Who this Book is For9
Confirm Fetal Life Before Using the Monitor10
Introducing the Avalon Family of Fetal Monitors11
2 What's New21
What's New in Release L.321
What's New in Release J.323
3 Basic Operation27
Supported Measurements28
Avalon FM20 and FM3029
Avalon FM40 and FM5030
Getting to Know Your Avalon FM20/FM3031
Getting to Know Your Avalon FM40/FM5034
Connecting the Monitor to AC Mains36
Wired Transducers36
Operating and Navigating38
Operating Modes44
Automatic Screen Layouts45
Settings45
Preparing to Monitor48
After Monitoring70
Switching the Monitor to the Standby Screen71
Disconnecting from Power71
Power On/Power Off Behavior71
Monitoring After a Power Failure72
Troubleshooting72
4 Cableless Monitoring73
Avalon CL Transducer System73
Basics of Cableless Systems74
Twins and Triplets Support75
Configuration of Cableless Systems75
Getting to Know Your Avalon CL76
Cableless Transducers84
CL Pods89
Calling Patients93
Telemetry94
Preparing to Monitor Cablelessly100
3
Battery Lifetime Management106
Patient Transport Within the Hospital107
Underwater Monitoring108
5 FM20/30 Battery Option109
External Power Supply M8023A109
Using Batteries110
Optimizing Battery Performance113
Storing the Battery115
Cableless Monitoring with FM20/30 Battery Option116
Patient Transport Within the Hospital116
6 Alarms117
Alarm Mode118
Nurse Call Systems118
Visual Alarm Indicators119
Audible Alarm Indicators119
Acknowledging Alarms121
Acknowledging Disconnect INOPs121
Alarm Reminder122
Pausing or Switching Off Alarms122
Alarm Limits124
Reviewing Alarms126
Latching Alarms126
Testing Alarms128
Alarm Behavior at Power On128
Admit/Discharge on the Monitor147
New Patient Check149
OB TraceVue/IntelliSpace Perinatal149
9 Non-S tress Test Timer151
Setting NST Autostart/Autostop151
Viewing the NST Timer151
Timer Expiry Notification151
Accessing the NST Setup Pop-up Keys152
10 Non-S tress Test Report153
Setting Up an NST Report153
NST Report Status Window154
NST Criteria157
4
11 Cross-Channel Verification (CCV)159
Misidentification of Heart Rates159
Cross-Channel Verification Functionality160
Overview of Cross-Channel Comparisons161
Coincidence Examples162
Recommended Actions for Coincidence INOP164
12 Monitoring FHR and FMP Using Ultrasound165
Technical Description165
Limitations of the Technology166
Misidentification of Maternal HR as FHR166
What You Need166
Cableless Monitoring - Important Considerations167
Preparing to Monitor168
Selecting Fetal Heart Sound169
Changing the Fetal Heart Sound Volume170
Fetal Movement Profile170
Troubleshooting172
Additional Information173
Important Considerations191
Monitoring Triplets192
Separating FHR Traces192
"Standard" Separation Order193
"Classic" Separation Order194
Switching Trace Separation On and Off195
When Trace Separation is On195
When Trace Separation is Off195
Troubleshooting196
15 Fetal Heart Rate Alarms197
Changing Alarm Settings197
Changing Signal Loss Delay198
16 Monitoring FHR Using DECG199
Misidentification of Maternal HR as FHR199
What You Need200
Making Connections202
What You Need207
External Toco Monitoring208
Toco Sensitivity208
Troubleshooting208
18 Monitoring Uterine Activity Internally211
What You Need211
Internal (IUP) Monitoring213
Troubleshooting214
19 Monitoring aFHR, aHR, and aToco215
Fetal Heart Rate aFHR215
Uterine Activity aToco216
What You Need217
At the Monitor217
Troubleshooting218
20 Monitoring Maternal Heart / Pulse Rate221
Priority for Maternal Heart / Pulse Rate221
Misidentification of Maternal HR for FHR222
Maternal HR from MECG Electrodes222
Monitoring MECG Wave225
Pulse Rate from Toco MP227
Pulse Rate from SpO2228
Adjusting the Heart Rate / Pulse Alarm Limits228
Average Pulse Rate from Noninvasive Blood Pressure228
Testing MECG Mode229
21 Printing the ECG Waveform231
22 Monitoring Noninvasive Blood Pressure235
Introducing the Oscillometric Noninvasive Blood Pressure Measurement235
Preparing to Measure Noninvasive Blood Pressure237
Starting and Stopping Measurements239
Enabling Automatic Mode and Setting Repetition Time239
Enabling Sequence Mode and Setting Up the Sequence240
Choosing the Alarm Source240
Assisting Venous Puncture241
Calibrating NBP241
Troubleshooting242
6
23 Monitoring SpO2243
Selecting an SpO2 Sensor243
Applying the Sensor243
Connecting SpO2 Cables244
Measuring SpO2244
SpO2 Signal Quality Indicator (FAST SpO2 only)245
Assessing a Suspicious SpO2 Reading246
Understanding SpO2 Alarms246
Setting Up Tone Modulation247
Setting the QRS Volume247
24 Monitoring Maternal Temperature249
Measuring Tympanic Temperature249
Entering Temperature Manually254
25 Paper Save Mode for Maternal Measurements255
26 Recovering Data257
Recovering Traces on Paper257
Recovering Traces on an OB TraceVue/IntelliSpace Perinatal System258
Manually Recording Stored Data258
27 Care and Cleaning261
General Points261
Cleaning and Disinfecting262
Cleaning and Disinfecting Monitoring Accessories263
Cleaning and Disinfecting the Tympanic Temperature Accessories264
Cleaning and Disinfecting CL Transducers and CL Pods264
Sterilizing265
28 Maintenance267
Inspecting the Equipment and Accessories267
Inspecting the Cables and Cords268
Maintenance Task and Test Schedule268
Recorder Maintenance269
Cleaning the Print Head273
Returning Equipment for Repair273
Disposing of the Monitor274
29 Accessories and Supplies275
Information on Latex275
Avalon CL Base Station275
Transducers276
Fetal Accessories276
Noninvasive Blood Pressure Accessories278
SpO2 Accessories280
