Puritan Bennett 740, 760, 700 User Manual

Page 1
.........................................................................
Operator’s Manual
Part No. G-061874-00 Rev. D September 2000
Page 2
Copyright Information
Copyright 2000 Mallinckrodt Inc. EasyCart, EasyNeb, 740, 760, and 700 Series are trademarks of Mallinckrodt Inc. All rights reserved. The 700 Series (including the 740 proprietary information, covered by one or more of the following U.S. Patents and foreign equivalents: 5,524,615; 5,540,222; 5,596,984; 5,632,270; 5,664,560; and 5,673,689.
The information contained in this manual is the sole property of Mallinckrodt Inc. and may not be duplicated without permission. This manual may be revised or replaced by Mallinckrodt Inc. at any time and without notice. You should ensure that you have the most current applicable version of this manual; if in doubt, contact the Technical Publications Department of Mallinckrodt Inc. While the information set forth herein is believed to be accurate, it is not a substitute for the exercise of professional judgment.
The ventilator should be operated and serviced only by trained professionals. Mallinckrodt’s sole responsibility with respect to the ventilator, and its use, is as stated in the limited warranty provided.
Nothing in this manual shall limit or restrict in any way Mallinckrodt’s right to revise or otherwise change or modify the equipment (including its software) described herein, without notice. In the absence of an express, written agreement to the contrary, Mallinckrodt Inc. has no obligation to furnish any such revisions, changes, or modifications to the owner or user of the equipment (including its software) described herein.
Ô
and 760Ôventilators) are manufactured in accordance with Mallinckrodt
Ô
Ventilator System
Definitions
This manual uses these special indicators to convey information of a specific nature:
Warning
Indicates a condition that can endanger the patient or the ventilator operator.
Caution
Indicates a condition that can damage the equipment.
NOTE:
Indicates points of particular emphasis that make operation of the ventilator more efficient or convenient.
700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)
Page 3
...............................................................................
.
Warnings, cautions, and notes
Please take the time to familiarize yourself with the following safety considerations, special handling requirements, and regulations that govern the use of the 700 Series Ventilator System.
Warning
To avoid an electrical shock hazard while servicing the ventilator, be sure to remove all power to the ventilator by disconnecting the power source and turning off all ventilator power switches.
Warning
To avoid a fire hazard, keep matches, lighted cigarettes, and all other sources of ignition (e.g., flammable anesthetics and/or heaters) away from the ventilator and oxygen hoses.
Do not use oxygen hoses that are worn, frayed, or contaminated by combustible materials such as grease or oils. (Textiles, oils, and other combustibles are easily ignited and burn with great intensity in air enriched with oxygen.)
In case of fire or a burning smell, immediately disconnect the ventilator from the oxygen supply and electrical power source.
Warning
Patients on life-support equipment should be appropriately monitored by competent medical personnel and suitable monitoring devices.
The 700 Series Ventilator is not intended to be a comprehensive monitoring device and does not activate alarms for all types of dangerous conditions for patients on life-support equipment.
Warning
Check the ventilator periodically as outlined in the service manual; do not use if defective. Immediately replace parts that are broken, missing, obviously worn, distorted, or contaminated.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual
iii
Page 4
Warning
An alternative source of ventilation should always be available when using the 700 Series Ventilator System.
Warning
To ensure proper servicing and avoid the possibility of physical injury, only qualified personnel should attempt to service or make authorized modifications to the ventilator.
The user of this product shall have sole responsibility for any ventilator malfunction due to operation or maintenance performed by anyone not trained by Mallinckrodt staff.
Warning
For a thorough understanding of ventilator operations, be sure to read the 700 Series Ventilator System Operator's Manual in its entirety before attempting to use the system.
Warning
Before activating any part of the ventilator, be sure to check the equipment for proper operation and, if appropriate, run the self­diagnostic short self test (SST) program described in this manual.
Caution
U.S. Federal law restricts this device to sale by or on the order of a physician.
Warranty
The 700 Series Ventilator System is warranted against defects in material and workmanship in accordance with Mallinckrodt Medical Equipment Warranty for a period of one year from the time of sale. To ensure the validity of the warranty, be sure to keep a maintenance record.
iv
700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)
Page 5
...............................................................................
.
M
Year of manufacture
The 700 Series Ventilator System’s year of manufacture is indicated by the fifth and sixth digits of the serial number which is located at the lower edge of the ventilator front panel.
Manufacturer
anufactured by
Nellcor Puritan Bennett Ireland A subsidiary of Mallinckrodt Inc. Mervue, Galway, Ireland
Phone: +353.91.753.771 Fax: +353.91.753.922
European Headquarters
Mallinckrodt Europe BV Hambakenwetering 1 5231 DD ’s-Hertogenbosch The Netherlands
Phone: +31.73.6485200 Fax: +31.73.6410915
Electromagnetic susceptibility
The 700 Series Ventilator System complies with the requirements of IEC 60601­1-2 (EMC Collateral Standard), which includes E-field susceptibility and ESD requirements. However, even though the device is compliant at the levels of immunity specified in the standard, certain transmitting devices (cellular phones, walkie-talkies, cordless phones, paging transmitters, etc.) emit radio frequencies that could interrupt ventilator operation if located in a range too close to the ventilator. It is difficult to determine when the field strength of these devices becomes excessive. Practitioners should be aware that radio frequency emissions are additive, and that the ventilator must be located a sufficient distance from transmitting devices to avoid interruption. Do not operate the ventilator in a magnetic resonance imaging (MRI) environment. The Alarm handling section of this manual describes possible ventilator alarms and what to do if they occur. Consult with your institution’s biomedical engineering department in case of interrupted ventilator operation, and before relocating any life support equipment.
Customer assistance
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual
For further assistance contact your local Mallinckrodt representative.
v
Page 6
vi
700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)
Page 7
Contents
..............................................................................
1 Introduction
1.1 Functionaldescription ......................................1-2
1.2 Symbolsandlabels ........................................1-8
1.3 Keyboard ...............................................1-13
1.3.1 VENTILATORSETTINGS .......................... 1-14
1.3.2 PATIENTDATA .................................. 1-24
1.3.3 VENTILATORSTATUS ............................1-28
2 Setting up the ventilator
2.1 Connectingandusinginternalandexternalbatteries ..............2-2
2.2 Connectingtheelectricalsupply............................... 2-6
2.3 Connecting the oxygen supply . . . . . . . . ........................ 2-8
2.4 Connectingtheventilatorbreathingcircuit ...................... 2-10
2.5 Installingthecollectorvial .................................. 2-13
2.6 Installingtheflexarm ......................................2-14
2.7 Installingthehumidifier..................................... 2-16
2.8 Usingtheventilatorcart ....................................2-17
3 Getting started
3.1 Poweringuptheventilator ...................................3-1
3.2 Selectingventilatorsettings .................................. 3-4
3.3 Viewingandchangingalarmsettings...........................3-6
3.4 Enteringandexitingstandbymode ............................ 3-8
4 Self tests (SST and EST)
4.1 Shortselftest(SST) ........................................ 4-3
4.2 Extended self test (EST) . . . . . . . . . . . . ....................... 4-13
5 Once ventilation begins
5.1 Changingsettings:aquickreview .............................5-1
5.1.1 Changingsettings.................................. 5-1
5.1.2 SwitchingbetweenVCV,PCV,andPSV ................ 5-2
5.1.3 Changingthemode ................................5-2
5.2 Viewingandchangingalarmsettings:aquickreview .............. 5-3
5.3 Adjustingapneaparameters ................................. 5-4
5.3.1 Adjustingtheapneainterval ..........................5-6
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual
vii
Page 8
5.4 Viewingpatientdata........................................ 5-6
5.5 The 100% O2 and MANUAL INSP keys . . . . . .................. 5-10
5.6 The EXP PAUSE and INSP PAUSE keys (760 only) ............. 5-11
6 The MENU key
6.1 Moreactivealarms......................................... 6-5
6.2 Autoresetalarms .......................................... 6-6
6.3 Selftests ................................................ 6-7
6.4 Usersettings ............................................. 6-7
6.4.1 Endotrachealtube ................................. 6-7
6.4.2 Humidifiertype .................................... 6-8
6.4.3 Dateandtimeset .................................. 6-8
6.4.4 Apneainterval(Ta)................................. 6-9
6.4.5 VCVflowpattern .................................. 6-9
6.4.6 Speakingvalvesetup .............................. 6-10
6.4.7 Alarmvolume .................................... 6-16
6.4.8 PCVtimingsetting ................................ 6-16
6.4.9 VolumeLEDbar.................................. 6-16
6.5 Oxygensensor........................................... 6-17
6.6 Standby mode . . . . . . . . . . . . ............................... 6-19
6.7 Batteryinfo ............................................. 6-20
6.8 Softwarerevision ......................................... 6-20
6.9 Servicesummary ......................................... 6-21
6.10 Nebulizer .............................................. 6-21
Contents
7 Alarm handling
7.1 Autoresetalarms .......................................... 7-3
7.2 Alarmsilence ............................................. 7-3
7.3 Alarmreset............................................... 7-4
7.4 Clinicalandtechnicalalarms ................................. 7-5
7.5 Poweralarm............................................. 7-17
7.5.1 LossofACPower ................................ 7-17
7.5.2 LossofPower ................................... 7-18
Appendix A Maintenance
A.1 Cleaning,disinfection,andsterilization .........................A-2
A.1.1 Cleaning: general guidelines . . . . . . ...................A-4
A.1.2 Disinfectionandsterilization .........................A-4
A.2 Preventivemaintenance ....................................A-5
viii
700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)
Page 9
.
Contents
...............................................................................
A.2.1 Daily or as required:
inspiratoryandexpiratorybacteriafilters.............. A-8
A.2.2 Dailyorasrequired:collectorvial .................... A-8
A.2.3 Dailyorasrequired:in-linewatertraps ................ A-9
A.2.4 Asnecessary:oxygensensorcalibration............... A-9
A.2.5 Every 250 hours (or 1 month of use): cooling fan filter . . . A-10 A.2.6 Every 1000 hours (or 3 months of use): air intake filter . . . A-11
A.2.7 Every2years:devicechecks ...................... A-12
A.2.8 Storage ....................................... A-13
A.2.9 Repacking ..................................... A-13
Appendix B Part numbers
Appendix C Specifications
C.1 Physical ................................................ C-2
C.2 Environmental ........................................... C-3
C.3 Power ................................................. C-3
C.4 Complianceandapprovals ................................. C-5
C.5 Technical ............................................... C-5
Appendix D Breath delivery
D.1 A/Cmode ............................................... D-3
D.2 SPONTmode ........................................... D-3
D.3 SIMVmode ............................................. D-4
D.3.1 Breathtiming .................................... D-4
Appendix E Alarm testing
Appendix F Pneumatic schematic
Appendix G Glossary
Index
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual
ix
Page 10
Contents
x
700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)
Page 11
Figures
..............................................................................
Figure 1-1 Block diagram: 700 Series Ventilatorfunction ............1-5
Figure 1-2 740 VentilatorSystemkeyboard......................1-13
Figure 1-3 760 VentilatorSystemkeyboard......................1-14
Figure2-1 Liftingtheventilator ................................ 2-2
Figure2-2 Internalbatterychargeindicator ....................... 2-3
Figure2-3 Pluggingtheexternalbatteryintotheventilator .......... 2-5
Figure2-4 Disconnectingtheexternalbattery .....................2-6
Figure2-5 Connectingtheventilatorpowercord ................... 2-7
Figure2-6 Storingthepowercordontheventilator ................. 2-7
Figure2-7 Connectingtheoxygensupply ........................ 2-9
Figure2-8 Connectingtheventilatorbreathingcircuit ..............2-12
Figure2-9 Installingthecollectorvial .......................... 2-13
Figure2-10 Installingtheflexarm ..............................2-14
Figure2-11 Shorteningtheflexarm ............................ 2-15
Figure2-12 Installingthehumidifier ............................2-16
Figure2-13 Lockingandunlockingthecart’sfrontwheels ...........2-17
Figure3-1 Turningthepowerswitchon(the“I”position)............. 3-2
Figure5-1 Viewingpatientdata ................................5-8
Figure 5-2 Volume bar graph (760 Ventilatoronly) ................. 5-9
Figure 6-1 Using the More active alarms menu function
to view active alarms 6-5
Figure 6-2 Using the Autoreset alarms menu function
to view autoreset alarms 6-6
Figure7-1 Viewingactivealarms ..............................7-2
FigureA-1 Removing/replacingthecollectorvial .................. A-9
FigureA-2 Coolingfancover ................................ A-10
FigureA-3 Airintakefilter ................................... A-12
FigureB-1 Ventilatoraccessories ............................. B-9
Figure C-1 Recommended ventilator breathing circuit configurations . . C-8
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual
xi
Page 12
Figures
FigureD-1 Flowwaveform ................................... D-2
FigureD-2 SIMVbreathperiodintervals ........................ D-4
Figure D-3 Synchronizing breath intervals with patient effort . . . . . . . . . D-5
FigureD-4 SpontaneousbreathsduringSIMV ................... D-5
FigureD-5 MandatorybreathsduringSIMV...................... D-5
FigureD-6 ManualinspirationduringSIMV ...................... D-6
xii
700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)
Page 13
Ta bl e s
..............................................................................
