4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s and Technical Reference Manual
Page 2
Page 3
Copyright Information
Copyright 2003, 2005, 2006 Puritan-Bennett Corporation. All rights reserved. The
TM
840
Ventilator System is manufactured in accordance with Puritan Bennett
proprietary information, covered by one or more of the following U.S. Patents and
foreign equivalents: 4,954,799; 5,161,525; 5,271,389; 5,301,921; 5,319,540;
5,339,807; 5,368,019; and 5,390,666. 840, 800 Series, DualView, SandBox, SmartAlert, Flow-by, and PTS 2000 are trademarks of Puritan-Bennett Corporation.
The information contained in this manual is the sole property of Puritan-Bennett
Corporation and may not be duplicated without permission. This manual may be
revised or replaced by Puritan-Bennett Corporation at any time and without
notice. You should ensure that you have the most current applicable version of
this manual; if in doubt, contact Puritan-Bennett Corporation or visit the
Puritan Bennett product manual web page at:
While the information set forth herein is believed to be accurate, it is not a
substitute for the exercise of professional judgment.
The ventilator should be operated and serviced only by trained professionals.
Puritan Bennett’s sole responsibility with respect to the ventilator, and its use, is as
stated in the limited warranty provided.
Nothing in this manual shall limit or restrict in any way Puritan Bennett’s right to
revise or otherwise change or modify the equipment (including its software)
described herein, without notice. In the absence of an express, written agreement
to the contrary, Puritan Bennett has no obligation to furnish any such revisions,
changes, or modifications to the owner or user of the equipment (including its
software) described herein.
Rx ONLY
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s and Technical Reference Manual
i
Page 4
Applicability
The information in this manual applies to
ventilator versions
840
manufactured or updated after August 2005. Some of this
information may not apply to earlier versions. Contact your
Puritan Bennett representative if in doubt.
Definitions
This manual uses three special indicators to convey information
of a specific nature. They include:
Warning
Indicates a condition that can endanger the patient or the
ventilator operator.
Caution
Indicates a condition that can damage the equipment.
NOTE:
Indicates points of particular emphasis that make
operation of the ventilator more efficient or
convenient.
840 Ventilator System Operator’s and Technical Reference Manual4-075609-00 Rev. G (10/06)
ii
Page 5
Warnings, cautions, and notes
Please take the time to familiarize yourself with the following
safety considerations, special handling requirements, and
regulations that govern the use of the
•To ensure proper servicing and avoid the possibility of physical
injury, only qualified personnel should attempt to service or
make authorized modifications to the ventilator.
The user of this product shall have sole responsibility for any
ventilator malfunction due to operation or maintenance
performed by anyone not trained by Puritan Bennett.
•To avoid an electrical shock hazard while servicing the
ventilator, be sure to remove all power to the ventilator by
disconnecting the power source and turning off all ventilator
power switches.
•To avoid a fire hazard, keep matches, lighted cigarettes, and all
other sources of ignition (e.g., flammable anesthetics and/or
heaters) away from the
Ventilator System and oxygen
840
hoses.
Do not use oxygen hoses that are worn, frayed, or
contaminated by combustible materials such as grease or oils.
Textiles, oils, and other combustibles are easily ignited and
burn with great intensity in air enriched with oxygen.
Ventilator System.
840
In case of fire or a burning smell, immediately disconnect the
ventilator from the oxygen supply, facility power, and backup
power source.
•When handling any part of the
Ventilator System, always
840
follow your hospital infection control guidelines for handling
infectious material.
Puritan Bennett recognizes that cleaning, sterilization,
sanitation, and disinfection practices vary widely among
health care institutions. It is not possible for Puritan Bennett
to specify or require specific practices that will meet all needs,
or to be responsible for the effectiveness of cleaning,
sterilization, and other practices carried out in the patient care
setting.
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s and Technical Reference Manual
iii
Page 6
•Patients on life-support equipment should be appropriately
monitored by competent medical personnel and suitable
monitoring devices.
The
Ventilator System is not intended to be a
840
comprehensive monitoring device and does not activate
alarms for all types of dangerous conditions for patients on
life-support equipment.
•For a thorough understanding of ventilator operations, be sure
to thoroughly read this manual before attempting to use the
system.
•Before activating any part of the ventilator, be sure to check
the equipment for proper operation and, if appropriate, run
SST as described in this manual.
•Do not use sharp objects to make selections on the graphic
user interface (GUI) display or keyboard.
•US federal law restricts this device to sale by or on the order of
a physician.
•Check the ventilator periodically as outlined in the
840 Ventilator System Service Manual
; do not use if defective.
Immediately replace parts that are broken, missing, obviously
worn, distorted, or contaminated.
•An alternative source of ventilation should always be available
when using the
Ventilator System.
840
Warranty
The
and workmanship in accordance with Puritan Bennett Medical
Equipment Warranty for 2 years from the time of sale. Keep a
maintenance record to ensure the validity of the warranty.
840 Ventilator System Operator’s and Technical Reference Manual4-075609-00 Rev. G (10/06)
iv
Ventilator System is warranted against defects in material
840
Page 7
Year of manufacture
The graphic user interface (GUI), breath delivery unit (BDU),
backup power source (BPS), and compressor contain a specific
year of manufacture applicable only for that assembly. The year of
manufacture is indicated by the fifth and sixth digits of the serial
number which is located at the back panel of the GUI, BDU, and
BPS, and the side panel of the compressor.
Manufacturer
Puritan-Bennett Corporation
4280 Hacienda Drive
Pleasanton, CA
USA
Electromagnetic susceptibility
The
IEC 60601-1-2:2004 (EMC Collateral Standard), including the Efield susceptibility requirements at a level of 10 volts per meter, at
frequencies from 80 MHz to 2.5 GHz, and the ESD requirements
of this standard.
Ventilator System complies with the requirements of
840
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s and Technical Reference Manual
v
Page 8
However, even at this level of device immunity, certain
transmitting devices (cellular phones, walkie-talkies, cordless
phones, paging transmitters, etc.) emit radio frequencies that
could interrupt ventilator operation if operated in a range too
close to the ventilator. It is difficult to determine when the field
strength of these devices becomes excessive.
Practitioners should be aware that radio frequency emissions are
additive, and that the ventilator must be located a sufficient
distance from transmitting devices to avoid interruption. Do not
operate the ventilator in a magnetic resonance imaging (MRI)
environment.
Warning
Accessory equipment connected to the power receptacle,
analog, and digital interfaces must be certified according
to IEC 60601-1. Furthermore, all configurations shall
comply with the system standard IEC 60601-1-1. Any
person who connects additional equipment to the power
receptacle, signal input part, or signal output part of the
840 ventilator configures a medical system, and is therefore
responsible for ensuring that the system complies with the
requirements of the system standard IEC 60601-1-1. If in
doubt, consult Puritan Bennett Technical Services at
1.800.255.6774 or your local representative.
This manual describes possible ventilator alarms and what to do if
they occur. Consult with your institution’s biomedical
engineering department in case of interrupted ventilator
operation, and before relocating any life support equipment.
Customer assistance
For further assistance call Puritan Bennett Technical Services at
1.800.255.6774 or contact your local Puritan Bennett
representative.
840 Ventilator System Operator’s and Technical Reference Manual4-075609-00 Rev. G (10/06)
vi
Page 9
Preface
This manual is divided into two parts: the operator’s manual and
the technical reference. The operator’s manual describes how to
operate the Puritan Bennett
provides product specifications and accessory order numbers. The
technical reference includes background information about how
the ventilator functions, including details on its operating modes,
self-tests, and other features. In the table of contents and index,
the prefix OP- identifies page numbers in the operator’s manual,
and the prefix TR- identifies page numbers in the technical
reference.
Any references to the software options
Plus
®
(VV+)
which includes VC+ and VS breath types
Proportional Assist Ventilation
that are made in this manual assume that the option has been
installed on the ventilator. If these options aren’t installed, then
references to their functions do not apply.
While this manual covers the ventilator configurations currently
supported by Puritan Bennett, it may not be all-inclusive and may
not be applicable to your ventilator. Within the USA, contact
Puritan Bennett at 1.800.255.6774 for questions about the
applicability of the information.
Ventilator System. It also
840
®
, Volume Ven t ilation
Tube Compensation (TC)
®
(PAV+),
BiLevel
and
, NeoMode®,
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s and Technical Reference Manual
vii
Page 10
This page is intentionally blank.
840 Ventilator System Operator’s and Technical Reference Manual4-075609-00 Rev. G (10/06)
840 Ventilator System Operator’s and Technical Reference Manual4-075609-00 Rev. G (10/06)
xxii
Page 25
C
HAPTER
1
Introduction
The intended use of the Puritan Bennett
for acute and subacute care of infant, pediatric, and adult patients.
Software options, available from Puritan Bennett, provide
additional ventilation functions.
The
management, offers selectable modes of breath delivery, and
assists the practitioner in the selection of the most appropriate
ventilator control parameters for the patient. The user interface is
intuitive and easy to operate for those with prior knowledge of
ventilator operation.
The user interface includes
monitored patient data, for easy assessment of the patient’s
condition. The touch screens also display the current ventilator
control parameters.
The
preview the selected ventilator control parameters prior to active
ventilation of the patient.
The
specific information about the cause, and prompts the user with
actions to resolve the reported condition(s).
Ventilator System facilitates work of breathing
840
SandBox
SmartAlert
1
Ventilator System is
840
DualView
™ area on the touch screen allows the practitioner to
™ system intercepts alarms, or events, provides
™ touch screens that display
The breath delivery unit (BDU) comprises the pneumatics and the
patient circuit.
The ventilator uses two independent Central Processing Units
(CPUs):
•Breath delivery unit (BDU) CPU
•Graphic user interface (GUI) CPU
The BDU CPU uses the ventilator control parameters, selected by
the practitioner, to deliver breaths to the patient. The BDU CPU
also runs continuous and extensive operational background
checks to ensure proper operation of the ventilator.
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 1-1
Page 26
OP 1
Introduction
The GUI CPU monitors the ventilator and the ventilator/patient
interaction. The GUI CPU also monitors the operation of the BDU
CPU and prevents simultaneous failure of control and monitor
functions when a single fault is reported.