7
Tympanic Temperature Accessories285
Recorder Paper285
Batteries285
30 Specifications and Standards Compliance287
Environmental Specifications287
Physical Specifications289
Interface Specifications293
Performance Specifications295
Recorder Specifications308
External Displays: FM40/FM50 Only310
Manufacturer's Information310
Trademark Acknowledgment310
Regulatory and Standards Compliance311
Environment321
Monitoring After a Loss of Power321
ESU, MRI, and Defibrillation321
Cardiac Pacemakers and Electrical Stimulators322
Fast Transients/Bursts322
Symbols on the System322
31 Default Settings Appendix327
Alarm and Measurement Default Settings327
Recorder Default Settings330
Index333
8
1Introduction
Who this Book is For
This book is for trained healthcare professionals using the Avalon FM20, FM30, FM40, and FM50
fetal/maternal monitors, and the Avalon CL Transducer System. It describes how to set up and use the
monitor and transducers. Familiarize yourself with all instructions including warnings and cautions
before starting to monitor patients. Read and keep the Instructions for Use that come with any
accessories, as these contain important information about application, care, and cleaning that is not
repeated in this book.
If you have received this Instruction for Use because your fetal monitor has been upgraded to a newer
software version L.3, be aware that the standards compliance information contained in the
Instructions for Use for L.3 does not apply to your fetal monitor. Refer to your original Instructions
for Use for standards compliance information.
You should be:
1
•Trained in the use of fetal heart rate (FHR) monitors.
•Trained in the interpretation of FHR traces.
•Familiar with using medical devices and with standard fetal monitoring procedures.
For information on how to configure and service the monitor, see the Configuration Guide and the
Service Guides, or contact your authorized service provider.
Your monitor may not have all of the features and options described in this guide. The exact
appearance of the monitor may differ slightly from that shown in the illustrations.
This guide may contain descriptions of functionality and features that are not implemented in the
equipment currently shipped to Japan and/or of products that are not currently sold in Japan due to
limitations and restrictions under the applicable local laws and regulations in Japan. Please contact your
local sales representative and/or Philips Customer Support for details.
In this guide:
•A warning alerts you to a potential serious outcome, adverse event, or safety hazard. Failure to
observe a warning may result in death or serious injury to the user or patient.
•A caution alerts you to where special care is necessary for the safe and effective use of the
product. Failure to observe a caution may result in:
–minor or moderate personal injury,
–damage to the product or other property,
–possibly in a remote risk of more serious injury.
9
1 Introduction
•Monitor refers to the entire fetal/maternal monitor. Display refers to the physical display unit.
Screen refers to everything you see on the monitor's display, such as measurements, alarms,
patient data, and so forth.
FM30 •Whenever a monitor’s identifier appears to the left of a heading or paragraph, it means that the
information applies to that monitor only. Where the information applies to all models, no
distinction is made.
Avalon CL
Avalon CTS
Whenever one of these identifiers appear to the left of a heading or paragraph, it means that the
information applies to that cableless transducer system. Where the information applies to both
systems, no distinction is made.
For installation instructions and technical description, see the corresponding Service Guide of the fetal
monitors.
Confirm Fetal Life Before Using the Monitor
Fetal monitoring technology available today is not always able to differentiate a fetal heart rate (FHR)
signal source from a maternal heart rate (MHR) source in all situations. Therefore, you should confirm
fetal life by independent means before starting to use the fetal monitor, for example, by palpation of fetal
movement or auscultation of fetal heart sounds using a fetoscope, stethoscope, or Pinard stethoscope.
If you cannot hear the fetal heart sounds, and you cannot confirm fetal movement by palpation,
confirm fetal life using obstetric ultrasonography. Continue to confirm that the fetus is the signal
source for the FHR during monitoring.
Be aware that:
•a maternal HR trace can exhibit features that are very similar to those of an FHR trace, even
including accelerations and decelerations. Do not rely solely on trace pattern features to identify a
fetal source.
•Fetal Movement Profile (FMP) annotations on a fetal trace alone may not always indicate that the
fetus is alive. The body of a deceased fetus can move and cause the monitor to annotate fetal body
movements.
10
Here are some examples where the maternal HR can be misidentified as the FHR.
When using an ultrasound transducer:
•It is possible to pick up maternal signal sources, such as the aorta or other large vessels.
•Misidentification may occur when the maternal HR is higher than normal (especially when it is
over 100 bpm).
When using a fetal scalp electrode:
•Electrical impulses from the maternal heart can sometimes be transmitted to the fetal monitor
through a recently deceased fetus via the spiral scalp electrode cable, appearing to be a fetal signal
source.