Table 1-1 Mode/breath type availability on 740/760 Ventilators ........1-2
Table 1-2 Changes to current settings in occlusion cycling mode . . . . . . . 1-7
Table 1-3 700 Series Ventilator keyboards: VENTILATOR SETTINGS . 1-15
Table 1-4 700 Series Ventilatorkeyboards:PATIENTDATA .........1-24
Table 1-5 700 Series Ventilator keyboards: VENTILATOR STATUS . . . 1-29
Table 4-1 700 Series Ventilatorselftests .........................4-2
Table4-2 SSTsequenceoftests................................4-9
Table4-3 OverallSSTresults .................................4-12
Table4-4 ESThardwarerequirements ..........................4-13
Table4-5 ESTsetupmessages................................ 4-14
Table4-6 ESTtestsequence..................................4-16
Table4-7 KeyfunctionsduringEST ............................ 4-21
Table4-8 PromptsduringEST.................................4-22
Table4-9 ESTcompletionstatus ..............................4-23
Table5-1 Breathtypeavailability ................................5-2
Table6-1 Menufunctionsummary............................... 6-2
Table7-1 Clinicalalarms ...................................... 7-5
Table7-2 Technicalalarms ...................................7-12
TableA-1 Cleaning,disinfection,andsterilization .................. A-3
TableA-2 Preventivemaintenanceschedule ..................... A-7
TableB-1 Ventilatoraccessories ............................... B-2
TableC-1 Physicalspecifications ............................... C-2
TableC-2 Environmentalspecifications .......................... C-3
TableC-3 Powerspecifications ................................ C-3
TableC-4 Complianceandapprovals ........................... C-5
TableC-5 Technicalspecifications .............................. C-5
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual
xiii
Page 14
Tables
xiv
700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)
Page 15
SECTION
Introduction 1
1
..............................................................................
The 700 Series Ventilator System (including the 740 and 760 Ventilators) provides respiratory support for a wide range of pediatric to adult patients for a wide variety of clinical conditions. The ventilator’s mixing technique allows it to ventilate critically ill patients at adjustable oxygen concentrations without the need for a blender, compressor, or hospital-grade wall air.
The 700 Series Ventilator System can be mains- or battery-powered. Each ventilator includes two microcontrollers: one for breath delivery (which controls ventilation), and one for the user interface (which monitors ventilator and patient data). Each microcontroller verifies that the other is functioning properly. Using two independent microcontrollers in this fashion prevents a single fault from causing a simultaneous failure of controlling and monitoring functions.
The 700 Series Ventilator System supplies mandatory or spontaneous breaths with a piston-based pneumatic system. Table 1-1 summarizes the modes and breath types offered by the 740 and 760 Ventilators. Mandatory breaths can be
volume control ventilation (VCV, available on 740 and 760 Ventilators) or pressure control ventilation (PCV, available on the 760 Ventilator only). VCV
delivers breaths to the patient at a preset tidal volume, peak flow, waveform, and oxygen concentration at a minimum respiratory rate. PCV delivers breaths to the patient at a preset inspiratory pressure, I:E ratio or inspiratory time, rise time factor (how quickly inspiratory pressure rises to achieve the set inspiratory pressure), and oxygen concentration at a minimum respiratory rate. A spontaneous breath allows the patient inspiratory flows of up to 300 L/min, with or without pressure support ventilation (PSV). On the 760 Ventilator, you can set the rise time factor and exhalation flow sensitivity (that is, the point at which the ventilator cycles from inspiration to exhalation) in PSV.
The ventilator begins apnea ventilation if no breath (patient- , ventilator-, or operator-initiated) is delivered within the selected apnea interval. Apnea ventilation is active during all modes. On the 740 Ventilator, only VCV breaths are available in apnea ventilation. On the 760 Ventilator, VCV or PCV breaths are available in apnea ventilation.
The 760 Ventilator also offers the ability to perform respiratory mechanics calculations and maneuvers as a standard feature using the EXP PAUSE (to
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual
1-1
Page 16
1 Introduction
calculate auto-PEEP) and INSP PAUSE (to calculate patient resistance and compliance) keys.
Table 1-1: Mode/breath type availability on 740/760 Ventilators
Mode/breath type 740 Ventilator 760 Ventilator
VCV breath type
PCV breath type
PSV breath type (support pressure setting)
PSV (rise time factor and exhalation sensitivity settings)
SIMV mode
Apnea ventilation (VCV breath type)
Apnea ventilation (choice of VCV or PCV breath type)
Respiratory mechanics (EXP PAUSE and INSP PAUSE)
••
This manual tells you how to operate and perform simple maintenance for the 700 Series Ventilator. Mallinckrodt recommends that you become familiar with this
manual and accompanying labels before attempting to operate or maintain the ventilator. If you need additional copies of this manual, contact your Mallinckrodt representative.
To ensure optimum performance of the 700 Series Ventilator System, Mallinckrodt recommends that a qualified service technician perform periodic maintenance on the ventilator. For more information, contact your Mallinckrodt representative.
1.1 Functional description
By pressing keys and turning the knob on the ventilator keyboard, the operator gives initial instructions and data to the ventilator (Figure 1-1). The user interface microcontroller processes this information and
1-2
700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)
Page 17
.
Introduction 1
...............................................................................
stores it in the ventilator’s memory. The breath delivery microcontroller uses this stored information to control and monitor the flow of gas to and from the patient.
The 700 Series Ventilator uses a flow trigger to recognize patient effort. The trigger monitors flow from the piston during exhalation. When the patient inhales, patient circuit pressure drops very slightly below end-expiratory pressure. At the same time, the piston moves forward to deliver flow to the ventilator breathing circuit and maintain the preset PEEP/CPAP level. The level of flow depends on the patient’s effort. If this flow exceeds the user-set level, the ventilator triggers. By design, the ventilator attempts to maintain PEEP in the presence of a circuit leak. Since a leak drives the piston to deliver flow to make up for pressure losses, a circuit leak can require an increase in the flow trigger level to avoid autocycling.
During exhalation, the ventilator’s piston retracts and draws air and oxygen into the cylinder. The ventilator uses room air, which means the ventilator can operate without a compressor or wall air source. Room air enters the ventilator through a protected user-replaceable air intake filter just inside the ventilator cabinet. This filter captures airborne particles.
Oxygen from a cylinder or wall supply enters the ventilator through a hose and oxygen fitting (the fitting is available in several versions). Once inside the ventilator, the oxygen is regulated to a pressure the ventilator can use, then mixed with air, according to the selected % O
The flow-triggered piston/cylinder system and motor controller circuit control the flow of gas to the patient. On the 760 Ventilator in PCV or PSV, the rate of flow is also determined by the preset rise time factor. This system is designed with a minute gap (about the size of a thin sheet of paper) between the piston and the cylinder wall. This design eliminates the friction between the piston and cylinder, allowing it to respond more rapidly than a “sealed” system.
.
2
A small amount of gas leaks through the gap between the piston and cylinder. Ventilator software and a continuous forward motion of the piston compensate for this leak.
The piston delivers the mixed air and oxygen through the inspiratory manifold system, and out to the patient. The oxygen concentration and temperature of the delivered gas are monitored here, using a galvanic oxygen sensor and a thermistor. The galvanic sensor generates a voltage proportional to the partial pressure of oxygen, from which the oxygen concentration is calculated. The ventilator alarms if the monitored oxygen concentration is more than ten percentage points above or below the % O
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual
setting. The inspiratory manifold
2
1-3
Page 18
1 Introduction
system also includes a safety valve to relieve patient pressure if necessary (for example, if the ventilator breathing circuit is kinked or occluded).
The patient system includes the components external to the ventilator that route gas between the ventilator and the patient. These components include the inspiratory filter (which protects against contamination between the ventilator and patient), a humidification device, ventilator breathing circuit (the tubing through which the gas travels), collector vial (which protects the exhalation system from moisture in the exhaled gas, and can be emptied without losing circuit PEEP), and an expiratory filter (which limits the bacteria in the patient’s exhaled gas from escaping to room air or contaminating the ventilator).
1-4
700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)
Page 19
.
Introduction 1
...............................................................................
+
-
740 Ventilator
Keyboard
Regulator
Piston/cylinder
system
Exhalation/
PEEP/CPAP
system
Filter
Oxygen
Room air
Inspiration manifold
Inspiratory filter
Gas intake system
Expiratory
filter
Humidification device
Collector
vial
Ventilator breathing circuit (inspiratory limb)
Patient
Ventilator breathing
circuit (expiratory limb)
7-00017
Figure 1-1. Block diagram: 700 Series Ventilator function
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual
1-5
Page 20
1 Introduction
The heated exhalation system monitors the flow of the patient’s exhaled gas using a differential pressure transducer. The patient exhales through the exhalation valve. During exhalation, the PEEP/CPAP system maintains user-selected pressure in the ventilator breathing circuit.
Throughout the respiratory cycle, pressure transducers monitor inspiratory, expiratory, and atmospheric pressures. The temperatures of the pneumatic compartment and inspiratory gas are also monitored. Information from these transducers is continuously used to update the calculations that control ventilation. (Appendix F provides a diagram of the ventilator’s pneumatic system and ventilator breathing circuit.)
Power to operate the ventilator comes from ac mains (wall) or battery power. The power supply is designed to protect against excessive voltages, temperatures, or current draws. A power cord retainer prevents the cord from accidental disconnection.
The ventilator includes an internal battery, and accommodates an optional external battery. Depending on the ventilator settings, battery backup power can be supplied for up to 2 ½ the external battery. Both batteries are recharged during operation from ac power. If both are installed, the external battery is used first when ac power is not present. If the external battery is depleted or not installed, the internal battery supplies power to the ventilator when ac power is not available. The keyboard indicates the source of power and battery charge level of the internal battery at all times.
hours using the internal battery, and up to 7 hours using
1-6
Emergency modes: The ventilator declares a ventilator inoperative (VENT INOP) condition if a hardware failure or critical software error that could compromise
safe ventilation occurs. In case of a ventilator inoperative condition, the VENT INOP indicator lights and the ventilator enters the safety valve open (SVO) state. To correct a ventilator inoperative condition, the ventilator must be turned off, then powered on again; at power-on, the operator must run extended self-test (EST). The ventilator must pass EST before normal ventilation can resume.
The safety valve allows the patient to breathe room air unassisted when the ventilator is in the SVO state. The ventilator remains in the SVO state until power-on self-test (POST) verifies that power levels to the ventilator are acceptable and that the motor controller and microcontrollers are functioning correctly, and until the user has confirmed ventilator settings.
If the ventilator enters the SVO state and POST is not running, the SAFETY VALVE OPEN indicator lights and a high-priority alarm sounds. The ventilator enters the SVO state if it detects a hardware or software failure that could
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 21
.
Introduction 1
...............................................................................
compromise safe ventilation. In case of a malfunction that prevents software from opening the safety valve, there is also an analog circuit that opens the safety valve when system pressure exceeds 115 cmH
If the ventilator detects an occlusion or a continuous high inspiratory pressure condition, it opens the safety and exhalation valves to vent excess pressure, then shuts them and begins occlusion cycling mode. In occlusion cycling mode the ventilator uses current settings except for those summarized in Table 1-2. If the ventilator again detects an occlusion or continuous high pressure condition, it again opens the safety and exhalation valves then resumes occlusion cycling mode. If the operator presses the alarm reset key or the ventilator does not detect an occlusion or continuous high pressure condition, it reverts to normal ventilation using the most recently accepted settings.
Table 1-2: Changes to current settings in occlusion cycling mode
Setting Change to setting
O (113 hPa).
2
HIGH PRESSURE alarm (VCV breath type)
PEEP (all modes) Set to 0 cmH2O
(all modes) Set to 100%
%O
2
SPONT mode (PSV breath type)
SUPPORT PRESSURE (PSV breath type)
SUPPORT PRESSURE (PCV breath type)
RISE TIME FACTOR (PCV breath type)
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
Setto30cmH
Breaths are delivered at a rate of 12/min with an inspiratory time of 2 seconds.
If less than 15 cmH If 15 cmH
If less than 15 cmH If 15 cmH2O or above: the current setting is used (no change).
Set to 70%
O
2
O:setto15cmH2O.
2
O or above: the current setting is used (no change).
2
O:setto15cmH2O.
2
1-7
Page 22
1 Introduction
1.2 Symbols and labels
These symbols and labels appear on the 700 Series Ventilator System:
Power switch positions per IEC 601-1.”Irepresents ON position; “O represents OFF position.
7-00421
Refer to manual per IEC 601-1. When this symbol appears on product, it means Refer to documentation for information.
7-00418
Potential equalization point, per IEC 601-1
7-00416
External battery connection
7-00426
Circuit breaker
1-8
7-00414
Serial number
SN
ac current
7-00427
Type B equipment, per IEC 601-1
7-00415
Indicates the degree of protection provided by enclosure (drip-proof).
IPX1
7-00403
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 23
.
Introduction 1
...............................................................................