The
breaths with a preset level of positive end-expiratory pressure
(PEEP), trigger sensitivity, and oxygen concentration. A
mandatory breath can either be pressure- or volume-controlled,
but it is always pressure-controlled in the optional
spontaneous breath allows patient inspiratory flows of up to
200 L/min, with or without pressure support.
The optional 806 Compressor unit provides compressed air to the
BDU, and can be used in place of wall or bottled air. The
compressor unit is powered through and communicates with the
BDU.
The 802 Backup Power Source (BPS) provides DC power to the
BDU and GUI in the event that AC power is lost. A new, fully
charged BPS runs the ventilator (without a compressor or a
humidifier) for at least 30 minutes, which allows transport of the
patient and the ventilator within the healthcare facility.
This manual tells you how to operate and perform simple
maintenance for the
this manual and accompanying labels before attempting to
operate or maintain the ventilator.
To ensure optimum performance of the
Puritan Bennett strongly recommends that certified biomedical
engineering technicians, or other personnel with equivalent
experience and training in the service of this type of equipment,
perform periodic maintenance on the ventilator. For more
information, contact Puritan Bennett Technical Services in the
U.S. at 1.800.255.6774. Outside of the U.S., contact your local
Puritan Bennett representative.
Ventilator System supplies mandatory or spontaneous
840
BiLevel
Ventilator System. Become familiar with
840
Ventilator System,
840
mode. A
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 1-2
Page 27
Introduction
1.1Technical description
1.1.1 General background
The practitioner uses the GUI touch screens, the off-screen keys,
and GUI knob to select the ventilator control parameters and
input data (see Figure 1-1). The GUI CPU processes this
information and stores it in ventilator memory. The BDU CPU
uses this stored information to control and monitor the flow of
gas to and from the patient. The two CPUs communicate to
transfer and verify any new ventilator control parameters or alarm
limits. Each CPU then performs continuous background
verification of operational and data integrity.
OP 1
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 1-3
Page 28
OP 1
Graphic user
interface (GUI)
Introduction
Collector
vial
Exhalation
module:
Active exhalation valve
Pressure transducer
Flow sensor
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 1-4
Page 29
Introduction
OP 1
1.1.2 Pressure and flow triggering
The ventilator uses flow or pressure triggering to recognize patient
effort. When
pressure triggering
pressure in the patient circuit. As the patient draws gas from the
circuit and airway pressure drops by at least the value selected for
pressure sensitivity, the ventilator delivers a breath.
When
flow triggering (Flow-by
monitors the difference between the inspiratory and expiratory
flow sensor measurements. As the patient inhales, the ventilator
measures less exhaled flow while the delivered flow remains
constant. The result is an increase in the difference between the
inspiratory and expiratory flows. When the difference is at least
the operator-selected value for flow sensitivity, the ventilator
delivers a breath.
If the patient is not inhaling, any difference between the delivered
and exhaled flow is due to sensor inaccuracy or leaks in the
patient system. To compensate for leaks in the patient system
which can cause autotriggering, the operator can increase the flow
sensitivity setting.
As a backup method of triggering inspiration, a pressure
sensitivity of 2 cmH
O is also in effect. This setting is the most
2
sensitive setting that is still large enough to avoid autotriggering,
yet will trigger with acceptable patient effort.
is in effect, the ventilator monitors
) is in effect, the ventilator
1.1.3 Breathing gas mixture
Air and oxygen from cylinders, wall supplies, or compressor (air
only) enter the ventilator through hoses and fittings (the fittings
are available in several versions). Once inside the ventilator, air
and oxygen are regulated to pressures appropriate for the
ventilator, then mixed according to the selected O
The ventilator delivers the mixed air and oxygen through the
inspiratory module
and out to the patient. The oxygen
concentration of the delivered gas is monitored here, using a
galvanic oxygen sensor. The galvanic sensor generates a voltage
proportional to the oxygen concentration. The ventilator reports
an alarm if the O
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
sensor is enabled and monitored oxygen
2
%.
2
OP 1-5
Page 30
OP 1
Introduction
concentration is more than seven percent above or below the
O
% setting, or below 18% after the concentration stabilizes.
2
The inspiratory manifold also includes a safety valve to relieve
patient pressure if necessary (for example, if the patient circuit is
kinked or occluded). The inspiratory module also corrects for gas
temperature and humidity, based on the practitioner-set
humidification type.
1.1.4 Inspiratory pneumatics
Ventilator inspiratory pneumatics consist of two parallel circuits:
one for oxygen and one for air. The primary elements of the
inspiratory pneumatics are two proportional solenoid valves
(PSOLs), which control the flow of gas delivered to the patient. Air
and oxygen flow sensors, along with pressure signals from the
patient circuit, provide feedback that the BDU CPU uses to
control the PSOLs.
As a result, the ventilator supplies mixed breathing gas to the
patient, based on the practitioner-set ventilator control
parameters. The mixed air and oxygen passes through the patient
circuit external to the ventilator. The system delivers the
breathing gas mixture to the patient at the patient wye, located in
the external patient circuit.
1.1.5 Patient circuit
The
patient circuit
ventilator that route gas between the ventilator and the patient.
These components include:
•an
•a humidification device (optional) in line with the patient
•the inspiratory and expiratory limbs of the patient circuit that
•a
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 1-6
inspiratory filter
between the patient and ventilator
circuit
conduct the breathing gas to and from the patient
collector vial
bulk moisture in the exhaled gas
comprises the components external to the
that protects against contamination
that protects the expiratory pneumatics from
Page 31
Introduction
OP 1
•an
expiratory filter
particulates in the patient’s exhaled gas into the room air or
inside the ventilator exhalation pneumatics
The ventilator actively controls the exhalation valve that the
software accurately positions throughout the patient’s inspiration
and exhalation. The exhalation valve allows the ventilator to
deliver aggressive breaths while pressure overshoots are
minimized, PEEP is controlled, and excess patient pressures are
relieved. The exhalation system monitors the exhaled gas leaving
the patient circuit for spirometry.
Throughout the respiratory cycle, pressure transducers monitor
inspiratory, expiratory, and atmospheric pressures. The
temperature of the exhaled gas is heated to a temperature above
its dew point to prevent condensation in the exhalation
compartment. Refer to Appendix C for a detailed diagram of the
ventilator’s pneumatic system and patient circuit.
that limits the escape of microorganisms and
1.1.6 AC mains and backup power system
The ventilator derives its power to operate from the AC mains
(wall) power or the backup power system (BPS). The design of the
BDU integral power supply protects against excessive voltages,
temperatures, or current draws. A power cord retainer prevents
accidental disconnection of the BDU from the AC mains. A power
switch cover on the front face of the BDU protects against spills
and accidental AC power-off.
The ventilator connects to the 802 BPS, which supplies DC power
to the ventilator if AC power is lost. A fully charged BPS operating
under nominal ambient conditions, can power the ventilator for
at least 30 minutes (the BPS does not power the compressor unit
or the humidifier, if present). The GUI indicates when the
ventilator is operating on the BPS, rather than AC mains.
When AC power is connected, it recharges the BPS. The BPS
continues to recharge from the AC power during normal
ventilator operation.
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 1-7
Page 32
OP 1
Introduction
1.1.7 Ventilator emergency states
Emergency states include
safety valve open (SVO)
always includes the
independent of a
The following describe the two ventilator emergency states:
•
Safety valve open (SVO)
air and oxygen supplies are lost, or an occlusion is detected, or
the ventilator enters the
The safety valve open (
room air unassisted by the ventilator. The ventilator remains
in the
SVO
state is corrected.
When the ventilator enters the
on the front face of the BDU illuminates, and a high-urgency
alarm sounds.
In case of a malfunction that prevents software from opening
the safety valve, there is also an analog circuit that opens the
safety valve if system pressure exceeds 100 to 120 cmH
•
Ventilator inoperative (Vent Inop)
ventilator inoperative condition if a hardware failure or critical
software error occurs that could compromise safe ventilation
of the patient.
Vent I n o p
state until the condition that caused the emergency
ventilator inoperative (Vent Inop)
. When a
state. A
SVO
: The ventilator enters a
Vent Ino p
SVO
condition.
Vent Inop
) state allows the patient to breathe
SVO
condition occurs, it
state can also occur
condition.
state, the
SVO
: The ventilator declares a
SVO
SVO
and
state if both
indicator
O.
2
When a
the front face of the BDU illuminates and the ventilator enters
the
If a
ventilator from use until qualified service personnel evaluate
and correct the
If the ventilator declares a
test (POST) must first verify that power levels to the ventilator
are acceptable and that the functions of the major electronics
systems are satisfactory before normal ventilation can resume.
Qualified service personnel must repair the ventilator to
correct the problem and execute EST successfully before
normal ventilation is allowed.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 1-8
Vent Inop
state, which in turns sounds a high-urgency alarm.
SVO
Vent Inop
condition occurs, the
condition occurs, immediately remove the
Vent Ino p
condition.
Vent Inop
Vent I n o p
state, the power on self
indicator on
Page 33
Introduction
1.2Graphic user interface
This section describes the graphic user interface (GUI), the GUI
keys, the GUI indicators, and the symbols you see on the GUI.
OP 1
Upper screen:
monitored
information
(alarms,
patient data)
Lower screen:
ventilator
control
parameters
The graphic user interface (GUI) of the
comprises the
DualView
touch screens, the off-screen keys located
Ventilator System
840
below the touch screens, and a knob. Use the knob to set a given
ventilator control parameter to its desired value. Press the
ACCEPT
key—the off-screen key above and right of the knob—to enter the
selected value or parameter into memory.
Figure 1-2 identifies the components of the GUI, and the location
of information on the
Vital patient data
Alarms and ventilator status
Assorted patient data,
including graphical displays
Active alarm log, if applicable
Primary patient parameters
Setup of ventilator control
parameters, alarm limits,
breath timing parameters,
and other parameters
Symbol definitions
DualView
touch screens.
