
I:E Inspiratory to expiratory ratio
O
2
O
2
1
O2% High delivered O2% alarm
3
O2%Low delivered O
Monitored oxygen percentage (patient data)
Oxygen percentage (ventilator control parameter)
% alarm
2
PC Pressure control (mandatory breath type)
P
P
2 P
3
4
P
MEAN
PEAK
PEAK
P
PEAK
P
PEAK
PEAK
Mean circuit pressure
High circuit pressure alarm
High circuit pressure alarm limit
Low circuit pressure alarm
Low circuit pressure alarm limit
Peak circuit pressure (patient data)
PEEP positive end-expiratory pressure (ventilator control
parameter)
PEEP
H
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
High PEEP (ventilator control parameter, BILEVEL mode only)
OP 1-18
Page 43
Introduction
OP 1
Table 1-3: 840 Ventilator System symbols and abbreviations (cont)
Symbol or
abbreviation
PEEP
I
PEEP
L
PEEP
TOT
PEEP
P
I
P
I END
P
PL
Intrinsic PEEP (patient data)
Low PEEP (ventilator control parameter, BILEVEL mode only)
Total PEEP (patient data)
End expiratory pressure (patient data)
Inspiratory pressure (ventilator control parameter)
End inspiratory pressure (patient data)
Plateau pressure (patient data)
Definition
POST Power on self test
PS Pressure Support (spontaneous breath type)
P
P
P
P
SENS
SUPP
-TRIG
VENT
Pressure sensitivity
Pressure support (ventilator control parameter)
Pressure triggering
High internal ventilator pressure alarm
R
STAT
Static resistance
SIMV Synchronous intermittent mandatory ventilation mode
SPONT Spontaneous ventilation mode
SST Short self test
T
A
T
E
T
H
T
I
1
T
I SPONT
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
Apnea interval
Expiratory time
High PEEP time (BILEVEL mode only)
Inspiratory time
High spontaneous inspiration time alarm
OP 1-19
Page 44
OP 1
T
Introduction
Table 1-3: 840 Ventilator System symbols and abbreviations (cont)
Symbol or
abbreviation
2
T
I SPONT
T
L
T
PL
V
E SET
High spontaneous inspiration time alarm limit
Low PEEP time (BILEVEL mode only)
Plateau time
Set minute volume (calculated from ventilator control
Definition
parameters)
V
E SPONT
1
V
3V
E TOT
E TO
Exhaled spontaneous minute volume
High exhaled minute volume alarm
Low exhaled minute volume alarm
VC Volume control (mandatory breath type)
V
MAX
V
SENS
V
1
V
T
TE
V
TE
Peak flow (ventilator control parameter)
Flow sensitivity
Tidal volume (ventilator control parameter)
Exhaled tidal volume
High exhaled tidal volume alarm
3
V
TE MAND
3
V
TE SPONT
1
V
TI
V
TI
Low exhaled mandatory tidal volume alarm
Low exhaled spontaneous tidal volume alarm
Inspired tidal volume
High inspired (mandatory or spontaneous) tidal volume alarm*
V
TI MAND
1
V
TI MAND
V
TI SPONT
1
V
TI SPONT
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
Inspired mandatory tidal volume
High inspired mandatory tidal volume alarm*
Inspired spontaneous tidal volume
High inspired spontaneous tidal volume alarm*
OP 1-20
Page 45
Introduction
OP 1
Table 1-3: 840 Ventilator System symbols and abbreviations (cont)
Symbol or
abbreviation
-TRIG
V
*Refer to Section 13.10 for information regarding inspired tidal volume alarms.
Flow triggering
Definition
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 1-21
Page 46
OP 1

1.4 Ventilator system labeling symbols

The following symbols appear on the various components of the
Ventilator System.
840
NOTE:
All labels shown are examples, and may not reflect the exact configuration of your ventilator.
Power switch positions: ON represents the power on position and represents the power off position. The power switch, located on the BDU front panel, turns ON/OFF the BDU and
8-10006
8-00400
the GUI. When the power switch is in the off position, the BPS continues to charge if AC power is present.
Refer to manual: When this symbol appears on the product, it means refer to documentation for information.
Type B equipment, per IEC 60601-1
Introduction
8-00445
Potential equalization point (ground): Provides a means of connection between the equipment and the potential equalization busbar of the electrical connection. A common grounding point for the entire ventilator.
8-00426
Indicates the degree of protection provided by enclosure (drip-proof)
8-00432
CSA certification mark that signifies the product has been evaluated to the applicable ANSI/Underwriters Laboratories Inc. (UL) and CSA standards for use in the US and Canada.
SN Serial number
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 1-22
Page 47
Introduction
Data Key Data key connection
OP 1
BPS charging indicator: When the ventilator is operating on
mains power, the top symbol (green indicator next to gray battery icon) on the front of the BPS indicates that the BPS is charged, and the bottom symbol (yellow indicator next to gray battery icon) on the front of the BPS indicates that the BPS is charging.
8-00462
Caution
Do not remove the data key. The data key enables software options, and stores ventilator operational hours, compressor unit operational hours, and the serial numbers for the BDU and GUI. The ventilator will not operate without its factory-installed data key.
TEST
PTS 2000
DISPLAY
(GUI)
Ventilator
circuit
breaker
Compressor &
humidifier
circuit
breaker
8-00405
Compressor outlet:
5.6 A max
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
TEST button. After you touch the Short Self Test (SST) onscreen key (available only during ventilator startup), you must press the TEST button within 5 seconds in order to access SST.
Puritan Bennett PTS 2000™ Performance Test System connection, for use by qualified service personnel only.
GUI connection
Circuit breaker for ventilator power supply, located in the BDU
Ventilator circuit breaker for compressor and humidifier
Alternating current (at AC inlet and AC power indicator)
Maximum allowed output to auxiliary mains socket (compressor electrical connection)
OP 1-23
Page 48
OP 1
Introduction
BPS electrical connection
8-00413
Exhalation filter latch unlock/lock
8-00434
Exhalation filter latch open indicator: This red indicator is located on the surface behind the closed latch, and is easily visible when the latch is open.
GUI mounting latch unlock/lock
8-00428
Remote alarm port
IOIOI
RS-232 port
Susceptible to electrostatic discharge
Electric shock hazard
8-00443
Explosion hazard
8-00450
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 1-24
Page 49
Introduction
Fire hazard
8-00417
802 BPS product information label
GUI product information label
OP 1
US/75427C
US/75422C
GUI ports label Remote alarm and RS-232 port (9.4-inch GUI only). Refer to Appendix E for GUI remote
alarm and RS-232 port specifications.
ENGLISH/75416A
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 1-25
Page 50
OP 1
Humidifier electrical label (This label not visible unless cover plate over humidifier electrical connection is
removed.)
BDU gas inlet label
Introduction
US/75415A
US/75421A
BDU to patient label
US/75417A
Compressor gas connection label
US/75418A
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 1-26
Page 51
Introduction
Compressor information label
OP 1
76505C
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 1-27
Page 52
OP 1
BDU information label
Introduction
US/75423C
BDU cooling vent label
US/75414A
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 1-28
Page 53
Introduction
BDU I/O disconnect label
BDU exhaust information label
OP 1
US/75419A
US/75420A
BPS electrical connection label
US/75334Lbl
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 1-29
Page 54
OP 1
Compressor lint filter label
Expiratory limb connector on exhalation filter
From patient
Introduction
US/75426A
8-00423
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 1-30
Page 55
C
HAPTER
2

