This chapter provides information on transcutaneous monitoring of
oxygen (tcpO
information on preparing the transducer for use, and applying the
transducer to the patient. This chapter includes the following sections:
•Introduction to the tcpO2/tcpCO2 Parameter Module . . . . . . . . 24-2
) and of carbon dioxide (tcpCO2). It also provides
2
tcpO2/tcpCO2 Module Section 24-1
Introduction to the tcpO2/tcpCO2 Parameter Module
Introduction to the tcpO2/tcpCO2 Parameter Module
What does it
Measure?
How the
tcpO2/
tcpCO2
Measureme
nt Works
Correlation
of
Transcut ane
ous with
Arterial
Blood Gas
Values
Section
tcpO2/tcpCO2 Module
The tcpO2/tcpCO2 module measures the partial pressure of the oxygen
and carbon dioxide that diffuses transcutaneously (through the skin).
The oxygen measurement is valid for an infant patient who is not under
gas anesthesia. These partial pressures provide a measure of the oxygen
and carbon dioxide levels in the blood.
The tcpO2/tcpCO2 transducer heats the skin. This has the effect of
increasing local blood perfusion so that oxygen and carbon dioxide can
diffuse to the skin surface more easily.
The transducer contains an electrolyte solution which is held in place by
two gas-permeable membranes. The oxygen and carbon dioxide that
diffuses out of the skin passes through these membranes into the
electrolyte solution where electrochemical reactions take place. These
reactions generate electrical signals which are representative of the
amounts of oxygen and carbon dioxide present. The electrical signals
are measured by the module to enable the tcpO
displayed as numerics on the standard display.
Transcutaneous measurements are not capable of replacing arterial
blood gas monitoring. However, transcutaneous monitoring can be used
to reduce the frequency of arterial sampling.
Transcutaneous values are representative of the carbon dioxide and
oxygen levels in the blood. The values at tissue level will not be the
same as those measured arterially because of the transcutaneous nature
of the measurement. However, provided that the transducer is properly
handled by following the procedures described in this chapter, the
transcutaneous values will correlate with (track closely) the arterial
values. For example, a drop in transcutaneous values will usually
indicate that arterial values have dropped as well.
/tcpCO2 values to be
2
Note that transcutaneous values will not always correlate with blood
samples taken from the capillary blood of the heel (heelsticks or
astrups).
24-2 tcpO2/tcpCO2 Module Section
Introduction to the tcpO2/tcpCO2 Parameter Module
For information on “Correlation of transcutaneous with arterial blood gas
values”, see references 5, 6 and 7 in Appendix on Analog Output)
Setting up
the Module
Transducer
Temperature
The optimum application period for the transducer at one site on a patient
is dependent on the transducer temperature and on the sensitivity of the
patient's skin. Before calibrating the transducer, (see the section
Calibrating the tcpO2/tcpCO2 Transducer in this chapter) you should
adjust the transducer temperature and also set the Site Timer.
WarningWarning
The temperature should be selected according to the patient’s age,
weight and physical condition.
Available temperatures for the transducer are 37.0°C, 41.0°C, 41.5°C,
42.0°C, 42.5°C, 43.0°C, 43.5°C, 44.0°C, 44.5°C and 45.0°C. Usually, a higher
transducer temperature results in both a better correlation and also a
shorter time delay between a change in arterial blood gas and its
detection by the transducer. However higher temperatures also increase
the risk of skin burns. A temperature between 42° and 44°C is preferred
by most physicians.
Note—The tcpO
/ tcpCO2 module is equipped with a temperature limiter
2
which prevents the sensor temperature from exceeding 46ºC. If the
temperature limiter fails, the following INOP message is displayed on the
monitor:
•tc EQUIP MALF.
Site TimerThe Site Timer helps you to guard against the risk of skin burn by
ensuring that the transducer is used at one site for no longer than a
predefined period. The transducer can be set to operate for any of the
following time periods: 0.5, 1, 1.5, 2, 2.5, 3, 4, 5, 6, 7 or 8 hours. The
selection is visible in the task window next to “Site Time##, and the
time remaining before the Site Timer expires appears next to “Time Left:”.