Prompt
area
Status
indicators
Off-screen
keys
CLEAR
key
ACCEPT
key
Knob
8-00025
Figure 1-2. 840 Ventilator System graphic user interface (GUI)
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 1-9
Page 34
OP 1
1.3User interface controls and indicators
Descriptions of the controls and indicators on the graphic user
interface are given in Table 1-1 below.
Table 1-1: 840 Ventilator System GUI controls and indicators
Control or indicatorFunction
Screen lock key: When the yellow light on the screen lock key
is lit, the screen or off-screen controls (including the knob and
ACCEPT key) have no effect when touched until you press
the screen lock key again. New alarms automatically unlock
the screen and controls.
The screen lock allows you to clean the touch screen and
prevents inadvertent changes to settings and displays.
Alarm volume key: Allows you to adjust the alarm volume
when you hold down this key while turning the knob. You
cannot turn off the alarm volume.
Alarm silence key: Turns off the audible alarm for two
minutes. The yellow light on the alarm silence key illuminates
during the silence period. An ALARM SILENCE IN PROGRESS
indicator displays on the lower touch screen, along with a
CANCEL button, if there is not a higher-priority alarm display
active. To exit out of the alarm silence, touch the CANCEL
button.
The system automatically exits the alarm silence when the
two-minute interval times out. High-urgency alarms such as
Device Alerts, Safety Valve Open, Occlusion, and loss of either
gas supply cancel the alarm silence.
Each time you press the alarm silence key, the silence period
resets to two minutes. Each time you press the alarm silence
key (whether or not there is an active alarm), the keypress is
recorded in the alarm log.
Introduction
Alarm reset key: Clears active alarms or resets high-urgency
alarms and cancels an active alarm silence. Each time you
press the reset key, it is recorded in the alarm log, if there is an
active alarm. You cannot reset a DEVICE ALERT alarm.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 1-10
Page 35
Introduction
OP 1
Table 1-1: 840 Ventilator System GUI controls and indicators (cont)
Control or indicatorFunction
Information key: Displays basic operating information about
the ventilator. Press the key to display a menu of information
topics, then touch the button corresponding to the desired
topic. Browse topical information using the , ,
, and buttons located in the information header.
GO BACK
Oxygen sensor calibration key: Delivers 100% oxygen (if
available) for two minutes and calibrates the oxygen sensor.
The green light on this key illuminates and a message (100%
Cal in Progress) on the lower touch screen indicates that
O
2
100% O
delivery is active. If you press the O2 key again, the
2
system restarts the two-minute delivery interval. Press
CANCEL to stop the calibration.
Use the procedure in Appendix D.2 to test the oxygen sensor
calibration.
Manual inspiration key: In A/C, SIMV, and SPONT modes,
delivers one manual breath to the patient in accordance with
the current mandatory breath parameters. In BILEVEL mode,
transitions from Low PEEP (PEEP
) to High PEEP (PEEPH) (or
L
vice versa). To avoid breath stacking, a manual inspiration is
not delivered during inspiration or during the restricted phase
of exhalation.
You can use the MANUAL INSP key to supplement minute
volume or to assist measurement of a patient data parameter,
such as peak inspiratory pressure, or to run an INSP PAUSE
maneuver in SPONT mode.
DIAGRAM
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 1-11
Page 36
OP 1
Introduction
Table 1-1: 840 Ventilator System GUI controls and indicators (cont)
Control or indicatorFunction
Expiratory pause key: Causes the ventilator to seal the
patient’s breathing circuit when the expiratory phase of a
designated breath, mandatory or spontaneous, is followed by
a time-cycled mandatory inspiration. An expiratory pause is
used to estimate PEEP
The ventilator performs two types of pause maneuver:
automatic, which you initiate by a momentary press of the
EXP PAUSE
continuous press on the key. An automatic pause performs the
maneuver until the pressure stabilizes, then takes its
measurements. The pause lasts at least 0.5 second and does
not exceed 3.0 seconds.
During a manual pause, the ventilator takes its measurements
as soon as the pressure stabilizes or the pause ends. The
ventilator continues the maneuver until you release the
PAU SE key. The pause cannot exceed 20 seconds.
Section 4.9 describes, in detail, how to use the EXP PAUSE
key.
key, and manual, which you control by a
and PEEPI (autoPEEP).
TOT
EXP
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 1-12
Page 37
Introduction
OP 1
Table 1-1: 840 Ventilator System GUI controls and indicators (cont)
Control or indicatorFunction
Inspiratory pause key: Causes the ventilator to seal the
patient’s breathing circuit at the conclusion of the gas delivery
phase of a designated, volume- or pressure-based mandatory
inspiration. This inspiratory pause maneuver provides a means
to measure the patient’s static lung-thoracic compliance
), static resistance (R
(C
STAT
The inspiratory pause maneuver maintains the inflated state
of the lungs.
The ventilator performs two types of pause maneuver:
automatic, which is initiated by the momentary press of the
INSP PAUSE
continuous press on the key.
An automatic pause performs the maneuver until the pressure
stabilizes, then the system takes its measurements. The pause
event lasts at least 0.5 second but no longer than
2.0 seconds.
In a manual pause, the maneuver continues until you release
the INSP PAUSE key, but cannot exceed 7 seconds. The
ventilator computes C
and displays the values at the end of the maneuver. P
computed and updated continuously during the plateau, and
its value is frozen at the end of the plateau. Section 4.10
describes, in detail, how to use the INSP PAUSE key.
key, and manual, which you control by a
STAT
), and plateau pressure (PPL).
STAT
and R
at the end of the plateau
STAT
is
PL
Knob: Adjusts the value of a setting. A highlighted button on
a touch screen means that the knob is linked to that setting.
Where applicable, a clockwise turn of the knob increases the
highlighted value, and a counterclockwise turn of the knob
decreases the highlighted value.
Clear: Cancels a proposed ventilator parameter value change.
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 1-13
Page 38
OP 1
Introduction
Table 1-1: 840 Ventilator System GUI controls and indicators (cont)
Control or indicatorFunction
Accept: Applies and saves new ventilator parameter value(s).
Red high-urgency alarm indicator ( ! ! ! ): This alarm
indicator blinks rapidly if active; it is steadily lit if autoreset.
Yellow medium-urgency alarm indicator ( ! ! ): This alarm
indicator blinks slowly if active; it turns off if autoreset.
Yellow low-urgency alarm indicator ( ! ): This indicator is
steadily lit if active; it turns off if autoreset.
Green normal ventilator operation indicator: When
ventilation is active and no alarm states exist, this indicator is
steadily lit. This indicator is off if the ventilator is not in a
ventilation mode, for example, during service mode or short
self test (SST).
VENT
INOP
SAFETY
VALVE
OPEN
BATTERY
READY
BATTERY
ON
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 1-14
Red ventilator inoperative (VENT INOP) indicator: The
ventilator cannot support ventilation and requires service. The
ventilator enters the safe state (safety ventilation) and
discontinues detection of new patient data or alarm
conditions. Qualified service personnel must repair the
ventilator to correct the problem and execute EST successfully
before normal ventilation is allowed. This indicator is
accompanied by an audio signal and cannot be reset.
Red safety valve open (SVO) indicator: The ventilator has
entered its safe state and opened its safety valve to allow the
patient to breathe unassisted from room air.
Green BPS ready indicator: The ventilator senses that the
BPS is installed, operational, and has at least two (2) minutes
of estimated run time.
Battery on power indicator: When the yellow bar to the
right of a lit BPS ready indicator is lit, the ventilator is
operating on BPS, and AC power is insufficient to support
ventilator operation. During BPS operation, power to the
compressor unit and the humidifier outlet is off.
Page 39
Introduction
OP 1
Table 1-1: 840 Ventilator System GUI controls and indicators (cont)
Control or indicatorFunction
COMPRESSOR
READY
COMPRESSOR
ON
The indicators on the breath delivery unit are shown in Table 1-2.
VENT
INOP
SAFETY
VALV E
OPEN
Green compressor ready indicator: The compressor cable
and air supply hose are connected to the ventilator. The
compressor is up to operating pressure but not supplying gas
to the ventilator. The compressor motor turns on
intermittently to keep the compressor chamber pressurized.
Green compressor operating indicator: When lit,
compressor is supplying air to the ventilator. This indicator
does not light unless the compressor is actually supplying air
to the ventilator.
Table 1- 2: BDU in dicators
Red ventilator inoperative (VENT INOP) indicator: The
ventilator cannot support ventilation and requires service. The
ventilator enters the safe state (safety ventilation) and
discontinues detection of new patient data or alarm
conditions. Qualified service personnel must repair the
ventilator to correct the problem and execute EST successfully
before normal ventilation is allowed. This indicator is
accompanied by an audio signal and cannot be reset.
Red safety valve open (SVO) indicator: The ventilator has
entered its safe state and opened its safety valve to allow the
patient to breathe unassisted from room air.
DISPLAY
(GUI)
INOP
Red loss of GUI indicator: The ventilator has detected a
malfunction that prevents the GUI from reliably displaying or
receiving information.
1.3.1 Onscreen symbols and abbreviations
Touch an onscreen symbol to display its definition in the lower
left corner of the lower screen. Table 1-3 summarizes the symbols
and abbreviations the ventilator uses.
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 1-15
Page 40
OP 1
For example, if you touch:
V
MAX
21.8
The symbol definition area shows this message:
V
MAX
Table 1-3: 840 Ventilator System symbols and abbreviations
L
min
= Peak flow
Introduction
Symbol or
abbreviation
(blinking)
P
%
Definition
Additional alarms that relate to the monitored information are
active. The symbol blinks when there is not enough screen
area to display all active alarms.