How to set up the 840 ventilator

Chapter 2 describes how to set up the ventilator:
How to connect the electrical supply
How to connect the air and oxygen supplies
How to connect the patient circuit and accessories
A Puritan Bennett Customer Service Engineer (CSE) must first install the ventilator and run an extended self test (EST), which calibrates the exhalation valve, flow sensors, and atmospheric pressure transducer, before you connect a patient to the ventilator for the first time.
Warning
• When you lift the ventilator, use assistance and appropriate safety precautions. Figure 2-1 shows the proper technique to lift each ventilator component.
• To avoid interrupted ventilator operation or possible damage to the ventilator, always use the ventilator on a level surface in its proper orientation.
• To avoid the possibility of injury to the patient and ensure proper ventilator operation, do not attach any device to the port labeled EXHAUST unless the device is specifically authorized by Puritan Bennett.
• To minimize the increased risk of fire due to an oxygen­enriched environment, do not use the ventilator in a hyperbaric chamber.
• To avoid raising the oxygen concentration of room air, use the ventilator in an adequately ventilated room.
2
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 2-1
Page 56
OP 2
Lift the BDU
from horizontal
surfaces as shown.
Use two people to lift
the compressor from
the base and the handles.
How to set up the 840 ventilator
Lift the GUI from the base and the handle.
8-00011
Figure 2-1. How to lift the ventilator components
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 2-2
Page 57
How to set up the 840 ventilator
Caution
• Do not connect or disconnect the ventilator’s graphic user
interface (GUI), backup power source (BPS), or compressor while the power switch is on or the ventilator is connected to AC power.
• All components must be securely mounted and connected by
qualified service personnel according to the appropriate Puritan Bennett installation instructions.
• Do not obstruct the breath delivery unit (BDU), GUI, or
compressor cooling vents or fan vents.
• To avoid possible damage to ventilator components, do not use
the horizontal surfaces of the ventilator to place or stack objects.
NOTE:
Before you use the ventilator for the first time, wipe the ventilator exterior clean and sterilize its components according to the instructions in Chapter 7 of this manual. Follow your institution’s protocol for cleaning and sterilizing the ventilator and its components.
OP 2
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 2-3
Page 58
OP 2
How to set up the 840 ventilator

2.1 How to connect the electrical supply

Warning
• To minimize the risk of electrical shock, always connect the ventilator power cord into a grounded AC power outlet.
• In the US, always connect the ventilator to an AC receptacle marked “Hospital Only” or “Hospital Grade” to ensure proper grounding of the ventilator.
• The 802 BPS must always be installed. Without the BPS, the ventilator is not protected against low or lost AC power. Do not use the ventilator unless a BPS with at least minimal charge is installed.
• When possible, connect the ventilator to an outlet connected to the hospital emergency back-up power system. Refer to Section A.4 for ventilator electrical specifications.
Normally the 802 BPS operates the ventilator when AC power is lost or drops below a minimum level.
A new, fully charged BPS can operate the ventilator (without the compressor or a humidifier) for at least 30 minutes; allowing the ventilator to be used for transport purposes within the healthcare facility.
Warning
The 802 BPS is intended for short-term use only, and is not intended as a primary alternative power source. The BPS is intended to power the BDU and GUI only. In case of AC power loss, power is not available to run either the compressor or the humidifier.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 2-4
Ventilator System is mains-powered. The
840
Page 59
How to set up the 840 ventilator
If you turn on the ventilator after it has been unplugged for an extended period, the occurs, recharge the 802 BPS by leaving it connected to a ventilator connected to AC power for up to 8 hours (ventilator does not need to be turned on). If, after turning the ventilator back on, the
INOPERATIVE BATTERY
LOW BATTERY
must replace the BPS battery.
Figure 2-2 shows how to connect the power cord to AC power. Built-in power cord retainer tabs protect against accidental disconnection. Ensure that the power cord is securely fastened into the AC receptacle prior to operation. To remove the cord, squeeze the tabs on the top and bottom of the plug and pull outward.
OP 2
LOW BATTERY
alarm is still active or if the
alarm is active, qualified service personnel
alarm may sound. If this
Power
cord
Power cord retainer tabs.
To AC power
8-00002
Squeeze tabs and pull outward to disconnect power cord.
Figure 2-2. How to connect the ventilator power cord
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 2-5
Page 60
OP 2
How to set up the 840 ventilator
Figure 2-3 shows the power switch and AC indicator. When illuminated, the AC indicator indicates that the ventilator is receiving AC power and that the 802 BPS will be recharged as needed. The AC indicator is independent of the power switch, and the power switch does not turn off AC power to the ventilator power supply. When both the power switch and AC indicator are on, power is available for the humidifier and compressor.
Ventilator power supply circuit breaker
ON
Ventilator
power switch
AC panel
AC indicator
Humidifier
and compressor
circuit breaker
Potential
equalization
(ground
point)
Compressor
connection
Figure 2-3. Ventilator power switch, AC indicator, and AC panel
If the ventilator power supply circuit breaker (located on the ventilator's AC panel, Figure 2-3) opens but AC power is still present and the ventilator is operating on BPS, power is still available to the humidifier and compressor connectors (although ventilator software disables compressor operation).
AC power connection
8-00045
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 2-6
Page 61
How to set up the 840 ventilator
When the power cord is not in use, wrap the power cord around the hook on the back of the cart for convenient storage (Figure 2-4).
OP 2
8-00004
Figure 2-4. Power cord storage on the cart
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 2-7
Page 62
OP 2
How to set up the 840 ventilator

