Philips M3535A HeartStart MRx, M3536A HeartStart MRx User manual

NBP and CO2 Calibration Instructions
M3535A/M3536A HeartStart MRx
PHILIPS
Notice
About This Edition
HeartStart MRx, product version indicated below. This information is subject to change without notice.
Philips shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Edition History
Pub. Number Ed. S/W Version Print Date
453564067081
1 9.00 and earlier Sept., 2007
Copyright
Copyright © 2007 Koninklijke Philips Electronics N.V.
All rights are reserved. Permission is granted to copy and distribute this document for your organization’s internal educational use. Reproduction and/or distribution outside your organization in whole or in part is prohibited without the prior written consent of the copyright holder.
Windows Media Player® is a registered trademark of Microsoft Corporation.
Use of supplies or accessories other than those recommended by Philips may compromise product performance.
Medical Device Directive
The HeartStart MRx complies with the requirements of the Medical Device Directive 93/42/EEC and carries the
mark accordingly.
0123
Manufacturer:
Philips Medical Systems 3000 Minuteman Road Andover, MA USA 01810-1099
(978) 687-1501
Authorized EU-representative:
Philips Medizin Systeme Böblingen GmbH Hewlett Packard Str. 2 71034 Böblingen Germany
U.S. FCC and Industry Canada Radio Compliance:
Contains FCC ID: PQC-WMTS-MODULE
Operation of this equipment requires the prior coordination with a frequency coordinator designated by the FCC for the Wireless Medical Telemetry Service. This device complies with Part 15 of the FCC rules and RSS-210 of Industry Canada. Operation is subject to the following conditions:
• This device may not cause harmful interference.
• This device must accept any interference received, including interference that may cause undesired operation.
Any changes or modifications to this equipment not expressly approved by Philips Medical Systems may cause harmful radio frequency interference and void your authority to operate this equipment.
Canada EMC: ICES-001
China: After-Sales Service: Beijing MEHECO-PHILIPS Medical
Equipment Service Center. After-Sales Service Address: No. 208, 2nd District, Wang Jing Li Ze Zhong Yuan, Chao Yang District, Beijing. Postal code: 100102. Telephone: 010-64392415. Registration number: SFDA(I)20043211207. Product Standard number: YZB/USA 52-21.
ii
Conventions Used in This Manual
These upgrade instructions contain the following conventions:
WARNING Warning statements describe conditions or actions that can result in personal injury or loss of life.
CAUTION Caution statements describe conditions or actions that can result in damage to the equipment or loss of
data.
NOTE Notes contain additional information on usage.
TIP Tips provide hands-on insight into servicing this product.
Text represents messages that appear on the display
[Softkey] represents softkey labels that appear on the display above or below the button
to which they correspond.
On-line viewing only
* Hypertext represents hypertext links, which will display as blue; click on the blue link to go
to that destination, then click on the blue destination to return.
Abbreviations
Name Abbreviation
HeartStart MRx Monitor/Defibrillator monitor/defibrillator
device
Carbon dioxide CO
End-tidal carbon dioxide EtCO
Non-invasive Blood Pressure NBP
2
2
iii
iv

Introduction

Introduction
CAUTION These instructions are intended for users trained in the safe and proper HeartStart MRx operation.
HeartStart MRx NBP and EtCO2 calibration is part of the HeartStart MRx maintenance routine that also includes operational checks, lithium ion battery maintenance, etc. described in HeartStart MRx Instructions for Use. NBP and EtCO not covered in the Instructions for Use.
This document covers the following calibration procedures:
• Yearly calibration (or every 10,000 cycles, whichever is sooner) of the Non-invasive Blood Pressure (NBP) module
• Yearly calibration (or every 4000 hours, whichever is sooner) of the End-tidal Carbon Dioxide (EtCO
Use these quick links to access the calibration procedures:
) module.
2
* NBP Module Calibration
calibration is the only HeartStart MRx maintenance procedure
2
* EtCO
Module Calibration
2

HeartStart MRx Calibration Overview

Consider reviewing the “HeartStart MRx Calibration” instructional video included in your calibration kit before proceeding with the calibration. To watch the DVD on your PC, you may need to install the latest version of Windows Media Player
http://www.microsoft.com/windows/windowsmedia/download/.
center The video is also available online at Perform calibration when prompted by the
Overdue
Fail/D (Versions B.05 and above) or Fail/NC (Versions prior to B.05). Regardless of your configuration settings, millimeters of mercury are the unit of measure for pressure in
the HeartStart MRx calibration. Use the conversion formulae from Ta b le 1 if necessary:
Table 1 Units of Pressure Conversion
inops. If a calibration is overdue, then the HeartStart MRx Op. Check fails with the code
1 unit = __ mmHg
1 kPa = 7.5 mmHg 1 mb 1 psi 1 atm. 1 inHg = 25.4 mmHg 30 inHg = 30 × 25.4 mmHg = 762 mmHg
.75 mmHg
=
51.7 mmHg
=
760 mmHg
=
®
available for free download from Windows Media download
http://theonlinelearningcenter.com/schtml/mrx/calibration/.
CO2 Calibration Overdue and NBP Calibration
Example:

Accessing Service Mode

The calibration activities are accessed through the Service Mode. To access the Service Mode:
1 From the Main Menu select Other, then Service. 2 Enter the “27689” password when prompted by the device.