U S E R M A N U A L
MODEL SERIES: PM3000
PM3100 (shown)
PM3200
PM3500
PM3600
SAVE THESE INSTRUCTIONS
|
Federal (USA) law restricts this device to sale by or |
|||
CAUTION |
||||
on the order of a physician. |
|
|||
|
|
|||
300 Held Drive |
|
Tel: (+001) |
610-262-6090 |
|
Northampton, PA 18067 USA |
Fax: (+001) |
610-262-6080 |
||
ISO 13485 Certified |
www.precisionmedical.com |
Contents |
|
RECEIVING / INSPECTION................................................... |
2 |
Intended Use..................................................................... |
2 |
READ ALL INSTRUCTIONS BEFORE USING....................... |
2 |
Explanation of Abbreviations ................................... |
2 |
SAFETY Information - Warnings and Cautions...... |
2 |
Specifications ................................................................. |
3 |
OPERATING INSTRUCTIONS............................................... |
4 |
Parts Description........................................................... |
5 |
Parts list............................................................................ |
5 |
Repair kits.......................................................................... |
6 |
DISASSEMBLY INSTRUCTIONS........................................... |
6 |
ASSEMBLY INSTRUCTIONS................................................. |
6 |
Vacuum regulator cleaning illustration............ |
7 |
Cleaning / decontamination....................................... |
8 |
Maintenance..................................................................... |
8 |
Returns.............................................................................. |
8 |
Disposal instructions.................................................. |
9 |
Troubleshooting........................................................... |
9 |
LIMITED WARRANTY .......................................................... |
10 |
DECLARATION OF CONFORMITY...................................... |
11 |
1
RECEIVING / INSPECTION
Remove the Precision Medical, Inc. Vacuum Regulator from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider.
Intended Use
The devices are intended to control and show the amount of vacuum from a central vacuum system used in various medical suctioning procedures.
READ ALL INSTRUCTIONS BEFORE USING
This manual instructs a Professional to install and operate the Vacuum Regulator. This is provided for your safety and to prevent damage to the Vacuum Regulator. If you do not understand this manual, DO NOT USE the Vacuum Regulator and contact your Provider.
Explanation of Abbreviations
l/min |
Liters Per Minute |
mmHg |
Millimeters of Mercury |
inHg |
Inches of Mercury |
kPa |
Kilopascal |
SAFETY Information - Warnings and Cautions
WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
CAUTION
0473
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
Used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage.
CONSULT ACCOMPANYING DOCUMENTS
Symbol indicates the device complies with the requirements of Directive 93/42/EEC concerning medical devices and all applicable International Standards. (On CE marked Devices ONLY)
WARNING
•DO NOT use this Vacuum Regulator for anything other than its Intended Use. Personal injury and/or damage to Regulator may result from misuse.
•Only personnel instructed and trained in its use should operate this Vacuum Regulator.