2.2 How to connect the air and oxygen supplies

The or wall supplies. Follow these steps to connect the air and oxygen supplies:
1
2
Ventilator System can use air and oxygen from cylinder
840
Ensure that the supply pressures are 35 to 100 psi (241 to 690 kPa).
Connect the supply hoses to the inlet connectors at the rear of the ventilator (see Figure 2-5).
Warning
• Connect only air to the air inlet, and only oxygen to the oxygen inlet. Do not attempt to switch air and oxygen or connect any other gas.
• Always connect at least two gas sources to the ventilator to ensure that a constant gas supply is available to the patient. There are three gas source connections: the compressor, air inlet, and oxygen inlet.
• Do not use conductive, high-pressure supply hoses. Doing so may negatively affect the ventilator’s ground isolation characteristics.
• Use only high-pressure supply hoses recommended by Puritan Bennett. Other hoses may be restrictive and may cause improper ventilator operation.
Caution
To prevent damage to the ventilator, ensure that the connections to the air and oxygen supplies are clean and unlubricated, and that there is no water in the air or oxygen supply gas.
If you suspect water in the air supply gas, use an external wall air water trap to prevent water damage to the ventilator or its components.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 2-8
Page 63
How to set up the 840 ventilator
NOTE:
When you connect a pressurized air or oxygen source, the ventilator air and oxygen regulators have a maximum bleed rate of 3 L/min, even when the ventilator is not in use. Always take this bleed rate into account when calculating air and oxygen usage.
When the air and oxygen hoses are not in use, you can wrap them around the hook on the back of the cart for convenient storage (Figure 2-5).
Air inlet
connector
Oxygen
inlet
connector
OP 2
Oxygen hose (from oxygen supply)
Air hose
(from air
supply)
8-00005
Figure 2-5. How to connect the air and oxygen supplies
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 2-9
Page 64
OP 2
How to set up the 840 ventilator

2.3 How to connect the patient circuit components

Warning
• To minimize the risk of bacterial contamination or component damage, inspiratory and expiratory filters must always be handled with care and connected to the ventilator during use.
• To minimize the risk of patient injury, use only patient circuits qualified for use in oxygen-enriched environments with the 840 Ventilator System. Do not use antistatic or electrically conductive tubing. To ensure a leak-tight connection, only use connectors and tubes with ISO-standard cone and socket fittings (or use adapters to connect barbed cuff fittings to ISO-standard fittings).
• If you use an external, pneumatically-powered nebulizer with the 840 ventilator, it adds flow to the patient circuit and can adversely affect spirometry, delivered O delivered tidal volumes, and breath triggering. Additionally, aerosolized particulates in the ventilator circuit can lead to an increase in exhalation filter resistance.
• Use one of the patient circuits listed in Appendix B to ensure that the maximum pressure/flow values specified by IEC 60601-2-12:2001 are not exceeded (see Table A-11 on page OP A-18 for patient circuit testing specifications). Using a circuit with a higher resistance does not prevent ventilation, but can cause a short self test (SST) fault or compromise the patient’s ability to breathe through the circuit.
2
%,
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 2-10
Page 65
How to set up the 840 ventilator
NOTE:
• Puritan Bennett recommends that you run SST (Short Self Test) every 15 days, and between patients, and when you change the patient circuit (particularly when you change the circuit type, for example, from adult to pediatric or neonatal).
• Puritan Bennett recognizes that the protocol for running SST varies widely among health care institutions. Puritan Bennett does not specify or require specific practices that will meet the needs of all institutions, nor is Puritan Bennett responsible for the effectiveness of institutional practices.

2.3.1 How to select and connect a patient circuit

Use low-compliance patient circuits to ensure optimum compliance compensation, and use pediatric patient circuits when the patient ideal body weight (IBW) is greater than 7 kg (15 lb) but less than or equal to 24 kg (53 lb). Use the software option and neonatal patient circuits for patients whose IBW is less than or equal to 7 kg.
For patients whose IBW is less than or equal to 24 kg, the compliance compensation volume limit is four times the set tidal volume, in addition to the set tidal volume. To avoid activating a severe occlusion alarm, only use neonatal patient circuits with the
NeoMode
Table 2-1 shows IBW values and patient circuit types. The “Allowed but not recommended” ranges require an override.
software option.
OP 2
NeoMode
Warning
Recommended ranges exist to ensure patient safety. Only those with the expertise to judge the appropriate circumstances should override the recommended ranges.
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 2-11
Page 66
OP 2
How to set up the 840 ventilator
Table 2-1: Patient circuit and IBW values
Recommendation Ideal body weight (IBW) in kg (lb)
Recommended Neonatal: 0.5-7.0 kg (1.1-15 lb)
Pediatric: 7.0-24 kg (15-53 lb) Adult: 25-150 kg (55-330 lb)
Allowed but not
recommended
Neonatal: Not applicable
Pediatric: 3.5-6.5 kg (7.7-14.3 lb), and
25-35 kg (55-77 lb)
Adult: 7-24 kg (15-53 lb)
Figure 2-6 shows how to connect the patient circuit, including the inspiratory filter, humidifier (if used), inspiratory limb, patient wye, expiratory limb, collector vial, and expiratory filter.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 2-12
Page 67
How to set up the 840 ventilator
(From patient)
Expiratory
filter
Expiratory limb
Patient wye
of patient circuit
Collector vial
OP 2
(To patient)
Inspiratory filter
Tubing
Inspiratory limb of patient circuit
Humidifier
8-00006
Figure 2-6. How to connect the patient circuit
Warning
To ensure that all patient circuit connections are leak-tight, always perform a circuit leak test by running SST each time you install the expiratory filter on the ventilator.
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 2-13
Page 68
OP 2
How to set up the 840 ventilator
Warning
Adding accessories to the ventilator can increase system resistance. Ensure that any changes to the recommended ventilator circuit configurations do not exceed the specified values for inspiratory and expiratory resistance (Appendix A). If you add accessories to the patient circuit, always run SST to measure circuit compliance before you begin ventilation of the patient.