•The recorded maternal HR (and any artifact) can be misinterpreted as an FHR (especially when it
is over 100 bpm).
When Fetal Movement Profile (FMP) is enabled:
FMP annotations in the absence of fetal life may be a result of:
•Movement of the deceased fetus during or following maternal movement.
•Movement of the deceased fetus during or following manual palpation of fetal movement
(especially if the pressure applied is too forceful).
1 Introduction
•Movement of the ultrasound transducer.
•The ultrasound transducer detecting a maternal movement source, such as the mother coughing.
See also “Monitoring FHR and FMP Using Ultrasound” on page 165 and “Monitoring FHR Using
DECG” on page 199.
To reduce the possibility of mistaking the maternal HR for an FHR, it is recommended that you
monitor both maternal and fetal heart rates. The monitor's cross-channel verification (CCV) feature
can help by automatically detecting when a maternal HR coincides with an FHR. For further details,
see “Cross-Channel Verification (CCV)” on page 159.
Introducing the Avalon Family of Fetal Monitors
The Avalon family of fetal monitors consists of the Avalon FM20, FM30, FM40, and FM50. While the
FM20/FM30 and the FM40/FM50 have different form factors, the method of operation is very
similar for all monitors. The Avalon fetal monitors also share transducers, accessories, software, and
are compatible with the Avalon CL, and Avalon CTS Fetal Transducer Systems.
Intended Use
The Philips Avalon FM20 (M2702A), FM30 (M2703A), FM40 (M2704A), and FM50 (M2705A) fetal/
maternal monitors are intended for:
•noninvasive monitoring of fetal heart rates and movements.
•invasive monitoring of direct fetal heart rate, intrauterine pressure, and for displaying and
recording of fetal and maternal electrocardiogram (ECG) (FM30 and FM50 only).
•displaying, storing, and recording patient data and parameter values, and for generating alarms
from fetal and maternal parameters.
•transmitting patient data and parameter values to a patient information and surveillance system.
•use by trained health care professionals.
•use in antepartum testing areas, in labor and delivery rooms, and during postpartum recovery in
the hospital environment. They are not intended for use in intensive care units or operating rooms.
+
•monitoring in a bath or shower (Avalon CL cableless transducers Toco
ECG/IUP only).
•transport situations in healthcare facilities, for healthcare facilities outside hospitals, such as
doctors’ offices, and for use in private households (FM20 and FM30 only).
MP, Ultrasound, and
11
1 Introduction
WARNING
The fetal/maternal monitors are not intended for:
•use during defibrillation, electro-surgery, or magnetic resonance imaging (MRI).
•Electrocardiography (ECG) measurements on patients connected to electrical stimulator or with
cardiac pacemakers.
•use of the invasive measurements IUP and fetal DECG, use of the patient module (M2738A) and
use of the Avalon CL system in domestic establishments, and those connected directly to the
public low-voltage supply network that supplies buildings used for domestic purposes.
•measuring the maternal temperature using the tympanic thermometer (866149) in private
households.
WARNING
No modification of the fetal monitors, transducers, and the Avalon CL base station is allowed.
CAUTION
U.S. federal law restricts this device to sale by, or on the order of, a physician.
All users must read the Instructions for Use before working with the fetal monitor. Disregarding the
contents of the Instructions for Use is considered abnormal use.
12
Indications for Use
Avalon Fetal/Maternal Monitor FM20
Indicated for use by trained health care professionals whenever there is a need for monitoring the
physiological parameters uterine activity, heart rate, oxygen saturation, noninvasive blood pressure,
pulse rate, and temperature of pregnant women, and the fetal heart rates of single fetuses, twins, and
triplets in labor and delivery rooms, in antepartum testing areas, in private households and during
transports in healthcare facilities.
Avalon Fetal/Maternal Monitor FM30
Indicated for use by trained health care professionals whenever there is a need for monitoring the
physiological parameters uterine activity, heart rate, electrocardiography (ECG), oxygen saturation,
noninvasive blood pressure, and pulse rate, and temperature of pregnant women, and the fetal heart
rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas, in
private households and during transports in healthcare facilities.
Avalon Fetal/Maternal Monitor FM40
Indicated for use by trained health care professionals whenever there is a need for monitoring the
physiological parameters uterine activity, heart rate, oxygen saturation, noninvasive blood pressure,
and pulse rate, and temperature of pregnant women, and the fetal heart rates of single fetuses, twins,
and triplets in labor and delivery rooms and in antepartum testing areas.
Avalon Fetal/Maternal Monitor FM50
Indicated for use by trained health care professionals whenever there is a need for monitoring the
physiological parameters uterine activity, heart rate, electrocardiography (ECG), oxygen saturation,
noninvasive blood pressure, and pulse rate, and temperature of pregnant women, and the fetal heart
rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas.
1 Introduction
Safety Information
In this guide:
•A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to
observe a warning may result in death or serious injury to the user or patient.
•A caution alerts you to where special care is necessary for the safe and effective use of the product.
Failure to observe a caution may result in minor or moderate personal injury or damage to the
product or other property, and possibly in a remote risk of more serious injury.
13
1 Introduction
No Host Monitoring
No Host Monitoring
cl NBP Disconnect
cl
SpO Disconnect
Electrical Hazards
WARNING
Electrical shock hazard: Do not open the monitor housing. Refer all servicing to qualified service
personnel.
•Always use the supplied power cord with the earthed mains plug to connect to an earthed AC
mains socket. Never adapt the mains plug from the fetal monitor to fit an unearthed AC mains
socket.
