Physio-Control LIFEPAK 15 User Manual

LIFEPAK® 15 MONITOR/ DEFIBRILLATOR
OPERATING INSTRUCTIONS
®
LIFEPAK
OPERATING INSTRUCTIONS
15
Important Information
!USA
Rx Only
!USA
LIFEPAK, LIFENET, QUIK-LOOK, and QUIK-COMBO are registered trademarks of Physio-Control, Inc. CODE SUMMARY, Shock Advisory System, CODE-STAT, DT EXPRESS, LIFEPAK CR, REDI-PAK, LIFEPAK EXPRESS, SunVue, and cprMAX are trademarks of Phys io-Contro Inc. Bluetooth is a registered trademark of Bluetooth SIG, Inc. CADEX is a registered trademark of Cadex Electronics Inc. Microstream, CapnoLine, and FilterLine are registered trademarks of Oridion Systems Ltd. The Oridion medical capnography in this product is covered b one or more of the following US patents: 6,428,483; 6,997,880; 5,300,859; 6,437,316 and their foreign equivalents. Additional paten applications pending. PC Card is a trademark of the Personal Computer Memory Card International Association. Masimo, the Radical logo Rainbow, and SET are registered trademarks of Masimo Corporation. EDGE System Technology is a trademark of Ludlow Technical Products. Formula 409 is a registered trademark of The Clorox Company. Specifications are subject to change without notice.
©2007-2013 Physio-Control, Inc. All rights reserved. Publication Date: 04/2013
MIN 3306222-005
Device Tracking
The U.S. Food and Drug Administration requires defibrillator manufacturers and distributors to track the location of their defibrillators. If the device is located somewhere other than the shipping address or the device has been sold, donated, lost, stolen, exported, destroyed, permanently retired from use, or if the device was not obtained directly from Physio-Control, please do one of the following: register the device at http://www.physio-control.com, call the device registration phone line at 1.800.426.4448, or use one of the postage-paid address change cards located in the back of this manual to update this vital tracking information.
Text Conventions
Throughout these operating instructions, special text characters (for example, CAPITAL LETTERS such as CHECK
PATIENT and SPEED DIAL) are used to indicate labels, screen messages, and voice prompts.
Version History
These operating instructions describe LIFEPAK 15 monitor/defibrillator devices with software revision 3306808-001 or later.
CONTENTS
1Preface
Introduction ...................................................................................................................1-3
Intended Use..................................................................................................................1-4
Modes of Operation .........................................................................................................1-5
2 Safety Information
Terms ............................................................................................................................2-3
General Dangers and Warnings .........................................................................................2-3
3 Basic Orientation
Front View ......................................................................................................................3-3
Back View ....................................................................................................................3-13
Batteries ......................................................................................................................3-15
Home Screen................................................................................................................3-17
Alarms .........................................................................................................................3-21
Options ........................................................................................................................3-23
Events..........................................................................................................................3-25
4 Monitoring
Monitoring the ECG .........................................................................................................4-3
Acquiring a 12-Lead ECG ..............................................................................................4-15
Monitoring SpO2, SpCO, and SpMet ...............................................................................4-25
Monitoring Noninvasive Blood Pressure ...........................................................................4-37
Monitoring ETCO2.........................................................................................................4-45
Monitoring Invasive Pressure ..........................................................................................4-54
Monitoring Continuous Temperature................................................................................4-61
Vital Sign and ST Segment Trends ..................................................................................4-66
5 Therapy
General Therapy Warnings and Cautions ............................................................................5-3
Therapy Electrode and Standard Paddle Placement ............................................................5-4
Automated External Defibrillation (AED)............................................................................5-7
Manual Defibrillation.....................................................................................................5-22
Synchronized Cardioversion Procedure ............................................................................5-27
Noninvasive Pacing .......................................................................................................5-32
Pediatric ECG Monitoring and Manual Mode Therapy Procedures.......................................5-39
6 Paddle Accessory Options
QUIK-COMBO Therapy Electrodes ....................................................................................6-3
Standard Paddles............................................................................................................6-6
Sterilizable Internal Defibrillation Paddles .......................................................................6-10
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
iii
7 Data Management
Patient Records and Reports ............................................................................................7-3
Memory Capacity ............................................................................................................7-9
Managing Current Patient Records ..................................................................................7-10
Managing Archived Patient Records ................................................................................7-11
8 Data Transmission
About Transmitting Patient Records and Reports................................................................8-3
Preparing the Monitor for Transmission .............................................................................8-4
Using Bluetooth Wireless Communication..........................................................................8-5
Using a Direct Connection..............................................................................................8-10
Transmitting Reports .....................................................................................................8-12
Considerations When Transmitting Data ..........................................................................8-14
Troubleshooting Tips ....................................................................................................8-15
9 Power Adapter
Basic Orientation ............................................................................................................9-3
Using the Power Adapter..................................................................................................9-5
General Maintenance.......................................................................................................9-8
Warranty ......................................................................................................................9-10
10 Maintaining the Equipment
General Maintenance and Testing ...................................................................................10-3
Battery Maintenance ...................................................................................................10-12
Cleaning the Device.....................................................................................................10-15
Storing the Device.......................................................................................................10-16
Loading Paper ............................................................................................................10-17
General Troubleshooting Tips .......................................................................................10-18
Service and Repair .....................................................................................................10-22
Product Recycling Information .....................................................................................10-23
Warranty ....................................................................................................................10-23
Accessories ................................................................................................................10-24
Appendix A: Specifications and Performance Characteristics
Appendix B: Screen Messages
Appendix C: Shock Advisory System
Appendix D: Electromagnetic Compatibility Guidance
Appendix E: Symbols
iv
LIFEPAK 15 Monitor/Defibrillator Operating Instructions

PREFACE

This chapter provides a brief introduction to the LIFEPAK® 15 monitor/defibrillator and describes the product’s intended use.
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-3
Intended Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1
Modes of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
1-1
PREFACE

Introduction

The LIFEPAK 15 monitor/defibrillator is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols.
These operating instructions include information and procedures related to all features of the LIFEPAK 15 monitor/defibrillator. Your LIFEPAK 15 monitor/defibrillator may not have all of these features.
These operating instructions describe the operation of the LIFEPAK 15 monitor/defibrillator when the factory default settings are used. The factory default settings for all setup options are identified in Table A-5 on page A-16. Your device may be set up with different default settings, based on your protocols. For information about changing default settings, see the LIFEPAK 15 Monitor/ Defibrillator Setup Options provided with your device.
IMPORTANT! Some LIFEPAK 15 monitor/defibrillator accessories are not interchangeable with
accessories that are used with other LIFEPAK monitor/defibrillators. Specific accessory incompatibilities are noted in the related sections.
1
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
1-3

Intended Use

Intended Use
The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in out-of­doors and indoor emergency care settings within the environmental conditions specified on page A-12. The LIFEPAK 15 monitor/defibrillator is designed to be used during ground transportation except when specified otherwise.
Monitoring and therapy functions may only be used on one patient at a time. Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older.
For additional intended use information, and information about the indications and contraindications of the monitoring and therapy functions, see the individual sections identified below.
• ECG Monitoring See page 4-3 Standard feature
• 12-Lead Electrocardiography See page 4-15 Optional
•SpO
• Noninvasive Blood Pressure Monitoring See page 4-37 Optional
• End-Tidal CO
• Invasive Pressure Monitoring See page 4-54 Optional
• Temperature Monitoring See page 4-61 Optional
• Vital Sign and ST Segment Trends See page 4-66 Optional
• Automated External Defibrillation See page 5-7 Standard feature
• Manual Defibrillation See page 5-22 Standard feature
• Noninvasive Pacing See page 5-32 Standard feature
SpCO, and SpMet Monitoring See page 4-25 Optional
2,
Monitoring See page 4-45 Optional
2
1-4
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
PREFACE

Modes of Operation

The LIFEPAK 15 monitor/defibrillator has the following modes of operation:
AED mode—for automated ECG analysis and a prompted treatment protocol for patients in cardiac arrest.
Manual mode—for performing manual defibrillation, synchronized cardioversion, noninvasive pacing, and ECG and vital sign monitoring.
Archive mode—for accessing stored patient information.
Setup mode—for changing default settings of the operating functions. For more information, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.
Demo mode—for simulated waveforms and trend graphs for demonstration purposes.
Service mode—for authorized personnel to perform diagnostic tests and calibrations. For more information, see the LIFEPAK 15 Monitor/Defibrillator Service Manual.
1
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
1-5

SAFETY INFORMATION

This chapter provides important information to help you operate the LIFEPAK 15 monitor/ defibrillator. Familiarize yourself with all of these terms and warnings.
Terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-3
General Dangers and Warnings . . . . . . . . . . . . . . . . . . . 2-3
2
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
2-1
SAFETY INFORMATION

Terms

The following terms are used either in these operating instructions or on the LIFEPAK 15 monitor/ defibrillator:
Danger: Immediate hazards that will result in serious personal injury or death.
2
Warning: Caution: Hazards or unsafe practices that may result in minor personal injury, product damage, or
property damage.
Hazards or unsafe practices that may result in serious personal injury or death.

General Dangers and Warnings

The following are general danger and warning statements. Other specific warnings and cautions are provided as needed in other sections of these operating instructions.
DANGER!
EXPLOSION HAZARD
Do not use this defibrillator in the presence of flammable gases or anesthetics.
WARNINGS
SHOCK OR FIRE HAZARDS
SHOCK HAZARD
The defibrillator delivers up to 360 joules of electrical energy. Unless properly used as described in these operating instructions, this electrical energy may cause serious injury or death. Do not attempt to operate this device unless thoroughly familiar with these operating instructions and the function of all controls, indicators, connectors, and accessories.
SHOCK HAZARD
Do not disassemble the defibrillator. It contains no operator serviceable components and dangerous high voltages may be present. Contact authorized service personnel for repair.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
2-3
General Dangers and Warnings
WARNINGS (CONTINUED)
SHOCK OR FIRE HAZARD
Do not immerse any portion of this defibrillator in water or other fluids. Avoid spilling any fluids on defibrillator or accessories. Spilled liquids may cause the defibrillator and accessories to perform inaccurately or fail. Do not clean with ketones or other flammable agents. Do not autoclave or sterilize this defibrillator or accessories unless otherwise specified.
POSSIBLE FIRE
Use care when operating this device close to oxygen sources (such as bag-valve-mask devices or ventilator tubing). Turn off gas source or move source away from patient during defibrillation.
ELECTRICAL INTERFERENCE HAZARDS
POSSIBLE ELECTRICAL INTERFERENCE WITH DEVICE PERFORMANCE
Equipment operating in close proximity may emit strong electromagnetic or radio frequency interference (RFI), which could affect the performance of this device. If use of equipment in close proximity is necessary, observe the device to verify normal operation in the configuration in which the device will be used. RFI may result in distorted ECG, incorrect ECG lead status, failure to detect a shockable rhythm, cessation of pacing, or incorrect vital sign measurements. Avoid operating the device near cauterizers, diathermy equipment, or other portable and mobile RF communications equipment. Do not rapidly key EMS radios on and off. Refer to Appendix D for recommended distances of equipment. Contact Physio-Control Technical Support if assistance is required.
2-4
POSSIBLE ELECTRICAL INTERFERENCE
Using cables, electrodes, or accessories not specified for use with this defibrillator may result in increased emissions or immunity from electromagnetic or radio frequency interference (RFI) which could affect the performance of this defibrillator or of equipment in close proximity. Use only parts and accessories specified in these operating instructions.
POSSIBLE ELECTRICAL INTERFERENCE
This defibrillator may cause electromagnetic interference (EMI) especially during charge and energy transfers. EMI may affect the performance of equipment operating in close proximity. Verify the effects of defibrillator discharge on other equipment prior to using the defibrillator in an emergency situation, if possible.
IMPROPER DEVICE PERFORMANCE HAZARDS
POSSIBLE IMPROPER DEVICE PERFORMANCE
Using other manufacturers’ cables, electrodes, power adapters, or batteries may cause the device to perform improperly and may invalidate the safety agency certifications. Use only the accessories that are specified in these operating instructions.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
SAFETY INFORMATION
WARNINGS (CONTINUED)
POSSIBLE IMPROPER DEVICE PERFORMANCE
Changing factory default settings will change the behavior of the device. Changes to the default settings must only be made by authorized personnel.
POSSIBLE DEVICE SHUTDOWN
Always have immediate access to a spare, fully charged, properly maintained battery. Replace the battery when the device displays a low battery warning.
SAFETY RISK AND POSSIBLE EQUIPMENT DAMAGE
POSSIBLE INJURY OR SKIN BURNS
Monitors, defibrillators, and their accessories (including electrodes and cables) contain ferromagnetic materials. As with all ferromagnetic equipment, these products must not be used in the presence of the high magnetic field created by a Magnetic Resonance Imaging (MRI) device. The high magnetic field created by an MRI device will attract the equipment with a force sufficient to cause death or serious personal injury to persons between the equipment and the MRI device. This magnetic attraction may also damage and affect the performance of the equipment. Skin burns will also occur due to heating of electrically conductive materials such as patient leads and pulse oximeter sensors. Consult the MRI manufacturer for more information.
2
POSSIBLE SKIN BURNS
A defect in the neutral electrode connection on HF surgical equipment could cause burns at the lead or sensor site and damage to the monitor/defibrillator. Do not apply patient leads or sensors when using high frequency (HF) surgical (electrocautery) equipment.
Note: The features of the LIFEPAK 15 monitor/defibrillator which could come in either direct or
casual contact with the patient or caregiver during normal use are not manufactured with materials that contain latex as an intentionally added component or expected impurity.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
2-5

BASIC ORIENTATION

This chapter provides a basic orientation to the LIFEPAK 15 monitor/defibrillator device and its controls, indicators, and connectors.
Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-3
Back View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3
Home Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
Events. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
3-1
BASIC ORIENTATION
CO2
SpO2
NIBP
ECG
P1
DANGER Explosion hazard. Do not use in the presence of flammable gases. WARNING Hazardous electric al output. For use only by qualified personnel.
Recommended
Adult VF Dose: 200-300-360J
P2
LIFEPAK 15 MONITOR/DEFIBRILLATOR
Self Test in progress...
©Physio-Control, Inc. 2008 All Rights Reserved 3207410-001
MONITOR/DEFBRILLATOR

Front View

Figure 3-1 shows the front of the LIFEPAK 15 monitor/defibrillator. The front of the device is described in the following sections.
3
Figure 3-1 Front View
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
3-3
Front View
ON
ENERGY
SELECT
CHARGE
CPR
SIZE
ANALYZE
LEAD
SYNC
Recommended
Adult VF Dose: 200-300-360J

Area 1

Figure 3-2 Area 1 Controls
3-4
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
BASIC ORIENTATION
Table 3-1 Area 1 Controls
CONTROL DESCRIPTION FOR MORE INFORMATION
3
VF dose label
1ON
ENERGY
2
SELECT
3 CHARGE Charges the defibrillator in Manual mode. See page 5-22
Physio-Control recommended energy dose for adult Ventricular Fibrillation (VF).
Turns device ON or OFF. LED illuminated when ON. Press and hold to turn device off.
Increases or decreases energy level in Manual mode.
Shock button. Initiates discharge of defibrillator energy to patient. LED flashes when charging is complete.
Auxiliary power indicator. LED illuminated when defibrillator is connected to auxiliary AC or DC power source, whether defibrillator is turned on or off.
Battery charging indicator. LED illuminated when installed batteries are fully charged. LED flashes when either battery is charging. LED is not illuminated when no batteries are installed or a battery is unable to be charged.
Illuminated Service LED indicates a condition exists that prevents or could prevent normal defibrillator operation.
See Biphasic Clinical Summaries at www.physio-control.com
See page 5-22
See page 5-22
See page 9-5
See page 9-5
See page 10-21
CPR
ANALYZE
LEAD Changes ECG lead. See page 4-4
SIZE Changes ECG size. See page 4-5
SYNC
©2007-2013 Physio-Control, Inc.
Controls CPR metronome. LED illuminated when metronome function is active.
Activates Shock Advisory System™ (AED mode). LED illuminated when AED is analyzing the ECG, and flashes when user is prompted to push
ANALYZE.
Activates Synchronized mode. LED illuminated when Sync mode is active and flashes with detection of each QRS.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
See page 5-25
See page 5-7
See page 5-27
3-5
Front View
PACER
RATE
CURRENT
PAUSE