2.3.2 How to install the expiratory filter and collector vial

Install the expiratory filter and collector vial as follows:
Place the expiratory filter latch in the up position (see
1
Figure 2-7).
Slide the expiratory filter into the housing area with the
2
expiratory limb connection facing you.
Push the expiratory filter latch down; it will position the filter
3
properly.
Attach the expiratory limb of the patient circuit to the filter’s
4
expiratory limb connection.
If you do not use a drain bag, be sure to cap the collector vial drain port on the expiratory filter (Figure 2-8).
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 2-14
Page 69
How to set up the 840 ventilator
Pull latch up to install filter; push down to hold filter and collector vial in place
Expiratory filter
Expiratory limb connection (from patient)
Figure 2-7. How to install the expiratory filter and collector vial
If you use a drain bag:
Install the expiratory filter. (Refer to the instructions above.)
1
Install the clamp on the drain bag tubing, making sure that
2
the clamp is closed.
Uncap collector vial drain port at the base of the collector vial.
3
Connect the collector bag tubing to the vial drain port.
4
Connect the other end of tubing to drain bag.
5
If the ventilator is mounted on the cart, place the drain bag in
6
the cart drawer (Figure 2-8).
Filter housing
hous area
Collector vial
OP 2
Slide the filter rim onto these tracks
8-00018
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 2-15
Page 70
OP 2
Drain bag
Tubing
Place the drain bag
in the cart drawer
Clamp
How to set up the 840 ventilator
The collector vial drain port must be capped if you do not use a drain bag
8-00007
Figure 2-8. How to use the collector vial with or without the drain bag
NOTE:
• The drain bag is designed to lie flat and should not be suspended.
• Check the inspiratory and expiratory limbs of the patient circuit, the collector vial, and the in-line water traps regularly for water buildup. Under certain conditions, they can fill quickly. Empty and clean the collector vial and in-line water traps as necessary.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 2-16
Page 71
How to set up the 840 ventilator

2.3.3 How to install the flex arm

The flex arm supports the patient circuit between the ventilator and the patient. Figure 2-9 shows how to install the flex arm onto one of the two threaded sockets on the ventilator cart.
Flex arm
Threaded
socket
OP 2
Threaded socket in the cart (one of two)
8-00008
Figure 2-9. How to install the flex arm
Caution
Use only the cart handles to move the ventilator. Do not pull or push the ventilator with the flex arm.
Flex arm replacement parts can be found in the
840 Ventilator System Service Manual
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
.
OP 2-17
Page 72
OP 2
How to set up the 840 ventilator

2.3.4 How to install the humidifier

An electrical outlet for a humidifier is located on the front of the BDU. Figure 2-10 shows how to install a Fisher & Paykel humidifier onto the ventilator.
Warning
• When using a Fisher & Paykel humidifier with the 840 ventilator, use the appropriate Fisher & Paykel humidifier chambers for adult, pediatric, and neonatal patients.
• Take proper precautions to prevent water/condensate from splashing into the patient circuit during circuit disconnects and high peak flow rate conditions.
• To avoid possible patient injury or damage to the ventilator system, follow your institution’s protocol for proper patient circuit condensate management.
Caution
• Qualified service personnel must first install the humidifier mounting hardware.
• To avoid equipment damage to the ventilator due to liquid ingress:
- Install the plug cover when the humidifier is plugged into the
ventilator.
- Install the flat cover plate over the humidifier electrical outlet on
the front of the BDU when the humidifier is not plugged into the ventilator.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 2-18
Page 73
How to set up the 840 ventilator
NOTE:
• To ensure uninterrupted ventilator operation, do not install a humidifier whose maximum current capabilities exceed 2.3 A, with a maximum power consumption of 270 VA.
• When you install a Fisher & Paykel humidifier, make sure the humidifier has a right-angle electrical plug. A short power cord is preferable.
•To ensure that ventilator occlusion detection operates properly, do not use Puritan Bennett Cascade humidifiers with the 840 Ventilator System.
• If you have further questions about humidifiers qualified for use with the 840 Ventilator System, contact Technical Services or your local Puritan Bennett representative.
OP 2
BDU
Plug cover
Humidifier
Mounting bracket on front of ventilator
8-00138
Figure 2-10. How to install the humidifier
(Fisher & Paykel version shown)
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 2-19
Page 74
OP 2
How to set up the 840 ventilator

2.3.5 How to use the ventilator cart

Figure 2-11 shows how to lock and unlock the cart’s front wheels.
Warning
To avoid interrupted ventilator operation or damage to ventilator components, use the cart to move the ventilator. Do not use the cables, the power cord, or patient circuit components to push or pull the ventilator.
Press down to lock
Press down to unlock
8-00010
Unlocked position Locked position
Figure 2-11. How to lock and unlock the cart’s front wheels
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 2-20
Page 75
C
HAPTER
3

How to run Short Self Test (SST)

Chapter 3 tells you:
When to run SST
Required equipment for SST
How to set up and run SST
The SST tests and their functions
How to understand the results of SST
3

3.1 Introduction to SST

SST uses an internal, programmed sequence of tests to:
Verify proper function of the flow and pressure sensors
Check the patient circuit for gas leaks
Measure the expiratory filter resistance
Measure patient circuit resistance
Measure patient circuit compliance
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 3-1
Page 76
OP 3
How to run Short Self Test (SST)
SST requires approximately three minutes to complete.
Warning
• Always disconnect the ventilator from the patient before you run SST. If you run SST while the ventilator is connected to the patient, physical injury to the patient may occur.
• An ALERT reported by SST indicates that the ventilator or a related component has a defect. Repair the ventilator or related component before you use the ventilator on a patient, unless you can determine with certainty that the defect cannot create a hazard for the patient, or add to the risks that may occur from other hazards.
• When you run SST, configure the patient circuit exactly as it will be used on the patient (for example, with same accessories). If you add accessories to the patient circuit after you run SST, you must rerun SST with the new accessories installed before you begin to ventilate the patient.