•Do not use AC mains extension cords or multiple portable socket-outlets.
•FM20/FM30 only: The protective earth conductor is required for EMC purposes. It has no
protective function against electric shock. Double and/or reinforced insulation protects this
device against electric shock.
•Do not connect any devices that are not supported as part of a system.
•Any non-medical device placed and operated in the patient’s vicinity must be powered with an
approved isolation transformer that ensures mechanical fixing of the power cords, and covering of
any unused power outlets.
•The fetal/maternal monitor is NOT intended for use during defibrillation, electro-surgery, or
MRI. Remove all transducers, sensors, and accessories before performing electro-surgery,
defibrillation, or MRI, otherwise harm to the patient or the user can result.
•Do not touch the charging contacts for the cableless transducers at the Avalon CL base station
while you are touching the patient.
Leakage currents: If several items of equipment used to monitor a patient are interconnected, the
resulting leakage current may exceed allowable limits.
Radio Frequency Interference
WARNING
•Short range radio connections are subject to interruption due to interference from other radio
sources in the vicinity, including microwaves, bluetooth devices, WLAN devices (802.11b,g,n), and
cordless phones. Depending on the strength and duration of the interference, the interruption may
occur for an extended period. A loss of connection, due to moving out-of-range, interference, or
for other reasons, is indicated with a
on the IntelliVue CL NBP or CL SpO
INOP at the fetal monitor. Correct channel configuration is important, refer to
the Service Guides and the Configuration Guide for details.
•To avoid magnetic interference affecting the mode of the pacemaker, ensure that the
Avalon CL base station does not come into close contact with implanted pacemakers.
•This equipment generates, uses, and radiates radio-frequency energy, and if it is not installed and
used in accordance with its accompanying documentation, may cause interference to radio
communications. Operation of this equipment in a residential area may cause interference, in
which case the users must take whatever measures may be required to correct the interference.
•Do not use cordless/mobile phones or any other portable RF communication system within the
patient vicinity, or within a 1.0 m radius of any part of the fetal monitoring system.
Pods, or a
2
INOP (here the host is the fetal monitor)
, or
, or
14
•For paced patients: The radiated SRR power of the CL SpO2, CL NBP Pods, CL F&M Pod, and
the CL Wide Range Pod, and other sources of radio-frequency energy, when used in very close
proximity of a pacemaker, might be sufficient to interfere with pacemaker performance. Due to
shielding effects of the body, internal pacemakers are somewhat less vulnerable than external
pacemakers. However, caution should be exercised when monitoring paced patients.
•In order to minimize the possibility of interference, avoid positioning and wearing the Cableless
Measurement Devices in very close proximity to a pacemaker. Consult the pacemaker
manufacturer for information on the RF susceptibility of their products.
Use Environment
WARNING
Explosion Hazard:
•Do not use in the presence of flammable anesthetics, such as a flammable anesthetic mixture with
air, oxygen, nitrous oxide, or in oxygen rich environment. Use of the devices in such an
environment may present an explosion hazard.
•Use only Philips batteries part number M4605A with the FM20 or FM30 with battery option. Use
of a different battery may present a risk of fire or explosion.
1 Introduction
Environmental Specifications:
•The performance specifications for the monitors, measurements, and accessories apply only for
use within the temperature, humidity, and altitude ranges specified in “Environmental
Specifications” on page 287.
Liquid Ingress:
•Do not operate the monitor if it is wet. If you spill liquid on the monitor, contact your service
personnel, or Philips service engineer.
•Never immerse the fetal monitor or the CL base station in liquid. You must protect them against
water sprays or splashes. Place the fetal monitor and the CL base station where there is no chance
of contact with, or falling into water or other liquids.
•Do not perform underwater monitoring (for example, in a bath or shower) using wired
transducers.
•The CL Fetal & Maternal Pod is not intended for underwater monitoring. The contacts between
the CL Fetal & Maternal Pod and the electrode patch have to be kept dry at all times. The CL Fetal
& Maternal Pod mounted on the electrode patch, can be worn underneath a shower, as long as the
CL Fetal & Maternal Pod stays mounted. Radio transmissions in the shower may be compromised.
Heat Exposure:
•Do not dry equipment using heating devices such as heaters, ovens (including microwave ovens),
hair dryers, and heating lamps.
•Do not put equipment or accessories in autoclave (for sterilization).
Positioning Equipment:
•The device should not be used adjacent to, or stacked with, other equipment unless otherwise
specified.
15
1 Introduction
INOP only
Alarms
Accessories
Prohibited Environments:
•The monitors and their transducers, Pods, and accessories are not intended for use in an MRI
environment or in an oxygen-enriched environment (for example, hyperbaric chambers).
WARNING
•Do not rely exclusively on the audible alarm system for fetal monitoring. Adjustment of alarm
volume to a low level or off during monitoring may result in a dangerous situation. Remember that
the most reliable method of fetal monitoring combines close personal surveillance with correct
operation of monitoring equipment.
•Alarm systems of the monitor and those of the connected obstetrical information and surveillance
system are independent and not synchronized.
•In
mode, no fetal/maternal patient alarms are enabled or indicated.
WARNING
Philips' approval: Use only Philips-approved accessories. Using non-Philips-approved accessories
may compromise device functionality and system performance, and cause a potential hazard.