Area 2

Figure 3-3 Area 2 Controls
Table 3-2 Area 2 Controls
CONTROL DESCRIPTION FOR MORE INFORMATION
PACER Activates pacer function. LED illuminated when function
is activated and flashes with each current pulse.
RATE Increases or decreases pacing rate. See page 5-32
CURRENT Increases or decreases pacing current. See page 5-32
PAUSE Temporarily slows pacing rate. See page 5-32
See page 5-32
3-6
LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Area 3

NIBP
ALARMS
OPTIONS
EVENT
HOME
SCREEN
Speed dial
BASIC ORIENTATION
3
Figure 3-4 Area 3 Controls
Table 3-3 Area 3 Controls
CONTROL DESCRIPTION FOR MORE INFORMATION
NIBP Initiates blood pressure measurement. LED
illuminated when BP measurement is being obtained.
ALARMS Activates and silences alarms. LED illuminated
when alarms are enabled and flashes when an alarm condition occurs.
OPTIONS Accesses optional functions. See page 3-23
EVENT Accesses user-defined events. See page 3-25
HOME SCREEN Returns to Home Screen display. See page 3-17
SPEED DIAL Scrolls through and selects screen or menu items. See page 3-19
Display mode button switches between color display and high contrast SunVue™ display.
See page 4-37
See page 3-21
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
3-7
Front View
12-LEAD
TRANSMIT
CODE
SUMMARY
PRINT

Area 4

Figure 3-5 Area 4 Controls
Table 3-4 Area 4 Controls
CONTROL DESCRIPTION FOR MORE INFORMATION
12-LEAD Initiates acquisition of 12-lead ECG. See page 4-15
TRANSMIT Initiates transmission of patient data. See page 8-12
CODE SUMMARY Prints CODE SUMMARY™ critical event record. See page 7-4
PRINT Starts and stops printer. See page 7-10
3-8
LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Area 5

Speaker
Printer
Therapy cable receptacle
CO
2
SpO2/ SpCO/
SpMet
NIBP
ECG
P1
P2
CO2
SpO2
NIBP
ECG
P1
P2
See warnings page 2-4
BASIC ORIENTATION
3
Figure 3-6 Area 5 Connectors, Speaker, and Printer
Table 3-5 Area 5 Connectors, Speaker, and Printer
LABEL DESCRIPTION FOR MORE INFORMATION
CO2 FilterLine® set port See page 4-45
SpO2/SpCO/SpMet Sensor cable port See page 4-25
NIBP Pneumatic tubing port See page 4-37
ECG Green electrically isolated ECG cable port See page 4-3
P1 Invasive pressure cable port See page 4-54
P2 Invasive pressure cable port See page 4-54
Speaker Projects device tones and voice prompts
Printer Door for 100 mm printer paper See page 10-17
Therapy cable receptacle
QUIK-COMBO paddles cable receptacle
®
therapy cable and standard (hard)
See page 3-12
Note: If your LIFEPAK 15 monitor/defibrillator is configured for temperature monitoring, P1 and P2
are replaced by a single port labeled TEMP. For more information about temperature monitoring, see page 4-61.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
3-9
Front View
CO2
SpO2
NIBP
ECG
P1
P2
CONNECTOR ACTION
CO2
Connect: Open CO
2
port door, insert FilterLine connector, and turn
clockwise until connector is firmly seated.
Disconnect: Rotate FilterLine connector counterclockwise and pull connector out.
SpO2/ SpCO/ SpMet
Connect: Align cable connector with SpO
2
port and push in until
connector clicks into place.
Disconnect: Press the gray buttons on each side of the cable connector simultaneously and pull connector out.
NIBP
Connect: Insert NIBP tubing connector into the NIBP port.
Disconnect: Press the latch on the left side of the port and pull
tubing connector out.
ECG
Connect: Align the green ECG connector with the ECG port; position the white line on the cable facing left. Insert the cable connector into the port until the connector is firmly seated.
Disconnect: Pull the ECG connector straight out.
P1/P2
Connect: Align the IP (invasive pressure) cable connector with the P1 or P2 port; position the gap on the connector facing up. Insert the cable connector into the port until the connector is firmly seated.
Disconnect: Grip the connector and pull straight out.
Connectors
Figure 3-7 Connectors for IP Monitoring Configuration
Note: If your LIFEPAK 15 monitor/defibrillator is configured for temperature monitoring, P1 and P2
are replaced by a single port labeled TEMP. For more information, see Figure 3-8 on page 3-11.
3-10
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
BASIC ORIENTATION
CONNECTOR ACTION
CO2
Connect: Open CO
2
port door, insert FilterLine connector, and
turn clockwise until connector is firmly seated.
Disconnect: Rotate FilterLine connector counterclockwise and pull connector out.
SpO2/ SpCO/
SpMet
Connect: Align cable connector with SpO
2
port and push in
until connector clicks into place.
Disconnect: Press the gray buttons on each side of the cable connector simultaneously and pull connector out.
NIBP
Connect: Insert NIBP tubing connector into the NIBP port.
Disconnect: Press the latch on the left side of the port and pull
tubing connector out.
ECG
Connect: Align the green ECG connector with the ECG port; position the white line on the cable facing left. Insert the cable connector into the port until the connector is firmly seated.
Disconnect: Pull the ECG connector straight out.
TEMP
Connect: Align the temperature adapter cable connector with the TEMP port. Insert the cable connector into the port until the connector is firmly seated.
Disconnect: Grip the connector and pull straight out.
CO2
SpO2
NIBP
ECG
TEMP
3
Figure 3-8 Connectors for Temperature Monitoring Configuration
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
3-11
Front View
1. Align the therapy cable connector with the receptacle.
2. Slide the therapy cable until you feel the connector lock in place. You will also hear a
“click.”
1. Press the release button on the therapy cable connector.
2. Slide the therapy cable connector out.
Connecting and Disconnecting the Therapy Cable
WARNING
POSSIBLE EQUIPMENT DAMAGE AND INABILITY TO DELIVER THERAPY
To help protect the therapy cable connector from damage or contamination, keep therapy cable connected to the defibrillator at all times. Inspect and test the therapy cable daily according to the Operator’s Checklist in the back of this manual. Physio­Control recommends replacement of therapy cables every three years to reduce the possibility of failure during patient use.
IMPORTANT!
(hard) paddles have the same type of connector and connect to the defibrillator at the same location. These therapy cables are not compatible with other LIFEPAK defibrillator/monitors.
To connect a therapy cable to the defibrillator:
Figure 3-9 Connect Therapy Cable
To disconnect the therapy cable from the defibrillator:
The LIFEPAK 15 monitor/defibrillator QUIK-COMBO therapy cable and standard
3-12
Figure 3-10 Disconnect Therapy Cable
LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Back View

Standard
paddle
wells
USB port
cover
2
1
System
connector
Battery
well 2
CO
2
exhaust
port
See
warnings
page 10-12
and
page 10-13
See
warnings
page 4-45
See warning
page 3-14
Battery
pins
Battery
well 1
Battery
contacts
Paddle
retainers
Paddle test
contacts
Auxiliary
power
connector
See warning
page 9-4 and
page 10-12
BASIC ORIENTATION
3
Figure 3-11 Back View
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
3-13
Back View
Table 3-6 Back View
LABEL DESCRIPTION FOR MORE INFORMATION
Battery wells, pins, and contacts
exhaust port Connects to a scavenger system when
CO
2
Standard paddle wells, retainers, and test contacts
USB port cover Protects USB port from the environment. For future use
System connector Connects device to a gateway or external
Auxiliary power connector
Each well holds one Lithium-ion battery. Two pins in each well transfer the battery power. Battery contacts transfer battery status information.
monitoring EtCO
Paddle wells stow standard (hard) paddles. Retainers provide secure retention and quick removal of the paddles. Test contacts allow complete paddles defibrillation checks according to the Operator’s Checklist.
computer for transfer of patient reports. Also provides real-time ECG output.
Connects to an optional AC or DC power adapter. Allows use of auxiliary power source.
during use of anesthetics.
2
See page 10-12
See page 4-45
See page 6-6 and Operator’s Checklist
the back of this manual
See page 7-3
See page 9-3
WARNING
SHOCK HAZARD
If you are monitoring a patient and using the system connector, all equipment connected to the system connector must be battery powered or electrically isolated from AC power according to EN 60601-1. If in doubt, disconnect the patient from the defibrillator before using the system connector. Only use Physio-Control recommended data transmission cables. For more information, contact Physio-Control Technical Support.
in
Note: To prevent inadvertent depletion of the defibrillator batteries, disconnect external devices
from the system connector when not in use.
3-14
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
BASIC ORIENTATION

Batteries

The LIFEPAK 15 monitor/defibrillator operates either on battery power using two Lithium-ion batteries, or with auxiliary power using the AC Power Adapter or DC Power Adapter. Batteries may be charged in the Station or Mobile Li-ion Battery Charger, the REDI-CHARGE™ Battery Charger, or in the monitor/defibrillator if it is connected to auxiliary power.
Note: Although the monitor/defibrillator can operate using auxiliary power with no batteries
installed, at least one battery should be installed at all times. If the monitor/defibrillator loses power for more than 30 seconds, the device reverts to the user-configured default settings and begins a new patient record.
IMPORTANT! The LIFEPAK 15 monitor/defibrillator Lithium-ion batteries are not interchangeable
with batteries that are used in other LIFEPAK defibrillators.
Routinely inspect batteries for damage or leakage. Recycle or discard damaged or leaking batteries.
Each battery has a fuel gauge that indicates the approximate charge level in the battery. Press the gray button above the battery symbol to check the battery’s charge level prior to installing it in the defibrillator. The four battery indicators shown here represent approximate charge—greater than 70%, greater than 50%, greater than 25%, and 25% or less, respectively.
3
Figure 3-12 Battery Charge Indicators
Battery warning indicators are shown below. A single flashing LED indicates that the battery is very low and needs to be charged. Any two or more flashing LEDs indicate that the battery is faulty and should be returned to your authorized service personnel.
Figure 3-13 Battery Warning Indicators
Note: Older or heavily used batteries lose charge capacity. If a battery fuel gauge indicates fewer
than four LEDs immediately after completing a charge cycle, the battery has reduced capacity. If the battery fuel gauge shows two or fewer LEDs after the battery completes a charge cycle, the battery should be replaced.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
3-15
Batteries
To install a battery:
1. Confirm that the battery is fully charged, unless the battery will be charged in the monitor/ defibrillator using the power adapter.
2. Inspect battery pins and contacts in the battery wells for signs of damage.
3. Align battery so battery clip is over the pins in the battery well.
4. Insert the end of the battery that is opposite the battery clip into the battery well.
5. Firmly press the clip end of the battery into the battery well until it clicks into place.
6. Repeat Step 1 through Step 5 to insert second battery.
To remove a battery, press the battery clip in and tilt the battery out of the battery well.
WARNING
POSSIBLE LOSS OF POWER DURING PATIENT CARE
Battery pins in the defibrillator may be damaged if batteries are dropped or forced into battery wells. Inspect pins routinely for signs of damage. Keep batteries installed at all times except when the device is removed from service for storage.
For information about battery maintenance, see "Battery Maintenance" on page 10-12.
3-16
LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Home Screen

Heart rate
SpO
2
/SpCO/
SpMet
EtCO2
IP1
IP2
NIBP
Channel 1
Channel 2
Channel 3
Message area
Heart symbol
Alarm limits
Bluetooth icon
Battery indicator
Selected energy
ECG Lead/Size
HR
SPO2 %
CO2 mmHg
ART mmHg
CVP mmHg
NIBP mmHg
J
Alarm indicator
Time
150 50
RR
BASIC ORIENTATION
3
Figure 3-14 Home Screen
The Home Screen is the main screen that displays ECG and other information. When a monitoring cable is attached to the device, the corresponding monitoring area on the screen is activated and the current patient values for that function are displayed. For example, when you connect an SpO cable, the SpO is connected. When the cable is disconnected, the SpO
area is activated on the screen. SpO2 values for the patient appear after the patient
2
patient values are replaced by dashes (--).
2
Separate controls do not activate the monitoring functions, except for NIBP.
Each vital sign monitoring area is colored to match its waveform. This color scheme aids in associating the displayed waveform with its vital sign value. When a function does not have a waveform displayed, the vital sign area is gray.
WARNING
FAILURE TO DETECT A CHANGE IN ECG RHYTHM
Heart rate meters may count internal pacing pulses during cardiac arrest or some arrhythmias. Do not rely entirely on heart rate meter alarms. Keep pacemaker patients under close surveillance.
2
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
3-17
Home Screen
Table 3-7 Home Screen
AREA DESCRIPTION FOR MORE INFORMATION
Alarm limits Limits display along the right side of the
parameter.
Heart symbol Flashes with detected QRS signals.
Alarm indicator Indicates whether alarms are on or silenced.
Absence of indicator means alarms are off.
Heart rate Device accurately detects and displays heart rates
between 20 and 300 beats per minute (bpm). If patient’s heart rate is below 20 bpm or above 300 bpm, or pacing is active, dashes (– – –) appear. If ECG is not active, the SpO monitor can display pulse rate, indicated by
(SPO2) or PR (NIBP).
or NIBP
2
PR
SpO2/SpCO/SpMet Oxygen saturation level displays as a percentage
from 50 to 100. Saturation below 50% displays as <50%. A fluctuating bar graph represents the pulse signal strength. When available and selected, the SpCO or SpMet value is displayed as a percent for 10 seconds, and then the SpO area reverts to the SpO
EtCO2 End-tidal CO
reading.
2
level displays in mmHg, Vol%, or
2
2
kPa. Respiratory rate (RR) displays in breaths per minute.
See page 3-21
See page 3-21
See page 4-25
See page 4-45
3-18
IP1/IP2 Displays systolic, diastolic, and mean invasive
See page 4-54 pressures in mmHg. Two channels are available; default labels are P1 and P2.
User-selectable labels include the following:
• ART (arterial pressure)
• PA (pulmonary artery pressure)
• CVP (central venous pressure)
• ICP (intracranial pressure)
• LAP (left atrial pressure)
Temp Displays
skin, esophageal, rectal, or bladder
See page 4-61
temperature.
NIBP Displays systolic, diastolic, and mean arterial
See page 4-37 pressures (MAP) in mmHg, and time to next BP, when interval is set.
Time Real or elapsed. See LIFEPAK 15
Monitor/Defibrillator
Setup Options provided
with your device.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
BASIC ORIENTATION
Waveform
Channel 3
CO2 SpO2 Trend
None
1. Rotate the SPEED DIAL to the desired setting.
2. Press the
SPEED DIAL to select the setting.
Table 3-7 Home Screen (Continued)
AREA DESCRIPTION FOR MORE INFORMATION
3
Bluetooth icon Indicates Bluetooth capability. The LED is
illuminated when a Bluetooth connection is established. Select this icon to access the Bluetooth setup menu.
Battery indicator Indicates presence of battery in battery well 1
and 2, relative level of charge, and battery in use.
Selected energy Selected defibrillation energy.
ECG Lead/Size Lead and size for ECG. See page 4-4
Channel 1 Displays the primary ECG waveform and is always
visible.
Channel 2 Displays an additional waveform, a continuation
of the Channel 1 ECG (cascading ECG), or a trend graph.
Channel 3 Displays an additional waveform or a trend graph. See page 4-70
Message area Displays up to two lines of status messages. See Appendix B
See page 8-3
See page 3-20
See page 4-4
See page 4-31

Navigating the Home Screen

Use the SPEED DIAL to navigate around the Home Screen. As you rotate the SPEED DIAL, the individual vital sign areas and waveform channels on the Home Screen are outlined. If you outline a vital sign area or channel and then press the
SPEED DIAL, a menu appears.
For example, rotate the following menu appears.
Whenever a menu is displayed, the ECG is always visible in Channel 1. To return to the Home Screen from any menu, press the
Rotate and press the
SPEED DIAL to outline Channel 3, and then press the SPEED DIAL. The
HOME SCREEN button.
SPEED DIAL to select an option in a menu.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
3-19
Home Screen

Battery Status Indicators

The Home Screen displays battery indicators that show the following information about the batteries installed in the defibrillator:
• Presence or absence of battery in battery well
• Battery in use
• Battery charge state
IMPORTANT! Always check the battery charge level and ensure batteries are adequately charged
before use.
When two batteries are installed, the defibrillator uses the battery with the lowest level of charge first. The battery in use is indicated by a white battery number in a black box. When a battery reaches the replace battery state, the defibrillator automatically switches to the other battery. When all battery capacity is exhausted, the defibrillator turns off. If you insert a charged battery and repower the device in less than 30 seconds, the defibrillator retains its settings. Table 3-8 provides a description of the various battery status indicators.
Table 3-8 Battery Status Indicators
INDICATOR MEANING DESCRIPTION
The defibrillator is using the battery in well 1 for power. Battery status
Active battery
Low battery
Very low battery
Unrecognized battery
No battery installed or fault detected
indicators display up to four green bars. Each green bar represents approximately 25% remaining charge. For example, three green bars indicate about 75% remaining charge.
Battery in well 1 is in use and is low. One yellow bar indicates 5% to 10% remaining charge.
Battery in well 1 is in use and is very low. One red flashing bar indicates 0 to 5% remaining charge. The defibrillator automatically switches to the other battery only if adequate charge is available. If both batteries show red bars, the occurs.
Battery in well 2 is not in use. Battery communication failed or a non­Physio-Control battery is installed. The battery may power the defibrillator but the level of charge is unknown and low battery messages and prompts will not occur.
No battery is installed in battery well 1, or a fault was detected in the battery in well 1 and the device will not use the battery.
REPLACE BATTERY voice prompt
Note: When the defibrillator is operating on auxiliary power using a power adapter, the battery
indicators show the battery charge level, but the well numbers are not highlighted. The and
REPLACE BATTERY messages and prompts do not occur when operating on auxiliary power.
3-20
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
LOW BATTERY
BASIC ORIENTATION
Select QUICK SET to activate the alarms for all active monitoring functions.
The Quick Set
limits automatically set high and
low limits based on the patient’s current vital sign values. For example, if the patient’s HR is 70, selecting
WIDE results in a high limit of
110 and a low limit of 45; selecting NARROW results in a high limit of 100 and a low limit of
50. The default is
WIDE.