3.2 When to run SST

NOTE:
Puritan Bennett recognizes that health care institutions may have their own ventilator protocols. However, Puritan Bennett is not responsible for the effectiveness of any institution’s protocols. Nor can Puritan Bennett specify, or require, specific practices that will meet the internal needs of every health care institution.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 3-2
Page 77
How to run Short Self Test (SST)
Puritan Bennett recommends that you run SST when one or more of the events below occurs:
When you replace the patient circuit and the exhalation filter after 15 days of use
When you are ready to connect a new patient to the ventilator
When you connect a different patient circuit to the ventilator
When you install a new or sterilized expiratory filter
When you change the patient circuit type
When you change the humidification device type
When you remove or add accessories to the patient circuit, such as a humidifier, water trap, or drain bag
OP 3
Use SST at any time, provided that a patient is ventilator, to:
Check the patient circuit for gas leaks
Calculate patient circuit compliance and resistance
Calculate expiratory filter resistance
After SST begins, the system prompts you to prepare the ventilator to conduct certain tests. The system waits indefinitely at a prompt until you take action and respond appropriately.
attached to the
not
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 3-3
Page 78
OP 3
How to run Short Self Test (SST)

3.3 SST components and requirements

When you conduct SST, you must have available the components and equipment that you will use on the patient:
Patient tubing
Expiratory filter and collector vial
•Inspiratory filter
Humidifier, as applicable
Other accessories (e.g. water traps, drain bag), as applicable
Additional requirements include:
A No. 1 rubber stopper to block the airway at the patient wye
Two gas sources (air and oxygen) connected to the ventilator
Each gas source pressure must be between 35-100 psi (241 to 690 kPa)
Caution
• To prevent SST failures due to leaks, ensure that any circuit components such as collector vial drain port cap (if not using a drain bag), the seal between the expiratory filter and collector vial, and water trap (if used) seals are properly installed.
• If you are using a drain bag, ensure that the tubing is properly installed on the collector vial drain port and that the tubing is clamped. If the drain bag tubing is not clamped during SST, large leaks and large compliance values are possible which may cause SST to report ALERTs or FAILURES.
Wait at least 10 minutes after you turn on the ventilator before you run SST. The warm up time of 10 minutes will stabilize the ventilator and ensure the accuracy of the SST tests.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 3-4
Page 79
How to run Short Self Test (SST)