Reuse: Never reuse disposable transducers, sensors, accessories, and so forth, that are intended for
single use, or single patient use only. Reuse may compromise device functionality and system
performance, and cause a potential hazard.
Electromagnetic compatibility: The use of accessories, transducers, and cables other than those
specified may result in increased electromagnetic emissions or decreased electromagnetic immunity of
the device.
Damage: Do not use a damaged sensor or one with exposed electrical contacts.
Cables and tubing: When connecting devices for acquiring measurements, always position cables and
NBP tubing carefully to avoid entanglement or potential strangulation.
16
Security Information
Erase All
Stored Data Recording
Protecting Personal Information
Protecting personal health information is a primary component of a security strategy. Each facility
using the monitors must provide the protective means necessary to safeguard personal information
consistent with country laws and regulations, and consistent with the facility’s policies for managing
this information. Protection can only be realized if you implement a comprehensive, multi-layered
strategy (including policies, processes, and technologies) to protect information and systems from
external and internal threats.
As per its intended use, the patient monitor operates in the patient vicinity and contains personal and
sensitive patient data. This also includes the trace print-outs at the monitor.
The monitor also includes controls to allow you to adapt the monitor to the patient's care model.
To ensure the patient's safety and protect their personal health information, you need a security
concept that includes:
•Physical security access measures - access to the monitor must be limited to authorized users.
It is essential that you consider physical security measures to ensure that unauthorized users
cannot gain access.
•Operational security measures - for example, ensuring that patients are discharged after
monitoring in order to remove their data from the monitor.
•Procedural security measures - for example, assigning only staff with a specific role the right to
use the monitors.
In addition, any security concept must consider the requirements of local country laws and regulations.
1 Introduction
Always consider data security aspects of the network topology and configuration when connecting
patient monitors to shared networks. Your medical facility is responsible for the security of the
network, where sensitive patient data from the monitor may be transferred.
When a monitor is returned for repair, disposed of, or removed from your medical facility for other
reasons, always ensure that all patient data is removed from the monitor by ending monitoring for the
last patient (see “Discharging a Patient” on page 148). Also select
menu, to erase all stored data.
NOTE
Log files generated by the monitors and measurement modules are used for system troubleshooting
and do not contain protected health data.
About HIPAA Rules
If applicable, your facility’s security strategy should include the standards set forth in the Health
Insurance Portability and Accountability Act of 1996 (HIPAA), introduced by the United States
Department of Health and Human Services. You should consider both the security and the privacy
rules and the HITECH Act when designing policies and procedures. For more information, please
visit:
http://www.hhs.gov/ocr/privacy/
in the
17
1 Introduction
About the EU Directives
If applicable, your facility’s security strategy should include the practices set forth in the Directive on
the protection of individuals with regard to the processing of personal data and on the free movement
of such data (Directive 95/46/EC of the European Parliament and of the Council of
24 October 1995). In addition, your facility should also take into account any additional, more
stringent standards put forward by any individual EU countries; that is, Germany, France, and so on.
Philips Product Security Policy Statement
Additional security and privacy information can be found on the Philips product security web site at:
http://www.healthcare.philips.com/main/support/equipment-performance/product-security/
index.wpd
Manufacturer Disclosure Statement for Medical Device Security – MDS2
You can view the Manufacturer Disclosure Statements for Medical Device Security (MDS2) for
specific devices at:
http://www.healthcare.philips.com/main/support/equipment-performance/product-security/
index.wpd
Overview of System Components
The Avalon CL system consists of the Avalon CL base station, the Avalon CL transducers, the
Avalon CL Fetal & Maternal Pod, the CL Wide Range Pod, and the IntelliVue CL Pods.
The IntelliVue CL Pods are only used for maternal measurements within the AvalonCL solution.
The following table provides an overview of all the devices.
Avalon FM20/FM30 and FM40/
FM50
Wired Transducers
FM20/FM30
M2702A and M2703A
FM40/FM50
M2704A and M2705A
Avalon CL Base Station
Avalon CL Transducers
Avalon CL Base Station
866074
CL Pods
18
1 Introduction
Avalon FM20/FM30 and FM40/
FM50
Wired Transducers
US transducer (wired)
M2736A
Toco/ Toco MP transducer
(wired)
Avalon CL Base Station
Avalon CL Transducers
Avalon CL US Transducer
(cableless)
866076
Avalon CL Toco+ MP Transducer
(cableless)
CL Pods
Avalon CL Fetal & Maternal Pod
866488
Avalon CL Wide Range Pod
M2734A and M2734B
866487
866075
19
1 Introduction
Avalon FM20/FM30 and FM40/
FM50
Wired Transducers
Toco+ transducer with ECG/IUP
capability (wired)
M2735A
Patient Module for ECG/IUP
M2738A
Avalon CL Base Station
Avalon CL Transducers
Avalon CL ECG/IUP Transducer
(cableless)
866077
CL Pods
IntelliVue CL NBP Pod
(cableless)
865216
IntelliVue CL SpO2 Pod
(cableless)
865215
20
2What's New
aFHR
aHR
aToco
cl F&M Electrode Status
cl F&M Status
This section lists the most important new features and improvements to the fetal monitors and their
user interface introduced with Release L.3.
You may not have all of these features, depending on the fetal monitor configuration purchased.
What's New in Release L.3
Avalon CL Fetal & Maternal Pod
The Avalon CL Fetal & Maternal Pod is an extension of the Avalon CL solution. The Avalon
CL Fetal & Maternal Pod is used together with a single use patch with five abdominal electrodes. Prior
to use, the Avalon CL Fetal & Maternal Pod is charged and assigned at the Avalon CL base station.