2
Specifications
Gauge Range: *Counterclockwise direction |
|
|
|
|
|||||||||
|
PM3000: |
0 - 200 mmHg - Full Vacuum |
PM3100HV: 0 - 300 mmHg - Full Vacuum |
||||||||||
*PM3000E: |
0 - 200 mmHg (0 - 26 kPa) |
*PM3200E: |
0 - 100 kPa (750 mmHg) - Full Vacuum |
||||||||||
|
PM3000HV: |
0 - 300 mmHg - Full Vacuum |
PM3500: |
0 - 150 mmHg |
|||||||||
|
PM3100: |
0 - 200 mmHg - Full Vacuum |
*PM3500E: |
0 - 200 mmHg (0 - 26 kPa) |
|||||||||
*PM3100E: 0 - 200 mmHg (0 - 26 kPa) |
PM3600: |
0 - 760 mmHg |
|||||||||||
*PM3100EHV: 0 - 300 mmHg (0 - 40kPa) |
*PM3600E: |
0 - 760 mmHg (0 - 100 kPa) |
|||||||||||
Gauge Accuracy: |
|
|
|
|
|
|
|
||||||
|
|
|
|
|
|
Analog: |
± 5% of MAX |
|
|||||
Digital/Analog, Dual Gauge: |
|
|
|
|
|
|
|||||||
|
|
|
|
|
|
Digital Display: |
± 1% of Full Scale |
|
|||||
|
|
|
|
|
|
Analog Gauge: |
± 5% of MAX within ref. Indicator |
||||||
Vacuum Ports: |
|
1/8 NPT Female |
|
||||||||||
Modes: |
|
|
|
|
|
|
|
|
|||||
PM3100, PM3200E3 & PM3600 Series: LINE |
- Provides maximum, continuous |
||||||||||||
|
|
|
|
|
|
|
|
|
|
|
vacuum from the vacuum source |
||
|
|
|
|
|
|
|
|
|
OFF |
- |
No Vacuum |
||
|
|
|
|
|
|
|
|
|
REG. |
- |
(Regulated) provides an adjustable, |
||
|
|
|
|
|
|
|
|
|
|
|
continuous vacuum level |
||
PM3000, PM3200E & PM3500 Series: |
OFF |
- |
No Vacuum |
||||||||||
|
|
|
|
|
|
|
|
|
REG. |
- |
(Regulated) provides an adjustable, |
||
|
|
|
|
|
|
|
|
|
|
|
continuous vacuum level |
||
|
|
|
|
Flow: Models |
|
|
|
|
Mode |
Max Flow |
|||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
*PM3000, PM3100, PM3200 & PM3500: |
|
REG |
|
51 l/min |
||||
|
|
|
|
|
|
|
|
|
|||||
|
LINE |
|
55 l/min |
||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
||
|
|
|
|
|
**PM3600 SERIES: |
|
|
|
REG |
|
71 l/min |
||
|
|
|
|
|
|
|
|
LINE |
|
82 l/min |
|||
|
|
|
|
|
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
|
|||||
|
|
|
* MAXIMUM FLOW is obtained with a vacuum source of 21” Hg (71.1 kPa) |
|
|||||||||
|
|
** MAXIMUM FLOW is obtained with a vacuum source of 25” Hg (84.6 kPa) |
|
||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Maximum Vacuum: |
|
|
|
|
|
|
|
||||
|
|
|
|
|
|
|
|||||||
|
|
|
|
|
PM3000 & PM3100: |
REG. Mode @ Full Vac - 396 mmHg (53 kPa) |
|||||||
|
|
|
|
|
|
|
|||||||
|
|
|
|
|
|
PM3200: REG. Mode @ Full Vac - 100 kPa (750 mmHg) |
|||||||
|
|
|
|
|
PM3500D: |
Restricted to 170 mmHg (23 kPa) |
|
± 10 mmHg (1.3 kPa) |
|||||
|
|
|
|
|
PM3500E: |
Restricted to 160 mmHg (21 kPa) |
|
||||||
|
|
|
|
|
|
|
|
||||||
|
|
|
|
|
|
PM3600: REG. Mode @ Full Vac - 760 mmHg (101 kPa) |
|||||||
|
|
|
|
|
|
|
|
|
|
||||
Operating Environmental Limits: |
|
|
|
0°F to 122°F (-18°C to 50°C) |
|||||||||
Recommended Environmental Operating Limits: 55°F to 85°F (13°C to 29°C) |
|||||||||||||
Storage Environmental Limits: |
|
|
|
|
|
|
|||||||
|
|
|
|
|
|
|
Temperature Range: |
|
-4°F to 140°F (-20°C to 60°C) |
||||
|
|
|
|
|
|
|
|
Humidity: |
|
Max 95% Noncondensing |
|||
Battery: |
3 Volt Lithium, AA (Digital Vacuum Gauge Models ONLY) |
Specifications are subject to change without prior notice.