3.4 SST Procedure

Warning
Always disconnect the ventilator from the patient before you run SST. If you run SST while the ventilator is connected to the patient, physical injury to the patient may occur.
Turn the power switch (located on the front of the BDU). The
1
system conducts the POST (power-on self test) and displays the
Ventilator Startup
Allow the ventilator to stabilize for ten minutes with the
2
power on.
Install the patient circuit, the expiratory and inspiratory filters
3
that you will use to ventilate the patient.
Caution
The patient circuit must be unobstructed and properly connected to the ventilator to ensure accurate circuit resistance measurement.
OP 3
screen.
At the
4
touch screen), then press the the BDU) within 5 seconds. (Refer to Figure 3-1 for location of
TEST
The system displays the
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
Ventilator Startup
button.)
screen, touch the SST button (lower
button (on the left side of
TEST
SST Setup
screen (lower touch screen).
OP 3-5
Page 80
OP 3
How to run Short Self Test (SST)
NOTE:
You must press the TEST button within 5 seconds of touching the SST button or SST will not start.
Tes t button
Figure 3-1. Test button location
Caution
Do not press the test button when powering up the ventilator. This may cause the ventilator to enter Service Mode. If you enter Service Mode, do not attempt to run Extended Self Test (EST) with a patient circuit. Doing so will cause EST to fail. If EST fails, the ventilator will remain in a Vent Inop state until EST successfully passes.
If you accidentally enter Service Mode, exit Service Mode by touching the EXIT button on the lower GUI screen and then pressing the ACCEPT key.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 3-6
Page 81
How to run Short Self Test (SST)
OP 3
Touch the
5
PATIENT CIRCUIT
key in the lower touch screen, then use the knob to select either Adult, Pediatric, or Neonatal (if
Touch the
6
NeoMode
software option is installed) patient circuit.
HUMIDIFICATION TYPE
key in the lower touch screen, then use the knob to select the humidification type that you will use for patient ventilation.
If you will not use a humidifier, set the humidification type to
.
HME
Press
7
ACCEPT
to complete your selection of the patient circuit
and humidification types.
Warning
Incorrectly specifying the patient circuit type or changing the patient circuit type after you have run SST can affect the accuracy of the compliance calculation, the measured exhaled tidal volume, and delivered/measured inspired tidal volumes. You must rerun SST when you change the circuit type. Compliance calculation and tidal volume accuracy may also be affected by incorrectly specifying or changing the humidifier after running SST. If you change humidifiers, ensure you change the humidification type as described in Section 4.8. For optimum accuracy, rerun SST using the new humidifier.
The ventilator automatically starts the test sequence. Refer to
8
Table 3-1 for details regarding each SST test step.
The SST
Flow Sensor, Expiratory Filter, Circuit Resistance
Compliance Calibration
tests require your intervention. The
, and
system will wait indefinitely for your response. Otherwise you don’t need to do anything unless a test result is
FAILURE
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
, or SST is complete.
ALERT
or
OP 3-7
Page 82
OP 3
How to run Short Self Test (SST)
As each test is performed, the
9
SST Status
screen shows test
results (see Table 3-2).
Warning
To ensure reliable SST results, do not repeat an individual test with a different patient circuit if the test result is FAILURE or ALERT. If you suspect a defective patient circuit, replace the patient circuit and restart SST from the beginning.
You can touch
10
touch
EXIT SST
EXIT SST
again to resume testing, or press
restart the ventilator (if SST has not detected an
FAILURE
).
during SST to halt testing. You can
ACCEPT
ALERT
Warning
To ensure correct compensation for circuit resistance and compliance, do not exit SST until the entire SST is successfully completed. Do not begin normal ventilation until the entire SST is successfully completed with the correct patient circuit installed.
When all of the tests in SST are complete, the
11
SST Status
displays all individual test results and SST outcome. Table 3-3 summarizes overall SST outcomes and how to proceed in each case.
To begin normal ventilation (if SST has not detected an
12
or
FAILURE
The ventilator reruns
13
The ventilator displays the
14
with
), touch
EXIT SST
Ventilator Startup
, then press
.
POST
Ventilator Startup
ACCEPT
.
screen. Proceed
to configure the system for the patient.
to
or
screen
ALERT
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 3-8
Page 83
How to run Short Self Test (SST)
Table 3-1: SST test sequence
Test step Function Comments
OP 3
SST Setup The system prompts you to
specify the patient circuit type and humidification type that you will use for the patient ventilation.
1 Specify the patient circuit
type.
2 Specify the humidification
type.
You can select one of three humidification types:
• Heated expiratory tube
• Non-heated expiratory tube
• HME (heat-moisture exchanger)
3 For non-HME humidifiers,
specify the dry humidifier volume. Use the specified volume, not the compressible volume, of the humidifier.
4 Press the
Warning
Select the correct patient circuit type and humidification type. Otherwise, faulty occlusion detection and erroneous expiratory spirometry can result.
ACCEPT
key.
NOTE:
HUMIDIFIER VOLUME
The HME.
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
button is not visible on the touch screen if you select
OP 3-9
Page 84
OP 3
How to run Short Self Test (SST)
Table 3-1: SST test sequence (cont)
Test step Function Comments
SST Flow Sensor Test The system prompts you to
connect the patient circuit to inspiratory filter.
Use Figure 2-6 on page OP 2-13 to connect the patient circuit.
1 Connect the patient circuit
to the inspiratory filter— but without the humidifier.
2 Press
ACCEPT
the test.
NOTE:
Do not run the Flow Sensor Test with a humidifier installed, even if you will use a humidifier when you begin patient ventilation.
The system prompts you to block the patient wye.
The system checks the accuracy of the inspiratory and expiratory flow sensors.
After the test completes, the system prompts you to connect the humidifier.
3 Block the wye with a
No. 1 stopper.
4 Press
If the status of the SST Flow Sensor Test is cannot use the function.
ACCEPT
FAILURE
NOTE:
If you will use a humidifier during patient ventilation, connect the humidifier to the patient circuit after the system passes the SST Flow Sensor Test. Refer to Figure 2-6 on page OP 2-13 for connection information.
to begin
.
, you
OVERRIDE
Circuit Pressure Test The system verifies proper
function of the BDU pressure sensors.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 3-10
If the status of the Circuit Pressure Test is FAILURE, you
cannot use the function.
OVERRIDE
Page 85
How to run Short Self Test (SST)
Table 3-1: SST test sequence (cont)
Test step Function Comments
OP 3
Circuit Leak Test The system determines the
ability of the circuit to hold pressure.
The system displays the drop in circuit pressure over a 10 second interval.
Expiratory Filter Resistance Test
The system prompts you to detach circuit tubing from the expiratory filter.
At the conclusion of the
Expiratory Filter Resistance Tes t, the system displays the
pressure drop across the expiratory filter.
ALERT
If the system reports and you choose to override the alert status, the result can be improper compliance compensation, inaccurate tidal volume delivery, or autotriggering during patient ventilation.
If the test detects excessive leaks, the system reports a
FAILURE
1 Detach the patient circuit
2 Press
If the system reports an for the Expiratory Filter Resistance Test and you override the inaccurate patient pressure estimation can result.
The system will report a
FAILURE
exhalation compartment occlusion or an expiratory filter occlusion.
If you do not correctly follow the prompts to disconnect and connect the patient circuit, the system will report a
FAILURE
.
from the expiratory filter.
ACCEPT
the test.
if the test detects an
.
ALERT
to begin
, an
ALERT
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 3-11
Page 86
OP 3
How to run Short Self Test (SST)
Table 3-1: SST test sequence (cont)
Test step Function Comments
Expiratory Filter Resistance Test
(cont)
Circuit Resistance The system prompts you to
The system prompts you to reattach the patient circuit.
unblock the patient wye.
The system displays the pressure drop across the inspiratory and expiratory limbs.
The reported pressure drop includes the effect of all devices installed on each limb, such as filters, water traps, or a humidifier.
3 Reattach the patient
circuit to the expiratory filter.
4 Press
1 Remove the stopper from
2 Press
If the system reports an for the pressure drop across the two limbs and you override the inaccurate patient pressure estimation can result.
The system reports a if the test detects excessive high or low limb resistance, or if you do not follow the prompt to unblock the wye.
ACCEPT
the next test.
the wye.
ACCEPT
the test.
ALERT
to begin
to begin
ALERT
, an
FAILURE
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 3-12
Page 87
How to run Short Self Test (SST)
Table 3-1: SST test sequence (cont)
Test step Function Comments
OP 3
Compliance Calibration
The system prompts you to block the patient wye.
If you selected a humidification type of either
Heated exp tube or Non­heated exp tube, the
ventilator prompts you to indicate if there is water in the humidifier.
The system displays the compliance of the patient circuit.
The system prompts you to unblock the patient wye.
1 Block the wye with a
No. 1 stopper.
2 Press
3 Press ACCEPT to indicate
If the system reports an for the patient circuit compliance and you override the compliance compensation or inaccurate tidal volume delivery can result.
The system reports a if the test detects an out-of­range compliance condition.
4 Remove the stopper from
5 Press
ACCEPT
the patient circuit compliance test.
YES
or CLEAR to indicate
NO
, as appropriate, to indicate whether or not there is water in the humidifier.
ALERT
, improper
the patient wye.
ACCEPT
complete the SST test sequence.
to begin
ALERT
FAILURE
to
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 3-13
Page 88
OP 3

3.5 SST Results

How to run Short Self Test (SST)
The
ventilator uses four status categories to characterize the
840
individual SST test results, and the overall SST outcome.
ALERT
You can override an
reported for an individual test if you
ALERT
can determine with certainty that the defect in the ventilator or related component cannot create a hazard for the patient, or add to the risks that may arise from other hazards.
NOTE:
If an ALERT is reported and you exit SST without overriding the ALERT, the ventilator will enter the safety valve open (SVO) state and cannot be used for normal ventilation until SST passes or the ALERT is overridden.
FAILURE
When the system declares a
FAILURE
for an individual test in the SST sequence, the ventilator enters the SVO state. When a ventilator experiences a
FAILURE
, immediately remove the equipment from clinical use until qualified service personnel have completed and verified the necessary repairs.
OVERRIDDEN
OVERRIDDEN
is a final status of the overall SST outcome and indicates that you used the override feature when the system reported an the test with an
PASS
condition. (The ventilator must have ended
ALERT
condition.)
ALERT
is the final status of the overall SST outcome in which no
PASS
alerts or failures were detected.
Refer to Table 3-2 and Table 3-3 to learn how to interpret and respond to each of these SST status categories.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 3-14
Page 89
How to run Short Self Test (SST)

