Precision Medical PM50 User Manual

USER MANUAL
Model: PM50
SAVE THESE INSTRUCTIONS
CAUTION
300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080
Federal (USA) law restricts this device to sale by or on the order of a physician.
CONTENTS
RECEIVING/INSPECTION ................................................... 2
INTENDED USE ................................................................... 2
READ ALL INSTRUCTIONS BEFORE USING .................... 2
EXPLANATION OF ABBREVIATIONS ................................ 2
SAFETY INFORMATION - WARNINGS AND CAUTIONS ... 2
SPECIFICATIONS ................................................................ 4
COMPONENT IDENTIFICATION ......................................... 5
GROUNDING INSTRUCTIONS ........................................... 6
OPERATING INSTRUCTIONS ............................................. 6
CLEANING/MAINTENANCE ................................................ 7
RETURNS ............................................................................ 8
DISPOSAL INSTRUCTIONS ................................................8
REPLACEMENT PARTS ...................................................... 8
LIMITED WARRANTY ........................................................ 10
1
RECEIVING/INSPECTION
Remove the Precision Medical, Inc. Easy Comp Compressor from packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider.
INTENDED USE
The basic function is to supply a portable compressed air source adequate to drive the high volume nebulizers.
READ ALL INSTRUCTIONS BEFORE USING
This manual instructs a Professional to install and operate the Easy Comp Compressor. This is provided for your safety and to prevent damage to the Compressor. If you do not understand this manual, DO NOT USE the Compressor and contact your Provider.
EXPLANATION OF ABBREVIATIONS
psi Pounds Per Square Inch l/min Liters Per Minute
SAFETY INFORMATION - WARNINGS AND CAUTIONS
DANGER
WARNING
CAUTION
CAUTION
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
Used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage.
CONSULT ACCOMPANYING DOCUMENTS
Type B Equipment
2
DANGER
• Caution must be exercised when using this Easy Comp Compressor where oxygen is being administered.
• DO NOT use near any type of ame or ammable/explosive
substances.
• DO NOT allow the Compressor to exceed Operating or Storage
Temperature specications.
TO REDUCE THE RISK OF ELECTROCUTION
• Always unplug the Compressor immediately after using.
• DO NOT use while bathing.
• DO NOT place or store the Compressor where it can fall or be
pulled into a tub or sink.
• DO NOT place or drop into water or other liquid.
• DO NOT reach for the Compressor that has fallen into water.
Unplug immediately.
• DO NOT attempt to repair the Easy Comp Compressor. Any attempt could result in electrical shock.
WARNING
To reduce risk of burns, electrocution, re, or injury to persons:
• Close supervision is necessary when this Easy Comp Compressor is used by, on, or near children or invalids.
• Use this Easy Comp Compressor only for its “Intended Use” as described in this manual.
• DO NOT use attachments not recommended by Precision Medical, Inc.
• DO NOT use the Compressor with any equipment that requires
“dry” air, including ventilators, gas blenders, or ow generators,
without a device to properly dehumidify the discharged air of the PM50.
• DO NOT use if the Compressor has a damaged cord or plug.
• DO NOT use if the Compressor has been dropped or
damaged.
• DO NOT drop or insert any object into any openings.
• DO NOT cover air vents. Restriction of air ow will cause
unit to overheat, causing the thermal reset to shut down the Compressor.
3
SPECIFICATIONS
Flow Output at 50 psig: ≥13 l/min
Maximum Rated Sustained Pressure:
Duty Cycle: 50%
Dimensions:
Length: 10.25 in (26.0 cm)
Width: 5.25 in (13.3 cm)
Height: 9.5 in (24.1 cm)
Weight: 10.8 lbs (4.9 kg)
Shipping Weight: 12.5 lbs (5.7 kg)
Electrical: 120 VAC, 60 Hz, 5 Amps
Operating Temperature Range:
Storage Temperature Range:
Outlet:
Safety Systems: High Temperature Motor Shutdown
Thermal Protector: Class B, Thermal Limit Automatic Reset
This Easy Comp Compressor is not suitable for use in the presence of a ammable anesthetic mixture with air, oxygen, or nitrous oxide.
Specications are subject to change without prior notice.
Grounded 9/16 - 18 Male Threaded Fitting
60 psi
50°F to 104°F (10°C to 40°C)
-4°F to 120°F (-20°C to 49°C)
4
COMPONENT IDENTIFICATION
CAUTION
Missing or illegible labels must be replaced, contact Precision Medical, Inc.