The upper alarm limit
The lower alarm limit
Press to access the alarm log
Alarm log contains events that you have not yet viewed
Rise time percent
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 1-16
Page 41
Introduction
OP 1
Table 1-3: 840 Ventilator System symbols and abbreviations (cont)
Symbol or
abbreviation
RAMP SQUARE
Definition
Flow pattern
The value you selected for a ventilator control parameter
exceeds its recommended limit (soft bound) and requires
acknowledgement to continue
or
The value selected exceeds its allowable minimum or
maximum limit (hard bound)
Press to view more patient data
Press to view patient data graphics
Press to view additional screens
X-axis (time or pressure) adjust of patient data graphics
Y-axis (pressure, volume, or flow) adjust of patient data
graphics
Baseline pressure (PEEP) adjust
A/CAssist control ventilation mode
AVApnea ventilation
C
STAT
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
Static compliance
OP 1-17
Page 42
OP 1
Introduction
Table 1-3: 840 Ventilator System symbols and abbreviations (cont)
Symbol or
abbreviation
E
SENS
Spontaneous expiratory sensitivity percentage
Definition
ESTExtended Self Test
fRespiratory rate (ventilator control parameter)
↑ f
f
TOT
TOT
Total respiratory rate (monitored)
High respiratory rate alarm
GUIGraphic user interface
HMEHeat-Moisture Exchanger
I:EInspiratory to expiratory ratio
O
2
O
2
1
O2%High delivered O2% alarm
3
O2%Low delivered O
Monitored oxygen percentage (patient data)
Oxygen percentage (ventilator control parameter)
% alarm
2
PCPressure control (mandatory breath type)
P
↑P
2 P
3
4
P
MEAN
PEAK
PEAK
P
PEAK
P
PEAK
PEAK
Mean circuit pressure
High circuit pressure alarm
High circuit pressure alarm limit
Low circuit pressure alarm
Low circuit pressure alarm limit
Peak circuit pressure (patient data)
PEEPpositive end-expiratory pressure (ventilator control
parameter)
PEEP
H
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
High PEEP (ventilator control parameter, BILEVEL mode only)
OP 1-18
Page 43
Introduction
OP 1
Table 1-3: 840 Ventilator System symbols and abbreviations (cont)
Symbol or
abbreviation
PEEP
I
PEEP
L
PEEP
TOT
PEEP
P
I
P
I END
P
PL
Intrinsic PEEP (patient data)
Low PEEP (ventilator control parameter, BILEVEL mode only)
Total PEEP (patient data)
End expiratory pressure (patient data)
Inspiratory pressure (ventilator control parameter)
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
Apnea interval
Expiratory time
High PEEP time (BILEVEL mode only)
Inspiratory time
High spontaneous inspiration time alarm
OP 1-19
Page 44
OP 1
T
Introduction
Table 1-3: 840 Ventilator System symbols and abbreviations (cont)
Symbol or
abbreviation
2
T
I SPONT
T
L
T
PL
V
E SET
High spontaneous inspiration time alarm limit
Low PEEP time (BILEVEL mode only)
Plateau time
Set minute volume (calculated from ventilator control
Definition
parameters)
V
E SPONT
1
V
3V
E TOT
E TO
Exhaled spontaneous minute volume
High exhaled minute volume alarm
Low exhaled minute volume alarm
VCVolume control (mandatory breath type)
V
MAX
V
SENS
V
1
V
T
TE
V
TE
Peak flow (ventilator control parameter)
Flow sensitivity
Tidal volume (ventilator control parameter)
Exhaled tidal volume
High exhaled tidal volume alarm
3
V
TE MAND
3
V
TE SPONT
1
V
TI
V
TI
Low exhaled mandatory tidal volume alarm
Low exhaled spontaneous tidal volume alarm
Inspired tidal volume
High inspired (mandatory or spontaneous) tidal volume
alarm*
V
TI MAND
1
V
TI MAND
V
TI SPONT
1
V
TI SPONT
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
Inspired mandatory tidal volume
High inspired mandatory tidal volume alarm*
Inspired spontaneous tidal volume
High inspired spontaneous tidal volume alarm*
OP 1-20
Page 45
Introduction
OP 1
Table 1-3: 840 Ventilator System symbols and abbreviations (cont)
Symbol or
abbreviation
-TRIG
V
*Refer to Section 13.10 for information regarding inspired tidal volume alarms.
Flow triggering
Definition
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 1-21
Page 46
OP 1
1.4Ventilator system labeling symbols
The following symbols appear on the various components of the
Ventilator System.
840
NOTE:
All labels shown are examples, and may not reflect the exact
configuration of your ventilator.
Power switch positions:ON represents the power on position
andrepresents the power off position. The power switch,
located on the BDU front panel, turns ON/OFF the BDU and
8-10006
8-00400
the GUI. When the power switch is in the off position, the BPS
continues to charge if AC power is present.
Refer to manual: When this symbol appears on the product, it
means refer to documentation for information.
Type B equipment, per IEC 60601-1
Introduction
8-00445
Potential equalization point (ground): Provides a means of
connection between the equipment and the potential
equalization busbar of the electrical connection. A common
grounding point for the entire ventilator.
8-00426
Indicates the degree of protection provided by enclosure
(drip-proof)
8-00432
CSA certification mark that signifies the product has been
evaluated to the applicable ANSI/Underwriters Laboratories
Inc. (UL) and CSA standards for use in the US and Canada.
SNSerial number
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 1-22
Page 47
Introduction
Data KeyData key connection
OP 1
BPS charging indicator: When the ventilator is operating on
mains power, the top symbol (green indicator next to gray
battery icon) on the front of the BPS indicates that the BPS is
charged, and the bottom symbol (yellow indicator next to gray
battery icon) on the front of the BPS indicates that the BPS is
charging.
8-00462
Caution
Do not remove the data key. The data key enables
software options, and stores ventilator operational
hours, compressor unit operational hours, and the
serial numbers for the BDU and GUI. The ventilator
will not operate without its factory-installed data
key.
TEST
PTS 2000
DISPLAY
(GUI)
Ventilator
circuit
breaker
Compressor &
humidifier
circuit
breaker
8-00405
Compressor outlet:
5.6 A max
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
TEST button. After you touch the Short Self Test (SST)
onscreen key (available only during ventilator startup), you
must press the TEST button within 5 seconds in order to access
SST.
Puritan Bennett PTS 2000™ Performance Test System
connection, for use by qualified service personnel only.
GUI connection
Circuit breaker for ventilator power supply, located in the BDU
Ventilator circuit breaker for compressor and humidifier
Alternating current (at AC inlet and AC power indicator)
Maximum allowed output to auxiliary mains socket
(compressor electrical connection)
OP 1-23
Page 48
OP 1
Introduction
BPS electrical connection
8-00413
Exhalation filter latch unlock/lock
8-00434
Exhalation filter latch open indicator: This red indicator is
located on the surface behind the closed latch, and is easily
visible when the latch is open.
GUI mounting latch unlock/lock
8-00428
Remote alarm port
IOIOI
RS-232 port
Susceptible to electrostatic discharge
Electric shock hazard
8-00443
Explosion hazard
8-00450
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 1-24
Page 49
Introduction
Fire hazard
8-00417
802 BPS product information label
GUI product information label
OP 1
US/75427C
US/75422C
GUI ports label
Remote alarm and RS-232 port (9.4-inch GUI only). Refer to Appendix E for GUI remote
alarm and RS-232 port specifications.
ENGLISH/75416A
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 1-25
Page 50
OP 1
Humidifier electrical label
(This label not visible unless cover plate over humidifier electrical connection is
removed.)
BDU gas inlet label
Introduction
US/75415A
US/75421A
BDU to patient label
US/75417A
Compressor gas connection label
US/75418A
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 1-26
Page 51
Introduction
Compressor information label
OP 1
76505C
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 1-27
Page 52
OP 1
BDU information label
Introduction
US/75423C
BDU cooling vent label
US/75414A
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 1-28
Page 53
Introduction
BDU I/O disconnect label
BDU exhaust information label
OP 1
US/75419A
US/75420A
BPS electrical connection label
US/75334Lbl
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 1-29
Page 54
OP 1
Compressor lint filter label
Expiratory limb connector on exhalation filter
From patient
Introduction
US/75426A
8-00423
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 1-30
Page 55
C
HAPTER
2
How to set up the 840 ventilator
Chapter 2 describes how to set up the ventilator:
•How to connect the electrical supply
•How to connect the air and oxygen supplies
•How to connect the patient circuit and accessories
A Puritan Bennett Customer Service Engineer (CSE) must first
install the ventilator and run an extended self test (EST), which
calibrates the exhalation valve, flow sensors, and atmospheric
pressure transducer, before you connect a patient to the ventilator
for the first time.
Warning
• When you lift the ventilator, use assistance and
appropriate safety precautions. Figure 2-1 shows the
proper technique to lift each ventilator component.
• To avoid interrupted ventilator operation or possible
damage to the ventilator, always use the ventilator on a
level surface in its proper orientation.
• To avoid the possibility of injury to the patient and
ensure proper ventilator operation, do not attach any
device to the port labeled EXHAUST unless the device is
specifically authorized by Puritan Bennett.
• To minimize the increased risk of fire due to an oxygenenriched environment, do not use the ventilator in a
hyperbaric chamber.
• To avoid raising the oxygen concentration of room air,
use the ventilator in an adequately ventilated room.
2
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 2-1
Page 56
OP 2
Lift the BDU
from horizontal
surfaces as shown.
Use two people to lift
the compressor from
the base and the handles.
How to set up the 840 ventilator
Lift the GUI from
the base and the handle.
8-00011
Figure 2-1. How to lift the ventilator components
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 2-2
Page 57
How to set up the 840 ventilator
Caution
• Do not connect or disconnect the ventilator’s graphic user
interface (GUI), backup power source (BPS), or compressor
while the power switch is on or the ventilator is connected to AC
power.
• All components must be securely mounted and connected by
qualified service personnel according to the appropriate
Puritan Bennett installation instructions.
• Do not obstruct the breath delivery unit (BDU), GUI, or
compressor cooling vents or fan vents.
• To avoid possible damage to ventilator components, do not use
the horizontal surfaces of the ventilator to place or stack
objects.
NOTE:
Before you use the ventilator for the first time, wipe the
ventilator exterior clean and sterilize its components
according to the instructions in Chapter 7 of this manual.
Follow your institution’s protocol for cleaning and sterilizing
the ventilator and its components.
OP 2
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 2-3
Page 58
OP 2
How to set up the 840 ventilator
2.1How to connect the electrical supply
Warning
• To minimize the risk of electrical shock, always connect
the ventilator power cord into a grounded AC power
outlet.