The Avalon CL Fetal & Maternal Pod is used with singleton pregnancies. This solution provides
especially benefits to patients with high body mass index (BMI), although it can be used for all patients.
2
The Avalon CL Fetal & Maternal Pod includes the following features:
•The Avalon CL Fetal & Maternal Pod measures fetal heart rate (
and uterine activity (
•The
contact status, when the CL Fetal & Maternal Pod is placed on the electrode patch, and the patch,
and the electrodes are applied to the patient's abdominal skin. The window can be opened by
selecting the new SmartKey
setup menu.
•The Avalon CL Fetal & Maternal Pod and the CL Wide Range Pod are assigned at the CL base
station in the same easy way as the IntelliVue CL Pods.
•The functions Finder LED and Call Patient work also with the Avalon CL Fetal & Maternal
Pod.
See “Avalon CL Fetal & Maternal Pod” on page 21, “CL Fetal & Maternal Electrode Patch” on
page 91, “Applying the CL Fetal & Maternal Patch and Pod” on page 100, and “CL Pod Assignment”
on page 99.
) from electrical signals.
window at the monitor gives an overview of the current electrode
, or selecting the corresponding function in the main
), maternal heart rate (
),
21
2 What's New
Enter Note
Type a note
Avalon CL Wide Range Pod
The Avalon CL Wide Range Pod is an extension of the Avalon CL solution. The Avalon CL Wide
Range Pod extends the signal range of the cableless measurements. It transmits the cableless
measurement signals via the hospital WLAN/WiFi to the fetal monitors. Wearing the CL Wide Range
Pod, the patient can walk freely within the hospital's WLAN/WiFi range during monitoring. Prior to
use, the Avalon CL Wide Range Pod is charged and assigned at the Avalon CL base station.
•A new WLAN symbol is printed on the trace to indicate the use of an assigned and active Avalon
CL Wide Range Pod.
•During monitoring with an active Avalon CL Wide Range Pod, the sound from the fetal heart rate
is replaced by an artificial QRS sound (when the patient is monitored with a CL US transducer or
CL F&M Pod).
See “CL Wide Range Pod” on page 91.
Dawes/Redman
Dawes/Redman is a method of fetal monitoring trace interpretation for helping to assess fetal
well-being. The algorithm based on the Dawes/Redman criteria uses the parameters fetal heart rate,
gestational age, fetal movement, accelerations and decelerations, and long, and short term variability to
generate a report.
Not available in the USA and territories relying on FDA market clearance.
Support of XDS Remote Display
The fetal monitor software supports the IntelliVue XDS “Remote Display” functionality only in
combination with IntelliSpace Perinatal revision K or higher. From the user interface of the
IntelliSpace Perinatal system, you can access all screen-operable functions of the currently connected
fetal monitor. These functions include for example, starting and stopping physiological measurements,
changing measurement modes, changing alarm limits, and silencing alarms.
New Alarm Behavior of CL Battery Empty INOP
The CL <xxx> Battery Empty INOP issued by the CL devices and the fetal monitor has a new alarm
behavior:
•At the fetal monitor, it is issued now with a cyan INOP alarm and tone.
•At CL devices (CL transducers and CL Pods), it is issued now with a tone.
See “Patient Alarms and INOPs” on page 129.
Using CL Pods with FM20/30 #E25
The CL Pods can now also be used with a battery operated FM20 or FM30 (option #E25). The
CL Pods can be assigned and activated. The CL F&M Pod can also be charged.
See “Using Batteries” on page 110.
Entering Notes - Type a Note
In the menu
pre-configured notes. To enter a note, select the new menu item
keypad opens. The typed note can be up to 30 characters long.
See “Typing Notes” on page 70.
, it is now possible to enter a note manually, instead of selecting one of the
. A window with a touch
22
What's New in Release J.3
Avalon CL Transducer System
The Avalon CL Transducer System provides cableless monitoring with the Avalon FM20/FM30 and
FM40/FM50 with the same functionality and performance as the wired measurement devices (e.g.
twin and triplets monitoring). The Avalon CL Transducer System has a straight-forward handling and
operating concept. The CL transducers are assigned by simply docking them at the CL base station, no
further configuration is necessary.
The Avalon CL Transducer System includes the following features:
•Cableless monitoring of twins and triplets (see “Monitoring Twin FHRs” on page 183 and
“Monitoring Triple FHRs” on page 191)
•Cableless maternal measurement Pods CL SpO
page 92)
•Maternal pulse from a CL Toco
Pulse Rate” on page 221)
•A cableless ECG/IUP transducer measuring IUP or fetal/maternal ECG (see “Monitoring
MECG Wave” on page 225 and “Monitoring FHR Using DECG” on page 199)
•Watertight cableless transducers that can be used to monitor in water (see “Underwater
Monitoring” on page 108)
+
MP (Smart Pulse) transducer (see “Monitoring Maternal Heart /
2 What's New
and CL NBP (see “IntelliVue CL Pods” on
2
•Patient call that pages an ambulating mother with an audible signal emitted by the worn CL
transducers (see “Calling Patients” on page 93)
•Out-of-range audible signal emitted by the worn CL transducers to inform an ambulating mother
that she has reached the limit of the active signal area-of-reach (see “Standard Radio Range of CL
Transducers” on page 87)
•A transducer finder LED on all CL transducers to help identify the assigned transducer (see
“Cableless Transducer LED Indication” on page 86)
Support For Use of Maternal Cableless Measurement Devices
The IntelliVue CL measurement Pods are patient-worn, battery-powered measurement devices for
and NBP. The devices provide measurement values on the built-in display and communicate
SpO
2
them to the fetal monitor using the wireless short range radio (SRR) interface of the Avalon CL base
station (see “IntelliVue CL Pods” on page 92).