Alarms

Quick Set
Limits Silence
VF/VT Alarm
Wide 2 Min
Off
Select LIMITS to change alarm limits to WIDE or
NARROW. See Table A-3 on page A-14.
Select SILENCE to turn off the audible alarm for up to 15 minutes. If an alarm limit is exceeded while the alarm is silenced, the violated vital sign flashes and an alarm message appears, but the alarm tone remains silent.
Alarms
Quick Set
Limits
Silence VF/VT Alarm
Wide
Narrow
Note: Older or heavily used batteries lose charge capacity. If a fully charged battery is installed in
the defibrillator and the battery status indicator shows less than four bars, the battery has reduced capacity. If a battery status indicator shows only one or two bars after a fully charged battery is installed, the battery has less than half the normal use time and should be recycled.
Alarms
LIFEPAK 15 monitor/defibrillator alarms can be set up to be ON or OFF when the defibrillator is turned on. For more information, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.
When alarms are set up to be ON, default limits are set. The limits temporarily appear to the right of the active vital signs. For all vital sign default alarm limits, see Table A-3 on page A-14.
3
If alarms are set up to be OFF, press ON or are enabled by pressing device. If power is lost for less than 30 seconds, for example due to a system reset or changing the only active battery, alarm settings are restored automatically.
ALARMS to enable the alarms. Whether alarms are set up to be
ALARMS, they can only be turned off by pressing ON to turn off the

Setting Alarms

When you press ALARMS, the following menu appears:
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
3-21
Alarms
Alarms
Quick Set
Limits
Silence VF/VT Alarm
Wide
2 Min Off
Select VF/VT ALARM to turn on continuous monitoring for ventricular fibrillation and ventricular tachycardia in Manual mode.
The VF/VT alarm indicator appears above the primary ECG when the alarm is ON.
When the alarm is silenced or suspended, a red X appears across the indicator .
Reselect
VF/VT to turn off this alarm.
Note: The heart rate display and corresponding heart rate alarm should not be relied upon to
provide an indication of ventricular fibrillation. Turn on the VF/VT alarm.
Note: When the VF/VT ALARM is ON, you are limited to PADDLES lead or Lead II in Channel 1. See
"Selecting ECG Lead" on page 4-4.
Note: The VF/VT alarm is suspended when the metronome is active, the noninvasive pacemaker is
on, or when standard paddles are attached and suspended when the monitor/defibrillator is charging or is charged.
PADDLES lead is selected. The alarm is also
3-22

Managing Alarms

The alarm bell symbol indicates when alarms are ON or OFF . All alarms that are controlled by tone sounds and the violated vital sign flashes.
To manage an alarm:
1. Press
2. Assess the cause of the alarm.
3. Assess the appropriateness of the limits settings (
If the patient is unstable, consider silencing the alarm for up to 15 minutes while attending to the patient. Do NOT reselect
WARNING
POSSIBLE FAILURE TO DETECT AN OUT OF RANGE CONDITION
4. After the patient is stable, reselect
QUICK SET have equal priority. When alarms are ON and an alarm limit is exceeded, a
ALARMS. This silences the alarm for 2 minutes.
WIDE or NARROW).
QUICK SET.
Reselecting QUICK SET resets the alarm limits around the patient’s current vital sign values, which may be outside the safe range for the patient.
QUICK SET, if necessary.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
BASIC ORIENTATION

Options

Patient...
Pacing...
Date/Time... Alarm Volume...
Archives...
Print...
User Test
When alarms are ON, you can silence them preemptively for up to 15 minutes.
To silence alarms preemptively:
3
1. Press
2. Select
3. Select
The message
ALARMS.
SILENCE.
SILENCE duration of 2, 5, 10, or 15 minutes.
ALARMS SILENCED appears in the message area at the bottom of the Home Screen.
Note: When you select SILENCE, the VF/VT alarm is not silenced.
Options
Press OPTIONS to display the Options menu. Rotate the SPEED DIAL to scroll through the choices. Press the
SPEED DIAL to make a selection.
Table 3-9 Options Menu Selections
SELECTION DESCRIPTION FOR MORE INFORMATION
Patient
Pacing
Date/Time
Alarm Volume
Archives Access archived patient records. See page 7-11
Enter patient name, patient ID, incident, age, and sex.
Select demand or nondemand pacing. Set internal pacer detection ON or OFF.
Set date and time. Cycle power for change to take effect.
Adjust volume for alarms, tones, voice prompts and CPR metronome.
See "Entering Patient
Data" in next section
See page 5-32
See LIFEPAK 15
Monitor/Defibrillator
Setup Options for time
display options.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
3-23
Options
Options
Patient...
Pacing...
Date/Time... Alarm Volume...
Archives...
Print...
User Test
1. Press OPTIONS.
2. Use the
SPEED DIAL to select PATIENT.
3. Select LAST NAME, FIRST NAME, PATIENT ID,
INCIDENT, AGE, or SEX. (LAST NAME is
selected in the example.)
Options / Patient
Last Name...
First Name... Patient ID...
Incident...
Age Sex
4. Rotate the SPEED DIAL to scroll through the characters and commands. Press the
SPEED
DIAL
to make a selection. The selected
character appears.
5. Repeat Step 4 until the name is complete.
6. Select
END.
Three additional commands are available:
SPACE—inserts blank space. BACKSPACE—deletes last character and
moves selection back one space.
CLEAR—clears all characters.
Options / Patient / Last Name
BCDEFGHIJKLMNOPQRSTUVWXYZ
Last Name: _ _ _ _ _ _ _ _ _ _ _ _ _ _
End
Space Backspace
Space
Clear
End
0 1 2 3 4 5 6 7 8 9 -
A
Table 3-9 Options Menu Selections (Continued)
SELECTION DESCRIPTION FOR MORE INFORMATION
Print
User Test Initiate device self-test. See page 10-5
Select report, format, mode, and speed for printing a current patient report.
See page 7-10

Entering Patient Data

To enter patient data:
3-24
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
BASIC ORIENTATION
1. Press EVENT to display the Events menu.
2. Rotate the
SPEED DIAL to scroll through the
choices. Press the
SPEED DIAL to make a
selection.
3. Select
MORE to display additional event
selections.

Events

Generic
Oxygen
IV Access
Nitroglycerin Morphine
Cancel Last
Intubation
CPR
Epinephrine
Atropine Lidocaine
More...
When an event is selected, the event and time stamp appear in the message area on the Home Screen.
Generic 12:20:30
Events
Use the Events menu to annotate patient events. A selected event appears in the Event log of the CODE SUMMARY critical event record. Events can be customized in Setup mode. For more information, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.
To select an event:
3
Notes:
• If you highlight an event but do not select it and the menu times out, a Generic event and time stamp are annotated in the event log.
• If you highlight an event but do not select it and then press time stamp are annotated in the event log.
• Select
©2007-2013 Physio-Control, Inc.
CANCEL LAST to indicate that an incorrect event was selected. A Cancel Last event and
time stamp print in the event log.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
HOME SCREEN, a Generic event and
3-25

MONITORING

This chapter describes the monitoring features of the LIFEPAK 15 monitor/defibrillator.
Monitoring the ECG . . . . . . . . . . . . . . . . . . . . . . . page 4-3
Acquiring a 12-Lead ECG . . . . . . . . . . . . . . . . . . . . . . 4-15
Monitoring SpO2, SpCO, and SpMet . . . . . . . . . . . . . . 4-25
4
Monitoring Noninvasive Blood Pressure . . . . . . . . . . . . 4-37
Monitoring ETCO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-45
Monitoring Invasive Pressure . . . . . . . . . . . . . . . . . . . . 4-54
Monitoring Continuous Temperature. . . . . . . . . . . . . . . 4-61
Vital Sign and ST Segment Trends . . . . . . . . . . . . . . . . 4-66
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-1
MONITORING

Monitoring the ECG

Intended Use

The electrocardiogram (ECG) is a recording of the electrical activity of the heart. ECG monitoring allows for identification and interpretation of cardiac rhythms or dysrhythmias and calculation of heart rate. The ECG is obtained by placing either electrodes or paddles on the patient and allows the heart’s electrical activity to be monitored and recorded.
ECG monitoring is a tool to be used in addition to patient assessment. Care should be taken to assess the patient at all times; do not rely solely on the ECG monitor.

ECG Monitoring Warning

WARNING
POSSIBLE MISINTERPRETATION OF ECG DATA
The frequency response of the monitor screen is intended only for basic ECG rhythm identification; it does not provide the resolution required for diagnostic and ST segment interpretation. For diagnostic or ST segment interpretation, or to enhance internal pacemaker pulse visibility, attach the multi-lead ECG cable. Then print the ECG rhythm in diagnostic frequency response (DIAG) or obtain a 12-lead ECG.
4
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-3
Monitoring the ECG
1. Press LEAD. If any ECG lead currently appears on the Home Screen, the lead changes to
PADDLES. If PADDLES lead is currently
displayed, the lead changes to Lead
II.
2. While the
LEAD menu is displayed, press LEAD
again or rotate the
SPEED DIAL to the desired
lead.
Note: If lead sets are predefined for Channels 2 and 3, the lead sets show on the menu. The ECG cable that is connected to the device, such as 3­lead or 5-wire, determines the leads you can select. For information about defining lead sets, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.
Lead
III
AVR
AVL
AVF II, CO2 I, II, III
AVR, AVL, AVF
Paddles
I
II
1. For the primary ECG, outline and select
CHANNEL 1 and then select LEAD.
2. Rotate the
SPEED DIAL to the desired ECG
lead.
3. Press the
SPEED DIAL to select the ECG lead.
4. Repeat this procedure to select or change displayed ECG waveforms for Channels 2 and 3.
Channel 1
Lead
Size 1.0

Selecting ECG Lead

The LIFEPAK 15 monitor/defibrillator includes two methods for selecting or changing the ECG lead.
To select or change the displayed ECG lead using the
:
LEAD button:
To select or change the displayed ECG lead using the SPEED DIAL:
Note: The ECG shows dashed lines until the electrodes are connected to the patient.
Note: When the VF/VT ALARM is ON, you are limited to PADDLES lead or Lead II in Channel 1. See
"Setting Alarms" on page 3-21.
4-4
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
1. Press SIZE.
2. While the
SIZE menu is displayed, press SIZE
again or rotate the
SPEED DIAL to the desired
size.
Size
1. For the primary ECG, outline and select
CHANNEL 1 and then select SIZE.
2. Rotate the
SPEED DIAL to the desired ECG
size.
3. Press the
SPEED DIAL to select the ECG size.
Channel 1
Lead
Size
II
1.0

Changing ECG Size

The LIFEPAK 15 monitor/defibrillator includes two methods for selecting or changing ECG size.
4
To select or change the displayed ECG size using the
To select or change the displayed ECG size using the
SIZE button:
SPEED DIAL:
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-5
Monitoring the ECG
1. Press the SPEED DIAL to select QRS VOLUME.
2. Rotate the
SPEED DIAL to the desired volume.
3. Press the
SPEED DIAL to set the volume.
HR
QRS Volume
Anterior
Lateral
QUIK-COMBO Therapy Electrodes
Standard Paddles
Sternum
Apex

Adjusting the Systole Volume

To adjust the systole beep volume, use the SPEED DIAL to outline and select the HR area on the Home Screen.
The following menu appears:
Note: The volume is reset to OFF each time the device is turned off.

Monitoring Using Paddle Accessories

To monitor ECG using paddles, you can use either QUIK-COMBO therapy electrodes or standard (hard) paddles. For more information about paddle accessories, see Chapter 6, "Paddle Accessory Options."
Anterior-Lateral Placement
Anterior-lateral placement is the only placement that should be used for ECG monitoring using paddle accessories.
To place the therapy electrodes or paddles:
1. Place either the therapy electrode or midaxillary line, with the center of the electrode in the midaxillary line, if possible (see Figure 4-1).
APEX paddle lateral to the patient’s left nipple in the
Figure 4-1 Anterior-Lateral Placement
4-6
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
2. Place the other therapy electrode or STERNUM paddle on the patient’s upper right torso, lateral to the sternum and below the clavicle as shown in Figure 4-1.
Special Situations for Electrode or Paddle Placement
When placing therapy electrodes or standard paddles, be aware of the special requirements in the following possible situations:
Obese Patients or Patients with Large Breasts
Apply therapy electrodes or standard paddles to a flat area on the chest, if possible. If skin folds or breast tissue prevent good adhesion, it may be necessary to spread skin folds apart to create a flat surface.
Thin Patients
Follow the contour of the ribs and spaces when pressing the therapy electrodes or standard paddles onto the torso. This limits air spaces or gaps under the electrodes and promotes good skin contact.
Patients with Implanted Devices Such as Pacemakers or Defibrillators
4
If possible, place therapy electrodes or standard paddles away from implanted device.
Paddles ECG Monitoring Procedure
To monitor using standard paddles or therapy electrodes:
1. Press
2. Prepare the patient’s skin:
3. Apply the standard paddles or therapy electrodes in the anterior-lateral position. For therapy
4. Connect the therapy electrodes to the therapy cable.
ON.
• Remove all clothing from the patient’s chest.
• Remove excessive chest hair as much as possible. Avoid nicking or cutting the skin if using a shaver or razor. If possible, avoid placing electrodes over broken skin.
• Clean and dry the skin, if necessary. Remove any medication patches and ointment on the patient’s chest.
• Briskly wipe the skin dry with a towel or gauze. This mildly abrades the skin and removes oils, dirt, and other debris for better electrode adhesion to the skin.
• Do not use alcohol, tincture of benzoin, or antiperspirant to prep the skin.
electrodes, confirm that the package is sealed and the Use By date is not passed. For standard paddles, apply conductive gel over the entire electrode surface.
5. Select
©2007-2013 Physio-Control, Inc.
PADDLES lead.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-7
Monitoring the ECG
12-Lead Cable
3-Lead Cable
5-Wire Cable
Precordial lead
attachment
Main cable
(4-wire)
Limb lead
attachment
Precordial lead
attachment
Main cable
12-Lead Cable