• In the US, always connect the ventilator to an AC
receptacle marked “Hospital Only” or “Hospital Grade”
to ensure proper grounding of the ventilator.
• The 802 BPS must always be installed. Without the BPS,
the ventilator is not protected against low or lost AC
power. Do not use the ventilator unless a BPS with at
least minimal charge is installed.
• When possible, connect the ventilator to an outlet
connected to the hospital emergency back-up power
system. Refer to Section A.4 for ventilator electrical
specifications.
Normally the
802 BPS operates the ventilator when AC power is lost or drops
below a minimum level.
A new, fully charged BPS can operate the ventilator (without the
compressor or a humidifier) for at least 30 minutes; allowing the
ventilator to be used for transport purposes within the healthcare
facility.
Warning
The 802 BPS is intended for short-term use only, and is not
intended as a primary alternative power source. The BPS is
intended to power the BDU and GUI only. In case of AC
power loss, power is not available to run either the
compressor or the humidifier.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 2-4
Ventilator System is mains-powered. The
840
Page 59
How to set up the 840 ventilator
If you turn on the ventilator after it has been unplugged for an
extended period, the
occurs, recharge the 802 BPS by leaving it connected to a
ventilator connected to AC power for up to 8 hours (ventilator
does not need to be turned on). If, after turning the ventilator
back on, the
INOPERATIVE BATTERY
LOW BATTERY
must replace the BPS battery.
Figure 2-2 shows how to connect the power cord to AC power.
Built-in power cord retainer tabs protect against accidental
disconnection. Ensure that the power cord is securely fastened
into the AC receptacle prior to operation. To remove the cord,
squeeze the tabs on the top and bottom of the plug and pull
outward.
OP 2
LOW BATTERY
alarm is still active or if the
alarm is active, qualified service personnel
alarm may sound. If this
Power
cord
Power cord retainer tabs.
To AC power
8-00002
Squeeze tabs and pull outward to
disconnect power cord.
Figure 2-2. How to connect the ventilator power cord
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 2-5
Page 60
OP 2
How to set up the 840 ventilator
Figure 2-3 shows the power switch and AC indicator. When
illuminated, the AC indicator indicates that the ventilator is
receiving AC power and that the 802 BPS will be recharged as
needed. The AC indicator is independent of the power switch, and
the power switch does not turn off AC power to the ventilator
power supply. When both the power switch and AC indicator are
on, power is available for the humidifier and compressor.
Ventilator
power supply
circuit breaker
ON
Ventilator
power switch
AC panel
AC indicator
Humidifier
and compressor
circuit breaker
Potential
equalization
(ground
point)
Compressor
connection
Figure 2-3. Ventilator power switch, AC indicator, and AC panel
If the ventilator power supply circuit breaker (located on the
ventilator's AC panel, Figure 2-3) opens but AC power is still
present and the ventilator is operating on BPS, power is still
available to the humidifier and compressor connectors (although
ventilator software disables compressor operation).
AC power
connection
8-00045
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 2-6
Page 61
How to set up the 840 ventilator
When the power cord is not in use, wrap the power cord around
the hook on the back of the cart for convenient storage
(Figure 2-4).
OP 2
8-00004
Figure 2-4. Power cord storage on the cart
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 2-7
Page 62
OP 2
How to set up the 840 ventilator
2.2How to connect the air and oxygen supplies
The
or wall supplies. Follow these steps to connect the air and oxygen
supplies:
1
2
Ventilator System can use air and oxygen from cylinder
840
Ensure that the supply pressures are 35 to 100 psi
(241 to 690 kPa).
Connect the supply hoses to the inlet connectors at the rear of
the ventilator (see Figure 2-5).
Warning
• Connect only air to the air inlet, and only oxygen to the
oxygen inlet. Do not attempt to switch air and oxygen or
connect any other gas.
• Always connect at least two gas sources to the ventilator
to ensure that a constant gas supply is available to the
patient. There are three gas source connections: the
compressor, air inlet, and oxygen inlet.
• Do not use conductive, high-pressure supply hoses.
Doing so may negatively affect the ventilator’s ground
isolation characteristics.
• Use only high-pressure supply hoses recommended by
Puritan Bennett. Other hoses may be restrictive and may
cause improper ventilator operation.
Caution
To prevent damage to the ventilator, ensure that the connections
to the air and oxygen supplies are clean and unlubricated, and
that there is no water in the air or oxygen supply gas.
If you suspect water in the air supply gas, use an external wall air
water trap to prevent water damage to the ventilator or its
components.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 2-8
Page 63
How to set up the 840 ventilator
NOTE:
When you connect a pressurized air or oxygen
source, the ventilator air and oxygen regulators
have a maximum bleed rate of 3 L/min, even when
the ventilator is not in use. Always take this bleed
rate into account when calculating air and oxygen
usage.
When the air and oxygen hoses are not in use, you can wrap them
around the hook on the back of the cart for convenient storage
(Figure 2-5).
Air inlet
connector
Oxygen
inlet
connector
OP 2
Oxygen hose
(from oxygen
supply)
Air hose
(from air
supply)
8-00005
Figure 2-5. How to connect the air and oxygen supplies
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 2-9
Page 64
OP 2
How to set up the 840 ventilator
2.3How to connect the patient circuit components
Warning
• To minimize the risk of bacterial contamination or
component damage, inspiratory and expiratory filters
must always be handled with care and connected to the
ventilator during use.
• To minimize the risk of patient injury, use only patient
circuits qualified for use in oxygen-enriched
environments with the 840 Ventilator System. Do not
use antistatic or electrically conductive tubing. To ensure
a leak-tight connection, only use connectors and tubes
with ISO-standard cone and socket fittings (or use
adapters to connect barbed cuff fittings to ISO-standard
fittings).
• If you use an external, pneumatically-powered nebulizer
with the 840 ventilator, it adds flow to the patient circuit
and can adversely affect spirometry, delivered O
delivered tidal volumes, and breath triggering.
Additionally, aerosolized particulates in the ventilator
circuit can lead to an increase in exhalation filter
resistance.
• Use one of the patient circuits listed in Appendix B to
ensure that the maximum pressure/flow values specified
by IEC 60601-2-12:2001 are not exceeded (see Table A-11
on page OP A-18 for patient circuit testing
specifications). Using a circuit with a higher resistance
does not prevent ventilation, but can cause a short self
test (SST) fault or compromise the patient’s ability to
breathe through the circuit.
2
%,
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 2-10
Page 65
How to set up the 840 ventilator
NOTE:
• Puritan Bennett recommends that you run SST
(Short Self Test) every 15 days, and between
patients, and when you change the patient
circuit (particularly when you change the circuit
type, for example, from adult to pediatric or
neonatal).
• Puritan Bennett recognizes that the protocol for
running SST varies widely among health care
institutions. Puritan Bennett does not specify or
require specific practices that will meet the needs
of all institutions, nor is Puritan Bennett
responsible for the effectiveness of institutional
practices.
2.3.1 How to select and connect a patient circuit
Use low-compliance patient circuits to ensure optimum
compliance compensation, and use pediatric patient circuits
when the patient ideal body weight (IBW) is greater than 7 kg
(15 lb) but less than or equal to 24 kg (53 lb). Use the
software option and neonatal patient circuits for patients whose
IBW is less than or equal to 7 kg.
For patients whose IBW is less than or equal to 24 kg, the
compliance compensation volume limit is four times the set tidal
volume, in addition to the set tidal volume. To avoid activating a
severe occlusion alarm, only use neonatal patient circuits with the
NeoMode
Table 2-1 shows IBW values and patient circuit types. The
“Allowed but not recommended” ranges require an override.
software option.
OP 2
NeoMode
Warning
Recommended ranges exist to ensure patient safety. Only
those with the expertise to judge the appropriate
circumstances should override the recommended ranges.
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 2-11
Page 66
OP 2
How to set up the 840 ventilator
Table 2-1: Patient circuit and IBW values
RecommendationIdeal body weight (IBW) in kg (lb)
RecommendedNeonatal: 0.5-7.0 kg (1.1-15 lb)
Pediatric: 7.0-24 kg (15-53 lb)
Adult: 25-150 kg (55-330 lb)
Allowed but not
recommended
Neonatal: Not applicable
Pediatric: 3.5-6.5 kg (7.7-14.3 lb), and
25-35 kg (55-77 lb)
Adult: 7-24 kg (15-53 lb)
Figure 2-6 shows how to connect the patient circuit, including the
inspiratory filter, humidifier (if used), inspiratory limb, patient
wye, expiratory limb, collector vial, and expiratory filter.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 2-12
Page 67
How to set up the 840 ventilator
(From patient)
Expiratory
filter
Expiratory limb
Patient wye
of patient circuit
Collector
vial
OP 2
(To patient)
Inspiratory
filter
Tubing
Inspiratory limb
of patient circuit
Humidifier
8-00006
Figure 2-6. How to connect the patient circuit
Warning
To ensure that all patient circuit connections are leak-tight,
always perform a circuit leak test by running SST each time
you install the expiratory filter on the ventilator.
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 2-13
Page 68
OP 2
How to set up the 840 ventilator
Warning
Adding accessories to the ventilator can increase system
resistance. Ensure that any changes to the recommended
ventilator circuit configurations do not exceed the specified
values for inspiratory and expiratory resistance
(Appendix A). If you add accessories to the patient circuit,
always run SST to measure circuit compliance before you
begin ventilation of the patient.
2.3.2 How to install the expiratory filter and
collector vial
Install the expiratory filter and collector vial as follows:
Place the expiratory filter latch in the up position (see
1
Figure 2-7).
Slide the expiratory filter into the housing area with the
2
expiratory limb connection facing you.
Push the expiratory filter latch down; it will position the filter
3
properly.
Attach the expiratory limb of the patient circuit to the filter’s
4
expiratory limb connection.
If you do not use a drain bag, be sure to cap the collector vial drain
port on the expiratory filter (Figure 2-8).
