Remove the Precision Medical, Inc. Easy Air Compressor (PM15-F, PM15EF,
PM15-P or PM15EP) from the packaging and inspect for damage. If there
is any damage, DO NOT USE and contact your Provider.
INTENDED USE
The basic function is to supply a portable compressed air source adequate
to drive the high volume nebulizers.
READ ALL INSTRUCTIONS BEFORE USING
This manual instructs a Professional to install and operate the Easy Air
Compressor (PM15-F, PM15EF, PM15-P or PM15EP). This is provided for
your safety and to prevent damage to the Easy Air Compressor. If you do
not understand this manual, DO NOT USE the Easy Air Compressor and
contact your Provider.
EXPLANATION OF ABBREVIATIONS
psiPounds Per Square Inch
l/minLiters Per Minute
SAFETY INFORMATION - WARNINGS AND CAUTIONS
DANGER
WARNING
CAUTION
CAUTION
Indicates an imminently hazardous situation which,
if not avoided, will result in death or serious injury.
Indicates a potentially hazardous situation which, if
not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if
not avoided, may result in minor or moderate injury.
Used without the safety alert symbol indicates a
potentially hazardous situation which, if not avoided,
may result in property damage.
CONSULT ACCOMPANYING DOCUMENTS
Type B Equipment
Protective Earth Ground
Power On
Power Off
2
DANGER
• Caution must be exercised when using this Compressor where
oxygen is being administered.
• DO NOT use near any type of ame or ammable/explosive
substances.
• DO NOT allow the Compressor to exceed Operating or Storage
Temperature specications.
TO REDUCE THE RISK OF ELECTROCUTION
• Always unplug the Compressor immediately after using.
• DO NOT use while bathing.
• DO NOT place or store the Compressor where it can fall or be
pulled into a tub or sink.
• DO NOT place or drop into water or other liquid.
• DO NOT reach for the Compressor that has fallen into water.
Unplug immediately.
• DO NOT attempt to repair the Easy Air Compressor. Any attempt
could result in electrical shock.
WARNING
To reduce risk of burns, electrocution, re, or injury to persons:
• Close supervision is necessary when this Compressor is used by,
on, or near children or invalids.
• Use this Easy Air Compressor only for its "Intended Use" as
described in this manual.
• DO NOT
• DO NOT use the Compressor with any equipment that requires "dry"
air, including ventilators, gas blenders, or ow generators, without a
device to properly dehumidify the discharged air of the PM15.
• DO NOT use if the Compressor has a damaged cord or plug.
• DO NOT use if the Compressor has been dropped or damaged.
• DO NOT drop or insert any object into any openings.
• DO NOT cover air vents. Restriction of air ow will cause unit to
overheat, causing the thermal reset to shut down the Compressor.
use attachments not recommended by Precision Medical, Inc.
(Models Equipped with a Pressure Gauge)
Maximum Rated Sustained Pressure: 60 psi
PM15 Output:
50 psi...................Minimum 13 l/min
PM15E/PM15EFGB Output:
50 psi...................Minimum 9 l/min
Operating Temperature Range: 50°F to 104°F (10°C to 40°C)
Storage Temperature Range: -4°F to 104°F (-20°C to 40°C)
(Models Equipped with a Flow Gauge)
PM15-F & PM15EF Output:
(± 1 l/min)
l/min - as set and read on the gauge
Storage/Operating Humidity: Maximum 95% Noncondensing
Safety Systems: High Temperature Motor Shutdown
Thermal Protector: Class B, Thermal Limit Automatic Reset
Equipment Classication:
Class I - with respect to protection from electrical shock
Type B - degree of protection against electrical shock
IPX0 - degree of protection against ingress of liquids
Mode of Operation - continuous
The Compressor meets the EMC requirements of EN 55011: Group 1 Class B.
The Compressor meets the requirements of CGA V-5: section 5.4.
This Compressor is not suitable for use in the presence of a ammable
anesthetic mixture with air, with oxygen or with nitrous oxide.
Specications are subject to change without prior notice.
4
COMPONENT IDENTIFICATION
CAUTION
Missing or illegible labels must be replaced, contact Precision Medical, Inc.
Carry Handle
On/Off (I/O)
Switch
Air
Outlet
Regulator
Knob
Gauge
Air Inlet
Filters
Power
Cord
Fuse Drawer
AC Inlet
Power Cord
Keeper
Hour Meter
5
GROUNDING INSTRUCTIONS
The Easy Air Compressor must be plugged into an electrical outlet that is
properly grounded. In the event of an electrical short circuit, grounding reduces
the risk of electrical shock. This product is equipped with a grounding circuit.
DANGER
Improper use of the grounding plug can result in a risk of electrical
shock.
WARNING
Ground reliability can only be achieved when the power cord is
connected to an equivalent receptacle marked “Hospital Grade”.
OPERATING INSTRUCTIONS
CAUTION
Inspect the Easy Air Compressor for visible damage, DO NOT USE
if damaged.
1.
Become familiar with the Compressor (see Component Identication).
2. Position the Compressor near a properly grounded electrical
receptacle. Allow a minimum of 4 in (10 cm) clearance on all sides to
ensure adequate air ow for cooling of the Compressor.