Monitoring Using ECG Cable Accessories

The following ECG cables, shown in Figure 4-2, are available for ECG monitoring with the LIFEPAK 15 monitor/defibrillator:
• 12-lead (either of 2 types)
•3-lead
•4-wire
• 5-wire
4-8
Figure 4-2 12-Lead, 3-Lead, 4-Wire, and 5-Wire ECG Cables
ECG Monitoring Procedure
To perform ECG monitoring:
1. Press
ON.
2. Attach the ECG cable to the green connector on the monitor.
3. Identify the appropriate electrode sites on the patient as shown in Figure 4-3.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
AHA Labels IEC Labels
RA Right Arm R Right LA Left Arm L Left *RL Right Leg N Negative LL Left Leg F Foot
*Note: Not used for 3-lead cable.
LA/LRA/R
LL/FRL/N
Figure 4-3 Limb Lead Electrode Placement
4. Prepare the patient’s skin for electrode application:
• Shave excessive hair at electrode site.
• For oily skin, clean skin with alcohol pad.
• Gently scrape skin to remove surface layer of dead cells and improve conduction of electrical signals.
4
• Avoid locating electrodes over tendons and major muscle masses.
• Clean and dry the skin.
5. Apply ECG electrodes:
• Confirm that the package is sealed and the Use By date is not passed.
• Attach an electrode to each of the lead wires.
• Grasp electrode tab and peel electrode from carrier.
• Inspect electrode gel and make sure gel is intact (discard electrode if gel is not intact).
• Hold electrode taut with both hands. Apply the electrode flat to the skin. Smooth tape outwardly. Avoid pressing the center of the electrode.
• Secure the trunk cable clasp to the patient’s clothing.
Note: Electrode quality is critical for obtaining an undistorted ECG signal. Always check the
date code on electrode packages for expiration date before using on a patient. Do not use electrodes that have expired. Disposable electrodes are intended for a single use.
6. Select the desired ECG lead on the monitor screen.
7. If necessary, adjust ECG size for accurate heart rate counting.
8. Press
PRINT to obtain an ECG printout.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-9
Monitoring the ECG
Precordial Lead ECG Monitoring
The precordial (chest) leads (see Table 4-1, "ECG Leads Color Codes") can be used for monitoring when using the 12-lead cable or 5-wire cable.
To perform precordial lead ECG monitoring:
1. Insert the precordial lead attachment into the main cable as shown in Figure 4-2 on page 4-8.
2. Place the precordial lead electrodes on the chest as described in the 12-lead ECG procedure and shown in Figure 4-5 on page 4-16.
Note: When using a 5-wire cable, attach the limb leads as described in "ECG Monitoring
Procedure" on page 4-8, and place the C-lead electrode on the chest in the precordial position desired. Note that the LIFEPAK 15 monitor labels the ECG for this lead as V1 on the screen and printout, regardless of the location of the C-lead electrode.
Leads Off
If an electrode or lead wire disconnects during ECG monitoring, the monitor emits an audible alarm and displays a messages continue until the electrode or lead wire is replaced.
LEADS OFF message. The ECG trace becomes a dashed line. The alarm and
Color Coding for ECG Leads
The lead wires and the electrode snaps for the patient ECG cable are color coded according to American Heart Association (AHA) or International Electrotechnical Commission (IEC) standards as listed in Table 4-1.
Table 4-1 ECG Leads Color Codes
LEADS AHA LABEL AHA COLOR IEC LABEL IEC COLOR
Limb Leads RA
LA
RL
LL
C
Precordial Leads V1
V2 V3 V4 V5 V6
White Black Green Red Brown
Red Yellow Green Blue Orange Violet
R L N F C
C1 C2 C3 C4 C5 C6
Red Yellow Black Green Brown
Red Yellow Green Brown Black Violet
4-10
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
Monitoring Patients Who Have Internal Pacemakers
The LIFEPAK 15 monitor/defibrillator internal pacemaker detection feature can be used to help identify internal pacemaker pulses on the printed ECG. When enabled, this feature uses lead V4 to detect internal pacemaker pulses. If V4 is not available because it is not attached or is too noisy, Lead II or Paddles Lead is used.
When the internal pacemaker detection feature is ON, the LIFEPAK 15 monitor/defibrillator annotates a hollow arrow on the printed ECG if internal pacemaker pulses are detected. Patient history and other ECG waveform data, such as wide QRS complexes, should be used to verify the presence of an internal pacemaker. False annotations of this arrow may occur if ECG artifacts mimic internal pacemaker pulses. If false annotations occur frequently, deactivate the detection feature using the
The LIFEPAK 15 monitor/defibrillator typically does not use internal pacemaker pulses to calculate the heart rate. However, when using therapy electrodes or standard paddles to monitor
PADDLES lead, the monitor may detect internal pacemaker pulses as QRS complexes, resulting
in in an inaccurate heart rate.
Large amplitude pacemaker pulses may overload the QRS complex detector circuitry so that no paced QRS complexes are counted. To help minimize ECG pickup of large unipolar pacemaker pulses, place ECG electrodes so the line between the positive and negative electrodes is perpendicular to the line between the pacemaker generator and the heart.
OPTIONS / PACING / INTERNAL PACER menu (see "Options" on page 3-23).
4
Smaller amplitude internal pacemaker pulses may not be distinguished clearly in For improved detection and visibility of internal pacemaker pulses, turn on the internal pacemaker detector function using the ECG cable, select an ECG lead, and print the ECG in diagnostic frequency response. For information about configuring internal pacemaker detection, see the Pacing Setup menu in the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.
OPTIONS / PACING / INTERNAL PACER menu or connect the
PADDLES lead.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-11
Monitoring the ECG

Troubleshooting Tips

If problems occur while monitoring the ECG, check Table 4-2 for aid in troubleshooting. For basic troubleshooting problems, such as no power, see "General Troubleshooting Tips" on page 10-18.
Table 4-2 Troubleshooting Tips for ECG Monitoring
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
Any of these messages displayed:
CONNECT ELECTRODES CONNECT ECG LEADS ECG LEADS OFF XX LEADS OFF
Therapy electrodes not connected
One or more ECG electrodes disconnected
ECG cable is not connected to monitor
Poor electrode-skin contact • Reposition cable or lead wires to
PACER was pressed. The
monitor automatically switched to Lead II, but ECG leads are not connected.
Broken ECG cable lead wire • Select another lead.
• Connect therapy electrode.
• Connect ECG electrode.
• Connect ECG cable.
prevent electrodes from pulling away from patient.
• Secure trunk cable clasp to patient’s clothing.
• Prepare skin and apply new electrodes.
• Connect ECG leads and initiate pacing.
• Select
• Check ECG cable continuity.
PADDLES lead, and use
standard paddles or therapy electrodes for ECG monitoring.
4-12
Screen blank and illuminated
Systole beeps not heard or do not occur with each QRS complex
Displayed heart rate (HR) different than pulse rate
Displayed heart rate (HR) different from displayed ECG waveform
ON LED
Screen not functioning properly
Volume too low • Adjust volume.
QRS amplitude too small to detect
ECG size set too high or too low
Monitor detecting the patient’s internal pacemaker pulses
ECG size set too high or too low
Monitor detecting the patient’s internal pacemaker pulses
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
• Print ECG on recorder as backup.
• Contact service personnel for repair.
• Adjust ECG size.
• Adjust ECG size
• Change monitor lead to reduce internal pacemaker pulse size.
• Adjust ECG size
• Change monitor lead to reduce internal pacemaker pulse size.
up or down.
up or down.
MONITORING
Table 4-2 Troubleshooting Tips for ECG Monitoring (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
4
Monitor displays dashes (---) instead of heart rate
Poor ECG signal quality Poor electrode-skin contact • Reposition cable or lead wires to
Heart rate is < 20 bpm • Use ECG printout to calculate
heart rate.
Heart rate is > 300 bpm • Use ECG printout to calculate
heart rate.
Pacing function is active • No corrective action needed.
prevent electrodes from pulling away from patient.
• Secure trunk cable clasp to patient’s clothing.
• Prepare skin and apply new electrodes.
Outdated, corroded, or dried­out electrodes
Loose connection. Damaged cable or connector/ lead wire
• Check Use By date on electrode packages.
• Use only unexpired silver/silver chloride electrodes. Leave electrodes in sealed pouch until time of use.
• Check or reconnect cable connections.
• Inspect ECG and therapy cables. Replace if damaged.
• Check cable with simulator and replace if malfunction observed.
Baseline wander (low frequency/high amplitude artifact)
Fine baseline artifact (high frequency/low amplitude)
©2007-2013 Physio-Control, Inc.
Noise because of radio frequency interference (RFI)
Inadequate skin preparation • Prepare skin and apply new
Poor electrode-skin contact • Check electrodes for proper
Diagnostic frequency response • Print ECG in monitor frequency
Inadequate skin preparation • Prepare skin and apply new
Isometric muscle tension in arms/legs
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
• Check for equipment causing RFI (such as a radio transmitter) and relocate or turn off equipment power.
electrodes.
adhesion.
response.
electrodes.
• Confirm that limbs are resting on a supportive surface.
• Check electrodes for proper adhesion.
4-13
Monitoring the ECG
Table 4-2 Troubleshooting Tips for ECG Monitoring (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
ECG amplitude too small Poor electrode-skin contact • Prepare skin and apply new
electrodes.
ECG lead selected • Increase ECG gain or change ECG
lead.
Monitor displays dashed lines with no ECG
OFF
messages
Monitor shows isoelectric (flat) line and selected
Internal pacemaker pulses difficult to see
LEADS
PADDLES lead
Patient condition (for example, significant myocardial muscle loss or tamponade)
PADDLES lead selected but
patient connected to ECG cable
The Test Load is connected to therapy cable
Pacemaker pulses are very small
Monitor frequency response limits visibility
• Increase ECG gain or change ECG
• Select one of the limb or
• Remove the Test Load and
• Connect ECG cable and select
• Turn on internal pacemaker
• Connect ECG cable and select a
• Print ECG in Diagnostic mode
For general troubleshooting tips, see Table 10-2 on page 10-18.
lead.
precordial leads.
connect therapy electrodes to cable.
another lead.
detector (see "Monitoring Patients Who Have Internal Pacemakers" on page 4-11).
lead other than
(see "How to Print a Current Report" on page 7-10).
PADDLES.
4-14
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING

Acquiring a 12-Lead ECG

Intended Use

The 12-lead ECG offers paramedics and emergency physicians significant advantages over the single lead ECG trace typically available in EMS. The 12-lead ECG not only provides a diagnostic quality ECG for use in the detection of ST elevation myocardial infarction (STEMI), but also allows the knowledgeable paramedic to determine the area of myocardial injury, anticipate associated potential complications, and implement treatment strategies accordingly. In addition, the 12-lead ECG provides a baseline for serial ECG evaluations.
The 12-lead ECG transmission to the emergency department (ED) is recommended by the AHA and ERC for patients with Acute Coronary Syndrome (ACS). When transmitted from the field, 12-lead ECG has been shown to shorten time to in-hospital treatment by an estimated 10 to 60 minutes. Patients may also benefit from triage and transport to the most appropriate facility. Documentation of transient or intermittent arrhythmias and other electrophysiologic events that occur in the prehospital setting can assist in diagnosis and treatment decisions in the ED.
4
Indications
The 12-lead electrocardiogram is used to identify, diagnose, and treat patients with cardiac disorders and is useful in the early detection and prompt treatment of patients with acute ST­elevation myocardial infarction (STEMI).
Contraindications
None known.

12-Lead ECG Warning

WARNING
POSSIBLE INABILITY TO OBTAIN A DIAGNOSTIC 12-LEAD ECG
Using previously unpackaged electrodes or electrodes past the Use By date may impair ECG signal quality. Remove electrodes from a sealed package immediately before use and follow the procedure for applying the electrodes.

Identifying Electrode Sites

To obtain a 12-lead ECG, place the electrodes on the limbs and the chest (precordium) as described in the following paragraph.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-15
Acquiring a 12-Lead ECG
RA/R
LA/L
LL/FRL/N
AHA Labels IEC Labels RA Right Arm R Right LA Left Arm L Left RL Right Leg N Negative LL Left Leg F Foot
LEAD LOCATION
V1 C1 Fourth intercostal space to the right of the sternum
V2 C2 Fourth intercostal space to the left of the sternum
V3 C3 Directly between leads V2/C2 and V4/C4
V4 C4 Fifth intercostal space at midclavicular line
V5 C5 Level with V4/C4 at left anterior axillary line
V6 C6 Level with V5/C5 at left midaxillary line
Angle of Louis
Limb Lead Electrode Sites
When acquiring a 12-lead ECG, limb lead electrodes are typically placed on the wrists and ankles as shown in Figure 4-4. The limb lead electrodes can be placed anywhere along the limbs. Do not place the limb lead electrodes on the torso when acquiring a 12-lead ECG.
Figure 4-4 Limb Lead Electrode Placement for 12-Lead ECG
Precordial Lead Electrode Sites
The six precordial (chest) leads are placed on specific locations as shown and summarized in Figure 4-5. Proper placement is important for accurate diagnosis and should be identified as follows: leads are V1 through V6 for AHA, or C1 through C6 for IEC. See "ECG Leads Color Codes" on page 4-10 for color codes.
Figure 4-5 Precordial Lead Electrode Placement
Locating the V1/C1 position (fourth intercostal space) is critically important, because it is the reference point for locating the placement of the remaining V/C leads.
4-16
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
Protective flap
(open)
Main cable
(4-wire)
Precordial lead
attachment
Precordial lead
attachment
Limb lead
attachment
Main cable
To locate the V1/C1 position:
1. Place your finger at the notch in the top of the sternum.
2. Move your finger slowly downward about 3.8 centimeters (1.5 inches) until you feel a slight horizontal ridge or elevation. This is the Angle of Louis where the manubrium joins the body of the sternum.
3. Locate the second intercostal space on the patient’s right side, lateral to and just below the Angle of Louis.
4. Move your finger down two more intercostal spaces to the fourth intercostal space, which is the V1/C1 position.
5. Continue locating other positions from V1/C1 (see Figure 4-5).
Other important considerations:
• When placing electrodes on female or obese patients, always place leads V3-V6 and C3-C6 under the breast rather than on the breast.
• Never use the nipples as reference points for locating the electrodes for men or women patients, because nipple locations vary widely.
4

12-Lead ECG Procedure

To acquire a 12-lead ECG:
1. Press
2. Insert the lead attachments into the main cable as shown in Figure 4-6.
3. Insert the cable connector into the green ECG connector on the monitor.
4. Prepare patient’s skin for electrode application (see page 4-9).
ON.
Figure 4-6 12-Lead ECG Cables
5. Apply ECG electrodes (see page 4-16).
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-17
Acquiring a 12-Lead ECG
6. Encourage the patient to remain as still as possible.
WARNING
POSSIBLE INACCURATE DIAGNOSIS
If age and sex are not entered when a 12-lead ECG is obtained, the interpretive statements are based on a default of a 50-year-old male and may provide incorrect analysis for that patient.
7. Press
8. The
The monitor acquires, analyzes, and automatically prints the 12-lead ECG. An ECG leads-off condition for any lead is indicated on the report by a dashed line.
12-LEAD. The 12-LEAD / AGE menu appears, prompting you to enter the patient’s age.
Use the or younger. If you do not enter an age, the default value of 50 years is used by the interpretive analysis program and annotated on the 12-lead ECG report.
Use the used by the interpretive analysis program and is annotated on the 12-lead ECG report.
SPEED DIAL to select the age. Always enter the patient’s age if the patient is 15 years old
12-LEAD / SEX menu appears, prompting you to enter the patient’s sex.
SPEED DIAL to select the patient’s sex. If you do not enter the sex, the default of male is
Note: If 15 years or less is entered for patient age, the 12-lead ECG prints at diagnostic frequency
response of 0.05–150 Hz, even when 0.05–40 Hz is set up as the print default.
Note: When 12-LEAD is pressed, internal pacemaker detection is automatically enabled, even if the
function is set up to be OFF.

ECG Override

If the monitor detects signal noise while acquiring data (such as patient motion or a disconnected electrode), the screen displays the message: remains and 12-lead ECG acquisition is interrupted until noise is eliminated. Take appropriate action to eliminate the signal noise. This message remains as long as signal noise is detected. When signal noise is eliminated, the monitor resumes acquiring data. To override the message and acquire the 12-lead ECG in spite of the signal noise, press acquired and printed with no interpretive statements. Any 12-lead ECG report acquired in this way is annotated with the following statement:
NOISY DATA! PRESS 12-LEAD TO ACCEPT. The message
12-LEAD again. The 12-lead ECG will be
ECG OVERRIDE: DATA QUALITY PROHIBITS INTERPRETATION.
If the signal noise persists for longer than 30 seconds, 12-lead ECG acquisition stops. The screen displays acquisition.
EXCESSIVE NOISE–12-LEAD CANCELLED. You must then press 12-LEAD to restart 12-lead ECG
Note: If 12-LEAD is pressed immediately after ECG electrodes are applied, the message NOISY DATA
may occur. This message is due to the temporary instability between the electrode gel and the patient’s skin that is not viewable on the ECG monitor screen, but is detected as noisy data. In
4-18
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
general, it is best to wait at least 30 seconds after applying the last electrode before pressing the
12-LEAD button, to allow for electrode/skin stabilization. Also, good skin preparation shortens the
stabilization time.