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 2-14
Page 69
How to set up the 840 ventilator
Pull latch up to install
filter; push down to
hold filter and collector
vial in place
Expiratory
filter
Expiratory limb
connection
(from patient)
Figure 2-7. How to install the expiratory filter and collector vial
If you use a drain bag:
Install the expiratory filter. (Refer to the instructions above.)
1
Install the clamp on the drain bag tubing, making sure that
2
the clamp is closed.
Uncap collector vial drain port at the base of the collector vial.
3
Connect the collector bag tubing to the vial drain port.
4
Connect the other end of tubing to drain bag.
5
If the ventilator is mounted on the cart, place the drain bag in
6
the cart drawer (Figure 2-8).
Filter
housing
hous
area
Collector vial
OP 2
Slide the filter rim
onto these tracks
8-00018
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 2-15
Page 70
OP 2
Drain bag
Tubing
Place the drain bag
in the cart drawer
Clamp
How to set up the 840 ventilator
The collector vial
drain port must
be capped if you
do not use a
drain bag
8-00007
Figure 2-8. How to use the collector vial with or without the drain bag
NOTE:
• The drain bag is designed to lie flat and should
not be suspended.
• Check the inspiratory and expiratory limbs of the
patient circuit, the collector vial, and the in-line
water traps regularly for water buildup. Under
certain conditions, they can fill quickly. Empty
and clean the collector vial and in-line water
traps as necessary.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 2-16
Page 71
How to set up the 840 ventilator
2.3.3 How to install the flex arm
The flex arm supports the patient circuit between the ventilator
and the patient. Figure 2-9 shows how to install the flex arm onto
one of the two threaded sockets on the ventilator cart.
Flex arm
Threaded
socket
OP 2
Threaded socket
in the cart
(one of two)
8-00008
Figure 2-9. How to install the flex arm
Caution
Use only the cart handles to move the ventilator. Do not pull or
push the ventilator with the flex arm.
Flex arm replacement parts can be found in the
840 Ventilator System Service Manual
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
.
OP 2-17
Page 72
OP 2
How to set up the 840 ventilator
2.3.4 How to install the humidifier
An electrical outlet for a humidifier is located on the front of the
BDU. Figure 2-10 shows how to install a Fisher & Paykel
humidifier onto the ventilator.
Warning
• When using a Fisher & Paykel humidifier with the
840 ventilator, use the appropriate Fisher & Paykel
humidifier chambers for adult, pediatric, and neonatal
patients.
• Take proper precautions to prevent water/condensate
from splashing into the patient circuit during circuit
disconnects and high peak flow rate conditions.
• To avoid possible patient injury or damage to the
ventilator system, follow your institution’s protocol for
proper patient circuit condensate management.
Caution
• Qualified service personnel must first install the humidifier
mounting hardware.
• To avoid equipment damage to the ventilator due to liquid
ingress:
- Install the plug cover when the humidifier is plugged into the
ventilator.
- Install the flat cover plate over the humidifier electrical outlet on
the front of the BDU when the humidifier is not plugged into the
ventilator.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 2-18
Page 73
How to set up the 840 ventilator
NOTE:
• To ensure uninterrupted ventilator operation, do
not install a humidifier whose maximum current
capabilities exceed 2.3 A, with a maximum
power consumption of 270 VA.
• When you install a Fisher & Paykel humidifier,
make sure the humidifier has a right-angle
electrical plug. A short power cord is preferable.
•To ensure that ventilator occlusion detection
operates properly, do not use Puritan Bennett
Cascade humidifiers with the 840 Ventilator
System.
• If you have further questions about humidifiers
qualified for use with the 840 Ventilator System,
contact Technical Services or your local
Puritan Bennett representative.
OP 2
BDU
Plug cover
Humidifier
Mounting
bracket on
front of
ventilator
8-00138
Figure 2-10. How to install the humidifier
(Fisher & Paykel version shown)
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 2-19
Page 74
OP 2
How to set up the 840 ventilator
2.3.5 How to use the ventilator cart
Figure 2-11 shows how to lock and unlock the cart’s front wheels.
Warning
To avoid interrupted ventilator operation or damage to
ventilator components, use the cart to move the ventilator.
Do not use the cables, the power cord, or patient circuit
components to push or pull the ventilator.
Press down
to lock
Press down
to unlock
8-00010
Unlocked positionLocked position
Figure 2-11. How to lock and unlock the cart’s front wheels
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 2-20
Page 75
C
HAPTER
3
How to run Short Self Test (SST)
Chapter 3 tells you:
•When to run SST
•Required equipment for SST
•How to set up and run SST
•The SST tests and their functions
•How to understand the results of SST
3
3.1Introduction to SST
SST uses an internal, programmed sequence of tests to:
•Verify proper function of the flow and pressure sensors
•Check the patient circuit for gas leaks
•Measure the expiratory filter resistance
•Measure patient circuit resistance
•Measure patient circuit compliance
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 3-1
Page 76
OP 3
How to run Short Self Test (SST)
SST requires approximately three minutes to complete.
Warning
• Always disconnect the ventilator from the patient before
you run SST. If you run SST while the ventilator is
connected to the patient, physical injury to the patient
may occur.
• An ALERT reported by SST indicates that the ventilator or
a related component has a defect. Repair the ventilator
or related component before you use the ventilator on a
patient, unless you can determine with certainty that the
defect cannot create a hazard for the patient, or add to
the risks that may occur from other hazards.
• When you run SST, configure the patient circuit exactly
as it will be used on the patient (for example, with same
accessories). If you add accessories to the patient circuit
after you run SST, you must rerun SST with the new
accessories installed before you begin to ventilate the
patient.
3.2When to run SST
NOTE:
Puritan Bennett recognizes that health care institutions may
have their own ventilator protocols. However, Puritan
Bennett is not responsible for the effectiveness of any
institution’s protocols. Nor can Puritan Bennett specify, or
require, specific practices that will meet the internal needs of
every health care institution.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 3-2
Page 77
How to run Short Self Test (SST)
Puritan Bennett recommends that you run SST when one or more
of the events below occurs:
•When you replace the patient circuit and the exhalation filter
after 15 days of use
•When you are ready to connect a new patient to the ventilator
•When you connect a different patient circuit to the ventilator
•When you install a new or sterilized expiratory filter
•When you change the patient circuit type
•When you change the humidification device type
•When you remove or add accessories to the patient circuit,
such as a humidifier, water trap, or drain bag
OP 3
Use SST at any time, provided that a patient is
ventilator, to:
•Check the patient circuit for gas leaks
•Calculate patient circuit compliance and resistance
•Calculate expiratory filter resistance
After SST begins, the system prompts you to prepare the ventilator
to conduct certain tests. The system waits indefinitely at a prompt
until you take action and respond appropriately.
attached to the
not
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 3-3
Page 78
OP 3
How to run Short Self Test (SST)
3.3SST components and requirements
When you conduct SST, you must have available the components
and equipment that you will use on the patient:
•Patient tubing
•Expiratory filter and collector vial
•Inspiratory filter
•Humidifier, as applicable
•Other accessories (e.g. water traps, drain bag), as applicable
Additional requirements include:
•A No. 1 rubber stopper to block the airway at the patient wye
•Two gas sources (air and oxygen) connected to the ventilator
•Each gas source pressure must be between 35-100 psi
(241 to 690 kPa)
Caution
• To prevent SST failures due to leaks, ensure that any circuit
components such as collector vial drain port cap (if not using a
drain bag), the seal between the expiratory filter and collector
vial, and water trap (if used) seals are properly installed.
• If you are using a drain bag, ensure that the tubing is properly
installed on the collector vial drain port and that the tubing is
clamped. If the drain bag tubing is not clamped during SST,
large leaks and large compliance values are possible which may
cause SST to report ALERTs or FAILURES.
Wait at least 10 minutes after you turn on the ventilator before
you run SST. The warm up time of 10 minutes will stabilize the
ventilator and ensure the accuracy of the SST tests.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 3-4
Page 79
How to run Short Self Test (SST)
3.4SST Procedure
Warning
Always disconnect the ventilator from the patient before
you run SST. If you run SST while the ventilator is
connected to the patient, physical injury to the patient may
occur.
Turn the power switch (located on the front of the BDU). The
1
system conducts the POST (power-on self test) and displays
the
Ventilator Startup
Allow the ventilator to stabilize for ten minutes with the
2
power on.
Install the patient circuit, the expiratory and inspiratory filters
3
that you will use to ventilate the patient.
Caution
The patient circuit must be unobstructed and properly connected
to the ventilator to ensure accurate circuit resistance
measurement.
OP 3
screen.
At the
4
touch screen), then press the
the BDU) within 5 seconds. (Refer to Figure 3-1 for location of
TEST
The system displays the
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
Ventilator Startup
button.)
screen, touch the SST button (lower
button (on the left side of
TEST
SST Setup
screen (lower touch screen).
OP 3-5
Page 80
OP 3
How to run Short Self Test (SST)
NOTE:
You must press the TEST button within 5 seconds of touching
the SST button or SST will not start.
Tes t
button
Figure 3-1. Test button location
Caution
Do not press the test button when powering up the ventilator. This
may cause the ventilator to enter Service Mode. If you enter
Service Mode, do not attempt to run Extended Self Test (EST) with
a patient circuit. Doing so will cause EST to fail. If EST fails, the
ventilator will remain in a Vent Inop state until EST successfully
passes.
If you accidentally enter Service Mode, exit Service Mode by
touching the EXIT button on the lower GUI screen and then
pressing the ACCEPT key.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 3-6
Page 81
How to run Short Self Test (SST)
OP 3
Touch the
5
PATIENT CIRCUIT
key in the lower touch screen,
then use the knob to select either Adult, Pediatric, or Neonatal
(if
Touch the
6
NeoMode
software option is installed) patient circuit.
HUMIDIFICATION TYPE
key in the lower touch
screen, then use the knob to select the humidification type
that you will use for patient ventilation.
If you will not use a humidifier, set the humidification type to
.
HME
Press
7
ACCEPT
to complete your selection of the patient circuit
and humidification types.
Warning
Incorrectly specifying the patient circuit type or changing
the patient circuit type after you have run SST can affect
the accuracy of the compliance calculation, the measured
exhaled tidal volume, and delivered/measured inspired
tidal volumes. You must rerun SST when you change the
circuit type.