Site time settings should be adapted to the patient's skin sensitivity.
tcpO2/tcpCO2 Module Section 24-3
Section
tcpO2/tcpCO2 Module
Introduction to the tcpO2/tcpCO2 Parameter Module
After the selected period of time has expired, an INOP message
“tcCHANGE SITE” is displayed and the INOP tone is sounded. The
Monitor can be configured either to switch off the transducer heating
when the Site Time has expired, or to continue monitoring. This should
be set as required by your hospital's policy, either by your biomedical
engineering department or by the Philips Service Engineer.
Correction for
tcpCO
Section
Values
2
If the choice
Off is available along with the above site times, you can
choose to disable the Site Timer so that no “CHANGE SITE” reminder is
displayed and the transducer monitors (and heats) indefinitely.
Availability of this choice will depend on your hospital's policy and can
be set by your biomedical engineering department or the Philips Service
Engineer.
Transcutaneous pCO
values tend to be higher than arterial values (due
2
to the metabolic processes of the skin and the effect of heating on the
blood under the transducer). This effect can be compensated for by
selecting the Severinghaus correction in a special Configuration Mode
on the monitor. (This has been selected if “corr.” appears under the
value for tcpCO
in the Task Window - consult your biomedical
2
engineering department or the Philips Service Engineer if you wish to
add or remove this correction.)
•The front of the module has two keys:
– The key for parameter setup. A light will appear
O2/CO2
above the key when you are in setup.
tcpO2/tcpCO2 Module
24-4 tcpO2/tcpCO2 Module Section
Introduction to the tcpO2/tcpCO2 Parameter Module
– The CAL key to start calibration procedures.
tcpO2 /tcpCO
M1018A
O2 /CO
2
CAL
2
T
80 x 70
The tcpO
Note—The “T” on the front of the module indicates that this module
transfers parameter settings from one monitor to another.
/tcpCO2 Module
2
Symbols to
Indicate Key
Functions
As detailed in the table below, some versions of this module now use
symbols, instead of words, to indicate the function of some keys. If the
monitor’s Reference Manual tells you, for example, to select the CAL key,
you should press the key marked with the corresponding “calibration”
symbol.
The design change also means that you will now find the module’s
product number on the rear of the module’s housing, not on the front.
Although the new modules do not show the letter “T” on the housing, all
modules retain their capability to transfer parameter settings from one
monitor to another.
tcpO2/tcpCO2 Module Section 24-5
Section
tcpO2/tcpCO2 Module
Introduction to the tcpO2/tcpCO2 Parameter Module
SymbolNameFunctionWhich Modules?
CalcalibrateSvO
Mainstream CO
(M1021A), tcpO2/tcpCO2 (M1018A)
2
(M1016A)
2
Section
tcpO2/tcpCO2 Module
24-6 tcpO2/tcpCO2 Module Section
Activating the Transducer
Follow the 24 hour activation procedure if:
•the electrolyte in your transducer has dried out (this occurs when
the transducer has been out of use for a long period),
•or you are about to use a new transducer (new transducers are
shipped dry).
Activate by following the Remembraning the Transducer procedure, then
leave the transducer in a safe place for 24 hours, unplugged from the
module and with the cap on. Remembrane the transducer again just
before carrying out the calibration procedure.
Activating the Transducer
tcpO2/tcpCO2 Module Section 24-7
Section
tcpO2/tcpCO2 Module
Preparing the transducer
Preparing the transducer
There are three steps to preparing a transducer for application to a
patient.
First, remembrane the transducer.
Then, calibrate the transducer (first checking the module settings
- Site Time, Transducer Temp and Alarm Limits).
Finally, prepare a site and apply the transducer to your patient.
The procedure for carrying out these steps is described in the following
three sections. Each section gives guidelines on when you should carry
out each step. Follow the procedures carefully to ensure best results.
Section
tcpO2/tcpCO2 Module
1. Remembrane the
tcpO2/
tcpCO2
Transducer
With normal use, the membranes will last approximately 1 week before
a change is required. Change the membrane on the transducer if:
•you are about to use a new transducer,
•the transducer is to be used with a new patient,
•the membranes of the transducer are damaged (scratched or
wrinkled),
•the transducer has been used for 5 days continuously,
•the transducer has been in storage for up to 28 days,
•you have a transducer which must be remembraned after
activating for 24 hours (see previous page, Activating the Transducer), or
•calibration has failed twice.
1. Check that you have the Philips tc Accessory Kit containing an Oring remover, absorbent paper, electrolyte solution and membrane
replacers.
24-8 tcpO2/tcpCO2 Module Section
Loading...
+ 17 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.