Computerized ECG Analysis

Computerized ECG analysis statements are automatically printed on 12-lead ECG reports. Printing of the interpretive statements is a setup option and may be turned off in Setup mode. For information on how to change this setup option, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.
The interpretative statements pertaining to myocardial injury, infarct, and ischemia are derived from measurements made on a signal-averaged beat (median beat) formed for each of the 12 leads. The computerized ECG analysis selects three representative beats from the ten seconds of data for each lead and averages the three beats to derive the median beat for that lead. The ECG analysis is always based on ECG data obtained at 0.05–150 Hz frequency response.
The analysis program is adjusted for patient age and sex. The 12-lead ECG interpretive algorithm used by the LIFEPAK 15 monitor/defibrillator is the University of Glasgow 12-Lead ECG Analysis Program. For more information, contact your Physio-Control representative for a copy of the Physio- Control Glasgow 12-Lead ECG Analysis Program Physician’s Guide.
4
WARNING
POSSIBLE INCORRECT TREATMENT WITH REPERFUSION THERAPY
Computerized ECG interpretive statements should not be used to withhold or prescribe patient treatment without review of the ECG data by qualified medical personnel. All 12-lead ECG interpretation statements provided by the LIFEPAK 15 monitor/ defibrillator include the printed message **UNCONFIRMED**. Always confirm interpretive statements by over-reading the ECG data.

Printed 12-Lead ECG Report Formats

Two 12-lead ECG report formats are available for printing: 3-channel or 4-channel. In addition, each of those formats can be printed in standard and cabrera styles.
The 3-Channel Format
The 3-channel format prints 2.5 seconds of data for each lead. Figure 4-7 is an example of a 12-lead ECG report printed in the 3-channel format, standard style. Figure 4-8 is an example of a 12-lead ECG report printed in the 3-channel format, cabrera style. The sequence in which the limb leads are presented differs between the standard and cabrera styles, as shown. The default format for printing 12-lead ECG reports is 3-channel standard. To change the printed format of
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-19
Acquiring a 12-Lead ECG
I II III
V1 V2 V3 V4 V5 V6
0.00 mm
0.00 mm
-1.10 mm
0.00 mm
0.00 mm
0.00 mm
0.00 mm
1.80 mm
2.50 mm
2.76 mm
0.00 mm
0.00 mm
Standard
measurement
Frequency
response
ECG size
Printer
speed
Device
number
Site
number
Software
version
Serial
number
Time/date
12-lead acquired
Computerized
ECG analysis
Patient
ID
Configuration
code
Lead
annotation
1 mV
reference
Report type
and number
Name:
Record ID: Patient ID: Incident: Age: 50
Lee, William
041495091422
528760224
BF382
Sex: M
12-Lead 1
HR 78 bpm
24 Apr 08 PR 0.170s QT/QTc P-QRS-T Axes
Abnormal ECG ** Unconfirmed** *** MEETS ST ELEVATION MI
CRITERIA ***
Sinus rhythm Anteroseptal ST elevation, CONSIDER
ACUTE INFARCT
aVR
aVL
aVF
QRS 0.094s
V1
V2
V3
V4
V5
V6
STJ Level
aVR aVL aVF
-
I II III
0.00 mm
0.00 mm
-1.10 mm
0.00 mm
0.00 mm
0.00 mm
0.00 mm
1.80 mm
2.50 mm
2.76 mm
0.00 mm
0.00 mm
Name: Lee, William
Record ID: Patient ID: Incident: Age: 50
12-Lead 1
HR 78 bpm
Abnormal ECG ** Unconfirmed** *** MEETS ST ELEVATION MI CRITERIA***
Sinus rhythm Anteroseptal ST elevation, CONSIDER
ACUTE INFARCT 24 Apr 08 PR 0.170s QT/QTc P-QRS-T Axes
aVL
aVR
aVF
V1
V2
V3
V4
V5
V6
Sex:M
QRS 0.094s
STJ Level
aVR aVL aVF V1 V2 V3 V4 V5 V6
12-lead ECG reports, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device. Alternatively, press
FORMAT.
OPTIONS, select PRINT, select REPORT: 12-LEAD, and then select
4-20
Figure 4-7 Example of Printed 3-Channel, Standard 12-Lead ECG Report
Figure 4-8 Example of Printed 3-Channel, Cabrera 12-Lead ECG Report
The 4-Channel Format
Figure 4-9 and Figure 4-10 are examples of 12-lead ECG reports printed in the 4-channel format. The 4-channel format consists of the median complex (or median beat) derived for each of the 12 leads and 10 seconds of data for Lead II.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
I II III
V1 V2 V3 V4 V5 V6
0.00 mm
0.00 mm
-1.10 mm
0.00 mm
0.00 mm
0.00 mm
0.00 mm
1.80 mm
2.50 mm
2.76 mm
0.00 mm
0.00 mm
Fiducial marks
Record ID: Patient ID: Incident: Age: 50
Sex: M
12-Lead 1
P-QRS-T Axes
24 Apr 08 PR .168s QT/QTc
HR 78 bpm
QRS 0.104s
Abnormal ECG **Unconfirmed** ***Meets ST ELEVATION MI
Sinus rhythm Anteroseptal ST elevation, CONSIDER ACUTE INFARCT
aVR
aVL
aVF
V1
V2
V3
V4
V5
V6
Lee, WilliamName:
STJ Level
aVR aVL aVF
I II III
V1 V2 V3 V4 V5 V6
0.00 mm
0.00 mm
-1.10 mm
0.00 mm
0.00 mm
0.00 mm
0.00 mm
1.80 mm
2.50 mm
2.76 mm
0.00 mm
0.00 mm
Record ID: Patient ID: Incident: Age: 50
Sex: M
12-Lead 1 24 Apr 08 PR .168s QT/QTc P-QRS-T Axes
Abnormal ECG **Unconfirmed** ***Meets ST ELEVATION MI CRITERIA***
Sinus rhythm Anteroseptal ST elevation, CONSIDER ACUTE INFARCT
Name:
Lee, William
HR 89 bpm
QRS 0.104s
aVL
-aVR
aVF
V1
V2
V3
V4
V5
V6
STJ Level
aVR aVL aVF
Note: The fiducial marks displayed in the 4-channel format identify the measurement intervals
used for the interpretive statements of the analysis program. These marks are part of the analysis program and cannot be turned off.
Figure 4-9 Example of Printed 4-Channel, Standard 12-Lead ECG Report
4
Figure 4-10 Example of Printed 4-Channel, Cabrera 12-Lead ECG Report

Printed 12-Lead ECG Frequency Response

The 12-lead ECG can be printed in two diagnostic frequency responses (or bandwidths):
0.05–40 Hz and 0.05–150 Hz. The frequency response of 0.05–150 Hz is the Association for the Advancement of Medical Instrumentation (AAMI) standard for diagnostic ECGs. The 0.05–40 Hz setting preserves the low frequency limit that is needed for the diagnosis of myocardial ischemia and infarction while reducing high frequency artifact (in particular from patient muscle tension) to help make the diagnostic printout less noisy and more readable.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-21
Acquiring a 12-Lead ECG
Note: The LIFEPAK 15 monitor/defibrillator acquires ECG data and performs the interpretive
analysis based on the full frequency of 0.05–150 Hz. The 0.05–40 Hz bandwidth affects only the printed appearance of the ECG data.
The 12-lead ECG printed in the 0.05–40 Hz setting can be used to diagnose acute myocardial ischemia and ST-segment elevation myocardial infarction (STEMI). This is because the low frequency limit of 0.05 Hz is not changed from the standard diagnostic setting of 0.05–150 Hz. The 0.05 Hz frequency provides accurate representation of low frequency signals, that is, the P, ST segment, and T waves. The presence or absence of ST segment changes indicative of myocardial ischemia or infarction will be accurately reproduced. In addition, the criteria for visual analysis and interpretation of cardiac rhythm and PR, QRS, and QT intervals are preserved, as is true with hospital cardiac monitors that have an upper frequency limit of 40 Hz.
However, in some adult patients, the amplitude (that is, voltage) of the QRS may be reduced when 12-lead ECGs are printed at the upper limit of 40 Hz rather than at 150 Hz. Therefore, certain diagnoses, which depend on R wave amplitude (for example, ventricular hypertrophy), should not be made using this setting. In the pediatric patient, this effect on R wave amplitude is particularly noticeable because QRS durations in children are typically quite narrow. Because R wave amplitude reduction is more likely with pediatric patients, the 12-lead ECG automatically prints at 0.05–150 Hz, overriding the 40 Hz limit, when a patient age of 15 years or younger is entered.
4-22
LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Troubleshooting Tips

Table 4-3 Troubleshooting Tips for the 12-Lead ECG
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
MONITORING
4
Any of these messages displayed:
CONNECT ECG LEADS ECG LEADS OFF XX LEADS OFF
Noisy signal and/or message displayed:
NOISY DATA! PRESS 12-LEAD TO ACCEPT
One or more ECG electrodes disconnected
ECG cable is not connected to monitor
Poor electrode-skin contact
Broken lead wire • Select another lead.
Noise in a lead other than the displayed lead
Poor electrode-skin contact
• Confirm ECG electrode connections.
• Confirm ECG cable connections.
• Reposition cable and/or lead wires to prevent electrodes from pulling away from patient.
• Secure trunk cable clasp to patient’s clothing.
• Prepare skin and apply new electrodes.
• Select
• Check ECG cable continuity.
•Press 12-LEAD again to override the
• Reposition cable or lead wires to prevent
• Secure trunk cable clasp to patient’s
• Prepare skin and apply new electrodes.
PADDLES lead, and use standard
paddles or therapy electrodes for ECG monitoring.
message. Examine the printout to determine leads affected by noise. Replace or reposition the affected electrodes and lead wires.
electrodes from pulling away from patient.
clothing.
©2007-2013 Physio-Control, Inc.
Loose connection • Check or reconnect cable connections.
Patient motion • Encourage patient to lie quietly.
• Support patient’s limbs.
Vehicle motion • Stop vehicle while acquiring 12-lead ECG
data.
Outdated, corroded, or dried-out electrodes
Radio Frequency Interference (RFI)
Damaged cable or connector/lead wire
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
• Check Use By date on electrode packages.
• Use only unexpired silver/silver chloride electrodes. Leave electrodes in sealed pouch until time of use.
• Check for equipment causing RFI (such as a radio transmitter) and relocate or turn off equipment power.
• Inspect main cable and attachments. Replace if damaged.
4-23
Acquiring a 12-Lead ECG
Table 4-3 Troubleshooting Tips for the 12-Lead ECG (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
Monitor does not complete 12-lead ECG operation sequence
Noisy signal and message displayed:
EXCESSIVE NOISE– 12-LEAD CANCELLED
Baseline wander (low frequency/high amplitude artifact)
Fine baseline artifact (high frequency/low amplitude)
Operator pressed another function button (such as
PRINT) before 12-lead
ECG sequence completed
Signal noise for more than 30 seconds
Inadequate skin preparation
Poor electrode-skin contact
Inadequate skin preparation
Isometric muscle tension in arms/legs
•Press
•Press 12-LEAD to acquire another 12-lead
• Prepare skin as described on page 4-8 and
• Check electrodes for proper adhesion.
• Prepare skin as described on page 4-9 and
• Confirm that limbs are resting on a
• Check electrodes for proper adhesion.
12-LEAD to acquire another 12-lead
ECG. Allow enough time for sequence to complete.
ECG.
apply new electrodes.
apply new electrodes.
supportive surface.
For general troubleshooting tips, see Table 10-2 on page 10-18.
4-24
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING

Monitoring SpO2, SpCO, and SpMet

SpO2, SpCO™, and SpMet™ are optional features for the LIFEPAK 15 monitor/defibrillator. When all three options (SpO oxygen saturation (SpO concentration (SpMet) in the blood.
, SpCO, and SpMet) are installed, the pulse oximeter measures functional
2
), carboxyhemoglobin concentration (SpCO), and methemoglobin
2
4
IMPORTANT! SpO
for use. Masimo monitor. Masimo Rainbow™ sensors are necessary to monitor SpCO and SpMet in addition to SpO
. These sensors are not compatible with other LIFEPAK defibrillator/monitors.
2
Nellcor SpO MNC adapter cable is used.
For a list of SpO monitor/defibrillator, see the Physio-Control web site. Carefully read the Directions for Use that are provided with the sensors and connector cables for a complete description, instructions, warnings, cautions, and specifications. To order sensors and connector cables, contact your Physio-Control representative or order online at store.physio-control.com.
sensors may be used with the LIFEPAK 15 monitor/defibrillator, if the Masimo Red™
2
-only sensors and combination SpO2, SpCO, and SpMet sensors are available
2
®
SpO2-only sensors that have a red connector are compatible with the LIFEPAK 15
sensors and connector cables that are intended for use with the LIFEPAK 15
2

Intended Use

A pulse oximeter is a noninvasive device that continuously measures functional oxygen saturations (SpO
), carboxyhemoglobin concentration (SpCO), and methemoglobin concentration (SpMet) in
2
the blood. Continuously monitoring SpO decreasing and can help the clinician act rapidly before the patient develops the later signs of hypoxemia. Previously, the blood parameters SpCO and SpMet could only be obtained from invasive blood gas samples. This new technology assists in identifying the often hidden conditions of carboxyhemoglobinemia (carbon monoxide poisoning) and methemoglobinemia (a condition that impedes delivery of oxygen to the tissues). Low levels of both SpCO and SpMet are normally found in the blood; however, early detection of significantly high levels can lead to proper diagnosis and treatment, and can help improve patient outcome.
can provide an early warning when oxygen saturation is
2
Pulse oximetry is a tool to be used in addition to patient assessment. Care should be taken to assess the patient at all times; do not rely solely on the SpO a trend toward patient deoxygenation is evident or carbon monoxide poisoning or methemoglobinemia is suspected, blood samples should also be analyzed using laboratory instruments to completely understand the patient’s condition.
Do not use the pulse oximeter to monitor patients for apnea, or as a replacement or substitute for ECG-based arrhythmia analysis.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
, SpCO, and SpMet measurements. If
2
4-25
Monitoring SpO2, SpCO, and SpMet
Indications
Pulse oximetry is indicated for use in any patient who is at risk of developing hypoxemia, carboxyhemoglobinemia, or methemoglobinemia. SpO motion and motion conditions, and in patients who are well or poorly perfused. SpCO and SpMet accuracies have not been validated under motion or low perfusion conditions.
Contraindications
None known.

SpO2, SpCO, and SpMet Warnings and Cautions

WARNINGS
SHOCK OR BURN HAZARDS
SHOCK OR BURN HAZARD
Before use, carefully read these operating instructions, the sensor and cable directions for use, and precautionary information.
monitoring may be used during no
2
SHOCK OR BURN HAZARD
Using other manufacturers’ sensors or cables may cause improper oximeter performance and invalidate safety agency certifications. Use only sensors and cables that are specified in these operating instructions.
INACCURATE READINGS HAZARDS
INACCURATE PULSE OXIMETER READINGS
Do not use a damaged sensor or cable. Do not alter the sensor or cable in any way. Alterations or modification may affect performance and/or accuracy. Never use more than one cable between the pulse oximeter and the sensor to extend the length.
INACCURATE PULSE OXIMETER READINGS
Sensors exposed to ambient light when incorrectly applied to a patient may exhibit inaccurate saturation readings. Securely place the sensor on the patient and check the sensor’s application frequently to help ensure accurate readings.
INACCURATE PULSE OXIMETER READINGS
Severe anemia, hypothermia, severe vasoconstriction, carboxyhemoglobin, methemoglobin, intravascular dyes that change usual blood pigmentation, elevated bilirubin, excessive patient movement, venous pulsations, electrosurgical interference, exposure to irradiation and placement of the sensor on an extremity that has a blood pressure cuff, intravascular line, or externally applied coloring (such as nail polish) may interfere with oximeter performance. The operator should be thoroughly familiar with the operation of the oximeter prior to use.
4-26
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
WARNINGS (CONTINUED)
INACCURATE PULSE OXIMETER READINGS
The pulsations from intra-aortic balloon support can be additive to the pulse rate on the oximeter pulse rate display. Verify patient’s pulse rate against the ECG heart rate.
POSSIBLE SKIN INJURY
Prolonged, continuous use of a sensor may cause irritation, blistering, or pressure necrosis of the skin. Check the sensor site regularly based on patient condition and type of sensor. Change the sensor site if skin changes occur. Do not use tape to hold the sensor in place as this may cause inaccurate readings or damage to the sensor or skin.
POSSIBLE STRANGULATION
Carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
4
CAUTIONS
EQUIPMENT HAZARDS
POSSIBLE EQUIPMENT DAMAGE
To avoid damage to the cable, always hold by the connector rather than the cable, when connecting or disconnecting either end.
POSSIBLE EQUIPMENT DAMAGE
Do not soak or immerse the sensors or cables in any liquid solution. Do not attempt to sterilize.