Signifies compliance with the Medical Device Directive, 93/42/EEC
7-00412
CSA and NRTL (Nationally Recognized Testing Laboratory) certification, granted by CSA
8-00417
The TUV Rheinland logo signifies TUV Rheinland Type Test approval to Annex III of the Medical Device Directive
7-00420
Exhaust port connector
EXHAUST
Inspiratory limb connector
TO
PATI EN T
Expiratory limb connector
FROM
PATI EN T
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
1-9
Page 24
1 Introduction
Oxygen inlet port label
O
2
.
V
max
61006
Air intake label
61010
Cooling fan label
General life support equipment warning label
1-10
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
61009
61011
Page 25
.
Introduction 1
...............................................................................
740 Ventilator back panel label
61012
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
1-11
Page 26
1 Introduction
760 Ventilator back panel label
1-12
61952
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 27
.
Introduction 1
...............................................................................
1.3 Keyboard
Caution
To avoid damaging the keyboard, do not press on it with sharp objects.
The keyboard (Figure 1-2 shows the 740 keyboard and Figure 1-3 shows the 760 keyboard) is grouped into three sections:
VENTILATOR SETTINGS: Where you set breath delivery variables.
PATIENT DATA: Where you set alarm limits and view the monitored pressures, breath timing, and volumes.
VENTILATOR STATUS: Where you see the alarm status and operating condition of the ventilator.
Patient data Ventilator status
Ventilator settings
Figure 1-2. 740 Ventilator System keyboard
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
7-00048
1-13
Page 28
1 Introduction
Patient data Ventilator status
1-14
Ventilator settings
7-00123
Figure 1-3. 760 Ventilator System keyboard
1.3.1 VENTILATOR SETTINGS
The VENTILATOR SETTINGS section of the keyboard allows you to select the ventilation mode, breath type, and settings. For more detail on ventilation modes and breath delivery, see Appendix D.
To change the mode and settings, select the mode, then the breath type, and then the ventilator settings. The keys flash during setup and mode changes to ensure that you review all pertinent settings. The keyboard is designed to minimize accidental or unintentional changes.
Table 1-3 summarizes the functions of the keys, knob, and indicators in the VENTILATOR SETTINGS section of the keyboard. Ventilator settings are also limited by these breath delivery boundaries:
I:E ratio £ 4:1 for PCV (760 Ventilator only), £ 3:1 for all other breath types
Inspiratory time = 0.2 to 8 seconds (excluding plateau)
Expiratory time ³ 0.2 seconds
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 29
.
Introduction 1
...............................................................................
PEEP/CPAP + SUPPORT PRESSURE or INSPIRATORY PRESSURE £ 80 cmH
O (80 hPa)
2
NOTE:
Maximum SUPPORT PRESSURE is 70 cmH INSPIRATORY PRESSURE is 80 cmH
2
SUPPORT PRESSURE or INSPIRATORY PRESSURE + PEEP/CPAP < HIGH PRESSURE - 2 cmH
O(2hPa)
2
HIGH PRESSURE (in A/C and SIMV modes) > PEEP/CPAP + 7cmH
O(7hPa)
2
HIGH PRESSURE (in SPONT mode) > PEEP/CPAP + SUPPORT PRESSURE + 2 cmH
O(2hPa)
2
HIGH PRESSURE > LOW INSP PRESSURE
Minute volume £ 50 L/min at an I:E ratio of 2:1
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
O.
O, maximum
2
Key/indicator Specifies... Range
Mode/breath type settings
A/C Assist/control mode VCV (volume control ventilation)
and PCV (pressure control ventilation) breath types.
(PCV available on 760 Ventilator only.)
SIMV Synchronous intermittent
mandatory ventilation mode
SPONT Spontaneous mode PSV breath type
VCV VCV breath type VCV available on 740 and 760
PCV PCV breath type PCV available on 760 Ventilator
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
VCV, PCV (760 only), and PSV (pressure support ventilation) breath types.
Ventilators in A/C or SIMV modes.
only in A/C or SIMV modes.
1-15
Page 30
1 Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)
Key/indicator Specifies... Range
PSV PSV breath type PSV available in SIMV or
SPONT modes.
Mandatory (VCV) settings
RESPIRATORY RATE
TIDAL VOLUME
PEAK FLOW Maximum flow of gas delivered
PLATEAU (s) Length of inspiratory pause after a
Mandatory (PCV) settings (760 Ventilator only)
RESPIRATORY RATE
The minimum number of mandatory breaths the patient receives per minute. During apnea ventilation the minimum RESPIRATORY RATE setting is 6 / minute.
Volume delivered to the patient during a mandatory breath, compliance-compensated and corrected to body temperature and pressure, saturated (BTPS).
during a mandatory breath (BTPS). (Combined with tidal volume, peak flow defines the active portion of inspiratory time.)
mandatory breath has been delivered, during which no gas is delivered.
The minimum number of mandatory breaths the patient receives per minute. During apnea ventilation the minimum RESPIRATORY RATE setting is 6 / minute.
1to70/minute Accuracy: ± (0.1 + 1%) /minute
40 to 2000 ml Accuracy: ± (10 ml + 10% of
setting)
3to150L/min Accuracy: ± (5 + 10% of setting)
L/min
0.0to2.0second Accuracy: ± 0.05 second
1to70/minute Accuracy: ± (0.1 + 1%) /minute
INSPIRATORY PRESSURE (760 only)
1-16
Pressure above PEEP during the inspiratory phase of a PCV breath.
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
5to80cmH Accuracy: ± (3 + 2.5% of setting)
O
cmH
2
O(5to80hPa)
2
Page 31
.
Introduction 1
...............................................................................
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)
Key/indicator Specifies... Range
TI/I:E RATIO (760 only)
RISE TIME FAC TOR (760 only)
NOTE:
To help determine the correct setting for inspiratory time, during PCV the message window displays peak inspiratory flow, end inspiratory flow, and end exhalation flow in L/min.
YoucanusetheMENUkeyto select inspiratory time (T ratio as the breath timing setting for a PCV breath. You can change the selected breath timing setting (T I:E ratio), but the setting remains constant when you change the respiratory rate in PCV.
Selecting I:E ratio makes the set ratio of inspiratory time to expiratory time for a PCV breath.
Thetimeforinspiratorypressureto rise from 0 to 95% of the target pressure level during a PCV breath.
A setting of 100 = a 100-ms rise time, and a setting of 5 = 80% of the inspiratory time or 2500 ms, whichever is less.
When you adjust this setting, the message window shows the actual time (in seconds) to reach 95% of target pressure.
)orI:E
I
or
I
Inspiratory time (T
0.2 to 8 seconds Accuracy: ± 0.05 second
I:E ratio: Accuracy: ± (0.1 + 2%)
5to100
£ 4:1
):
I
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
1-17
Page 32
1 Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)
Key/indicator Specifies... Range
Spontaneous (PSV) settings
SUPPORT PRESSURE
RISE TIME FAC TOR (760 only)
Pressure above PEEP maintained during spontaneous inspiration. Support pressure is terminated when inspiratory flow falls to 25% of peak inspiratory flow, or to the exhalation sensitivity setting (760 only), or 10 L/min or 25% of peak flow, whichever is lower (740 only). Maximum inspiratory time is 3.5 seconds for adults, and 2.5 seconds for pediatric patients.
Thetimeforinspiratorypressureto rise from 0 to 95% of the target pressure level during a PSV breath.
A setting of 100 = a 100-ms rise time, and a setting of 5 = 80% of the inspiratory time or 1500 ms (when adult ventilator breathing circuit is selected) or 600 ms (when pediatric circuit is selected), whichever is less.
When you adjust this setting, the message window shows the actual time (in seconds) to reach 95% of target pressure.
0to70cmH Accuracy: ± (3 + 2.5% of setting)
O
cmH
2
5to100
O(0to70hPa)
2
1-18
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 33
.
Introduction 1
...............................................................................
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)
Key/indicator Specifies... Range
EXH SENSITIVITY (760 only)
Common settings
PEEP/CPAP Positive end expiratory pressure/
The percent of peak expiratory flow at which the ventilator cycles from inspiration to exhalation for spontaneous breaths. The flow at which the ventilator cycles from inspiration to exhalation for PSV breaths. Exhalation begins when the inspiratory flow is less than the set value.
To help set EXH SENSITIVITY appropriately, the peak inspiratory flow and end inspiratory flow are displayed in the message window in PSV. (To ensure accurate estimates of these flows, it is important to run SST so that tubing compliance calculations are correct.)
continuous positive airway pressure. Minimum pressure maintained during inspiratory and expiratory phases.
1to80%
0to35cmH Accuracy:
± (2 cmH
O + 4% of setting)
2
O(0to35hPa)
2
TRIGGER SENSITIVITY (L/min)
%O
2
Inspiratory flow required to trigger the ventilator to deliver a breath.
Percentage of inspired oxygen of the gas delivered to the patient.
1to20L/min
21 to 100% Accuracy: ± 3% full scale
NOTE:
It may take several minutes for the oxygen percentage to stabilize.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
1-19
Page 34
1 Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)
Key/indicator Specifies... Range
Other keys, knobs, and indicators
APNEA PA RA MS k ey
AllowsyoutoselectVCVorPCV(760 Ventilator only) apnea ventilation, and the apnea ventilation settings. The apnea interval is adjustable from 10 to 60 seconds. Apnea ventilation is available in all modes. For apnea parameter ranges, see mandatory VCV settings and mandatory PCV (760 Ventilator only) settings.
MENU Allows you to view active and reset alarms, run SST and EST, adjust
certain settings (including alarm volume, PCV timing, volume bar graph display, endotracheal tube size, and date and time), access oxygen sensor functions (calibrate, enable or disable % O
alarm limits, and
2
enable or disable display of oxygen sensor reading), enter standby mode, view battery information, display software revision, display service information, and enter EasyNeb nebulizer functions. (Section 6 tells you how to use the menu function.)
100% O
2
Switches the % O2to 100% for 2 minutes, then returns to the current %
setting. The 2-minute interval restarts every time you press 100%
O
2
O
. Once the 100% O2has started, you can press CLEAR to stop the
2
maneuver (unless you have entered a MENU key function or selected a setting).
MANUAL INSP Delivers one mandatory breath to the patient according to the current
mandatory settings (in A/C or SIMV) or the current apnea parameters (in SPONT). You can deliver a MANUAL INSP at any time during the exhalation phase of a breath as long as the exhaled flow is less than 30% of the peak exhaled flow, except during apnea ventilation.
1-20
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 35
.
Introduction 1
...............................................................................
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)
Key/indicator Specifies... Range
EXP PAUSE (760 only)
Allows you to measure the patient’s auto-PEEP. An EXP PAUSE maneuver causes the ventilator to close the exhalation valve at the end of the expiratory phase, and not deliver the next mandatory breath. At the end of the maneuver, the message window shows the calculated value for auto-PEEP (expiratory pressure at the beginning of the maneuver minus expiratory pressure at the end of the maneuver) and total PEEP for 30 seconds.
The message window shows the end expiratory flow in L/min at the beginning of each breath. If there is expiratory flow when the ventilator delivers the next breath, auto-PEEP is present.
The EXP PAUSE continues as long as you hold down the key, and should last only until expirator y pressure stabilizes. An EXP PAUSE maneuver ends when you release the key, the patient initiates a breath, an alarm occurs, the expiratory phase (including the maneuver) lasts more than 20 seconds, or the ventilator detects a leak.
Auto-PEEP: Range: 1 to 35 cmH Accuracy: ± (1 cmH The EXP PAUSE maneuver is unavailable when the RESPIRATORY
RATE setting is less than 3 /minute.
O.
2
O + 3% of reading).
2
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
1-21
Page 36
1 Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)
Key/indicator Specifies... Range
Other keys, knobs, and indicators (continued)
INSP PAUSE (760 only)
Allows you to measure the patient’s compliance and resistance. An extended inspiratory pause also allows you to expand the patients lungs forupto10seconds.
Pressing INSP PAUSE momentarily causes the ventilator to wait until the end of the inspiratory phase of the current or next mandatory breath (in SPONT mode, the ventilator delivers a mandatory breath using the MANUAL INSP key according to the apnea settings), stop breath delivery, and keep the exhalation valve closed. The INSP PAUSE continues until the ventilator detects a stable plateau pressure or 2 seconds have elapsed. An INSP PAUSE maneuver ends when a stable plateau is reached or an alarm occurs.
Pressing INSP PAUSE for 2 or more seconds after the pause begins causes the ventilator to deliver an inspiratory pause for as long as you hold down the key. An extended INSP PAUSE maneuver ends when you release the key or 10 seconds have elapsed.
You can press CLEAR or release the INSP PAUSE key at any time to cancel an INSP PAUSE maneuver, and alarms cancel the maneuver.
At the end of the breath, the message window shows the calculated value for compliance and resistance (if the mandatory breath was a VCV breath) or compliance (if the mandatory breath was a PCV breath) for 30 seconds.
Compliance: Range: 1 to 150 mL/H Accuracy: ± (1 mL/cmH
O.
2
O + 20% of reading).
2
Resistance: Range: 0 to 150 cmH Accuracy: ± (3 cmH
O/L/second.
2
O/L/second + 20% of reading).
