PolyScience 6260T11A110B, 6360T11A120C, 6560T11A120C, 6760T21A130D, 6160T21A130D Data sheet

...
Chillers and Coolers | Chillers | Portable
Portable Chillers — 6000 Series – ¼ to 1 HP
Key Specifications
Working Temperature: -10° to +40°C,
-10° to +70°C with heater option
Temperature Stability: ±0.1°C
Cooling Capacity: Up to 2900 W @ 20°C
Reservoir Capacity: 1.1 gallons/4.2 liters
Overall Dimensions 27.6 x 14.5 x 22.6" (L x W x H): 70.2 x 36.8 x 57.5 cm
Features:
• Stable and reliable cooling for many common heat removal applications, including laser etching, AA furnaces, ICP, rotary evaporators, vacuum systems, reaction vessels, plasma etching, and condenser cooling
• Large, dual displays present temperature and pressure or flow rate simultaneously
• Compact, portable design takes up less floor space
• Cooling at ambient temperatures as high as 35°C
• Choice of pumps and compressor sizes
• User-adjustable temperature, pressure, and flow rate alarms
• Heater option extends cooling temperature to +50°C, provides the ability to apply heat up to 70°C; ideal for applications, such as lasers, that must be brought to a temperature above ambient before operation can begin
• External temperature tracking and communications capability (optional)
• Cool Command™, WhisperCool™ (3/4 and 1 HP models)
C 155859 US
Large digital readouts provide information on temperature as well as the process flow rate or pressure at a glance.
See pages 124 and 125 for considerations when choosing a chiller.
102
| Phone: 800-229-7569 • +1-847-647-0611
6000 Series Chillers with Turbine Pump
6200
1/4 HP
Temperature Range -10° to +40°C (-10° to +70°C with heater option)
Temperature Stability ±0.1°C
1
Cooling Capacity (W)
Maximum Pressure psi (bar)
Maximum Flow gpm (l/min) 3.5 (13.2) 2.9 (11) 3.5 (13.2) 2.9 (11) 3.5 (13.2) 2.9 (11) 3.5 (13.2) 2.9 (11) 3.5 (13.2) 2.9 (11)
Power Requirements (V/Hz) 120/60 240/50 120/60 240/50 120/60 240/50 208-230/60 240/50 208-230/60 240/50
Part Number 120 VAC/60 Hz 6260T11A110B 6360T11A120C 6560T11A120C 6760T21A130D 6160T21A130D
Part Number 240 VAC/50 Hz 6250T21A130E 6350T21A130E 6550T21A130E 6750T21A130E 6150T21A130E
@
0°C
+10°C
+20°C
400 600 850
100 (6.9) 83 (5.7) 100 (6.9) 83 (5.7) 100 (6.9) 83 (5.7) 100 (6.9) 83 (5.7) 100 (6.9) 83 (5.7)
300 500 700
530 990
1400
6300
1/3 HP
485 935
1280
750 1150
1700
6500
1/2 HP
765 1140 1775
875
1550
2300
6700
3/4 HP
975
1500
2200
1000
1925
2900
6100
1 HP
1200 1900
2650
6000 Series Chillers with Positive Displacement Pump
6200
1/4 HP
Temperature Range -10° to +40°C (-10° to +70°C with heater option)
Temperature Stability ±0.1°C
Cooling Capacity (W)
Maximum Pressure psi (bar)
Maximum Flow gpm (l/min)
Power Requirements (V/Hz)
Part Number 120 VAC/60 Hz 6260P11A110B 6360P11A1120C 6560P11A120C 6760P41A130D 6160P41A130D
Part Number 240 VAC/50 Hz 6250P21A130E 6350P21A130E 6550P21A130E 6750P41A130E 6150P41A130E
1
@ 0°C
+10°C
+20°C
400 600 850
100 (6.9) 83 (5.7) 100 (6.9) 83 (5.7) 100 (6.9) 83 (5.7) 100 (6.9) 83 (5.7) 100 (6.9) 83 (5.7)
1.0 (3.8) 0.95 (3.6) 1.0 (3.8) 0.95 (3.6) 1.0 (3.8) 0.95 (3.6) 3.5 (13.2) 2.9 (11) 3.5 (13.2) 2.9 (11)
120/60 240/50 120/60 240/50 120/60 240/50 208-230/60 240/50 208-230/60 240/50
300 500 700
530 990
1400
6300
1/3 HP
485 935
1280
750 1150
1700
6500
1/2 HP
765 1140 1775
875
1550
2300
6700
3/4 HP
975
1500
2200
1000
1925
2900
6100
1 HP
1200 1900
2650
6000 Series Chillers with Magnetic Drive Centrifugal Pump
6200
1/4 HP
Temperature Range -10° to +40°C (-10° to +70°C with heater option)
Temperature Stability ±0.1°C
Cooling Capacity (W)
Maximum Pressure psi (bar)
Maximum Flow gpm (l/min)
Power Requirements (V/Hz)
Part Number 120 VAC/60 Hz 6260M11A110B 6360M11A110B 6560M11A120C 6760M21A130D 6160M21A130D
Part Number 240 VAC/50 Hz 6250M21A130E 6350M21A130E 6550M21A130E 6750M21A130E 6150M21A130E
1. Cooling Capacity based on 20°C (68°F) ambient temperature and a 50%/50% mix of ethylene glycol and distilled water as coolant See pages 109 through 112 for 6000 Series Chiller options and accessories. Electrical plugs for the part numbers listed are standard U.S. and European types. See page 128 for additional plug types and part numbers.
1
@ 0°C
+10°C
+20°C
300 700 950
10 (.69) 9.5 (0.66) 10 (.69) 9.5 (.66) 10 (.69) 9.5 (.66) 10 (.69) 9.5 (.66) 10 (.69) 9.5 (.66)
4.1 (15.5) 3.9 (14.7) 4.1 (15.5) 3.9 (14.7) 4.1 (15.5) 3.9 (14.7) 4.1 (15.5) 3.9 (14.7) 4.1 (15.5) 3.9 (14.7)
120/60 240/50 120/60 240/50 120/60 240/50 208-230/60 240/50 208-230/60 240/50
400 600 800
700
1000
1430
6300
1/3 HP
600 960 1180
850 1250 1800
6500
1/2 HP
765 1140 1775
975 1550 2350
6700
3/4 HP
1075 1600 2250
www.polyscience.com |www.polyscience.com |
6100
1 HP
1100
1835
2900
1400
2050
2750
103