No Implied License

Possession or purchase of the pulse oximeter does not convey any expressed or implied license to use the pulse oximeter with unauthorized sensors or cables which would, alone or in combination with this device, fall within the scope of one or more of the patents relating to this device.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-27
Monitoring SpO2, SpCO, and SpMet
Sensor (holds LEDs and detector)
Red
Infrared
Light-emitting diodes
Light-receiving detector
Cable

How a Pulse Oximeter Works

A pulse oximeter sensor directs light through a patient’s fleshy body site (usually a finger or toe). The sensor sends wavelengths of light from the emitter to the receiving detector as shown in Figure 4-11.
Figure 4-11 How a Pulse Oximeter Works
The pulse oximeter translates the amount of light received by the detector to the various forms of hemoglobin saturation levels and displays them as SpO values for SpO 9% (the higher range of normal is often seen in smokers). Normal values for SpMet are typically less than 2% and may be caused by exposure to some pharmaceuticals including local anesthetic agents and chemical agents such as nitrites.

SpO2, SpCO, and SpMet Monitoring Considerations

The quality of the SpO2, SpCO, and SpMet readings depends on correct sensor size and placement, adequate blood flow through the sensor site, and limiting patient motion and sensor exposure to ambient light. For example, with very low perfusion at the sensor site, readings may be lower than core arterial oxygen saturation. Test methods for accuracy are available by contacting your local Physio-Control representative.
Use the following criteria to select the appropriate pulse oximeter sensor:
• Patient size (adult, pediatric, infant) and weight
• Patient perfusion to extremities
• Patient activity level
• Available application sites on the patient’s body
• Sterility requirements
• Anticipated duration of monitoring
, SpCO, and SpMet percentages. Normal
2
typically range from 95% to 100%. Normal values for SpCO are typically less than
2
4-28
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
To help ensure optimal performance:
• Use a dry and appropriately sized sensor.
• Choose a site that is well perfused. The ring finger is preferred.
• Choose a site that least restricts patient movement, such as finger of the non-dominant hand.
• Be sure the fleshy part of the digit completely covers the detector.
• Keep the sensor site at the same level as the patient’s heart.
• Apply the sensor according to the Directions for Use provided with the sensor.
• Observe all warnings and cautions noted in the sensor’s Directions for Use.
Sensor Application
The preferred site for sensor application is the ring finger of the non-dominant hand. To position the sensor:
1. Orient the sensor so the cable is on the back of the patient’s hand.
2. Place the finger in the sensor until the tip of the finger touches the “raised digit stop.”
4
3. The hinged tabs of the sensor should open to evenly distribute the grip pressure of the sensor along the length of the finger. Check the arrangement of the sensor to verify correct positioning. Complete coverage of the detector window is needed to ensure accurate data.
The sensors are sensitive to light. If excessive ambient light is present, remove or reduce lighting, cover the sensor site with an opaque material to block the light, and check appropriateness of sensor site. Failure to do so could result in inaccurate measurements.
If excessive movement presents a problem during SpCO/SpMet monitoring, consider the following possible solutions:
• Be sure the sensor is secure and properly aligned.
• Use a disposable adhesive sensor.
• If possible, move the sensor to a less active site.
Note: Wrapping the sensor too tightly or using supplemental tape to hold the sensor in place
may cause inaccurate oximeter readings.
Note: Circulation distal to the sensor site should be checked routinely.
IMPORTANT! Masimo Rainbow sensors are necessary to monitor SpCO and SpMet and are not
compatible with other LIFEPAK defibrillator/monitors.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-29
Monitoring SpO2, SpCO, and SpMet

Oximeter Monitoring Procedure

Power to the pulse oximeter is controlled by the LIFEPAK 15 monitor/defibrillator. When the defibrillator is turned on, the oximeter turns on and performs a calibration and self-test that requires approximately 20 seconds. During the calibration and self-test, the screen does not display SpO
, SpCO, or SpMet information.
2
To conserve battery power, the pulse oximeter goes into “sleep mode” when not in use. Sleep mode is activated within 10 seconds of disconnecting the sensor. During sleep mode, the screen does not display SpO oximeter performs a self-test and then returns to normal mode.
, SpCO, or SpMet information. When a sensor or patient signal is detected, the
2
The pulse oximeter measures and displays SpO than 50% are displayed as <50. When SpO
levels between 50 and 100%. SpO2 levels less
2
levels are between 70 and 100%, oximeter
2
measurements are accurate ±3 digits. The pulse oximeter measures and displays SpCO in the range of 0–40% with accuracy of ±3 digits. The pulse oximeter measures and displays SpMet in the range of 0–15% with accuracy of ± 1 digit.
To monitor SpO
1. Press
ON.
:
2
2. Connect the pulse oximeter cable to the monitor and sensor.
3. Attach the sensor to the patient.
4. Observe the pulse bar for fluctuation. Amplitude of the pulse bar indicates relative signal quality.
5. Confirm that the SpO
6. Use the
SPEED DIAL to adjust volume, sensitivity, and averaging time, as necessary.
reading appears and is stable.
2
To monitor SpCO or SpMet:
1. Perform Step 2 through Step 5 above.
2. Verify that an SpCO/SpMet sensor is in use. Only Rainbow sensors are capable of reading SpCO/ SpMet.
3. Encourage the patient to remain still.
4. To quickly obtain SpCO or SpMet value, press
PRINT. If dashes (---) appear on printout instead of
values for SpCO or SpMet, allow a few more seconds for measurement to be obtained.
or
To display SpCO or SpMet:
•Use the
•Select
•Select
SPEED DIAL to select the SpO
PARAMETER from menu.
SPCO or SPMET. Selected value displays for 10 seconds.
area.
2
Note: SpCO and SpMet monitoring are not intended for use under patient motion or low perfusion
conditions.
4-30
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING

SpCO/SpMet Advisory

If the SpCO or SpMet reading is above normal limits, indicating a dangerous amount of carboxyhemoglobin or methemoglobin, an Advisory occurs.
During an Advisory:
4
• The elevated SpCO or SpMet value is displayed instead of SpO
• The elevated value flashes and the alarm tone sounds.
• One of the following Advisory messages appears in the message area:
Advisory: SpCO > 10%
Advisory: SpMet > 3%
To cancel the Advisory, press message remains on the screen until the elevated value returns to within normal limits or the device is turned off.
ALARMS. The SpO
area reverts to the SpO2 reading. The Advisory
2
.
2
WARNING
INACCURATE SPO2 READINGS
Carboxyhemoglobin and methemoglobin may erroneously increase SpO2 readings. The amount that SpO carboxyhemoglobin or methemoglobin that is present.
INACCURATE SPCO AND SPMET READINGS
Very low arterial oxygen saturation levels may cause inaccurate SpCO and SpMET readings.
increases is approximately equal to the amount of
2

The Pleth Waveform

You can display the plethysmographic (pleth) waveform in Channel 2 or 3.
To display the pleth waveform:
1. Rotate the
2. Press the
3. Select The waveform is automatically sized for optimum waveform viewing.
©2007-2013 Physio-Control, Inc.
SPEED DIAL to outline waveform CHANNEL 2 or 3.
SPEED DIAL. The Channel menu appears.
WAVEFORM and then select SPO2. The SpO
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
waveform appears in the selected channel.
2
4-31
Monitoring SpO2, SpCO, and SpMet
1. Rotate the SPEED DIAL to outline the SpO2 area on the Home Screen.
2. Press the
SPEED DIAL.
3. Highlight and select SPO2 VOLUME.
4. Rotate the
SPEED DIAL to the desired
volume.
5. Press the
SPEED DIAL to set the volume.
SpO2_SpCo_SpMet
Parameter SpO2 Volume
Sensitivity
Averaging Time
SpO2
Normal
8 Seconds
Volume
To adjust the pulse tone volume:
Sensitivity
The sensitivity setting allows you to adjust the oximeter to either NORMAL or HIGH for differing perfusion states.
To adjust sensitivity:
4-32
1. Outline and select the SpO
2. Select
SENSITIVITY and then select NORMAL or HIGH.
area on the Home Screen.
2
Note: NORMAL sensitivity is recommended for most patients. The HIGH sensitivity setting allows
SpO
monitoring under low perfusion states, such as the severe hypotension of shock. However,
2
when SpO patient closely when using the
sensitivity is set to HIGH, the signal is more susceptible to artifact. Monitor the
2
HIGH sensitivity setting.
Averaging Time
Averaging time allows you to adjust the time period that is used to average the SpO2 value.
To adjust averaging time:
1. Outline and select the SpO
2. Select
AVERAGING TIME and then select one of the following:
• 4 Seconds
• 8 Seconds
•12 Seconds
•16 Seconds
area on the Home Screen.
2
Note: Averaging time of 8 seconds is recommended for most patients. For patients with rapidly
changing SpO artifact is affecting the performance of the pulse oximeter.
values, 4 seconds is recommended. Use a 12- or 16-second time period when
2
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING

Pulse Rate Monitoring

If ECG monitoring is not active, the SpO2 sensor can be used to monitor the patient’s pulse rate. The pulse rate value is indicated by
Pulse rate monitoring is a tool to be used in addition to patient assessment. Care should be taken to assess the patient at all times. Check pulse manually if patient shows signs of abnormal pulse rate.

Cleaning

Pulse oximetry sensors may be adhesive (single-patient use) or reusable.
To clean the reusable sensor and connector cable:
1. Disconnect the sensor and cable from the monitor. Inspect the cable for damage.
2. Use a clean, soft cloth dampened with 70% isopropyl alcohol to wipe clean.
3. Allow to dry thoroughly before placing the sensor on a patient or reconnecting the cable to the monitor.
Note: Do not attempt to sterilize. Do not soak or immerse in any liquid solution. For information
about cleaning the device, see "Cleaning the Device" on page 10-15.
PR (SPO2).
4

Troubleshooting Tips

Table 4-4 Troubleshooting Tips for SpO2, SpCO, and SpMet
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
The monitor measures a pulse, but there is no oxygen saturation or pulse rate
or pulse rate changes
SpO
2
rapidly, pulse amplitude is erratic
Excessive patient motion • Keep patient still.
• Check that sensor is secure.
• Relocate sensor.
• Apply adhesive sensor.
Patient perfusion may be too low
Excessive patient motion • Keep patient still.
An electrosurgical unit (ESU) may be interfering with performance
• Check patient.
• Increase sensitivity.
• Check that sensor is secure.
• Relocate sensor.
• Apply adhesive sensor.
• Increase sensitivity.
• Move the monitor as far as possible from the ESU.
• Plug the ESU and monitor into different circuits.
• Move the ESU ground pad as close to the surgical site as possible.
©2007-2013 Physio-Control, Inc.
Sensor may be damp • Replace sensor.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-33
Monitoring SpO2, SpCO, and SpMet
Table 4-4 Troubleshooting Tips for SpO
, SpCO, and SpMet (Continued)
2
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
SPO2: NO SENSOR DETECTED
message appears
No SpO value (---) is displayed
, SpCO, or SpMet
2
Sensor not connected to patient or cable disconnected from monitor/defibrillator
Damaged cable or sensor • Replace damaged cable or
Sensor may be too tight • Reposition sensor.
Patient is in cardiac arrest or shock
Oximeter may be performing self-calibration or self-test
Defibrillator shock just delivered
• Check that sensor and cable are connected properly.
• Check that appropriate sensor is in use.
sensor.
• Relocate sensor.
• Check patient.
• Wait for completion.
• If values do not display within 30 seconds, disconnect and reconnect sensor. If values do not display within another 30 seconds, replace sensor.
• None. If values do not display within 30 seconds, disconnect and reconnect sensor. If values do not display within another 30 seconds, replace sensor.
4-34
Different SpCO or SpMet measurements on same patient
XXX appears in place of SpO
reading
2
High intensity lights (such as pulsating strobe lights) may be interfering with performance
Damaged cable or sensor • Replace damaged cable or
Every measurement, even on the same patient, can be different
SpO2 module failed. Internal cable failed.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
• Cover sensor with opaque material, if necessary.
sensor.
• Confirm by taking three measurements: ring finger, middle finger, and then index finger; average the results.
• Turn device off and then on again.
• If problem persists, contact qualified service personnel.
MONITORING
4
Table 4-4 Troubleshooting Tips for SpO
, SpCO, and SpMet (Continued)
2
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
SPO2: CHECK SENSOR message
appears
SPO2: UNKNOWN SENSOR
message appears
Sensor is disconnected from patient or cable
Excessive ambient light • Remove or block light
Faulty or defective sensor • Replace sensor.
Patient has a weak pulse or low blood pressure, or the sensor is not properly placed
A sensor that is not Physio­Control approved is connected to the device
• Attach the sensor.
• Check that sensor is secure.
source, if possible.
• Cover sensor with opaque material, if necessary.
• Change sensor location.
• Check if patient perfusion is adequate for sensor location.
• Check that sensor is secure and not too tight.
• Check that sensor is not on extremity with blood pressure cuff or intravascular line.
• Test sensor on someone else.
• Check that the sensor is approved by Physio-Control.
• If using Nellcor sensor, check that it is connected to monitor using Red MNC adapter cable.
Masimo
SPO2: SEARCHING FOR PULSE
message appears
SPO2: LOW PERFUSION message
appears
SP02: POOR QUALITY SIGNAL
message appears
SPCO: POOR QUALITY SIGNAL
message appears
SPMET: POOR QUALITY SIGNAL
message appears
©2007-2013 Physio-Control, Inc.
A sensor is connected to the patient and is searching for a pulse
Patient has a weak pulse • Change sensor location.
When the signal quality is low, the accuracy of the measurement may be compromised
When the signal quality is low, the accuracy of the measurement may be compromised
When the signal quality is low, the accuracy of the measurement may be compromised
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
• Wait for completion.
• Check that sensor and cable are connected properly.
• Move sensor to a better perfused site.
• Check that sensor and cable are connected properly.
• Move sensor to a better perfused site.
• Check that sensor and cable are connected properly.
• Move sensor to a better perfused site.
4-35
Monitoring SpO2, SpCO, and SpMet
Table 4-4 Troubleshooting Tips for SpO
, SpCO, and SpMet (Continued)
2
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
SPCO/SPMET: POOR QUALITY
message appears
SIGNAL
SPO2: SENSOR DOES NOT SUPPORT SPCO OR SPMET
appears
message
Note: Most Rainbow sensor messages (SpO
The
POOR QUALITY SIGNAL message indicates the specific parameter affected.
When the signal quality is low, the accuracy of the measurement may be compromised
SpO
-only sensor used with
2
SpCO/SpMet capable device
, SpCO, and SpMet) are reported as SPO2: (MESSAGE).
2
• Check that sensor and cable are connected properly.
• Move sensor to a better perfused site.
• None necessary, or use Rainbow sensor to measure SpCO or SpMet.
For general troubleshooting tips, see Table 10-2 on page 10-18.
4-36
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING

Monitoring Noninvasive Blood Pressure

Intended Use

The LIFEPAK 15 noninvasive blood pressure (NIBP) monitor measures blood pressure (BP) using the oscillometric measurement technique to determine systolic, diastolic, and mean arterial pressures, and pulse rate. The measurement can be initiated manually or set to recur automatically at predetermined intervals.
Blood pressure measurements determined using this device are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method, within the limits prescribed by the American National Standard Electronic or automated sphygmomanometers (AAMI SP-10).
NIBP is a tool to be used in addition to patient assessment. Care should be taken to assess the patient at all times; do not rely solely on the NIBP monitor.
Indications
4
Noninvasive blood pressure monitoring is intended for detection of hypertension or hypotension and monitoring BP trends in patient conditions such as, but not limited to, shock, acute dysrhythmia, or major fluid imbalance.
Contraindications
None known.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-37
Monitoring Noninvasive Blood Pressure

NIBP Monitoring Warnings and Caution

WARNINGS
POSSIBLE LOSS OF INTRAVENOUS ACCESS AND INACCURATE INFUSION RATE
Do not apply the blood pressure cuff on an extremity that is used for an intravenous infusion. Patency of the intravenous infusion may be affected by blood pressure measurement due to the occlusion of blood flow.
POSSIBLE CIRCULATION IMPAIRMENT
Prolonged, continuous use of a blood pressure cuff may impair blood flow to the extremity. Check circulation regularly and loosen or reposition the cuff if changes in circulation occur.
INACCURATE READINGS HAZARDS
POSSIBLE INACCURATE BLOOD PRESSURE READINGS
Do not alter the NIBP monitor’s pneumatic tubing. Altering NIBP tubing may cause improper performance and may void the warranty. Avoid compression or restriction of pressure tubes.
POSSIBLE INACCURATE BLOOD PRESSURE READINGS
Using NIBP accessories not recommended by Physio-Control may cause the device to perform improperly and invalidate the safety agency certifications. Use only the accessories that are specified in these operating instructions.
POSSIBLE INACCURATE OXYGEN SATURATION READINGS
Do not perform NIBP measurement on an extremity used for oxygen saturation monitoring. Oxygen saturation measurement is affected by blood pressure measurement due to the occlusion of blood flow.
CAUTION
EQUIPMENT DAMAGE
Do not inflate a cuff unless it is placed on an extremity.