2
CLEAR Pressing CLEAR before accepting a setting cancels the proposed
setting. Pressing CLEAR does not cancel accepted settings. Pressing CLEAR twice returns the ventilator to its previous state (unless
you have entered a MENU key function that requires you to press CLEAR more times).
Pressing CLEAR during a 100% O2maneuver cancels the maneuver.
1-22
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 37
.
Introduction 1
...............................................................................
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)
Key/indicator Specifies... Range
Other keys, knobs, and indicators (continued)
ACCEPT Makes changes to settings effective. If you dont press ACCEPT within
30 seconds of proposing a new setting, the user interface returns to its previous state.
Knob Adjusts the value of a setting or selects a menu option. A setting value
that flashes means that the knob is linked to that setting. Turning the knob clockwise increases the value, and turning the knob counterclockwise decreases the value.
CURRENT Lights when the ventilator is operating according to the displayed
settings, or during apnea ventilation. (There is one indicator for mandatory breaths, and one for spontaneous breaths.)
PROPOSED Lights when you propose a mode or breath type, or you are setting
apnea parameters. Once a proposed setting is accepted, it becomes effective at the next breath.
APNEA PA RA MS indicator
Message window
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
Lights when apnea ventilation is active. Lights with PROPOSED indicator when you are setting apnea
parameters, and both indicators turn off once apnea parameters are accepted.
Shows up to four lines of information (20 characters per line). First line: Reserved for the highest-priority active or autoreset alarm. On
the 760 Ventilator only, if no alarm is active and the display of the oxygen sensor reading is enabled, the % O VCV is the current or proposed breath type and there are no active or autoreset alarms, shows the selected VCV flow pattern (ramp or square).
Second line: Information about the menu function or settings, alarm silence time remaining, or current date and time. During normal ventilation, shows Flow (L/min).”
Third and fourth lines: Reserved for other messages. For every breath type, peak and end inspiratory flows are displayed on the third line, and end expiratory flow is displayed on the fourth line (except that inspiratory flow is not displayed during VCV breaths or VCV apnea ventilation).
is displayed here. Otherwise, if
2
1-23
Page 38
1 Introduction
1.3.2 PATIENT DATA
The PATIENT DATA section of the keyboard allows you to view the pressure, breath timing, and volume of the patient’s breath. You can also view the alarm settings. A lighted key indicates that a measurement is selected, and its value appears in the display window. Values are continuously displayed and updated during ventilation.
Table 1-4 summarizes the functions of the keys and indicators in the PATIENT DATA section of the keyboard.
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA
Key/
indicator
Pressure
MEAN PRESSURE
PEAK PRESSURE
PLATEAU PRESSURE
(760 only)
PEEP/CPAP (760 only)
Function Range
Shows the calculated value of ventilator breathing circuit pressure over an entire respiratory cycle. Updated at the beginning of each breath.
Shows the maximum pressure measured during inspiration. Updated at the beginning of each expiratory phase. (Default pressure display.)
Shows the pressure measured at the end of the plateau period of a mandatory inspiration (whether the inspiration is in a regular VCV breath or is part of an inspiratory pause maneuver). Updated at the beginning of each expiratory phase. The PRESSURE display shows a blank if the ventilator does not detect a stable plateau pressure.
Shows the pressure measured at the expiratory limb before any inspiratory effort. Updated at the beginning of each inspiratory phase.
0to99cmH Accuracy:
± (3 + 4% of reading) cmH2O
0to140cmH Accuracy:
± (3 + 4% of reading) cmH
0to140cmH Accuracy:
± (3 + 4% of reading) cmH
0to140cmH Accuracy:
± (3 + 4% of reading) cmH
O(0to99hPa)
2
O(0to140hPa)
2
O(0to140hPa)
2
O(0to140hPa)
2
2
2
2
O
O
O
1-24
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 39
.
Introduction 1
...............................................................................
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA (continued)
Key/
indicator
Breath timing
RATE (/min) Shows the calculated value of the total
respiratory rate, based on the previous 60 seconds or 8 breaths (whichever interval is shorter). Updated at the beginning of each breath. (Default breath timing display.)
The calculation is reset (and display is blank) when ventilation starts, when apnea ventilation starts or autoresets, when you change the mode, breath type, or RESPIRATORY RATE setting, and whenyoupressthealarmresetkey.
I:E RATIO Shows the ratio of measured inspiratory
time to measured expiratory time. Updated at the beginning of each breath.
INSP TIME(s) (760 only)
The measured inspiratory time, including breaths that are truncated due to a HIGH PRESSURE alarm. Updated at the beginning of each expiratory phase.
Function Range
1 to 199 /minute Accuracy:
± (0.1 +1% of reading)/minute
1:99.9 to 9.9:1 Accuracy:
± (0.1 + 2%)
0.1 to 9.90 seconds Accuracy:
± 0.05 seconds
Vol ume
EXHALED VOLUME (ml)
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
Shows the patients measured expiratory tidal volume averaged over the last 5 breaths (for A/C VCV breaths, ventilator­initiated PCV breaths, and PCV apnea breaths) or for the just-completed breath (for all other breaths). Corrected to BTPS and compliance-compensated. Updated at the beginning of each inspiration. (Default volume display.)
The calculation is reset when ventilation starts, when apnea ventilation starts or autoresets, when you change the mode or breath type, and when you press the alarm reset key.
0to9L Accuracy:
± (10 ml + 10% of reading)
1-25
Page 40
1 Introduction
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA (continued)
Key/
indicator
Volume (continued)
TOTAL MINUTE VOLUME (L)
DELIVERED VOLUME (ml) (760 only)
SPONT MINUTE VOLUME (L) (760 only)
Shows the patients measured expiratory minute volume, based on the previous 60 seconds or 8 breaths (whichever interval is shorter). Updated at the beginning of each breath.
The calculation is reset when ventilation starts, when apnea ventilation starts or autoresets, when you change the mode or breath type, and when you press the alarm reset key.
Shows the measured inspiratory tidal volume for the just-completed PCV or PSV breath. Corrected to BTPS and compliance-compensated. Updated at the beginning of each inspiration for PCV and PSV breath types.
Shows the patients measured expiratory minute volume for all spontaneous breaths, based on the previous 60 seconds or 8 breaths (whichever interval is shorter). Updated at the beginning of each breath.
The calculation is reset when ventilation starts, when apnea ventilation starts or autoresets, when you change the mode or breath type, and when you press the alarm reset key.
Function Range
0to99L Accuracy:
± (10 ml + 10% of reading)
0 to 3000 ml Accuracy:
± (10 ml + 10% of reading)
0to99L Accuracy:
± (10 ml + 10% of reading)
1-26
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 41
.
Introduction 1
...............................................................................
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA (continued)
Key/
indicator
Alarm settings
HIGH RATE An active alarm indicates that measured
respiratory rate is higher than the alarm setting.
HIGH TIDAL VOLUME
LOW INSP PRESSURE
LOW TIDAL VOLUME
HIGH PRESSURE
An active alarm indicates that exhaled volume for three out of four consecutive breaths was above the alarm setting.
An active alarm indicates that monitored circuit pressure is below the alarm setting at the end of inspiration. Inactive in for any spontaneous breath.
An active alarm indicates that delivered volume for three out of four consecutive breaths were below the alarm setting. (If this alarm is set to 0 ml and breath type is PCV, an active alarm indicates that delivered volume is less than 3 ml for three out of four consecutive breaths.)
An active alarm indicates that two consecutive breaths were truncated because circuit pressure reached the alarm setting.
Function Range
3 to 100 /minute Accuracy:
± (0.1 +1% of setting)/minute
20 to 6000 ml Accuracy:
± (10 ml + 10% of setting)
3to60cmH Accuracy:
± (1 + 3% of setting)
0 to 2000 ml Accuracy:
± (10 ml + 10% of setting)
10 to 90 cmH (10to90hPa)
Accuracy: ± (1 + 3% of setting)
O(3to60hPa)
2
O
2
LOW MINUTE VOLUME
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
An active alarm indicates that monitored minute volume is less than the alarm setting, based on an eight-breath running average or the previous minute, whichever is less.
0to50L Accuracy:
± (10 ml + 10% of setting)
1-27
Page 42
1 Introduction
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA (continued)
Key/
indicator
Other indicators
Pressure bar graph
Volume bar graph (760 only)
MAND Lights at the start of each breath to
Shows real-time pressures in centimeters of water (cmH2O) or hectopascals (hPa).
LEDs show the current HIGH PRESSURE alarm setting and the peak pressure of the last breath during exhalation.
Shows real-time exhaled volume in milliliters (ml). Volumes are compliance­compensated and corrected to BTPS.
The active scale is determined by the HIGH TIDAL VOLUME alarm setting. LEDs show the current HIGH TIDAL VOLUME and LOW TIDAL VOLUME alarm settings. You can use the MENU key to enable or disable the volume bar graph. During exhalation, LEDs show the maximum exhaled volume of the last breath.
indicate a ventilator- or operator-initiated (time or manually triggered) mandatory breath is being delivered.
Function Range
-10to90cmH (-10to90hPa)
Resolution: 1 cmH2O(1hPa)
If HIGH TIDAL VOLUME setting < 500 ml: 0 to 500 ml
Resolution: 5 ml
If HIGH TIDAL VOLUME setting >
Resolution: 20 ml
Not applicable
O
2
500ml:0to2000ml
ASSIST Lights at the start of each breath to
indicate a patient-initiated mandatory (flow triggered) breath is being delivered.
SPONT Lights at the start of each breath to
indicate a patient-initiated spontaneous (flow triggered) breath is being delivered.
1.3.3 VENTILATOR STATUS
The VENTILATOR STATUS section of the keyboard shows the operating condition of the ventilator, and is continuously updated during operation. Table
1-28
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Not applicable
Not applicable
Page 43
.
Introduction 1
...............................................................................
1-5 summarizes the functions of the keys and indicators in the VENTILATOR STATUS section of the keyboard.
Table 1-5: 700 Series Ventilator keyboards: VENTILATOR STATUS
Key/indicator Color Function
ALARM Red
(high priority)
CAUTION Yellow
(medium priority)
NORMAL Green Lights when no alarm condition is present.
VENT INOP Red
(high priority)
SAFETY VALVE OPEN
Red (high priority)
Flashes when a high-priority alarm is active. A repeating sequence of three, then two beeps sounds. Lights steadily when a high-priority alarm has been autoreset.
Flashes when a medium-priority alarm is active. A repeating sequence of three beeps sounds. Lights steadily when a medium-priority alarm has been autoreset.
Lights to indicate that the ventilator is inoperative, and the ventilator safety valve is open. A qualified service technician must run and pass the extended self-test (EST) before normal ventilation can resume.
If the condition that caused the safety valve to open no longer exists, and the VENT INOP indicator is off, press the alarm reset key to resume ventilation.
Lights when the ventilators safety valve and exhalation valve open and only room air is available to the patient.
Can indicate that the ventilator is inoperative, or there is an occlusion in the ventilator breathing circuit. If possible, the message window shows the alarm that triggered the safety valve open condition and how much time has elapsed since the last breath was triggered.
ON AC/ BATTERY CHARGING
ON INTERNAL BATTERY
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
Green Lights when the ventilator is running on ac
power and the battery is charging.
Yellow Flashes when the ventilator is running on the
internal battery.
1-29
Page 44
1 Introduction
Table 1-5: 700 Series Ventilator keyboards: VENTILATOR STATUS (continued)
Key/indicator Color Function
ON EXTERNAL BATTERY
INTERNAL BATTERY LEVEL
7-00423
7-00424
Yellow Flashes when the ventilator is running on the
external battery.
Green Shows the relative charge level of the internal
battery.
Ye l l o w Alarm silence: Silences the alarm sound for 2
minutes from the most recent key press.
Not applicable Alarm reset: Reestablishes all alarm indicators,
cancels the alarm silence period, and resets the patient data displays. If the condition that caused the alarm still exists, the alarm reactivates. Cancels apnea ventilation, if active. Reestablishes previous settings and ventilation resumes, unless the ventilator is inoperative.
1-30
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 45
SECTION
Setting up the ventilator 2
2
..............................................................................
This section tells you how to set up the ventilator, including:
Connecting and using the internal and external batteries
Connecting the electrical supply
Connecting the oxygen supply
Connecting the ventilator breathing circuit
Installing the collector vial
Installing the flex arm
Installing the humidifier
Using the ventilator cart
Warning
To avoid tipping or damaging the ventilator, do not stack other equipment on the ventilator. 700 Series Ventilators are designed to be mounted on either a cart or a shelf by a qualified service technician. When lifting the ventilator, lift from the base, and use assistance and appropriate safety precautions. Figure 2-1 shows proper lifting technique.
To avoid the possibility of injury to the patient and to ensure proper ventilator operation, do not attach any device to the port labeled EXHAUSTunless the device is specifically authorized by Mallinckrodt.
To minimize the increased risk of fire due to an oxygen-enriched environment, do not use the ventilator in a hyperbaric chamber.
To avoid raising the oxygen concentration of room air, use the ventilator in an adequately ventilated room.
Caution
Do not obstruct the cooling fan
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
.