Compliance calculation and tidal volume accuracy may also
be affected by incorrectly specifying or changing the
humidifier after running SST. If you change humidifiers,
ensure you change the humidification type as described in
Section 4.8. For optimum accuracy, rerun SST using the
new humidifier.
The ventilator automatically starts the test sequence. Refer to
8
Table 3-1 for details regarding each SST test step.
system will wait indefinitely for your response. Otherwise you
don’t need to do anything unless a test result is
FAILURE
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
, or SST is complete.
ALERT
or
OP 3-7
Page 82
OP 3
How to run Short Self Test (SST)
As each test is performed, the
9
SST Status
screen shows test
results (see Table 3-2).
Warning
To ensure reliable SST results, do not repeat an individual
test with a different patient circuit if the test result is
FAILURE or ALERT. If you suspect a defective patient circuit,
replace the patient circuit and restart SST from the
beginning.
You can touch
10
touch
EXIT SST
EXIT SST
again to resume testing, or press
restart the ventilator (if SST has not detected an
FAILURE
).
during SST to halt testing. You can
ACCEPT
ALERT
Warning
To ensure correct compensation for circuit resistance and
compliance, do not exit SST until the entire SST is
successfully completed. Do not begin normal ventilation
until the entire SST is successfully completed with the
correct patient circuit installed.
When all of the tests in SST are complete, the
11
SST Status
displays all individual test results and SST outcome. Table 3-3
summarizes overall SST outcomes and how to proceed in each
case.
To begin normal ventilation (if SST has not detected an
12
or
FAILURE
The ventilator reruns
13
The ventilator displays the
14
with
), touch
EXIT SST
Ventilator Startup
, then press
.
POST
Ventilator Startup
ACCEPT
.
screen. Proceed
to configure the system for the patient.
to
or
screen
ALERT
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 3-8
Page 83
How to run Short Self Test (SST)
Table 3-1: SST test sequence
Test stepFunctionComments
OP 3
SST SetupThe system prompts you to
specify the patient circuit
type and humidification type
that you will use for the
patient ventilation.
1Specify the patient circuit
type.
2Specify the humidification
type.
You can select one of three
humidification types:
• Heated expiratory tube
• Non-heated expiratory
tube
• HME (heat-moisture
exchanger)
3For non-HME humidifiers,
specify the dry humidifier
volume. Use the specified
volume, not the
compressible volume, of
the humidifier.
4Press the
Warning
Select the correct
patient circuit type
and humidification
type. Otherwise,
faulty occlusion
detection and
erroneous expiratory
spirometry can result.
ACCEPT
key.
NOTE:
HUMIDIFIER VOLUME
The
HME.
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
button is not visible on the touch screen if you select
OP 3-9
Page 84
OP 3
How to run Short Self Test (SST)
Table 3-1: SST test sequence (cont)
Test stepFunctionComments
SST Flow Sensor TestThe system prompts you to
connect the patient circuit to
inspiratory filter.
Use Figure 2-6 on page OP
2-13 to connect the patient
circuit.
1Connect the patient circuit
to the inspiratory filter—
but without the
humidifier.
2Press
ACCEPT
the test.
NOTE:
Do not run the Flow Sensor Test with a humidifier
installed, even if you will use a humidifier when you
begin patient ventilation.
The system prompts you to
block the patient wye.
The system checks the
accuracy of the inspiratory
and expiratory flow sensors.
After the test completes, the
system prompts you to
connect the humidifier.
3Block the wye with a
No. 1 stopper.
4Press
If the status of the SST Flow Sensor Test is
cannot use the
function.
ACCEPT
FAILURE
NOTE:
If you will use a humidifier during patient ventilation, connect the humidifier
to the patient circuit after the system passes the SST Flow Sensor Test. Refer to
Figure 2-6 on page OP 2-13 for connection information.
to begin
.
, you
OVERRIDE
Circuit Pressure TestThe system verifies proper
function of the BDU pressure
sensors.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 3-10
If the status of the Circuit
Pressure Test is FAILURE, you
cannot use the
function.
OVERRIDE
Page 85
How to run Short Self Test (SST)
Table 3-1: SST test sequence (cont)
Test stepFunctionComments
OP 3
Circuit Leak TestThe system determines the
ability of the circuit to hold
pressure.
The system displays the drop
in circuit pressure over a
10 second interval.
Expiratory Filter
Resistance Test
The system prompts you to
detach circuit tubing from
the expiratory filter.
At the conclusion of the
Expiratory Filter Resistance
Tes t, the system displays the
pressure drop across the
expiratory filter.
ALERT
If the system reports
and you choose to override
the alert status, the result can
be improper compliance
compensation, inaccurate
tidal volume delivery, or
autotriggering during patient
ventilation.
If the test detects excessive
leaks, the system reports a
FAILURE
1Detach the patient circuit
2Press
If the system reports an
for the Expiratory Filter Resistance Test and you
override the
inaccurate patient pressure
estimation can result.
The system will report a
FAILURE
exhalation compartment
occlusion or an expiratory
filter occlusion.
If you do not correctly follow
the prompts to disconnect
and connect the patient
circuit, the system will report a
FAILURE
.
from the expiratory filter.
ACCEPT
the test.
if the test detects an
.
ALERT
to begin
, an
ALERT
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 3-11
Page 86
OP 3
How to run Short Self Test (SST)
Table 3-1: SST test sequence (cont)
Test stepFunctionComments
Expiratory Filter
Resistance Test
(cont)
Circuit ResistanceThe system prompts you to
The system prompts you to
reattach the patient circuit.
unblock the patient wye.
The system displays the
pressure drop across the
inspiratory and expiratory
limbs.
The reported pressure drop
includes the effect of all
devices installed on each
limb, such as filters, water
traps, or a humidifier.
3Reattach the patient
circuit to the expiratory
filter.
4Press
1Remove the stopper from
2Press
If the system reports an
for the pressure drop across
the two limbs and you
override the
inaccurate patient pressure
estimation can result.
The system reports a
if the test detects excessive
high or low limb resistance, or
if you do not follow the
prompt to unblock the wye.
ACCEPT
the next test.
the wye.
ACCEPT
the test.
ALERT
to begin
to begin
ALERT
, an
FAILURE
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 3-12
Page 87
How to run Short Self Test (SST)
Table 3-1: SST test sequence (cont)
Test stepFunctionComments
OP 3
Compliance
Calibration
The system prompts you to
block the patient wye.
If you selected a
humidification type of either
Heated exp tube or Nonheated exp tube, the
ventilator prompts you to
indicate if there is water in
the humidifier.
The system displays the
compliance of the patient
circuit.
The system prompts you to
unblock the patient wye.
1Block the wye with a
No. 1 stopper.
2Press
3Press ACCEPT to indicate
If the system reports an
for the patient circuit
compliance and you override
the
compliance compensation or
inaccurate tidal volume
delivery can result.
The system reports a
if the test detects an out-ofrange compliance condition.
4Remove the stopper from
5Press
ACCEPT
the patient circuit
compliance test.
YES
or CLEAR to indicate
NO
, as appropriate, to
indicate whether or not
there is water in the
humidifier.
ALERT
, improper
the patient wye.
ACCEPT
complete the SST test
sequence.
to begin
ALERT
FAILURE
to
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 3-13
Page 88
OP 3
3.5SST Results
How to run Short Self Test (SST)
The
ventilator uses four status categories to characterize the
840
individual SST test results, and the overall SST outcome.
ALERT
You can override an
reported for an individual test if you
ALERT
can determine with certainty that the defect in the ventilator or
related component cannot create a hazard for the patient, or add
to the risks that may arise from other hazards.
NOTE:
If an ALERT is reported and you exit SST without overriding the
ALERT, the ventilator will enter the safety valve open (SVO)
state and cannot be used for normal ventilation until SST
passes or the ALERT is overridden.
FAILURE
When the system declares a
FAILURE
for an individual test in the
SST sequence, the ventilator enters the SVO state. When a
ventilator experiences a
FAILURE
, immediately remove the
equipment from clinical use until qualified service personnel have
completed and verified the necessary repairs.
OVERRIDDEN
OVERRIDDEN
is a final status of the overall SST outcome and
indicates that you used the override feature when the system
reported an
the test with an
PASS
condition. (The ventilator must have ended
ALERT
condition.)
ALERT
is the final status of the overall SST outcome in which no
PASS
alerts or failures were detected.
Refer to Table 3-2 and Table 3-3 to learn how to interpret and
respond to each of these SST status categories.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 3-14
Page 89
How to run Short Self Test (SST)
3.5.1 How to interpret individual SST test results
SST reports a test result status for each of the individual tests. Use
Table 3-2 to interpret SST test results and to determine how to
respond.
Table 3-2: Individual SST test results
OP 3
If the test
status is:
PASSED
ALERT
FAILURE
It means:Do this:
The system did not
detect a fault for the
You do not need to do anything, unless you
are prompted by the ventilator.
individual test.
The test result is not
ideal, but is not
When the system prompts you, touch one of
these buttons, then press ACCEPT:
critical.
If SST is in progress, it
EXIT SST
halts further testing
and prompts you to
make a decision.
A critical problem has
been detected, and
RESTART SST
REPEAT
Touch one of these buttons, then press
ACCEPT:
SST cannot complete
until the ventilator
EXIT SST
passes the failed test.
RESTART SST
NEXT
Discontinue SST
Repeat SST from the
beginning
Proceed to the next test
Repeat the test
Discontinue SST
Repeat SST from the
beginning
REPEAT
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
Repeat the test
OP 3-15
Page 90
OP 3
How to run Short Self Test (SST)
3.5.2 SST outcomes
When SST has completed all of the tests, use Table 3-3 to
determine how to proceed.