How NIBP Monitoring Works

The NIBP monitor uses the oscillometric measurement technique. The oscillometric technique does not use Korotkoff sounds to determine blood pressure; rather, it monitors the changes in pressure pulses that are caused by the flow of blood through the artery. The NIBP monitor inflates the cuff around the patient’s arm to a value that occludes the artery, and then deflates the cuff in steps. When blood starts to flow through the artery, the increasing blood flow causes the amplitude of the pressure pulses in the cuff to increase. As the NIBP monitor steps the pressure down, the pulses
4-38
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
reach a peak amplitude and then start to decrease. The rising and falling amplitude values form a curve that is analyzed to yield systolic pressure, diastolic pressure, and mean arterial pressure (MAP).
The NIBP monitor measures the pulse rate by tracking the number of pulses over time. The NIBP monitor uses artifact rejection techniques to provide accurate results under most operating conditions. When a patient is experiencing arrhythmias during a measurement, the accuracy of the pulse determination may be affected or the time needed to complete a measurement may be extended. In shock conditions, the low amplitude of blood pressure waveforms makes it difficult for the monitor to accurately determine the systolic and diastolic pressures.

NIBP Monitoring Considerations

As with any noninvasive oscillometric blood pressure monitor, clinical conditions can affect the accuracy of the measurements obtained, including the following:
• The patient’s physiological condition. For example, shock may result in a blood pressure waveform that has a low amplitude, making it difficult for the monitor to accurately determine the systolic and diastolic pressures.
4
• The position of the patient.
• Motion may prolong the measurement process since motion artifacts have to be rejected in the data stream. Motion that affects measurement can include patient movement, patient seizure, bumping the cuff, and flexing the extremity under the cuff.
• The presence of other medical devices. The NIBP monitor does not operate effectively if the patient is connected to a heart/lung machine.
• When a patient is experiencing arrhythmias, pulse rate accuracy may be affected or the time needed to complete an NIBP measurement may be extended. The device automatically deflates if a blood pressure measurement cannot be obtained in 120 seconds.
• Blood pressure and pulse can fluctuate greatly between measurements; the monitor cannot alert the operator of changes in vital signs that occur between measurement cycles.
• There may be some difference between readings taken manually and readings from the NIBP monitor due to the differing sensitivity of the two methods. The NIBP monitor meets the ANSI/ SP10 AAMI standard that requires a mean difference of ±5 mmHg, with a standard deviation no greater than 8 mmHg, compared to auscultatory readings.
• When using the NIBP monitor during defibrillation, the NIBP monitor is not available when the defibrillator is being charged. Upon shock, the monitor resets and dashes (– – –) appear in place of pressure readings. After defibrillation, you can resume blood pressure measurement according to "NIBP Monitoring Procedure" on page 4-40.
• If the blood pressure cuff fails to deflate for any reason or causes undue discomfort to the patient, remove the cuff from the arm or disconnect the tubing from the defibrillator.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-39
Monitoring Noninvasive Blood Pressure
1. Rotate the SPEED DIAL to outline the NIBP area.
2. Press the
SPEED DIAL. The NIBP menu
appears.
3. Select
INITIAL PRESSURE.
4. Rotate the
SPEED DIAL to the desired
pressure.
5. Press the
SPEED DIAL to set the initial
pressure.
NIBP
Start Interval
Initial Pressure
Off
160 mmHg

Cuff Selection

The use of properly designed and sized cuffs is essential for the accurate measurement of blood pressure. The cuff must fit snugly around the extremity to occlude the artery. For a list of BP cuffs that are intended for use with the LIFEPAK 15 monitor/defibrillator, see the LIFEPAK 15 Monitor/ Defibrillator Accessories Catalog at store.physio-control.com.

NIBP Monitoring Procedure

The NIBP monitor inflates an occluding cuff and determines systolic and diastolic pressures, mean arterial pressure (MAP), and pulse rate. Pressure measurements are reported in mmHg and pulse rate in beats per minute (bpm).
Both single-measurement and specified-interval (timer-controlled) methods of blood pressure reading are available.
The NIBP monitor draws power from the defibrillator. When the defibrillator is turned on, the NIBP monitor conducts a self-test that takes approximately three seconds.
IMPORTANT! The LIFEPAK 15 monitor NIBP port and tubing are not compatible or
interchangeable with the NIBP tubing that is used with other LIFEPAK monitor/defibrillators.
4-40
Changing the Initial Inflation Pressure
The initial cuff pressure should be set approximately 30 mmHg higher than the patient’s anticipated systolic pressure. The factory default initial inflation pressure for the first measurement is 160 mmHg. For pediatric patients, the initial cuff pressure may need to be lowered. Initial inflation settings are 80, 100, 120, 140, 160, or 180 mmHg.
Caution should be taken not to lower the initial pressure below the adult patient’s systolic measurement. Doing so may cause the cuff to reinflate and cause patient discomfort. For subsequent measurements, the monitor inflates approximately 30 mmHg higher than the previously determined systolic pressure.
To select an initial pressure:
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
Countdown timer—displays time until next measurement
Systolic pressure
Diastolic pressure
Mean arterial pressure (MAP)
Note: Measurement data is recorded in the LIFEPAK 15 monitor/defibrillator Vital Sign Log. For
more information about the Vital Sign Log and its use, see Chapter 7, "Data Management."
Manual Single-Measurement Procedure
The NIBP measurement typically takes 40 seconds to complete. If the measurement is not completed within 120 seconds, the cuff automatically deflates.
To obtain a manual single measurement:
4
1. Press
2. Select the appropriately-sized cuff.
3. Properly align the cuff artery markings, if present, and apply snugly to the extremity.
4. Connect the tubing to the cuff and to the NIBP port on the monitor.
5. Change the initial inflation pressure, if necessary.
6. Position the extremity in a relaxed and supported position at approximately the same level as
7. Press
To cancel a measurement, press
Note: NIBP pulse rate is displayed only when ECG or SpO
ON.
the patient’s heart. Inform the patient that the cuff will inflate and cause a “big squeeze” around the arm and that the patient’s fingers may tingle.
NIBP to start the measurement, and check that the patient’s arm is not moving. When
the measurement is complete, systolic, diastolic, and mean arterial pressures are displayed.
NIBP again.
is not active.
2
Timer-Controlled Measurement Procedure
When the timer is set, the monitor performs recurring measurements at a fixed interval. When using timer-controlled measurement, the interval is counted from the start of the measurement to the start of the next measurement. Choices are minutes.
OFF (factory default), 2, 3, 5, 10, 15, 30, and 60
To take a manual measurement between timer-controlled measurements, press interval is counted from the beginning of the manual measurement.
Figure 4-12 NIBP Measurements and Timer
©2007-2013 Physio-Control, Inc.
NIBP. The next
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-41
Monitoring Noninvasive Blood Pressure
To set timer-controlled measurements:
1. Press
2. Select the appropriately-sized cuff.
3. Properly align the cuff artery markings, if present, and apply snugly to the extremity.
4. Connect the tubing to the cuff and to the NIBP port on the monitor.
5. Rotate the
6. Press the
7. Select
8. Position the extremity in a relaxed and supported position at approximately the same level as
9. Press
To cancel a measurement in progress, press
ON.
SPEED DIAL to outline the NIBP area.
SPEED DIAL. The NIBP menu appears.
INTERVAL and then select the desired time interval.
the patient’s heart. Inform the patient that the cuff will inflate and cause a “big squeeze” around the arm and that the patient’s fingers may tingle.
NIBP to start the measurement, and check that the patient’s arm is not moving. When
the measurement is complete, systolic, diastolic, and mean arterial pressures are displayed. The countdown timer shows the time to the next automatic NIBP measurement.
NIBP again.
Note: If at any time the cuff pressure exceeds 290 mmHg or there is a system failure of the
NIBP module, timer-controlled NIBP is terminated. To reactivate, follow the Timer-Controlled Measurement Procedure.

Cleaning

To clean the cuff and pneumatic tubing:
1. Disconnect the tubing from the cuff and monitor. Use a clean, soft cloth dampened with a germicidal solution to wipe clean.
2. Inspect the tubing for cracks or kinks. If any damage is noted, replace the tubing.
3. Inspect the cuff for damage or excessive wear. If any damage is noted, replace the cuff.
4. Allow both to dry before placing the cuff on a patient or reconnecting the tubing to the monitor.
For information about cleaning the device, see "Cleaning the Device" on page 10-15.
4-42
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING

Troubleshooting Tips

Table 4-5 Troubleshooting Tips for NIBP Monitoring
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
4
NIBP AIR LEAK message appears Cuff applied too loosely
Leak in cuff/monitor pneumatic system
NIBP FLOW ERROR message
appears
NIBP FAILED message appears The monitor cannot establish
NIBP INITIALIZING message
appears
NIBP MOTION message appears The patient extremity moved
The pneumatic system is not maintaining stable cuff pressure
zero-pressure reference
NIBP requested while NIBP module is still initializing
too much for the monitor to accurately complete the measurement
• Check cuff for snug fit on patient.
• Check that the cuff/monitor connection is secure.
• Check cuff for leaks. Do not use a cuff that exhibits a leak.
• Deflate or remove cuff.
• Check tubing for leaks.
• Replace cuff.
• Check tubing for kink or blockage.
• If this message persists, remove monitor from use and obtain service. Use another method to measure the patient’s blood pressure.
• Wait until message disappears and request NIBP.
• Have patient lie quietly with extremity relaxed and supported.
• Check that patient’s arm does not move during NIBP measurement.
NIBP OVERPRESSURE message
appears
©2007-2013 Physio-Control, Inc.
Cuff pressure exceeded 290 mmHg
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
• Disconnect tubing or remove cuff.
• Avoid very rapid squeezing of the cuff.
• If this message persists, remove the cuff from use and obtain service.
4-43
Monitoring Noninvasive Blood Pressure
Table 4-5 Troubleshooting Tips for NIBP Monitoring (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
NIBP TIME OUT message appears The monitor did not complete a
measurement in 120 seconds
NIBP WEAK PULSE message
appears
XXX appears in place of
NIBP readings
NIBP CHECK CUFF message
appears
The monitor did not detect any pulses
NIBP module failed. NIBP module failed to calibrate successfully.
Cuff is not connected to patient or device
• Check cuff for snug fit on patient.
• Check that cuff artery markings are aligned with the artery.
• Repeat measurement.
• Try a higher initial pressure.
• If this message persists, use another method to measure the patient’s blood pressure.
• Check pulses distal to the cuff.
• Check cuff for snug fit on patient.
• Check that cuff artery markings are aligned with the artery.
• Turn device off and then on again.
• If problem persists, contact qualified service personnel.
• Check cuff for snug fit on patient.
• Check that cuff artery markings are aligned with the artery.
• Check cuff tubing connection to device.
Unable to connect NIBP tubing to device
Cuff not deflating Internal valves fail to open • Disconnect NIBP tubing.
Cuff not inflating Cuff is not connected to the
The LIFEPAK 12 NIBP tubing connector is not compatible with the LIFEPAK 15 NIBP port
device
Leak in tubing, cuff, or connector
For general troubleshooting tips, see Table 10-2 on page 10-18.
4-44
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
• Obtain correct NIBP tubing that is compatible with LIFEPAK 15 monitor/ defibrillator.
• Remove cuff from patient.
• Check tubing connection to device and cuff.
• Replace NIBP tubing or cuff.
MONITORING

Monitoring ETCO2

Intended Use

The end-tidal CO2 (EtCO2) monitor is a capnometric device that uses non-dispersive infrared spectroscopy to continuously measure the amount of CO present at the end of exhalation (EtCO can be used with intubated or nonintubated patients. Respiration rate is also measured and displayed in breaths per minute.
). The sample is obtained by the side stream method and
2
during each breath and report the amount
2
4
The EtCO assess the patient at all times; do not rely solely on the EtCO
monitor is a tool to be used in addition to patient assessment. Care should be taken to
2
monitor.
2
Indications
EtCO2 monitoring is used to detect trends in the level of expired CO2. It is used for monitoring breathing efficacy and treatment effectiveness in acute cardiopulmonary care, for example, to determine if adequate compressions are being performed during CPR or to rapidly detect whether an endotracheal tube has been placed successfully.
Contraindications
None known.

EtCO2 Monitoring Warnings

WARNINGS
FIRE HAZARDS
FIRE HAZARD
Before use, carefully read these operating instructions, the FilterLine directions for use, and precautionary information.
®
tubing
FIRE HAZARD
The FilterLine tubing may ignite in the presence of O electrosurgical devices, or high heat. Use with caution to prevent flammability of the FilterLine tubing.
FIRE HAZARD
Flammable anesthetics become mixed with the patient’s air that is sampled by the capnometer. When using the EtCO as nitrous oxide or certain other anesthetics, connect the EtCO scavenger system.
©2007-2013 Physio-Control, Inc.
when directly exposed to laser,
2
monitor in the presence of flammable gases, such
2
gas port to a
2
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-45
Monitoring ETCO2
WARNINGS (CONTINUED)
INACCURATE READINGS HAZARDS
POSSIBLE INACCURATE PATIENT ASSESSMENT
The EtCO be used as a diagnostic apnea monitor. An apnea message appears if a valid breath has not been detected for 30 seconds and indicates the time elapsed since the last valid breath. It must be used in conjunction with clinical signs and symptoms.
POSSIBLE INACCURATE CO2 READINGS
Using other manufacturers’ CO improperly and invalidate the safety agency certifications. Use only the accessories that are specified in these operating instructions.
HEALTH HAZARDS
POSSIBLE STRANGULATION
Carefully route the patient tubing (FilterLine) to reduce the possibility of patient entanglement or strangulation.
monitor is intended only as an adjunct in patient assessment and is not to
2
accessories may cause the device to perform
2
INFECTION HAZARD
Do not reuse, sterilize, or clean Microstream
®
CO2 accessories as they are designed for
single-patient one-time use.

How Capnography Works

An EtCO2 sensor continuously monitors carbon dioxide (CO2) that is inspired and exhaled by the patient. The sensor employs Microstream non-dispersive infrared (IR) spectroscopy to measure the concentration of CO
The CO
FilterLine system delivers a sample of the exhaled gases directly from the patient into the
2
LIFEPAK 15 monitor for CO and secretion accumulation and prevents obstruction, which maintains the shape of the CO waveform.
The CO
sensor captures a micro sample (15 microliters). This extremely small volume allows for
2
fast rise time and accurate CO
The Microbeam IR source illuminates the sample cell and the reference cell. This proprietary IR light source generates only the specific wavelengths characteristic of the CO Therefore, no compensations are required when concentrations of O vapor are present in the exhaled breath.
molecules that absorb infrared light.
2
measurement. The low sampling flow rate (50 ml/min) reduces liquid
2
readings, even at high respiration rates.
2
2
absorption spectrum.
2
, anesthetic agent, or water
2
4-46
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
IV
V
II
III
I
You can set up the LIFEPAK 15 monitor/defibrillator to use the capnography Body Temperature Pressure Saturated (BTPS) conversion method. This option corrects for the difference in temperature and moisture between the sampling site and alveoli. The correction formula is 0.97 × the measured EtCO your device.