2-1
Page 46
2 Setting up the ventilator
Use two people to lift
Lift ventilator from base only
7-00053
Figure 2-1. Lifting the ventilator
2.1 Connecting and using internal and external batteries
Warning
A Mallinckrodt battery must always be installed in the ventilator. Without a battery, the ventilator is not protected against low or lost ac power. Do not use the ventilator unless a battery with at least minimal charge is installed.
2-2
NOTE:
If the ventilator has been stored for an extended period, allow it to acclimate to its environment before turning it on. This helps ensure that the ventilator powers up correctly.
Every 700 Series Ventilator System includes an internal battery as a backup power supply that comes standard with the ventilator. The internal battery can provide up to 2
½ hours of backup power, depending on ventilator settings and
battery charge level. The ventilator operates on backup power when ac power is lost or drops below minimum. The internal battery’s charge level is continuously
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 47
.
Setting up the ventilator 2
-
+
...............................................................................
indicated on the keyboard (see Figure 2-2). The ventilator alarms when it determines that only 5 minutes of power remain using the current settings.
Internal battery level
7-00054
Figure 2-2. Internal battery charge indicator
The optional external battery supplies up to 7 additional hours of backup power in case ac power is lost or falls below minimum. (See Appendix B for external battery ordering information.) If installed, the external battery is the first source of backup power. The ventilator uses the internal battery if the external battery is depleted or is not connected.
NOTE:
When the ventilator switches off because battery charge is inadequate, turn off the power switch to prevent the battery from being fully depleted. Connect the ventilator to ac power as soon as possible.
The ventilator charges the internal and external batteries during ac power operation, or in standby mode when attached to ac power. (See Appendix C for battery specifications.) When the ventilator is operating on battery, you can use the MENU key to check the estimated operational time remaining until recharge is required for the internal and external batteries. (Section 6 tells you how to use the MENU key.) The ventilator must operate for 2 minutes before it can estimate battery time remaining.
You can disconnect and connect the external battery during normal operation. When operating the ventilator on battery power, you can install the external
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
2-3
Page 48
2 Setting up the ventilator
battery on the cart for easy transport. Follow these steps to connect the external battery:
1. Mount the external battery on the base of the cart (or other suitable location).
2. Plug the external battery’s cable into the connector (Figure 2-3).
3. Figure 2-4 shows you how to disconnect the external battery.
To ensure that the batteries retain their charge, store the ventilator in standby mode, with its power switch on, and connected to ac power. If the ventilator is not stored in this manner, check battery charge levels before using the ventilator.
Caution
If you plan to store the ventilator for more than 6 months, remove batteries before storage. Replace the internal battery before using the ventilator again. A qualified service technician must replace the battery according to the instructions in the 700 Series Ventilator System Service Manual.
If you turn on the ventilator after it has been unplugged for an extended period, the LOW EXT BATTERY and LOW INT BATTERY alarms may become active. If so, recharge the internal battery by leaving the ventilator plugged in and turned on in standby mode (with no patient connected) for up to 8 hours.
2-4
If the LOW EXT BATTERY, LOW INT BATTERY, or BAT NOT CHARGING alarm is still active after 8 hours, replace the batteries. If you turn off the ventilator when the battery is absent or inadequately charged, a LOSS OF POWER alarm sounds for at least 2 minutes.
NOTE:
The ON BATTERY indicator flashes when the ventilator is operating on battery power.
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 49
.
Setting up the ventilator 2
...............................................................................
External
battery
connector
External
battery
7-00055
Figure 2-3. Plugging the external battery into the ventilator
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
2-5
Page 50
2 Setting up the ventilator
7-00051
Figure 2-4. Disconnecting the external battery
2.2 Connecting the electrical supply
Warning
To avoid electrical shock hazard, connect the ventilator power cord into a grounded ac power outlet. If the integrity of the ac ground is in question, operate the ventilator from the internal or external battery.
If used in the U.S., connect the ventilator to an ac receptacle marked Hospital Onlyor Hospital Gradeto ensure grounding reliability.
2-6
Every 700 Series Ventilator System is supplied with a power cord. A power cord retainer covers the connector and socket to protect against liquid spills or accidental disconnection.
When the ventilator is ready for operation on ac power, connect the power cord to ac power (see Figure 2-5).
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 51
.
Setting up the ventilator 2
...............................................................................
NOTE:
During operation on ac power, the power cord retainer must always be in place.
To ac power
Power cord retainer
Power cord
Figure 2-5. Connecting the ventilator power cord
When the power cord is not in use, you can wrap the power cord around the brackets on the back of the ventilator for convenient storage (see Figure 2-6).
Figure 2-6. Storing the power cord on the ventilator
7-00056
7-00057
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
2-7
Page 52
2 Setting up the ventilator
2.3 Connecting the oxygen supply
Warning
To ensure proper oxygen concentration, do not obstruct the ventilators air intake.
To ensure adequate oxygen delivery to the patient, use Mallinckrodt-supplied oxygen hoses only. Use of other oxygen hoses could result in inadequate or inappropriate oxygen pressures or leaks at the oxygen inlet.
When using a cylinder oxygen supply, point the cylinder's pressure relief device away from the ventilator air intake. This helps avoid creating an oxygen-rich environment within the ventilator in the event that the cylinder oxygen regulator malfunctions.
The 700 Series Ventilator System can use oxygen from a cylinder or wall supply. Follow these steps to connect the oxygen supply:
1. Ensure that the oxygen supply pressure is between 40 and 90 psi (275 and 620 kPa) (50 psi (345 kPa) recommended).
2-8
Warning
Due to excessive restriction of certain hose assemblies (listed in Table B-1), reduced FIO pressures levels, make sure that oxygen inlet pressure is ³ 50 psi (345 kPa) when using these hose assemblies.
2. Connect the oxygen supply to the oxygen inlet connector on the side of the ventilator (see Figure 2-7). Make sure ventilator is configured with adapter(s) as required.
Caution
To prevent damage to the ventilator, ensure that the connection to the oxygen supply is clean and unlubricated.
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
< 50 psi (345 kPa) are used. To maintain correct FIO
levels may result when oxygen inlet
2
2
Page 53
.
Setting up the ventilator 2
...............................................................................
NOTE:
Whenever a pressurized oxygen source is connected to the ventilator, the oxygen regulator has a maximum bleed rate of 3 L/min, even when the ventilator is not in use. Always take this bleed rate into account when calculating oxygen cylinder volume and duration. This is especially important when using the ventilator and cart while transporting patients who are being ventilated at above-ambient
levels.
FIO
2
Oxygen inlet connector
Air intake
Oxygen hose (to oxygen supply)
7-00058
Figure 2-7. Connecting the oxygen supply
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
2-9
Page 54
2 Setting up the ventilator
2.4 Connecting the ventilator breathing circuit
Warning
To minimize the risk of bacterial contamination or component damage, inspiratory and expiratory filters must always be handled with care and connected to the ventilator during use.
Use of an external pneumatically powered nebulizer with a 700 Series Ventilator results in added flow to the ventilator breathing
circuit. This additional flow can affect delivered tidal volumes, spirometry and spirometry alarms, delivered FIO triggering. (However, the Mallinckrodt EasyNebNebulizer is designed to deliver aerosol medications to the patient without affecting ventilator performance or patient data.)
To minimize the risk of patient injury, use only ventilator breathing circuits qualified for use in oxygen-enriched environments with 700 Series Ventilators. Do not use antistatic or electrically conductive tubing. To ensure a leak-tight connection, only use connectors and tubes with ISO-standard cone and socket fittings (or use adapters to connect barbed cuff fittings to ISO-standard fittings).
Use ventilator breathing circuits identified by Mallinckrodt for use with 700 Series Ventilators or circuits that ensure that the maximum resistance values specified in Table C-5 are not exceeded. Using a circuit with a higher resistance does not prevent the ventilator from ventilating as long as the ventilator passes SST.
, and patient
2
2-10
Caution
Adding accessories to the ventilator can increase system resistance. Ensure that any changes to the recommended ventilator circuit configurations do not exceed the values for resistance provided in
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Appendix C.
Page 55
.
Setting up the ventilator 2
...............................................................................
NOTE:
Mallinckrodt recommends that you run SST every 15 days, between patients, and when you change the ventilator breathing circuit. Mallinckrodt recognizes that the protocol for running SST varies widely among healthcare institutions. It is not possible for Mallinckrodt to specify or require specific practices that will meet all needs, or to be responsible for the effectiveness of those practices.
Theventilatorusesanoxygensensortotriggeranalarmifthe delivered % O setting. You can view the ventilators%O2measurement by using the User settings function of the MENU key.
Check the inspiratory and expiratory limbs of the ventilator breathing circuit and the collector vial and in-line water traps regularly for water buildup. Under certain conditions, they can fill quickly. In case of a partial occlusion alarm, check for water in the ventilator breathing circuit. In very severe cases, such a water occlusion can potentially cause the ventilator to enter occlusion cycling mode. Empty and clean the collector vial and in-line water traps as necessary.
Mallinckrodt recommends using pediatric circuits when ventilating patients with 5-mm or smaller internal-diameter artificial airways.
is 10 percentage points above or below the % O
2
2
Figure 2-8 shows how to connect the ventilator breathing circuit, including the inspiratory filter, humidifier (if used), inspiratory limb, patient wye, expiratory limb, collector vial, and expiratory filter.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
2-11
Page 56
2 Setting up the ventilator
Ventilator
(from patient)
Expiratory filter
(arrow on filter
indicates direction
of flow)
Collector vial
Expiratory
limb of the
ventilator
breathing circuit
Ventilator (to patient)
Inspiratory filter (arrow on filter indicates direction of flow)
Humidifier
Inspiratory limb of the ventilator breathing circuit (install EasyNeb between inspiratory filter and patient wye)
Patient wye
7-00059
Ô
2-12
Figure 2-8. Connecting the ventilator breathing circuit
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 57
.
Setting up the ventilator 2
...............................................................................
2.5 Installing the collector vial
Figure 2-9 shows you how to slide the collector vial into the bracket on the cart or shelf mounting plate.
Bracket (on cart or shelf-mounting plate)
Collector
vial
7-00060
Figure 2-9. Installing the collector vial
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
2-13
Page 58
2 Setting up the ventilator
2.6 Installing the flex arm
Figure 2-10 shows you how to install the flex arm onto the threaded socket on the ventilator.
Flex arm
2-14
7-00061
Figure 2-10. Installing the flex arm
NOTE:
If the flex arm is too large to fit into a tight space, you can shorten the arm by removing the middle segment (Figure 2-11).
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 59
.
Setting up the ventilator 2
...............................................................................
Remove
Remove
Remove
7-00062
Figure 2-11. Shortening the flex arm
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
2-15
Page 60
2 Setting up the ventilator
2.7 Installing the humidifier
Figure 2-12 shows you how to slide the humidifier into the bracket on the cart or shelf mounting plate.
NOTE:
The humidifier must be powered independently of the ventilator (regardless of whether the ventilator is running on ac or battery power).
Bracket (on cart or shelf-mounting plate)
2-16
Humidifier
7-00063
Figure 2-12. Installing the humidifier
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 61
.
Setting up the ventilator 2
...............................................................................
2.8 Using the ventilator cart
The ventilator cart is equipped with a storage area for the external battery. You can lock and unlock the cart’s front wheels with your foot, as shown in Figure 2-13.
Caution
To prevent equipment damage, pull, rather than push, the ventilator over high thresholds when using the cart to transport the ventilator. Ensure that the flex arm is positioned at the front of the ventilator.
Ventilator cart base
Unlocked
Locked
Unlocked
7-00064
Figure 2-13. Locking and unlocking the carts front wheels
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
2-17
Page 62
2 Setting up the ventilator
2-18
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 63
SECTION
Getting started 3
3
..............................................................................
This section tells you how to:
Power up the ventilator
Select ventilator settings
View and change alarm limits
Enter and exit standby mode
NOTE:
Mallinckrodt recommends that before using the ventilator for the first time, you wipe the ventilator exterior clean and sterilize its components according to the instructions in Appendix A. Follow your institutions protocol for cleaning and sterilizing the ventilator and its components.
3.1 Powering up the ventilator
NOTE:
To ensure battery charge(s), store the ventilator in standby mode with the power cord plugged into the mains and the power switch in the on (I) position whenever possible. The power switch controls ac mains and battery power to the ventilator.
1. Turn the power switch on (see Figure 3-1). The ventilator displays this message:
POST running... NPB 7x0 S/W Rev x PM due xxxxx hrs
“PM due” refers to the number of hours until a routine preventive maintenance procedure is due, based on the ventilator’s operational hours.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
3-1
Page 64
3 Getting started
Warning
If a single beep is not audible during POST, the main audible alarm may be malfunctioning. Remove the ventilator from use and contact service.
NOTE:
If the ventilator is running on ac but the battery is low, upon power up it may begin ventilating using the last valid settings and an ABNORMAL RESTART alarm may occur.
Do not switch off ventilator power during POST.
3-2
7-00065
Figure 3-1. Turning the power switch on (the “I” position)
2. Once the power-on self-test (POST) is complete (several seconds), the ventilator displays the message:
Accept settings to start ventilation
The last valid settings are proposed, and the appropriate indicator lights to show the ventilator’s source of power.