3. Ensure the On/Off (I/O) Switch is in the “OFF” (O) position.
4. Ensure the Power Cord is properly connected.
5. Attach appropriate device to the Outlet Fitting securely, without
overtightening.
6. Pull Regulator Knob “OUT” to unlock regulator, then turn Knob
fully counterclockwise. (Failure to do so may cause the device
not to start.)
7. Press the On/Off (I/O) Switch to the “ON” (I) position.
8. Adjust pressure/ow.
To INCREASE pressure / ow - Turn Knob CLOCKWISE
To DECREASE pressure / ow - Turn Knob COUNTERCLOCKWISE
11. When the Compressor is not in use, press the On/Off (I/O) Switch to
the “OFF” (O) position.
6
CLEANING / MAINTENANCE
WARNING
• Turn the Easy Air Compressor off and disconnect from any
external power source before cleaning.
• DO NOT attempt to repair. There are no serviceable parts inside.
To clean the lter (weekly):
1. Unplug Compressor.
2. Check air inlet lters for dust buildup.
3. If dust buildup occurs, remove air inlet lters and wash in warm, soapy
water and rinse.
4. Absorb excess water from lter with a dry towel.
5. Return the air inlet lters.
Cleaning for Exterior Components:
Use a clean cloth or sponge dampened with soap and water. Wipe dry
with a clean cloth or paper towel.
CAUTION
• Excessive dust buildup on lter will reduce performance of
Compressor. If this occurs, clean or replace with a new lter.
• The air lter does not lter out saline. Treatments using nebulized
saline will draw the saline mist back into the Compressor. Saline
causes rapid deterioration of the compressor seals and will void the
warranty.
At ~8500 Hours (or as needed):
Return the Easy Air Compressor to your Provider for Compressor
Maintenance (including rebuilding pump).
Contact Precision Medical, Inc. for Preventative Maintenance Services.
7
RETURNS
Returned products require a Return Goods Authorization (RGA) number,
contact Precision Medical, Inc. All returns must be packaged in sealed
containers to prevent damage. Precision Medical, Inc. will not be responsible
for goods damaged in transit. Refer to Precision Medical Inc. Return Policy
available on the Internet, www.precisionmedical.com.
Manuals available on our website; www.precisionmedical.com
NOTE: When returning this Easy Air Compressor, disconnect the power cord from the back of the device.
DISPOSAL INSTRUCTIONS
This device and its packaging contain no hazardous materials. No special
precautions need to be taken when disposing the device and/or its packaging.
Please Recycle
REPLACEMENT PARTS
DESCRIPTIONPART #
Large Filter503237
Small Filter502391
8
TROUBLESHOOTING
If the Easy Air Compressor fails to function, consult the Troubleshooting
Guide below. If problem cannot be solved, consult your Provider.
5. Turn off Compressor.
When Power restored,
repeat Start Up
Procedure.
6. BEFORE STARTING
COMPRESSOR.
Pull “OUT” regulator
Knob to unlock,
then turn Knob fully
counterclockwise.
1. Pull Regulator Knob
out
1. Allow a minimum of
4 in (10 cm) clearance
on all sides to ensure
adequate air ow
for cooling of the
Compressor.
2. Clean Air Filters
1. Tighten connection
2. Re-adjust regulator
9
LIMITED WARRANTY
AND
LIMITATION OF LIABILITY
Precision Medical, Inc. warrants that EASY AIR 15 Compressor
(the Product) will be free of defects in workmanship and/or material
for the following period:
One (1) year from shipment
Should any failure to conform to this warranty appear within the
applicable period, Precision Medical, Inc. shall, upon written
notication thereof and substantiation that the goods have been
stored, installed, maintained and operated in accordance with
Precision Medical, Inc.’s instructions and standard industry practice,
and that no modications, substitutions, or alterations have been
made to the goods, correct such defect by suitable repair or
replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
The representatives of Precision Medical, Inc. or any retailers are
not authorized to make oral warranties about the merchandise
described in this contract, and any such statements shall not be
relied upon and are not part of the contract for sale. Thus, this
writing is a nal, complete and exclusive statement of the terms
of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY
WARRANTY OF MERCHANTABILITY, FITNESS FOR A
PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY,
WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable
for special, incidental or consequential damages including but not
limited to lost prots, lost sales, or injury to person or property.
Correction of non-conformities as provided above shall constitute
fulllment of all liabilities of Precision Medical, Inc. whether based
on contract, negligence, strict tort or otherwise. Precision Medical,
Inc. reserves the right to discontinue manufacture of any product or
change product materials, designs, or specications without notice.
Precision Medical, Inc. reserves the right to correct clerical or
typographical errors without penalty.
10
Precision Medical, Inc.
300 Held Drive
Northampton, PA 18067 USA
Toll Free Tel: 800-272-7285
Toll Free Fax: 800-353-1240
Tel:(+001) 610-262-6090
Fax:(+001) 610-262-6080
ISO 13485 Certied
Tell us how we are doing!
Visit us at www.precisionmedical.com
502597 Rev17 07/13 (?M) Printed in USA
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.