EtCO2 Monitoring Waveform Analysis

Valuable information concerning the patient’s expired CO2 can be acquired by examination and interpretation of the waveform.
The Phases of the Waveform
Figure 4-13 is a graphic representation of a normal capnograph waveform. Four phases of the waveform require analysis. The flat I–II baseline segment (Respiratory Baseline) represents continued inhalation of CO Upstroke), a sharp rise, represents exhalation of a mixture of dead space gases and alveolar gases from acini with the shortest transit times. Phase III–IV (Expiratory Plateau) represents the alveolar plateau, characterized by exhalation of mostly alveolar gas. Point IV is the end-tidal (EtCO
) value that is recorded and displayed by the monitor. Phase IV–V (Inspiratory
2
Downstroke), a sharp fall, reflects the inhalation of gases that are CO normal capnograph or EtCO ventilation, or equipment function.
value. See the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with
2
-free gas. This value normally is zero. The II–III segment (Expiratory
2
-free. Alterations of the
2
values are the result of changes in metabolism, circulation,
2
4
Figure 4-13 Phases of the Respiratory Waveform
Respiratory Baseline Elevation of the waveform baseline (I–II segment) usually represents
rebreathing CO Rebreathing CO hypoventilation. Precipitous rises in both baseline and EtCO contamination of the sensor.
Expiratory Upstroke In the normal waveform, the rising phase (II–III segment) is usually steep.
When this segment becomes less steep, CO site. The causes of this delay can be physiologic or mechanical and include bronchospasm, obstruction of the upper airway, or obstruction (or kinking) of an endotracheal tube (ETT).
Expiratory Plateau The plateau of the waveform, which represents the remainder of expiration
(III-IV segment), should be nearly horizontal. The end of the plateau represents the EtCO Upward slanting of the expiratory plateau occurs when there is uneven emptying of the alveoli.
©2007-2013 Physio-Control, Inc.
. This elevation usually is accompanied by gradual increases in the EtCO2 value.
2
is common in circumstances of artificially produced increased dead space and
2
values usually indicate
2
delivery is delayed from the lungs to the sampling
2
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
value.
2
4-47
Monitoring ETCO2
Similar to the diminished slope of the Expiratory Upstroke, this pattern can occur in asthma, chronic obstructive pulmonary disease (COPD), partial upper-airway obstruction, or partial mechanical obstruction such as a partially kinked ETT.
Inspiratory Downstroke The fall to baseline (IV-V segment) is a nearly vertical drop. This slope
can be prolonged and can blend with the expiratory plateau in cases of leakage in the exhale portion of the breathing circuit. The peak EtCO numeric end-tidal value without observing the breathing waveform may obscure the presence of a leak.

EtCO2 Monitoring Procedure

When activated, the EtCO2 monitor draws power from the defibrillator. The LIFEPAK 15 monitor/ defibrillator activates the EtCO Initialization, self-test, and warm up of the EtCO may take up to two-and-one-half minutes.
CAUTION
monitor when it senses the attachment of the FilterLine set.
2
value (IV) is often not reached. Relying on the
2
monitor is typically less than 30 seconds, but
2
POSSIBLE EQUIPMENT DAMAGE
Failure to replace a broken or missing CO2 port door may allow water or particulate contamination of the internal CO
sensor. This may cause the CO2 module to
2
malfunction.
To monitor EtCO
1. Press
ON.
2. Select the appropriate EtCO
3. Open the CO
4. Verify that the CO
:
2
accessory for the patient.
2
port door and insert the FilterLine connector; turn connector clockwise until tight.
2
area is displayed. The EtCO2 monitor performs the autozero routine as part
2
of the initialization self-test.
Note: If you use a ventilation system, do not connect the FilterLine set to the patient/ventilation
system until the EtCO
5. Display CO
6. Connect the CO
waveform in Channel 2 or 3.
2
2
7. Confirm that the EtCO scale for the best visualization of the waveform. You can change the scale, if desired, as described in the next section.
monitor has completed its self-test and warm-up.
2
FilterLine set to the patient.
value and waveform are displayed. The monitor automatically selects the
2
Note: It is possible for the FilterLine set to become loose at the device connection and still have an
EtCO
value and CO2 waveform, but they may be erroneously low. Make sure the FilterLine
2
connection is firmly seated and tight.
4-48
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
Note: The capnography module performs self-maintenance within the first hour of monitoring and
once an hour during continuous monitoring. The self-maintenance includes “auto-zeroing.” Self­maintenance is also initiated when the surrounding temperature changes 8°C (14.4°F) or more, or the surrounding pressure changes greater than 20 mmHg. The CO attempts to purge the tubing. To clear the messages, remove the FilterLine tubing and reconnect it to the monitor.
CO2 FILTERLINE PURGING or CO2 FILTERLINE BLOCKAGE

CO2 Display

The following scales are available to display the CO2 waveform. The LIFEPAK 15 monitor/ defibrillator automatically selects the scale based on the measured EtCO scale, outline and select the CO the scale menu.
• Autoscale (default)
• 0–20 mmHg (0–4 Vol% or kPa)
• 0–50 mmHg (0–7 Vol% or kPa)
• 0–100 mmHg (0–14 Vol% or kPa)
area using the SPEED DIAL and then select the desired scale from
2
module detects this change and
2
value. To change the CO2
2
4
The CO the 4-second screen. There is a slight delay between when the breath occurs and when it appears on the screen. Printouts are at 25 mm/sec. Continuous print may be changed to 12.5 mm/sec, if desired.
The monitor shows the maximum CO increasing, the change can be seen with every breath. However, if the values are continually decreasing, it will take up to 20 seconds for a lower numerical value to be displayed. Because of this, the EtCO
waveform is compressed (displayed at 12.5 mm/sec sweep speed) to provide more data in
2
value over the last 20 seconds. If the EtCO2 values are
2
value may not always match the level of the CO2 waveform.
2

CO2 Alarms

The EtCO2 monitor provides:
•EtCO
•FiCO
• Apnea alarm (automatic and not adjustable)
Note: The apnea alarm occurs when a breath has not been detected for 30 seconds. The message
ALARM APNEA appears in the message area along with the time since the last detected breath.
high and low alarms controlled by activating ALARMS (see "Alarms" on page 3-21)
2
(inspired CO2) alarm (automatic and not adjustable)
2
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-49
Monitoring ETCO2

CO2 Detection

A CO2 waveform appears when any CO2 is detected, but CO2 must be greater than 3.5 mmHg for a numerical value to be displayed. However, the CO is at least 8 mmHg (1.0% or kPa). Valid breaths must be detected in order for the apnea alarm to function and to count the respiratory rate (RR). The RR represents an average over the last eight breaths.
module will not recognize a breath until the CO2
2
When CO dashes “---” or a flat solid line at or near zero—several factors must be quickly evaluated. Assess for the following causes:
Equipment issues
• Disconnection of the FilterLine set from the endotracheal tube (ETT)
• System is purging due to fluid in the patient/sensor connection from ET administration of
• System is auto-zeroing
• Shock was delivered and system is resetting
• Loose FilterLine set to device connection
Loss of airway function
• Improper placement of ETT
• ETT dislodgment
• ETT obstruction
Physiological factors
• Apnea • Loss of perfusion • Exsanguination
• Massive pulmonary embolism • Inadequate CPR
is not detected in the cardiac arrest situation—for example, the CO2 waveform is either
2
medications

Cleaning

Accessories for CO2 monitoring are disposable and are intended for single-patient use. Do not clean and reuse a FilterLine set. Dispose of the contaminated waste according to local protocols.
For information about cleaning the device, see "Cleaning the Device" on page 10-15.
4-50
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING

Troubleshooting Tips

Table 4-6 Troubleshooting Tips for EtCO2 Monitoring
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
4
ALARM APNEA message appears
and waveform is solid line at or near zero
CO2 FILTERLINE OFF message
appears and waveform is “---”
CO2 FILTERLINE PURGING
message appears and waveform is “---”
CO2 FILTERLINE BLOCKAGE
message appears and waveform is “---”
No breath has been detected for 30 seconds since last valid breath
FilterLine connection to device is loose
FilterLine set is disconnected from patient or ETT
FilterLine set disconnected or not securely connected to device
FilterLine set is kinked or clogged with fluid, or rapid altitude change occurred
The message appears after 30 seconds of unsuccessful purging
• Check the patient.
• Twist FilterLine connector clockwise until tight and firmly seated.
• Check ventilation equipment (if used) for leaks or disconnected tubing.
• Connect FilterLine set to device port.
• Twist FilterLine connector clockwise until tight and firmly seated.
• Disconnect and then reconnect the FilterLine set.
• Twist FilterLine connector clockwise until tight and firmly seated.
• Disconnect and then reconnect the FilterLine set.
• Change the FilterLine set.
CO2 INITIALIZING message
appears and waveform is “---”
AUTO ZEROING message appears
and waveform is “---”
©2007-2013 Physio-Control, Inc.
FilterLine set is kinked or clogged
FilterLine set connected to device while module is initializing
Defibrillation shock delivered • None. System resets
Module is performing self­maintenance
Defibrillation shock delivered • None. System resets
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
• Twist FilterLine connector clockwise until tight and firmly seated.
• None.
automatically within 20 seconds.
• None.
automatically within 20 seconds.
4-51
Monitoring ETCO2
Table 4-6 Troubleshooting Tips for EtCO
Monitoring (Continued)
2
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
EtCO2 values are erratic FilterLine connection to device
is loose
• Twist FilterLine connector clockwise until tight and firmly seated.
A leak in the FilterLine set • Check for connection leaks
and line leaks to patient, and correct, if necessary.
A mechanically ventilated
• No action required.
patient breathes spontaneously or patient is talking
values are consistently
EtCO
2
higher than expected
Physiological cause such as COPD
• None.
Inadequate ventilation • Check ventilator, increase
ventilatory rate/bagging.
Patient splinting during breathing
• Supporting measures such as pain relief.
Improper calibration • Contact qualified service
personnel.
values are consistently
EtCO
2
lower than expected
FilterLine connection to device is loose
• Twist FilterLine connector clockwise until tight and firmly seated.
waveform stays elevated
CO
2
for several seconds
Sudden extreme increase in EtCO
2
XXX appears instead of EtCO
value
There is no EtCO CO
waveform is flat
2
value and the
2
Physiological cause • See Physiological factors in
"CO2 Detection" on page 4-50.
Hyperventilation • Check ventilator, decrease
ventilatory rate/bagging.
Improper calibration • Contact qualified service
personnel.
Expiration is prolonged due to bagging technique
• Release bag reservoir completely with expiration. Observe that elevated baseline returns to normal level.
Fluid has entered CO2 module • Contact qualified service
personnel.
CO2 module malfunction • Turn device off and then on
2
again.
• If problem persists, contact qualified service personnel.
Measured CO
3.5 mmHg
is less than
2
• See "CO2 Detection" on page 4-50.
4-52
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
Note: To decrease the likelihood of the FilterLine connection coming loose during use, hand-
straighten the tubing after removal from the package before connecting to patient or device.
For general troubleshooting tips, see Table 10-2 on page 10-18.
4
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-53

Monitoring Invasive Pressure

Monitoring Invasive Pressure

Intended Use

The LIFEPAK 15 invasive pressure (IP) monitor is intended for measuring arterial, venous, intracranial, and other physiological pressures using an invasive catheter system with a compatible transducer.
The IP monitor is a tool to be used in addition to patient assessment. Care should be taken to assess the patient at all times; do not rely solely on the IP monitor.
Indications
Invasive pressure monitoring is indicated for use in patients who require continuous monitoring of physiological pressures in order to rapidly assess changes in the patient’s condition or response to therapy. It may also be used to aid in medical diagnosis.
Contraindications
None known.

IP Monitoring Warnings

WARNINGS
INACCURATE READINGS HAZARDS
POSSIBLE INACCURATE PRESSURE READINGS, AIR EMBOLISM, BLOOD LOSS, OR LOSS OF STERILITY
Before use, carefully read these operating instructions, and the transducer and infusion set instructions for use and precautionary information.
INACCURATE PRESSURE READINGS
Pressure readings should correlate with the patient’s clinical presentation. If readings do not correlate, verify that the zeroing stopcock is positioned at the patient’s zero reference, rezero the transducer, and/or check the transducer with a known or calibrated pressure. Manually check cuff blood pressure.
INACCURATE PRESSURE READINGS
Changing the patient’s position changes the zero reference level. Relevel the transducer’s zeroing stopcock any time the patient’s position is changed.
4-54
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MONITORING
ART systolic pressure
ART diastolic pressure
CVP mean pressure
ART mean pressure
WARNINGS (CONTINUED)
HEALTH HAZARDS
POSSIBLE LETHAL ARRHYTHMIA
Ventricular fibrillation may be induced if the isoelectric barrier of the transducer is disrupted. The isoelectric barrier within the transducer may be disrupted if the transducer body is damaged. Do not use a transducer that is visibly damaged or leaking fluid.
INCREASED INTRACRANIAL PRESSURE
Do not use a continuous flush device with transducers used for intracranial monitoring.

IP Monitoring

Two channels are available for invasive pressure monitoring, with default labels P1 and P2 and the user-selectable labels shown in Table 4-7.
4
Table 4-7 IP Labels and Descriptions
LABEL DESCRIPTION
ART Arterial Pressure
PA Pulmonary Artery Pressure
CVP Central Venous Pressure
ICP Intracranial Pressure
LAP Left Atrial Pressure
When the default labels P1 and P2 are used, the IP monitoring area displays systolic, diastolic, and mean pressures. When ICP, LAP, or CVP labels are used, the IP monitoring area displays mean pressure in large type. Systolic and diastolic pressures are not displayed.
Figure 4-14 IP Labels
Because pressures can change in a short time, data should be checked regularly during vital sign monitoring.
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-55
Monitoring Invasive Pressure

How IP Monitoring Works

IP monitoring involves the conversion of fluid pressure into an electrical signal. The conversion is accomplished with a pressure transducer. The transducer is connected to a patient’s indwelling pressure catheter using a special assembly of tubing, stopcocks, adapters, flush valves, and fluids, commonly known as a flush system. The transducer translates the pressure wave into an electrical signal. A well-functioning flush system is essential for obtaining undistorted waveforms and accurate information.
IP monitoring is available on either Channel 2 or 3. The IP connector (6-pin type 3102A-14S-6S) is compatible with industry standard (60601-2-34 and AAMI-BP22) pressure transducers with 5μV/V/ mmHg sensitivity. The customer is responsible for determining whether transducers comply with standards and are compatible with the monitor.
The IP connector pinout has the following configuration, counterclockwise from 12 o’clock, viewed from the front of the LIFEPAK 15 monitor/defibrillator.
A pin = - signal B pin = + excitation C pin = + signal
D pin = - excitation E pin = shield F pin = unlabeled
An invasive pressure adapter cable is used to connect the transducer to the monitor.

IP Monitoring Procedure

Prepare a flush system according to local protocols. Position the transducer at the patient’s phlebostatic axis (zero-reference level).
To avoid offset errors, a zero reference must be established before any meaningful pressure readings are obtained. This is done by opening the transducer stopcock to air so that atmospheric pressure becomes the reference.
The P1 or P2 connector and Channel 2 or 3 can be used for IP monitoring. P1 and Channel 2 are used in these instructions.
To monitor IP:
1. Prepare the transducer system according to the operating instructions provided with the transducer and your local protocol.
2. Press
3. Connect the IP cable to the transducer and to the P1 port on the monitor.
4. Use the default label
ON.
P1 or select ART, PA, CVP, ICP, or LAP. To change the label, select the P1
area. From the menu, select
P1. Select a label from the list.
4-56
5. Use the menu, select
SPEED DIAL to outline and select CHANNEL 2 on the Home Screen. From the Channel 2
WAVEFORM and then select the label that is desired for the waveform.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Loading...