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 65
.
Getting started 3
...............................................................................
Warning
Always let the ventilator run for 10 minutes before placing it on a patient to allow heaters to warm up. Failure to warm up the ventilator could result in failure of the SST flow sensor test, and higher measured exhaled tidal volumes.
Caution
Mallinckrodt recommends running short self-test (SST) and testing alarms to check proper operation of the ventilator before ventilation begins or according to your institution’s protocol. Section 4 tells you how to run SST. Appendix E tells you how to test alarms.
NOTE:
If the ventilator detects low oxygen pressure, a low battery level, or an occluded or missing air intake filter during POST, POST continues uninterrupted. These alarms become active when POST is complete, and you can override them by pressing the alarm reset key.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
3-3
Page 66
3 Getting started
3.2 Selecting ventilator settings
NOTE:
Each keypress should result in a blipsound; otherwise, the main alarm speaker may be faulty. If a keypress does not result in this sound, turn the ventilator off then on to run POST and verify the proper functioning of the speaker.
1. Select the mode: To select a mode, press A/C, SIMV, or SPONT. (Appendix D gives a detailed explanation of these ventilation modes.)
In assist/control (A/C) mode, you can only select volume
A/C
SIMV
SPONT
control ventilation (VCV) or pressure control ventilation (PCV) settings (PCV is only available on the 760 Ventilator).
In synchronous intermittent mandatory ventilation (SIMV) mode, you can select both VCV or PCV (760 only) and pressure support ventilation (PSV) settings.
In spontaneous (SPONT) mode, you can only select PSV settings.
3-4
2. Confirm or change settings: For every flashing key light, you must touch the key (and adjust the setting if necessary) before you can apply the settings. You can confirm or change settings in any order.
Once you press a key, the proposed setting flashes to indicate that the knob is linked to the selected setting. The setting’s LCD window shows the setting and the message window displays the current setting, range, units, name of the setting, and associated parameter. Turn the knob to change the setting or press another key
Press each flashing key and turn the knob to adjust.
7-00066
to leave the setting as is. (The setting then stops flashing.)
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 67
.
Getting started 3
...............................................................................
3. Press CLEAR to cancel the most recently changed setting and return to the last valid setting. (This only works if you haven’t pressed ACCEPT yet.) Pressing CLEAR after you’ve pressed 100% O maneuver. The ventilator will beep if you attempt to change a setting to a value above or below the setting range.
stops the 100% O
2
2
4. PEEP/CPAP, TRIGGER SENSITIVITY, and % O
: You can
2
change these settings in any mode. These keys don’t flash when you select a new mode, and you are not required to confirm these settings.
NOTE:
The O
% alarm is delayed for 30 seconds and 12 L of tidal volume
2
following a change in setting so that the oxygen concentration can stabilize.
5. Apply the settings: Press ACCEPT.
ACCEPT
NOTE:
If apnea ventilation is possible, the ventilator displays this message when normal ventilation begins:
PEEP/CPAP
TRIGGER
SENSITIVITY
L/min
%O2
7-00067
Review Ta = xx s
where Ta is the apnea interval and xx is the current setting in seconds (from 10 to 60 s). Section 5 tells you how to adjust apnea ventilation parameters. You are not required to review apnea parameters before normal ventilation begins.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
3-5
Page 68
3 Getting started
Ventilation begins according to the displayed settings and the existing alarm limits. You should review all alarm limits and change them if necessary. (Section 3.3 tells you how.) You can change alarm limits or ventilator settings at any time during normal ventilation.
Warning
In PSV it is impossible to set support pressure + PEEP to deliberately cause a HIGH PRESSURE alarm. Although Mallinckrodt does not recommend doing so, it is possible in VCV to set a tidal volume that causes breath truncation and a HIGH PRESSURE alarm. To avoid activating a HIGH PRESSURE alarm, select the appropriate TIDAL VOLUME and HIGH PRESSURE alarm settings.
3.3 Viewing and changing alarm settings
You can view alarm settings at any time without affecting ventilation by pressing the appropriate key. The ALARM SETTINGS window displays the value of the alarm being viewed or changed. The alarm setting flashes for up to 30 seconds, and you can change an alarm setting any time its setting is flashing in the window.
3-6
All alarm settings are in effect during ventilation (except for LOW INSP PRESSURE, which is not active in SPONT mode). You can change alarm settings at any time during ventilation. You can only change one alarm setting at a time.
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 69
.
Getting started 3
...............................................................................
To change a setting: touch, turn, ACCEPT.
1. Touch the key: The key light turns on, and the current alarm setting flashes in the window for 30 seconds. The message window shows the current setting, range, and units of the alarm.
2. Turn the knob to adjust the alarm setting (this only works when the alarm setting is flashing in the window). The ventilator will beep if you attempt to select a setting above or below the setting range.
3. Press ACCEPT to apply the new alarm setting. The new alarm setting stops flashing and the message window displays this message:
Setting(s) accepted
Warning
Inadvertent extubation in which the endotracheal tube remains connected to the breathing circuit could result in pressurization above the LOW INSP PRESSURE alarm setting. The LOW INSP PRESSURE and low volume alarms should be set appropriately.
To ensure patient safety, always set the LOW TIDAL VOLUME and HIGH RATE alarms.
ACCEPT
7-00066
7-00068
7-00067
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
3-7
Page 70
3 Getting started
NOTE:
The O alarm limit is 10 percentage points above the set % O O measured % O enabled (using the Oxygen sensor menu function, described in Section 6).
3.4 Entering and exiting standby mode
Standby mode is a waiting state in which there is no ventilation, but the ventilator maintains its settings and battery charge(s).
Warning
To avoid raising the oxygen concentration of room air, ensure that
To avoid patient injury due to lack of ventilatory support, do not
% alarm limits are based on the set % O2. The high O2%
2
% alarm limit is 10 percentage points below the set % O2.The
2
isshowninthemessagewindowifitsdisplayis
2
, and the low
2
the ventilator is in an adequately ventilated room if connected to an oxygen supply while in standby mode. Disconnect the oxygen supply if you do not intend to use the ventilator immediately.
enter standby mode with a patient attached to the ventilator. You must confirm that no patient is attached before entering standby mode.
3-8
NOTE:
To maintain battery charge(s), the ventilator must be plugged into ac power with the power switch on while in standby mode.
Follow these steps to enter standby mode:
1. Turn on the ventilator. If ventilation has already begun, you must turn off the ventilator, then turn it back on without starting ventilation. This allows you access to the Standby mode menu function.
2. Press MENU.
3. Turn the knob to select standby mode.
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 71
.
Getting started 3
...............................................................................
4. Press ACCEPT. The ventilator displays this message:
Is pt disconnected? ACCEPT to proceed
5. Confirm that no patient is attached by pressing ACCEPT. (If you don’t press ACCEPT while this message is displayed, the ventilator will not enter standby mode.)
6. This message tells you that the ventilator is in standby mode:
In standby mode Clear to exit
The ON AC/BATTERY CHARGING, INTERNAL BATTERY LEVEL, and SAFETY VALVE OPEN indicators turn on. All other key lights and displays are off during standby mode. The only alarms that can become active during standby mode are FAN FAILED ALERT, BAT NOT CHARGING, LOSS AC POWER, and LOSS OF POWER.
Follow these steps to exit from standby mode and resume ventilation:
1. Press CLEAR.
2. The ventilator displays this message:
POST running...
3. After a few seconds, POST is complete and the ventilator begins ventilation using the last valid settings.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
3-9
Page 72
3 Getting started
3-10
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 73
SECTION
Self tests (SST and EST) 4
4
..............................................................................
All 700 Series Ventilators include these built-in self tests:
Short self test (SST) is a short (about 2 to 3 minutes) and simple sequence of
tests that verifies proper ventilator operation, checks the ventilator breathing circuit for leaks, and measures the circuit compliance. Power on self test (POST) is part of SST.
Extended self test (EST) is a thorough test of the operational integrity of the
ventilator, both the electronics and the pneumatics. POST and SST are part of EST.
Warning
Before running SST or EST, you must disconnect the ventilator from the patient. Running SST or EST while the ventilator is connected to the patient can injure the patient.
A fault identified in SST or EST indicates that the ventilator or an associated component is defective. A defective ventilator or associated component should be repaired before the ventilator is returned to service, unless it can be determined with certainty that the defect cannot create a hazard for the patient, or add to the risks which may arise from other hazards.
Caution
To ensure accurate SST and EST operation, run SST and EST in room air (% O at least 10 minutes. (EST also requires an oxygen supply.)
Table 4-1 describes SST and EST.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
is 21%) after the ventilator has been powered on for
2
4-1
Page 74
4 Self tests (SST and EST)
Tabl e 4-1: 700 Series Ventilator self tests
Test
name
Short self test (SST)
Extended self test (EST)
Purpose When test is run
An abbreviated version of EST to be run primarily by the operator. Makes detailed checks of the pneumatics and electronics. It also characterizes system leaks and system/ tubing compliance to compensate during breath delivery. POST is run as part of SST.
Thoroughly tests the operational integrity of the ventilator, both the electronics and pneumatics. POST and SST are part of EST.
Before patient is connected to the ventilator or after ventilator breathing circuit or humidifier is changed.
When the ventilator is serviced.
As part of the ventilator performance verification.
Following a ventilator inoperative.
NOTE:
The performance verification, described in the 700 Series Ventilator System Service Manual, is a more thorough test of the ventilator to
verify specifications are met. The technician runs a partial or full performance verification after servicing the ventilator and at regular intervals.
4-2
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 75
.
Self tests (SST and EST) 4
...............................................................................
4.1 Short self test (SST)
Table 4-2 summarizes the functions of SST.
NOTE:
Mallinckrodt recommends that you run SST every 15 days, between patients, and when you change the ventilator breathing circuit. Mallinckrodt recognizes that the protocol for running SST varies widely among healthcare institutions. It is not possible for Mallinckrodt to specify or require specific practices that will meet all needs, or to be responsible for the effectiveness of those practices.
When SST asks you to do something (for example, press ACCEPT or block the patient wye), the ventilator will wait indefinitely for your response.
Follow these steps to run SST:
1. Turn on the ventilator. If ventilation has already begun, you must turn off the ventilator, then turn it back on without starting ventilation. This allows you access to the Self test menu function.
If you select the Self test function during ventilation the window displays this message:
N/A in ventilation
2. Press MENU, turn the knob to select Self test, then press ACCEPT. The usual 30-second timeout for your input is disabled during the entry to the Self test function.
3. The window displays this message:
Shortselftest Turn knob to view
4. Turn the knob to select Short self test (you can also select Extended self test). The window displays this message:
Shortselftest Vent warming 09:59 CLEAR to bypass
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
4-3
Page 76
4 Self tests (SST and EST)
If the ventilator has already been running for more than 10 minutes before you select Short self test, this message is bypassed (see step 5). If the ventilator was recently running and is already warmed up, you can press CLEAR to override the Ve nt w ar min g message and begin SST. Mallinckrodt cannot guarantee the accuracy of test results in this case, however.
5. The window displays this message:
Is pt disconnected? ACCEPT to proceed
6. Press ACCEPT to confirm that no patient is connected to the ventilator. (Or press CLEAR to back up through the menu.) The window displays this message:
Block wye
7. Block the patient wye and then press ACCEPT.
8. The ventilator runs power-on self-test (POST) and displays this message for a few seconds:
POST running...
Warning
If a single beep is not audible during POST, the main audible alarm may be malfunctioning. Remove the ventilator from use and contact service.
4-4
NOTE:
If the ventilator is running on ac but the battery is low, upon power up it may begin ventilating using the last valid settings and an ABNORMAL RESTART alarm may occur.
During POST it is recommended that you not switch off ventilator power.
9. The message window then displays this message:
humidification device type
Choose humidifier
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 77
.
Self tests (SST and EST) 4
...............................................................................
where the humidification device type is the most recently selected type, as follows:
HME: Heat moisture exchanger (“artificial nose”)
Dual heated wire: Humidifier with heated wire on expiratory limb, or on both inspiratory and expiratory limbs
No heated wire: Conventional humidifier without heated wire circuit on the expiratory limb
Turn the knob to select the humidifier type attached, then press ACCEPT.
Warning
Incorrectly specifying the humidifier type during SST can affect the accuracy of spirometry calculations.
10. The message window then displays this message:
tubing type
Choose tubing type
where the tubing type is the most recently selected ventilator breathing circuit type.
Turn the knob to select the ventilator breathing circuit type (adult or pediatric tubing) attached to the ventilator, then press ACCEPT.
Warning
Incorrectly specifying the ventilator breathing circuit type during SST can cause an inappropriate sensitivity for the leak test and occlusion alarm. Mallinckrodt recommends using pediatric circuits when ventilating patients with 5-mm or smaller internal-diameter artificial airways.
NOTE:
To ensure that compliance compensation functions correctly, you must run SST with the circuit configured as intended for use on the patient.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
4-5
Page 78
4 Self tests (SST and EST)
11. The message window then displays this message:
ET size: x.x mm Choose ET size
Turn the knob to select the appropriate ET (endotracheal tube) size, then press ACCEPT.