Maternal Temperature Measurement
To measure maternal temperature, the new optional tympanic thermometer (866149) is available for
the Avalon fetal monitors (“Monitoring Maternal Temperature” on page 249). The measurement data
is:
•documented and printed out at the local recorder, and transmitted to the obstetrical information
and surveillance system.
•displayed as a numeric on the screen.
23
2 What's New
Record on Trace
Send to OB Sys
Start ECG
Record ECG
Call Patient
Tele Info
Tele Info
Tele Info
Enter Temp
NBP Modes
QuickAdmit
Coincidence
Coincidence
Manually Entered Maternal Temperature Measurements
Manually measured temperatures can be entered at the fetal monitor. They are displayed as a numeric
on screen, and are printed out on the recorder trace (“Monitoring Maternal Temperature” on
page 249).
SpO2 Recordings and Transmissions
SpO2 annotation on local recorder
You can now configure the interval for printing the SpO
the new configuration setting
transmission to an obstetrical information and surveillance system
SpO
2
You can now configure the interval for transmitting the SpO
information and surveillance system. With the new configuration setting
interval to 1 or 5 minutes.
New Design for the User Interface
The user interface for the fetal monitors has been redesigned to bring the presented information into
the foreground, letting the structural elements such as keys and frames retreat into the background.
Additionally special regard was given to making the "look and feel" similar to that of standard software
products (see “Operating and Navigating” on page 38).
New SmartKeys
•The
•With the
transducers.
•With the
the
connected Avalon CL base station.
•With the
manually measured maternal temperature values.
•With the
start and stop a measurement.
•With the
All new SmartKeys are optional, and have to be configured in Configuration Mode for use (see
“SmartKeys” on page 42).
SmartKey and menu item is renamed to
SmartKey, you can now page patients who are ambulating wearing Avalon CL
SmartKey, you can call up the
window, you can control and view the status of the cableless transducers from the
SmartKey, a pop-up window opens showing a numeric pad for entering
SmartKey, you can access the NBP Mode selection and setup, and can directly
SmartKey, you can quick admit a patient to the monitor.
numeric values on the recorder trace. With
2
, you can set the interval to 1 or 5 minutes.
numeric values to an obstetrical
2
, you can set the
.
window on the fetal monitor display. In
Coincidence INOP Tone
When the cross-channel verification detects that the signal of the maternal heart rate coincides with the
fetal heart rate, the
INOP tone has a configurable delay see “Cross-Channel Verification (CCV)” on page 159.
24
INOP is now issued with a tone at the fetal monitor. The
Increased Internal Back-up Memory
ECG Leads Off, Cuff Overpress, Cuff NotDeflated, Battery Empty, No
Pulse
The internal back-up memory is now able to store traces and data from at least the last 3.5 hours with
the software revision J.3, and minimum 7 hours with the new mainboard hardware revision A 00.18
(see “Manually Recording Stored Data” on page 258 and “Recovering Data” on page 257).
Dual System Interface Support
If the fetal monitor is connected via a LAN connection to OB TraceVue/IntelliSpace Perinatal, the
RS232 interface can be used independently to connect e.g. an EMR system on read-only basis. The
system connected to the RS232 interface in this case cannot alter any data (such as ADT data, or the
date and time setting), or interfere with functions of the monitor, but is able to read output data. The
obstetrical information and surveillance system connected via LAN has priority.
USB Interface
An optional USB interface allows the use of bar code readers and input devices such as a keyboard, or
mouse (see “Getting to Know Your Avalon FM20/FM30”/“Bottom” on page 32 and “Getting to
Know Your Avalon FM40/FM50”/“Rear” on page 33).
Flexible Nurse Call Interface
An optional Flexible Nurse Call interface allows the connection of a nurse call device to the fetal
monitors (see “Getting to Know Your Avalon FM20/FM30”/“Bottom” on page 32 and “Getting to
Know Your Avalon FM40/FM50”/“Rear” on page 33).
2 What's New
DHCP Support
The DHCP support offers an alternative to BOOTP. DHCP (Dynamic Host Configuration Protocol)
enables the fetal monitors to request an IP address (internet protocol address) from the connected
network (OB TraceVue/IntelliSpace Perinatal) automatically.
Data Export Support
You can now export measurement values from the monitor to other devices via the LAN interface, or
with the optional MIB RS232 interface (see “Getting to Know Your Avalon FM20/FM30”/“Bottom”
on page 32 and “Getting to Know Your Avalon FM40/FM50”/“Rear” on page 33).
NBP Configurable Measurement Sequence
Up to four measurement cycles can be set up which will run consecutively. For each cycle, you can set
the number of measurements and the interval between them. By setting the last cycle to run
continuously, you can have regular measurements continue after the sequence has run (see “Enabling
Sequence Mode and Setting Up the Sequence” on page 240.
Alarms Enhancements
In addition to the standard cyan INOPs, some INOPs can now be configured as red or yellow INOPs
to provide a severity indication (
) (see “Alarms” on page 117).