3.5.1 How to interpret individual SST test results

SST reports a test result status for each of the individual tests. Use Table 3-2 to interpret SST test results and to determine how to respond.
Table 3-2: Individual SST test results
OP 3
If the test status is:
PASSED
ALERT
FAILURE
It means: Do this:
The system did not detect a fault for the
You do not need to do anything, unless you are prompted by the ventilator.
individual test.
The test result is not ideal, but is not
When the system prompts you, touch one of these buttons, then press ACCEPT:
critical. If SST is in progress, it
EXIT SST
halts further testing and prompts you to make a decision.
A critical problem has been detected, and
RESTART SST
REPEAT
Touch one of these buttons, then press
ACCEPT: SST cannot complete until the ventilator
EXIT SST
passes the failed test.
RESTART SST
NEXT
Discontinue SST
Repeat SST from the beginning
Proceed to the next test
Repeat the test
Discontinue SST
Repeat SST from the beginning
REPEAT
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
Repeat the test
OP 3-15
Page 90
OP 3
How to run Short Self Test (SST)

3.5.2 SST outcomes

When SST has completed all of the tests, use Table 3-3 to determine how to proceed.
Table 3-3: Overall SST outcomes
If the SST outcome is:
PASSED
ALERT
It means: Do this:
All tests passed. Touch one of these buttons, then press
ACCEPT:
Exit SST and begin normal ventilation
Repeat SST from the beginning
Discontinue SST
One or more faults were detected. If you can determine with certainty that this
EXIT SST
RESTART SST
Touch one of these buttons, then press ACCEPT:
EXIT SST
cannot create a hazard for the patient, or add to the risks which may arise from other hazards, you can choose to override the ALERT status and authorize ventilation.
RESTART SST
OVERRIDE
Repeat SST from the beginning
Press ACCEPT to override the ALERT, as allowed by your institution’s protocol. Touch EXIT SST, then press ACCEPT to begin normal ventilation.
FAILURE
One or more critical faults were detected. The ventilator enters the SVO state and cannot be used for
Restart SST with a different patient circuit. Touch one of these buttons, then press ACCEPT:
EXIT SST
Discontinue SST
normal ventilation until SST passes. Service is required.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
RESTART SST
Press ACCEPT to repeat SST from the beginning. If the failure persists, contact qualified service personnel.
OP 3-16
Page 91
C
HAPTER
4

How to use the 840 ventilator

Chapter 4 tells you:
How the ventilator user interface is structured
How to start up the ventilator for a new or previous patient
How to change main settings
How to change other settings
How to set the humidification type, expiratory sensitivity, and disconnect sensitivity
How to enable or disable the oxygen sensor
How to select and set the variable that remains constant when the breath rate setting is changed
How to set the alarm limits
How to perform inspiratory and expiratory pause maneuvers
How to interpret inspiratory pause maneuver displays
How to use Non-invasive ventilation (NIV)
4
NOTE:
The DualView touch screens use light beams to detect where you touch the screen. To avoid a
DEVICE ALERT
substances or objects on the screen.
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
alarm, do not place any foreign
OP 4-1
Page 92
OP 4

4.1 Structure of user interface

The following buttons are available on the upper and lower touch screens. These buttons appear across the bottom portion of each of the two touch screens.
Upper screen
How to use the 840 ventilator
Display More patient graphics
vent setup (vent type, mode, breath types, trigger type, settings)
data (e.g. O2%,
P
I END
Diagnostic code log (system diagnostic, system information, EST/SST diagnostic logs)
VENT
SETUP
)
Current/proposed apnea setup
Alarm log Active alarms (time, event, urgency, alarm, analysis)
Operational time log (compressor, ventilator hours)
Lower screen
APNEA SETUP
Communication setup (printer/DCI, baud rate, data bits, parity mode)
SST result log
ALARM SETUP
Current/proposed alarm settings
Ventilator configuration (revisions, serial numbers, part numbers, installed options)
Time/date change
Other screens
Test summary (time, date, outcome of SST, EST)
Other screensCurrent/proposed
More settings (humidification type,
sensor enable/
O
2
disable, disconnect sensitivity, humidifier volume)
Figure 4-1. Touch screen user interface
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 4-2
Page 93
How to use the 840 ventilator

4.2 Patient setup

Warning
Always complete the patient setup before you attach a patient to the ventilator. If you attach a patient before the setup procedure is complete, the ventilator issues a procedure error and initiates the safety ventilation mode.
When you turn on the ventilator, the ventilator automatically runs POST (Power On Self Test). After POST passes, the system displays the screen. The prompt area, located in the lower right corner of the lower screen, contains setup instructions.
Ventilator Startup
OP 4
screen (Figure 4-2) on the lower
Figure 4-2. Ventilator Startup screen
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
OP 4-3
Page 94
OP 4
How to use the 840 ventilator

4.2.1 How to ventilate with most recent control parameters

To continue ventilation with the most recent ventilator control parameters, touch does not begin until a patient is connected. A flashing reminder arrow prompts you to consider the previous Tube I.D. and Tube Type if the prior Spontaneous Type used these parameters.
Same Patient
and press
ACCEPT
. Ventilation