Warning
Specifying an ET tube size that is too large can cause premature termination of breaths on very small pediatric patients. During normal ventilation, you can use the MENU key’s User settings function to change the ET tube size (see Section 6).
12. The ventilator automatically starts the test sequence. A message indicates which test is being performed. Once started, a test cannot be interrupted. In most cases you don’t need to do anything unless a test result is fault or fail, or your action is required. When a test passes, SST automatically starts the next test.
NOTE:
The PRESSURE bar graph and the PRESSURE display window show the real-time expiratory pressure throughout SST.
4-6
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 79
.
Self tests (SST and EST) 4
...............................................................................
These keys are valid during SST:
To confirm that a requested action is complete.
ACCEPT
To repeat a test, return to the prompt at the start of a test, or indicate an
CLEAR
inaudible alarm (during the Main alarm speaker test and Backup alarm test only).
MANUAL
INSP
7-00424
7-00423 7-00072
To override an Incomplete or Fault test result.
To restart from the beginning of SST.
TostoptestingandskiptotheendofSST.
NOTE:
If the tubing or humidifier type or ET tube size that youve selected has changed from the previous time SST was run, this key is temporarily disabled until the Circuit comp test is complete.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
4-7
Page 80
4 Self tests (SST and EST)
13. The I/E filter test prompts you to complete these actions (press ACCEPT when each action is complete):
Display:
Disconnect I tubing Reconnect I tubing
Display:
Unblock wye Block wye
Display:
Disconnect E tubing Reconnect E tubing
(Do not disconnect inspiratory filter.)
+
-
740 Ventilator
7-00069
+
-
740 Ventilator
7-00070
(Do not disconnect expiratory filter.)
+
-
4-8
740 Ventilator
7-00071
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 81
.
Self tests (SST and EST) 4
...............................................................................
14. When the final test in the sequence is complete, this message is displayed along with the overall SST result:
SST finished testing
Table 4-3 summarizes overall SST results and how to proceed in each case.
15. Unblock the patient wye, then press ACCEPT. The ventilator reruns POST.
Table 4-2: SST sequence of tests
Test Function Comments
Is pt disconnected?
Leak test Checks the ventilator breathing
P
ePcyl
compare
Auto-zero sol test
Circuit comp test
Safety valve test
Asks for confirmation that a patient is not connected to the ventilator.
circuit for leaks. Displays expiratory pressure on pressure bar graph.
Verifies proper function of pressure transducers in internal pneumatics and ventilator breathing circuit. Displays expiratory pressure on pressure bar graph.
Verifies the proper function (ability to open and close) of
(expiratory) and P
the P
e
(cylinder) solenoids.
Determines the compliance of the ventilator breathing circuit. Displays expiratory pressure on pressure bar graph.
Verifies that safety valve relieves excess circuit pressure. Displays expiratory pressure on pressure bar graph.
cyl
SST cannot proceed until you confirm that a patient is not connected.
Fault if not passed. Overriding a fault could cause
improper compliance compensation, inaccurate tidal volume delivery, or autocycling.
Fault if not passed. Overriding a fault could cause
improper compliance compensation, or autocycling.
Failu re if not passed.
Fault if not passed.
Overriding a fault could cause improper compliance compensation or inaccurate tidal volume delivery.
Failu re if not passed.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
4-9
Page 82
4 Self tests (SST and EST)
Table 4-2: SST sequence of tests (continued)
Test Function Comments
PEEP system test
Flow/O
2
sensor test
I/E filter test Checks the pressure drop
Heaters test Verifies correct operation of
Verifies that the PEEP system can generate and maintain preset PEEP levels. Displays expiratory pressure on pressure bar graph.
Checks the accuracy of the exhalation flow sensor and the oxygen sensor.
across the inspiratory and expiratory limbs of the entire patient system. Checks the pressure drop across the filters.
the heaters in the exhalation subsystem.
Failu re if not passed.
Failu re if flow sensor test does not
pass. Fault if oxygen sensor test does not
pass. Use the MENU key’sOxygen sensor function to recalibrate the oxygen sensor (see Section 6).
Overriding a fault (oxygen sensor test only) could result in inaccurate % O measurement.
Fault if not passed. This test is optional: pressing CLEAR at the initial prompt skips this test.
Skipping this test or overriding a fault could result in inadequate bacteria protection or excessive resistance to inspiration or exhalation.
Fault if not passed. Overriding a fault could cause
inaccurate monitoring of exhaled volume or lead to damage to the ventilators flow sensor or exhalation valve.
2
4-10
Main alarm test
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Sounds alarm at three volumes and verifies that user reports hearing it.
Failu re if not passed. If alarm is audible press ACCEPT. If
not, press CLEAR.
Page 83
.
Self tests (SST and EST) 4
...............................................................................
Table 4-2: SST sequence of tests (continued)
Test Function Comments
Backup alarm test
Checks the operation of the backup alarm circuit.
NOTE:
Faults detected during SST may require removing the ventilator from clinical use immediately and servicing it. A trained operator, however, must evaluate the situation and determine whether the ventilator can be used. Failures detected during SST require rerunning SST (using the alarm reset key). It is not possible to exit SST if SST fails. If SST continues to fail, the ventilator must be removed from service
Fault if not passed. If alarm is audible pressACCEPT.Ifnot,pressCLEAR.
Overriding a fault could cause no backupaudiblealarmincasethemain audible alarm fails.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
4-11
Page 84
4 Self tests (SST and EST)
Table 4-3: Overall SST results
If the SST
result is:
SST passed All tests passed. Press ACCEPT to exit SST or press
SST incomplete
SST fault One or more tests did not pass.
SST failed One or more critical problems
*If you press MANUAL INSP to exit SST, you will be prompted to confirm that you wish to override the fault or incomplete test. Confirm by pressing MANUAL INSP again.
All tests performed were passed, but some were skipped. The skipped tests were passed on a previous run.
These tests might not compromise the ventilators ability to ventilate safely, based on the operator’s evaluation of the situation.
were detected.
It means: Do this:
the alarm reset key to restart SST from the beginning.
Press the alarm reset key to restart SST from the beginning or press MANUAL INSP to override this result and exit SST.*
Press the alarm reset key to restart SST from the beginning or press MANUAL INSP to override this result and exit SST.*
Press the alarm reset key to rerun SST. It is not possible to exit SST if SST fails. If SST continues to fail, remove the ventilator from use and contact service.
4-12
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 85
.
Self tests (SST and EST) 4
...............................................................................
4.2 Extended self test (EST)
To run EST, you must have the hardware listed in Table 4-4.
Table 4-4: EST hardware requirements
Hardware description
Ventilator breathing circuit
Stopper, wye (no. 2) G-061574-00 or local supplier
Stopper, inspiration port (no. 3)
Oxygen source, 40 to 90 psi (275 to 620 kPa)
Warning
Due to excessive restriction of Air Liquide, Australian, and Dräger hose assemblies, reduced FIO Make sure oxygen inlet pressure is ³ 50 psi (345 kPa) when using these hose assemblies, to maintain correct FIO
levels can result if you use oxygen inlet pressures < 50 psi (345 kPa).
2
To use as test circuit: G-061208-00 or equivalent (adult, reusable, without heated wire)
NOTE:
To ensure that compliance compensation functions correctly, the user must run EST or SST with the circuit configured as intended for use on the patient.
G-061575-00 or local supplier
Local supplier
Manufacturer/model or Mallinckrodt
part number
levels.
2
NOTE:
Mallinckrodt recommends that you always run the full EST before placing the ventilator into operation following service and as part of the ventilators routine performance verification. For more information on performance verification, see the 700 Series Ventilator System Service Manual.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
4-13
Page 86
4 Self tests (SST and EST)
Follow these steps to run EST:
1. Set up the ventilator as for normal operation, complete with humidifier, if applicable, and leak-tight ventilator breathing circuit.
2. Turn on the ventilator. If ventilation has already begun, you must turn off the ventilator, then turn it back on without starting ventilation. This allows you access to the Self test menu function.
3. Press MENU, turn the knob to select Self test,thenpressACCEPT.
4. Turn the knob to select Extended self test. The window displays the messages shown in Table 4-5.
Table 4-5: EST setup messages
Message Description
Extended self test Vent war mi ng xx:xx CLEAR to bypass
Is pt disconnected? Confirm that patient is disconnected by pressing ACCEPT; or press
Block wye Install no. 2 stopper and then press ACCEPT.
POST running... Wait a few seconds until POST is completed.
This message is displayed (followed by a countdown) if you just turned on the ventilator. The ventilator must be on for 10 minutes before running EST so that component temperatures can stabilize.
If the ventilator was recently running and is already warmed up, you can press CLEAR to override the Vent warming message and begin SST. Mallinckrodt cannot guarantee the accuracy of test results in this case, however.
CLEARtoreturntoservicemenu.
NOTE:
If a single beep is not audible during POST, the main audible alarm may be malfunctioning. Contact service.
During POST do not switch off ventilator power.
4-14
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 87
.
Self tests (SST and EST) 4
...............................................................................
Table 4-5: EST setup messages (continued)
Message Description
humidification device type
Choose humidifier
tubing type
Choose tubing type
Turn knob to select desired humidification device, then press ACCEPT. Device types include: HME (heat-moisture exchanger or artificial nose), Dual heated wire (humidifier with heated wires on both inspiratory and expiratory limbs), or No heated wire (humidifier without a heated wire on expiratory limb).
Warning
Incorrectly specifying the humidifier type during EST can affect the accuracy of spirometry calculations.
Turn knob to select either Adult tubing or Pediatric tubing and then press ACCEPT.
Warning
Incorrectly specifying the ventilator breathing circuit type during EST can cause an inappropriate sensitivity for the leak test and occlusion alarm. Mallinckrodt recommends using pediatric circuits when ventilating patients with 5-mm or smaller internal-diameter artificial airways.
NOTE:
To ensure that compliance compensation functions correctly, run EST with the circuit configured as intended for use on the patient.
ET size: x.x mm Choose ET size
Turn knob to select appropriate ET (endotracheal tube) size, and then press ACCEPT.
Warning
Specifying an ET tube size that is too large can cause premature termination of breaths on very small pediatric patients. During normal ventilation, you can use the MENU key’s User settings function to change the ET tube size (see Section 6).
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
4-15
Page 88
4 Self tests (SST and EST)
5. The ventilator automatically starts the test sequence (Table 4-6). Unless a test requires your intervention (and waits indefinitely for your response), you don’t need to do anything until a test result is fault or failed, or EST is complete. Use the keys listed in Table 4-7 as needed.
Table 4-6: EST test sequence
Test number
and name
1. DAC-ADC loop test
2. Safety valve test
3. Motor sensor test
4. Zeroing of Po
5. O
2
solenoids test
Function Comments
A loopback test to check digital-to-analog (D/A) and analog-to-digital (A/D) convertersoperation.
Verifies that safety valve relieves excess circuit pressure.
Checks operation of rotary encoder.
Zeroes oxygen regulator pressure transducer.
Disconnect and reconnect oxygen supply, as directed.
Press ACCEPT to signify you have performed each requested action.
Checks oxygen solenoids and oxygen sensor.
Failure if not passed. Not in SST.
Failure if not passed. Also in SST.
Failure if not passed. Not in SST.
Faultifnotpassed.NotinSST.
NOTE:
Overriding this fault could result in inaccurate oxygen percentage delivery.
Faultifnotpassed.NotinSST.
NOTE:
Overriding this fault could result in inaccurate oxygen percentage delivery.
4-16
6. Flow sensor test
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Checks accuracy of exhalation flow sensor.
Failure if not passed. Also in SST.
Page 89
.
Self tests (SST and EST) 4
...............................................................................
Table 4-6: EST test sequence (continued)
Test number
and name
7. Leak test Checks ventilator breathing circuit for leaks.
7. PeP
cyl
compare (also part of test 7)
7. Auto-zero
sol test (also part of test 7)
Verifies proper functioning of pressure transducers and ventilator breathing circuit.
Verifies the proper function of the
and P
P
e
Function Comments
solenoids.
cyl
Faultifnotpassed.AlsoinSST.
NOTE:
Overriding this fault could cause improper compliance compensation, inaccurate tidal volume delivery, or autocycling.
Faultifnotpassed.AlsoinSST.
NOTE:
Overriding this fault could cause improper compliance compensation, inaccurate tidal volume delivery, or autocycling.
Failure if not passed. Also in SST.
NOTE:
Overriding this fault could cause improper compliance compensation, inaccurate tidal volume delivery, or autocycling.
8. Circuit
comp test
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
Determines ventilator breathing circuit compliance.
Faultifnotpassed.AlsoinSST.
4-17
Page 90
4 Self tests (SST and EST)
Table 4-6: EST test sequence (continued)
Test number
and name
9. I/E filter test Checks pressure drop across inspiratory and expiratory limbs of entire patient system. Checks pressure drop across filters.
Disconnect and reconnect inspiration (I) tube from/to inspiratory filter outlet, and exhalation (E) tube from/to expiratory filter inlet, as directed. Block and unblock wye with a no. 2 stopper, as directed. Press ACCEPT to signify you have performed each requested action.