Pressure Gauge
Air Outlet
Fitting
Regulator Knob
Power Cord Keeper
On / Off
Switch
Suction Feet
Filter End Cap
Air Filter Assembly
Power Cord
5
GROUNDING INSTRUCTIONS
The Easy Comp Compressor must be plugged into an electrical outlet that is properly grounded. In the event of an electrical short circuit, grounding reduces the risk of electrical shock. This product is equipped with a grounding circuit.
DANGER
Improper use of the grounding plug can result in a risk of electrical shock.
WARNING
Ground reliability can only be achieved when the power cord is connected to a properly grounded receptacle.
OPERATING INSTRUCTIONS
CAUTION
Inspect the Easy Comp Compressor for visual damage, DO NOT USE if damaged.
1. Become familiar with the Compressor (see Compo-
nent Identication).
2. Place Compressor near a properly grounded electrical
receptacle on a table, oor or other horizontal, smooth,
hard surface. Allow a minimum of 4 in. (10 cm) clearance on all sides to
ensure adequate air ow for cooling of product.
3. Press down on Compressor until Suction Feet are secure.
4. Ensure the On/Off Switch is in the “OFF” position.
5. Plug the Compressor into a properly grounded receptacle.
6. Attach appropriate device to the Outlet Fitting, securely, without overtightening.
7. Press the On/Off Switch to the “ON” position.
8. Adjust pressure.
To INCREASE pressure - Turn Knob CLOCKWISE
To DECREASE pressure - Turn Knob COUNTERCLOCKWISE
6
9. Begin procedure.
10. When procedure is complete, press the On/Off Switch to the “OFF” position.
CLEANING/MAINTENANCE
WARNING
Turn the Easy Comp Compressor off and disconnect
from power source before cleaning
To clean the lter (weekly):
1. Unplug Compressor.
2. Pry off lter end cap to check air lter for dust buildup.
3. If dust buildup occurs, remove air lter and wash in warm,
soapy water and rinse.
4. Absorb excess water from lter with a dry towel.
5. Return lter and end cap.
Cleaning for Exterior Components:
Use a clean cloth or sponge dampened with soap and water. Wipe dry with a clean cloth or paper towel.
.
CAUTION
Excessive dust buildup on lter will reduce performance of
Compressor. If this occurs, clean or replace with a new lter.
The air lter does not lter out saline. Treatments using nebulized saline will draw the saline mist back into the
Compressor
compressor seals and will void the warranty.
. Saline cause rapid deterioration of the
7
RETURNS
Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the Internet, www.precisionmedical.com.
Manuals available on our website; www.precisionmedical.com
DISPOSAL INSTRUCTIONS
This device and its packaging contain no hazardous materials. No special precautions need to be taken when disposing the device and/or its packaging.
Please Recycle
REPLACEMENT PARTS
Description Part Number
Filter Assembly 1834 Filter (Case -10 ea.) 1841-10
8
TROUBLESHOOTING
If the PM50 Easy Comp Compressor fails to function, consult the Troubleshooting Guide below. If the problem cannot be corrected, consult your Provider.
Problem Probable Cause Remedy Compressor will
not operate
Unable to adjust
pressure
Overheating 1. Poor Ventilation 1. Move
Low pressure 1. Loose
1. On/Off switch in “OFF” position
2. Power cord not connected properly
3. Compressor overheated
1. Defective
regulator
connection at
outlet tting
2. Improper
regulator adjustment
1. Push On/Off switch to “ON” position
2. Check power cord
3. Shut Compressor off, wait 15 minutes, Ensure vents are clear
1.
Contact Provider
Compressor to a cooler, better ventilated area
1. Tighten
connection
2. Readjust
regulator
9
LIMITED WARRANTY
AND LIMITATION OF LIABILITY
Precision Medical, Inc. warrants that the PM50 Easy Comp Compressor (the Product) will be free of defects in workmanship and/or material for the following period:
One (1) year from shipment Should any failure to conform to this warranty appear within the
applicable period, Precision Medical, Inc. shall, upon written
notication thereof and substantiation that the goods have been
stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry
practice, and that no modications, substitutions, or alterations
have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES. The representatives of Precision Medical, Inc. or any retailers are
not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus,
this writing is a nal, complete and exclusive statement of the
terms of that contract. THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF
ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages
including but not limited to lost prots, lost sales, or injury to
person or property. Correction of non-conformities as provided
above shall constitute fulllment of all liabilities of Precision
Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product
materials, designs, or specications without notice.
Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.
10
Precision Medical, Inc.
300 Held Drive
Northampton, PA 18067 USA
Toll Free Tel: 800-272-7285 Toll Free Fax: 800-353-1240 Tel:(+001) 610-262-6090 Fax:(+001) 610-262-6080
ISO 13485 Certied