Table 3-3: Overall SST outcomes
If the SST
outcome is:
PASSED
ALERT
It means:Do this:
All tests passed.Touch one of these buttons, then press
ACCEPT:
Exit SST and begin normal
ventilation
Repeat SST from the
beginning
Discontinue SST
One or more faults
were detected. If you
can determine with
certainty that this
EXIT SST
RESTART SST
Touch one of these buttons, then press
ACCEPT:
EXIT SST
cannot create a
hazard for the patient,
or add to the risks
which may arise from
other hazards, you
can choose to
override the ALERT
status and authorize
ventilation.
RESTART SST
OVERRIDE
Repeat SST from the
beginning
Press ACCEPT to override
the ALERT, as allowed by
your institution’s protocol.
Touch EXIT SST, then press
ACCEPT to begin normal
ventilation.
FAILURE
One or more critical
faults were detected.
The ventilator enters
the SVO state and
cannot be used for
Restart SST with a different patient circuit.
Touch one of these buttons, then press
ACCEPT:
EXIT SST
Discontinue SST
normal ventilation
until SST passes.
Service is required.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
RESTART SST
Press ACCEPT to repeat
SST from the beginning. If
the failure persists, contact
qualified service personnel.
OP 3-16
Page 91
C
HAPTER
4
How to use the 840 ventilator
Chapter 4 tells you:
•How the ventilator user interface is structured
•How to start up the ventilator for a new or previous patient
•How to change main settings
•How to change other settings
•How to set the humidification type, expiratory sensitivity, and
disconnect sensitivity
•How to enable or disable the oxygen sensor
•How to select and set the variable that remains constant when
the breath rate setting is changed
•How to set the alarm limits
•How to perform inspiratory and expiratory pause maneuvers
•How to interpret inspiratory pause maneuver displays
•How to use Non-invasive ventilation (NIV)
4
NOTE:
The DualView touch screens use light beams to
detect where you touch the screen. To avoid a
DEVICE ALERT
substances or objects on the screen.
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
alarm, do not place any foreign
OP 4-1
Page 92
OP 4
4.1 Structure of user interface
The following buttons are available on the upper and lower touch
screens. These buttons appear across the bottom portion of each
of the two touch screens.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 4-2
Page 93
How to use the 840 ventilator
4.2Patient setup
Warning
Always complete the patient setup before you attach a
patient to the ventilator. If you attach a patient before the
setup procedure is complete, the ventilator issues a
procedure error and initiates the safety ventilation mode.
When you turn on the ventilator, the ventilator automatically
runs POST (Power On Self Test). After POST passes, the system
displays the
screen. The prompt area, located in the lower right corner of the
lower screen, contains setup instructions.
Ventilator Startup
OP 4
screen (Figure 4-2) on the lower
Figure 4-2. Ventilator Startup screen
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
OP 4-3
Page 94
OP 4
How to use the 840 ventilator
4.2.1 How to ventilate with most recent control
parameters
To continue ventilation with the most recent ventilator control
parameters, touch
does not begin until a patient is connected. A flashing reminder
arrow prompts you to consider the previous Tube I.D. and Tube
Type if the prior Spontaneous Type used these parameters.
Same Patient
and press
ACCEPT
. Ventilation
4.2.2 How to ventilate with new control parameters
Refer to Table A-12 in Appendix Afor the descriptions, ranges,
resolutions, accuracies, and new patient values of the available
ventilator control parameters.
ouch the
1
T
parameters for patient ventilation.
Touch the
2
to adjust the IBW. The proposed value is highlighted.
Warning
Always enter the IBW that is appropriate for the patient.
The system uses the patient’s IBW to automatically set
certain values, alarm limits, and parameter boundary limits
for several initial parameters. (The IBW values that
correlate with patient height are listed in Table 4-1.)
New Patient
(ideal body weight) button, then turn the knob
IBW
button to select new ventilator control
Touch
3
the
If you want to return to the
RESTART
The system displays the
4
following buttons, and uses the rotary knob or drop-down
menus to display the available selections.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 4-4
CONTINUE.
CONTINUE
button.
(You must touch the
button will display.)
Ventilator Startup
New Patient Settings
button first before
IBW
screen, touch the
screen with the
Page 95
How to use the 840 ventilator
OP 4
Vent Type
•
INVASIVE
Determines the ventilation type
:
— conventional ventilation using either
endotracheal (ET) or tracheostomy (trach) tubes
•
non-invasive
NIV (
ventilation using full-face masks,
) —
nasal masks, infant nasal prongs, or uncuffed ET tubes
(see Section 4.12 for specific information on how to use
NIV)
Determines the type and sequence of breath delivery
Mode
:
•
(Assist/Control)
A/C
•
•
•
(Synchronous Intermittent Mandatory Ventilation)
SIMV
SPONT
BILEVEL
(Spontaneous)
(available only with
BiLevel
software option
when Vent Type is INVASIVE)
Mandatory Type:
Determines the type of mandatory breath
control
•
(Pressure Control)
PC
•
(Volume Control)
VC
• VC+ (Volume Control Plus available only with
Ventilation Plus (VV+
) software option when
Volume
Vent Type
INVASIVE)
is
(If the selected
Mode
is
SPONT
, the
Mandatory Type
applies to
manual inspirations only.)
Determines type of support for
Spontaneous Type
:
spontaneous breaths
•
(Pressure Support)
PS
•
(Tube Compensation available only with TC software
TC
option when Vent Type is INVASIVE)
•
(Volume Support available only with
VS
VV+
software
option when Vent Type is INVASIVE)
•
(Proportional Assist available only with
PA
PAV+
software option when Vent Type is INVASIVE)
•
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
NONE
OP 4-5
Page 96
OP 4
How to use the 840 ventilator
(If the selected
Mode
is
A/C,
the
Spontaneous Type
button
does not appear.)
Determines the method used to detect patient
Trigger Type
:
inspiratory effort
•
•V-
(Pressure) (not available when Vent Type is NIV or
-
P
TRIG
when using the
(Flow)
TRIG
NeoMode
option)
Touch the button and turn the knob to adjust the desired
settings. When you complete your settings changes, touch
CONTINUE
The final
5
.
New Patient Settings
screen appears. Touch the button
of each parameter you want to change, then turn the knob to
select its value. To cancel this change, press the
cancel all changes and start over, touch the
CLEAR
RESTART
NOTE:
The ventilator control parameter you are setting may be
dependent upon other ventilator settings that determine its
boundaries. Refer to the prompt area on the lower GUI screen
(Figure 1-2) for more information.
Press
6
ACCEPT
to put all of your ventilation control settings
into effect. Normal ventilation begins once a patient is
connected.
The
7
Apnea Setup
screen appears. Apnea settings are
automatically determined based on IBW, circuit type, and
mandatory breath type, but you can change them. If you
change any apnea settings, press
ACCEPT
to apply.
Although you are not required to change or confirm apnea
settings, you should verify that they are appropriate for the
patient prior to ventilation.
Press the
8
ALARM SETUP
limit settings on the
button to review the current alarm
Alarm Settings
screen. Ensure that they
are appropriate for the patient. To change any limit, touch the
button and turn the knob. To cancel, touch
ALARM
. To apply the settings, press the
PROPOSED
ACCEPT
key.
key. To
button.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 4-6
Page 97
How to use the 840 ventilator
You may choose to calibrate the ventilator’s oxygen sensor at
9
this point. Press the
keyboard below the touch screens.
During the oxygen sensor calibration, the ventilator delivers
100% oxygen (if available) for two minutes and calibrates the
oxygen sensor in the Breath Delivery Unit (BDU).
The ventilator always monitors the delivery of oxygen to the
patient unless you disable the oxygen sensor. Touch the
SETTINGS
button to access oxygen sensor disable or enable
functions.
After you accept the ventilation control parameters, you can
10
attach a patient to the ventilator. Ventilation only begins
when the ventilator senses that a patient is attached.
If you attach a patient before completing setup, the ventilator
initiates safety ventilation mode and annunciates a
PROCEDURE ERROR
patient setup.
Warning
Each patient circuit type is appropriate for a specified range
of IBW values. This information is summarized in Table 4-2.
The recommended ranges exist to ensure patient safety.
Only those with expertise to judge the appropriate
circumstances should override the recommended ranges.
OP 4
100% O2 / CAL 2 min
alarm that is reset once you complete the
key located on the
MORE
4.2.3 Patient data and current settings
The top of the upper screen shows vital patient data. (Out-ofrange data flashes to alert you.) The current breath type is
indicated in the upper left corner:
•
= Control
C
•
=Spontaneous
S
•
= Assist
A
You can access additional patient data when you touch the
PATIENT DATA
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
button.
MORE
OP 4-7
Page 98
OP 4
How to use the 840 ventilator
You can display the definitions for any symbol used in the patient
data, alarm log, or settings areas by touching the symbol. The
symbol definitions appear at the bottom of the lower touch
screen.
Current ventilator control settings are displayed across the top of
the lower touch screen (Figure 4-3). If you press the
key, the lower touch screen automatically displays the
2 min
PROGRESS
PROGRESS
indicator. If you touch the
indicator will appear if there is no other higher-
priority display active. Press the
CANCEL
Alarm Silence
button for either
100% O
key, the
/CAL
2
IN
IN
indicator to cancel the alarm silence or oxygen sensor calibration
in progress.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 4-8
Page 99
How to use the 840 ventilator
Breath type
(C = Control)
OP 4
Vital patient data area
Alarm area
Patient data
(upper screen)
Subscreen area
Main ventilator
control settings
Ventilator settings
(lower screen)
3VTE MAND
2 of last 4 mand breaths ≤ set limit. Check for leaks, changes in
patient'sR&C.
11:2025 Jun 2000
Subscreen area
Figure 4-3. Touch screen appearance during normal ventilation (shown with
alarm silence and 100% O
4-075609-00 Rev. G (10/06)840 Ventilator System Operator’s Manual
/CAL in progress)
2
OP 4-9
Page 100
OP 4
How to use the 840 ventilator
4.2.4 Ideal Body Weight (IBW)
The system initially sets most upper and lower alarm limits based
on the patient’s IBW. After you enter the IBW, you can review and
change these alarm settings, if needed. Table 4-1 below provides
the information needed to determine the patient’s IBW using the
patient’s height.
840 Ventilator System Operator’s Manual4-075609-00 Rev. G (10/06)
OP 4-10
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.