10. PEEP
system test
Verifies that PEEP system can generate and maintain preset PEEP levels within either of two sets of limits. If results are within the inner set of limits, test passes the first time. If the results are outside of the outer set of limits, test fails the first time. If the results are outside the inner limits but within the outer limits, the ventilator adjusts its calibration table in nonvolatile RAM (NVRAM) and repeats the test to verify the calibration. The test fails if the calibration cannot be verified after five attempts.
Function Comments
Faultifnotpassed.AlsoinSST. Mandatory in EST, but optional in SST.
NOTE:
Overriding this fault could result in inadequate bacteria protection or excessive resistance to inspiration or exhalation.
Failure if not passed. Also in SST, but not identical.
4-18
11. Check
valve test
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Checks check valves on inlet and outlet por ts of cylinder and on exhalation limb of ventilator breathing circuit.
Failure if not passed. Not in SST.
Page 91
.
Self tests (SST and EST) 4
...............................................................................
Table 4-6: EST test sequence (continued)
Test number
and name
12. Piston leak test
13. Lights­displays test
14. Keys test Verifies that every row and
Checks piston leak against a table of leak values stored in NVRAM.
First remove inspiratory filter; then block and unblock inspiratory filter port as directed, using no. 3 stopper. Press ACCEPT to signify you have performed each requested action. Replace inspiratory tubing at end of test.
Verifies that lights and displays function.
Observe each named display, and press ACCEPT to verify that it is lit. Press CLEAR if a display is not lit.
column of key matrix can be correctly read.
Press each named key (each keys LED will also be lit). Press CLEAR if a key’s LED is not lit.
Function Comments
Failure if not passed. Not in SST.
Failure if not passed. Not in SST.
Failure if not passed. Not in SST.
15. Main alarm test
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
Checks whether various alarm sounds are audible (at maximum volume).
Listen for each named sound, then press ACCEPT to verify that it is audible. Press CLEAR if a sound is not audible.
Failure if not passed. Also in SST.
4-19
Page 92
4 Self tests (SST and EST)
Table 4-6: EST test sequence (continued)
Test number
and name
16. Backup alarm test
17. Fan test Verifies that main fan is operating
18. Heaters test
Checks operation of backup alarm circuit.
Press ACCEPT if alarm is audible or press CLEAR if alarm is not audible.
correctly by reading fan status bits.
Verifies correct operation of heaters in exhalation system.
Function Comments
Faultifnotpassed.AlsoinSST.
NOTE:
Overriding this fault could result in no audible alarm if the main speaker fails.
Failure if not passed. Not in SST.
Faultifnotpassed.AlsoinSST.
NOTE:
Overriding this fault could result in inaccurate monitoring of exhaled volume or damage to the ventilator’sflowsensoror exhalation valve.
4-20
At the end of each test, the test name and pass/fail/fault status are displayed. When a test passes, EST automatically starts the next test. You do not need to do anything unless a test result is Fault or Fa il, or your action is required. If a test does not pass, you can still continue EST, although the overall EST result is not passed until the error condition is corrected. Since information on multiple errors can help troubleshooting, it can be useful to complete EST even with errors.
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 93
.
Self tests (SST and EST) 4
...............................................................................
Table 4-7: Key functions during EST
Key Function
To confirm that a requested action is complete.
ACCEPT
Repeat a test or return to prompt at start of a test.
CLEAR
MANUAL
INSP
7-00424
7-00423 7-00072
To override an Incomplete or Fault test result.
To restart from the beginning of SST.
To stop testing and skip to the end of SST.
NOTE:
IfthetubingorhumidifiertypeorETtubesizethatyou’ve selected has changed from the previous time SST was run, this key is temporarily disabled until the Circuit comp test is complete.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
4-21
Page 94
4 Self tests (SST and EST)
6. Some tests display additional prompts, some of which require your response. These are listed in Table 4-8.
Table 4-8: Prompts during EST
Prompt Your response
Disconnect O2 supply Disconnect ventilator from oxygen supply. Press ACCEPT.
Reconnect O2 supply Reconnect ventilator to oxygen supply. Press ACCEPT.
Disconnect I tubing Disconnect tubing from inspiratory filter outlet. Leave inspiratory
filter in place. Press ACCEPT.
Idrop=xx.xat 100
or
Idrop=xx.xat 40 Reconnect I tubing
Unblock wye Remove stopper from patient wye. Press ACCEPT.
Block wye Insert a stopper into patient wye. Press ACCEPT.
Disconnect E tubing Reconnect tubing to inspiratory filter outlet, and disconnect tubing
Edrop=x.x at 100
or
Edrop=xx.xat 40 Reconnect E tubing
Block insp filt port Remove inspiratory filter. Insert no. 3 stopper into inspiratory filter
Reconnect I tubing Remove stopper from inspiratory filter outlet. Reinstall inspiratory
Name of a section of display
Upper MDW = blocks? Press ACCEPT to confirm that all pixels are turned on in upper
This is the inspiratory filter pressure drop in cmH 100 L/min (adult tubing) or 40 L/min (pediatric tubing). Reconnect tubing to inspiratory filter outlet. Press ACCEPT.
from expiratory filter inlet. Leave expiratory filter in place. Press ACCEPT.
This is the expiratory filter pressure drop in cmH2Oataflowof100 L/min (adult tubing) or 40 L/min (pediatric tubing). Press ACCEPT. Reconnect tubing to expiratory filter outlet.
outlet. Press ACCEPT.
filter. Press ACCEPT.
Press ACCEPT to confirm that all LEDs/LCDs in named section are on.
half.
Oataflowof
2
Lower MDW = blocks? Press ACCEPT to confirm that all pixels are turned on in lower
half.
Name of a key Press named key.
4-22
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Page 95
.
Self tests (SST and EST) 4
...............................................................................
Table 4-8: Prompts during EST (continued)
Prompt Your response
High alarm sound Confirm that you hear alarm by pressing ACCEPT.
Medium alarm sound Confirm that you hear alarm by pressing ACCEPT.
No sound Confirm that you do not hear alarm by pressing ACCEPT.
Is backup alarm ON? Confirm that you hear alarm by pressing ACCEPT.
7. When the last test in the sequence is complete, the EST completion status is displayed (Table 4-9). Respond as indicated.
If certain tests do not pass, safe ventilation may be compromised, and the ventilator is rendered inoperative until it is repaired. This is called a ventilator failure. If certain other tests do not pass, but ventilation might not be compromised (subject to a trained operator’s evaluation), this is called a ventilator fault.
NOTE:
Faults detected during EST may not require the ventilator to be serviced or removed from use immediately. A trained operator, however, must evaluate the situation and determine whether the ventilator can be used. Failures detected during EST require immediate servicing and removal of the ventilator from clinical use.
Table 4-9: EST completion status
If the EST result is:
EST passed
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
It means: Do this:
All tests were performed and all passed.
To retest from start of test sequence,
press alarm reset key. To exit EST and resume ventilation,press
ACCEPT. Unblock the wye as directed, then press ACCEPT. POST now runs, and ventilation resumes at last valid settings.
4-23
Page 96
4 Self tests (SST and EST)
Table 4-9: EST completion status (continued)
If the EST result is:
EST incomplete
EST fault One or more tests did not
It means: Do this:
All tests passed, but some tests were skipped. The skipped tests were passed on a previous run.
pass. These tests might not compromise the ventilator’s ability to ventilate safely, based on the operator’s evaluation of the situation.
Warning
Do not use a ventilator that has completed EST with a fault status without verifying its operational integrity by means other than EST and determining that the patient will not be placed at risk.
To retest from the start of the test sequence, press the alarm reset key.
To exit EST and resume ventilation, press
MANUAL INSP. You are then asked whether you want to use the override feature. Press MANUAL INSP again to complete the override. Unblock the wye as directed, then press ACCEPT. POST now runs, and ventilation resumes at last valid settings.
To retest from the start of the test sequence, press the alarm reset key.
To exit EST and resume ventilation, press
MANUAL INSP. You are asked whether you want to use the override feature. Press MANUAL INSP again to complete the override. Unblock the wye as directed, then press ACCEPT. POST now runs, and ventilation resumes at last valid settings.
EST failed One or more tests failed that
might compromise the ventilators ability to ventilate safely. A VENT INOP condition is declared when youcyclepowertothe ventilator.
4-24
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Contact service.
To retest from the start of the test sequence, press the alarm reset key.
Page 97
SECTION
Once ventilation begins 5
5
..............................................................................
This section tells you:
How to change settings (a quick review)
How to view and change alarm limits (a quick review)
How to adjust apnea parameters
Howtoviewpatientdata
About the 100% O
About the EXP PAUSE and INSP PAUSE keys
5.1 Changing settings: a quick review
This section describes changing settings, switching between volume control ventilation (VCV), pressure control ventilation (PCV, available on the 760 Ventilator only) and pressure support ventilation (PSV), and changing modes.
5.1.1 Changing settings
and MANUAL INSP keys
2
You can only change a setting that is part of the current or proposed mode and breath type, or is one of the three common settings (PEEP/CPAP, TRIGGER SENSITIVITY, and % O cannot change TIDAL VOLUME. You can change several settings at once, or one at a time.
To change a setting: touch, turn, ACCEPT.
1. Touch the key: The key lights, the selected setting flashes, and the message window shows the current setting, range, units, and name of the setting. (If applicable, associated parameter information is also displayed.)
2. Turn the knob to adjust the setting.
3. Repeat steps 1 and 2 for every setting you want to change. Press CLEAR to cancel the most recent setting.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
). For example, in SPONT mode, you
2
5-1
Page 98
5 Once ventilation begins
4. Press ACCEPT to apply the new settings. The key lights turn off, new settings are displayed, and the message window displays this message:
Setting(s) accepted
5.1.2 Switching between VCV, PCV, and PSV
To switch between breath types, you must first select a mode that allows it. Follow the steps in Section 5.1.3 to change the mode.
If you try to select a breath type that is not allowed in the current mode, the ventilator displays this message:
Choose mode before choosing type
Table 5-1 summarizes which breath types are available on the 740 and 760 Ventilators, and when those breath types are selectable.
Table 5-1: Breath type availability
Ventilator Breath types Comments
740 Ventilator Vol ume co ntrol
ventilation (VCV)
Pressure support ventilation (PSV)
760 Ventilator VCV
PSV
Pressure control
ventilation (PCV)
5.1.3 Changing the mode
During ventilation, the current mode key is lit and settings are displayed. To change the mode:
1. Select the mode: Press A/C, SIMV, or SPONT. The key lights for the applicable settings flash.
2. Select the breath type: If you selected A/C or SIMV modes, select VCV or PCV (760 only). If you selected SIMV or SPONT modes, select PSV.
5-2
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
VCV not available in SPONT mode
PSV not available in A/C mode
All breath types available in SIMV mode.
VCV and PCV not available in SPONT
mode
PSV not available in A/C mode
All breath types available in SIMV mode.
Page 99
.
Once ventilation begins 5
...............................................................................
3. Select the settings: For every flashing key light, you must touch the key (and adjust the setting if necessary) before the new mode can be applied.
You can change PEEP/CPAP, TRIGGER SENSITIVITY, and % O any time, but you are not required to acknowledge or change these settings when you change the mode.
4. Apply the settings: Press ACCEPT.
NOTE:
See Appendix D for a detailed explanation of mode and breath types.
at
2
5.2 Viewing and changing alarm settings: a quick review
You can view alarm settings at any time without affecting ventilation: just touch the appropriate key.
You can change alarm settings at any time during ventilation. You can only change one alarm setting at a time. To change a setting: touch, turn,
ACCEPT.
1. Touch the key: The key lights steadily and the alarm setting flashes in its window for up to 30 seconds.
2. Turn the knob to adjust the alarm setting. (You can adjust the setting as long as the setting flashes.)
3. Press ACCEPT to apply the new alarm setting. The new alarm setting stops flashing, and the message window displays this message:
Setting(s) accepted
You cannot change alarm settings until ventilator settings have been accepted.
G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual
5-3
Page 100
5 Once ventilation begins
5.3 Adjusting apnea parameters
Warning
To ensure that the patient is ventilated appropriately in case of apnea, you must review apnea parameters, and adjust them if necessary.
NOTE:
During an APNEA alarm, you can change APNEA PARAMS but not other settings.
You can view or change apnea parameters only when apnea ventilation is possible (in SPONT mode, or in A/C or SIMV when RESPIRATORY RATE is less than 6 /minute).
1. Press APNEA PARAMS to view apnea parameter settings. When apnea ventilation is possible, the ventilator shows this message when you have accepted settings and normal ventilation has started:
Review Ta = xx s
APNEA
PARAMS
7-00086
5-4
where Ta is the apnea inteval, and xx is the current apnea interval setting in seconds.
For every flashing key light, touch the key (and adjust the setting if necessary), then apply the settings.
7-00066
On the 740 Ventilator, only VCV
apnea ventilation settings (RESPIRATORY RATE, TIDAL VOLUME, and PEAK FLOW) are available.
On the 760 Ventilator, you can select VCV apnea settings or PCV
apnea settings (RESPIRATORY RATE, INSPIRATORY PRESSURE, and I:E RATIO or T
;RISETIMEFACTORisfixedat
I
50%).
700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)
Loading...