25
2 What's New
Alarm Reminder
Alarm Reminder
Auto Free
Alarm Reminder
In Configuration Mode, you can set now an
reminder of alarm conditions that remain active after the alarm is acknowledged. This reminder may
take the form of a repetition of the alarm tone for a limited time, or an unlimited repetition of the
alarm tone (this is the same as a new alarm). The interval between silencing the alarm and sounding the
reminder tone can be set to one, two, or three minutes (see “Alarm Reminder” on page 122).
Auto Free
In Configuration Mode, you can now set an
when the fetal monitor has been powered off, or is in standby mode for a set time. Only the
demographic patient data is deleted, the trace data is not affected.
setting which discharges a patient automatically
. The
emits an audible
26
3Basic Operation
This chapter gives you an overview of the monitor and its functions. It tells you how to perform tasks
that are common to all measurements (such as entering data, switching on a measurement, changing
some monitor settings, and setting up the recorder). The alarms section gives an overview of alarms.
The remaining sections tell you how to perform individual measurements, and how to care for and
maintain the equipment.
3
27
3 Basic Operation
Supported Measurements
Different measurements for the same physiological parameter may have a different appearance on the
trace, due to: variability (HR), averaging, delay, amplitude, or artifacts. Before interpreting the trace,
regard the fetal monitor setup and transducers used.
The following Fetal measurements are supported:
MeasurementsFM20FM30FM40FM50
Fetal Heart Rate (FHR) via
US (including Twins)
Triple FHR via US
dFHR via Direct ECG
(DECG)
aFHR via CL F&M Pod
To co
aToco via CL F&M Pod
Intrauterine Pressure (IUP)
The following Maternal measurements are supported:
MeasurementsFM20FM30FM40FM50
Maternal Heart Rate (MHR)
via Maternal ECG Electrodes
Maternal ECG (MECG wave)
aHR via CL F&M Pod
Maternal Pulse from Toco
StandardStandardStandardStandard
OptionalOptionalOptionalOptional
-Standard-Standard
OptionalOptionalOptionalOptional
StandardStandardStandardStandard
OptionalOptionalOptionalOptional
-Standard-Standard
StandardStandardStandardStandard
-Standard-Standard
OptionalOptionalOptionalOptional
StandardStandardStandardStandard
Noninvasive Blood Pressure
with Pulse Rate
Pulse Oximetry (Maternal
SpO
) with Pulse Rate
2
Maternal Temperature
28
OptionalOptionalOptionalOptional
OptionalOptionalOptionalOptional
OptionalOptionalOptionalOptional
Avalon FM20 and FM30
This section outlines the capabilities of your monitor.
3 Basic Operation
Avalon FM20
The Avalon FM20 fetal/maternal monitor provides a solution for external fetal monitoring
applications, and optional noninvasive maternal vital signs.
You can monitor fetal heart rates (FHRs) externally using ultrasound, uterine activity and maternal
pulse using an external Toco transducer, and the maternal heart rate (MHR) with maternal ECG
electrodes, and optionally, noninvasive blood pressure and maternal oxygen saturation (SpO
Measurements are displayed on a 6.5 inch color display as numerics. The display is a touchscreen, and
you operate the monitor using this touchscreen interface. The integrated recorder documents fetal and
maternal measurements as well as the user-defined annotations.
You can connect the monitor to an OB TraceVue/IntelliSpace Perinatal system via the RS232
connection, or over a LAN connection (with OB TraceVue Revision E.00.00 and later, or IntelliSpace
Perinatal Revision H.0 and later).
Avalon FM30
The Avalon FM30 fetal/maternal monitor offers a solution for both external and internal fetal
monitoring applications, and optional noninvasive maternal vital signs.
The Avalon FM30 shares all the features and capabilities of the Avalon FM20. In addition, you can
monitor one FHR internally with a direct fetal electrocardiogram (DECG), uterine activity internally
using an intrauterine pressure (IUP) catheter together with a Toco
The Avalon FM30 carries the IP label, indicating that it is capable of intrapartum monitoring.
2
+
transducer or patient module.
).
29
3 Basic Operation
FM20/30 with
Battery
Option #E25
The battery option for the FM20/30 provides support for the in-transport monitoring of all
measurements when disconnected from a power supply. Existing data storage is automatically
uploaded to OB TraceVue or IntelliSpace Perinatal after reconnecting it to the system. Trace printing
Only
during transport is also possible.
Avalon FM40 and FM50
This section outlines the capabilities of your monitor.
Avalon FM40
The Avalon FM40 fetal/maternal monitor provides a solution for external fetal monitoring
applications, and noninvasive maternal vital signs.
You can monitor fetal heart rates (FHRs) externally using ultrasound, uterine activity using an external
Toco transducer, and the maternal heart rate (MHR) via maternal ECG electrodes, and noninvasive
blood pressure and maternal oxygen saturation (SpO
Measurements are displayed on a 6.5 inch color display as numerics. The display is a touchscreen, and
you operate the monitor using this touchscreen interface. The integrated recorder documents fetal and
maternal measurements as well as the user-defined annotations.
You can connect the monitor to an OB TraceVue/IntelliSpace Perinatal system with the RS232
connection, or over a LAN connection (with OB TraceVue Revision E.00.00 and later, or IntelliSpace
Perinatal Revision H.0 and later).
).
2
30
Loading...
+ 308 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.