4.2.2 How to ventilate with new control parameters

Refer to Table A-12 in Appendix Afor the descriptions, ranges, resolutions, accuracies, and new patient values of the available ventilator control parameters.
ouch the
1
T
parameters for patient ventilation.
Touch the
2
to adjust the IBW. The proposed value is highlighted.
Warning
Always enter the IBW that is appropriate for the patient. The system uses the patient’s IBW to automatically set certain values, alarm limits, and parameter boundary limits for several initial parameters. (The IBW values that correlate with patient height are listed in Table 4-1.)
New Patient
(ideal body weight) button, then turn the knob
IBW
button to select new ventilator control
Touch
3
the
If you want to return to the
RESTART
The system displays the
4
following buttons, and uses the rotary knob or drop-down menus to display the available selections.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 4-4
CONTINUE.
CONTINUE
button.
(You must touch the
button will display.)
Ventilator Startup
New Patient Settings
button first before
IBW
screen, touch the
screen with the
Page 95
How to use the 840 ventilator
OP 4
Vent Type
INVASIVE
Determines the ventilation type
:
— conventional ventilation using either
endotracheal (ET) or tracheostomy (trach) tubes
non-invasive
NIV (
ventilation using full-face masks,
) —
nasal masks, infant nasal prongs, or uncuffed ET tubes (see Section 4.12 for specific information on how to use NIV)
Determines the type and sequence of breath delivery
Mode
:
(Assist/Control)
A/C
(Synchronous Intermittent Mandatory Ventilation)
SIMV
SPONT
BILEVEL
(Spontaneous)
(available only with
BiLevel
software option
when Vent Type is INVASIVE)
Mandatory Type:
Determines the type of mandatory breath
control
(Pressure Control)
PC
(Volume Control)
VC
• VC+ (Volume Control Plus available only with
Ventilation Plus (VV+
) software option when
Volume
Vent Type
INVASIVE)
is
(If the selected
Mode
is
SPONT
, the
Mandatory Type
applies to
manual inspirations only.)
Determines type of support for
Spontaneous Type
:
spontaneous breaths
(Pressure Support)
PS
(Tube Compensation available only with TC software
TC
option when Vent Type is INVASIVE)
(Volume Support available only with
VS
VV+
software
option when Vent Type is INVASIVE)
(Proportional Assist available only with
PA
PAV+
software option when Vent Type is INVASIVE)
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
NONE
OP 4-5
Page 96
OP 4
How to use the 840 ventilator
(If the selected
Mode
is
A/C,
the
Spontaneous Type
button
does not appear.)
Determines the method used to detect patient
Trigger Type
:
inspiratory effort
•V-
(Pressure) (not available when Vent Type is NIV or
-
P
TRIG
when using the
(Flow)
TRIG
NeoMode
option)
Touch the button and turn the knob to adjust the desired settings. When you complete your settings changes, touch
CONTINUE
The final
5
.
New Patient Settings
screen appears. Touch the button of each parameter you want to change, then turn the knob to select its value. To cancel this change, press the cancel all changes and start over, touch the
CLEAR
RESTART
NOTE:
The ventilator control parameter you are setting may be dependent upon other ventilator settings that determine its boundaries. Refer to the prompt area on the lower GUI screen (Figure 1-2) for more information.
Press
6
ACCEPT
to put all of your ventilation control settings into effect. Normal ventilation begins once a patient is connected.
The
7
Apnea Setup
screen appears. Apnea settings are automatically determined based on IBW, circuit type, and mandatory breath type, but you can change them. If you change any apnea settings, press
ACCEPT
to apply.
Although you are not required to change or confirm apnea settings, you should verify that they are appropriate for the patient prior to ventilation.
Press the
8
ALARM SETUP
limit settings on the
button to review the current alarm
Alarm Settings
screen. Ensure that they are appropriate for the patient. To change any limit, touch the button and turn the knob. To cancel, touch
ALARM
. To apply the settings, press the
PROPOSED
ACCEPT
key.
key. To
button.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 4-6
Page 97
How to use the 840 ventilator
You may choose to calibrate the ventilator’s oxygen sensor at
9
this point. Press the keyboard below the touch screens.
During the oxygen sensor calibration, the ventilator delivers 100% oxygen (if available) for two minutes and calibrates the oxygen sensor in the Breath Delivery Unit (BDU).
The ventilator always monitors the delivery of oxygen to the patient unless you disable the oxygen sensor. Touch the
SETTINGS
button to access oxygen sensor disable or enable
functions.
After you accept the ventilation control parameters, you can
10
attach a patient to the ventilator. Ventilation only begins when the ventilator senses that a patient is attached.
If you attach a patient before completing setup, the ventilator initiates safety ventilation mode and annunciates a
PROCEDURE ERROR
patient setup.
Warning
Each patient circuit type is appropriate for a specified range of IBW values. This information is summarized in Table 4-2.
The recommended ranges exist to ensure patient safety. Only those with expertise to judge the appropriate circumstances should override the recommended ranges.
OP 4
100% O2 / CAL 2 min
alarm that is reset once you complete the
key located on the
MORE

4.2.3 Patient data and current settings

The top of the upper screen shows vital patient data. (Out-of­range data flashes to alert you.) The current breath type is indicated in the upper left corner:
= Control
C
=Spontaneous
S
= Assist
A
You can access additional patient data when you touch the
PATIENT DATA
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
button.
MORE
OP 4-7
Page 98
OP 4
How to use the 840 ventilator
You can display the definitions for any symbol used in the patient data, alarm log, or settings areas by touching the symbol. The symbol definitions appear at the bottom of the lower touch screen.
Current ventilator control settings are displayed across the top of the lower touch screen (Figure 4-3). If you press the
key, the lower touch screen automatically displays the
2 min
PROGRESS PROGRESS
indicator. If you touch the indicator will appear if there is no other higher-
priority display active. Press the
CANCEL
Alarm Silence
button for either
100% O
key, the
/CAL
2
IN
IN
indicator to cancel the alarm silence or oxygen sensor calibration in progress.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 4-8
Page 99
How to use the 840 ventilator
Breath type (C = Control)
OP 4
Vital patient data area
Alarm area
Patient data
(upper screen)
Subscreen area
Main ventilator control settings
Ventilator settings
(lower screen)
3VTE MAND
2 of last 4 mand breaths set limit. Check for leaks, changes in patient'sR&C.
11:20 25 Jun 2000
Subscreen area
Figure 4-3. Touch screen appearance during normal ventilation (shown with
alarm silence and 100% O
4-075609-00 Rev. G (10/06) 840 Ventilator System Operator’s Manual
/CAL in progress)
2
OP 4-9
Page 100
OP 4
How to use the 840 ventilator

4.2.4 Ideal Body Weight (IBW)

The system initially sets most upper and lower alarm limits based on the patient’s IBW. After you enter the IBW, you can review and change these alarm settings, if needed. Table 4-1 below provides the information needed to determine the patient’s IBW using the patient’s height.
840 Ventilator System Operator’s Manual 4-075609-00 Rev. G (10/06)
OP 4-10
Loading...